| Class 2 Device Recall PDM ( patient data monitor) | |
Date Initiated by Firm | September 19, 2014 |
Date Posted | October 09, 2014 |
Recall Status1 |
Terminated 3 on September 13, 2016 |
Recall Number | Z-0060-2015 |
Recall Event ID |
69207 |
510(K)Number | K071073 |
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product | GE Healthcare Carescape Patient Data Module |
Code Information |
Mfg Lot or Serial # SA307020379GR SA307020383GR SA307020386GR SA307020391GR SA307020393GA SA307020396GA SA307020397GA SA307020399GR SA307020400GA SA307030405GR SA307030411GR SA307030412GR SA307030413GR SA307030417GA SA307030419GA SA307030420GA SA307030423GA SA307050429GA SA307070436GR SA307070438GR SA307070441GR SA307070442GR SA307070446GA SA307070447GR SA307070448GA SA307070449GR SA307070450GR SA307070453GR SA307070454GR SA307070455GR SA307080459GA SA307080462GR SA307080463GR SA307080466GA SA307080467GA SA307080469GA SA307080472GA SA307080473GR SA307080477GR SA307080478GR SA307080484GA SA307080488GA SA307080489GA SA307080491GA SA307080493GR SA307080497GA SA307080498GR SA307080499GA SA307080502GR SA307080504GR SA307080508GR SA307080509GA SA307080510GR SA307080513GA SA307080520GR SA307080521GA SA307080522GR SA307080524GA SA307080525GR SA307080526GR SA307080533GR SA307105145GA SA307110544GA SA307110548GR SA307110556GA SA307110559GR SA307110560GR SA307110572GR SA307110575GR SA307110579GR SA307120587GR SA307140598GA SA307140599GR SA307140606GA SA307140607GA SA307140610GA SA307140611GA SA307140614GA SA307140615GA SA307140618GR SA307140619GA SA307140621GA SA307140623GA SA307140627GA SA307140628GA SA307140631GA SA307140637GA SA307140640GR SA307140642GA SA307150645GA SA307150651GA SA307150655GA SA307150656GA SA307150671GR SA307150675GA SA307150681GR SA307180694GR SA307180695GA SA307180696GR SA307180697GR SA307180698GA SA307180699GR SA307180701GA SA307180702GR SA307180704GR SA307180705GR SA307180706GR SA307180707GA SA307180708GA SA307190717GR SA307190722GA SA307190727GA SA307190729GR SA307190730GR SA307190731GR SA307190732GR SA307190733GA SA307190734GR SA307190736GR SA307190737GR SA307190738GR SA307190739GR SA307190741GR SA307190742GR SA307190743GR SA307190744GR SA307190745GR SA307190749GR SA307190750GR SA307190751GA SA307190752GA SA307190753GA SA307190754GA SA307190755GR SA307190756GR SA307190757GR SA307190759GA SA307190760GA SA307190761GA SA307190762GA SA307190764GR SA307190768GR SA307190769GR SA307190770GR SA307190771GR SA307190772GA SA307190773GA SA307190774GR SA307190775GR SA307200776GA SA307200777GR SA307200784GR SA307200788GR SA307200791GA SA307200792GA SA307200793GA SA307200794GA SA307200795GA SA307200796GA SA307200797GA SA307200798GA SA307200799GA SA307200800GR SA307200801GA SA307200802GA SA307200803GA SA307200804GA SA307200805GA SA307210815GA SA307210818GA SA307210819GR SA307210820GR SA307210821GA SA307210824GA SA307210825GA SA307210826GA SA307210827GA SA307210828GR SA307210829GA SA307210832GA SA307210833GA SA307210836GR SA307210837GA SA307210838GA SA307210839GR SA307210841GR SA307210843GA SA307210858GA SA307210863GA SA307220869GA SA307220870GA SA307220872GA SA307220875GA SA307220877GA SA307220878GA SA307220879GA SA307220880GR SA307220888GA SA307220890GR SA307220892GR SA307220893GA SA307220898GR SA307220900GR SA307220901GR SA307220902GA SA307220903GR SA307220904GR SA307220908GA SA307220914GR SA307220920GR SA307220923GA SA307220931GR SA307220932GR SA307220935GA SA307220936GA SA307220937GA SA307220938GA SA307220941GA SA307220942GA SA307220944GA SA307220945GA SA307220946GA SA307220947GR SA307220948GA SA307220950GA SA307220951GA SA307220952GA SA307220953GA SA307220954GA SA307220955GA SA307220956GA SA307220957GA SA307220958GA SA307220959GA SA307220961GA SA307220963GA SA307220978GA SA307220983GR SA307220986GA SA307220987GA SA307220988GA SA307220990GR SA307220996GR SA307220999GR SA307221000GA SA307221001GA SA307221003GA SA307221004GA SA307221006GA SA307221007GR SA307221008GA SA307221009GA SA307221010GA SA307221011GA SA307221012GA SA307221013GR SA307221014GA SA307221016GA SA307221018GA SA307221026GR SA307221027GA SA307221028GA SA307221030GA SA307221033GA SA307221035GA SA307221036GA SA307221037GA SA307221038GA SA307221039GA SA307221040GA SA307221041GA SA307221042GA SA307221043GA SA307221045GA SA307221047GA SA307221048GA SA307221049GA SA307221051GR SA307221052GA SA307221055GA SA307221056GA SA307221057GA SA307221058GR SA307221059GA SA307221060GA SA307221061GA SA307221064GA SA307221065GA SA307221066GA SA307221068GA SA307221069GA SA307221070GA SA307221071GA SA307221072GA SA307221076GA SA307221077GA SA307221078GA SA307221079GA SA307221080GA SA307221081GR SA307221084GA SA307221089GA SA307221092GA SA307221096GR SA307221097GA SA307221100GA SA307221101GA SA307221102GA SA307221103GA SA307221104GR SA307221106GR SA307221107GR SA307221108GA SA307221109GA SA307221110GA SA307221113GA SA307221114GR SA307221115GR SA307221116GA SA307221117GA SA307221118GA SA307221119GA SA307221120GR SA307221121GA SA307221122GA SA307221123GA SA307221124GA SA307221125GR SA307221126GR SA307221127GA SA307221128GA SA307221129GR SA307221131GA SA307221132GR SA307221133GA SA307221134GA SA307221135GA SA307221136GA SA307221138GA SA307221139GA SA307221140GA SA307221141GA SA307221142GA SA307221144GA SA307221145GA SA307221152GA SA307221154GR SA307221156GR SA307221158GA SA307221160GA SA307221161GA SA307221162GA SA307221163GA SA307221164GA SA307221165GA SA307221168GR SA307221170GA SA307221171GR SA307221172GR SA307221173GR SA307221175GA SA307221176GR SA307221177GA SA307221178GA SA307221179GA SA307221180GA SA307221181GA SA307221183GR SA307221186GA SA307231191GA SA307231194GA SA307231196GA SA307231197GA SA307231199GR SA307231200GA SA307231201GA SA307231202GA SA307231203GA SA307231204GA SA307231205GR SA307231207GA SA307231209GR SA307231210GR SA307231211GA SA307231212GR SA307231215GA SA307231217GA SA307231218GA SA307231219GA SA307231220GA SA307231222GA SA307231223GA SA307231224GA SA307231225GA SA307231226GA SA307231227GA SA307231228GA SA307231229GA SA307231230GA SA307231231GA SA307231232GA SA307231233GA SA307231234GA SA307231235GA SA307231236GA SA307231237GA SA307231240GA SA307231243GA SA307231244GA SA307231245GA SA307231246GA SA307231247GA SA307231248GA SA307231250GA SA307231251GA SA307231252GA SA307231255GA SA307231256GR SA307231257GR SA307231259GA SA307231260GA SA307231262GR SA307231265GA SA307231266GA SA307231267GA SA307231268GA SA307231270GA SA307231271GA SA307231274GR SA307231275GR SA307231276GA SA307231277GA SA307231278GA SA307231279GA SA307231281GA SA307231282GA SA307231285GA SA307231286GA SA307231291GA SA307231292GA SA307231294GA SA307231295GA SA307231296GA SA307231298GA SA307231299GA SA307231300GA SA307231301GA SA307231302GA SA307231303GA SA307231304GA SA307231305GA SA307231306GA SA307231307GA SA307231310GA SA307231311GA SA307231313GA SA307231314GA SA307231315GA SA307231316GA SA307231317GA SA307231318GA SA307231319GA SA307231321GA SA307231322GA SA307231323GA SA307231324GA SA307231325GA SA307231329GA SA307231330GA SA307231333GA SA307231334GA SA307231335GA SA307231336GA SA307231337GA SA307231338GA SA307231341GA SA307231343GA SA307231344GA SA307231345GA SA307231346GA SA307231347GA SA307231348GA SA307231349GA SA307231350GA SA307231352GR SA307231355GA SA307231356GA SA307231359GA SA307231362GR SA307231363GA SA307231364GA SA307231365GA SA307231366GR SA307231368GA SA307231370GR SA307231372GA SA307231373GA SA307231375GA SA307231376GR SA307231381GA SA307231382GA SA307231383GA SA307231384GA SA307231385GR SA307231386GA SA307231387GA SA307231388GA SA307231389GA SA307231390GA SA307231391GA SA307231392GA SA307231393GA SA307231394GR SA307231395GA SA307231396GA SA307231397GR SA307231398GA SA307231399GA SA307231402GA SA307231403GA SA307231404GR SA307231405GA SA307231406GR SA307231408GA SA307231410GA SA307231411GA SA307231412GA SA307231413GR SA307231414GR SA307231416GA SA307231417GA SA307231418GA
SA307231420GA SA307231422GA SA307231423GR SA307231424GR SA307231426GA SA307231427GA SA307231428GR SA307231429GA SA307231430GA SA307231431GR SA307231432GA SA307231433GA SA307231434GA SA307231435GA SA307231436GA SA307231437GA SA307231438GA SA307231439GA SA307231440GA SA307231442GA SA307231443GA SA307231444GR SA307231445GR SA307231448GA SA307231449GA SA307231450GA SA307231451GA SA307231453GA SA307231454GA SA307231455GA SA307231456GA SA307231457GA SA307231458GA SA307231459GA SA307231462GA SA307231463GR SA307231464GA SA307231465GA SA307231466GA SA307231467GA SA307231468GA SA307231469GA SA307231470GA SA307231471GA SA307231472GA SA307231473GR SA307231475GR SA307231476GR SA307231477GA SA307231478GA SA307231479GR SA307231483GA SA307231484GA SA307231485GA SA307231486GA SA307231488GA SA307231489GA SA307231490GA SA307231491GA SA307231492GA SA307231494GA SA307231495GA SA307231496GA SA307231497GA SA307231498GA SA307231499GA SA307231500GA SA307231501GA SA307231502GA SA307241503GA SA307241504GA SA307241505GA SA307241506GA SA307241507GA SA307241510GR SA307241512GR SA307241513GR SA307241518GA SA307241519GR SA307241521GA SA307241523GA SA307241527GA SA307241528GA SA307241529GA SA307241530GA SA307241532GR SA307241533GA SA307241536GA SA307241538GA SA307241539GA SA307241540GA SA307241541GA SA307241542GA SA307241544GA SA307241546GA SA307241547GA SA307241549GA SA307241550GA SA307241551GA SA307241556GR SA307241557GA SA307241559GR SA307241560GA SA307241561GA SA307241562GA SA307241563GA SA307241565GA SA307241567GA SA307241568GA SA307241573GA SA307241574GA SA307241575GA SA307241577GR SA307241578GR SA307241579GA SA307241581GA SA307241583GR SA307241585GA SA307241586GR SA307241587GA SA307241588GA SA307241589GA SA307241590GA SA307241591GA SA307241592GA SA307241594GA SA307241600GA SA307241602GA SA307241604GA SA307241605GA SA307241608GA SA307241610GA SA307241611GA SA307241612GA SA307241615GA SA307241618GR SA307241621GA SA307241622GA SA307241623GA SA307241625GA SA307241626GA SA307241628GA SA307241630GA SA307241632GA SA307241633GA SA307241640GA SA307241668GR SA307241670GA SA307241673GR SA307241674GR SA307241676GR SA307241677GR SA307241680GA SA307241681GA SA307241682GR SA307241683GA SA307241685GA SA307241686GR SA307241689GA SA307241690GR SA307241691GR SA307241695GR SA307241697GA SA307241701GR SA307241705GA SA307241706GA SA307241708GR SA307251710GR SA307251713GR SA307251715GR SA307251716GR SA307251718GA SA307251719GR SA307251721GR SA307251722GA SA307251724GA SA307251726GR SA307251727GR SA307251729GR SA307251734GA SA307251735GR SA307251736GA SA307251738GA SA307251740GR SA307251741GR SA307251742GA SA307251743GR SA307251744GA SA307251745GA SA307251746GR SA307251747GR SA307251748GA SA307251750GA SA307251751GA SA307251752GA SA307251753GA SA307251754GA SA307251755GA SA307251756GA SA307251757GR SA307251758GA SA307251759GA SA307251760GA SA307251762GA SA307251763GA SA307251764GA SA307251765GA SA307251766GA SA307251767GA SA307251768GA SA307251769GA SA307251770GA SA307251771GA SA307251772GA SA307251773GA SA307251774GA SA307251775GA SA307251776GA SA307251777GA SA307251778GR SA307251779GA SA307251780GA SA307251781GA SA307251782GR SA307251783GA SA307251784GR SA307251785GR SA307251786GA SA307251790GA SA307251791GR SA307251794GR SA307251795GR SA307251796GR SA307251797GR SA307251798GR SA307251800GR SA307251801GA SA307251803GA SA307261811GA SA307261813GA SA307261814GR SA307261816GA SA307261817GA SA307261818GA SA307261820GA SA307261821GA SA307261822GA SA307261823GA SA307261824GA SA307261825GA SA307261826GA SA307261827GA SA307261828GR SA307261829GA SA307261830GA SA307261831GA SA307261836GR SA307261837GA SA307261839GA SA307261841GA SA307261842GR SA307261843GA SA307261844GA SA307261845GA SA307261847GR SA307291862GA SA307291863GA SA307291864GA SA307291869GA SA307291872GA SA307291874GA SA307291876GA SA307291877GA SA307291878GA SA307291879GA SA307291880GA SA307291881GA SA307291882GA SA307291883GA SA307291884GA SA307291886GA SA307291887GA SA307291894GA SA307291897GR SA307291898GA SA307291899GR SA307291900GA SA307291901GA SA307291905GR SA307291907GR SA307291910GR SA307291911GR SA307291912GA SA307291918GR SA307291920GR SA307291924GA SA307291925GA SA307291926GA SA307291928GA SA307291929GA SA307291930GA SA307291933GA SA307291934GA SA307291935GR SA307291936GA SA307291943GA SA307291946GA SA307291950GA SA307291951GA SA307291955GR SA307291959GA SA307291961GA SA307291962GA SA307291967GA SA307291968GA SA307291969GA SA307291973GA SA307291975GA SA307291976GA SA307291977GR SA307291978GA SA307291979GA SA307291980GA SA307291981GA SA307291983GA SA307291984GA SA307301988GA SA307301992GA SA307301993GA SA307301994GA SA307301996GA SA307301998GA SA307302000GA SA307302002GA SA307302003GA SA307302004GA SA307302006GA SA307302008GA SA307302010GA SA307302011GA SA307302012GA SA307302015GA SA307302016GR SA307302017GA SA307302021GA SA307302022GA SA307302023GA SA307302028GA SA307302035GA SA307302043GR SA307302046GA SA307302047GA SA307302050GA SA307302051GA SA307302052GA SA307302053GA SA307302054GA SA307302055GA SA307302056GA SA307302057GA SA307302059GA SA307302063GA SA307302064GA SA307302066GR SA307302067GA SA307302068GA SA307302070GA SA307302071GA SA307302072GA SA307302074GA SA307302075GA SA307302076GA SA307302077GA SA307302078GA SA307302079GA SA307302080GA SA307302081GA SA307302082GA SA307302083GA SA307302084GA SA307302085GA SA307312086GA SA307312087GA SA307312088GA SA307312089GA SA307312091GA SA307312092GA SA307312093GA SA307312094GA SA307312095GA SA307312096GA SA307312097GA SA307312098GA SA307312099GA SA307312100GA SA307312101GA SA307312102GA SA307312103GA SA307312106GA SA307312108GR SA307312109GA SA307322123GA SA307322125GA SA307322126GR SA307322133GR SA307322146GR SA307322147GA SA307322150GA SA307322152GA SA307322153GA SA307322155GA SA307322156GA SA307322157GA SA307322158GA SA307322165GR SA307322172GA SA307322176GA SA307322179GA SA307322183GA SA307322184GA SA307322185GA SA307322188GA SA307322189GA SA307322194GA SA307322195GA SA307322196GA SA307332206GA SA307332209GR SA307332210GA SA307332213GR SA307332215GR SA307332216GR SA307332221GR SA307332223GA SA307332224GR SA307332225GA SA307332227GR SA307332228GR SA307332229GR SA307332230GA SA307332231GA SA307332232GA SA307332233GA SA307332234GA SA307332235GA SA307332236GA SA307332237GA SA307332239GA SA307332240GA SA307332241GR SA307332243GR SA307332244GA SA307332245GR SA307332246GR SA307332247GR SA307332249GA SA307332250GR SA307332251GA SA307332252GA SA307332253GA SA307332254GA SA307332255GA SA307332256GA SA307332257GA SA307332258GA SA307332259GA SA307332260GA SA307332261GA SA307332262GA SA307332263GA SA307332264GA SA307332266GA SA307332267GA SA307332268GA SA307332269GA SA307332270GA SA307332271GA SA307332272GA SA307332273GA SA307332274GA SA307332275GA SA307332276GA SA307332277GA SA307332278GA SA307332279GA SA307332280GR SA307332281GA SA307332282GA SA307332284GA SA307332285GA SA307332286GA SA307332287GA SA307332289GA SA307332290GA SA307332292GA SA307342296GA SA307342297GA SA307342298GA SA307342299GA SA307342300GA SA307342301GA SA307342302GA SA307342305GA SA307342306GA SA307342307GA SA307342308GA SA307342309GA SA307342311GA SA307342312GR SA307342313GR SA307342314GA SA307342315GA SA307342316GA SA307342318GA SA307342319GA SA307342320GA SA307342321GA SA307342322GA SA307342323GA SA307342324GA SA307342325GA SA307342328GA SA307342329GA SA307342330GA SA307342331GA SA307342332GA SA307342333GA SA307342334GA SA307342335GA SA307342336GA SA307342337GA SA307342338GA SA307342339GA SA307342340GA SA307342341GA SA307342342GA SA307342343GA SA307342344GA SA307342345GA SA307342346GA SA307342347GA SA307342348GA SA307342349GA SA307342350GA SA307352351GA SA307352352GA SA307352353GA SA307352354GA SA307352355GA SA307352356GA SA307352357GA SA307352360GA SA307352363GA SA307352364GA SA307352368GA SA307352369GA SA307352370GA SA307352373GA SA307352374GA SA307362375GA SA307362376GA SA307362377GA SA307362378GA SA307362379GA SA307362380GA SA307362381GA SA307362382GA SA307362383GA SA307362384GA SA307362385GA SA307362387GA SA307362391GR SA307362392GR SA307362393GA SA307362394GA SA307362395GA SA307362396GA SA307362398GA SA307362399GA SA307362402GA SA307362403GA SA307362404GA SA307362405GA SA307362406GA SA307362407GA SA307362408GR SA307362412GR SA307362413GR SA307362414GA SA307362415GA SA307362417GA SA307362419GA SA307362421GA SA307362423GA SA307362429GA SA307362430GA SA307362434GA SA307362435GA SA307362439GA SA307362440GA SA307362441GA SA307362442GA SA307382446GA SA307382447GA SA307382448GR SA307382449GA SA307382450GA SA307382451GR SA307382453GR SA307382454GR SA307382458GA SA307382460GA SA307382461GR SA307382463GA SA307382464GA SA307382465GR SA307382466GA SA307382469GA SA307382470GA SA307382471GR SA307382472GA SA307382473GA SA307382475GA SA307382476GR SA307382477GA SA307382478GR SA307382479GA SA307382480GA SA307382482GA SA307382483GA SA307382484GA SA307382486GA SA307382488GA SA307382489GA SA307382490GR SA307382491GA SA307382492GA SA307382493GA SA307382494GA SA307382495GA SA307382497GA SA307382499GA SA307382500GA SA307382502GA SA307382504GA SA307382505GA SA307382506GR SA307382507GA SA307382510GA SA307382511GA SA307382516GA SA307382517GA SA307382518GA SA307382520GA SA307382524GA SA307382526GA SA307382527GA SA307382528GA SA307382530GA SA307382531GA SA307382534GA SA307382535GA SA307382536GA SA307382537GA SA307382538GA SA307382539GA SA307382540GA SA307382541GA SA307382542GA SA307382544GA SA307382545GA SA307382546GA SA307382547GA SA307382549GA SA307382550GA SA307382551GA SA307382552GA SA307382553GA SA307382554GA SA307382555GA SA307382557GA SA307382558GR SA307382559GA SA307382560GA SA307382561GA SA307382563GA SA307382564GA SA307382565GA SA307382566GA SA307382567GA SA307382568GA SA307382569GR SA307382573GA SA307382574GA SA307382575GA SA307382576GA SA307382577GA SA307382578GA SA307382579GA SA307382581GA SA307382582GA SA307382583GA SA307382584GA SA307382586GA SA307382587GR SA307382589GA SA307382591GA SA307382593GA SA307382594GA SA307382595GA SA307382597GA SA307382599GA SA307382600GR SA307382601GA SA307382609GA SA307382610GA SA307382611GA SA307382612GA SA307382613GA SA307382615GA SA307382616GA SA307382618GA SA307382620GA SA307382621GA SA307382622GA SA307382623GA SA307382624GA SA307382625GA SA307382626GA SA307382627GA SA307382629GR SA307382630GA SA307382631GA SA307382632GR SA307382633GA SA307382634GA SA307382635GR SA307382636GA SA307382637GR SA307382638GA SA307382639GR SA307382640GA SA307382641GA SA307382642GR SA307382644GA SA307382645GR SA307382646GA SA307382647GR SA307382648GA SA307382649GA SA307382650GA SA307382651GR SA307382652GR SA307382653GA SA307382654GA SA307382655GA SA307382656GA SA307382657GA SA307382658GA SA307382660GA SA307382661GA SA307382662GR SA307382663GA SA307382664GR SA307382665GA SA307382666GA SA307392667GA SA307392669GA SA307392671GR SA307392672GR SA307392673GA SA307392674GA SA307392675GR SA307392676GA SA307392677GA SA307392678GA SA307392679GA SA307392680GA SA307392681GA SA307392682GA SA307392683GA SA307392684GA SA307392685GA SA307392686GA SA307392688GA SA307392690GA SA307392691GA SA307392692GA SA307392694GA SA307392695GA SA307392696GA SA307392697GA SA307392698GA SA307392699GA SA307392700GA SA307392701GA SA307392702GA SA307392703GA SA307392704GA SA307392705GA
SA307392706GA SA307392707GA SA307392708GA SA307392709GA SA307392710GA SA307392711GA SA307392714GA SA307392715GA SA307392717GA SA307392718GA SA307392719GA SA307392720GA SA307392721GA SA307392722GA SA307392724GA SA307392725GA SA307392726GA SA307392727GA SA307392728GA SA307392729GA SA307392730GA SA307392731GA SA307392732GA SA307392734GA SA307392735GA SA307392736GA SA307392737GA SA307392738GA SA307392739GA SA307392740GA SA307392741GA SA307392743GA SA307392744GA SA307392745GA SA307392746GA SA307392747GA SA307392748GA SA307392749GA SA307392750GA SA307392751GA SA307392752GA SA307392753GA SA307392754GA SA307392758GA SA307392759GA SA307392760GA SA307392761GA SA307392762GA SA307392763GA SA307392764GA SA307392765GA SA307392766GA SA307392768GA SA307392769GA SA307392820GA SA307392821GA SA307392822GA SA307392824GA SA307392825GA SA307392826GA SA307392827GA SA307392828GA SA307392829GR SA307392830GA SA307392832GA SA307392833GA SA307392834GA SA307392835GA SA307392836GA SA307392837GA SA307392838GA SA307392839GA SA307392840GA SA307392842GA SA307392843GA SA307392844GA SA307392845GA SA307392846GA SA307392847GA SA307392849GA SA307392850GA SA307392851GA SA307392852GA SA307392853GA SA307392854GA SA307392855GA SA307392856GA SA307392857GA SA307392858GA SA307392860GA SA307392861GA SA307392862GA SA307392863GA SA307392864GR SA307392866GA SA307392867GA SA307392869GA SA307392932GA SA307392933GA SA307392934GA SA307392936GA SA307392937GA SA307392938GA SA307392939GA SA307392941GA SA307392943GA SA307392944GA SA307392947GA SA307392948GA SA307392949GA SA307392950GA SA307392951GA SA307392953GA SA307392954GA SA307392955GA SA307392956GA SA307392957GA SA307392958GA SA307392959GA SA307392960GA SA307392961GA SA307392962GA SA307392963GA SA307392964GA SA307392965GA SA307392967GA SA307392969GA SA307392970GA SA307392973GA SA307392974GA SA307392975GA SA307392976GA SA307392977GA SA307392978GA SA307392979GA SA307392980GA SA307392982GA SA307392983GA SA307392984GA SA307392986GA SA307392987GA SA307392988GA SA307392989GR SA307392991GA SA307392994GA SA307392996GA SA307392997GA SA307392998GA SA307393001GA SA307393003GA SA307393004GA SA307393005GA SA307393007GA SA307393011GA SA307393012GA SA307393014GA SA307393016GA SA307393018GA SA307393019GA SA307393020GA SA307393021GA SA307393022GA SA307393024GA SA307393025GA SA307393026GA SA307393031GA SA307393032GA SA307393035GA SA307403040GA SA307403042GA SA307403044GA SA307403047GA SA307403048GA SA307403049GA SA307403051GA SA307403052GA SA307403053GA SA307403055GA SA307403057GA SA307403060GR SA307403061GA SA307403062GA SA307403063GA SA307403067GA SA307403068GA SA307403070GA SA307403071GA SA307403072GA SA307403074GA SA307403075GA SA307403079GA SA307403082GA SA307403087GA SA307403101GA SA307403103GA SA307403104GA SA307403106GA SA307403108GA SA307403110GA SA307403112GA SA307403115GA SA307403116GA SA307403120GA SA307403123GA SA307403124GA SA307403127GA SA307403128GA SA307403131GA SA307403134GA SA307403135GA SA307403136GA SA307403137GA SA307403138GA SA307403139GA SA307403140GR SA307403141GA SA307403146GA SA307403149GA SA307403152GA SA307413100GA SA307413101GA SA307413103GA SA307413104GA SA307413105GA SA307413106GR SA307413107GA SA307413108GA SA307413111GA SA307413112GA SA307413113GA SA307413114GA SA307413115GA SA307413116GA SA307413117GA SA307413118GA SA307413119GA SA307413120GA SA307413122GA SA307413124GA SA307423126GA SA307423127GA SA307423128GA SA307423129GA SA307423130GA SA307423131GA SA307423133GA SA307423135GA SA307423136GA SA307423138GA SA307423140GA SA307423141GA SA307423142GA SA307423145GA SA307423146GA SA307423148GA SA307423149GA SA307423150GA SA307423151GA SA307423152GA SA307423153GA SA307423154GA SA307423156GR SA307423157GA SA307423161GA SA307423165GA SA307423167GA SA307423168GA SA307423170GA SA307423175GA SA307423176GA SA307423178GA SA307423181GA SA307443185GA SA307443186GA SA307443187GA SA307443191GA SA307443192GA SA307443193GA SA307443194GA SA307443196GA SA307443197GA SA307443198GA SA307443200GA SA307443202GA SA307443204GA SA307443206GA SA307443207GA SA307443208GR SA307443209GA SA307443210GA SA307443211GA SA307443212GA SA307443214GA SA307443215GA SA307443216GA SA307443217GA SA307443218GA SA307443221GA SA307443224GA SA307443225GA SA307443226GA SA307443227GA SA307443228GA SA307443229GA SA307443230GA SA307443231GA SA307443232GA SA307443233GA SA307443234GA SA307443235GA SA307443236GA SA307443237GA SA307443239GA SA307443240GA SA307443241GA SA307443244GA SA307443245GA SA307443246GA SA307443248GA SA307443250GA SA307443251GA SA307443252GA SA307443253GA SA307443254GA SA307443255GA SA307443256GA SA307443258GA SA307443259GA SA307443260GA SA307443261GA SA307443262GR SA307443263GA SA307443264GA SA307443265GA SA307443266GA SA307443268GA SA307443269GA SA307443270GA SA307443271GA SA307443272GA SA307443273GA SA307443305GA SA307443307GA SA307443309GA SA307443310GA SA307443312GA SA307443313GA SA307443314GA SA307443315GA SA307443316GA SA307443318GA SA307443320GA SA307443324GA SA307443325GR SA307443326GA SA307443327GA SA307443328GA SA307443330GA SA307443332GA SA307443333GA SA307443335GA SA307443340GA SA307443341GA SA307443342GA SA307443344GA SA307443345GA SA307443346GA SA307443348GR SA307443349GA SA307443355GA SA307443358GA SA307443359GA SA307443360GA SA307443361GR SA307443362GA SA307443363GA SA307443365GA SA307443366GA SA307443367GA SA307443368GA SA307443369GA SA307443370GA SA307443371GR SA307443372GA SA307443373GA SA307443375GA SA307443376GA SA307443377GA SA307443379GA SA307443380GA SA307443381GA SA307443382GA SA307443383GA SA307443384GA SA307443385GA SA307443386GR SA307443387GA SA307453390GA SA307453391GA SA307453392GA SA307453398GA SA307453399GA SA307453400GR SA307453401GA SA307453402GA SA307453404GA SA307453405GA SA307453407GA SA307453409GA SA307453411GA SA307453413GA SA307453414GA SA307453415GA SA307453416GA SA307453417GR SA307453418GA SA307453419GA SA307453420GR SA307453421GA SA307453423GA SA307453424GA SA307453425GA SA307453426GA SA307453427GA SA307453428GR SA307453429GA SA307453430GR SA307453435GA SA307453436GA SA307453438GR SA307453439GR SA307453440GA SA307453441GA SA307453442GA SA307453446GA SA307453447GA SA307453450GR SA307453451GA SA307453452GR SA307453454GA SA307453455GR SA307453456GA SA307453457GA SA307453459GA SA307453460GR SA307453462GR SA307453463GA SA307453466GA SA307453467GA SA307453468GR SA307453470GA SA307453471GR SA307453473GR SA307453476GR SA307453477GA SA307453478GR SA307453479GA SA307453480GA SA307453482GA SA307453484GR SA307453485GA SA307453486GA SA307453487GA SA307453488GA SA307453490GA SA307453491GA SA307453492GA SA307453493GA SA307453494GA SA307453495GA SA307453496GR SA307453497GR SA307453519GA SA307453521GA SA307453522GA SA307453523GA SA307453524GA SA307453525GA SA307453526GA SA307453527GA SA307453528GA SA307453529GA SA307453531GA SA307453532GA SA307453533GA SA307453534GA SA307453535GA SA307453536GA SA307453537GA SA307453538GA SA307453540GA SA307453541GA SA307453542GA SA307453543GA SA307453545GA SA307453547GA SA307453548GA SA307453549GA SA307453551GA SA307453556GA SA307453557GR SA307453558GA SA307453559GA SA307453561GA SA307453562GR SA307453563GR SA307453564GA SA307453567GA SA307453569GA SA307453570GA SA307453571GA SA307453572GA SA307453573GA SA307453574GA SA307453575GA SA307453578GA SA307453579GA SA307453580GA SA307453581GA SA307453582GA SA307453583GA SA307453584GA SA307453585GA SA307453586GA SA307453587GA SA307453588GR SA307453589GA SA307453592GA SA307453593GA SA307453594GA SA307453597GR SA307453598GA SA307453599GR SA307453600GR SA307453601GA SA307453602GR SA307453603GA SA307453604GA SA307453606GA SA307453608GA SA307453609GA SA307453610GA SA307453611GA SA307453612GA SA307453613GR SA307453614GR SA307453615GA SA307453616GR SA307453617GA SA307453618GA SA307463649GA SA307463650GA SA307463651GA SA307463652GR SA307463653GA SA307463654GA SA307463655GA SA307463656GA SA307463657GA SA307463658GA SA307483659GA SA307483660GA SA307483661GA SA307483662GA SA307483663GA SA307483664GA SA307483666GA SA307483667GA SA307483668GA SA307483669GA SA307483670GA SA307483671GA SA307483672GA SA307483673GA SA307483674GA SA307483675GA SA307483676GA SA307483677GA SA307483678GA SA307483679GA SA307483680GA SA307483681GA SA307483682GA SA307483683GA SA307483684GA SA307483685GA SA307483686GA SA307483687GA SA307483688GA SA307483689GA SA307483690GA SA307483691GA SA307483692GA SA307483693GA SA307483694GA SA307483695GA SA307483696GA SA307483697GA SA307483698GA SA307483699GA SA307483700GA SA307483701GA SA307483702GA SA307483703GA SA307483704GA SA307483705GA SA307483706GA SA307483707GA SA307483709GA SA307483710GA SA307483711GA SA307483712GA SA307483713GA SA307483714GA SA307483715GA SA307483716GA SA307483717GA SA307483719GA SA307483720GA SA307483721GA SA307483722GA SA307483723GA SA307483724GA SA307483725GA SA307483726GA SA307483727GA SA307483728GA SA307483729GA SA307483730GA SA307483731GA SA307483732GA SA307483733GA SA307483734GA SA307483735GA SA307483736GA SA307483737GA SA307483738GA SA307483739GA SA307483740GA SA307483741GA SA307483742GA SA307483743GA SA307483744GA SA307483745GA SA307483746GA SA307483747GA SA307483749GA SA307483750GA SA307483751GA SA307483752GA SA307483753GA SA307483754GA SA307483755GA SA307483756GA SA307483757GA SA307483758GA SA307483759GA SA307483760GA SA307483761GA SA307483762GA SA307483763GA SA307483764GA SA307483765GA SA307483766GA SA307483767GA SA307483768GA SA307483769GA SA307483770GA SA307483771GA SA307483772GA SA307483773GA SA307483774GA SA307483775GA SA307483776GA SA307483777GA SA307483778GA SA307483780GR SA307483782GA SA307483783GA SA307483784GA SA307483786GA SA307483788GA SA307483789GA SA307483790GA SA307483792GA SA307483794GA SA307483795GA SA307483796GA SA307483797GA SA307483798GA SA307483800GA SA307483801GA SA307483803GA SA307483804GA SA307483805GA SA307483806GA SA307483808GA SA307483809GA SA307483810GA SA307483811GA SA307483812GA SA307483813GA SA307483814GA SA307483816GA SA307483817GA SA307483818GA SA307483819GA SA307483821GA SA307483822GA SA307483823GA SA307483824GA SA307483825GA SA307483826GA SA307483827GA SA307483828GA SA307483829GA SA307483830GA SA307483831GA SA307483832GR SA307483834GA SA307483835GA SA307483836GA SA307483837GA SA307483838GA SA307483840GA SA307483841GA SA307483843GA SA307493845GA SA307493846GA SA307493847GA SA307493848GA SA307493849GA SA307493850GA SA307493851GA SA307493853GA SA307493854GA SA307493855GA SA307493856GA SA307493857GA SA307493858GA SA307493859GA SA307493860GA SA307493861GA SA307493863GA SA307493864GA SA307493865GA SA307493866GA SA307493867GA SA307493868GA SA307493869GA SA307493870GA SA307493871GA SA307493872GA SA307493873GA SA307493874GA SA307493875GA SA307493876GA SA307493877GA SA307493878GA SA307493879GA SA307493880GA SA307493881GA SA307493882GA SA307493883GA SA307493884GA SA307493887GA SA307493888GA SA307493890GA SA307493891GA SA307493892GA SA307493893GA SA307493894GA SA307493895GA SA307493896GA SA307493897GA SA307493898GA SA307493899GA SA307493900GA SA307493901GA SA307493902GA SA307493903GA SA307493904GA SA307493905GA SA307493906GA SA307493907GA SA307493908GA SA307493909GA SA307493910GA SA307493911GA SA307493912GA SA307493913GA SA307493914GA SA307493915GA SA307493916GA SA307493917GA SA307493918GA SA307493919GA SA307493920GA SA307493921GA SA307493922GA
SA307493923GA SA307493924GA SA307493925GA SA307493926GA SA307493927GA SA307493928GA SA307493929GA SA307493930GA SA307493931GA SA307493932GA SA307493933GA SA307493934GA SA307493935GA SA307493936GA SA307493937GA SA307493938GA SA307493939GA SA307493940GA SA307493941GA SA307493943GA SA307493944GA SA307493945GA SA307493946GA SA307493947GA SA307493948GA SA307493949GA SA307493950GA SA307493951GA SA307493952GA SA307493953GA SA307493954GA SA307493955GA SA307493956GA SA307493957GA SA307493958GA SA307493959GA SA307493960GA SA307493961GA SA307493962GA SA307493963GA SA307493964GA SA307493965GA SA307493966GA SA307493967GA SA307493969GA SA307493970GA SA307493971GA SA307493972GA SA307493973GA SA307493974GA SA307493975GA SA307493976GA SA307493977GA SA307493980GA SA307493981GA SA307493982GA SA307493983GA SA307493984GA SA307493985GA SA307493987GA SA307493988GA SA307493989GR SA307493990GA SA307493991GA SA307493992GA SA307493994GA SA307493995GA SA307493996GA SA307493997GA SA307493999GA SA307494000GA SA307494001GA SA307494002GA SA307494003GA SA307494004GA SA307494005GA SA307494006GA SA307494007GA SA307494008GA SA307494009GA SA307494010GA SA307494011GA SA307494012GA SA307494013GA SA307494014GR SA307494016GA SA307494017GA SA307494018GA SA307494019GA SA307494020GA SA307494021GA SA307494022GA SA307494023GA SA307494024GA SA307494025GA SA307494026GA SA307494027GA SA307494028GA SA307494029GA SA307494030GA SA307494031GA SA307494032GA SA307494033GA SA307494034GA SA307494035GA SA307494037GA SA307494038GA SA307494039GA SA307494040GA SA307494042GA SA307494043GA SA307494044GA SA307494045GA SA307494047GA SA307494048GA SA307494049GA SA307494050GA SA307494051GA SA307504066GA SA307504067GA SA307504068GA SA307504069GA SA307504070GA SA307504071GA SA307504072GA SA307504073GA SA307504074GA SA307504075GA SA307504076GA SA307504077GA SA307504078GA SA307504079GA SA307504080GA SA307504081GA SA307504082GA SA307504083GA SA307504084GA SA307504085GA SA307504086GA SA307504087GA SA307504089GA SA307504090GA SA307504092GA SA307504093GA SA307504094GA SA307504095GA SA307504096GA SA307504102GA SA307504104GA SA307504106GA SA307504107GA SA307504108GA SA307504109GA SA307504110GA SA307504111GA SA307504112GA SA307504114GA SA307504117GA SA307504118GA SA307504119GA SA307504120GA SA307504122GA SA307504123GA SA307504125GA SA307504126GA SA307504127GA SA307504128GA SA307504129GA SA307504130GA SA307504132GA SA307504133GA SA307504134GA SA307504135GA SA307504136GA SA307504137GA SA307504138GA SA307504139GA SA307504140GA SA307504141GA SA307504142GA SA307504143GA SA307504144GA SA307504145GA SA307504147GA SA307504148GA SA307504149GA SA307504150GA SA307504151GA SA307504152GA SA307504153GA SA307504154GA SA307504155GA SA307504156GA SA307504157GA SA307504159GA SA307504160GA SA307504161GA SA307504162GA SA307504163GA SA307504164GA SA307504165GA SA307504166GA SA307504168GA SA307504170GA SA307504171GA SA307504172GA SA307504173GA SA307504174GA SA307504175GA SA307504176GR SA307504177GA SA307504178GA SA307504179GA SA307504180GA SA307504181GA SA307504182GA SA307504183GA SA307504184GA SA307504185GA SA307504186GA SA307504190GA SA307504193GA SA307504196GA SA307504197GA SA307504198GA SA307504199GA SA307504200GA SA307504201GA SA307504203GA SA307504206GA SA307504207GA SA307504208GA SA307504209GA SA307504213GA SA307504215GA SA307504216GA SA307504217GA SA307504218GA SA307504219GA SA307504220GA SA307504221GA SA307504223GA SA307504233GR SA307514308GA SA307514321GA SA307514325GA SA307514328GA SA307514330GA SA307514331GA SA307514332GA SA307514333GR SA307514336GA SA307514337GA SA307514340GA SA307514342GA SA307514352GA SA307514355GR SA307514356GR SA307514357GR SA307514358GA SA307514359GR SA307514360GA SA307514363GR SA307514364GR SA307514365GR SA307514366GR SA307514367GR SA307514368GR SA307514370GR SA307514371GA SA307514372GR SA307514373GR SA307514376GA SA307514377GA SA307514378GA SA307514379GA SA307514380GR SA307514381GA SA307514382GA SA307514383GR SA307514384GR SA307514385GR SA307514386GA SA307514387GR SA307514388GR SA307514389GR SA307514391GR SA307514410GA SA307514414GA SA307514415GA SA307514418GA SA307514423GA SA307514424GA SA307514434GR SA308044513GR SA308044517GR SA308044520GR SA308044521GR SA308044522GR SA308044523GR SA308074538GA SA308074539GA SA308074540GR SA308074541GA SA308074542GR SA308074543GA SA308074547GA SA308074548GA SA308074549GA SA308074550GA SA308074551GR SA308074552GA SA308074553GR SA308074554GA SA308074555GA SA308074557GA SA308074559GR SA308074561GR SA308074564GA SA308074567GR SA308074569GR SA308074570GR SA308074571GR SA308074575GR SA308074576GR SA308074577GA SA308074578GA SA308074579GR SA308074580GA SA308074583GA SA308074585GA SA308074587GA SA308074588GA SA308074589GA SA308074593GR SA308074594GR SA308074596GR SA308074597GA SA308074598GA SA308074603GA SA308074604GR SA308074605GA SA308074606GA SA308074611GR SA308084612GA SA308084615GA SA308084616GA SA308084619GA SA308084620GA SA308084621GA SA308084622GA SA308084623GA SA308084643GA SA308084646GA SA308084652GA SA308084653GA SA308084658GA SA308084662GA SA308084664GA SA308084667GA SA308084670GA SA308084671GR SA308084673GR SA308084674GA SA308084675GA SA308084677GA SA308084679GA SA308084680GA SA308084681GA SA308084682GA SA308084684GA SA308084685GA SA308084686GA SA308084688GR SA308084690GA SA308084691GA SA308084692GA SA308084693GA SA308084694GR SA308084695GR SA308084699GA SA308084701GA SA308084702GA SA308084703GA SA308084704GA SA308084706GA SA308084707GA SA308084708GR SA308084710GA SA308084712GA SA308084713GA SA308084715GA SA308084717GA SA308084718GR SA308084719GR SA308084720GR SA308084721GA SA308084722GA SA308084724GR SA308084729GA SA308084735GA SA308084736GA SA308084737GA SA308084738GA SA308084740GA SA308084742GA SA308084743GA SA308084749GA SA308084750GA SA308084751GA SA308084753GA SA308084754GA SA308084756GA SA308084758GA SA308084759GA SA308084761GA SA308084762GA SA308084766GA SA308084769GA SA308084770GA SA308084774GA SA308084776GR SA308084777GA SA308084778GA SA308084779GA SA308084783GA SA308084784GR SA308084786GA SA308084787GA SA308084788GA SA308084789GA SA308084790GA SA308084791GA SA308084792GA SA308084793GA SA308084794GA SA308084795GA SA308094796GA SA308094797GA SA308094800GR SA308094801GA SA308094803GA SA308094807GA SA308094808GR SA308094809GA SA308094811GA SA308094814GA SA308094815GA SA308094816GA SA308094818GA SA308094819GA SA308094820GA SA308094832GA SA308094836GA SA308094837GA SA308094839GA SA308094841GA SA308094842GA SA308094843GA SA308094844GA SA308094845GA SA308094847GA SA308094850GR SA308094851GA SA308094852GA SA308094853GA SA308094855GA SA308094857GA SA308094858GA SA308094859GA SA308094860GA SA308094861GA SA308094862GA SA308094863GA SA308094864GA SA308094865GA SA308094867GA SA308094869GA SA308094871GA SA308094872GA SA308094875GA SA308094877GA SA308094878GA SA308094880GA SA308094884GA SA308094891GA SA308094894GA SA308094903GA SA308094915GA SA308094921GA SA308094925GA SA308094927GA SA308094929GA SA308094934GA SA308094938GA SA308094940GA SA308094941GR SA308094942GA SA308094943GA SA308094944GA SA308094949GA SA308094954GA SA308094955GA SA308094958GA SA308094959GA SA308094960GA SA308094965GA SA308094968GA SA308094978GA SA308094979GA SA308094984GA SA308094989GA SA308094991GA SA308094999GA SA308095000GA SA308095002GA SA308095013GA SA308095014GA SA308095015GA SA308095016GA SA308095017GA SA308095018GA SA308095019GA SA308095020GA SA308095021GA SA308095022GA SA308095023GR SA308095025GA SA308095026GA SA308095027GR SA308095028GA SA308095029GA SA308095030GA SA308095031GR SA308095032GA SA308095033GA SA308095034GR SA308095035GA SA308095036GR SA308095038GA SA308095040GA SA308095041GA SA308095042GA SA308095043GA SA308095044GA SA308095054GA SA308095055GA SA308095058GA SA308095060GR SA308095063GA SA308095069GA SA308095070GA SA308095071GR SA308095072GR SA308095073GR SA308095074GA SA308105075GA SA308105076GA SA308105077GA SA308105078GA SA308105079GA SA308105080GA SA308105082GA SA308105086GA SA308105088GA SA308105091GR SA308105092GA SA308105093GA SA308105094GA SA308105095GR SA308105097GR SA308105098GA SA308105100GA SA308105101GR SA308105102GA SA308105103GA SA308105107GR SA308105108GA SA308105109GA SA308105110GA SA308105111GA SA308105115GR SA308105116GA SA308105119GA SA308105120GA SA308105121GA SA308105122GA SA308105123GA SA308105124GA SA308105125GA SA308105126GA SA308105127GA SA308105128GA SA308105130GA SA308105131GA SA308105132GA SA308105133GA SA308105134GA SA308105135GA SA308105136GA SA308105137GA SA308105138GA SA308105139GR SA308105140GA SA308105141GA SA308105145GR SA308105165GR SA308105173GA SA308105176GR SA308105178GR SA308105185GA SA308105191GA SA308105193GA SA308105196GA SA308105198GA SA308105203GA SA308115247GR SA308115265GA SA308115276GA SA308115287GA SA308115288GA SA308115289GA SA308115290GA SA308115291GA SA308115292GA SA308115294GA SA308115295GA SA308115296GA SA308115297GA SA308115298GA SA308115300GA SA308115301GA SA308115303GA SA308115305GA SA308115306GA SA308125325GA SA308125326GA SA308125330GR SA308125335GA SA308125341GA SA308125344GA SA308125348GA SA308125349GA SA308125350GA SA308125354GA SA308125355GA SA308125356GA SA308125360GA SA308125364GA SA308125369GA SA308125374GA SA308125375GA SA308125376GA SA308125377GA SA308125379GA SA308125380GA SA308125382GA SA308125386GA SA308125390GA SA308125393GA SA308125394GA SA308125395GA SA308125396GA SA308125397GA SA308125399GA SA308125400GA SA308125401GA SA308125402GA SA308125403GA SA308125404GA SA308125405GA SA308125406GA SA308125408GA SA308125409GR SA308125410GA SA308125413GA SA308125414GA SA308125416GA SA308125418GA SA308125419GA SA308125420GA SA308125421GA SA308125423GA SA308125424GA SA308125425GA SA308125426GA SA308125427GA SA308125428GA SA308125429GA SA308135430GR SA308135431GA SA308135432GA SA308135433GA SA308135434GA SA308135435GA SA308135436GA SA308135437GA SA308135438GA SA308135439GA SA308135440GA SA308135441GA SA308135442GA SA308135444GA SA308135445GA SA308135446GA SA308135447GA SA308135448GA SA308135449GA SA308135450GA SA308135451GA SA308135452GA SA308135454GA SA308135455GA SA308135456GA SA308135457GA SA308135458GA SA308135459GA SA308135460GA SA308135462GA SA308135463GA SA308135465GA SA308135466GR SA308135467GA SA308135468GA SA308135471GA SA308135472GA SA308135474GA SA308135475GA SA308135476GA SA308135477GA SA308135478GA SA308135479GA SA308135480GA SA308135481GA SA308135482GA SA308135483GA SA308135484GA SA308135485GA SA308135486GR SA308135488GA SA308135489GA SA308135490GA SA308135491GA SA308135492GA SA308135493GA SA308135494GA SA308135495GA SA308135496GA SA308135497GA SA308135498GA SA308135499GA SA308135500GA SA308135501GA SA308135502GA SA308135503GA SA308135504GR SA308135505GA SA308135506GA SA308135507GA SA308135508GA SA308135509GA SA308135510GA SA308135511GA SA308135512GA SA308135513GA SA308135514GA SA308135518GR SA308135519GA SA308135521GA SA308135527GR SA308135533GR SA308135541GA SA308135545GA SA308135546GA SA308135547GA SA308135548GA SA308135549GA SA308135550GA SA308135551GA SA308135552GA SA308135553GA SA308135555GA SA308135556GA SA308135557GA SA308135559GA SA308135561GA SA308135563GA SA308135565GA SA308135566GA SA308135567GA SA308135568GA SA308135569GA SA308135571GA SA308135572GA SA308135573GA SA308135576GA SA308135577GA SA308135579GA SA308135581GA SA308135582GA
SA308135584GA SA308135586GA SA308135587GA SA308135588GA SA308135590GA SA308135591GA SA308135592GA SA308135593GA SA308135594GA SA308135595GA SA308135597GA SA308135598GA SA308135599GA SA308135601GA SA308135602GA SA308135603GA SA308135605GA SA308135606GA SA308135608GA SA308135609GA SA308135611GA SA308135613GA SA308135614GA SA308135615GR SA308135616GA SA308135617GA SA308135618GR SA308135619GA SA308135621GA SA308135622GA SA308135625GA SA308135627GA SA308135629GA SA308135630GA SA308135633GA SA308135639GA SA308135648GA SA308145655GR SA308145657GA SA308145661GA SA308145668GA SA308145675GA SA308145681GA SA308145684GA SA308145685GA SA308145687GR SA308145688GA SA308145690GA SA308145692GA SA308145693GA SA308145694GA SA308145695GA SA308145696GA SA308145697GA SA308145698GA SA308145700GA SA308145703GA SA308145704GA SA308145705GA SA308145707GA SA308145708GA SA308145714GA SA308145717GA SA308145722GA SA308145723GA SA308145724GA SA308145725GA SA308145726GA SA308145727GA SA308145729GA SA308145730GA SA308145731GA SA308145732GA SA308145734GA SA308145738GA SA308145739GA SA308145740GA SA308145741GA SA308145742GA SA308145743GA SA308145744GA SA308145746GA SA308145747GA SA308145748GA SA308145749GA SA308145750GA SA308145753GA SA308145754GR SA308145756GA SA308145758GA SA308145759GA SA308145760GA SA308145761GA SA308145762GA SA308145763GA SA308145764GA SA308145765GA SA308145766GA SA308145767GA SA308145768GA SA308145770GA SA308145771GA SA308145772GA SA308145773GA SA308145774GA SA308145776GR SA308145780GA SA308145781GA SA308145786GR SA308145787GA SA308145797GA SA308145803GA SA308145806GA SA308145807GA SA308145810GA SA308145812GA SA308145813GA SA308145814GA SA308145815GA SA308145816GA SA308145817GA SA308145818GA SA308145820GA SA308145822GA SA308145823GA SA308145824GA SA308145827GA SA308145828GA SA308145831GA SA308145832GA SA308145833GA SA308145834GA SA308145835GA SA308155840GA SA308155845GA SA308155846GR SA308155847GA SA308155850GA SA308155851GA SA308155854GR SA308155857GA SA308155858GA SA308155860GA SA308155863GA SA308155864GA SA308155865GA SA308155866GR SA308155867GA SA308155868GR SA308155869GA SA308155872GA SA308155873GA SA308155874GR SA308155875GA SA308155876GA SA308155878GA SA308155882GA SA308155885GA SA308155886GA SA308155887GA SA308155888GA SA308155892GA SA308155897GA SA308155898GA SA308155899GA SA308155900GR SA308155903GA SA308155904GA SA308155906GA SA308155908GA SA308155909GA SA308155914GA SA308155915GA SA308155916GA SA308155917GA SA308155918GA SA308155919GA SA308155920GA SA308155921GA SA308155923GA SA308155925GA SA308155928GA SA308155929GA SA308155931GA SA308155932GA SA308155933GA SA308155936GA SA308155937GA SA308155938GA SA308155939GA SA308155942GA SA308155944GA SA308155945GA SA308155946GA SA308155947GA SA308155949GA SA308155950GA SA308155953GA SA308155954GA SA308155955GA SA308155956GA SA308155957GA SA308155959GA SA308155960GA SA308155961GR SA308155964GA SA308155965GA SA308155966GA SA308155967GA SA308155968GA SA308155969GR SA308155970GA SA308155973GA SA308155975GR SA308155978GR SA308155980GA SA308155981GA SA308155983GA SA308155984GA SA308155985GA SA308155987GA SA308155989GA SA308155990GA SA308165991GA SA308165993GA SA308165994GA SA308165996GA SA308165997GA SA308165999GA SA308166000GA SA308166001GA SA308166002GA SA308166005GA SA308166006GA SA308166007GR SA308166008GA SA308166009GA SA308166010GR SA308166011GA SA308166012GA SA308166013GR SA308166014GA SA308166015GA SA308166017GA SA308166018GA SA308166019GA SA308166020GA SA308166022GR SA308166023GA SA308166024GR SA308166025GA SA308166026GR SA308166028GA SA308166029GA SA308166030GA SA308166031GA SA308166033GA SA308166034GA SA308166037GA SA308166039GA SA308166040GR SA308166041GA SA308166042GR SA308166043GA SA308166050GA SA308166052GR SA308166053GA SA308166054GA SA308166057GA SA308166058GA SA308166059GA SA308166060GA SA308166061GA SA308166062GA SA308166063GA SA308166066GR SA308166067GR SA308166068GA SA308166069GA SA308166072GA SA308166073GA SA308166074GR SA308166075GA SA308166077GA SA308166079GA SA308166080GA SA308166081GR SA308166082GR SA308166083GR SA308166085GA SA308166086GA SA308166088GA SA308166090GR SA308166091GA SA308166092GA SA308166095GA SA308166096GA SA308166097GA SA308166098GA SA308166100GA SA308166101GA SA308166102GR SA308166103GA SA308166107GR SA308166108GA SA308166110GA SA308166111GR SA308166112GR SA308166113GA SA308166114GA SA308166115GR SA308166118GA SA308166119GR SA308166120GA SA308166121GA SA308166122GA SA308166123GA SA308166124GA SA308166126GA SA308166127GA SA308166128GA SA308166129GA SA308166130GR SA308166131GA SA308166132GA SA308166133GA SA308166134GA SA308166135GA SA308166137GA SA308166138GA SA308166139GA SA308166141GA SA308166142GA SA308166143GA SA308166144GA SA308166145GA SA308166146GA SA308166147GA SA308166148GA SA308166149GA SA308166151GA SA308166152GA SA308166153GA SA308166154GA SA308166156GA SA308166157GA SA308166159GA SA308166160GA SA308166161GA SA308166162GR SA308166163GA SA308166164GA SA308166165GA SA308166166GA SA308166167GA SA308166168GR SA308166169GA SA308166170GA SA308166171GA SA308166172GA SA308166173GA SA308166174GA SA308166175GA SA308166177GA SA308166178GA SA308166179GA SA308166180GA SA308166181GA SA308166182GA SA308166183GA SA308166184GA SA308166185GA SA308166186GR SA308166189GA SA308166190GA SA308166193GA SA308166194GA SA308166195GA SA308166196GA SA308166197GA SA308166198GA SA308166199GA SA308166200GA SA308166201GR SA308166203GA SA308166204GA SA308166205GA SA308166207GA SA308166208GA SA308166209GA SA308166210GA SA308166211GA SA308166212GA SA308166213GA SA308166214GA SA308166215GA SA308166216GA SA308176218GA SA308176219GA SA308176220GA SA308176221GA SA308176223GA SA308176224GA SA308176225GA SA308176227GA SA308176229GA SA308176230GA SA308176232GA SA308176234GA SA308176235GA SA308176237GA SA308176238GA SA308176239GA SA308176242GA SA308176243GA SA308176244GA SA308176245GR SA308176246GA SA308176247GA SA308176248GA SA308176250GA SA308176251GA SA308176253GA SA308176254GA SA308176255GA SA308176258GA SA308176259GA SA308176260GA SA308176261GA SA308176262GA SA308176263GA SA308176264GA SA308176265GA SA308176266GA SA308176267GA SA308176268GA SA308176269GA SA308176270GA SA308176271GA SA308176272GA SA308176273GA SA308176274GA SA308176275GA SA308176277GA SA308176279GA SA308176280GA SA308176281GA SA308176282GA SA308176283GA SA308176284GA SA308176285GA SA308176286GA SA308176287GA SA308176288GA SA308176289GA SA308176290GA SA308176291GA SA308176295GA SA308176296GR SA308176297GA SA308176298GA SA308176299GA SA308176300GA SA308176301GA SA308176302GA SA308176303GA SA308176304GA SA308176305GA SA308176306GA SA308176308GA SA308186310GA SA308186312GA SA308186313GA SA308186314GA SA308186316GA SA308186317GA SA308186318GA SA308186319GA SA308186320GA SA308186323GA SA308186324GA SA308186326GA SA308186327GA SA308186331GA SA308186333GA SA308186334GA SA308186335GR SA308186337GA SA308186350GA SA308186351GA SA308186352GA SA308186353GA SA308186354GA SA308186357GA SA308186359GA SA308186360GA SA308186362GA SA308186364GA SA308186365GA SA308186366GA SA308186368GA SA308186369GA SA308186370GA SA308186371GA SA308186372GR SA308186373GA SA308186374GA SA308186375GA SA308186376GA SA308186377GA SA308186378GA SA308186380GA SA308186381GA SA308186382GR SA308186383GA SA308186385GA SA308186388GA SA308186389GA SA308186390GA SA308186391GA SA308186392GA SA308186394GA SA308186395GA SA308186396GA SA308186397GA SA308186398GA SA308186401GA SA308186402GA SA308186403GA SA308186404GA SA308186405GA SA308186406GA SA308186407GA SA308186410GA SA308186416GA SA308186417GA SA308186422GR SA308186423GA SA308186424GA SA308186426GA SA308186427GA SA308186429GR SA308186431GA SA308186432GA SA308186433GR SA308186437GA SA308186445GA SA308196446GA SA308196447GA SA308196448GA SA308196449GA SA308196450GA SA308196451GA SA308196452GA SA308196453GA SA308196454GR SA308196455GR SA308196456GA SA308196457GR SA308196459GA SA308196464GR SA308196466GA SA308196467GA SA308196468GR SA308196471GR SA308196472GA SA308196473GA SA308196474GR SA308196475GA SA308196477GA SA308196478GA SA308196480GA SA308196485GA SA308196486GA SA308196487GA SA308196488GA SA308196491GA SA308196492GA SA308196493GA SA308196494GA SA308196495GA SA308196497GA SA308196498GA SA308196499GA SA308196500GA SA308196501GA SA308196502GA SA308196503GA SA308196505GA SA308196506GA SA308196510GA SA308196511GA SA308196512GA SA308196513GA SA308196514GA SA308196515GA SA308196516GA SA308196517GA SA308196518GA SA308196520GA SA308196521GA SA308196522GA SA308196523GA SA308196525GA SA308196526GA SA308196527GA SA308196528GA SA308196529GA SA308196530GA SA308196531GA SA308196532GA SA308196533GA SA308196535GA SA308196536GA SA308196537GA SA308196538GA SA308196540GA SA308196542GA SA308196545GA SA308196546GA SA308196547GA SA308196549GA SA308196550GA SA308196551GA SA308196553GA SA308196555GA SA308196557GA SA308196559GA SA308196560GR SA308196561GA SA308196562GA SA308196563GA SA308196564GA SA308196568GR SA308196569GA SA308196574GA SA308196575GA SA308196579GA SA308196581GA SA308196582GA SA308196584GR SA308196585GR SA308206587GA SA308206591GA SA308206592GA SA308206593GA SA308206594GR SA308206595GA SA308206596GA SA308206597GA SA308206598GA SA308206604GA SA308206605GA SA308206612GA SA308206613GA SA308206614GA SA308206615GA SA308206618GA SA308206625GA SA308206629GA SA308206634GA SA308206647GA SA308206650GA SA308206651GA SA308206652GA SA308206653GA SA308206654GA SA308206655GA SA308206656GA SA308206657GA SA308206658GA SA308206659GA SA308206660GA SA308206662GA SA308206666GA SA308206667GA SA308206668GA SA308206669GA SA308206671GA SA308206672GA SA308206673GA SA308206674GA SA308206675GA SA308206676GA SA308206677GA SA308206678GA SA308206680GA SA308206681GA SA308206682GA SA308206683GA SA308206684GA SA308206685GA SA308206686GA SA308206687GA SA308206688GA SA308206689GA SA308206690GA SA308206691GA SA308206692GA SA308206693GA SA308206694GA SA308206695GA SA308206696GA SA308206698GA SA308206700GA SA308206701GA SA308206702GA SA308206703GA SA308206704GA SA308206705GA SA308206706GA SA308206707GA SA308206708GA SA308206709GR SA308206710GA SA308206711GA SA308206712GA SA308206713GA SA308206714GA SA308206715GA SA308206716GA SA308206717GA SA308206718GA SA308206719GA SA308206720GA SA308206722GA SA308206724GA SA308206725GA SA308206726GA SA308206727GA SA308206728GA SA308206729GA SA308206730GA SA308206731GA SA308206732GA SA308206733GA SA308206735GA SA308206736GA SA308206739GA SA308206740GA SA308206741GA SA308206744GA SA308206745GA SA308206746GA SA308206747GA SA308206748GA SA308206749GA SA308206750GA SA308206752GA SA308206754GA SA308206755GA SA308206756GA SA308206758GA SA308206759GA SA308206760GA SA308206761GA SA308216762GA SA308216763GA SA308216764GA SA308216765GA SA308216766GA SA308216768GA SA308216769GA SA308216770GA SA308216772GA SA308216773GA SA308216774GA SA308216776GA SA308216777GA SA308216779GA SA308216780GA SA308266784GA SA308266789GA SA308266811GA SA308266814GA SA308266815GA SA308266818GA SA308266819GA SA308266820GR SA308266821GR SA308266822GA SA308266824GA SA308266825GA SA308266826GA SA308266829GA SA308266830GA SA308266831GA SA308266833GA SA308266834GA SA308266835GA SA308266836GR SA308266838GA SA308266839GA SA308266840GA SA308266841GA SA308266842GA SA308266844GA SA308266845GA SA308266846GA SA308266847GA
SA308266849GR SA308266850GA SA308266851GA SA308266853GA SA308266854GA SA308266856GA SA308266857GA SA308266858GA SA308266859GA SA308266862GA SA308266863GA SA308266864GA SA308266865GR SA308266866GA SA308266867GA SA308266868GR SA308266869GA SA308266870GA SA308266872GA SA308266874GA SA308266878GR SA308266879GA SA308266880GA SA308266881GA SA308266882GA SA308266885GA SA308266886GA SA308266888GR SA308266890GR SA308266892GA SA308266894GA SA308266895GA SA308266896GA SA308266898GA SA308266899GR SA308266900GR SA308266902GA SA308286903GA SA308286906GA SA308286909GA SA308286911GR SA308286913GA SA308286915GA SA308286918GA SA308286919GA SA308286920GA SA308286922GA SA308286923GA SA308286925GA SA308286928GA SA308286929GR SA308286930GA SA308286932GA SA308286933GA SA308286937GA SA308286939GA SA308286945GA SA308286949GA SA308286950GA SA308286951GA SA308286953GA SA308286955GA SA308286960GA SA308286961GA SA308286962GA SA308286966GA SA308286967GA SA308286968GA SA308286969GA SA308286970GA SA308286971GA SA308286972GA SA308286973GA SA308286975GA SA308286977GA SA308286978GA SA308286979GA SA308286981GA SA308286982GA SA308286984GA SA308286985GA SA308286986GA SA308286987GA SA308286988GA SA308286990GA SA308286993GA SA308286994GA SA308286995GA SA308286996GA SA308287000GA SA308287006GA SA308287007GA SA308287008GA SA308287010GA SA308287011GA SA308287013GA SA308287016GA SA308287017GA SA308287018GA SA308287019GA SA308287020GA SA308287021GA SA308287023GA SA308287024GA SA308297041GA SA308297059GA SA308297061GA SA308297062GA SA308297063GA SA308317065GA SA308317066GR SA308317067GA SA308317068GA SA308317069GA SA308317070GA SA308317071GA SA308317072GA SA308317073GA SA308317074GR SA308317075GA SA308317076GA SA308317078GA SA308317079GA SA308317081GA SA308317082GA SA308317083GA SA308317084GA SA308317087GA SA308317088GA SA308317089GA SA308317092GA SA308317093GA SA308317095GA SA308317096GA SA308317098GR SA308317100GA SA308317101GA SA308317104GA SA308317107GA SA308317108GA SA308317109GA SA308317111GA SA308317113GA SA308317114GA SA308317115GR SA308317118GA SA308317121GA SA308317122GA SA308317124GA SA308317125GA SA308317126GA SA308317127GA SA308317128GA SA308317129GA SA308317130GA SA308317131GA SA308317132GA SA308317133GA SA308317134GA SA308317135GA SA308317136GA SA308317137GA SA308317138GA SA308317140GA SA308317141GA SA308317142GA SA308317143GA SA308317144GA SA308317145GA SA308317146GA SA308317148GA SA308317150GA SA308317151GA SA308317152GA SA308317153GA SA308317154GA SA308317156GA SA308317157GA SA308317158GA SA308317159GA SA308317161GA SA308317162GA SA308317163GA SA308317165GA SA308317168GA SA308317170GA SA308317171GA SA308317172GA SA308317174GA SA308317177GA SA308317178GA SA308317179GA SA308317180GA SA308317181GA SA308317182GA SA308317184GA SA308317185GA SA308317187GA SA308317188GA SA308317189GA SA308317190GA SA308317191GA SA308317193GR SA308317194GA SA308317195GA SA308317197GA SA308317199GA SA308317200GA SA308317201GA SA308317202GA SA308317203GA SA308317207GA SA308317208GR SA308317210GA SA308317211GA SA308317213GA SA308317224GA SA308317225GA SA308317227GA SA308317230GA SA308317231GA SA308317232GA SA308317233GA SA308317234GR SA308317236GA SA308317237GA SA308317238GA SA308317239GA SA308317240GA SA308317241GA SA308317242GA SA308317243GA SA308317244GA SA308317245GA SA308317246GA SA308317247GA SA308317250GA SA308317251GA SA308317252GA SA308317256GR SA308317257GA SA308317260GA SA308317261GA SA308317263GA SA308317265GA SA308317266GR SA308317267GA SA308317268GA SA308317269GA SA308317270GA SA308317272GA SA308317273GA SA308317274GA SA308317275GA SA308317276GA SA308317278GA SA308317279GA SA308317281GA SA308317282GA SA308317287GA SA308317290GA SA308317291GA SA308317292GA SA308317293GA SA308317294GA SA308317297GA SA308317298GA SA308317299GA SA308317300GA SA308317301GA SA308317303GA SA308317304GR SA308317305GA SA308317308GA SA308317309GA SA308317310GA SA308317311GA SA308327312GA SA308327313GA SA308327316GA SA308327317GA SA308327319GA SA308327320GA SA308327323GA SA308327324GA SA308327325GA SA308327326GA SA308327328GA SA308327329GA SA308327330GA SA308327331GA SA308327333GA SA308327334GR SA308327335GA SA308327336GA SA308327337GA SA308327339GA SA308327341GA SA308327342GA SA308327343GA SA308327345GA SA308327346GA SA308327347GA SA308327348GA SA308327349GA SA308327350GA SA308327351GA SA308327355GA SA308327360GA SA308327362GA SA308327363GA SA308327364GA SA308327365GA SA308327366GA SA308327368GA SA308327369GA SA308327371GA SA308327373GA SA308327374GA SA308327375GA SA308327377GA SA308327378GA SA308327379GA SA308327380GA SA308327381GA SA308327382GA SA308327383GA SA308327384GA SA308327386GA SA308327402GA SA308327403GA SA308327420GA SA308327421GA SA308327424GA SA308327425GA SA308327426GA SA308327428GA SA308327429GA SA308327431GA SA308327432GA SA308327433GA SA308327434GA SA308327435GA SA308327436GA SA308327437GA SA308327438GA SA308327439GA SA308327440GA SA308327441GA SA308327444GA SA308327445GA SA308327446GA SA308327447GA SA308327448GR SA308327449GA SA308327450GR SA308327451GR SA308327452GA SA308327453GA SA308327455GA SA308327456GA SA308327457GA SA308327458GA SA308327459GA SA308327460GA SA308327461GA SA308327464GA SA308327465GA SA308327468GA SA308327469GA SA308327470GA SA308327471GA SA308327472GA SA308327473GA SA308327474GA SA308327475GA SA308327477GA SA308327478GA SA308327479GA SA308327480GA SA308327481GA SA308327482GA SA308327483GA SA308327485GA SA308327486GA SA308327487GA SA308327488GA SA308327491GR SA308327492GA SA308327493GA SA308327494GR SA308327495GA SA308327497GA SA308327499GA SA308327500GA SA308327501GA SA308327503GA SA308327505GA SA308327506GA SA308327507GA SA308327508GA SA308327509GR SA308327511GR SA308327512GA SA308327513GA SA308327514GA SA308327515GA SA308327516GA SA308327517GA SA308327518GA SA308327519GA SA308327520GA SA308327521GA SA308327523GA SA308327524GA SA308327525GA SA308327527GA SA308327528GR SA308327529GA SA308327530GA SA308327531GA SA308327532GA SA308327533GA SA308327536GA SA308327537GA SA308327542GA SA308327543GA SA308327544GA SA308327547GR SA308327548GA SA308327549GR SA308327550GA SA308327551GA SA308327554GA SA308327555GA SA308327556GA SA308327557GA SA308327558GA SA308337563GA SA308337564GA SA308337565GA SA308337566GA SA308337567GA SA308337568GA SA308337569GA SA308337571GA SA308337572GA SA308337573GA SA308337574GA SA308337575GA SA308337576GA SA308337577GA SA308337578GA SA308337579GA SA308337581GA SA308337582GA SA308337583GA SA308337584GA SA308337585GA SA308337586GA SA308337587GA SA308337588GR SA308337589GA SA308337590GA SA308337591GA SA308337592GA SA308337594GR SA308337595GA SA308337596GA SA308337597GA SA308337598GR SA308337599GR SA308337601GA SA308337602GR SA308337603GA SA308337604GA SA308337605GA SA308337606GA SA308337607GA SA308337608GA SA308337609GA SA308337610GA SA308337611GA SA308337612GA SA308337613GA SA308337614GA SA308337615GA SA308337616GA SA308337617GA SA308337618GA SA308337619GA SA308337620GA SA308337621GA SA308337622GA SA308337623GA SA308337624GR SA308337625GA SA308337626GA SA308337628GA SA308337629GA SA308337630GA SA308337632GR SA308337633GA SA308337635GA SA308337637GA SA308337638GA SA308337639GA SA308337640GA SA308337641GA SA308337642GA SA308337643GA SA308337644GA SA308337645GA SA308337649GA SA308337650GA SA308337651GA SA308337653GA SA308337654GA SA308337655GA SA308337656GR SA308337658GA SA308337659GA SA308337660GA SA308337661GA SA308337662GA SA308337664GA SA308337665GA SA308337668GA SA308337669GR SA308337670GA SA308337671GR SA308337672GA SA308337678GA SA308337679GA SA308337680GA SA308337681GA SA308337682GA SA308337683GA SA308337684GA SA308337685GA SA308337687GR SA308337689GA SA308337690GA SA308337692GA SA308337693GA SA308337694GA SA308337695GA SA308337696GA SA308337698GA SA308337699GA SA308337700GA SA308337701GA SA308337702GA SA308337705GA SA308337708GA SA308347723GA SA308347724GA SA308347725GR SA308347726GA SA308347727GR SA308347728GA SA308347729GA SA308347730GA SA308347731GA SA308347732GA SA308347737GA SA308347741GA SA308347742GA SA308347743GA SA308347744GA SA308347746GA SA308347747GA SA308347748GA SA308347749GR SA308347750GA SA308347752GA SA308347757GA SA308347762GA SA308347764GA SA308347765GA SA308347766GA SA308347768GA SA308347769GA SA308347770GA SA308347771GA SA308347773GA SA308347774GA SA308347775GA SA308347776GA SA308347777GA SA308347778GA SA308347779GA SA308347780GA SA308347781GA SA308347782GA SA308347783GA SA308347784GA SA308347785GA SA308347786GA SA308347787GA SA308347788GA SA308347790GA SA308347793GA SA308347794GA SA308347795GA SA308347796GA SA308347797GA SA308347799GA SA308347802GA SA308347803GA SA308347804GA SA308347806GA SA308347807GA SA308347808GA SA308347810GA SA308347813GA SA308347814GA SA308347815GA SA308347817GA SA308347818GA SA308347819GA SA308347820GA SA308347821GA SA308347822GA SA308347823GA SA308347825GA SA308347826GA SA308347827GR SA308347828GA SA308347831GA SA308347832GA SA308347833GA SA308347834GA SA308347835GA SA308347837GA SA308347838GA SA308347839GA SA308347840GA SA308347842GA SA308347843GA SA308347852GA SA308347853GA SA308347854GA SA308347855GA SA308347856GA SA308347857GA SA308347858GA SA308347859GA SA308347860GA SA308357862GA SA308357864GA SA308357865GA SA308357866GA SA308357868GA SA308357871GA SA308357874GA SA308357875GA SA308357876GA SA308357877GA SA308357879GA SA308357881GA SA308357882GA SA308357883GA SA308357885GA SA308357886GA SA308357887GA SA308357889GA SA308357891GA SA308357892GA SA308357894GA SA308357896GA SA308357897GA SA308357898GA SA308357899GA SA308357906GA SA308357909GA SA308357910GA SA308357911GA SA308357913GA SA308357914GA SA308357916GA SA308357917GA SA308357918GA SA308357919GA SA308357920GA SA308357921GA SA308357922GA SA308357923GA SA308357925GA SA308357926GA SA308357927GA SA308357928GR SA308357930GA SA308357933GA SA308357934GA SA308357935GA SA308357936GA SA308357937GA SA308357938GA SA308357939GA SA308357940GA SA308357942GA SA308357947GA SA308357948GA SA308357949GA SA308357950GA SA308357951GA SA308357952GA SA308357953GA SA308357954GA SA308357955GA SA308357956GA SA308357957GA SA308357958GA SA308357959GA SA308357960GA SA308357961GA SA308357962GA SA308357963GA SA308357964GA SA308357965GA SA308357966GA SA308357968GA SA308357969GA SA308357970GA SA308357971GA SA308357972GA SA308357973GA SA308357974GA SA308357975GA SA308357978GA SA308357979GA SA308357980GA SA308357981GA SA308357982GA SA308357983GA SA308357984GA SA308357985GA SA308357986GA SA308357987GA SA308357988GA SA308357989GA SA308357990GA SA308357991GR SA308357992GA SA308357993GA SA308357994GA SA308357995GA SA308357996GA SA308357997GA SA308357998GA SA308357999GA SA308358000GA SA308358001GA SA308358002GA SA308358003GA SA308358004GA SA308358005GA SA308358006GA SA308358007GA SA308358008GA SA308358009GA SA308358010GA SA308358011GA SA308358012GA SA308358013GA SA308358014GA SA308358015GA SA308358016GA SA308358017GA SA308358018GA SA308358019GA SA308358020GA SA308358021GA SA308358022GA SA308358023GA SA308358024GA SA308358025GA SA308358026GA SA308358027GA SA308358028GA SA308358029GA SA308358031GA SA308358032GA SA308358033GA SA308358034GA SA308358035GA SA308358036GA SA308358037GA SA308358038GA SA308358039GA SA308358040GA SA308358041GA SA308358042GA SA308358043GA SA308358044GA SA308358045GA SA308358046GA SA308368047GA SA308368048GA
SA308368049GA SA308368051GA SA308368052GA SA308368053GA SA308368054GA SA308368055GA SA308368056GA SA308368057GA SA308368058GA SA308368059GA SA308368061GA SA308368063GA SA308368064GA SA308368065GA SA308368066GA SA308368067GA SA308368068GA SA308368069GA SA308368071GA SA308368072GA SA308368073GA SA308368074GA SA308368075GA SA308368076GA SA308368078GA SA308368079GA SA308368080GA SA308368081GA SA308368082GA SA308368083GA SA308368084GA SA308368085GA SA308368086GA SA308368087GA SA308368088GA SA308368089GA SA308368090GA SA308368091GA SA308368092GA SA308368093GR SA308368094GA SA308368095GA SA308368096GA SA308368097GA SA308368098GA SA308368099GA SA308368100GA SA308368101GA SA308368102GA SA308368103GA SA308368104GA SA308368105GA SA308368106GA SA308368107GA SA308368108GA SA308368109GA SA308368110GA SA308368112GA SA308368114GA SA308368118GA SA308368119GA SA308368123GA SA308368125GA SA308368126GA SA308368128GA SA308368129GA SA308368130GA SA308368132GA SA308368133GA SA308368134GA SA308368135GA SA308368136GA SA308368137GA SA308368138GA SA308368139GA SA308368140GR SA308368142GA SA308368144GA SA308368145GA SA308368146GA SA308368147GA SA308368148GA SA308368149GA SA308368150GA SA308368151GA SA308368152GR SA308368154GA SA308368155GA SA308368156GA SA308368157GA SA308368158GA SA308368159GA SA308368160GA SA308368161GA SA308368162GA SA308368163GA SA308368164GA SA308368165GA SA308368166GA SA308368167GA SA308368168GA SA308368169GA SA308368170GA SA308368171GA SA308368172GA SA308368173GA SA308368174GA SA308368175GA SA308368177GA SA308368178GA SA308368179GA SA308368180GA SA308368181GA SA308368182GA SA308368183GR SA308368184GA SA308368185GA SA308368186GA SA308368187GA SA308368188GA SA308368189GA SA308368190GA SA308368191GA SA308368192GA SA308368193GA SA308368194GA SA308368196GA SA308368197GA SA308368198GA SA308368199GA SA308368200GA SA308368201GR SA308368203GA SA308368205GA SA308368206GA SA308368207GA SA308368208GA SA308368209GA SA308368210GA SA308368212GA SA308368213GA SA308368214GA SA308368215GA SA308368216GA SA308368217GA SA308368218GA SA308368219GA SA308368220GA SA308368221GA SA308368222GA SA308368223GA SA308368224GA SA308368225GA SA308368226GA SA308368227GA SA308368228GA SA308368229GA SA308368230GA SA308368233GA SA308368235GA SA308368236GA SA308368237GA SA308368238GA SA308368241GA SA308368244GA SA308368245GA SA308368248GA SA308378251GA SA308378252GA SA308378253GA SA308378254GA SA308378255GA SA308378256GA SA308378257GA SA308378258GA SA308378259GA SA308378260GA SA308378261GA SA308378262GA SA308378263GA SA308378264GA SA308378265GA SA308378266GA SA308378267GA SA308378268GA SA308378269GA SA308378270GA SA308378271GA SA308378272GA SA308378273GA SA308378275GA SA308378277GA SA308378278GA SA308378279GA SA308378282GA SA308378283GA SA308378284GA SA308378285GA SA308378287GA SA308378288GA SA308378289GA SA308378290GA SA308378291GA SA308378292GA SA308378293GA SA308378294GA SA308378295GA SA308378298GA SA308378299GA SA308378300GA SA308378301GA SA308378302GA SA308378303GA SA308378304GA SA308378305GA SA308378307GA SA308378308GA SA308378309GA SA308378310GA SA308378311GR SA308378312GA SA308378313GA SA308378314GA SA308378315GA SA308378316GA SA308378317GA SA308378318GA SA308378319GA SA308378320GA SA308378322GA SA308378323GA SA308378325GA SA308378326GR SA308378327GA SA308378328GA SA308378329GA SA308378330GA SA308378331GA SA308378332GA SA308378333GA SA308378334GA SA308378335GA SA308388365GA SA308388366GA SA308388407GA SA308388408GA SA308388409GA SA308388411GA SA308388412GA SA308388413GA SA308388414GA SA308388415GA SA308388416GA SA308388417GA SA308388418GA SA308388419GA SA308388420GA SA308388421GA SA308388423GA SA308388424GA SA308388425GA SA308388426GA SA308388427GA SA308388428GA SA308388429GA SA308388430GA SA308388431GA SA308388432GA SA308388433GA SA308388434GA SA308388435GA SA308388436GA SA308388437GA SA308388438GA SA308388439GA SA308388440GA SA308388441GA SA308388442GA SA308388443GR SA308388444GA SA308388445GA SA308388447GA SA308388448GA SA308388449GA SA308388450GA SA308388451GA SA308388453GA SA308388454GA SA308388455GA SA308388457GA SA308388458GA SA308388459GA SA308388460GA SA308388461GA SA308388462GA SA308388463GA SA308388464GA SA308388465GA SA308388466GA SA308388467GA SA308388468GR SA308388469GA SA308388470GA SA308388471GA SA308388472GA SA308388473GA SA308388474GA SA308388475GA SA308388476GR SA308388477GA SA308388478GA SA308388479GA SA308388480GA SA308388482GA SA308388483GA SA308388484GA SA308388485GA SA308388486GA SA308388487GA SA308388489GA SA308388490GA SA308388491GA SA308388492GA SA308388493GA SA308388494GA SA308388495GA SA308388496GA SA308388497GA SA308388498GA SA308388499GA SA308388500GA SA308388501GA SA308388502GA SA308388503GA SA308388504GA SA308388505GA SA308388507GA SA308388508GA SA308388509GA SA308388510GA SA308388511GA SA308388512GA SA308388513GA SA308388514GA SA308388515GA SA308388516GA SA308388517GA SA308388518GR SA308388519GA SA308388520GA SA308388521GA SA308388522GA SA308388523GA SA308388524GA SA308388525GA SA308388526GA SA308388527GA SA308388528GA SA308388529GA SA308388530GA SA308388531GA SA308388532GA SA308388534GA SA308388535GR SA308388536GA SA308388537GA SA308388538GA SA308388539GA SA308388541GA SA308388542GA SA308388543GA SA308388544GA SA308388545GA SA308388546GA SA308388547GA SA308388549GA SA308388551GA SA308388552GA SA308388553GA SA308388555GA SA308388556GA SA308388557GA SA308388559GA SA308388560GA SA308388561GA SA308388562GA SA308388565GA SA308388567GA SA308388568GA SA308388570GA SA308388577GA SA308388578GA SA308388579GA SA308388580GR SA308388582GA SA308388583GA SA308388584GA SA308388585GA SA308388586GA SA308388587GA SA308388588GA SA308388590GA SA308388595GA SA308388600GA SA308388602GA SA308388604GA SA308388605GA SA308388606GA SA308388607GA SA308388608GA SA308388609GA SA308388610GA SA308388613GA SA308388614GA SA308388615GA SA308388616GA SA308388618GR SA308388619GA SA308388620GA SA308388622GA SA308388625GA SA308388626GA SA308388628GA SA308388629GA SA308388635GA SA308398639GA SA308398640GA SA308398641GA SA308398642GA SA308398643GA SA308398644GA SA308398645GA SA308398646GA SA308398647GA SA308398648GA SA308398649GA SA308398650GA SA308398651GA SA308398652GA SA308398653GA SA308398654GA SA308398655GA SA308398657GA SA308398658GA SA308398660GA SA308398661GA SA308398665GA SA308398670GA SA308398676GA SA308398677GA SA308398681GA SA308398683GA SA308398687GA SA308398694GA SA308398695GA SA308398696GA SA308398697GA SA308398698GA SA308398699GA SA308398700GA SA308398701GA SA308398702GR SA308398703GA SA308398704GA SA308398705GA SA308398707GA SA308398708GA SA308398711GA SA308398714GA SA308398716GA SA308398717GA SA308398719GR SA308398720GA SA308398721GA SA308398722GA SA308398725GA SA308398728GA SA308398729GA SA308398730GA SA308398731GA SA308398732GA SA308398733GA SA308398738GA SA308398740GA SA308398741GA SA308398742GR SA308398743GA SA308398745GA SA308398749GA SA308398752GA SA308398753GA SA308398754GA SA308398755GA SA308398756GA SA308398761GA SA308398762GA SA308398763GA SA308398765GA SA308398769GA SA308398770GA SA308398771GA SA308398772GA SA308398773GA SA308398777GR SA308398779GA SA308398781GA SA308398783GR SA308398784GR SA308398785GA SA308398795GA SA308398802GA SA308398803GA SA308398810GA SA308398812GA SA308398813GA SA308398817GR SA308398821GA SA308398822GA SA308398824GA SA308398825GA SA308398826GA SA308398828GA SA308398830GA SA308398841GA SA308398847GA SA308398848GA SA308398853GA SA308398855GA SA308398857GA SA308398858GA SA308398860GA SA308398861GR SA308398865GR SA308398866GA SA308398868GA SA308398872GA SA308398873GA SA308398874GA SA308398875GA SA308398876GA SA308398877GA SA308398878GA SA308398879GA SA308398880GR SA308398881GA SA308398882GA SA308398883GA SA308398884GA SA308398885GR SA308398886GR SA308398887GA SA308398888GA SA308398889GA SA308398891GR SA308398892GA SA308398893GA SA308398894GA SA308398895GA SA308398896GR SA308398897GA SA308398898GA SA308398899GA SA308398900GA SA308398901GA SA308398903GA SA308408906GA SA308408907GA SA308408910GA SA308408911GA SA308408912GA SA308408915GR SA308408916GA SA308408917GA SA308408924GA SA308408925GA SA308408934GA SA308408938GA SA308408946GA SA308408948GA SA308408949GA SA308408957GA SA308408960GA SA308408965GA SA308408966GA SA308408967GA SA308408968GA SA308408969GA SA308408970GA SA308408971GA SA308408973GA SA308408974GA SA308408975GA SA308408976GA SA308408977GR SA308408979GA SA308408980GA SA308408981GA SA308408982GA SA308408983GA SA308408984GA SA308408985GA SA308408986GA SA308408987GA SA308408988GA SA308408990GR SA308408991GR SA308408993GR SA308408996GR SA308409001GA SA308409002GA SA308409010GA SA308409012GA SA308409013GA SA308409016GA SA308409017GA SA308409019GA SA308409020GA SA308409022GA SA308409025GA SA308409027GA SA308409030GA SA308409032GR SA308409033GA SA308409034GA SA308409036GR SA308409037GA SA308409038GA SA308409039GA SA308409041GA SA308409042GA SA308409043GA SA308409044GA SA308409045GA SA308409046GA SA308409047GA SA308409050GA SA308409056GA SA308409057GA SA308409058GA SA308409059GA SA308409060GA SA308409063GA SA308409064GA SA308409065GA SA308409069GA SA308409071GA SA308409073GA SA308409074GA SA308409075GA SA308409076GA SA308409078GR SA308409079GA SA308409081GR SA308409082GA SA308409083GA SA308409087GA SA308409091GA SA308409092GR SA308409093GA SA308409094GA SA308409096GA SA308409099GR SA308409107GA SA308409108GR SA308409109GA SA308409110GA SA308409111GA SA308409113GA SA308409114GR SA308409115GR SA308409116GR SA308409117GA SA308409118GR SA308409119GA SA308409121GR SA308409122GA SA308409123GA SA308409124GA SA308409125GA SA308409126GA SA308409127GA SA308409128GA SA308409130GA SA308409131GA SA308409132GA SA308409133GA SA308409134GA SA308409135GR SA308409136GA SA308409137GA SA308409139GR SA308409140GA SA308409141GR SA308409142GR SA308409143GA SA308409145GR SA308409146GR SA308409147GR SA308409148GR SA308409149GR SA308409150GR SA308409152GR SA308409154GA SA308409163GA SA308409166GA SA308409167GA SA308409168GA SA308409171GA SA308409172GR SA308409176GA SA308409179GA SA308409180GR SA308409181GA SA308409182GA SA308409185GA SA308409186GA SA308409190GA SA308409191GA SA308409192GA SA308409193GA SA308409195GA SA308409199GA SA308409201GA SA308409202GR SA308409203GR SA308409205GA SA308409206GA SA308409208GA SA308409209GA SA308409210GA SA308409211GA SA308419213GA SA308419214GA SA308419216GA SA308419217GA SA308419219GA SA308419220GR SA308419221GA SA308419222GA SA308419223GA SA308419224GA SA308419225GA SA308419226GA SA308419227GA SA308419228GA SA308419234GA SA308419235GA SA308419237GR SA308419241GA SA308419247GA SA308419248GA SA308419252GA SA308419253GA SA308419254GA SA308419256GA SA308419261GA SA308419265GA SA308419269GA SA308419270GA SA308419275GA SA308419276GA SA308419277GA SA308419278GA SA308419279GA SA308419280GR SA308419283GR SA308419284GR SA308419285GA SA308419286GA SA308419287GA SA308419288GR SA308419289GR SA308419290GA SA308419291GR SA308419292GR SA308419293GA SA308419294GR SA308419296GA SA308419297GR SA308419298GR SA308419299GR SA308419301GR SA308419302GA SA308419303GR SA308419304GA SA308419305GR SA308419306GR SA308419307GA SA308419308GA SA308429309GA SA308429310GA SA308429311GA SA308429313GA SA308429314GA SA308429315GA SA308429316GA SA308429317GA SA308429318GA SA308429319GA SA308429320GA
SA308429322GA SA308429323GA SA308429324GA SA308429326GA SA308429329GR SA308429330GA SA308429332GA SA308429333GA SA308429334GA SA308429335GA SA308429336GA SA308429337GA SA308429338GA SA308429339GA SA308429343GA SA308429344GA SA308429345GA SA308429346GA SA308429348GA SA308429352GR SA308429356GA SA308429358GA SA308429361GA SA308429362GR SA308429364GA SA308429365GA SA308429366GR SA308429367GR SA308429368GA SA308429369GA SA308429370GA SA308429371GA SA308429372GA SA308429373GA SA308429374GA SA308429375GA SA308429376GA SA308429377GA SA308429378GA SA308429379GA SA308429380GR SA308429381GR SA308429382GR SA308429383GR SA308429384GA SA308429385GA SA308429386GR SA308429387GR SA308429388GA SA308429389GA SA308429390GA SA308429391GR SA308429392GA SA308429393GA SA308429394GR SA308429395GA SA308429396GR SA308429397GR SA308429398GA SA308429400GA SA308429401GR SA308429402GA SA308429403GA SA308429404GA SA308429405GR SA308429406GA SA308429407GA SA308429408GR SA308429409GA SA308429410GA SA308429411GA SA308429412GA SA308429413GA SA308429414GA SA308429415GA SA308429416GA SA308429417GA SA308429418GA SA308429419GA SA308429420GA SA308429421GA SA308429422GA SA308429423GA SA308429424GA SA308439425GA SA308439426GA SA308439427GA SA308439429GA SA308439431GA SA308439433GA SA308439434GA SA308439435GA SA308439438GA SA308439439GA SA308439442GA SA308439444GA SA308439445GA SA308439446GA SA308439452GA SA308439453GA SA308439454GA SA308439455GA SA308439456GA SA308439457GA SA308439458GR SA308439459GA SA308439460GA SA308439461GA SA308439462GA SA308439463GA SA308439464GA SA308439465GA SA308439466GA SA308439468GA SA308439469GA SA308439470GA SA308439471GA SA308439472GA SA308439473GA SA308439474GA SA308439475GA SA308439476GA SA308439477GA SA308439478GR SA308439479GA SA308449480GA SA308449481GA SA308449483GA SA308449484GR SA308449485GA SA308449486GA SA308449487GA SA308449488GA SA308449489GA SA308449490GA SA308449491GA SA308449494GA SA308449495GA SA308449496GA SA308449498GA SA308449499GA SA308449500GR SA308449501GA SA308449502GA SA308449504GA SA308449505GA SA308449507GA SA308449511GA SA308449513GR SA308449515GR SA308449516GA SA308449517GA SA308449523GA SA308449524GA SA308449526GA SA308449528GA SA308449529GA SA308449530GA SA308449531GR SA308449532GA SA308449533GA SA308449534GA SA308449535GA SA308449536GR SA308449537GA SA308449538GA SA308449540GA SA308449541GA SA308449542GA SA308449543GA SA308449544GA SA308449545GA SA308449546GA SA308449547GA SA308449548GA SA308449549GA SA308449550GA SA308449551GA SA308449552GA SA308449554GA SA308449555GA SA308449556GA SA308449557GA SA308449558GA SA308449559GA SA308449560GA SA308449561GR SA308449562GA SA308449563GA SA308449564GA SA308449565GA SA308449566GR SA308449568GR SA308449569GA SA308449570GR SA308449571GR SA308449572GA SA308449574GA SA308449576GA SA308449577GR SA308449579GA SA308449581GA SA308449583GA SA308449585GA SA308449589GA SA308449590GA SA308449591GA SA308449593GR SA308449596GA SA308449597GA SA308449598GA SA308449599GA SA308449600GA SA308449601GA SA308449602GA SA308449603GA SA308449604GA SA308449606GA SA308449607GA SA308449608GA SA308449609GA SA308449610GA SA308449611GR SA308449612GA SA308449613GA SA308449614GA SA308449615GA SA308449616GA SA308449617GA SA308449618GA SA308449619GA SA308449620GA SA308449621GR SA308449622GA SA308449623GA SA308449624GA SA308449625GA SA308449626GA SA308449627GA SA308449629GA SA308449630GA SA308449631GA SA308449632GA SA308449633GA SA308449634GA SA308449635GA SA308449636GA SA308449637GA SA308459638GR SA308459639GA SA308459642GR SA308459643GA SA308459644GA SA308459645GA SA308459648GA SA308459649GA SA308459650GA SA308459652GA SA308459653GA SA308459654GA SA308459655GA SA308459656GA SA308459657GA SA308459658GA SA308459659GA SA308459661GA SA308459665GA SA308459666GA SA308459670GA SA308459674GA SA308459676GR SA308459678GR SA308459679GA SA308459690GA SA308459691GA SA308459692GA SA308459693GA SA308459694GA SA308459695GA SA308459696GA SA308459698GA SA308459699GR SA308459700GA SA308459701GA SA308459702GA SA308459703GA SA308459704GA SA308459705GA SA308459706GA SA308459707GA SA308459708GA SA308459709GA SA308459710GA SA308459711GA SA308459712GA SA308459713GA SA308459714GA SA308459715GA SA308459717GA SA308459718GA SA308459719GA SA308459720GA SA308459721GA SA308459722GA SA308459723GA SA308459725GA SA308459726GA SA308459727GA SA308459728GA SA308459729GA SA308459730GA SA308459731GA SA308459732GA SA308459733GA SA308459734GA SA308459735GA SA308459736GA SA308459737GA SA308459738GA SA308459739GA SA308459740GA SA308459741GA SA308459742GA SA308459743GA SA308459744GR SA308459745GA SA308459746GA SA308459747GA SA308459748GA SA308459749GA SA308459750GA SA308459751GA SA308459752GA SA308459754GA SA308459755GA SA308459756GA SA308459757GR SA308459758GA SA308459759GA SA308459760GA SA308459761GA SA308459762GA SA308459763GA SA308459765GA SA308459766GA SA308459767GA SA308459768GA SA308459769GA SA308459770GA SA308459771GA SA308459772GR SA308459773GA SA308459774GA SA308459776GA SA308459777GA SA308459779GA SA308459780GA SA308459781GA SA308459782GA SA308459783GA SA308459784GA SA308459785GA SA308459786GA SA308459787GA SA308459788GA SA308459789GA SA308459790GA SA308459791GA SA308459792GA SA308459793GA SA308459794GR SA308459796GA SA308459797GA SA308459798GA SA308459799GA SA308459800GA SA308459801GA SA308459802GA SA308459803GA SA308459804GA SA308459805GA SA308459806GA SA308459807GA SA308459808GA SA308459809GA SA308459810GA SA308459811GA SA308459812GA SA308459813GA SA308459814GA SA308459815GA SA308459816GA SA308459817GA SA308459818GA SA308459819GA SA308459820GA SA308459821GA SA308459822GA SA308459823GA SA308459824GA SA308459825GA SA308459826GA SA308459828GA SA308459829GA SA308459830GA SA308459831GA SA308459832GA SA308459833GA SA308459834GA SA308459835GA SA308459836GA SA308459837GA SA308459838GA SA308459839GA SA308459840GA SA308459841GA SA308459843GR SA308459844GA SA308459845GA SA308459846GR SA308459850GA SA308459862GA SA308459863GA SA308459864GA SA308459865GA SA308459866GA SA308459869GA SA308459871GA SA308459876GR SA308459877GR SA308469879GA SA308469885GR SA308469888GR SA308479892GR SA308479895GR SA308479900GA SA308481005GR SA308481014GA SA308481035GA SA308481036GA SA308481038GA SA308481041GA SA308481044GA SA308481046GA SA308481052GA SA308481053GA SA308481054GA SA308481055GA SA308481056GA SA308481057GA SA308481058GA SA308481059GA SA308481060GA SA308481061GA SA308481063GA SA308481068GA SA308489911GA SA308489912GA SA308489917GA SA308489918GA SA308489921GR SA308489922GA SA308489927GA SA308489948GA SA308489954GR SA308489957GA SA308489960GA SA308489962GA SA308489964GA SA308489971GR SA308489973GA SA308489976GR SA308489979GA SA308489983GA SA308489994GR SA308491071GA SA308491072GA SA308491073GA SA308491074GA SA308491075GA SA308491076GA SA308491077GA SA308491078GA SA308491079GA SA308491080GA SA308491081GA SA308491082GA SA308491083GA SA308491084GA SA308491085GA SA308491086GA SA308491087GA SA308491088GA SA308491089GA SA308491090GA SA308491091GR SA308491092GA SA308491093GA SA308491094GR SA308491095GR SA308491096GA SA308491097GA SA308491098GR SA308491099GR SA308491100GA SA308491101GA SA308491102GA SA308491103GA SA308491104GA SA308491105GA SA308491106GA SA308491107GA SA308491108GA SA308491109GA SA308491110GA SA308491111GR SA308491112GA SA308491113GA SA308491114GA SA308491115GA SA308491116GA SA308491117GA SA308491118GA SA308491119GA SA308491120GA SA308491121GA SA308491122GA SA308491123GA SA308491124GA SA308491125GA SA308491126GA SA308491127GA SA308491129GA SA308491130GA SA308491131GA SA308491132GA SA308491133GA SA308491134GA SA308491135GA SA308491136GA SA308491137GA SA308491138GA SA308491139GA SA308491140GA SA308491141GA SA308491142GA SA308491143GR SA308491144GA SA308491145GR SA308491146GR SA308491147GR SA308491148GR SA308491149GR SA308491151GR SA308491159GA SA308491161GA SA308491167GA SA308491168GR SA308491169GA SA308491170GA SA308491171GR SA308491172GA SA308491173GA SA308491174GA SA308491177GA SA308491179GA SA308491180GA SA308491181GA SA308491182GA SA308491183GA SA308491184GA SA308491185GA SA308491186GA SA308491187GA SA308491188GA SA308491189GA SA308491190GA SA308491191GA SA308491192GA SA308491193GA SA308491194GA SA308491195GA SA308491196GA SA308491197GA SA308491198GA SA308491199GA SA308491200GA SA308491201GA SA308491202GA SA308491203GA SA308491204GA SA308491205GA SA308491206GA SA308491207GA SA308491208GA SA308491209GA SA308491210GA SA308491211GA SA308491212GA SA308491214GA SA308491215GA SA308491217GA SA308491218GA SA308491219GR SA308491220GA SA308491223GA SA308491224GA SA308491225GA SA308491226GA SA308491227GA SA308491228GA SA308491230GA SA308491233GA SA308491234GA SA308491235GA SA308491236GA SA308491237GA SA308491238GA SA308491239GA SA308491240GA SA308491241GR SA308491243GA SA308491244GA SA308491245GA SA308491246GA SA308491247GA SA308491248GA SA308491249GA SA308491250GR SA308491251GA SA308501252GA SA308501254GA SA308501255GA SA308501256GA SA308501257GA SA308501258GA SA308501259GA SA308501260GA SA308501261GA SA308501262GA SA308501263GA SA308501264GA SA308501265GA SA308501266GA SA308501267GA SA308501268GA SA308501269GA SA308501270GA SA308501271GA SA308501272GA SA308501273GR SA308501274GA SA308501275GA SA308501276GA SA308501277GR SA308501278GR SA308501280GR SA308501281GR SA308501282GA SA308501283GA SA308501284GA SA308501285GA SA308501286GR SA308501287GA SA308501288GR SA308501289GR SA308501290GA SA308501291GA SA308501292GA SA308501293GA SA308501294GA SA308501295GA SA308501296GA SA308501297GA SA308501298GA SA308501301GA SA308501302GA SA308501303GA SA308501304GA SA308501305GA SA308501306GA SA308501307GA SA308501308GA SA308501309GR SA308501310GA SA308501311GA SA308501312GA SA308501313GA SA308501314GR SA308501315GA SA308501316GA SA308501317GA SA308501318GA SA308501319GA SA308501320GA SA308501321GA SA308501322GR SA308501323GR SA308501324GA SA308501325GA SA308501326GA SA308501327GA SA308501328GA SA308501329GA SA308501331GA SA308501332GA SA308501333GA SA308501334GA SA308501335GA SA308501336GA SA308501337GA SA308501339GA SA308501340GA SA308501341GA SA308501342GA SA308501343GA SA308501344GA SA308501349GA SA308501350GA SA308501351GA SA308501353GA SA308501354GA SA308501355GA SA308501356GA SA308501357GA SA308501358GA SA308501359GA SA308501360GA SA308501361GA SA308501362GA SA308501363GA SA308501364GA SA308501365GA SA308501366GA SA308501367GA SA308501368GA SA308501369GA SA308501370GA SA308501371GA SA308501372GA SA308501373GA SA308501374GA SA308501375GA SA308501376GA SA308501377GA SA308501378GA SA308501379GA SA308501380GA SA308501381GA SA308501382GR SA308501384GA SA308501386GA SA308501387GA SA308501388GA SA308501389GA SA308501390GA SA308501391GA SA308501392GA SA308501393GA SA308501394GA SA308501395GA SA308501396GA SA308501397GA SA308501398GA SA308501399GA SA308501401GA SA308501402GA SA308501404GA SA308501405GA SA308501406GA SA308501408GA SA308501409GA SA308501411GA SA308501413GA SA308501414GR SA308501415GA SA308501416GA SA308501419GA SA308501421GR SA308501422GA SA308501423GA SA308501424GA SA308501425GR SA308501427GA SA308501428GA SA308501430GA SA308501431GA SA308501432GA SA308501433GA SA308511435GA SA308511436GA SA308511437GA SA308511438GA SA308511439GA SA308511440GA SA308511441GA SA308511442GA SA308511444GA SA308511446GA SA308511447GR
SA308511448GA SA308511449GA SA308511451GA SA308511452GA SA308511453GA SA308511454GA SA308511455GR SA308511456GR SA308511457GA SA308511458GA SA308511459GR SA308511460GR SA308511461GR SA308511462GA SA308511464GR SA308511466GA SA308511467GR SA308511468GR SA308511469GR SA308511471GA SA308511472GA SA308511473GR SA308511474GR SA308511475GR SA308511476GA SA308511477GA SA308511478GA SA308511479GR SA308511480GA SA308511481GR SA308511483GR SA308511484GA SA308511485GA SA308511486GA SA308511487GR SA308511488GA SA308511489GR SA308511490GA SA308511491GA SA308511492GA SA308511493GA SA308511495GA SA308511496GR SA308511499GA SA308511500GR SA308511501GR SA308511502GA SA308511503GA SA308511504GA SA308511505GA SA308511506GR SA308511508GA SA308511509GA SA308511510GR SA308511511GR SA308511512GR SA308511513GR SA308511514GA SA308511517GA SA308511519GA SA308511520GR SA308511521GR SA308511522GR SA308511523GA SA308511524GA SA308511525GA SA308511526GA SA308511527GA SA308511529GR SA308511530GA SA308511531GR SA308511532GR SA308511533GA SA308511534GA SA308511535GA SA308511536GA SA308511537GA SA308511538GA SA308511540GR SA308511541GA SA308511542GA SA308511545GR SA308511547GA SA308511548GA SA308511551GR SA308511552GA SA308511554GA SA308511555GA SA308511556GA SA308511557GA SA308511558GA SA308511559GR SA308511560GA SA308511561GA SA308511562GA SA308511563GA SA308511564GA SA308511565GA SA308511566GA SA308511567GA SA308511568GA SA308511569GA SA308511571GA SA308511572GA SA308511573GA SA308511574GA SA308511575GA SA308511576GA SA308511577GA SA308511578GA SA308511579GA SA308511580GA SA308511581GA SA308511582GA SA308511583GA SA308511584GA SA308511585GA SA308511586GA SA308511587GA SA308511588GA SA308511589GA SA308511590GA SA308511591GA SA308511592GA SA308511593GA SA308511594GA SA308511595GA SA308511596GA SA308511597GA SA308511598GA SA308511599GA SA308511600GA SA308511601GA SA308511602GA SA308511603GA SA308511604GA SA308511605GA SA308511606GA SA308511607GA SA308511608GA SA308511609GA SA308511610GA SA308511611GA SA308511612GA SA308511613GA SA308511614GA SA308511615GA SA308511616GA SA308511617GA SA308511618GA SA308511619GA SA308511620GA SA308511621GA SA308511622GA SA308511623GA SA308511624GA SA308511625GR SA308511626GA SA308511627GA SA308511628GR SA308511629GA SA308511630GA SA308511631GR SA308511634GA SA308511636GA SA308511637GA SA308511638GA SA308511640GR SA308511641GA SA308511643GR SA308511644GA SA308511645GA SA308511646GA SA308511648GA SA308511649GA SA308511650GA SA308511651GA SA308511652GA SA308511653GA SA308511654GA SA308511656GA SA308511657GA SA308511658GA SA308511659GA SA308511660GA SA308511661GA SA308511662GA SA308511663GA SA308511664GA SA308511666GA SA308511667GA SA308511668GA SA308511669GA SA308511671GA SA308511672GA SA308511674GR SA308511675GA SA308511677GA SA308511678GA SA308511679GA SA308511680GA SA308511681GA SA308511682GA SA308511684GA SA308511685GA SA308511686GA SA308511687GA SA308511689GA SA308511690GR SA308511693GA SA308511695GA SA308511696GA SA308511697GA SA308511698GA SA308511699GA SA308511702GA SA308511703GA SA308511704GA SA308511705GA SA308511708GA SA308511710GA SA308511711GA SA308511715GA SA308511717GA SA308511719GA SA308511723GA SA308511724GA SA308511725GA SA308511727GA SA308511728GA SA308511746GA SA308511750GA SA308511752GA SA308511753GA SA308511754GA SA308511755GA SA308511757GA SA308511763GA SA308511764GA SA308511765GA SA308511769GA SA308511771GA SA308521773GA SA308521774GA SA308521775GA SA308521776GA SA308521777GA SA308521778GR SA308521779GA SA308521780GA SA308521781GA SA308521782GA SA308521783GA SA308521784GA SA308521785GA SA308521786GA SA308521787GA SA308521788GR SA308521789GA SA308521790GR SA308521791GA SA308521792GA SA308521794GA SA308521795GA SA308521796GA SA308521798GA SA308521799GA SA308521801GA SA308521802GA SA308521803GA SA308521804GA SA308521805GA SA308521807GA SA308521809GA SA308521810GA SA308521811GA SA308521812GA SA308521813GA SA308521814GA SA308521815GA SA308521817GA SA308521818GA SA308521819GA SA308521821GA SA308521822GA SA308521823GA SA308521824GA SA308521825GR SA308521826GA SA308521827GA SA308521828GA SA308521829GA SA308521830GA SA308521831GA SA308521832GA SA308521833GA SA309011834GA SA309011835GA SA309011836GA SA309011837GA SA309011838GA SA309011839GA SA309011840GA SA309011841GA SA309011842GA SA309011843GA SA309011844GA SA309011845GA SA309011846GA SA309011848GA SA309011849GA SA309011850GA SA309011851GA SA309011852GA SA309011854GA SA309011855GA SA309011856GA SA309011857GA SA309011858GR SA309011859GA SA309011860GR SA309011861GR SA309011862GA SA309011863GA SA309011864GA SA309011865GA SA309011866GA SA309011869GA SA309011870GA SA309011871GA SA309011872GA SA309011873GA SA309011875GA SA309011876GA SA309011877GA SA309011878GA SA309011879GA SA309011881GA SA309011883GA SA309011884GA SA309011885GA SA309011886GA SA309011887GA SA309011888GA SA309011889GA SA309011890GA SA309011891GA SA309011892GA SA309011894GA SA309011895GA SA309011896GA SA309011897GA SA309011898GA SA309011899GA SA309011900GA SA309011901GA SA309011902GA SA309011903GA SA309021904GA SA309021905GA SA309021906GA SA309021907GA SA309021908GA SA309021909GR SA309021910GA SA309021911GR SA309021912GA SA309021913GA SA309021914GA SA309021915GA SA309021916GA SA309021917GA SA309021918GA SA309021919GR SA309021920GA SA309021921GA SA309021922GA SA309021923GA SA309021926GA SA309021927GA SA309021929GA SA309021930GA SA309021931GA SA309021932GA SA309021934GA SA309021935GA SA309021940GA SA309021941GA SA309021942GA SA309021943GA SA309021944GA SA309021945GA SA309021946GA SA309021947GA SA309021949GA SA309021950GA SA309021951GA SA309021952GA SA309021954GA SA309021955GA SA309021956GA SA309021958GA SA309021959GA SA309021960GA SA309021961GA SA309021962GA SA309021963GA SA309021964GA SA309021965GA SA309021966GA SA309021967GA SA309021968GA SA309021969GA SA309021970GA SA309021971GA SA309021972GA SA309021973GA SA309021974GA SA309021976GA SA309021977GR SA309021978GR SA309021979GA SA309021980GA SA309021981GA SA309021982GA SA309021983GA SA309021984GA SA309021985GA SA309021986GA SA309021988GR SA309031989GA SA309031990GA SA309031991GA SA309031992GA SA309031993GA SA309031995GA SA309031997GA SA309031998GA SA309031999GA SA309032000GA SA309032001GA SA309032002GA SA309032003GA SA309032004GA SA309032005GA SA309032006GA SA309032007GA SA309032008GA SA309032010GA SA309032014GR SA309032015GA SA309032016GR SA309032018GA SA309032019GA SA309032020GA SA309032022GA SA309032024GA SA309032025GA SA309032026GA SA309032027GA SA309032028GA SA309032029GA SA309032030GA SA309032031GA SA309032032GA SA309032033GA SA309032034GA SA309032037GA SA309032039GA SA309032040GA SA309032041GA SA309032042GA SA309032043GA SA309032048GA SA309032050GA SA309032052GA SA309032064GA SA309032066GA SA309032070GA SA309032071GA SA309032072GA SA309032073GA SA309032074GR SA309032075GA SA309032076GA SA309032077GA SA309032078GA SA309032080GA SA309032083GR SA309032086GA SA309032087GA SA309032091GA SA309032092GA SA309032105GA SA309032106GA SA309032108GA SA309032110GA SA309032113GA SA309042116GA SA309042118GA SA309042122GA SA309042123GA SA309042127GA SA309042131GA SA309042133GA SA309042145GA SA309042150GA SA309042169GA SA309042170GA SA309042173GA SA309042180GA SA309042183GA SA309042184GA SA309042185GA SA309042186GA SA309042187GA SA309042189GA SA309042195GA SA309042201GA SA309042207GA SA309042208GA SA309042213GA SA309042214GA SA309052216GA SA309052219GA SA309052220GA SA309052222GA SA309052224GA SA309052225GA SA309052227GA SA309052228GA SA309052229GA SA309052233GR SA309052236GA SA309052239GA SA309062266GA SA309062267GA SA309062269GA SA309062271GA SA309062273GA SA309062274GA SA309062277GA SA309062278GA SA309062279GA SA309062281GA SA309062282GA SA309062283GA SA309062284GR SA309062285GA SA309062286GR SA309062292GA SA309062296GA SA309062298GA SA309062299GA SA309062300GA SA309062302GA SA309062303GA SA309062304GA SA309062305GA SA309062308GA SA309062309GA SA309062310GA SA309062312GR SA309062313GA SA309062315GA SA309062318GA SA309062325GR SA309062330GR SA309062336GA SA309072343GA SA309072344GA SA309072347GA SA309072348GA SA309072350GA SA309072351GA SA309072352GA SA309072355GA SA309072356GA SA309072357GA SA309072362GA SA309072363GA SA309072364GA SA309072365GA SA309072366GA SA309072368GA SA309072370GA SA309072372GA SA309072374GA SA309072375GA SA309072377GA SA309072379GA SA309072380GA SA309072382GA SA309072383GA SA309072384GA SA309072385GA SA309072387GA SA309072391GA SA309072393GR SA309072400GR SA309072401GR SA309082415GR SA309082423GA SA309082429GA SA309082443GA SA309082449GA SA309082452GA SA309082454GA SA309082456GA SA309082457GA SA309082462GA SA309092474GA SA309092477GA SA309092478GA SA309092479GA SA309092480GA SA309092481GA SA309092482GA SA309092483GA SA309092484GA SA309092485GA SA309092486GA SA309092487GA SA309092488GR SA309092489GA SA309092490GA SA309092491GA SA309092492GA SA309092493GA SA309092494GR SA309092496GA SA309092500GA SA309092502GA SA309092503GA SA309092505GA SA309092506GA SA309092507GA SA309092508GA SA309092510GA SA309092511GA SA309092512GA SA309092513GA SA309092514GA SA309092515GA SA309092517GA SA309092518GA SA309092519GA SA309092520GA SA309092521GA SA309092522GA SA309092524GA SA309092525GA SA309092526GA SA309092527GA SA309092528GA SA309092529GA SA309092530GA SA309092531GA SA309092532GA SA309092533GA SA309092534GA SA309092535GA SA309092536GA SA309092537GA SA309092538GA SA309092539GA SA309092540GA SA309092541GA SA309092542GA SA309092543GA SA309112544GA SA309112545GA SA309112546GA SA309112547GA SA309112548GA SA309112549GA SA309112550GA SA309112551GA SA309112552GA SA309112553GA SA309112554GA SA309112555GA SA309112556GA SA309112557GA SA309112558GA SA309112559GA SA309112560GA SA309112562GA SA309112563GA SA309112565GA SA309112566GR SA309112567GA SA309112568GA SA309112569GA SA309112571GR SA309112572GA SA309112573GA SA309112574GA SA309112576GA SA309112577GA SA309112578GA SA309112579GA SA309112580GA SA309112581GA SA309112582GA SA309112583GA SA309112584GA SA309112585GA SA309112586GA SA309112587GA SA309112588GA SA309112589GA SA309112590GA SA309112591GA SA309112592GA SA309112593GA SA309112594GA SA309112595GA SA309112596GA SA309112597GA SA309112598GR SA309112599GA SA309112600GR SA309112602GA SA309112603GA SA309112604GA SA309112605GA SA309112606GA SA309112607GR SA309112608GR SA309112609GA SA309112611GR SA309112612GA SA309112613GA SA309112614GR SA309112615GR SA309112616GR SA309112617GA SA309112618GA SA309122619GR SA309122620GA SA309122621GA SA309122622GA SA309122623GA SA309122624GA SA309122625GA SA309122626GA SA309122627GA SA309122628GA SA309122629GA SA309122630GA SA309122631GA SA309122632GA SA309122633GA SA309122634GA SA309122635GA SA309122636GA SA309122638GA SA309122639GA SA309122640GR SA309122643GA SA309122644GA SA309122645GA SA309122646GA SA309122647GA SA309122648GA SA309122649GA SA309122650GA SA309122652GA SA309122653GA SA309122654GA SA309122655GA SA309122656GA SA309122657GA SA309122658GA SA309122659GA SA309122661GA SA309122662GA SA309122663GA SA309122664GR SA309122665GA SA309122666GA SA309122667GA SA309122669GA SA309122670GA SA309122671GA SA309122672GA SA309122673GA SA309122674GA SA309122676GA SA309122677GA SA309122678GA SA309122680GA SA309122681GA SA309122683GA SA309122684GA SA309122685GA SA309122686GA SA309122687GA SA309122688GA
