| | Class 2 Device Recall Philips Expression IP5 Information Portal |  |
| Date Initiated by Firm | June 30, 2014 |
| Date Posted | August 19, 2014 |
| Recall Status1 |
Terminated 3 on December 14, 2016 |
| Recall Number | Z-2270-2014 |
| Recall Event ID |
68956 |
| 510(K)Number | K121424 |
| Product Classification |
Monitor,physiological,patient(without arrhythmia detection or alarms) - Product Code MWI
|
| Product | Philips IP5 Information Portal
The Expression Information Portal (Model IP5), a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professionals to remotely monitor the vital signs of patients undergoing MRI procedures. |
| Code Information |
Product: 865471 Number 453564251511 Serial Numbers: US23200121, US23200122, US23200123, US23200124, US23200125, US23200126, US23200127, US23200128, US23200129, US23200130, US23200131, US23200132, US23200133, US23200134, US23200135, US23200136, US23200137, US23200138, US23200139, US23200140, US23200141, US23200142, US23200143, US23200144, US23200145, US23200146, US23200147, US23200148, US23200149, US23200150, US23200151, US23200152, US23200153, US23200154, US23200155, US23200156, US23200157, US23200158, US23200159, US23200160, US23200161, US23200162, US23200163, US23200164, US23200165, US23200166, US23200167, US23200168, US23200169, US23200170, US23200172, US23200173, US23200176, US23200177, US23200178, US23200179, US23200180, US23200181, US23200182, US23200183, US23200184, US23200188, US23200189, US23200190, US23200191, US23200192, US23200193, US23200194, US23200195, US23200196, US23200197, US23200198, US23200199, US23200200, US23200201, US23200202, US23200203, US23200204, US23200205, US23200206, US23200207, US23200208, US23200209, US23200210, US23200211, US23200212, US23200213, US23200214, US23200215, US23200216, US23200217, US23200218, US23200219, US23200220, US23200221, US23200222, US23200223, US23200224, US23200225, US23200226, US23200227, US23200228, US23200229, US23200230, US23200231, US23200232, US23200233, US23200234, US23200235, US23200236, US23200237, US23200238, US23200239, US23200240, US23200241, US23200242, US23200243, US23200244, US23200245, US23200246, US23200247, US23200248, US23200249, US23200250, US23200251, US23200252, US23200253, US23200254, US23200255, US23200256, US23200257, US23200258, US23200259, US23200260, US23200261, US23200262, US23200263, US23200264, US23200265, US23200266, US23200267, US23200268, US23200269, US23200270, US23200271, US23200272, US23200273, US23200274, US23200275, US23200276, US23200277, US23200278, US23200279, US23200280, US23200281, US23200282, US23200283, US23200284, US23200285, US23200286, US23200287, US23200288, US23200289, US23200290, US23200291, US23200292, US23200293, US23200294, US23200295, US23200296, US23200297, US23200298, US23200299, US23200300, US23200301, US23200302, US23200303, US23200304, US23200305, US23200306, US23200307, US23200308, US23200309, US23200310, US23200311, US23200312, US23200313, US23200314, US23200315, US23200316, US23200317, US23200318, US23200319, US23200320, US23200321, US23200322, US23200323, US23200324, US23200325, US23200326, US23200327, US23200328, US23200329, US23200330, US23200331, US23200332, US23200333, US23200334, US23200335, US23200336, US23200337, US23200338, US23200339, US23200340, US23200341, US23200342, US23200343, US23200344, US23200345, US23200346, US23200347, US23200349, US23200350, US23200351, US23200352, US23200353, US23200354, US23200356, US23200357, US23200358, US23200359, US23200360, US23200361, US23200362, US23200363, US23200364, US23200365, US23200366, US23200367, US23200368, US23200369, US23200370, US23200371, US23200372, US23200373, US23200374, US23200375, US23200376, US23200377, US23200378, US23200379, US23200380, US23200381, US23200382, US23200383, US23200385, US23200386, US23200387, US23200388, US23200389, US23200390, US23200391, US23200393, US23200394, US23200395, US23200397, US23200398, US23200399, US23200400, US23200401, US23200402, US23200403, US23200405, US23200406, US23200407, US23200408, US23200409, US23200410, US23200411, US23200413, US23200415, US23200416, US23200419, US23200421, US23200422, US23200423, US23200424, US23200425, US23200427, US23200428, US23200429, US23200431, US23200434, US34600510, US34600511, US34600512, US34600513, US34600514, US34600515, US34600516, US34600517, US34600518, US34600519, US34600520, US34600521, US34600522, US34600523, US34600524, US34600525, US34600526, US34600528, US34600529