| | Class 2 Device Recall The Rush Endobronchial Tube set |  |
| Date Initiated by Firm | February 28, 2014 |
| Date Posted | March 19, 2014 |
| Recall Status1 |
Terminated 3 on September 28, 2016 |
| Recall Number | Z-1231-2014 |
| Recall Event ID |
67630 |
| 510(K)Number | K951091 |
| Product Classification |
Tube, bronchial (w/wo connector) - Product Code BTS
|
| Product | Rusch Brochopart and Rusch Brochopart White
The Rush Endobronchial Tube set is indicated for uses commonly associated with oral endobronchial tubes, including, but not limited to, thoracic surgery, bronchospirometry, administration of endobronchial anesthesia, etc. The endobronchial tube allows ventilation and selective inflation or deflation of either lung. |
| Code Information |
Product Code - 116200260, Batch 09LE51, 10GE31, 11LE42, 12DE17, 12IE37, 13CT26, 13DT32; Product Code - 116200280 - Batch 09HE34, 09JE41, 10BE07, 11HE33, 11JE41, 13ET12; Product Code - 116200350 - Batch 0918M, 0915M, 09JE43, 09JE42, 0917M, 11AE03, 11DE15, 11GE27, 09HE34, 11JE43, 11JE42, 12AE02, 12BE07, 12DE17, 12EE19, 12GE27, 12HE33, 13AE01, 13CT26, 13DT33, 13ET12; Product Code 116200370 - Batch 0915M, 0917M, 0914M, 09JE41, 09JE43, 11DE15, 11EE20, 11GE27, 11JE42, 11HE32, 11JE43, 11IE39, 12AE01, 12AE03, 12GE31, 12BE06, 12DE18, 12IE39, 12JE44, 13AT44, 13BT44, 13AE01, 12LE49, 13DT33; Product Code 116200390 - Batch 0914M, 0922M, 0917M, 09JE44, 09FE23, 09HE35, 09JE42, 10GE29, 10AE02, 09JE43, 09KE46, 09LE51, 10AE01, 10CE11, 10DE18, 10EE19, 10JE41, 10EE20, 11DE15, 11KE42, 11GE30, 11HH33, 11IE37, 11IE38, 11IE39, 13AT36, 13BT25, 13BT29; Product Code 116200410 - Batch 0922M, 0915M, 0918M, 09IE36, 09JE43, 10AE05, 10BE06, 10DE16, 10GE28, 10JE41, 11DE15, 11GE27, 11HE34, 11JE43, 13BT44, 13DT32; Product Code 116201350 - Batch 0921M, 09JE40, 09FE23, 13GT05; Product Code 116201370 - Batch 0922M, 0915M, 13AE01, 13ET18, 13HT12, 09FE24; Product Code 116201390 - Batch 09JE44, 10CE10; Product Code 116201410 - Batch 10KE47 |
| FEI Number |
3005747797
|
Recalling Firm/ Manufacturer |
Teleflex Medical 4024 Stirrup Creek Dr Durham NC 27703-9000
|
| For Additional Information Contact | Michael Taggart 919-433-4816 |
Manufacturer Reason for Recall | The product may fail to achieve seal of right lung due to the cuff inflating to one side |
FDA Determined Cause 2 | Device Design |
| Action | Teleflex sent a Urgent Medical Device Recall Notification dated February 28, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers were instructed:
1. If you have affected stock, immediately discontinue use and quarantine any products with the catalog numbers listed above.
2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-866-804-9881, Attn: Customer Service. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical.
3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-866-804-9881, Attn: Customer Service. This will allow us to document your receipt of this letter.
Teleflex Medical is committed to providing high quality, safe and effective products. We sincerely apologize for any inconvenience this action may cause your operations. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990. |
| Quantity in Commerce | 63,568 ea. |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MD, MA, MI, MN, MS, MO, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WV and WI,and the countries of United Arab Emirates, Austria, Australia, Azerbaijan, Belgium, Belgrade, Bahrain, Canada, Switzerland, Cameroon, China, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Dominican Republic, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Great Britain, Greece, Guyana, Croatia, Hong Kong, Hungary, Ireland, Israel, India, Iraq, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Latvia, Libya, Morocco, Serbia, Malaysia, Martinique, Nigeria, Netherlands, Norway, Oman, Panama, French Polynesia, Philippines, Poland, Portugal, Qatar, Reunion, Russia, San Salvador, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, San Marino, Thailand, Trinidad, Tunisia, Turkey, Venezuela, Yemen and Zambia |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = BTS
|
|
|
|