SA309122689GA SA309122691GR SA309122692GA SA309122694GA SA309122695GA SA309122696GA SA309122697GA SA309122699GA SA309122712GA SA309122713GA SA309122715GA SA309122717GR SA309122719GA SA309122720GA SA309122721GR SA309122724GR SA309122727GA SA309122729GA SA309122730GA SA309122732GA SA309122733GR SA309122735GA SA309122752GA SA309122757GA SA309122764GR SA309122765GA SA309122766GA SA309122769GA SA309122770GR SA309122771GA SA309122773GA SA309122774GA SA309122775GA SA309122776GA SA309122778GA SA309122779GA SA309122782GA SA309122784GA SA309122791GA SA309132792GA SA309132793GA SA309132794GA SA309132795GA SA309132796GA SA309132797GA SA309132798GA SA309132799GA SA309132802GA SA309132804GA SA309132807GA SA309132808GA SA309132810GA SA309132811GA SA309132813GR SA309132815GA SA309132816GA SA309132818GA SA309132819GA SA309132820GA SA309132821GA SA309132824GA SA309132825GA SA309132826GR SA309132827GA SA309132828GA SA309132829GA SA309132830GA SA309132831GR SA309132832GR SA309132833GR SA309132834GA SA309132835GR SA309132836GR SA309132837GA SA309132838GA SA309132839GA SA309132844GA SA309132846GA SA309132856GA SA309132858GA SA309132859GA SA309132860GA SA309132861GA SA309132862GA SA309132863GA SA309132866GA SA309132867GA SA309132868GA SA309132869GA SA309142872GA SA309142873GA SA309142875GA SA309142876GA SA309142878GA SA309142879GA SA309142881GA SA309142882GA SA309142883GA SA309142884GA SA309142885GA SA309142886GA SA309142887GA SA309142888GA SA309142890GA SA309142891GA SA309142893GR SA309142894GA SA309142896GA SA309142903GA SA309142904GA SA309142910GA SA309142911GA SA309142912GA SA309142913GA SA309142914GA SA309142915GA SA309142917GA SA309142918GA SA309142919GA SA309142926GA SA309142928GA SA309142929GA SA309142930GA SA309142931GA SA309142932GA SA309142934GA SA309142936GA SA309142938GA SA309142939GA SA309142940GA SA309142941GA SA309142942GA SA309142943GA SA309142944GA SA309142945GA SA309142946GA SA309142947GA SA309142948GA SA309142950GA SA309142951GA SA309142953GA SA309142954GA SA309142955GA SA309142956GA SA309142957GA SA309142958GA SA309142959GA SA309142960GA SA309142961GA SA309142962GA SA309142964GA SA309142966GA SA309142967GA SA309142969GA SA309142970GA SA309142971GA SA309142972GA SA309142973GA SA309142974GA SA309142975GA SA309142976GA SA309142977GA SA309142978GA SA309142979GA SA309142980GA SA309142993GA SA309142994GA SA309142995GA SA309142996GA SA309142997GA SA309142998GA SA309142999GA SA309143000GA SA309143001GA SA309143002GA SA309143003GA SA309143004GA SA309143005GA SA309143006GA SA309143007GA SA309143008GA SA309143009GA SA309143010GA SA309143011GA SA309143012GA SA309143013GA SA309143014GA SA309143015GA SA309143016GR SA309143017GA SA309143018GA SA309143019GA SA309143020GA SA309143021GA SA309143022GA SA309143023GA SA309153029GA SA309153035GA SA309153045GA SA309153055GA SA309153056GA SA309153058GA SA309153059GA SA309153061GA SA309153064GA SA309153065GA SA309153066GA SA309153067GA SA309153068GA SA309153069GA SA309153075GA SA309153079GA SA309153081GA SA309153082GA SA309153083GA SA309153084GA SA309153086GA SA309153087GA SA309153088GA SA309153090GA SA309153091GA SA309153092GA SA309153093GA SA309153094GA SA309153095GA SA309153096GA SA309153097GA SA309153098GA SA309153099GA SA309173100GA SA309173101GA SA309173105GA SA309173110GA SA309173111GA SA309173117GA SA309173126GA SA309173127GA SA309173128GA SA309173140GA SA309173145GA SA309173146GA SA309173147GA SA309173149GA SA309173155GA SA309173157GA SA309173160GA SA309173163GA SA309173165GA SA309173170GA SA309173171GA SA309173174GA SA309173175GA SA309173178GA SA309173180GA SA309173184GA SA309183190GA SA309183191GA SA309183192GA SA309183193GA SA309183194GA SA309183195GA SA309183197GA SA309183198GA SA309183201GA SA309183205GA SA309183206GA SA309183208GA SA309183211GA SA309183213GA SA309183214GA SA309183215GA SA309183217GA SA309183218GA SA309183219GA SA309183220GA SA309183221GA SA309183222GA SA309183223GA SA309183224GA SA309183225GA SA309183226GA SA309183227GA SA309183228GA SA309183229GA SA309183230GA SA309183231GA SA309183232GA SA309183233GA SA309183234GA SA309183235GA SA309183236GA SA309183238GA SA309183239GA SA309183241GA SA309183243GA SA309183244GA SA309183245GA SA309183246GA SA309183247GA SA309183248GA SA309183251GA SA309183252GA SA309193253GA SA309193254GA SA309193255GA SA309193256GA SA309193257GA SA309193258GA SA309193259GA SA309193260GA SA309193262GA SA309193263GA SA309193264GA SA309193265GA SA309193267GA SA309193269GA SA309193272GA SA309193273GA SA309193276GA SA309193277GA SA309193278GA SA309193280GA SA309193281GA SA309193282GA SA309193283GA SA309193285GA SA309193286GA SA309193287GA SA309193288GA SA309193289GA SA309193290GA SA309193291GA SA309193292GA SA309193293GA SA309193295GA SA309193300GA SA309193302GA SA309193306GA SA309193309GA SA309193310GA SA309193311GR SA309193312GA SA309193314GA SA309193315GA SA309193318GA SA309193322GA SA309193328GA SA309193329GA SA309193331GA SA309193336GA SA309193337GA SA309193339GA SA309193340GA SA309193343GA SA309193346GA SA309193352GA SA309193356GA SA309193357GA SA309193360GA SA309193362GA SA309203379GR SA309203384GA SA309203388GA SA309203390GA SA309203391GA SA309203393GA SA309203394GA SA309203403GR SA309203405GR SA309203406GR SA309203407GA SA309203409GA SA309203417GA SA309203421GR SA309203424GA SA309203425GR SA309203427GA SA309203429GR SA309203430GA SA309203433GA SA309223435GR SA309223442GA SA309223443GR SA309223444GA SA309223445GA SA309223450GA SA309223455GA SA309223456GA SA309223458GR SA309223462GA SA309223465GR SA309223466GR SA309223469GR SA309223471GR SA309223472GA SA309223476GR SA309223482GR SA309223484GR SA309223485GA SA309223489GA SA309223503GA SA309223508GA SA309223516GA SA309233517GA SA309233518GA SA309233522GA SA309233523GR SA309233524GA SA309233525GR SA309233526GR SA309233527GR SA309233528GA SA309233529GA SA309233530GA SA309233531GR SA309233533GR SA309233534GA SA309233536GA SA309233539GA SA309233547GA SA309233550GA SA309233561GA SA309233562GA SA309233563GA SA309233564GA SA309233565GA SA309233566GA SA309233567GA SA309233568GA SA309233569GA SA309233570GA SA309233571GA SA309233572GA SA309233573GA SA309233574GA SA309233575GA SA309233576GA SA309233577GA SA309233578GA SA309233579GR SA309233580GA SA309233581GA SA309233582GA SA309233583GA SA309233584GA SA309233585GA SA309233586GA SA309233587GA SA309233588GA SA309233589GA SA309233590GA SA309233591GA SA309243592GA SA309243593GA SA309243594GA SA309243595GA SA309243596GA SA309243597GA SA309243599GA SA309243600GA SA309243601GA SA309243602GA SA309243603GA SA309243604GA SA309243605GA SA309243606GA SA309243607GA SA309243608GA SA309243609GA SA309243610GA SA309243611GA SA309243612GA SA309243613GA SA309243614GA SA309243615GA SA309243616GA SA309243617GA SA309243618GA SA309243619GA SA309243620GA SA309243621GA SA309243622GA SA309243624GA SA309243625GA SA309243626GA SA309243627GA SA309243628GA SA309243629GA SA309243630GA SA309243631GA SA309243634GA SA309243648GA SA309243658GA SA309243663GA SA309243665GA SA309243667GA SA309243668GA SA309243669GA SA309243671GA SA309243673GA SA309243674GA SA309243675GR SA309243677GA SA309243678GA SA309243679GA SA309243680GA SA309243681GA SA309243682GA SA309243684GA SA309243685GA SA309243686GA SA309243687GA SA309243688GR SA309243689GA SA309243690GA SA309243691GA SA309243692GA SA309243693GA SA309243694GA SA309243695GR SA309243696GA SA309243697GA SA309243698GA SA309243700GA SA309243701GA SA309243703GA SA309243706GA SA309243708GA SA309243710GA SA309253714GA SA309253715GA SA309253716GA SA309253717GA SA309253718GA SA309253719GA SA309253720GA SA309253723GA SA309253726GA SA309253728GA SA309253730GA SA309253731GA SA309253732GA SA309253743GA SA309253744GA SA309253750GA SA309253753GA SA309253755GA SA309253756GA SA309253757GA SA309253760GA SA309253761GA SA309253763GA SA309253765GA SA309253767GR SA309253771GA SA309253773GA SA309253774GR SA309253777GA SA309253778GA SA309253781GA SA309253782GR SA309253784GA SA309253790GR SA309253792GA SA309253807GA SA309253812GA SA309253833GA SA309253834GA SA309253837GR SA309253839GR SA309253840GA SA309253845GR SA309253848GA SA309253849GA SA309253850GR SA309253852GR SA309253855GA SA309253856GA SA309253861GR SA309253863GA SA309253866GR SA309253867GA SA309253869GA SA309253871GA SA309253873GA SA309253874GA SA309253878GA SA309253880GA SA309253881GA SA309253882GR SA309253885GR SA309253886GA SA309253887GA SA309253888GA SA309253890GA SA309253891GA SA309253894GA SA309253895GR SA309253896GA SA309253899GR SA309253906GA SA309253909GA SA309253911GA SA309263914GA SA309263917GA SA309263918GA SA309263919GA SA309263921GA SA309263923GA SA309263924GA SA309263925GA SA309263929GA SA309263930GA SA309263933GA SA309263935GA SA309263937GA SA309263938GA SA309263939GA SA309263940GA SA309263941GA SA309263943GA SA309263945GA SA309263946GA SA309263948GA SA309263949GA SA309263951GA SA309263952GA SA309263953GA SA309263954GA SA309263957GA SA309263958GA SA309263969GA SA309283976GA SA309283977GA SA309283978GA SA309283983GR SA309283989GA SA309283991GA SA309283992GA SA309283995GA SA309283999GR SA309284001GA SA309284004GA SA309284005GA SA309284007GR SA309284008GA SA309284009GA SA309284011GA SA309284012GA SA309284013GA SA309294028GA SA309294029GA SA309294030GA SA309294035GA SA309294037GA SA309294044GA SA309294047GA SA309294048GA SA309294050GA SA309294051GA SA309294053GA SA309294055GA SA309294057GA SA309294058GA SA309294060GA SA309294061GA SA309294064GA SA309294065GA SA309294067GA SA309294068GA SA309294071GA SA309294079GA SA309294084GA SA309294085GA SA309294087GA SA309294089GA SA309294090GA SA309294092GA SA309294093GA SA309294103GA SA309294112GA SA309294113GA SA309294116GR SA309304179GR SA309304201GR SA309304218GA SA309304220GA SA309304229GA SA309304234GA SA309304237GA SA309304240GA SA309304244GR SA309304245GA SA309304248GA SA309304280GR SA309304283GR SA309304284GA SA309324303GA SA309324317GA SA309324321GA SA309324342GR SA309324355GA SA309324358GA SA309324359GA SA309324361GR SA309324362GA SA309324363GA SA309324364GA SA309324366GR SA309324367GA SA309324370GA SA309324372GA SA309324374GA SA309324378GA SA309324379GA SA309324381GA SA309324388GA SA309324390GA SA309324394GA SA309324395GR SA309324397GA SA309324398GA SA309324403GA SA309324405GA SA309324406GA SA309324407GA SA309324411GR SA309324413GA SA309324419GA SA309324420GA SA309324421GA SA309324422GA SA309324423GA SA309324424GR SA309324425GA SA309324435GA SA309324436GA SA309324437GA SA309324438GA SA309324442GA SA309324443GA SA309324445GA SA309324470GA SA309324481GA SA309324482GA SA309324485GA SA309324504GA SA309324515GA SA309324516GA SA309324519GA SA309324520GA SA309324521GA SA309324527GA SA309324532GA SA309324533GA SA309324534GA SA309324537GA SA309324538GA SA309324539GA SA309324540GA SA309324541GA SA309324542GA SA309324543GA SA309324544GA SA309324545GA SA309324546GA SA309324547GA SA309324548GR SA309324549GA SA309324551GA SA309334557GA SA309334558GA SA309334560GA SA309334561GA SA309334562GA SA309334563GA SA309334564GA SA309334566GA SA309334569GA SA309334571GA SA309334574GA SA309334575GA SA309334576GA SA309334577GA SA309334579GA SA309334581GA SA309334582GR SA309334583GA SA309334587GA SA309334588GA SA309334590GR SA309334591GA SA309334592GA SA309334593GA SA309334596GA SA309334597GA SA309334598GA SA309334602GA SA309334604GA SA309334612GR
SA309334621GA SA309334638GA SA309334639GR SA309334642GA SA309334650GA SA309334652GA SA309334664GA SA309334665GA SA309334669GA SA309334675GR SA309334681GA SA309334700GA SA309334701GA SA309334705GA SA309334706GA SA309334713GA SA309334721GA SA309334747GA SA309334753GA SA309334756GR SA309334757GA SA309334758GA SA309334759GA SA309334760GA SA309334761GR SA309334762GA SA309334763GA SA309334764GA SA309334766GA SA309334767GA SA309334768GA SA309334770GA SA309334771GA SA309334772GA SA309334774GA SA309334776GA SA309334777GA SA309334778GA SA309334779GA SA309334780GA SA309334781GA SA309334782GA SA309334783GA SA309334784GA SA309334785GR SA309334786GA SA309334789GR SA309334790GR SA309334793GR SA309334794GA SA309334795GA SA309334797GA SA309334799GA SA309334800GA SA309334801GA SA309334802GA SA309334803GA SA309334804GA SA309334806GA SA309334807GA SA309334808GA SA309334809GR SA309334810GA SA309334811GA SA309334813GA SA309334814GA SA309334815GA SA309334816GA SA309334823GA SA309334826GA SA309334828GA SA309334829GA SA309334830GA SA309334831GA SA309344834GA SA309344850GR SA309344857GA SA309344859GA SA309344862GR SA309344863GR SA309344864GR SA309344865GR SA309344869GA SA309344871GA SA309344872GA SA309344873GA SA309344874GA SA309344875GA SA309344876GA SA309344877GA SA309344878GA SA309344879GA SA309344880GA SA309354882GA SA309354883GR SA309354884GA SA309354885GA SA309354886GA SA309354887GA SA309354888GA SA309354889GA SA309354891GA SA309354892GR SA309354893GA SA309354897GA SA309354898GR SA309354900GR SA309354903GR SA309354904GA SA309354905GA SA309354906GR SA309354907GR SA309354909GA SA309354910GA SA309354914GA SA309354916GR SA309354919GR SA309354922GA SA309354924GA SA309354926GR SA309354928GA SA309354930GA SA309354933GA SA309354935GR SA309354936GA SA309354937GA SA309354939GA SA309354941GA SA309354942GR SA309354945GR SA309354948GA SA309354949GA SA309354950GA SA309354952GA SA309354953GA SA309354955GA SA309354956GA SA309354957GA SA309354959GA SA309354962GA SA309354963GA SA309354964GA SA309354965GA SA309354966GR SA309354967GR SA309354968GA SA309354969GA SA309354970GA SA309354971GA SA309354972GR SA309354974GA SA309354976GA SA309354977GA SA309354978GA SA309354979GA SA309354980GA SA309354982GA SA309354985GA SA309354986GA SA309354987GA SA309354988GA SA309354989GA SA309354990GA SA309354991GA SA309354993GA SA309354995GA SA309355000GA SA309355002GA SA309355003GA SA309355004GA SA309355006GA SA309355007GA SA309355008GA SA309355009GR SA309355010GA SA309355012GA SA309355013GA SA309355014GA SA309355015GA SA309355016GA SA309355017GA SA309355019GA SA309355020GA SA309355022GA SA309355023GR SA309355024GR SA309355025GA SA309355026GA SA309355027GA SA309365028GA SA309365029GA SA309365032GR SA309365033GA SA309365034GR SA309365036GR SA309365037GR SA309365038GR SA309365039GA SA309365041GR SA309365042GR SA309365043GR SA309365045GA SA309365046GR SA309365047GA SA309365050GR SA309365051GR SA309365053GR SA309365054GA SA309365055GA SA309365056GR SA309365057GR SA309365059GA SA309365060GA SA309365061GA SA309365062GA SA309365063GA SA309365064GA SA309365065GA SA309365067GA SA309365068GA SA309365069GR SA309365071GR SA309365072GA SA309365073GR SA309365074GR SA309365075GA SA309365076GA SA309365077GR SA309365078GA SA309365079GA SA309365082GA SA309365083GR SA309365084GA SA309365086GA SA309365091GA SA309365092GA SA309365093GR SA309365094GR SA309365095GA SA309365096GR SA309365098GA SA309365099GA SA309365103GA SA309365105GA SA309365108GA SA309365110GA SA309365111GA SA309365112GA SA309365113GA SA309365118GA SA309365120GA SA309365122GR SA309365123GR SA309365127GA SA309365129GR SA309365130GR SA309365131GA SA309365134GA SA309365141GA SA309365143GA SA309365147GR SA309365150GA SA309365151GA SA309365154GA SA309365155GR SA309365157GA SA309365158GA SA309365162GA SA309365163GA SA309365165GA SA309365166GA SA309365167GA SA309365168GR SA309365171GA SA309365173GA SA309365174GA SA309365178GA SA309365179GA SA309365180GA SA309365181GA SA309365182GA SA309365183GA SA309365184GR SA309365185GA SA309365186GA SA309365187GR SA309365189GA SA309365193GA SA309365197GA SA309365198GA SA309365199GR SA309365203GA SA309365204GA SA309365205GR SA309365206GA SA309365207GR SA309365214GA SA309365215GR SA309365216GA SA309365217GA SA309365221GA SA309365223GR SA309365225GR SA309365226GR SA309365228GA SA309365229GR SA309365230GR SA309365231GR SA309365233GR SA309365234GR SA309365235GR SA309365236GR SA309365237GR SA309365239GR SA309365240GR SA309365241GA SA309365243GA SA309365245GR SA309365246GA SA309365248GR SA309365249GR SA309365250GA SA309365251GA SA309365252GR SA309365254GR SA309365257GA SA309365260GA SA309365261GA SA309365262GA SA309365263GA SA309365264GA SA309365266GA SA309365268GR SA309365270GA SA309365271GR SA309365272GR SA309365273GR SA309365274GR SA309365275GR SA309365276GA SA309365277GR SA309365278GR SA309365279GA SA309365280GA SA309365283GA SA309365285GA SA309365286GR SA309365287GA SA309365288GA SA309365289GA SA309365291GA SA309365292GA SA309365293GA SA309365294GA SA309365295GA SA309365296GA SA309365297GA SA309365298GR SA309365299GA SA309365300GR SA309365301GA SA309365302GA SA309365303GR SA309365304GR SA309365305GA SA309365306GA SA309365307GA SA309365308GR SA309365309GR SA309365310GA SA309365311GR SA309365312GA SA309365315GR SA309365316GR SA309365318GA SA309365319GA SA309365321GR SA309365322GR SA309365323GA SA309365324GA SA309365325GR SA309365326GA SA309365327GR SA309365329GA SA309365331GR SA309365332GA SA309365333GR SA309365335GA SA309365336GR SA309365337GR SA309365338GR SA309365339GR SA309365340GR SA309365341GA SA309365342GA SA309365343GR SA309365344GA SA309365345GA SA309365346GA SA309365348GR SA309365349GA SA309365350GR SA309365351GR SA309365352GR SA309365353GR SA309365355GA SA309365356GR SA309365357GR SA309365358GR SA309365359GR SA309365360GR SA309365361GA SA309365362GA SA309375363GA SA309375364GR SA309375366GA SA309375367GR SA309375368GR SA309375369GA SA309375370GA SA309375371GR SA309375372GR SA309375373GR SA309375375GR SA309375376GR SA309375377GR SA309375378GA SA309375379GA SA309375380GR SA309375381GR SA309375382GA SA309375383GA SA309375384GA SA309375392GA SA309375393GA SA309375394GA SA309375395GA SA309375396GA SA309375397GA SA309375399GA SA309375400GA SA309375401GA SA309375402GA SA309375403GA SA309375404GA SA309375406GA SA309375407GA SA309375408GA SA309375409GA SA309375410GA SA309375411GA SA309375412GA SA309375413GA SA309375414GA SA309375415GA SA309375416GA SA309375418GR SA309375419GR SA309375420GA SA309375422GA SA309375423GR SA309375424GA SA309375425GR SA309375426GR SA309375428GR SA309375429GR SA309375430GR SA309375432GR SA309375433GR SA309375434GA SA309375435GA SA309375436GR SA309375440GR SA309375441GR SA309375442GR SA309375444GR SA309375447GR SA309375448GR SA309375449GR SA309375451GR SA309375452GR SA309375453GR SA309375454GR SA309375455GR SA309375456GR SA309375457GA SA309375458GR SA309375459GR SA309375460GA SA309375463GA SA309375465GA SA309375466GA SA309375467GR SA309375468GR SA309375469GA SA309375470GR SA309375471GR SA309375473GR SA309375474GA SA309375475GA SA309375477GR SA309375482GR SA309375483GA SA309375488GR SA309375492GA SA309375494GA SA309375495GR SA309375502GR SA309375506GR SA309375507GA SA309385510GA SA309385514GA SA309385524GA SA309385528GR SA309385538GA SA309385544GA SA309385546GA SA309385547GA SA309385550GR SA309385552GA SA309385553GA SA309385554GA SA309385556GA SA309385561GR SA309385567GA SA309385574GA SA309385579GA SA309385584GA SA309385605GA SA309385607GA SA309385608GA SA309385611GA SA309385627GA SA309385628GR SA309385637GA SA309385640GA SA309395705GR SA309395708GA SA309395710GR SA309395717GR SA309395741GA SA309395746GR SA309395754GR SA309395755GA SA309395756GA SA309395777GA SA309395792GR SA309395793GA SA309395794GR SA309395812GR SA309405821GR SA309405824GA SA309405825GR SA309405827GA SA309405832GA SA309405834GA SA309405836GA SA309405837GA SA309405838GA SA309405841GA SA309405844GA SA309405845GR SA309405848GA SA309405851GR SA309405852GA SA309405856GA SA309405857GA SA309405858GA SA309405859GA SA309405860GA SA309405862GA SA309405864GA SA309405865GA SA309405866GA SA309405867GA SA309405868GA SA309405869GA SA309405871GA SA309405873GA SA309405874GA SA309405875GA SA309405876GA SA309405877GA SA309405879GA SA309405880GA SA309405881GA SA309405883GA SA309405884GA SA309405885GA SA309405886GA SA309405887GA SA309405888GA SA309405889GA SA309405890GA SA309405891GA SA309405892GA SA309405893GA SA309405894GA SA309405895GA SA309405896GA SA309405897GA SA309405898GA SA309405899GA SA309405900GA SA309405901GA SA309405902GA SA309405903GA SA309405904GA SA309405906GA SA309405909GA SA309405912GA SA309405913GA SA309405914GA SA309405915GA SA309405917GA SA309405918GA SA309405919GA SA309405921GA SA309415922GA SA309415924GA SA309415925GA SA309415926GA SA309415927GA SA309415928GA SA309415929GA SA309415931GA SA309415933GA SA309415934GA SA309415935GA SA309415936GA SA309415937GA SA309415938GA SA309415939GA SA309415949GA SA309415950GA SA309415951GA SA309415952GA SA309415953GA SA309415954GA SA309415955GA SA309415956GA SA309415957GA SA309415959GA SA309415960GA SA309415961GA SA309415962GA SA309415964GA SA309415966GA SA309415967GA SA309415968GA SA309415969GA SA309415970GA SA309415971GA SA309415987GA SA309415988GA SA309415989GA SA309415990GA SA309415991GA SA309415992GA SA309415993GA SA309415994GA SA309415995GA SA309415996GA SA309415997GA SA309415998GA SA309415999GA SA309416000GA SA309416001GA SA309416002GA SA309416003GA SA309416004GA SA309416005GR SA309416006GA SA309416007GA SA309416008GA SA309416009GA SA309416011GA SA309416012GA SA309416013GA SA309416014GA SA309416015GA SA309416016GA SA309416017GA SA309416018GA SA309416020GA SA309426022GA SA309426023GA SA309426025GA SA309426026GA SA309426027GA SA309426028GA SA309426029GA SA309426031GA SA309426032GA SA309426033GA SA309426034GA SA309426035GA SA309426036GA SA309426038GA SA309426039GR SA309426040GA SA309426041GR SA309426042GA SA309426043GA SA309426045GA SA309426046GA SA309426047GA SA309426048GA SA309426049GA SA309426051GA SA309426052GA SA309426054GA SA309426055GA SA309426056GA SA309426057GA SA309426058GA SA309426059GA SA309426060GA SA309426061GA SA309426063GA SA309426064GA SA309426069GA SA309426070GA SA309426075GA SA309426077GA SA309426079GA SA309426081GA SA309426083GA SA309426084GA SA309426085GA SA309426089GA SA309426090GA SA309426091GA SA309426093GA SA309426096GA SA309426099GA SA309426102GA SA309426103GA SA309426104GA SA309426105GA SA309426106GA SA309426107GA SA309426109GA SA309426110GA SA309426111GA SA309426113GA SA309426114GA SA309426115GA SA309426116GA SA309426117GA SA309426120GA SA309426121GR SA309426122GR SA309426124GA SA309426125GA SA309426127GA SA309426128GA SA309426129GA SA309426130GA SA309426131GA SA309426132GA SA309426133GA SA309426134GA SA309426136GA SA309426137GR SA309426139GA SA309426140GA SA309426141GA SA309426144GA SA309426146GA SA309426148GA SA309426149GA SA309426150GA SA309426152GA SA309426154GA SA309426156GA SA309426157GA SA309426158GA SA309426159GA SA309426160GA SA309426161GA SA309426162GA SA309426163GA SA309426165GA SA309426166GA SA309436167GA SA309436170GA SA309436172GA SA309436176GA SA309436178GA SA309436179GA SA309436181GA SA309436182GA SA309436183GA SA309436187GR SA309436188GR SA309436189GA SA309436190GR SA309436191GA SA309436192GR
SA309436194GA SA309436195GR SA309436196GA SA309436197GR SA309436198GA SA309436199GA SA309436200GR SA309436201GA SA309436202GR SA309436203GR SA309436204GR SA309436205GR SA309436206GA SA309436207GA SA309436208GR SA309436210GA SA309436212GA SA309436215GA SA309436217GA SA309436218GA SA309436219GA SA309436220GA SA309436223GR SA309436224GA SA309436226GA SA309436268GA SA309436269GA SA309436270GR SA309436271GA SA309436273GA SA309436274GA SA309436275GA SA309436277GA SA309436280GA SA309436282GA SA309436283GA SA309436284GA SA309436287GA SA309436288GA SA309436289GA SA309436290GA SA309436291GA SA309436293GA SA309436294GA SA309436295GA SA309436296GA SA309436297GR SA309436298GA SA309436299GA SA309436300GA SA309436301GA SA309436303GA SA309436304GA SA309436305GA SA309436306GA SA309436307GA SA309436308GA SA309436309GA SA309436310GA SA309436311GA SA309436312GA SA309436313GA SA309436314GA SA309436316GA SA309436317GA SA309436320GA SA309436323GA SA309436325GA SA309436326GA SA309436330GA SA309436332GA SA309436333GA SA309436334GA SA309436335GA SA309436336GA SA309436337GA SA309436338GA SA309436339GA SA309436340GA SA309436341GA SA309436342GA SA309436343GA SA309436344GA SA309436345GA SA309436346GA SA309436347GA SA309436349GA SA309436351GA SA309436352GA SA309436353GA SA309436354GA SA309436355GA SA309436356GA SA309436357GA SA309436358GA SA309436359GA SA309436361GA SA309436364GA SA309436365GA SA309436366GA SA309436367GA SA309436368GA SA309436369GA SA309436370GA SA309436371GA SA309436372GA SA309436373GA SA309436374GA SA309436375GA SA309446376GA SA309446377GA SA309446378GA SA309446380GA SA309446381GR SA309446382GA SA309446384GA SA309446387GA SA309446388GA SA309446391GA SA309446393GA SA309446394GA SA309446395GA SA309446396GA SA309446397GA SA309446398GA SA309446402GA SA309446403GA SA309446404GR SA309446405GA SA309446406GA SA309446407GR SA309446409GA SA309446410GA SA309446411GA SA309446413GA SA309446414GA SA309446415GA SA309446416GA SA309446417GA SA309446418GA SA309446420GA SA309446422GA SA309446425GA SA309446426GA SA309446427GA SA309446429GA SA309446430GA SA309446431GA SA309446433GA SA309446434GA SA309446435GA SA309446436GA SA309446439GA SA309446440GA SA309446441GR SA309446442GA SA309446444GA SA309446445GA SA309446446GA SA309446447GA SA309446449GA SA309446451GA SA309446452GA SA309446454GA SA309446456GA SA309446457GA SA309446458GA SA309446459GR SA309446461GA SA309446462GA SA309446464GA SA309446465GA SA309446467GA SA309446468GR SA309446469GA SA309446470GA SA309446471GA SA309446472GR SA309446473GA SA309446475GA SA309446477GA SA309446478GA SA309446479GA SA309446480GA SA309446484GR SA309446486GA SA309446487GR SA309446488GR SA309446491GA SA309446492GA SA309446496GA SA309446498GA SA309446501GA SA309446502GA SA309446503GR SA309446505GR SA309446506GA SA309446507GA SA309446512GA SA309446517GA SA309446519GA SA309446520GA SA309446521GA SA309446523GA SA309446524GA SA309446526GA SA309446527GA SA309446531GA SA309446532GR SA309446533GA SA309446534GA SA309456536GA SA309456539GA SA309456540GR SA309456541GA SA309456542GA SA309456546GR SA309456550GA SA309456551GR SA309456552GR SA309456553GR SA309456554GR SA309456555GR SA309456556GA SA309456557GA SA309456558GR SA309456559GA SA309456560GA SA309456562GA SA309456564GA SA309456599GA SA309456600GA SA309456601GA SA309456602GA SA309456603GA SA309456604GA SA309456605GA SA309456606GA SA309456607GA SA309456608GA SA309456609GA SA309456610GA SA309456611GA SA309456612GA SA309456613GA SA309456614GA SA309456616GA SA309456617GA SA309456618GA SA309456619GA SA309456620GA SA309456621GA SA309456622GA SA309456624GA SA309456625GA SA309456626GA SA309456627GA SA309456628GA SA309456629GA SA309456630GA SA309456631GA SA309456632GA SA309456636GA SA309456639GA SA309456641GA SA309456642GA SA309456644GA SA309456645GA SA309456651GA SA309456654GA SA309456658GA SA309456662GA SA309456667GA SA309456668GA SA309456669GA SA309456670GA SA309456671GA SA309456672GA SA309456673GA SA309456674GA SA309456677GA SA309456680GA SA309456681GA SA309456683GA SA309456684GA SA309456685GA SA309456687GA SA309456689GA SA309456690GA SA309456691GA SA309456692GA SA309456693GA SA309456694GA SA309456696GA SA309456698GA SA309456703GA SA309456704GA SA309456705GA SA309456706GA SA309456707GA SA309456708GA SA309456710GA SA309456714GA SA309456716GA SA309456717GA SA309456721GA SA309456723GA SA309456726GA SA309456727GA SA309456732GA SA309466741GA SA309466752GA SA309466755GA SA309466756GA SA309466768GA SA309466770GA SA309466771GA SA309466772GA SA309466791GA SA309466792GA SA309466801GA SA309466803GA SA309466804GA SA309466805GA SA309466807GA SA309466808GA SA309466809GA SA309466810GA SA309466812GA SA309466813GA SA309466814GA SA309466815GA SA309466816GA SA309466817GA SA309466818GA SA309466819GA SA309466820GA SA309466821GA SA309466822GA SA309466823GA SA309466824GA SA309466825GA SA309466826GA SA309466827GA SA309466828GA SA309466829GA SA309466830GA SA309466831GA SA309466832GA SA309466833GA SA309466834GA SA309466835GA SA309466836GA SA309466837GA SA309466840GA SA309466841GR SA309466842GA SA309466843GA SA309466844GA SA309466845GR SA309466846GA SA309466847GA SA309466849GA SA309466850GA SA309466851GA SA309466852GA SA309466853GA SA309466854GA SA309466856GA SA309466857GA SA309466858GA SA309466859GA SA309466860GA SA309476862GA SA309476868GA SA309476872GA SA309476874GR SA309476875GA SA309476878GA SA309476879GA SA309476885GR SA309476886GA SA309476893GR SA309476894GA SA309476897GA SA309476901GA SA309476902GA SA309476903GA SA309476904GA SA309476905GA SA309476906GA SA309476907GA SA309476908GA SA309476909GA SA309476910GA SA309476912GA SA309476913GA SA309476914GA SA309476916GR SA309476918GA SA309476919GA SA309476920GA SA309476921GA SA309476923GA SA309476924GA SA309476928GA SA309476929GA SA309476930GA SA309476932GA SA309476934GA SA309476939GR SA309476940GA SA309476941GR SA309476943GR SA309476947GA SA309476948GA SA309476949GA SA309476953GA SA309476954GA SA309476955GA SA309476957GR SA309476958GA SA309476962GR SA309476968GA SA309476970GR SA309476972GR SA309476973GA SA309476974GA SA309476975GA SA309476976GR SA309476978GR SA309476990GA SA309476997GA SA309476998GA SA309477000GA SA309477001GA SA309477002GA SA309477004GA SA309477005GA SA309477006GR SA309477007GR SA309477009GR SA309477010GA SA309477011GR SA309477012GR SA309477013GA SA309477015GR SA309477016GA SA309477017GR SA309477018GA SA309477020GA SA309477022GR SA309477025GA SA309477026GA SA309477027GA SA309477028GR SA309477029GR SA309477030GA SA309477031GA SA309477034GA SA309477035GA SA309477036GA SA309477037GA SA309477039GA SA309477040GA SA309477041GA SA309477042GA SA309477043GA SA309477044GA SA309477046GA SA309477047GA SA309477048GA SA309477049GA SA309477051GA SA309477052GA SA309477053GA SA309477054GA SA309477056GA SA309477057GA SA309477058GA SA309477059GA SA309477060GA SA309477061GA SA309477062GA SA309477065GA SA309477066GA SA309477067GA SA309477069GA SA309477070GA SA309477071GA SA309477072GA SA309477075GA SA309477076GA SA309477078GA SA309477079GA SA309477080GA SA309477081GA SA309477082GA SA309477083GA SA309477084GA SA309477086GA SA309477087GA SA309477088GA SA309477089GA SA309477090GA SA309477094GR SA309477095GA SA309477096GA SA309477097GA SA309477099GA SA309477100GA SA309477101GA SA309477102GA SA309477104GA SA309477106GA SA309477107GA SA309477109GA SA309477110GA SA309477111GA SA309477112GA SA309477113GA SA309477114GA SA309477115GA SA309477117GA SA309477119GA SA309477120GA SA309477121GA SA309477122GA SA309477123GA SA309477124GA SA309477125GA SA309477126GA SA309477127GA SA309477128GA SA309477129GA SA309477130GA SA309477132GA SA309477134GA SA309477135GA SA309477138GA SA309477139GA SA309477140GA SA309477141GA SA309477142GA SA309497143GA SA309497144GA SA309497145GA SA309497147GA SA309497148GA SA309497149GR SA309497150GA SA309497151GA SA309497153GA SA309497154GA SA309497156GA SA309497157GA SA309497158GA SA309497159GA SA309497160GA SA309497161GA SA309497163GA SA309497164GA SA309497165GA SA309497166GA SA309497167GA SA309497168GA SA309497169GA SA309497170GA SA309497172GA SA309497173GA SA309497174GA SA309497175GR SA309497176GA SA309497177GA SA309497278GR SA309507312GA SA309507314GA SA309507343GA SA309507348GA SA309507358GR SA309507360GR SA309507370GA SA309507371GR SA309507382GA SA309507383GR SA309507385GR SA309507396GR SA309507407GA SA309507417GR SA309507420GA SA309507432GA SA309507434GA SA309507435GR SA309507442GA SA309507445GR SA309517452GA SA309517454GA SA309517456GA SA309517458GR SA309517460GA SA309517464GA SA309517465GA SA309517467GA SA309517469GA SA309517470GA SA309517471GA SA309517472GR SA309517473GA SA309517474GA SA309517476GA SA309517477GA SA309517478GA SA309517480GA SA309517481GR SA309517483GA SA309517484GR SA309517487GA SA309517488GA SA309517489GA SA309517490GA SA309517491GA SA309517493GA SA309517494GA SA309517495GA SA309517496GA SA309517497GR SA309517499GR SA309517500GR SA309517503GA SA309517504GR SA309517505GA SA309517506GA SA309517507GR SA309517508GR SA309517511GA SA309517513GA SA309517514GA SA309517515GR SA309517516GA SA309517517GA SA309517518GA SA309517519GR SA309517520GA SA309517521GR SA309517523GA SA309517525GR SA309517527GA SA309517528GA SA309517529GA SA309517531GA SA309517533GR SA309517534GA SA309517535GA SA309517537GR SA309517538GA SA309517540GR SA309517541GA SA309517543GA SA309517544GA SA309517545GA SA309517546GA SA309517547GA SA309517549GA SA309517550GA SA309517552GA SA309517553GA SA309517554GA SA309517555GA SA309517556GA SA309517557GA SA309517558GA SA309517559GR SA309517560GA SA309517561GA SA309517562GA SA309517563GA SA309517564GA SA309517565GA SA309517566GA SA309517567GR SA309517568GA SA309517569GA SA309517570GA SA309517571GA SA309517572GR SA309517573GA SA309517574GA SA309517575GA SA309517576GA SA309517577GA SA309517578GA SA309517579GA SA309517580GA SA309517581GA SA309517582GA SA309517583GA SA309517584GA SA309517585GA SA309517586GA SA309517587GA SA309517588GA SA309517589GA SA309517590GA SA309517591GA SA309517592GA SA309517593GA SA309517594GA SA309517595GA SA309517597GA SA309517598GA SA309517599GA SA309517600GA SA309517601GA SA309517602GA SA309517603GA SA309517604GA SA309517605GA SA309517606GA SA309517607GA SA309517608GA SA309517609GA SA309517610GA SA309517611GA SA309517613GA SA309517616GA SA309517617GA SA309517618GA SA309517619GA SA309517620GA SA309517621GA SA309517622GA SA309517623GA SA309517624GA SA309517625GA SA309517626GR SA309517628GA SA309517629GR SA309517630GA SA309517632GA SA309517635GA SA309517637GA SA309517638GA SA309517639GR SA309517640GA SA309517644GA SA309517646GA SA309517647GA SA309517648GA SA309517649GA SA309517650GA SA309517651GA SA309517652GA SA309517653GA SA309517656GR SA309517657GR SA309517658GA SA309517659GA SA309517660GA SA309517661GR SA309517662GA SA309517663GA SA309517664GA SA309517666GA SA309517668GA SA309517669GA SA309517671GA SA309517672GA SA309517673GA SA309517674GA SA309517675GR SA309517676GA SA309517678GA SA309517679GR SA309517681GA SA309517684GA SA309517686GA SA309517687GA SA309517688GA SA309517689GA SA309517690GA SA309517691GA SA309517692GR SA309517694GA SA309517698GA SA309517700GA SA309517701GR SA309517703GA SA309517704GA SA309517706GR SA309517707GR SA309517709GA SA309517711GR SA309517713GR SA309517715GR SA309517718GR SA309517722GR SA309517727GR SA309517728GR SA309517731GA SA309517733GR
SA309517734GR SA309517735GR SA309517736GA SA309517737GR SA309517763GA SA309517765GA SA309517772GR SA309517773GR SA309517775GR SA309517776GA SA309517777GR SA309517778GR SA309517780GR SA309517786GR SA309517787GA SA309517788GR SA309517789GA SA309517793GR SA309517794GR SA309517795GA SA309517796GR SA309517797GR SA309517798GR SA309517799GR SA309517800GA SA309517803GR SA309517804GA SA309517805GR SA309517807GR SA309517809GR SA309517810GR SA309517811GR SA309517813GR SA309517814GA SA309517815GR SA309517816GR SA309517817GR SA309517819GA SA309517820GA SA309517821GR SA309517822GA SA309517823GR SA309517824GA SA309517826GA SA309517830GR SA309517831GA SA309517832GR SA309517834GR SA309517837GA SA309517838GA SA309517842GA SA309517847GA SA309517848GA SA309517849GA SA309517851GA SA309517852GR SA309517853GR SA309517854GA SA309517855GA SA309517859GA SA309517862GA SA309517865GA SA309517867GA SA309517869GA SA309517871GA SA309517873GR SA309517875GA SA309517877GA SA309517878GA SA309517880GA SA309517882GR SA309517883GA SA309517885GA SA309517890GA SA309517891GR SA309517892GA SA309517894GR SA309517898GA SA309517899GA SA309517900GA SA309517901GA SA309517902GA SA309517903GA SA309517904GR SA309517905GR SA309517906GA SA309527910GR SA309527911GA SA309527912GA SA309527913GR SA309527914GA SA309527916GA SA309527918GR SA309527920GR SA309527921GR SA309527922GR SA309527923GR SA309527927GR SA309527930GR SA309527933GR SA309527934GR SA309527937GR SA309527939GA SA309527941GR SA309527942GR SA309527943GA SA309527945GR SA309527946GR SA309527948GA SA309527955GA SA309527956GA SA309527957GA SA309527961GR SA309527962GA SA309527965GR SA309527969GR SA309527986GR SA309527987GA SA309527992GA SA309527994GA SA309527995GA SA309527996GA SA309527997GA SA309527998GA SA309527999GA SA309528000GA SA309528002GA SA309528003GA SA309528004GA SA309528008GA SA309528010GR SA309528011GA SA309528012GA SA309528013GR SA309528014GA SA309528015GA SA309538026GA SA309538029GA SA309538031GA SA309538034GA SA309538043GA SA309538045GA SA309538046GA SA309538048GA SA309538049GA SA309538050GA SA309538052GA SA309538056GA SA309538058GA SA309538059GA SA309538060GA SA309538061GA SA309538062GA SA309538063GA SA309538070GA SA309538079GA SA309538080GA SA309538081GA SA309538084GA SA309538085GA SA309538087GA SA309538090GA SA309538092GA SA309538093GA SA309538094GA SA309538095GA SA309538096GA SA309538098GA SA309538103GA SA309538104GA SA309538106GA SA310028112GA SA310028114GR SA310028115GA SA310028120GA SA310028122GA SA310028126GA SA310028127GA SA310028129GA SA310028130GA SA310028132GA SA310028134GA SA310028135GA SA310028138GA SA310028140GR SA310028142GA SA310028145GA SA310028148GA SA310028149GR SA310028154GA SA310028155GR SA310028171GA SA310028180GA SA310028181GA SA310028182GR SA310028184GA SA310028186GR SA310028187GA SA310028189GA SA310028191GR SA310028195GA SA310028196GA SA310028197GA SA310028199GR SA310028201GA SA310028203GA SA310028211GA SA310028213GR SA310028214GR SA310028217GA SA310028218GA SA310028221GA SA310028222GA SA310028223GR SA310028252GA SA310028262GA SA310028265GR SA310028266GR SA310028268GA SA310028269GR SA310028272GA SA310028273GA SA310028274GR SA310028275GR SA310028280GR SA310028284GA SA310028290GA SA310028291GA SA310028294GA SA310028300GA SA310028306GA SA310028307GA SA310028315GA SA310028317GA SA310028319GA SA310028326GA SA310028331GA SA310028337GA SA310038343GA SA310038348GA SA310038352GR SA310038359GA SA310038380GA SA310038386GA SA310038391GA SA310038395GA SA310038398GR SA310038408GA SA310038427GR SA310038430GA SA310038440GA SA310038441GA SA310038444GA SA310038449GA SA310038462GA SA310038476GA SA310038483GA SA310038487GA SA310038492GA SA310038497GA SA310038503GA SA310038507GA SA310038510GA SA310038513GA SA310038514GA SA310038520GA SA310038524GA SA310038525GA SA310038527GA SA310038530GA SA310038534GA SA310038537GA SA310048541GA SA310048542GA SA310048553GA SA310048558GA SA310048562GA SA310048564GA SA310048567GA SA310048579GA SA310048592GR SA310048604GA SA310048613GA SA310048616GA SA310048619GA SA310048620GA SA310048624GR SA310048633GA SA310048634GA SA310048640GA SA310048641GA SA310048643GA SA310048647GA SA310048649GA SA310048658GA SA310048660GA SA310048663GA SA310048664GA SA310048665GA SA310048666GA SA310048667GA SA310048668GA SA310048670GA SA310048673GA SA310048676GA SA310048679GA SA310048680GA SA310048681GA SA310048683GA SA310048684GA SA310048686GA SA310048687GA SA310058690GA SA310058691GA SA310058692GA SA310058694GA SA310058696GA SA310058697GA SA310058698GA SA310058700GA SA310058701GA SA310058704GA SA310058705GA SA310058707GA SA310058708GA SA310058709GA SA310058716GA SA310058717GA SA310058720GA SA310058721GA SA310058722GA SA310058725GA SA310058726GA SA310058730GA SA310058733GA SA310058735GR SA310058737GA SA310058739GR SA310058743GA SA310058745GA SA310058749GA SA310058750GA SA310058752GA SA310058753GA SA310058756GA SA310058758GA SA310058759GA SA310058760GA SA310058761GA SA310058762GA SA310058764GR SA310058765GA SA310068797GA SA310068798GA SA310068799GR SA310068800GA SA310068801GA SA310068803GA SA310068804GA SA310068806GA SA310068809GA SA310068810GA SA310068812GA SA310068814GA SA310068816GA SA310068818GA SA310068819GA SA310068823GA SA310068824GA SA310068833GA SA310068838GA SA310068839GA SA310068844GA SA310068848GA SA310068852GA SA310068853GA SA310068858GA SA310068868GA SA310068870GA SA310068875GR SA310068880GA SA310068884GA SA310068886GA SA310068887GA SA310068892GA SA310068893GA SA310068895GA SA310068898GA SA310068899GA SA310068902GA SA310068914GA SA310068915GA SA310068919GA SA310068921GA SA310068922GA SA310068925GA SA310068926GA SA310068927GA SA310068929GA SA310068938GA SA310068941GA SA310068943GA SA310068947GA SA310118960GA SA310118961GA SA310118962GR SA310118963GA SA310118964GA SA310118966GA SA310118967GA SA310118968GA SA310118970GA SA310118971GA SA310118972GA SA310118973GA SA310118974GA SA310118976GA SA310118977GA SA310118978GA SA310118979GA SA310118982GA SA310118983GA SA310118984GA SA310118985GA SA310118986GA SA310118988GA SA310118989GA SA310118990GA SA310118991GA SA310118993GA SA310118995GA SA310118996GA SA310118997GA SA310118998GA SA310118999GA SA310119001GA SA310119002GA SA310119003GA SA310119006GA SA310119007GA SA310119013GA SA310119014GA SA310119017GA SA310139041GA SA310139044GA SA310139045GA SA310139046GA SA310139047GA SA310139051GR SA310139052GA SA310139054GA SA310139055GA SA310139056GA SA310139058GA SA310139059GA SA310139061GR SA310139062GA SA310139063GR SA310139065GA SA310139066GR SA310139068GA SA310139069GA SA310139070GA SA310139071GA SA310139073GA SA310139074GA SA310139075GA SA310139076GA SA310139077GA SA310139080GA SA310139082GA SA310139083GA SA310139085GR SA310139086GA SA310139087GA SA310139088GA SA310139089GA SA310139092GA SA310139093GA SA310139094GA SA310139095GA SA310139096GA SA310139098GA SA310139104GA SA310139106GA SA310139107GA SA310139114GA SA310139115GA SA310139119GA SA310139120GA SA310139121GA SA310139122GA SA310139124GR SA310139126GA SA310139127GA SA310139132GR SA310139133GA SA310139134GA SA310139135GA SA310139141GA SA310139142GA SA310139143GA SA310139144GA SA310139145GA SA310139146GA SA310139148GA SA310139149GA SA310139150GA SA310139151GA SA310139152GA SA310139153GA SA310139154GA SA310139155GA SA310139156GR SA310139157GA SA310139158GA SA310139159GA SA310139160GA SA310139161GA SA310139162GA SA310139163GA SA310139164GA SA310139165GA SA310139166GA SA310139167GA SA310139168GA SA310139172GA SA310139173GA SA310139174GA SA310139175GA SA310139177GA SA310139178GA SA310139179GA SA310139180GA SA310139183GA SA310139184GA SA310139185GA SA310139186GA SA310139188GA SA310139190GA SA310139191GA SA310139193GA SA310139194GA SA310139198GA SA310139200GA SA310139201GA SA310139202GA SA310139203GA SA310139205GA SA310139206GA SA310139207GA SA310139209GA SA310139210GA SA310139212GA SA310139213GA SA310139215GA SA310139218GA SA310139219GA SA310139222GR SA310139231GA SA310139234GA SA310139244GR SA310139245GA SA310139248GA SA310139249GA SA310139250GA SA310139251GA SA310139254GA SA310139255GA SA310139256GA SA310139257GA SA310139259GA SA310139260GA SA310139262GA SA310139263GA SA310139264GA SA310139265GA SA310139266GA SA310139267GA SA310139268GA SA310139269GA SA310139270GA SA310139271GA SA310139272GA SA310139273GA SA310139274GA SA310139275GA SA310139276GA SA310139277GA SA310139278GA SA310139280GA SA310139281GA SA310139282GA SA310139283GA SA310139284GA SA310139285GA SA310139286GA SA310139287GA SA310139288GA SA310139290GA SA310139291GA SA310139293GA SA310139294GA SA310139296GA SA310139298GA SA310139303GA SA310139305GA SA310139306GA SA310139308GA SA310149309GA SA310149316GA SA310149319GA SA310149321GA SA310149328GA SA310149333GA SA310149334GA SA310149336GA SA310149339GA SA310149340GA SA310149341GA SA310149342GA SA310149344GA SA310149345GA SA310149346GA SA310149348GA SA310149349GA SA310149350GA SA310149351GA SA310149352GA SA310149354GA SA310149355GR SA310149356GA SA310149357GA SA310159359GA SA310159367GA SA310159368GA SA310159373GA SA310159374GA SA310159375GA SA310159378GA SA310159380GA SA310159383GA SA310159386GR SA310159423GA SA310159424GA SA310159427GA SA310159429GA SA310159430GA SA310159431GA SA310159433GA SA310159435GA SA310159436GA SA310159437GA SA310159438GA SA310159439GA SA310159440GA SA310159441GA SA310159449GA SA310159450GA SA310159454GR SA310159455GA SA310159476GA SA310169480GA SA310169481GA SA310169483GA SA310169485GA SA310169491GA SA310169498GA SA310169499GA SA310169500GR SA310169501GR SA310169502GA SA310169503GA SA310169505GA SA310169506GR SA310169507GR SA310169523GA SA310169526GA SA310169529GR SA310169532GA SA310199562GA SA310199564GR SA310199565GR SA310199566GA SA310199567GR SA310199570GA SA310199572GA SA310199573GA SA310199576GA SA310199578GA SA310199579GA SA310199580GR SA310199581GA SA310199583GA SA310199587GA SA310199588GA SA310199589GA SA310199590GA SA310199591GA SA310199592GA SA310199593GA SA310199594GA SA310199595GA SA310199596GA SA310199597GA SA310199598GA SA310199599GA SA310199600GA SA310199601GA SA310199602GA SA310199603GA SA310199604GA SA310199605GA SA310199606GA SA310199607GA SA310199608GA SA310199609GA SA310199610GA SA310199612GA SA310199613GA SA310199614GA SA310199615GA SA310199616GA SA310199617GA SA310199618GA SA310199619GA SA310199620GA SA310199621GA SA310199622GA SA310199623GA SA310199626GA SA310199627GA SA310199640GA SA310199642GA SA310199643GA SA310199644GA SA310199645GR SA310199646GA SA310199647GR SA310199649GA SA310199650GA SA310199651GA SA310199652GA SA310199653GA SA310199654GA SA310199655GA SA310199656GA SA310199657GA SA310199658GA SA310199660GA SA310199663GR SA310199664GA SA310199667GA SA310199668GA SA310199669GR SA310199670GA SA310199672GA SA310199673GA SA310199674GA SA310199675GA SA310199676GA SA310199677GA SA310199678GA SA310199679GA SA310199680GA SA310199683GA SA310199686GA SA310199691GR SA310199702GA SA310199703GA SA310199704GA SA310199713GA SA310199715GA SA310199716GA SA310199717GA SA310199719GA SA310199723GA SA310199727GA SA310199730GA SA310199735GA SA310199737GA SA310199738GA SA310199739GA SA310199741GA SA310199745GA SA310199752GA SA310199753GA SA310199757GR SA310199759GR SA310219782GA SA310219783GR SA310219785GA SA310219787GA SA310219788GA SA310219790GA SA310219791GA SA310219792GA SA310219793GR SA310219794GA SA310219797GA
SA310219798GA SA310219799GR SA310219800GR SA310219801GA SA310219802GA SA310219803GA SA310219804GA SA310219805GA SA310219807GA SA310219808GA SA310219809GA SA310219810GA SA310219811GA SA310219812GA SA310219813GA SA310219814GA SA310219815GA SA310219816GA SA310219817GA SA310219818GA SA310219819GA SA310219820GA SA310219821GR SA310219822GA SA310219823GA SA310219825GA SA310219826GA SA310219827GA SA310219828GA SA310219830GA SA310219831GR SA310219832GA SA310219833GR SA310219834GA SA310219835GA SA310219836GA SA310219837GR SA310219838GR SA310219839GR SA310219840GA SA310219841GA SA310219843GA SA310219844GA SA310219845GA SA310219846GA SA310219847GA SA310219854GA SA310259855GA SA310259856GA SA310259857GA SA310259858GA SA310259859GA SA310259860GR SA310259861GR SA310259862GA SA310259863GA SA310259864GA SA310259865GA SA310259866GA SA310259867GA SA310259868GA SA310259869GR SA310259870GR SA310259871GA SA310259872GA SA310259873GA SA310259874GA SA310259875GA SA310259876GA SA310259877GA SA310259878GA SA310259879GA SA310259880GA SA310259881GA SA310259882GA SA310259883GA SA310259884GR SA310259885GA SA310259886GA SA310259887GA SA310259888GA SA310259889GA SA310259890GA SA310259891GA SA310259892GA SA310259893GA SA310259894GA SA310259911GA SA310259912GA SA310259913GA SA310259914GA SA310259916GA SA310259920GA SA310259923GA SA310259928GA SA310259929GA SA310259932GR SA310259944GA SA310259951GR SA310259960GA SA310259964GA SA310259971GA SA310259972GR SA310259973GA SA310259974GA SA310259975GA SA310259976GA SA310259977GA SA310259979GA SA310259980GA SA310259981GA SA310259982GA SA310259985GA SA310259987GA SA310259988GA SA310259989GA SA310260001GA SA310260003GA SA310260004GA SA310260006GA SA310260007GA SA310260008GA SA310260009GA SA310260010GA SA310260011GA SA310260012GA SA310260013GR SA310260014GA SA310260015GA SA310260016GA SA310260017GA SA310260018GA SA310260019GA SA310260020GA SA310260021GA SA310260022GA SA310260023GA SA310260024GA SA310260025GA SA310260026GA SA310260027GA SA310260028GA SA310260030GA SA310260031GA SA310260032GA SA310260033GA SA310260034GA SA310260035GA SA310260036GA SA310260038GA SA310260039GA SA310260040GA SA310260041GA SA310260042GA SA310260043GA SA310260044GA SA310260046GA SA310260047GA SA310260049GA SA310260050GA SA310260051GA SA310260053GA SA310260054GA SA310260055GA SA310260056GA SA310260057GA SA310260059GA SA310260060GA SA310260061GA SA310260062GA SA310260063GA SA310260064GA SA310260065GA SA310260066GA SA310260067GA SA310260068GA SA310260069GA SA310260071GA SA310260072GA SA310260073GA SA310260074GA SA310260075GA SA310260078GA SA310260079GA SA310260080GA SA310260081GA SA310260082GA SA310260214GA SA310260215GA SA310260216GA SA310260217GA SA310260218GA SA310260219GA SA310260221GA SA310260224GA SA310260225GA SA310260226GA SA310260229GA SA310260230GA SA310260231GA SA310260232GA SA310260233GA SA310260236GA SA310260237GA SA310260239GA SA310260240GA SA310260241GA SA310260242GA SA310260243GA SA310260244GA SA310260245GA SA310260246GA SA310260247GA SA310260248GA SA310260249GA SA310260250GA SA310260252GA SA310260253GA SA310260254GA SA310260255GA SA310260257GA SA310260297GA SA310260303GR SA310260305GA SA310260307GA SA310260308GR SA310260309GA SA310260310GA SA310260311GA SA310260312GA SA310260313GA SA310260314GA SA310260320GA SA310260322GA SA310260324GA SA310260325GA SA310260326GA SA310260327GA SA310260328GA SA310260330GA SA310260331GA SA310260332GA SA310260334GA SA310260335GA SA310260336GA SA310260337GA SA310260339GA SA310260341GA SA310260342GA SA310260343GA SA310260346GA SA310260347GA SA310260348GA SA310260350GA SA310260351GA SA310260352GA SA310260353GA SA310260354GA SA310260355GA SA310260359GR SA310260360GA SA310260362GA SA310260365GA SA310260367GR SA310260372GA SA310260373GA SA310260374GA SA310260375GA SA310260376GA SA310260377GA SA310260379GA SA310260381GR SA310260382GA SA310260383GA SA310260384GA SA310260385GA SA310260387GA SA310260389GA SA310260390GR SA310260392GA SA310260393GR SA310260395GA SA310260396GA SA310260397GA SA310260398GR SA310260399GA SA310260401GA SA310260402GA SA310260404GA SA310260405GA SA310260406GA SA310260407GA SA310260408GA SA310260409GA SA310260411GA SA310260412GA SA310260414GA SA310260415GR SA310260417GA SA310260420GA SA310260421GA SA310260422GA SA310260423GA SA310260424GA SA310260428GA SA310260429GA SA310260430GA SA310260434GA SA310260436GA SA310260437GA SA310260438GA SA310260439GA SA310260469GA SA310260470GA SA310260471GA SA310260472GA SA310260473GA SA310260474GA SA310260479GA SA310260480GA SA310260482GA SA310260486GA SA310260488GA SA310260490GA SA310260491GA SA310260493GA SA310260494GA SA310260495GA SA310260496GA SA310260499GA SA310260521GA SA310260522GR SA310260523GA SA310260524GA SA310260526GR SA310260528GA SA310260529GA SA310260531GA SA310260532GA SA310260533GA SA310260534GA SA310260536GA SA310260539GA SA310260541GA SA310260542GA SA310260543GA SA310260544GA SA310260546GA SA310260547GA SA310260553GR SA310260554GA SA310270582GA SA310270584GA SA310270585GA SA310270586GA SA310270587GA SA310270588GA SA310290619GA SA310290622GA SA310290649GA SA310290650GA SA310290653GA SA310290655GA SA310290657GA SA310290658GA SA310290659GA SA310290660GA SA310290661GA SA310290663GR SA310290668GR SA310290686GA SA310290699GA SA310290704GA SA310290705GA SA310290706GA SA310290707GA SA310290708GA SA310290709GA SA310290710GA SA310290711GA SA310290712GA SA310290713GA SA310290714GA SA310290715GA SA310290716GA SA310290717GA SA310290719GA SA310290720GA SA310290721GA SA310290722GA SA310290723GA SA310290724GA SA310290725GA SA310290726GA SA310290727GA SA310290728GA SA310290729GA SA310290730GA SA310290732GA SA310290733GA SA310290735GR SA310290739GA SA310290745GA SA310290746GA SA310290747GA SA310290748GA SA310290749GA SA310290750GA SA310290752GA SA310290753GA SA310290754GA SA310290755GA SA310290756GA SA310290757GA SA310290758GA SA310290759GR SA310290760GA SA310300763GA SA310300766GA SA310300768GA SA310300769GA SA310300770GA SA310300771GA SA310300773GA SA310300774GA SA310300776GA SA310300777GA SA310300778GA SA310300779GA SA310300780GA SA310300781GA SA310300783GA SA310300784GA SA310300785GA SA310300786GA SA310300788GA SA310300789GA SA310300790GA SA310300791GA SA310300792GA SA310300793GA SA310300794GA SA310300795GA SA310300796GA SA310300797GA SA310300798GA SA310300799GA SA310300800GA SA310300801GA SA310300802GA SA310300803GA SA310300804GA SA310300805GA SA310300806GA SA310310840GA SA310310845GA SA310310848GA SA310310852GA SA310310854GA SA310310857GA SA310310859GA SA310310861GA SA310310862GA SA310310864GA SA310310865GA SA310310866GA SA310310867GA SA310310868GA SA310310871GA SA310310872GA SA310310875GA SA310310876GA SA310310883GR SA310310885GA SA310310890GR SA310310895GA SA310310898GA SA310310901GA SA310310902GA SA310310903GR SA310310904GA SA310310907GA SA310310908GA SA310310916GA SA310310918GA SA310310920GA SA310310921GA SA310310922GA SA310310924GA SA310310925GA SA310310926GA SA310310952GR SA310310956GA SA310310973GR SA310310974GA SA310310978GA SA310310993GA SA310321001GA SA310321007GA SA310321014GA SA310321019GA SA310321020GA SA310321024GA SA310321040GA SA310321041GA SA310321042GR SA310321046GA SA310321047GA SA310321048GA SA310321049GA SA310321055GA SA310321056GA SA310321057GA SA310321061GA SA310321062GA SA310321063GA SA310321066GA SA310321068GA SA310321069GA SA310321070GA SA310321072GA SA310321073GA SA310331079GR SA310331080GA SA310331081GA SA310331092GA SA310331098GA SA310331104GR SA310331109GA SA310331115GA SA310331128GA SA310331136GA SA310331138GA SA310331139GA SA310331145GA SA310331149GA SA310331154GA SA310331159GA SA310331160GA SA310331162GA SA310331217GA SA310331218GA SA310331219GA SA310331220GA SA310331221GA SA310331222GA SA310331223GA SA310331224GA SA310331228GA SA310331230GA SA310331231GA SA310331232GA SA310331234GA SA310331238GA SA310331242GA SA310331243GA SA310331245GA SA310331248GA SA310331249GR SA310331251GA SA310331253GA SA310331255GA SA310331256GA SA310331259GA SA310331261GA SA310331264GA SA310331265GA SA310331266GA SA310331270GA SA310331271GA SA310331272GA SA310331273GA SA310331274GA SA310331276GA SA310331277GA SA310331278GA SA310331279GA SA310331281GA SA310331283GA SA310331284GA SA310331288GA SA310331289GA SA310331292GA SA310331293GA SA310331294GA SA310331295GA SA310331296GA SA310331297GA SA310331298GA SA310331300GA SA310331301GA SA310331304GA SA310331305GA SA310331313GA SA310331316GA SA310331322GA SA310331323GA SA310331326GA SA310331327GA SA310331329GA SA310331331GA SA310331332GA SA310331334GR SA310331335GA SA310331337GA SA310331341GA SA310331342GA SA310331343GA SA310331345GA SA310331347GA SA310331348GA SA310331349GA SA310331350GA SA310331351GA SA310331352GA SA310331356GA SA310341357GA SA310341358GA SA310341360GA SA310341362GA SA310341363GA SA310341364GA SA310341365GA SA310341366GA SA310341367GA SA310341368GA SA310341369GA SA310341370GA SA310341371GR SA310341372GA SA310341373GA SA310341374GA SA310341376GA SA310341377GA SA310341379GA SA310341380GA SA310341381GA SA310341382GA SA310341385GA SA310341386GA SA310341387GA SA310341388GA SA310341391GA SA310341398GA SA310341399GA SA310341400GA SA310341401GA SA310341402GA SA310341403GA SA310341404GA SA310341406GA SA310341408GA SA310341410GA SA310341411GA SA310341412GA SA310341415GA SA310341417GA SA310341419GA SA310341422GA SA310341424GA SA310341425GA SA310341426GA SA310341427GA SA310341428GA SA310341429GA SA310341430GA SA310341431GA SA310341432GA SA310341434GA SA310341435GA SA310341436GA SA310341451GA SA310341458GA SA310351467GA SA310351468GA SA310351469GA SA310351471GA SA310351472GA SA310351473GA SA310351474GA SA310351475GA SA310351476GA SA310351477GA SA310351478GA SA310351479GA SA310351480GA SA310351481GA SA310351482GA SA310351483GA SA310351484GA SA310351485GA SA310351486GA SA310351487GA SA310351490GA SA310351491GA SA310351492GA SA310351493GA SA310351494GA SA310351495GA SA310351496GA SA310351497GA SA310351499GA SA310351500GA SA310351501GA SA310351502GA SA310351503GA SA310351504GA SA310351505GA SA310351506GA SA310351507GA SA310351508GA SA310351509GA SA310351510GA SA310351513GA SA310351514GA SA310351516GA SA310351517GA SA310351518GA SA310351519GA SA310351520GA SA310351521GA SA310351523GA SA310351524GA SA310351525GA SA310351526GA SA310351529GA SA310351530GA SA310351531GA SA310351532GA SA310351533GA SA310351534GA SA310351535GA SA310351537GA SA310351538GA SA310351541GA SA310351542GA SA310351543GR SA310351545GA SA310351546GA SA310351547GA SA310351548GR SA310351550GA SA310351551GR SA310351552GR SA310351554GR SA310351555GA SA310351556GA SA310351558GA SA310351559GA SA310351560GA SA310351561GA SA310351562GA SA310351563GA SA310351564GA SA310351565GA SA310351566GA SA310351574GA SA310351584GA SA310351587GA SA310351588GA SA310361589GA SA310361590GA SA310361591GA SA310361592GA SA310361593GA SA310361594GA SA310361595GA SA310361596GA SA310361597GA SA310361598GA SA310361599GA SA310361601GA SA310361602GA SA310361603GA SA310361604GA SA310361605GA SA310361606GA SA310361607GA SA310361610GA SA310361611GA SA310361612GA SA310361613GA SA310361614GA SA310361615GA SA310361616GA SA310361617GA SA310361618GA SA310361619GA SA310361620GA SA310361621GA SA310361622GA SA310361623GA SA310361624GA SA310361625GA SA310361626GA SA310361627GA SA310361630GA SA310361633GA SA310361634GA SA310361637GA
SA310361639GA SA310361640GA SA310361641GA SA310361643GA SA310361644GA SA310361647GA SA310361648GA SA310361649GA SA310361650GA SA310361651GR SA310361652GA SA310361653GA SA310361654GA SA310361655GA SA310361659GA SA310361666GA SA310361667GA SA310361668GA SA310361669GA SA310361671GA SA310361672GA SA310361673GA SA310361674GA SA310361678GA SA310361679GA SA310361680GA SA310361681GA SA310361684GA SA310361688GA SA310361690GA SA310361691GA SA310361692GA SA310361693GA SA310361694GA SA310361695GA SA310361696GA SA310361697GA SA310361698GA SA310361699GA SA310361700GA SA310361701GA SA310361702GA SA310361703GA SA310361704GA SA310361705GA SA310361706GA SA310361707GA SA310361708GA SA310361709GA SA310361710GA SA310361741GA SA310361742GA SA310361743GA SA310361745GA SA310361746GA SA310361747GA SA310361748GA SA310361749GA SA310361750GA SA310361751GA SA310361752GA SA310361753GA SA310361755GA SA310361756GA SA310361757GA SA310361758GA SA310361759GA SA310361760GA SA310361761GA SA310361762GA SA310361763GA SA310361764GA SA310361765GA SA310361766GA SA310361768GA SA310361770GA SA310361771GR SA310361772GR SA310361773GR SA310361774GR SA310361776GA SA310361777GR SA310361778GR SA310361779GR SA310361780GA SA310361781GA SA310361783GR SA310361784GR SA310361785GA SA310361787GR SA310361788GR SA310361789GR SA310361791GR SA310361792GA SA310361794GA SA310361795GA SA310361796GR SA310361797GA SA310361798GA SA310361799GA SA310361800GA SA310361801GA SA310361802GR SA310361803GA SA310361804GA SA310361805GA SA310361807GA SA310361808GA SA310361809GR SA310361810GA SA310361811GR SA310361812GA SA310361814GA SA310361815GA SA310361817GA SA310361820GA SA310361821GA SA310361822GA SA310361823GA SA310361824GA SA310361825GA SA310361826GA SA310361828GA SA310361829GA SA310361830GA SA310361831GA SA310361832GA SA310361833GA SA310361834GA SA310361835GA SA310361838GA SA310361839GA SA310361840GA SA310361841GA SA310361843GA SA310361844GA SA310361850GA SA310361852GA SA310361853GA SA310361857GA SA310361859GA SA310361863GA SA310361864GA SA310361872GR SA310361873GA SA310361874GR SA310361876GA SA310361877GA SA310361880GR SA310361886GA SA310361887GA SA310361890GA SA310361891GA SA310361892GA SA310361893GA SA310361894GA SA310361895GA SA310361897GA SA310361898GA SA310361899GA SA310361900GA SA310361901GA SA310361902GA SA310361906GA SA310361907GA SA310361912GA SA310361914GA SA310361915GA SA310361916GA SA310361918GA SA310361919GA SA310361920GA SA310361921GA SA310361922GA SA310361923GA SA310361924GA SA310361925GA SA310361926GA SA310361927GA SA310361928GA SA310361929GA SA310361930GA SA310361932GA SA310361933GA SA310361934GA SA310361935GA SA310361936GA SA310361937GA SA310361938GA SA310361939GA SA310361945GA SA310361946GA SA310361948GA SA310361949GA SA310361952GA SA310361954GA SA310361955GA SA310361956GA SA310361957GA SA310361958GA SA310361961GA SA310371965GA SA310371966GA SA310371967GA SA310371968GA SA310371969GA SA310371970GA SA310371971GA SA310371972GA SA310371973GA SA310371974GA SA310371975GA SA310371976GA SA310371977GA SA310371978GA SA310371979GA SA310371980GA SA310371981GA SA310371982GA SA310371983GA SA310371984GA SA310371985GA SA310371986GA SA310371987GA SA310371988GA SA310371989GA SA310371990GA SA310371991GA SA310371992GA SA310371993GA SA310371995GA SA310371996GA SA310371997GA SA310371998GA SA310371999GA SA310372001GA SA310372003GA SA310372004GA SA310372005GA SA310372007GA SA310372010GA SA310372012GA SA310372015GA SA310372016GA SA310372017GA SA310372018GA SA310372021GA SA310372022GA SA310372023GA SA310372024GA SA310372025GA SA310372026GA SA310372027GR SA310372029GA SA310372030GA SA310372035GA SA310372036GA SA310372037GA SA310372039GA SA310372040GA SA310372043GA SA310372048GA SA310372049GA SA310372053GA SA310372054GA SA310372055GA SA310372058GA SA310372059GR SA310372063GR SA310372064GA SA310372065GA SA310372067GR SA310372069GR SA310372072GA SA310372076GA SA310372077GA SA310372079GR SA310372082GA SA310372084GA SA310372088GR SA310372090GR SA310372092GR SA310372096GA SA310372097GA SA310372098GA SA310372099GA SA310372101GA SA310372102GR SA310372104GA SA310372105GA SA310372107GA SA310372108GA SA310372118GR SA310372139GA SA310372144GA SA310372145GA SA310372149GR SA310372151GA SA310372152GA SA310372155GA SA310372156GA SA310372158GA SA310372159GA SA310372163GA SA310372167GR SA310372168GR SA310372172GR SA310372179GA SA310372189GA SA310372193GA SA310372199GA SA310372202GA SA310372203GR SA310372204GA SA310372275GA SA310372280GA SA310372289GA SA310372294GA SA310372302GA SA310372303GA SA310372304GA SA310372307GA SA310372310GA SA310382325GR SA310382326GR SA310382329GA SA310382330GA SA310382331GR SA310382332GA SA310382333GA SA310382334GA SA310382335GA SA310382336GR SA310382343GA SA310382347GR SA310382359GA SA310382364GA SA310382371GA SA310382378GA SA310382385GA SA310382386GA SA310382387GA SA310382391GA SA310382408GA SA310382413GA SA310382415GA SA310382418GA SA310382420GA SA310382421GA SA310382424GA SA310382425GA SA310382427GA SA310382428GA SA310382429GA SA310382432GA SA310382438GA SA310382442GA SA310382446GA SA310382447GR SA310382449GA SA310382450GA SA310382454GA SA310382458GA SA310382465GA SA310382466GA SA310382468GA SA310382469GA SA310382470GA SA310382472GA SA310382473GA SA310382474GA SA310382476GA SA310382477GA SA310382478GA SA310382479GA SA310382483GA SA310382484GA SA310382485GA SA310382492GA SA310382494GA SA310382497GA SA310382499GA SA310382501GA SA310382504GA SA310382508GA SA310382511GA SA310382513GA SA310382516GA SA310382517GA SA310382524GA SA310382527GR SA310382530GA SA310382531GA SA310382532GA SA310382534GA SA310382535GA SA310382538GR SA310382540GA SA310382541GA SA310382542GA SA310382543GA SA310382544GA SA310382545GA SA310382547GA SA310382548GR SA310382550GA SA310382552GA SA310382553GR SA310382554GA SA310382555GA SA310382556GA SA310382557GA SA310382558GA SA310382561GA SA310382562GA SA310382563GA SA310382565GA SA310382566GA SA310382568GA SA310382570GA SA310382571GA SA310382572GA SA310382574GR SA310382575GA SA310382576GR SA310382577GA SA310382578GA SA310382580GA SA310382581GA SA310382583GA SA310382584GA SA310382587GA SA310382590GA SA310382591GA SA310382592GR SA310382593GA SA310382596GA SA310382598GA SA310382599GA SA310382601GA SA310382602GA SA310382603GA SA310382604GA SA310382605GA SA310382606GA SA310382607GA SA310382608GA SA310382609GA SA310382610GA SA310382611GA SA310382612GA SA310382613GA SA310382614GA SA310382615GA SA310382616GA SA310382617GA SA310382618GA SA310382619GA SA310382621GA SA310382622GA SA310382626GA SA310382627GA SA310382628GA SA310382630GR SA310382631GA SA310382632GA SA310382633GA SA310382634GA SA310382635GA SA310382636GA SA310382638GA SA310382639GA SA310382640GA SA310382642GA SA310382643GA SA310382644GA SA310382645GA SA310382646GA SA310382647GA SA310382648GA SA310382649GA SA310382650GA SA310382651GR SA310382652GA SA310382654GA SA310382655GA SA310382656GA SA310382657GR SA310382658GA SA310382659GA SA310382660GA SA310382661GA SA310382662GA SA310382663GA SA310382664GA SA310382665GA SA310382667GA SA310382670GA SA310382671GA SA310382673GA SA310382674GA SA310382682GA SA310382684GA SA310382685GA SA310382686GA SA310382689GA SA310382690GA SA310382691GR SA310382693GA SA310382696GA SA310382698GR SA310382700GA SA310382704GA SA310382706GR SA310382708GR SA310382709GR SA310382710GA SA310382712GR SA310382714GR SA310382715GR SA310382716GR SA310382717GA SA310382719GR SA310382720GR SA310382721GR SA310382722GA SA310382723GA SA310382724GR SA310382725GA SA310382726GR SA310382727GA SA310382729GA SA310382732GA SA310382733GA SA310382734GA SA310382736GA SA310382737GA SA310382739GR SA310382740GA SA310382741GR SA310382743GA SA310382744GA SA310382745GA SA310382746GA SA310382747GA SA310382748GR SA310382750GA SA310382751GA SA310382752GA SA310382754GA SA310382755GR SA310382756GA SA310382759GA SA310382761GA SA310382762GA SA310382764GA SA310382765GA SA310382766GR SA310382768GA SA310382769GR SA310382777GA SA310382782GA SA310382784GA SA310382786GA SA310382790GA SA310382791GR SA310382792GA SA310382794GR SA310382798GA SA310382800GA SA310382805GA SA310382813GA SA310382814GA SA310382821GA SA310382822GA SA310382830GA SA310382832GA SA310382836GA SA310382838GA SA310382839GR SA310382843GA SA310382847GA SA310382849GA SA310392866GA SA310392886GA SA310392887GA SA310392889GR SA310392895GA SA310392897GA SA310392898GR SA310392899GR SA310392903GA SA310392904GA SA310392911GA SA310392912GA SA310392934GA SA310392935GA SA310392936GA SA310392937GR SA310392938GA SA310392939GA SA310392942GA SA310392943GA SA310392944GA SA310392946GA SA310392947GA SA310392949GR SA310392950GA SA310392951GA SA310392952GA SA310392953GA SA310392954GA SA310392955GA SA310392957GA SA310392959GA SA310392962GA SA310392993GA SA310392994GA SA310392995GA SA310392996GA SA310392997GA SA310392998GA SA310392999GA SA310393000GA SA310393001GA SA310393003GA SA310393004GA SA310393006GA SA310393007GA SA310393008GA SA310393009GA SA310393010GA SA310393011GA SA310393012GA SA310393013GA SA310393014GA SA310393015GA SA310393017GA SA310393019GA SA310393020GA SA310393023GA SA310393024GR SA310393025GA SA310393026GA SA310393027GR SA310393028GR SA310393029GA SA310393030GR SA310393031GA SA310393032GA SA310393033GR SA310393034GR SA310393035GR SA310393036GA SA310393037GA SA310393038GA SA310393039GA SA310393040GA SA310393041GA SA310393042GA SA310393044GA SA310393045GA SA310393048GA SA310393049GR SA310393050GA SA310393051GA SA310393052GA SA310393053GA SA310393054GA SA310393055GA SA310393056GR SA310393057GR SA310393058GA SA310393059GA SA310393060GR SA310393061GR SA310393062GR SA310393063GR SA310393064GA SA310393065GA SA310393066GR SA310393067GR SA310393068GR SA310393070GA SA310393071GA SA310393072GA SA310393073GR SA310393074GA SA310393075GA SA310393076GR SA310393077GA SA310393080GA SA310393083GA SA310393086GA SA310393088GR SA310393094GA SA310393097GA SA310393106GA SA310393108GA SA310393109GA SA310393114GA SA310393130GA SA310393131GA SA310393138GR SA310393139GR SA310393140GR SA310393142GA SA310393143GA SA310393144GA SA310393145GR SA310393146GR SA310393147GR SA310393148GR SA310393149GR SA310393150GA SA310393152GR SA310393154GR SA310393155GA SA310393156GR SA310393158GR SA310393162GA SA310393164GR SA310393165GA SA310393166GR SA310393167GA SA310393168GR SA310393169GR SA310393170GA SA310393171GA SA310393172GR SA310393173GA SA310393174GR SA310393175GA SA310393177GR SA310393178GR SA310393188GA SA310393190GA SA310393192GR SA310393196GR SA310393197GA SA310393200GA SA310393202GA SA310393212GA SA310393213GA SA310393216GA SA310393219GA SA310393222GA SA310393229GA SA310393230GR SA310393233GA SA310393234GA SA310393235GR SA310393237GA SA310393243GR SA310393246GR SA310393249GA SA310393250GA SA310393252GA SA310393254GA SA310393255GA SA310393260GA SA310393261GA SA310393262GA SA310393263GR SA310393267GA SA310393268GA SA310393269GA SA310393270GA SA310393274GA SA310393279GA SA310393280GA SA310393281GA SA310393283GR SA310393284GA SA310393294GA SA310393295GA SA310393296GA SA310393303GR SA310393304GA SA310393309GA SA310393310GA SA310403311GA SA310403312GA SA310403319GA SA310403320GA SA310403321GA SA310403322GA SA310403325GA SA310403328GA SA310403332GA SA310403341GA SA310403346GA SA310403348GA SA310403349GR SA310403350GA SA310403363GA SA310403364GA SA310403372GA SA310403374GA SA310403379GA
SA310403383GR SA310403384GR SA310403385GR SA310403390GA SA310403396GR SA310403398GA SA310403403GR SA310403405GA SA310403413GR SA310403414GA SA310403418GA SA310403419GR SA310403420GR SA310403423GR SA310403426GA SA310403427GR SA310403433GA SA310403435GA SA310403440GR SA310403442GA SA310413451GA SA310413458GA SA310413471GA SA310413473GR SA310413476GA SA310413491GA SA310413493GA SA310413495GR SA310413497GR SA310413498GA SA310413499GA SA310413500GA SA310413507GA SA310413509GA SA310413511GA SA310413514GA SA310413524GA SA310413534GA SA310413535GA SA310413537GA SA310413538GA SA310413544GA SA310413545GA SA310413556GA SA310413559GA SA310413560GA SA310413561GA SA310413563GR SA310423572GA SA310423573GR SA310423577GA SA310423583GA SA310423587GA SA310423593GA SA310423600GR SA310423604GA SA310423606GA SA310423607GA SA310423611GA SA310423613GA SA310423614GA SA310423617GA SA310423618GA SA310423619GA SA310423621GA SA310423622GA SA310423626GA SA310423628GA SA310423635GR SA310423637GA SA310423639GA SA310423640GA SA310423649GA SA310423650GA SA310423651GA SA310423652GA SA310423657GA SA310423662GA SA310423663GA SA310423666GA SA310423667GA SA310423668GA SA310423670GA SA310423671GA SA310423672GA SA310423673GR SA310423674GA SA310423675GA SA310423677GA SA310423678GR SA310423680GR SA310423681GA SA310423682GR SA310423683GA SA310423684GR SA310423685GA SA310423688GA SA310423690GR SA310423694GA SA310423696GA SA310423697GA SA310423698GA SA310423701GA SA310423704GA SA310423705GA SA310423708GA SA310423709GA SA310423712GA SA310423713GA SA310423715GA SA310423720GA SA310423721GA SA310423723GA SA310423730GA SA310423736GA SA310423741GA SA310423742GA SA310423744GA SA310423745GA SA310423748GA SA310423755GA SA310423756GA SA310423758GA SA310423759GA SA310423761GA SA310423766GA SA310423769GA SA310423770GR SA310423771GA SA310423772GA SA310423773GA SA310423777GA SA310423778GA SA310423779GA SA310423780GA SA310423785GA SA310423787GA SA310423793GA SA310423794GA SA310423799GA SA310423800GA SA310423801GR SA310423806GR SA310423807GR SA310423809GA SA310423811GA SA310423813GR SA310423814GA SA310423816GR SA310423818GA SA310423819GA SA310423821GA SA310423822GA SA310423825GA SA310423836GA SA310423838GA SA310423839GA SA310423841GA SA310423842GA SA310423845GA SA310423862GA SA310423864GA SA310423866GA SA310423869GA SA310423871GA SA310423872GR SA310423873GA SA310423874GA SA310423879GA SA310423882GA SA310423883GA SA310423887GA SA310423888GA SA310423893GA SA310423895GA SA310423896GA SA310423903GA SA310423904GA SA310423906GR SA310423907GA SA310423910GA SA310423912GA SA310423914GA SA310423916GA SA310423917GA SA310423919GA SA310423921GA SA310423929GA SA310423933GA SA310423935GA SA310423937GA SA310423940GA SA310423942GA SA310423943GA SA310423944GA SA310423946GA SA310423947GR SA310423949GA SA310423950GA SA310423951GA SA310423952GR SA310423953GA SA310423956GA SA310423957GA SA310423958GA SA310423959GA SA310423960GA SA310423961GA SA310423962GA SA310423967GA SA310423971GA SA310423973GA SA310423978GR SA310423979GA SA310423980GA SA310423982GR SA310423985GA SA310423986GA SA310423987GA SA310423988GR SA310423989GA SA310423994GA SA310423995GR SA310424000GA SA310424001GA SA310424002GA SA310424003GA SA310424004GA SA310424006GA SA310424008GA SA310424010GA SA310424012GR SA310424017GA SA310424019GA SA310424025GA SA310424027GA SA310424028GR SA310424031GA SA310424032GA SA310424033GA SA310424034GA SA310424037GR SA310424038GA SA310424039GA SA310424044GA SA310424046GA SA310434048GA SA310434051GR SA310434053GA SA310434058GR SA310434059GA SA310434063GA SA310434064GA SA310434066GA SA310434069GA SA310434070GA SA310434071GA SA310434074GA SA310434076GA SA310434077GA SA310434078GA SA310434081GA SA310434084GA SA310434085GA SA310434088GR SA310434089GA SA310434090GA SA310434091GA SA310434092GA SA310434094GA SA310434097GA SA310434098GA SA310434099GR SA310434100GR SA310434102GR SA310434106GR SA310434108GA SA310434110GA SA310434114GA SA310434118GA SA310434119GA SA310434120GA SA310434122GA SA310434123GA SA310434124GA SA310434132GA SA310434134GR SA310434137GR SA310434138GA SA310434139GA SA310434146GA SA310434148GA SA310434151GA SA310434153GA SA310434157GA SA310434170GR SA310434180GA SA310434181GA SA310434182GA SA310434183GA SA310434185GA SA310434186GA SA310434188GA SA310434194GA SA310434199GA SA310434200GA SA310434204GA SA310434205GA SA310434207GA SA310434208GA SA310434210GA SA310434211GA SA310434212GA SA310434213GA SA310434227GA SA310434238GA SA310434242GA SA310444247GA SA310444248GA SA310444251GA SA310444253GA SA310444258GA SA310444259GA SA310444266GA SA310444269GA SA310444270GA SA310444272GR SA310444275GA SA310444278GA SA310444287GA SA310444289GA SA310444291GA SA310444292GA SA310444294GA SA310444296GA SA310444301GA SA310444306GA SA310444312GA SA310444324GA SA310444325GA SA310444326GA SA310444364GA SA310444369GA SA310444494GA SA310444496GA SA310444499GA SA310444503GA SA310444504GA SA310444509GA SA310444513GA SA310444521GA SA310444531GA SA310444839GA SA310444840GA SA310444842GA SA310444846GR SA310444849GR SA310444851GR SA310444856GA SA310444859GA SA310444863GA SA310444864GA SA310444893GA SA310454915GA SA310454918GA SA310454924GA SA310454925GA SA310454926GA SA310454927GA SA310454928GA SA310454929GA SA310454934GA SA310454935GA SA310454936GA SA310454937GA SA310454938GA SA310454939GA SA310454940GA SA310454943GA SA310454944GA SA310454945GA SA310454946GA SA310454947GA SA310454948GA SA310454949GA SA310454950GA SA310454952GA SA310454953GA SA310454954GA SA310454975GA SA310454976GA SA310454981GA SA310454982GA SA310454983GA SA310454984GA SA310454986GA SA310454987GA SA310454988GA SA310454989GA SA310454990GA SA310454991GR SA310454992GA SA310454994GA SA310454995GA SA310454996GR SA310455000GA SA310455004GA SA310455006GR SA310455009GA SA310455012GA SA310455015GA SA310455016GA SA310455017GR SA310455018GA SA310455024GA SA310455025GA SA310455026GR SA310455028GA SA310455030GA SA310455031GA SA310455033GA SA310455034GA SA310455046GA SA310455047GA SA310455048GA SA310455052GA SA310455053GA SA310455055GA SA310455056GA SA310455057GR SA310455058GA SA310455061GR SA310455062GA SA310455065GA SA310455066GR SA310455067GR SA310455069GA SA310455084GA SA310455091GA SA310455118GR SA310455119GA SA310455120GR SA310455121GA SA310455122GA SA310455123GA SA310455125GA SA310455126GA SA310455127GA SA310455130GA SA310455131GA SA310455142GA SA310455146GA SA310455147GA SA310455148GA SA310455149GA SA310455150GA SA310455151GA SA310455152GA SA310455153GA SA310455154GA SA310455155GA SA310455156GA SA310455157GA SA310455159GA SA310455160GA SA310455161GA SA310455163GA SA310455164GA SA310455170GA SA310455173GA SA310455174GA SA310455227GA SA310455228GA SA310455230GA SA310455232GA SA310455233GA SA310455248GA SA310455251GR SA310455254GA SA310455256GR SA310455257GA SA310455259GA SA310455264GA SA310455266GA SA310455268GA SA310455272GA SA310455275GA SA310455278GA SA310455282GA SA310455286GA SA310455289GA SA310465293GA SA310465294GA SA310465295GR SA310465298GA SA310465305GA SA310465307GA SA310465310GR SA310465312GA SA310465313GA SA310465314GA SA310465315GA SA310465317GR SA310465318GA SA310465350GA SA310465352GA SA310465359GA SA310465366GA SA310465367GA SA310465383GA SA310465394GA SA310465396GA SA310475464GA SA310475481GA SA310475485GA SA310475489GA SA310475501GA SA310475510GA SA310475517GA SA310475525GR SA310475536GA SA310475537GR SA310475540GA SA310475548GR SA310475557GA SA310475566GA SA310475575GA SA310475579GA SA310475581GA SA310475582GA SA310475585GA SA310475591GA SA310475592GA SA310475598GA SA310475600GA SA310475608GA SA310475609GA SA310475610GA SA310475611GR SA310475612GR SA310475613GA SA310475626GA SA310475628GA SA310475629GA SA310475630GA SA310475631GA SA310475634GA SA310475637GR SA310475639GA SA310475649GA SA310475663GA SA310475681GA SA310475682GA SA310475684GA SA310475685GA SA310475686GA SA310475695GA SA310475696GA SA310475697GA SA310475698GR SA310475701GA SA310475703GA SA310475707GA SA310475709GR SA310475710GA SA310475719GA SA310475728GA SA310475730GA SA310475732GA SA310475733GA SA310475734GA SA310485768GA SA310485782GA SA310485785GA SA310485786GA SA310485788GA SA310485789GA SA310485790GA SA310485792GA SA310485793GR SA310485794GA SA310485795GA SA310485796GA SA310485800GA SA310485801GA SA310485804GA SA310485806GA SA310485807GA SA310485808GR SA310485810GA SA310485811GA SA310485812GA SA310485814GA SA310485816GA SA310485818GA SA310485819GA SA310485822GA SA310485824GA SA310485828GA SA310485833GA SA310485834GA SA310485836GA SA310485843GA SA310485847GA SA310485848GA SA310485851GA SA310485853GA SA310485856GA SA310485859GA SA310485863GA SA310485867GA SA310485881GA SA310485882GR SA310495890GA SA310495913GR SA310495916GA SA310495919GA SA310495945GA SA310495946GR SA310495955GA SA310495970GA SA310496134GR SA310496135GR SA310496142GR SA310496143GR SA310496148GA SA310496156GA SA310496159GR SA310496210GR SA310496211GA SA310496212GA SA310496214GA SA310496215GA SA310496220GA SA310496221GA SA310496225GR SA310496226GA SA310496229GA SA310496232GA SA310496274GR SA310496276GA SA310496277GR SA310496283GR SA310496284GR SA310496288GR SA310496290GR SA310496291GR SA310496292GR SA310496294GR SA310496295GR SA310496296GA SA310496299GR SA310496302GR SA310496303GA SA310496304GR SA310496306GA SA310496310GR SA310496314GA SA310496317GA SA310496322GA SA310496324GA SA310496326GR SA310496329GA SA310496330GR SA310496331GA SA310496332GR SA310496340GA SA310496347GR SA310496349GA SA310496351GR SA310496352GR SA310496353GR SA310496354GR SA310496358GR SA310496360GR SA310496361GR SA310496362GR SA310496364GR SA310496366GA SA310496368GA SA310496370GA SA310496371GA SA310496372GR SA310496373GR SA310496375GA SA310496376GA SA310496378GA SA310496383GA SA310496384GA SA310496385GR SA310496386GA SA310496387GA SA310496388GA SA310496390GR SA310496393GA SA310496394GA SA310496401GA SA310496403GR SA310496410GR SA310496411GR SA310496415GA SA310496418GA SA310496421GR SA310496425GR SA310496426GR SA310496428GA SA310496450GA SA310496454GR SA310496456GA SA310496457GA SA310496461GR SA310496462GA SA310496464GR SA310496467GA SA310496469GA SA310496470GA SA310496471GA SA310496472GA SA310496474GA SA310496476GR SA310496479GR SA310496481GR SA310496483GA SA310496485GR SA310496489GA SA310496490GA SA310496492GA SA310496496GR SA310496499GA SA310496500GR SA310496501GR SA310496503GR SA310496511GR SA310496513GA SA310496514GR SA310496517GA SA310496518GR SA310496519GR SA310496520GR SA310496522GR SA310496524GR SA310496527GR SA310496528GR SA310496529GA SA310496532GR SA310496614GA SA310496615GA SA310496618GA SA310496622GA SA310496623GA SA310496624GA SA310496626GA SA310496630GA SA310496631GA SA310496633GR SA310496634GR SA310496636GR SA310496638GA SA310496643GR SA310496646GR SA310496649GR SA310496650GA SA310496651GA SA310496653GR SA310506654GA SA310506655GA SA310506656GA SA310506658GA SA310506665GA SA310506672GA SA310506674GA SA310506675GA SA310506676GR SA310506682GA SA310506683GA SA310506691GA SA310506692GA SA310506696GA SA310506700GA SA310506703GA SA310506706GA SA310506707GA SA310506708GA SA310506710GR SA310506713GA SA310506717GA SA310506725GR SA310506726GA SA310506729GA SA310506731GR SA310506732GA SA310506733GA SA310506734GR SA310506736GA SA310506739GA SA310506741GA SA310506744GA SA310506746GR SA310506747GA SA310506749GA
SA310506750GR SA310506753GR SA310506754GA SA310506762GA SA310506771GA SA310506772GR SA310506773GR SA310506775GR SA310506782GR SA310506784GA SA310506785GA SA310506786GA SA310506787GA SA310506788GA SA310506801GA SA310506808GA SA310506809GR SA310506810GA SA310506811GR SA310506812GA SA310506813GA SA310506814GA SA310506817GR SA310506818GR SA310506820GA SA310506821GA SA310506822GA SA310506824GR SA310506826GR SA310506830GR SA310506831GA SA310506832GA SA310506836GA SA310506838GA SA310506841GA SA310506843GR SA310506844GR SA310506848GR SA310506849GR SA310506852GR SA310506853GA SA310506854GA SA310506859GA SA310506860GR SA310506867GR SA310506868GR SA310506869GA SA310506870GA SA310506874GA SA310506877GA SA310506881GA SA310506882GA SA310506895GA SA310506989GA SA310506990GA SA310506991GR SA310506995GR SA310506999GA SA310507000GA SA310507003GR SA310507005GA SA310507011GR SA310507012GA SA310507014GA SA310507015GR SA310507016GA SA310507018GA SA310507021GA SA310507022GA SA310507024GA SA310507025GA SA310507027GA SA310507036GA SA310507038GA SA310507040GR SA310507045GA SA310507047GR SA310507050GA SA310507056GA SA310507058GR SA310507062GA SA310507064GA SA310507067GA SA310507069GA SA310507070GA SA310507072GA SA310507073GA SA310507074GA SA310507075GA SA310507076GR SA310507077GA SA310507078GA SA310507080GA SA310507082GA SA310507083GA SA310507084GA SA310507089GA SA310507090GA SA310507092GR SA310507093GR SA310507094GA SA310507095GR SA310507096GA SA310507100GA SA310507104GA SA310507106GA SA310507107GA SA310507108GA SA310507110GR SA310507111GA SA310507113GA SA310507114GA SA310507115GR SA310507116GR SA310507119GA SA310507121GA SA310507122GA SA310507123GA SA310507125GA SA310507127GA SA310507128GR SA310507129GA SA310507131GR SA310507132GA SA310507135GA SA310507138GA SA310507139GA SA310507140GA SA310507141GA SA310507143GA SA310507145GR SA310507146GR SA310507149GA SA310507152GR SA310507153GR SA310507154GA SA310507156GA SA310507157GA SA310507161GA SA310507162GA SA310507163GA SA310507164GA SA310507170GR SA310507171GA SA310507174GR SA310507176GA SA310507181GR SA310507183GR SA310507185GR SA310507196GA SA310507197GA SA310507199GA SA310507200GA SA310507201GA SA310507202GA SA310507203GA SA310507205GR SA310507212GA SA310507215GR SA310507216GA SA310507218GA SA310507219GA SA310507225GR SA310507228GA SA310507230GA SA310507231GR SA310507232GR SA310507235GR SA310507236GR SA310507237GA SA310507238GA SA310507240GR SA310507241GR SA310507242GR SA310507247GR SA310507262GA SA310507266GA SA310507267GA SA310507270GA SA310507272GA SA310507275GR SA310507276GA SA310507278GR SA310507279GR SA310507281GR SA310507288GA SA310507292GA SA310507296GA SA310507298GA SA310507301GA SA310507302GA SA310507303GA SA310507304GA SA310507305GA SA310507307GA SA310507308GA SA310507309GA SA310507310GA SA310507311GA SA310507314GR SA310507316GR SA310507317GR SA310507319GA SA310507326GA SA310507328GA SA310507329GA SA310507330GA SA310507332GA SA310507336GA SA310507337GA SA310507338GA SA310507344GA SA310507346GA SA310507356GA SA310507357GA SA310507366GR SA310507367GA SA310507371GA SA310507372GA SA310507373GR SA310507374GR SA310507375GA SA310507377GA SA310507379GA SA310507380GA SA310507381GA SA310507382GR SA310507383GA SA310507384GA SA310507385GA SA310507386GA SA310507387GA SA310507388GA SA310507389GA SA310507392GR SA310507395GA SA310507396GR SA310507397GA SA310507398GA SA310507399GA SA310507400GR SA310507401GA SA310507404GA SA310507406GR SA310507407GA SA310507408GA SA310507409GA SA310507410GR SA310507412GA SA310507413GR SA310507414GA SA310507415GA SA310507416GR SA310507417GA SA310507419GR SA310507420GA SA310507421GA SA310507422GR SA310507426GR SA310507427GA SA310507429GA SA310507430GA SA310507432GA SA310507433GR SA310507436GA SA310507439GA SA310507441GR SA310507442GA SA310507443GA SA310507444GA SA310507445GA SA310507447GR SA310507448GA SA310507449GA SA310507451GA SA310507452GA SA310507454GA SA310507455GA SA310507456GR SA310507458GA SA310507460GR SA310507461GA SA310507463GA SA310507464GA SA310507469GA SA310507473GA SA310507475GA SA310507476GA SA310507477GA SA310507478GA SA310507480GA SA310507481GA SA310507482GA SA310507483GA SA310507484GA SA310507485GA SA310507486GA SA310507488GA SA310507489GA SA310507490GA SA310507491GA SA310507492GA SA310507493GA SA310507494GA SA310507496GA SA310507497GA SA310507499GA SA310507503GA SA310507504GA SA310507505GA SA310507506GA SA310507507GA SA310507508GA SA310507509GA SA310507511GA SA310507512GA SA310507516GA SA310507517GA SA310507518GA SA310507519GA SA310507521GA SA310507523GA SA310507525GA SA310507526GA SA310507527GA SA310507529GA SA310507533GA SA310507534GA SA310507535GA SA310507536GA SA310507537GA SA310507539GA SA310507541GA SA310507543GA SA310507546GA SA310507547GA SA310507548GA SA310507550GA SA310507551GA SA310507552GA SA310507553GA SA310507554GA SA310507555GA SA310507556GA SA310517558GA SA310517559GA SA310517560GA SA310517563GA SA310517564GA SA310517565GA SA310517566GA SA310517567GA SA310517568GA SA310517570GA SA310517574GA SA310517581GA SA310517585GA SA310517587GA SA310517588GA SA310517589GA SA310517591GA SA310517595GA SA310517596GA SA310517597GA SA310517602GA SA310517605GA SA310517606GA SA310517607GA SA310517608GA SA310517610GR SA310517612GA SA310517613GA SA310517614GA SA310517616GA SA310517619GA SA310517633GA SA310517634GR SA310517635GA SA310517636GA SA310517638GA SA310517641GA SA310517645GR SA310517646GR SA310517650GA SA310517651GR SA310517652GR SA310517656GA SA310517657GA SA310517659GA SA310517660GA SA310517661GA SA310517662GA SA310517663GR SA310517664GR SA310517667GA SA310517670GA SA310517678GR SA310517679GA SA310517680GR SA310517696GR SA310517697GA SA310517704GA SA310517706GA SA310517707GR SA310517708GR SA310517711GA SA310517712GA SA310517717GR SA310517718GR SA310517719GA SA310517720GA SA310517721GR SA310517723GR SA310517724GR SA310517728GA SA310517729GA SA310517731GR SA310517732GA SA310517736GR SA310517737GA SA310517738GA SA310517739GA SA310517742GA SA310517744GR SA310517745GA SA310517746GA SA310517747GR SA310517751GA SA310517752GR SA310517753GA SA310517754GR SA310517755GA SA310517757GA SA310517758GA SA310517760GA SA310517761GR SA310517762GA SA310517763GA SA310517764GA SA310517765GR SA310517766GR SA310517767GA SA310517768GA SA310517769GR SA310517770GR SA310517771GR SA310517772GA SA310517773GA SA310517774GR SA310517776GA SA310517777GA SA310517778GR SA310517780GA SA310517782GA SA310517783GA SA310517785GA SA310517786GA SA310517789GA SA310517790GA SA310517791GA SA310517792GR SA310517818GA SA310517821GA SA310517823GR SA310517824GR SA310517825GA SA310517828GA SA310517830GA SA310517838GA SA310517839GA SA310517841GA SA310517865GA SA310517866GA SA310517875GR SA310517877GA SA310517878GA SA310517982GA SA310517983GR SA310517985GR SA310517986GA SA310517987GR SA310517988GA SA310517990GA SA310517991GR SA310517993GR SA310517994GA SA310517995GR SA310517996GA SA310517997GA SA310517998GA SA310518001GR SA310518003GA SA310518004GA SA310518009GA SA310518014GR SA310518147GA SA310518149GA SA310518150GR SA310518154GA SA310518156GA SA310518157GA SA310518159GA SA310518161GA SA310518162GA SA310518163GA SA310518164GA SA310518168GR SA310518169GR SA310518170GR SA310518176GR SA310518179GR SA310518181GA SA310518185GA SA310518187GA SA310518190GA SA310518194GA SA310518198GA SA310518203GA SA310518204GA SA310518205GR SA310518207GR SA310518208GA SA310518209GA SA310518210GA SA310518214GA SA310518216GR SA310518222GR SA310518223GA SA310518225GR SA310518228GR SA310518229GR SA310518230GA SA310518232GR SA310518233GR SA310518235GR SA310518236GR SA310518237GR SA310518240GR SA310518241GA SA310518242GA SA310518247GA SA310518251GA SA310518252GA SA310518255GA SA310518256GA SA310518257GA SA310518258GA SA310518259GA SA310518261GR SA310518262GA SA310518263GA SA310518265GA SA310518266GA SA310518268GA SA310518269GA SA310518270GA SA310518271GA SA310518272GA SA310518274GA SA310518275GR SA310518276GA SA310518278GA SA310518280GA SA310518282GA SA310518286GA SA310518288GA SA310518289GA SA310518291GA SA310518292GA SA310518294GA SA310518296GA SA310518297GA SA310518299GA SA310518303GA SA310518304GA SA310518305GA SA310518307GA SA310518312GA SA310518319GA SA310518320GA SA310518324GA SA310518325GA SA310518326GA SA310518330GA SA310518332GA SA310518333GA SA310518335GA SA310518336GA SA310518337GA SA310518341GA SA310518342GA SA310518343GA SA310518345GA SA310518348GA SA310518349GA SA310518350GA SA310518351GA SA310518352GA SA310518353GA SA310518356GA SA310518358GA SA310518359GA SA310518360GA SA310518361GA SA310518364GR SA310518365GR SA310518366GA SA310518367GA SA310518368GA SA310518369GA SA310518370GA SA310518371GA SA310518373GA SA310518376GA SA310518384GR SA310518385GA SA310518386GA SA310518388GA SA310518390GR SA310518391GA SA310518393GA SA310518394GR SA310518396GA SA310518397GA SA310518398GA SA310518399GA SA310518400GA SA310518401GA SA310518402GA SA310518403GA SA310518407GA SA310518408GA SA310518409GA SA310518410GA SA310518411GA SA310518412GA SA310518414GA SA310518415GA SA310518416GA SA310518417GA SA310518419GR SA310518421GR SA310518423GR SA310518424GA SA310518430GA SA310518433GA SA310518434GA SA310518436GA SA310518437GA SA310518439GA SA310518441GA SA310518444GA SA310518446GA SA310518447GA SA310518449GA SA310518450GA SA310518451GA SA310518452GA SA310518453GA SA310518457GA SA310518458GA SA310518459GA SA310518460GA SA310518461GA SA310518462GA SA310518463GA SA310518464GA SA310518466GR SA310518467GR SA310518471GA SA310518472GA SA310518473GA SA310518474GA SA310518475GA SA310518476GA SA310518477GR SA310518479GA SA310518480GA SA310518482GA SA310518483GA SA310518484GA SA310518485GA SA310518487GA SA310518491GA SA310518492GA SA310518493GA SA310518495GA SA310518499GA SA310518501GA SA310518503GA SA310518507GR SA310518509GA SA310518511GR SA310518521GA SA310518523GA SA310518527GR SA310518528GA SA310518530GA SA310518533GA SA310518535GA SA310518537GA SA310518539GR SA310518541GA SA310518543GA SA310518544GR SA310518548GA SA310518551GA SA310518552GR SA310518553GA SA310518554GA SA310518555GA SA310518561GA SA310518615GR SA310518617GA SA310518618GA SA310518621GR SA310518623GA SA310518624GR SA310518625GR SA310518626GA SA310518627GA SA310518628GA SA310518630GR SA310518631GR SA310518632GA SA310518633GA SA310518634GA SA310518635GR SA310518636GA SA310518637GR SA310518639GA SA310518640GA SA310518642GR SA310518643GR SA310518645GA SA310518648GA SA310518649GA SA310518651GA SA310518652GA SA310518653GA SA310518654GR SA310518656GA SA310518658GR SA310518660GR SA310518661GR SA310518662GR SA310518663GA SA310518664GA SA310518665GR SA310518666GR SA310518670GA SA310518671GA SA310518672GR SA310518673GA SA310518674GA SA310518675GA SA310518676GA SA310518677GA SA310518678GA SA310518679GA SA310518680GR SA310518681GA SA310518684GR SA310518685GA SA310518686GA SA310518687GA SA310518688GR SA310518690GA SA310518691GA SA310518692GA SA310518693GA SA310518694GR SA310518695GA SA310518696GA SA310518697GR SA310518698GR SA310518699GA SA310518700GA SA310518702GA SA310518703GR SA310518704GR SA310518705GA SA310518706GR SA310518710GA SA310518711GA SA310518712GA SA310518713GA SA310518714GA SA310518715GR SA310518716GR SA310518717GA SA310518718GA SA310518720GA SA310518723GA SA310518724GA SA310518725GA SA310528726GR SA310528733GA SA310528736GA SA310528744GA