, US34600530, US34600531, US34600532, US34600533, US34600534, US34600535, US34600536, US34600537, US34600538, US34600539, US34600540, US34600541, US34600542, US34600543, US34600544, US34600545, US34600546, US34600547, US34600548, US34600549, US34600550, US34600551, US34600552, US34600553, US34600554, US34600555, US34600556, US34600557, US34600558, US34600559, US34600560, US34600561, US34600562, US34600563, US34600564, US34600565, US34600566, US34600567, US34600568, US34600569, US34600570, US34600571, US34600572, US34600573, US34600574, US34600575, US34600576, US34600577, US34600578, US34600579, US34600580, US34600581, US34600582, US34600583, US34600584, US34600585, US34600586, US34600587, US34600588, US34600589, US34600590, US34600591, US34600592, US34600592, US34600593, US34600594, US34600595, US34600596, US34600597, US34600598, US34600599, US34600600, US34600601, US34600602, US34600603, US34600604, US34600605, US34600606, US34600607, US34600608, US34600609, US34600610, US34600611, US34600612, US34600613, US34600614, US34600615, US34600616, US34600617, US34600618, US34600619, US34600620, US34600621, US34600622, US34600623, US34600624, US34600625, US34600626, US34600627, US34600628, US34600629, US34600630, US34600631, US34600632, US34600633, US34600634, US34600635, US34600636, US34600637, US34600638, US34600639, US34600640, US34600641, US34600642, US34600643, US34600644, US34600645, US34600646, US34600647, US34600648, US34600649, US34600650, US34600651, US34600652, US34600653, US34600654, US34600655, US34600656, US34600657, US34600658, US34600659, US34600660, US34600661, US34600662, US34600663, US34600664, US34600665, US34600666, US34600667, US34600668, US34600669, US34600670, US34600671, US34600672, US34600673, US34600674, US34600675, US34600676, US34600677, US34600678, US34600679, US34600680, US34600681, US34600682, US34600683, US34600684, US34600685, US34600686, US34600687, US34600688, US34600689, US34600690, US34600691, US34600692, US34600693, US34600694, US34600695, US34600696, US34600697, US34600698, US34600699, US34600700, US34600701, US34600702, US34600703, US34600704, US34600705, US34600706, US34600707, US34600708, US34600709, US34600710, US34600711, US34600712, US34600713, US34600714, US34600715, US34600716, US34600717, US34600718, US34600719, US34600720, US34600721, US34600722, US34600723, US34600724, US34600729, US34600730, US34600731, US34600732, US34600733, US34600734, US34600735, US34600736, US34600737, US34600738, US34600739, US34600740, US34600741, US34600742, US34600743, US34600744, US34600745, US34600747, US34600748, US34600749, US34600750, US34600751, US34600752, US34600753, US34600754, US34600755, US34600756, US34600757, US34600758, US34600759, US34600760, US34600761, US34600762, US34600763, US34600764, US34600765, US34600766, US34600767, US34600768, US34600769, US34600770, US34600771, US34600772, US34600773, US34600774, US34600775, US34600776, US34600777, US34600778, US34600779, US34600780, US34600781, US34600782, US34600783, US34600784, US34600786, US34600787, US34600788, US34600789, US34600790, US34600791, US34600792, US34600794, US34600795, and US34600796 |
| FEI Number |
1051786
|
Recalling Firm/ Manufacturer |
Invivo Corporation 12151 Research Pkwy Suite 200 Orlando FL 32826-3222
|
| For Additional Information Contact | Debbie Levasseur 407-275-3220 Ext. 171 |
Manufacturer Reason for Recall | The wireless function of the IP5 radio could be interrupted causing loss of wireless communication between the IP5 and host monitoring system. |
FDA Determined Cause 2 | Process design |
| Action | The firm, Philips Healthcare, sent an "URGENT - Field Safety Notice" dated June 30, 2014 to its customers. The notice describes the product, problem and actions to be taken. The notice also provides the customers with instructions for use while waiting for the update to the IP5. A Philips representative will contact you regarding your affected device.
If you need any further information or support concerning this issue, please contact your local Philips representative: 877-468-4861 option 1, then option 2. |
| Quantity in Commerce | 573 devices. |
| Distribution | Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, MN, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, Hawaii, Alaska, and District of Columbia; and countries of: Australia, Austria, Denmark, France, Germany, Ireland, Netherlands, Norway, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MWI
|
|
|
|