 Class 2 Device Recall Integra LifeSciences Corporationsee related information
Date Initiated by FirmDecember 05, 2013
Date PostedFebruary 20, 2014
Recall Status1 Terminated 3 on January 26, 2017
Recall NumberZ-0994-2014
Recall Event ID 67406
ProductIntegra Cusa Excel Rx Only CEM" Nosecone for CUSA Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.
Code Information All unexpired lot numbers for CEM Nosecone Gatalouge No. C6623 and C6636 from 1/1/2010 to present.  CEM" Nosecone, Catalogue No. C6623: 1083655, 1084219, 1084398, 1084399, 1084400, 1084682, 1084683, 1090937, 1091712, 1091714, 1091715, 1093125, 1093126, 1093128, 1093129, 1093471, 1093847, 1094575, 1094740, 1095587, 1095873, 1100237, 1100238, 1101217, 1101547, 1101548, 1102069, 1102070, 1102544, 1102545, 1102546, 1103102, 1103392, 1103444, 1103445, 1103446, 1104264, 1104265, 1105065, 1105066, 1105067, 1110208, 1110311, 1110348, 1110349, 1110473, 1110474, 1110892, 1111760, 1112182, 1112808, 1112809, 1113175, 1113176, 1113177, 1113178, 1113662, 1114112, 1114654, 1114655, 1114656, 1115224, 1115225, 1115499, 1115500, 1115501, 1115853, 1115854, 1120771, 1120772, 1120773, 1120774, 1121141, 1121142, 1121383, 1121627, 1121628, 1122257, 1122259, 1122260, 1123050, 1123312, 1123915, 1124642, 1124643, 1124980, 1125393, 1125394, 1125933, 1130036, 1130037, 1130038, 1130039, 1130218, 1130219, 1130663, 1130664, 1131340, 1131344, 1131986, 1132207, 1132326, 1132327, 1132328, 1132880  CEM" Nosecone, Catalogue No. C6636:  1084216, 1084293, 1084377, 1084680, 1084681, 1090123, 1090124, 1090125, 1090126, 1090660, 1090661, 1090826, 1091395, 1091434, 1091700, 1092388, 1092389, 1092390, 1092391, 1092392, 1093012, 1093013, 1093130, 1093131, 1093472, 1093473, 1093474, 1093824, 1093825, 1093826, 1093827, 1094741, 1095104, 1095105, 1095106, 1095414, 1095415, 1095418, 1095419, 1100239, 1100240, 1100241, 1101220, 1101221, 1102647, 1102648, 1102649, 1102650, 1102651, 1103103, 1103394, 1103395, 1103437, 1103848, 1103851, 1103916, 1103917, 1103919, 1103920, 1103921, 1103922, 1104258, 1104259, 1104260, 1104261, 1104262, 1105068, 1105069, 1105070, 1110333, 1110975, 1110976, 1110977, 1111491, 1111492, 1111493, 1112186, 1112187, 1112211, 1112212, 1112769, 1112770, 1112771, 1113180, 1113181, 1113182, 1113183, 1113184, 1113664, 1113665, 1113666, 1114358, 1114359, 1114657, 1114658, 1114659, 1114660, 1115036, 1115037, 1115038, 1115039, 1115503, 1115504, 1115505, 1115855, 1115856, 1115857, 1115858, 1115859, 1115860, 1120360, 1120361, 1120362, 1120775, 1120776, 1120777, 1121568, 1121729, 1122344, 1122346, 1122347, 1123053, 1123054, 1123497, 1123498, 1123499, 1123500, 1123501, 1124054, 1124055, 1124703, 1124704, 1124705, 1124706, 1124707, 1124710, 1125395, 1125396, 1125397, 1125398, 1125399, 1125400, 1125476, 1125934, 1125935, 1125936, 1125937, 1125938, 1125939, 1130220, 1130221, 1130222, 1130223, 1130834, 1131243, 1131987, 1132777, 1132778, 1132779, 1132780, 1132879, 1133061, 1103850, 1103852, 1103917, 1103918, 1103922, 1104257, 1104263, 1105071, 1110475, 1110476, 1110978, 1111490, 1111493, 1112187, 1112188, 1112212, 1113179, 1113180, 1113663, 1114357, 1114358, 1114659, 1114660, 1115039, 1115502, 1115504, 1115505, 1120776, 1120777, 1121167, 1121168, 1121169, 1121565, 1121566, 1121567, 1122344, 1122743, 1123051, 1123052, 1123497, 1123498, 1123499, 1123500, 1123501, 1123508, 1123509, 1123510, 1123511, 1123512, 1124052, 1124053, 1124707, 1124709
Recalling Firm/
Manufacturer
Integra LifeSciences Corp.
311 Enterprise Dr
Plainsboro NJ 08536-3344
For Additional Information Contact
609-275-0500
Manufacturer Reason
for Recall
The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.
FDA Determined
Cause 2
Material/Component Contamination
ActionIntegra sent an Urgent: Medical Device Correction Letter/Acknowledgement Return Form dated December 5, 2013 via traceable courier to their customers. The notification: advises customers of the nature of the issue and how it could affect CEM Nosecone use, advises customers what to do if the issue occurs and requests customers complete and return the Acknowledgment Return Form. The form should be faxed to rc@integralife.com or 1-609-275-9445. Any questions can be directed to customer service at 855-532-1723.
Quantity in Commerce17253 boxes
DistributionWorldwide Distribution - USA Nationwide