SA310528746GA SA310528749GA SA310528751GR SA310528766GR SA310528773GA SA310528775GA SA310528778GA SA310528787GA SA310528795GA SA310528801GA SA310528802GA SA310528803GA SA310528810GR SA310528812GR SA310528813GA SA310528822GR SA310528825GR SA310528842GA SA310528844GA SA310528845GR SA310528849GA SA310528852GA SA310528854GA SA310528855GA SA310528856GA SA310528858GA SA310528859GA SA310528861GA SA310528862GA SA310528863GA SA310528864GA SA310528865GA SA310528868GA SA310528869GA SA310528870GA SA310528871GA SA310528875GA SA310528879GA SA310528880GA SA310528883GA SA310528887GA SA310528888GA SA310528889GA SA310528891GA SA310528895GA SA310528896GA SA310528897GA SA310528898GA SA310528899GA SA310528901GR SA310528902GR SA310528906GA SA310528911GA SA310528912GA SA310528913GR SA310528914GA SA310528915GA SA310528917GA SA310528918GA SA310528919GR SA310528920GA SA310528922GA SA310528923GA SA310528925GA SA310528926GR SA310528928GA SA310528930GA SA310528931GR SA310528932GA SA310528933GA SA310528934GR SA310528935GA SA310528936GR SA310528937GA SA310528939GA SA310528940GA SA310528943GA SA310528944GA SA310528947GA SA310528949GA SA310528951GA SA310528952GA SA310528953GA SA310528955GA SA310528957GA SA310528958GR SA310528959GA SA310528960GA SA310528961GR SA310528962GR SA310528963GR SA310528965GA SA310528967GA SA310528968GA SA310528969GA SA310528972GA SA310528975GA SA310528977GA SA310528979GR SA310528980GA SA310528981GR SA310528984GA SA310528985GR SA310528987GA SA310528989GR SA310528990GA SA310528992GA SA310528994GA SA310528997GA SA310528998GA SA310528999GA SA310529000GA SA310529001GA SA310529003GA SA310529004GA SA310529005GA SA310529006GA SA310529007GA SA310529008GA SA310529009GA SA310529010GR SA310529011GR SA310529013GA SA310529027GA SA310529028GA SA310529029GA SA310529030GA SA310529031GA SA310529032GA SA310529033GA SA310529034GA SA310529035GA SA310529036GR SA310529038GA SA310529040GA SA310529041GA SA310529042GA SA310529043GA SA310529044GA SA310529046GA SA310529047GA SA310529048GA SA310529049GA SA310529050GA SA310529053GA SA310529054GA SA310529057GA SA310529058GA SA310529060GA SA310529061GA SA310529063GR SA310529068GR SA310529069GR SA310529071GA SA310529073GR SA310529075GR SA310529079GA SA310529081GA SA310529083GR SA310529085GR SA310529086GA SA310529087GA SA310529088GA SA310529092GR SA310529093GA SA310529095GA SA310529096GA SA310529097GA SA310529099GA SA310529101GA SA310529102GA SA310529103GA SA310529106GA SA310529107GA SA310529108GA SA310529109GR SA310529112GA SA310529113GA SA310529114GA SA310529115GR SA310529116GR SA310529118GR SA310529120GR SA310529122GA SA310529156GA SA310529157GA SA310529158GA SA310529159GA SA310529160GA SA310529161GA SA310529163GA SA310529164GA SA310529165GA SA310529197GA SA310529199GA SA310529200GA SA310529201GA SA310529202GA SA310529203GA SA310529204GA SA310529206GA SA310529207GA SA310529208GA SA310529211GA SA310529214GA SA310529216GA SA310529218GA SA310529219GA SA310529220GA SA310539239GA SA310539306GA SA310539311GA SA310539316GA SA310539321GA SA310539322GA SA310539347GA SA310539349GA SA310539351GA SA311029363GA SA311029364GA SA311029365GA SA311029366GA SA311039370GA SA311039372GA SA311039377GA SA311039385GA SA311039387GA SA311039389GA SA311039391GA SA311039392GA SA311039393GA SA311039405GA SA311039409GA SA311039410GA SA311039412GA SA311039413GA SA311039414GR SA311039416GA SA311039417GA SA311039418GA SA311039419GA SA311039421GA SA311039422GA SA311039423GA SA311039424GA SA311039425GA SA311039429GA SA311039430GA SA311039432GA SA311039439GA SA311039441GA SA311039442GA SA311039449GA SA311039450GA SA311039452GA SA311039453GA SA311039458GA SA311039459GA SA311039463GA SA311039464GA SA311039467GA SA311039468GA SA311039469GR SA311039472GA SA311039473GA SA311039475GA SA311039477GA SA311039478GA SA311039479GR SA311039480GA SA311039483GA SA311039484GR SA311039485GA SA311039486GA SA311039488GA SA311039490GA SA311039494GA SA311039495GR SA311039510GR SA311039518GA SA311039520GA SA311039529GA SA311039537GR SA311039539GA SA311039540GA SA311039545GA SA311039547GA SA311039551GA SA311039552GA SA311049576GA SA311049577GA SA311049578GA SA311059580GA SA311059581GR SA311059582GR SA311059584GA SA311059585GR SA311059587GA SA311059588GR SA311059589GA SA311059590GR SA311059593GA SA311059595GR SA311059596GR SA311059597GA SA311059598GA SA311059599GA SA311059600GA SA311059601GA SA311059602GA SA311059603GR SA311059870GA SA311059871GA SA311059872GA SA311059873GA SA311059874GA SA311059875GA SA311059876GA SA311059877GA SA311059878GA SA311059879GA SA311059880GA SA311059881GA SA311059882GA SA311059883GA SA311059884GA SA311059885GA SA311059886GA SA311059887GA SA311059888GA SA311059889GA SA311059890GA SA311059891GA SA311059892GA SA311059893GA SA311059894GA SA311059895GA SA311059896GA SA311059897GA SA311059898GA SA311059899GA SA311059900GA SA311059901GA SA311059902GA SA311059903GA SA311059904GA SA311059905GA SA311059906GA SA311059907GA SA311059908GA SA311059909GA SA311060004GA SA311060007GA SA311060008GA SA311060011GA SA311060012GA SA311060013GA SA311060015GA SA311060016GR SA311060017GA SA311060020GR SA311060023GA SA311060027GA SA311060028GA SA311060035GA SA311060037GA SA311060038GA SA311060039GR SA311060041GA SA311060050GR SA311060067GA SA311060069GA SA311060070GA SA311060074GA SA311060076GA SA311060077GR SA311060084GA SA311060085GR SA311060086GR SA311060088GA SA311060092GA SA311060093GR SA311060095GR SA311060096GA SA311060097GR SA311060098GA SA311060102GA SA311060103GA SA311060104GA SA311060106GA SA311060108GA SA311060112GA SA311060115GA SA311060116GA SA311060117GA SA311060118GA SA311060119GA SA311060120GA SA311060122GA SA311060123GA SA311060124GA SA311060125GA SA311060127GA SA311060128GA SA311060129GA SA311060130GA SA311060131GA SA311060132GA SA311060133GA SA311060134GA SA311060135GA SA311060137GA SA311060139GA SA311060141GA SA311060143GA SA311060144GA SA311060145GA SA311060150GA SA311060151GA SA311060152GA SA311060153GA SA311060158GA SA311060159GA SA311060160GR SA311060165GA SA311060166GR SA311060169GA SA311060175GA SA311060176GA SA311060177GA SA311060178GA SA311060179GA SA311060182GA SA311060184GA SA311060187GA SA311060188GA SA311060190GA SA311060191GA SA311060192GR SA311060194GA SA311060195GA SA311060199GA SA311060200GA SA311060203GA SA311060206GA SA311060209GR SA311060210GA SA311060211GA SA311060215GA SA311060216GA SA311060217GA SA311060218GA SA311060220GR SA311060222GA SA311060223GA SA311060226GA SA311060231GA SA311060234GA SA311060236GA SA311060239GA SA311060241GA SA311060242GA SA311060244GR SA311060245GA SA311060247GA SA311060248GA SA311060249GA SA311060253GA SA311060257GA SA311060258GA SA311060259GA SA311060261GA SA311060262GA SA311060263GA SA311060264GA SA311060268GA SA311060272GA SA311060275GA SA311060276GA SA311060278GA SA311060279GA SA311060282GA SA311060283GR SA311060285GA SA311060286GA SA311060287GA SA311060288GA SA311060289GA SA311060290GA SA311060292GA SA311060294GR SA311060295GR SA311060296GA SA311060299GR SA311060300GR SA311060301GA SA311060302GA SA311060304GR SA311060306GA SA311060307GA SA311060309GA SA311060310GR SA311060312GA SA311060313GA SA311060314GA SA311060315GA SA311060318GR SA311060319GA SA311060321GA SA311060323GA SA311060325GR SA311060328GA SA311060329GA SA311060334GR SA311060337GR SA311060344GA SA311060346GA SA311060348GA SA311060349GA SA311060351GR SA311060353GA SA311060354GR SA311060355GA SA311060357GA SA311060358GA SA311060359GR SA311060361GA SA311060362GA SA311060363GA SA311060366GA SA311060367GA SA311060368GR SA311060370GR SA311060374GA SA311060376GA SA311060377GA SA311060381GA SA311060386GA SA311060389GR SA311060390GA SA311060392GA SA311060393GA SA311060395GA SA311060396GA SA311069606GA SA311069619GA SA311069625GA SA311069627GA SA311069630GA SA311069636GR SA311069637GA SA311069638GA SA311069639GA SA311069640GR SA311069641GA SA311069644GA SA311069645GA SA311069648GA SA311069649GA SA311069650GR SA311069651GA SA311069652GA SA311069653GA SA311069654GA SA311069655GA SA311069656GA SA311069657GR SA311069658GA SA311069659GA SA311069663GA SA311069665GA SA311069666GA SA311069667GA SA311069668GA SA311069669GR SA311069670GR SA311069674GA SA311069675GA SA311069676GA SA311069677GR SA311069678GA SA311069679GA SA311069681GR SA311069682GR SA311069683GA SA311069684GA SA311069685GR SA311069687GR SA311069688GA SA311069689GA SA311069692GA SA311069694GA SA311069695GA SA311069696GA SA311069697GA SA311069699GA SA311069701GR SA311069702GA SA311069704GA SA311069705GA SA311069706GR SA311069708GA SA311069709GA SA311069710GA SA311069711GA SA311069712GA SA311069713GA SA311069714GA SA311069715GA SA311069716GA SA311069718GA SA311069720GA SA311069721GA SA311069725GA SA311069726GA SA311069727GA SA311069729GA SA311069730GA SA311069733GA SA311069735GA SA311069742GR SA311069744GA SA311069745GA SA311069747GA SA311069749GR SA311069750GR SA311069751GR SA311069753GA SA311069755GA SA311069758GA SA311069759GA SA311069762GA SA311069763GA SA311069765GA SA311069767GA SA311069768GR SA311069769GA SA311069770GA SA311069771GA SA311069772GA SA311069774GA SA311069777GA SA311069779GR SA311069780GA SA311069781GA SA311069782GA SA311069785GA SA311069786GA SA311069787GA SA311069788GA SA311069789GR SA311069792GR SA311069793GR SA311069794GA SA311069795GR SA311069796GA SA311069798GA SA311069799GR SA311069801GR SA311069803GA SA311069804GR SA311069807GR SA311069813GA SA311069814GA SA311069815GR SA311069816GA SA311069817GA SA311069818GR SA311069819GR SA311069820GA SA311069822GA SA311069824GA SA311069827GA SA311069828GA SA311069830GA SA311069833GR SA311069835GA SA311069836GR SA311069837GR SA311069839GA SA311069840GA SA311069841GA SA311069842GA SA311069843GR SA311069844GA SA311069845GA SA311069847GA SA311069849GA SA311069850GR SA311069851GA SA311069853GA SA311069854GA SA311069856GA SA311069857GA SA311069860GA SA311070409GR SA311070443GA SA311070473GR SA311070474GR SA311070475GR SA311070476GA SA311070477GA SA311070478GA SA311070479GA SA311070480GA SA311070481GR SA311070482GA SA311070509GA SA311070517GA SA311070518GA SA311070519GA SA311070520GA SA311070521GR SA311070522GA SA311070523GA SA311070527GA SA311070528GA SA311070529GA SA311070532GA SA311070533GA SA311070534GA SA311070537GA SA311070538GA SA311070539GA SA311070540GA SA311070542GA SA311070544GA SA311070545GA SA311070546GA SA311070547GA SA311070549GA SA311070550GA SA311070551GA SA311070552GA SA311070554GA SA311070555GA SA311070561GA SA311070563GA SA311070565GA SA311070566GA SA311070567GA SA311070568GA SA311070569GA SA311070570GA SA311070572GA SA311070574GA SA311070575GA SA311070577GA SA311070578GA SA311070579GA SA311070581GA SA311070582GA SA311070585GA SA311070586GA SA311070587GA SA311070588GA SA311070593GA SA311070594GA SA311070596GA SA311070597GA SA311080599GA SA311080600GA SA311080601GA SA311080603GA SA311080606GA SA311080608GA SA311080610GA SA311080611GA SA311080613GR SA311080614GA SA311080617GA SA311080618GA SA311080619GA SA311080620GA SA311080621GA SA311080622GA SA311080623GA SA311080624GA SA311080625GA SA311080626GR SA311080627GA SA311080628GA SA311080631GA SA311080632GA SA311080634GR SA311080635GA SA311080636GR SA311080637GA SA311080638GR SA311080639GA SA311080640GR SA311080641GA SA311080642GA SA311080643GR SA311080645GA SA311080646GA SA311080649GA SA311080652GR SA311080653GR SA311080655GA
SA311080656GA SA311080657GA SA311080658GA SA311080661GA SA311080662GR SA311080663GA SA311080664GA SA311080665GA SA311080666GA SA311080667GA SA311080668GA SA311080669GA SA311080671GA SA311080680GA SA311080682GR SA311080683GA SA311080684GA SA311080685GA SA311080687GA SA311080688GA SA311080689GA SA311080690GA SA311080692GA SA311080694GA SA311080695GR SA311080698GA SA311080699GR SA311080700GA SA311080701GA SA311080702GA SA311080703GA SA311080704GA SA311080705GA SA311080706GA SA311080707GA SA311080709GA SA311080710GA SA311080711GA SA311080712GA SA311080714GA SA311080716GA SA311080717GA SA311080718GA SA311080719GA SA311080720GR SA311080721GA SA311080722GA SA311080723GR SA311080724GA SA311080725GA SA311080726GR SA311080729GA SA311080730GA SA311080731GA SA311080733GA SA311080735GA SA311080738GA SA311080739GA SA311080740GA SA311080742GA SA311080743GA SA311080744GA SA311090747GR SA311090748GA SA311090750GR SA311090751GR SA311090752GA SA311090753GA SA311090754GR SA311090755GA SA311090756GA SA311090757GR SA311090758GR SA311090759GA SA311090760GR SA311090762GA SA311090763GA SA311090764GA SA311090765GA SA311090766GA SA311090769GA SA311090770GR SA311090771GA SA311090772GA SA311090774GA SA311090775GA SA311090776GR SA311090777GA SA311090781GR SA311090782GA SA311090784GA SA311090785GA SA311090786GA SA311090787GA SA311090788GA SA311090790GA SA311090791GA SA311090793GA SA311090794GA SA311090795GA SA311090796GA SA311090801GA SA311090802GR SA311090806GA SA311090807GA SA311090808GA SA311090809GA SA311090810GA SA311090811GA SA311090812GA SA311090813GR SA311090814GA SA311090815GR SA311090816GA SA311090817GA SA311090818GA SA311090819GR SA311090820GR SA311090822GA SA311090824GA SA311090825GA SA311090826GA SA311090829GA SA311090832GA SA311090833GA SA311090836GA SA311090837GA SA311090838GR SA311090839GA SA311090840GA SA311090841GA SA311090844GA SA311090847GA SA311090848GA SA311090852GA SA311090854GA SA311090858GA SA311090865GA SA311090866GA SA311090867GA SA311090868GA SA311090871GA SA311090873GA SA311090874GA SA311090875GR SA311090878GA SA311090880GA SA311090881GA SA311090882GR SA311090883GA SA311090884GA SA311090885GA SA311090888GA SA311090889GA SA311090890GA SA311090892GA SA311090893GR SA311090894GA SA311090895GA SA311090896GR SA311090897GA SA311090898GA SA311090899GA SA311090901GA SA311090902GA SA311090903GR SA311090904GA SA311090905GR SA311090906GR SA311090908GA SA311090909GR SA311090910GA SA311090912GR SA311090917GA SA311090919GR SA311090923GR SA311090925GR SA311090926GA SA311090927GA SA311090929GR SA311090930GR SA311090931GA SA311090932GA SA311090936GA SA311090937GA SA311090939GR SA311090940GA SA311090941GA SA311090942GA SA311090943GR SA311090945GR SA311090946GA SA311090947GR SA311090950GA SA311090951GA SA311090952GA SA311090953GR SA311090954GA SA311090956GA SA311090957GA SA311090958GA SA311090959GA SA311090960GA SA311090961GA SA311090962GA SA311090963GA SA311090965GA SA311090966GA SA311090967GA SA311090969GA SA311090970GR SA311090971GA SA311090974GA SA311090975GA SA311090977GA SA311090978GA SA311090980GA SA311090981GA SA311090982GA SA311090983GA SA311090984GA SA311090985GA SA311090986GR SA311090987GA SA311090988GA SA311090989GA SA311090990GA SA311090991GA SA311090993GA SA311090994GA SA311090996GA SA311090997GA SA311090998GR SA311091000GA SA311091001GA SA311091002GR SA311091005GA SA311091008GA SA311091009GA SA311091010GA SA311091011GA SA311091013GA SA311091014GA SA311091015GA SA311091016GA SA311091017GA SA311091018GA SA311091019GA SA311091020GA SA311091021GA SA311091022GA SA311091023GA SA311091024GA SA311091025GA SA311091026GA SA311091027GA SA311091028GA SA311091030GA SA311091031GA SA311091032GA SA311091037GA SA311091038GA SA311091041GA SA311091042GR SA311091043GA SA311091044GA SA311091045GA SA311091048GA SA311091050GA SA311091051GR SA311091054GA SA311091055GA SA311091056GA SA311091057GA SA311091058GA SA311091059GA SA311091060GA SA311091061GA SA311091062GA SA311101063GA SA311101064GA SA311101065GA SA311101066GA SA311101067GA SA311101068GA SA311101069GA SA311101070GA SA311101071GA SA311101074GA SA311101076GA SA311101078GA SA311101079GA SA311101081GA SA311101082GA SA311101083GA SA311101086GA SA311101087GA SA311101091GA SA311101095GR SA311101096GA SA311101097GR SA311101101GA SA311101102GA SA311101104GA SA311101107GA SA311101108GA SA311101109GA SA311101110GA SA311101111GR SA311101112GR SA311101113GA SA311101115GA SA311101116GA SA311101117GA SA311101118GA SA311101119GA SA311101120GA SA311101122GA SA311101124GA SA311101125GA SA311101126GA SA311101128GA SA311101130GA SA311101131GA SA311101132GA SA311101134GA SA311101136GA SA311101143GA SA311101145GA SA311101147GA SA311101148GA SA311101149GA SA311101150GA SA311101155GA SA311101158GA SA311101163GA SA311101164GA SA311101165GA SA311101167GA SA311101168GA SA311101171GA SA311101174GA SA311101175GA SA311101177GA SA311101179GA SA311101180GA SA311101182GA SA311101184GA SA311101186GR SA311101187GA SA311101190GA SA311101191GA SA311101192GA SA311101193GA SA311101194GR SA311101195GA SA311101197GA SA311101198GA SA311101200GA SA311101201GA SA311101202GA SA311101203GA SA311101204GA SA311101206GA SA311101208GA SA311101210GA SA311101212GA SA311101214GA SA311101216GA SA311101217GA SA311101218GA SA311101219GR SA311101222GA SA311101228GA SA311101229GA SA311101230GA SA311101231GA SA311101232GA SA311101233GA SA311101234GA SA311101236GA SA311101237GA SA311101238GA SA311101239GA SA311101240GA SA311101242GA SA311101244GA SA311101245GA SA311101246GA SA311101247GA SA311101249GA SA311101250GA SA311101253GA SA311101256GA SA311101261GA SA311101262GA SA311101263GA SA311101264GA SA311101265GA SA311101266GA SA311101267GA SA311101268GA SA311101269GA SA311101273GA SA311101276GA SA311101277GA SA311101280GR SA311101281GA SA311101283GA SA311101284GA SA311101285GA SA311101287GR SA311101289GA SA311101290GA SA311101291GA SA311101296GR SA311101298GA SA311101300GR SA311101301GA SA311101302GA SA311101303GA SA311101304GA SA311101306GA SA311101307GA SA311101308GA SA311101309GA SA311101310GA SA311101311GA SA311101313GA SA311101316GA SA311101317GA SA311101318GA SA311101319GA SA311101320GA SA311101321GA SA311101322GA SA311101323GA SA311101324GA SA311101325GA SA311101326GA SA311101327GA SA311101328GA SA311101330GA SA311101331GA SA311101332GA SA311101333GA SA311101334GA SA311101335GA SA311101337GA SA311101338GA SA311101339GA SA311101340GA SA311101341GA SA311101342GA SA311101343GA SA311101344GA SA311101345GA SA311101346GA SA311101347GA SA311101348GA SA311101349GR SA311101350GA SA311101351GA SA311101352GA SA311101353GA SA311101354GA SA311101355GA SA311101356GA SA311101357GA SA311101358GA SA311101359GA SA311101360GA SA311101361GA SA311101363GA SA311101364GA SA311101365GA SA311101366GA SA311101367GR SA311101369GA SA311101370GA SA311101371GR SA311101372GA SA311101374GA SA311101375GA SA311101376GA SA311101377GA SA311101379GA SA311101380GA SA311101381GA SA311101382GA SA311101383GA SA311101387GA SA311101388GA SA311101390GR SA311101391GA SA311101392GA SA311101398GA SA311101399GA SA311101401GA SA311101402GA SA311101404GR SA311101405GA SA311101406GA SA311101407GA SA311101409GA SA311101411GA SA311101412GA SA311101413GA SA311101414GA SA311101415GA SA311101416GA SA311101417GA SA311101419GA SA311101421GA SA311101423GA SA311101425GA SA311101426GA SA311101427GR SA311101428GA SA311101429GA SA311101430GA SA311101432GA SA311101433GA SA311101435GR SA311101436GA SA311101437GA SA311101438GA SA311101441GA SA311101442GA SA311101446GA SA311101447GA SA311101448GA SA311101450GR SA311101451GA SA311101453GA SA311101454GA SA311101456GA SA311101458GA SA311101459GA SA311101461GA SA311101462GA SA311101463GA SA311101464GA SA311101465GA SA311101467GR SA311101468GA SA311101469GA SA311101470GA SA311101471GA SA311101472GA SA311101473GA SA311101474GA SA311101475GA SA311101476GA SA311101477GA SA311101478GA SA311101479GA SA311101480GA SA311101482GA SA311101483GA SA311101485GA SA311101486GA SA311101487GA SA311101488GA SA311101489GA SA311101491GA SA311101493GA SA311101494GA SA311101495GA SA311101496GA SA311101497GA SA311101498GA SA311101499GA SA311101502GA SA311101503GA SA311101504GA SA311101505GA SA311101506GA SA311101507GA SA311101510GA SA311101511GA SA311101512GA SA311101513GA SA311101514GA SA311101515GA SA311101516GA SA311101517GA SA311101519GA SA311101520GA SA311101521GA SA311101522GA SA311101523GA SA311101525GA SA311101526GA SA311101527GA SA311101528GA SA311101529GA SA311101531GA SA311101532GA SA311101534GA SA311101535GA SA311101536GA SA311101537GA SA311101538GA SA311101539GA SA311101540GA SA311101541GA SA311101542GA SA311101543GA SA311101544GA SA311101545GA SA311101546GA SA311101547GA SA311101549GA SA311101550GA SA311101552GA SA311101554GA SA311101555GA SA311101556GA SA311111557GA SA311111558GA SA311111559GA SA311111560GA SA311111561GA SA311111562GA SA311111563GA SA311111564GA SA311111566GR SA311111567GA SA311111568GA SA311111569GA SA311111570GA SA311111571GA SA311111572GA SA311111573GA SA311111574GA SA311111575GA SA311111576GA SA311111577GA SA311111578GA SA311111579GR SA311111580GA SA311111581GA SA311111582GA SA311111585GA SA311111587GA SA311111588GR SA311111589GA SA311111590GA SA311111591GA SA311111592GA SA311111593GA SA311111594GA SA311111596GA SA311111597GR SA311111598GR SA311111600GA SA311111601GA SA311111603GA SA311111604GA SA311111605GA SA311111607GA SA311111608GA SA311111609GA SA311111610GA SA311111612GA SA311111613GR SA311111623GA SA311111624GA SA311111625GA SA311111626GA SA311111629GA SA311111631GA SA311111633GA SA311111637GA SA311111638GR SA311111640GR SA311111641GA SA311111654GA SA311111655GA SA311111657GA SA311111659GA SA311111662GR SA311111663GR SA311111665GA SA311111666GA SA311111667GR SA311111668GA SA311111669GR SA311111670GA SA311111671GR SA311111674GR SA311111675GR SA311111676GR SA311111677GR SA311111678GA SA311111679GR SA311111680GA SA311111681GA SA311111682GA SA311111683GA SA311111684GA SA311111685GR SA311111686GR SA311111687GA SA311111688GA SA311111689GR SA311111690GA SA311111692GA SA311111693GA SA311111694GR SA311111695GR SA311111696GR SA311121698GA SA311121700GA SA311121701GA SA311121703GA SA311121708GA SA311121712GA SA311121714GA SA311121717GA SA311121718GA SA311121721GR SA311121723GA SA311121724GA SA311121725GA SA311121727GA SA311121728GR SA311121729GA SA311121730GA SA311121733GA SA311121734GR SA311121735GA SA311121738GA SA311121740GA SA311121741GA SA311121742GA SA311121743GA SA311121746GA SA311121747GR SA311121748GA SA311121749GA SA311121750GR SA311121751GA SA311121752GA SA311121755GR SA311121758GR SA311121759GA SA311121760GR SA311121761GR SA311121763GA SA311121764GR SA311121765GA SA311121766GA SA311121767GA SA311121768GR SA311121769GA SA311121770GA SA311121772GA SA311121773GR SA311121775GA SA311121776GA SA311121780GA SA311121782GR SA311121783GA SA311121785GA SA311121788GA SA311121789GR SA311121790GA SA311121791GA SA311121792GA SA311121793GA SA311121794GA SA311121795GA SA311121796GA SA311121798GA SA311121799GR SA311121800GA SA311121801GA SA311121802GA SA311121803GA SA311121804GA SA311121806GA SA311121807GR SA311121808GA SA311121809GR SA311121813GR SA311121814GA SA311121815GA SA311121816GR SA311121817GA SA311121819GA SA311121820GA SA311121821GR SA311121822GA SA311121823GA SA311121824GA SA311121825GA SA311121827GR SA311121828GR
SA311121829GA SA311121830GA SA311121831GA SA311121832GA SA311121833GA SA311121835GA SA311121837GR SA311121838GA SA311121839GA SA311121840GA SA311121843GR SA311121844GA SA311121845GA SA311121847GA SA311121848GR SA311121851GA SA311121856GA SA311121858GA SA311121867GA SA311121868GA SA311121876GA SA311121877GA SA311121880GA SA311121885GA SA311121890GR SA311121891GA SA311121892GA SA311121897GA SA311121899GA SA311121909GA SA311121919GR SA311121943GR SA311121946GR SA311121949GR SA311121957GR SA311121975GR SA311121977GR SA311121981GA SA311121983GA SA311121992GA SA311121999GR SA311122001GA SA311122003GA SA311122007GR SA311122008GA SA311122011GA SA311122012GA SA311122014GR SA311122016GA SA311122017GA SA311122020GR SA311122022GR SA311122023GA SA311122024GA SA311122025GA SA311122027GA SA311122032GR SA311122033GR SA311122034GA SA311122035GR SA311122036GR SA311122039GA SA311122041GR SA311122042GA SA311122048GA SA311122049GA SA311122050GR SA311122051GA SA311122060GR SA311122061GR SA311122065GA SA311122067GA SA311122069GR SA311122072GR SA311122074GA SA311122076GR SA311122079GA SA311122082GA SA311122084GA SA311122085GA SA311122087GR SA311122088GA SA311122089GR SA311122095GA SA311122098GA SA311122104GA SA311122105GA SA311122108GA SA311122109GA SA311122113GR SA311122116GA SA311122117GA SA311122118GA SA311122119GA SA311122120GA SA311122123GA SA311122124GA SA311122125GA SA311122127GA SA311122128GA SA311122129GR SA311122131GA SA311122132GA SA311122133GA SA311122135GA SA311122136GA SA311122138GA SA311122149GA SA311122150GA SA311122155GA SA311122156GA SA311122159GA SA311122161GA SA311122163GR SA311122166GA SA311122167GA SA311122168GA SA311122170GA SA311122171GA SA311122172GA SA311122175GA SA311122176GA SA311122178GA SA311122181GA SA311122182GR SA311122183GA SA311122184GA SA311122185GA SA311122186GA SA311122191GA SA311122194GA SA311122197GA SA311122200GA SA311122201GA SA311122202GA SA311122204GA SA311122205GR SA311122209GA SA311122210GA SA311122211GA SA311122213GA SA311122214GA SA311122215GA SA311122216GA SA311122217GA SA311122218GA SA311122223GA SA311122225GA SA311122228GA SA311122231GA SA311122232GA SA311122233GA SA311122234GA SA311122237GA SA311122239GA SA311122240GA SA311122241GA SA311122242GA SA311122243GA SA311122245GA SA311122246GA SA311122247GA SA311122249GA SA311122250GA SA311122252GA SA311122253GA SA311122254GA SA311122255GA SA311122256GA SA311122260GA SA311122261GR SA311122263GA SA311122264GA SA311122265GA SA311122266GA SA311122268GA SA311122269GA SA311122271GA SA311122272GA SA311122273GR SA311122274GA SA311132276GA SA311132278GA SA311132281GA SA311132282GA SA311132285GA SA311132289GA SA311132292GA SA311132293GR SA311132294GA SA311132296GA SA311132298GA SA311132300GA SA311132301GA SA311132302GA SA311132303GA SA311132304GR SA311132305GA SA311132306GA SA311132307GA SA311132309GA SA311132310GA SA311132311GA SA311132312GA SA311132315GA SA311132316GA SA311132317GA SA311132319GA SA311132320GA SA311132323GA SA311132324GA SA311132325GA SA311132327GA SA311132328GA SA311132329GA SA311132331GR SA311132332GA SA311132333GA SA311132334GA SA311132335GA SA311132336GA SA311132337GA SA311132338GA SA311132339GA SA311132341GA SA311132342GA SA311132343GA SA311132344GA SA311132345GA SA311132346GA SA311132347GA SA311132349GA SA311132350GR SA311132351GA SA311132353GA SA311132354GA SA311132355GA SA311132356GR SA311132357GA SA311132358GA SA311132359GA SA311132360GA SA311132361GA SA311132362GA SA311132363GA SA311132364GA SA311132365GA SA311132368GA SA311132370GA SA311132371GR SA311132372GA SA311132374GA SA311132375GA SA311132376GA SA311132378GR SA311132381GA SA311132382GA SA311132383GA SA311132384GA SA311132385GA SA311132386GA SA311132389GA SA311132390GA SA311132391GA SA311132392GA SA311132393GA SA311132394GA SA311132395GA SA311132396GA SA311132397GA SA311132400GA SA311132401GA SA311132402GA SA311132403GA SA311132405GA SA311132407GA SA311132408GA SA311132409GA SA311132410GA SA311132411GA SA311132413GR SA311132414GR SA311132415GA SA311132416GR SA311132417GA SA311132418GA SA311132421GA SA311132422GA SA311132425GA SA311132427GR SA311132430GA SA311132433GR SA311132434GA SA311132435GA SA311132438GA SA311132439GR SA311132440GA SA311132441GA SA311132444GA SA311132447GR SA311132448GA SA311132449GR SA311132452GA SA311132453GR SA311132454GR SA311132455GA SA311132457GA SA311132458GA SA311132459GR SA311132460GA SA311132461GA SA311132463GA SA311132464GA SA311132465GR SA311132467GA SA311132468GA SA311132472GA SA311132476GA SA311132479GA SA311132481GA SA311132482GA SA311132483GA SA311132488GA SA311132490GA SA311132491GA SA311132492GA SA311132494GA SA311132495GA SA311132496GR SA311132498GA SA311132499GA SA311132500GA SA311132502GA SA311132503GR SA311132504GA SA311132505GA SA311132506GA SA311132507GA SA311132508GR SA311132509GA SA311132510GA SA311132512GA SA311132514GA SA311132515GA SA311132517GA SA311132518GR SA311132521GA SA311132522GA SA311132523GA SA311132526GA SA311132527GA SA311132529GA SA311132533GA SA311132534GA SA311132537GA SA311132545GR SA311132549GA SA311132550GA SA311132552GA SA311132553GA SA311132555GA SA311132558GA SA311132560GR SA311132561GA SA311132562GA SA311132563GA SA311132565GA SA311132570GA SA311132572GR SA311132573GA SA311132574GA SA311132577GA SA311132578GA SA311132579GR SA311132581GA SA311132582GA SA311132585GA SA311132586GA SA311132587GA SA311132589GA SA311132590GA SA311132591GA SA311132592GR SA311132594GR SA311132595GA SA311132597GA SA311132600GR SA311132604GA SA311132606GA SA311132610GA SA311132612GA SA311132616GA SA311132620GA SA311132623GA SA311132630GA SA311132636GA SA311132639GR SA311132651GR SA311142678GA SA311142679GR SA311142686GR SA311142689GA SA311142696GR SA311142700GA SA311142706GR SA311142707GA SA311142713GA SA311142714GR SA311142716GA SA311142717GA SA311142718GA SA311142719GA SA311142720GA SA311142721GR SA311142723GR SA311142724GR SA311142725GR SA311142727GA SA311142730GR SA311142736GR SA311142739GA SA311142745GA SA311142746GR SA311142748GA SA311142749GA SA311142752GR SA311142754GA SA311142756GA SA311142758GA SA311142759GA SA311142760GA SA311142763GA SA311142764GA SA311142765GA SA311142767GA SA311142768GA SA311142769GA SA311142775GA SA311142779GA SA311142780GA SA311142781GR SA311142782GA SA311142783GA SA311142784GA SA311142786GA SA311142788GA SA311142789GR SA311142790GA SA311142791GA SA311142792GA SA311142793GA SA311142795GR SA311142798GA SA311142801GR SA311142802GR SA311142804GA SA311142806GA SA311142815GA SA311142816GA SA311142821GR SA311142823GA SA311142824GA SA311142852GA SA311142853GR SA311142854GR SA311142856GR SA311142872GR SA311142873GA SA311142874GA SA311142876GA SA311142877GA SA311142879GR SA311142880GR SA311142882GA SA311142883GA SA311142885GA SA311142886GA SA311142890GR SA311142891GR SA311142892GA SA311142894GA SA311142896GA SA311142898GA SA311142900GA SA311142901GA SA311142903GR SA311142904GA SA311142905GA SA311142907GA SA311142909GR SA311142910GA SA311142911GA SA311142913GA SA311142920GA SA311142921GA SA311142922GA SA311142929GA SA311142931GA SA311142935GA SA311142936GA SA311142937GA SA311142938GA SA311142940GA SA311142941GA SA311142942GA SA311142943GA SA311142944GA SA311142945GA SA311142946GA SA311142954GA SA311142958GA SA311142963GA SA311142964GA SA311142966GA SA311142969GA SA311142971GA SA311142972GA SA311142973GR SA311142974GA SA311142976GA SA311142979GA SA311142982GA SA311142984GR SA311142990GR SA311142996GA SA311142997GA SA311143001GR SA311143004GR SA311143007GA SA311143010GR SA311143012GR SA311153032GR SA311153036GA SA311153037GR SA311153039GR SA311153043GR SA311153044GR SA311153046GR SA311153047GR SA311153048GA SA311153050GR SA311153052GR SA311153054GR SA311153055GA SA311153059GR SA311153060GR SA311153063GR SA311153065GR SA311153067GR SA311153068GA SA311153069GA SA311153071GR SA311153073GA SA311153080GA SA311153084GA SA311153088GA SA311153089GA SA311153090GA SA311153094GA SA311153096GA SA311153101GA SA311153104GA SA311153108GA SA311153110GA SA311153111GA SA311153112GA SA311153114GA SA311153115GA SA311153117GA SA311153118GA SA311153119GA SA311153120GA SA311153124GA SA311153129GA SA311153132GA SA311153136GA SA311153138GA SA311153139GA SA311153144GA SA311153146GA SA311153149GA SA311153150GA SA311153154GA SA311153157GA SA311153161GA SA311153162GA SA311153163GA SA311153164GA SA311153165GA SA311153169GA SA311153170GA SA311153172GA SA311153176GA SA311153181GA SA311153188GR SA311153189GA SA311153190GA SA311153192GA SA311153194GA SA311153195GA SA311153196GA SA311153197GA SA311153202GR SA311153206GA SA311153209GA SA311153210GA SA311153213GA SA311153216GA SA311153217GA SA311153219GA SA311153223GA SA311153225GA SA311153226GR SA311153228GA SA311153230GA SA311153231GA SA311153233GA SA311153234GA SA311153236GA SA311153238GA SA311153240GA SA311153241GA SA311153242GA SA311153243GA SA311153244GA SA311153245GA SA311153246GA SA311153247GA SA311153248GA SA311153249GA SA311153250GA SA311153251GA SA311153252GA SA311153253GA SA311153255GR SA311153256GA SA311153257GA SA311153261GA SA311153263GA SA311153264GA SA311153266GR SA311153267GA SA311153269GA SA311153270GA SA311153272GA SA311153273GA SA311153274GA SA311153275GR SA311153276GA SA311153277GA SA311153278GA SA311153279GA SA311153280GA SA311153281GA SA311153282GR SA311153285GA SA311153287GA SA311153288GA SA311153290GA SA311153291GA SA311153292GA SA311153293GA SA311153294GA SA311153295GA SA311153296GA SA311153297GA SA311153298GA SA311153299GA SA311153300GA SA311153301GA SA311153302GA SA311153303GA SA311153304GA SA311153309GR SA311153311GR SA311153316GA SA311153317GR SA311153318GA SA311153319GA SA311153320GA SA311153323GA SA311153325GR SA311153326GA SA311153327GA SA311153328GA SA311153329GA SA311153330GA SA311153331GR SA311153333GR SA311153334GA SA311153335GA SA311153336GR SA311153339GA SA311153340GA SA311153341GA SA311153342GR SA311153343GA SA311153344GA SA311153346GA SA311153347GA SA311153350GA SA311153351GA SA311153352GA SA311153353GA SA311153355GR SA311153356GR SA311153357GA SA311153358GR SA311153359GA SA311153360GR SA311153362GA SA311153365GA SA311163369GA SA311163371GR SA311163376GA SA311163389GR SA311163390GA SA311163393GA SA311163398GA SA311163401GA SA311163403GA SA311163406GA SA311163410GA SA311163412GA SA311163415GR SA311163416GA SA311163417GA SA311163427GA SA311163428GA SA311163430GR SA311163431GR SA311163433GR SA311163435GR SA311163436GR SA311163437GA SA311163440GA SA311163441GR SA311163442GA SA311163443GR SA311163445GR SA311163447GA SA311163448GA SA311163449GA SA311163453GA SA311163454GR SA311163455GR SA311163459GR SA311163460GR SA311163461GA SA311163464GA SA311163466GR SA311163468GA SA311163469GR SA311163470GR SA311163473GA SA311163479GA SA311163480GA SA311163483GR SA311163484GR SA311163487GA SA311163488GR SA311163489GA SA311163532GR SA311163533GA SA311163542GA SA311163543GA SA311163545GA SA311163547GR SA311163551GR SA311163552GA SA311163559GA SA311163560GA SA311163561GA SA311163562GA SA311163563GA SA311163565GA SA311163567GR SA311163574GA SA311163576GA SA311163579GA SA311163580GA SA311163584GR SA311163585GA SA311163586GA SA311163587GR SA311163592GR SA311163594GR SA311163598GR SA311163601GA SA311163607GA SA311163609GR SA311163612GR SA311163613GA SA311163617GR SA311163620GA SA311163623GA SA311163624GA SA311163625GA SA311163627GA SA311163628GA
SA311163629GA SA311163630GA SA311163633GA SA311163638GA SA311163639GA SA311163640GA SA311163641GA SA311163642GR SA311163647GA SA311163649GR SA311163650GA SA311163651GR SA311163696GR SA311163698GA SA311163703GR SA311163710GR SA311163718GA SA311163723GR SA311163726GR SA311163729GA SA311163734GA SA311163776GR SA311163779GR SA311163782GA SA311163783GA SA311163784GR SA311163790GA SA311163794GR SA311163796GR SA311163799GR SA311163800GR SA311163801GA SA311163812GR SA311163815GA SA311163817GA SA311163818GA SA311163823GA SA311163824GA SA311163825GR SA311163830GR SA311163832GR SA311163836GR SA311163838GR SA311163840GA SA311163843GR SA311163844GA SA311163846GA SA311163857GR SA311163858GA SA311163860GA SA311163862GA SA311163868GR SA311163871GA SA311163872GA SA311163876GA SA311163878GA SA311163880GA SA311163881GA SA311163882GA SA311163884GA SA311163886GA SA311163891GA SA311163894GA SA311163896GA SA311163897GR SA311163899GA SA311163900GA SA311163901GA SA311163902GA SA311163903GA SA311163904GA SA311163905GA SA311163906GA SA311163907GA SA311163911GA SA311163913GR SA311163915GR SA311163916GA SA311163917GA SA311163919GA SA311163922GA SA311163923GA SA311163924GA SA311163926GA SA311163927GA SA311163932GA SA311163934GA SA311163935GR SA311163936GA SA311163937GR SA311163938GR SA311163942GR SA311163943GA SA311163944GR SA311163948GR SA311163949GA SA311163950GR SA311163952GR SA311163953GR SA311163954GR SA311163957GR SA311163959GR SA311163965GA SA311163966GR SA311163967GR SA311163968GA SA311163969GR SA311163970GA SA311163971GA SA311163981GR SA311163982GR SA311163985GA SA311163986GA SA311163989GR SA311163990GR SA311163991GA SA311163992GA SA311174002GA SA311174004GR SA311174007GA SA311174009GA SA311174019GR SA311174025GR SA311174031GA SA311174032GA SA311174035GR SA311174037GR SA311174038GA SA311174040GA SA311174041GR SA311174049GA SA311174053GA SA311174059GA SA311174060GA SA311174062GR SA311174064GA SA311174066GA SA311174071GA SA311174072GA SA311174075GA SA311174077GR SA311174080GR SA311174083GR SA311174084GA SA311174085GR SA311174087GA SA311174088GA SA311174089GA SA311174093GA SA311174095GA SA311174096GA SA311174098GA SA311174100GA SA311174101GR SA311174107GA SA311174119GA SA311174120GA SA311174121GA SA311174122GA SA311174123GA SA311174125GA SA311174126GA SA311174130GA SA311174131GR SA311174133GR SA311174134GR SA311174140GR SA311174141GA SA311174142GA SA311174143GA SA311174144GA SA311174149GR SA311174150GR SA311174151GA SA311174153GA SA311174156GA SA311174158GR SA311174159GA SA311174164GA SA311174165GA SA311174166GR SA311174168GA SA311174172GA SA311174176GA SA311174180GA SA311174181GA SA311174186GA SA311174187GA SA311174189GA SA311174191GA SA311174194GA SA311174195GA SA311174196GA SA311174200GA SA311174205GA SA311174208GA SA311174213GA SA311174214GA SA311174215GR SA311174216GA SA311174217GA SA311174219GA SA311174221GA SA311174222GA SA311174225GA SA311174226GA SA311174227GR SA311174229GR SA311174231GA SA311174232GA SA311174233GA SA311174236GR SA311174240GA SA311174243GR SA311174244GA SA311174246GA SA311174247GA SA311174248GA SA311174249GA SA311174250GA SA311174251GA SA311174253GR SA311174254GA SA311184336GR SA311184343GA SA311184346GA SA311184350GA SA311184351GA SA311184352GR SA311184354GR SA311184355GA SA311184356GA SA311184357GA SA311184358GA SA311184359GA SA311184360GA SA311184361GA SA311184364GA SA311184365GA SA311184366GA SA311184367GA SA311184368GA SA311184369GA SA311184371GR SA311184374GA SA311184378GA SA311184381GA SA311184382GR SA311184385GR SA311184389GA SA311184390GR SA311184392GA SA311184394GA SA311184395GA SA311184396GA SA311184397GA SA311184403GA SA311184404GA SA311184405GA SA311184407GA SA311184408GA SA311184409GA SA311184411GA SA311184412GA SA311184414GA SA311184415GR SA311184416GR SA311184417GA SA311184418GR SA311184419GA SA311184420GR SA311184421GA SA311184422GA SA311184423GA SA311184424GR SA311184425GR SA311184426GR SA311184427GA SA311184428GA SA311184429GA SA311184430GA SA311184431GA SA311184432GR SA311184433GR SA311184434GA SA311184436GR SA311184437GA SA311184439GA SA311184440GA SA311184441GR SA311184443GR SA311184444GR SA311184445GR SA311184446GR SA311184448GR SA311184449GA SA311184450GR SA311184451GR SA311184452GR SA311184453GA SA311184454GA SA311184455GR SA311184456GA SA311184458GR SA311184459GA SA311184460GA SA311184461GA SA311184462GR SA311184463GA SA311184465GR SA311184466GA SA311184467GA SA311184468GA SA311184469GA SA311184472GA SA311184475GA SA311184477GR SA311184478GA SA311184479GR SA311184481GA SA311184483GA SA311184484GA SA311184485GR SA311184487GA SA311184489GR SA311184491GR SA311184492GA SA311184493GR SA311184494GA SA311184495GR SA311184496GA SA311184499GA SA311184500GA SA311184502GA SA311184503GA SA311184504GA SA311184505GA SA311184506GA SA311184507GR SA311184509GA SA311184513GR SA311184516GA SA311184522GA SA311184523GA SA311184527GA SA311184529GA SA311184531GA SA311184535GA SA311184536GR SA311184540GA SA311184548GA SA311184549GA SA311184550GA SA311184555GA SA311184561GA SA311184563GA SA311184574GA SA311184580GA SA311184586GA SA311184588GA SA311184596GA SA311184601GA SA311184602GA SA311184605GA SA311184609GA SA311184616GA SA311184619GR SA311184620GR SA311184621GR SA311184624GA SA311184626GA SA311184627GA SA311184628GA SA311184629GA SA311184630GA SA311184631GA SA311184635GR SA311184637GA SA311184638GA SA311184639GA SA311184640GA SA311184642GA SA311184643GA SA311184644GA SA311184647GR SA311184652GA SA311184654GA SA311184655GA SA311184656GA SA311184657GA SA311184658GA SA311184659GA SA311184662GA SA311184663GA SA311184665GA SA311184666GA SA311184667GA SA311184668GA SA311184669GR SA311184671GA SA311184672GR SA311184673GR SA311184674GA SA311184675GR SA311184676GR SA311184677GA SA311184678GR SA311184679GA SA311184680GA SA311184681GA SA311184682GA SA311184683GA SA311184685GA SA311184686GA SA311184687GA SA311184688GA SA311184689GR SA311184691GR SA311184692GA SA311184693GR SA311184694GR SA311184695GA SA311184697GA SA311184698GR SA311184699GR SA311184700GR SA311184701GR SA311184702GR SA311184703GA SA311184704GA SA311184705GA SA311184706GR SA311184707GA SA311184708GA SA311184709GR SA311184710GR SA311184711GA SA311184712GA SA311184713GA SA311184714GA SA311184715GA SA311184717GA SA311184718GA SA311184720GR SA311184721GR SA311184723GA SA311184724GA SA311184726GA SA311184727GA SA311184728GR SA311184731GR SA311184733GA SA311184734GR SA311184740GA SA311184741GR SA311184742GR SA311184743GR SA311184744GR SA311184745GR SA311184746GA SA311184748GA SA311184750GR SA311184755GR SA311184756GR SA311184757GA SA311184758GA SA311184762GR SA311184764GR SA311184765GR SA311184766GR SA311184767GA SA311184770GR SA311184771GR SA311184772GR SA311184773GR SA311184774GR SA311184775GR SA311184777GR SA311184778GR SA311184779GR SA311184780GR SA311184781GA SA311184782GA SA311184783GR SA311184784GR SA311184785GR SA311184786GA SA311184787GR SA311184792GR SA311184793GA SA311184794GR SA311184795GA SA311184796GA SA311184801GA SA311184802GR SA311184804GR SA311184806GR SA311184807GA SA311184809GR SA311184811GA SA311184812GA SA311184814GR SA311184815GA SA311184817GR SA311184818GR SA311184819GA SA311184820GA SA311184821GA SA311184822GR SA311184823GR SA311184827GA SA311184828GA SA311184830GA SA311184833GA SA311184834GA SA311184835GA SA311184836GA SA311184837GA SA311184839GA SA311184841GR SA311184842GA SA311184844GR SA311184846GA SA311184847GR SA311184851GR SA311184857GR SA311184858GA SA311184859GA SA311184860GR SA311184861GA SA311184862GA SA311184864GA SA311184865GR SA311184867GA SA311184868GR SA311184873GA SA311184874GA SA311184876GA SA311184877GA SA311184878GA SA311194880GA SA311194881GA SA311194883GA SA311194884GA SA311194885GA SA311194887GA SA311194889GA SA311194891GA SA311194893GA SA311194894GA SA311194895GR SA311194897GA SA311194898GR SA311194900GR SA311194901GA SA311194902GR SA311194903GR SA311194906GA SA311194907GA SA311194908GR SA311194909GA SA311194910GR SA311194912GR SA311194913GA SA311194914GR SA311194915GA SA311194916GA SA311194917GA SA311194918GR SA311194919GR SA311194921GA SA311194922GA SA311194926GA SA311194928GR SA311194929GA SA311194932GA SA311194936GR SA311194939GA SA311194940GR SA311194941GA SA311194944GA SA311194945GR SA311194946GA SA311194947GA SA311194948GA SA311194949GA SA311194950GR SA311194951GR SA311194952GA SA311194953GA SA311194954GA SA311194955GR SA311194956GR SA311194958GA SA311194959GR SA311194960GR SA311194963GR SA311194964GR SA311194965GR SA311194966GA SA311194967GR SA311194969GR SA311194970GR SA311194971GA SA311194972GA SA311194974GA SA311194975GA SA311194976GA SA311194979GA SA311194981GA SA311194982GR SA311194983GA SA311194988GR SA311194990GA SA311194992GA SA311194995GA SA311194996GR SA311194997GR SA311194999GA SA311195002GA SA311195005GA SA311195008GA SA311195010GR SA311195011GR SA311195012GR SA311195013GA SA311195014GR SA311195016GR SA311195017GA SA311195018GA SA311195019GR SA311195020GA SA311195021GA SA311195037GR SA311195039GR SA311195040GR SA311195041GA SA311195045GA SA311195046GA SA311195050GA SA311195053GA SA311195054GR SA311195055GA SA311195056GA SA311195057GA SA311195058GA SA311195059GA SA311195060GA SA311195064GA SA311195066GA SA311195067GR SA311195069GA SA311195070GA SA311195073GA SA311195077GA SA311195078GA SA311195079GA SA311195080GA SA311195081GR SA311195082GA SA311195083GA SA311195086GA SA311195088GA SA311195089GA SA311195090GA SA311195091GA SA311195092GA SA311195094GR SA311195096GA SA311195097GA SA311195098GA SA311195099GA SA311195100GA SA311195101GA SA311195102GA SA311195104GA SA311195105GR SA311195106GA SA311195107GA SA311195108GA SA311195109GA SA311195110GA SA311195111GA SA311195112GR SA311195113GA SA311195115GA SA311195117GA SA311195118GA SA311195119GA SA311195120GA SA311195121GA SA311195122GA SA311195123GA SA311195124GA SA311195125GA SA311195126GA SA311195127GA SA311195128GA SA311195129GA SA311195130GR SA311195132GA SA311195133GA SA311195135GA SA311195137GA SA311195139GA SA311195140GA SA311195141GA SA311195142GR SA311195143GA SA311195144GA SA311195145GA SA311195148GA SA311195149GA SA311195151GA SA311195157GA SA311195160GA SA311195161GA SA311195162GA SA311195163GA SA311195164GA SA311195165GA SA311195166GA SA311195168GA SA311195169GA SA311195173GA SA311195174GR SA311195175GR SA311195176GR SA311195177GR SA311195178GR SA311195179GR SA311195180GR SA311195181GA SA311195183GR SA311195184GA SA311195185GA SA311195186GR SA311195187GR SA311195188GR SA311195189GR SA311195192GA SA311195193GA SA311195194GR SA311195195GA SA311195196GR SA311195197GA SA311195200GA SA311195201GA SA311195202GR SA311195203GR SA311195204GA SA311195205GR SA311195206GR SA311195207GR SA311195209GA SA311195210GR SA311195211GR SA311195212GA SA311195213GR SA311195215GA SA311195216GA SA311195217GA SA311195218GA SA311195219GR SA311195220GA SA311195221GA SA311195222GA SA311195223GA SA311195224GA SA311195225GA SA311195226GA SA311195227GA SA311195228GR SA311195229GA SA311195230GA SA311195231GA SA311195232GA SA311195233GA SA311195234GA SA311195236GA SA311195237GA SA311195238GA SA311195239GA SA311195240GR SA311195241GA SA311195243GA SA311195244GA SA311195245GA SA311195246GA SA311195247GA SA311195253GA SA311195259GA SA311195260GA SA311195261GR SA311195262GA SA311195263GA SA311195267GR SA311195269GA SA311195273GA SA311195277GR
SA311195279GA SA311195280GA SA311195281GA SA311195282GA SA311195283GR SA311195285GR SA311195286GA SA311205290GR SA311205291GR SA311205300GR SA311205304GA SA311205305GA SA311205308GR SA311205310GA SA311205312GR SA311205314GA SA311205315GA SA311205318GA SA311205320GR SA311205321GR SA311205331GR SA311205335GR SA311205336GR SA311205341GA SA311205343GA SA311205345GA SA311205348GA SA311205350GA SA311205353GA SA311205355GA SA311205356GA SA311205357GA SA311205359GA SA311205360GA SA311205362GA SA311205363GA SA311205364GA SA311205367GA SA311205373GA SA311205376GA SA311205379GA SA311205381GA SA311205390GA SA311205391GR SA311205392GA SA311205395GA SA311205397GR SA311205398GR SA311205399GA SA311205400GA SA311205401GR SA311205402GR SA311205408GA SA311205409GA SA311205410GA SA311205411GA SA311205412GR SA311205414GR SA311205420GA SA311205423GA SA311205424GR SA311205427GR SA311205429GA SA311205430GR SA311205432GR SA311205434GA SA311205439GA SA311205440GR SA311205443GA SA311205445GA SA311205446GA SA311205448GA SA311205449GA SA311205450GA SA311205451GA SA311205452GA SA311205453GA SA311205455GA SA311205456GA SA311205457GA SA311205458GA SA311205460GR SA311205461GA SA311205464GA SA311205465GA SA311205466GA SA311205467GR SA311205469GA SA311205471GA SA311205472GA SA311205473GA SA311205474GA SA311205475GA SA311205476GA SA311205477GA SA311205478GA SA311205479GA SA311205480GR SA311205481GA SA311205482GA SA311205483GA SA311205484GA SA311205486GA SA311205487GA SA311205488GA SA311205490GA SA311205491GA SA311205492GR SA311205494GA SA311205495GA SA311205499GA SA311205500GA SA311205501GA SA311205562GA SA311205563GR SA311205567GR SA311205568GR SA311205569GR SA311205573GA SA311205574GA SA311205575GR SA311205577GR SA311205579GR SA311205581GR SA311205582GR SA311205584GR SA311205585GR SA311205587GA SA311205588GA SA311205591GR SA311205592GA SA311205593GR SA311205597GA SA311205601GA SA311205602GR SA311205604GA SA311205605GA SA311205607GA SA311205608GA SA311205609GA SA311205611GA SA311205613GA SA311205615GA SA311205616GA SA311205617GA SA311205618GA SA311205619GA SA311215682GA SA311215688GA SA311215689GA SA311215691GA SA311215692GA SA311215697GR SA311215699GA SA311215701GR SA311215703GR SA311215704GA SA311215705GA SA311215706GA SA311215707GA SA311215708GA SA311215709GA SA311215710GA SA311215711GA SA311215712GA SA311215713GA SA311215716GA SA311215719GA SA311215720GA SA311215721GA SA311215722GA SA311215724GA SA311215725GA SA311215726GA SA311215727GA SA311215728GA SA311215730GA SA311215732GA SA311215735GA SA311215737GA SA311215738GA SA311215741GA SA311215804GR SA311215806GA SA311215808GA SA311215809GR SA311215834GA SA311215835GA SA311215836GA SA311215838GR SA311215839GA SA311215845GR SA311215851GR SA311215852GR SA311215853GR SA311215856GA SA311215857GA SA311215860GA SA311215861GA SA311225892GA SA311225893GA SA311225896GA SA311225900GR SA311225901GR SA311225902GA SA311225903GA SA311225904GA SA311225905GA SA311225906GA SA311225910GR SA311225913GR SA311225915GR SA311225916GR SA311225917GR SA311225919GA SA311225920GR SA311225921GA SA311225923GR SA311225924GR SA311225926GR SA311225927GR SA311225931GR SA311225933GR SA311225934GR SA311225936GR SA311225937GR SA311225942GA SA311225944GR SA311225947GR SA311225949GR SA311225950GR SA311225952GA SA311225955GR SA311225959GR SA311225960GR SA311225961GR SA311225963GR SA311225965GR SA311225966GR SA311225969GR SA311225970GR SA311225971GR SA311225972GR SA311225976GR SA311225980GA SA311225987GA SA311225993GA SA311226000GA SA311226002GR SA311236004GR SA311236009GR SA311236010GR SA311236011GR SA311236013GR SA311236015GA SA311236016GA SA311236023GA SA311236025GA SA311236028GR SA311236029GA SA311236030GR SA311236032GA SA311236035GA SA311236036GA SA311236037GR SA311236040GA SA311236043GA SA311236047GA SA311236050GA SA311236054GR SA311236060GA SA311236061GR SA311236064GA SA311236066GA SA311236067GA SA311236068GA SA311236069GA SA311236074GA SA311236075GA SA311236076GR SA311236077GA SA311236078GA SA311236079GA SA311236081GA SA311236082GR SA311236085GR SA311236087GA SA311236088GA SA311236090GA SA311236092GR SA311236094GA SA311236097GA SA311236098GA SA311236100GA SA311236101GA SA311236102GA SA311236104GA SA311236106GA SA311236110GA SA311236111GA SA311236116GR SA311236122GR SA311236182GR SA311236183GR SA311236184GA SA311236185GA SA311236186GA SA311236187GA SA311236191GR SA311236206GA SA311236210GA SA311236212GR SA311236216GA SA311236218GR SA311236222GA SA311236227GR SA311236228GA SA311236231GR SA311236233GA SA311236234GR SA311236235GR SA311236236GA SA311236237GR SA311236238GR SA311236239GA SA311236240GA SA311236241GA SA311236242GA SA311236243GR SA311236244GA SA311236245GA SA311236246GA SA311236247GR SA311236248GA SA311236249GA SA311236251GA SA311236256GA SA311236260GA SA311236261GA SA311236262GA SA311236263GA SA311236264GA SA311236265GA SA311236268GR SA311236269GA SA311236270GA SA311236271GR SA311236273GA SA311236274GA SA311236275GA SA311236276GA SA311236277GA SA311236278GA SA311236279GA SA311236281GA SA311236282GR SA311236285GA SA311236286GA SA311236288GA SA311236290GA SA311236291GA SA311236292GA SA311236293GA SA311236294GA SA311236295GR SA311236296GA SA311236297GA SA311236303GA SA311236307GA SA311236309GA SA311236312GA SA311236314GA SA311236315GR SA311236317GR SA311236319GA SA311236322GR SA311236323GA SA311236325GA SA311236326GA SA311236327GA SA311236328GA SA311236329GA SA311236330GA SA311236332GR SA311236334GA SA311236335GA SA311236336GA SA311236337GA SA311236338GA SA311236339GA SA311236340GA SA311236341GA SA311236342GA SA311236344GA SA311236346GR SA311236347GA SA311236348GA SA311236349GA SA311236355GA SA311236356GR SA311236358GR SA311236360GR SA311236361GR SA311236363GA SA311236367GR SA311236369GA SA311236371GR SA311236372GR SA311236376GA SA311236380GR SA311246383GA SA311246384GR SA311246386GR SA311246388GA SA311246390GR SA311246392GR SA311246393GR SA311246397GA SA311246398GA SA311246399GA SA311246404GA SA311246411GA SA311246412GR SA311246418GA SA311246419GA SA311246425GA SA311246426GA SA311246429GR SA311246436GA SA311246440GA SA311246442GA SA311246443GA SA311246444GA SA311246450GA SA311246452GA SA311246453GR SA311246456GR SA311246457GR SA311246460GA SA311246461GA SA311246462GR SA311246463GA SA311246465GA SA311246466GA SA311246470GR SA311246472GA SA311246474GA SA311246480GA SA311246483GR SA311246484GR SA311246485GA SA311246490GR SA311246493GA SA311246497GR SA311246499GA SA311246500GA SA311246502GA SA311246506GA SA311246507GA SA311246509GR SA311246510GR SA311246512GR SA311246513GA SA311246515GA SA311246516GA SA311246517GA SA311246518GA SA311246520GA SA311246521GR SA311246522GA SA311246523GA SA311246524GA SA311246525GA SA311246526GA SA311246527GA SA311246528GA SA311246529GA SA311246530GA SA311246531GA SA311246534GA SA311246537GA SA311246542GA SA311246543GA SA311246545GA SA311246546GA SA311246547GA SA311246549GR SA311246551GA SA311246552GA SA311246553GR SA311246559GA SA311246560GA SA311246561GR SA311246562GA SA311246565GR SA311246566GA SA311246567GA SA311246568GA SA311246570GA SA311246571GA SA311246572GA SA311246573GA SA311246576GA SA311246581GA SA311246583GR SA311246585GR SA311246586GA SA311246587GR SA311246588GA SA311246591GA SA311246592GA SA311246593GA SA311246594GR SA311246595GR SA311246596GA SA311246597GA SA311246598GA SA311246599GA SA311246600GR SA311246601GA SA311246602GR SA311246603GA SA311246604GA SA311246605GA SA311246606GA SA311246607GA SA311246608GA SA311246609GR SA311246611GA SA311246613GA SA311246614GA SA311246615GR SA311246616GA SA311246617GA SA311246618GA SA311246619GR SA311246620GA SA311246621GA SA311246622GA SA311246624GA SA311246625GR SA311246626GA SA311246627GA SA311246628GR SA311246629GR SA311246631GA SA311246632GR SA311246633GA SA311246634GA SA311246635GA SA311246637GA SA311246638GA SA311246639GA SA311246640GA SA311246641GA SA311246642GR SA311246643GA SA311246644GA SA311246645GR SA311246646GA SA311246647GA SA311246648GA SA311246649GA SA311246650GA SA311246651GA SA311246652GA SA311246653GR SA311246654GA SA311246655GA SA311246656GR SA311246658GA SA311246659GA SA311246660GA SA311246662GA SA311246663GA SA311246664GA SA311246665GA SA311246666GR SA311246669GA SA311246670GA SA311246672GA SA311246673GA SA311246674GR SA311246676GR SA311246677GR SA311246678GA SA311246679GA SA311246681GA SA311256687GA SA311256688GR SA311256694GA SA311256695GA SA311256699GR SA311256702GR SA311256706GR SA311256708GA SA311256709GA SA311256716GR SA311256718GA SA311256719GA SA311256721GA SA311256723GR SA311256725GA SA311256726GA SA311256727GA SA311256728GR SA311256729GR SA311256730GA SA311256731GA SA311256732GA SA311256733GA SA311256734GA SA311256735GR SA311256737GA SA311256738GA SA311256739GR SA311256740GA SA311256742GA SA311256743GA SA311256744GR SA311256745GA SA311256747GA SA311256748GA SA311256750GA SA311256751GA SA311256752GA SA311256753GA SA311256755GR SA311256756GA SA311256757GA SA311256758GA SA311256759GA SA311256760GR SA311256761GR SA311256762GR SA311256763GA SA311256765GR SA311256766GA SA311256767GA SA311256768GA SA311256769GR SA311256770GR SA311256771GA SA311256772GR SA311256773GA SA311256774GR SA311256775GA SA311256776GA SA311256777GR SA311256778GA SA311256782GR SA311256783GA SA311256785GA SA311256786GA SA311256787GR SA311256788GA SA311256789GA SA311256790GR SA311256791GA SA311256792GA SA311256794GR SA311256795GA SA311256796GA SA311256797GR SA311256799GR SA311256800GA SA311256801GA SA311256804GA SA311256805GA SA311256807GA SA311256812GR SA311256814GR SA311256815GA SA311256816GR SA311256819GA SA311256822GA SA311256823GA SA311256825GA SA311256826GR SA311256829GA SA311256830GR SA311256831GR SA311256832GA SA311256833GA SA311256834GA SA311256835GA SA311256838GA SA311256839GA SA311256840GR SA311256841GA SA311256844GR SA311256847GA SA311256855GA SA311256856GA SA311256868GA SA311256869GR SA311256870GR SA311266872GA SA311266878GA SA311266879GA SA311266880GA SA311266881GA SA311266882GA SA311266883GA SA311266886GR SA311266889GA SA311266890GA SA311266894GA SA311266895GA SA311266896GA SA311266898GA SA311266899GA SA311266901GA SA311266902GR SA311266903GA SA311266906GA SA311266907GA SA311266911GA SA311266915GR SA311266919GR SA311266922GA SA311266925GR SA311266938GA SA311266942GR SA311266946GR SA311266947GR SA311266952GR SA311266953GA SA311266954GR SA311266955GR SA311266956GA SA311266957GR SA311266958GR SA311266959GA SA311266960GR SA311266962GR SA311266963GR SA311266964GA SA311266965GR SA311266967GA SA311266968GR SA311266969GA SA311266970GR SA311266971GA SA311266972GR SA311266973GR SA311266974GR SA311266975GR SA311266976GR SA311266977GR SA311266978GR SA311266979GR SA311266980GR SA311266981GR SA311266982GR SA311266983GR SA311266984GR SA311266985GR SA311266987GR SA311266988GR SA311266989GR SA311266990GR SA311266991GR SA311266992GR SA311266997GR SA311267000GA SA311267003GR SA311267004GA SA311267006GR SA311267007GA SA311267009GR SA311267010GR SA311267011GR SA311287012GA SA311287013GA SA311287014GR SA311287015GA SA311287016GA SA311287019GR SA311287020GA SA311287021GR SA311287022GA SA311287023GA SA311287024GR SA311287025GR SA311287026GA SA311287027GA SA311287028GA SA311287029GR SA311287030GR SA311287031GA SA311287033GA SA311287036GR SA311287037GA SA311287038GA SA311287039GA SA311287040GA SA311287041GR SA311287042GR SA311287044GA
SA311287045GA SA311287046GR SA311287047GR SA311287048GR SA311287049GA SA311287050GA SA311287051GA SA311287053GR SA311287054GA SA311287055GA SA311287056GA SA311287057GR SA311287058GA SA311287059GA SA311287060GA SA311287061GA SA311287064GR SA311287065GR SA311287066GA SA311287067GR SA311287069GR SA311287070GR SA311287071GR SA311287072GR SA311287073GR SA311287074GA SA311287075GA SA311287076GA SA311287077GA SA311287078GR SA311287079GA SA311287080GA SA311287082GR SA311287083GA SA311287084GR SA311287085GR SA311287087GR SA311287088GA SA311287089GR SA311287090GR SA311287093GA SA311287094GA SA311287097GA SA311287099GR SA311287100GR SA311287101GR SA311287102GR SA311287103GA SA311287104GA SA311287105GR SA311287106GA SA311287107GR SA311287108GR SA311287109GR SA311287111GA SA311287112GA SA311287114GR SA311287115GA SA311287116GA SA311287117GR SA311287120GR SA311287121GR SA311287122GR SA311287123GR SA311287124GA SA311287125GR SA311287128GA SA311287129GR SA311287130GR SA311287131GA SA311287132GA SA311287133GR SA311287134GR SA311287135GR SA311287136GA SA311287137GR SA311287138GR SA311287140GA SA311287143GR SA311287146GR SA311287147GR SA311287149GR SA311287150GR SA311287151GR SA311287152GR SA311287153GR SA311287154GR SA311287156GA SA311287158GR SA311287159GA SA311287160GR SA311287161GR SA311287162GR SA311287165GA SA311287166GA SA311287167GA SA311287168GR SA311287169GA SA311287170GA SA311287171GR SA311287172GR SA311287174GR SA311287175GA SA311287177GR SA311287178GA SA311287179GA SA311287180GA SA311287181GR SA311287182GA SA311287183GA SA311287184GA SA311287185GR SA311287186GA SA311287187GA SA311287190GA SA311287191GR SA311287192GA SA311287194GR SA311287195GA SA311287196GA SA311287197GR SA311287198GA SA311287199GA SA311287200GA SA311287201GA SA311287202GA SA311287205GR SA311287206GR SA311287208GA SA311287209GA SA311287210GR SA311287211GA SA311287212GA SA311287213GA SA311287214GR SA311287215GR SA311287216GR SA311287217GR SA311287218GA SA311297219GR SA311297220GR SA311297221GA SA311297222GA SA311297223GA SA311297225GA SA311297226GA SA311297227GA SA311297228GR SA311297229GA SA311297231GR SA311297232GA SA311297234GA SA311297236GA SA311297237GR SA311297238GA SA311297239GR SA311297240GR SA311297242GR SA311297243GA SA311297246GA SA311297249GA SA311297250GR SA311297253GR SA311297254GA SA311297255GR SA311297256GR SA311297257GA SA311297258GA SA311297259GA SA311297260GA SA311297261GR SA311297263GA SA311297265GR SA311297266GR SA311297267GA SA311297268GR SA311297269GR SA311297270GA SA311297272GR SA311297273GA SA311297275GA SA311297276GA SA311297277GA SA311297278GA SA311297279GA SA311297280GA SA311297281GA SA311297283GR SA311297284GR SA311297285GR SA311297286GR SA311297287GR SA311297289GR SA311297291GR SA311297292GR SA311297293GA SA311297294GR SA311297299GA SA311297301GA SA311297302GR SA311297303GR SA311297304GR SA311297306GR SA311297307GR SA311297311GA SA311297316GR SA311297321GR SA311297322GA SA311297324GR SA311297325GR SA311297326GR SA311297327GA SA311297328GR SA311297329GR SA311297330GA SA311297334GA SA311297335GA SA311297336GR SA311297337GR SA311297338GR SA311297339GR SA311297340GA SA311297341GR SA311297342GR SA311297343GA SA311297344GA SA311297345GR SA311297347GR SA311297348GA SA311297351GA SA311297353GA SA311297354GA SA311297355GA SA311297356GA SA311297359GR SA311297361GA SA311297362GA SA311297364GA SA311297365GR SA311297366GA SA311297367GR SA311297368GA SA311297369GR SA311297370GA SA311297371GR SA311297372GR SA311297373GR SA311297374GR SA311297375GR SA311297376GA SA311297377GR SA311297378GA SA311297379GA SA311297380GR SA311297381GR SA311297383GA SA311297384GA SA311297386GA SA311297387GA SA311297389GR SA311297390GR SA311297391GR SA311297392GR SA311297393GR SA311297394GA SA311297395GA SA311297396GR SA311297397GR SA311297398GA SA311297399GR SA311297400GA SA311297401GR SA311297402GA SA311297403GA SA311297404GA SA311297405GA SA311297406GR SA311297407GA SA311297408GR SA311297409GR SA311307410GA SA311307411GA SA311307412GA SA311307415GR SA311307417GR SA311307418GA SA311307419GA SA311307421GA SA311307423GA SA311307424GA SA311307425GA SA311307427GA SA311307428GA SA311307432GR SA311307436GA SA311307437GA SA311307438GA SA311307440GA SA311307441GA SA311307443GA SA311307444GA SA311307445GA SA311307446GA SA311307448GA SA311307449GA SA311307450GA SA311307453GA SA311307454GA SA311307456GA SA311307463GR SA311307464GA SA311307466GA SA311307468GR SA311307471GA SA311307472GR SA311307475GA SA311307476GA SA311307479GA SA311307481GR SA311307482GR SA311307484GA SA311307491GR SA311307492GR SA311307493GA SA311307494GR SA311307495GA SA311307496GA SA311307499GA SA311307500GA SA311307501GA SA311307502GA SA311307503GA SA311307506GA SA311307507GA SA311307508GA SA311307510GA SA311307511GA SA311307512GA SA311307514GA SA311307523GR SA311307529GA SA311307531GA SA311307532GR SA311307541GA SA311307542GA SA311307543GA SA311307544GR SA311307545GR SA311307548GA SA311307549GA SA311307550GA SA311307551GR SA311307552GA SA311307555GA SA311307556GA SA311307557GR SA311307559GR SA311307560GA SA311307561GA SA311307563GR SA311307564GA SA311307565GA SA311307566GA SA311307567GA SA311307571GR SA311307572GR SA311307574GA SA311307575GR SA311307577GA SA311307578GA SA311307579GA SA311307581GA SA311307582GA SA311307585GA SA311307586GA SA311307587GA SA311307588GA SA311307590GA SA311307592GA SA311307593GA SA311307596GR SA311307597GA SA311317599GR SA311317600GR SA311317602GA SA311317606GA SA311317608GA SA311317613GA SA311317614GA SA311317615GA SA311317618GA SA311317619GR SA311317620GA SA311317621GA SA311317623GA SA311317627GR SA311317628GA SA311317629GR SA311317635GA SA311317637GA SA311317638GA SA311317639GA SA311317640GA SA311317641GA SA311317643GA SA311317647GA SA311317648GR SA311317652GA SA311317656GA SA311317657GA SA311317658GA SA311317662GA SA311317663GA SA311317664GA SA311317667GA SA311317670GA SA311317671GA SA311317672GA SA311317676GR SA311317678GR SA311317679GR SA311317683GR SA311317685GR SA311317687GR SA311317690GR SA311317691GR SA311317692GA SA311317693GR SA311317696GR SA311317697GR SA311317699GR SA311317700GR SA311317701GR SA311317702GR SA311317703GA SA311317704GA SA311317707GR SA311317708GA SA311317709GA SA311317716GA SA311317718GA SA311317720GA SA311317724GA SA311317729GA SA311317730GR SA311317732GA SA311317733GR SA311317734GA SA311317735GA SA311317736GA SA311317740GA SA311317754GA SA311317755GR SA311317765GA SA311317772GR SA311317773GA SA311327807GA SA311327808GA SA311327809GA SA311327810GA SA311327814GR SA311327816GA SA311327817GA SA311327818GR SA311327819GA SA311327820GR SA311327821GR SA311327824GA SA311327827GA SA311327829GA SA311327830GR SA311327832GR SA311327835GR SA311327836GA SA311327837GA SA311327838GR SA311327839GR SA311327840GA SA311327841GA SA311327842GA SA311327843GR SA311327845GR SA311327846GA SA311327849GR SA311327850GR SA311327851GA SA311327853GA SA311327855GR SA311327857GA SA311327858GR SA311327861GR SA311327863GR SA311327864GR SA311327866GA SA311327873GA SA311327875GA SA311327878GA SA311327879GA SA311327881GR SA311327882GR SA311327884GR SA311327887GR SA311327888GR SA311327889GA SA311327891GR SA311327893GR SA311327895GA SA311327896GA SA311327897GA SA311327898GA SA311327899GA SA311327901GR SA311327902GA SA311327903GR SA311327904GA SA311327906GA SA311327907GA SA311327908GA SA311327909GA SA311327912GA SA311327916GR SA311327917GA SA311327918GA SA311327921GA SA311327922GR SA311327923GA SA311327924GR SA311327926GA SA311327927GR SA311327928GR SA311327929GR SA311327930GA SA311327931GR SA311327933GA SA311327935GR SA311327937GA SA311327939GA SA311327941GA SA311327943GA SA311327944GA SA311327945GR SA311327946GR SA311327947GR SA311327948GA SA311327949GA SA311327950GA SA311327951GA SA311327952GA SA311327954GA SA311327955GA SA311327956GA SA311327957GA SA311327960GA SA311327961GA SA311327963GA SA311327964GA SA311327965GA SA311327967GA SA311327968GA SA311327970GA SA311337971GA SA311337972GA SA311337973GA SA311337974GA SA311337977GA SA311337979GA SA311337982GA SA311337983GA SA311337984GR SA311337987GA SA311337988GA SA311337990GR SA311337991GA SA311337992GR SA311337994GA SA311337995GA SA311337998GR SA311338000GA SA311338001GA SA311338002GA SA311338005GA SA311428006GR SA311428007GR SA311428008GR SA311428009GR SA311428010GR SA311428014GR SA311428016GR SA311428017GA SA311428019GR SA311428021GA SA311428022GR SA311428024GR SA311428025GA SA311428027GA SA311428029GR SA311428030GA SA311428034GA SA311428036GR SA311428037GR SA311428038GR SA311428039GA SA311428040GA SA311428041GR SA311428042GR SA311428044GR SA311428045GA SA311428046GR SA311428047GA SA311428048GR SA311428049GR SA311428051GA SA311428052GA SA311428053GR SA311428054GR SA311428055GA SA311428058GA SA311428067GR SA311428071GA SA311428074GR SA311428075GR SA311428076GA SA311428080GA SA311428082GR SA311428083GR SA311428085GA SA311428087GR SA311428089GA SA311428092GA SA311428093GA SA311428095GR SA311428097GA SA311428098GR SA311428099GR SA311428100GR SA311428101GA SA311428102GR SA311428103GR SA311428104GA SA311428106GA SA311428107GR SA311428108GA SA311428109GR SA311428110GR SA311428111GR SA311428113GA SA311428114GR SA311428116GR SA311428117GR SA311428118GR SA311428120GA SA311428122GA SA311428123GR SA311428126GR SA311428127GA SA311428128GA SA311428130GA SA311428132GA SA311428133GR SA311428135GR SA311428136GA SA311428137GA SA311428138GR SA311428139GA SA311428140GR SA311428141GR SA311428143GR SA311428146GA SA311428147GR SA311428149GR SA311428150GA SA311428151GA SA311428152GR SA311428153GR SA311428155GR SA311428156GR SA311428157GR SA311428160GR SA311428161GA SA311428163GR SA311428164GA SA311428165GR SA311428166GR SA311428167GR SA311428168GA SA311428169GR SA311428170GA SA311428171GA SA311448177GA SA311448188GA SA311448189GR SA311448190GR SA311448201GR SA311458217GA SA311458221GA SA311458222GA SA311458223GR SA311458226GA SA311458229GA SA311458230GA SA311458237GA SA311458239GA SA311458240GR SA311458244GR SA311458249GA SA311458252GR SA311458258GA SA311458260GA SA311458272GA SA311458282GA SA311458292GA SA311458333GR SA311458341GR SA311458344GR SA311458345GR SA311458347GR SA311458348GR SA311458350GA SA311458351GR SA311458352GR SA311458353GR SA311458354GA SA311458358GR SA311458362GR SA311458366GR SA311458367GA SA311458369GR SA311458370GR SA311458371GA SA311458372GR SA311458373GR SA311458374GA SA311468375GA SA311468377GA SA311468380GR SA311468382GR SA311468383GA SA311468384GR SA311468388GR SA311468394GR SA311468395GA SA311468396GA SA311468397GA SA311468402GR SA311468404GR SA311468407GA SA311468408GR SA311468409GA SA311468414GA SA311468415GA SA311468419GR SA311468420GR SA311468424GA SA311468432GA SA311468433GR SA311468438GR SA311468443GA SA311468445GA SA311468446GA SA311468448GA SA311488454GR SA311488455GR SA311488456GR SA311488458GR SA311488459GR SA311488460GR SA311488462GA SA311488464GA SA311488465GA SA311488468GA SA311488469GA SA311488470GA SA311488471GA SA311488472GA SA311488473GA SA311488474GA SA311488476GA SA311488477GA SA311488478GA SA311488479GA SA311488480GA SA311488481GA SA311488482GA SA311488484GA SA311488485GA SA311488486GA SA311488487GA SA311488488GA SA311508560GA SA311508561GR SA311508563GR SA311508564GA SA311508565GR SA311508566GR SA311508567GR SA311508568GR SA311508569GA SA311508570GA SA311508572GR SA311508573GA
SA311508574GR SA311508575GA SA311508576GA SA311508578GR SA311508581GA SA311508582GA SA311508584GR SA311508585GA SA311508586GA SA311508588GA SA311508589GR SA311508590GA SA311508591GR SA311508592GA SA311508593GA SA311508596GA SA311508597GA SA311508598GA SA311508599GA SA311508600GA SA311508614GA SA311508623GA SA311508624GA SA311508636GR SA311508637GA SA311508640GR SA311508643GA SA311508645GR SA311508646GA SA311508647GA SA311508648GA SA311508650GA SA311508652GA SA311508653GA SA311508654GA SA311508655GA SA311508656GA SA311508657GA SA311508658GR SA311508660GR SA311508661GR SA311508663GR SA311508664GA SA311508665GA SA311508667GR SA311508668GA SA311508670GR SA311508673GA SA311508674GA SA311508675GA SA311508676GR SA311508677GR SA311508678GR SA311508679GR SA311508680GR SA311518681GA SA311518682GR SA311518683GA SA311518684GR SA311518685GR SA311518686GR SA311518687GA SA311518688GR SA311518689GA SA311518690GR SA311518692GR SA311518693GA SA311518694GR SA311518695GA SA311518697GA SA311518699GA SA311518700GA SA311518701GA SA311518702GA SA311518703GA SA311518704GA SA311518705GA SA311518706GA SA311518707GR SA311518708GA SA311518709GA SA311518710GA SA311518712GR SA311518713GA SA311518714GR SA311518715GR SA311518716GA SA311518717GA SA311518718GA SA311518719GA SA311518720GR SA311518721GA SA311518722GA SA311518723GA SA311518724GA SA311518727GR SA311518730GA SA311518731GA SA311518732GA SA311518734GA SA311518735GA SA311518736GA SA311528737GA SA311528738GA SA311528739GA SA311528740GR SA311528741GA SA311528742GA SA311528743GA SA311528744GA SA312018746GA SA312018747GR SA312018751GR SA312018754GA SA312018755GA SA312018808GA SA312018809GA SA312018810GR SA312018811GR SA312018812GR SA312018813GR SA312018814GR SA312018815GR SA312018817GR SA312018819GR SA312018820GR SA312018822GR SA312018823GA SA312018824GA SA312018826GA SA312018827GR SA312018828GA SA312018830GR SA312018831GR SA312018835GR SA312018836GR SA312018837GR SA312018839GA SA312018840GA SA312018841GA SA312018843GA SA312018844GA SA312018845GA SA312018846GA SA312018847GA SA312018848GA SA312018850GR SA312018851GA SA312018852GA SA312018856GA SA312018857GA SA312018861GA SA312018862GA SA312018863GA SA312018865GA SA312018866GA SA312018867GA SA312018868GA SA312018869GR SA312018870GA SA312018871GA SA312018874GR SA312018875GA SA312018881GA SA312018883GR SA312018884GR SA312018886GR SA312018887GR SA312018888GR SA312018889GA SA312018893GA SA312018895GA SA312018897GA SA312018898GA SA312018899GR SA312018902GA SA312018904GA SA312018905GR SA312028910GA SA312028914GA SA312028915GA SA312028916GA SA312028917GA SA312028919GA SA312028921GA SA312028922GA SA312028926GA SA312028927GR SA312028928GA SA312028929GA SA312028932GA SA312028933GA SA312028936GA SA312028938GR SA312028940GR SA312028941GR SA312028948GR SA312028950GA SA312028966GR SA312028978GR SA312028979GA SA312028980GA SA312028982GA SA312028984GR SA312028986GA SA312028987GR SA312028988GR SA312028989GR SA312028991GA SA312028992GR SA312028993GR SA312028994GR SA312028995GR SA312028997GR SA312028998GR SA312029001GR SA312029002GA SA312029003GR SA312029004GR SA312029005GR SA312029006GR SA312029007GR SA312029009GR SA312029010GR SA312029011GR SA312029012GA SA312029014GA SA312029016GA SA312029017GR SA312029020GA SA312029021GR SA312029022GA SA312029023GA SA312029024GA SA312029025GA SA312029027GA SA312029028GR SA312029030GA SA312029031GR SA312029032GR SA312029033GA SA312029034GA SA312029035GR SA312029036GR SA312029038GR SA312029039GR SA312029043GA SA312029045GR SA312029046GA SA312029047GR SA312029052GR SA312029053GR SA312029054GA SA312029055GR SA312029056GA SA312029057GA SA312029058GR SA312029059GA SA312029060GA SA312029061GA SA312029062GR SA312029063GR SA312029064GA SA312029065GR SA312029066GR SA312029069GA SA312039076GR SA312039077GA SA312039078GA SA312039080GR SA312039081GR SA312039082GA SA312039083GR SA312039084GR SA312039085GR SA312039086GR SA312039087GR SA312039089GR SA312039090GA SA312039091GR SA312039092GA SA312039093GR SA312039094GR SA312039098GR SA312039100GA SA312039101GR SA312039102GR SA312039103GR SA312039105GA SA312039106GA SA312039107GR SA312039108GA SA312039109GA SA312039111GR SA312039112GA SA312039113GR SA312039114GA SA312039117GA SA312039118GA SA312039120GA SA312039122GR SA312039123GA SA312039125GA SA312039126GA SA312039127GR SA312039128GA SA312039129GR SA312039130GA SA312039131GR SA312039133GR SA312039134GA SA312039136GR SA312039137GA SA312039138GA SA312039140GA SA312039142GA SA312039143GA SA312039144GR SA312039147GA SA312039148GA SA312039149GA SA312039151GR SA312039152GA SA312039153GA SA312039154GR SA312039157GR SA312039158GR SA312039159GA SA312039161GA SA312039162GR SA312039163GR SA312039164GR SA312039165GA SA312039166GA SA312039168GA SA312039169GA SA312039172GA SA312039173GA SA312039174GA SA312039175GA SA312039177GA SA312039179GA SA312039180GR SA312039181GR SA312039183GA SA312039185GR SA312039190GA SA312039193GA SA312039194GR SA312039196GA SA312039197GR SA312039198GR SA312039199GR SA312039201GR SA312039205GA SA312039207GR SA312039211GR SA312039220GA SA312039221GA SA312039223GA SA312039225GA SA312039228GR SA312039230GA SA312039231GA SA312039232GA SA312049235GR SA312049236GR SA312049237GA SA312049238GA SA312049242GA SA312049243GA SA312049244GR SA312049246GR SA312049247GR SA312049248GR SA312049249GR SA312049250GR SA312049253GR SA312049254GA SA312049255GA SA312049256GR SA312049258GR SA312049259GA SA312049260GA SA312049261GR SA312049263GR SA312049264GR SA312049265GR SA312049267GA SA312049268GR SA312049270GR SA312049271GR SA312049272GA SA312049275GA SA312049276GA SA312049278GA SA312049279GR SA312049280GA SA312049281GA SA312049283GA SA312049287GA SA312049288GA SA312049289GR SA312049292GA SA312049293GA SA312049294GR SA312049296GR SA312049297GR SA312049298GR SA312049299GA SA312049300GA SA312049302GR SA312049307GR SA312049309GR SA312049312GA SA312049316GR SA312049317GR SA312049318GR SA312049320GA SA312049324GA SA312049326GA SA312049329GR SA312049331GA SA312049332GA SA312049336GA SA312049339GR SA312049341GR SA312049342GA SA312049344GA SA312049346GR SA312049347GA SA312049348GR SA312049350GR SA312049353GA SA312049356GA SA312049358GR SA312049359GR SA312049360GR SA312049361GA SA312049362GA SA312049363GR SA312049365GA SA312049368GA SA312049369GA SA312049370GA SA312049371GA SA312049372GA SA312049373GA SA312049374GA SA312049377GA SA312049378GA SA312049382GR SA312049383GR SA312049385GA SA312049386GA SA312049387GR SA312049388GA SA312049389GA SA312049390GA SA312049392GR SA312049393GA SA312049394GA SA312049395GR SA312049396GR SA312049399GR SA312049400GA SA312049401GR SA312049402GA SA312049403GA SA312049405GA SA312049407GA SA312049408GR SA312049409GA SA312049410GA SA312049411GR SA312049414GR SA312049416GA SA312049418GA SA312049419GA SA312049423GA SA312049426GA SA312049427GA SA312049428GR SA312049429GA SA312049430GA SA312049432GA SA312049433GA SA312049434GA SA312049436GA SA312049437GA SA312049438GR SA312049441GA SA312049442GA SA312049443GA SA312049446GA SA312049447GA SA312049448GA SA312049449GA SA312049451GA SA312049452GR SA312049453GA SA312049454GA SA312049457GA SA312049458GR SA312049459GA SA312049461GA SA312049462GA SA312049463GA SA312049466GA SA312049467GA SA312049468GA SA312049469GA SA312059471GA SA312059472GR SA312059473GA SA312059474GA SA312059475GR SA312059479GA SA312059480GA SA312059481GR SA312059483GA SA312059486GA SA312059489GA SA312059490GA SA312059493GR SA312059494GA SA312059496GA SA312059497GR SA312059500GA SA312059501GR SA312059502GA SA312059503GA SA312059507GA SA312059510GA SA312059511GR SA312059512GA SA312059513GA SA312059515GA SA312059516GA SA312059517GR SA312059518GR SA312059519GA SA312059520GA SA312059521GA SA312059522GA SA312059523GA SA312059524GA SA312059525GA SA312059526GA SA312059528GA SA312059530GA SA312059533GA SA312059534GA SA312059535GR SA312059536GR SA312059537GA SA312059538GR SA312059539GR SA312059540GA SA312059542GR SA312059543GA SA312059547GR SA312059549GR SA312059550GR SA312059551GR SA312059552GA SA312059553GA SA312059555GR SA312059556GR SA312059557GA SA312059558GR SA312059559GA SA312059561GR SA312059562GR SA312059563GR SA312059564GR SA312059565GR SA312059566GR SA312059569GR SA312059578GR SA312059581GR SA312059584GR SA312059587GR SA312059591GR SA312059608GA SA312059609GA SA312059610GA SA312059611GA SA312059612GR SA312059613GA SA312059614GA SA312059615GR SA312059616GA SA312059617GA SA312059618GA SA312059619GA SA312059620GA SA312059625GA SA312059630GR SA312059631GR SA312059632GA SA312059633GA SA312059635GA SA312059636GA SA312059637GA SA312059638GA SA312059639GA SA312069640GA SA312069642GA SA312069643GA SA312069646GA SA312069648GA SA312069651GA SA312069652GA SA312069653GA SA312069654GA SA312069656GA SA312069657GA SA312069659GA SA312069661GA SA312069663GA SA312069664GA SA312069665GA SA312069666GA SA312069667GA SA312069668GA SA312069669GA SA312069672GA SA312069674GA SA312069675GA SA312069677GA SA312069678GR SA312069679GA SA312069683GA SA312069684GR SA312069685GR SA312069686GR SA312069688GA SA312069689GA SA312069690GR SA312069691GR SA312069692GA SA312069694GR SA312069696GR SA312069697GA SA312069698GA SA312069700GA SA312069707GA SA312069709GA SA312069711GA SA312069712GA SA312069713GA SA312069714GR SA312069715GA SA312069716GA SA312069717GR SA312069719GR SA312069720GA SA312069721GA SA312069722GA SA312069725GR SA312069726GR SA312069728GR SA312069730GR SA312069732GR SA312069734GA SA312069736GR SA312069737GR SA312069739GR SA312069741GR SA312069742GR SA312069743GA SA312069745GR SA312069746GR SA312069749GA SA312069750GR SA312069754GA SA312069757GR SA312069759GR SA312069760GA SA312069764GR SA312069766GA SA312069767GR SA312069768GR SA312069769GR SA312069772GR SA312069775GA SA312069778GA SA312069779GA SA312069780GR SA312069782GR SA312069785GR SA312069787GA SA312069790GA SA312069792GA SA312069793GA SA312069794GA SA312069795GA SA312069796GA SA312069799GA SA312069801GA SA312069802GA SA312069803GA SA312069804GA SA312069810GA SA312069812GR SA312069815GR SA312069816GA SA312069818GA SA312069823GA SA312069826GA SA312069829GR SA312069832GR SA312069834GA SA312069837GA SA312069838GA SA312069839GA SA312069844GA SA312069846GA SA312069849GA SA312069850GA SA312069852GR SA312069855GA SA312069856GR SA312069857GA SA312069860GA SA312069861GA SA312069862GA SA312069863GA SA312069864GA SA312069865GR SA312069868GA SA312069869GA SA312069870GR SA312069872GR SA312069873GR SA312069874GR SA312069875GA SA312069876GR SA312069877GR SA312069878GR SA312069880GR SA312069881GA SA312069883GR SA312069884GR SA312070001GR SA312070002GR SA312070004GA SA312070005GA SA312070006GA SA312070007GA SA312070008GA SA312070009GA SA312070010GA SA312070011GA SA312070013GA SA312070014GA SA312070017GR SA312070018GA SA312070019GA SA312070020GA SA312070022GA SA312070023GA SA312070024GA SA312070025GA SA312070026GA SA312070027GR SA312070028GA SA312070030GA SA312070031GA SA312070032GA SA312070033GR SA312070034GA SA312070035GA SA312070036GA SA312070037GA SA312070038GA SA312070039GR SA312070040GA SA312070041GR SA312070042GA SA312070043GA SA312070044GR SA312070045GR SA312070046GA SA312070047GR SA312070048GA SA312070049GA SA312070051GR SA312070052GA SA312070053GR SA312070054GA SA312070055GA SA312070056GR SA312070057GA SA312070058GA SA312070059GA SA312070060GA SA312070061GA SA312070062GA
SA312070064GA SA312070065GR SA312070066GR SA312070067GA SA312070068GA SA312070069GA SA312070070GA SA312070071GA SA312070072GR SA312070073GA SA312070074GA SA312070075GA SA312070076GA SA312070079GA SA312070080GA SA312070081GA SA312070082GR SA312070083GR SA312070084GA SA312070085GA SA312070086GR SA312070087GR SA312070089GR SA312070090GA SA312070091GR SA312070092GR SA312070093GR SA312070094GR SA312070095GA SA312070096GR SA312070097GR SA312070098GR SA312070100GA SA312070101GA SA312070102GA SA312070105GA SA312070106GA SA312070107GA SA312070108GR SA312070109GA SA312070110GA SA312070111GR SA312070112GA SA312070113GA SA312070115GA SA312070116GA SA312070117GR SA312070119GR SA312070120GA SA312070121GR SA312070123GA SA312070124GA SA312070126GA SA312070129GR SA312070130GA SA312070131GA SA312070132GA SA312070133GA SA312070135GA SA312070136GR SA312070139GA SA312070140GR SA312070141GA SA312070142GA SA312070143GA SA312070145GA SA312070146GA SA312070147GR SA312070148GA SA312070149GR SA312070150GA SA312070151GA SA312070152GA SA312070153GR SA312070154GR SA312070155GA SA312070156GA SA312070157GA SA312070158GA SA312070159GR SA312070161GA SA312070162GA SA312070163GA SA312070164GA SA312070165GA SA312070166GA SA312070167GA SA312070168GR SA312070169GR SA312070170GA SA312070171GA SA312070172GA SA312070173GR SA312070175GA SA312070176GA SA312070177GA SA312070178GA SA312070179GA SA312070182GA SA312070183GA SA312070184GA SA312070187GR SA312070188GA SA312070190GA SA312070191GA SA312070193GR SA312070195GR SA312070196GA SA312070197GR SA312070198GA SA312070199GA SA312070200GR SA312070201GR SA312070202GR SA312070206GA SA312070208GA SA312070209GR SA312070210GA SA312070212GR SA312070214GR SA312070215GA SA312070216GA SA312070217GA SA312070219GR SA312070220GA SA312070221GA SA312070225GA SA312070226GA SA312070227GA SA312070230GA SA312070231GR SA312070232GA SA312070233GA SA312070236GA SA312070237GA SA312070241GA SA312070242GA SA312070243GR SA312070244GA SA312070245GR SA312070246GA SA312070247GA SA312070248GA SA312070249GA SA312070250GA SA312070253GR SA312070254GA SA312070255GA SA312070256GA SA312070257GA SA312070258GA SA312070259GR SA312070260GR SA312070261GA SA312070262GA SA312070263GA SA312070264GA SA312070265GA SA312070266GA SA312070267GA SA312070269GA SA312070270GA SA312070271GR SA312070272GA SA312070273GA SA312070274GR SA312070275GA SA312070276GA SA312070278GA SA312070279GA SA312070280GR SA312070281GA SA312070282GR SA312070283GR SA312070284GA SA312070285GA SA312070286GR SA312070287GR SA312070288GA SA312070289GA SA312070290GR SA312070291GR SA312070292GA SA312070293GA SA312070294GA SA312070295GR SA312070296GA SA312070297GA SA312070298GA SA312070299GR SA312070300GR SA312070301GA SA312070303GA SA312070304GA SA312070305GA SA312070306GA SA312070307GR SA312070308GR SA312070309GR SA312070310GA SA312070311GR SA312070312GR SA312070313GA SA312070314GR SA312070315GR SA312070317GR SA312070318GR SA312070319GR SA312070320GA SA312070321GR SA312070322GA SA312070323GR SA312070324GA SA312070325GR SA312070326GR SA312070327GA SA312070328GA SA312070329GA SA312070330GR SA312070331GR SA312070332GR SA312070333GA SA312070334GA SA312070335GA SA312070336GA SA312070338GR SA312070339GA SA312070342GA SA312070343GA SA312070345GR SA312070346GA SA312070347GA SA312070349GA SA312070350GA SA312070351GR SA312070353GR SA312070354GA SA312070355GR SA312070356GR SA312070357GA SA312070359GR SA312070361GA SA312070363GR SA312070364GA SA312070365GA SA312070366GA SA312070367GA SA312070368GA SA312070369GR SA312070370GR SA312070371GA SA312070372GR SA312070373GA SA312070374GR SA312070375GA SA312070376GA SA312070377GA SA312070378GA SA312070380GA SA312070381GA SA312070382GR SA312070384GR SA312070385GA SA312079887GA SA312079888GR SA312079890GR SA312079891GR SA312079892GR SA312079894GR SA312079895GR SA312079897GR SA312079899GA SA312079900GA SA312079902GA SA312079907GR SA312079908GA SA312079914GR SA312079915GA SA312079917GA SA312079918GA SA312079919GA SA312079920GA SA312079922GA SA312079924GA SA312079926GA SA312079927GA SA312079928GA SA312079929GR SA312079930GR SA312079933GA SA312079934GA SA312079935GA SA312079936GA SA312079937GA SA312079938GA SA312079940GA SA312079941GA SA312079942GA SA312079943GR SA312079944GA SA312079945GR SA312079946GA SA312079948GA SA312079950GR SA312079952GA SA312079953GA SA312079956GA SA312079957GA SA312079958GR SA312079959GA SA312079960GA SA312079961GR SA312079962GA SA312079963GR SA312079966GA SA312079967GA SA312079968GA SA312079970GA SA312079971GA SA312079972GA SA312079974GA SA312079975GA SA312079977GA SA312079979GA SA312079980GA SA312079982GA SA312079983GA SA312079985GR SA312079987GA SA312079988GA SA312079989GA SA312079991GA SA312079992GA SA312079993GA SA312079994GA SA312079995GA SA312079996GA SA312079998GA SA312080386GA SA312080388GA SA312080389GA SA312080390GA SA312080396GA SA312080397GA SA312080398GA SA312080399GA SA312080401GA SA312080402GA SA312080403GA SA312080409GA SA312080410GA SA312080412GA SA312080415GA SA312080416GA SA312080417GA SA312080419GA SA312080420GR SA312080422GA SA312080423GA SA312080424GA SA312080425GA SA312080426GA SA312080427GA SA312080429GA SA312080431GA SA312080432GA SA312080433GA SA312080434GA SA312080462GA SA312080463GA SA312080464GA SA312080465GA SA312080466GA SA312080467GA SA312080471GA SA312080472GA SA312080473GA SA312080476GA SA312080477GA SA312080478GA SA312080479GA SA312080486GA SA312080487GA SA312080488GA SA312080489GR SA312080494GA SA312080496GA SA312080497GA SA312080498GA SA312080500GA SA312080505GA SA312080506GA SA312080508GA SA312080509GA SA312080510GA SA312080512GA SA312080513GA SA312080514GA SA312080515GA SA312080517GA SA312080520GA SA312080522GA SA312080523GA SA312080525GA SA312080526GA SA312080528GA SA312080529GR SA312080532GA SA312080533GR SA312080534GA SA312080535GA SA312080536GA SA312080537GA SA312080538GA SA312080539GR SA312080540GA SA312080541GA SA312080543GA SA312080544GR SA312080545GR SA312080548GA SA312080551GA SA312080552GA SA312080553GR SA312080554GR SA312080555GR SA312080556GR SA312080558GR SA312080559GR SA312080560GA SA312080563GA SA312080564GA SA312080566GA SA312080570GA SA312080571GA SA312080573GA SA312080574GA SA312080575GA SA312080577GR SA312080579GA SA312080584GA SA312080587GR SA312080590GA SA312080591GA SA312080594GA SA312080596GR SA312080597GA SA312080598GR SA312080604GR SA312080608GR SA312080610GA SA312080612GA SA312080622GA SA312080624GA SA312080625GR SA312080626GR SA312080629GA SA312080630GR SA312080631GR SA312080635GA SA312080645GA SA312080646GR SA312080647GA SA312080648GR SA312080649GR SA312080653GA SA312080655GR SA312080656GR SA312080658GR SA312080659GA SA312080662GR SA312080663GR SA312080666GA SA312080667GR SA312080668GA SA312080670GA SA312080671GA SA312080674GR SA312080677GR SA312080678GR SA312080679GR SA312080681GA SA312080683GA SA312080684GA SA312080686GR SA312080688GR SA312080689GA SA312080690GR SA312080691GA SA312080692GR SA312080693GA SA312080695GR SA312080700GR SA312080701GR SA312080704GA SA312080705GA SA312080706GA SA312080707GR SA312080708GR SA312080711GA SA312080713GA SA312080714GA SA312080716GA SA312080717GA SA312080719GR SA312080720GA SA312080721GA SA312080724GR SA312080725GA SA312080727GA SA312080728GA SA312080729GA SA312080731GA SA312080732GA SA312080743GR SA312080744GA SA312080748GR SA312080749GA SA312080751GA SA312080753GA SA312080754GA SA312080756GA SA312080758GA SA312080759GA SA312080761GA SA312080763GR SA312080764GA SA312080769GA SA312080770GA SA312080771GA SA312080775GR SA312080777GA SA312080779GA SA312080781GA SA312080783GR SA312080784GR SA312080787GR SA312080790GR SA312080791GR SA312080792GA SA312080794GR SA312080796GA SA312080797GR SA312080798GA SA312080802GA SA312080803GA SA312080804GA SA312080805GR SA312080806GA SA312080808GA SA312080809GR SA312080810GA SA312080811GA SA312080812GR SA312080813GA SA312080814GA SA312080817GA SA312080819GA SA312080820GA SA312080822GA SA312080824GA SA312080826GA SA312080827GA SA312080828GR SA312080830GA SA312080831GR SA312080833GR SA312080834GA SA312080837GA SA312080839GA SA312080840GA SA312080845GA SA312080848GA SA312080852GA SA312080856GR SA312080858GA SA312080860GA SA312080864GR SA312080871GA SA312080872GA SA312080873GA SA312080874GR SA312080876GA SA312080877GA SA312080878GA SA312080881GA SA312080882GA SA312080884GA SA312080885GA SA312080886GR SA312080888GA SA312080890GR SA312080892GA SA312080893GA SA312080894GA SA312080899GA SA312080900GR SA312080901GA SA312080903GA SA312080905GA SA312080909GA SA312080912GA SA312080914GA SA312080916GA SA312090917GA SA312090921GA SA312090923GA SA312090927GA SA312090928GA SA312090931GR SA312090934GA SA312090942GR SA312090948GA SA312090966GR SA312090967GA SA312090969GA SA312090970GA SA312090972GR SA312090974GR SA312090978GA SA312090981GR SA312090982GA SA312090983GA SA312090985GA SA312090987GA SA312090988GA SA312090989GA SA312090990GA SA312090993GA SA312090994GA SA312090995GR SA312090996GA SA312090997GA SA312090998GA SA312090999GA SA312091000GA SA312091001GR SA312091002GA SA312091003GR SA312091004GA SA312091006GR SA312091007GR SA312091009GA SA312091010GA SA312091011GA SA312091013GA SA312091014GA SA312091017GA SA312091018GA SA312091019GR SA312091020GR SA312091021GA SA312091022GR SA312091024GA SA312091025GA SA312091026GA SA312091027GA SA312091028GA SA312091030GA SA312091031GA SA312091032GA SA312091033GR SA312091034GA SA312091035GA SA312091036GA SA312091037GA SA312091038GA SA312091039GA SA312091041GA SA312091042GA SA312091044GA SA312091045GA SA312091046GA SA312091048GA SA312091052GA SA312091053GR SA312091054GA SA312091055GA SA312091057GA SA312091058GA SA312091060GA SA312091061GA SA312091062GA SA312091063GA SA312091064GA SA312091066GA SA312091067GA SA312091068GA SA312091069GA SA312091070GA SA312091071GA SA312091072GA SA312091073GA SA312091074GR SA312091075GR SA312091076GA SA312091077GA SA312091078GA SA312091079GA SA312091081GA SA312091082GA SA312091083GA SA312091084GA SA312091085GA SA312091086GA SA312091087GA SA312091088GA SA312091090GA SA312091091GA SA312091092GA SA312091094GA SA312091095GA SA312091096GA SA312091103GA SA312091104GA SA312091107GA SA312091108GA SA312091109GA SA312091110GA SA312091111GR SA312091113GA SA312091114GR SA312091116GR SA312091125GA SA312091137GA SA312091145GA SA312091146GA SA312091148GA SA312091153GA SA312091154GA SA312091178GA SA312091190GA SA312091196GA SA312091197GA SA312091201GA SA312091202GA SA312091204GA SA312091206GA SA312091207GA SA312091208GA SA312091211GR SA312091215GA SA312091216GR SA312091218GA SA312091219GA SA312091220GA SA312091221GA SA312091222GA SA312091223GA SA312091224GA SA312091225GR SA312091226GA SA312091227GA SA312091229GA SA312091230GA SA312091231GA SA312091232GA SA312091233GA SA312091234GA SA312091236GA SA312091237GA SA312091238GA SA312091239GA SA312091240GA SA312091241GA SA312091242GA SA312091243GA SA312091244GA SA312091245GA SA312091246GA SA312091247GA SA312091248GA SA312091250GA SA312091251GR SA312091252GA SA312091253GR SA312091254GA SA312091256GA SA312091257GA SA312091258GA SA312091259GA SA312091260GR SA312091261GA SA312091262GR SA312091265GA SA312091266GA SA312091268GA SA312091269GR SA312091270GR SA312091271GA SA312091272GA SA312091274GR SA312091275GA SA312091276GA SA312091277GR SA312091278GR SA312091279GR SA312091280GR SA312091281GR SA312091282GR
SA312091283GR SA312091286GA SA312091287GA SA312091289GA SA312091293GA SA312091294GA SA312091297GR SA312091298GA SA312091299GR SA312091300GR SA312091302GA SA312091304GR SA312091306GR SA312091307GR SA312091308GR SA312091309GR SA312091310GA SA312091311GR SA312091312GR SA312091313GR SA312091314GR SA312091315GR SA312091316GR SA312091317GR SA312091319GR SA312091320GR SA312091322GR SA312091323GR SA312091325GR SA312091326GR SA312091328GR SA312091329GR SA312091330GR SA312091331GR SA312091332GA SA312091333GA SA312091334GR SA312091335GA SA312091336GR SA312091337GA SA312091338GR SA312091339GR SA312091340GR SA312091341GR SA312091342GR SA312091343GR SA312091345GR SA312091346GR SA312091347GR SA312091348GR SA312091349GR SA312091350GR SA312091352GR SA312091353GR SA312091355GA SA312091356GR SA312091359GR SA312091362GR SA312091365GA SA312101374GR SA312101375GA SA312101377GR SA312101386GA SA312101389GR SA312101391GR SA312101394GR SA312101396GR SA312101397GA SA312101399GR SA312101401GR SA312101406GA SA312101408GR SA312101411GR SA312101412GR SA312101414GR SA312101416GA SA312101418GA SA312101421GR SA312101428GA SA312101430GA SA312101433GR SA312101434GA SA312101438GR SA312101440GA SA312101441GA SA312101443GA SA312101445GR SA312101446GA SA312101448GA SA312101449GA SA312101450GA SA312101451GA SA312101452GA SA312101453GA SA312101454GR SA312101455GA SA312101456GA SA312101457GA SA312101458GR SA312111462GA SA312111469GA SA312111470GR SA312111473GA SA312111476GA SA312111482GA SA312111483GR SA312111484GR SA312111485GR SA312111486GA SA312111487GA SA312111489GR SA312111491GA SA312111493GA SA312111495GA SA312111499GA SA312111502GA SA312111504GA SA312111510GA SA312111512GA SA312111513GR SA312111514GA SA312111516GA SA312111517GA SA312111518GA SA312111519GA SA312111521GA SA312111522GA SA312111523GA SA312111524GA SA312111525GA SA312111526GA SA312111528GA SA312111529GA SA312111530GA SA312111531GA SA312111533GA SA312111534GA SA312111535GA SA312111536GA SA312111537GA SA312111538GA SA312111540GA SA312111541GR SA312111542GA SA312111543GA SA312111544GA SA312111545GA SA312111546GA SA312111548GA SA312111549GA SA312111552GA SA312111554GA SA312111556GA SA312111557GA SA312111558GA SA312111559GR SA312111560GA SA312111561GA SA312111563GA SA312111564GR SA312111565GA SA312111566GA SA312111567GA SA312111568GA SA312111569GA SA312111571GR SA312111572GA SA312111573GA SA312111575GA SA312111577GA SA312111578GA SA312111580GA SA312111581GA SA312111582GA SA312111584GA SA312111585GA SA312111586GA SA312111588GA SA312111589GR SA312111590GA SA312111629GR SA312111652GR SA312111654GR SA312111655GR SA312111656GA SA312111664GA SA312111666GA SA312111667GR SA312111669GR SA312111682GR SA312111685GR SA312111688GA SA312111690GR SA312121694GR SA312121695GR SA312121698GR SA312121700GA SA312121703GR SA312121704GR SA312121705GA SA312121706GA SA312121707GA SA312121708GA SA312121709GR SA312121711GR SA312121712GA SA312121713GR SA312121715GA SA312121716GA SA312121717GA SA312121718GA SA312121719GA SA312121721GR SA312121726GA SA312121729GA SA312121732GR SA312121739GR SA312121740GR SA312121742GR SA312121746GR SA312121751GR SA312121752GA SA312121753GA SA312121754GR SA312121755GR SA312121757GA SA312121758GA SA312121759GA SA312121762GA SA312121764GA SA312121767GA SA312121768GA SA312121771GA SA312121772GA SA312121774GA SA312121777GA SA312121778GA SA312121779GA SA312121781GR SA312121782GA SA312121783GA SA312121784GA SA312121785GA SA312121786GA SA312121787GA SA312121788GA SA312121789GA SA312121790GA SA312121792GR SA312121793GR SA312121794GR SA312121796GA SA312121797GA SA312121798GR SA312121799GA SA312121800GA SA312121802GA SA312121804GA SA312121805GA SA312121810GR SA312121811GA SA312121812GA SA312121813GA SA312121814GA SA312121817GA SA312121818GA SA312121821GA SA312121822GA SA312121827GA SA312121828GR SA312121829GA SA312121831GA SA312121833GA SA312121834GA SA312121835GR SA312121836GR SA312121837GA SA312121838GA SA312121839GA SA312121840GA SA312121841GR SA312121845GR SA312121847GA SA312121849GA SA312121851GA SA312121853GA SA312121854GR SA312121855GA SA312121856GA SA312121857GA SA312121859GA SA312121860GR SA312121861GR SA312121862GR SA312121863GR SA312121864GR SA312121867GA SA312121869GR SA312121870GA SA312121873GA SA312121874GR SA312121876GA SA312121877GA SA312121880GR SA312121881GR SA312121882GR SA312121883GR SA312121884GR SA312121885GA SA312121886GA SA312121887GA SA312121888GR SA312121889GA SA312121891GA SA312121892GA SA312121893GA SA312121894GA SA312121897GA SA312121898GR SA312121899GA SA312121902GR SA312121903GA SA312121904GA SA312121905GR SA312121906GR SA312121907GR SA312121909GR SA312121910GA SA312121911GR SA312121912GR SA312121913GA SA312121915GR SA312121917GA SA312121918GA SA312121920GA SA312121921GR SA312121923GA SA312121924GA SA312121926GA SA312121927GR SA312121928GA SA312121930GA SA312121932GA SA312121933GA SA312121934GA SA312121936GA SA312121937GA SA312121938GA SA312121939GR SA312121940GA SA312121941GA SA312121944GA SA312121946GA SA312121949GA SA312121951GA SA312121952GA SA312121953GR SA312121955GA SA312121956GA SA312121957GA SA312121958GA SA312121959GA SA312121960GA SA312121962GA SA312121963GR SA312121964GA SA312121966GA SA312121967GA SA312121973GA SA312121974GA SA312121976GA SA312121980GA SA312121981GA SA312121982GA SA312121983GA SA312121984GA SA312121986GR SA312121988GA SA312121990GA SA312121992GR SA312121994GA SA312121997GA SA312121998GR SA312122000GR SA312122001GA SA312122003GA SA312122004GR SA312122005GR SA312122007GA SA312122009GA SA312122013GA SA312122014GA SA312122016GR SA312122018GA SA312122025GR SA312122026GA SA312122028GR SA312122029GR SA312122030GA SA312122031GA SA312122036GA SA312122037GA SA312122038GA SA312122039GA SA312122040GA SA312122041GA SA312122042GA SA312122043GA SA312122048GA SA312122050GA SA312122051GR SA312122052GR SA312122054GA SA312122055GA SA312122057GA SA312122058GA SA312122060GA SA312122063GA SA312122064GA SA312122065GA SA312122067GA SA312122068GA SA312122069GR SA312122071GA SA312122076GA SA312122079GA SA312122080GA SA312122082GA SA312122083GA SA312122084GA SA312122085GA SA312122087GA SA312132089GA SA312132090GA SA312132091GA SA312132093GA SA312132095GA SA312132100GA SA312132101GR SA312132102GA SA312132103GA SA312132105GR SA312132106GR SA312132107GA SA312132109GA SA312132110GA SA312132113GR SA312132115GR SA312132116GA SA312132118GA SA312132119GA SA312132123GA SA312132124GR SA312132126GA SA312132128GR SA312132130GA SA312132131GR SA312132132GA SA312132134GA SA312132135GA SA312132136GA SA312132138GA SA312132139GA SA312132140GA SA312132142GA SA312132143GR SA312132144GR SA312132145GA SA312132146GR SA312132147GR SA312132148GA SA312132149GA SA312132150GR SA312132151GA SA312132164GA SA312132165GA SA312132166GA SA312132167GA SA312132168GA SA312132170GA SA312132172GR SA312132173GA SA312132174GA SA312132175GR SA312132176GA SA312132178GA SA312142250GA SA312152309GR SA312152310GA SA312152311GR SA312152314GA SA312152315GA SA312152317GR SA312152320GR SA312152321GR SA312152323GA SA312152324GA SA312152325GA SA312152326GA SA312152328GR SA312152331GR SA312152332GR SA312152333GR SA312152334GR SA312152336GR SA312152337GR SA312152338GA SA312152339GR SA312152340GA SA312152341GR SA312152342GA SA312152343GA SA312152348GA SA312152350GR SA312152352GR SA312152353GA SA312152354GA SA312152355GR SA312152357GA SA312152358GR SA312152359GA SA312152362GA SA312152363GR SA312152365GA SA312152366GR SA312152371GR SA312152372GR SA312152373GA SA312152374GA SA312152375GR SA312152376GR SA312152377GA SA312152378GA SA312152379GA SA312152380GA SA312152382GR SA312152386GA SA312152387GA SA312152388GA SA312152390GR SA312152391GR SA312152395GR SA312152398GA SA312152399GA SA312152400GA SA312152404GA SA312152406GA SA312152408GA SA312152409GA SA312152410GA SA312152411GA SA312152413GA SA312152414GA SA312152415GA SA312152416GA SA312152417GA SA312152418GA SA312152419GR SA312152420GA SA312152421GA SA312152422GR SA312152424GA SA312152425GA SA312152426GA SA312152427GA SA312152429GA SA312152430GA SA312152431GA SA312152432GA SA312152433GR SA312152434GA SA312152436GA SA312152439GA SA312152440GA SA312152441GA SA312152444GA SA312152445GA SA312152446GA SA312152447GA SA312152448GR SA312152449GA SA312152450GA SA312152451GR SA312152452GR SA312152453GA SA312152454GA SA312152455GA SA312152456GA SA312152457GA SA312152458GA SA312152459GA SA312152462GA SA312152463GR SA312152464GA SA312152466GA SA312152467GA SA312152468GA SA312152471GA SA312152474GA SA312152475GA SA312152476GA SA312152477GR SA312152478GA SA312152479GA SA312152486GA SA312152488GR SA312152490GA SA312152491GA SA312152492GA SA312152495GA SA312152496GA SA312152498GA SA312152499GA SA312152500GA SA312152501GA SA312152502GR SA312152503GA SA312152504GA SA312152505GA SA312152506GA SA312152507GR SA312152509GA SA312152510GA SA312152511GR SA312152512GA SA312152513GA SA312152515GA SA312152516GA SA312152517GA SA312152518GA SA312152519GA SA312152520GA SA312152521GA SA312152522GR SA312152523GA SA312152524GA SA312152525GA SA312152526GA SA312152527GA SA312152528GA SA312152529GA SA312152530GA SA312152531GA SA312152532GA SA312152533GA SA312152534GA SA312152535GA SA312152536GA SA312152537GA SA312152538GA SA312152539GA SA312152540GA SA312152545GR SA312152546GA SA312152548GR SA312152549GR SA312152551GR SA312152552GA SA312152553GA SA312152554GA SA312152555GA SA312152557GA SA312152559GA SA312152562GA SA312192565GR SA312192566GR SA312192567GR SA312192568GR SA312192569GR SA312192570GA SA312192571GR SA312192572GA SA312192573GR SA312192574GA SA312192575GR SA312192576GR SA312192577GR SA312192578GA SA312192579GR SA312192580GR SA312192581GR SA312192582GR SA312192583GR SA312192584GA SA312192585GR SA312192586GA SA312192587GA SA312192590GA SA312192591GR SA312192592GR SA312192593GR SA312192595GA SA312192596GR SA312192598GR SA312192599GA SA312192601GR SA312192602GA SA312192603GA SA312192604GA SA312192605GA SA312192606GR SA312192607GR SA312192608GR SA312192609GA SA312192610GR SA312192611GA SA312192613GR SA312192614GR SA312192615GR SA312192617GA SA312192618GR SA312192619GR SA312192621GR SA312192622GR SA312192624GR SA312192625GR SA312192626GA SA312192627GR SA312192628GR SA312192630GR SA312192631GA SA312192632GA SA312192633GA SA312192634GA SA312192635GA SA312192636GR SA312192638GR SA312192639GA SA312192640GR SA312192641GR SA312192642GR SA312192643GA SA312192645GA SA312192646GA SA312192647GR SA312192649GR SA312192650GA SA312192651GA SA312192652GA SA312192653GA SA312192654GR SA312192655GA SA312192656GA SA312192657GA SA312192658GA SA312192659GA SA312192661GA SA312192662GA SA312192663GA SA312192664GA SA312192665GA SA312192666GA SA312192667GR SA312192668GA SA312192669GA SA312192672GA SA312192673GA SA312192674GA SA312192675GA SA312192676GA SA312192677GR SA312192678GA SA312192679GA SA312192681GA SA312192682GA SA312192685GA SA312192687GR SA312192688GA SA312192689GA SA312192690GA SA312192691GA SA312192692GA SA312192693GA SA312192694GA SA312192695GR SA312192699GA SA312192700GA SA312192702GA SA312192703GA SA312192704GA SA312192705GA SA312192706GA SA312192707GA SA312192708GA SA312192709GA SA312192710GA SA312192711GA SA312192712GA SA312192713GA SA312192715GA SA312192716GA SA312192718GA SA312192719GA SA312192720GA SA312192721GA SA312192722GA SA312192724GA SA312192725GA SA312192726GA SA312192727GA
