| | Class 2 Device Recall ZYM B Reagent (REF 70493) |  |
| Date Initiated by Firm | March 06, 2014 |
| Date Posted | April 18, 2014 |
| Recall Status1 |
Terminated 3 on March 14, 2017 |
| Recall Number | Z-1475-2014 |
| Recall Event ID |
67820 |
| Product Classification |
Device, general purpose, microbiology, diagnostic - Product Code LIB
|
| Product | ZYM B Reagent (REF 70493).
The ZYM B reagent is an additional test used to reveal the results of some miniaturized biochemical test included in the API strips. The API product line is a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms. |
| Code Information |
Codes - 1002706740, Exp. Date 10/14/2014, 1002717860, Exp. Date 10/24/2014, 1002742610, Exp. Date 10/30/2014, 1002760700, Exp. Date 11/4/2014, 1002821420, Exp. Date 11/26/2014, 1002857840, Exp. Date 12/12/2014, 1002860530, Exp. Date 12/12/2014 |
| FEI Number |
1000611339
|
Recalling Firm/ Manufacturer |
BioMerieux SA
Chemin De L'Orme
Marcy L'Etoile
France
|
| For Additional Information Contact | Ryan LeMelle 314-731-8582 |
Manufacturer Reason for Recall | bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent |
FDA Determined Cause 2 | Other |
| Action | Letters were issued to customers on March 10, 2014. The letter contained instructions to be used if they still were using the product. The letter also included a report form which was to be returned. |
| Quantity in Commerce | 33,853 |
| Distribution | Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WV, WA, WI, and WY and the country of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Benin, Bolivia, Bosnia, Brazil, Brunel, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Central African Republic, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Erythrea, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Haiti, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jamaica, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|