| | Class 2 Device Recall Medtronic Select 3D Arterial Cannula |  |
| Date Initiated by Firm | January 30, 2014 |
| Date Posted | February 20, 2014 |
| Recall Status1 |
Terminated 3 on October 06, 2014 |
| Recall Number | Z-0995-2014 |
| Recall Event ID |
67400 |
| 510(K)Number | K043179 |
| Product Classification |
Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
|
| Product | Medtronic Select 3D Arterial Cannulae. Vented, models 78618 (18 Fr.) and 78620 (20 Fr.). Non-Vented Cannulae, models 78718 (18 Fr.) and 78720 (20 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic.
The Select 3D Arterial Cannulae both have clear flexible, thin wall wire-wound PVC bodies with angled, beveled tips. The proximal end of the cannula includes a 3/8" (0.95 cm) vented or non-vented connector with a peel cap. The vented connector allows air to be vented from the cannula before connection to the perfusion line.
The Select 3D Arterial Cannula tip has three integrated flutes which help diffuse and disperse blood flow. The cannula body features a tip orientation line indicating direction of the cannula tip during cannulation. Overall cannula length is 11.5" (29.2 cm).
These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass. |
| Code Information |
Model 78618, Lot numbers: 2011041199, 2011062306, 2011100971, 2011102092, 2011120307, 2012051764, 2012061644, 2012080274, 2012110107, 2012120373, 2013010071, 2013050197, 2013081170, 2013102358, 201203C853. Model 78620, Lot numbers: 2011030255, 2011040939, 2011041341, 2011101160 2011111043, 2012011041, 2012020540, 2012070617, 2012110017, 2012111257, 2013031297, 2013050408, 2013060024, 2013060816, 2013070896, 2013090420, 2013091570, 2013101359. Model 78718, Lot numbers: 2011020219, 2011051461, 2011051501, 2011060279, 2011081265, 2011091551, 2011100022, 2011111305, 2012011048, 2012032171, 2012040645, 2012070832, 2012071313, 2013031493, 2013041737, 2013051367, 2013060409, 2013091960, 2013102157, 2013110588, 201110C116, 201203C855. Model 78720, Lot numbers: 2011010067, 2011021825, 2011062606, 2011062607, 2011070536, 2011071117, 2011080605, 2011081029, 2011081501, 2011082463, 2011090046, 2011090047, 2011100092, 2012011142, 2012011424, 2012051863, 2012060210, 2012061344, 2012061868, 2012120133, 2012120781, 2012120941, 2012121072, 2013030331, 2013030943, 2013080633, 2013081206, 2013110894, 2013120553. |
| FEI Number |
2182208
|
Recalling Firm/ Manufacturer |
Medtronic Inc. Cardiac Rhythm Disease Management 8200 Coral Sea St NE Saint Paul MN 55112-4391
|
| For Additional Information Contact | Customer Service 800-854-3570 |
Manufacturer Reason for Recall | Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | The firm, Medtronic, sent an "Urgent Medical Device Recall" letter dated January, 2014 via UPS on January 30, 2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify and quarantine all listed product in your inventory; return all listed product to Medtronic and contact Customer Service at 800 854-3570 and reference RADAR #167826 to initiate a return and credit of product; complete and return the attached Customer Confirmation Certificate via fax to Medtronic at 651-367-0612 to the attention of Customer Focused Quality, and share this notification with others in your organization as appropriate.
Select 3D and Select CAP Arterial Cannula are not available for purchase at this time. Please contact your Medtronic sales representative or Medtronic LifeLine Technical Services at 1 877-526-7890 or (+1 763) 526-7890 to assist with alternative cannula selection and any additional questions you may have related to this notification. |
| Quantity in Commerce | 19483 total (7669 US, 11814 OUS) |
| Distribution | Worldwide distribution: US (Nationwide) including Puerto Rico, and Internationally to: Algeria, Argentina, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Cyprus, El Salvador, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Jordan, Kuwait, Netherlands, Oman, Paraguay, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Venezuela, and Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DWF
|
|
|
|