SA312192728GA SA312192729GA SA312192731GA SA312232792GA SA312232793GA SA312232794GA SA312232800GA SA312232803GA SA312232804GA SA312232808GA SA312232809GA SA312232811GA SA312232812GA SA312232813GA SA312232816GA SA312232817GA SA312232818GA SA312232819GA SA312232820GA SA312232822GA SA312232824GA SA312232825GA SA312232826GA SA312232827GA SA312232831GA SA312232832GA SA312232834GA SA312232835GA SA312232837GA SA312232839GA SA312232842GA SA312232843GA SA312232845GA SA312232847GA SA312232848GR SA312232852GA SA312232853GA SA312232854GA SA312232855GA SA312232856GA SA312232857GA SA312232858GA SA312232861GA SA312232862GA SA312232863GA SA312232864GA SA312232866GA SA312232867GA SA312232871GA SA312232872GA SA312232874GA SA312232875GA SA312232876GA SA312232877GA SA312232879GA SA312232880GA SA312232881GA SA312232882GA SA312232883GA SA312232886GA SA312232888GA SA312232892GA SA312232895GA SA312232897GA SA312232898GA SA312232899GA SA312232900GA SA312232901GA SA312232902GA SA312232904GA SA312232905GA SA312232907GA SA312232910GA SA312242953GA SA312242954GA SA312242955GR SA312242956GR SA312242957GA SA312242958GA SA312242959GA SA312242960GA SA312242962GA SA312242963GA SA312242964GR SA312242965GR SA312242966GA SA312242967GA SA312242969GA SA312242970GA SA312242971GA SA312242972GR SA312242974GR SA312242975GR SA312242976GA SA312242977GR SA312242978GR SA312242979GA SA312242980GA SA312242981GA SA312242982GA SA312242986GA SA312242987GA SA312242988GA SA312242989GA SA312242990GA SA312242991GA SA312242992GA SA312242994GA SA312242997GR SA312242998GA SA312242999GA SA312243000GR SA312243001GA SA312243002GR SA312243003GA SA312243004GA SA312243005GA SA312243007GA SA312243008GA SA312243009GA SA312243010GA SA312243011GA SA312243012GA SA312243013GA SA312243014GA SA312243015GA SA312243017GA SA312243018GA SA312243019GA SA312243020GA SA312243021GA SA312243022GA SA312243023GA SA312243024GA SA312243025GA SA312243026GA SA312243027GA SA312243028GA SA312243029GA SA312243030GA SA312243031GA SA312243032GA SA312243033GR SA312243034GA SA312243036GA SA312243037GA SA312243038GA SA312243039GA SA312243040GA SA312243041GA SA312243042GA SA312243044GA SA312243045GA SA312243046GA SA312243047GA SA312243048GA SA312243049GA SA312243050GA SA312243051GA SA312243052GA SA312243053GA SA312243054GA SA312243056GA SA312243057GA SA312253060GA SA312253061GA SA312253062GA SA312253063GA SA312253064GA SA312253065GA SA312253067GA SA312253068GA SA312253070GA SA312253072GA SA312253074GA SA312253075GA SA312253076GA SA312253079GA SA312253080GA SA312253081GA SA312253084GA SA312253085GA SA312253086GA SA312253087GA SA312253088GA SA312253089GA SA312253091GA SA312253092GA SA312253093GA SA312253094GA SA312253095GA SA312253096GA SA312253097GA SA312253098GA SA312253099GA SA312253100GA SA312253101GA SA312253103GA SA312253104GR SA312253106GA SA312253107GR SA312253108GA SA312253109GA SA312253110GA SA312253111GA SA312253112GA SA312253113GA SA312253114GA SA312253115GA SA312253118GA SA312253122GA SA312253123GR SA312253124GA SA312253125GA SA312253126GA SA312253127GA SA312253128GA SA312253129GA SA312253130GA SA312253132GA SA312253133GR SA312253135GA SA312253136GA SA312253137GA SA312253138GR SA312253139GA SA312253141GA SA312253144GR SA312253146GA SA312253147GR SA312253148GA SA312253149GA SA312253151GA SA312253152GA SA312253153GA SA312253154GA SA312253155GA SA312253156GR SA312253158GA SA312253159GR SA312253160GA SA312253164GA SA312253165GA SA312253166GA SA312253167GA SA312253168GA SA312253169GA SA312253170GA SA312253171GA SA312253172GR SA312253173GA SA312253174GA SA312253175GA SA312253176GA SA312253177GA SA312253178GA SA312253179GA SA312253180GA SA312253181GA SA312253182GA SA312253183GR SA312253186GA SA312253187GA SA312253188GA SA312253190GA SA312253191GR SA312253194GR SA312253195GR SA312253196GA SA312253198GA SA312253200GA SA312253201GA SA312253207GR SA312253209GA SA312253212GA SA312253216GA SA312253224GA SA312253225GA SA312253226GA SA312253232GA SA312253234GA SA312253235GA SA312253236GA SA312253241GA SA312253244GA SA312253245GA SA312253248GA SA312253251GA SA312253252GA SA312253254GA SA312253258GA SA312253259GA SA312253260GA SA312253261GA SA312253263GR SA312253265GA SA312253267GA SA312253268GA SA312253269GR SA312253270GA SA312253274GA SA312253275GA SA312253276GR SA312253277GA SA312253278GR SA312253280GA SA312253281GA SA312253283GA SA312253285GA SA312253286GA SA312253288GA SA312253289GA SA312253290GA SA312253291GR SA312253292GA SA312253300GR SA312253302GR SA312253303GA SA312253304GA SA312253305GA SA312253306GA SA312253309GA SA312253310GA SA312253314GR SA312253316GA SA312253317GR SA312253319GA SA312253320GA SA312253321GA SA312253322GA SA312253323GA SA312253324GA SA312253326GA SA312253327GR SA312253328GA SA312253329GA SA312253330GR SA312253331GA SA312253332GR SA312253334GA SA312253335GA SA312253336GA SA312253337GA SA312253338GA SA312253340GA SA312253342GA SA312253343GR SA312253344GA SA312253345GA SA312253346GA SA312253347GA SA312253348GA SA312253349GR SA312253350GA SA312253352GR SA312253353GA SA312253354GA SA312253355GA SA312253356GA SA312253357GA SA312253358GA SA312253359GA SA312253360GA SA312253361GA SA312253365GA SA312253366GA SA312253370GA SA312253371GA SA312253372GA SA312253373GA SA312253374GA SA312253375GA SA312253376GA SA312253377GA SA312253378GA SA312253380GA SA312253382GA SA312253383GA SA312253385GR SA312253386GA SA312253388GA SA312253390GA SA312253391GA SA312253394GA SA312253396GR SA312253398GA SA312253400GA SA312253401GR SA312253402GA SA312253404GA SA312253407GA SA312253408GA SA312253409GA SA312253410GA SA312253412GA SA312253414GA SA312253415GA SA312253418GA SA312253419GA SA312253420GA SA312253421GA SA312253424GA SA312253433GA SA312253434GA SA312253435GA SA312253436GA SA312253437GA SA312253438GA SA312253439GA SA312253442GA SA312253443GA SA312253444GA SA312253447GA SA312253449GA SA312253450GA SA312253461GA SA312253464GA SA312253466GA SA312253468GA SA312253470GA SA312253472GA SA312253474GA SA312253477GA SA312253478GR SA312253479GA SA312253481GA SA312253482GA SA312253483GA SA312253484GA SA312253485GA SA312253486GA SA312253489GA SA312253490GA SA312253491GA SA312253492GA SA312253494GA SA312253495GA SA312253496GA SA312253497GA SA312253498GA SA312253500GA SA312253501GA SA312253502GA SA312253503GA SA312253504GA SA312253506GA SA312253507GR SA312253508GA SA312253509GA SA312253510GA SA312253511GA SA312253514GA SA312253515GA SA312253518GA SA312253520GA SA312253523GA SA312253525GA SA312253526GA SA312253527GA SA312253528GA SA312253529GA SA312253530GA SA312253531GA SA312253532GA SA312253533GA SA312253534GA SA312253535GA SA312253536GA SA312253537GA SA312253538GA SA312253539GA SA312253540GA SA312253541GA SA312253542GA SA312253543GA SA312253544GA SA312253545GA SA312253546GA SA312253547GA SA312253548GA SA312253549GA SA312253550GA SA312253551GA SA312253552GA SA312253553GA SA312253554GA SA312253555GA SA312253556GA SA312253557GA SA312253558GA SA312253560GR SA312253561GA SA312253562GA SA312253564GA SA312253565GA SA312253566GA SA312253567GA SA312253568GA SA312253569GA SA312253570GA SA312253571GA SA312253572GA SA312253573GA SA312253574GA SA312253575GA SA312253576GA SA312253577GA SA312253578GA SA312253579GA SA312253580GA SA312253581GA SA312253582GA SA312253583GA SA312253584GA SA312253585GA SA312253586GA SA312253587GA SA312253588GA SA312253590GA SA312253592GA SA312253595GR SA312253596GA SA312253597GA SA312253598GA SA312253599GA SA312253600GA SA312253601GA SA312253602GA SA312253603GA SA312253605GA SA312253606GA SA312263613GA SA312263615GA SA312263616GA SA312263618GA SA312263619GA SA312263620GA SA312263621GA SA312263622GA SA312263623GA SA312263624GA SA312263625GA SA312263627GA SA312263628GA SA312263629GA SA312263631GA SA312263634GA SA312263635GA SA312263638GA SA312263640GA SA312263641GA SA312263642GA SA312263643GA SA312263644GR SA312263645GA SA312263646GA SA312263648GA SA312263649GA SA312263650GA SA312263651GA SA312263652GA SA312263653GA SA312263654GA SA312263656GA SA312263657GA SA312263658GA SA312263659GR SA312263660GA SA312263661GA SA312263662GA SA312263663GA SA312263664GA SA312263666GA SA312263667GA SA312263669GA SA312263670GA SA312263672GA SA312263673GA SA312263674GA SA312263675GA SA312263676GA SA312263678GA SA312263679GA SA312263680GA SA312263681GA SA312263682GA SA312263684GA SA312263685GA SA312263687GA SA312263688GA SA312263689GA SA312263690GA SA312263691GA SA312263692GA SA312263694GA SA312263695GA SA312263696GR SA312263697GA SA312263698GA SA312263699GA SA312263701GA SA312263703GA SA312263704GA SA312263706GA SA312263707GA SA312263708GA SA312263709GA SA312263710GR SA312263711GA SA312263712GA SA312263713GA SA312263714GA SA312263715GA SA312263716GA SA312263717GA SA312263718GA SA312263719GA SA312263720GA SA312263722GA SA312263723GA SA312263724GA SA312263725GA SA312263726GA SA312263728GA SA312263729GA SA312263730GA SA312263731GA SA312263732GA SA312263733GA SA312263734GA SA312263735GR SA312263736GA SA312263738GA SA312263739GA SA312263740GA SA312263741GA SA312263742GA SA312263743GA SA312263744GA SA312263745GA SA312263746GA SA312263747GA SA312263748GA SA312263749GA SA312263750GA SA312263752GR SA312263753GA SA312263754GA SA312263755GA SA312263756GA SA312263757GA SA312263758GA SA312263759GA SA312263760GA SA312263761GA SA312263762GA SA312263764GA SA312263767GA SA312263768GA SA312263769GA SA312263770GA SA312263771GA SA312263772GA SA312263774GA SA312263775GA SA312263776GA SA312263777GA SA312263778GA SA312263779GA SA312263780GA SA312263781GA SA312263782GA SA312263783GA SA312263784GA SA312263785GA SA312263786GA SA312263787GA SA312263789GA SA312263790GA SA312263792GA SA312263794GA SA312263795GA SA312263797GA SA312263798GA SA312263799GA SA312263800GA SA312263802GA SA312263804GA SA312263805GA SA312263806GA SA312263807GA SA312263808GR SA312263809GA SA312263810GA SA312263812GA SA312263813GA SA312263816GA SA312263817GA SA312263821GA SA312263822GA SA312263828GA SA312263829GA SA312263830GA SA312263831GA SA312263832GA SA312263833GA SA312263834GA SA312263836GA SA312263837GA SA312263838GA SA312263839GA SA312263840GA SA312263841GA SA312263842GA SA312263843GR SA312263844GA SA312263845GA SA312263846GA SA312263847GA SA312263849GA SA312263850GA SA312263851GA SA312263852GA SA312263854GA SA312263855GA SA312263856GA SA312263857GA SA312263858GA SA312263859GA SA312263860GA SA312263861GA SA312263862GA SA312263863GA SA312263864GA SA312263865GA SA312263867GA SA312263868GA SA312263869GA SA312263870GA SA312263871GA SA312263872GA SA312263873GA SA312263875GA SA312263876GA SA312263877GA SA312263878GA SA312263879GA SA312263880GA SA312263881GA SA312263882GA SA312263883GA SA312263884GA SA312263885GA SA312263886GA SA312263888GA SA312263889GA SA312263890GA SA312263892GA SA312263893GA SA312263894GA SA312263895GA SA312263896GA SA312263897GA SA312263899GA SA312263900GA SA312263901GA SA312263902GA SA312263903GA SA312263906GA SA312263907GA SA312263908GR SA312263912GA SA312263913GA SA312263914GR SA312263915GA SA312263916GA SA312263917GA SA312263918GA SA312263919GA SA312263920GR SA312263923GR SA312263924GA SA312263927GA SA312263930GA SA312263935GR SA312263951GA SA312263952GA SA312263953GR SA312263954GA SA312263955GA SA312263956GA SA312263958GA SA312263959GA SA312263960GA SA312263961GA SA312263962GA SA312263963GA SA312263964GA SA312263965GA SA312263967GA
SA312263968GA SA312263969GA SA312263970GA SA312263971GA SA312263972GA SA312263974GA SA312263977GA SA312263983GA SA312263990GR SA312264000GR SA312264002GR SA312264005GA SA312264007GA SA312264010GA SA312264014GA SA312264015GA SA312264016GA SA312264017GR SA312264018GA SA312264019GA SA312264020GA SA312264022GA SA312264024GA SA312264025GA SA312264027GA SA312264028GA SA312264029GA SA312264031GA SA312264033GA SA312264034GA SA312264035GA SA312264037GA SA312264040GA SA312264042GA SA312264044GA SA312264045GA SA312264046GA SA312264048GA SA312264049GA SA312264050GA SA312264051GA SA312264052GA SA312264053GA SA312264055GA SA312264056GA SA312264058GA SA312264059GA SA312264061GA SA312264062GA SA312264063GA SA312264064GA SA312264065GA SA312264071GA SA312264087GR SA312264089GA SA312264093GA SA312264094GA SA312264095GA SA312264096GA SA312264100GA SA312284116GA SA312284123GA SA312284124GA SA312284126GA SA312284130GA SA312284132GA SA312284139GR SA312284155GA SA312284159GA SA312284165GR SA312284179GA SA312284180GA SA312284182GA SA312284183GA SA312284217GA SA312284219GA SA312284220GA SA312284222GA SA312284223GA SA312284226GA SA312284227GA SA312284228GA SA312284229GA SA312284234GA SA312284236GA SA312284237GA SA312284238GA SA312284239GA SA312284240GA SA312284241GA SA312284242GA SA312284243GA SA312284245GA SA312284246GA SA312284247GA SA312284248GA SA312284249GA SA312284250GA SA312284251GA SA312284252GA SA312284253GA SA312284254GA SA312284255GA SA312284256GA SA312284257GA SA312284258GA SA312284259GA SA312284261GA SA312284263GA SA312284264GA SA312284265GA SA312284266GA SA312284267GA SA312284268GA SA312284269GR SA312284270GA SA312284272GA SA312284273GA SA312284274GA SA312284276GA SA312284278GR SA312284279GA SA312284284GR SA312284285GR SA312284292GR SA312284297GR SA312284298GA SA312284299GA SA312284301GR SA312284302GA SA312284303GR SA312284304GR SA312284305GA SA312284313GR SA312284314GR SA312284315GR SA312284316GA SA312284317GR SA312284318GR SA312284319GR SA312284320GR SA312284321GR SA312284322GR SA312284324GR SA312284325GR SA312284326GA SA312284327GR SA312284328GA SA312284329GA SA312284330GR SA312284331GR SA312284332GR SA312284333GR SA312284334GR SA312284335GA SA312284338GR SA312284339GR SA312284340GR SA312284341GR SA312284342GR SA312294343GA SA312294344GR SA312294345GA SA312294348GA SA312294349GR SA312294350GR SA312294353GR SA312294354GR SA312294355GR SA312294356GA SA312294357GA SA312294360GA SA312294361GA SA312294362GA SA312294363GA SA312294365GA SA312294367GA SA312294368GA SA312294369GA SA312294371GA SA312294372GA SA312294375GA SA312294376GA SA312294377GR SA312294378GA SA312294379GA SA312294381GR SA312294382GA SA312294384GA SA312294385GR SA312294386GA SA312294388GR SA312294390GR SA312294391GR SA312294392GR SA312294393GA SA312294394GA SA312294397GR SA312294399GR SA312294400GA SA312294402GA SA312294404GA SA312294405GR SA312294406GA SA312294407GR SA312294408GR SA312294409GR SA312294410GA SA312294412GR SA312294413GA SA312294414GR SA312294416GR SA312294417GA SA312294420GR SA312294422GR SA312294424GR SA312294425GR SA312294426GA SA312294427GA SA312294428GA SA312294429GR SA312294430GA SA312294431GA SA312294432GA SA312294433GR SA312294436GA SA312294438GR SA312294439GA SA312294440GR SA312294443GR SA312294445GA SA312294447GA SA312294448GA SA312294449GA SA312294450GR SA312294451GR SA312294454GR SA312294456GR SA312294460GA SA312294461GR SA312294463GR SA312294464GR SA312294467GR SA312294471GR SA312294473GR SA312294475GA SA312294476GA SA312294477GA SA312294478GA SA312294481GA SA312294482GR SA312294483GA SA312294484GA SA312294485GR SA312294487GA SA312294488GA SA312294489GA SA312294490GA SA312294491GR SA312294492GA SA312294494GA SA312294496GA SA312294497GA SA312294499GA SA312294500GA SA312294501GA SA312294502GA SA312294503GA SA312294505GA SA312294506GA SA312294507GA SA312294510GA SA312294517GR SA312294518GA SA312294519GA SA312294520GA SA312294522GA SA312294524GA SA312294525GA SA312294527GA SA312294529GA SA312294535GA SA312294544GA SA312294546GA SA312294548GA SA312294551GR SA312294556GR SA312294560GA SA312294566GA SA312294568GA SA312294571GA SA312294578GA SA312294579GA SA312294581GA SA312294582GA SA312294584GA SA312294586GA SA312294587GA SA312294588GA SA312294589GA SA312294590GA SA312294591GA SA312294592GA SA312294593GA SA312294594GA SA312294595GA SA312294597GA SA312294598GA SA312294600GA SA312294601GA SA312294602GA SA312294603GA SA312294604GA SA312294605GA SA312294606GA SA312294607GA SA312294608GA SA312294609GA SA312294612GA SA312294613GA SA312294614GR SA312294616GR SA312294617GA SA312294618GA SA312294621GA SA312294623GA SA312294625GA SA312294626GA SA312294630GA SA312294632GA SA312294634GA SA312294635GA SA312294641GR SA312294642GA SA312294644GR SA312294646GA SA312294657GA SA312294658GA SA312294660GA SA312294661GA SA312294662GA SA312294663GA SA312294665GR SA312294666GA SA312294667GA SA312294669GA SA312294670GA SA312294673GA SA312294674GR SA312294675GA SA312294677GR SA312294678GA SA312294679GA SA312294680GA SA312294681GA SA312294682GA SA312294683GA SA312294684GA SA312294685GA SA312294686GA SA312294687GA SA312294688GA SA312294689GA SA312294690GA SA312294691GA SA312294692GA SA312294693GA SA312294694GA SA312294695GA SA312294697GA SA312294698GA SA312294699GA SA312294700GA SA312294701GA SA312294702GA SA312294703GA SA312294704GA SA312294705GA SA312294706GA SA312294707GA SA312294708GA SA312294709GA SA312294711GA SA312294712GA SA312294713GA SA312294714GA SA312294715GA SA312294716GA SA312294717GA SA312294718GA SA312294720GA SA312294721GA SA312294722GA SA312294723GA SA312294724GA SA312294725GA SA312294726GA SA312294727GA SA312294728GA SA312294729GA SA312294730GA SA312294731GA SA312294732GA SA312294753GA SA312294754GA SA312294755GA SA312294756GA SA312294757GA SA312294758GA SA312294759GA SA312294760GA SA312294762GA SA312294763GA SA312294764GA SA312294765GA SA312294766GA SA312294767GA SA312294768GA SA312294769GA SA312294770GA SA312294771GA SA312294772GA SA312294773GA SA312294775GA SA312294776GA SA312294777GA SA312294778GA SA312294779GA SA312294780GA SA312294781GA SA312294782GA SA312294783GA SA312294784GA SA312294785GA SA312294786GA SA312294787GA SA312294788GA SA312294789GA SA312294790GA SA312294791GA SA312314793GA SA312314794GA SA312314795GA SA312314796GA SA312314797GA SA312314798GA SA312314799GA SA312314800GA SA312314801GA SA312314802GA SA312314803GA SA312314804GA SA312314805GA SA312314806GA SA312314807GA SA312314808GA SA312314809GA SA312314810GA SA312314811GA SA312314812GA SA312314813GA SA312314814GA SA312314815GA SA312314816GA SA312314817GA SA312314818GA SA312314819GA SA312314820GA SA312314821GA SA312314822GR SA312314823GR SA312314824GA SA312314825GA SA312314826GA SA312314827GA SA312314828GA SA312324832GR SA312344847GA SA312344852GR SA312344868GA SA312344881GA SA312344908GA SA312344919GA SA312344922GA SA312344936GA SA312344940GA SA312344941GA SA312344942GA SA312344945GA SA312344946GA SA312344954GA SA312344955GA SA312344959GA SA312344965GA SA312344968GA SA312344970GA SA312344973GA SA312344976GA SA312344988GA SA312344989GA SA312344995GA SA312345002GA SA312345004GA SA312345007GA SA312345012GA SA312345013GA SA312345014GA SA312345015GA SA312345020GA SA312345028GA SA312345034GA SA312345036GA SA312345038GA SA312345040GA SA312345042GA SA312345044GA SA312345045GA SA312345047GA SA312345048GA SA312345050GA SA312345052GA SA312345054GR SA312345055GA SA312345057GA SA312345058GA SA312345060GA SA312345062GA SA312345065GA SA312345067GA SA312345069GA SA312345070GA SA312345071GA SA312345076GR SA312345078GA SA312345081GA SA312345085GA SA312345087GA SA312345090GA SA312345092GA SA312345100GA SA312345105GA SA312345106GA SA312345108GA SA312345109GR SA312345116GA SA312345117GA SA312345119GA SA312345121GA SA312345122GA SA312345123GA SA312345124GA SA312345125GA SA312345126GA SA312345127GA SA312345128GA SA312345129GA SA312345132GA SA312345133GA SA312345134GA SA312345140GA SA312345144GA SA312345147GA SA312345151GA SA312345158GA SA312345168GA SA312345169GA SA312345170GA SA312345176GA SA312345189GA SA312345190GA SA312345193GA SA312355195GA SA312355196GA SA312355197GA SA312355199GA SA312355201GA SA312355202GA SA312355204GA SA312355205GA SA312355207GA SA312355210GA SA312355211GA SA312355212GA SA312355213GA SA312355214GA SA312355215GA SA312355216GA SA312355218GA SA312355220GA SA312355221GA SA312355223GA SA312355224GA SA312355225GA SA312355227GA SA312355231GA SA312355234GA SA312355235GA SA312355241GA SA312355318GA SA312355319GA SA312355322GA SA312355324GA SA312355327GA SA312355329GA SA312355330GA SA312355331GA SA312355332GA SA312355335GA SA312355336GA SA312355339GA SA312355342GA SA312355343GA SA312355344GA SA312355375GA SA312355376GA SA312355377GA SA312355378GA SA312355379GA SA312355381GA SA312355382GA SA312355383GA SA312355384GA SA312355387GA SA312355388GA SA312355389GA SA312355390GA SA312355391GA SA312355392GA SA312355393GA SA312355394GA SA312355396GA SA312355397GA SA312355398GA SA312355399GA SA312355400GA SA312355401GA SA312355402GA SA312355403GA SA312355404GA SA312355406GA SA312355407GA SA312355408GA SA312355409GA SA312355414GA SA312355442GA SA312355447GA SA312355448GA SA312355449GA SA312355451GR SA312355453GA SA312355456GA SA312355459GA SA312355460GA SA312355461GA SA312355462GA SA312355463GA SA312355468GR SA312355470GA SA312355473GR SA312355477GA SA312355502GA SA312355503GA SA312355512GA SA312355517GR SA312355530GA SA312365531GA SA312365573GA SA312365589GA SA312365595GA SA312365598GA SA312365602GA SA312365605GA SA312365618GA SA312365621GA SA312365625GA SA312365628GA SA312365634GA SA312365647GA SA312365648GA SA312365650GA SA312365653GA SA312365655GA SA312365663GA SA312365664GA SA312365670GA SA312365675GA SA312365678GA SA312365680GA SA312365711GA SA312365713GA SA312365716GA SA312365718GA SA312365719GA SA312365720GA SA312365721GA SA312365722GA SA312365723GA SA312365725GA SA312365726GA SA312365727GA SA312365728GA SA312365729GA SA312365730GA SA312365731GA SA312365734GA SA312365735GR SA312365736GA SA312365737GA SA312365738GA SA312365740GA SA312365741GA SA312365742GA SA312365743GA SA312365744GA SA312365745GA SA312365749GA SA312365751GA SA312365758GA SA312365759GA SA312365760GA SA312365762GA SA312365763GA SA312365765GA SA312365774GA SA312365790GA SA312365802GA SA312365804GA SA312365810GA SA312365811GA SA312365813GA SA312365831GA SA312365832GA SA312365833GA SA312365835GA SA312365837GA SA312365838GA SA312365839GA SA312365841GA SA312365842GA SA312365844GA SA312365845GA SA312365847GA SA312365848GA SA312365849GA SA312365850GA SA312365855GA SA312375867GA SA312375869GA SA312375874GA SA312375876GA SA312375884GA SA312375886GA SA312375887GA SA312375888GA SA312375890GA SA312375895GA SA312375898GA SA312375900GA SA312375901GA SA312375904GA SA312375907GA SA312375913GA SA312375916GA SA312375920GA SA312375923GA SA312375924GA SA312375926GA SA312375928GA SA312375930GA SA312375932GA SA312375935GA SA312375937GA SA312375938GA SA312375939GA SA312375940GA SA312375941GA SA312375942GA SA312375944GA SA312375947GA SA312375950GA SA312375953GA SA312375954GA SA312375962GA SA312375972GA SA312375975GA SA312375981GA SA312375982GA SA312375987GA SA312375988GA SA312375989GA SA312375990GA SA312375991GA SA312375993GA SA312375994GA SA312375996GA SA312375998GA SA312376000GA SA312376001GA SA312376003GA SA312376004GA
SA312376008GA SA312376010GA SA312376011GA SA312376013GA SA312376016GA SA312376018GA SA312376019GA SA312376020GA SA312376021GA SA312376022GA SA312376024GA SA312376026GA SA312376027GA SA312376029GA SA312376030GA SA312376031GA SA312376032GA SA312376033GA SA312376034GA SA312376035GA SA312376036GA SA312376038GA SA312376039GA SA312376040GA SA312376041GA SA312376042GA SA312376043GA SA312376044GA SA312376045GA SA312376046GA SA312376047GA SA312376048GA SA312376049GA SA312376051GA SA312376053GA SA312376056GA SA312376057GA SA312376059GA SA312376060GA SA312376061GA SA312376062GA SA312376063GA SA312376064GA SA312376065GA SA312376066GA SA312376067GA SA312376068GA SA312376069GA SA312376070GA SA312376072GA SA312376074GA SA312376083GA SA312376085GA SA312376086GA SA312376087GA SA312376089GA SA312376102GA SA312376103GA SA312376106GA SA312376116GA SA312376120GA SA312376122GA SA312376123GR SA312376125GA SA312376126GA SA312376128GA SA312376129GA SA312376130GA SA312376131GA SA312376132GA SA312376133GA SA312376134GA SA312376135GA SA312376136GA SA312376137GA SA312376139GR SA312376140GA SA312376141GA SA312376142GA SA312376143GA SA312376145GA SA312376146GA SA312376147GA SA312376149GA SA312376150GR SA312376151GA SA312376152GA SA312376153GA SA312376154GA SA312376155GA SA312376156GA SA312376157GA SA312376158GA SA312376159GA SA312376160GA SA312376161GA SA312376164GA SA312376165GA SA312376167GA SA312376169GA SA312376170GA SA312376171GA SA312376172GA SA312376173GA SA312376174GA SA312376175GA SA312376176GA SA312376177GA SA312376179GA SA312376181GA SA312376182GA SA312376183GA SA312376184GA SA312376185GA SA312376186GA SA312376187GA SA312376188GA SA312376189GA SA312376192GA SA312376193GA SA312376195GA SA312376196GA SA312376197GA SA312376198GA SA312376201GA SA312376202GA SA312376204GA SA312376205GA SA312376206GA SA312376207GA SA312376208GA SA312376209GA SA312376210GA SA312376211GA SA312376212GA SA312376214GA SA312376215GA SA312376216GA SA312376217GA SA312376219GA SA312376220GA SA312376221GA SA312376222GA SA312376223GA SA312376224GA SA312376225GA SA312376226GA SA312376228GA SA312376229GA SA312376230GA SA312376233GA SA312376234GA SA312376235GA SA312376236GA SA312376237GA SA312376239GA SA312376240GA SA312376241GA SA312376242GA SA312376243GA SA312376244GA SA312376245GA SA312376246GA SA312376247GA SA312376248GA SA312376249GA SA312376250GA SA312376281GA SA312376287GA SA312376292GA SA312376295GA SA312376297GA SA312376300GA SA312378006GA SA312378008GA SA312378009GA SA312378019GA SA312378021GA SA312378024GR SA312378029GA SA312378030GA SA312378032GA SA312378033GA SA312378034GA SA312378036GA SA312378038GA SA312378040GA SA312378043GA SA312378048GA SA312378051GA SA312378059GA SA312378063GA SA312386311GA SA312386326GA SA312386330GA SA312386331GA SA312386333GA SA312386337GR SA312386347GA SA312386350GA SA312386353GA SA312386356GA SA312386359GA SA312386360GA SA312386361GA SA312386362GA SA312386363GA SA312386364GA SA312386365GA SA312386367GA SA312386368GA SA312386369GA SA312386372GA SA312386373GA SA312386376GA SA312386377GA SA312386378GA SA312386380GA SA312386381GA SA312386382GA SA312386383GA SA312386384GA SA312386385GA SA312386387GA SA312386388GA SA312386389GA SA312386390GA SA312386393GA SA312386394GA SA312386396GA SA312386399GA SA312386400GA SA312386401GA SA312386405GA SA312386406GA SA312386407GA SA312386408GA SA312386411GA SA312386413GA SA312386414GA SA312386415GA SA312386416GA SA312386417GA SA312386418GA SA312386419GA SA312386420GA SA312386421GA SA312386423GA SA312386425GA SA312386428GA SA312386431GA SA312386433GA SA312386434GA SA312386435GA SA312386436GA SA312386437GA SA312386438GA SA312386439GA SA312386440GA SA312386441GA SA312386442GA SA312386443GA SA312386444GA SA312386445GA SA312386446GA SA312386447GA SA312386448GA SA312386449GA SA312386450GA SA312386451GA SA312386452GA SA312386453GA SA312386454GA SA312386455GA SA312386456GA SA312386458GA SA312386459GA SA312386460GA SA312386461GA SA312386463GA SA312386464GA SA312386465GA SA312386466GA SA312386468GA SA312386470GA SA312386471GA SA312386472GA SA312386473GA SA312386474GA SA312386475GA SA312386476GA SA312386478GA SA312386479GR SA312386484GA SA312386485GA SA312386486GA SA312386487GA SA312386488GA SA312386490GA SA312386493GA SA312386495GA SA312386498GA SA312386499GA SA312386500GA SA312386504GA SA312386507GA SA312386508GA SA312386510GA SA312386512GA SA312386513GA SA312386514GA SA312386516GA SA312386518GA SA312386520GA SA312386522GA SA312386524GA SA312386525GA SA312386527GA SA312386528GA SA312386531GA SA312386533GA SA312386536GA SA312386537GA SA312386538GA SA312386539GA SA312386540GA SA312386542GA SA312386543GA SA312386544GA SA312386545GA SA312386546GA SA312386548GA SA312386549GA SA312386550GA SA312386551GA SA312386553GA SA312386554GA SA312386555GA SA312386556GA SA312386558GR SA312386559GA SA312386560GA SA312386563GR SA312386564GA SA312386565GA SA312386566GA SA312386568GA SA312386569GA SA312386570GA SA312386571GA SA312386572GA SA312386575GA SA312386577GA SA312386578GA SA312386579GA SA312386581GA SA312386582GA SA312386583GA SA312386584GA SA312386585GA SA312386586GA SA312386587GA SA312386588GA SA312386589GA SA312386591GA SA312386592GA SA312386594GA SA312386595GA SA312386596GA SA312386597GA SA312386598GA SA312386599GA SA312386601GA SA312386602GA SA312386603GA SA312386604GA SA312386605GA SA312386606GA SA312386607GA SA312386608GA SA312386609GA SA312386610GA SA312386613GA SA312386614GA SA312386615GA SA312386617GA SA312386620GA SA312386622GA SA312386623GA SA312386625GA SA312386626GA SA312386627GA SA312386628GA SA312386629GA SA312386630GA SA312386631GA SA312386633GA SA312386637GA SA312386641GA SA312386642GA SA312386644GA SA312386646GA SA312386647GA SA312386648GA SA312386649GA SA312386650GA SA312386652GA SA312386653GA SA312386654GA SA312386655GA SA312386656GA SA312386657GA SA312386658GA SA312386659GA SA312386660GA SA312386662GA SA312386663GA SA312386664GA SA312386666GA SA312386667GA SA312386668GA SA312386669GA SA312386670GA SA312386671GA SA312386673GA SA312386676GA SA312386677GA SA312386678GA SA312386680GA SA312386681GA SA312386682GA SA312386683GA SA312386684GA SA312386686GA SA312386687GA SA312386689GA SA312386692GA SA312386693GA SA312386694GA SA312386695GA SA312386696GA SA312386700GA SA312396701GA SA312396702GA SA312396703GA SA312396704GA SA312396707GA SA312396708GA SA312396709GA SA312396710GA SA312396711GA SA312396712GA SA312396713GA SA312396714GA SA312396715GA SA312396716GA SA312396717GA SA312396718GA SA312396719GA SA312396720GA SA312396721GA SA312396723GA SA312396724GA SA312396725GA SA312396726GA SA312396727GA SA312396728GA SA312396729GA SA312396730GA SA312396731GA SA312396732GA SA312396733GA SA312396734GA SA312396735GA SA312396737GA SA312396738GA SA312396739GA SA312396741GA SA312396742GA SA312396743GA SA312396744GA SA312396745GA SA312396746GA SA312396747GA SA312396748GA SA312396749GA SA312396750GA SA312396751GA SA312396752GA SA312396753GA SA312396754GA SA312396755GA SA312396756GA SA312396757GA SA312396758GA SA312396759GA SA312396760GA SA312396761GA SA312396762GA SA312396763GA SA312396764GA SA312396765GA SA312396767GA SA312396768GA SA312396769GA SA312396770GA SA312396771GA SA312396772GA SA312396773GA SA312396774GA SA312396776GA SA312396777GA SA312396778GA SA312396779GA SA312396780GA SA312396782GA SA312396783GA SA312396784GA SA312396785GA SA312396786GA SA312396788GA SA312396789GA SA312396790GA SA312396791GA SA312396792GA SA312396793GA SA312396794GA SA312396795GA SA312396796GA SA312396797GA SA312396798GA SA312396799GA SA312396801GA SA312396802GA SA312396803GA SA312396804GA SA312396806GA SA312396808GA SA312396809GA SA312396811GA SA312396812GA SA312396813GA SA312396814GA SA312396815GA SA312396816GA SA312396817GA SA312396819GA SA312396820GA SA312396821GA SA312396822GA SA312396825GA SA312396826GA SA312396828GA SA312396829GA SA312396830GA SA312396831GA SA312396832GA SA312396833GA SA312396835GA SA312396836GA SA312396838GA SA312396839GA SA312396840GA SA312396841GA SA312396843GA SA312396844GA SA312396845GA SA312396846GA SA312396847GA SA312396848GA SA312396849GA SA312396850GA SA312396851GA SA312396852GA SA312396854GA SA312396855GA SA312396862GA SA312396863GA SA312396865GA SA312396866GA SA312396867GA SA312396869GA SA312396872GA SA312396874GA SA312396875GA SA312396876GA SA312396877GA SA312396878GA SA312396879GA SA312396880GA SA312396882GA SA312396885GA SA312396887GA SA312396890GA SA312396891GA SA312396893GA SA312396896GA SA312396900GA SA312396901GA SA312396906GA SA312396907GA SA312396909GA SA312396911GA SA312396913GA SA312396914GA SA312396915GA SA312396916GA SA312396918GA SA312396919GA SA312396920GA SA312396921GA SA312396922GA SA312396924GA SA312396925GA SA312396926GA SA312396927GA SA312396929GA SA312396931GA SA312396932GA SA312396933GA SA312396934GA SA312396935GA SA312396936GA SA312396937GA SA312396938GA SA312396939GA SA312396940GA SA312396942GA SA312396950GA SA312396956GA SA312396961GA SA312396963GA SA312396965GA SA312396971GA SA312396973GA SA312396974GA SA312396975GA SA312396977GA SA312396981GA SA312396983GA SA312396985GA SA312396986GA SA312396987GA SA312396988GA SA312396989GA SA312396990GA SA312396992GA SA312396993GA SA312396994GR SA312396995GA SA312396996GA SA312396997GA SA312397000GA SA312397001GA SA312397002GA SA312397003GA SA312397004GA SA312397005GA SA312397006GA SA312397007GA SA312397008GA SA312397009GA SA312397011GA SA312397012GA SA312397013GA SA312397014GR SA312397016GA SA312397017GA SA312397019GA SA312397020GA SA312397021GA SA312397022GA SA312397023GA SA312397024GA SA312397026GA SA312397027GA SA312397028GA SA312397029GA SA312397030GA SA312397039GA SA312397042GA SA312397044GA SA312397045GA SA312397054GA SA312397055GA SA312397062GA SA312397064GA SA312397065GA SA312397066GA SA312397067GA SA312397068GA SA312397069GA SA312397070GA SA312397071GA SA312397072GA SA312397075GA SA312397076GA SA312397078GA SA312397080GA SA312397082GA SA312397084GA SA312397085GA SA312397086GA SA312397089GA SA312397093GA SA312397100GA SA312397101GA SA312397103GA SA312397138GA SA312397140GA SA312397147GA SA312397155GA SA312397157GA SA312397159GA SA312397160GA SA312397161GA SA312397164GA SA312397165GA SA312397168GA SA312397169GA SA312397171GA SA312397173GA SA312397174GA SA312397175GA SA312397176GA SA312397177GA SA312397178GA SA312397180GA SA312397181GA SA312397182GA SA312397183GA SA312397184GA SA312397185GA SA312397186GA SA312397187GA SA312397191GA SA312397195GA SA312397196GA SA312397197GA SA312397201GA SA312397202GA SA312397204GA SA312397205GA SA312397206GA SA312397207GA SA312397209GA SA312397210GA SA312397224GA SA312397234GA SA312397246GA SA312397252GA SA312397254GA SA312397272GA SA312397273GA SA312407278GA SA312407281GA SA312407282GA SA312407288GA SA312407289GA SA312407290GA SA312407291GA SA312407292GA SA312407294GA SA312407296GA SA312407297GA SA312407300GA SA312417301GA SA312417313GA SA312417320GA SA312417321GA SA312417330GA SA312417334GA SA312417335GA SA312417336GA SA312417337GA SA312417339GA SA312417340GA SA312417341GA SA312417342GA SA312417343GA SA312417347GA SA312417348GA SA312417351GA SA312417355GA SA312417359GA SA312417367GA SA312417368GA SA312417369GA SA312417371GA SA312417372GA SA312417373GA SA312417375GA SA312417378GA SA312417379GA SA312417380GA SA312417384GA SA312417385GA SA312417387GA SA312417391GA SA312417412GA
SA312447419GA SA312447420GR SA312447422GA SA312447424GA SA312447425GA SA312447426GA SA312447427GA SA312447428GA SA312447431GA SA312447432GA SA312447438GA SA312447440GA SA312447441GA SA312447444GA SA312447447GA SA312447448GA SA312447450GA SA312447452GA SA312447454GA SA312447455GA SA312447456GA SA312447457GA SA312447460GA SA312447461GA SA312447462GA SA312447463GA SA312447464GA SA312447465GA SA312447467GA SA312447469GA SA312447471GA SA312447474GA SA312447475GA SA312447476GA SA312447478GA SA312447512GA SA312447513GA SA312447515GA SA312447517GA SA312447518GA SA312447521GA SA312447522GA SA312447523GA SA312447524GR SA312447526GA SA312447527GA SA312447528GA SA312447529GR SA312447530GA SA312447531GA SA312447532GA SA312447534GR SA312447537GA SA312447539GA SA312447540GA SA312447543GA SA312447544GA SA312447546GA SA312447547GR SA312447549GA SA312447551GA SA312447553GA SA312447555GR SA312447557GA SA312447560GA SA312447566GA SA312447573GA SA312447574GA SA312447576GA SA312447577GA SA312447579GA SA312457626GA SA312457657GA SA312457679GA SA312457683GA SA312457685GA SA312457692GA SA312457703GA SA312457711GA SA312457719GA SA312457740GA SA312457759GA SA312457768GA SA312457810GR SA312457812GR SA312457814GA SA312457878GA SA312457891GA SA312457910GA SA312457915GA SA312457927GA SA312457964GA SA312457968GA SA312457993GA SA312457995GA SA312458001GA SA312458007GR SA312458015GA SA312458018GA SA312458020GA SA312458039GA SA312458047GA SA312458055GA SA312458060GA SA312458073GA SA312458099GA SA312458104GA SA312458130GA SA312458139GA SA312458145GA SA312458147GA SA312458151GA SA312458156GA SA312458157GA SA312458160GA SA312458167GA SA312458168GA SA312458176GA SA312458181GA SA312458195GA SA312458226GA SA312458228GA SA312458232GA SA312458233GA SA312458246GA SA312458256GA SA312458264GA SA312458397GA SA312458433GA SA312458455GA SA312458460GA SA312458464GA SA312458467GA SA312458468GA SA312458469GA SA312468535GA SA312468536GA SA312468539GA SA312468544GA SA312468546GA SA312468554GA SA312468556GA SA312468558GA SA312468560GA SA312468561GA SA312468574GA SA312468576GA SA312468578GA SA312468580GA SA312468586GA SA312468590GA SA312468592GA SA312468593GA SA312468594GA SA312468595GA SA312468596GA SA312468597GA SA312468599GA SA312468601GA SA312468602GA SA312468603GA SA312468604GA SA312468606GA SA312468608GA SA312468609GA SA312468610GA SA312468611GA SA312468612GA SA312468614GA SA312468615GA SA312468616GA SA312468617GA SA312468618GA SA312468619GA SA312468621GA SA312468652GA SA312468653GA SA312468654GA SA312468656GA SA312468657GA SA312468664GA SA312468665GA SA312468666GA SA312468667GA SA312468668GA SA312468672GA SA312468675GA SA312468677GA SA312468678GA SA312468679GA SA312468681GA SA312468682GA SA312468683GA SA312468684GA SA312468685GA SA312468686GA SA312468687GA SA312468688GA SA312468689GA SA312468690GA SA312468693GA SA312468694GA SA312468695GA SA312468696GA SA312468698GA SA312468701GA SA312468702GA SA312468704GA SA312468707GA SA312468708GA SA312468709GA SA312468710GA SA312468711GA SA312468712GA SA312468713GA SA312468714GA SA312468715GA SA312468716GA SA312468719GA SA312468721GA SA312468722GA SA312468723GA SA312468724GA SA312468725GA SA312468726GA SA312468727GA SA312468728GA SA312468729GA SA312468730GA SA312468731GA SA312468732GA SA312468733GA SA312468734GA SA312468735GA SA312468736GA SA312468737GA SA312468738GA SA312468739GA SA312468740GA SA312468741GA SA312468785GA SA312468790GA SA312468795GA SA312468799GA SA312468804GA SA312468809GA SA312468810GA SA312468811GA SA312468823GA SA312468825GA SA312468830GA SA312468832GA SA312468834GA SA312468836GA SA312468843GA SA312468847GA SA312468855GA SA312468860GA SA312468863GA SA312468869GA SA312468877GA SA312468880GA SA312468890GA SA312468891GA SA312468894GA SA312468900GA SA312468901GA SA312468910GA SA312468922GA SA312468923GA SA312468926GA SA312468927GA SA312468929GA SA312468931GA SA312468934GA SA312468936GA SA312468941GA SA312468945GA SA312468947GA SA312468950GA SA312468951GA SA312468955GA SA312468958GA SA312468960GA SA312468961GA SA312468965GA SA312468966GA SA312468985GA SA312468997GA SA312468998GA SA312469008GA SA312469013GA SA312469014GA SA312469015GA SA312469016GA SA312469017GA SA312469018GA SA312469019GA SA312469020GA SA312469021GA SA312469022GA SA312469023GA SA312469024GA SA312469025GA SA312469026GA SA312469027GA SA312469028GA SA312469033GA SA312469037GA SA312469043GA SA312469044GA SA312469045GA SA312469048GA SA312469053GA SA312469056GA SA312469057GA SA312469058GA SA312469065GA SA312469069GA SA312469075GA SA312469076GA SA312469079GA SA312469080GA SA312469082GA SA312469083GA SA312469084GA SA312469086GR SA312469087GA SA312469088GA SA312469089GA SA312469091GA SA312469092GA SA312469093GA SA312469094GA SA312469095GA SA312469096GA SA312469097GA SA312469098GA SA312469099GA SA312469100GA SA312469101GA SA312469102GA SA312469109GA SA312469112GA SA312469126GA SA312469133GA SA312469148GA SA312469159GA SA312469161GA SA312469165GA SA312469168GA SA312469178GA SA312469179GA SA312469180GA SA312469208GA SA312469209GA SA312469212GA SA312469214GA SA312469216GA SA312469219GA SA312469220GA SA312469222GA SA312469223GA SA312469224GA SA312469226GA SA312469227GA SA312469229GA SA312469248GA SA312469249GA SA312469250GA SA312469261GA SA312469263GA SA312469265GA SA312469268GA SA312469280GA SA312469281GA SA312469304GA SA312469310GA SA312469347GA SA312469354GA SA312469377GR SA312469385GA SA312469387GA SA312469388GA SA312469389GA SA312469392GA SA312469393GR SA312469394GA SA312469396GR SA312469397GR SA312469398GR SA312469402GA SA312469404GA SA312469411GA SA312469413GA SA312469415GA SA312469416GA SA312469417GA SA312469420GA SA312469425GA SA312469426GA SA312469428GA SA312469431GA SA312469434GA SA312469436GA SA312469438GA SA312469439GA SA312469440GA SA312469442GA SA312469446GA SA312469447GA SA312469451GA SA312469452GA SA312469453GA SA312469456GA SA312469457GA SA312469459GA SA312469460GA SA312469461GA SA312469462GA SA312469464GR SA312469466GA SA312469468GA SA312469469GA SA312469470GA SA312469471GA SA312469473GR SA312469474GR SA312469475GA SA312469476GR SA312469477GR SA312469478GA SA312469480GA SA312469481GR SA312469482GA SA312469483GA SA312469484GA SA312469485GA SA312469486GR SA312469487GR SA312469488GA SA312469489GA SA312469490GA SA312469491GR SA312469492GA SA312469495GA SA312469501GA SA312469504GA SA312469506GA SA312469508GA SA312469509GA SA312469511GA SA312469512GA SA312469518GA SA312469519GA SA312469521GA SA312470005GA SA312470009GA SA312470010GA SA312470011GA SA312470013GA SA312470020GA SA312470021GA SA312470022GA SA312470023GA SA312470025GA SA312470027GA SA312470028GA SA312470029GA SA312470030GA SA312470031GA SA312470032GA SA312470036GA SA312470039GA SA312470040GA SA312470045GA SA312470049GA SA312470051GA SA312470052GA SA312470053GA SA312470061GA SA312470063GA SA312470068GA SA312470072GA SA312470074GA SA312470075GA SA312470078GA SA312470079GA SA312470081GA SA312470084GA SA312470085GA SA312479532GA SA312479537GA SA312479538GA SA312479540GA SA312479546GA SA312479547GA SA312479555GA SA312479556GA SA312479578GA SA312479579GA SA312479580GA SA312479582GA SA312479583GA SA312479585GA SA312479593GA SA312479595GA SA312479596GA SA312479603GA SA312479606GA SA312479607GA SA312479608GA SA312479612GA SA312479613GA SA312479614GR SA312479615GA SA312479618GA SA312479622GA SA312479623GA SA312479624GA SA312479625GA SA312479628GA SA312479629GA SA312479631GA SA312479633GA SA312479650GA SA312479652GA SA312479653GA SA312479654GR SA312479656GA SA312479659GA SA312479660GA SA312479663GA SA312479664GA SA312479668GA SA312479669GA SA312479672GR SA312479673GR SA312479674GA SA312479675GA SA312479676GR SA312479678GA SA312479682GR SA312479684GA SA312479688GA SA312479692GA SA312479697GR SA312479698GA SA312479705GA SA312479706GA SA312479709GA SA312479729GA SA312479731GA SA312479732GA SA312479733GR SA312479735GA SA312479736GA SA312479737GR SA312479738GA SA312479739GA SA312479741GR SA312479743GR SA312479744GA SA312479745GA SA312479746GA SA312479747GR SA312479748GR SA312479749GA SA312479753GA SA312479756GA SA312479757GA SA312479760GA SA312479761GA SA312479763GA SA312479764GA SA312479771GA SA312479772GA SA312479773GA SA312479774GA SA312479775GR SA312479776GA SA312479778GR SA312479779GA SA312479780GA SA312479781GA SA312479782GA SA312479783GA SA312479785GA SA312479786GA SA312479790GA SA312479792GA SA312479799GA SA312479800GR SA312479802GR SA312479805GA SA312479806GA SA312479807GA SA312479809GA SA312479810GA SA312479813GA SA312479820GA SA312479822GA SA312479823GA SA312479824GA SA312479826GA SA312479827GA SA312479831GA SA312479834GA SA312479836GA SA312479841GA SA312479846GA SA312479847GA SA312479848GA SA312479849GA SA312479851GA SA312479852GA SA312479854GA SA312479855GA SA312479856GA SA312479858GA SA312479859GA SA312479860GA SA312479866GA SA312479872GA SA312479874GA SA312479875GA SA312479876GA SA312479884GA SA312479891GA SA312479892GA SA312479893GA SA312479895GA SA312479896GA SA312479897GA SA312479899GA SA312479900GA SA312479902GA SA312479903GA SA312479904GA SA312479905GA SA312479906GA SA312479907GA SA312479908GA SA312479909GA SA312479910GA SA312479911GA SA312479912GA SA312479913GA SA312479914GA SA312479915GA SA312479916GA SA312479917GA SA312479919GA SA312479920GA SA312479921GA SA312479922GA SA312479923GA SA312479924GA SA312479927GA SA312479928GA SA312479929GA SA312479931GA SA312479932GA SA312479933GA SA312479934GA SA312479935GA SA312479936GA SA312479937GA SA312479938GA SA312479939GA SA312479940GA SA312479941GA SA312479942GA SA312479944GA SA312479946GA SA312479947GA SA312479948GA SA312479953GA SA312479954GA SA312479963GA SA312479966GA SA312479968GA SA312479969GA SA312480091GA SA312480092GA SA312480093GA SA312480095GA SA312480098GA SA312480099GA SA312480101GA SA312480102GA SA312480105GA SA312480106GA SA312480107GA SA312480110GA SA312480114GA SA312480121GA SA312480122GA SA312480123GA SA312480128GA SA312480146GA SA312480147GA SA312480161GA SA312480165GA SA312480167GA SA312480176GA SA312480177GA SA312480178GA SA312480183GA SA312480185GA SA312480187GA SA312480192GA SA312480193GA SA312480197GA SA312480200GA SA312480204GA SA312480207GA SA312480208GR SA312480211GA SA312480215GA SA312480216GA SA312480217GA SA312480218GA SA312480220GA SA312480222GA SA312480224GA SA312480227GA SA312480228GA SA312480233GA SA312480234GA SA312480236GA SA312480237GA SA312480239GA SA312480240GA SA312480241GA SA312480245GA SA312480246GA SA312480247GA SA312480249GA SA312480250GA SA312480251GA SA312480252GA SA312480254GA SA312480256GA SA312480259GA SA312480260GA SA312480262GA SA312480263GA SA312480264GA SA312480267GA SA312480268GA SA312480269GA SA312480270GR SA312480273GA SA312480274GA SA312480276GA SA312480278GA SA312480279GA SA312480281GA SA312480282GA SA312480283GA SA312480284GA SA312480285GA SA312480286GA SA312480288GA SA312480289GA SA312480290GA SA312480291GA SA312480292GA SA312480293GA SA312480294GA SA312480295GA SA312480296GA SA312480298GA SA312480299GA SA312480300GA SA312480301GA SA312480302GA SA312480305GA SA312480306GA SA312480308GA SA312480309GA SA312480311GA SA312480312GA SA312480314GA SA312480316GA SA312480318GA SA312480319GA SA312480320GA SA312480321GA SA312480322GA SA312480323GA SA312480326GA SA312480328GA SA312480330GA SA312480332GA SA312480335GA SA312480339GA SA312480340GA SA312480348GA
SA312480351GA SA312480355GA SA312480356GA SA312480360GA SA312480363GA SA312480364GA SA312480365GA SA312480366GA SA312480368GA SA312480370GA SA312480371GA SA312480372GA SA312480374GA SA312480375GA SA312480376GA SA312480378GA SA312480380GA SA312480381GA SA312480384GA SA312480385GA SA312480390GA SA312480392GA SA312480393GA SA312480394GA SA312480395GA SA312480396GA SA312480397GA SA312480398GA SA312480399GA SA312480408GA SA312480409GA SA312480416GA SA312480417GA SA312480423GA SA312480425GA SA312480427GA SA312480432GA SA312480436GA SA312480437GA SA312480438GA SA312480440GA SA312480442GA SA312480446GA SA312480449GA SA312480451GA SA312480452GA SA312480453GA SA312480456GA SA312480457GA SA312480458GA SA312480460GA SA312480462GA SA312480464GA SA312480465GA SA312480466GA SA312480468GA SA312480470GA SA312480471GA SA312480472GR SA312480477GA SA312480479GA SA312480480GA SA312480481GA SA312480482GA SA312480485GA SA312480487GA SA312480488GA SA312480490GA SA312480493GA SA312480494GA SA312480499GA SA312480504GA SA312480509GA SA312480510GA SA312480511GA SA312480512GA SA312480513GA SA312480515GA SA312480516GA SA312480517GA SA312480518GA SA312480519GA SA312480520GA SA312480522GA SA312480523GA SA312480524GA SA312480525GA SA312480526GA SA312480528GA SA312480530GA SA312480531GA SA312480532GA SA312480533GA SA312480535GA SA312480536GA SA312480537GA SA312480538GA SA312480539GA SA312480540GA SA312480541GA SA312480542GA SA312480547GA SA312480551GA SA312480554GA SA312480560GA SA312480561GA SA312480563GA SA312480565GA SA312480566GA SA312480568GA SA312480570GA SA312480573GA SA312480575GA SA312480579GA SA312480582GA SA312480585GA SA312480587GR SA312480596GA SA312480598GA SA312480600GR SA312480602GA SA312480603GA SA312480609GA SA312480612GA SA312480622GA SA312480628GA SA312480629GA SA312480631GA SA312480632GA SA312480638GA SA312480642GA SA312480644GA SA312480645GA SA312480651GA SA312480657GA SA312480661GA SA312480662GA SA312480663GA SA312480664GA SA312480665GA SA312480666GA SA312480667GA SA312480668GA SA312480669GA SA312480670GA SA312480671GA SA312480673GA SA312480674GA SA312480676GA SA312480677GA SA312480678GA SA312480679GA SA312480680GA SA312480681GA SA312480683GA SA312480684GA SA312480685GA SA312480686GA SA312480687GA SA312480688GA SA312480689GA SA312480690GA SA312480691GA SA312480694GA SA312480695GA SA312480705GA SA312480708GA SA312480709GA SA312480711GA SA312480717GA SA312480719GA SA312480721GA SA312480722GA SA312480723GA SA312480724GA SA312480725GA SA312480726GA SA312480728GA SA312480729GA SA312480733GA SA312480734GA SA312480735GA SA312480736GA SA312480737GA SA312480738GA SA312480739GA SA312480740GA SA312480741GA SA312480742GA SA312480743GA SA312480744GA SA312480745GA SA312480746GA SA312480748GA SA312480749GA SA312480750GA SA312480751GA SA312480752GA SA312480753GA SA312480754GA SA312480755GA SA312480756GA SA312480757GA SA312480758GA SA312480759GA SA312480760GA SA312480762GA SA312480763GA SA312480764GA SA312480765GA SA312480766GA SA312480767GA SA312480768GA SA312480769GA SA312480770GA SA312480771GA SA312480772GA SA312480773GA SA312480774GA SA312480775GA SA312480776GA SA312480778GA SA312480779GA SA312480780GA SA312480782GA SA312480783GA SA312480784GA SA312480786GA SA312480787GA SA312480788GA SA312480789GA SA312480793GA SA312480794GA SA312480800GA SA312480802GA SA312480804GA SA312480805GA SA312480806GA SA312480808GR SA312480809GA SA312480810GA SA312480811GA SA312480813GA SA312480814GA SA312480815GA SA312480817GA SA312480818GA SA312480819GA SA312480821GA SA312480822GA SA312480823GA SA312480824GR SA312480825GA SA312480826GA SA312480827GA SA312480829GA SA312480830GR SA312480831GA SA312480833GA SA312480835GR SA312480837GA SA312480839GA SA312480840GA SA312480841GA SA312480842GA SA312480843GA SA312480844GA SA312480845GA SA312480846GA SA312480847GA SA312480848GA SA312480849GA SA312480850GA SA312480851GA SA312480852GA SA312480854GA SA312480855GA SA312480858GA SA312480859GA SA312480860GA SA312480861GA SA312480862GA SA312480863GA SA312480864GA SA312480865GA SA312480866GA SA312480869GA SA312490872GA SA312490873GA SA312490876GA SA312490877GA SA312490878GA SA312490879GA SA312490880GA SA312490881GA SA312490882GA SA312490885GA SA312490886GA SA312490887GA SA312490888GA SA312490890GA SA312490891GA SA312490892GR SA312490894GA sa312490895ga SA312490896GA SA312490897GA SA312490898GA SA312490900GR SA312490901GA SA312490902GA SA312490903GA SA312490904GA SA312490905GA SA312490906GA SA312490909GA SA312490910GA SA312490911GA SA312490912GA SA312490913GA SA312490915GA SA312490916GA SA312490917GA SA312490918GA SA312490919GA SA312490921GA SA312490922GA SA312490923GA SA312490924GA SA312490926GA SA312490927GA SA312490928GA SA312490929GA SA312490930GA SA312490931GA SA312490932GA SA312490933GA SA312490934GA SA312490935GA SA312490936GA SA312490938GA SA312490939GA SA312490940GA SA312490942GA SA312490943GA SA312490947GA SA312490948GA SA312490949GA SA312490951GA SA312490955GA SA312490958GA SA312490959GA SA312490963GA SA312490964GA SA312490965GA SA312490966GA SA312490972GA SA312490976GA SA312490977GA SA312490978GA SA312490979GA SA312490980GA SA312490982GA SA312490983GA SA312490984GA SA312490987GA SA312490988GA SA312490989GA SA312490990GA SA312490993GA SA312490994GA SA312490995GA SA312490996GA SA312490997GA SA312490998GA SA312490999GA SA312491000GA SA312491001GA SA312491003GA SA312491004GA SA312491005GA SA312491006GA SA312491007GA SA312491008GA SA312491009GA SA312491010GA SA312491011GA SA312491013GA SA312491015GA SA312491016GA SA312491017GA SA312491018GA SA312491019GA SA312491022GA SA312491023GA SA312491024GA SA312491025GA SA312491026GA SA312491027GA SA312491030GA SA312491031GA SA312491033GA SA312491034GA SA312491035GA SA312491036GA SA312491038GA SA312491039GA SA312491040GA SA312491041GA SA312491042GA SA312491043GA SA312491046GA SA312491047GA SA312491049GA SA312491050GA SA312491051GA SA312491052GA SA312491053GA SA312491054GA SA312491055GA SA312491056GA SA312491057GA SA312491058GA SA312491059GA SA312491060GA SA312491061GA SA312491062GA SA312491063GA SA312491068GA SA312491070GA SA312491071GA SA312491072GA SA312491075GA SA312491076GR SA312491077GA SA312491078GA SA312491080GA SA312491082GA SA312491083GA SA312491084GA SA312491085GA SA312491086GA SA312491087GA SA312491088GA SA312491089GA SA312491090GA SA312491092GA SA312491094GA SA312491095GA SA312491096GA SA312491097GA SA312491099GA SA312491100GA SA312491101GA SA312491102GA SA312491103GA SA312491104GA SA312491105GA SA312491106GA SA312491107GA SA312491108GA SA312491109GA SA312491110GA SA312491111GA SA312491112GA SA312491114GA SA312491115GA SA312491116GA SA312491117GA SA312491119GR SA312491121GA SA312491122GA SA312491125GA SA312491126GA SA312491129GA SA312491130GA SA312491132GA SA312491133GA SA312491135GA SA312491136GA SA312491137GA SA312491138GA SA312491142GA SA312491143GA SA312491144GA SA312491147GA SA312491148GA SA312491150GA SA312491151GA SA312491152GA SA312491153GA SA312491154GA SA312491155GA SA312491157GA SA312491158GA SA312491161GR SA312491163GA SA312491166GA SA312491167GR SA312491170GA SA312491171GA SA312491175GA SA312491181GA SA312491182GA SA312491184GA SA312491185GA SA312491188GA SA312491196GA SA312491198GA SA312491199GA SA312491200GR SA312491202GA SA312491204GA SA312491211GA SA312491213GA SA312491218GA SA312491220GA SA312491233GR SA312491234GR SA312491240GA SA312491245GA SA312491247GA SA312491249GA SA312491252GA SA312491257GA SA312491258GA SA312491259GA SA312491262GA SA312491266GA SA312491310GA SA312491317GA SA312491323GA SA312491330GA SA312491336GA SA312491357GA SA312491362GA SA312491364GA SA312491365GA SA312491367GA SA312491369GA SA312491372GA SA312491376GA SA312491377GA SA312491378GA SA312491382GA SA312491384GA SA312491385GA SA312491386GA SA312491388GA SA312491389GA SA312491390GA SA312491392GA SA312491398GA SA312491399GA SA312491402GA SA312491403GA SA312491404GA SA312491416GA SA312491417GA SA312491426GA SA312491447GA SA312491458GA SA312491461GA SA312491463GA SA312491470GA SA312491473GA SA312491474GA SA312491479GA SA312491481GA SA312491482GA SA312491486GA SA312491490GA SA312491492GA SA312491493GA SA312491503GA SA312491504GA SA312491506GA SA312491511GA SA312491512GA SA312491517GA SA312491519GR SA312491520GA SA312491522GA SA312491527GA SA312491531GA SA312491539GA SA312501557GA SA312501558GA SA312501560GA SA312501562GA SA312501580GA SA312501592GA SA312501595GA SA312501597GA SA312501599GA SA312501600GA SA312501603GA SA312501616GA SA312501621GA SA312501628GA SA312501630GR SA312501631GA SA312501634GA SA312501652GA SA312501695GA SA312501696GA SA312501726GA SA312501732GA SA312501743GR SA312501758GA SA312501762GA SA312501769GA SA312501772GA SA312501778GA SA312501784GA SA312501785GA SA312501816GR SA312501818GA SA312501820GA SA312501821GA SA312501822GA SA312501823GA SA312501824GA SA312501826GA SA312501830GA SA312501831GA SA312501837GA SA312501838GA SA312501845GA SA312501856GA SA312501876GA SA312501890GR SA312501946GA SA312501948GA SA312501951GA SA312501952GA SA312501953GA SA312501954GA SA312501955GA SA312501957GA SA312501960GA SA312501961GA SA312501962GA SA312501963GA SA312501966GA SA312501968GA SA312501970GA SA312501971GA SA312501972GA SA312501973GA SA312501974GA SA312501976GA SA312501978GA SA312501979GA SA312501980GA SA312501981GA SA312501982GA SA312501983GA SA312501986GA SA312501987GA SA312501988GA SA312501989GA SA312501991GA SA312501992GA SA312501993GA SA312501994GA SA312501995GA SA312501996GA SA312501997GA SA312501998GA SA312501999GA SA312502000GA SA312502002GA SA312502003GA SA312502005GA SA312502006GA SA312502007GA SA312502008GA SA312502009GA SA312502010GA SA312502011GA SA312502013GA SA312502014GA SA312502016GA SA312502018GA SA312502020GA SA312502022GA SA312502023GA SA312502026GA SA312502027GA SA312502028GA SA312502029GA SA312502030GA SA312502031GA SA312502032GA SA312502033GA SA312502034GA SA312502035GA SA312502036GA SA312502037GA SA312502038GA SA312502040GA SA312502042GA SA312502044GA SA312502046GA SA312502047GA SA312502048GA SA312502049GA SA312502051GA SA312502052GA SA312502061GA SA312502062GA SA312502064GA SA312502065GA SA312502071GA SA312502072GA SA312502079GA SA312502081GA SA312502082GA SA312502086GA SA312502087GA SA312502090GA SA312502091GA SA312502092GA SA312502099GA SA312502100GA SA312502101GA SA312502103GA SA312502105GA SA312502106GA SA312502111GA SA312502112GA SA312502113GA SA312502114GA SA312502116GA SA312502120GA SA312502122GA SA312502123GA SA312502124GA SA312502126GA SA312502127GA SA312502129GA SA312502130GA SA312502131GA SA312502134GA SA312502138GA SA312502140GA SA312502141GA SA312502143GA SA312502149GA SA312502153GR SA312502154GA SA312502158GA SA312502160GA SA312502162GA SA312502163GA SA312502164GA SA312502167GA SA312502171GA SA312502172GA SA312502173GA SA312502174GA SA312502175GA SA312512179GA SA312512180GA SA312512181GR SA312512183GA SA312512184GA SA312512185GA SA312512186GA SA312512188GA SA312512189GA SA312512190GA SA312512192GA SA312512194GA SA312512195GA SA312512196GA SA312512197GA SA312512198GA SA312512199GA SA312512200GA SA312512201GA SA312512204GR SA312512205GA SA312512206GA SA312512207GA SA312512210GA SA312512211GA SA312512214GA SA312512215GA SA312512216GA SA312512219GA SA312512220GA SA312512222GA SA312512225GA SA312512226GA SA312512227GA SA312512229GA SA312512230GA SA312512231GA
SA312512232GA SA312512233GA SA312512234GA SA312512235GR SA312512236GA SA312512237GA SA312512238GA SA312512239GA SA312512240GA SA312512241GA SA312512242GA SA312512243GA SA312512244GA SA312512245GA SA312512246GA SA312512247GA SA312512248GA SA312512249GA SA312512250GA SA312512251GA SA312512252GA SA312512253GA SA312512254GA SA312512255GA SA312512256GA SA312512259GA SA312512260GA SA312512261GA SA312512262GA SA312512263GA SA312512264GA SA312512265GA SA312512266GA SA312512267GA SA312512268GA SA312512269GA SA312512270GA SA312512271GA SA312512272GA SA312512274GA SA312512276GA SA312512277GA SA312512278GA SA312512279GA SA312512280GA SA312512281GA SA312512282GA SA312512283GA SA312512284GA SA312512285GA SA312512286GA SA312512287GA SA312512288GA SA312512289GA SA312512290GA SA312512291GA SA312512292GA SA312512294GA SA312512297GR SA312512298GA SA312512299GR SA312512301GA SA312512302GR SA312512303GR SA312512304GA SA312512305GA SA312512308GR SA312512312GR SA312512313GA SA312512314GA SA312512316GA SA312512317GR SA312512318GA SA312512319GA SA312512320GR SA312512322GA SA312512323GA SA312512324GA SA312512325GA SA312512326GA SA312512329GA SA312512330GA SA312512331GA SA312512333GA SA312512334GA SA312512335GA SA312512336GA SA312512338GA SA312512339GA SA312512340GA SA312512341GA SA312512342GA SA312512343GR SA312512344GA SA312512345GA SA312512347GA SA312512348GA SA312512349GA SA312512350GA SA312512351GA SA312512353GA SA312512354GA SA312512355GA SA312512356GA SA312512357GA SA312512358GA SA312512359GA SA312512360GA SA312512361GA SA312512362GA SA312512363GA SA312512365GR SA312512366GA SA312512367GR SA312512369GA SA312512370GA SA312512371GA SA312512372GA SA312512373GA SA312512374GA SA312512375GA SA312512376GA SA312512377GA SA312512378GR SA312512379GA SA312512380GA SA312522382GA SA312522383GA SA312522384GA SA312522385GA SA312522387GA SA312522390GR SA312522394GA SA312522395GA SA312522397GR SA312522398GA SA312522401GA SA312522403GA SA312522405GA SA312522410GA SA312522413GA SA312522416GA SA312522417GA SA312522420GR SA312522421GA SA312522432GA SA312522435GA SA312522438GA SA312522472GA SA312522474GA SA312522475GR SA312522478GA SA312522479GA SA312522480GA SA312522484GA SA312522495GA SA312522496GA SA312522505GA SA312522510GA SA312522511GA SA312522512GA SA312522513GA SA312522516GA SA312522518GA SA312522520GA SA312522523GA SA312522526GA SA312522529GA SA312522530GA SA312522534GA SA312522536GA SA312522538GA SA312522542GA SA312522544GA SA312522546GA SA312522550GA SA312522552GA SA312522553GA SA312522554GA SA312522555GA SA312522570GA SA312522576GA SA312522582GA SA312522584GA SA312522590GA SA312522598GA SA312522600GA SA312522601GA SA312522607GA SA312522610GA SA312522617GA SA312522620GA SA313002678GA SA313002684GA SA313002695GA SA313002702GA SA313002704GA SA313002708GA SA313002710GA SA313002712GA SA313002716GA SA313002719GA SA313002720GA SA313012741GA SA313012746GA SA313012752GA SA313012753GA SA313012766GR SA313012783GA SA313012794GA SA313012815GA SA313012853GA SA313012870GA SA313012919GA SA313012922GA SA313012923GA SA313012927GA SA313012934GA SA313012943GA SA313012949GA SA313012950GA SA313012964GA SA313012974GA SA313012975GA SA313012978GA SA313012980GA SA313013004GA SA313013008GA SA313013011GA SA313013013GA SA313013015GA SA313013016GA SA313013020GA SA313013021GA SA313013022GA SA313013026GA SA313013036GA SA313013048GA SA313013049GA SA313013059GA SA313013066GA SA313013068GA SA313013075GA SA313013077GA SA313013079GA SA313013081GA SA313013084GA SA313013088GA SA313013090GA SA313013102GA SA313013107GA SA313013111GA SA313013112GA SA313013113GA SA313013115GA SA313013117GA SA313013118GA SA313013120GA SA313013122GA SA313013123GA SA313013124GA SA313013125GR SA313013127GA SA313013132GA SA313013133GA SA313013135GA SA313013136GA SA313013142GA SA313013145GA SA313013146GA SA313013147GA SA313013148GA SA313013149GA SA313013151GA SA313013152GA SA313013154GA SA313013155GA SA313013156GA SA313013157GA SA313013158GA SA313013159GA SA313013160GA SA313013161GA SA313013162GA SA313013163GA SA313013164GA SA313013165GA SA313013166GA SA313023169GR SA313023192GA SA313023197GA SA313023199GA SA313023204GA SA313023207GA SA313023208GA SA313023210GA SA313023211GA SA313023212GA SA313023224GA SA313023228GA SA313023232GA SA313023239GA SA313023240GA SA313023244GA SA313023247GA SA313023249GA SA313023253GA SA313023262GA SA313023264GA SA313023266GA SA313023274GR SA313023275GA SA313023278GA SA313023279GA SA313023280GA SA313023281GA SA313023282GA SA313023283GA SA313023284GA SA313023285GA SA313023286GA SA313023288GA SA313023290GA SA313023291GA SA313023292GA SA313023294GA SA313023295GA SA313023299GA SA313023300GR SA313023306GA SA313023310GA SA313023311GA SA313023339GA SA313023343GA SA313023352GA SA313023360GA SA313023364GA SA313023374GA SA313023375GA SA313023382GA SA313023390GA SA313023392GA SA313023394GA SA313023399GA SA313023402GA SA313023404GA SA313023406GA SA313023407GR SA313023408GA SA313023409GR SA313023412GR SA313023415GA SA313023417GR SA313023418GR SA313023422GR SA313023423GR SA313023424GR SA313023426GR SA313023429GR SA313023430GA SA313023431GR SA313023433GA SA313023434GA SA313023437GR SA313023449GA SA313023495GA SA313023497GA SA313023498GA SA313023502GA SA313023503GR SA313023504GA SA313023506GA SA313033525GR SA313033527GR SA313033529GA SA313033530GR SA313033532GA SA313033565GA SA313033567GA SA313033569GA SA313033571GA SA313033573GA SA313033575GA SA313033576GA SA313033579GA SA313033580GA SA313033582GA SA313033586GR SA313033591GA SA313033592GA SA313033594GA SA313033596GR SA313033601GA SA313033607GA SA313033608GA SA313033609GA SA313033614GA SA313033615GA SA313033619GA SA313033620GA SA313033623GA SA313033627GA SA313033629GR SA313033632GA SA313033633GA SA313033634GA SA313033635GA SA313033636GA SA313033637GA SA313033638GA SA313033639GR SA313033640GA SA313033641GA SA313033642GA SA313033644GA SA313033645GA SA313033649GA SA313033651GA SA313033653GR SA313033655GA SA313033656GR SA313033657GR SA313033658GA SA313033659GA SA313033660GR SA313033665GR SA313033669GR SA313033670GA SA313033671GA SA313033672GR SA313033674GA SA313033676GA SA313033677GA SA313033679GA SA313033680GR SA313033682GA SA313033685GA SA313033686GA SA313033687GA SA313033688GA SA313033689GA SA313033690GA SA313033691GA SA313033692GA SA313033693GA SA313033694GA SA313033695GA SA313033696GA SA313033697GA SA313033702GA SA313033709GA SA313033712GA SA313033716GA SA313033717GA SA313033718GA SA313033719GA SA313033720GA SA313033721GA SA313033722GA SA313033723GA SA313033724GA SA313033725GA SA313033726GA SA313033727GA SA313033728GA SA313033729GA SA313033730GA SA313033731GA SA313033732GA SA313033733GA SA313033734GA SA313033735GA SA313033736GA SA313033737GA SA313033739GA SA313033740GA SA313033741GA SA313033742GA SA313033743GA SA313033744GA SA313033748GA SA313033760GA SA313033762GA SA313033766GA SA313033774GA SA313033808GA SA313033810GA SA313033811GA SA313033812GA SA313033817GA SA313033819GA SA313033820GA SA313033821GA SA313033823GA SA313033824GA SA313033827GA SA313033828GA SA313033829GA SA313033832GA SA313033836GA SA313033837GA SA313033838GA SA313033839GA SA313033840GA SA313033841GA SA313033843GA SA313033844GA SA313033845GA SA313033846GA SA313043867GA SA313043869GA SA313043870GA SA313043871GA SA313043875GA SA313043876GA SA313043878GA SA313043879GA SA313043881GA SA313043885GA SA313043886GA SA313043890GA SA313043893GA SA313043895GA SA313043896GA SA313043898GA SA313043911GA SA313043926GA SA313043936GA SA313043947GA SA313043948GA SA313043949GA SA313043959GA SA313043963GA SA313043964GA SA313043967GA SA313043970GA SA313043972GR SA313043977GR SA313043985GA SA313043988GA SA313043990GA SA313043998GA SA313044004GA SA313044005GA SA313044006GA SA313044007GA SA313044008GA SA313044009GA SA313044010GA SA313044011GA SA313044012GA SA313044013GA SA313044014GA SA313044015GA SA313044017GA SA313044018GA SA313044019GA SA313044021GA SA313044022GA SA313044023GA SA313044024GR SA313044025GA SA313044026GA SA313044027GA SA313044029GA SA313044032GA SA313044033GA SA313044034GA SA313044035GA SA313044036GR SA313044037GA SA313044038GA SA313044041GA SA313044042GR SA313044043GA SA313044044GA SA313044045GA SA313044046GA SA313044047GA SA313044049GA SA313044050GA SA313044051GA SA313044052GA SA313044053GA SA313044054GR SA313044055GR SA313044056GA SA313044057GA SA313044058GA SA313044061GA SA313044062GA SA313044063GA SA313044064GA SA313044065GA SA313044066GA SA313044067GA SA313044068GA SA313044070GR SA313044071GA SA313044072GA SA313044073GA SA313044075GA SA313044076GA SA313044081GA SA313044083GA SA313044084GA SA313044085GA SA313044086GA SA313044087GA SA313044088GA SA313044089GA SA313044090GA SA313044092GA SA313044093GA SA313044095GA SA313044097GA SA313044098GA SA313044099GA SA313044102GA SA313044103GA SA313044104GA SA313044105GA SA313044106GA SA313044107GA SA313044108GA SA313044109GA SA313044110GA SA313044112GA SA313044113GA SA313044114GA SA313044115GA SA313044116GA SA313044117GA SA313044118GA SA313044119GA SA313044123GA SA313054155GA SA313054156GA SA313054157GA SA313054158GA SA313054159GA SA313054160GA SA313054161GA SA313054162GA SA313054163GA SA313064164GA SA313064166GA SA313064167GA SA313064174GA SA313064175GA SA313064176GA SA313064180GA SA313064183GA SA313064187GA SA313064188GA SA313064189GA SA313064190GA SA313064192GA SA313064194GA SA313064195GA SA313064198GA SA313064200GA SA313064201GA SA313064203GA SA313064205GR SA313064209GA SA313064211GR SA313064222GR SA313064223GR SA313064225GR SA313064230GR SA313064236GA SA313064249GA SA313064254GA SA313064255GA SA313064256GA SA313064257GA SA313064258GA SA313064259GA SA313064260GA SA313064261GA SA313064262GA SA313064263GA SA313064266GA SA313064277GA SA313064283GA SA313064293GA SA313064294GA SA313064300GR SA313064307GR SA313064315GR SA313064318GR SA313064320GR SA313064321GR SA313064322GR SA313064323GR SA313064325GA SA313064330GR SA313064331GR SA313064336GA SA313064337GA SA313064348GA SA313064354GR SA313064356GR SA313064358GR SA313064359GR SA313064360GA SA313064362GR SA313064363GA SA313064368GR SA313064369GA SA313064370GR SA313064374GR SA313064383GR SA313064386GR SA313064394GR SA313064426GA SA313064428GA SA313064429GA SA313064430GA SA313064431GA SA313064432GA SA313064434GA SA313064435GA SA313064436GA SA313064437GA SA313064438GA SA313064439GA SA313064441GA SA313064444GA SA313084447GA SA313084452GA SA313084470GA SA313084480GA SA313084482GA SA313084486GA SA313084488GA SA313084492GA SA313084494GA SA313084495GA SA313084496GA SA313084497GA SA313084498GA SA313084499GA SA313084502GA SA313084504GA SA313084505GA SA313084517GA SA313084521GA SA313084522GA SA313084524GA SA313084526GR SA313084531GA SA313084533GA SA313084536GA SA313084537GA SA313084538GA SA313084539GA SA313084540GA SA313084541GA SA313084543GA SA313084544GA SA313084545GA SA313084546GA SA313084547GA SA313084549GA SA313084550GA SA313084552GA SA313084553GA SA313084554GA SA313084555GA SA313084556GA SA313084557GA SA313084558GA SA313084561GA SA313084562GA SA313084564GA SA313084566GA SA313084569GA SA313084570GA SA313084571GA SA313084572GA SA313084576GA SA313084577GA SA313084578GA SA313084579GA SA313084581GA SA313084589GA SA313084590GA SA313084591GA SA313084593GA SA313084594GA SA313084627GA SA313084628GA SA313084631GA SA313084643GA SA313084655GA SA313084657GA SA313084659GA SA313084660GA
SA313084661GA SA313084663GA SA313084664GA SA313084665GA SA313084668GA SA313084669GA SA313084671GA SA313084674GA SA313084678GA SA313084679GA SA313084681GA SA313084682GA SA313084683GA SA313084684GA SA313084686GA SA313084693GA SA313084695GA SA313084698GA SA313084700GA SA313084701GA SA313084705GA SA313084706GA SA313084711GA SA313084715GA SA313084717GA SA313084719GA SA313084720GA SA313084723GA SA313084724GA SA313084725GA SA313084726GA SA313084727GA SA313084728GA SA313084731GA SA313084733GA SA313084734GA SA313084735GA SA313084738GA SA313084740GA SA313084742GA SA313084743GA SA313084744GA SA313084746GA SA313084747GA SA313084748GA SA313084750GA SA313084751GA SA313084752GA SA313084754GA SA313084755GA SA313084757GA SA313084759GA SA313084762GA SA313084763GA SA313084764GA SA313084769GA SA313084773GA SA313084774GA SA313084775GA SA313084779GA SA313084780GR SA313084782GA SA313084783GA SA313084784GA SA313084785GA SA313084786GA SA313084788GA SA313084789GA SA313084791GR SA313084792GA SA313084796GA SA313084797GA SA313084800GA SA313084801GA SA313084804GA SA313084806GA SA313084808GA SA313084809GA SA313084812GR SA313084813GA SA313084815GA SA313084822GR SA313084823GA SA313084824GR SA313084826GA SA313084827GA SA313084828GR SA313084830GR SA313084831GA SA313084832GR SA313084833GA SA313084837GA SA313084838GA SA313084839GA SA313084841GR SA313084842GR SA313084844GA SA313084846GA SA313084848GR SA313084849GA SA313084850GR SA313084853GA SA313084855GR SA313084859GA SA313084872GA SA313084881GR SA313084882GR SA313084884GA SA313084886GR SA313084887GA SA313084889GA SA313084890GA SA313084891GA SA313084892GA SA313084893GA SA313084894GA SA313084897GA SA313084898GA SA313084899GA SA313084902GA SA313084903GA SA313084907GA SA313084909GA SA313084910GA SA313084911GA SA313084912GR SA313084913GA SA313084914GA SA313084916GA SA313084917GA SA313084918GA SA313084919GA SA313084920GA SA313084921GA SA313084922GA SA313084923GA SA313084926GA SA313084927GA SA313084928GA SA313084931GA SA313084932GA SA313084933GA SA313084936GR SA313084937GA SA313084942GA SA313084943GR SA313084944GA SA313084946GA SA313084950GR SA313084953GA SA313084954GA SA313084955GA SA313084964GA SA313084966GA SA313084968GA SA313084969GA SA313084970GA SA313084971GA SA313084972GA SA313084974GA SA313084979GA SA313094983GA SA313094986GA SA313094987GA SA313094988GA SA313094989GA SA313094990GA SA313094991GA SA313094993GA SA313094994GA SA313094998GA SA313095001GA SA313095002GA SA313095003GA SA313095004GA SA313095005GA SA313095006GA SA313095007GA SA313095008GA SA313095009GA SA313095011GA SA313095012GA SA313095013GA SA313095014GA SA313095016GA SA313095028GA SA313095029GA SA313095030GA SA313095032GA SA313095033GA SA313095035GA SA313095036GA SA313095038GA SA313095040GA SA313095041GA SA313095043GA SA313095044GA SA313095045GA SA313095046GA SA313095047GA SA313095049GA SA313095050GA SA313095054GA SA313095055GA SA313095059GA SA313095060GA SA313095061GR SA313095064GA SA313095070GA SA313095072GA SA313095085GA SA313095086GA SA313095087GA SA313095092GA SA313095097GA SA313095098GA SA313095099GA SA313095100GA SA313095103GA SA313095105GA SA313095106GA SA313095107GA SA313095108GA SA313095109GA SA313095110GA SA313095113GA SA313095114GA SA313095116GA SA313095124GR SA313095126GA SA313095127GR SA313095128GA SA313095129GA SA313095131GA SA313095132GA SA313095135GA SA313095136GA SA313095137GA SA313095138GA SA313095139GA SA313095140GR SA313095141GA SA313095142GA SA313095143GA SA313095144GA SA313095145GA SA313095146GA SA313095147GA SA313095148GA SA313095150GA SA313095151GA SA313095152GA SA313095153GA SA313095154GA SA313095156GA SA313095157GA SA313095158GA SA313095159GA SA313095160GA SA313095162GA SA313095163GA SA313095164GA SA313095165GA SA313095168GA SA313095170GA SA313095172GA SA313095174GA SA313095175GA SA313095176GA SA313095177GA SA313095179GA SA313095182GA SA313095186GA SA313095193GA SA313095244GA SA313095258GR SA313095264GA SA313095271GA SA313095274GA SA313095282GA SA313095284GR SA313095285GA SA313095286GA SA313095287GR SA313095288GA SA313095289GA SA313095293GA SA313095295GA SA313095297GA SA313095306GA SA313095308GA SA313095309GA SA313095311GA SA313095312GA SA313095317GR SA313095319GR SA313095320GR SA313095325GA SA313095327GA SA313095329GR SA313095332GA SA313095333GR SA313095341GA SA313095343GA SA313095347GA SA313095348GA SA313095349GA SA313095359GR SA313095363GR SA313095365GA SA313095369GR SA313095371GA SA313095375GR SA313095376GA SA313095377GR SA313095378GR SA313095380GA SA313095381GA SA313095383GA SA313095384GA SA313095391GA SA313095392GA SA313095394GA SA313095395GA SA313095396GA SA313095399GA SA313095401GA SA313095403GA SA313095406GA SA313095410GA SA313095411GR SA313095412GA SA313095414GR SA313095415GA SA313095416GA SA313095417GA SA313095418GA SA313095419GA SA313095420GA SA313095422GA SA313095429GA SA313095431GA SA313095434GA SA313095435GA SA313095436GA SA313095437GA SA313095438GA SA313095439GA SA313095440GA SA313095443GR SA313095444GA SA313095445GA SA313095447GR SA313095448GA SA313095450GA SA313095452GA SA313095453GA SA313095456GA SA313095458GA SA313095460GR SA313095465GA SA313095467GA SA313095468GR SA313095471GA SA313095473GA SA313095478GA SA313095483GA SA313095484GR SA313095485GR SA313095486GA SA313095487GA SA313095489GR SA313095490GR SA313095494GA SA313095495GA SA313095506GR SA313095508GA SA313095511GA SA313095514GA SA313095519GA SA313095527GA SA313095531GA SA313095532GA SA313095534GR SA313095535GA SA313095536GA SA313095537GA SA313095538GR SA313095540GA SA313095541GR SA313095543GA SA313095545GR SA313095546GR SA313095547GA SA313095548GR SA313095549GR SA313095552GA SA313095553GA SA313095555GA SA313095557GA SA313095558GA SA313095559GA SA313095560GA SA313095561GA SA313095563GA SA313095565GA SA313095567GA SA313095568GA SA313095569GA SA313095571GA SA313095572GA SA313095580GA SA313095581GA SA313095582GA SA313095584GA SA313095585GA SA313095588GA SA313095596GA SA313095598GA SA313095608GA SA313095611GA SA313095615GA SA313095616GA SA313095617GA SA313095618GA SA313095621GA SA313095622GA SA313095623GR SA313095624GA SA313095625GA SA313095628GR SA313095633GA SA313095634GA SA313095636GA SA313095637GA SA313095641GR SA313095645GA SA313095646GR SA313095647GA SA313095648GA SA313095649GA SA313095650GA SA313105652GR SA313105667GR SA313105669GA SA313105670GA SA313105683GA SA313105687GA SA313105689GA SA313105690GA SA313105691GA SA313105695GR SA313105698GA SA313105699GA SA313105700GA SA313105702GA SA313105703GA SA313105705GA SA313105707GA SA313105708GR SA313105710GA SA313105712GA SA313105713GA SA313105715GA SA313105716GA SA313105718GA SA313105720GA SA313105721GA SA313105723GA SA313105724GA SA313105725GA SA313105726GA SA313105727GR SA313105729GR SA313105730GA SA313105734GA SA313105735GA SA313105739GR SA313105740GR SA313105742GA SA313105745GR SA313105746GA SA313105748GA SA313105749GA SA313105750GA SA313105751GA SA313105752GA SA313105753GA SA313105757GA SA313105760GA SA313105761GA sa313105766ga SA313105767GA SA313105768GA SA313105770GA SA313105771GA SA313105773GA SA313105775GA SA313105776GA SA313105778GA SA313105779GR SA313105780GA SA313105781GA SA313105782GA SA313105783GA SA313105784GA SA313105785GR SA313105787GA SA313105788GA SA313105789GR SA313105790GA SA313105791GA SA313105792GA SA313105794GA SA313105795GA SA313105797GA SA313105798GA SA313105800GA SA313105801GR SA313105803GA SA313105804GA SA313105805GA SA313105807GA SA313105808GA SA313105809GA SA313105810GA SA313105811GA SA313105812GA SA313105813GA SA313105814GA SA313105815GA SA313105816GA SA313105817GA SA313105818GA SA313105819GR SA313105821GA SA313105822GA SA313105823GA SA313105824GA SA313105825GA SA313105826GR SA313105827GR SA313105829GA SA313105830GR SA313105832GA SA313105834GA SA313105835GR SA313105837GA SA313105838GR SA313105839GA SA313105840GR SA313105842GA SA313105843GA SA313105844GA SA313105845GA SA313105846GA SA313105847GA SA313105848GA SA313105849GA SA313105850GA SA313105851GR SA313105852GA SA313105853GA SA313105854GR SA313105855GA SA313105856GA SA313105857GR SA313105858GA SA313105859GA SA313105860GA SA313105861GR SA313105862GA SA313105864GR SA313105865GA SA313105866GA SA313105868GA SA313105869GA SA313105870GA SA313105871GA SA313105873GA SA313105875GA SA313105876GA SA313105878GA SA313105880GR SA313105882GA SA313105884GA SA313105885GA SA313105886GR SA313105888GA SA313105889GR SA313105890GA SA313105892GA SA313105893GA SA313105894GA SA313105895GA SA313105896GR SA313105900GR SA313105901GR SA313105902GA SA313105904GA SA313105905GA SA313105908GR SA313105909GA SA313105910GR SA313105912GA SA313105915GA SA313105919GA SA313105921GA SA313105923GR SA313105938GR SA313105941GR SA313105942GA SA313105947GA SA313105951GR SA313105962GA SA313105967GA SA313105979GA SA313105982GA SA313105984GA SA313105986GA SA313105992GA SA313105994GA SA313105995GA SA313105996GA SA313105997GA SA313105998GA SA313106000GA SA313106001GA SA313106002GA SA313106010GA SA313106013GR SA313106014GA SA313106015GA SA313106017GA SA313106021GA SA313106025GA SA313106027GA SA313106028GA SA313106030GA SA313106031GA SA313106032GA SA313106033GR SA313106034GA SA313106036GA SA313106039GR SA313106040GR SA313106041GR SA313106042GA SA313106043GA SA313106046GA SA313106047GA SA313106051GA SA313106052GA SA313106053GA SA313106054GA SA313106055GR SA313106057GA SA313106058GA SA313106059GA SA313106060GA SA313106061GA SA313106062GA SA313106064GA SA313106068GA SA313106070GA SA313106071GA SA313106072GA SA313106073GA SA313106074GA SA313106076GA SA313106079GR SA313106081GA SA313106082GA SA313106083GA SA313106084GA SA313106085GA SA313106086GA SA313106089GA SA313106091GA SA313106094GA SA313106095GA SA313106096GA SA313106097GR SA313106098GR SA313106102GA SA313106106GA SA313106107GA SA313106108GA SA313106110GA SA313106121GA SA313106122GA SA313106123GA SA313106127GA SA313106128GA SA313106130GA SA313106132GA SA313106133GA SA313106134GA SA313106137GA SA313106139GA SA313106143GA SA313106145GA SA313106149GA SA313106150GA SA313106153GA SA313106154GA SA313106156GA SA313106157GR SA313106160GR SA313106161GA SA313106163GA SA313106164GR SA313106165GA SA313106167GA SA313106168GR SA313106170GA SA313106171GA SA313106172GA SA313106173GR SA313106175GA SA313106176GA SA313106177GA SA313106178GA SA313106179GA SA313106180GA SA313106182GA SA313106183GA SA313106184GA SA313106185GR SA313106187GA SA313106188GA SA313106189GR SA313106190GA SA313106191GA SA313106192GR SA313106193GA SA313106194GA SA313106195GR SA313106196GA SA313106198GA SA313106199GA SA313106200GA SA313106201GA SA313106202GA SA313106203GA SA313106205GA SA313106208GA SA313106209GA SA313106210GA SA313106211GA SA313106213GA SA313106215GA SA313106216GA SA313106217GA SA313106218GA SA313106219GA SA313106220GR SA313106221GR SA313106222GR SA313106223GR SA313106224GA SA313106225GA SA313106226GR SA313106227GA SA313106228GA SA313106231GA SA313106232GA SA313106233GA SA313106235GA SA313106237GA SA313106238GR SA313106241GA SA313106242GR SA313106243GA SA313106246GA SA313106248GA SA313106249GA SA313106250GA sa313106251ga SA313106252GA sa313106254ga SA3131062550GA SA313106256GA SA313106257GA SA313106258GA SA313106259GA SA313106260GA SA313106261GA SA313106264GA SA313106268GR SA313106269GR SA313106270GR SA313106271GA sa313106273ga SA313106276GA SA313106277GA SA313106278GA SA313106279GA SA313106280GA SA313106282GA SA313106283GA SA313106284GA SA313106285GA SA313106286GA SA313106287GA SA313106289GA SA313106290GA SA313106292GA SA313106293GA SA313106297GA SA313106298GA SA313106300GA SA313106301GA SA313106302GA SA313106303GA SA313106304GA SA313106305GA SA313106306GR SA313106307GA SA313106309GA SA313106311GR SA313106313GR SA313106315GR SA313106316GR SA313106317GR SA313106318GA SA313106319GR SA313106320GR SA313106321GR SA313106323GR SA313106326GR SA313106327GR SA313106328GR SA313106329GR SA313106331GA SA313106333GA SA313106334GR SA313106337GA SA313106338GA SA313106340GA SA313106344GA SA313106345GA SA313106346GA SA313106347GA SA313106355GA SA313106360GA SA313106361GA SA313106362GA SA313106363GA SA313106374GA SA313106375GA SA313106382GR SA313106383GA SA313106384GA SA313106386GA SA313106389GA SA313106391GA SA313106396GA SA313106397GA SA313106400GA SA313106405GA SA313106408GA SA313106409GA SA313106419GA SA313106420GA SA313106421GA SA313106422GA SA313106424GA SA313106425GA SA313106427GA SA313106433GA SA313106440GA SA313116446GA SA313116448GA SA313116451GA SA313116455GA SA313116458GA SA313116459GA SA313116463GR SA313116466GA SA313116467GA SA313116470GA SA313116474GA SA313116475GA SA313116476GA SA313116477GA SA313116482GA SA313116483GA SA313116485GA SA313116486GA SA313116488GA SA313116489GA SA313116490GA SA313116495GA SA313116497GA SA313116498GA SA313116500GA SA313116503GA SA313116504GA SA313116509GA SA313116510GA SA313116511GA SA313116514GA SA313116518GA SA313116519GA SA313116521GA SA313116525GA SA313116530GA SA313116531GA SA313116533GA SA313116534GA SA313116537GA SA313116540GA SA313116541GA SA313116543GA SA313116592GA SA313116594GA SA313116597GA SA313116600GA SA313116609GA SA313116610GA SA313116611GA SA313116612GA SA313116617GA SA313116619GA SA313116621GR SA313116622GA SA313116626GA SA313116631GA SA313116646GA SA313116649GA SA313116652GA SA313116656GA SA313116659GA SA313116661GA SA313116663GA SA313116664GA SA313116665GA SA313116670GA SA313116671GA SA313116672GA SA313116674GA SA313116675GA SA313116676GA SA313116677GA SA313116682GA SA313116683GR SA313116684GA SA313116685GA SA313116686GR SA313116687GA SA313116690GA SA313116693GA SA313116695GA SA313116697GA SA313116699GA SA313116701GA SA313116703GA SA313116707GR SA313116708GR SA313116712GA SA313116713GA SA313116715GA SA313116719GA SA313116720GR SA313116726GA SA313116729GA SA313116730GA SA313116732GA SA313116735GA SA313116738GA SA313116740GA SA313116743GA SA313116744GA SA313116745GA SA313116746GA SA313116747GA SA313116749GA SA313116750GA SA313116751GA SA313116752GA SA313116753GA SA313116754GA SA313116755GA SA313116759GA SA313116766GR SA313116769GR SA313116771GR SA313116774GA SA313116779GR SA313116783GA SA313116784GA SA313116788GR SA313116792GA SA313116793GA SA313116800GR SA313116805GA SA313116807GA SA313116808GA SA313116811GR SA313116815GA SA313116818GR SA313116822GR SA313116827GA SA313116828GA SA313116831GR SA313116835GA SA313116836GA SA313116840GA SA313116842GA SA313116843GA SA313116845GA SA313116847GA SA313116848GA SA313116849GR SA313116850GR SA313116851GR SA313116852GA SA313116854GR SA313116855GA SA313116856GR SA313116858GA SA313116860GR SA313116862GA SA313116863GR SA313116880GA SA313116887GR SA313116890GA SA313116895GA SA313116896GA SA313116905GR SA313116908GA SA313116909GA SA313116911GA SA313116914GR SA313116918GA SA313116919GA SA313116920GA SA313116921GA SA313116922GA SA313116926GA SA313116928GA SA313116929GA SA313116930GA SA313116931GA SA313116934GA SA313116938GA SA313116942GA SA313116944GA SA313116947GA SA313116948GA SA313116949GA SA313116950GA SA313116953GA SA313116955GA SA313116958GA SA313116963GA SA313116966GA SA313116968GA SA313116970GA SA313116972GA SA313116975GR SA313116977GA SA313116978GA SA313116983GA SA313116984GA SA313116985GA SA313116986GA SA313116988GA SA313116989GA SA313116992GA SA313116993GR SA313116995GA SA313116996GA SA313117000GA SA313117001GA SA313117002GA SA313117004GA SA313117006GA SA313117009GA SA313117012GA SA313117013GA SA313117016GA SA313117019GA SA313117022GA SA313117023GA SA313117028GA SA313117029GA SA313117031GA SA313117033GA SA313117034GA SA313117036GR SA313117038GR SA313117041GA SA313117043GR SA313117047GR SA313117049GA SA313117050GR SA313117058GA SA313117059GR SA313117062GA SA313117065GA SA313117068GA SA313117076GA SA313117080GR SA313117086GR SA313117091GA SA313117098GA SA313117106GA SA313117114GA SA313117117GA SA313117119GA SA313117121GA SA313117122GR SA313117123GR SA313117124GR SA313117125GR SA313117135GR SA313117136GR SA313117138GA SA313117139GA SA313117142GA SA313117148GR SA313117156GA SA313117158GA SA313117172GA SA313117184GA SA313117186GA SA313117188GA SA313117198GA SA313117199GA SA313117206GR SA313117207GR SA313127226GA SA313127231GA SA313127233GR SA313127239GR SA313127257GR SA313127264GR SA313127265GR SA313127280GA SA313127298GA SA313127299GA SA313127303GA SA313127305GA SA313127324GA SA313127326GA SA313127364GA SA313127365GA SA313127368GA SA313127376GA SA313127377GA SA313127379GA SA313127385GA SA313127386GA SA313127392GA SA313127395GA SA313127397GA SA313127402GA SA313127405GA SA313127408GA SA313127412GA SA313127417GA SA313127423GR SA313127425GR SA313127427GR SA313127428GR SA313127430GR SA313127431GR SA313127432GR SA313127443GR SA313127446GR SA313127448GR SA313127460GR SA313127461GA SA313127462GA SA313127466GR SA313127469GR SA313127479GA SA313127481GR SA313127489GA SA313127491GA SA313127504GA SA313127507GA SA313127510GA SA313127511GA SA313127512GR SA313127516GA SA313127521GA SA313127523GA SA313127531GR SA313127537GA SA313127552GA SA313127563GA SA313127565GA SA313127566GR SA313127567GA SA313127568GA SA313127570GA SA313127572GA SA313127573GA SA313127574GA SA313127579GA SA313127581GA SA313127586GA SA313127588GA SA313127592GR SA313127594GA SA313127595GA SA313127596GA SA313127598GA SA313127601GA SA313127603GR SA313127605GR SA313127606GR SA313127608GR SA313127610GR SA313127611GR SA313127612GR SA313127614GR SA313127615GR SA313127616GR SA313127619GR SA313127620GR SA313127621GR SA313127622GR SA313127623GA SA313127624GA SA313127625GR SA313127626GA SA313127627GR SA313127628GA SA313127629GA SA313127632GR SA313127637GA SA313127640GA SA313127646GR SA313127649GA SA313127657GR SA313127658GR SA313127663GR SA313127665GR SA313127666GR SA313127668GR SA313127670GR SA313127672GR SA313127674GA SA313127684GA SA313127685GA SA313127686GA SA313127687GA SA313127689GA SA313127690GR SA313127694GA SA313127695GA SA313127696GA SA313127697GA SA313127698GA SA313127699GA SA313127701GA sa313127703ga SA313127713GA SA313127714GA SA313127717GA SA313127718GA SA313127722GR SA313127724GR SA313127726GA SA313127737GR SA313137748GA SA313137782GR SA313147943GR SA313147950GA SA313147953GR SA313147957GA SA313147975GA SA313147977GR SA313147984GA SA313148000GA SA313148001GA SA313148009GA SA313148010GA SA313148012GA SA313148016GA SA313148022GA SA313198175GR SA313198187GR SA313198213GA SA313198229GA SA313198236GR SA313198243GR SA313198246GR SA313198247GR SA313198251GR SA313198253GR SA313198254GR SA313198267GA SA313198269GR SA313198276GA SA313198277GA SA313198278GA SA313198279GR SA313198284GR SA313198289GR SA313198292GA SA313198294GA SA313198295GA SA313198305GR SA313198315GA SA313198319GA SA313198372GA SA313198401GA SA313198433GR SA313198476GA SA313198488GA SA313198515GA SA313198517GA SA313198533GA SA313198624GA SA313198700GA SA313198702GA SA313198704GA SA313198706GA SA313198738GA SA313208790GA SA313208794GR SA313208800GA SA313208821GA SA313208828GA SA313208841GA SA313208852GA SA313208873GA SA313208875GA SA313208878GA SA313208879GA SA313208880GA SA313208886GA SA313208887GA SA313208888GA SA313208889GA SA313208890GA SA313208892GA SA313208927GA SA313208929GA SA313208930GA SA313208932GA SA313208936GA SA313208937GA SA313208938GA SA313208948GA SA313208952GA SA313208959GA SA313208960GA SA313208961GA SA313208963GA SA313208964GA SA313208966GA SA313208969GA SA313208970GA SA313208971GA SA313208972GA SA313208973GA SA313208974GA SA313208975GA SA313208976GA SA313208978GA SA313208979GA SA313208980GA SA313208991GA SA313208992GA SA313208994GA SA313208997GA SA313208998GA SA313209000GA SA313219011GA SA313219015GR SA313219017GA SA313219035GA SA313219037GA SA313219041GA SA313219047GA SA313219050GA SA313219052GR SA313219059GA SA313219060GA SA313219061GR SA313219063GR SA313219064GA SA313219067GA SA313219078GA SA313219090GA SA313219095GR SA313219097GR SA313219098GR SA313219102GR SA313219115GR SA313219122GR SA313219124GA SA313219130GA SA313219135GA SA313219138GA SA313219145GA SA313219148GA SA313219150GA SA313219153GA SA313219154GA SA313219159GA SA313219163GA SA313219164GA SA313219166GA SA313219167GA SA313219168GA SA313219169GA SA313219172GA SA313219173GA SA313219175GA SA313219178GA SA313219189GA SA313219190GA SA313219192GA SA313219194GA SA313219195GA SA313219197GA SA313219198GA SA313219200GA SA313219201GA SA313219205GA SA313219206GR SA313219208GA SA313219209GA SA313219210GA SA313219212GA SA313219214GA SA313219216GA SA313219219GA SA313219221GA SA313219222GA SA313219224GA SA313219225GA SA313219229GA SA313219231GA SA313219232GA SA313219235GA SA313219237GA SA313219240GA SA313219245GA SA313219260GA SA313219282GA SA313219291GA SA313219294GA SA313219298GA SA313220004GA SA313220006GA SA313220007GA SA313220008GA SA313220010GA SA313220011GA SA313220016GA SA313220017GA SA313220018GA SA313229328GA SA313229329GA SA313229338GA SA313229368GA SA313229386GA SA313229448GA SA313229469GR SA313229492GA SA313229509GA SA313229513GA SA313229555GA SA313229741GR SA313229745GR SA313229748GA SA313229755GR SA313229756GR SA313229764GR SA313229765GR SA313229769GR SA313229770GR SA313229773GR SA313229775GR SA313229780GA SA313229781GR SA313229788GA SA313229805GA SA313229818GA SA313229819GA SA313229821GR SA313229824GA SA313229825GA SA313229827GA SA313229828GR SA313229848GA SA313229852GR SA313229854GA SA313229855GR SA313229856GR SA313229857GA SA313229858GA SA313229860GA SA313229861GA SA313229864GA SA313229867GA SA313229868GA SA313229869GA SA313229870GA SA313229871GA SA313229876GA SA313229877GA SA313229879GA SA313229880GA SA313229881GA SA313229883GA SA313229884GA SA313229894GR SA313229898GA SA313229899GA SA313229902GA SA313229905GR SA313229911GA SA313229947GA SA313229952GA SA313229954GA SA313229962GA SA313229978GA SA313229983GA SA313229985GA SA313230031GR SA313230039GR SA313230050GA SA313230054GR SA313230065GR SA313230067GA SA313230069GR SA313230072GA SA313230075GR SA313230079GR SA313230080GR SA313230086GA SA313230091GA SA313230095GR SA313230101GA SA313230103GR SA313230107GR SA313230112GR SA313230113GA SA313230120GA SA313230123GA SA313230145GR SA313230155GA SA313230157GR SA313230163GR SA313230164GR SA313230168GR SA313230175GR SA313230182GA SA313230204GA SA313230229GR SA313230244GA SA313230246GA SA313230249GA SA313230252GA SA313230253GA SA313230254GA SA313230256GA SA313230259GA SA313230260GA SA313230261GA SA313230264GA SA313230266GA SA313230267GA SA313230272GA SA313230274GR SA313230276GA SA313230284GR SA313230285GR SA313230289GR SA313230313GR SA313230327GR SA313230335GR SA313230341GR SA313230343GA SA313230350GA SA313230351GA SA313230360GR SA313230371GR SA313230372GR SA313230377GA SA313230380GA SA313230387GR SA313230393GA SA313230395GR SA313230396GR SA313230398GR SA313230404GR SA313230405GA SA313230406GR SA313230407GR SA313230408GR SA313230409GR SA313230410GR SA313230411GR SA313230416GR SA313230419GR SA313230422GA SA313230428GR SA313230429GR SA313230430GR SA313230434GR SA313230437GR SA313230439GR SA313230440GR SA313230443GR SA313230446GR SA313230448GR SA313230452GR SA313230453GR SA313230455GR SA313230456GR SA313230457GR SA313230458GR SA313230459GR SA313230460GR SA313230464GR SA313230467GR SA313230468GA SA313230470GR SA313230473GR SA313230474GA SA313230487GR SA313230488GR SA313230491GA SA313230492GA SA313230494GR SA313230497GR SA313230500GA SA313230506GR SA313230508GR SA313230512GR SA313230520GA SA313230531GR SA313230533GR SA313230538GA SA313230541GA SA313230544GA SA313230568GA SA313230569GR SA313230579GA SA313230633GA SA313230666GR SA313230672GA SA313230676GR SA313230688GA SA313230689GA SA313230704GA SA313240941GR SA313240960GR SA313240966GA SA313240968GA SA313240983GA SA313240988GA SA313240989GA SA313241002GA SA313241050GA SA313241051GA SA313241089GA SA313251112GA SA313251113GA SA313251114GA SA313251116GA SA313251117GA SA313251118GA SA313251122GA SA313251123GA SA313251124GA SA313251125GA SA313251126GA SA313251365GA SA313251414GA SA313251420GA SA313251465GA SA313251494GA SA313251501GA SA313251514GA SA313251548GA SA313251584GA SA313251619GA SA313251660GA SA313251673GA SA313251684GA SA313251685GA SA313261701GA SA313261717GA SA313261720GA SA313261723GA SA313261724GA SA313261726GA SA313261767GA SA313261768GA SA313261797GA SA313261799GA SA313261816GA SA313261817GA SA313261819GA SA313261832GR SA313261835GA SA313261862GA SA313261880GA SA313261881GA SA313261903GA SA313261912GA SA313261915GA SA313271998GA SA313272001GA SA313272002GA SA313272005GA SA313272006GA SA313272010GA SA313272011GA SA313272014GA SA313272016GA SA313272018GA SA313272019GA SA313272023GA SA313272026GA SA313272057GA SA313272059GA SA313272060GA SA313272064GA SA313272065GA SA313272066GA SA313272067GR SA313272069GA SA313272070GA SA313272071GA SA313272072GA SA313272084GA SA313272086GA SA313272088GA SA313272089GA SA313272091GA SA313272094GA SA313272095GA SA313272096GA SA313272097GA SA313272100GA SA313272101GA SA313272102GA SA313272103GA SA313272104GA SA313272105GA SA313272106GA SA313272107GA SA313272108GA SA313272109GA SA313272111GA SA313272112GA SA313272126GA SA313272151GA SA313272157GA SA313272169GA SA313272178GA SA313272190GA SA313272207GA SA313272208GA SA313272210GA SA313272211GA SA313272213GA SA313272214GA SA313272215GA SA313272216GA SA313272217GA SA313272218GA SA313272223GA SA313272224GA SA313272226GA SA313272227GA SA313272230GA SA313272233GA SA313272236GA SA313272241GA SA313272243GA SA313272250GA SA313272252GA SA313272254GA SA313272257GA SA313272258GA SA313272261GA SA313272277GA SA313272297GA SA313272298GA SA313272299GA SA313272300GA SA313272301GA SA313272302GA SA313272306GA SA313272307GA SA313272311GA SA313272312GA SA313272313GA SA313272314GA SA313272317GA SA313272318GA SA313272319GA SA313272320GA SA313272321GA SA313272322GA SA313272325GA SA313272340GA SA313272341GA SA313272344GA SA313272346GA SA313272363GA SA313272364GA SA313272365GA SA313272368GA SA313272370GA SA313272373GA SA313272378GA SA313272380GA SA313272381GA SA313272382GR SA313272383GA SA313272384GA SA313272385GA SA313272388GA SA313272389GA SA313272392GA SA313272403GR SA313272407GR SA313272413GA SA313272414GR SA313272416GA SA313272420GA SA313272431GA SA313272432GA SA313272436GA SA313272440GA SA313272441GA SA313282453GA SA313282458GA SA313282459GA SA313282462GA SA313282463GA SA313282467GA SA313282475GA SA313282482GA SA313282488GA SA313282510GA SA313282518GA SA313282520GA SA313282521GR SA313282522GR SA313282527GA SA313282528GA SA313282529GA SA313282532GA SA313282533GA SA313282537GA SA313282539GA SA313282540GA SA313282542GR SA313282544GA SA313282547GR SA313282548GA SA313282550GA SA313282551GA SA313282554GA SA313282559GA SA313282562GA SA313282564GA SA313282565GA SA313282566GA SA313282568GA SA313282569GA SA313282571GA SA313282572GA SA313282573GR SA313282574GA SA313282577GA SA313282578GA SA313282579GA SA313282580GA SA313282581GA SA313282582GA SA313282584GA SA313282585GA SA313282586GA SA313282587GA SA313282588GA SA313282589GA SA313282590GA SA313282594GA SA313282607GA SA313282608GA SA313282609GA SA313282613GA SA313282614GA SA313282616GR SA313282619GA SA313282626GR SA313282627GA SA313282633GA SA313282634GA SA313282644GA SA313282653GA SA313282659GA SA313282662GA SA313282697GA SA313282703GA SA313282711GA SA313282712GA SA313282713GA SA313282714GA SA313282715GA SA313282716GA SA313282717GA SA313282721GA SA313282723GA SA313282725GA SA313282726GA SA313282730GA SA313282731GA SA313282732GR SA313282734GA SA313282735GA SA313282738GA SA313282739GA SA313282744GA SA313282745GA SA313282747GA SA313282756GA SA313282757GA SA313282762GA SA313282768GA SA313282770GR SA313282771GR SA313282772GA SA313282773GA SA313282774GA SA313282775GA SA313282777GA SA313282778GA SA313282779GA SA313282780GA SA313282781GR SA313282782GA SA313282783GA SA313282784GA SA313282785GA SA313282786GA SA313282787GA SA313282788GA SA313282789GR SA313282790GA SA313282791GA SA313282792GA SA313282793GA SA313282794GR SA313282795GA SA313282796GA SA313282797GA SA313282799GA SA313282800GA SA313282801GR SA313282802GA SA313282803GA SA313282804GA SA313282805GA SA313282807GA SA313282808GA SA313282809GA SA313282810GA SA313282811GA SA313282812GA SA313282816GA SA313282817GA SA313282819GR SA313282820GA SA313282821GA SA313282822GR SA313282823GR SA313282824GA SA313282825GA SA313282826GA SA313282827GA SA313282828GR SA313282830GA SA313282831GA SA313282832GA SA313282833GA SA313282834GA SA313282835GA SA313282836GA SA313282837GA SA313282839GA SA313282840GA SA313282841GA SA313282842GA SA313282845GA SA313282847GA SA313282849GA SA313282850GA SA313282853GA SA313282854GA SA313282855GA SA313282856GA SA313282858GA SA313282860GA SA313282861GA SA313282862GA SA313282863GA SA313282864GA SA313282866GA SA313282879GA SA313282881GA SA313292904GA SA313292905GA SA313292906GA SA313292907GA SA313292909GA SA313292910GA SA313292911GA SA313292912GA SA313292913GA SA313292914GA SA313292916GA SA313292917GA SA313292919GA SA313292920GA SA313292922GA SA313292924GA SA313292925GA SA313292927GA SA313292928GA SA313292929GA SA313292930GA SA313292931GA SA313292933GA SA313292935GA SA313292936GA SA313292938GA SA313292941GA SA313292942GA SA313292947GA SA313292950GA SA313292952GA SA313292955GA SA313292960GA SA313292963GA SA313292964GA SA313292968GA SA313292973GA SA313292975GR SA313292976GA SA313292977GA SA313292978GA SA313292979GA SA313292983GA SA313292984GA SA313292986GA SA313292987GA SA313292989GA SA313292990GA SA313292991GA SA313292993GA SA313292995GA SA313292996GA SA313292998GR SA313293001GA SA313293005GA SA313293008GA SA313293009GA SA313293011GA SA313293012GA SA313293013GA SA313293015GA SA313293016GA SA313293017GA SA313293018GA SA313293019GA SA313293020GA SA313293021GA SA313293031GA SA313293035GA SA313293048GA SA313293049GA SA313293052GA SA313293053GA SA313293056GA SA313293058GA SA313293063GA SA313293066GA SA313293073GA SA313293076GA SA313293080GA SA313293082GA SA313293083GA SA313293089GA SA313293091GA SA313293093GA SA313293095GA SA313293098GA SA313293100GA SA313293102GA SA313293104GA SA313293111GA SA313293112GA SA313293116GA SA313293117GA SA313293119GA SA313293123GA SA313293125GA SA313293126GA SA313293134GA SA313293135GA SA313293136GA SA313293137GA SA313293138GA SA313293139GA SA313293140GA SA313293141GA SA313293143GA SA313293146GA SA313293162GA SA313293167GA SA313293169GA SA313293170GA SA313293174GA SA313293175GA SA313293178GA SA313293180GA SA313293183GA SA313293185GA SA313293186GA SA313293188GA SA313293190GA SA313293194GA SA313293196GA SA313293197GA SA313293198GA SA313293209GA SA313293211GA SA313293215GA SA313293216GA SA313293224GA SA313293227GA SA313293229GA SA313293231GA SA313293234GA SA313293240GA SA313293244GA SA313293245GA SA313293247GA SA313293249GA SA313293250GA SA313293252GA SA313293257GA SA313293258GA SA313293259GA SA313293262GA SA313293268GA SA313293269GA SA313293272GA SA313293273GA SA313293279GA SA313293282GA SA313293283GA SA313293290GA SA313293292GA SA313293299GR SA313293307GR SA313293310GA SA313293387GR SA313293457GR SA313293464GR SA313293466GR SA313293499GR SA313293530GR SA313293534GA SA313303599GA SA313303625GA SA313303638GA SA313303686GR SA313303688GR SA313303697GR SA313303708GA SA313303713GR SA313303727GR SA313303729GA SA313303731GR SA313303735GA SA313303765GR SA313303766GR SA313303796GR SA313303819GA SA313303831GR SA313303842GA SA313303908GA SA313303912GR SA313303934GR SA313303935GR SA313303987GR SA313303988GR SA313303994GR SA313304000GR SA313304009GR SA313304024GR SA313304032GA SA313304058GA SA313304080GA SA313304114GR SA313304145GA SA313304152GR SA313304171GR SA313304219GA SA313304234GA SA313304267GR SA313304274GR SA313304290GA SA313304305GR SA313304307GR SA313304318GR SA313304319GR SA313304334GA SA313304359GA SA313304362GR SA313304365GR SA313314413GA SA313314414GR SA313314423GR SA313314425GR SA313314435GA SA313314438GA SA313314439GA SA313314442GA SA313314443GA SA313314444GA SA313314445GA SA313314446GA SA313314448GA SA313314450GA SA313314452GA SA313314456GA SA313314457GA SA313314458GA SA313314460GA SA313314461GA SA313314462GA SA313314464GA SA313314466GA SA313314467GA SA313314469GA SA313314470GA SA313314474GA SA313314478GA SA313314481GA SA313314482GA SA313314487GA SA313314489GA SA313314494GA SA313314506GA SA313314509GA SA313314512GA SA313314518GA SA313314525GA SA313314526GA SA313314529GA SA313314550GA SA313314554GA SA313314572GA SA313314585GA SA313314586GA SA313314587GA SA313314588GA SA313314589GA SA313314591GA SA313314592GR SA313314597GA SA313314598GA SA313314599GA SA313314600GR SA313314601GA SA313314602GA SA313314603GA SA313314605GA SA313314606GA SA313314607GA SA313314609GA SA313314610GA SA313314613GA SA313314614GA SA313314615GA SA313314616GA SA313314617GA SA313314619GA SA313314620GA SA313314621GA SA313314622GA SA313314623GA SA313314624GA SA313314626GA SA313314628GR SA313314629GA SA313314630GA SA313314632GA SA313314633GA SA313314634GA SA313314635GA SA313314636GA SA313314638GA SA313314639GA SA313314640GA SA313314641GA SA313314642GA SA313314643GA SA313314644GA SA313314645GA SA313314646GA SA313314647GA SA313314648GA SA313314649GA SA313314650GA SA313314651GA SA313314652GA SA313314653GA SA313314654GA SA313314655GA SA313314656GA SA313314657GA SA313314658GA SA313314659GA SA313314660GA SA313314661GA SA313314662GA SA313314663GA SA313314664GA SA313314665GR SA313314666GA SA313314667GA SA313314669GA SA313314670GA SA313314671GA SA313314735GA SA313314736GA SA313314737GA SA313314738GA SA313314739GA SA313314742GA SA313314743GA SA313314744GA SA313314745GA SA313314748GR SA313314749GA SA313314750GA SA313314751GA SA313314753GA SA313314756GA SA313314758GA SA313314760GA SA313314761GA SA313314762GA SA313314763GA SA313314765GA SA313314768GA SA313314769GA SA313314770GA SA313314771GA SA313314772GA SA313314773GA SA313314774GA SA313314775GA SA313314776GA SA313314779GA SA313314780GA SA313314781GA SA313314782GA SA313314783GA SA313314784GA SA313314785GA SA313314786GA SA313314787GA SA313314788GA SA313314789GA SA313314790GA SA313314791GA SA313314792GA SA313314793GA SA313314794GA SA313314795GA SA313314798GA SA313314799GA SA313314800GA SA313314801GA SA313314802GA SA313314803GA SA313314804GA SA313314806GA SA313314809GA SA313314810GA SA313314812GA SA313314817GA SA313314818GA SA313314821GA SA313314822GA SA313314824GA SA313314826GA SA313314828GA SA313314829GA SA313314830GA SA313314833GA SA313314837GA SA313314838GA SA313314843GA SA313314845GA SA313314847GA SA313314848GA SA313314851GA SA313314856GA SA313314858GA SA313314862GA SA313314865GA SA313314887GA SA313324896GA SA313324897GA SA313324898GA SA313324899GA SA313324900GA SA313324902GA SA313324903GA SA313324904GA SA313324905GA SA313324906GA SA313324907GA SA313324908GA SA313324910GA SA313324912GA SA313324913GA SA313324914GA SA313324915GA SA313324916GA SA313324917GA SA313324918GA SA313324919GA SA313324920GA SA313324921GA SA313324922GA SA313324924GA SA313324929GA SA313324930GA SA313324931GA SA313324933GA SA313324936GA SA313324938GA SA313324953GA SA313324955GA SA313324958GA SA313324959GA SA313324960GA SA313324961GA SA313324962GA SA313324963GA SA313324964GA SA313324967GA SA313324969GA SA313324970GA SA313324972GA SA313324973GA SA313324974GA SA313324975GA SA313324976GA SA313324977GA SA313324979GA SA313324980GA SA313324983GA SA313324984GA SA313324986GA SA313324988GA SA313324989GA SA313324990GA SA313324996GA SA313324997GA SA313335003GA SA313335006GA SA313335007GA SA313335009GA SA313335011GA SA313335013GA SA313335014GA SA313335017GA SA313335020GA SA313335021GA SA313335022GA SA313335030GA SA313335031GA SA313335032GA SA313335035GA SA313335040GA SA313335043GA SA313335045GA SA313335047GA SA313335049GA SA313335050GR SA313335051GR SA313335052GA SA313335054GA SA313335055GA SA313335056GA SA313335057GA SA313335059GA SA313335061GA SA313335062GR SA313335064GA SA313335065GA SA313335066GA SA313335067GA SA313335068GA SA313335069GR SA313335070GR SA313335071GR SA313335072GA SA313335073GA SA313335074GA SA313335075GR SA313335076GR SA313335077GA SA313335078GA SA313335079GR SA313335080GA SA313335081GA SA313335082GA SA313335083GR SA313335084GR SA313335085GA SA313335086GA SA313335088GA SA313335092GA SA313335093GR SA313335095GA SA313335096GR SA313335097GR SA313335098GA SA313335099GA SA313335100GA SA313335101GA SA313335102GA SA313335103GA SA313335104GA SA313335106GR SA313335107GR SA313335108GA SA313335109GA SA313335113GA SA313335114GA SA313335115GR SA313335116GA SA313335117GA SA313335119GA SA313335120GA SA313335121GR SA313335122GA SA313335123GA SA313335126GR SA313335127GA SA313335128GA SA313335129GA SA313335130GR SA313335131GA SA313335133GA SA313335137GA SA313335139GA SA313335145GA SA313335152GA SA313335154GR SA313335157GA SA313335165GA SA313335169GA SA313335171GA SA313335174GA SA313335176GA SA313335177GA SA313335186GA SA313335187GA SA313335194GA SA313335196GA SA313335197GA SA313335199GA SA313335200GA SA313335207GA SA313335208GA SA313335210GA SA313335212GA SA313335213GA SA313335214GA SA313335216GA SA313335217GA SA313335221GA SA313335222GA SA313335225GA SA313335226GA SA313335228GA SA313335230GA SA313335231GA SA313335232GA SA313335236GA SA313335237GA SA313335238GA SA313335239GA SA313335240GA SA313335247GA SA313335249GA SA313335250GA SA313335251GA SA313335254GA SA313335256GA SA313335258GA SA313335259GA SA313335261GA SA313335262GA SA313335265GA SA313335268GA SA313335269GA SA313335270GA SA313335272GR SA313335273GA SA313335274GA SA313335275GA SA313335276GR SA313335277GA SA313335280GA SA313335281GA SA313335282GA SA313335285GA SA313335286GA SA313335287GA SA313335288GA SA313335289GA SA313335290GA SA313335293GA SA313335295GA SA313335296GA SA313335301GA SA313335302GA SA313335304GA SA313335305GA SA313335306GA SA313335308GA SA313335309GA SA313335310GA SA313335312GR SA313335313GA SA313335314GA SA313335315GA SA313335316GA SA313335317GA SA313335318GA SA313335319GA SA313335320GR SA313335321GA SA313335322GA SA313335323GA SA313335324GA SA313335325GA SA313335326GA SA313335328GA SA313335329GA SA313335331GA SA313335333GA SA313335334GA SA313335336GA SA313335337GA SA313335338GA SA313335339GA SA313335340GA SA313335342GA SA313335344GA SA313335345GR SA313335346GA SA313335348GA SA313335349GA SA313335352GA SA313335354GA SA313335355GA SA313335356GA SA313335359GA SA313335361GA SA313335363GA SA313335365GR SA313335368GA SA313335374GA SA313335376GA SA313335378GA SA313335379GA SA313335380GA SA313335382GA SA313335383GA SA313335384GA SA313335385GA SA313335386GA SA313335387GA SA313335388GA SA313335389GA SA313335392GA SA313335395GA SA313335396GA SA313335398GA SA313335401GA SA313335402GA SA313335403GA SA313335404GA SA313335405GA SA313335406GA SA313335408GA SA313335409GA SA313335410GA SA313335414GA SA313335415GA SA313335416GA SA313335417GA SA313335418GA SA313335420GA SA313335421GA SA313335422GA SA313335423GA SA313335425GA SA313335429GA SA313335431GA SA313335433GA SA313335434GA SA313335436GA SA313335437GA SA313335438GA SA313335440GA SA313335441GA SA313335444GA SA313335448GA SA313335450GA SA313335452GA SA313335453GA SA313335457GA SA313335459GA SA313335460GA SA313335461GA SA313335462GA SA313335463GA SA313335464GA SA313335496GA SA313335497GA SA313335498GA SA313335500GA SA313335501GA SA313335502GA SA313335503GA SA313335504GA SA313335505GA SA313335507GA SA313335509GA SA313335512GA SA313335516GA SA313335518GA SA313335519GA SA313335520GA SA313335521GA SA313345525GA SA313345526GA SA313345528GA SA313345529GA SA313345532GA SA313345533GA SA313345536GA SA313345537GA SA313345538GA SA313345540GA SA313345541GA SA313345542GA SA313345543GA SA313345544GA SA313345545GA SA313345546GA SA313345548GA SA313345550GA SA313345551GA SA313345552GA SA313345555GA SA313345556GA SA313345557GA SA313345558GA SA313345559GA SA313345560GA SA313345561GA SA313345563GA SA313345564GA SA313345565GA SA313345566GA SA313345568GA SA313345569GA SA313345570GA SA313345571GA SA313345572GA SA313345573GA SA313345574GA SA313345578GA SA313345579GA SA313345580GA SA313345581GA SA313345583GA SA313345584GA SA313345585GA SA313345587GA SA313345588GA SA313345589GA SA313345591GA SA313345592GA SA313345597GA SA313345598GA SA313345600GA SA313345601GA SA313345602GA SA313345603GA SA313345604GA SA313345605GA SA313345606GA SA313345607GA SA313345608GA SA313345609GA SA313345610GA SA313345611GA SA313345612GA SA313345613GA SA313345614GA SA313345615GA SA313345616GA SA313345617GA SA313345619GA SA313345623GA SA313345626GA SA313345627GA SA313345633GA SA313345635GA SA313345636GA SA313345638GA SA313345639GA SA313345640GA SA313345642GA SA313345643GA SA313345644GR SA313345645GA SA313345646GA SA313345647GR SA313345649GA SA313345650GA SA313345651GA SA313345652GA SA313345653GA SA313345654GA SA313345655GA SA313345656GA SA313345657GA SA313345658GA SA313345659GA SA313345660GA SA313345662GA SA313345663GA SA313345664GA SA313345665GA SA313345666GA SA313345667GA SA313345668GA SA313345669GA SA313345670GA SA313345671GA SA313345672GA SA313345673GA SA313345674GA SA313345677GA SA313345678GA SA313345679GA SA313345680GA SA313345681GA SA313345682GA SA313345683GA SA313345685GA SA313345687GA SA313345688GA SA313345689GA SA313345690GA SA313345691GA SA313345692GA SA313345693GA SA313345695GA SA313345697GA SA313345700GA SA313345701GA SA313345702GA SA313345703GA SA313345706GA SA313345707GA SA313345708GA SA313345709GA SA313345710GR SA313345711GA SA313345712GR SA313345713GA SA313345714GR SA313345715GA SA313345721GA SA313345722GR SA313345724GA SA313345725GR SA313345727GA SA313345728GA SA313345729GA SA313345731GA SA313345733GR SA313345734GA SA313345735GA SA313345736GA SA313345737GA SA313345739GA SA313345740GA SA313345741GA SA313345742GA SA313345743GA SA313345744GA SA313345745GA SA313345746GA SA313345748GA SA313345749GA SA313345750GA SA313345751GA SA313345752GA SA313345753GA SA313345754GA SA313345756GA SA313345757GR SA313345759GA SA313345761GA SA313345763GA SA313345764GA SA313345766GA SA313345767GR SA313345768GA SA313345769GA SA313345770GA SA313345771GA SA313345772GA SA313345773GA SA313345774GA SA313345776GA SA313345778GA SA313345779GA SA313345781GA SA313345782GR SA313345783GA SA313345784GA SA313345785GA SA313345786GA SA313345787GA SA313345788GA SA313345789GA SA313345790GA SA313345792GA SA313345793GA SA313345794GA SA313345795GA SA313345796GA SA313345797GA SA313345798GA SA313345799GA SA313345800GA SA313345801GA SA313345802GA SA313345803GA SA313345804GA SA313345805GA SA313345806GA SA313345807GA SA313345808GA SA313345809GA SA313345810GA SA313345811GA SA313345812GA SA313345813GA SA313345814GA SA313345815GA SA313345816GA SA313345817GA SA313345818GA SA313345819GA SA313345820GA SA313345821GA SA313345822GA SA313345823GA SA313345824GA SA313345825GA SA313345826GA SA313345827GA SA313345828GA SA313345829GA SA313345830GA SA313345831GA SA313345832GA SA313345833GA SA313345834GA SA313345835GA SA313345836GA SA313345837GA SA313345838GA SA313345839GA SA313345840GA SA313345841GA SA313345842GA SA313345843GA SA313345844GA SA313345845GA SA313345846GA SA313345847GA SA313345848GA SA313345849GA SA313345850GA SA313345851GA SA313345852GA SA313345853GA SA313345854GA SA313345855GA SA313345856GA SA313345857GA SA313345858GA SA313345859GA SA313345861GA SA313345862GA SA313345863GA SA313345864GA SA313345865GA SA313345866GA SA313345867GA SA313345868GA SA313345869GA SA313345870GA SA313345871GA SA313345872GA SA313345873GA SA313345874GA SA313345875GA SA313345876GA SA313345877GA SA313345878GA SA313345879GA SA313345880GA SA313345881GA SA313345882GA SA313345883GA SA313345884GA SA313345885GA SA313345886GA SA313345887GA SA313345888GA SA313345889GA SA313345890GA SA313345891GA SA313345892GA SA313345893GA SA313345894GA SA313345895GA SA313345896GA SA313345897GA SA313345898GA SA313345899GA SA313345900GA SA313345901GA SA313345902GA SA313345903GA SA313345904GA SA313345905GA SA313345906GA SA313345907GA SA313345908GA SA313345909GA SA313345910GA SA313345911GA SA313345912GR SA313345913GA SA313345914GA SA313345915GA SA313345916GA SA313345917GA SA313345918GA SA313345919GA SA313345920GA SA313345921GA SA313345922GA SA313345923GA SA313345924GA SA313345925GA SA313345926GA SA313345927GA SA313345928GA SA313345929GA SA313345930GA SA313345931GA SA313345932GA SA313345933GA SA313345934GA SA313345935GA SA313345936GA SA313345937GA SA313345938GA SA313345939GA SA313345940GA SA313345941GA SA313345942GA SA313345944GA SA313345945GA SA313345946GA SA313345947GA SA313345948GA SA313345949GA SA313345950GA SA313345951GA SA313345952GA SA313345953GA SA313345954GA SA313345955GA SA313345956GA SA313345957GA SA313345958GA SA313345959GA SA313345960GA SA313345961GA SA313345962GA SA313345963GA SA313345964GA SA313345965GA SA313345966GA SA313345967GA SA313345968GA SA313345969GR SA313345970GA SA313345971GA SA313345972GA SA313345973GA SA313345974GA SA313345975GA SA313345976GA SA313345977GA SA313345978GA SA313345979GA SA313345980GA SA313345981GA SA313345982GA SA313345983GA SA313345984GA SA313345985GA SA313345986GA SA313345987GA SA313345988GA SA313345989GA SA313345991GA SA313345992GA SA313345993GA SA313345994GA SA313345995GA SA313345996GA SA313345997GA SA313345998GA SA313345999GA SA313346000GA SA313346001GA SA313346003GA SA313346004GA SA313346005GA SA313346006GA SA313346007GA SA313346008GA SA313346009GA SA313346010GA SA313346011GA SA313346012GA SA313346013GA SA313346014GA SA313346015GA SA313346016GR SA313346017GA SA313346018GA SA313346019GA SA313346020GA SA313346021GA SA313346023GA SA313346024GA SA313346025GA SA313346026GA SA313346027GA SA313346028GA SA313346029GA SA313346030GA SA313346031GA SA313346032GA SA313346033GA SA313346034GA SA313346035GA SA313346036GA SA313346037GA SA313346038GA SA313346039GA SA313346040GA SA313346041GA SA313346043GA SA313346044GA SA313346045GA SA313346046GA SA313346047GA SA313346048GA SA313346049GA SA313346050GA SA313346051GA SA313346052GA SA313346053GA SA313346054GA SA313346055GA SA313346056GA SA313346057GA SA313346058GA SA313346059GA SA313346060GA SA313346061GA SA313346062GA SA313346063GA SA313346065GA SA313346066GA SA313346067GA SA313346068GA SA313346069GA SA313346070GA SA313346071GA SA313346072GA SA313346073GA SA313346074GA SA313346075GA SA313346076GA SA313346078GA SA313346079GA SA313346080GA SA313346081GA SA313346082GA SA313346083GA SA313346084GA SA313346085GA SA313346086GA SA313346087GA SA313346088GA SA313346089GA SA313346090GA SA313346091GA SA313346092GA SA313346093GA SA313346094GA SA313346095GA SA313346096GA SA313346097GA SA313346098GA SA313346099GA SA313346100GA SA313346101GA SA313346102GA SA313346103GA SA313346104GA SA313346105GA SA313346106GA SA313346107GA SA313346108GA SA313346109GA SA313346110GA SA313346111GA SA313346112GA SA313346113GA SA313346114GA SA313346115GA SA313346116GA SA313346117GA SA313346118GA SA313346119GA SA313346120GA SA313346121GA SA313346122GA SA313346123GA SA313346124GA SA313346125GA SA313356126GA SA313356128GA SA313356129GA SA313356130GA SA313356131GA SA313356132GA SA313356133GA SA313356134GA SA313356135GA SA313356136GA SA313356137GA SA313356139GA SA313356140GA SA313356141GA SA313356142GA SA313356143GA SA313356144GA SA313356145GA SA313356146GA SA313356147GA SA313356148GA SA313356149GA SA313356150GA SA313356151GA SA313356152GA SA313356153GA SA313356154GA SA313356155GA SA313356156GA SA313356157GA SA313356158GA SA313356159GA SA313356160GA SA313356161GA SA313356162GA SA313356163GA SA313356164GA SA313356165GA SA313356166GA SA313356167GA SA313356168GA SA313356169GR SA313356170GA SA313356171GA SA313356172GA SA313356173GA SA313356174GA SA313356175GA SA313356176GA SA313356177GA SA313356178GA SA313356179GA SA313356181GA SA313356182GA SA313356183GA SA313356184GR SA313356185GA SA313356186GA SA313356187GA SA313356188GA SA313356189GA SA313356190GA SA313356191GA SA313356192GA SA313356194GA SA313356195GA SA313356196GA SA313356197GA SA313356198GA SA313356199GA SA313356200GA SA313356201GA SA313356202GA SA313356203GA SA313356204GA SA313356205GA SA313356206GA SA313356207GA SA313356209GA SA313356210GA SA313356211GA SA313356212GA SA313356213GA SA313356214GA SA313356215GA SA313356216GA SA313356217GA SA313356218GA SA313356219GA SA313356220GA SA313356221GA SA313356222GA SA313356223GA SA313356224GA SA313356225GA SA313356226GA SA313356227GA SA313356228GA SA313356229GA SA313356230GA SA313356231GA SA313356232GA SA313356233GA SA313356234GA SA313356235GA SA313356236GA SA313356237GA SA313356238GA SA313356239GA SA313356240GA SA313356241GA SA313356242GA SA313356243GA SA313356244GR SA313356245GA SA313356246GA SA313356247GA SA313356248GA SA313356249GA SA313356250GA SA313356251GA SA313356252GA SA313356253GA SA313356254GA SA313356255GA SA313356256GA SA313356257GA SA313356258GA SA313356259GA SA313356260GA SA313356261GA SA313356262GA SA313356263GA SA313356265GR SA313356266GA SA313356267GA SA313356268GA SA313356269GA SA313356270GA SA313356271GA SA313356272GR SA313356273GA SA313356274GR SA313356275GA SA313356276GA SA313356277GA SA313356278GA SA313356279GA SA313356280GA SA313356281GA SA313356282GA SA313356283GA SA313356284GA SA313356285GA SA313356286GA SA313356287GA SA313356288GA SA313356289GA SA313356290GA SA313356291GA SA313356292GA SA313356293GA SA313356294GA SA313356295GA SA313356296GA SA313356297GA SA313356298GA SA313356299GA SA313356300GA SA313356301GA SA313356302GA SA313356303GA SA313356304GA SA313356305GA SA313356306GA SA313356307GA SA313356308GA SA313356309GA SA313356310GA SA313356311GA SA313356312GA SA313356313GA SA313356315GA SA313356316GA SA313356317GA SA313356318GA SA313356319GA SA313356320GA SA313356321GA SA313356322GA SA313356323GA SA313356324GA SA313356325GA SA313356326GA SA313356327GR SA313356328GA SA313356329GA SA313356330GA SA313356331GA SA313356332GA SA313356333GA SA313356334GA SA313356335GA SA313356337GA SA313356338GA SA313356339GA SA313356340GA SA313356341GA SA313356342GA SA313356343GA SA313356344GA SA313356345GA SA313356346GA SA313356347GA SA313356348GA SA313356349GA SA313356350GA SA313356351GA SA313356352GA SA313356354GA SA313356355GA SA313356356GA SA313356357GA SA313356358GA SA313356359GA SA313356360GA SA313356362GA SA313356363GA SA313356364GA SA313356365GA SA313356366GA SA313356367GA SA313356368GA SA313356369GA SA313356370GA SA313356371GA SA313356372GA SA313356373GA SA313356374GA SA313356375GA SA313356376GA SA313356377GA SA313356378GA SA313356380GA SA313356381GA SA313356382GA SA313356383GA SA313356384GA SA313356385GA SA313356386GA SA313356387GA SA313356389GA SA313356390GA SA313356391GA SA313356392GA SA313356393GA SA313356394GA SA313356395GA SA313356396GA SA313356397GA SA313356398GA SA313356399GA SA313356400GA SA313356401GA SA313356402GA SA313356403GA SA313356404GA SA313356405GA SA313356406GA SA313356407GA SA313356408GA SA313356409GA SA313356410GA SA313356411GA SA313356412GA SA313356413GA SA313356414GA SA313356415GA SA313356416GA SA313356417GA SA313356418GA SA313356419GA SA313356420GA SA313356421GA SA313356422GA SA313356423GA SA313356424GA SA313356425GA SA313356426GR SA313356427GA SA313356428GA SA313356429GA SA313356430GA SA313356431GA SA313356432GA SA313356433GA SA313356434GA SA313356435GA SA313356436GA SA313356437GA SA313356438GA SA313356439GA SA313356440GA SA313356441GA SA313356442GA SA313356443GA SA313356444GA SA313356445GA SA313356446GA SA313356447GA SA313356448GA SA313356449GA SA313356450GA SA313356451GA SA313356452GA SA313356453GA SA313356454GA SA313356455GA SA313356456GA SA313356457GA SA313356458GA SA313356459GA SA313356460GA SA313356461GA SA313356462GA SA313356463GA SA313356464GA SA313356465GA SA313356467GA SA313356468GA SA313356469GA SA313356470GA SA313356471GA SA313356472GA SA313356473GA SA313356474GA SA313356475GA SA313356476GA SA313356477GA SA313356478GA SA313356479GA SA313356480GA SA313356481GA SA313356482GA SA313356483GA SA313356484GA SA313356485GA SA313356486GA SA313356487GA SA313356488GA SA313356489GA SA313356490GA SA313356491GA SA313356492GA SA313356493GA SA313356494GA SA313356495GA SA313356496GA SA313356497GA SA313356498GA SA313356499GA SA313356500GA SA313356501GA SA313356502GA SA313356503GA SA313356504GA SA313356505GA SA313356506GA SA313356507GA SA313356508GA SA313356509GA SA313356510GA SA313356511GA SA313356512GA SA313356513GA SA313356514GA SA313356515GA SA313356516GA SA313356517GA SA313356518GR SA313356519GA SA313356520GA SA313356521GA SA313356522GA SA313356523GR SA313356524GA SA313356525GA SA313356526GR SA313356527GA SA313356528GA SA313356529GA SA313356530GA SA313356531GA SA313356532GA SA313356533GA SA313356534GA SA313356535GA SA313356536GA SA313356537GA SA313356539GA SA313356540GA SA313356541GA SA313356542GA SA313356543GA SA313356544GA SA313356545GA SA313356546GA SA313356547GA SA313356548GA SA313356549GA SA313356550GA SA313356551GA SA313356552GA SA313356553GA SA313356554GA SA313356555GA SA313356556GA SA313356557GA SA313356558GA SA313356559GA SA313356560GA SA313356561GA SA313356562GA SA313356563GA SA313356564GA SA313356565GA SA313356566GA SA313356567GA SA313356568GA SA313356569GA SA313356570GA SA313356571GA SA313356572GA SA313356573GA SA313356574GA SA313356575GA SA313356576GA SA313356577GA SA313356578GA SA313356579GA SA313356580GA SA313356581GA SA313356582GA SA313356583GA SA313356584GA SA313356585GA SA313356586GA SA313356587GA SA313356588GA SA313356589GA SA313356590GA SA313356591GA SA313356592GA SA313356593GA SA313356594GA SA313356595GA SA313356596GA SA313356597GA SA313356598GA SA313356599GA SA313356600GA SA313356601GA SA313356602GA SA313356603GA SA313356604GA SA313356605GA SA313356606GA SA313356607GA SA313356608GA SA313356609GA SA313356610GA SA313356611GA SA313356612GA SA313356613GA SA313356614GA SA313356615GA SA313356616GA SA313356617GA SA313356618GA SA313356619GA SA313356620GA SA313356621GA SA313356622GA SA313356623GA SA313356624GA SA313356625GA SA313356627GA SA313356629GA SA313356630GA SA313356631GA SA313356632GA SA313356633GA SA313356634GA SA313356635GA SA313356636GA SA313356637GA SA313356638GA SA313356639GA SA313356640GA SA313356641GA SA313356642GA SA313356643GA SA313356644GA SA313356645GA SA313356646GA SA313356647GA SA313356648GA SA313356649GA SA313356650GA SA313356651GA SA313356652GA SA313356653GA SA313356654GA SA313356655GA SA313356656GA SA313356657GA SA313356658GA SA313356659GA SA313356660GA SA313356661GA SA313356662GA SA313356663GA SA313356664GA SA313356665GA SA313356666GA SA313356667GA SA313356668GA SA313356669GA SA313356670GA SA313356671GA SA313356672GA SA313356674GA SA313356675GA SA313356676GA SA313356677GA SA313356678GA SA313356679GA SA313356680GA SA313356681GA SA313356682GA SA313356683GA SA313356684GA SA313356685GA SA313356686GA SA313356687GA SA313356688GA SA313356689GA SA313356690GA SA313356691GA SA313356692GA SA313356693GA SA313356694GA SA313356695GA SA313356697GA SA313356698GA SA313356699GA SA313356700GA SA313366701GA SA313366702GA SA313366703GA SA313366704GA SA313366705GA SA313366706GA SA313366707GA SA313366708GA SA313366709GA SA313366710GA SA313366711GA SA313366712GA SA313366713GA SA313366714GA SA313366715GA SA313366716GA SA313366718GA SA313366720GA SA313366721GA SA313366723GA SA313366724GA SA313366725GA SA313366726GA SA313366727GA SA313366728GA SA313366730GA SA313366731GA SA313366732GA SA313366733GA SA313366734GA SA313366735GA SA313366736GA SA313366737GA SA313366738GA SA313366739GA SA313366740GA SA313366741GA SA313366742GA SA313366743GA SA313366744GA SA313366745GA SA313366746GA SA313366747GA SA313366748GA SA313366749GA SA313366750GA SA313366751GA SA313366752GA SA313366753GA SA313366754GA SA313366755GA SA313366758GA SA313366759GA SA313366760GA SA313366761GA SA313366762GA SA313366763GA SA313366764GA SA313366765GA SA313366766GA SA313366767GA SA313366768GA SA313366769GA SA313366770GA SA313366771GA SA313366772GA SA313366773GA SA313366774GA SA313366775GA SA313366776GA SA313366777GA SA313366778GA SA313366779GA SA313366780GA SA313366781GA SA313366782GA SA313366783GA SA313366784GA SA313366785GA SA313366787GA SA313366788GA SA313366789GA SA313366790GA SA313366791GA SA313366792GA SA313366793GA SA313366794GA SA313366795GA SA313366796GA SA313366797GA SA313366798GA SA313366800GA SA313366801GA SA313366802GA SA313366803GA SA313366804GA SA313366805GA SA313366806GA SA313366807GA SA313366808GA SA313366809GA SA313366810GA SA313366811GA SA313366812GA SA313366813GA SA313366814GA SA313366815GA SA313366816GA SA313366817GA SA313366818GA SA313366819GA SA313366820GA SA313366822GA SA313366823GA SA313366824GA SA313366825GA SA313366826GA SA313366827GA SA313366828GA SA313366829GA SA313366830GA SA313366831GA SA313366832GA SA313366833GA SA313366834GA SA313366835GA SA313366836GA SA313366837GA SA313366838GA SA313366839GA SA313366840GA SA313366841GA SA313366842GR SA313366843GA SA313366844GA SA313366845GA SA313366846GA SA313366847GA SA313366848GA SA313366849GA SA313366850GA SA313366851GA SA313366852GA SA313366853GA SA313366854GA SA313366855GA SA313366856GA SA313366857GA SA313366858GA SA313366859GA SA313366860GA SA313366861GA SA313366862GA SA313366863GA SA313366864GA SA313366865GA SA313366866GA SA313366867GA SA313366868GA SA313366869GA SA313366870GA SA313366872GA SA313366873GA SA313366874GA SA313366875GA SA313366876GA SA313366877GA SA313366878GA SA313366879GA SA313366880GA SA313366881GA SA313366882GA SA313366883GA SA313366884GA SA313366885GA SA313366886GA SA313366887GA SA313366888GA SA313366889GA SA313366890GA SA313366891GA SA313366892GA SA313366893GA SA313366894GA SA313366895GA SA313366896GA SA313366897GA SA313366898GA SA313366899GA SA313366900GA SA313366901GA SA313366902GA SA313366903GA SA313366904GA SA313366905GA SA313366906GA SA313366907GA SA313366908GA SA313366909GA SA313366910GA SA313366911GA SA313366912GA SA313366913GA SA313366914GA SA313366915GA SA313366916GA SA313366917GA SA313366918GA SA313366919GA SA313366920GA SA313366921GA SA313366922GA SA313366923GA SA313366924GA SA313366926GA SA313366927GA SA313366928GA SA313366929GA SA313366930GA SA313366931GA SA313366932GA SA313366933GA SA313366934GA SA313366935GA SA313366936GA SA313366937GA SA313366938GA SA313366939GA SA313366940GA SA313366941GA SA313366942GA SA313366943GA SA313366944GA SA313366945GA SA313366946GA SA313366947GA SA313366948GA SA313366949GA SA313366950GA SA313366951GA SA313366952GA SA313366953GA SA313366954GA SA313366955GR SA313366956GA SA313366957GA SA313366958GA SA313366959GA SA313366960GA SA313366961GA SA313366962GA SA313366963GA SA313366964GA SA313366965GA SA313366966GA SA313366967GA SA313366968GA SA313366969GA SA313366970GA SA313366971GA SA313366972GA SA313366973GA SA313366974GA SA313366975GA SA313366976GA SA313366977GA SA313366978GA SA313366979GA SA313366980GA SA313366981GA SA313366982GA SA313366983GA SA313366984GA SA313366985GA SA313366986GA SA313366987GA SA313366988GA SA313366989GA SA313366990GA SA313366991GA SA313366992GA SA313366993GA SA313366994GA SA313366995GA SA313366996GA SA313366997GA SA313366998GA SA313366999GA SA313367000GA SA313367001GA SA313367002GA SA313367003GA SA313367004GA SA313367005GA SA313367006GA SA313367007GA SA313367008GA SA313367009GA SA313367010GA SA313367011GA SA313367012GA SA313367013GA SA313367014GA SA313367015GA SA313367016GA SA313367017GA SA313367018GA SA313367019GA SA313367020GA SA313367021GA SA313367022GA SA313367023GA SA313367024GA SA313367025GA SA313367026GA SA313367027GA SA313367028GA SA313367029GA SA313367030GA SA313367031GA SA313367032GA SA313367033GA SA313367034GA SA313367035GA SA313367036GA SA313367037GA SA313367038GA SA313367039GA SA313367040GA SA313367041GA SA313367042GA SA313367043GA SA313367044GA SA313367045GA SA313367046GA SA313367047GR SA313367048GA SA313367049GA SA313367050GA SA313367051GA SA313367052GA SA313367053GA SA313367054GA SA313367055GA SA313367056GA SA313367057GA SA313367058GR SA313367059GA SA313367060GA SA313367061GA SA313367062GR SA313367063GA SA313367064GA SA313367065GA SA313367066GA SA313367067GA SA313367068GA SA313367069GA SA313367070GA SA313367071GA SA313367072GA SA313367073GA SA313367074GA SA313367075GA SA313367076GA SA313367077GA SA313367078GA SA313367079GA SA313367080GA SA313367081GR SA313367082GA SA313367083GA SA313367084GA SA313367085GA SA313367086GA SA313367087GA SA313367088GA SA313367089GA SA313367090GA SA313367091GA SA313367092GA SA313367093GA SA313367094GA SA313367095GA SA313367096GA SA313367097GA SA313367098GA SA313367099GA SA313367100GA SA313367101GA SA313367102GA SA313367103GA SA313367104GA SA313367105GA SA313367106GA SA313367107GA SA313367108GA SA313367109GA SA313367110GA SA313367111GA SA313367112GA SA313367113GA SA313367114GA SA313367115GA SA313367116GA SA313367117GA SA313367118GA SA313367119GA SA313367120GA SA313367121GA SA313367122GA SA313367123GA SA313367124GA SA313367125GA SA313367126GA SA313367127GA SA313367128GA SA313367129GA SA313367130GA SA313367131GA SA313367132GA SA313367133GA SA313367134GA SA313367135GA SA313367136GA SA313367137GA SA313367138GA SA313367139GA SA313367140GA SA313367141GA SA313367142GA SA313367143GA SA313367144GA SA313367145GA SA313367146GA SA313367147GA SA313367148GA SA313367149GA SA313367150GA SA313367151GA SA313367152GA SA313367153GA SA313367154GA SA313367155GA SA313367156GA SA313367157GA SA313367158GA SA313367159GA SA313367160GA SA313367161GA SA313367162GA SA313367163GA SA313367164GA SA313367165GA SA313367166GA SA313367167GA SA313367168GA SA313367169GA SA313367170GA SA313367171GA SA313367172GA SA313367173GA SA313367174GA SA313367175GA SA313367176GA SA313367177GR SA313367178GA SA313367179GA SA313367180GA SA313367181GA SA313367182GR SA313367183GA SA313367184GA SA313367185GA SA313367186GA SA313367187GA SA313367188GA SA313367189GA SA313367190GR SA313367191GA SA313367192GA SA313367193GA SA313367194GA SA313367195GA SA313367196GA SA313367197GA SA313367198GA SA313367199GA SA313367200GA SA313367201GA SA313367202GA SA313367203GA SA313367204GA SA313367205GA SA313367206GA SA313367207GR SA313367208GA SA313367209GA SA313367210GA SA313367211GA SA313367212GA SA313367213GR SA313367214GA SA313367215GA SA313367216GR SA313367217GA SA313367218GA SA313367219GA SA313367220GA SA313367221GA SA313367222GR SA313367223GA SA313367224GR SA313367225GR SA313367226GA SA313367227GA SA313367228GA SA313367229GA SA313367230GA SA313367231GA SA313367232GA SA313367233GA SA313367234GA SA313367235GR SA313367236GA SA313367237GA SA313367238GA SA313367239GA SA313367240GA SA313367241GA SA313367242GA SA313367243GA SA313367244GA SA313367245GA SA313367246GA SA313367247GA SA313367248GA SA313367249GA SA313367250GA SA313367251GA SA313367252GA SA313367253GA SA313367254GA SA313367255GA SA313367256GA SA313367257GA SA313367258GA SA313367259GA SA313367260GA SA313367261GA SA313367262GA SA313367263GA SA313367264GA SA313367265GA SA313367496GA SA313367497GA SA313367498GA SA313367499GA SA313367500GA SA313367502GA SA313367504GA SA313367505GA SA313367506GA SA313367507GA SA313367508GA SA313367509GA SA313367512GA SA313367513GA SA313367514GA SA313367515GA SA313367516GA SA313367518GA SA313367519GA SA313367521GA SA313367522GA SA313367523GA SA313367524GA SA313367527GA SA313367528GA SA313367530GA SA313367532GA SA313367533GA SA313367536GR SA313367537GA SA313367540GA SA313367543GA SA313367544GA SA313367546GA SA313367547GA SA313367548GR SA313367549GA SA313367550GA SA313367552GA SA313367553GA SA313367554GA SA313367555GA SA313367556GA SA313367563GA SA313367565GA SA313367566GA SA313367569GA SA313367572GA SA313367573GA SA313367575GA SA313367576GA SA313367579GA SA313367580GA SA313367582GA SA313367586GA SA313367587GA SA313367588GA SA313367589GA SA313367590GA SA313367591GA SA313367593GA SA313367594GA SA313367595GA SA313367596GA SA313367597GA SA313367598GA SA313367599GA SA313367600GA SA313377601GA SA313377602GA SA313377603GA SA313377604GA SA313377605GA SA313377606GA SA313377607GA SA313377608GA SA313377609GA SA313377610GA SA313377611GA SA313377612GA SA313377613GA SA313377614GA SA313377615GA SA313377616GA SA313377617GA SA313377618GA SA313377619GA SA313377620GA SA313377621GA SA313377622GA SA313377623GA SA313377624GA SA313377625GA SA313377626GA SA313377627GA SA313377628GA SA313377629GA SA313377630GA SA313377631GA SA313377632GA SA313377633GA SA313377634GA SA313377635GA SA313377636GA SA313377637GA SA313377638GA SA313377639GA SA313377640GA SA313377641GA SA313377642GA SA313377643GA SA313377644GA SA313377645GA SA313377650GA SA313377665GA SA313377667GR SA313377685GA SA313377687GA SA313377688GA SA313377689GA SA313377691GA SA313377693GA SA313377694GA SA313377695GA SA313377699GA SA313377701GA SA313377702GA SA313377703GA SA313377704GA SA313377705GA SA313377706GA SA313377707GA SA313377708GA SA313377709GA SA313377710GR SA313377711GA SA313377712GA SA313377716GA SA313377717GA SA313377719GA SA313377720GA SA313377721GA SA313377722GA SA313377723GA SA313377724GA SA313377725GA SA313377726GA SA313377727GA SA313377728GA SA313377729GA SA313377730GA SA313377731GA SA313377732GA SA313377733GA SA313377734GA SA313377735GA SA313377736GA SA313377737GA SA313377738GR SA313377739GR SA313377740GR SA313377741GA SA313377742GA SA313377743GA SA313377744GA SA313377745GA SA313377746GA SA313377747GA SA313377748GA SA313377749GR SA313377750GR SA313377751GA SA313377752GA SA313377753GA SA313377754GA SA313377755GA SA313377756GA SA313377757GA SA313377758GA SA313377760GA SA313377761GA SA313377762GA SA313377763GA SA313377764GA SA313377765GA SA313377766GA SA313377767GA SA313377768GA SA313377769GA SA313377770GA SA313377771GA SA313377772GR SA313377773GA SA313377774GA SA313377775GR SA313377776GA SA313377777GA SA313377778GR SA313377779GR SA313377780GA SA313377781GA SA313377782GR SA313377783GA SA313377784GA SA313377785GA SA313377786GA SA313377787GA SA313377788GA SA313377789GA SA313377790GA SA313377791GA SA313377792GA SA313377793GR SA313377794GA SA313377795GR SA313377796GA SA313377797GA SA313377798GA SA313377799GA SA313377800GA SA313377801GA SA313377802GA SA313377803GR SA313377804GA SA313377805GA SA313377806GA SA313377807GA SA313377808GA SA313377809GA SA313377810GA SA313377811GA SA313377812GR SA313377813GR SA313377814GR SA313377815GA SA313377816GA SA313377817GA SA313377818GA SA313377819GA SA313377820GR SA313377821GA SA313377822GA SA313377823GA SA313377824GA SA313377825GR SA313377826GA SA313377827GA SA313377828GA SA313377829GR SA313377830GA SA313377831GA SA313377832GA SA313377833GA SA313377834GA SA313377835GA SA313377836GA SA313377837GA SA313377838GA SA313377839GA SA313377840GA SA313377841GA SA313377842GA SA313377844GA SA313377845GA SA313377846GA SA313377847GA SA313377848GA SA313377849GA SA313377850GA SA313377851GA SA313377852GA SA313377853GA SA313377854GA SA313377855GA SA313377856GA SA313377857GA SA313377858GA SA313377859GA SA313377860GA SA313377861GA SA313377862GA SA313377863GA SA313377864GA SA313377865GA SA313377866GA SA313377867GA SA313377868GA SA313377869GA SA313377870GA SA313377871GA SA313377872GA SA313377873GA SA313377874GA SA313377875GA SA313377876GA SA313377877GA SA313377878GA SA313377879GA SA313377880GA SA313377881GA SA313377882GA SA313377883GA SA313377884GA SA313377885GA SA313377886GA SA313377887GA SA313377888GA SA313377889GA SA313377890GA SA313377891GA SA313377892GA SA313377893GA SA313377894GA SA313377895GA SA313377896GA SA313377897GA SA313377898GA SA313377899GA SA313377900GA SA313377901GA SA313377902GA SA313377903GA SA313377904GA SA313377905GA SA313377906GA SA313377908GA SA313377909GA SA313377910GA SA313377911GA SA313377912GA SA313377913GA SA313377914GA SA313377915GA SA313377916GA SA313377917GA SA313377918GA SA313377920GA SA313377921GA SA313377922GA SA313377923GA SA313377924GA SA313377925GA SA313377926GA SA313377927GA SA313377928GA SA313377929GA SA313377930GA SA313377931GA SA313377932GA SA313377933GA SA313377934GA SA313377935GA SA313377936GA SA313377937GA SA313377938GA SA313377939GA SA313377940GA SA313377941GA SA313377942GA SA313377943GA SA313377944GA SA313377945GA SA313377946GA SA313377947GA SA313377948GA SA313377949GA SA313377950GA SA313377951GA SA313377952GA SA313377953GA SA313377954GA SA313377955GA SA313377956GA SA313377957GA SA313377958GA SA313377959GA SA313377960GA SA313377961GA SA313377962GA SA313377963GA SA313377964GA SA313377965GA SA313377966GA SA313377967GA SA313377968GA SA313377969GA SA313377970GA SA313377971GA SA313377972GA SA313377973GA SA313377974GA SA313377975GA SA313377976GA SA313377977GA SA313377978GA SA313377979GA SA313377980GA SA313377981GA SA313377982GA SA313377983GA SA313377984GA SA313377985GA SA313377986GA SA313377987GA SA313377988GA SA313377989GA SA313377990GA SA313377991GA SA313377992GA SA313377993GA SA313377994GA SA313377995GA SA313377996GA SA313377997GA SA313377998GA SA313377999GA SA313378000GA SA313378001GA SA313378002GA SA313378003GA SA313378004GA SA313378005GA SA313378006GA SA313378007GA SA313378008GA SA313378009GA SA313378010GA SA313378011GA SA313378012GA SA313378013GA SA313378014GA SA313378015GA SA313378016GA SA313378017GA SA313378018GA SA313378019GA SA313378020GA SA313378021GA SA313378022GA SA313378023GA SA313378024GA SA313378025GA SA313378026GA SA313378027GA SA313378029GA SA313378030GA SA313378032GA SA313378033GA SA313378034GA SA313378036GA SA313378038GA SA313378039GA SA313378040GA SA313378041GA SA313378044GA SA313378046GA SA313378047GA SA313378048GA SA313378049GA SA313378050GA SA313378051GR SA313378052GA SA313378053GA SA313378054GA SA313378055GA SA313378056GA SA313378057GA SA313378058GA SA313378059GA SA313378060GA SA313378061GA SA313378062GA SA313378063GA SA313378065GA SA313378066GA SA313378068GA SA313378069GA SA313378070GA SA313378071GA SA313378072GA SA313378073GA SA313378074GA SA313378075GA SA313378076GA SA313378077GA SA313378078GA SA313378079GA SA313378080GA SA313378081GA SA313378082GA SA313378083GA SA313378084GA SA313378085GA SA313378086GA SA313378087GA SA313378088GA SA313378089GA SA313378090GA SA313378091GA SA313378092GA SA313378094GA SA313378095GA SA313378096GA SA313378097GA SA313378098GA SA313378099GA SA313378100GA SA313378101GA SA313378102GA SA313378103GA SA313378104GA SA313378105GA SA313378106GA SA313378107GA SA313378108GA SA313378109GA SA313378110GA SA313378111GA SA313378112GA SA313378113GA SA313378114GA SA313378115GA SA313378116GA SA313378117GA SA313378118GA SA313378119GA SA313378120GA SA313378121GA SA313378122GA SA313378123GA SA313378124GA SA313378125GA SA313378126GA SA313378127GA SA313378128GA SA313378129GA SA313378130GA SA313378131GA SA313378132GA SA313378133GA SA313378134GA SA313378135GA SA313378136GA SA313378137GA SA313378138GA SA313378139GA SA313378140GA SA313378141GA SA313378142GA SA313378143GA SA313378144GA SA313378145GA SA313378146GA SA313378147GA SA313378148GA SA313378150GA SA313378151GA SA313378152GA SA313378153GA SA313378154GA SA313378155GA SA313378156GA SA313378157GA SA313378158GA SA313378159GA SA313378160GA SA313378161GA SA313378162GA SA313378163GA SA313378164GA SA313378165GA SA313378166GA SA313378167GA SA313378168GA SA313378169GA SA313378170GA SA313378171GA SA313378172GA SA313378173GA SA313378174GA SA313378175GA SA313378176GA SA313378177GA SA313378178GA SA313378179GA SA313378180GA SA313378181GA SA313378182GA SA313378183GA SA313378184GA SA313378185GA SA313378186GA SA313378187GA SA313378188GA SA313378189GA SA313378190GA SA313378191GA SA313378192GA SA313378193GA SA313378194GA SA313378195GA SA313378196GA SA313378197GA SA313378198GA SA313378199GA SA313378200GA SA313378201GA SA313378202GA SA313378203GA SA313378204GA SA313378205GA SA313378206GA SA313378207GA SA313378208GA SA313378209GA SA313378210GA SA313378211GA SA313378212GA SA313378213GA SA313378214GA SA313378215GA SA313378216GA SA313378217GA SA313378218GA SA313378219GA SA313378220GA SA313378221GA SA313378223GA SA313378224GA SA313378225GA SA313378228GA SA313378231GA SA313378233GA SA313378235GA SA313378238GA SA313378240GA SA313378253GA SA313378254GA SA313378255GA SA313378259GA SA313378260GA SA313378261GA SA313378262GA SA313378267GA SA313378270GA SA313378271GA SA313378272GA SA313378273GA SA313378274GA SA313378277GA SA313378278GA SA313378280GA SA313378283GA SA313378286GA SA313378293GA SA313378297GA SA313378299GA SA313378300GA SA313378302GA SA313378303GA SA313378304GA SA313378305GA SA313378306GA SA313378309GA SA313378317GA SA313378319GA SA313378320GA SA313378324GA SA313378328GA SA313378329GA SA313378332GA SA313378333GA SA313378334GA SA313378335GA SA313378337GA SA313378342GA SA313378343GA SA313378344GA SA313378347GA SA313378348GA SA313378349GA SA313378350GA SA313378354GA SA313378355GA SA313378357GA SA313378360GA SA313378362GA SA313378363GA SA313378364GA SA313378365GA SA313378366GA SA313378367GA SA313378370GA SA313378372GA SA313378373GA SA313378375GA SA313378378GA SA313378386GA SA313378390GA SA313378391GA SA313378394GA SA313378395GA SA313378396GA SA313378397GA SA313378398GA SA313378399GA SA313378400GA SA313378409GA SA313378414GA SA313378416GA SA313378422GA SA313378429GA SA313378431GA SA313378436GA SA313378439GA SA313378446GA SA313378455GA SA313378456GA SA313378459GA SA313378461GA SA313378462GA SA313378463GA SA313378464GA SA313378467GA SA313378468GA SA313378469GA SA313378470GA SA313378471GA SA313378473GA SA313378474GA SA313378477GA SA313378480GA SA313388495GA SA313388503GA SA313388505GA SA313388507GA SA313388508GA SA313388509GA SA313388510GA SA313388511GA SA313388512GA SA313388523GA SA313388530GA SA313388535GA SA313388549GA SA313388552GA SA313388558GA SA313388562GA SA313388563GA SA313388564GA SA313388565GA SA313388566GA SA313388567GA SA313388568GA SA313388569GA SA313388570GA SA313388571GA SA313388572GA SA313388573GA SA313388574GA SA313388575GA SA313388576GA SA313388577GA SA313388578GR SA313388579GA SA313388580GA SA313388581GA SA313388582GA SA313388583GA SA313388584GA SA313388585GA SA313388586GA SA313388587GA SA313388588GA SA313388589GA SA313388590GA SA313388591GA SA313388592GA SA313388593GA SA313388594GA SA313388595GA SA313388596GA SA313388597GA SA313388598GA SA313388599GA SA313388600GA SA313388601GA SA313388602GA SA313388603GA SA313388604GA SA313388605GA SA313388606GA SA313388607GA SA313388608GA SA313388609GA SA313388610GA SA313388611GA SA313388612GA SA313388613GA SA313388615GA SA313388616GA SA313388617GA SA313388618GR SA313388619GA SA313388620GA SA313388621GA SA313388622GA SA313388623GR SA313388624GA SA313388625GA SA313388626GA SA313388627GA SA313388628GA SA313388629GA SA313388630GA SA313388631GA SA313388632GA SA313388633GA SA313388634GA SA313388635GA SA313388636GA SA313388637GA SA313388638GA SA313388639GA SA313388640GA SA313388641GA SA313388642GA SA313388643GA SA313388644GA SA313388645GA SA313388646GA SA313388647GA SA313388648GA SA313388649GA SA313388650GA SA313388651GA SA313388652GA SA313388653GA SA313388654GA SA313388655GA SA313388656GA SA313388657GA SA313388658GA SA313388659GA SA313388661GA SA313388662GA SA313388663GA SA313388664GA SA313388665GA SA313388666GA SA313388667GA SA313388668GA SA313388669GA SA313388670GA SA313388671GA SA313388672GA SA313388673GA SA313388674GA SA313388675GA SA313388676GA SA313388677GA SA313388678GA SA313388679GA SA313388680GA SA313388681GA SA313388682GA SA313388689GA SA313388690GA SA313388693GA SA313388694GA SA313388695GA SA313388696GA SA313388699GA SA313388702GA SA313388703GA SA313388704GA SA313388707GA SA313388709GA SA313388710GA SA313388711GA SA313388712GA SA313388713GA SA313388714GA SA313388715GA SA313388716GA SA313388717GA SA313388718GA SA313388719GA SA313388720GA SA313388721GA SA313388722GA SA313388723GA SA313388724GA SA313388725GA SA313388726GA SA313388727GA SA313388728GA SA313388729GA SA313388730GA SA313388731GA SA313388732GA SA313388733GA SA313388734GA SA313388735GA SA313388736GA SA313388737GA SA313388738GA SA313388739GA SA313388740GA SA313388741GA SA313388742GA SA313388743GA SA313388744GA SA313388745GA SA313388746GA SA313388747GA SA313388748GA SA313388749GA SA313388750GA SA313388751GA SA313388752GA SA313388753GA SA313388754GA SA313388755GA SA313388756GA SA313388757GA SA313388758GA SA313388759GA SA313388760GA SA313388761GA SA313388762GA SA313388763GA SA313388764GA SA313388765GA SA313388766GA SA313388767GA SA313388768GA SA313388769GA SA313388770GA SA313388771GA SA313398772GA SA313398773GA SA313398774GA SA313398775GA SA313398776GA SA313398778GA SA313398779GA SA313398780GA SA313398781GA SA313398782GA SA313398783GA SA313398784GA SA313398785GA SA313398786GA SA313398787GA SA313398788GA SA313398789GA SA313398791GA SA313398792GA SA313398793GA SA313398794GA SA313398795GA SA313398796GA SA313398797GA SA313398798GA SA313398800GA SA313398801GA SA313398802GA SA313398803GA SA313398804GA SA313398805GA SA313398806GA SA313398807GA SA313398808GA SA313398809GA SA313398810GA SA313398811GA SA313398812GA SA313398813GA SA313398814GA SA313398815GA SA313398817GA SA313398819GA SA313398820GA SA313398821GA SA313398822GA SA313398823GA SA313398824GA SA313398825GA SA313398826GA SA313398827GA SA313398828GA SA313398829GA SA313398830GA SA313398831GA SA313398832GA SA313398833GA SA313398834GA SA313398835GA SA313398836GA SA313398837GA SA313398838GA SA313398839GA SA313398840GR SA313398841GA SA313398842GA SA313398843GA SA313398844GR SA313398845GA SA313398846GA SA313398847GA SA313398848GA SA313398849GA SA313398850GA SA313398851GA SA313398852GA SA313398853GA SA313398854GA SA313398855GA SA313398856GA SA313398857GA SA313398858GA SA313398859GA SA313398860GA SA313398861GA SA313398862GA SA313398863GA SA313398864GA SA313398865GA SA313398866GA SA313398867GA SA313398868GA SA313398869GA SA313398871GA SA313398872GA SA313398873GA SA313398874GA SA313398877GA SA313398878GA SA313398879GA SA313398880GA SA313398881GA SA313398882GA SA313398883GR SA313398884GA SA313398885GA SA313398887GA SA313398888GA SA313398889GA SA313398890GA SA313398891GA SA313398892GA SA313398893GA SA313398894GA SA313398895GA SA313398896GA SA313398897GA SA313398898GA SA313398899GA SA313398900GR SA313398901GA SA313398903GA SA313398904GA SA313398905GR SA313398906GA SA313398907GA SA313398908GA SA313398909GA SA313398910GA SA313398911GA SA313398912GA SA313398913GR SA313398914GA SA313398915GA SA313398916GA SA313398917GA SA313398919GA SA313398920GA SA313398921GA SA313398922GA SA313398923GR SA313398924GR SA313398925GA SA313398926GA SA313398927GA SA313398928GA SA313398929GA SA313398930GA SA313398931GA SA313398932GA SA313398933GA SA313398934GA SA313398935GA SA313398936GA SA313398937GA SA313398938GA SA313398940GA SA313398941GA SA313398942GA SA313398946GA SA313398947GA SA313398948GA SA313398949GA SA313398950GA SA313398951GA SA313398952GA SA313398953GA SA313398954GA SA313398955GR SA313398956GA SA313398957GA SA313398958GA SA313398959GA SA313398960GA SA313398961GA SA313398962GA SA313398963GR SA313398964GA SA313398965GA SA313398966GA SA313398967GA SA313398968GA SA313398969GR SA313398970GA SA313398971GA SA313398972GA SA313398973GA SA313398974GR SA313398975GA SA313398976GA SA313398977GA SA313398978GA SA313398979GA SA313398980GA SA313398981GA SA313408982GA SA313408984GA SA313408985GA SA313408986GA SA313408987GA SA313408988GR SA313408989GR SA313408990GA SA313408991GA SA313408992GR SA313408993GR SA313408994GA SA313408995GA SA313408996GA SA313408997GA SA313408998GA SA313409000GA SA313409001GA SA313409002GA SA313409003GA SA313409004GA SA313409005GA SA313409008GA SA313409010GA SA313409012GA SA313409013GA SA313409014GA SA313409015GA SA313409016GA SA313409017GA SA313409018GA SA313409019GA SA313409020GA SA313409022GA SA313409023GA SA313409024GA SA313409025GR SA313409026GR SA313409027GA SA313409028GA SA313409031GA SA313409033GA SA313409035GA SA313409036GA SA313409037GA SA313409038GA SA313409040GA SA313409041GA SA313409047GA SA313409051GA SA313409053GA SA313409054GA SA313409055GA SA313409056GA SA313409057GA SA313409058GA SA313409059GA SA313409060GA SA313409061GA SA313409062GA SA313409063GA SA313409064GA SA313409065GA SA313409066GA SA313409067GA SA313409068GA SA313409069GA SA313409070GA SA313409071GA SA313409072GA SA313409073GA SA313409074GA SA313409075GA SA313409076GA SA313409077GA SA313409078GA SA313409079GA SA313409080GA SA313409081GA SA313409082GA SA313409083GA SA313409084GA SA313409085GA SA313409086GA SA313409087GA SA313409088GA SA313409089GA SA313409090GA SA313409091GA SA313409092GA SA313409093GA SA313409094GA SA313409095GA SA313409096GA SA313409097GA SA313409098GA SA313409099GA SA313409100GA SA313409101GA SA313409102GA SA313409103GA SA313409104GA SA313409105GA SA313409106GA SA313409107GA SA313409108GA SA313409109GA SA313409110GA SA313409111GA SA313409112GA SA313409113GA SA313409114GA SA313409115GA SA313409116GA SA313409117GA SA313409118GA SA313409119GA SA313409120GA SA313409121GA SA313409122GA SA313409123GA SA313409124GA SA313409125GA SA313409126GA SA313409127GA SA313409128GA SA313409129GA SA313409130GA SA313409131GA SA313409132GA SA313409133GA SA313409134GA SA313409135GA SA313409136GA SA313409137GA SA313409139GA SA313409140GA SA313409141GA SA313409142GA SA313409144GA SA313409145GA SA313409146GA SA313409147GA SA313409148GA SA313409149GA SA313409150GA SA313409151GA SA313409152GA SA313409153GA SA313409154GA SA313409155GA SA313409156GA SA313409157GA SA313409158GA SA313409159GA SA313409161GA SA313409162GA SA313409163GA SA313409164GA SA313409165GA SA313409166GA SA313409168GA SA313409169GA SA313409170GA SA313409171GA SA313409173GA SA313409174GA SA313409175GA SA313409176GA SA313409177GA SA313409178GA SA313409179GA SA313409180GA SA313409181GA SA313409182GA SA313409183GA SA313409184GA SA313409185GA SA313409186GA SA313409187GA SA313409188GA SA313409189GA SA313409190GA SA313409191GA SA313409222GA SA313409223GA SA313409224GA SA313409225GA SA313409227GA SA313409230GA SA313409232GA SA313409233GA SA313409234GA SA313409235GA SA313409237GA SA313409238GA SA313409239GA SA313409240GA SA313409242GA SA313409244GA SA313409245GA SA313409248GA SA313409249GA SA313409250GA SA313409253GA SA313409255GA SA313409256GA SA313409258GA SA313409259GA SA313409261GA SA313409262GA SA313409265GA SA313409266GA SA313409273GA SA313409274GA SA313409276GA SA313409277GA SA313409278GA SA313409280GA SA313409282GA SA313409283GA SA313409285GA SA313409286GA SA313409287GA SA313409288GA SA313409289GA SA313409290GA SA313409291GA SA313409292GA SA313409293GA SA313409295GA SA313409296GA SA313409297GA SA313409298GA SA313409299GA SA313409300GA SA313409301GA SA313409302GA SA313409303GA SA313409305GA SA313409306GA SA313409307GA SA313409308GA SA313409309GA SA313409310GA SA313409311GA SA313409312GA SA313409313GA SA313409314GA SA313409315GA SA313409316GA SA313409317GA SA313409318GA SA313409319GA SA313409320GA SA313409321GA SA313409322GA SA313409324GA SA313409325GA SA313409326GA SA313409327GA SA313409328GA SA313409329GA SA313409330GA SA313409331GA SA313409332GA SA313409333GA SA313409334GA SA313409335GA SA313409336GA SA313409337GA SA313409338GA SA313409339GA SA313409340GA SA313409341GA SA313409342GA SA313409344GA SA313409345GA SA313409346GA SA313409347GA SA313409349GA SA313409350GA SA313409351GA SA313409352GA SA313409353GA SA313409354GA SA313409357GA SA313409358GA SA313409359GA SA313409360GA SA313409361GA SA313409363GA SA313409364GA SA313409365GA SA313409366GA SA313409367GA SA313409368GA SA313409369GA SA313409370GA sa313409371ga SA313409372GA SA313409373GA SA313409374GA SA313409375GA SA313409376GA SA313409377GA SA313409378GA SA313409379GA SA313409380GA SA313409381GA SA313409382GA SA313409383GA SA313409384GA SA313409385GA SA313409386GA SA313409387GA SA313409388GA SA313409389GA SA313409390GA SA313409391GA SA313409392GA SA313409393GA SA313409394GA SA313409396GA SA313409397GA SA313409398GA SA313409399GA SA313409400GA SA313409403GA SA313409404GA SA313409405GA SA313409407GA SA313409408GA SA313409409GA SA313409410GA SA313409411GA SA313409412GA SA313409413GA SA313409414GA SA313409415GA SA313409416GA SA313409417GA SA313409419GA SA313409421GA SA313409422GA SA313409423GA SA313409426GA SA313409427GA SA313409429GA SA313409431GA SA313410001GA SA313410002GA SA313410003GA SA313410004GA SA313410005GA SA313410006GA SA313410007GA SA313410008GA SA313410009GA SA313410010GA SA313410011GA SA313410012GA SA313410013GA SA313410014GA SA313410015GA SA313410016GA SA313410017GA SA313410018GA SA313410019GA SA313410020GA SA313410021GA SA313410022GA SA313410023GA SA313410024GA SA313410025GA SA313410026GA SA313410027GA SA313410028GA SA313410029GA SA313419433GA SA313419434GA SA313419438GA SA313419439GA SA313419441GA SA313419442GA SA313419443GA SA313419447GA SA313419449GA SA313419453GA SA313419454GA SA313419455GA SA313419456GA SA313419457GA SA313419458GA SA313419461GA SA313419462GA SA313419463GA SA313419464GA SA313419465GA SA313419466GA SA313419467GA SA313419468GA SA313419469GA SA313419470GA SA313419471GA SA313419472GA SA313419473GA SA313419474GA SA313419475GA SA313419476GA SA313419478GA SA313419479GA SA313419480GA SA313419481GA SA313419482GA SA313419483GA SA313419484GA SA313419485GA SA313419486GA SA313419487GA SA313419488GA SA313419489GA SA313419490GA SA313419491GA SA313419492GA SA313419494GA SA313419496GA SA313419497GA SA313419498GA SA313419499GA SA313419500GA SA313419503GA SA313419504GA SA313419505GA SA313419506GA SA313419507GA SA313419509GA SA313419510GA SA313419512GA SA313419513GA SA313419514GA SA313419516GA SA313419517GA SA313419520GA SA313419521GA SA313419522GA SA313419524GA SA313419525GA sa313419529ga SA313419531GA SA313419532GA SA313419533GA SA313419535GA SA313419536GA SA313419537GA SA313419538GA SA313419539GA SA313419540GA SA313419542GA SA313419543GA SA313419545GA SA313419547GA sa313419550ga SA313419551GA SA313419552GA SA313419553GA SA313419554GA SA313419555GA SA313419556GA SA313419557GA SA313419558GA SA313419559GA SA313419560GA SA313419561GA SA313419562GA SA313419563GA SA313419564GA SA313419565GA SA313419566GA SA313419567GA SA313419568GA SA313419569GA SA313419570GA SA313419571GA SA313419572GA SA313419573GA SA313419574GA SA313419575GA SA313419576GA SA313419577GA SA313419578GA SA313419579GA SA313419580GA SA313419581GA SA313419583GA SA313419584GA SA313419586GA SA313419587GA SA313419588GA SA313419589GA SA313419590GA SA313419591GA SA313419592GA SA313419593GA SA313419594GA SA313419595GA SA313419597GA SA313419598GA SA313419599GA SA313419600GA SA313419601GA SA313419602GA SA313419603GA SA313419604GA SA313419605GA SA313419606GA SA313419607GA SA313419610GA SA313419611GA SA313419612GA SA313419613GA SA313419614GA SA313419616GA SA313419617GA SA313419618GA SA313419619GA SA313419620GA SA313419621GA SA313419622GA SA313419623GA SA313419624GA SA313419625GA SA313419626GA SA313419627GA SA313419628GA SA313419629GA SA313419630GA SA313419631GA SA313419632GA SA313419633GA SA313419634GA SA313419635GA SA313419636GA SA313419637GA SA313419638GA SA313419639GA SA313419640GA SA313419641GA SA313419642GA SA313419644GA SA313419645GA SA313419646GA SA313419647GA SA313419648GA SA313419649GA SA313419650GA SA313419651GA SA313419652GA SA313419653GA SA313419654GA SA313419655GA SA313419656GA SA313419657GA SA313419658GA SA313419659GA SA313419660GA SA313419661GA SA313419662GA SA313419664GA SA313419665GA SA313419666GA SA313419667GA SA313419668GA SA313419669GA SA313419670GA SA313419671GA SA313419672GA SA313419673GA SA313419674GA SA313419675GA SA313419676GA SA313419677GA SA313419678GA SA313419679GA SA313419680GA SA313419681GA SA313419682GA SA313419683GA SA313419684GA SA313419685GA SA313419686GA SA313419687GA SA313419688GA SA313419689GA SA313419690GA SA313419691GA SA313419692GA SA313419693GA SA313419694GA SA313419695GA SA313419696GA SA313419697GA SA313419698GA SA313419699GA SA313419700GA SA313419701GA SA313419702GA SA313419703GA SA313419704GA SA313419705GA SA313419708GA SA313419709GA SA313419710GA SA313419711GA SA313419712GA SA313419713GA SA313419714GA SA313419715GA SA313419716GA SA313419717GA SA313419718GA SA313419719GA SA313419720GA SA313419721GA SA313419722GA SA313419723GA SA313419724GA SA313419725GA SA313419726GA SA313419727GA SA313419728GA SA313419729GA SA313419730GA SA313419731GA SA313419732GA SA313419733GA SA313419735GA SA313419744GA SA313419748GA SA313419758GA SA313419764GA SA313419765GA SA313419766GA SA313419767GA SA313419768GA SA313419769GA SA313419770GA SA313419771GA SA313419772GA SA313419773GA SA313419774GA SA313419775GA SA313419776GA SA313419777GA SA313419778GA SA313419780GA SA313419781GA SA313419782GA SA313419783GA SA313419784GA SA313419785GA SA313419787GA SA313419788GA SA313419789GA SA313419790GA SA313419793GA SA313419794GA SA313419795GA SA313419796GA SA313419797GA SA313419798GA SA313419799GA SA313419800GA SA313419801GA SA313419802GA SA313419803GA SA313419804GA SA313419805GA SA313419806GA SA313419807GA SA313419808GA SA313419809GA SA313419810GA SA313419811GA SA313419812GA SA313419813GA SA313419814GA SA313419815GA SA313419816GA SA313419817GA SA313419818GA SA313419819GA SA313419820GA SA313419821GA SA313419822GA SA313419823GA SA313419824GA SA313419825GA SA313419826GA SA313419827GA SA313419828GA SA313419829GA SA313419830GA SA313419831GA SA313419832GA SA313419833GA SA313419834GA SA313419835GA SA313419836GA SA313419837GA SA313419839GA SA313419840GA SA313419841GA SA313419842GA SA313419843GA SA313419844GA SA313419845GA SA313419846GA SA313419847GA SA313419848GA SA313419849GA SA313419850GA SA313419851GA SA313419852GA SA313419853GA SA313419854GA SA313419855GA SA313419856GA SA313419857GA SA313419858GA SA313419859GA SA313419860GA SA313419861GA SA313419862GA SA313419863GA SA313419864GA SA313419865GA SA313419866GA SA313419867GA SA313419868GA SA313419869GA SA313419870GA SA313419871GA SA313419872GA SA313419873GA SA313419874GA SA313419876GA SA313419877GA SA313419878GA SA313419879GA SA313419880GA SA313419881GA SA313419882GA SA313419883GA SA313419884GA SA313419885GA SA313419886GA SA313419887GA SA313419888GA SA313419889GA SA313419890GA SA313419891GA SA313419892GA SA313419893GA SA313419894GA SA313419895GA SA313419896GA SA313419897GA SA313419898GA SA313419899GA SA313419900GA SA313419901GA SA313419902GA SA313419903GA SA313419905GA SA313419906GA SA313419907GA SA313419908GA SA313419909GA SA313419910GA SA313419911GA SA313419912GA SA313419913GA SA313419914GA SA313419915GA SA313419916GA SA313419917GA SA313419918GA SA313419919GA SA313419920GA SA313419921GA SA313419922GA SA313419923GA SA313419924GA SA313419925GA SA313419926GA SA313419927GA SA313419928GA SA313419929GA SA313419930GA SA313419931GA SA313419932GA SA313419933GA SA313419934GA SA313419935GA SA313419936GA SA313419937GA SA313419938GA SA313419939GA SA313419940GA SA313419941GA SA313419942GA SA313419944GA SA313419945GA SA313419946GA SA313419947GA SA313419948GA SA313419949GA SA313419950GA SA313419951GA SA313419952GA SA313419953GA SA313419954GA SA313419955GA SA313419956GA SA313419957GA SA313419958GA SA313419959GA SA313419960GA SA313419961GA SA313419962GA SA313419963GA SA313419964GA SA313419965GA SA313419966GA SA313419967GA SA313419968GA SA313419969GA SA313419970GA SA313419971GA SA313419972GA SA313419973GA SA313419974GA SA313420036GA SA313420037GA SA313420038GA SA313420039GA SA313420040GA SA313420041GA SA313420042GA SA313420044GA SA313420045GA SA313420047GA SA313420051GA SA313420053GA SA313420054GA SA313420055GA SA313420057GA SA313420058GA SA313420060GA SA313420061GA SA313420064GA SA313420066GA SA313420068GA SA313420070GA SA313420072GA SA313420073GA SA313420074GA SA313420076GA SA313420077GA SA313420079GA SA313420080GA SA313420081GA SA313420082GA SA313420084GA SA313420085GA SA313420087GA SA313420088GA SA313420090GA SA313420092GA SA313420093GA SA313420094GA SA313420095GA SA313420096GA SA313420097GA SA313420098GA SA313420099GA SA313420102GA SA313420103GA SA313420104GA SA313420105GA SA313420107GA SA313420109GA SA313420110GA SA313420111GA SA313420112GA SA313420114GA SA313420115GA SA313420117GA SA313420118GA SA313420119GA SA313420120GA SA313420121GA SA313420122GA SA313420123GA SA313420124GA SA313420125GA SA313420126GA SA313420129GA SA313420133GA SA313420134GA SA313420135GA SA313420136GA SA313420137GA SA313420138GA SA313420139GA SA313420140GA SA313420143GA SA313420145GA SA313420146GA SA313420147GA SA313420151GA SA313420152GA SA313420154GA SA313420155GA SA313420156GA SA313420157GA SA313420158GA SA313420159GA SA313420160GA SA313420161GA SA313420162GA SA313420163GA SA313420164GA SA313420165GA SA313420167GA SA313420168GA SA313420169GA SA313420170GA SA313420171GA SA313420172GA SA313420173GA SA313420174GA SA313420175GA SA313420176GA SA313420178GA SA313420180GA SA313420182GA SA313420183GA SA313420184GA SA313420185GA SA313420186GA SA313420187GA SA313420188GA SA313420191GA SA313420192GA SA313420193GA SA313420197GA SA313420198GA SA313420199GA SA313420200GA SA313420201GA SA313420202GA SA313420208GA SA313420209GA SA313420210GA SA313420214GA SA313420217GA SA313420219GA SA313420222GA SA313420223GA SA313420224GA SA313420229GA SA313420233GA SA313420234GA SA313420237GA SA313420240GA SA313420243GA SA313420249GA SA313420252GA SA313420254GA SA313420257GA SA313420258GA SA313420262GA SA313420267GA SA313420274GA SA313420279GA SA313420282GA SA313420286GA SA313420287GA SA313420289GA SA313420290GA SA313420295GA SA313420297GA SA313420303GA SA313420307GA SA313420313GA SA313420314GA SA313420315GA SA313420319GA SA313420326GA SA313420328GA SA313420335GA SA313420343GA SA313420354GA SA313420359GA SA313420363GA SA313420366GA SA313420378GA SA313420382GA SA313420384GA SA313420385GA SA313420391GA SA313420393GA SA313420404GA SA313420414GA SA313420418GA SA313420421GA SA313420423GA SA313420427GA SA313420430GA SA313420431GA SA313420436GA SA313420438GA SA313420440GA SA313420442GA SA313420443GA SA313420444GA SA313420448GA SA313420450GA SA313420456GA SA313420457GA SA313420458GA SA313420459GA SA313420465GA SA313420468GA SA313420471GA SA313420473GA SA313420475GA SA313420476GA SA313420479GA SA313420480GA SA313420482GA SA313420483GA SA313420488GA SA313420498GA SA313420504GA SA313420506GA SA313420507GA SA313420509GA SA313420513GA SA313420518GA SA313420522GA SA313420523GA SA313420526GA SA313420531GA SA313420533GA SA313420534GA SA313420544GA SA313420548GA SA313420550GA SA313420555GA SA313420556GA SA313420557GA SA313420558GA SA313420559GA SA313420562GA SA313420564GA SA313420568GA SA313420571GA SA313420575GA SA313420577GA SA313420578GA SA313420579GA SA313420581GA SA313420582GA SA313420587GA SA313420588GA SA313420591GA SA313420595GA SA313420596GA SA313420597GA SA313420600GA SA313420607GA SA313420614GA SA313420616GA SA313430620GA SA313430621GA SA313430622GA SA313430623GA SA313430624GA SA313430625GA SA313430627GA SA313430645GA SA313430646GA SA313430647GA SA313430648GA SA313430649GA SA313430650GA SA313430654GA SA313430659GA SA313430661GA SA313430662GA SA313430665GA SA313430667GA SA313430669GA SA313430673GA SA313430677GA SA313430683GA SA313430685GA SA313430687GA SA313430688GA SA313430691GA SA313430692GA SA313430693GA SA313430697GA SA313430699GA SA313430700GA SA313430705GA SA313430706GA SA313430707GA SA313430709GA SA313430711GA SA313430713GA SA313430715GA SA313430718GA SA313430719GA SA313430721GA SA313430723GA SA313430724GA SA313430726GA SA313430727GA SA313430730GA SA313430731GA SA313430735GA SA313430743GA SA313430746GA SA313430751GA SA313430752GA SA313430753GA SA313430754GA SA313430755GA SA313430757GA SA313430759GA SA313430761GA SA313430762GA SA313430763GA SA313430764GA SA313430766GA SA313430767GA SA313430768GA SA313430771GA SA313430772GA SA313430774GA SA313430776GA SA313430777GA SA313430779GA SA313430780GA SA313430781GA SA313430782GA SA313430783GA SA313430784GA SA313430785GA SA313430789GA SA313430790GA SA313430791GA SA313430792GA SA313430793GA SA313430794GA SA313430795GA SA313430796GA SA313430797GA SA313430798GA SA313430799GA SA313430800GA SA313430801GA SA313430802GA SA313430804GA SA313430805GA SA313430806GA SA313430807GA SA313430808GA SA313430809GA SA313430810GA SA313430811GA SA313430812GA SA313430813GA SA313430815GA SA313430818GA SA313430819GA SA313430820GA SA313430821GA SA313430822GA SA313430823GA SA313430824GA SA313430825GA SA313430827GA SA313430828GA SA313430875GA SA313430907GA SA313430915GA SA313430916GA SA313430936GA SA313430954GA SA313430955GA SA313430957GA SA313430958GA SA313430961GA SA313430966GA SA313430967GA SA313430970GA SA313430971GA SA313430973GA SA313430974GA SA313430975GA SA313430977GA SA313430982GA SA313430987GR SA313430992GA SA313430993GA SA313430994GA SA313430997GA SA313430998GA SA313431000GA SA313431001GA SA313431002GA SA313431003GA SA313431004GA SA313431005GA SA313431006GA SA313431007GA SA313431008GA SA313431009GA SA313431010GA SA313431012GA SA313431013GA SA313431014GA SA313431016GA SA313431017GA SA313431018GA SA313431019GA SA313431020GA SA313431022GA SA313431025GA SA313431026GA SA313431027GA SA313431028GA SA313431030GA SA313431031GA SA313431032GA SA313431033GA SA313431034GA SA313431036GA SA313431037GA SA313431039GA SA313431040GA SA313431041GA SA313431042GA SA313431043GA SA313431044GA SA313431045GA SA313431047GA SA313431048GA SA313431050GA SA313431053GA SA313431054GA SA313431055GA SA313431056GA SA313431058GA SA313431059GA SA313431060GA SA313431061GA SA313431062GA SA313431063GA SA313431064GA SA313431065GA SA313431066GA SA313431067GA SA313431068GA SA313431069GA SA313431072GA SA313431073GA SA313431074GA SA313431079GA SA313431081GA SA313431082GA SA313431083GA SA313431084GA SA313431085GA SA313431088GA SA313431089GA SA313431091GA SA313431093GA SA313431095GA SA313431097GA SA313431098GA SA313431099GA SA313431101GA SA313431104GA SA313431105GA SA313431111GA SA313431115GA SA313431123GA SA313431125GA SA313431126GA SA313431127GA SA313431128GA SA313431129GA SA313431131GA SA313431136GA SA313431143GA SA313431148GA SA313431152GA SA313431155GA SA313431158GA SA313431168GA SA313431171GA SA313431172GA SA313431201GA SA313431206GA SA313431228GA SA313431237GA SA313441304GA SA313441305GA SA313441306GA SA313441307GA SA313441308GA SA313441310GA SA313441311GA SA313441312GA SA313441314GA SA313441315GA SA313441320GA SA313441321GA SA313441334GA SA313441335GA SA313441337GA SA313441338GA SA313441344GA SA313441345GA SA313441350GA SA313441352GA SA313441354GA SA313441357GA SA313441358GA SA313441359GA SA313441361GA SA313441362GA SA313441363GA SA313441365GA SA313441367GA SA313441368GA SA313441369GA SA313441370GA SA313441371GA SA313441372GA SA313441374GA SA313441375GA SA313441377GA SA313441378GA SA313441382GA SA313441383GA SA313441385GA SA313441386GA SA313441388GA SA313441389GA SA313441391GA SA313441392GA SA313441393GA SA313441394GA SA313441395GA SA313441397GA SA313441398GA SA313441399GA SA313441400GA SA313441401GA SA313441402GA SA313441403GA SA313441404GA SA313441405GA SA313441406GA SA313441407GA SA313441408GA SA313441409GA SA313441410GA SA313441411GA SA313441412GA SA313441413GA SA313441414GA SA313441415GA SA313441416GA SA313441417GA SA313441418GA SA313441420GA SA313441421GA SA313441437GA SA313441438GA SA313441439GA SA313441440GA SA313441442GA SA313441443GA SA313441444GA SA313441446GA SA313441447GA SA313441448GA SA313441450GA SA313441451GA SA313441453GA SA313441454GA SA313441455GA SA313441456GA SA313441457GA SA313441458GA SA313441459GA SA313441460GA SA313441461GA SA313441462GA SA313441463GA SA313441464GA SA313441465GA SA313441466GA SA313441468GA SA313441469GA SA313441470GA SA313441471GA SA313441472GA SA313441473GA SA313441474GA SA313441475GA SA313441476GA SA313441477GA SA313441478GA SA313441479GA SA313441480GA SA313441481GA SA313441482GA SA313441483GA SA313441484GA SA313441485GA SA313441486GA SA313441487GA SA313441490GA SA313441492GA SA313441493GA SA313441498GA SA313441501GA SA313441503GA SA313441505GA SA313441507GA SA313441509GA SA313441510GA SA313441511GA SA313441512GA SA313441514GA SA313441520GA SA313441521GA SA313441522GA SA313441524GA SA313441526GA SA313441527GA SA313441528GA SA313441531GA SA313441532GA SA313441534GA SA313441537GA SA313441538GA SA313441539GA SA313441545GA SA313441547GA SA313441551GA SA313441552GA SA313441553GA SA313441555GA SA313441558GA SA313441560GA SA313441561GA SA313441563GA SA313441565GA SA313441566GA SA313441599GA SA313441601GA SA313441602GA SA313441603GA SA313441604GA SA313441605GA SA313441607GA SA313441609GA SA313441610GA SA313441611GA SA313441612GA SA313441613GR SA313441614GA SA313441615GA SA313441616GA SA313441617GA SA313441618GA SA313441620GA SA313441621GA SA313441622GA SA313441623GA SA313441624GA SA313441626GA SA313441632GA SA313441639GA SA313441640GA SA313441641GA SA313441643GA SA313441644GA SA313441647GA SA313441648GA SA313441649GA SA313441650GA SA313441653GA SA313441654GA SA313441655GA SA313441656GA SA313441659GA SA313441660GA SA313441661GA SA313441662GA SA313441663GA SA313441664GA SA313441665GA SA313441666GA SA313441668GA SA313441670GA SA313441671GA SA313441673GA SA313441674GA SA313441675GA SA313441677GA SA313441678GA SA313441679GA SA313441681GA SA313441682GA SA313441683GA SA313441684GA SA313441685GA SA313441686GA SA313441687GA SA313441688GA SA313441690GA SA313441697GA SA313441699GA SA313441700GA SA313441702GA SA313441704GA SA313441706GA SA313441707GA SA313451710GA SA313451711GA SA313451713GA SA313451714GA SA313451716GA SA313451720GA SA313451722GA SA313451725GA SA313451737GA SA313451743GA SA313451745GA SA313451746GA SA313451780GA SA313451784GA SA313451786GA SA313451787GA SA313451791GA SA313451792GA SA313451793GA SA313451794GA SA313451796GA SA313451800GA SA313451801GA SA313451805GA SA313451808GA SA313451813GA SA313451820GA SA313451824GA SA313451826GA SA313451827GA SA313451828GA SA313451829GA SA313451833GA SA313451837GA SA313451838GA SA313451844GA SA313451846GA SA313451849GA SA313451850GA SA313451851GA SA313451853GA SA313451854GA SA313451856GA SA313451871GA SA313451872GA SA313451874GA SA313451875GA SA313451876GA SA313451880GA SA313451881GA SA313451893GA SA313451910GA SA313451911GA SA313451912GA SA313451923GA SA313451924GA SA313451934GA SA313451936GA SA313451941GA SA313451942GA SA313451945GA SA313451946GA SA313451951GA SA313451952GA SA313451954GA SA313451958GA SA313451961GA SA313451966GA SA313451970GA SA313451971GA SA313451974GA SA313451977GA SA313451978GA SA313451986GA SA313451987GA SA313451988GA SA313451990GA SA313451991GA SA313451994GA SA313451996GA SA313451999GA SA313452000GA SA313452004GA SA313452015GA SA313452021GA SA313452026GA SA313452028GA SA313452031GA SA313452036GA SA313452037GA SA313452038GA SA313452040GA SA313452044GA SA313452046GA SA313452049GA SA313452050GA SA313452051GA SA313452053GA SA313452054GA SA313452064GA SA313452070GA SA313452074GA SA313452081GA SA313452083GA SA313452085GA SA313452088GA SA313452091GA SA313452093GA SA313452095GA SA313452097GA SA313452099GA SA313452100GA SA313452101GA SA313452103GA SA313452104GA SA313452106GA SA313452107GA SA313452108GA SA313452109GA SA313452110GA SA313452111GA SA313452112GA SA313452116GA SA313452119GA SA313452122GA SA313452123GA SA313452125GA SA313452126GA SA313452127GA SA313452130GA SA313452131GA SA313452132GA SA313452133GA SA313452135GA SA313452141GA SA313452142GA SA313452144GA SA313452145GA SA313452146GA SA313452147GA SA313452148GA SA313452150GA SA313452152GA SA313452155GA SA313462157GA SA313462158GA SA313462166GA SA313462178GA SA313462188GA SA313462193GA SA313462194GA SA313462195GA SA313462196GA SA313462221GA SA313462225GA SA313462229GA SA313462230GA SA313462232GA SA313462233GA SA313462234GA SA313462236GA SA313462244GA SA313462247GA SA313462256GA SA313462259GA SA313462263GA SA313462264GA SA313462275GA SA313462277GA SA313462280GA SA313462281GA SA313462283GA SA313462284GA SA313462285GA SA313462286GA SA313462288GA SA313462289GA SA313462290GA SA313462292GA SA313462293GA SA313462294GA SA313462295GA SA313462296GA SA313462297GA SA313462298GA SA313462299GA SA313462300GA SA313462301GA SA313462303GA SA313462304GA SA313462305GA SA313462306GA SA313462308GA SA313462309GA SA313462310GA SA313462311GA SA313462314GA SA313462315GA SA313462316GA SA313462317GA SA313462324GA SA313462325GA SA313462328GA SA313462329GA SA313462330GA SA313462331GA SA313462333GA SA313462334GA SA313462336GA SA313462339GA SA313462340GA SA313462342GA SA313462343GA SA313462344GA SA313462345GA SA313462347GA SA313462348GA SA313462349GA SA313462350GA SA313462351GA SA313462352GA SA313462353GA SA313462354GA SA313462355GA SA313462356GA SA313462357GA SA313462358GA SA313462359GA SA313462360GA SA313462361GA SA313462362GA SA313462363GA SA313462364GA SA313462365GA SA313462396GA SA313462397GA SA313462398GA SA313462399GA SA313462401GA SA313462402GA SA313462403GA SA313462404GA SA313462405GA SA313462406GA SA313462407GA SA313462408GA SA313462409GA SA313462411GA SA313462412GA SA313462413GA SA313462414GA SA313462416GA SA313462417GA SA313462418GA SA313462419GA SA313462420GA SA313462422GA SA313462424GA SA313462425GA SA313462426GA SA313462428GA SA313462429GA SA313462430GA SA313462431GA SA313462432GA SA313462433GA SA313462434GA SA313462435GA SA313462436GA SA313462437GA SA313462438GA SA313462439GA SA313462440GA SA313462441GA SA313462442GA SA313462443GA SA313462445GA SA313462446GA SA313462447GA SA313462448GA SA313462449GA SA313462450GR SA313462451GA SA313462452GA SA313462453GA SA313462454GA SA313462455GA SA313462456GA SA313462457GA SA313462458GR SA313462459GA SA313462460GA SA313462461GR SA313462462GA SA313462463GA SA313462464GA SA313462465GA SA313462466GA SA313462467GA SA313462468GA SA313462469GA SA313462470GA SA313462472GA SA313462473GA SA313462474GA SA313462475GA SA313462476GA SA313462477GA SA313462478GA SA313462479GA SA313462480GA SA313462483GA SA313462484GA SA313462485GR SA313462486GA SA313462487GA SA313462488GA SA313462489GA SA313462490GA SA313462491GA SA313462492GA SA313462493GA SA313462494GA SA313462495GA SA313462496GA SA313462497GA SA313462498GA SA313462499GA SA313462500GA SA313462503GA SA313462504GA SA313462505GA SA313462506GA SA313462507GA SA313462508GA SA313462509GA SA313462510GA SA313462511GA SA313462512GA SA313462513GA SA313462514GA SA313462515GA SA313462516GA SA313462517GA SA313462518GA SA313462519GA SA313462520GA SA313462521GA SA313462522GA SA313462523GA SA313462525GA SA313462526GA SA313462527GA SA313462528GA SA313462529GA SA313462530GA SA313462531GA SA313462532GA SA313462533GA SA313462534GA SA313462535GA SA313462536GA SA313462537GA SA313462538GA SA313462539GA SA313462540GA SA313462541GA SA313462542GA SA313462543GA SA313462544GA SA313462545GA SA313462546GA SA313462547GA SA313462549GA SA313462550GA SA313462551GA SA313462552GA SA313462553GA SA313462554GA SA313462555GA SA313462556GA SA313462557GA SA313462558GA SA313462559GA SA313462561GA SA313462562GA SA313462563GA SA313462564GA SA313462565GA SA313462566GA SA313462567GA SA313462568GA SA313462569GA SA313462570GA SA313462571GA SA313462572GA SA313462573GA SA313462574GA SA313462575GA SA313462576GA SA313462577GA SA313462578GA SA313462579GA SA313462580GA SA313462581GA SA313462582GR SA313462583GA SA313462584GA SA313462585GA SA313462586GA SA313462587GA SA313462588GA SA313462589GA SA313462590GA SA313462591GA SA313462592GA SA313462593GA SA313462594GA SA313462595GR SA313462596GA SA313462597GA SA313462598GA SA313462599GA SA313462600GA SA313462601GA SA313462602GA SA313462603GA SA313462604GA SA313462605GA SA313482606GA SA313482609GA SA313482611GA SA313482613GA SA313482617GA SA313482618GA SA313482619GA SA313482622GA SA313482623GA SA313482624GA SA313482626GA SA313482632GA SA313482636GA SA313482638GA SA313482639GA SA313482641GA SA313482642GA SA313482643GA SA313482644GA SA313482645GA SA313482649GA SA313482650GA SA313482657GA SA313482662GA SA313482663GA SA313482666GA SA313482667GA SA313482670GA SA313482671GA SA313482674GA SA313482676GA SA313482677GA SA313482678GA SA313482680GA SA313482682GA SA313482683GA SA313482686GA SA313482687GA SA313482688GA SA313482690GA SA313482691GA SA313482692GA SA313482693GA SA313482695GA SA313482696GA SA313482701GA SA313482703GA SA313482706GA SA313482707GA SA313482708GA SA313482710GA SA313482712GA SA313482713GA SA313482715GA SA313482716GA SA313482717GA SA313482718GA SA313482719GA SA313482720GA SA313482721GA SA313482722GA SA313482723GA SA313482724GA SA313482728GA SA313482729GA SA313482730GA SA313482732GA SA313482733GA SA313482736GA SA313482740GA SA313482742GA SA313482745GA SA313482755GA SA313482768GA SA313482769GA SA313482772GA SA313482778GA SA313482779GA SA313482787GA SA313482788GA SA313482790GA SA313482791GA SA313482792GA SA313482793GA SA313482794GA SA313482796GA SA313482797GA SA313482799GA SA313482800GA SA313482801GA SA313482802GA SA313482803GA SA313482805GA SA313482806GA SA313482807GA SA313482808GA SA313482809GA SA313482810GA SA313482811GA SA313482815GA SA313482846GA SA313482848GA SA313482849GA SA313482851GA SA313482852GA SA313482853GA SA313482854GA SA313482855GA SA313482856GA SA313482858GA SA313482859GA SA313482860GA SA313482861GA SA313482862GA SA313482866GA SA313482867GA SA313482868GA SA313482869GA SA313482870GA SA313482871GA SA313482873GA SA313482874GA SA313482875GA SA313492907GA SA313492908GA SA313492909GA SA313492910GA SA313492911GA SA313492912GA SA313492913GA SA313492914GA SA313492915GA SA313492916GA SA313492917GA SA313492918GA SA313492919GA SA313492920GA SA313492921GA SA313492922GA SA313492923GA SA313492925GA SA313492926GA SA313492927GA SA313492928GA SA313492929GA SA313492930GA SA313492931GA SA313492932GA SA313492933GA SA313492934GA SA313492935GA SA313492936GA SA313492937GA SA313492938GA SA313492939GA SA313492940GA SA313492941GA SA313492942GA SA313492943GA SA313492944GA SA313492945GA SA313492946GA SA313492947GA SA313492948GA SA313492949GA SA313492950GA SA313492951GA SA313492952GA SA313492953GA SA313492954GA SA313492957GA SA313492958GA SA313492959GA SA313492960GA SA313492963GA SA313492964GA SA313492965GA SA313493002GA SA313493006GA SA313493007GA SA313493008GA SA313493009GA SA313493010GA SA313493011GA SA313493012GA SA313493013GA SA313493014GA SA313493015GA SA313493016GA SA313493017GA SA313493019GA SA313493032GA SA313493033GA SA313493034GA SA313493036GA SA313493037GA SA313493044GA SA313493045GA SA313493046GA SA313493047GA SA313493048GA SA313493081GA SA313493082GA SA313493083GA SA313493084GA SA313493086GA SA313493087GA SA313493088GA SA313493090GA SA313493091GA SA313493092GA SA313493093GA SA313493094GA SA313493095GA SA313493096GA SA313493097GA SA313493098GA SA313493099GA SA313493100GA SA313493101GA SA313493102GA SA313493104GA SA313493105GA SA313493107GA SA313493108GA SA313493109GA SA313493182GA SA313493183GA SA313493184GA SA313493185GA SA313493187GA SA313493188GA SA313493191GA SA313493192GA SA313493193GA SA313493195GA SA313493196GA SA313493197GA SA313493198GA SA313493199GA SA313493200GA SA313493201GA SA313493202GA SA313493204GA SA313493205GA SA313493206GA SA313493207GA SA313493208GA SA313493209GA SA313493213GA SA313493218GA SA313493240GA SA313493241GA SA313493242GA SA313493243GA SA313493245GA SA313493250GA SA313493251GA SA313493252GR SA313493254GA SA313493262GA SA313493263GA SA313493264GA SA313493392GA SA313493393GA SA313493394GA SA313493395GA SA313493396GA SA313493397GA SA313493398GA SA313493399GA SA313493400GA SA313493401GA SA313493402GA SA313493403GA SA313493404GA SA313493405GA SA313493406GA SA313493407GA SA313493408GA SA313493409GA SA313493410GA SA313493411GA SA313493413GA SA313493414GA SA313493415GA SA313493416GA SA313493417GA SA313493418GA SA313493419GA SA313493420GA SA313493422GA SA313493423GA SA313493424GA SA313493425GA SA313493426GA SA313493427GA SA313493428GA SA313493429GA SA313493430GA SA313493431GA SA313493432GA SA313493433GA SA313493434GA SA313493435GA SA313493436GA SA313493437GA SA313493438GA SA313493439GA SA313493441GA SA313493442GA SA313493444GA SA313493445GA SA313493446GA SA313493447GA SA313493448GA SA313493449GA SA313493450GA SA313493451GA SA313493452GA SA313493454GA SA313493455GA SA313493457GA SA313493458GA SA313493459GA SA313493460GA SA313493463GA SA313493464GA SA313493465GA SA313493466GA SA313493467GA SA313493468GA SA313493470GA SA313493542GA SA313493543GA SA313493545GA SA313493546GA SA313493547GA SA313493548GA SA313493549GA SA313493550GA SA313493552GA SA313493554GA SA313493555GR SA313493556GA SA313493558GA SA313493559GA SA313493561GA SA313493562GA SA313493563GA SA313493564GA SA313493565GA SA313493566GA SA313493567GA SA313493568GA SA313493570GA SA313503572GA SA313503573GA SA313503574GA SA313503575GA SA313503576GA SA313503577GA SA313503578GA SA313503579GA SA313503580GA SA313503581GA SA313503582GA SA313503583GA SA313503585GA SA313503587GA SA313503588GA SA313503589GA SA313503591GA SA313503592GA SA313503593GA SA313503604GA SA313503605GA SA313503608GA SA313503609GA SA313503610GA SA313503611GA SA313503671GA SA313503672GA SA313503673GA SA313503674GA SA313503677GA SA313503679GA SA313503680GA SA313503681GA SA313503682GA SA313503683GA SA313503684GA SA313503685GA SA313503686GA SA313503687GA SA313503688GA SA313503689GA SA313503690GA SA313503691GA SA313503692GA SA313503693GA SA313503694GA SA313503695GA SA313503697GA SA313503698GA SA313503699GA SA313503700GA SA313503701GA SA313503706GA SA313503707GA SA313503711GA SA313503716GA SA313503717GA SA313503718GA SA313503720GA SA313503722GA SA313503723GA SA313503724GA SA313503725GA SA313503726GA SA313503728GA SA313513745GA SA313513747GA SA313513748GA SA313513749GA SA313513750GA SA313513753GA SA313513754GA SA313513755GA SA313513757GA SA313513758GA SA313513759GA SA313513760GA SA313513762GA SA313513764GA SA313513766GA SA313513767GA SA313513768GA SA313513769GA SA313513770GA SA313513772GA SA313513773GA SA313513774GA SA313513775GA SA313513776GA SA313513777GA SA313513779GA SA313513780GA SA313513781GA SA313513782GA SA313513783GA SA313513784GA SA313513785GA SA313513786GA SA313513787GA SA313513788GA SA313513789GA SA313513790GA SA313513791GR SA313513792GA SA313513793GA SA313513794GA SA313513795GA SA313513796GA SA313513797GA SA313513798GA SA313513799GA SA313513800GA SA313513801GA SA313513802GA SA313513803GA SA313513804GA SA313513805GA SA313513806GA SA313513807GA SA313513808GA SA313513809GA SA313513810GA SA313513811GA SA313513812GA SA313513813GA SA313513815GA SA313513816GA SA313513817GA SA313513818GA SA313513819GA SA313513820GA SA313513821GA SA313513822GA SA313513823GA SA313513825GA SA313513826GA SA313513827GA SA313513828GA SA313513830GA SA313513831GR SA313513833GA SA313513835GA SA313513840GA SA313513841GA SA313513842GA SA313513844GA SA313513845GA SA313513847GR SA313513852GA SA313513854GR SA313513856GA SA313513857GA SA313513858GA SA313513859GA SA313513860GA SA313513864GA SA313513865GA SA313513869GA SA313513872GA SA313513875GA SA313513876GA SA313513883GA SA313513885GA SA313513886GA SA313513889GA SA313513890GA SA313513894GA SA313513898GA SA313513900GA SA313513902GA SA313513903GA SA313513904GA SA313513905GA SA313513906GA SA313513907GA SA313513908GA SA313513909GA SA313513910GA SA313513911GA SA313513912GA SA313513913GA SA313513914GA SA313513915GA SA313513918GA SA313513919GA SA313513922GR SA313513924GA SA313513925GA SA313513927GA SA313513928GA SA313513936GA SA313513937GA SA313513938GA SA313513939GA SA313513943GA SA313513944GA SA313523950GA SA313523953GA SA313523956GA SA313523957GA SA313523958GA SA313523959GA SA313523960GA SA313523961GA SA313523962GA SA313523964GA SA313523965GA SA313523966GA SA313523967GA SA313523968GA SA313523969GA SA313523976GA SA313523977GA SA313523978GA SA313523979GA SA314004051GA SA314004052GA SA314004053GA SA314004054GA SA314004055GA SA314004056GA SA314004057GA SA314004058GA SA314004059GA SA314004101GA SA314004105GA SA314004106GA SA314004107GA SA314004108GA SA314004112GA SA314004114GA SA314004125GA SA314004127GA SA314004128GA SA314004130GA SA314004176GA SA314004179GA SA314004185GA SA314004187GA SA314004188GA SA314004189GA SA314004190GA SA314004196GA SA314004198GA SA314004200GA SA314004207GA SA314004244GA SA314004247GA SA314004248GA SA314004263GA SA314004292GA SA314004298GA SA314004301GA SA314014322GA SA314014327GA SA314014328GA SA314014329GA SA314014330GA SA314014331GA SA314014332GA SA314014333GA SA314014334GA SA314014335GA SA314014337GA SA314014338GA SA314014339GA SA314014341GA SA314014342GA SA314014354GA SA314014356GA SA314014359GA SA314014366GA SA314014367GA SA314014368GA SA314014369GA SA314014371GA SA314014374GA SA314014375GA SA314014376GA SA314014377GA SA314014378GA SA314014380GA SA314014381GA SA314014382GA SA314014383GA SA314014384GA SA314014385GA SA314014386GA SA314014389GA SA314014390GA SA314014391GA SA314014392GA SA314014393GA SA314014394GA SA314014395GA SA314014396GA SA314014397GA SA314014398GA SA314014401GA SA314014402GA SA314014405GA SA314014406GA SA314014407GA SA314014408GA SA314014409GA SA314014412GA SA314014413GA SA314014415GA SA314014416GA SA314014417GA SA314014418GA SA314014419GA SA314014420GA SA314014421GA SA314014423GA SA314014424GA SA314014425GA SA314014426GA SA314014428GA SA314014429GA SA314014430GA SA314014431GA SA314014433GA SA314014435GA SA314014437GA SA314014440GA SA314014441GA SA314014442GA SA314014445GA SA314014446GA SA314014447GA SA314014449GA SA314014454GA SA314014455GA SA314014456GA SA314014458GA SA314014459GA SA314014461GA SA314014462GA SA314014463GA SA314014465GA SA314014466GA SA314014469GA SA314014470GA SA314014471GA SA314014474GA SA314014475GA SA314014479GA SA314014481GA SA314014483GA SA314014484GA SA314014487GA SA314014488GA SA314014490GA SA314014491GA SA314014492GA SA314014493GA SA314014496GA SA314014498GA SA314014499GA SA314014501GA SA314014503GA SA314014506GA SA314014507GA SA314014508GA SA314014510GA SA314014511GA SA314014512GA SA314014514GA SA314014515GA SA314014516GA SA314014518GA SA314014519GA SA314014521GA SA314014525GA SA314014526GA SA314014528GA SA314014529GA SA314014530GA SA314014532GA SA314014534GA SA314014535GA SA314014537GA SA314014541GA SA314014542GA SA314014543GA SA314014544GA SA314014545GA SA314014547GA SA314014549GA SA314014552GA SA314014555GA SA314014556GA SA314014557GA SA314014558GA SA314014559GA SA314014560GA SA314014562GA SA314014564GA SA314014565GA SA314014566GA SA314014568GA SA314014569GA SA314014570GA SA314014571GA SA314014573GA SA314014574GA SA314014576GA SA314014577GA SA314014580GA SA314014581GA SA314014582GA SA314014585GA SA314014586GA SA314014588GA SA314014589GA SA314014590GA SA314014591GA SA314014592GA SA314014593GA SA314014594GA SA314014596GA SA314014597GA SA314014603GA SA314014604GA SA314014605GA SA314014607GA SA314014610GA SA314014612GA SA314014613GA SA314014614GA SA314014615GA SA314014616GA SA314014617GA SA314014619GA SA314014620GA SA314014621GA SA314014623GA SA314014624GA SA314014625GA SA314014626GA SA314014627GA SA314014628GA SA314014629GA SA314014630GA SA314014631GA SA314014633GA SA314014634GA SA314014635GA SA314014636GA SA314014637GA SA314014638GA SA314014639GA SA314014640GA SA314014641GA SA314014642GA SA314014643GA SA314014644GA SA314014646GA SA314014648GA SA314014653GA SA314014655GA SA314014656GA SA314014657GA SA314014659GA SA314014660GA SA314014661GA SA314014662GA SA314014663GA SA314014665GA SA314014666GA SA314014667GA SA314014668GA SA314014669GA SA314014670GA SA314014672GA SA314014675GA SA314014676GA SA314014678GA SA314014679GA SA314014681GA SA314014682GA SA314014683GA SA314014684GA SA314014686GA SA314014687GA SA314014689GA SA314014690GA SA314014692GA SA314014695GA SA314014696GA SA314014697GA SA314014699GA SA314014700GA SA314014704GA SA314014706GA SA314014707GA SA314014709GA SA314014711GA SA314014715GA SA314014717GA SA314014718GA SA314014719GA SA314014722GA SA314014723GA SA314014724GA SA314014725GA SA314014726GA SA314014727GA SA314014728GA SA314014729GA SA314014730GA SA314014731GA SA314014732GA SA314014733GA SA314014734GA SA314014735GA SA314014737GA SA314014738GA SA314014739GA SA314014742GA SA314014745GA SA314014748GA SA314014750GA SA314014755GA SA314014757GA SA314014758GA SA314014760GA SA314014765GA SA314014766GA SA314024781GA SA314024783GA SA314024787GA SA314024792GA SA314024793GA SA314024794GA SA314024795GA SA314024797GA SA314024802GA SA314024805GA SA314024814GA SA314024817GA SA314024821GA SA314024822GA SA314024825GA SA314024829GA SA314024831GA SA314024833GA SA314024834GA SA314024835GA SA314024837GA SA314024844GA SA314024846GA SA314024850GA SA314024854GA SA314024857GA SA314024858GA SA314024859GA SA314024864GA SA314024865GA SA314024867GA SA314024868GA SA314024870GA SA314024871GA SA314024872GA SA314024873GA SA314024874GA SA314024876GA SA314024878GA SA314024879GA SA314024881GA SA314024882GA SA314024884GA SA314024885GA SA314024886GA SA314024887GA SA314024888GA SA314024891GA SA314024892GA SA314024894GA SA314024896GA SA314024899GA SA314024900GA SA314024901GA SA314024903GA SA314024904GA SA314024908GA SA314024909GA SA314024910GA SA314024912GA SA314024913GA SA314024914GA SA314024915GA SA314024920GA SA314024921GA SA314024922GA SA314024923GA SA314024930GA SA314024931GA SA314024932GA SA314024933GA SA314024935GA SA314024937GA SA314024943GA SA314024944GA SA314024945GA SA314024946GA SA314024947GA SA314024948GA SA314024949GA SA314024950GA SA314024951GA SA314024952GA SA314024953GA SA314024954GA SA314024955GA SA314024957GA SA314024958GA SA314024961GA SA314024962GA SA314024964GA SA314024966GA SA314024967GA SA314024968GA SA314024969GA SA314024970GA SA314024971GA SA314024972GA SA314024973GA SA314024974GA SA314024975GA SA314024976GA SA314024977GA SA314024978GA SA314024979GA SA314024980GA SA314024981GA SA314024982GA SA314024983GA SA314024984GA SA314024985GA SA314024986GA SA314024987GA SA314024988GA SA314024989GA SA314024990GA SA314024992GA SA314024993GA SA314024994GA SA314024997GA SA314024998GA SA314024999GA SA314025000GA SA314025001GA SA314025002GA SA314025003GA SA314025004GA SA314025005GA SA314025006GA SA314025008GA SA314025009GA SA314025010GA SA314025012GA SA314025013GA SA314025014GA SA314025016GA SA314025017GA SA314025018GA SA314025019GA SA314025020GA SA314025021GA SA314025022GA SA314025024GA SA314025025GA SA314025026GA SA314025027GA SA314025028GA SA314025032GA SA314025033GA SA314025034GA SA314025035GA SA314025036GA SA314025037GA SA314025038GA SA314025039GA SA314025040GA SA314025041GA SA314025042GA SA314025044GA SA314025045GA SA314025048GA SA314025049GA SA314025050GA SA314025051GA SA314025052GA SA314025053GA SA314025055GA SA314025056GA SA314025057GA SA314025058GA SA314025059GA SA314025061GA SA314025062GA SA314025063GA SA314025066GA SA314025067GA SA314025071GA SA314025073GA SA314025074GA SA314025078GA SA314025080GA SA314025082GA SA314025083GA SA314025084GA SA314025085GA SA314025086GA SA314025088GA SA314025089GA SA314025091GA SA314025092GA SA314025094GA SA314025095GA SA314025096GA SA314025098GA SA314025100GA SA314025103GA SA314025107GA SA314025109GA SA314025111GA SA314025114GA SA314025116GA SA314025123GA SA314025127GA SA314025128GA SA314025136GA SA314025139GA SA314025140GA SA314025142GA SA314025143GA SA314025144GA SA314025145GA SA314025146GA SA314025147GA SA314025148GA SA314025149GA SA314025150GA SA314025151GA SA314025152GA SA314025153GA SA314025154GA SA314025155GA SA314025156GA SA314025158GA SA314025159GA SA314025160GA SA314025161GA SA314025162GA SA314025163GA SA314025164GA SA314025165GA SA314025166GA SA314025168GA SA314025170GA SA314025171GA SA314025172GA SA314025173GA SA314025175GA SA314025177GA SA314025179GA SA314025181GA SA314025183GA SA314025187GA SA314025196GA SA314025201GA SA314025204GA SA314025205GA SA314025206GA SA314025209GA SA314025211GA SA314025213GA SA314025214GA SA314025215GA SA314025216GA SA314025217GA SA314025218GA SA314025219GA SA314025221GA SA314025222GA SA314025223GA SA314025224GA SA314025225GA SA314025226GA SA314025227GA SA314025229GA SA314025230GA SA314025231GA SA314035233GA SA314035235GA SA314035238GA SA314035239GA SA314035242GA SA314035246GA SA314035249GA SA314035254GA SA314035257GA SA314035261GA SA314035263GA SA314035265GA SA314035266GA SA314035267GA SA314035268GA SA314035269GA SA314035274GA SA314035276GA SA314035277GA SA314035281GA SA314035282GA SA314035283GA SA314035284GA SA314035285GA SA314035286GA SA314035288GA SA314035289GA SA314035291GA SA314035292GA SA314035293GA SA314035294GA SA314035295GA SA314035296GA SA314035297GA SA314035301GA SA314035303GA SA314035304GA SA314035305GA SA314035306GA SA314035307GA SA314035308GA SA314035309GA SA314035311GA SA314035312GA SA314035313GA SA314035315GA SA314035317GA SA314035318GA SA314035319GA SA314035321GA SA314035322GA SA314035323GA SA314035324GA SA314035325GA SA314035328GA SA314035329GA SA314035331GA SA314035333GA SA314035334GA SA314035336GA SA314035337GA SA314035339GA SA314035340GA SA314035342GA SA314035343GA SA314035344GA SA314035346GA SA314035347GA SA314035349GA SA314035350GA SA314035351GA SA314035352GA SA314035355GA SA314035356GA SA314035357GA SA314035359GA SA314035360GA SA314035362GA SA314035363GA SA314035364GA SA314035365GA SA314035366GA SA314035367GA SA314035368GA SA314035369GA SA314035370GA SA314035371GA SA314035372GA SA314035373GA SA314035374GA SA314035375GA SA314035376GA SA314035377GA SA314035379GA SA314035380GA SA314035381GA SA314035384GA SA314035385GA SA314035386GA SA314035387GA SA314035388GA SA314035389GA SA314035390GA SA314035392GA SA314035393GA SA314035394GA SA314035395GA SA314035397GA SA314035399GA SA314035400GA SA314035402GA SA314035404GA SA314035408GA SA314035409GA SA314035411GA SA314035412GA SA314035413GA SA314035414GA SA314035415GA SA314035416GA SA314035417GA SA314035418GA SA314035419GA SA314035421GA SA314035422GA SA314035424GA SA314035426GA SA314035427GA SA314035429GA SA314035430GA SA314035431GA SA314035432GA SA314035433GA SA314035434GA SA314035435GA SA314035436GA SA314035437GA SA314035438GA SA314035439GA SA314035440GA SA314035441GA SA314035442GA SA314045443GA SA314045444GA SA314045445GA SA314045446GA SA314045447GA SA314045448GA SA314045449GA SA314045450GA SA314045451GA SA314045452GA SA314045453GA SA314045454GA SA314045455GA SA314045456GA SA314045457GA SA314045458GA SA314045459GA SA314045460GA SA314045461GA SA314045462GA SA314045463GA SA314045464GA SA314045465GA SA314045466GA SA314045467GA SA314045468GA SA314045469GA SA314045470GA SA314045471GA SA314045472GA SA314045473GA SA314045474GA SA314045476GA SA314045477GA SA314045478GA SA314045479GA SA314045480GA SA314045481GA SA314045482GA SA314045484GA SA314045485GA SA314045486GA SA314045487GA SA314045488GA SA314045489GA SA314045490GA SA314045492GA SA314045493GA SA314045494GA SA314045495GA SA314045496GA SA314045497GA SA314045498GA SA314045499GA SA314045500GA SA314045501GA SA314045503GA SA314045504GA SA314045505GA SA314045507GA SA314045508GA SA314045511GA SA314045513GA SA314045514GA SA314045515GA SA314045516GA SA314045518GA SA314045520GA SA314045523GA SA314045524GA SA314045525GA SA314045527GA SA314045528GA SA314045529GA SA314045530GA SA314045531GA SA314045533GA SA314045535GA SA314055538GA SA314055540GA SA314055541GA SA314055542GA SA314055546GA SA314055547GA SA314055549GA SA314055550GA SA314055551GA SA314055553GA SA314055557GA SA314055558GA SA314055559GA SA314055560GA SA314055561GA SA314055564GA SA314055567GA SA314055569GA SA314055571GA SA314055572GA SA314055573GA SA314055579GA SA314055580GA SA314055581GA SA314055582GA SA314055584GA SA314055585GA SA314055586GA SA314055587GA SA314055588GA SA314055589GA SA314055590GA SA314055591GA SA314055592GA SA314055593GA SA314055595GA SA314055596GA SA314055600GA SA314055601GA SA314055605GA SA314055606GA SA314055608GA SA314055610GA SA314055611GA SA314055612GA SA314055614GA SA314055615GA SA314055616GA SA314055618GA SA314055621GA SA314055622GA SA314055623GA SA314055624GA SA314055625GA SA314055627GA SA314055628GA SA314055629GA SA314055630GA SA314055631GA SA314055632GA SA314055633GA SA314055634GA SA314055635GA SA314055636GA SA314055638GA SA314055639GA SA314055640GA SA314055641GA SA314055642GA SA314055643GA SA314055644GA SA314055645GA SA314055646GA SA314055648GA SA314055650GA SA314055651GA SA314055652GA SA314055653GA SA314055654GA SA314055655GA SA314055656GA SA314055660GA SA314055661GA SA314055662GA SA314055663GA SA314055669GA SA314055670GA SA314055671GA SA314055676GR SA314055677GA SA314055678GA SA314055681GA SA314055682GA SA314055683GA SA314055686GA SA314055688GA SA314055690GA SA314055691GA SA314055696GA SA314055698GA SA314055699GA SA314055703GA SA314055708GA SA314055712GA SA314055715GA SA314055717GA SA314055718GA SA314055720GA SA314055721GA SA314055722GA SA314055724GA SA314055725GA SA314055726GA SA314055727GA SA314055728GA SA314055729GA SA314055730GA SA314055731GA SA314055733GA SA314055734GA SA314065735GA SA314065742GA SA314065747GA SA314065750GA SA314065751GA SA314065752GA SA314065753GA SA314065754GA SA314065756GA SA314065757GA SA314065759GA SA314065760GA SA314065761GA SA314065762GA SA314065763GA SA314065765GA SA314065766GA SA314065767GA SA314065768GA SA314065769GA SA314065770GA SA314065771GA SA314065774GA SA314065778GA SA314065780GA SA314065781GA SA314065784GA SA314065785GA SA314065788GA SA314065789GA SA314065791GA SA314065793GA SA314065794GA SA314075796GA SA314075797GA SA314075798GA SA314075799GA SA314075801GA SA314075802GA SA314075803GA SA314075804GA SA314075806GA SA314075807GA SA314075808GA SA314075809GA SA314075810GA SA314075811GA SA314075812GA SA314075813GA SA314075814GA SA314075815GA SA314075816GA SA314075818GA SA314075819GA SA314075821GA SA314075823GA SA314075824GA SA314075825GA SA314075826GA SA314075827GA SA314075828GA SA314075829GA SA314075830GA SA314075831GA SA314075833GA SA314075834GA SA314075835GA SA314075837GA SA314075838GA SA314075839GA SA314075841GA SA314075848GA SA314075849GA SA314075850GA SA314075851GA SA314075852GA SA314075853GA SA314075854GA SA314075855GA SA314075857GA SA314075858GA SA314075859GA SA314075860GA SA314075861GA SA314075862GA SA314075863GA SA314075864GA SA314075865GA SA314075866GA SA314075867GA SA314075868GA SA314075869GA SA314075870GA SA314075871GA SA314075872GA SA314075873GA SA314075874GA SA314075875GA SA314075877GA SA314075878GA SA314075880GA SA314075881GA SA314075882GA SA314075883GA SA314075947GA SA314075948GA SA314075952GA SA314075979GA SA314075980GA SA314075981GA SA314075983GA SA314075987GA SA314075989GA SA314075991GA SA314075993GA SA314075997GA SA314075998GA SA314076001GA SA314076002GA SA314076003GA SA314076008GA SA314076017GA SA314076018GA SA314076021GA SA314076024GA SA314076030GA SA314076032GA SA314076033GA SA314076034GA SA314076066GA SA314076069GA SA314076070GA SA314076072GA SA314076076GA SA314076077GA SA314076078GA SA314076080GA SA314076082GA SA314076087GA SA314076090GA SA314076093GA SA314076094GA SA314076103GA SA314076104GA SA314076112GA SA314076114GA SA314076115GA SA314076118GA SA314076119GA SA314076120GA SA314076121GA SA314076122GA SA314076127GA SA314076128GA SA314076134GA SA314076137GA SA314076141GA SA314076150GA SA314076156GA SA314076160GA SA314076168GA SA314076170GA SA314076171GA SA314076176GA SA314076179GA SA314076180GA SA314076182GA SA314076183GA SA314076187GA SA314076188GA SA314076190GA SA314076191GA SA314076192GA SA314076196GA SA314076197GA SA314076199GR SA314076201GA SA314076202GA SA314076203GR SA314076205GA SA314076208GA SA314076210GA SA314076213GA SA314076215GA SA314076216GA SA314076217GA SA314076219GA SA314076220GA SA314076222GA SA314076223GA SA314076225GA SA314076226GA SA314076227GA SA314076229GA SA314076232GA SA314076233GA SA314076234GA SA314076236GA SA314076238GA SA314076240GA SA314076241GA SA314076242GA SA314076244GA SA314076245GA SA314076250GA SA314076254GA SA314076255GA SA314076257GA SA314076259GA SA314076261GA SA314076263GA SA314076264GA SA314076265GA SA314076266GA SA314076267GA SA314076268GA SA314076269GA SA314076272GA SA314076275GA SA314076277GA SA314076278GA SA314076279GA SA314076280GA SA314076283GA SA314076288GA SA314076292GA SA314076297GA SA314076298GA SA314076299GA SA314076301GA SA314076302GA SA314076303GA SA314076307GA SA314076309GA SA314076313GA SA314076314GA SA314076320GA SA314076321GA SA314076322GA SA314076323GA SA314076329GA SA314076332GA SA314076339GA SA314086347GA SA314086350GA SA314086352GA SA314086356GA SA314086357GA SA314086360GA SA314086366GA SA314086367GA SA314086369GA SA314086370GA SA314086378GA SA314086383GA SA314086384GA SA314086385GA SA314086386GA SA314086387GA SA314086388GA SA314086389GA SA314086390GA SA314086391GA SA314086393GA SA314086394GA SA314086404GA SA314086405GA SA314086406GA SA314086407GA SA314086408GA SA314086410GA SA314086411GA SA314086419GA SA314086421GA SA314086422GA SA314086435GA SA314086440GA SA314086441GA SA314086443GA SA314086448GA SA314086450GA SA314086452GA SA314086461GA SA314086464GA SA314086468GA SA314086469GA SA314086471GA SA314086474GA SA314086475GA SA314086477GA SA314086480GA SA314086481GA SA314086491GA SA314086492GA SA314086496GA SA314086497GA SA314086498GA SA314086499GA SA314086502GA SA314086504GA SA314086507GA SA314086508GA SA314086509GA SA314086510GA SA314086512GA SA314086513GA SA314086514GA SA314086515GA SA314086516GA SA314086517GA SA314086518GA SA314086519GA SA314086520GA SA314086523GA SA314086524GA SA314086525GA SA314086526GA SA314086544GA SA314086548GA SA314086554GA SA314086555GA SA314086556GA SA314086558GA SA314086559GA SA314086565GA SA314086566GA SA314086578GA SA314086580GA SA314086589GA SA314086599GA SA314086604GA SA314086605GA SA314086606GA SA314086608GA SA314086609GA SA314086612GA SA314086613GA SA314086614GA SA314086615GA SA314086616GA SA314086618GA SA314086620GA SA314086623GA SA314086627GA SA314086628GA SA314086629GA SA314086635GA SA314086638GA SA314086639GA SA314086640GA SA314086641GA SA314086642GA SA314086644GA SA314086646GA SA314086647GA SA314086651GA SA314086655GA SA314086656GA SA314086658GA SA314086661GA SA314086664GA SA314086667GA SA314086668GA SA314086670GA SA314086671GA SA314086672GA SA314086674GA SA314086675GA SA314086676GA SA314086677GA SA314086678GA SA314086679GA SA314086680GA SA314086687GA SA314086688GA SA314086689GA SA314086692GA SA314086693GA SA314086694GA SA314086697GA SA314086698GA SA314086699GA SA314086701GA SA314086702GA SA314086703GA SA314086704GA SA314086705GA SA314086706GA SA314086707GA SA314086708GA SA314086710GA SA314086711GA SA314086715GA SA314086716GA SA314086717GA SA314086719GA SA314086720GA SA314086721GA SA314086722GA SA314086723GA SA314086724GA SA314086725GA SA314086726GA SA314086728GA SA314086729GA SA314086730GA SA314086731GA SA314086732GA SA314086735GA SA314086736GA SA314086737GA SA314086738GA SA314086740GA SA314086742GA SA314096743GA SA314096744GA SA314096746GA SA314096747GA SA314096749GA SA314096750GA SA314096752GA SA314096753GA SA314096754GA SA314096755GA SA314096756GA SA314096757GA SA314096758GA SA314096760GA SA314096761GA SA314096762GA SA314096764GA SA314096765GA SA314096766GA SA314096767GA SA314096768GA SA314096769GA SA314096770GA SA314096771GA SA314096772GA SA314096773GA SA314096774GA SA314096775GA SA314096776GA SA314096778GA SA314096779GA SA314096780GA SA314096781GA SA314096782GA SA314096783GA SA314096784GA SA314096785GA SA314096786GA SA314096787GA SA314096788GA SA314096789GA SA314096790GA SA314096791GA SA314096792GA SA314096793GA SA314096794GA SA314096795GA SA314096796GA SA314096797GA SA314096799GA SA314096800GA SA314096801GA SA314096804GA SA314096805GA SA314096808GA SA314096809GA SA314096810GA SA314096811GA SA314096812GA SA314096814GA SA314096817GA SA314096818GA SA314096819GA SA314096820GA SA314096821GA SA314096822GA SA314096823GA SA314096825GA SA314096826GA SA314096828GA SA314096829GA SA314096830GA SA314096831GA SA314096834GA SA314096837GA SA314096838GA SA314096839GA SA314096840GA SA314096842GA SA314096843GA SA314096846GA SA314096852GA SA314096853GA SA314096854GA SA314096855GA SA314096861GA SA314096864GA SA314096866GA SA314096868GA SA314096870GA SA314096876GA SA314096877GA SA314096901GA SA314096911GA SA314096929GA SA314096933GA SA314096934GA SA314096935GA SA314096936GA SA314096937GA SA314096938GA SA314096939GA SA314096942GA SA314096943GA SA314096944GA SA314096946GA SA314096947GA SA314096948GA SA314096949GA SA314096950GA SA314096951GA SA314096952GA SA314096955GA SA314096957GA SA314096958GA SA314096959GA SA314096960GA SA314096962GA SA314096964GA SA314096966GA SA314096967GA SA314096970GA SA314096972GA SA314096973GA SA314096975GA SA314096976GA SA314096978GA SA314096980GA SA314096981GA SA314096983GA SA314096984GA SA314096985GA SA314096987GA SA314096989GA SA314096992GA SA314096993GA SA314096995GA SA314096997GA SA314096998GA SA314096999GA SA314097001GA SA314097003GA SA314097005GA SA314097006GA SA314097007GA SA314097009GA SA314097010GA SA314097011GA SA314097014GA SA314097015GA SA314097016GA SA314097018GA SA314097019GA SA314097020GA SA314097023GA SA314097024GA SA314097025GA SA314097028GA SA314097033GA SA314097035GA SA314097036GA SA314097037GA SA314097038GA SA314097042GA SA314097047GA SA314097050GA SA314097059GA SA314097061GA SA314097063GA SA314097064GA SA314097069GA SA314097070GA SA314097071GA SA314097078GA SA314097080GA SA314097085GA SA314097086GA SA314097087GA SA314097090GA SA314097091GA SA314097095GA SA314097096GA SA314097100GA SA314097103GA SA314097104GA SA314097110GA SA314097111GA SA314097113GA SA314097115GA SA314097118GA SA314097119GA SA314097121GA SA314097122GA SA314097123GA SA314097124GA SA314097125GA SA314097128GA SA314097129GA SA314097132GA SA314097133GA SA314097136GA SA314097141GA SA314097143GA SA314097144GA SA314097149GA SA314097152GA SA314097154GA SA314097155GA SA314097156GA SA314097158GA SA314097159GA SA314097160GA SA314097161GA SA314097162GA SA314097163GA SA314097164GA SA314097165GA SA314097170GA SA314097171GA SA314097172GA SA314097173GA SA314097174GA SA314097177GA SA314097178GA SA314097243GA SA314097251GA SA314097255GA SA314097256GA SA314097263GA SA314097264GA SA314097268GA SA314097280GA SA314097282GA SA314097290GA SA314097298GA SA314097300GA SA314097302GA SA314097307GA SA314097309GA SA314097310GA SA314097311GA SA314097313GA SA314097315GA SA314097316GA SA314097318GA SA314097320GA SA314097321GA SA314097322GA SA314097323GA SA314097324GA SA314097325GA SA314097326GA SA314097327GA SA314097329GA SA314097331GA SA314097332GA SA314097333GA SA314097344GA SA314097352GA SA314097355GA SA314097357GA SA314097358GA SA314107366GA SA314107371GA SA314107372GA SA314107373GA SA314107374GA SA314107376GA SA314107380GA SA314107381GA SA314107382GA SA314107384GA SA314107385GA SA314107391GA SA314107392GA SA314107393GA SA314107394GA SA314107395GA SA314107396GA SA314107397GA SA314107398GA SA314107399GA SA314107400GA SA314107401GA SA314107402GA SA314107408GA SA314107411GA SA314107412GA SA314107415GA SA314107418GA SA314107420GA SA314107421GA SA314107422GA SA314107423GA SA314107424GA SA314107429GA SA314107432GA SA314107436GA SA314107439GA SA314107440GA SA314107444GA SA314107445GA SA314107451GA SA314107455GA SA314107460GA SA314107470GA SA314107471GA SA314107472GA SA314107473GA SA314107474GA SA314107475GA SA314107476GA SA314107477GA SA314107478GA SA314107479GA SA314107480GA SA314107481GA SA314107482GA SA314107483GA SA314107484GA SA314107486GA SA314107487GA SA314107488GA SA314107489GA SA314107490GA SA314107491GA SA314107492GA SA314107494GA SA314107495GA SA314107496GA SA314107497GA SA314107498GA SA314107499GA SA314107500GA SA314107501GA SA314107502GA SA314107503GA SA314107504GA SA314107505GA SA314107506GA SA314107507GA SA314107508GA SA314107509GA SA314107510GA SA314107511GA SA314107512GA SA314107513GA SA314107514GA SA314107515GA SA314107516GA SA314107517GA SA314107518GA SA314107519GA SA314107520GA SA314107521GA SA314107522GA SA314107523GA SA314107524GA SA314107525GA SA314107526GA SA314107527GA SA314107528GA SA314107529GA SA314107530GA SA314107531GA SA314107532GA SA314107533GA SA314107535GA SA314107536GA SA314107537GA SA314107538GA SA314107539GA SA314107540GA SA314107541GA SA314107542GA SA314107543GA SA314107544GA SA314107545GA SA314107546GA SA314107547GA SA314107548GA SA314107549GA SA314107550GA SA314107551GA SA314107552GA SA314107553GA SA314107554GA SA314107555GA SA314107556GA SA314107557GA SA314107558GA SA314107559GA SA314107560GA SA314107561GA SA314107562GA SA314107563GA SA314107564GA SA314107566GA SA314107571GA SA314107572GA SA314107574GA SA314107575GA SA314107576GA SA314107578GA SA314107579GA SA314107580GA SA314107581GA SA314107582GA SA314107583GA SA314107584GA SA314107585GA SA314107586GA SA314107588GA SA314107589GA SA314107591GA SA314107593GA SA314107594GA SA314107597GA SA314107598GA SA314107599GA SA314107603GA SA314107606GA SA314107607GA SA314107608GA SA314107612GA SA314107614GA SA314107615GA SA314107616GA SA314107618GA SA314107619GA SA314107650GA SA314107653GA SA314107654GA SA314107655GA SA314107661GA SA314107663GA SA314107664GA SA314107666GA SA314107669GA SA314107670GA SA314107671GA SA314107675GA SA314107678GA SA314107680GA SA314107686GA SA314107688GA SA314107689GA SA314107690GA SA314107692GA SA314107694GA SA314107695GA SA314107696GA SA314107697GA SA314107698GA SA314107699GA SA314107700GA SA314107701GA SA314107702GA SA314107703GA SA314107704GA SA314107705GA SA314107707GA SA314107740GA SA314107741GA SA314107742GA SA314107743GA SA314107744GA SA314107745GA SA314107746GA SA314107747GA SA314107748GA SA314107749GA SA314107750GA SA314107751GA SA314107752GA SA314107753GA SA314107754GA SA314107755GA SA314107756GA SA314107757GA SA314107758GA SA314107759GA SA314107760GA SA314107761GA SA314107762GA SA314107763GA SA314107764GA SA314107765GA SA314107766GA SA314107767GA SA314107768GA SA314107769GA SA314107778GA SA314107781GA SA314107782GA SA314107783GA SA314107784GA SA314107785GA SA314107786GA SA314107787GA SA314107788GA SA314107789GA SA314107790GA SA314107791GA SA314107792GA SA314107793GA SA314107794GA SA314107795GA SA314107796GA SA314107797GA SA314107798GA SA314107799GA SA314107800GA SA314107801GA SA314107802GA SA314107803GA SA314107804GA SA314107805GA SA314107806GA SA314107807GA SA314107838GA SA314107840GA SA314107841GA SA314107842GA SA314107843GA SA314107844GA SA314107845GA SA314107846GA SA314107847GA SA314107849GA SA314107850GA SA314107851GA SA314107852GA SA314107854GA SA314107856GA SA314107858GA SA314107859GA SA314107861GA SA314107862GA SA314107863GA SA314107864GA SA314107865GA SA314107869GA SA314107872GA SA314107873GA SA314107874GA SA314107882GA SA314107885GA SA314107889GA SA314107890GA SA314107892GA SA314107896GA SA314107897GA SA314107898GA SA314107902GA SA314107904GA SA314107905GA SA314107906GA SA314107907GA SA314107909GA SA314107910GA SA314107913GA SA314107917GA SA314107918GA SA314107919GA SA314107922GA SA314107924GA SA314107926GA SA314107927GA SA314107928GA SA314107929GA SA314107930GA SA314107931GA SA314107932GA SA314107933GA SA314107934GA SA314107935GA SA314107936GA SA314107937GA SA314107938GA SA314107939GA SA314107940GA SA314107941GA SA314107942GA SA314107943GA SA314107944GA SA314107945GA SA314107946GA SA314107947GA SA314107948GA SA314107949GA SA314107950GA SA314107951GA SA314107953GA SA314107955GA SA314107956GA SA314107957GA SA314107958GA SA314107959GA SA314107960GA SA314107961GA SA314107963GA SA314107964GA SA314107966GA SA314107967GA SA314107968GA SA314107969GA SA314107970GA SA314107973GA SA314107975GA SA314107976GA SA314107977GA SA314107979GA SA314107981GA SA314107983GA SA314107984GA SA314107985GA SA314107986GA SA314107987GA SA314107988GA SA314107989GA SA314107990GA SA314107991GA SA314107992GA SA314107994GA SA314107995GA SA314107996GA SA314107997GA SA314107999GA SA314108000GA SA314108005GA SA314108006GA SA314108008GA SA314108010GA SA314108011GA SA314108012GA SA314108013GA SA314108014GA SA314108015GA SA314108016GA SA314108017GA SA314108023GA SA314108027GA SA314108029GA SA314108034GA SA314108042GA SA314108045GA SA314108051GA SA314108054GA SA314108058GA SA314108059GA SA314108061GA SA314108068GA SA314108071GA SA314108073GA SA314108074GA SA314108075GA SA314108076GA SA314108077GA SA314108078GA SA314108079GA SA314108080GA SA314108081GA SA314108082GA SA314108086GA SA314108087GA SA314108088GA SA314108089GA SA314108093GA SA314108094GA SA314108095GA SA314108096GA SA314108097GA SA314108098GA SA314108099GA SA314108102GA SA314108103GA SA314108105GA SA314108106GA SA314108107GA SA314108109GA SA314108110GA SA314108111GA SA314108112GA SA314108113GA SA314108114GA SA314108115GA SA314108117GA SA314108118GA SA314108119GA SA314108120GA SA314108121GA SA314108123GA SA314108124GA SA314108127GA SA314108128GA SA314108147GA SA314108152GA SA314108162GA SA314108183GA SA314108201GA SA314108215GA SA314108256GA SA314108270GA SA314108274GA SA314108275GA SA314108279GA SA314108281GA SA314108283GA SA314108284GA SA314108285GA SA314108286GA SA314108287GA SA314108290GA SA314108291GA SA314108293GA SA314108294GA SA314108295GA SA314108299GA SA314108301GA SA314108302GA SA314108304GA SA314108308GA SA314108309GA SA314108310GA SA314118347GA SA314118349GA SA314118352GA SA314118354GA SA314118356GA SA314118357GA SA314118361GA SA314118362GA SA314118363GA SA314118364GA SA314118365GA SA314118366GA SA314118368GA SA314118370GA SA314118373GA SA314118375GA SA314118377GA SA314118380GA SA314118382GA SA314118384GA SA314118385GA SA314118388GA SA314118391GA SA314118393GA SA314118395GA SA314118396GA SA314118399GA SA314118406GA SA314118407GA SA314118408GA SA314118412GA SA314118413GA SA314118450GA SA314118451GA SA314118452GA SA314118453GA SA314118454GA SA314118455GA SA314118456GA SA314118457GA SA314118458GA SA314118459GA SA314118461GA SA314118462GA SA314118463GA SA314118464GA SA314118465GA SA314118466GA SA314118467GA SA314118468GA SA314118469GA SA314118470GA SA314118471GA SA314118472GA SA314118473GA SA314118474GA SA314118475GA SA314118477GA SA314118478GA SA314118479GA SA314118480GA SA314118481GA SA314118482GA SA314118483GA SA314118484GA SA314118485GA SA314118486GA SA314118487GA SA314118488GA SA314118489GA SA314118490GA SA314118491GA SA314118492GA SA314118493GA SA314118495GA SA314118496GA SA314118497GA SA314118498GA SA314118499GA SA314118500GA SA314118501GA SA314118502GA SA314118503GA SA314118504GA SA314118505GA SA314118506GA SA314118507GA SA314118508GA SA314118509GA SA314118517GA SA314118525GA SA314118527GA SA314118529GA SA314118532GA SA314118533GA SA314118534GA SA314118535GA SA314118539GA SA314118559GA SA314118560GA SA314118574GA SA314118575GA SA314118576GA SA314118578GA SA314118579GA SA314118580GA SA314118581GA SA314118582GA SA314118583GA SA314118584GA SA314118585GA SA314118587GA SA314118588GA SA314118589GA SA314118590GA SA314118591GA SA314118592GA SA314118593GA SA314118594GA SA314118595GA SA314118596GA SA314118597GA SA314118598GA SA314118599GA SA314118600GA SA314118602GA SA314118603GA SA314118605GA SA314118606GA SA314118607GA SA314118608GA SA314118609GA SA314118611GA SA314118612GA SA314118613GA SA314118614GA SA314118615GA SA314118616GA SA314118617GA SA314118619GA SA314118621GA SA314118622GA SA314118623GA SA314118624GA SA314118625GA SA314118626GA SA314118627GA SA314118629GA SA314118630GA SA314118631GA SA314118632GA SA314118633GA SA314118634GA SA314118635GA SA314118636GA SA314118637GA SA314118638GA SA314118639GA SA314118640GA SA314118642GA SA314118643GA SA314118644GA SA314118645GA SA314118646GA SA314118647GA SA314118648GA SA314118649GA SA314118651GA SA314118652GA SA314118653GA SA314118654GA SA314118655GA SA314118657GA SA314118658GA SA314118659GA SA314118661GA SA314118662GA SA314118663GA SA314118694GA SA314118695GA SA314118696GA SA314118697GA SA314118698GA SA314118699GA SA314118700GA SA314118701GA SA314118702GA SA314118703GA SA314118704GA SA314118705GA SA314118707GA SA314118708GA SA314118709GA SA314118710GA SA314118711GA SA314118712GA SA314118713GA SA314118714GA SA314118715GA SA314118716GA SA314118717GA SA314118718GA SA314118719GA SA314118720GA SA314118721GA SA314118722GA SA314118723GA SA314118724GA SA314118725GA SA314118726GA SA314118727GA SA314118728GA SA314118729GA SA314118730GA SA314118731GA SA314118736GA SA314118740GA SA314118743GA SA314118746GA SA314118748GA SA314118749GA SA314118750GA SA314118752GA SA314118753GA SA314118816GA SA314118817GA SA314118819GA SA314118825GA SA314118826GA SA314118827GA SA314118828GA SA314118829GA SA314118830GA SA314118831GA SA314118832GA SA314118833GA SA314118834GA SA314118836GA SA314118839GA SA314118840GA SA314118841GA SA314118842GA SA314118844GA SA314118845GA SA314118846GA SA314118848GA SA314118849GA SA314118851GA SA314118852GA SA314118853GA SA314118854GA SA314118855GA SA314118857GA SA314118858GA SA314118859GA SA314118860GA SA314118861GA SA314118862GA SA314118863GA SA314118864GA SA314118866GA SA314118868GA SA314118871GA SA314118882GA SA314118894GA SA314118895GA SA314118903GA SA314118951GA SA314118957GA SA314118961GA SA314118962GA SA314118967GA SA314118972GA SA314118973GA SA314118982GA SA314118983GA SA314118984GA SA314118986GA SA314118987GA SA314118988GA SA314118989GA SA314118990GA SA314118991GA SA314118992GA SA314118993GA SA314118994GA SA314118996GA SA314118997GA SA314119000GA SA314119002GA SA314119004GA SA314119005GA SA314119006GA SA314119008GA SA314119009GA SA314119010GA SA314119011GA SA314119013GA SA314119014GA SA314119016GA SA314119019GA SA314119020GA SA314119023GA SA314119027GA SA314119028GA SA314119029GA SA314119031GA SA314119032GA SA314119034GA SA314119035GA SA314119036GA SA314119037GA SA314119039GA SA314119040GA SA314119041GA SA314119042GA SA314119043GA SA314119044GA SA314119045GA SA314119084GA SA314119085GA SA314119091GA SA314119094GA SA314119099GA SA314119106GA SA314119107GA SA314119110GA SA314119114GA SA314119117GA SA314119119GA SA314119128GA SA314129136GA SA314129139GA SA314129143GA SA314129145GA SA314129149GA SA314129151GA SA314129155GA SA314129159GA SA314129160GA SA314129168GA SA314129173GA SA314129174GA SA314129176GA SA314129191GA SA314129192GA SA314129193GA SA314129194GA SA314129195GA SA314129199GA SA314129205GA SA314129207GA SA314129209GA SA314129224GA SA314129228GA SA314129265GA SA314129267GA SA314129275GA SA314129276GA SA314129286GA SA314129287GA SA314129292GA SA314129295GA SA314129296GA SA314129297GA SA314129298GA SA314129299GA SA314129304GA SA314129310GA SA314129315GA SA314139347GA SA314139348GA SA314139352GA SA314139354GA SA314139360GA SA314139362GA SA314139363GA SA314139367GA SA314139369GA SA314139370GA SA314139371GA SA314139378GA SA314139386GA SA314139387GA SA314139389GA SA314139401GA SA314139407GA SA314139419GA SA314139421GA SA314139424GA SA314139427GA SA314139429GA SA314139431GA SA314139434GA SA314139442GA SA314139450GA SA314139451GA SA314139456GA SA314139462GA SA314139464GA SA314139468GA SA314139472GA SA314139473GA SA314139474GA SA314139476GA SA314139477GA SA314139478GA SA314139479GA SA314139480GA SA314139481GA SA314139482GA SA314139488GA SA314139491GA SA314139493GA SA314139494GA SA314139495GA SA314139496GA SA314139501GA SA314139503GA SA314139517GA SA314139526GA SA314139531GA SA314139538GA SA314139546GA SA314139551GA SA314139558GA SA314139564GA SA314139565GA SA314139568GA SA314139572GA SA314139574GA SA314139575GA SA314139583GA SA314139585GA SA314139587GA SA314139592GA SA314139594GA SA314139596GA SA314139598GA SA314149618GA SA314149619GA SA314149620GA SA314149622GA SA314149623GA SA314149624GA SA314149627GA SA314149628GA SA314149629GA SA314149631GA SA314149663GA SA314149664GA SA314149665GA SA314149666GA SA314149668GA SA314149669GA SA314149670GA SA314149671GA SA314149672GA SA314149673GA SA314149674GA SA314149675GA SA314149676GA SA314149677GA SA314149678GA SA314149679GA SA314149680GA SA314149681GA SA314149682GA SA314149683GA SA314149684GA SA314149685GA SA314149686GA SA314149687GA SA314149688GA SA314149689GA SA314149690GA SA314149691GA SA314149692GA SA314149693GA SA314149694GA SA314149695GA SA314149696GA SA314149697GA SA314149698GA SA314149699GA SA314149700GA SA314149701GA SA314149702GA SA314149703GA SA314149705GA SA314149706GA SA314149707GA SA314149708GA SA314149709GA SA314149710GA SA314149711GA SA314149712GA SA314149713GA SA314149714GA SA314149715GA SA314149716GA SA314149717GA SA314149718GA SA314149719GA SA314149720GA SA314149721GA SA314149722GA SA314149723GA SA314149724GA SA314149725GA SA314149726GA SA314149727GA SA314149728GA SA314149729GA SA314149730GA SA314149731GA SA314149732GA SA314149733GA SA314149734GA SA314149735GA SA314149736GA SA314149737GA SA314149738GA SA314149739GA SA314149740GA SA314149741GA SA314149742GA SA314149743GA SA314149744GA SA314149745GA SA314149746GA SA314149748GA SA314149749GA SA314149750GA SA314149751GA SA314149752GA SA314149753GA SA314149754GA SA314149755GA SA314149756GA SA314149758GA SA314149759GA SA314149760GA SA314149761GA SA314159762GA SA314159763GA SA314159764GA SA314159765GA SA314159766GA SA314159767GA SA314159768GA SA314159769GA SA314159770GA SA314159771GA SA314159772GA SA314159773GA SA314159774GA SA314159775GA SA314159776GA SA314159777GA SA314159778GA SA314159779GA SA314159780GA SA314159781GA SA314159782GA SA314159783GA SA314159784GA SA314159785GA SA314159786GA SA314159787GA SA314159788GA SA314159789GA SA314159790GA SA314159791GA SA314159792GA SA314159793GA SA314159794GA SA314159795GA SA314159796GA SA314159797GA SA314159798GA SA314159799GA SA314159801GA SA314159802GA SA314159803GA SA314159804GA SA314159805GA SA314159806GA SA314159807GA SA314159808GA SA314159809GA SA314159810GA SA314159811GA SA314159812GA SA314159813GA SA314159815GA SA314159816GA SA314159817GA SA314159818GA SA314159820GA SA314159821GA SA314159822GA SA314160001GA SA314160003GA SA314160004GA SA314160007GA SA314160011GA SA314160012GA SA314160013GA SA314160018GA SA314160019GA SA314160020GA SA314160021GA SA314160023GA SA314160024GA SA314160025GA SA314160026GA SA314160027GA SA314160028GA SA314160029GA SA314160030GA SA314160031GA SA314160032GA SA314160033GA SA314160040GA SA314160043GA SA314160053GA SA314160056GA SA314160057GA SA314160060GA SA314160061GA SA314160063GA SA314160065GA SA314160067GA SA314160074GA SA314160077GA SA314160079GA SA314160082GA SA314160088GA SA314160091GA SA314160095GA SA314160096GA SA314160098GA SA314160099GA SA314160102GA SA314160104GA SA314160105GA SA314160107GA SA314160108GA SA314160110GA SA314160114GA SA314160118GA SA314169824GA SA314169825GA SA314169829GA SA314169840GA SA314169843GA SA314169844GA SA314169848GA SA314169853GA SA314169854GA SA314169855GA SA314169856GA SA314169857GA SA314169858GA SA314169859GA SA314169860GA SA314169861GA SA314169862GA SA314169864GA SA314169866GA SA314169867GA SA314169868GA SA314169869GA SA314169871GA SA314169873GA SA314169874GA SA314169875GA SA314169876GA SA314169877GA SA314169878GA SA314169879GA SA314169880GA SA314169883GA SA314169884GA SA314169885GA SA314169886GA SA314169887GA SA314169888GA SA314169889GA SA314169890GA SA314169891GA SA314169892GA SA314169893GA SA314169896GA SA314169897GA SA314169898GA SA314169899GA SA314169900GA SA314169902GA SA314169904GA SA314169905GA SA314169906GA SA314169908GA SA314169911GA SA314169913GA SA314169914GA SA314169916GA SA314169920GA SA314169921GA SA314169923GA SA314169924GA SA314169925GA SA314169926GA SA314169927GA SA314169929GA SA314169930GA SA314169931GA SA314169933GA SA314169935GA SA314169936GA SA314169937GA SA314169938GA SA314169939GA SA314169940GA SA314169941GA SA314170122GA SA314170123GA SA314170124GA SA314170125GA SA314170127GA SA314170128GA SA314170129GA SA314170130GA SA314170132GA SA314170133GA SA314170134GA SA314170135GA SA314170136GA SA314170137GA SA314170139GA SA314170140GA SA314170141GA SA314170142GA SA314170143GA SA314170145GA SA314170146GA SA314170147GA SA314170148GA SA314170149GA SA314170151GA SA314170152GA SA314170154GA SA314170155GA SA314170156GA SA314170157GA SA314170158GA SA314170160GA SA314170161GA SA314170163GA SA314170166GA SA314170168GA SA314170169GA SA314170170GA SA314170171GA SA314170173GA SA314170175GA SA314170176GA SA314170177GA SA314170178GA SA314170179GA SA314170181GA SA314170182GA SA314170186GA SA314170192GA SA314170196GA SA314170204GA SA314170205GA SA314170207GA SA314170208GA SA314170210GA SA314170211GA SA314170214GA SA314170216GA SA314170217GA SA314170220GA SA314170225GA SA314170230GA SA314170231GA SA314170240GA SA314170242GA SA314170243GA SA314170244GA SA314170245GA SA314170246GA SA314170248GA SA314170249GA SA314170250GA SA314170251GA SA314170252GA SA314170253GA SA314170254GA SA314170256GA SA314170257GA SA314170258GA SA314170259GA SA314170260GA SA314170261GA SA314170262GA SA314170263GA SA314170264GA SA314170265GA SA314170266GA SA314170277GA SA314170278GA SA314170279GA SA314170287GA SA314170288GA SA314170292GA SA314170293GA SA314170295GA SA314170296GA SA314170301GA SA314170309GA SA314170311GA SA314170312GA SA314170313GA SA314170314GA SA314170316GA SA314170317GA SA314170318GA SA314170319GA SA314170320GA SA314170321GA SA314170322GA SA314170323GA SA314170324GA SA314170325GA SA314170326GA SA314170328GA SA314170329GA SA314170330GA SA314170331GA SA314170332GA SA314170333GA SA314170334GA SA314170335GA SA314170336GA SA314170337GA SA314170338GA SA314170339GA SA314170340GA SA314170341GA SA314170355GA SA314170361GA SA314170363GA SA314170364GA SA314170365GA SA314170367GA SA314170368GA SA314170369GA SA314170370GA SA314170371GA SA314170372GA SA314170373GA SA314170374GA SA314170375GA SA314170376GA SA314170377GA SA314170378GA SA314170379GA SA314170380GA SA314170381GA SA314170382GA SA314170383GA SA314170385GA SA314170390GA SA314170391GA SA314170392GA SA314170393GA SA314170394GA SA314170395GA SA314170397GA SA314170401GA SA314170402GA SA314170406GA SA314170410GA SA314170413GA SA314170414GA SA314170415GA SA314170419GA SA314170420GA SA314170421GA SA314170422GA SA314170425GA SA314170426GA SA314170427GA SA314170428GA SA314170430GA SA314170431GA SA314170434GA SA314170436GA SA314170447GA SA314170448GA SA314170449GA SA314170450GA SA314170451GA SA314170452GA SA314170453GA SA314170455GA SA314170456GA SA314170458GA SA314170459GA SA314170460GA SA314170461GA SA314170465GA SA314170466GA SA314170470GA SA314180472GA SA314180473GA SA314180476GA SA314180477GA SA314180478GA SA314180480GA SA314180484GA SA314180493GA SA314180497GA SA314180498GA SA314180503GA SA314180504GA SA314180506GA SA314180508GA SA314180510GA SA314180511GA SA314180513GA SA314180518GA SA314180519GA SA314180522GA SA314180525GA SA314180527GA SA314180528GA SA314180529GA SA314180530GA SA314180531GA SA314180534GA SA314180535GA SA314180537GA SA314180538GA SA314180539GA SA314180543GA SA314180544GA SA314180545GA SA314180546GA SA314180547GA SA314180549GA SA314180551GA SA314180555GA SA314180558GA SA314180559GA SA314180561GA SA314180562GA SA314180566GA SA314180567GA SA314180568GA SA314180577GA SA314180584GA SA314180585GA SA314180589GA SA314180590GA SA314180599GA SA314180603GA SA314180604GA SA314180606GA SA314180609GA SA314180614GA SA314180615GA SA314180616GA SA314180618GA SA314180620GA SA314180621GA SA314180622GA SA314180623GA SA314180625GA SA314180628GA SA314180633GA SA314180634GA SA314180642GA SA314180646GA SA314180650GA SA314180652GA SA314180665GA SA314180667GA SA314180669GA SA314180670GA SA314180671GA SA314180673GA SA314180674GA SA314180675GA SA314180676GA SA314180677GA SA314180679GA SA314180680GA SA314180681GA SA314180682GA SA314180683GA SA314180684GA SA314180685GA SA314180686GA SA314180687GA SA314180688GA SA314180689GA SA314180690GA SA314180691GA SA314180692GA SA314180693GA SA314180694GA SA314180695GA SA314180697GA SA314180703GA SA314180704GA SA314180708GA SA314180709GA SA314180710GA SA314180711GA SA314180713GA SA314180714GA SA314180715GA SA314180718GA SA314180719GA SA314180722GA SA314180723GA SA314180724GA SA314180726GA SA314180728GA SA314180729GA SA314180731GA SA314180735GA SA314180736GA SA314180737GA SA314180738GA SA314180740GA SA314180742GA SA314180743GA SA314180744GA SA314180745GA SA314180746GA SA314180747GA SA314180748GA SA314180749GA SA314180750GA SA314180753GA SA314180754GA SA314180755GA SA314180756GA SA314180757GA SA314180758GA SA314180759GA SA314180760GA SA314180762GA SA314180764GA SA314180766GA SA314180768GA SA314180769GA SA314180770GA SA314180771GA SA314180772GA SA314180773GA SA314180775GA SA314180777GA SA314180778GA SA314180779GA SA314180782GA SA314180783GA SA314180785GA SA314180786GA SA314180787GA SA314190788GA SA314190789GA SA314190790GA SA314190792GA SA314190793GA SA314190794GA SA314190795GA SA314190796GA SA314190799GA SA314190800GA SA314190801GA SA314190802GA SA314190803GA SA314190804GA SA314190806GA SA314190807GA SA314190808GA SA314190809GA SA314190810GA SA314190811GA SA314190812GA SA314190813GA SA314190814GA SA314190815GA SA314190816GA SA314190817GA SA314190818GA SA314190819GA SA314190820GA SA314190821GA SA314190822GA SA314190823GA SA314190824GA SA314190825GA SA314190826GA SA314190827GA SA314190828GA SA314190829GA SA314190830GA SA314190831GA SA314190832GA SA314190833GA SA314190834GA SA314190835GA SA314190836GA SA314190837GA SA314190838GA SA314190839GA SA314190840GA SA314190841GA SA314190842GA SA314190843GA SA314190844GA SA314190845GA SA314190846GA SA314190847GA SA314190848GA SA314190849GA SA314190851GA SA314190852GA SA314190853GA SA314190854GA SA314190855GA SA314190856GA SA314190858GA SA314190860GA SA314190861GA SA314190863GA SA314190864GA SA314190865GA SA314190866GA SA314190867GA SA314190868GA SA314190869GA SA314190871GA SA314190873GA SA314190874GA SA314190875GA SA314190876GA SA314190877GA SA314190878GA SA314190879GA SA314190880GA SA314190882GA SA314190883GA SA314190884GA SA314190886GA SA314190887GA SA314190888GA SA314190889GA SA314190890GA SA314190891GA SA314190892GA SA314190893GA SA314190894GA SA314190895GA SA314190896GA SA314190897GA SA314190898GA SA314190899GA SA314190900GA SA314190901GA SA314190903GA SA314190904GA SA314190905GA SA314190907GA SA314190909GA SA314190910GA SA314190911GA SA314190912GA SA314190913GA SA314190914GA SA314190915GA SA314190916GA SA314190918GA SA314190919GA SA314190920GA SA314190921GA SA314190922GA SA314190924GA SA314190927GA SA314190928GA SA314190929GA SA314190930GA SA314190931GA SA314190934GA SA314190935GA SA314190937GA SA314190938GA SA314190939GA SA314190941GA SA314190943GA SA314190944GA SA314190945GA SA314190946GA SA314190947GA SA314190948GA SA314190949GA SA314190950GA SA314190953GA SA314190954GA SA314190955GA SA314190956GA SA314190957GA SA314190958GA SA314190959GA SA314190960GA SA314190961GA SA314190962GA SA314190963GA SA314190966GA SA314190967GA SA314190968GA SA314190970GA SA314190971GA SA314190972GA SA314190973GA SA314190974GA SA314190975GA SA314190976GA SA314190977GA SA314190978GA SA314190979GA SA314190980GA SA314190982GA SA314190984GA SA314190985GA SA314190986GA SA314190987GA SA314190988GA SA314190989GA SA314190990GA SA314190991GA SA314190992GA SA314190993GA SA314190994GA SA314190995GA SA314190996GA SA314190997GA SA314190998GA SA314190999GA SA314191000GA SA314191001GA SA314191002GA SA314191003GA SA314191004GA SA314191005GA SA314191006GA SA314191007GA SA314191008GA SA314191009GA SA314191010GA SA314191011GA SA314191012GA SA314191013GA SA314191014GA SA314191015GA SA314191016GA SA314191017GA SA314191020GA SA314191021GA SA314191022GA SA314191024GA SA314191025GA SA314191026GA SA314191027GA SA314191028GA SA314191029GA SA314191030GA SA314191031GA SA314191032GA SA314191033GA SA314191034GA SA314191035GA SA314191036GA SA314191037GA SA314191038GA SA314191039GA SA314191040GA SA314191041GA SA314191042GA SA314191043GA SA314191044GA SA314191046GA SA314191047GA SA314191048GA SA314191049GA SA314191050GA SA314191051GA SA314191052GA SA314191053GA SA314191054GA SA314191055GA SA314191056GA SA314191057GA SA314191058GA SA314191059GA SA314191060GA SA314191061GA SA314191062GA SA314191063GA SA314191064GA SA314191065GA SA314191066GA SA314191067GA SA314191068GA SA314191069GA SA314191070GA SA314191071GA SA314191072GA SA314191073GA SA314191074GA SA314191075GA SA314191076GA SA314191077GA SA314191078GA SA314191079GA SA314191080GA SA314191081GA SA314191082GA SA314191083GA SA314191084GA SA314191085GA SA314191086GA SA314191087GA SA314191088GA SA314191089GA SA314191090GA SA314191091GA SA314191092GA SA314191093GA SA314191094GA SA314191095GA SA314191096GA SA314191097GA SA314191098GA SA314191099GA SA314191100GA SA314191101GA SA314191102GA SA314191103GA SA314191104GA SA314191105GA SA314191106GA SA314191107GA SA314191108GA SA314191109GA SA314191110GA SA314191111GA SA314191112GA SA314191113GA SA314191114GA SA314191115GA SA314191116GA SA314191117GA SA314191118GA SA314191119GA SA314191120GA SA314191121GA SA314191122GA SA314191124GA SA314191126GA SA314191128GA SA314191129GA SA314191131GA SA314191132GA SA314191133GA SA314191134GA SA314191135GA SA314191136GA SA314191137GA SA314191138GA SA314191139GA SA314191140GA SA314191141GA SA314191144GA SA314191145GA SA314191146GA SA314191148GA SA314191150GA SA314191151GA SA314191152GA SA314191153GA SA314191154GA SA314191155GA SA314191157GA SA314191159GA SA314191161GA SA314191163GA SA314191164GA SA314191165GA SA314191167GA SA314191168GA SA314191169GA SA314191170GA SA314191171GA SA314191172GA SA314191174GA SA314191175GA SA314191176GA SA314191177GA SA314191179GA SA314191180GA SA314191181GA SA314191182GA SA314191183GA SA314191184GA SA314191185GA SA314191186GA SA314191187GA SA314191188GA SA314191189GA SA314191190GA SA314191191GA SA314191192GA SA314191196GA SA314191197GA SA314191198GA SA314201207GA SA314201208GA SA314201209GA SA314201210GA SA314201211GA SA314201212GA SA314201213GA SA314201214GA SA314201215GA SA314201216GA SA314201217GA SA314201218GA SA314201219GA SA314201220GA SA314201221GA SA314201222GA SA314201224GA SA314201225GA SA314201227GA SA314201229GA SA314201230GA SA314201231GA SA314201232GA SA314201233GA SA314201234GA SA314201235GA SA314201236GA SA314201237GA SA314201238GA SA314201239GA SA314201240GA SA314201241GA SA314201242GA SA314201243GA SA314201244GA SA314201245GA SA314201246GA SA314201247GA SA314201248GA SA314201249GA SA314201250GA SA314201251GA SA314201253GA SA314201254GA SA314201255GA SA314201256GA SA314201257GA SA314201258GA SA314201259GA SA314201260GA SA314201261GA SA314201262GA SA314201263GA SA314201264GA SA314201265GA SA314201266GA SA314201267GA SA314201269GA SA314201270GA SA314201271GA SA314201272GA SA314201273GA SA314201274GA SA314201276GA SA314201277GA SA314201278GA SA314201279GA SA314201280GA SA314201281GA SA314201282GA SA314201283GA SA314201284GA SA314201285GA SA314201287GA SA314201288GA SA314201289GA SA314201290GA SA314201291GA SA314201292GA SA314201293GA SA314201294GA SA314201295GA SA314201296GA SA314201297GA SA314201298GA SA314201299GA SA314201300GA SA314201301GA SA314201302GA SA314201303GA SA314201304GA SA314201305GA SA314201306GA SA314201307GA SA314201308GA SA314201309GA SA314201310GA SA314201311GA SA314201312GA SA314201313GA SA314201314GA SA314201315GA SA314201316GA SA314201317GA SA314201318GA SA314201319GA SA314201320GA SA314201321GA SA314201322GA SA314201323GA SA314201324GA SA314201325GA SA314201326GA SA314201327GA SA314201328GA SA314201329GA SA314201330GA SA314201331GA SA314201332GA SA314201333GA SA314201334GA SA314201335GA SA314201336GA SA314201338GA SA314201339GA SA314201340GA SA314201341GA SA314201342GA SA314201343GA SA314201344GA SA314201346GA SA314201347GA SA314201348GA SA314201349GA SA314201351GA SA314201352GA SA314201353GA SA314201354GA SA314201355GA SA314201356GA SA314201357GA SA314201358GA SA314201359GA SA314201360GA SA314201361GA SA314201362GA SA314201363GA SA314201364GA SA314201366GA SA314201367GA SA314201368GA SA314201369GA SA314201370GA SA314201371GA SA314201372GA SA314201373GA SA314201374GA SA314201375GA SA314201376GA SA314201377GA SA314201378GA SA314201379GA SA314201380GA SA314201400GA SA314201401GA SA314201406GA SA314201410GA SA314201412GA SA314201426GA SA314201441GA SA314201442GA SA314201443GA SA314201444GA SA314201445GA SA314201446GA SA314201447GA SA314201448GA SA314201449GA SA314201450GA SA314201451GA SA314201452GA SA314201453GA SA314201454GA SA314201455GA SA314201456GA SA314201457GA SA314201458GA SA314201459GA SA314201460GA SA314201461GA SA314201462GA SA314201463GA SA314201464GA SA314201465GA SA314201466GA SA314201467GA SA314201468GA SA314201469GA SA314201470GA SA314201478GA SA314201482GA SA314201497GA SA314201499GA SA314201503GA SA314201507GA SA314201509GA SA314201510GA SA314201512GA SA314201516GA SA314201517GA SA314201522GA SA314201523GA SA314201525GA SA314201526GA SA314201527GA SA314201530GA SA314201531GA SA314201532GA SA314201533GA SA314201534GA SA314201535GA SA314201536GA SA314201537GA SA314201538GA SA314201539GA SA314201540GA SA314201541GA SA314201542GA SA314201543GA SA314201544GA SA314201545GA SA314201546GA SA314201547GA SA314201548GA SA314201549GA SA314201550GA SA314201551GA SA314201552GA SA314201553GA SA314201554GA SA314201555GA SA314201556GA SA314201557GA SA314201558GA SA314201559GA SA314201560GA SA314201561GA SA314201562GA SA314201563GA SA314201564GA SA314201565GA SA314201566GA SA314201567GA SA314201568GA SA314201569GA SA314201570GA SA314201571GA SA314201572GA SA314201573GA SA314201574GA SA314201575GA SA314201576GA SA314201577GA SA314201578GA SA314201579GA SA314201580GA SA314201581GA SA314201582GA SA314201583GA SA314201584GA SA314201585GA SA314201586GA SA314201587GA SA314201589GA SA314201590GA SA314201592GA SA314201596GA SA314201597GA SA314201599GA SA314201600GA SA314201601GA SA314201602GA SA314201603GA SA314201604GA SA314201605GA SA314201606GA SA314201611GA SA314201613GA SA314201614GA SA314201615GA SA314201618GA SA314201619GA SA314201621GA SA314201622GA SA314201623GA SA314201624GA SA314201625GA SA314201626GA SA314201627GA SA314201628GA SA314201629GA SA314201630GA SA314201632GA SA314201633GA SA314201634GA SA314201635GA SA314201636GA SA314201637GA SA314201638GA SA314201639GA SA314201640GA SA314201641GA SA314201642GA SA314201643GA SA314201644GA SA314201645GA SA314201646GA SA314201647GA SA314201648GA SA314201649GA SA314201650GA SA314201651GA SA314201652GA SA314201653GA SA314201654GA SA314201655GA SA314201656GA SA314201657GA SA314201658GA SA314201659GA SA314201660GA SA314201661GA SA314201662GA SA314201663GA SA314201666GA SA314201667GA SA314201669GA SA314201671GA SA314201672GA SA314201673GA SA314201674GA SA314201675GA SA314201677GA SA314201678GA SA314201680GA SA314201681GA SA314201682GA SA314201683GA SA314201685GA SA314201686GA SA314201687GA SA314201688GA SA314201689GA SA314201690GA SA314201691GA SA314201692GA SA314201695GA SA314201696GA SA314201697GA SA314201698GA SA314201700GA SA314201702GA SA314201703GA SA314201704GA SA314201706GA SA314201708GA SA314201710GA SA314201711GA SA314201713GA SA314201714GA SA314201715GA SA314201717GA SA314201718GA SA314201719GA SA314201723GA SA314201724GA SA314201725GA SA314201727GA SA314201730GA SA314201732GA SA314201737GA SA314201738GA SA314201741GA SA314201742GA SA314201743GA SA314201744GA SA314201745GA SA314201746GA SA314201747GA SA314201748GA SA314201750GA SA314201751GA SA314201752GA SA314201754GA SA314201755GA SA314201757GA SA314201758GA SA314201760GA SA314201761GA SA314201762GA SA314201763GA SA314201765GA SA314201766GA SA314201767GA SA314201768GA SA314201769GA SA314201770GA SA314201771GA SA314201772GA SA314201773GA SA314201774GA SA314201775GA SA314201776GA SA314201777GA SA314201778GA SA314201779GA SA314201783GA SA314201784GA SA314201785GA SA314201786GA SA314201790GA SA314201791GA SA314201795GA SA314201796GA SA314201798GA SA314201799GA SA314201800GA SA314201801GA SA314201802GA SA314201803GA SA314201806GA SA314201807GA SA314201808GA SA314201809GA SA314201810GA SA314201812GA SA314201813GA SA314201814GA SA314201816GA SA314201818GA SA314201819GA SA314201820GA SA314201821GA SA314201823GA SA314201824GA SA314201825GA SA314201826GA SA314201827GA SA314201828GA SA314201829GA SA314201830GA SA314201832GA SA314201834GA SA314201835GA SA314201836GA SA314201837GA SA314201839GA SA314201841GA SA314211850GA SA314211852GA SA314211855GA SA314211856GA SA314211859GA SA314211860GA SA314211861GA SA314211862GA SA314211863GA SA314211864GA SA314211865GA SA314211866GA SA314211867GA SA314211868GA SA314211869GA SA314211870GA SA314211871GA SA314211872GA SA314211873GA SA314211874GA SA314211875GA SA314211876GA SA314211877GA SA314211878GA SA314211879GA SA314211880GA SA314211881GA SA314211882GA SA314211883GA SA314211884GA SA314211885GA SA314211886GA SA314211887GA SA314211888GA SA314211889GA SA314211890GA SA314211891GA SA314211893GA SA314211894GA SA314211895GA SA314211896GA SA314211897GA SA314211898GA SA314211899GA SA314211900GA SA314211901GA SA314211902GA SA314211903GA SA314211904GA SA314211906GA SA314211907GA SA314211909GA SA314211910GA SA314211911GA SA314211912GA SA314211913GA SA314211914GA SA314211915GA SA314211916GA SA314211917GA SA314211918GA SA314211919GA SA314211920GA SA314211922GA SA314211925GA SA314211927GA SA314211930GA SA314211931GA SA314211933GA SA314211935GA SA314211937GA SA314211939GA SA314211941GA SA314211951GA SA314211952GA SA314211954GA SA314211955GA SA314211956GA SA314211959GA SA314211960GA SA314211961GA SA314211962GA SA314211963GA SA314211964GA SA314211965GA SA314211966GA SA314211967GA SA314211968GA SA314211969GA SA314211970GA SA314211971GA SA314211972GA SA314211973GA SA314211974GA SA314211975GA SA314211976GA SA314211977GA SA314211978GA SA314211979GA SA314211980GA SA314211981GA SA314211983GA SA314211984GA SA314211986GA SA314211987GA SA314211988GA SA314211989GA SA314211992GA SA314211993GA SA314211994GA SA314211995GA SA314211996GA SA314211997GA SA314211998GA SA314211999GA SA314212000GA SA314212001GA SA314212002GA SA314212003GA SA314212004GA SA314212005GA SA314212006GA SA314212009GA SA314212010GA SA314212013GA SA314212015GA SA314212016GA SA314212018GA SA314212020GA SA314212025GA SA314212026GA SA314212027GA SA314212029GA SA314212031GA SA314212033GA SA314212037GA SA314212042GA SA314212043GA SA314212044GA SA314212046GA SA314212047GA SA314212048GA SA314212051GA SA314212054GA SA314212055GA SA314212056GA SA314212059GA SA314212060GA SA314212061GA SA314212063GA SA314212102GA SA314212109GA SA314212117GA SA314212124GA SA314212126GA SA314212127GA SA314212128GA SA314212129GA SA314212130GA SA314212131GA SA314212134GA SA314212135GA SA314212136GA SA314212137GA SA314212138GA SA314212140GA SA314212141GA SA314212142GA SA314212144GA SA314212147GA SA314212148GA SA314212149GA SA314212150GA SA314212151GA SA314212152GA SA314212153GA SA314212154GA SA314212155GA SA314212156GA SA314212157GA SA314212158GA SA314212160GA SA314212161GA SA314212162GA SA314212164GA SA314212165GA SA314212166GA SA314212167GA SA314212169GA SA314212170GA SA314212171GA SA314212173GA SA314212174GA SA314212176GA SA314212177GA SA314212178GA SA314212179GA SA314212180GA SA314212181GA SA314212182GA SA314212183GA SA314212185GA SA314212186GA SA314212189GA SA314212190GA SA314212191GA SA314212193GA SA314212194GA SA314212195GA SA314212196GA SA314212197GA SA314212198GA SA314212200GA SA314212203GA SA314212205GA SA314212206GA SA314212207GA SA314212208GA SA314212209GA SA314212210GA SA314212215GA SA314212216GA SA314212218GA SA314212219GA SA314212221GA SA314212223GA SA314212224GA SA314212227GA SA314212230GA SA314212237GA SA314212242GA SA314212243GA SA314212244GA SA314212245GA SA314212247GA SA314212249GA SA314212251GA SA314212252GA SA314212253GA SA314212256GA SA314222274GA SA314222275GA SA314222276GA SA314222278GA SA314222281GA SA314222286GA SA314222289GA SA314222295GA SA314222296GA SA314222297GA SA314222298GA SA314222300GA SA314222301GA SA314222302GA SA314222303GA SA314222304GA SA314222305GA SA314222306GA SA314222313GA SA314222315GA SA314222316GA SA314222317GA SA314222323GA SA314222325GA SA314222335GA SA314222349GA SA314222350GA SA314222351GA SA314222352GA SA314222353GA SA314222355GA SA314222358GA SA314222359GA SA314222361GA SA314222362GA SA314222363GA SA314222365GA SA314222369GA SA314222372GA SA314222373GA SA314222374GA SA314222375GA SA314222376GA SA314222381GA SA314222383GA SA314222385GA SA314222386GA SA314222391GA SA314222393GA SA314222395GA SA314222396GA SA314222398GA SA314222399GA SA314222400GA SA314222401GA SA314222404GA SA314222405GA SA314222406GA SA314222407GA SA314222410GA SA314222411GA SA314222416GA SA314222417GA SA314222418GA SA314222419GA SA314222422GA SA314222423GA SA314222424GA SA314222425GA SA314222428GA SA314222429GA SA314222430GA SA314222432GA SA314222433GA SA314222434GA SA314222435GA SA314222436GA SA314222437GA SA314222438GA SA314222439GA SA314222440GA SA314222441GA SA314222442GA SA314222443GA SA314222444GA SA314222445GA SA314222447GA SA314222448GA SA314222449GA SA314222451GA SA314222452GA SA314222454GA SA314222455GA SA314222456GA SA314222458GA SA314222460GA SA314222461GA SA314222462GA SA314222463GA SA314222464GA SA314222465GA SA314222466GA SA314222467GA SA314222468GA SA314222471GA SA314222472GA SA314222473GA SA314222476GA SA314222477GA SA314222478GA SA314222479GA SA314222480GA SA314222481GA SA314222482GA SA314222483GA SA314222484GA SA314222485GA SA314222486GA SA314222487GA SA314222488GA SA314222489GA SA314222490GA SA314222493GA SA314222495GA SA314222496GA SA314222497GA SA314222498GA SA314222499GA SA314222500GA SA314222501GA SA314222502GA SA314222503GA SA314222504GA SA314222505GA SA314222506GA SA314222507GA SA314222508GA SA314222509GA SA314222510GA SA314222511GA SA314222512GA SA314222514GA SA314222515GA SA314222516GA SA314222518GA SA314222519GA SA314222521GA SA314222522GA SA314222523GA SA314222524GA SA314222525GA SA314222526GA SA314222527GA SA314222528GA SA314222529GA SA314222531GA SA314222532GA SA314222533GA SA314222534GA SA314222535GA SA314222536GA SA314222537GA SA314222538GA SA314222539GA SA314222540GA SA314222542GA SA314222543GA SA314222544GA SA314222545GA SA314222546GA SA314222547GA SA314222548GA SA314222549GA SA314222550GA SA314222551GA SA314222552GA SA314222553GA SA314222554GA SA314222555GA SA314222556GA SA314222557GA SA314222558GA SA314222559GA SA314222560GA SA314222561GA SA314222562GA SA314222563GA SA314222565GA SA314222568GA SA314222569GA SA314222570GA SA314222571GA SA314222572GA SA314222573GA SA314222574GA SA314222576GA SA314222577GA SA314222579GA SA314222580GA SA314222581GA SA314222582GA SA314222583GA SA314222584GA SA314222585GA SA314222588GA SA314222591GA SA314222592GA SA314222593GA SA314222594GA SA314222599GA SA314222600GA SA314222601GA SA314222602GA SA314222603GA SA314222604GA SA314222607GA SA314222608GA SA314222610GA SA314222611GA SA314222612GA SA314222613GA SA314222614GA SA314222615GA SA314222616GA SA314222617GA SA314222618GA SA314222619GA SA314222621GA SA314222623GA SA314222624GA SA314222625GA SA314222626GA SA314222627GA SA314222628GA SA314222629GA SA314222630GA SA314222631GA SA314222633GA SA314222634GA SA314222635GA SA314222637GA SA314222638GA SA314222639GA SA314222640GA SA314222641GA SA314222642GA SA314222643GA SA314222644GA SA314222645GA SA314222648GA SA314222650GA SA314222651GA SA314222652GA SA314222653GA SA314222654GA SA314222655GA SA314222656GA SA314222657GA SA314222658GA SA314222659GA SA314222660GA SA314222661GA SA314222662GA SA314222663GA SA314222664GA SA314222665GA SA314222666GA SA314222667GA SA314222668GA SA314222669GA SA314222670GA SA314222671GA SA314222672GA SA314222673GA SA314222674GA SA314222675GA SA314222676GA SA314222677GA SA314222678GA SA314222709GA SA314222711GA SA314222715GA SA314222719GA SA314222722GA SA314222724GA SA314222725GA SA314222729GA SA314222732GA SA314222735GA SA314222736GA SA314222737GA SA314222738GA SA314222739GA SA314222761GA SA314222764GA SA314222765GA SA314222771GA SA314222776GA SA314222780GA SA314222786GA SA314222789GA SA314222791GA SA314222792GA SA314222797GA SA314222803GA SA314222808GA SA314222813GA SA314222815GA SA314222816GA SA314222818GA SA314222822GA SA314222827GA SA314222832GA SA314222837GA SA314222840GA SA314222852GA SA314222853GA SA314222854GA SA314222855GA SA314222856GA SA314222857GA SA314222859GA SA314222860GA SA314222861GA SA314222862GA SA314222863GA SA314222864GA SA314222866GA SA314222867GA SA314222868GA SA314222870GA SA314222872GA SA314222873GA SA314222875GA SA314222876GA SA314222878GA SA314222880GA SA314222881GA SA314222882GA SA314222883GA SA314222884GA SA314222885GA SA314222886GA SA314222887GA SA314222888GA SA314222889GA SA314222890GA SA314222892GA SA314222893GA SA314222894GA SA314222895GA SA314222896GA SA314222897GA SA314222899GA SA314222900GA SA314222901GA SA314222902GA SA314222903GA SA314222905GA SA314222906GA SA314222907GA SA314222908GA SA314222909GA SA314222910GA SA314222911GA SA314222913GA SA314222914GA SA314222915GA SA314222916GA SA314222917GA SA314222918GA SA314222919GA SA314222920GA SA314222921GA SA314222922GA SA314222923GA SA314222924GA SA314222925GA SA314222926GA SA314222927GA SA314222928GA SA314222929GA SA314222930GA SA314222931GA SA314222932GA SA314222933GA SA314222934GA SA314222935GA SA314222936GA SA314222937GA SA314222938GA SA314222939GA SA314222940GA SA314222941GA SA314222942GA SA314222943GA SA314222944GA SA314222945GA SA314222946GA SA314222947GA SA314222948GA SA314222949GA SA314222950GA SA314222951GA SA314222952GA SA314222953GA SA314222954GA SA314222955GA SA314222957GA SA314222958GA SA314222959GA SA314222960GA SA314222961GA SA314222962GA SA314222963GA SA314222964GA SA314222965GA SA314222966GA SA314222967GA SA314222968GA SA314222969GA SA314222970GA SA314232972GA SA314232974GA SA314232976GA SA314232977GA SA314232978GA SA314232981GA SA314232982GA SA314232983GA SA314232984GA SA314232986GA SA314232987GA SA314232989GA SA314232992GA SA314232993GA SA314232994GA SA314232996GA SA314232997GA SA314232998GA SA314232999GA SA314233000GA SA314233001GA SA314233002GA SA314233003GA SA314233005GA SA314233006GA SA314233008GA SA314233009GA SA314233012GA SA314233013GA SA314233014GA SA314233015GA SA314233016GA SA314233017GA SA314233018GA SA314233019GA SA314233022GA SA314233023GA SA314233024GA SA314233025GA SA314233026GA SA314233028GA SA314233029GA SA314233030GA SA314233031GA SA314233032GA SA314233033GA SA314233034GA SA314233035GA SA314233036GA SA314233037GA SA314233038GA SA314233039GA SA314233040GA SA314233042GA SA314233043GA SA314233045GA SA314233047GA SA314233048GA SA314233049GA SA314233050GA SA314233051GA SA314233052GA SA314233053GA SA314233054GA SA314233055GA SA314233056GA SA314233058GA SA314233059GA SA314233061GA SA314233062GA SA314233063GA SA314233064GA SA314233065GA SA314233066GA SA314233067GA SA314233068GA SA314233069GA SA314233070GA SA314233071GA SA314233072GA SA314233073GA SA314233074GA SA314233075GA SA314233076GA SA314233077GA SA314233079GA SA314233080GA SA314233081GA SA314233082GA SA314233083GA SA314233084GA SA314233085GA SA314233086GA SA314233087GA SA314233088GA SA314233089GA SA314233090GA SA314233091GA SA314233092GA SA314233093GA SA314233094GA SA314233095GA SA314233096GA SA314233097GA SA314233099GA SA314233100GA SA314233101GA SA314233102GA SA314233103GA SA314233104GA SA314233105GA SA314233106GA SA314233107GA SA314233108GA SA314233109GA SA314233110GA SA314233111GA SA314233112GA SA314233113GA SA314233114GA SA314233115GA SA314233116GA SA314233117GA SA314233118GA SA314233119GA SA314233120GA SA314233121GA SA314233122GA SA314233123GA SA314233124GA SA314233125GA SA314233126GA SA314233127GA SA314233128GA SA314233129GA SA314233130GA SA314233131GA SA314233132GA SA314233133GA SA314233134GA SA314233135GA SA314233137GA SA314233138GA SA314233139GA SA314233140GA SA314233141GA SA314233142GA SA314233143GA SA314233144GA SA314233145GA SA314233146GA SA314233147GA SA314233148GA SA314233149GA SA314233150GA SA314243151GA SA314243152GA SA314243153GA SA314243154GA SA314243155GA SA314243156GA SA314243157GA SA314243158GA SA314243159GA SA314243160GA SA314243161GA SA314243162GA SA314243163GA SA314243164GA SA314243165GA SA314243166GA SA314243167GA SA314243168GA SA314243169GA SA314243170GA SA314243171GA SA314243172GA SA314243173GA SA314243174GA SA314243175GA SA314243176GA SA314243177GA SA314243178GA SA314243179GA SA314243180GA SA314243181GA SA314243182GA SA314243183GA SA314243184GA SA314243185GA SA314243186GA SA314243187GA SA314243188GA SA314243189GA SA314243190GA SA314243191GA SA314243192GA SA314243194GA SA314243195GA SA314243196GA SA314243197GA SA314243198GA SA314243199GA SA314243200GA SA314243201GA SA314243202GA SA314243203GA SA314243204GA SA314243205GA SA314243206GA SA314243207GA SA314243208GA SA314243209GA SA314243210GA SA314243213GA SA314243216GA SA314243217GA SA314243218GA SA314243221GA SA314243224GA SA314243226GA SA314243227GA SA314243228GA SA314243229GA SA314243231GA SA314243239GA SA314243240GA SA314243243GA SA314243244GA SA314243245GA SA314243246GA SA314243250GA SA314243256GA SA314243257GA SA314243258GA SA314243259GA SA314243260GA SA314243266GA SA314243267GA SA314243268GA SA314243270GA SA314243271GA SA314243272GA SA314243274GA SA314243275GA SA314243276GA SA314243277GA SA314243278GA SA314243280GA SA314243281GA SA314243282GA SA314243283GA SA314243284GA SA314243285GA SA314243287GA SA314243288GA SA314243289GA SA314243291GA SA314243292GA SA314243293GA SA314243294GA SA314243296GA SA314243297GA SA314243298GA SA314243299GA SA314243300GA SA314243301GA SA314243302GA SA314243303GA SA314243304GA SA314243305GA SA314243306GA SA314243307GA SA314243308GA SA314243309GA SA314243310GA SA314243311GA SA314243312GA SA314243313GA SA314243314GA SA314243315GA SA314243316GA SA314243318GA SA314243319GA SA314243320GA SA314243321GA SA314243322GA SA314243323GA SA314243324GA SA314243325GA SA314243326GA SA314243327GA SA314243328GA SA314243329GA SA314243330GA SA314243331GA SA314243332GA SA314243333GA SA314243334GA SA314243335GA SA314243336GA SA314243337GA SA314243338GA SA314243339GA SA314243340GA SA314243341GA SA314243342GA SA314243343GA SA314243344GA SA314243346GA SA314243347GA SA314243348GA SA314243349GA SA314243350GA SA314243351GA SA314243352GA SA314243353GA SA314243354GA SA314243355GA SA314243356GA SA314243357GA SA314243358GA SA314243359GA SA314243360GA SA314243361GA SA314243362GA SA314243363GA SA314243364GA SA314243365GA SA314243366GA SA314243367GA SA314243368GA SA314243369GA SA314243370GA SA314243371GA SA314243372GA SA314243373GA SA314243374GA SA314243375GA SA314243376GA SA314243377GA SA314243378GA SA314243379GA SA314243380GA SA314243381GA SA314243382GA SA314243383GA SA314243384GA SA314243385GA SA314243386GA SA314243387GA SA314243388GA SA314243389GA SA314243390GA SA314243391GA SA314243392GA SA314243393GA SA314243394GA SA314243395GA SA314243397GA SA314243398GA SA314243400GA SA314243401GA SA314243402GA SA314243403GA SA314243404GA SA314243405GA SA314243406GA SA314243407GA SA314243408GA SA314243409GA SA314243410GA SA314243411GA SA314243412GA SA314243413GA SA314243414GA SA314243415GA SA314243416GA SA314243417GA SA314243419GA SA314243420GA SA314243421GA SA314243422GA SA314243423GA SA314243424GA SA314243425GA SA314243426GA SA314243427GA SA314243428GA SA314243429GA SA314243430GA SA314243431GA SA314243432GA SA314243433GA SA314243434GA SA314243435GA SA314243436GA SA314243437GA SA314243438GA SA314243439GA SA314243440GA SA314243441GA SA314243442GA SA314243443GA SA314243444GA SA314243445GA SA314243446GA SA314243447GA SA314243448GA SA314243449GA SA314243450GA SA314243451GA SA314243452GA SA314243453GA SA314243454GA SA314243455GA SA314243456GA SA314243457GA SA314243458GA SA314243459GA SA314243462GA SA314243464GA SA314243466GA SA314243468GA SA314243469GA SA314243470GA SA314243471GA SA314243472GA SA314243473GA SA314243474GA SA314243475GA SA314243476GA SA314243477GA SA314243478GA SA314243479GA SA314243480GA SA314253481GA SA314253482GA SA314253483GA SA314253484GA SA314253485GA SA314253486GA SA314253487GA SA314253488GA SA314253489GA SA314253490GA SA314253491GA SA314253492GA SA314253493GA SA314253494GA SA314253495GA SA314253496GA SA314253497GA SA314253499GA SA314253500GA SA314253501GA SA314253502GA SA314253503GA SA314253504GA SA314253505GA SA314253506GA SA314253507GA SA314253508GA SA314253509GA SA314253510GA SA314253511GA SA314253512GA SA314253513GA SA314253514GA SA314253515GA SA314253516GA SA314253517GA SA314253518GA SA314253519GA SA314253520GA SA314253521GA SA314253522GA SA314253523GA SA314253524GA SA314253525GA SA314253526GA SA314253527GA SA314253528GA SA314253529GA SA314253530GA SA314253531GA SA314253532GA SA314253533GA SA314253534GA SA314253535GA SA314253536GA SA314253537GA SA314253538GA SA314253539GA SA314253540GA SA314253543GA SA314253544GA SA314253545GA SA314253546GA SA314253547GA SA314253548GA SA314253549GA SA314253550GA SA314253551GA SA314253552GA SA314253553GA SA314253555GA SA314253556GA SA314253557GA SA314253558GA SA314253559GA SA314253561GA SA314253563GA SA314253565GA SA314253566GA SA314253569GA SA314253570GA SA314253571GA SA314253572GA SA314253574GA SA314253575GA SA314253576GA SA314253577GA SA314253578GA SA314253579GA SA314253580GA SA314253581GA SA314253584GA SA314253585GA SA314253587GA SA314253588GA SA314253589GA SA314253590GA SA314253591GA SA314253592GA SA314253593GA SA314253595GA SA314253596GA SA314253597GA SA314253598GA SA314253599GA SA314253601GA SA314253602GA SA314253603GA SA314253604GA SA314253606GA SA314253607GA SA314253609GA SA314253611GA SA314253613GA SA314253618GA SA314253619GA SA314253620GA SA314253622GA SA314253623GA SA314253625GA SA314253627GA SA314253628GA SA314253630GA SA314253632GA SA314253633GA SA314253634GA SA314253635GA SA314253636GA SA314253637GA SA314253639GA SA314253640GA SA314253641GA SA314253644GA SA314253645GA SA314253646GA SA314253647GA SA314253649GA SA314253650GA SA314253652GA SA314253654GA SA314253655GA SA314253656GA SA314253657GA SA314253660GA SA314253661GA SA314253664GA SA314253665GA SA314253666GA SA314253669GA SA314253671GA SA314253673GA SA314253676GA SA314253679GA SA314253682GA SA314253683GA SA314253684GA SA314253687GA SA314253688GA SA314253689GA SA314253690GA SA314253691GA SA314253692GA SA314253693GA SA314253694GA SA314253695GA SA314253696GA SA314253697GA SA314253715GA SA314253716GA SA314253722GA SA314253724GA SA314253727GA SA314253731GA SA314253732GA SA314253734GA SA314253743GA SA314253745GA SA314253751GA SA314263753GA SA314263754GA SA314263755GA SA314263756GA SA314263757GA SA314263758GA SA314263759GA SA314263760GA SA314263762GA SA314263763GA SA314263764GA SA314263765GA SA314263766GA SA314263767GA SA314263769GA SA314263770GA SA314263771GA SA314263772GA SA314263773GA SA314263775GA SA314263776GA SA314263777GA SA314263778GA SA314263779GA SA314263780GA SA314263781GA SA314263790GA SA314263792GA SA314263809GA SA314263810GA SA314263813GA SA314263814GA SA314263815GA SA314263816GA SA314263817GA SA314263818GA SA314263819GA SA314263820GA SA314263821GA SA314263822GA SA314263823GA SA314263825GA SA314263826GA SA314263827GA SA314263828GA SA314263830GA SA314263832GA SA314273845GA SA314273846GA SA314273847GA SA314273848GA SA314273849GA SA314273850GA SA314273851GA SA314273852GA SA314273853GA SA314273854GA SA314273855GA SA314273856GA SA314273857GA SA314273858GA SA314273859GA SA314273860GA SA314273861GA SA314273862GA SA314273863GA SA314273864GA SA314273865GA SA314273867GA SA314273869GA SA314273870GA SA314273871GA SA314273872GA SA314273874GA SA314273875GA SA314273876GA SA314273877GA SA314273878GA SA314273879GA SA314273880GA SA314273881GA SA314273883GA SA314273884GA SA314273885GA SA314273886GA SA314273887GA SA314273888GA SA314273889GA SA314273890GA SA314273891GA SA314273892GA SA314273893GA SA314273894GA SA314273895GA SA314273897GA SA314273898GA SA314273899GA SA314273900GA SA314273901GA SA314273907GA SA314273909GA SA314273910GA SA314273912GA SA314273915GA SA314273923GA SA314273931GA SA314273938GA SA314273941GA SA314273943GA SA314273945GA SA314273949GA SA314273950GA SA314273952GA SA314273953GA SA314273958GA SA314273959GA SA314273960GA SA314273961GA SA314273965GA SA314273969GA SA314273970GA SA314273974GA SA314273977GA SA314273981GA SA314273983GA SA314273984GA SA314273985GA SA314273986GA SA314273987GA SA314273990GA SA314273996GA SA314273998GA SA314273999GA SA314274000GA SA314274001GA SA314274002GA SA314274003GA SA314274006GA SA314274009GA SA314274010GA SA314274011GA SA314274012GA SA314274013GA SA314274014GA SA314274015GA SA314274016GA SA314274017GA SA314274021GA SA314274026GA SA314274027GA SA314274028GA SA314274029GA SA314274030GA SA314274031GA SA314274032GA SA314274033GA SA314274034GA SA314274035GA SA314274036GA SA314274038GA SA314274039GA SA314274040GA SA314274041GA SA314274042GA SA314274044GA SA314274045GA SA314274046GA SA314274049GA SA314274051GA SA314274052GA SA314274055GA SA314274059GA SA314274061GA SA314274063GA SA314274067GA SA314274069GA SA314274070GA SA314274071GA SA314274072GA SA314274084GA SA314274087GA SA314274088GA SA314274090GA SA314274093GA SA314274095GA SA314274098GA SA314274100GA SA314274102GA SA314274114GA SA314274115GA SA314274116GA SA314274117GA SA314274119GA SA314274120GA SA314274123GA SA314274124GA SA314274126GA SA314274129GA SA314274132GA SA314274134GA SA314274135GA SA314274138GA SA314274139GA SA314274140GA SA314274142GA SA314274143GA SA314274145GA SA314274149GA SA314274152GA SA314274156GA SA314274166GA SA314274168GA SA314274170GA SA314274171GA SA314274173GA SA314274174GA SA314274177GA SA314274179GA SA314274182GA SA314274183GA SA314274184GA SA314274185GA SA314274191GA SA314274196GA SA314274197GA SA314274198GA SA314274199GA SA314274201GA SA314274202GA SA314274203GA SA314274204GA SA314274206GA SA314274207GA SA314274210GA SA314274212GA SA314274213GA SA314274215GA SA314274217GA SA314274218GA SA314274220GA SA314274222GA SA314274223GA SA314274225GA SA314274226GA SA314274229GA SA314274230GA SA314274231GA SA314274232GA SA314274235GA SA314274236GA SA314274237GA SA314274238GA SA314274239GA SA314274240GA SA314274242GA SA314274243GA SA314274244GA SA314274245GA SA314274246GA SA314274248GA SA314274249GA SA314274250GA SA314274251GA SA314274253GA SA314274254GA SA314274255GA SA314274256GA SA314274257GA SA314274258GA SA314274259GA SA314274260GA SA314274262GA SA314274263GA SA314274264GA SA314274265GA SA314274266GA SA314274267GA SA314274269GA SA314274270GA SA314274271GA SA314274272GA SA314274274GA SA314274275GA SA314274276GA SA314274277GA SA314274278GA SA314274279GA SA314274280GA SA314274282GA SA314274283GA SA314274284GA SA314274285GA SA314274286GA SA314274289GA SA314274290GA SA314274291GA SA314274292GA SA314274293GA SA314274294GA SA314274295GA SA314274296GA SA314274297GA SA314274299GA SA314274300GA SA314274301GA SA314274304GA SA314274305GA SA314274306GA SA314274307GA SA314274309GA SA314274310GA SA314274311GA SA314274312GA SA314274313GA SA314274314GA SA314274316GA SA314274317GA SA314274318GA SA314274319GA SA314274320GA SA314274321GA SA314274323GA SA314274324GA SA314274325GA SA314274326GA SA314274327GA SA314274329GA SA314274338GA SA314274339GA SA314274342GA SA314284346GA SA314284347GA SA314284348GA SA314284349GA SA314284350GA SA314284351GA SA314284352GA SA314284355GA SA314284356GA SA314284359GA SA314284360GA SA314284362GA SA314284363GA SA314284366GA SA314284367GA SA314284368GA SA314284369GA SA314284371GA SA314284372GA SA314284373GA SA314284374GA SA314284376GA SA314284379GA SA314284380GA SA314284381GA SA314284382GA SA314284383GA SA314284385GA SA314284386GA SA314284387GA SA314284388GA SA314284389GA SA314284390GA SA314284391GA SA314284392GA SA314284393GA SA314284394GA SA314284395GA SA314284396GA SA314284397GA SA314284398GA SA314284399GA SA314284400GA SA314284401GA SA314284403GA SA314284404GA SA314284405GA SA314284406GA SA314284408GA SA314284409GA SA314284411GA SA314284413GA SA314284414GA SA314284417GA SA314284418GA SA314284419GA SA314284420GA SA314284421GA SA314284422GA SA314284423GA SA314284424GA SA314284425GA SA314284426GA SA314284427GA SA314284428GA SA314284430GA SA314284431GA SA314284432GA SA314284434GA SA314284435GA SA314284436GA SA314284437GA SA314284438GA SA314284439GA SA314284440GA SA314284441GA SA314284442GA SA314284443GA SA314284445GA SA314284446GA SA314284447GA SA314284448GA SA314284449GA SA314284450GA SA314284452GA SA314284453GA SA314284455GA SA314284456GA SA314284457GA SA314284458GA SA314284459GA SA314284460GA SA314284462GA SA314284465GA SA314284466GA SA314284467GA SA314284468GA SA314284469GA SA314284470GA SA314284471GA SA314284472GA SA314284473GA SA314284474GA SA314284475GA SA314284476GA SA314284477GA SA314284478GA SA314284480GA SA314284482GA SA314284483GA SA314284484GA SA314284485GA SA314284486GA SA314284487GA SA314284489GA SA314284491GA SA314284493GA SA314284494GA SA314284495GA SA314284496GA SA314284497GA SA314284498GA SA314284499GA SA314284500GA SA314284501GA SA314284502GA SA314284503GA SA314284505GA SA314284506GA SA314284510GA SA314284511GA SA314284512GA SA314284513GA SA314284514GA SA314284517GA SA314284519GA SA314284520GA SA314284522GA SA314284523GA SA314284524GA SA314284526GA SA314284527GA SA314284528GA SA314284530GA SA314284531GA SA314284533GA SA314284534GA SA314284535GA SA314284536GA SA314284538GA SA314284539GA SA314284540GA SA314284541GA SA314284543GA SA314284544GA SA314284545GA SA314284546GA SA314284547GA SA314284549GA SA314284550GA SA314284551GA SA314284553GA SA314284554GA SA314284555GA SA314284556GA SA314284557GA SA314284558GA SA314284560GA SA314284561GA SA314284562GA SA314284564GA SA314284565GA SA314284566GA SA314284567GA SA314284568GA SA314284569GA SA314284571GA SA314284573GA SA314284574GA SA314284575GA SA314284576GA SA314284577GA SA314284578GA SA314284579GA SA314284580GA SA314284581GA SA314284582GA SA314284583GA SA314284584GA SA314284585GA SA314284586GA SA314284587GA SA314284589GA SA314284594GA SA314284596GA SA314284597GA SA314284598GA SA314284599GA SA314284600GA SA314284601GA SA314284602GA SA314284603GA SA314284604GA SA314284605GA SA314284606GA SA314284609GA SA314284610GA SA314284611GA SA314284612GA SA314284613GA SA314284614GA SA314284615GA SA314284616GA SA314284619GA SA314284620GA SA314284622GA SA314284623GA SA314284624GA SA314284625GA SA314284626GA SA314284627GA SA314284628GA SA314284629GA SA314284630GA SA314284631GA SA314284632GA SA314284633GA SA314284634GA SA314284635GA SA314284636GA SA314284638GA SA314284639GA SA314284641GA SA314284642GA SA314284643GA SA314284646GA SA314284648GA SA314284649GA SA314284650GA SA314284651GA SA314284652GA SA314284653GA SA314284654GA SA314284655GA SA314284656GA SA314284657GA SA314284658GA SA314284659GA SA314284660GA SA314284661GA SA314284662GA SA314284663GA SA314284664GA SA314284665GA SA314284666GA SA314284668GA SA314284669GA SA314284671GA SA314284672GA SA314284673GA SA314284674GA SA314284675GA SA314284676GA SA314284677GA SA314284678GA SA314284680GA SA314284681GA SA314284682GA SA314284684GA SA314284685GA SA314284686GA SA314284687GA SA314284688GA SA314284689GA SA314284690GA SA314284691GA SA314284692GA SA314284693GA SA314284695GA SA314284696GA SA314284697GA SA314284698GA SA314284699GA SA314284700GA SA314284701GA SA314284702GA SA314284703GA SA314284704GA SA314284705GA SA314284706GA SA314284708GA SA314284710GA SA314284711GA SA314284712GA SA314284713GA SA314284714GA SA314284715GA SA314284716GA SA314284717GA SA314284718GA SA314284719GA SA314284720GA SA314284722GA SA314284724GA SA314284725GA SA314284726GA SA314284727GA SA314284728GA SA314284730GA SA314284732GA SA314284733GA SA314284734GA SA314284737GA SA314284739GA SA314284740GA SA314284742GA SA314284744GA SA314284747GA SA314284749GA SA314284750GA SA314284751GA SA314284752GA SA314284753GA SA314284754GA SA314284755GA SA314284757GA SA314284760GA SA314284762GA SA314284763GA SA314284764GA SA314284765GA SA314284766GA SA314284767GA SA314284768GA SA314284769GA SA314294770GA SA314294771GA SA314294772GA SA314294773GA SA314294774GA SA314294775GA SA314294777GA SA314294778GA SA314294779GA SA314294780GA SA314294781GA SA314294782GA SA314294783GA SA314294784GA SA314294785GA SA314294786GA SA314294787GA SA314294788GA SA314294789GA SA314294790GA SA314294791GA SA314294792GA SA314294793GA SA314294795GA SA314294797GA SA314294798GA SA314294799GA SA314294800GA SA314294801GA SA314294802GA SA314294803GA SA314294804GA SA314294805GA SA314294806GA SA314294807GA SA314294808GA SA314294809GA SA314294810GA SA314294811GA SA314294812GA SA314294813GA SA314294814GA SA314294815GA SA314294816GA SA314294817GA SA314294818GA SA314294820GA SA314294821GA SA314294822GA SA314294823GA SA314294824GA SA314294825GA SA314294826GA SA314294828GA SA314294829GA SA314294831GA SA314294834GA SA314294835GA SA314294836GA SA314294838GA SA314294839GA SA314294841GA SA314294842GA SA314294843GA SA314294844GA SA314294845GA SA314294846GA SA314294847GA SA314294848GA SA314294849GA SA314294850GA SA314294851GA SA314294853GA SA314294855GA SA314294856GA SA314294858GA SA314294873GA SA314294880GA SA314294888GA SA314294890GA SA314294892GA SA314294893GA SA314294894GA SA314294895GA SA314294896GA SA314294897GA SA314294898GA SA314294900GA SA314294901GA SA314294902GA SA314294905GA SA314294906GA SA314294907GA SA314294908GA SA314294909GA SA314294911GA SA314294912GA SA314294914GA SA314294915GA SA314294917GA SA314294918GA SA314294920GA SA314294921GA SA314294922GA SA314294923GA SA314294924GA SA314294925GA SA314294926GA SA314294927GA SA314294928GA SA314294929GA SA314294930GA SA314294931GA SA314294932GA SA314294938GA SA314294939GA SA314294940GA SA314294941GA SA314294942GA SA314294945GA SA314294946GA SA314294947GA SA314294949GA SA314294954GA SA314294957GA SA314294958GA SA314294961GA SA314294965GA SA314294970GA SA314294975GA SA314294977GA SA314294981GA SA314294983GA SA314294992GA SA314294995GA SA314295001GA SA314295002GA SA314295006GA SA314295010GA SA314295017GA SA314295018GA SA314295025GA SA314295038GA SA314295041GA SA314295046GA SA314295057GA SA314295060GA SA314295061GA SA314295062GA SA314295063GA SA314295064GA SA314295065GA SA314295067GA SA314295068GA SA314295072GA SA314295080GA SA314295082GA SA314295084GA SA314295086GA SA314295087GA SA314295088GA SA314295089GA SA314295090GA SA314295091GA SA314295093GA SA314295098GA SA314295100GA SA314295101GA SA314295104GA SA314295105GA SA314295107GA SA314295112GA SA314295115GA SA314295116GA SA314295120GA SA314295123GA SA314295129GA SA314295147GA SA314295178GA SA314295182GA SA314295200GA SA314295214GA SA314295219GA SA314295223GA SA314295224GA SA314295242GA SA314295243GA SA314295259GA SA314295265GA SA314295268GA SA314295276GA SA314295277GA SA314295279GA SG510420048GA SG510420049GA SG510420051GA SG510420053GA SA313462201GA SA312522514GA SA312522522GA SA312522540GA SA312522614GA SA312522642GA SA312522643GA SA312522650GA SA312522651GA SA307020377GA SA307030403GA SA307060433GA SA307070451GA SA307080474GA SA307080541GA SA307110543GA SA307110547GA SA307110550GA SA307110558GA SA307110562GA SA307110563GA SA307110564GA SA307110565GA SA307110566GA SA307110567GA SA307110576GA SA307110580GA SA307110581GA SA307200778GA SA307200781GA SA307200783GA SA307200786GA SA307200787GA SA307210806GA SA307210810GA SA307210812GA SA307210813GA SA307220876GA SA307220894GA SA307220896GA SA307220907GA SA307220910GA SA307220911GA SA307220913GA SA307220915GA SA307220916GA SA307220917GA SA307220921GA SA307220925GA SA307220928GA SA307220933GA SA307220967GA SA307220968GA SA307220969GA SA307220975GA SA307220985GA SA307220989GA SA307221017GA SA307221020GA SA307221021GA SA307221023GA SA307221024GA SA307221031GA SA307221085GA SA307221086GA SA307221087GA SA307221088GA SA307221094GA SA307231487GA SA307251792GA SA307372443GA SA307382603GA SA307382605GA SA307382619GA SA307443238GA SA307443242GA SA307443247GA SA307443249GA SA307453389GA SA307453394GA SA307453396GA SA307453403GA SA307453406GA SA307453408GA SA307453410GA SA307453412GA SA307453553GA SA307453554GA SA307453576GA SA307453577GA SA307453591GA SA307453605GA SA307453607GA SA307483665GA SA307483708GA SA307483718GA SA307483779GA SA307483791GA SA307483793GA SA307483807GA SA307483815GA SA307483820GA SA307483833GA SA307483839GA SA307493862GA SA307493993GA SA307493998GA SA307504189GA SA307504191GA SA307504202GA SA307504204GA SA307504211GA SA307514323GA SA307514335GA SA307514350GA SA307514353GA SA307514399GA SA307514400GA SA307514402GA SA307514411GA SA307514416GA SA307514419GA SA307514420GA SA307514421GA SA307514422GA SA307514425GA SA307514426GA SA307514427GA SA307514428GA SA307514430GA SA307514433GA SA307514438GA SA307514456GA SA308044511GA SA308074586GA SA308074602GA SA308084618GA SA308084624GA SA308084635GA SA308084636GA SA308084637GA SA308084640GA SA308084655GA SA308084659GA SA308084660GA SA308084739GA SA308084744GA SA308084747GA SA308084752GA SA308084755GA SA308084775GA SA308094870GA SA308094873GA SA308094908GA SA308094911GA SA308094933GA SA308094937GA SA308094947GA SA308094953GA SA308094980GA SA308094981GA SA308094998GA SA308095024GA SA308095047GA SA308095051GA SA308095056GA SA308095059GA SA308095061GA SA308095064GA SA308095065GA SA308095066GA SA308095067GA SA308095068GA SA308115277GA SA308115278GA SA308115281GA SA308115293GA SA308125324GA SA308125332GA SA308125333GA SA308125338GA SA308125339GA SA308125340GA SA308125342GA SA308125343GA SA308125352GA SA308125365GA SA308125366GA SA308125371GA SA308135580GA SA308135585GA SA308145680GA SA308145711GA SA308145715GA SA308145769GA SA308145799GA SA308145800GA SA308145809GA SA308155881GA SA308155905GA SA308155943GA SA308155951GA SA308155958GA SA308155963GA SA308165995GA SA308166078GA SA308166125GA SA308186349GA SA308186355GA SA308186358GA SA308186361GA SA308186363GA SA308186367GA SA308186384GA SA308186399GA SA308186409GA SA308186412GA SA308186413GA SA308186414GA SA308186415GA SA308186420GA SA308186442GA SA308196465GA SA308196469GA SA308196470GA SA308196541GA SA308266827GA SA308266852GA SA308286935GA SA308286944GA SA308286976GA SA308297025GA SA308297027GA SA308297030GA SA308297036GA SA308297039GA SA308297043GA SA308297049GA SA308337666GA SA308347733GA SA308347734GA SA308347736GA SA308347745GA SA308347753GA SA308347792GA SA308347801GA SA308347805GA SA308347809GA SA308347816GA SA308347830GA SA308347836GA SA308357861GA SA308357863GA SA308357867GA SA308357869GA SA308357870GA SA308357872GA SA308357873GA SA308357880GA SA308357884GA SA308357890GA SA308357893GA SA308357900GA SA308357901GA SA308357902GA SA308357903GA SA308357904GA SA308357905GA SA308357907GA SA308357908GA SA308357912GA SA308357915GA SA308357924GA SA308357929GA SA308357931GA SA308357932GA SA308368195GA SA308368234GA SA308368239GA SA308368240GA SA308368242GA SA308368246GA SA308368247GA SA308368249GA SA308378276GA SA308388569GA SA308408913GA SA308408929GA SA308408935GA SA308408945GA SA308408964GA SA308409138GA SA308409162GA SA308409196GA SA308439443GA SA308439451GA SA308439467GA SA308449522GA SA308459667GA SA308459668GA SA308459683GA SA308459685GA SA308459686GA SA308459687GA SA308459688GA SA308459689GA SA308459716GA SA308459764GA SA308459775GA SA308459778GA SA308459842GA SA308459848GA SA308459851GA SA308459852GA SA308459853GA SA308459855GA SA308459858GA SA308459867GA SA308459870GA SA308459872GA SA308459873GA SA308459875GA SA308469878GA SA308469880GA SA308469881GA SA308469882GA SA308469883GA SA308479897GA SA308479902GA SA308479905GA SA308479907GA SA308479909GA SA308481009GA SA308481010GA SA308481011GA SA308481012GA SA308481015GA SA308481020GA SA308481026GA SA308481028GA SA308481029GA SA308481031GA SA308481032GA SA308481034GA SA308481043GA SA308481050GA SA308489910GA SA308489913GA SA308489914GA SA308489915GA SA308489916GA SA308489923GA SA308489937GA SA308489943GA SA308489950GA SA308489963GA SA308489993GA SA308489995GA SA308489997GA SA308489998GA SA308501410GA SA308501417GA SA308501426GA SA309011847GA SA309011874GA SA309011893GA SA309021957GA SA309031994GA SA309032011GA SA309042130GA SA309042134GA SA309042135GA SA309042136GA SA309042137GA SA309042138GA SA309042139GA SA309042140GA SA309042141GA SA309042142GA SA309042143GA SA309042146GA SA309042147GA SA309042148GA SA309042181GA SA309042182GA SA309042191GA SA309042192GA SA309042193GA SA309042197GA SA309042198GA SA309042200GA SA309042202GA SA309042203GA SA309042205GA SA309042206GA SA309042210GA SA309042211GA SA309052217GA SA309052218GA SA309052221GA SA309052226GA SA309052237GA SA309052238GA SA309052240GA SA309052241GA SA309052242GA SA309052243GA SA309052244GA SA309052245GA SA309052246GA SA309052247GA SA309052248GA SA309052249GA SA309052250GA SA309052251GA SA309052252GA SA309052253GA SA309052254GA SA309052255GA SA309052256GA SA309052257GA SA309052258GA SA309052259GA SA309052260GA SA309052261GA SA309052262GA SA309052263GA SA309052264GA SA309052265GA SA309062275GA SA309062295GA SA309062297GA SA309062317GA SA309062319GA SA309062320GA SA309062321GA SA309062322GA SA309062323GA SA309062324GA SA309062327GA SA309062329GA SA309062332GA SA309062333GA SA309062334GA SA309062335GA SA309062337GA SA309062338GA SA309062339GA SA309062340GA SA309062341GA SA309072342GA SA309072346GA SA309072349GA SA309072359GA SA309072360GA SA309072378GA SA309072399GA SA309072410GA SA309082413GA SA309082417GA SA309082418GA SA309082419GA SA309082420GA SA309082421GA SA309082422GA SA309082424GA SA309082425GA SA309082426GA SA309082427GA SA309082428GA SA309082430GA SA309082431GA SA309082432GA SA309082433GA SA309082434GA SA309082435GA SA309082436GA SA309082437GA SA309082438GA SA309082439GA SA309082440GA SA309082441GA SA309082442GA SA309082453GA SA309082459GA SA309082468GA SA309092469GA SA309092470GA SA309092471GA SA309092472GA SA309092495GA SA309092497GA SA309092498GA SA309092499GA SA309092501GA SA309122675GA SA309122679GA SA309122698GA SA309173161GA SA309183237GA SA309193284GA SA309203383GA SA309203387GA SA309203395GA SA309203401GA SA309294124GA SA309324328GA SA309334791GA SA309446481GA SA309446500GA SA309446510GA SA309456634GA SA309456655GA SA309456657GA SA309466753GA SA309476865GA SA309476917GA SA309476969GA SA309476982GA SA309476999GA SA309477021GA SA309477024GA SA309477108GA SA309497155GA SA309497223GA SA309497224GA SA309497225GA SA309497227GA SA309497228GA SA309497229GA SA309497231GA SA309497233GA SA309497234GA SA309497268GA SA309497269GA SA309497270GA SA309497271GA SA309497272GA SA309497273GA SA309497275GA SA309497276GA SA309497277GA SA309497279GA SA309497283GA SA309497284GA SA309497285GA SA309497286GA SA309497288GA SA309497301GA SA309507354GA SA309507365GA SA309507373GA SA309507379GA SA309507380GA SA309507381GA SA309507388GA SA309507390GA SA309507393GA SA309507395GA SA309507397GA SA309507402GA SA309507406GA SA309507410GA SA309507418GA SA309507422GA SA309507423GA SA309507428GA SA309507431GA SA309507438GA SA309507446GA SA309517453GA SA309517459GA SA309517475GA SA309517479GA SA309517802GA SA309517812GA SA309517850GA SA309517872GA SA309517876GA SA309517879GA SA309517888GA SA309527908GA SA309527928GA SA309527951GA SA309527959GA SA309527968GA SA309528001GA SA309538030GA SA309538091GA SA309538107GA SA310028174GA SA310028185GA SA310028206GA SA310028225GA SA310028258GA SA310028289GA SA310028292GA SA310028327GA SA310028332GA SA310028340GA SA310038357GA SA310038358GA SA310038371GA SA310038373GA SA310038375GA SA310038385GA SA310038387GA SA310038396GA SA310038399GA SA310038402GA SA310038403GA SA310038407GA SA310038411GA SA310038412GA SA310038418GA SA310038422GA SA310038428GA SA310038437GA SA310038442GA SA310038451GA SA310038452GA SA310038456GA SA310038458GA SA310038461GA SA310038463GA SA310038469GA SA310038478GA SA310038479GA SA310038486GA SA310038494GA SA310038496GA SA310038501GA SA310038504GA SA310038517GA SA310038519GA SA310038521GA SA310038526GA SA310038533GA SA310038535GA SA310038538GA SA310048544GA SA310048548GA SA310048550GA SA310048554GA SA310048555GA SA310048556GA SA310048566GA SA310048568GA SA310048571GA SA310048576GA SA310048583GA SA310048587GA SA310048591GA SA310048599GA SA310048600GA SA310048601GA SA310048602GA SA310048603GA SA310048605GA SA310048606GA SA310048608GA SA310048609GA SA310048610GA SA310048611GA SA310048612GA SA310048615GA SA310048617GA SA310048618GA SA310048625GA SA310048627GA SA310048628GA SA310048629GA SA310048648GA SA310048651GA SA310048656GA SA310048675GA SA310058693GA SA310058746GA SA310068813GA SA310068834GA SA310068842GA SA310068877GA SA310068924GA SA310119015GA SA310139067GA SA310139181GA SA310139246GA SA310139261GA SA310149314GA SA310149327GA SA310149353GA SA310149358GA SA310159362GA SA310159363GA SA310159369GA SA310159371GA SA310159456GA SA310169511GA SA310219824GA SA310259943GA SA310259946GA SA310259948GA SA310259949GA SA310259959GA SA310260290GA SA310260292GA SA310260293GA SA310260294GA SA310260295GA SA310260296GA SA310260298GA SA310260299GA SA310260300GA SA310260301GA SA310260302GA SA310260304GA SA310260306GA SA310260315GA SA310260316GA SA310260318GA SA310260319GA SA310260321GA SA310260388GA SA310290624GA SA310290626GA SA310290627GA SA310290630GA SA310290635GA SA310290643GA SA310290654GA SA310290665GA SA310290702GA SA310290703GA SA310331123GA SA310331328GA SA310331333GA SA310331336GA SA310331338GA SA310331339GA SA310331344GA SA310331346GA SA310331353GA SA310341359GA SA310341361GA SA310341375GA SA310341384GA SA310371994GA SA310372068GA SA310372138GA SA310372140GA SA310372161GA SA310372180GA SA310372194GA SA310372288GA SA310372308GA SA310372309GA SA310382623GA SA310382625GA SA310382629GA SA310382641GA SA310382705GA SA310382730GA SA310382731GA SA310382735GA SA310382738GA SA310382753GA SA310382758GA SA310382760GA SA310382767GA SA310382770GA SA310393141GA SA310413517GA SA310423646GA SA310423654GA SA310423805GA SA310424026GA SA310434052GA SA310454933GA SA310475539GA SA310475573GA SA310475583GA SA310475615GA SA310495891GA SA310495892GA SA310495893GA SA310495894GA SA310495896GA SA310495897GA SA310495899GA SA310495900GA SA310495901GA SA310495903GA SA310495904GA SA310495906GA SA310495908GA SA310495909GA SA310495911GA SA310495912GA SA310495918GA SA310495932GA SA310495933GA SA310495934GA SA310495941GA SA310495942GA SA310495943GA SA310495950GA SA310495954GA SA310495956GA SA310495957GA SA310495959GA SA310495973GA SA310496128GA SA310496131GA SA310496144GA SA310496145GA SA310496146GA SA310496152GA SA310496209GA SA310496216GA SA310496280GA SA310496286GA SA310496293GA SA310496297GA SA310496301GA SA310496305GA SA310496307GA SA310496308GA SA310496311GA SA310496313GA SA310496316GA SA310496343GA SA310496356GA SA310496380GA SA310496381GA SA310496382GA SA310496391GA SA310496392GA SA310496397GA SA310496398GA SA310496399GA SA310496406GA SA310496408GA SA310496412GA SA310496414GA SA310496416GA SA310496417GA SA310496419GA SA310496422GA SA310496423GA SA310496427GA SA310496431GA SA310496432GA SA310496433GA SA310496446GA SA310496448GA SA310496452GA SA310496477GA SA310496506GA SA310496510GA SA310496530GA SA310496632GA SA310506670GA SA310506689GA SA310506697GA SA310506722GA SA310506738GA SA310506766GA SA310506800GA SA310506873GA SA310507102GA SA310507118GA SA310507313GA SA310507324GA SA310507325GA SA310507353GA SA310507376GA SA310507524GA SA310507538GA SA310507549GA SA310517604GA SA310517618GA SA310517621GA SA310517625GA SA310517626GA SA310517629GA SA310517632GA SA310517637GA SA310517673GA SA310517676GA SA310517681GA SA310517682GA SA310517844GA SA310517870GA SA310517871GA SA310518000GA SA310518146GA SA310518177GA SA310518182GA SA310518184GA SA310518186GA SA310518191GA SA310518192GA SA310518193GA SA310518195GA SA310518197GA SA310518202GA SA310518219GA SA310518234GA SA310518306GA SA310518311GA SA310518486GA SA310518629GA SA310528791GA SA310528867GA SA310528872GA SA310528884GA SA311039378GA SA311039380GA SA311039384GA SA311039532GA SA311039533GA SA311039544GA SA311060021GA SA311060101GA SA311060105GA SA311060114GA SA311060168GA SA311060213GA SA311060224GA SA311060229GA SA311060233GA SA311060252GA SA311060260GA SA311060281GA SA311060305GA SA311060326GA SA311060331GA SA311060332GA SA311060335GA SA311060338GA SA311060345GA SA311060356GA SA311060365GA SA311060369GA SA311060371GA SA311060372GA SA311060375GA SA311060379GA SA311060380GA SA311060383GA SA311060398GA SA311060399GA SA311069661GA SA311069724GA SA311069738GA SA311069790GA SA311069797GA SA311069800GA SA311069802GA SA311069805GA SA311069806GA SA311069808GA SA311069810GA SA311069829GA SA311069831GA SA311069832GA SA311069848GA SA311070525GA SA311070553GA SA311070556GA SA311070557GA SA311070571GA SA311070573GA SA311070583GA SA311070584GA SA311070590GA SA311080598GA SA311080607GA SA311080609GA SA311080616GA SA311080629GA SA311080647GA SA311080648GA SA311080651GA SA311080660GA SA311080713GA SA311080715GA SA311080727GA SA311080736GA SA311080737GA SA311080745GA SA311090789GA SA311090799GA SA311090803GA SA311090805GA SA311090831GA SA311090843GA SA311090845GA SA311090846GA SA311090855GA SA311090869GA SA311090872GA SA311090876GA SA311090886GA SA311101084GA SA311101127GA SA311101129GA SA311101161GA SA311101176GA SA311101183GA SA311101185GA SA311101188GA SA311101196GA SA311101205GA SA311101207GA SA311101209GA SA311101211GA SA311101215GA SA311101223GA SA311101255GA SA311101258GA SA311101259GA SA311101270GA SA311101271GA SA311101288GA SA311101305GA SA311101385GA SA311101395GA SA311101397GA SA311101422GA SA311101440GA SA311101445GA SA311101449GA SA311101455GA SA311101553GA SA311111565GA SA311121699GA SA311121705GA SA311121709GA SA311121711GA SA311121720GA SA311121726GA SA311121861GA SA311121869GA SA311121872GA SA311121889GA SA311121895GA SA311121911GA SA311122043GA SA311122045GA SA311122059GA SA311122062GA SA311122064GA SA311122099GA SA311122126GA SA311122140GA SA311122145GA SA311122148GA SA311122154GA SA311122160GA SA311122180GA SA311122195GA SA311122219GA SA311122222GA SA311132366GA SA311132566GA SA311142908GA SA311142915GA SA311142916GA SA311142919GA SA311142930GA SA311142932GA SA311142933GA SA311142955GA SA311142980GA SA311153083GA SA311153097GA SA311153098GA SA311153099GA SA311153102GA SA311153105GA SA311153106GA SA311153113GA SA311153116GA SA311153125GA SA311153126GA SA311153127GA SA311153134GA SA311153135GA SA311153143GA SA311153145GA SA311153148GA SA311153158GA SA311153175GA SA311153205GA SA311153211GA SA311153212GA SA311153227GA SA311163975GA SA311174185GA SA311184386GA SA311184530GA SA311184553GA SA311184576GA SA311184604GA SA311184641GA SA311184646GA SA311184648GA SA311184649GA SA311184650GA SA311184651GA SA311184660GA SA311184661GA SA311195095GA SA311195138GA SA311195146GA SA311195152GA SA311195153GA SA311195154GA SA311195155GA SA311195158GA SA311195159GA SA311195167GA SA311195249GA SA311195251GA SA311195252GA SA311195257GA SA311195265GA SA311205340GA SA311205342GA SA311205344GA SA311205346GA SA311205347GA SA311205349GA SA311205351GA SA311205352GA SA311205354GA SA311205361GA SA311205365GA SA311205366GA SA311205368GA SA311205374GA SA311205375GA SA311205377GA SA311205380GA SA311205382GA SA311205383GA SA311205384GA SA311205385GA SA311205387GA SA311205388GA SA311205389GA SA311205442GA SA311205444GA SA311205462GA SA311215846GA SA311225898GA SA311225914GA SA311225918GA SA311225922GA SA311225925GA SA311225932GA SA311225935GA SA311225939GA SA311225940GA SA311225941GA SA311225943GA SA311225946GA SA311225956GA SA311225958GA SA311225962GA SA311225964GA SA311225967GA SA311225968GA SA311225973GA SA311225975GA SA311225996GA SA311225998GA SA311236005GA SA311236012GA SA311236017GA SA311236049GA SA311236055GA SA311236058GA SA311236062GA SA311236108GA SA311236114GA SA311236117GA SA311236120GA SA311236190GA SA311236250GA SA311236252GA SA311236254GA SA311236267GA SA311236283GA SA311236284GA SA311236350GA SA311236362GA SA311236365GA SA311236374GA SA311236379GA SA311246410GA SA311246479GA SA311246544GA SA311256689GA SA311256707GA SA311256711GA SA311256713GA SA311256781GA SA311256793GA SA311256871GA SA311266873GA SA311266876GA SA311266892GA SA311266897GA SA311266900GA SA311266910GA SA311266951GA SA311266993GA SA311266995GA SA311266996GA SA311266998GA SA311266999GA SA311267001GA SA311267002GA SA311267005GA SA311287110GA SA311287127GA SA311287164GA SA311297282GA SA311297290GA SA311307447GA SA311307505GA SA311317649GA SA311317725GA SA311327833GA SA311327848GA SA311327852GA SA311428018GA SA311428043GA SA311508641GA SA311508642GA SA311508644GA SA311508649GA SA311508651GA SA311508666GA SA311508669GA SA311508671GA SA312018745GA SA312018752GA SA312018753GA SA312018756GA SA312018757GA SA312018763GA SA312018764GA SA312018765GA SA312018767GA SA312018773GA SA312018774GA SA312018775GA SA312018776GA SA312018816GA SA312018825GA SA312018833GA SA312018838GA SA312018842GA SA312018849GA SA312018882GA SA312018891GA SA312018892GA SA312018900GA SA312018903GA SA312018907GA SA312028999GA SA312029008GA SA312029018GA SA312029019GA SA312029051GA SA312029071GA SA312039214GA SA312039216GA SA312039217GA SA312039218GA SA312039219GA SA312039222GA SA312039226GA SA312039227GA SA312039229GA SA312069701GA SA312069718GA SA312069777GA SA312070103GA SA312070104GA SA312070122GA SA312070125GA SA312070341GA SA312070348GA SA312070360GA SA312080391GA SA312080393GA SA312080394GA SA312080400GA SA312080408GA SA312080411GA SA312080421GA SA312080435GA SA312080492GA SA312080495GA SA312080499GA SA312080501GA SA312080502GA SA312080503GA SA312080504GA SA312080507GA SA312080511GA SA312080583GA SA312080589GA SA312080603GA SA312080618GA SA312080627GA SA312080628GA SA312080636GA SA312080641GA SA312080680GA SA312080709GA SA312080712GA SA312080718GA SA312080726GA SA312080818GA SA312080821GA SA312080846GA SA312080855GA SA312091267GA SA312091273GA SA312091292GA SA312111511GA SA312111532GA SA312111653GA SA312121756GA SA312121765GA SA312121801GA SA312121807GA SA312121832GA SA312121846GA SA312121858GA SA312121868GA SA312121896GA SA312121996GA SA312122010GA SA312122012GA SA312122020GA SA312122022GA SA312122027GA SA312132117GA SA312132133GA SA312152345GA SA312152485GA SA312253308GA SA312253311GA SA312253313GA SA312253367GA SA312253368GA SA312263939GA SA312284151GA SA312284169GA SA312284172GA SA312284173GA SA312284176GA SA312284184GA SA312284282GA SA312284296GA SA312284306GA SA312294498GA SA312294628GA SA312344921GA SA312355236GA SA312355475GA SA312355483GA SA312355505GA SA312355511GA SA312355523GA SA312355525GA SA312365569GA SA312365570GA SA312365591GA SA312365592GA SA312365593GA SA312365594GA SA312365596GA SA312365597GA SA312365599GA SA312365600GA SA312365601GA SA312365603GA SA312365604GA SA312365606GA SA312365607GA SA312365608GA SA312365609GA SA312365610GA SA312365611GA SA312365612GA SA312365614GA SA312365615GA SA312365616GA SA312365619GA SA312365620GA SA312365627GA SA312365638GA SA312365641GA SA312365691GA SA312375992GA SA312386374GA SA312386391GA SA312386402GA SA312386404GA SA312386409GA SA312386422GA SA312386424GA SA312386427GA SA312386430GA SA312386462GA SA312386467GA SA312447430GA SA312447453GA SA312447466GA SA312447470GA SA312447472GA SA312447473GA SA312447477GA SA312447479GA SA312447480GA SA312447516GA SA312447542GA SA312447548GA SA312447554GA SA312447558GA SA312447568GA SA312447581GA SA312457686GA SA312457723GA SA312457727GA SA312457728GA SA312457736GA SA312457786GA SA312457795GA SA312457800GA SA312457807GA SA312457811GA SA312457858GA SA312457923GA SA312458187GA SA312468673GA SA312468680GA SA312468878GA SA312468977GA SA312469134GA SA312470080GA SA312480209GA SA312480261GA SA312480331GA SA312480338GA SA312480343GA SA312480353GA SA312480373GA SA312480377GA SA312480383GA SA312480388GA SA312480412GA SA312480413GA SA312480415GA SA312480420GA SA312480424GA SA312480439GA SA312480572GA SA312480581GA SA312480583GA SA312480682GA SA312480761GA SA312490974GA SA312491074GA SA312491091GA SA312491162GA SA312491169GA SA312491179GA SA312491208GA SA312491231GA SA312491248GA SA312491256GA SA312491272GA SA312491274GA SA312491290GA SA312491293GA SA312491476GA SA312491487GA SA312491491GA SA312491494GA SA312491496GA SA312491502GA SA312491507GA SA312491513GA SA312491518GA SA312491535GA SA312491536GA SA312491538GA SA312491541GA SA312491542GA SA312491546GA SA312501817GA SA312501819GA SA312501825GA SA312501827GA SA312501828GA SA312501829GA SA312501832GA SA312501833GA SA312501834GA SA312501835GA SA312501836GA SA312501840GA SA312501841GA SA312501842GA SA312501843GA SA312501844GA SA312501875GA SA312501964GA SA312501965GA SA312501967GA SA312501977GA SA312501984GA SA312501985GA SA312502004GA SA312502012GA SA312502015GA SA312502017GA SA312502019GA SA312502024GA SA312502025GA SA312502041GA SA312502045GA SA312502053GA SA312502055GA SA312502078GA SA312502089GA SA312502093GA SA312502094GA SA312502117GA SA312502118GA SA312502119GA SA312502132GA SA312502150GA SA312502169GA SA312512293GA SA312522481GA SA312522539GA SA312522547GA SA313012866GA SA313012963GA SA313013082GA SA313013083GA SA313013114GA SA313013116GA SA313013119GA SA313013121GA SA313013128GA SA313013153GA SA313023236GA SA313023260GA SA313023261GA SA313023263GA SA313023265GA SA313023267GA SA313023268GA SA313023269GA SA313023270GA SA313023271GA SA313023272GA SA313023273GA SA313023276GA SA313023287GA SA313023301GA SA313023305GA SA313023309GA SA313023314GA SA313023344GA SA313023365GA SA313023366GA SA313023368GA SA313023369GA SA313023370GA SA313023373GA SA313023377GA SA313023379GA SA313023381GA SA313023383GA SA313023384GA SA313023385GA SA313023387GA SA313023389GA SA313023395GA SA313023396GA SA313023397GA SA313023401GA SA313023405GA SA313023410GA SA313023413GA SA313023414GA SA313033646GA SA313033647GA SA313043913GA SA313043924GA SA313043925GA SA313043976GA SA313043994GA SA313044131GA SA313044133GA SA313084450GA SA313084462GA SA313084548GA SA313084582GA SA313084588GA SA313084592GA SA313084647GA SA313084666GA SA313084667GA SA313084675GA SA313084696GA SA313084811GA SA313084875GA SA313084896GA SA313084900GA SA313084904GA SA313084924GA SA313084929GA SA313095228GA SA313095230GA SA313095231GA SA313095242GA SA313095246GA SA313095248GA SA313095250GA SA313095254GA SA313095292GA SA313095294GA SA313095303GA SA313095323GA SA313095336GA SA313095337GA SA313095340GA SA313095342GA SA313095344GA SA313095350GA SA313095354GA SA313095355GA SA313095356GA SA313095358GA SA313095367GA SA313095368GA SA313095373GA SA313095374GA SA313095382GA SA313095386GA SA313095389GA SA313095390GA SA313095397GA SA313095402GA SA313095404GA SA313095407GA SA313095408GA SA313095409GA SA313095413GA SA313095421GA SA313095423GA SA313095425GA SA313095426GA SA313095427GA SA313095428GA SA313095430GA SA313095432GA SA313095459GA SA313095464GA SA313095474GA SA313095475GA SA313095479GA SA313095480GA SA313095481GA SA313095482GA SA313095492GA SA313095496GA SA313095500GA SA313095504GA SA313095507GA SA313095509GA SA313095520GA SA313095523GA SA313095524GA SA313095525GA SA313095529GA SA313095530GA SA313095590GA SA313105891GA SA313105911GA SA313105987GA SA313116681GA SA313116694GA SA313116711GA SA313116728GA SA313116875GA SA313116884GA SA313127465GA SA313127590GA SA313127607GA SA313198299GA SA313198418GA SA313198426GA SA313198432GA SA313198434GA SA313198436GA SA313198437GA SA313198443GA SA313198507GA SA313198524GA SA313198525GA SA313198530GA SA313198744GA SA313208803GA SA313208835GA SA313208881GA SA313208885GA SA313230550GA SA313240908GA SA313271984GA SA313272267GA SA313272270GA SA313272358GA SA313272362GA SA313272371GA SA313272412GA SA313272415GA SA313282477GA SA313282545GA SA313282555GA SA313282557GA SA313282560GA SA313282561GA SA313282567GA SA313282575GA SA313282576GA SA313292945GA SA313292962GA SA313292972GA SA313292982GA SA313292992GA SA313292997GA SA313292999GA SA313293039GA SA313293044GA SA313293045GA SA313293051GA SA313293087GA SA313293092GA SA313293113GA SA313293151GA SA313293152GA SA313293156GA SA313293176GA SA313293182GA SA313293203GA SA313293205GA SA313293206GA SA313293207GA SA313293218GA SA313293222GA SA313293226GA SA313293228GA SA313293239GA SA313293246GA SA313293255GA SA313293264GA SA313293265GA SA313293270GA SA313293274GA SA313293275GA SA313293277GA SA313293278GA SA313293280GA SA313293284GA SA313293286GA SA313314465GA SA313314534GA SA313314546GA SA313314553GA SA313314579GA SA313335364GA SA313335456GA SA313335508GA SA313345620GA SA313345629GA SA313345630GA SA313345631GA SA313345648GA SA313345696GA SA313345698GA SA313345780GA SA313345860GA SA313356138GA SA313356336GA SA313356379GA SA313356538GA SA313398799GA SA313398816GA SA313409203GA SA313409356GA SA313419432GA SA313419444GA SA313419448GA SA313419451GA SA313419460GA SA313419582GA SA313420030GA SA313420033GA SA313420034GA SA313420043GA SA313420059GA SA313420063GA SA313420065GA SA313420069GA SA313420071GA SA313420075GA SA313420083GA SA313420089GA SA313420100GA SA313420127GA SA313420132GA SA313420144GA SA313420166GA SA313420181GA SA313420189GA SA313420190GA SA313420194GA SA313420232GA SA313420251GA SA313420263GA SA313420269GA SA313420277GA SA313420293GA SA313431173GA SA313431175GA SA313431176GA SA313431177GA SA313431179GA SA313431191GA SA313431196GA SA313431199GA SA313431200GA SA313431202GA SA313431203GA SA313431209GA SA313431216GA SA313431217GA SA313431218GA SA313431220GA SA313431221GA SA313431222GA SA313431223GA SA313431224GA SA313431229GA SA313431232GA SA313431233GA SA313431234GA SA313431236GA SA313431238GA SA313431240GA SA313441244GA SA313441268GA SA313441276GA SA313441277GA SA313441278GA SA313441279GA SA313441280GA SA313441282GA SA313441317GA SA313441318GA SA313441319GA SA313441341GA SA313441373GA SA313441384GA SA313441502GA SA313441600GA SA313441651GA SA313441676GA SA313451729GA SA313451730GA SA313451731GA SA313451734GA SA313451735GA SA313451738GA SA313451741GA SA313451744GA SA313451747GA SA313451785GA SA313451802GA SA313451814GA SA313451825GA SA313451835GA SA313451852GA SA313451878GA SA313451879GA SA313451882GA SA313451884GA SA313451927GA SA313451933GA SA313451937GA SA313451938GA SA313451963GA SA313451979GA SA313451985GA SA313452002GA SA313452003GA SA313452005GA SA313452010GA SA313452027GA SA313452032GA SA313452039GA SA313452047GA SA313452048GA sa313452055ga SA313452058GA SA313452060GA SA313452062GA SA313452066GA SA313452068GA SA313452075GA SA313452076GA SA313452078GA SA313452084GA SA313452086GA SA313452090GA SA313452096GA SA313452120GA SA313452137GA SA313452153GA SA313462160GA SA313462162GA SA313462164GA SA313462165GA SA313462168GA SA313462169GA SA313462172GA SA313462177GA SA313462184GA SA313462187GA SA313462189GA SA313462198GA SA313462202GA SA313462205GA SA313462206GA SA313462210GA SA313462215GA SA313462228GA SA313462235GA SA313462245GA SA313462335GA SA313462421GA SA313462502GA SA313482648GA SA313482689GA SA313482694GA SA314014495GA SA314014538GA SA314024995GA SA314024996GA SA314025031GA SA314025043GA SA314025046GA SA314025060GA SA314045519GA SA314045526GA SA314055566GA SA314055597GA sa314055716ga SA307220981GA SA307231400GA SA308409120GA SA308429399GA SA310331340GA SA310403380GA SA310423975GA SA310434087GA SA310434121GA SA310434147GA SA310507177GA SA310507291GA SA310517622GA SA310517627GA SA310517640GA SA310517654GA SA310517665GA SA310517666GA SA310517668GA SA310517669GA SA310517671GA SA310517674GA SA310518148GA SA310518166GA SA310518180GA SA310518196GA SA310518201GA SA310518215GA SA310518218GA SA310518224GA SA310518238GA SA310518243GA SA310518246GA SA310518249GA SA310518250GA SA310518264GA SA310518267GA SA310518273GA SA310518279GA SA310518293GA SA310518295GA SA310518301GA SA310518310GA SA310518313GA SA310518314GA SA310518318GA SA310518321GA SA310518327GA SA310518334GA SA311111661GA SA311111672GA SA311122134GA SA311132525GA SA311428062GA SA308105217GA SA309507427GA SA310361911GA SA310372062GA SA310382382GA SA310382384GA SA310382390GA SA310382392GA SA310382393GA SA310382394GA SA310382396GA SA310382436GA SA310382439GA SA310382441GA SA310382448GA SA310382457GA SA310382467GA SA310382471GA SA310382475GA SA310382480GA SA310382481GA SA310382482GA SA310382486GA SA310382489GA SA310382490GA SA310382493GA SA310382496GA SA310382500GA SA310382503GA SA310382506GA SA310382507GA SA310382509GA SA310382510GA SA310382512GA SA310382514GA SA310382515GA SA310382519GA SA310382520GA SA310382521GA SA310382522GA SA310393236GA SA310393259GA SA310393272GA SA310393275GA SA310393277GA SA310393288GA SA310403336GA SA310403337GA SA310413454GA SA310413455GA SA310413456GA SA310413457GA SA310413459GA SA310413462GA SA310413463GA SA310413464GA SA310413465GA SA310413466GA SA310413469GA SA310413508GA SA310413541GA SA310413542GA SA310413548GA SA310413549GA SA310413555GA SA310413557GA SA310413558GA SA310413564GA SA310413565GA SA310423585GA SA310423586GA SA310423588GA SA310423589GA SA310423591GA SA310423592GA SA310423602GA SA310423608GA SA310423609GA SA310423610GA SA310423615GA SA310423616GA SA310423623GA SA310423656GA SA310423659GA SA310423661GA SA310423664GA SA310423691GA SA310423740GA SA310423746GA SA310423752GA SA310423767GA SA310423768GA SA310496139GA SA310496141GA SA310496157GA SA310496460GA SA310496465GA SA310496498GA SA310496515GA SA310518647GA SA310528976GA SA311069811GA SA311101481GA SA311101484GA SA311101490GA SA311101492GA SA311101500GA SA311101501GA SA311101508GA SA311101509GA SA311101518GA SA311101524GA SA311101530GA SA311101533GA SA311101548GA SA311101551GA SA311111611GA SA311121704GA SA311121706GA SA311121710GA SA311121715GA SA311121716GA SA311174209GA SA311174223GA SA311184377GA SA311184537GA SA311184541GA SA311184577GA SA311184585GA SA311184645GA SA311225978GA SA311225983GA SA311225984GA SA311225990GA SA311236063GA SA311236072GA SA311236073GA SA311236089GA SA311236091GA SA311236095GA SA311236099GA SA311236105GA SA311236107GA SA311236109GA SA311236112GA SA311236113GA SA311236118GA SA311266961GA SA311266986GA SA312059541GA SA312069641GA SA312253058GA SA312253192GA SA312253293GA SA312253296GA SA312253301GA SA312253315GA SA312253341GA SA312253351GA SA312345141GA SA312407280GA SA312407283GA SA312447511GA SA312447519GA SA312447520GA SA312447525GA SA312447536GA SA312447538GA SA312447550GA SA312447556GA SA312447561GA SA312447562GA SA312447563GA SA312447565GA SA312447571GA SA312447572GA SA312447575GA SA312447580GA SA312447582GA SA312447600GA SA312457969GA SA312457998GA SA312458096GA SA312468555GA SA312469054GA SA312469443GA SA312469454GA SA312470024GA SA312470060GA SA312479559GA SA312479560GA SA312479562GA SA312479563GA SA312479564GA SA312479565GA SA312479567GA SA312479571GA SA312479574GA SA312479576GA SA312479587GA SA312479591GA SA312479597GA SA312479598GA SA312479601GA SA312479602GA SA312479604GA SA312479605GA SA312479609GA SA312479611GA SA312479630GA SA312479632GA SA312479634GA SA312479635GA SA312479636GA SA312479637GA SA312479639GA SA312479640GA SA312479680GA SA312479683GA SA312479687GA SA312479689GA SA312479691GA SA312479699GA SA312479702GA SA312479703GA SA312479710GA SA312479715GA SA312479722GA SA312479723GA SA312479728GA SA312479759GA SA312479770GA SA312479796GA SA312479798GA SA312479808GA SA312479812GA SA312479821GA SA312479828GA SA312479830GA SA312479832GA SA312479835GA SA312479843GA SA312479857GA SA312479862GA SA312479863GA SA312479864GA SA312479865GA SA312479867GA SA312479869GA SA312479870GA SA312479873GA SA312479877GA SA312479878GA SA312479879GA SA312479880GA SA312479881GA SA312479882GA SA312479885GA SA312479886GA SA312479888GA SA312479889GA SA312479890GA SA312479894GA SA312480109GA SA312480116GA SA312480117GA SA312480419GA SA312491440GA SA313043962GA SA313044016GA SA313044031GA SA313044039GA SA313044040GA SA313044048GA SA313044059GA SA313044060GA SA313044082GA SA313044091GA SA313044100GA SA313044101GA SA313044111GA SA313064196GA SA313095065GA SA313095123GA SA313095166GA SA313095169GA SA313095180GA SA313095262GA SA313095263GA SA313095267GA SA313095268GA SA313095270GA SA313095272GA SA313095279GA SA309334765GA SA309334788GA SA309334812GA SA309365114GA SA309365190GA SA310403317GA SA309304258GA SA309304275GA SA309304277GA SA309324453GA SA309324455GA SA309324480GA SA309324494GA SA309324496GA SA309324507GA SA309324508GA SA309324509GA SA309324512GA SA313127214GA SA313230677GA SA313261855GA SA313272282GA SA313272366GA SA313272374GA SA313272376GA SA313272377GA SA313272387GA SA313272442GA SA313282443GA SA313282446GA SA313293055GA SA313293086GA SA313293109GA SA313293145GA SA313293157GA SA313293179GA SA313293187GA SA313293199GA SA313293202GA SA313293281GA SA313293287GA SA313293291GA SA313314447GA SA313314468GA SA313314472GA SA313314520GA SA313314522GA SA313314528GA SA313314530GA SA313314531GA SA313314532GA SA313314533GA SA313314536GA SA313314537GA SA313314539GA SA313314541GA SA313314542GA SA313314543GA SA313314544GA SA313314547GA SA313314548GA SA313314549GA SA313314551GA SA313314552GA SA313314560GA SA313314562GA SA313314566GA SA313314568GA SA313314580GA SA313314583GA SA313314584GA SA313324911GA SA313346077GA SA313409243GA SA313409343GA SA313409362GA SA313409395GA SA313409401GA SA313409402GA SA313409424GA SA313409428GA SA313409430GA SA313419440GA SA313419446GA SA313419450GA SA313420046GA SA313420056GA SA313420062GA SA313420067GA SA313420078GA SA313420086GA SA313430725GA SA313430738GA SA313431147GA SA313431164GA SA313431167GA SA313431215GA SA313452092GA SA314014602GA SA314075846GA SA314119087GA SA314139553GA SA314160035GA SA314160038GA SA314160049GA SA314160054GA SA314160058GA SA314180629GA SA314180632GA SA314190969GA SA314190981GA SA314190983GA SA314191018GA SA314191023GA SA314201223GA SA314201756GA SA314273918GA SA314273921GA SA314273935GA SA314273954GA SA314273956GA SA314273962GA SA314273980GA SA314273991GA SA309263915GA SA309354983GA SA309354994GA SA309538053GA SA309538073GA SA310149311GA SA310159370GA SA310159381GA SA310159463GA SA310159464GA SA310159465GA SA310159466GA SA310159467GA SA310159468GA SA310159469GA SA310259904GA SA310403377GA SA310403412GA SA310403430GA SA310423791GA SA310434060GA SA310434101GA SA310434112GA SA310434117GA SA310496130GA SA310496136GA SA310496150GA SA310496158GA SA310496312GA SA310496320GA SA310507348GA SA310518422GA SA310518489GA SA310518494GA SA310518496GA SA310518506GA SA310518525GA SA310539289GA SA311069863GA SA311080674GA SA311090900GA SA311090907GA SA311090918GA SA311090920GA SA311090933GA SA311091034GA SA311091035GA SA311091049GA SA311091053GA SA311101099GA SA311101114GA SA311101123GA SA312049310GA SA312069828GA SA312070021GA SA312070050GA SA312079964GA SA312079973GA SA312079986GA SA312079990GA SA312091159GA SA312091175GA SA312091176GA SA312091177GA SA312091179GA SA312091188GA SA312091189GA SA312091193GA SA312091195GA SA312091198GA SA312091199GA SA312091205GA SA312091212GA SA312091214GA SA312091217GA SA312091228GA SA312192684GA SA312232891GA SA312253202GA SA312253218GA SA312253460GA SA312253480GA SA312253524GA SA312263617GA SA312263978GA SA312264013GA SA312264030GA SA312264038GA SA312264054GA SA312264060GA SA312376178GA SA312386314GA SA312417390GA SA312417417GA SA312480111GA SA312480112GA SA313064298GA SA313084680GA SA313095243GA SA313095334GA SA313095335GA SA313095388GA SA313095470GA SA313095501GA SA313335058GA SA313378229GA SA313378285GA SA313378292GA SA313378295GA SA313378323GA SA313378389GA SA313388531GA SA313388540GA SA313430750GA SA313431011GA SA313431046GA SA313431049GA SA313431087GA SA313431090GA SA313431103GA SA313431106GA SA313431110GA SA313431113GA SA313431116GA SA313431119GA SA313431122GA SA313431124GA SA313431133GA SA313431135GA SA313431139GA SA313431144GA SA313431157GA SA313431161GA SA313431170GA SA313441347GA SA313452007GA SA314086652GA SA314086683GA SA314201337GA SA314201345GA SA314201350GA SA314222713GA SA314222781GA SA312480405GA SA312480747GA SA312501602GA SA312501605GA SA312501606GA SA312501619GA SA312501623GA SA312501626GA SA312501640GA SA312501641GA SA312501642GA SA312501645GA SA312501649GA SA312501651GA SA312501653GA SA312501654GA SA308511515GA SA310434165GA SA311184634GA SA311287145GA SA311307504GA SA311317705GA SA311317710GA SA311317712GA SA311317713GA SA311317723GA SA311317738GA SA311327805GA SA311327806GA SA311327813GA SA311327815GA SA311327828GA SA311327831GA SA311508587GA SA313106136GA SA313261761GA SA313272119GA SA313335042GA SA313335044GA SA313335135GA SA313335136GA SA313335141GA SA313335143GA SA313335150GA SA313335151GA SA313335161GA SA313335166GA SA313335167GA SA313335170GA SA313335173GA SA313335178GA SA313335184GA SA313335188GA SA313335190GA SA313335191GA SA313335192GA SA313335202GA SA313335209GA SA313335220GA SA313335234GA SA313335242GA SA313335244GA SA313335248GA SA313335375GA SA313430686GA SA313441529GA SA313451739GA SA313462312GA SA314004180GA SA314004197GA SA314004202GA SA314004203GA SA314004204GA SA314004208GA SA314004210GA SA314004212GA SA314004215GA SA314004216GA SA314004245GA SA314004246GA SA314004293GA SA314004294GA SA314004297GA SA314004309GA SA314004310GA SA314004311GA SA314004316GA SA314014680GA SA314024768GA SA314024776GA SA314024785GA SA314024790GA SA314024796GA SA314024845GA SA314024851GA SA314024883GA SA314024889GA SA314024918GA SA314024925GA SA314024926GA SA314024927GA SA314024938GA SA314024939GA SA314024940GA SA314024942GA SA314076218GA SA314201390GA SA314201392GA SA314201421GA SA314201427GA SA314201475GA SA314201476GA SA314201491GA SA314201506GA SA314201511GA SA314201838GA SA314201840GA SA314201845GA SA314253610GA SA314273968GA SA314274315GA SA314284433GA SA314284444GA SA314284461GA SA314284492GA SA314284529GA SA314284532GA SA314284542GA SA314284552GA SA314284563GA SA314284570GA SA314284590GA SA314284591GA SA314284593GA SA314284595GA SA314284608GA SA314284621GA SA314284640GA SA314284644GA SA314284645GA SA307020374GA SA307020375GA SA307020381GA SA307020385GA SA307020387GA SA307030404GA SA307030408GA SA307060435GA SA307080481GA SA307190714GA SA307190715GA SA307190719GA SA307190723GA SA307190725GA SA307190735GA SA307190740GA SA307190747GA SA307200779GA SA307200790GA SA307210807GA SA307210809GA SA307210811GA SA307210817GA SA307210860GA SA307210861GA SA307210862GA SA307220874GA SA307220881GA SA307220882GA SA307220883GA SA307220886GA SA307220889GA SA307220895GA SA307220899GA SA307220912GA SA307220918GA SA307220919GA SA307220922GA SA307220926GA SA307220929GA SA307220930GA SA307220934GA SA307220962GA SA307220965GA SA307220971GA SA307220972GA SA307220973GA SA307220974GA SA307220980GA SA307220994GA SA307220995GA SA307220998GA SA307221015GA SA307221025GA SA307221029GA SA307221082GA SA307221083GA SA307221098GA SA307221146GA SA307221151GA SA307221155GA SA307231216GA SA307231380GA SA307251730GA SA307291889GA SA307291941GA SA307291942GA SA307291947GA SA307291956GA SA307291960GA SA307302009GA SA308155839GA SA308155842GA SA308155848GA SA308155859GA SA308155871GA SA308166109GA SA308166116GA SA308409068GA SA308409080GA SA308409086GA SA308409178GA SA308419273GA SA308511450GA SA309334726GA SA309375498GA SA309385509GA SA309385569GA SA309385580GA SA309426108GA SA309436324GA SA309436360GA SA309507342GA SA309507376GA SA309507387GA SA309507413GA SA309517634GA SA309517685GA SA309517693GA SA309517708GA SA309517833GA SA309517895GA SA309517907GA SA309527964GA SA309538028GA SA309538032GA SA309538035GA SA309538036GA SA309538037GA SA309538039GA SA309538041GA SA309538047GA SA309538088GA SA310028151GA SA310028153GA SA310028173GA SA310028175GA SA310028178GA SA310028179GA SA310028188GA SA310028190GA SA310028205GA SA310028209GA SA310028238GA SA310028239GA SA310028241GA SA310028245GA SA310028247GA SA310028248GA SA310028249GA SA310028251GA SA310028253GA SA310028254GA SA310028257GA SA310028259GA SA310028260GA SA310028264GA SA310028279GA SA310028283GA SA310028295GA SA310028296GA SA310068888GA SA310068913GA SA310068917GA SA310331263GA SA310423665GA SA310423669GA SA310423679GA SA310423692GA SA310423700GA SA310423703GA SA310423706GA SA310423707GA SA310423710GA SA310423711GA SA310423714GA SA310423716GA SA310423718GA SA310423719GA SA310423722GA SA310423727GA SA310423927GA SA310423941GA SA310423954GA SA310423964GA SA310423969GA SA310424014GA SA311121969GA SA311121972GA SA311121974GA SA311121976GA SA311121988GA SA311122006GA SA311132539GA SA311132607GA SA311132668GA SA311142807GA SA311142978GA SA311163458GA SA311163467GA SA311163485GA SA311163553GA SA311174067GA SA311184438GA SA311184474GA SA311184498GA SA311184515GA SA311184518GA SA311184519GA SA311184533GA SA311184544GA SA311184554GA SA311184558GA SA311184559GA SA311184560GA SA311184564GA SA311184571GA SA311184572GA SA311184581GA SA311184582GA SA311184584GA SA311184590GA SA311184592GA SA311184608GA SA311184610GA SA311184611GA SA311184612GA SA311307439GA SA311448205GA SA312480406GA SA312491525GA SA312491540GA SA307140602GA SA307140624GA SA307140626GA SA307140633GA SA307140634GA SA307140635GA SA307140643GA SA307221157GA SA307231261GA SA307291858GA SA307291865GA SA307291870GA SA307301991GA SA307302014GA SA307302024GA SA307382452GA SA307423179GA SA307504158GA SA307514347GA SA307514351GA SA307514392GA SA307514393GA SA307514394GA SA307514395GA SA307514396GA SA307514401GA SA307514403GA SA307514406GA SA307514408GA SA307514412GA SA307514429GA SA307514436GA SA307514437GA SA307514439GA SA307514440GA SA307514441GA SA307514442GA SA307514443GA SA307514447GA SA307514448GA SA307514449GA SA307514452GA SA307514454GA SA307514455GA SA307514457GA SA307524459GA SA307524474GA SA307524475GA SA307524478GA SA307524481GA SA307524482GA SA307524483GA SA308044496GA SA308135528GA SA308368077GA SA308459854GA SA309375490GA SA310139243GA SA310149318GA SA310159365GA SA310159372GA SA310159384GA SA310159388GA SA310159390GA SA310159471GA SA310290656GA SA310341378GA SA310351470GA SA310351488GA SA310351489GA SA310351498GA SA310351511GA SA310351512GA SA310351515GA SA310351522GA SA310351571GA SA310351586GA SA310361896GA SA310392940GA SA310392941GA SA310392945GA SA310392948GA SA310392956GA SA310392958GA SA310392960GA SA310392961GA SA310393016GA SA310393018GA SA310393021GA SA310393022GA SA310393043GA SA310393046GA SA310393047GA SA310403318GA SA310423754GA SA310423782GA SA310423870GA SA310423884GA SA310423891GA SA310423894GA SA310423898GA SA310423905GA SA310423972GA SA310423984GA SA310423992GA SA310424041GA SA310434047GA SA310434061GA SA310434073GA SA310434080GA SA310434086GA SA310434116GA SA310434126GA SA310434127GA SA310434128GA SA310434129GA SA310434130GA SA310434144GA SA310434149GA SA310434159GA SA310434161GA SA310434163GA SA310434164GA SA310434167GA SA310434169GA SA310434175GA SA310434176GA SA310434177GA SA310434184GA SA310434191GA SA310455068GA SA310455116GA SA310455136GA SA310455231GA SA310455238GA SA310455253GA SA310455270GA SA310455274GA SA310455288GA SA310475513GA SA310475570GA SA310506756GA SA310506759GA SA310506761GA SA310506781GA SA310506791GA SA310506792GA SA310506806GA SA310507088GA SA310507158GA SA310507206GA SA310507221GA SA310507263GA SA310507268GA SA310507287GA SA310507290GA SA310507391GA SA310507405GA SA310507425GA SA310507440GA SA310507446GA SA310507466GA SA310507472GA SA310507530GA SA310517727GA SA310517740GA SA310517741GA SA310518322GA SA310518455GA SA310518478GA SA311039521GA SA311059579GA SA311059591GA SA311059604GA SA311060121GA SA311060246GA SA311069662GA SA311069672GA SA311069691GA SA311069752GA SA311069756GA SA311070526GA SA311070530GA SA311101133GA SA311101172GA SA311111628GA SA311121910GA SA311122096GA SA311122189GA SA311122198GA SA311122203GA SA311122220GA SA311132279GA SA311132286GA SA311132471GA SA311132497GA SA311132501GA SA311132564GA SA311132568GA SA311142884GA SA311142893GA SA311142897GA SA311142902GA SA311142906GA SA311184380GA SA311184391GA SA311184393GA SA311184570GA SA311194896GA SA311195071GA SA311195208GA SA311205378GA SA311205386GA SA311256703GA SA311266909GA SA311508616GA SA312018853GA SA312018854GA SA312018855GA SA312018858GA SA312018859GA SA312049239GA SA312049240GA SA312049266GA SA312080469GA SA312080530GA SA312080547GA SA312080572GA SA312080576GA SA312080580GA SA312080767GA SA312080829GA SA312121803GA SA312253237GA SA312253266GA SA312253271GA SA312253282GA SA312294389GA SA312294396GA SA312294398GA SA312458245GA SA312480367GA SA312480369GA SA312480732GA SA312480856GA SA312480857GA SA312480868GA SA312490871GA SA313095587GA SA313117175GA SA313240917GA SA313293069GA SA313293071GA SA313293074GA SA313293075GA SA313293150GA SA313314441GA SA313314637GA SA313324895GA SA313324965GA SA313324971GA SA313345723GA SA313345726GA SA313388513GA SA313388526GA SA313420452GA SA313420502GA SA313420520GA SA313420542GA SA307210808GA SA307210814GA SA307210816GA SA307210830GA SA307210834GA SA307210859GA SA307220964GA SA307220979GA SA307221091GA SA307221174GA SA307221187GA SA307231206GA SA307231214GA SA307231258GA SA307241511GA SA307241576GA SA307241692GA SA307251725GA SA307251804GA SA307261807GA SA307261808GA SA307261809GA SA307261810GA SA307261815GA SA307261832GA SA307261834GA SA307291867GA SA307291890GA SA307291896GA SA307291902GA SA307291906GA SA307291908GA SA307291909GA SA307291915GA SA307291916GA SA307291917GA SA307291919GA SA307291921GA SA307291922GA SA307291944GA SA307301999GA SA307302013GA SA307302020GA SA307302025GA SA307302031GA SA307342304GA SA307352361GA SA307352365GA SA307352372GA SA307362409GA SA307362410GA SA307362411GA SA307362422GA SA307362425GA SA307362428GA SA307362432GA SA307362433GA SA307362437GA SA307382462GA SA307382487GA SA307382512GA SA307382533GA SA307382562GA SA307382570GA SA307382571GA SA307382572GA SA307382580GA SA307382585GA SA307382588GA SA307382590GA SA307382598GA SA307382602GA SA307382606GA SA307382607GA SA307393028GA SA307443257GA SA307443267GA SA307443378GA SA307453393GA SA307453431GA SA307453552GA SA307453595GA SA307504194GA SA307514322GA SA307514324GA SA307514327GA SA307514344GA SA308094868GA SA308094876GA SA308094948GA SA308094952GA SA308094957GA SA308094970GA SA308115308GA SA308135534GA SA308135544GA SA308155940GA SA308155948GA SA308166071GA SA308186356GA SA308186418GA SA308196507GA SA308196508GA SA308196519GA SA308196543GA SA308196544GA SA308196566GA SA308206590GA SA308206599GA SA308206600GA SA308206627GA SA308206642GA SA308206663GA SA308206699GA SA308206753GA SA308216771GA SA308216778GA SA308297029GA SA308297034GA SA308297035GA SA308297037GA SA308297038GA SA308297040GA SA308297042GA SA308297044GA SA308297045GA SA308297046GA SA308297047GA SA308297048GA SA308297051GA SA308297052GA SA308297053GA SA308297056GA SA308297057GA SA308297058GA SA308327422GA SA308327423GA SA308327443GA SA308327476GA SA308327484GA SA308327489GA SA308327490GA SA308327498GA SA308327504GA SA308327522GA SA308327534GA SA308327538GA SA308327539GA SA308327540GA SA308327541GA SA308327552GA SA308327553GA SA308337559GA SA308337560GA SA308337561GA SA308337593GA SA308337631GA SA308337636GA SA308357878GA SA308368050GA SA308368062GA SA308368202GA SA308378296GA SA308388446GA SA308388596GA SA308388632GA SA308398659GA SA308398689GA SA308398724GA SA308398726GA SA308398737GA SA308398748GA SA308398767GA SA308398774GA SA308398776GA SA308398856GA SA308398862GA SA308398863GA SA308398867GA SA308398902GA SA308408905GA SA308408908GA SA308408921GA SA308408937GA SA308408941GA SA308409015GA SA308409023GA SA308409028GA SA308409031GA SA308409055GA SA308409072GA SA308409084GA SA308409088GA SA308409156GA SA308409157GA SA308409160GA SA308409164GA SA308409165GA SA308409187GA SA308419245GA SA308419246GA SA308419249GA SA308419251GA SA308419271GA SA308511518GA SA309082444GA SA309082445GA SA309082446GA SA309082448GA SA309082451GA SA309082460GA SA309082464GA SA309082465GA SA309082466GA SA309082467GA SA309112601GA SA309132857GA SA309142880GA SA309142921GA SA309142922GA SA309142923GA SA309142924GA SA309142925GA SA309142927GA SA309142933GA SA309142935GA SA309142937GA SA309142949GA SA309142952GA SA309142963GA SA309173150GA SA309193334GA SA309193354GA SA309193358GA SA309203365GA SA309203369GA SA309203380GA SA309203382GA SA309203402GA SA309223477GA SA309223479GA SA309223480GA SA309223507GA SA309223515GA SA309253828GA SA309283993GA SA309283994GA SA309283996GA SA309283998GA SA309284000GA SA309284006GA SA309304225GA SA309304228GA SA309304233GA SA309304239GA SA309304241GA SA309304242GA SA309304246GA SA309304247GA SA309304250GA SA309304253GA SA309304254GA SA309304255GA SA309304256GA SA309304257GA SA309304260GA SA309304265GA SA309304267GA SA309304268GA SA309304270GA SA309304271GA SA309304273GA SA309304274GA SA309304276GA SA309304279GA SA309304281GA SA309304285GA SA309304286GA SA309304288GA SA309324446GA SA309324447GA SA309324448GA SA309324450GA SA309324451GA SA309324452GA SA309324454GA SA309324457GA SA309324458GA SA309324459GA SA309324460GA SA309324461GA SA309324462GA SA309324463GA SA309324464GA SA309324465GA SA309324467GA SA309324468GA SA309324469GA SA309324471GA SA309324474GA SA309324476GA SA309324477GA SA309324478GA SA309324479GA SA309324484GA SA309324487GA SA309324488GA SA309324489GA SA309324491GA SA309324493GA SA309324495GA SA309324497GA SA309324498GA SA309324499GA SA309324500GA SA309324501GA SA309324503GA SA309324505GA SA309324510GA SA309324511GA SA309324513GA SA309324514GA SA309324517GA SA309324518GA SA309324522GA SA309324523GA SA309324524GA SA309324526GA SA309324528GA SA309324529GA SA309324530GA SA309324531GA SA309324535GA SA309324536GA SA309324550GA SA309324552GA SA309324553GA SA309324554GA SA309324555GA SA309334671GA SA309334691GA SA309334698GA SA309334709GA SA309334743GA SA309334749GA SA309334825GA SA309344845GA SA309344848GA SA309354901GA SA309365088GA SA309365090GA SA309365097GA SA309365102GA SA309365104GA SA309365124GA SA309365125GA SA309365126GA SA309365128GA SA309365132GA SA309365133GA SA309365135GA SA309365136GA SA309365137GA SA309365138GA SA309365139GA SA309365140GA SA309365142GA SA309365144GA SA309365290GA SA309385582GA SA309385589GA SA309385676GA SA309385689GA SA309395693GA SA309395694GA SA309395698GA SA309395700GA SA309395703GA SA309395706GA SA309395707GA SA309395713GA SA309395714GA SA309395715GA SA309395720GA SA309395727GA SA309395732GA SA309395735GA SA309395736GA SA309395745GA SA309395748GA SA309395752GA SA309395762GA SA309436272GA SA309436281GA SA309436292GA SA309436315GA SA309436348GA SA309456635GA SA309456646GA SA309456653GA SA309456675GA SA309456688GA SA309497146GA SA309497162GA SA309497171GA SA309497230GA SA309497232GA SA309497282GA SA309497308GA SA309517835GA SA310038388GA SA310038393GA SA310038423GA SA310038436GA SA310048614GA SA310048621GA SA310068907GA SA310068931GA SA310139084GA SA310159453GA SA310199559GA SA310199560GA SA310199563GA SA310199568GA SA310199569GA SA310199571GA SA310199574GA SA310199577GA SA310199582GA SA310199584GA SA310199585GA SA310199586GA SA310199641GA SA310199659GA SA310199662GA SA310199671GA SA310199681GA SA310199685GA SA310199687GA SA310199688GA SA310199692GA SA310199693GA SA310199694GA SA310199696GA SA310199697GA SA310199698GA SA310199700GA SA310199706GA SA310199714GA SA310199718GA SA310199724GA SA310199726GA SA310199729GA SA310199731GA SA310199732GA SA310199733GA SA310199734GA SA310199736GA SA310199740GA SA310199742GA SA310199743GA SA310199746GA SA310199747GA SA310199749GA SA310199751GA SA310199754GA SA310199755GA SA310199758GA SA310199760GA SA310199761GA SA310270579GA SA310270580GA SA310270581GA SA310300761GA SA310300762GA SA310300764GA SA310300765GA SA310300767GA SA310300787GA SA310331330GA SA310351539GA SA310393238GA SA310403315GA SA310403323GA SA310403329GA SA310403339GA SA310403354GA SA310403356GA SA310403358GA SA310403359GA SA310403360GA SA310403361GA SA310403362GA SA310403367GA SA310403370GA SA310403378GA SA310403382GA SA310403387GA SA310403434GA SA310413453GA SA310413516GA SA310413543GA SA310423580GA SA310423584GA SA310423693GA SA310423699GA SA310423732GA SA310423743GA SA310423747GA SA310423751GA SA310423802GA SA310423846GA SA310423856GA SA310423867GA SA310423877GA SA310423881GA SA310423889GA SA310423899GA SA310423909GA SA310423911GA SA310423976GA SA310423977GA SA310423991GA SA310423993GA SA310424035GA SA310424040GA SA310424043GA SA310434049GA SA310434057GA SA310434065GA SA310434068GA SA310434095GA SA310434105GA SA310434230GA SA310444268GA SA310444271GA SA310444273GA SA310444274GA SA310444279GA SA310444280GA SA310444282GA SA310444285GA SA310444288GA SA310444290GA SA310444293GA SA310444295GA SA310444297GA SA310444299GA SA310444300GA SA310444302GA SA310444303GA SA310444307GA SA310444309GA SA310444310GA SA310444313GA SA310444315GA SA310444316GA SA310444322GA SA310444323GA SA310444366GA SA310444367GA SA310444368GA SA310444487GA SA310444489GA SA310444490GA SA310444491GA SA310444492GA SA310444493GA SA310444497GA SA310444500GA SA310444501GA SA310444502GA SA310444511GA SA310444514GA SA310444517GA SA310444519GA SA310444520GA SA310444522GA SA310444526GA SA310444532GA SA310444533GA SA310444534GA SA310444843GA SA310444844GA SA310444845GA SA310444847GA SA310444848GA SA310444852GA SA310444853GA SA310444854GA SA310444857GA SA310444860GA SA310444861GA SA310444862GA SA310444865GA SA310444869GA SA310444890GA SA310444891GA SA310454930GA SA310454931GA SA310454932GA SA310454941GA SA310454942GA SA310454951GA SA310454978GA SA310454979GA SA310454980GA SA310454998GA SA310454999GA SA310455001GA SA310455002GA SA310455005GA SA310455007GA SA310455008GA SA310455010GA SA310455013GA SA310455014GA SA310455019GA SA310455020GA SA310455021GA SA310455023GA SA310455027GA SA310455032GA SA310455049GA SA310455050GA SA310455051GA SA310455054GA SA310455059GA SA310455063GA SA310455064GA SA310455070GA SA310455071GA SA310455072GA SA310455073GA SA310455074GA SA310455089GA SA310455117GA SA310455129GA SA310455135GA SA310455137GA SA310455141GA SA310455158GA SA310455162GA SA310455167GA SA310455169GA SA310455171GA SA310455225GA SA310455226GA SA310455229GA SA310455239GA SA310455240GA SA310455241GA SA310455242GA SA310455243GA SA310455244GA SA310455246GA SA310455250GA SA310455252GA SA310455255GA SA310455258GA SA310455269GA SA310455276GA SA310455281GA SA310455285GA SA310455287GA SA310455290GA SA310465296GA SA310465320GA SA310465336GA SA310465340GA SA310465345GA SA310465351GA SA310465353GA SA310465355GA SA310465356GA SA310465357GA SA310465358GA SA310465360GA SA310465362GA SA310465363GA SA310465364GA SA310465365GA SA310465372GA SA310465373GA SA310465375GA SA310465378GA SA310465382GA SA310465384GA SA310465387GA SA310465388GA SA310465389GA SA310465390GA SA310465392GA SA310465393GA SA310465395GA SA310465399GA SA310465401GA SA310465402GA SA310465404GA SA310465405GA SA310465406GA SA310465407GA SA310465409GA SA310465413GA SA310465415GA SA310465416GA SA310465417GA SA310465418GA SA310465421GA SA310465422GA SA310465424GA SA310465425GA SA310465426GA SA310465428GA SA310465429GA SA310465430GA SA310465434GA SA310465440GA SA310465442GA SA310465444GA SA310465447GA SA310465448GA SA310475450GA SA310475453GA SA310475454GA SA310475455GA SA310475456GA SA310475457GA SA310475459GA SA310475461GA SA310475462GA SA310475463GA SA310475469GA SA310475472GA SA310475475GA SA310475476GA SA310475478GA SA310475486GA SA310475491GA SA310475492GA SA310475502GA SA310475503GA SA310475506GA SA310475507GA SA310475512GA SA310475514GA SA310475521GA SA310496336GA SA310506673GA SA310506679GA SA310506680GA SA310506684GA SA310506687GA SA310506694GA SA310506695GA SA310506714GA SA310506716GA SA310506724GA SA310507066GA SA310507071GA SA310507079GA SA310507081GA SA310507101GA SA310507136GA SA310507390GA SA310507423GA SA310507450GA SA310517573GA SA310517590GA SA310517698GA SA310517699GA SA310517701GA SA310517702GA SA310517705GA SA310517713GA SA310517722GA SA310517725GA SA310517826GA SA310517827GA SA310517832GA SA310517867GA SA310517868GA SA310517874GA SA310518005GA SA310518012GA SA310518244GA SA310518248GA SA310518253GA SA310518254GA SA310518281GA SA310518284GA SA310518287GA SA310518290GA SA310518302GA SA310518316GA SA310518328GA SA310518329GA SA310518331GA SA310518338GA SA310518339GA SA310518340GA SA310518344GA SA310518405GA SA310518469GA SA310518646GA SA310518650GA SA310518667GA SA310518668GA SA310518669GA SA310518701GA SA310518707GA SA310518708GA SA310518709GA SA310518722GA SA310529077GA SA311039383GA SA311039426GA SA311039427GA SA311039487GA SA311039492GA SA311121713GA SA311121812GA SA311121917GA SA311132283GA SA311153049GA SA311163405GA SA311163859GA SA311163863GA SA311163867GA SA311163877GA SA311163883GA SA311163889GA SA311163898GA SA311163909GA SA311163918GA SA311163920GA SA311163940GA SA311163945GA SA311163958GA SA311163960GA SA311163964GA SA311163972GA SA311163976GA SA311163978GA SA311163983GA SA311163984GA SA311163987GA SA311163988GA SA311163994GA SA311163997GA SA311174177GA SA311174207GA SA311174210GA SA311174211GA SA311184625GA SA311184872GA SA313271992GA SA313282553GA SA313293122GA SA313293132GA SA313293165GA SA313293168GA SA313314437GA SA313314440GA SA313314484GA SA313523963GA SA314201477GA SA314201502GA SA314201504GA SA314201514GA SA314201518GA SA314201520GA SA314201521GA SA314201528GA SA314273928GA SA314284416GA SA314284735GA SA314294860GA SA307504124GA SA309062288GA SA309072367GA SA309072369GA SA309072371GA SA309072373GA SA309072381GA SA309072386GA SA309072388GA SA309072389GA SA309072390GA SA309072392GA SA309072394GA SA309072397GA SA309072402GA SA309072403GA SA309072404GA SA309072405GA SA309072406GA SA309072407GA SA309072408GA SA309072409GA SA309072411GA SA309082412GA SA309082414GA SA309082416GA SA309082447GA SA309082455GA SA309082458GA SA309082461GA SA309082463GA SA309112561GA SA309122668GA SA309122718GA SA309122723GA SA309122749GA SA309122767GA SA309132845GA SA309132847GA SA309132848GA SA309132849GA SA309132850GA SA309132851GA SA309132852GA SA309132853GA SA309132855GA SA309142984GA SA309142985GA SA309142986GA SA309142992GA SA309153024GA SA309153025GA SA309153027GA SA309153032GA SA309153033GA SA309153047GA SA309294117GA SA309324354GA SA309334594GA SA309334673GA SA309334731GA SA309334733GA SA309334736GA SA309334738GA SA309334742GA SA309334746GA SA309334750GA SA309344835GA SA309344840GA SA309344841GA SA309344844GA SA309344846GA SA309344849GA SA309344858GA SA309344868GA SA309354894GA SA309354895GA SA309354902GA SA309354908GA SA309354912GA SA309354918GA SA309354920GA SA309354925GA SA309354938GA SA309354943GA SA309354944GA SA309354947GA SA313229389GA SA313229392GA SA313229417GA SA313229427GA SA313229439GA SA313229447GA SA313229453GA SA313229456GA SA314180635GA SA314180651GA SA311090891GA SA311297300GA SA307120586GA SA307190721GA SA307190746GA SA307190758GA SA307322182GA SA307332207GA SA308084642GA SA308084733GA SA308155877GA SA308155883GA SA308327338GA SA309173185GA SA309304230GA SA309304235GA SA309324449GA SA309324456GA SA309324466GA SA309324475GA SA309324483GA SA309324486GA SA309324490GA SA309324492GA SA309324506GA SA309375421GA SA309375431GA SA309385684GA SA309395704GA SA309395753GA SA309395809GA SA309405829GA SA309405831GA SA309405835GA SA309446483GA SA309456733GA SA309466749GA SA309466762GA SA309466764GA SA309466765GA SA309466766GA SA309466788GA SA309476945GA SA309476951GA SA309476952GA SA309517784GA SA309517889GA SA309517893GA SA309517896GA SA309527929GA SA309527958GA SA310028170GA SA310028246GA SA310038351GA SA310038370GA SA310038384GA SA310038392GA SA310038401GA SA310038404GA SA310038405GA SA310048598GA SA310048607GA SA310048626GA SA310068851GA SA310068855GA SA310068856GA SA310068860GA SA310068862GA SA310068867GA SA310068874GA SA310068891GA SA310068894GA SA310068903GA SA310068906GA SA310068912GA SA310068920GA SA310068935GA SA310118980GA SA310139252GA SA310159460GA SA310159462GA SA310159474GA SA310169509GA SA310169514GA SA310169516GA SA310169518GA SA310169519GA SA310169521GA SA310169522GA SA310169524GA SA310169525GA SA310169528GA SA310169530GA SA310169531GA SA310169534GA SA310169535GA SA310169536GA SA310169537GA SA310260256GA SA310290636GA SA310290700GA SA310310963GA SA310321050GA SA310361677GA SA310372305GA SA310413523GA SA310434226GA SA310444267GA SA310444317GA SA310444318GA SA310444319GA SA310444320GA SA310444321GA SA310444356GA SA310444357GA SA310444359GA SA310444363GA SA310444488GA SA310444505GA SA310444508GA SA310444512GA SA310444515GA SA310444516GA SA310444525GA SA310444528GA SA310444529GA SA310444530GA SA310444535GA SA310444841GA SA310444855GA SA310444858GA SA310444866GA SA310454977GA SA310465344GA SA310475632GA SA310475711GA SA310485823GA SA310485825GA SA310485827GA SA310485830GA SA310485837GA SA310485849GA SA310485854GA SA310485855GA SA310485879GA SA310485886GA SA310496213GA SA310496217GA SA310496219GA SA310496234GA SA310496235GA SA310496239GA SA310496240GA SA310507085GA SA310507500GA SA310507510GA SA310517562GA SA310517572GA SA310517575GA SA310517576GA SA310539359GA SA311039434GA SA311091004GA SA311132288GA SA311163887GA SA311163956GA SA311327969GA SA311458279GA SA312049417GA SA312059527GA SA312070174GA SA312080722GA SA312080747GA SA312080766GA SA312080773GA SA312080793GA SA312080816GA SA312080832GA SA312101395GA SA312101413GA SA312101442GA SA312111503GA SA312152346GA SA312202740GA SA312376050GA SA312378025GA SA312378031GA SA312447567GA SA312480672GA SA312480675GA SA312491255GA SA312502133GA SA313012806GA SA313023307GA SA313023315GA SA313023388GA SA313023393GA SA313105777GA SA313117126GA SA313117137GA SA313127506GA SA313127569GA SA313127576GA SA313127599GA SA313148003GA SA313229498GA SA313229510GA SA313229518GA SA313229554GA SA313229572GA SA313229574GA SA313230132GA SA313240838GA SA313251270GA SA313251693GA SA313293289GA SA313314505GA SA313314517GA SA313314574GA SA313419779GA SA313452072GA sa313462186ga SA313482771GA SA313493003GA SA314004304GA SA314004315GA SA314014355GA SA314014400GA SA314014410GA SA314014452GA SA314014664GA SA314014673GA SA314014674GA SA314024812GA SA314024823GA SA314024826GA SA314024863GA SA314025130GA SA314055613GA SA314055709GA SA314065777GA SA314075805GA SA314075820GA SA314075822GA SA314075840GA SA314075879GA SA314076228GA SA314076253GA SA314076258GA SA314076262GA SA314076286GA SA314076306GA SA314076325GA SA314076326GA SA314076327GA SA314076331GA SA314076334GA SA314076336GA SA314076338GA SA314076341GA SA314086359GA SA314086372GA SA314086379GA SA314086445GA SA314118974GA SA314119076GA SA314119079GA SA314119093GA SA314119111GA SA314119120GA SA314119123GA SA314119135GA SA314180610GA SA314201524GA SA314222357GA SA314253672GA SA314273844GA SA314076081GA SA314170201GA SA314170218GA SA314233098GA SA314233136GA SA307220887GA SA307220909GA SA307220924GA SA307221063GA SA307221149GA SA307241696GA SA307251720GA SA307251731GA SA307251732GA SA307382596GA SA307382608GA SA307382614GA SA307382617GA SA308095048GA SA308095057GA SA308196539GA SA308206611GA SA308206635GA SA308206636GA SA308206649GA SA308206751GA SA308216775GA SA308297026GA SA308297028GA SA308297031GA SA308297033GA SA308297054GA SA308297055GA SA308347849GA SA308368120GA SA308388364GA SA308408978GA SA308408989GA SA308408992GA SA308408994GA SA308408995GA SA308408997GA SA308409000GA SA308409003GA SA308409005GA SA308409006GA SA308409007GA SA308409008GA SA308409095GA SA308409097GA SA308409098GA SA308409100GA SA308409103GA SA308409104GA SA308419281GA SA308469884GA SA308491128GA SA308491164GA SA308491165GA SA308491166GA SA308491176GA SA308491221GA SA308491222GA SA308491231GA SA308491232GA SA308501412GA SA308511655GA SA308511670GA SA308511673GA SA308511713GA SA308511751GA SA308511758GA SA308511759GA SA308511760GA SA308511761GA SA308511762GA SA308511766GA SA308511767GA SA308521793GA SA308521797GA SA308521808GA SA309011868GA SA309011880GA SA309021924GA SA309021925GA SA309021928GA SA309021933GA SA309021936GA SA309021937GA SA309021938GA SA309021939GA SA309021948GA SA309021953GA SA309032023GA SA309032035GA SA309032036GA SA309032038GA SA309032044GA SA309032045GA SA309032046GA SA309032047GA SA309032049GA SA309032051GA SA309032084GA SA309032085GA SA309032088GA SA309032089GA SA309032090GA SA309032093GA SA309032094GA SA309032095GA SA309032096GA SA309032097GA SA309032098GA SA309032099GA SA309032100GA SA309032101GA SA309032102GA SA309032103GA SA309032104GA SA309032107GA SA309032111GA SA309042114GA SA309042115GA SA309042117GA SA309042119GA SA309042120GA SA309042121GA SA309042124GA SA309042125GA SA309042126GA SA309042149GA SA309042151GA SA309042152GA SA309042153GA SA309042154GA SA309042155GA SA309042159GA SA309042160GA SA309042161GA SA309042167GA SA309042175GA SA309042176GA SA309042177GA SA309042178GA SA309042179GA SA309042190GA SA309042196GA SA309042199GA SA309082450GA SA309092504GA SA309112564GA SA309112575GA SA309122637GA SA309122651GA SA309122708GA SA309122709GA SA309122710GA SA309122711GA SA309122714GA SA309122716GA SA309122725GA SA309122728GA SA309122731GA SA309122734GA SA309122736GA SA309122737GA SA309122738GA SA309122739GA SA309122740GA SA309122741GA SA309122742GA SA309122743GA SA309122744GA SA309122745GA SA309122746GA SA309122747GA SA309122748GA SA309122750GA SA309122751GA SA309122753GA SA309122754GA SA309122755GA SA309122756GA SA309122758GA SA309122759GA SA309122760GA SA309122761GA SA309122762GA SA309122763GA SA309122772GA SA309122777GA SA309122780GA SA309122781GA SA309122783GA SA309122785GA SA309122786GA SA309122787GA SA309122788GA SA309122789GA SA309122790GA SA309132840GA SA309132841GA SA309132842GA SA309132843GA SA309142870GA SA309142871GA SA309142874GA SA309142889GA SA309142892GA SA309142897GA SA309142898GA SA309142899GA SA309142900GA SA309142902GA SA309142905GA SA309142908GA SA309142909GA SA309142920GA SA309173151GA SA309173159GA SA309173166GA SA309173186GA SA309173187GA SA309193301GA SA309193304GA SA309193327GA SA309193330GA SA309193341GA SA309193349GA SA309193363GA SA309203366GA SA309203370GA SA309203371GA SA309203375GA SA309203404GA SA309203420GA SA309203432GA SA309223439GA SA309223453GA SA309223478GA SA309223514GA SA309233544GA SA309233552GA SA309233553GA SA309233554GA SA309233555GA SA309233556GA SA309233557GA SA309233558GA SA309233559GA SA309233560GA SA309243623GA SA309243633GA SA309243635GA SA309243636GA SA309243637GA SA309243638GA SA309243639GA SA309243640GA SA309243641GA SA309243644GA SA309243647GA SA309243649GA SA309243656GA SA309243657GA SA309243660GA SA309243661GA SA309243662GA SA309243664GA SA309243666GA SA309243672GA SA309243676GA SA309253724GA SA309253737GA SA309253739GA SA309253740GA SA309253747GA SA309253748GA SA309253758GA SA309253762GA SA309253764GA SA309253768GA SA309253769GA SA309253770GA SA309253776GA SA309253779GA SA309253780GA SA309253783GA SA309253785GA SA309253786GA SA309253787GA SA309253788GA SA309253789GA SA309253791GA SA309253793GA SA309253794GA SA309253795GA SA309253796GA SA309253797GA SA309253798GA SA309253799GA SA309253800GA SA309253801GA SA309253802GA SA309253803GA SA309253804GA SA309253805GA SA309253806GA SA309253809GA SA309253810GA SA309253811GA SA309253813GA SA309253814GA SA309253815GA SA309253816GA SA309253817GA SA309253818GA SA309253820GA SA309253822GA SA309253824GA SA309253825GA SA309253826GA SA309253827GA SA309253829GA SA309253830GA SA309253831GA SA309253832GA SA309253835GA SA309253851GA SA309253853GA SA309253854GA SA309253858GA SA309253859GA SA309253860GA SA309253862GA SA309253872GA SA309253879GA SA309253883GA SA309253901GA SA309253902GA SA309253903GA SA309253904GA SA309253905GA SA309253907GA SA309263926GA SA309263927GA SA309263928GA SA309263931GA SA309263932GA SA309263934GA SA309263936GA SA309263942GA SA309263944GA SA309263947GA SA309263956GA SA309263959GA SA309263972GA SA309294015GA SA309294017GA SA309294018GA SA309294019GA SA309294020GA SA309294021GA SA309294022GA SA309294024GA SA309294026GA SA309294027GA SA309294032GA SA309294034GA SA309294036GA SA309294038GA SA309294040GA SA309294041GA SA309294042GA SA309294045GA SA309294046GA SA309294049GA SA309294054GA SA309294056GA SA309294059GA SA309294062GA SA309294063GA SA309294066GA SA309294070GA SA309294072GA SA309294073GA SA309294074GA SA309294075GA SA309294076GA SA309294078GA SA309294080GA SA309294081GA SA309294082GA SA309294083GA SA309294086GA SA309294088GA SA309294091GA SA309294095GA SA309294096GA SA309294097GA SA309294098GA SA309294099GA SA309294100GA SA309294101GA SA309294102GA SA309294104GA SA309294105GA SA309294106GA SA309294107GA SA309294108GA SA309294109GA SA309294110GA SA309294111GA SA309294114GA SA309294115GA SA309294118GA SA309294119GA SA309294120GA SA309294121GA SA309294122GA SA309294123GA SA309294125GA SA309294126GA SA309294127GA SA309294128GA SA309294129GA SA309294130GA SA309294131GA SA309294132GA SA309294133GA SA309294134GA SA309304135GA SA309304136GA SA309304137GA SA309304138GA SA309304139GA SA309304140GA SA309304141GA SA309304142GA SA309304143GA SA309304144GA SA309304145GA SA309304146GA SA309304147GA SA309304148GA SA309304149GA SA309304150GA SA309304151GA SA309304152GA SA309304153GA SA309304154GA SA309304155GA SA309304156GA SA309304157GA SA309304158GA SA309304159GA SA309304160GA SA309304162GA SA309304163GA SA309304164GA SA309304165GA SA309304166GA SA309304167GA SA309304168GA SA309304169GA SA309304170GA SA309304171GA SA309304173GA SA309304174GA SA309304175GA SA309304176GA SA309304177GA SA309304178GA SA309304180GA SA309304181GA SA309304182GA SA309304183GA SA309304184GA SA309304185GA SA309304186GA SA309304187GA SA309304188GA SA309304189GA SA309304190GA SA309304191GA SA309304192GA SA309304193GA SA309304194GA SA309304195GA SA309304196GA SA309304197GA SA309304198GA SA309304199GA SA309304200GA SA309304202GA SA309304203GA SA309304204GA SA309304205GA SA309304206GA SA309304207GA SA309304208GA SA309304209GA SA309304210GA SA309304211GA SA309304212GA SA309304213GA SA309304214GA SA309304215GA SA309304216GA SA309304217GA SA309304219GA SA309304223GA SA309304231GA SA309304249GA SA309304252GA SA309304261GA SA309304264GA SA309304266GA SA309304269GA SA309314290GA SA309314292GA SA309314293GA SA309314294GA SA309314295GA SA309314296GA SA309314297GA SA309314298GA SA309324299GA SA309324301GA SA309324302GA SA309324305GA SA309324306GA SA309324307GA SA309324308GA SA309324309GA SA309324311GA SA309324312GA SA309324313GA SA309324314GA SA309324315GA SA309324316GA SA309324318GA SA309324319GA SA309324320GA SA309324322GA SA309324323GA SA309324324GA SA309324325GA SA309324326GA SA309324327GA SA309324329GA SA309324330GA SA309324331GA SA309324333GA SA309324334GA SA309324335GA SA309324336GA SA309324337GA SA309324338GA SA309324340GA SA309324341GA SA309324344GA SA309324345GA SA309324347GA SA309324348GA SA309324349GA SA309324350GA SA309324351GA SA309324353GA SA309324356GA SA309324357GA SA309324360GA SA309324365GA SA309324368GA SA309324369GA SA309324371GA SA309324373GA SA309324375GA SA309324376GA SA309324377GA SA309324380GA SA309324382GA SA309324384GA SA309324385GA SA309324386GA SA309324387GA SA309324389GA SA309324391GA SA309324392GA SA309324393GA SA309324396GA SA309324399GA SA309324401GA SA309324402GA SA309324404GA SA309324408GA SA309324409GA SA309324410GA SA309324412GA SA309324414GA SA309324415GA SA309324416GA SA309324418GA SA309324426GA SA309324427GA SA309324428GA SA309324429GA SA309324430GA SA309324431GA SA309324432GA SA309324433GA SA309324439GA SA309324440GA SA309324441GA SA309324444GA SA309334556GA SA309334559GA SA309334565GA SA309334567GA SA309334568GA SA309334570GA SA309334572GA SA309334578GA SA309334580GA SA309334584GA SA309334585GA SA309334589GA SA309334595GA SA309334599GA SA309334600GA SA309334601GA SA309334603GA SA309334605GA SA309334606GA SA309334607GA SA309334608GA SA309334609GA SA309334611GA SA309334614GA SA309334615GA SA309334616GA SA309334617GA SA309334618GA SA309334619GA SA309334620GA SA309334622GA SA309334623GA SA309334624GA SA309334625GA SA309334626GA SA309334627GA SA309334628GA SA309334629GA SA309334630GA SA309334631GA SA309334632GA SA309334633GA SA309334634GA SA309334635GA SA309334636GA SA309334637GA SA309334641GA SA309334643GA SA309334644GA SA309334645GA SA309334646GA SA309334647GA SA309334648GA SA309334649GA SA309334651GA SA309334653GA SA309334654GA SA309334656GA SA309334657GA SA309334658GA SA309334659GA SA309334660GA SA309334661GA SA309334662GA SA309334663GA SA309334668GA SA309334672GA SA309334674GA SA309334676GA SA309334677GA SA309334678GA SA309334679GA SA309334680GA SA309334682GA SA309334683GA SA309334685GA SA309334687GA SA309334688GA SA309334689GA SA309334690GA SA309334692GA SA309334693GA SA309334694GA SA309334695GA SA309334696GA SA309334697GA SA309334702GA SA309334703GA SA309334707GA SA309334708GA SA309334710GA SA309334711GA SA309334712GA SA309334714GA SA309334715GA SA309334716GA SA309334717GA SA309334718GA SA309334719GA SA309334720GA SA309334723GA SA309334724GA SA309334725GA SA309334727GA SA309334729GA SA309334730GA SA309334737GA SA309334740GA SA309334741GA SA309334744GA SA309334748GA SA309334751GA SA309334752GA SA309334755GA SA309344838GA SA309344839GA SA309344842GA SA309344843GA SA309344847GA SA309344851GA SA309354915GA SA309354917GA SA309355021GA SA309365044GA SA309365048GA SA309375478GA SA309375479GA SA309375480GA SA309375481GA SA309375484GA SA309375486GA SA309375487GA SA309375489GA SA309375491GA SA309375493GA SA309375496GA SA309375497GA SA309375499GA SA309375500GA SA309375501GA SA309375504GA SA309375505GA SA309385508GA SA309385511GA SA309385512GA SA309385513GA SA309385515GA SA309385516GA SA309385517GA SA309385518GA SA309385519GA SA309385520GA SA309385521GA SA309385523GA SA309385526GA SA309385527GA SA309385529GA SA309385530GA SA309385531GA SA309385532GA SA309385533GA SA309385534GA SA309385536GA SA309385537GA SA309385539GA SA309385540GA SA309385541GA SA309385542GA SA309385545GA SA309385548GA SA309385549GA SA309385555GA SA309385558GA SA309385559GA SA309385560GA SA309385562GA SA309385563GA SA309385565GA SA309385566GA SA309385570GA SA309385571GA SA309385572GA SA309385573GA SA309385575GA SA309385577GA SA309385578GA SA309385583GA SA309385585GA SA309385586GA SA309385587GA SA309385588GA SA309385590GA SA309385591GA SA309385592GA SA309385593GA SA309385594GA SA309385596GA SA309385597GA SA309385598GA SA309385599GA SA309385600GA SA309385601GA SA309385602GA SA309385603GA SA309385604GA SA309385606GA SA309385609GA SA309385610GA SA309385612GA SA309385613GA SA309385615GA SA309385616GA SA309385618GA SA309385619GA SA309385620GA SA309385621GA SA309385622GA SA309385623GA SA309385625GA SA309385626GA SA309385629GA SA309385631GA SA309385632GA SA309385633GA SA309385634GA SA309385635GA SA309385636GA SA309385638GA SA309385639GA SA309385641GA SA309385642GA SA309385643GA SA309385645GA SA309385646GA SA309385647GA SA309385648GA SA309385649GA SA309385650GA SA309385651GA SA309385652GA SA309385653GA SA309385654GA SA309385655GA SA309385656GA SA309385657GA SA309385659GA SA309385660GA SA309385661GA SA309385663GA SA309385665GA SA309385668GA SA309385669GA SA309385671GA SA309385672GA SA309385673GA SA309385677GA SA309385679GA SA309385680GA SA309385681GA SA309385682GA SA309385683GA SA309385685GA SA309385686GA SA309385687GA SA309385688GA SA309385690GA SA309385691GA SA309385692GA SA309395722GA SA309395726GA SA309395737GA SA309395738GA SA309395739GA SA309395742GA SA309395747GA SA309395760GA SA309395786GA SA309395787GA SA309395790GA SA309395798GA SA309395799GA SA309395802GA SA309395805GA SA309395806GA SA309395807GA SA309395810GA SA309446379GA SA309446383GA SA309446386GA SA309446389GA SA309446390GA SA309446408GA SA309446412GA SA309446421GA SA309456633GA SA309456637GA SA309456638GA SA309456640GA SA309456643GA SA309456647GA SA309456649GA SA309456650GA SA309456652GA SA309456656GA SA309456659GA SA309456660GA SA309456663GA SA309456664GA SA309456666GA SA309456682GA SA309456695GA SA309456697GA SA309456701GA SA309456702GA SA309456709GA SA309456711GA SA309456712GA SA309456713GA SA309456715GA SA309517548GA SA309517840GA SA309538065GA SA309538066GA SA309538072GA SA310028136GA SA310028141GA SA310028144GA SA310028309GA SA310028310GA SA310028311GA SA310028312GA SA310028314GA SA310028318GA SA310028322GA SA310028325GA SA310028330GA SA310028334GA SA310028339GA SA310038345GA SA310038346GA SA310038482GA SA310038485GA SA310038489GA SA310038491GA SA310038498GA SA310038502GA SA310038505GA SA310038509GA SA310038511GA SA310038515GA SA310038516GA SA310038518GA SA310038528GA SA310038529GA SA310038531GA SA310038532GA SA310038536GA SA310048539GA SA310048540GA SA310048543GA SA310048545GA SA310048546GA SA310048547GA SA310048549GA SA310048551GA SA310048561GA SA310048569GA SA310048570GA SA310048573GA SA310048574GA SA310048575GA SA310048578GA SA310048581GA SA310048582GA SA310048630GA SA310048631GA SA310048632GA SA310048636GA SA310048639GA SA310048642GA SA310048650GA SA310048654GA SA310048655GA SA310048657GA SA310048659GA SA310118969GA SA310118981GA SA310118987GA SA310139279GA SA310149312GA SA310149320GA SA310149329GA SA310149335GA SA310159361GA SA310159376GA SA310159377GA SA310159379GA SA310159457GA SA310159461GA SA310159477GA SA310259898GA SA310259918GA SA310259986GA SA310260002GA SA310260037GA SA310260222GA SA310260333GA SA310260527GA SA310260556GA SA310290632GA SA310290642GA SA310290737GA SA310310841GA SA310310846GA SA310310853GA SA310310855GA SA310310878GA SA310310880GA SA310310882GA SA310310887GA SA310310888GA SA310310897GA SA310310899GA SA310310900GA SA310310913GA SA310310919GA SA310310923GA SA310310927GA SA310310928GA SA310310929GA SA310310931GA SA310310932GA SA310310934GA SA310310935GA SA310310936GA SA310310937GA SA310310938GA SA310310946GA SA310310947GA SA310310948GA SA310310949GA SA310310950GA SA310310955GA SA310310958GA SA310310960GA SA310310962GA SA310310965GA SA310310966GA SA310310967GA SA310310968GA SA310310969GA SA310310970GA SA310310971GA SA310310972GA SA310310975GA SA310310976GA SA310310977GA SA310310979GA SA310310980GA SA310310981GA SA310310982GA SA310310983GA SA310310984GA SA310310985GA SA310310986GA SA310310988GA SA310310989GA SA310310990GA SA310310991GA SA310310992GA SA310310994GA SA310310995GA SA310310996GA SA310310997GA SA310310998GA SA310310999GA SA310321000GA SA310321002GA SA310321004GA SA310321005GA SA310321006GA SA310321008GA SA310321009GA SA310321011GA SA310321012GA SA310321013GA SA310321015GA SA310321016GA SA310321017GA SA310321018GA SA310321021GA SA310321022GA SA310321025GA SA310321026GA SA310321027GA SA310321028GA SA310321029GA SA310321030GA SA310321031GA SA310321032GA SA310321033GA SA310321034GA SA310321035GA SA310321036GA SA310321037GA SA310321038GA SA310321039GA SA310321044GA SA310321045GA SA310321051GA SA310321053GA SA310321054GA SA310321058GA SA310321059GA SA310321064GA SA310321074GA SA310321077GA SA310321078GA SA310331117GA SA310331118GA SA310331126GA SA310331127GA SA310361632GA SA310361636GA SA310361638GA SA310361888GA SA310361889GA SA310361904GA SA310361909GA SA310361910GA SA310403441GA SA310403449GA SA310413452GA SA310413460GA SA310413461GA SA310413467GA SA310413468GA SA310413470GA SA310413472GA SA310413474GA SA310413475GA SA310413477GA SA310413478GA SA310413480GA SA310413482GA SA310413483GA SA310413484GA SA310413487GA SA310413488GA SA310413489GA SA310413490GA SA310413492GA SA310413494GA SA310413496GA SA310413501GA SA310413502GA SA310413503GA SA310413506GA SA310413510GA SA310413512GA SA310413513GA SA310413515GA SA310413519GA SA310413520GA SA310413525GA SA310413527GA SA310413528GA SA310413531GA SA310413532GA SA310413533GA SA310413539GA SA310413540GA SA310413546GA SA310413547GA SA310413550GA SA310413552GA SA310413553GA SA310413554GA SA310413562GA SA310413566GA SA310413567GA SA310413568GA SA310413569GA SA310423574GA SA310423575GA SA310423578GA SA310423579GA SA310423581GA SA310423582GA SA310423590GA SA310423594GA SA310423597GA SA310423599GA SA310423601GA SA310423603GA SA310423605GA SA310423612GA SA310423620GA SA310423624GA SA310423625GA SA310423629GA SA310423630GA SA310423631GA SA310423632GA SA310423633GA SA310423634GA SA310423636GA SA310423638GA SA310423641GA SA310423645GA SA310423647GA SA310423660GA SA310423724GA SA310423725GA SA310423728GA SA310423729GA SA310423731GA SA310423734GA SA310423735GA SA310423823GA SA310423828GA SA310423830GA SA310423832GA SA310423833GA SA310423844GA SA310423848GA SA310423849GA SA310423852GA SA310423854GA SA310423855GA SA310423857GA SA310423858GA SA310423859GA SA310423860GA SA310423861GA SA310423863GA SA310423865GA SA310423920GA SA310423925GA SA310423931GA SA310423932GA SA310423934GA SA310423936GA SA310423938GA SA310423939GA SA310423990GA SA310423997GA SA310423999GA SA310434055GA SA310434075GA SA310434103GA SA310434133GA SA310434141GA SA310434171GA SA310434206GA SA310434214GA SA310434215GA SA310434216GA SA310434217GA SA310434218GA SA310434219GA SA310434220GA SA310434221GA SA310434222GA SA310434223GA SA310434224GA SA310434225GA SA310434228GA SA310434229GA SA310434231GA SA310434232GA SA310434233GA SA310434234GA SA310434235GA SA310434236GA SA310434237GA SA310434239GA SA310434240GA SA310434241GA SA310434243GA SA310434244GA SA310434245GA SA310444246GA SA310444249GA SA310444250GA SA310444252GA SA310444254GA SA310444255GA SA310444257GA SA310444260GA SA310444261GA SA310444263GA SA310444264GA SA310444265GA SA310444276GA SA310444281GA SA310444283GA SA310444284GA SA310444305GA SA310444308GA SA310455082GA SA310455087GA SA310455107GA SA310455145GA SA310455165GA SA310455172GA SA310455175GA SA310455283GA SA310465304GA SA310465308GA SA310465324GA SA310465335GA SA310465337GA SA310465348GA SA310465349GA SA310465354GA SA310465368GA SA310465369GA SA310465370GA SA310465374GA SA310465385GA SA310465391GA SA310465400GA SA310465408GA SA310465414GA SA310465423GA SA310465437GA SA310496126GA SA310517815GA SA310517873GA SA310518277GA SA310518283GA SA310518378GA SA310518420GA SA310518426GA SA310518427GA SA310528809GA SA310529148GA SA310529153GA SA310539348GA SA311039433GA SA311060002GA SA311060009GA SA311060202GA SA311060221GA SA311060266GA SA311060273GA SA311060277GA SA311069690GA SA311069700GA SA311069703GA SA311090911GA SA311090935GA SA311090949GA SA311090955GA SA311090968GA SA311090972GA SA311090979GA SA311090992GA SA311090995GA SA311090999GA SA311091040GA SA311101072GA SA311101073GA SA311101075GA SA311101092GA SA311101100GA SA311101103GA SA311101144GA SA311101153GA SA311101156GA SA311101159GA SA311101160GA SA311101299GA SA311111649GA SA311121841GA SA311121846GA SA311121971GA SA311121973GA SA311121979GA SA311121982GA SA311121986GA SA311121987GA SA311121989GA SA311121993GA SA311121994GA SA311121997GA SA311121998GA SA311122002GA SA311122005GA SA311122009GA SA311122010GA SA311122066GA SA311122071GA SA311122073GA SA311122093GA SA311122094GA SA311122101GA SA311122102GA SA311122111GA SA311122112GA SA311122115GA SA311132419GA SA311132420GA SA311132426GA SA311132429GA SA311132431GA SA311132437GA SA311132445GA SA311132446GA SA311132450GA SA311132462GA SA311132485GA SA311132486GA SA311132487GA SA311132513GA SA311132520GA SA311132528GA SA311132543GA SA311132605GA SA311132613GA SA311132614GA SA311132619GA SA311142677GA SA311142680GA SA311142683GA SA311142684GA SA311142688GA SA311142693GA SA311142698GA SA311142703GA SA311142729GA SA311142737GA SA311142741GA SA311142742GA SA311142744GA SA311142797GA SA311142855GA SA311143011GA SA311143017GA SA311143018GA SA311143020GA SA311153030GA SA311153038GA SA311153053GA SA311153232GA SA311153239GA SA311153268GA SA311153283GA SA311153284GA SA311153289GA SA311153306GA SA311153314GA SA311153322GA SA311163432GA SA311163434GA SA311163444GA SA311163451GA SA311163456GA SA311163462GA SA311163463GA SA311163465GA SA311163471GA SA311163472GA SA311163474GA SA311163475GA SA311163476GA SA311163477GA SA311163478GA SA311163481GA SA311163482GA SA311163486GA SA311163534GA SA311163536GA SA311163537GA SA311163538GA SA311163540GA SA311163544GA SA311163546GA SA311163548GA SA311163549GA SA311163554GA SA311163555GA SA311163558GA SA311163564GA SA311163581GA SA311163589GA SA311163591GA SA311163595GA SA311163614GA SA311163621GA SA311163636GA SA311163655GA SA311163702GA SA311163704GA SA311163712GA SA311163714GA SA311163717GA SA311163720GA SA311163722GA SA311163732GA SA311163780GA SA311163791GA SA311163803GA SA311163806GA SA311163811GA SA311163820GA SA311163829GA SA311163837GA SA311163839GA SA311163841GA SA311163848GA SA311163853GA SA311173999GA SA311174000GA SA311174003GA SA311174005GA SA311174006GA SA311174015GA SA311174017GA SA311174018GA SA311174021GA SA311174033GA SA311174044GA SA311184716GA SA311184843GA SA311184845GA SA311194890GA SA311194911GA SA311194924GA SA311205426GA SA311205437GA SA311246680GA SA311287204GA SA311297346GA SA311307515GA SA311307516GA SA311307517GA SA311307518GA SA311307520GA SA311307522GA SA311307525GA SA311307526GA SA311307527GA SA311307528GA SA311307533GA SA311307534GA SA311307535GA SA311307537GA SA311307595GA SA311317660GA SA311317668GA SA311317737GA SA311327823GA SA311327825GA SA311327874GA SA311448182GA SA311448183GA SA311448184GA SA311448185GA SA311448186GA SA311448187GA SA311448191GA SA311458215GA SA311458216GA SA311458227GA SA311458228GA SA311458232GA SA311458233GA SA311458235GA SA311458238GA SA311458242GA SA311458246GA SA311458247GA SA311458248GA SA311458250GA SA311458251GA SA311458254GA SA311458255GA SA311458261GA SA311458264GA SA311458265GA SA311458266GA SA311468387GA SA311508577GA SA311518698GA SA311518711GA SA311518725GA SA312018873GA SA312018876GA SA312018877GA SA312018878GA SA312018879GA SA312018880GA SA312018885GA SA312018894GA SA312028912GA SA312028913GA SA312028920GA SA312028923GA SA312028924GA SA312028925GA SA312028930GA SA312028931GA SA312028934GA SA312028937GA SA312028939GA SA312028942GA SA312028943GA SA312028944GA SA312028945GA SA312028946GA SA312028947GA SA312028951GA SA312028952GA SA312028953GA SA312028954GA SA312028955GA SA312028956GA SA312028957GA SA312028958GA SA312028959GA SA312028960GA SA312028961GA SA312028962GA SA312028963GA SA312028964GA SA312028965GA SA312028968GA SA312028969GA SA312028970GA SA312028971GA SA312028972GA SA312028973GA SA312028974GA SA312028975GA SA312028976GA SA312028977GA SA312028981GA SA312028983GA SA312028985GA SA312029000GA SA312029013GA SA312029029GA SA312029037GA SA312029041GA SA312029042GA SA312029044GA SA312029049GA SA312029070GA SA312069784GA SA312070277GA SA312079978GA SA312080870GA SA312080879GA SA312090918GA SA312090919GA SA312090920GA SA312090922GA SA312090924GA SA312090925GA SA312090926GA SA312090929GA SA312090930GA SA312090932GA SA312090933GA SA312090935GA SA312090936GA SA312090937GA SA312090938GA SA312090939GA SA312090940GA SA312090941GA SA312090943GA SA312090944GA SA312090945GA SA312090946GA SA312090947GA SA312090949GA SA312090950GA SA312090951GA SA312090952GA SA312090953GA SA312090954GA SA312090955GA SA312090956GA SA312090957GA SA312090958GA SA312090959GA SA312090960GA SA312090961GA SA312090962GA SA312090963GA SA312090964GA SA312090965GA SA312090968GA SA312090971GA SA312090973GA SA312090975GA SA312090976GA SA312090977GA SA312090979GA SA312090980GA SA312090984GA SA312090986GA SA312090991GA SA312090992GA SA312091005GA SA312091008GA SA312091012GA SA312091016GA SA312101432GA SA312111464GA SA312111474GA SA312111475GA SA312111477GA SA312111478GA SA312111480GA SA312111481GA SA312111490GA SA312111494GA SA312111496GA SA312111497GA SA312111498GA SA312111501GA SA312111592GA SA312111593GA SA312111594GA SA312111595GA SA312111596GA SA312111597GA SA312111599GA SA312111600GA SA312111601GA SA312111602GA SA312111603GA SA312111604GA SA312111605GA SA312111606GA SA312111607GA SA312111608GA SA312111609GA SA312111610GA SA312111611GA SA312111612GA SA312111613GA SA312111614GA SA312111615GA SA312111616GA SA312111617GA SA312111618GA SA312111619GA SA312111620GA SA312111621GA SA312111622GA SA312111623GA SA312111624GA SA312111625GA SA312111626GA SA312111627GA SA312111628GA SA312111634GA SA312111635GA SA312111637GA SA312111638GA SA312111639GA SA312111641GA SA312242968GA SA312253295GA SA312263874GA SA312263891GA SA312263898GA SA312263909GA SA312263929GA SA312263932GA SA312263941GA SA312263942GA SA312263945GA SA312263947GA SA312294374GA SA312294387GA SA312294418GA SA312294421GA SA312294423GA SA312294434GA SA312294455GA SA312294466GA SA312294470GA SA312294472GA SA312294495GA SA312294508GA SA312294509GA SA312294511GA SA312294512GA SA312294513GA SA312294514GA SA312294515GA SA312294521GA SA312294523GA SA312294534GA SA312294536GA SA312294537GA SA312294538GA SA312294539GA SA312294540GA SA312294541GA SA312294542GA SA312294543GA SA312294545GA SA312294547GA SA312294549GA SA312294550GA SA312294553GA SA312294554GA SA312294557GA SA312294562GA SA312294563GA SA312294565GA SA312294567GA SA312294572GA SA312294573GA SA312294574GA SA312294575GA SA312294576GA SA312294577GA SA312294580GA SA312294583GA SA312294585GA SA312294596GA SA312294610GA SA312294620GA SA312294624GA SA312294629GA SA312294633GA SA312294637GA SA312294638GA SA312294639GA SA312294640GA SA312294643GA SA312294645GA SA312294647GA SA312294648GA SA312294652GA SA312294668GA SA312355479GA SA312376200GA SA312396705GA SA312396722GA SA312396736GA SA312396740GA SA312396766GA SA312396818GA SA312396868GA SA312396895GA SA312396958GA SA312397061GA SA312397081GA SA312397113GA SA312397116GA SA312397117GA SA312397128GA SA312397132GA SA312397134GA SA312397148GA SA312397150GA SA312397203GA SA312397211GA SA312397212GA SA312397213GA SA312397214GA SA312397215GA SA312397217GA SA312397218GA SA312397219GA SA312397220GA SA312397222GA SA312397223GA SA312397225GA SA312397226GA SA312397227GA SA312397228GA SA312397229GA SA312397230GA SA312397231GA SA312397232GA SA312397233GA SA312397235GA SA312397236GA SA312397237GA SA312397238GA SA312397239GA SA312397240GA SA312397242GA SA312397243GA SA312397244GA SA312397245GA SA312397247GA SA312397248GA SA312397249GA SA312397250GA SA312397253GA SA312397255GA SA312397256GA SA312397258GA SA312397259GA SA312397260GA SA312397261GA SA312397262GA SA312397263GA SA312397264GA SA312397265GA SA312397266GA SA312397267GA SA312397268GA SA312397270GA SA312407295GA SA312417302GA SA312417303GA SA312417304GA SA312417305GA SA312417307GA SA312417308GA SA312417309GA SA312417310GA SA312417312GA SA312417314GA SA312417315GA SA312417319GA SA312417322GA SA312417324GA SA312417325GA SA312417326GA SA312417327GA SA312417331GA SA312417333GA SA312417338GA SA312417345GA SA312417350GA SA312417352GA SA312417353GA SA312417354GA SA312417356GA SA312417357GA SA312417358GA SA312417360GA SA312417361GA SA312417363GA SA312417370GA SA312417376GA SA312417377GA SA312417381GA SA312417383GA SA312417388GA SA312417389GA SA312417394GA SA312417396GA SA312417411GA SA312417413GA SA312417414GA SA312417415GA SA312417416GA SA312458083GA SA312458128GA SA312458129GA SA312458131GA SA312458132GA SA312458133GA SA312458134GA SA312458135GA SA312458136GA SA312458137GA SA312458140GA SA312458141GA SA312458142GA SA312458143GA SA312458144GA SA312458146GA SA312458148GA SA312458149GA SA312458150GA SA312458152GA SA312458153GA SA312458154GA SA312458155GA SA312458158GA SA312458159GA SA312458161GA SA312458162GA SA312458163GA SA312458164GA SA312458165GA SA312458166GA SA312458169GA SA312458170GA SA312458171GA SA312458172GA SA312458173GA SA312458174GA SA312458175GA SA312458177GA SA312468605GA SA312468607GA SA312468613GA SA312468655GA SA312468662GA SA312468663GA SA312468670GA SA312468671GA SA312468674GA SA312468692GA SA312468697GA SA312468699GA SA312468700GA SA312468705GA SA312468706GA SA312468717GA SA312468718GA SA312468808GA SA312468812GA SA312468827GA SA312468829GA SA312468833GA SA312468837GA SA312468839GA SA312468840GA SA312468845GA SA312468846GA SA312468865GA SA312468868GA SA312468870GA SA312468873GA SA312468874GA SA312468875GA SA312468876GA SA312468879GA SA312468881GA SA312468882GA SA312468883GA SA312468884GA SA312468885GA SA312468886GA SA312468887GA SA312468888GA SA312468889GA SA312468892GA SA312468893GA SA312468895GA SA312468896GA SA312468897GA SA312468898GA SA312468899GA SA312468903GA SA312468904GA SA312468905GA SA312468906GA SA312468908GA SA312468909GA SA312468911GA SA312468912GA SA312468913GA SA312468914GA SA312468915GA SA312468916GA SA312468917GA SA312468918GA SA312468920GA SA312468921GA SA312468928GA SA312468939GA SA312468942GA SA312468943GA SA312468948GA SA312468949GA SA312469012GA SA312469074GA SA312469228GA SA312469230GA SA312469231GA SA312469233GA SA312469234GA SA312469235GA SA312469236GA SA312469237GA SA312469238GA SA312469239GA SA312469240GA SA312469241GA SA312469242GA SA312469245GA SA312469247GA SA312469255GA SA312469256GA SA312469259GA SA312469269GA SA312469270GA SA312469271GA SA312469272GA SA312469273GA SA312469275GA SA312469276GA SA312469277GA SA312469278GA SA312469372GA SA312469376GA SA312469378GA SA312469409GA SA312469458GA SA312469465GA SA312469472GA SA312479718GA SA312479740GA SA312479815GA SA312522431GA SA313012810GA SA313012842GA SA313012960GA SA313013091GA SA313013092GA SA313013093GA SA313013095GA SA313013108GA SA313013126GA SA313013134GA SA313023201GA SA313023342GA SA313023348GA SA313023353GA SA313023376GA SA313023380GA SA313023416GA SA313084905GA SA313084938GA SA313084939GA SA313084941GA SA313095080GA SA313106063GA SA313106093GA SA313116449GA SA313127676GA SA313127677GA SA313127678GA SA313127679GA SA313127680GA SA313127681GA SA313127683GA SA313127688GA SA313127691GA SA313127692GA SA313127723GA SA313137868GA SA313147948GA SA313147951GA SA313147969GA SA313147985GA SA313198205GA SA313198228GA SA313198230GA SA313198327GA SA313198339GA SA313198344GA SA313198350GA SA313198405GA SA313198681GA SA313198682GA SA313198687GA SA313198688GA SA313198691GA SA313208914GA SA313208915GA SA313208918GA SA313208919GA SA313208942GA SA313208949GA SA313208950GA SA313261834GA SA313261902GA SA313261905GA SA313272049GA SA313314593GA SA313314608GA SA313335427GA SA313378287GA SA313378330GA SA313378476GA SA313420130GA SA313420131GA SA313420142GA SA313420149GA SA313420179GA SA313420230GA SA313420239GA SA313420241GA SA313420242GA SA313420245GA SA313420256GA SA313420260GA SA313420264GA SA313420301GA SA313420302GA SA313441309GA SA313441313GA SA313441515GA SA313441516GA SA313441525GA SA313441530GA SA313441546GA SA313441550GA SA313441557GA SA313441629GA SA313441633GA SA313441692GA SA313441693GA SA313513743GA SA313513744GA SA313513814GA SA313513850GA SA313513862GA SA313513880GA SA313513897GA SA313513899GA SA313513917GA SA313513920GA SA313513930GA SA313513931GA SA313513932GA SA313513933GA SA313513934GA SA313513935GA SA313513941GA SA313513942GA SA313523946GA SA313523949GA SA313523951GA SA313523952GA SA313523955GA SA314075977GA SA314075978GA SA314075992GA SA314076012GA SA314076014GA SA314076025GA SA314076028GA SA314076029GA SA314076071GA SA314076079GA SA314076083GA SA314076085GA SA314076086GA SA314076088GA SA314076102GA SA314076106GA SA314076107GA SA314076123GA SA314076124GA SA314076125GA SA314076132GA SA314076136GA SA314076138GA SA314076139GA SA314076145GA SA314076149GA SA314076151GA SA314076152GA SA314076153GA SA314076155GA SA314076159GA SA314076172GA SA314076178GA SA314076185GA SA314076204GA SA314076206GA SA314086543GA SA314086557GA SA314086643GA SA314086659GA SA314086709GA SA314096751GA SA314097067GA SA314107413GA SA314107419GA SA314107435GA SA314107485GA SA314107568GA SA314107577GA SA314107587GA SA314107662GA SA314108043GA SA314118494GA SA314118732GA SA314118737GA SA314118741GA SA314118742GA SA314118745GA SA314118751GA SA314118814GA SA314118821GA SA314118838GA SA314118867GA SA314118875GA SA314118949GA SA314118964GA SA314118971GA SA314118978GA SA314118998GA SA314119021GA SA314119033GA SA314119078GA SA314119082GA SA314119086GA SA314119089GA SA314119095GA SA314119096GA SA314119097GA SA314119098GA SA314119104GA SA314119105GA SA314119115GA SA314119118GA SA314119124GA SA314119127GA SA314119129GA SA314119131GA SA314119132GA SA314119134GA SA314129166GA SA314129167GA SA314129169GA SA314129170GA SA314129175GA SA314129177GA SA314129178GA SA314129179GA SA314129180GA SA314129181GA SA314129183GA SA314129185GA SA314129186GA SA314129187GA SA314129188GA SA314129189GA SA314129196GA SA314129197GA SA314129198GA SA314129200GA SA314129201GA SA314129202GA SA314129204GA SA314129206GA SA314129210GA SA314129211GA SA314129213GA SA314129214GA SA314129218GA SA314129219GA SA314129220GA SA314129222GA SA314129225GA SA314129258GA SA314129261GA SA314129262GA SA314129264GA SA314129268GA SA314129269GA SA314129272GA SA314129277GA SA314129278GA SA314129279GA SA314129281GA SA314129283GA SA314129290GA SA314129291GA SA314129300GA SA314129305GA SA314129313GA SA314129314GA SA314139346GA SA314139355GA SA314139357GA SA314139358GA SA314139361GA SA314139372GA SA314139374GA SA314139391GA SA314139392GA SA314139394GA SA314139395GA SA314139396GA SA314139399GA SA314139400GA SA314139403GA SA314139404GA SA314139405GA SA314139406GA SA314139408GA SA314139409GA SA314139410GA SA314139413GA SA314139415GA SA314139416GA SA314139423GA SA314139425GA SA314139428GA SA314139430GA SA314139432GA SA314139433GA SA314139435GA SA314139448GA SA314139449GA SA314139452GA SA314139453GA SA314139454GA SA314139455GA SA314139457GA SA314139459GA SA314139460GA SA314139461GA SA314139463GA SA314139465GA SA314139467GA SA314139469GA SA314139483GA SA314139499GA SA314139502GA SA314139506GA SA314139507GA SA314139508GA SA314139512GA SA314139513GA SA314139515GA SA314139516GA SA314139518GA SA314139525GA SA314139527GA SA314139528GA SA314139530GA SA314139533GA SA314139534GA SA314139541GA SA314139543GA SA314139544GA SA314139545GA SA314139547GA SA314139548GA SA314139552GA SA314139562GA SA314139571GA SA314139580GA SA314139582GA SA314139584GA SA314139588GA SA314139595GA SA314139600GA SA314139601GA SA314149602GA SA314149603GA SA314149604GA SA314149605GA SA314149606GA SA314149607GA SA314149608GA SA314149609GA SA314149612GA SA314149614GA SA314149616GA SA314149632GA SA314149633GA SA314149635GA SA314149636GA SA314149637GA SA314149638GA SA314149639GA SA314149640GA SA314149641GA SA314149642GA SA314149643GA SA314149644GA SA314149645GA SA314149646GA SA314149647GA SA314149648GA SA314149650GA SA314149651GA SA314149652GA SA314149653GA SA314149654GA SA314149656GA SA314149657GA SA314149658GA SA314149659GA SA314180509GA SA314180533GA SA314180583GA SA314180591GA SA314180607GA SA314190906GA SA314190964GA SA314201593GA SA314201598GA SA314201610GA SA314201620GA SA314201759GA SA314201817GA SA314212188GA SA314212204GA SA314212211GA SA314212212GA SA314212239GA SA314212240GA SA314212254GA SA314222283GA SA314222564GA SA314222590GA SA314222622GA SA314222647GA SA314273939GA SA314273988GA SA314273997GA SA314274107GA SA314274178GA SA314274214GA SA312376284GA SA312376308GA SA312386624GA SA312397010GA SA312397015GA SA312397063GA SA312397073GA SA312397079GA SA312397090GA SA312397158GA SA312397170GA SA308501330GA SA308501345GA SA308501348GA SA308501352GA SA308501383GA SA310507217GA SA311246454GA SA311246482GA SA311246503GA SA311256803GA SA311256836GA SA311256857GA SA311256865GA SA311317761GA SA312501604GA SA313012751GA SA313012816GA SA313023316GA SA313023496GA SA313033583GA SA313043951GA SA313043952GA SA313043953GA SA313043954GA SA313043955GA SA309446497GA SA309466779GA SA309466786GA SA309466787GA SA309466796GA SA309466798GA SA310038397GA SA310260545GA SA310290693GA SA310321003GA SA310361608GA SA310372044GA SA310372070GA SA310372142GA SA310372150GA SA310372153GA SA310372162GA SA310372296GA SA310382375GA SA310382376GA SA310382377GA SA310382379GA SA310382380GA SA310382381GA SA310382383GA SA310382388GA SA310382395GA SA310382397GA SA310382399GA SA310382400GA SA310382401GA SA310382402GA SA310382403GA SA310382404GA SA310382405GA SA310382406GA SA310382407GA SA310382409GA SA310382410GA SA310382411GA SA310382412GA SA310382414GA SA310382416GA SA310382417GA SA310382419GA SA310382422GA SA310382423GA SA310382426GA SA310382430GA SA310382431GA SA310382434GA SA310382435GA SA310382437GA SA310382440GA SA310382443GA SA310382444GA SA310382451GA SA310382452GA SA310382453GA SA310382455GA SA310382459GA SA310382460GA SA310382461GA SA310382462GA SA310382463GA SA310382464GA SA311184370GA SA311184525GA SA311184557GA SA311184613GA SA312253387GA SA312253392GA SA312253397GA SA312253403GA SA312253405GA SA312253406GA SA312253411GA SA312253416GA SA312253417GA SA312253422GA SA312253423GA SA312253427GA SA312253428GA SA312253429GA SA312253430GA SA312253431GA SA312253432GA SA312253446GA SA312253448GA SA312253451GA SA312253452GA SA312253453GA SA312253455GA SA312253457GA SA312253458GA SA312253462GA SA312253471GA SA312253473GA SA312253487GA SA312253516GA SA312376203GA SA312386316GA SA313084756GA SA313084814GA SA314253638GA SA314253746GA SA309477038GA SA310260029GA SA310290644GA SA310300772GA SA310310839GA SA310310881GA SA310310945GA SA310403343GA SA310403353GA SA310403431GA SA310403439GA SA310423795GA SA310423876GA SA308286943GA SA308388363GA SA308409129GA SA309042129GA SA309042144GA SA310372132GA SA310372171GA SA310372205GA SA310382533GA SA310382600GA SA310382785GA SA310382826GA SA310382850GA SA310403381GA SA310403389GA SA310403402GA SA310403407GA SA310423784GA SA310423788GA SA310423868GA SA310423878GA SA310434067GA SA310434096GA SA310434152GA SA310455115GA SA310455236GA SA310455260GA SA310455271GA SA310465297GA SA310465306GA SA310465311GA SA310465316GA SA310496504GA SA311256818GA SA311287017GA SA311287018GA SA311287035GA SA311287043GA SA311287052GA SA311287081GA SA311287086GA SA311287091GA SA311287096GA SA312263966GA SA312264011GA SA312264057GA SA312264099GA SA312345041GA SA312386315GA SA312386317GA SA312386329GA SA312386332GA SA312386334GA SA312386370GA SA312386672GA SA312386674GA SA312386675GA SA312386688GA SA312386690GA SA312386691GA SA312386697GA SA312386698GA SA312386699GA SA312480633GA SA312480836GA SA312501956GA SA312512187GA SA312396976GA SA312396978GA SA312396979GA SA312396982GA SA312396984GA SA312396998GA SA312397151GA SA312397152GA SA312397153GA SA312397154GA SA312397156GA SA312397162GA SA312397166GA SA312397179GA SA312397188GA SA312397189GA SA312397192GA SA312397193GA SA312397194GA SA312397199GA SA312397200GA SA312397208GA SA310507120GA SA310507487GA SA311080633GA SA312069695GA SA312069723GA SA312069822GA SA312069841GA SA312070003GA SA312070012GA SA312070099GA SA312079954GA SA312079969GA SA312079997GA SA312091152GA SA312101398GA SA312101400GA SA312101403GA SA312111459GA SA312192670GA SA312192680GA SA312192683GA SA312192686GA SA312192697GA SA312192698GA SA312192701GA SA312192732GA SA312253059GA SA312253066GA SA312253121GA SA312253131GA SA312253299GA SA312263926GA SA312263928GA SA312263979GA SA312284337GA SA312294446GA SA312294627GA SA312294792GA SA312386679GA SA312396810GA SA312396827GA SA312396834GA SA312396837GA SA312396842GA SA312396857GA SA312396858GA SA312396864GA SA312396871GA SA312396883GA SA312396886GA SA312396889GA SA312396892GA SA312396894GA SA312396898GA SA312396902GA SA312396904GA SA312396905GA SA313084956GA SA313084965GA SA313095034GA SA313095089GA SA313095120GA SA313095259GA SA313095564GA SA313377759GA SA313419786GA SA313419838GA SA313419943GA SA313420113GA SA313441494GA SA313482616GA SA313482628GA SA313482640GA SA313482665GA SA313482673GA SA313482681GA SA313482697GA SA313482702GA SA313482705GA SA313482774GA SA314065744GA SA314065749GA SA314075842GA SA314211892GA SA314211905GA SA314211908GA SA314211928GA SA314211936GA SA314211944GA SA314211946GA SA314212012GA SA314222354GA SA314222382GA SA314222389GA SA314222392GA SA314222394GA SA314222397GA SA310260487GA SA310393002GA SA310424029GA SA310444262GA SA310455003GA SA310507086GA SA310507479GA SA310517577GA SA310518683GA SA310528894GA SA310529145GA SA311039390GA SA311060161GA SA311060162GA SA311060170GA SA311060197GA SA311060269GA SA311069722GA SA311069734GA SA311122057GA SA311142747GA SA311142934GA SA311153086GA SA311153152GA SA311153156GA SA311153159GA SA311153166GA SA311153171GA SA311153179GA SA311153180GA SA311153191GA SA311153199GA SA311153200GA SA311153201GA SA311153204GA SA311163372GA SA311163385GA SA311163414GA SA311163422GA SA311163424GA SA311163425GA SA311163426GA SA311184528GA SA311184594GA SA311184636GA SA311184664GA SA311256701GA SA311256705GA SA311266875GA SA311266891GA SA311266893GA SA311266994GA SA311297360GA SA311307513GA SA311307519GA SA311307521GA SA311307530GA SA311317645GA SA311317646GA SA311317655GA SA311327812GA SA311327936GA SA311327942GA SA311458234GA SA311458340GA SA311458346GA SA312018749GA SA312018750GA SA312018760GA SA312018761GA SA312018766GA SA312018770GA SA312018772GA SA312018901GA SA312039074GA SA312039202GA SA312039204GA SA312232829GA SA312232833GA SA312232846GA SA312355230GA SA312386469GA SA312386480GA SA312386481GA SA312386482GA SA312386489GA SA312386535GA SA312386685GA SA312396928GA SA312396930GA SA312397274GA SA312468691GA SA312469445GA SA312470004GA SA312470014GA SA312470015GA SA312470017GA SA312470018GA SA312470033GA SA312470037GA SA312470070GA SA312470073GA SA312470088GA SA312479542GA SA312479784GA SA312479926GA SA312479955GA SA312479957GA SA312479962GA SA312479965GA SA312480096GA SA312480113GA SA312480115GA SA312491168GA SA312491225GA SA312491227GA SA312491228GA SA312491232GA SA312491237GA SA312491244GA SA312491373GA SA312491537GA SA312501613GA SA312501620GA SA312501622GA SA312501635GA SA312501638GA SA312501643GA SA312501646GA SA312501647GA SA312501650GA SA312502085GA SA312502135GA SA312502139GA SA312502161GA SA312512309GA SA312522497GA SA312522586GA SA313012865GA SA313064238GA SA312069671GA SA312069673GA SA312069682GA SA313023398GA SA313043944GA SA313043946GA SA313043950GA SA313043956GA SA313043958GA SA313043960GA SA313043971GA SA313043973GA SA314284592GA SA310475451GA SA310475460GA SA310475530GA SA310475532GA SA310475534GA SA311184578GA SA312294401GA SA312294411GA SA312294442GA SA312294480GA SA312294486GA SA312294504GA SA312294533GA SA312294555GA SA312355437GA SA312355438GA SA312355452GA SA312355454GA SA312355457GA SA312355458GA SA312355465GA SA312355469GA SA312376077GA SA312376091GA SA312376282GA SA312376283GA SA312376285GA SA312376288GA SA312376290GA SA312376291GA SA312376296GA SA312376298GA SA312376301GA SA312376302GA SA312376304GA SA312376305GA SA312376306GA SA312376309GA SA312386491GA SA312386502GA SA312386506GA SA312386511GA SA312386515GA SA312386519GA SA312386523GA SA312386532GA SA312386534GA SA312386547GA SA312080621GA SA312522440GA SA312522473GA SA312522476GA SA312522482GA SA312522483GA SA312522488GA SA312522503GA SA312522504GA SA312522524GA SA312522537GA SA312522556GA SA312522560GA SA312522561GA SA313398818GA SA313451832GA SA314222891GA SA307241679GA SA307301985GA SA307483748GA SA307483781GA SA307483785GA SA307483787GA SA307483799GA SA307483842GA SA307493844GA SA307504188GA SA307504195GA SA307504212GA SA307504214GA SA307504232GA SA307514235GA SA307514238GA SA307514245GA SA307514248GA SA307514256GA SA307514276GA SA307514277GA SA307514289GA SA307514290GA SA307514302GA SA307514311GA SA307514312GA SA307514316GA SA307514338GA SA307514343GA SA307514346GA SA307514349GA SA307514390GA SA307514397GA SA307514398GA SA307514404GA SA307514405GA SA307514417GA SA307514431GA SA307514432GA SA307514435GA SA307514444GA SA307514446GA SA307514450GA SA307514451GA SA307524470GA SA308044501GA SA308044525GA SA308084649GA SA308084741GA SA308105214GA SA308105223GA SA308115234GA SA308115235GA SA308115237GA SA308115242GA SA308115248GA SA308115250GA SA308115252GA SA308115255GA SA308115256GA SA308115257GA SA308115258GA SA308115262GA SA308115263GA SA308115264GA SA308115266GA SA308115267GA SA308115270GA SA308115271GA SA308115272GA SA308115273GA SA308115274GA SA308115275GA SA308115279GA SA308115280GA SA308115283GA SA308115284GA SA308115285GA SA308115286GA SA308125309GA SA308125310GA SA308125311GA SA308125313GA SA308125315GA SA308125320GA SA308125321GA SA308125322GA SA308125323GA SA308125329GA SA308125331GA SA308125334GA SA308125336GA SA308125337GA SA308125346GA SA308125347GA SA308125353GA SA308125357GA SA308125358GA SA308125359GA SA308125361GA SA308125363GA SA308125367GA SA308125368GA SA308125370GA SA308125372GA SA308125373GA SA308135516GA SA308135517GA SA308135520GA SA308135522GA SA308135525GA SA308135526GA SA308135529GA SA308135530GA SA308135531GA SA308135532GA SA308135535GA SA308135536GA SA308135537GA SA308135538GA SA308135539GA SA308135540GA SA308135542GA SA308145656GA SA308145658GA SA308145660GA SA308145663GA SA308145664GA SA308145665GA SA308145666GA SA308145667GA SA308145669GA SA308145670GA SA308145672GA SA308145673GA SA308145676GA SA308145678GA SA308145679GA SA308145682GA SA308145683GA SA308145691GA SA308155971GA SA308166056GA SA308166070GA SA308186393GA SA308196558GA SA308206603GA SA308216767GA SA308297032GA SA308409048GA SA308409054GA SA308409077GA SA308429355GA SA308439450GA SA308449509GA SA308449527GA SA308449573GA SA308449578GA SA308449580GA SA308449582GA SA308449584GA SA308449586GA SA308449587GA SA308449592GA SA308459646GA SA308459660GA SA308459662GA SA308459663GA SA308459664GA SA308459680GA SA308459681GA SA308491213GA SA308491216GA SA308511756GA SA309042194GA SA309122660GA SA309122722GA SA309142983GA SA309142987GA SA309142988GA SA309142990GA SA309142991GA SA309153026GA SA309153028GA SA309153030GA SA309153031GA SA309153036GA SA309153037GA SA309153039GA SA309153040GA SA309153041GA SA309153043GA SA309153044GA SA309153049GA SA309153050GA SA309153051GA SA309153052GA SA309153054GA SA309153057GA SA309153060GA SA309153062GA SA309153063GA SA309153070GA SA309153074GA SA309153077GA SA309153078GA SA309153085GA SA309173106GA SA309173162GA SA309193298GA SA309193299GA SA309263916GA SA310048669GA SA310058695GA SA310465361GA SA310506657GA SA310506661GA SA310506678GA SA310506702GA SA310506728GA SA310506737GA SA310506846GA SA310506862GA SA310506864GA SA310506865GA SA310506871GA SA310506872GA SA310506875GA SA310506879GA SA310506880GA SA310506892GA SA310506893GA SA310506894GA SA310506896GA SA310506897GA SA310506992GA SA310506993GA SA310506997GA SA310506998GA SA310507001GA SA310507002GA SA310507004GA SA310507006GA SA310507007GA SA310507008GA SA310507009GA SA310507010GA SA310507017GA SA310507020GA SA310507023GA SA310507028GA SA310507031GA SA310507032GA SA310507037GA SA310507048GA SA310507049GA SA310507055GA SA311090976GA SA312018748GA SA312018758GA SA312018762GA SA312018768GA SA312018769GA SA312152330GA SA312152351GA SA312284293GA SA312284294GA SA312284295GA SA312284323GA SA312386403GA SA312386410GA SA312386426GA SA312386432GA SA312386457GA SA312386477GA SA312447559GA SA312447564GA SA312480337GA SA312480342GA SA312480352GA SA312480382GA SA312480387GA SA312480402GA SA312480403GA SA312480411GA SA312480414GA SA313272147GA sa313272193ga SA313420236GA SA313420268GA SA313441543GA SA313451712GA SA313451715GA SA313451728GA SA313451736GA SA313451740GA SA313452154GA SA313462174GA SA313462175GA SA313462183GA SA313462219GA SA313493085GA SA314024991GA SA314025101GA SA314076129GA SA314076340GA SA314086433GA SA314086472GA SA314170247GA SA314170255GA SA314222850GA SA314222851GA SA314222879GA SA314233011GA SA314222322GA SA314222333GA SA307150647GA SA307150649GA SA308337627GA SA308347735GA SA308347738GA SA308347739GA SA308347740GA SA308347751GA SA308347755GA SA308347759GA SA308347760GA SA308347761GA SA308347811GA SA308347812GA SA308347824GA SA308491070GA SA309142916GA SA309436285GA SA310259942GA SA310372085GA SA310382820GA SA310392853GA SA310392864GA SA310392868GA SA310392870GA SA310392871GA SA310392873GA SA310392879GA SA310392915GA SA310392919GA SA310392923GA SA310393084GA SA310393292GA SA310403352GA SA310403391GA SA310403393GA SA310403395GA SA310403399GA SA310403410GA SA310403421GA SA310403429GA SA310403432GA SA310403443GA SA310403447GA SA310403448GA SA310403450GA SA310423763GA SA310423781GA SA310423808GA SA310423875GA SA310423880GA SA310423892GA SA310423908GA SA310423913GA SA310424045GA SA310434054GA SA310434056GA SA310434104GA SA310434109GA SA310434111GA SA310434113GA SA310434142GA SA310434150GA SA310434154GA SA310434155GA SA310434158GA SA310434166GA SA310434178GA SA310434179GA SA310434189GA SA310496236GA SA310496478GA SA310518404GA SA310518432GA SA310518470GA SA310518519GA SA310518526GA SA310518532GA SA310528731GA SA310528741GA SA310528774GA SA310528942GA SA310529052GA SA311132380GA SA311287062GA SA311287068GA SA312152460GA SA312152465GA SA312192730GA SA312232860GA SA312253204GA SA312253205GA SA312253206GA SA312253211GA SA312253214GA SA312253217GA SA312253219GA SA312253227GA SA312253240GA SA312253253GA SA312253255GA SA312253257GA SA312253273GA SA312344842GA SA312344857GA SA312344890GA SA312344974GA SA312355317GA SA312355413GA SA312355482GA SA312355507GA SA312355508GA SA312355515GA SA312375951GA SA312376119GA SA312376138GA SA312376144GA SA312376148GA SA312376162GA SA312376163GA SA312376180GA SA312376190GA SA312376194GA SA312376218GA SA312376231GA SA312376232GA SA312417395GA SA312470035GA SA312470058GA SA312479651GA SA312479943GA SA312479960GA SA312491176GA SA312501694GA SA312501729GA SA312501733GA SA312501735GA SA312501740GA SA312501745GA SA312501763GA SA312501789GA SA312501791GA SA312501793GA SA312501802GA SA312501806GA SA312501808GA SA312501812GA SA312501851GA SA312501864GA SA313023167GA SA313023209GA SA313033714GA SA313043921GA SA313084551GA SA313084559GA SA313084560GA SA313084906GA SA313084940GA SA313084947GA SA313084948GA SA313084952GA SA313095455GA SA313095521GA SA313105736GA SA313105796GA SA313106158GA SA313106174GA SA313106186GA SA313106214GA SA313106310GA SA313116821GA SA313117131GA SA313127271GA SA313127282GA SA313127287GA SA313127297GA SA313127575GA SA313147904GA SA313148004GA SA313208763GA SA313219034GA SA313251228GA SA313251606GA SA313251686GA SA313261722GA SA313314612GA SA313314618GA SA313314668GA SA313314672GA SA313314673GA SA313314740GA SA313314741GA SA313314778GA SA313324925GA SA313324928GA SA313324937GA SA313324942GA SA313324945GA SA313324949GA SA313324985GA SA313324991GA SA313324995GA SA313334998GA SA313335000GA SA313335118GA SA313335125GA SA313335257GA SA313335260GA SA313335279GA SA313335291GA SA313335292GA SA313335330GA SA313335332GA SA313335358GA SA313335367GA SA313335370GA SA313335371GA SA313335372GA SA313345527GA SA313345554GA SA313345618GA SA313345621GA SA313345625GA SA313345634GA SA313345641GA SA313345675GA SA313345676GA SA313345684GA SA313345686GA SA313345694GA SA313345704GA SA313345732GA SA313378239GA SA313378312GA SA313378315GA SA313378325GA SA313388522GA SA313388660GA SA313398790GA SA313420050GA SA313420052GA SA313431153GA SA313431180GA SA314014533GA SA314014563GA SA314014606GA SA314014618GA SA314025202GA SA314035279GA SA314035383GA SA314045536GA SA314065740GA SA314065746GA SA314065764GA SA314065775GA SA314086500GA SA314170159GA SA314191156GA SA314212143GA SA314212168GA SA314222364GA SA314222367GA SA314222368GA SA314222371GA SA314222377GA SA314222378GA SA314222606GA SA314222649GA SA314222898GA SA314263803GA SA314273994GA SA314274005GA SA314274074GA SA314274085GA SA314274109GA SA314284410GA SA314294830GA SA314294840GA SA308145689GA SA308176252GA SA309426087GA SA309426100GA SA309436286GA SA309466737GA SA309466738GA SA309466757GA SA309466761GA SA309466778GA SA309466790GA SA309476933GA SA309476938GA SA309476942GA SA309476944GA SA309476959GA SA309476964GA SA309476977GA SA309476981GA SA309476983GA SA309476989GA SA309476991GA SA309476992GA SA310259938GA SA310259939GA SA310259941GA SA310259950GA SA310259953GA SA310259954GA SA310259955GA SA310259957GA SA310259958GA SA310259963GA SA310259965GA SA310259967GA SA310259970GA SA310331299GA SA310361813GA SA310361816GA SA310361908GA SA310372052GA SA310372066GA SA310372154GA SA310372157GA SA310372160GA SA310372277GA SA310372278GA SA310372279GA SA310372281GA SA310372282GA SA310372284GA SA310372286GA SA310372287GA SA310372293GA SA310372295GA SA310372297GA SA310372301GA SA310372313GA SA310382487GA SA310382488GA SA310382491GA SA310382498GA SA310382523GA SA310517831GA SA311039446GA SA311236287GA SA311236316GA SA311236321GA SA311236324GA SA311508659GA SA311508662GA SA311508672GA SA312018890GA SA312028908GA SA312028909GA SA312028918GA SA312028935GA SA312049464GA SA312059548GA SA312080567GA SA312080742GA SA312080778GA SA312080807GA SA312080835GA SA312080836GA SA312080843GA SA312080844GA SA312080880GA SA312080891GA SA312080898GA SA312080913GA SA312121865GA SA312253184GA SA312253297GA SA312253312GA SA312253339GA SA312375934GA SA312375997GA SA312376002GA SA312376005GA SA312376015GA SA312376017GA SA312376054GA SA312376055GA SA312376058GA SA312396917GA SA312396923GA SA312397190GA SA312397198GA SA312457988GA SA312468822GA SA312480347GA SA312480386GA SA312501593GA SA312502021GA SA312512202GA SA312512203GA SA312512208GA SA312512209GA SA312512212GA SA312512213GA SA312512217GA SA312512218GA SA312512221GA SA312512223GA SA312512224GA SA312512228GA SA313219215GA SA313271980GA SA313293003GA SA313293007GA SA313293059GA SA313293061GA SA313293081GA SA313293088GA SA313293094GA SA313293096GA SA313293099GA SA313293101GA SA313293105GA SA313293106GA SA313293107GA SA313293114GA SA313293120GA SA313293124GA SA313293127GA SA313293128GA SA313293130GA SA313293142GA SA313293172GA SA313314454GA SA313314473GA SA313314480GA SA313314486GA SA313314488GA SA313314491GA SA313314492GA SA313314493GA SA313314752GA SA313314754GA SA313314759GA SA313314767GA SA313314813GA SA313314863GA SA313335138GA SA313335144GA SA313335162GA SA313335168GA SA313335175GA SA313335180GA SA313335181GA SA313335183GA SA313335185GA SA313335201GA SA313335203GA SA313335223GA SA313335224GA SA313335227GA SA313335235GA SA313335245GA SA313335253GA SA313335424GA SA313419495GA SA313419511GA SA313420306GA SA313420309GA SA313420312GA SA313420325GA SA313420329GA SA313420331GA SA313420333GA SA313420336GA SA313420337GA SA313420340GA SA313420346GA SA313420347GA SA313420348GA SA313420350GA SA313420351GA SA313420352GA SA313420353GA SA313420355GA SA313420358GA SA313420360GA SA313420369GA SA313420371GA SA313420373GA SA313420374GA SA313420376GA SA313420386GA SA313420389GA SA313420396GA SA313420400GA SA313420403GA SA313420406GA SA313420432GA SA313420527GA SA313420529GA SA313420536GA SA313430657GA SA313430684GA SA313441467GA SA313441533GA SA313441535GA SA313441554GA SA313441556GA SA313441559GA SA313441634GA SA313441636GA SA313441694GA SA313441701GA SA313451732GA SA313451733GA SA313451748GA SA313451788GA SA313451789GA SA313451795GA SA313451798GA SA313451799GA SA313451804GA SA313451806GA SA313451809GA SA313451810GA SA313451812GA SA313451815GA SA313451817GA SA313451818GA SA313451819GA SA313451823GA SA313451830GA SA313451836GA SA313451842GA SA313451890GA SA313451894GA SA313451895GA SA313451896GA SA313451899GA SA313451900GA SA313451901GA SA313451914GA SA313451919GA SA313451929GA SA313451932GA SA313451940GA SA313451949GA SA313451959GA SA313451965GA SA313451972GA SA313451976GA SA313451980GA SA313451981GA SA313451989GA SA313452001GA SA313452006GA SA313452009GA SA313452012GA SA313452014GA SA313452016GA SA313452017GA SA313452024GA SA313452034GA SA313452057GA SA313462226GA SA313462258GA SA313462276GA SA313462319GA SA314014404GA SA314014677GA SA314024767GA SA314024769GA SA314024773GA SA314024779GA SA314024789GA SA314024800GA SA314024827GA SA314024916GA SA314024929GA SA314097260GA SA314097262GA SA314097281GA SA314097283GA SA314097285GA SA314097287GA SA314097292GA SA314097299GA SA314097317GA SA314097328GA SA314107431GA SA314107433GA SA314107437GA SA314107464GA SA314107857GA SA314108004GA SA314170437GA SA314170443GA SA314170469GA SA314180479GA SA314180512GA SA314180516GA SA314180548GA SA314180552GA SA314180557GA SA314180563GA SA314180572GA SA314180573GA SA314180576GA SA314180592GA SA314180596GA SA314180601GA SA314180627GA SA314180643GA SA314180645GA SA314180647GA SA314180648GA SA314180649GA SA314180662GA SA314180664GA SA314180666GA SA314180712GA SA314180717GA SA314180720GA SA314180727GA SA314180733GA SA314211923GA SA314211949GA SA314211950GA SA314273917GA SA314273922GA SA314273924GA SA314273925GA SA314273932GA SA314274165GA SA314274167GA SA314274169GA SA314274211GA SA314274228GA SA314274233GA SA314274234GA SA314295113GA SA309122768GA SA310517842GA SA310517876GA SA311060146GA SA311060181GA SA312039213GA SA312080687GA SA312080694GA SA312284291GA SA312080911GA SA308501253GR SA309324417GR SA309446476GR SA309476956GR SA309476961GR SA309476994GR SA309477014GR SA309497152GR SA309517705GR SA309517781GR SA309517856GR SA309517863GR SA309527993GR SA309528005GR SA309528007GR SA310290679GR SA310403409GR SA310423738GR SA310423900GR SA310506711GR SA310507043GR SA310507402GR SA310507428GR SA310507431GR SA310507522GR SA310517756GR SA310517817GR SA310517869GR SA310518002GR SA310518007GR SA310518010GR SA310518152GR SA310518217GR SA310518465GR SA310518638GR SA310518657GR SA310518659GR SA310518719GR SA310518721GR SA310529222GR SA311060072GR SA311060075GR SA311069660GR SA311080605GR SA311080741GR SA311090928GR SA311121777GR SA311122056GR SA311122153GR SA311122208GR SA311132308GR SA311132557GR SA311163490GR SA311174016GR SA311195242GR SA311215736GR SA311297298GR SA311297314GR SA311317598GR SA311317680GR SA311468428GR SA311488475GR SA312039097GR SA312080661GR SA312080800GR SA312132141GR SA312152313GR SA312152461GR SA312152483GR SA312152484GR SA312192589GR SA312192594GR SA312192612GR SA312192616GR SA312192629GR SA312294457GR SA312345051GR SA312365724GR SA312378035GR SA312386336GR SA312386492GR SA312386494GR SA312386541GR SA312386580GR SA312457980GR SA312458427GA SA312458470GR SA312468569GR SA312469073GR SA312479522GR SA312479840GR SA312480834GR SA312490884GR SA312512337GR SA312512364GR SA313013098GA SA313023289GR SA313023293GR SA313023296GR SA313023297GR SA313023298GR SA313023323GA SA313023421GR SA313023427GR SA313023440GA SA313023448GA SA313023509GR SA313033572GR SA313033661GR SA313033678GR SA313084478GA SA313084515GA SA313084523GA SA313084638GA SA313084656GA SA313084677GA SA313084685GR SA313084716GA SA313084722GA SA313084729GA SA313084730GA SA313084732GA SA313084736GA SA313084737GA SA313084739GA SA313084741GA SA313084805GA SA313084930GA SA313084957GA SA313084960GA SA313084980GA SA313095015GA SA313095039GA SA313095052GA SA313095073GA SA313095149GA SA313095155GA SA313095188GA SA313095229GA SA313095304GR SA313095345GA SA313095387GA SA313095400GA SA313095449GA SA313095461GA SA313095463GA SA313095469GA SA313095488GA SA313095498GA SA313095499GA SA313095502GA SA313095513GA SA313095515GR SA313095516GA SA313095518GR SA313095522GA SA313095533GA SA313095539GA SA313095544GR SA313095551GA SA313095554GA SA313095592GA SA313095593GA SA313095594GA SA313095595GA SA313095605GA SA313095607GA SA313095620GA SA313095629GA SA313095630GA SA313095640GA SA313095644GA SA313105658GA SA313105659GA SA313105681GA SA313105688GA SA313105693GA SA313105696GA SA313105697GA SA313105706GA SA313105709GA SA313105711GA SA313105717GA SA313105728GA SA313105731GA SA313105732GA SA313105737GA SA313105738GA SA313105741GA SA313105744GA SA313105755GA SA313105756GA SA313105759GA SA313105762GA SA313105763GA SA313105764GA SA313105765GA SA313105769GA SA313105772GA SA313105774GA SA313105786GA SA313105799GA SA313105828GA SA313105833GA SA313105836GR SA313105841GR SA313105863GA SA313105867GA SA313105874GA SA313105877GA SA313105887GA SA313105897GA SA313105898GA SA313105914GA SA313105916GA SA313105925GA SA313105929GA SA313105934GA SA313105954GA SA313105956GA SA313105963GA SA313105965GA SA313105970GA SA313105983GA SA313106003GA SA313106004GA SA313106008GA SA313106012GA SA313106018GA SA313106019GA SA313106035GA SA313106037GA SA313106038GA SA313106044GA SA313106056GA SA313106077GA SA313106078GA SA313106080GA SA313106087GA SA313106090GA SA313106092GA SA313106100GA SA313106101GA SA313106104GA SA313106109GA SA313106111GA SA313106112GA SA313106116GA SA313106117GA SA313106118GA SA313106119GA SA313106120GA SA313106125GA SA313106126GA SA313106129GA SA313106135GA SA313106144GA SA313106151GA SA313106155GA SA313106159GA SA313106162GA SA313106166GA SA313106169GA SA313106181GA SA313106204GA SA313106207GA SA313106212GA SA313106229GA SA313106230GA SA313106234GA SA313106236GA SA313106239GA SA313106240GA SA313106247GA SA313106253GA SA313106267GA SA313106272GA SA313106274GA SA313106281GA SA313106288GA SA313106294GA SA313106295GA SA313106299GA SA313106308GA SA313106314GA SA313106322GA SA313106324GA SA313106332GA SA313106335GA SA313106368GA SA313106370GA SA313106372GA SA313106373GA SA313106376GA SA313106385GA SA313106387GA SA313106388GA SA313106390GA SA313106393GA SA313106398GA SA313106413GA SA313106417GA SA313106430GA SA313106431GA SA313106432GA SA313106436GA SA313106441GA SA313106444GA SA313116452GA SA313116453GA SA313116454GA SA313116457GA SA313116462GA SA313116465GA SA313116468GA SA313116469GA SA313116471GA SA313116478GA SA313116487GA SA313116492GA SA313116496GA SA313116501GA SA313116508GA SA313116512GA SA313116513GA SA313116517GA SA313116520GA SA313116522GA SA313116523GA SA313116524GA SA313116527GA SA313116528GA SA313116532GA SA313116535GA SA313116536GA SA313116538GA SA313116542GA SA313116581GA SA313116582GA SA313116584GA SA313116585GA SA313116587GA SA313116588GA SA313116589GA SA313116590GA SA313116591GA SA313116596GA SA313116599GA SA313116602GA SA313116604GA SA313116606GA SA313116608GA SA313116613GA SA313116614GA SA313116615GA SA313116618GA SA313116623GA SA313116624GA SA313116627GA SA313116628GA SA313116630GA SA313116635GA SA313116636GA SA313116637GA SA313116638GA SA313116639GA SA313116643GA SA313116645GA SA313116653GA SA313116655GA SA313116657GA SA313116662GA SA313116667GA SA313116669GA SA313116679GA SA313116717GA SA313116727GA SA313116731GA SA313116734GA SA313116736GA SA313116737GA SA313116739GA SA313116742GA SA313116748GA SA313116756GA SA313116760GA SA313116770GA SA313116777GA SA313116778GA SA313116786GA SA313116797GA SA313116819GR SA313116826GA SA313116876GA SA313116882GA SA313116892GA SA313116899GA SA313116904GA SA313116912GA SA313116915GA SA313116916GA SA313116917GA SA313116923GA SA313116925GA SA313116932GA SA313116933GA SA313116935GA SA313116936GA SA313116937GA SA313116939GA SA313116940GA SA313116941GA SA313116943GA SA313116945GA SA313116946GA SA313116956GA SA313116957GA SA313116959GA SA313116960GA SA313116961GA SA313116962GA SA313116964GA SA313116974GA SA313116979GA SA313116980GA SA313116981GA SA313116982GA SA313116987GA SA313116990GA SA313116991GA SA313116997GA SA313116999GA SA313117005GA SA313117007GA SA313117008GA SA313117010GA SA313117011GA SA313117017GA SA313117018GA SA313117021GA SA313117024GA SA313117025GA SA313117026GA SA313117027GA SA313117030GA SA313117037GA SA313117045GA SA313117051GA SA313117055GA SA313117056GA SA313117061GA SA313117063GA SA313117066GA SA313117069GA SA313117070GA SA313117071GA SA313117073GA SA313117074GA SA313117075GA SA313117078GA SA313117079GA SA313117081GA SA313117082GA SA313117083GA SA313117084GA SA313117085GA SA313117087GA SA313117088GA SA313117092GA SA313117094GA SA313117095GA SA313117102GA SA313117103GA SA313117108GA SA313117109GA SA313117110GA SA313117111GA SA313117115GA SA313117116GA SA313117118GA SA313117120GA SA313117128GA SA313117129GA SA313117141GA SA313117146GA SA313117149GA SA313117162GA SA313117164GA SA313117170GA SA313117173GA SA313117177GA SA313117180GA SA313117181GA SA313117182GA SA313117183GA SA313117185GA SA313117189GA SA313117190GA SA313117195GA SA313117196GA SA313117197GA SA313117200GA SA313117201GA SA313117202GA SA313117203GA SA313117204GA SA313117208GA SA313117209GA SA313127212GA SA313127213GA SA313127215GA SA313127217GA SA313127220GA SA313127222GA SA313127223GA SA313127225GA SA313127228GA SA313127229GA SA313127237GA SA313127242GA SA313127243GA SA313127246GA SA313127248GA SA313127251GA SA313127252GA SA313127254GA SA313127255GA SA313127259GA SA313127260GA SA313127266GA SA313127267GA SA313127269GA SA313127270GA SA313127273GA SA313127276GA SA313127278GA SA313127284GA SA313127286GA SA313127288GA SA313127290GA SA313127291GA SA313127294GA SA313127317GA SA313127318GA SA313127330GA SA313127380GA SA313127396GA SA313127401GA SA313127416GA SA313127420GA SA313127424GA SA313127426GA SA313127429GA SA313127434GA SA313127435GA SA313127436GA SA313127437GA SA313127439GA SA313127440GA SA313127445GA SA313127449GA SA313127450GA SA313127451GA SA313127452GA SA313127454GA SA313127456GA SA313127459GA SA313127463GA SA313127464GA SA313127470GA SA313127471GA SA313127472GA SA313127473GA SA313127474GA SA313127475GA SA313127476GA SA313127477GA SA313127478GA SA313127480GA SA313127482GA SA313127484GA SA313127490GR SA313127494GA SA313127499GA SA313127500GR SA313127513GA SA313127518GA SA313127519GR SA313127520GR SA313127522GR SA313127524GA SA313127525GR SA313127528GR SA313127530GR SA313127532GR SA313127534GR SA313127535GR SA313127536GR SA313127538GR SA313127539GR SA313127542GR SA313127543GR SA313127545GA SA313127547GR SA313127550GA SA313127551GR SA313127554GR SA313127556GR SA313127558GR SA313127561GR SA313127562GR SA313127571GR SA313127582GR SA313127583GA SA313127589GA SA313127602GA SA313127604GA SA313127609GA SA313127636GA SA313127639GA SA313127643GA SA313127645GA SA313127650GA SA313127653GA SA313127660GA SA313127664GA SA313127667GA SA313127669GA SA313127671GA SA313127673GA SA313127675GA SA313127693GA SA313127702GA SA313127704GA SA313127711GA SA313127725GA SA313127728GA SA313127729GA SA313127730GA SA313127738GA SA313127741GA SA313127742GA SA313137744GA SA313137751GA SA313137752GA SA313137754GA SA313137758GA SA313137760GA SA313137764GA SA313137766GA SA313137768GA SA313137769GA SA313137770GA SA313137772GA SA313137774GA SA313137777GA SA313137778GA SA313137781GA SA313137843GA SA313137844GA SA313137845GA SA313137846GA SA313137847GA SA313137848GA SA313137849GA SA313137850GA SA313137851GA SA313137852GA SA313137853GA SA313137855GA SA313137856GA SA313137857GA SA313137858GA SA313137859GA SA313137860GA SA313137861GA SA313137862GA SA313137863GA SA313137865GA SA313137866GA SA313137867GA SA313137869GA SA313137872GA SA313137873GA SA313137874GA SA313147903GA SA313147905GA SA313147906GA SA313147907GA SA313147908GA SA313147909GA SA313147910GA SA313147911GA SA313147912GA SA313147916GA SA313147918GA SA313147919GA SA313147921GA SA313147922GA SA313147923GA SA313147924GA SA313147925GA SA313147926GA SA313147927GA SA313147928GA SA313147929GA SA313147930GA SA313147931GA SA313147932GA SA313147933GA SA313147934GA SA313147935GA SA313147937GA SA313147938GA SA313147939GA SA313147941GA SA313147942GA SA313147944GA SA313147945GA SA313147947GA SA313147949GA SA313147952GA SA313147954GA SA313147955GA SA313147956GA SA313147958GA SA313147959GA SA313147960GA SA313147961GA SA313147962GA SA313147963GA SA313147964GA SA313147966GA SA313147967GA SA313147968GA SA313147970GA SA313147972GA SA313147973GA SA313147976GA SA313147978GA SA313147979GA SA313147980GA SA313147981GA SA313147982GA SA313147986GA SA313147987GA SA313147988GA SA313147989GA SA313147990GA SA313147991GA SA313147992GA SA313147993GA SA313147994GA SA313147997GA SA313147998GA SA313148002GA SA313148005GA SA313148006GA SA313148011GA SA313148014GA SA313148015GA SA313148017GA SA313148019GA SA313148021GA SA313188108GA SA313188113GA SA313198128GA SA313198129GA SA313198130GA SA313198131GA SA313198132GA SA313198133GA SA313198134GA SA313198135GA SA313198136GA SA313198137GA SA313198138GA SA313198139GA SA313198140GA SA313198141GA SA313198142GA SA313198143GA SA313198144GA SA313198145GA SA313198146GA SA313198147GA SA313198148GA SA313198149GA SA313198150GA SA313198151GA SA313198152GA SA313198153GA SA313198154GA SA313198155GA SA313198156GA SA313198157GA SA313198158GA SA313198159GA SA313198160GA SA313198161GA SA313198163GA SA313198164GA SA313198166GA SA313198167GA SA313198168GA SA313198169GA SA313198170GA SA313198171GA SA313198172GA SA313198181GA SA313198182GA SA313198191GA SA313198193GA SA313198194GA SA313198198GA SA313198202GA SA313198204GA SA313198208GA SA313198212GA SA313198215GA SA313198218GA SA313198219GA SA313198220GA SA313198221GA SA313198222GA SA313198224GA SA313198225GA SA313198226GA SA313198227GA SA313198231GA SA313198232GA SA313198233GA SA313198234GA SA313198238GA SA313198245GA SA313198248GA SA313198256GA SA313198263GA SA313198264GA SA313198273GA SA313198281GA SA313198285GA SA313198288GA SA313198293GA SA313198297GA SA313198298GA SA313198300GA SA313198302GA SA313198303GA SA313198304GA SA313198306GA SA313198307GA SA313198308GA SA313198309GA SA313198310GA SA313198311GA SA313198312GA SA313198313GA SA313198314GA SA313198316GA SA313198317GA SA313198318GA SA313198320GA SA313198321GA SA313198322GA SA313198323GA SA313198324GA SA313198325GA SA313198326GA SA313198328GA SA313198329GA SA313198330GA SA313198331GA SA313198332GA SA313198335GA SA313198336GA SA313198337GA SA313198338GA SA313198340GA SA313198341GA SA313198342GA SA313198343GA SA313198345GA SA313198346GA SA313198347GA SA313198348GA SA313198349GA SA313198351GA SA313198352GA SA313198353GA SA313198354GA SA313198355GA SA313198356GA SA313198357GA SA313198358GA SA313198359GA SA313198361GA SA313198362GA SA313198363GA SA313198365GA SA313198366GA SA313198367GA SA313198369GA SA313198370GA SA313198371GA SA313198373GA SA313198375GA SA313198377GA SA313198378GA SA313198379GA SA313198380GA SA313198381GA SA313198382GA SA313198385GA SA313198386GA SA313198387GA SA313198388GA SA313198390GA SA313198391GA SA313198392GA SA313198393GA SA313198395GA SA313198398GA SA313198402GA SA313198404GA SA313198406GA SA313198407GA SA313198408GA SA313198409GA SA313198410GA SA313198411GA SA313198412GA SA313198413GA SA313198414GA SA313198416GA SA313198417GA SA313198419GA SA313198421GA SA313198422GA SA313198423GA SA313198424GA SA313198425GA SA313198427GA SA313198428GA SA313198429GA SA313198430GA SA313198435GA SA313198438GA SA313198439GA SA313198440GA SA313198441GA SA313198442GA SA313198444GA SA313198445GA SA313198446GA SA313198447GA SA313198448GA SA313198449GA SA313198450GA SA313198451GA SA313198452GA SA313198453GA SA313198454GA SA313198455GA SA313198456GA SA313198457GA SA313198458GA SA313198459GA SA313198461GA SA313198462GA SA313198463GA SA313198464GA SA313198465GA SA313198466GA SA313198467GA SA313198468GA SA313198469GA SA313198470GA SA313198471GA SA313198473GA SA313198474GA SA313198475GA SA313198477GA SA313198478GA SA313198479GA SA313198480GA SA313198481GA SA313198482GA SA313198483GA SA313198484GA SA313198485GA SA313198486GA SA313198487GA SA313198489GA SA313198490GA SA313198491GA SA313198492GA SA313198493GA SA313198494GA SA313198495GA SA313198496GA SA313198497GA SA313198498GA SA313198499GA SA313198500GA SA313198501GA SA313198502GA SA313198503GA SA313198504GA SA313198505GA SA313198506GA SA313198508GA SA313198509GA SA313198510GA SA313198511GA SA313198512GA SA313198513GA SA313198514GA SA313198516GA SA313198518GA SA313198519GA SA313198520GA SA313198521GA SA313198522GA SA313198523GA SA313198526GA SA313198527GA SA313198528GA SA313198529GA SA313198531GA SA313198532GA SA313198534GA SA313198535GA SA313198536GA SA313198537GA SA313198538GA SA313198539GA SA313198540GA SA313198541GA SA313198542GA SA313198543GA SA313198544GA SA313198545GA SA313198546GA SA313198547GA SA313198548GA SA313198549GA SA313198550GA SA313198551GA SA313198552GA SA313198553GA SA313198554GA SA313198555GA SA313198556GA SA313198557GA SA313198558GA SA313198559GA SA313198560GA SA313198561GA SA313198562GA SA313198563GA SA313198564GA SA313198565GA SA313198566GA SA313198567GA SA313198568GA SA313198569GA SA313198570GA SA313198571GA SA313198572GA SA313198573GA SA313198574GA SA313198575GA SA313198576GA SA313198577GA SA313198578GA SA313198579GA SA313198580GA SA313198581GA SA313198582GA SA313198583GA SA313198584GA SA313198586GA SA313198587GA SA313198588GA SA313198589GA SA313198590GA SA313198592GA SA313198593GA SA313198594GA SA313198597GA SA313198598GA SA313198600GA SA313198601GA SA313198602GA SA313198603GA SA313198604GA SA313198605GA SA313198606GA SA313198607GA SA313198608GA SA313198610GA SA313198611GA SA313198612GA SA313198613GA SA313198614GA SA313198615GA SA313198616GA SA313198617GA SA313198618GA SA313198619GA SA313198620GA SA313198621GA SA313198622GA SA313198623GA SA313198626GA SA313198627GA SA313198628GA SA313198629GA SA313198630GA SA313198631GA SA313198632GA SA313198633GA SA313198634GA SA313198635GA SA313198636GA SA313198638GA SA313198639GA SA313198640GA SA313198641GA SA313198642GA SA313198643GA SA313198644GA SA313198646GA SA313198647GA SA313198648GA SA313198649GA SA313198651GA SA313198652GA SA313198653GA SA313198654GA SA313198655GA SA313198656GA SA313198657GA SA313198658GA SA313198659GA SA313198660GA SA313198661GA SA313198662GA SA313198663GA SA313198664GA SA313198665GA SA313198666GA SA313198668GA SA313198669GA SA313198670GA SA313198671GA SA313198672GA SA313198673GA SA313198674GA SA313198675GA SA313198676GA SA313198677GA SA313198678GA SA313198679GA SA313198680GA SA313198683GA SA313198684GA SA313198685GA SA313198689GA SA313198690GA SA313198692GA SA313198693GA SA313198694GA SA313198695GA SA313198696GA SA313198697GA SA313198698GA SA313198699GA SA313198701GA SA313198703GA SA313198705GA SA313198708GA SA313198709GA SA313198710GA SA313198711GA SA313198712GA SA313198713GA SA313198714GA SA313198715GA SA313198716GA SA313198717GA SA313198718GA SA313198719GA SA313198720GA SA313198721GA SA313198722GA SA313198723GA SA313198724GA SA313198725GA SA313198726GA SA313198727GA SA313198728GA SA313198729GA SA313198730GA SA313198731GA SA313198732GA SA313198733GA SA313198734GA SA313198735GA SA313198736GA SA313198737GA SA313198739GA SA313198740GA SA313198741GA SA313198742GA SA313198743GA SA313198745GA SA313198746GA SA313198748GA SA313198749GA SA313198750GA SA313208751GA SA313208752GA SA313208753GA SA313208754GA SA313208755GA SA313208756GA SA313208757GA SA313208758GA SA313208759GA SA313208760GA SA313208762GA SA313208764GA SA313208765GA SA313208766GA SA313208767GA SA313208768GA SA313208769GA SA313208770GA SA313208771GA SA313208772GA SA313208773GA SA313208774GA SA313208775GA SA313208776GA SA313208777GA SA313208778GA SA313208780GA SA313208781GA SA313208782GA SA313208783GA SA313208784GA SA313208785GA SA313208786GA SA313208787GA SA313208789GA SA313208791GA SA313208792GA SA313208793GA SA313208795GA SA313208796GA SA313208797GA SA313208798GA SA313208799GA SA313208801GA SA313208802GA SA313208804GA SA313208805GA SA313208806GA SA313208807GA SA313208808GA SA313208809GA SA313208810GA SA313208811GA SA313208812GA SA313208813GA SA313208814GA SA313208815GA SA313208816GA SA313208817GA SA313208818GA SA313208819GA SA313208820GA SA313208822GA SA313208823GA SA313208824GA SA313208825GA SA313208827GA SA313208829GA SA313208830GA SA313208831GA SA313208832GA SA313208833GA SA313208834GA SA313208836GA SA313208837GA SA313208838GA SA313208839GA SA313208840GA SA313208842GA SA313208843GA SA313208844GA SA313208845GA SA313208846GA SA313208848GA SA313208849GA SA313208850GA SA313208851GA SA313208853GA SA313208854GA SA313208855GA SA313208856GA SA313208857GA SA313208858GA SA313208860GA SA313208861GA SA313208862GA SA313208863GA SA313208864GA SA313208865GA SA313208866GA SA313208867GA SA313208869GA SA313208870GA SA313208871GA SA313208872GA SA313208874GA SA313208876GA SA313208877GA SA313208882GA SA313208883GA SA313208884GA SA313208891GA SA313208893GA SA313208894GA SA313208895GA SA313208896GA SA313208897GA SA313208898GA SA313208899GA SA313208901GA SA313208902GA SA313208903GA SA313208906GA SA313208907GA SA313208908GA SA313208909GA SA313208912GA SA313208916GA SA313208917GA SA313208920GA SA313208922GA SA313208923GA SA313208926GA SA313208934GA SA313208935GA SA313208945GA SA313208946GA SA313208947GA SA313208953GA SA313208955GA SA313208957GA SA313208958GA SA313208962GA SA313208965GA SA313208967GA SA313208968GA SA313208982GA SA313208990GA SA313208999GA SA313209007GA SA313209008GA SA313219012GA SA313219013GA SA313219014GA SA313219016GA SA313219018GA SA313219022GA SA313219023GA SA313219025GA SA313219026GA SA313219028GA SA313219029GA SA313219030GA SA313219031GA SA313219036GA SA313219038GA SA313219039GA SA313219040GA SA313219042GA SA313219043GA SA313219044GA SA313219045GA SA313219046GA SA313219048GA SA313219049GA SA313219053GA SA313219054GA SA313219055GA SA313219056GA SA313219057GA SA313219065GA SA313219066GA SA313219069GA SA313219070GA SA313219071GA SA313219072GA SA313219074GA SA313219077GA SA313219079GA SA313219080GA SA313219082GA SA313219083GA SA313219084GA SA313219085GA SA313219086GA SA313219087GA SA313219088GA SA313219089GA SA313219091GA SA313219092GA SA313219093GA SA313219094GA SA313219096GA SA313219099GA SA313219100GA SA313219101GA SA313219103GA SA313219104GA SA313219105GA SA313219106GA SA313219107GA SA313219108GA SA313219110GA SA313219111GA SA313219112GA SA313219113GA SA313219114GA SA313219116GA SA313219117GA SA313219118GA SA313219119GA SA313219120GA SA313219123GA SA313219125GA SA313219126GA SA313219127GA SA313219128GA SA313219129GA SA313219131GA SA313219132GA SA313219133GA SA313219134GA SA313219136GA SA313219137GA SA313219139GA SA313219140GA SA313219141GA SA313219142GA SA313219143GA SA313219144GA SA313219146GA SA313219147GA SA313219151GA SA313219152GA SA313219155GA SA313219157GA SA313219158GA SA313219160GA SA313219161GA SA313219162GA SA313219165GA SA313219170GA SA313219171GA SA313219174GA SA313219176GA SA313219177GA SA313219179GA SA313219180GA SA313219181GA SA313219182GA SA313219183GA SA313219184GA SA313219185GA SA313219186GA SA313219187GA SA313219188GA SA313219191GA SA313219193GA SA313219196GA SA313219199GA SA313219202GA SA313219203GA SA313219204GA SA313219207GA SA313219211GA SA313219213GA SA313219217GA SA313219218GA SA313219220GA SA313219223GA SA313219226GA SA313219227GA SA313219228GA SA313219230GA SA313219233GA SA313219234GA SA313219236GA SA313219238GA SA313219239GA SA313219241GA SA313219242GA SA313219243GA SA313219244GA SA313219246GA SA313219247GA SA313219248GA SA313219249GA SA313219250GA SA313219251GA SA313219253GA SA313219254GA SA313219255GA SA313219256GA SA313219257GA SA313219258GA SA313219259GA SA313219261GA SA313219262GA SA313219263GA SA313219264GA SA313219265GA SA313219266GA SA313219267GA SA313219268GA SA313219269GA SA313219270GA SA313219271GA SA313219272GA SA313219273GA SA313219274GA SA313219275GA SA313219276GA SA313219277GA SA313219278GA SA313219279GA SA313219280GA SA313219281GA SA313219283GA SA313219284GA SA313219285GA SA313219286GA SA313219287GA SA313219288GA SA313219289GA SA313219290GA SA313219292GA SA313219293GA SA313219295GA SA313219296GA SA313219299GA SA313219300GA SA313219301GA SA313219302GA SA313219303GA SA313219304GA SA313219305GA SA313219306GA SA313219307GA SA313219308GA SA313220001GA SA313220002GA SA313220003GA SA313220005GA SA313220009GA SA313220012GA SA313220013GA SA313220014GA SA313220015GA SA313220019GA SA313220020GA SA313220021GA SA313220022GA SA313220023GA SA313220024GA SA313220025GA SA313220026GA SA313220027GA SA313220028GA SA313220029GA SA313220030GA SA313229309GA SA313229310GA SA313229311GA SA313229312GA SA313229313GA SA313229314GA SA313229315GA SA313229316GA SA313229317GA SA313229318GA SA313229320GA SA313229321GA SA313229322GA SA313229323GA SA313229325GA SA313229327GA SA313229334GA SA313229335GA SA313229337GA SA313229339GA SA313229340GA SA313229341GA SA313229342GA SA313229343GA SA313229344GA SA313229345GA SA313229346GA SA313229347GA SA313229348GA SA313229349GA SA313229350GA SA313229351GA SA313229352GA SA313229353GA SA313229354GA SA313229355GA SA313229356GA SA313229357GA SA313229358GA SA313229359GA SA313229360GA SA313229361GA SA313229362GA SA313229363GA SA313229364GA SA313229365GA SA313229366GA SA313229367GA SA313229369GA SA313229370GA SA313229371GA SA313229372GA SA313229374GA SA313229375GA SA313229376GA SA313229377GA SA313229378GA SA313229379GA SA313229380GA SA313229381GA SA313229382GA SA313229383GA SA313229384GA SA313229385GA SA313229387GA SA313229388GA SA313229390GA SA313229391GA SA313229393GA SA313229394GA SA313229395GA SA313229396GA SA313229397GA SA313229398GA SA313229399GA SA313229400GA SA313229401GA SA313229402GA SA313229404GA SA313229405GA SA313229406GA SA313229407GA SA313229409GA SA313229410GA SA313229411GA SA313229413GA SA313229414GA SA313229415GA SA313229416GA SA313229418GA SA313229419GA SA313229420GA SA313229421GA SA313229422GA SA313229429GA SA313229431GA SA313229432GA SA313229433GA SA313229434GA SA313229435GA SA313229438GA SA313229440GA SA313229441GA SA313229442GA SA313229446GA SA313229450GA SA313229452GA SA313229454GA SA313229455GA SA313229457GA SA313229458GA SA313229459GA SA313229460GA SA313229461GA SA313229462GA SA313229463GA SA313229464GA SA313229465GA SA313229466GA SA313229467GA SA313229468GA SA313229470GA SA313229471GA SA313229472GA SA313229473GA SA313229474GA SA313229475GA SA313229476GA SA313229477GA SA313229478GA SA313229479GA SA313229480GA SA313229481GA SA313229482GA SA313229483GA SA313229484GA SA313229485GA SA313229486GA SA313229487GA SA313229488GA SA313229493GA SA313229499GA SA313229507GA SA313229508GA SA313229511GA SA313229515GA SA313229517GA SA313229549GA SA313229551GA SA313229556GA SA313229557GA SA313229560GA SA313229563GA SA313229569GA SA313229573GA SA313229575GA SA313229579GA SA313229580GA SA313229581GA SA313229582GA SA313229584GA SA313229585GA SA313229586GA SA313229587GA SA313229588GA SA313229589GA SA313229590GA SA313229591GA SA313229593GA SA313229594GA SA313229595GA SA313229597GA SA313229598GA SA313229599GA SA313229600GA SA313229602GA SA313229603GA SA313229604GA SA313229606GA SA313229607GA SA313229610GA SA313229611GA SA313229612GA SA313229613GA SA313229614GA SA313229615GA SA313229617GA SA313229619GA SA313229620GA SA313229621GA SA313229622GA SA313229623GA SA313229624GA SA313229627GA SA313229630GA SA313229632GA SA313229633GA SA313229634GA SA313229635GA SA313229637GA SA313229638GA SA313229639GA SA313229640GA SA313229641GA SA313229642GA SA313229645GA SA313229646GA SA313229647GA SA313229648GA SA313229649GA SA313229650GA SA313229651GA SA313229652GA SA313229655GA SA313229656GA SA313229659GA SA313229662GA SA313229663GA SA313229664GA SA313229665GA SA313229666GA SA313229667GA SA313229669GA SA313229673GA SA313229678GA SA313229679GA SA313229683GA SA313229685GA SA313229688GA SA313229692GA SA313229693GA SA313229694GA SA313229697GA SA313229698GA SA313229701GA SA313229702GA SA313229716GA SA313229717GA SA313229718GA SA313229719GA SA313229720GA SA313229721GA SA313229722GA SA313229723GA SA313229724GA SA313229725GA SA313229726GA SA313229732GA SA313229738GA SA313229739GA SA313229740GA SA313229744GA SA313229746GA SA313229747GA SA313229751GA SA313229753GA SA313229754GA SA313229757GA SA313229758GA SA313229759GA SA313229761GA SA313229762GA SA313229763GA SA313229766GA SA313229767GA SA313229768GA SA313229772GA SA313229774GA SA313229777GA SA313229779GA SA313229782GA SA313229783GA SA313229784GA SA313229785GA SA313229786GA SA313229787GA SA313229789GA SA313229790GA SA313229792GA SA313229794GA SA313229796GA SA313229797GA SA313229804GA SA313229807GA SA313229808GA SA313229810GA SA313229812GA SA313229814GA SA313229815GA SA313229816GA SA313229817GA SA313229820GA SA313229822GA SA313229823GA SA313229826GA SA313229829GA SA313229830GA SA313229831GA SA313229832GA SA313229833GA SA313229834GA SA313229835GA SA313229836GA SA313229837GA SA313229838GA SA313229839GA SA313229840GA SA313229841GA SA313229843GA SA313229844GA SA313229845GA SA313229846GA SA313229847GA SA313229849GA SA313229850GA SA313229851GA SA313229853GA SA313229859GA SA313229862GA SA313229863GA SA313229865GA SA313229866GA SA313229872GA SA313229873GA SA313229874GA SA313229875GA SA313229878GA SA313229882GA SA313229885GA SA313229886GA SA313229887GA SA313229888GA SA313229889GA SA313229890GA SA313229891GA SA313229892GA SA313229893GA SA313229895GA SA313229896GA SA313229897GA SA313229900GA SA313229901GA SA313229903GA SA313229904GA SA313229906GA SA313229907GA SA313229908GA SA313229909GA SA313229910GA SA313229912GA SA313229913GA SA313229914GA SA313229915GA SA313229916GA SA313229917GA SA313229938GA SA313229939GA SA313229941GA SA313229942GA SA313229943GA SA313229944GA SA313229945GA SA313229946GA SA313229948GA SA313229949GA SA313229950GA SA313229951GA SA313229953GA SA313229956GA SA313229957GA SA313229958GA SA313229959GA SA313229960GA SA313229961GA SA313229963GA SA313229964GA SA313229965GA SA313229966GA SA313229967GA SA313229968GA SA313229969GA SA313229970GA SA313229971GA SA313229972GA SA313229973GA SA313229974GA SA313229976GA SA313229977GA SA313229979GA SA313229980GA SA313229981GA SA313229982GA SA313229984GA SA313229986GA SA313229987GA SA313230032GA SA313230033GA SA313230035GA SA313230037GA SA313230038GA SA313230040GA SA313230041GA SA313230042GA SA313230043GA SA313230044GA SA313230045GA SA313230046GA SA313230047GA SA313230048GA SA313230049GA SA313230051GA SA313230052GA SA313230053GA SA313230055GA SA313230056GA SA313230057GA SA313230058GA SA313230059GA SA313230060GA SA313230061GA SA313230062GA SA313230064GA SA313230066GA SA313230068GA SA313230070GA SA313230071GA SA313230073GA SA313230076GA SA313230077GA SA313230078GA SA313230082GA SA313230085GA SA313230088GA SA313230090GA SA313230093GA SA313230094GA SA313230096GA SA313230097GA SA313230098GA SA313230100GA SA313230102GA SA313230106GA SA313230108GA SA313230109GA SA313230111GA SA313230114GA SA313230115GA SA313230116GA SA313230118GA SA313230119GA SA313230121GA SA313230122GA SA313230124GA SA313230125GA SA313230127GA SA313230128GA SA313230129GA SA313230130GA SA313230131GA SA313230133GA SA313230134GA SA313230135GA SA313230136GA SA313230137GA SA313230138GA SA313230139GA SA313230140GA SA313230141GA SA313230142GA SA313230143GA SA313230144GA SA313230146GA SA313230147GA SA313230148GA SA313230149GA SA313230150GA SA313230153GA SA313230160GA SA313230170GA SA313230172GA SA313230174GA SA313230241GA SA313230242GA SA313230243GA SA313230245GA SA313230247GA SA313230248GA SA313230250GA SA313230251GA SA313230255GA SA313230257GA SA313230258GA SA313230262GA SA313230263GA SA313230265GA SA313230268GA SA313230269GA SA313230270GA SA313230271GA SA313230273GA SA313230277GA SA313230278GA SA313230279GA SA313230280GA SA313230281GA SA313230282GA SA313230283GA SA313230286GA SA313230287GA SA313230288GA SA313230290GA SA313230291GA SA313230292GA SA313230293GA SA313230294GA SA313230295GA SA313230296GA SA313230297GA SA313230298GA SA313230299GA SA313230300GA SA313230301GA SA313230302GA SA313230303GA SA313230304GA SA313230305GA SA313230306GA SA313230308GA SA313230309GA SA313230310GA SA313230311GA SA313230312GA SA313230314GA SA313230315GA SA313230316GA SA313230317GA SA313230318GA SA313230319GA SA313230320GA SA313230321GA SA313230322GA SA313230323GA SA313230324GA SA313230325GA SA313230326GA SA313230328GA SA313230329GA SA313230330GA SA313230331GA SA313230332GA SA313230333GA SA313230334GA SA313230336GA SA313230337GA SA313230338GA SA313230339GA SA313230340GA SA313230342GA SA313230344GA SA313230345GA SA313230346GA SA313230347GA SA313230348GA SA313230349GA SA313230352GA SA313230353GA SA313230354GA SA313230356GA SA313230357GA SA313230359GA SA313230361GA SA313230363GA SA313230366GA SA313230367GA SA313230368GA SA313230369GA SA313230370GA SA313230375GA SA313230379GA SA313230381GA SA313230384GA SA313230388GA SA313230390GA SA313230391GA SA313230392GA SA313230394GA SA313230397GA SA313230399GA SA313230400GA SA313230401GA SA313230402GA SA313230403GA SA313230412GA SA313230413GA SA313230414GA SA313230415GA SA313230417GA SA313230418GA SA313230420GA SA313230421GA SA313230424GA SA313230427GA SA313230432GA SA313230433GA SA313230435GA SA313230436GA SA313230441GA SA313230442GA SA313230444GA SA313230449GA SA313230454GA SA313230461GA SA313230463GA SA313230465GA SA313230469GA SA313230471GA SA313230472GA SA313230475GA SA313230479GA SA313230482GA SA313230483GA SA313230484GA SA313230485GA SA313230486GA SA313230489GA SA313230490GA SA313230493GA SA313230495GA SA313230496GA SA313230499GA SA313230502GA SA313230503GA SA313230505GA SA313230507GA SA313230509GA SA313230510GA SA313230511GA SA313230513GA SA313230514GA SA313230515GA SA313230521GA SA313230523GA SA313230525GA SA313230529GA SA313230532GA SA313230534GA SA313230540GA SA313230542GA SA313230545GA SA313230547GA SA313230548GA SA313230552GA SA313230553GA SA313230555GA SA313230557GA SA313230558GA SA313230560GA SA313230561GA SA313230562GA SA313230563GA SA313230565GA SA313230572GA SA313230580GA SA313230583GA SA313230600GA SA313230606GA SA313230612GA SA313230613GA SA313230617GA SA313230623GA SA313230626GA SA313230627GA SA313230630GA SA313230631GA SA313230632GA SA313230634GA SA313230642GA SA313230644GA SA313230645GA SA313230646GA SA313230649GA SA313230651GA SA313230655GA SA313230657GA SA313230658GA SA313230659GA SA313230661GA SA313230662GA SA313230663GA SA313230665GA SA313230671GA SA313230675GA SA313230678GA SA313230679GA SA313230680GA SA313230685GA SA313230686GA SA313230687GA SA313230693GA SA313230697GA SA313230698GA SA313230699GA SA313230703GA SA313230706GA SA313230707GA SA313230711GA SA313240794GA SA313240902GA SA313240903GA SA313240909GA SA313240910GA SA313240911GA SA313240915GA SA313240916GA SA313240918GA SA313240919GA SA313240920GA SA313240924GA SA313240927GA SA313240928GA SA313240929GA SA313240930GA SA313240931GA SA313240933GA SA313240934GA SA313240936GA SA313240937GA SA313240939GA SA313240940GA SA313240943GA SA313240946GA SA313240948GA SA313240953GA SA313240954GA SA313240957GA SA313240958GA SA313240970GA SA313240973GA SA313240974GA SA313240975GA SA313240981GA SA313240985GA SA313240986GA SA313240991GA SA313241000GA SA313241023GA SA313241044GA SA313251115GA SA313251121GA SA313251149GA SA313251212GA SA313251219GA SA313251309GA SA313251327GA SA313251360GA SA313251498GA SA313251515GA SA313251549GA SA313251550GA SA313251553GA SA313251555GA SA313251560GA SA313251566GA SA313251568GA SA313251570GA SA313251571GA SA313251576GA SA313251577GA SA313251578GA SA313251579GA SA313251580GA SA313251581GA SA313251586GA SA313251587GA SA313251591GA SA313251592GA SA313251593GA SA313251595GA SA313251596GA SA313251598GA SA313251600GA SA313251612GA SA313251617GA SA313251618GA SA313251620GA SA313251621GA SA313251625GA SA313251629GA SA313251630GA SA313251631GA SA313251636GA SA313251658GA SA313251667GA SA313251676GA SA313251679GA SA313251680GA SA313251681GA SA313251683GA SA313251687GA SA313251688GA SA313251689GA SA313251690GA SA313251691GA SA313251692GA SA313251694GA SA313251695GA SA313251696GA SA313261697GA SA313261698GA SA313261699GA SA313261700GA SA313261702GA SA313261703GA SA313261704GA SA313261706GA SA313261707GA SA313261708GA SA313261709GA SA313261710GA SA313261711GA SA313261712GA SA313261713GA SA313261714GA SA313261715GA SA313261716GA SA313261718GA SA313261721GA SA313261727GA SA313261728GA SA313261729GA SA313261731GA SA313261732GA SA313261733GA SA313261734GA SA313261735GA SA313261736GA SA313261737GA SA313261738GA SA313261739GA SA313261740GA SA313261741GA SA313261742GA SA313261743GA SA313261744GA SA313261745GA SA313261746GA SA313261748GA SA313261749GA SA313261751GA SA313261752GA SA313261753GA SA313261754GA SA313261755GA SA313261756GA SA313261757GA SA313261758GA SA313261760GA SA313261762GA SA313261763GA SA313261764GA SA313261765GA SA313261766GA SA313261769GA SA313261770GA SA313261771GA SA313261772GA SA313261773GA SA313261774GA SA313261775GA SA313261776GA SA313261777GA SA313261778GA SA313261779GA SA313261780GA SA313261781GA SA313261782GA SA313261783GA SA313261784GA SA313261785GA SA313261786GA SA313261787GA SA313261788GA SA313261789GA SA313261790GA SA313261791GA SA313261792GA SA313261793GA SA313261794GA SA313261795GA SA313261796GA SA313261798GA SA313261800GA SA313261801GA SA313261802GA SA313261803GA SA313261804GA SA313261805GA SA313261806GA SA313261807GA SA313261808GA SA313261809GA SA313261810GA SA313261811GA SA313261812GA SA313261813GA SA313261814GA SA313261815GA SA313261818GA SA313261820GA SA313261821GA SA313261822GA SA313261823GA SA313261824GA SA313261825GA SA313261827GA SA313261828GA SA313261829GA SA313261830GA SA313261831GA SA313261833GA SA313261836GA SA313261837GA SA313261838GA SA313261839GA SA313261840GA SA313261841GA SA313261842GA SA313261843GA SA313261844GA SA313261845GA SA313261846GA SA313261847GA SA313261848GA SA313261849GA SA313261850GA SA313261851GA SA313261852GA SA313261853GA SA313261854GA SA313261856GA SA313261857GA SA313261858GA SA313261859GA SA313261860GA SA313261861GA SA313261863GA SA313261864GA SA313261865GA SA313261866GA SA313261867GA SA313261868GA SA313261869GA SA313261870GA SA313261871GA SA313261872GA SA313261873GA SA313261875GA SA313261876GA SA313261877GA SA313261878GA SA313261879GA SA313261882GA SA313261883GA SA313261885GA SA313261886GA SA313261887GA SA313261888GA SA313261890GA SA313261891GA SA313261892GA SA313261893GA SA313261894GA SA313261895GA SA313261896GA SA313261897GA SA313261898GA SA313261899GA SA313261900GA SA313261901GA SA313261904GA SA313261906GA SA313261907GA SA313261908GA SA313261909GA SA313261910GA SA313261911GA SA313261913GA SA313261914GA SA313261916GA SA313261917GA SA313261918GA SA313261919GA SA313261920GA SA313261921GA SA313261922GA SA313261923GA SA313261924GA SA313261925GA SA313261927GA SA313261928GA SA313261930GA SA313261931GA SA313261932GA SA313261933GA SA313261934GA SA313261935GA SA313261936GA SA313271967GA SA313271968GA SA313271969GA SA313271971GA SA313271972GA SA313271973GA SA313271974GA SA313271975GA SA313271976GA SA313271977GA SA313271978GA SA313271979GA SA313271981GA SA313271982GA SA313271983GA SA313271985GA SA313271986GA SA313271987GA SA313271988GA SA313271989GA SA313271990GA SA313271991GA SA313271993GA SA313271994GA SA313271995GA SA313271996GA SA313271997GA SA313271999GA SA313272000GA SA313272003GA SA313272004GA SA313272007GA SA313272008GA SA313272009GA SA313272012GA SA313272013GA SA313272015GA SA313272017GA SA313272020GA SA313272021GA SA313272022GA SA313272024GA SA313272025GA SA313272027GA SA313272028GA SA313272029GA SA313272030GA SA313272031GA SA313272032GA SA313272033GA SA313272034GA SA313272035GA SA313272036GA SA313272037GA SA313272038GA SA313272039GA SA313272040GA SA313272041GA SA313272042GA SA313272043GA SA313272044GA SA313272045GA SA313272048GA SA313272051GA SA313272052GA SA313272053GA SA313272054GA SA313272055GA SA313272056GA SA313272061GA SA313272062GA SA313272063GA SA313272068GA SA313272073GA SA313272074GA SA313272075GA SA313272076GA SA313272077GA SA313272079GA SA313272080GA SA313272081GA SA313272082GA SA313272083GA SA313272085GA SA313272087GA SA313272090GA SA313272092GA SA313272093GA SA313272098GA SA313272099GA SA313272110GA SA313272113GA SA313272114GA SA313272115GA SA313272116GA SA313272117GA SA313272118GA SA313272120GA SA313272122GA SA313272123GA SA313272124GA SA313272125GA SA313272127GA SA313272129GA SA313272130GA SA313272131GA SA313272132GA SA313272133GA SA313272134GA SA313272135GA SA313272136GA SA313272137GA SA313272139GA SA313272140GA SA313272141GA SA313272142GA SA313272143GA SA313272144GA SA313272145GA SA313272146GA SA313272149GA SA313272152GA SA313272153GA SA313272154GA SA313272155GA SA313272159GA SA313272160GA SA313272162GA SA313272163GA SA313272164GA SA313272165GA SA313272166GA SA313272168GA SA313272170GA SA313272171GA SA313272172GA SA313272173GA SA313272175GA SA313272176GA SA313272177GA SA313272179GA SA313272181GA SA313272182GA SA313272183GA SA313272184GA SA313272187GA SA313272188GA SA313272189GA SA313272194GA SA313272195GA SA313272196GA SA313272197GA SA313272198GA SA313272199GA SA313272200GA SA313272202GA SA313272203GA SA313272204GA SA313272205GA SA313272206GA SA313272209GA SA313272212GA SA313272219GA SA313272220GA SA313272221GA SA313272222GA SA313272225GA SA313272228GA SA313272229GA SA313272231GA SA313272232GA SA313272234GA SA313272235GA SA313272237GA SA313272238GA SA313272239GA SA313272240GA SA313272242GA SA313272244GA SA313272245GA SA313272246GA SA313272247GA SA313272248GA SA313272249GA SA313272251GA SA313272253GA SA313272255GA SA313272256GA SA313272259GA SA313272260GA SA313272262GA SA313272263GA SA313272264GA SA313272265GA SA313272266GA SA313272268GA SA313272269GA SA313272271GA SA313272272GA SA313272275GA SA313272276GA SA313272278GA SA313272280GA SA313272281GA SA313272283GA SA313272284GA SA313272285GA SA313272287GA SA313272288GA SA313272289GA SA313272290GA SA313272291GA SA313272292GA SA313272293GA SA313272294GA SA313272295GA SA313272296GA SA313272303GA SA313272304GA SA313272305GA SA313272308GA SA313272309GA SA313272310GA SA313272315GA SA313272316GA SA313272323GA SA313272324GA SA313272326GA SA313272327GA SA313272328GA SA313272329GA SA313272330GA SA313272331GA SA313272332GA SA313272333GA SA313272334GA SA313272335GA SA313272336GA SA313272337GA SA313272338GA SA313272339GA SA313272342GA SA313272343GA SA313272345GA SA313272347GA SA313272348GA SA313272349GA SA313272350GA SA313272351GA SA313272353GA SA313272354GA SA313272355GA SA313272356GA SA313272357GA SA313272359GA SA313272361GA SA313272372GA SA313272391GA SA313272396GA SA313272408GA SA313272410GA SA313272419GA SA313272422GA SA313272423GA SA313272430GA SA313272437GA SA313282444GA SA313282451GA SA313282456GA SA313282457GA SA313282460GA SA313282464GA SA313282469GA SA313282470GA SA313282472GA SA313282474GA SA313282483GA SA313282493GA SA313282498GA SA313282500GA SA313282503GA SA313282506GA SA313282512GA SA313282513GA SA313282515GA SA313282516GA SA313282519GA SA313282523GA SA313282524GA SA313282525GA SA313282526GA SA313282534GA SA313282535GA SA313282541GA SA313282546GA SA313282549GA SA313282552GA SA313282556GA SA313282593GA SA313282612GA SA313282621GA SA313282631GA SA313282640GA SA313282655GA SA313282656GA SA313282658GA SA313282664GA SA313282665GA SA313282666GA SA313282667GA SA313282668GA SA313282669GA SA313282670GA SA313282671GA SA313282672GA SA313282673GA SA313282674GA SA313282676GA SA313282677GA SA313282678GA SA313282679GA SA313282680GA SA313282681GA SA313282682GA SA313282683GA SA313282684GA SA313282685GA SA313282686GA SA313282687GA SA313282688GA SA313282689GA SA313282690GA SA313282691GA SA313282692GA SA313282719GA SA313282765GA SA313282798GA SA313282806GA SA313282846GA SA313282852GA SA313282877GA SA313282878GA SA313292908GA SA313292915GA SA313292918GA SA313292923GA SA313292926GA SA313292932GA SA313292934GA SA313292937GA SA313292939GA SA313292946GA SA313292954GA SA313292957GA SA313292958GA SA313292967GA SA313292970GA SA313293030GA SA313293033GA SA313293037GA SA313293046GA SA313293064GA SA313293068GA SA313293072GA SA313293108GA SA313293129GA SA313293154GA SA313293181GA SA313293261GA SA313293294GA SA313293295GA SA313293296GA SA313293297GA SA313293298GA SA313293300GA SA313293301GA SA313293302GA SA313293303GA SA313293304GA SA313293305GA SA313293306GA SA313293308GA SA313293309GA SA313293311GA SA313293312GA SA313293313GA SA313293314GA SA313293315GA SA313293316GA SA313293317GA SA313293318GA SA313293319GA SA313293320GA SA313293321GA SA313293322GA SA313293323GA SA313293324GA SA313293325GA SA313293326GA SA313293327GA SA313293328GA SA313293329GA SA313293330GA SA313293331GA SA313293332GA SA313293333GA SA313293334GA SA313293335GA SA313293336GA SA313293337GA SA313293338GA SA313293339GA SA313293340GA SA313293341GA SA313293342GA SA313293343GA SA313293344GA SA313293345GA SA313293346GA SA313293347GA SA313293348GA SA313293349GA SA313293350GA SA313293351GA SA313293352GA SA313293353GA SA313293354GA SA313293355GA SA313293356GA SA313293357GA SA313293358GA SA313293359GA SA313293360GA SA313293361GA SA313293362GA SA313293363GA SA313293364GA SA313293365GA SA313293366GA SA313293367GA SA313293368GA SA313293369GA SA313293370GA SA313293371GA SA313293372GA SA313293373GA SA313293374GA SA313293375GA SA313293376GA SA313293377GA SA313293378GA SA313293379GA SA313293380GA SA313293381GA SA313293382GA SA313293383GA SA313293384GA SA313293385GA SA313293386GA SA313293388GA SA313293389GA SA313293390GA SA313293391GA SA313293392GA SA313293393GA SA313293394GA SA313293395GA SA313293396GA SA313293397GA SA313293398GA SA313293399GA SA313293400GA SA313293401GA SA313293402GA SA313293403GA SA313293404GA SA313293405GA SA313293406GA SA313293407GA SA313293408GA SA313293409GA SA313293410GA SA313293411GA SA313293412GA SA313293413GA SA313293414GA SA313293415GA SA313293416GA SA313293417GA SA313293418GA SA313293419GA SA313293420GA SA313293421GA SA313293422GA SA313293423GA SA313293424GA SA313293425GA SA313293426GA SA313293427GA SA313293428GA SA313293429GA SA313293430GA SA313293431GA SA313293432GA SA313293433GA SA313293434GA SA313293435GA SA313293436GA SA313293437GA SA313293438GA SA313293439GA SA313293440GA SA313293441GA SA313293442GA SA313293443GA SA313293444GA SA313293445GA SA313293446GA SA313293447GA SA313293448GA SA313293449GA SA313293450GA SA313293451GA SA313293452GA SA313293453GA SA313293454GA SA313293455GA SA313293456GA SA313293458GA SA313293459GA SA313293460GA SA313293461GA SA313293462GA SA313293463GA SA313293465GA SA313293467GA SA313293468GA SA313293469GA SA313293470GA SA313293471GA SA313293472GA SA313293473GA SA313293474GA SA313293475GA SA313293476GA SA313293477GA SA313293478GA SA313293479GA SA313293480GA SA313293481GA SA313293482GA SA313293483GA SA313293484GA SA313293485GA SA313293486GA SA313293487GA SA313293488GA SA313293489GA SA313293490GA SA313293491GA SA313293492GA SA313293493GA SA313293494GA SA313293495GA SA313293496GA SA313293497GA SA313293498GA SA313293500GA SA313293501GA SA313293502GA SA313293503GA SA313293504GA SA313293505GA SA313293506GA SA313293507GA SA313293509GA SA313293510GA SA313293511GA SA313293512GA SA313293513GA SA313293514GA SA313293515GA SA313293516GA SA313293517GA SA313293518GA SA313293519GA SA313293520GA SA313293521GA SA313293522GA SA313293523GA SA313293524GA SA313293525GA SA313293526GA SA313293527GA SA313293528GA SA313293529GA SA313293531GA SA313293532GA SA313293533GA SA313293535GA SA313293536GA SA313293537GA SA313293538GA SA313293539GA SA313293540GA SA313293541GA SA313293542GA SA313293543GA SA313293544GA SA313293545GA SA313293546GA SA313293547GA SA313293548GA SA313303549GA SA313303550GA SA313303551GA SA313303552GA SA313303553GA SA313303554GA SA313303555GA SA313303556GA SA313303557GA SA313303558GA SA313303559GA SA313303560GA SA313303561GA SA313303562GA SA313303563GA SA313303564GA SA313303565GA SA313303566GA SA313303567GA SA313303568GA SA313303569GA SA313303570GA SA313303571GA SA313303572GA SA313303573GA SA313303574GA SA313303575GA SA313303576GA SA313303577GA SA313303578GA SA313303579GA SA313303580GA SA313303581GA SA313303582GA SA313303583GA SA313303584GA SA313303585GA SA313303586GA SA313303587GA SA313303588GA SA313303589GA SA313303590GA SA313303591GA SA313303592GA SA313303593GA SA313303594GA SA313303595GA SA313303596GA SA313303597GA SA313303598GA SA313303600GA SA313303601GA SA313303602GA SA313303603GA SA313303604GA SA313303605GA SA313303606GA SA313303607GA SA313303608GA SA313303609GA SA313303610GA SA313303611GA SA313303612GA SA313303613GA SA313303614GA SA313303615GA SA313303616GA SA313303617GA SA313303618GA SA313303619GA SA313303620GA SA313303621GA SA313303622GA SA313303623GA SA313303624GA SA313303626GA SA313303627GA SA313303628GA SA313303629GA SA313303630GA SA313303631GA SA313303632GA SA313303633GA SA313303634GA SA313303635GA SA313303636GA SA313303637GA SA313303639GA SA313303640GA SA313303641GA SA313303642GA SA313303643GA SA313303644GA SA313303645GA SA313303646GA SA313303647GA SA313303648GA SA313303649GA SA313303650GA SA313303651GA SA313303652GA SA313303653GA SA313303654GA SA313303655GA SA313303656GA SA313303657GA SA313303658GA SA313303659GA SA313303660GA SA313303661GA SA313303662GA SA313303663GA SA313303664GA SA313303665GA SA313303666GA SA313303667GA SA313303668GA SA313303669GA SA313303670GA SA313303671GA SA313303672GA SA313303673GA SA313303674GA SA313303675GA SA313303676GA SA313303677GA SA313303678GA SA313303679GA SA313303680GA SA313303681GA SA313303682GA SA313303683GA SA313303684GA SA313303685GA SA313303687GA SA313303689GA SA313303690GA SA313303691GA SA313303692GA SA313303693GA SA313303694GA SA313303695GA SA313303696GA SA313303698GA SA313303699GA SA313303700GA SA313303701GA SA313303702GA SA313303703GA SA313303704GA SA313303705GA SA313303706GA SA313303707GA SA313303709GA SA313303710GA SA313303711GA SA313303712GA SA313303714GA SA313303715GA SA313303716GA SA313303717GA SA313303718GA SA313303719GA SA313303720GA SA313303721GA SA313303722GA SA313303723GA SA313303724GA SA313303725GA SA313303726GA SA313303728GA SA313303730GA SA313303732GA SA313303733GA SA313303734GA SA313303736GA SA313303737GA SA313303738GA SA313303739GA SA313303740GA SA313303741GA SA313303742GA SA313303743GA SA313303744GA SA313303745GA SA313303746GA SA313303747GA SA313303748GA SA313303749GA SA313303750GA SA313303751GA SA313303752GA SA313303753GA SA313303754GA SA313303755GA SA313303756GA SA313303757GA SA313303758GA SA313303759GA SA313303760GA SA313303761GA SA313303762GA SA313303763GA SA313303764GA SA313303767GA SA313303768GA SA313303769GA SA313303770GA SA313303771GA SA313303772GA SA313303773GA SA313303774GA SA313303775GA SA313303776GA SA313303777GA SA313303778GA SA313303779GA SA313303780GA SA313303781GA SA313303782GA SA313303783GA SA313303784GA SA313303785GA SA313303786GA SA313303787GA SA313303788GA SA313303789GA SA313303790GA SA313303791GA SA313303792GA SA313303793GA SA313303794GA SA313303795GA SA313303797GA SA313303798GA SA313303799GA SA313303800GA SA313303801GA SA313303802GA SA313303803GA SA313303804GA SA313303805GA SA313303806GA SA313303807GA SA313303808GA SA313303809GA SA313303810GA SA313303811GA SA313303812GA SA313303813GA SA313303814GA SA313303815GA SA313303816GA SA313303817GA SA313303818GA SA313303820GA SA313303821GA SA313303822GA SA313303823GA SA313303824GA SA313303825GA SA313303826GA SA313303827GA SA313303828GA SA313303829GA SA313303830GA SA313303832GA SA313303833GA SA313303834GA SA313303835GA SA313303836GA SA313303837GA SA313303838GA SA313303839GA SA313303840GA SA313303841GA SA313303843GA SA313303844GA SA313303845GA SA313303846GA SA313303847GA SA313303848GA SA313303849GA SA313303850GA SA313303851GA SA313303852GA SA313303853GA SA313303854GA SA313303855GA SA313303856GA SA313303857GA SA313303858GA SA313303859GA SA313303860GA SA313303861GA SA313303862GA SA313303863GA SA313303864GA SA313303865GA SA313303866GA SA313303867GA SA313303868GA SA313303869GA SA313303870GA SA313303871GA SA313303872GA SA313303873GA SA313303874GA SA313303875GA SA313303876GA SA313303877GA SA313303878GA SA313303879GA SA313303880GA SA313303881GA SA313303882GA SA313303883GA SA313303884GA SA313303885GA SA313303886GA SA313303887GA SA313303888GA SA313303889GA SA313303890GA SA313303891GA SA313303892GA SA313303893GA SA313303894GA SA313303895GA SA313303896GA SA313303897GA SA313303898GA SA313303899GA SA313303900GA SA313303901GA SA313303902GA SA313303903GA SA313303904GA SA313303905GA SA313303906GA SA313303907GA SA313303909GA SA313303910GA SA313303911GA SA313303913GA SA313303914GA SA313303915GA SA313303916GA SA313303917GA SA313303918GA SA313303919GA SA313303920GA SA313303921GA SA313303922GA SA313303923GA SA313303924GA SA313303925GA SA313303926GA SA313303927GA SA313303928GA SA313303929GA SA313303930GA SA313303931GA SA313303932GA SA313303933GA SA313303936GA SA313303937GA SA313303938GA SA313303939GA SA313303940GA SA313303942GA SA313303943GA SA313303944GA SA313303945GA SA313303946GA SA313303947GA SA313303948GA SA313303949GA SA313303950GA SA313303951GA SA313303952GA SA313303953GA SA313303954GA SA313303955GA SA313303956GA SA313303957GA SA313303958GA SA313303959GA SA313303960GA SA313303961GA SA313303962GA SA313303963GA SA313303964GA SA313303965GA SA313303966GA SA313303967GA SA313303968GA SA313303969GA SA313303970GA SA313303971GA SA313303972GA SA313303973GA SA313303974GA SA313303975GA SA313303976GA SA313303977GA SA313303978GA SA313303979GA SA313303980GA SA313303981GA SA313303982GA SA313303983GA SA313303984GA SA313303985GA SA313303986GA SA313303989GA SA313303990GA SA313303991GA SA313303992GA SA313303993GA SA313303995GA SA313303996GA SA313303997GA SA313303998GA SA313303999GA SA313304001GA SA313304002GA SA313304003GA SA313304004GA SA313304006GA SA313304007GA SA313304008GA SA313304010GA SA313304011GA SA313304012GA SA313304013GA SA313304014GA SA313304015GA SA313304016GA SA313304017GA SA313304018GA SA313304019GA SA313304020GA SA313304021GA SA313304022GA SA313304023GA SA313304025GA SA313304026GA SA313304027GA SA313304028GA SA313304029GA SA313304030GA SA313304031GA SA313304033GA SA313304034GA SA313304035GA SA313304036GA SA313304037GA SA313304038GA SA313304039GA SA313304040GA SA313304041GA SA313304042GA SA313304043GA SA313304044GA SA313304045GA SA313304046GA SA313304047GA SA313304048GA SA313304049GA SA313304050GA SA313304051GA SA313304052GA SA313304053GA SA313304054GA SA313304055GA SA313304056GA SA313304057GA SA313304059GA SA313304060GA SA313304061GA SA313304062GA SA313304063GA SA313304064GA SA313304065GA SA313304066GA SA313304067GA SA313304068GA SA313304069GA SA313304070GA SA313304071GA SA313304072GA SA313304073GA SA313304074GA SA313304075GA SA313304076GA SA313304077GA SA313304078GA SA313304079GA SA313304081GA SA313304082GA SA313304083GA SA313304084GA SA313304085GA SA313304086GA SA313304087GA SA313304088GA SA313304089GA SA313304090GA SA313304091GA SA313304092GA SA313304093GA SA313304094GA SA313304095GA SA313304096GA SA313304097GA SA313304098GA SA313304099GA SA313304100GA SA313304101GA SA313304102GA SA313304103GA SA313304104GA SA313304105GA SA313304106GA SA313304107GA SA313304108GA SA313304109GA SA313304110GA SA313304111GA SA313304112GA SA313304113GA SA313304115GA SA313304116GA SA313304117GA SA313304118GA SA313304119GA SA313304121GA SA313304122GA SA313304123GA SA313304124GA SA313304125GA SA313304126GA SA313304127GA SA313304128GA SA313304129GA SA313304130GA SA313304131GA SA313304132GA SA313304133GA SA313304134GA SA313304135GA SA313304136GA SA313304137GA SA313304138GA SA313304139GA SA313304140GA SA313304141GA SA313304142GA SA313304143GA SA313304144GA SA313304146GA SA313304147GA SA313304148GA SA313304149GA SA313304150GA SA313304151GA SA313304153GA SA313304154GA SA313304155GA SA313304157GA SA313304158GA SA313304159GA SA313304160GA SA313304161GA SA313304162GA SA313304163GA SA313304164GA SA313304165GA SA313304166GA SA313304167GA SA313304168GA SA313304169GA SA313304170GA SA313304172GA SA313304173GA SA313304174GA SA313304175GA SA313304176GA SA313304177GA SA313304178GA SA313304179GA SA313304180GA SA313304181GA SA313304182GA SA313304183GA SA313304184GA SA313304185GA SA313304186GA SA313304187GA SA313304188GA SA313304189GA SA313304190GA SA313304191GA SA313304192GA SA313304193GA SA313304194GA SA313304195GA SA313304196GA SA313304197GA SA313304198GA SA313304199GA SA313304200GA SA313304201GA SA313304202GA SA313304203GA SA313304204GA SA313304205GA SA313304206GA SA313304207GA SA313304208GA SA313304209GA SA313304210GA SA313304211GA SA313304212GA SA313304213GA SA313304214GA SA313304215GA SA313304216GA SA313304217GA SA313304218GA SA313304220GA SA313304221GA SA313304222GA SA313304223GA SA313304225GA SA313304226GA SA313304228GA SA313304229GA SA313304230GA SA313304231GA SA313304232GA SA313304233GA SA313304235GA SA313304236GA SA313304237GA SA313304238GA SA313304239GA SA313304240GA SA313304241GA SA313304242GA SA313304243GA SA313304244GA SA313304245GA SA313304246GA SA313304247GA SA313304248GA SA313304249GA SA313304250GA SA313304251GA SA313304252GA SA313304253GA SA313304254GA SA313304255GA SA313304256GA SA313304257GA SA313304258GA SA313304259GA SA313304260GA SA313304261GA SA313304262GA SA313304263GA SA313304264GA SA313304265GA SA313304266GA SA313304268GA SA313304269GA SA313304270GA SA313304271GA SA313304272GA SA313304273GA SA313304275GA SA313304276GA SA313304277GA SA313304278GA SA313304279GA SA313304280GA SA313304281GA SA313304282GA SA313304283GA SA313304284GA SA313304285GA SA313304286GA SA313304287GA SA313304288GA SA313304289GA SA313304291GA SA313304292GA SA313304293GA SA313304294GA SA313304295GA SA313304296GA SA313304297GA SA313304298GA SA313304299GA SA313304300GA SA313304301GA SA313304302GA SA313304303GA SA313304304GA SA313304306GA SA313304308GA SA313304309GA SA313304310GA SA313304311GA SA313304312GA SA313304313GA SA313304314GA SA313304315GA SA313304316GA SA313304317GA SA313304320GA SA313304321GA SA313304322GA SA313304323GA SA313304324GA SA313304325GA SA313304326GA SA313304327GA SA313304328GA SA313304329GA SA313304330GA SA313304331GA SA313304332GA SA313304333GA SA313304335GA SA313304336GA SA313304337GA SA313304338GA SA313304339GA SA313304340GA SA313304341GA SA313304342GA SA313304343GA SA313304344GA SA313304345GA SA313304346GA SA313304347GA SA313304348GA SA313304349GA SA313304350GA SA313304351GA SA313304352GA SA313304353GA SA313304354GA SA313304355GA SA313304356GA SA313304357GA SA313304358GA SA313304360GA SA313304361GA SA313304363GA SA313304364GA SA313304366GA SA313304367GA SA313304368GA SA313304369GA SA313304370GA SA313304371GA SA313304372GA SA313304373GA SA313304374GA SA313304375GA SA313304376GA SA313304377GA SA313304378GA SA313304379GA SA313304380GA SA313304381GA SA313304382GA SA313304383GA SA313304384GA SA313304385GA SA313304386GA SA313304387GA SA313304388GA SA313304389GA SA313304390GA SA313304391GA SA313304392GA SA313304393GA SA313304394GA SA313304395GA SA313304396GA SA313304397GA SA313304398GA SA313304399GA SA313304400GA SA313304401GA SA313304402GA SA313304403GA SA313304404GA SA313314405GA SA313314406GA SA313314407GA SA313314408GA SA313314409GA SA313314410GA SA313314411GA SA313314412GA SA313314415GA SA313314416GA SA313314417GA SA313314418GA SA313314419GA SA313314420GA SA313314421GA SA313314422GA SA313314424GA SA313314426GA SA313314427GA SA313314428GA SA313314429GA SA313314430GA SA313314431GA SA313314432GA SA313314433GA SA313314453GA SA313314459GA SA313314485GA SA313314497GA SA313314510GA SA313314513GA SA313314514GA SA313314561GA SA313314570GA SA313314575GA sa313452143ga SA312080395GA SA312294403GA SA312294444GA SA312294552GA SA312294631GA SA312294653GA SA312479765GA SA312479861GA SA312479971GA SA312491415GA SA313023411GA SA313095322GA SA313095542GA SA308459672GA SA313105989GA SA313106007GS SA311069748GA SA311069761GA SA311069838GA SA311132559GA SA311132567GA SA311132576GA SA311142889GA SA311142895GA SA311142899GA SA312080772GA SA313292985GA SA313314842GA SA313314867GA SA313314873GA SA313314876GA SA313314883GA sa313409167ga sa313419544ga SA313431015GA SA313431023GA SA313431024GA SA313431029GA SA313431035GA SA313431038GA SA313431051GA SA313431052GA SA313431057GA SA313431070GA SA313431071GA SA313431075GA SA313431076GA SA313431078GA SA313431080GA SA313431092GA SA313431102GA SA313431112GA SA313431114GA SA313441441GA SA313441452GA sa313452121ga SA314045522GA SA314055556GA SA314075999GA SA314295102GA SA314295118GA SA307514445GA SA307514458GA SA308084687GA SA308084716GA SA308084723GA SA308094846GA SA308094885GA SA308094899GA SA308094913GA SA308094914GA SA308094919GA SA308094935GA SA308094946GA SA308135524GA SA308378250GA SA308378274GA SA308388506GA SA308388540GA SA308398656GA SA308398663GA SA308398713GA SA308398718GA SA308398739GA SA308398751GA SA308398764GA SA308398864GA SA308398869GA SA308398870GA SA308479903GA SA308479906GA SA308481001GA SA308481006GA SA308481007GA SA308481008GA SA308481013GA SA308481016GA SA308481017GA SA308481018GA SA308481022GA SA308481023GA SA308481027GA SA308481030GA SA308481033GA SA308481048GA SA308489928GA SA308489959GA SA308489961GA SA308489974GA SA308489990GA SA308489991GA SA308491175GA SA308491242GA SA308501429GA SA309032109GA SA309062268GA SA309062270GA SA309062272GA SA309062287GA SA309062289GA SA309062290GA SA309062291GA SA309062293GA SA309062294GA SA309062301GA SA309062306GA SA309062307GA SA309062311GA SA309062314GA SA309062316GA SA309062326GA SA309062328GA SA309072354GA SA309072361GA SA309072376GA SA309253857GA SA309263950GA SA309263955GA SA309294043GA SA309294052GA SA309294094GA SA309324472GA SA309324473GA SA309436319GA SA309446424GA SA309446466GA SA309446489GA SA309446499GA SA309446518GA SA309456537GA SA309456544GA SA309456561GA SA309456648GA SA309456661GA SA309456678GA SA309456679GA SA309456686GA SA309456722GA SA309456724GA SA309456725GA SA309456734GA SA309466751GA SA309476936GA SA309476963GA SA309476987GA SA309477033GA SA309477045GA SA309477064GA SA309477085GA SA309477091GA SA309477092GA SA309477093GA SA309477098GA SA309477103GA SA309477116GA SA309477133GA SA309497226GA SA309497274GA SA309497280GA SA309497281GA SA309497287GA SA309497302GA SA309497303GA SA309497304GA SA309497305GA SA309497306GA SA309497307GA SA309497310GA SA309497311GA SA309507337GA SA309517468GA SA309517482GA SA309517485GA SA309517512GA SA309517536GA SA309517665GA SA309517680GA SA309517682GA SA309517699GA SA309517712GA SA309517717GA SA309517720GA SA309517725GA SA309517726GA SA309517729GA SA309517769GA SA309517774GA SA309517779GA SA309517782GA SA309517783GA SA309517785GA SA309517792GA SA309527990GA SA309538022GA SA309538064GA SA309538071GA SA309538077GA SA309538082GA SA310028150GA SA310028152GA SA310028157GA SA310028158GA SA310028161GA SA310028163GA SA310028164GA SA310028165GA SA310028167GA SA310028169GA SA310028177GA SA310028183GA SA310028202GA SA310028215GA SA310028216GA SA310028220GA SA310028240GA SA310028243GA SA310028255GA SA310028256GA SA310028263GA SA310028267GA SA310028286GA SA310028287GA SA310028298GA SA310028304GA SA310028308GA SA310038350GA SA310038353GA SA310038355GA SA310038360GA SA310038363GA SA310038364GA SA310038365GA SA310038367GA SA310038368GA SA310038369GA SA310038376GA SA310038381GA SA310038394GA SA310038433GA SA310038434GA SA310038445GA SA310038448GA SA310038450GA SA310038466GA SA310038484GA SA310038488GA SA310038490GA SA310038493GA SA310038500GA SA310038506GA SA310048552GA SA310048559GA SA310048560GA SA310048572GA SA310048577GA SA310048580GA SA310048638GA SA310048645GA SA310048646GA SA310048678GA SA310068827GA SA310068845GA SA310068846GA SA310068857GA SA310068859GA SA310068863GA SA310068865GA SA310068866GA SA310068871GA SA310068872GA SA310068876GA SA310068881GA SA310068882GA SA310068885GA SA310068889GA SA310068890GA SA310068896GA SA310068897GA SA310068900GA SA310068901GA SA310068904GA SA310068908GA SA310068911GA SA310068916GA SA310068918GA SA310068923GA SA310068930GA SA310068933GA SA310068934GA SA310068936GA SA310068940GA SA310068942GA SA310068945GA SA310068948GA SA310068950GA SA310119016GA SA310139081GA SA310139169GA SA310139187GA SA310139189GA SA310139192GA SA310139195GA SA310139197GA SA310139199GA SA310139208GA SA310139211GA SA310139221GA SA310139225GA SA310139226GA SA310139227GA SA310139229GA SA310139230GA SA310139236GA SA310139237GA SA310139239GA SA310139240GA SA310139241GA SA310139253GA SA310149330GA SA310149331GA SA310149332GA SA310149337GA SA310159443GA SA310169484GA SA310169504GA SA310199750GA SA310219806GA SA310219842GA SA310259915GA SA310259917GA SA310259919GA SA310259921GA SA310259922GA SA310259924GA SA310259925GA SA310259926GA SA310259927GA SA310259930GA SA310259978GA SA310259983GA SA310259984GA SA310259990GA SA310260045GA SA310260048GA SA310260052GA SA310260058GA SA310260070GA SA310260076GA SA310260077GA SA310260220GA SA310260228GA SA310260234GA SA310260235GA SA310260238GA SA310260251GA SA310260476GA SA310260477GA SA310260478GA SA310260481GA SA310260483GA SA310260484GA SA310260525GA SA310260530GA SA310260535GA SA310260537GA SA310260538GA SA310260548GA SA310260549GA SA310260550GA SA310260551GA SA310260552GA SA310310842GA SA310310843GA SA310310847GA SA310310850GA SA310310851GA SA310310856GA SA310310860GA SA310310863GA SA310310879GA SA310310889GA SA310310906GA SA310321052GA SA310331225GA SA310331227GA SA310331229GA SA310331233GA SA310331236GA SA310331239GA SA310331241GA SA310331244GA SA310331246GA SA310331252GA SA310331254GA SA310331258GA SA310331262GA SA310331269GA SA310331275GA SA310331282GA SA310331286GA SA310331290GA SA310331291GA SA310341407GA SA310341409GA SA310341416GA SA310341418GA SA310341421GA SA310341423GA SA310341433GA SA310341452GA SA310361600GA SA310361646GA SA310361905GA SA310372000GA SA310372002GA SA310372006GA SA310372011GA SA310372013GA SA310372014GA SA310372019GA SA310372020GA SA310372031GA SA310372033GA SA310372038GA SA310372041GA SA310372042GA SA310372045GA SA310372046GA SA310372047GA SA310372050GA SA310372051GA SA310372056GA SA310372057GA SA310372141GA SA310372143GA SA310372148GA SA310372187GA SA310372200GA SA310382357GA SA310382445GA SA310382505GA SA310403355GA SA310444298GA SA310475466GA SA310475488GA SA310475519GA SA310475527GA SA310475529GA SA310475542GA SA310475543GA SA310475551GA SA310475586GA SA310475597GA SA310495895GA SA310495898GA SA310495902GA SA310495905GA SA310495907GA SA310495910GA SA310495914GA SA310495915GA SA310495917GA SA310495937GA SA310495944GA SA310495947GA SA310495948GA SA310495949GA SA310495951GA SA310495952GA SA310495958GA SA310495960GA SA310495961GA SA310495969GA SA310495971GA SA310495972GA SA310495974GA SA310495975GA SA310517710GA SA310517787GA SA310517833GA SA311039444GA SA311060255GA SA311060270GA SA311060271GA SA311060297GA SA311060298GA SA311060303GA SA311060308GA SA311060343GA SA311060385GA SA311060397GA SA311080670GA SA311080679GA SA311122199GA SA311142953GA SA311174163GA SA311174197GA SA311205447GA SA311205489GA SA311246481GA SA311256842GA SA311307416GA SA311307451GA SA311307573GA SA311317601GA SA311317604GA SA311317636GA SA311317665GA SA311317721GA SA311317726GA SA311317731GA SA311327844GA SA311327854GA SA311327859GA SA311327860GA SA311327862GA SA311327877GA SA311327894GA SA311458275GA SA311458310GA SA311458317GA SA311458322GA SA311468423GA SA311468434GA SA311488452GA SA311488461GA SA311488489GA SA311508571GA SA311508579GA SA311508580GA SA311508583GA SA311508594GA SA311508595GA SA311508601GA SA311508602GA SA311508603GA SA311508604GA SA311508605GA SA311508606GA SA311508607GA SA311508608GA SA311508609GA SA311508610GA SA311508611GA SA311508612GA SA311508613GA SA311508615GA SA311508617GA SA311508618GA SA311508619GA SA311508620GA SA311508621GA SA311508622GA SA311508628GA SA311508629GA SA311508630GA SA311508631GA SA311508632GA SA311508633GA SA311508634GA SA311508635GA SA311508638GA SA311508639GA SA311518691GA SA311518696GA SA312018821GA SA312069699GA SA312069702GA SA312069703GA SA312069704GA SA312069708GA SA312069710GA SA312069811GA SA312070138GA SA312080387GA SA312080404GA SA312080406GA SA312080413GA SA312080414GA SA312080474GA SA312080586GA SA312080588GA SA312080619GA SA312080637GA SA312080644GA SA312080650GA SA312080654GA SA312080665GA SA312080669GA SA312080672GA SA312080676GA SA312080697GA SA312080703GA SA312080738GA SA312121890GA SA312344858GA SA312344867GA SA312344869GA SA312344876GA SA312344879GA SA312344880GA SA312344882GA SA312344885GA SA312344888GA SA312344889GA SA312344893GA SA312344899GA SA312344906GA SA312344907GA SA312344918GA SA312344920GA SA312344926GA SA312344927GA SA312344962GA SA312344966GA SA312365564GA SA312365584GA SA312365617GA SA312365624GA SA312365626GA SA312365629GA SA312365637GA SA312365645GA SA312365646GA SA312386338GA SA312386366GA SA312417393GA SA312447459GA SA312457707GA SA312470077GA SA312479524GA SA312479853GA SA312479967GA SA312480730GA SA312480731GA SA312491081GA SA312491174GA SA312491180GA SA312491216GA SA312491222GA SA312491239GA SA312491301GA SA312491304GA SA312491354GA SA312491359GA SA312491370GA SA312491383GA SA312491406GA SA312491409GA SA312491412GA SA312491413GA SA312491414GA SA312491419GA SA312491421GA SA312491424GA SA312491428GA SA312491429GA SA312491430GA SA312491433GA SA312491438GA SA312491442GA SA312491444GA SA312491450GA SA312491452GA SA312491454GA SA312501959GA SA312502057GA SA312502058GA SA312502067GA SA312502069GA SA312502073GA SA312502075GA SA312502088GA SA312502097GA SA312502108GA SA312502109GA SA312502110GA SA312502115GA SA312502121GA SA312502125GA SA312502136GA SA312502146GA SA312502147GA SA312502148GA SA312502165GA SA312502166GA SA313043889GA SA313043900GA SA313043910GA SA313043914GA SA313043930GA SA313043932GA SA313043935GA SA313043942GA SA313064173GA SA313064208GA SA313064212GA SA313064216GA SA313064274GA SA313064280GA SA313064314GA SA313064334GA SA313064351GA SA313064378GA SA313084795GA SA313095167GA SA313095173GA SA313095181GA SA313095183GA SA313095184GA SA313095187GA SA313095189GA SA313095190GA SA313095191GA SA313095192GA SA313095194GA SA313095589GA SA313105660GA SA313105666GA SA313105672GA SA313105872GA SA313116461GA SA313116499GA SA313116505GA SA313116506GA SA313116507GA SA313116539GA SA313116642GA SA313127296GA SA313127390GA SA313127391GA SA313127393GA SA313127394GA SA313127409GA SA313127578GA SA313127593GA SA313127613GA SA313127618GA SA313127630GA SA313127631GA SA313127633GA SA313127634GA SA313127635GA SA313127638GA SA313127641GA SA313127644GA SA313127647GA SA313127648GA SA313127651GA SA313127652GA SA313127654GA SA313127655GA SA313127656GA SA313127659GA SA313127661GA SA313127662GA SA313137776GA SA313147965GA SA313208859GA SA313208868GA SA313208951GA SA313208954GA SA313208956GA SA313251551GA SA313251552GA SA313251554GA SA313251556GA SA313251558GA SA313251559GA SA313251561GA SA313251563GA SA313251564GA SA313251565GA SA313251569GA SA313251572GA SA313251573GA SA313251574GA SA313251575GA SA313251583GA SA313251588GA SA313251589GA SA313251590GA SA313251594GA SA313251597GA SA313251601GA SA313251602GA SA313251604GA SA313251608GA SA313251613GA SA313251615GA SA313251616GA SA313251624GA SA313251674GA SA313261705GA SA313261725GA SA313261826GA SA313272273GA SA313282468GA SA313282629GA SA313282654GA SA313282696GA SA313282708GA SA313282718GA SA313282728GA SA313282776GA SA313282818GA SA313282829GA SA313282838GA SA313282880GA SA313314503GA SA313314674GA SA313335010GA SA313335012GA SA313335053GA SA313335063GA SA313335087GA SA313335105GA SA313335132GA SA313335158GA SA313335164GA SA313335204GA SA313335215GA SA313335243GA SA313335369GA SA313335381GA SA313335394GA SA313335397GA SA313335399GA SA313335443GA SA313335510GA SA313335511GA SA313335514GA SA313335515GA SA313335517GA SA313335522GA SA313335524GA SA313345530GA SA313345534GA SA313345582GA SA313345594GA SA313345596GA SA313345624GA SA313345628GA SA313345632GA SA313345637GA SA313345661GA SA313345699GA SA313345705GA SA313345719GA SA313345738GA SA313345747GA SA313345760GA SA313345762GA SA313345775GA SA313345777GA SA313345791GA SA313366729GA SA313378226GA sa313378252ga sa313388537ga SA313419534GA SA313420248GA SA313420304GA SA313420305GA SA313420308GA SA313420316GA SA313420324GA SA313420327GA SA313420372GA SA313420375GA SA313420381GA SA313420514GA SA313420584GA SA313430708GA SA313430733GA SA313430756GA SA313430775GA SA313430803GA SA313430814GA SA313430816GA SA313430817GA SA313430826GA SA313430829GA SA313430830GA SA313430868GA SA313430869GA SA313430870GA SA313430871GA SA313430872GA SA313430873GA SA313430874GA SA313430876GA SA313430878GA SA313430879GA SA313430880GA SA313430881GA SA313430882GA SA313430883GA SA313430884GA SA313430885GA SA313430886GA SA313430887GA SA313430888GA SA313430889GA SA313430890GA SA313430891GA SA313430892GA SA313430893GA SA313430897GA SA313430900GA SA313430901GA SA313430905GA SA313430906GA SA313430908GA SA313430909GA SA313430910GA SA313430911GA SA313430912GA SA313430913GA SA313430914GA SA313430918GA SA313430937GA SA313430938GA SA313430940GA SA313430947GA SA313430948GA SA313430952GA SA313430953GA SA313430956GA SA313430959GA SA313430960GA SA313430962GA SA313430963GA SA313430964GA SA313430965GA SA313430968GA SA313430969GA SA313430978GA SA313430979GA SA313430984GA SA313430988GA SA313430989GA SA313430990GA SA313430991GA SA313430995GA SA313430996GA SA313430999GA SA313431096GA SA313431100GA SA313431109GA SA313431117GA SA313431130GA SA313431151GA SA313431160GA SA313431163GA SA313431165GA SA313431181GA SA313431230GA SA313441489GA SA313441491GA SA313441495GA SA313441499GA SA313441500GA SA313441504GA SA313441506GA SA313441508GA SA313441606GA SA313441608GA SA313441619GA SA313441625GA SA313441627GA SA313441628GA SA313441667GA SA313441669GA SA313441672GA SA313441680GA SA313451843GA SA313451845GA SA313451885GA SA313451886GA SA313451888GA SA313451889GA SA313451891GA SA313451892GA SA313451902GA SA313451903GA SA313451904GA SA313451915GA SA313451917GA SA313451918GA SA313451920GA SA313451921GA SA313451926GA SA313451930GA SA313451931GA SA313451944GA SA313451947GA SA313451950GA SA313451956GA SA313451967GA SA313462156GA SA313462161GA SA313462163GA SA313462167GA SA313462170GA SA313462176GA SA313462179GA SA313462180GA SA313462185GA sa313462190ga SA313462191GA SA313462192GA SA313462207GA SA313462212GA SA313462216GA SA313462251GA SA313462252GA SA313462254GA SA313462260GA SA313462261GA SA313462265GA SA313462268GA SA313462270GA SA313462271GA SA313462272GA SA313462274GA SA313482607GA SA313482610GA SA313482612GA SA313482625GA SA313482627GA SA313482631GA SA313482634GA SA313482646GA SA313482647GA SA313482652GA SA313482655GA SA313482660GA SA313482668GA SA313482672GA SA313482684GA SA313482699GA SA313482704GA SA313482749GA SA313482784GA SA313482795GA SA313513839GA SA313513861GA SA313513867GA SA313513895GA SA313513929GA SA313523947GA SA314014485GA SA314014486GA SA314014489GA SA314014497GA SA314014500GA SA314014505GA SA314014509GA SA314014536GA SA314014539GA SA314014553GA SA314014645GA SA314014688GA SA314014691GA sa314024902ga SA314024959GA SA314024960GA SA314024963GA SA314025108GA SA314025131GA SA314025133GA SA314025137GA SA314025138GA SA314025169GA SA314045475GA SA314045506GA SA314055544GA SA314055568GA SA314055570GA SA314055575GA SA314055583GA SA314055602GA SA314055604GA SA314055609GA SA314055620GA SA314055637GA SA314055657GA SA314055658GA SA314055659GA SA314055664GA SA314055668GA SA314055685GA SA314055687GA SA314055689GA SA314055692GA SA314055694GA SA314055695GA SA314055697GA SA314055700GA SA314055704GA SA314055705GA SA307070444GA SA307190726GA SA307190767GA SA307231253GA SA307231415GA SA307231421GA SA307291873GA SA307302007GA SA307302019GA SA307302027GA SA307302030GA SA307302032GA SA307302033GA SA307302034GA SA307302038GA SA307302040GA SA307302041GA SA307302042GA SA307302044GA SA307302045GA SA307322127GA SA307322128GA SA307322129GA SA307322130GA SA307322131GA SA307322132GA SA307322134GA SA307322135GA SA307322136GA SA307322167GA SA307322168GA SA307322169GA SA307322170GA SA307322171GA SA307322173GA SA307322174GA SA307322177GA SA307322178GA SA307322180GA SA307322181GA SA307332208GA SA307332211GA SA307332212GA SA307332214GA SA307332217GA SA307332218GA SA307332219GA SA307332222GA SA307332226GA SA307332248GA SA307382548GA SA308297060GA SA308297064GA SA308337646GA SA308337647GA SA308337648GA SA308337652GA SA308337667GA SA308337673GA SA308337703GA SA308337706GA SA308337707GA SA308337709GA SA308337710GA SA308337711GA SA308337712GA SA308337713GA SA308337714GA SA308337715GA SA308337716GA SA308337717GA SA308337718GA SA308337719GA SA308337720GA SA308337721GA SA308337722GA SA308347772GA SA308347798GA SA309062280GA SA309253725GA SA309375462GA SA309385568GA SA309385576GA SA309385595GA SA309385614GA SA309385630GA SA309385658GA SA309385662GA SA309385666GA SA309385667GA SA309395712GA SA309395716GA SA309395718GA SA309395719GA SA309395723GA SA309395725GA SA309395731GA SA309395734GA SA309395749GA SA309395750GA SA309395771GA SA309395789GA SA309405820GA SA309405833GA SA309405839GA SA309426024GA SA309426062GA SA309426068GA SA309426072GA SA309426073GA SA309426074GA SA309426076GA SA309426080GA SA309426082GA SA309426088GA SA309426092GA SA309426094GA SA309426097GA SA309426098GA SA309426101GA SA309426118GA SA309426123GA SA309426126GA SA309426135GA SA309426138GA SA309426142GA SA309426143GA SA309426145GA SA309426151GA SA309426153GA SA309426164GA SA309436169GA SA309436173GA SA309436174GA SA309436175GA SA309436185GA SA309436193GA SA309436209GA SA309436211GA SA309436213GA SA309436214GA SA309436216GA SA309436221GA SA309436222GA SA309436225GA SA309436267GA SA309436276GA SA309517631GA SA309517633GA SA309517732GA SA310038455GA SA310048594GA SA310199722GA SA310199748GA SA310260223GA SA310260349GA SA310300775GA SA310300782GA SA310310844GA SA310310849GA SA310310858GA SA310310869GA SA310310870GA SA310310873GA SA310310874GA SA310310884GA SA310310891GA SA310310892GA SA310310893GA SA310310894GA SA310310911GA SA310321065GA SA310321071GA SA310413485GA SA310413486GA SA310423653GA SA310423687GA SA310423843GA SA310423928GA SA310423948GA SA310423955GA SA310423965GA SA310423968GA SA310423998GA SA310424005GA SA310424007GA SA310424011GA SA310424015GA SA310424016GA SA310424018GA SA310424020GA SA310424021GA SA310424022GA SA310424023GA SA310424024GA SA310475493GA SA310475500GA SA310475559GA SA310475560GA SA310475561GA SA310475563GA SA310475605GA SA310475659GA SA310475712GA SA310485769GA SA310485803GA SA310485817GA SA310485831GA SA310485852GA SA310485864GA SA310485865GA SA310485866GA SA310485872GA SA310485875GA SA310485887GA SA310496137GA SA310496160GA SA310496287GA SA310496318GA SA310496344GA SA310496463GA SA310496473GA SA310496487GA SA310496491GA SA310496494GA SA310496495GA SA310506663GA SA310506745GA SA310506845GA SA310506851GA SA310507019GA SA310507029GA SA310507051GA SA310507057GA SA310507097GA SA310507117GA SA310507148GA SA311060156GA SA311060225GA SA311101275GA SA311122037GA SA311122038GA SA311132583GA SA311287095GA SA311287163GA SA311297230GA SA311297248GA SA311297252GA SA311297295GA SA311297296GA SA311297305GA SA311297308GA SA311297309GA SA311297312GA SA311297315GA SA311297317GA SA311297318GA SA311297319GA SA311297320GA SA311297323GA SA311307459GA SA311307477GA SA311307546GA SA311317609GA SA311317632GA SA311317674GA SA311317686GA SA311317744GA SA311317753GA SA311317756GA SA311317760GA SA311327856GA SA311327905GA SA311327914GA SA311327919GA SA311327925GA SA311337978GA SA311428023GA SA311468393GA SA311468398GA SA311468401GA SA311468403GA SA312039073GA SA312039116GA SA312039119GA SA312039124GA SA312039145GA SA312039146GA SA312039160GA SA312039170GA SA312039176GA SA312080842GA SA312080875GA SA312152493GA SA312294452GA SA312294462GA SA312365798GA SA312375891GA SA312375892GA SA312375893GA SA312375894GA SA312375896GA SA312375897GA SA312375899GA SA312375902GA SA312375903GA SA312375905GA SA312375906GA SA312375908GA SA312375909GA SA312375910GA SA312375911GA SA312375915GA SA312375917GA SA312375918GA SA312375919GA SA312375921GA SA312375922GA SA312375925GA SA312375929GA SA312375931GA SA312375933GA SA312375936GA SA312375943GA SA312375945GA SA312375946GA SA312375948GA SA312375949GA SA312376025GA SA312376289GA SA312376294GA SA312378012GA SA312378014GA SA312378015GA SA312378018GA SA312378020GA SA312378022GA SA312378026GA SA312378028GA SA312397074GA SA312397083GA SA312397087GA SA312447451GA SA312447458GA SA312447514GA SA312458094GA SA312458466GA SA312470006GA SA312479883GA SA312479887GA SA312490875GA SA312490899GA SA312491318GA SA312491331GA SA312491338GA SA312491352GA SA312491366GA SA312491427GA SA312491434GA SA312491441GA SA312491453GA SA312491465GA SA312491483GA SA312491495GA SA312491510GA SA312491543GA SA313023259GA SA313084718GA SA313084761GA SA313094985GA SA313094999GA SA313095010GA SA313095031GA SA313095042GA SA313095066GA SA313095067GA SA313095068GA SA313095069GA SA313095071GA SA313095075GA SA313095090GA SA313095093GA SA313095094GA SA313095096GA SA313095265GA SA313095277GA SA313105945GA SA313106049GA SA313106105GA SA313116586GA SA313116593GA SA313116595GA SA313116605GA SA313116644GA SA313116913GA SA313116924GA SA313116927GA SA313116965GA SA313116967GA SA313117100GA SA313117193GA SA313127444GA SA313198162GA SA313198203GA SA313198206GA SA313198207GA SA313198209GA SA313198210GA SA313198211GA SA313198214GA SA313198216GA SA313198217GA SA313198360GA SA313198364GA SA313198376GA SA313198394GA SA313198396GA SA313198397GA SA313198399GA SA313198400GA SA313198403GA SA313261747GA SA313261926GA SA313272138GA SA313272156GA SA313272158GA SA313272180GA SA313272185GA SA313282481GA SA313282517GA SA313282843GA SA313292898GA SA313292900GA SA313292956GA SA313292959GA SA313314495GA SA313314500GA SA313314523GA SA313314527GA SA313314746GA SA313335024GA SA313335039GA SA313335146GA SA313335147GA SA313335218GA SA313335393GA SA313335447GA SA313335523GA SA313409418GA SA313419742GA SA313419760GA SA313419875GA SA313419904GA SA313420091GA SA313420108GA SA313420153GA SA313420177GA SA313420195GA SA313420196GA SA313420204GA SA313420205GA SA313420206GA SA313420207GA SA313420212GA SA313420213GA SA313420215GA SA313420216GA SA313420218GA SA313420221GA SA313420244GA SA313420311GA SA313420487GA SA313420505GA SA313420508GA SA313420537GA SA313420565GA SA313420566GA SA313420569GA SA313420570GA SA313420583GA SA313420585GA SA313420586GA SA313420589GA SA313420590GA SA313420592GA SA313420593GA SA313420594GA SA313420599GA SA313420604GA SA313420605GA SA313420611GA SA313420612GA SA313420615GA SA313430867GA SA313430941GA SA313451723GA SA313451742GA SA314014323GA SA314014364GA SA314014427GA sa314014432ga SA314014434GA SA314014472GA SA314014476GA SA314014685GA SA314014694GA SA314014721GA SA314014746GA SA314014753GA SA314024775GA SA314025069GA SA314025099GA SA314025104GA SA314025118GA SA314075975GA SA314076022GA SA314086363GA SA314086490GA SA314086522GA SA314097242GA SA314097250GA SA314097253GA SA314097269GA SA314097286GA SA314097293GA SA314097303GA SA314097312GA SA314097319GA SA314107459GA SA314107461GA SA314107467GA SA314129257GA SA314129259GA SA314129263GA SA314129266GA SA314129280GA SA314129289GA SA314129293GA SA314129294GA SA314201594GA SA314211940GA SA307231249GA SA308409049GA SA308409053GA SA308409070GA SA308409153GA SA308409155GA SA308409158GA SA308409159GA SA308409169GA SA307302029GA SA307352358GA SA307352362GA SA307352366GA SA307362390GA SA308206697GA SA308206721GA SA308388611GA SA308398854GA SA309153042GA SA309253738GA SA312501863GA SA312501942GA SA312501945GA SA312501947GA SA312501950GA SA312501958GA SA312502056GA SA312502059GA SA312502066GA SA312502070GA SA312502084GA SA312502096GA SA312502102GA SA312502107GA SA312502145GA SA308084763GA SA308094973GA SA308125388GA SA308125389GA SA308125391GA SA308125392GA SA308125398GA SA308125415GA SA308125417GA SA308125422GA SA308135487GA SA308155880GA SA308155901GA SA308186379GA SA308186408GA SA308186411GA SA308186419GA SA308186428GA SA308186430GA SA308186434GA SA308186435GA SA308186436GA SA308186438GA SA308186439GA SA308186440GA SA308186443GA SA308196460GA SA308196461GA SA308196462GA SA308196479GA SA308196481GA SA308196482GA SA308196483GA SA308196484GA SA308196489GA SA308196490GA SA308196496GA SA308196534GA SA308196552GA SA308196554GA SA308196565GA SA308429321GA SA308429328GA SA308429340GA SA308449503GA SA308449506GA SA308449510GA SA308449514GA SA308449520GA SA308459651GA SA308459675GA SA308459860GA SA308459868GA SA308479891GA SA308479894GA SA308479896GA SA308479899GA SA309072345GA SA309072358GA SA309132809GA SA310506760GA SA310506780GA SA310506795GA SA310506805GA SA310506834GA SA310507168GA SA310507169GA SA310507184GA SA310507186GA SA310507239GA SA310507248GA SA310507250GA SA310507256GA SA310507286GA SA310507295GA SA310507368GA SA310507370GA SA310528850GA SA311142787GA SA311317624GA SA311317677GA SA311317684GA SA311317746GA SA311428031GA SA311448178GA SA312069789GA SA312070077GA SA312070127GA SA312070128GA SA312070144GA SA312070160GA SA312070181GA SA312070185GA SA312070189GA SA312070192GA SA312070203GA SA312070205GA SA312070207GA SA312070213GA SA312070218GA SA312070223GA SA312070224GA SA312070228GA SA312070229GA SA312070234GA SA312070235GA SA312070238GA SA312070239GA SA312070240GA SA307080531GA SA307150677GA SA309062331GA SA309203368GA SA309203378GA SA311060186GA SA311060212GA SA311060219GA SA311060250GA SA311060256GA SA311060311GA SA311060316GA SA311060317GA SA311060341GA SA311060350GA SA311070516GA SA311070548GA SA311070592GA SA309507339GA SA309517667GA SA309517677GA SA309517697GA SA309517702GA SA309517719GA SA309517721GA SA309517723GA SA309517768GA SA309517771GA SA309517791GA SA309517806GA SA309517818GA SA309517825GA SA309517836GA SA309517868GA SA309517874GA SA309517884GA SA309517886GA SA309517887GA SA309517897GA SA309527925GA SA309527926GA SA309527935GA SA309527936GA SA309527944GA SA309527947GA SA309527949GA SA309527950GA SA309527952GA SA309527960GA SA309527963GA SA309527966GA SA309527970GA SA309527971GA SA309527988GA SA309538017GA SA309538023GA SA309538025GA SA309538027GA SA309538033GA SA309538040GA SA309538042GA SA309538044GA SA309538100GA SA309538108GA SA310028162GA SA310028166GA SA310028176GA SA310028192GA SA310028212GA SA310028226GA SA310413521GA SA310413526GA SA310413529GA SA310413530GA SA310413551GA SA310413570GA SA310423598GA SA310423642GA SA310423643GA SA310423644GA SA310423648GA SA310423655GA SA310423717GA SA310423726GA SA310423733GA SA310423824GA SA310423826GA SA310423827GA SA310423835GA SA310423837GA SA310423840GA SA310423847GA SA310423850GA SA310423853GA SA310423922GA SA310423923GA SA310423924GA SA310423926GA SA310423930GA SA310423945GA SA310423963GA SA310423966GA SA310424009GA SA310454993GA SA310475467GA SA310506748GA SA311060174GA SA311060183GA SA311060193GA SA311060196GA SA311060320GA SA311060322GA SA311060339GA SA311070531GA SA311070535GA SA311070559GA SA311070560GA SA311070589GA SA311080708GA SA311090768GA SA311090797GA SA311090821GA SA311090827GA SA311121697GA SA311121702GA SA311121719GA SA311121722GA SA311121736GA SA311121739GA SA311121745GA SA311142950GA SA311142957GA SA311142965GA SA311142981GA SA311142983GA SA311142985GA SA311153153GA SA311153155GA SA311153167GA SA311153168GA SA311153174GA SA311153177GA SA311153178GA SA311153183GA SA311153184GA SA311153187GA SA311153193GA SA311153198GA SA311153203GA SA311153221GA SA311153224GA SA311215729GA SA311215733GA SA311215734GA SA311297251GA SA312049456GA SA312049465GA SA312069655GA SA312069676GA SA312080896GA SA312232807GA SA312386600GA SA312386632GA SA312386643GA SA312396800GA SA312396824GA SA312397135GA SA312397141GA SA312397271GA SA312407286GA SA312491508GA SA312491524GA SA312491529GA SA312491533GA SA312522492GA SA312522519GA SA312522592GA SA312522593GA SA312522602GA SA312522604GA SA312522606GA SA312522608GA SA312522611GA SA312522615GA SA312522619GA SA313012958GA SA313013106GA SA313023277GA SA313023513GA SA313033568GA SA313033570GA SA313033574GA SA313033595GA SA313033597GA SA313033598GA SA313033599GA SA313033602GA SA313272121GA SA313293166GA SA313293213GA sa313409355ga SA313419501GA SA313419502GA SA313420203GA SA313420284GA SA313420320GA SA313430628GA SA313430663GA SA313430664GA SA313430666GA SA313430672GA SA313430676GA SA313430678GA SA313430679GA SA314014414GA SA314014693GA SA314014743GA SA314014747GA SA314014749GA SA314024832GA SA314076224GA SA314076231GA SA314076237GA SA314076246GA SA314076247GA SA314076248GA SA314076271GA SA314076281GA SA314076284GA SA314076285GA SA314076291GA SA314076308GA SA314076316GA SA314086348GA SA314086349GA SA314086354GA SA314086358GA SA314086361GA SA314086362GA SA314086364GA SA314086374GA SA314086375GA SA314086377GA SA314139359GA SA314170188GA SA314201275GA SA314211858GA SA314263761GA SA314263768GA SA314263774GA SA314263782GA SA314273866GA SA314273868GA SA314273902GA SA314273904GA SA314273906GA SA314273908GA SA314273911GA SA314273913GA SA314273926GA SA314273927GA SA314273929GA SA314273930GA SA314273934GA SA314273936GA SA314273948GA SA314273951GA SA314274216GA SA314274298GA SA314274303GA SA314284357GA SA314284358GA SA314284370GA SA314284375GA SA314284378GA SA314284384GA SA311317689GA SA311428015GA SA311428050GA SA311428084GA SA311428088GA SA312049319GA SA312049327GA SA312049431GA SA312049440GA SA312049450GA SA312069644GA SA312069647GA SA312069649GA SA312069650GA SA312069660GA SA312069662GA SA312080521GA SA312080524GA SA312080569GA SA312080581GA SA312080593GA SA312080606GA SA312080609GA SA312121791GA SA307231192GA SA308196570GA SA308196572GA SA308196573GA SA308196578GA SA308196580GA SA308196583GA SA308206586GA SA308206589GA SA308206601GA SA308206602GA SA308206606GA SA308206607GA SA308206608GA SA308206609GA SA308206610GA SA308206616GA SA308206617GA SA308206619GA SA308206620GA SA308206621GA SA308206622GA SA308206623GA SA308206624GA SA308206626GA SA308206628GA SA308206631GA SA308206632GA SA308206633GA SA308206637GA SA308206638GA SA308206639GA SA308206640GA SA308206641GA SA308206643GA SA308206644GA SA308206645GA SA308206646GA SA308206661GA SA308206664GA SA308206670GA SA311225894GA SA311225899GA SA311225907GA SA311236351GA SA311246496GA SA308125387GA SA308196576GA SA309183249GA SA309193261GA SA309193270GA SA309193271GA SA309193316GA SA309193320GA SA309193326GA SA309517466GA SA309517486GA SA309517714GA SA309517716GA SA309517730GA SA309517766GA SA309517770GA SA310028133GA SA310028208GA SA310028244GA SA310028270GA SA310028277GA SA310028282GA SA310028288GA SA310028293GA SA310028305GA SA310038361GA SA310038379GA SA310038382GA SA310038390GA SA310038400GA SA310038409GA SA310038416GA SA310038447GA SA310038460GA SA310038471GA SA310048586GA SA310058723GA SA310058727GA SA310058728GA SA310058729GA SA310058742GA SA310058748GA SA310068815GA SA310068836GA SA310068837GA SA310068861GA SA310382537GA SA310382559GA SA310382573GA SA310382585GA SA310382817GA SA310392900GA SA310392902GA SA310392924GA SA310393227GA SA310393242GA SA310393248GA SA310393271GA SA310393282GA SA310393299GA SA310403351GA SA310403366GA SA310413536GA SA310424036GA SA310434079GA SA310455124GA SA310455139GA SA310455234GA SA310507187GA SA310507259GA SA310528776GA SA310528780GA SA310528782GA SA310528783GA SA310528793GA SA311039523GA SA311039527GA SA311060327GA SA311060364GA SA311060378GA SA311069754GA SA311069773GA SA311069784GA SA311069826GA SA311069846GA SA311069858GA SA311069861GA SA311070515GA SA311070591GA SA311080612GA SA311080644GA SA311090851GA SA311090879GA SA311101089GA SA311101093GA SA311101121GA SA311101169GA SA311101170GA SA311101173GA SA311101251GA SA311101286GA SA311101293GA SA311101314GA SA311101336GA SA311101362GA SA311101368GA SA311101439GA SA311101443GA SA311101444GA SA311101452GA SA311111583GA SA311111584GA SA311111586GA SA311111595GA SA311111602GA SA311111615GA SA311111673GA SA311111691GA SA311121754GA SA311121756GA SA311121757GA SA311121771GA SA311121774GA SA311121778GA SA311121784GA SA311121787GA SA311121810GA SA311121811GA SA311121818GA SA311121826GA SA311121834GA SA311121836GA SA311122019GA SA311122030GA SA311122040GA SA311132569GA SA311132571GA SA311132580GA SA311132584GA SA311142772GA SA311142875GA SA311142878GA SA311142887GA SA311163626GA SA311163700GA SA311163705GA SA311163719GA SA311163778GA SA311163892GA SA311184442GA SA311184543GA SA311184696GA SA311205498GA SA311236211GA SA311266966GA SA311267008GA SA311297244GA SA311327811GA SA311327867GA SA311458357GA SA311468391GA SA311468411GA SA312018860GA SA312039178GA SA312039184GA SA312039191GA SA312039195GA SA312039206GA SA312039208GA SA312039215GA SA312039224GA SA312039233GA SA312049245GA SA312049257GA SA312059531GA SA312059544GA SA312059546GA SA312059560GA SA312059567GA SA312059568GA SA312059605GA SA312059606GA SA312059607GA SA312059621GA SA312059622GA SA312059623GA SA312059626GA SA312059627GA SA312059634GA SA312070180GA SA312070252GA SA312080592GA SA312080602GA SA312080614GA SA312080615GA SA312091015GA SA312091097GA SA312091098GA SA312091105GA SA312091112GA SA312091235GA SA312091285GA SA312091295GA SA312091301GA SA312091303GA SA312091354GA SA312091357GA SA312111460GA SA312111467GA SA312111539GA SA312121826GA SA312121830GA SA312121866GA SA312121878GA SA312121900GA SA312121914GA SA312121916GA SA312121931GA SA312121993GA SA312122015GA SA312122017GA SA312122033GA SA312122044GA SA312132104GA SA312132129GA SA312152394GA SA312152412GA SA312192660GA SA312192714GA SA312192717GA SA312253185GA SA312253189GA SA312253242GA SA312253389GA SA312284232GA SA312284260GA SA312284287GA SA312284289GA SA312284336GA SA312294346GA SA312294347GA SA312294351GA SA312294352GA SA312294358GA SA312294359GA SA312294364GA SA312294366GA SA312294370GA SA312294671GA SA312294696GA SA312294710GA SA312294761GA SA312344931GA SA312345111GA SA312345131GA SA312345157GA SA312365771GA SA312365772GA SA312365773GA SA312365775GA SA312365779GA SA312365781GA SA312365782GA SA312365787GA SA312365789GA SA312365792GA SA312365793GA SA312365794GA SA312365796GA SA312365797GA SA312365800GA SA312365803GA SA312365805GA SA312365806GA SA312365809GA SA312365812GA SA312365814GA SA312365815GA SA312365817GA SA312365818GA SA312365819GA SA312365820GA SA312365821GA SA312365822GA SA312365823GA SA312365825GA SA312365827GA SA312365828GA SA312365829GA SA312365830GA SA312396912GA SA312397221GA SA312397241GA SA312457889GA SA312457914GA SA312469173GA SA312470048GA SA312480223GA SA312480401GA SA312480407GA SA312491172GA SA312491219GA SA312491326GA SA312491341GA SA312491343GA SA312491348GA SA312491363GA SA312491418GA SA312491445GA SA312491475GA SA312491477GA SA312491480GA SA312491484GA SA312491485GA SA312491488GA SA312491498GA SA312491499GA SA312491500GA SA312491505GA SA312491509GA SA312491516GA SA312501975GA SA312522381GA SA312522391GA SA312522436GA SA312522501GA SA312522515GA SA312522535GA SA312522543GA SA312522612GA SA313002714GA SA313012845GA SA313012917GA SA313012918GA SA313012972GA SA313013006GA SA313013028GA SA313013033GA SA313013035GA SA313013045GA SA313013110GA SA313023180GA SA313023372GA SA313043891GA SA313043892GA SA313043897GA SA313043899GA SA313043902GA SA313043903GA SA313043905GA SA313043906GA SA313043908GA SA313043912GA SA313043916GA SA313043917GA SA313043923GA SA313043957GA SA313044028GA SA313044132GA SA313084461GA SA313084673GA SA313084713GA SA313084721GA SA313084799GA SA313084836GA SA313084854GA SA313084858GA SA313084864GA SA313084865GA SA313084866GA SA313084868GA SA313084871GA SA313084876GA SA313084877GA SA313084901GA SA313095321GA SA313095393GA SA313117143GA SA313127486GA SA313127515GA SA313261884GA SA313293160GA SA313293210GA SA313293263GA SA313314611GA SA313314627GA SA313324947GA SA313324968GA SA313324981GA SA313345717GA SA313356127GA SA313366799GA SA313378393GA SA313388524GA SA313388691GA SA313419477GA SA313419734GA SA313419739GA SA313419741GA SA313419759GA SA313420128GA SA313420428GA SA313420574GA SA313430745GA SA313431132GA SA313431154GA SA313431162GA SA313441517GA SA313451724GA SA313451726GA SA313451727GA SA313451790GA SA313451803GA SA313451870GA SA313451873GA SA313451993GA SA313452052GA SA313482813GA SA313482863GA SA313492955GA SA313492961GA SA313493004GA SA313493005GA SA313493186GA SA313493189GA SA313493190GA SA313513863GA SA313513896GA SA313513901GA SA313523980GA SA314004102GA SA314004104GA SA314004111GA SA314004113GA SA314004115GA SA314004117GA SA314004120GA SA314004121GA SA314004123GA SA314004126GA SA314004131GA SA314004172GA SA314004173GA SA314004174GA SA314004175GA SA314004177GA SA314004181GA SA314004182GA SA314004192GA SA314004193GA SA314004194GA SA314004195GA SA314004199GA SA314004201GA SA314004205GA SA314004206GA SA314004209GA SA314004211GA SA314004213GA SA314004214GA SA314004217GA SA314004218GA SA314004219GA SA314004220GA SA314004221GA SA314004222GA SA314004223GA SA314004249GA SA314004264GA SA314004295GA SA314004296GA SA314004299GA SA314004300GA SA314004302GA SA314004303GA SA314004305GA SA314004306GA SA314004307GA SA314004308GA SA314004312GA SA314004314GA SA314004317GA SA314004319GA SA314014324GA SA314014438GA SA314014439GA SA314014457GA SA314014478GA SA314014531GA SA314014611GA SA314014671GA SA314014741GA SA314014751GA SA314014752GA SA314024770GA SA314024778GA SA314024782GA SA314024798GA SA314024804GA SA314024842GA SA314024848GA SA314024849GA SA314024852GA SA314024855GA SA314024860GA SA314024890GA SA314024919GA SA314024934GA SA314035341GA SA314035410GA SA314055599GA SA314055675GA SA314055707GA SA314055713GA SA314065748GA SA314075817GA SA314076084GA SA314076091GA SA314076113GA SA314076130GA SA314076135GA SA314076144GA SA314076146GA SA314076157GA SA314076200GA SA314076207GA SA314086718GA SA314096867GA SA314097356GA SA314107914GA SA314108056GA SA314108084GA SA314108116GA SA314118706GA SA314118870GA SA314119092GA SA314129208GA SA314129215GA SA314129253GA SA314129273GA SA314139377GA SA314139393GA SA314139412GA SA314139484GA SA314139497GA SA314139556GA SA314169909GA SA314170215GA SA314201435GA SA314201749GA SA314253568GA SA314253667GA SA314253698GA SA314253699GA SA314253704GA SA314253709GA SA314253712GA SA314253714GA SA314253717GA SA314253725GA SA314253728GA SA314253730GA SA314253737GA SA314253739GA SA314253740GA SA314253741GA SA314253748GA SA314253749GA SA314273843GA SA314273972GA SA314284607GA SA314295117GA SA314295126GA SA314295197GA SG511480128GA SG511480129GA SG511480130GA SA309354975GA SA307110546GA SA307110570GA SA307110573GA SA307150661GA SA307220885GA SA307231254GA SA307241687GA SA307241694GA SA307392687GA SA307392993GA SA307403056GA SA307403064GA SA307403077GA SA307443317GA SA307443323GA SA308074582GA SA308074584GA SA308074599GA SA308084714GA SA308094810GA SA308094840GA SA308094848GA SA308094854GA SA308094879GA SA308094883GA SA308094887GA SA308094888GA SA308094889GA SA308094890GA SA308094893GA SA308094895GA SA308094896GA SA308094898GA SA308094900GA SA308094907GA SA308094916GA SA308115302GA SA308115304GA SA308115307GA SA308317302GA SA308327314GA SA308327526GA SA308327545GA SA308327546GA SA308337570GA SA308337600GA SA308337634GA SA308337688GA SA308337697GA SA308408998GA SA308408999GA SA308409101GA SA308409102GA SA308409105GA SA309042188GA SA309052215GA SA309052231GA SA309052234GA SA309052235GA SA309072353GA SA309344881GA SA309446494GA SA310048557GA SA310048595GA SA310139247GA SA310199561GA SA310290621GA SA310290623GA SA310290625GA SA310290628GA SA310290629GA SA310290631GA SA310290633GA SA310290634GA SA310290638GA SA310290639GA SA310290640GA SA310290641GA SA310290645GA SA310290646GA SA310290647GA SA310290648GA SA310290652GA SA310290662GA SA310290664GA SA310290666GA SA310290669GA SA310290670GA SA310290671GA SA310290672GA SA310290674GA SA310290675GA SA310290676GA SA310290677GA SA310290678GA SA310290680GA SA310290681GA SA310290682GA SA310290683GA SA310290684GA SA310290685GA SA310290687GA SA310290688GA SA310290689GA SA310290690GA SA310290691GA SA310290692GA SA310290694GA SA310290695GA SA310290697GA SA310290698GA SA310290701GA SA310341389GA SA310361754GA SA310361786GA SA310361818GA SA310361819GA SA310361827GA SA310361837GA SA310361842GA SA310361845GA SA310361846GA SA310361847GA SA310361848GA SA310361851GA SA310361854GA SA310361855GA SA310361858GA SA310361860GA SA310361861GA SA310361862GA SA310361865GA SA310361866GA SA310361867GA SA310361868GA SA310361869GA SA310361870GA SA310361871GA SA310361875GA SA310361878GA SA310361881GA SA310361882GA SA310361884GA SA310361943GA SA310361959GA SA310372091GA SA310372114GA SA310372116GA SA310372122GA SA310372123GA SA310372125GA SA310372135GA SA310372137GA SA310372164GA SA310372177GA SA310372178GA SA310372183GA SA310372186GA SA310372188GA SA310372190GA SA310372192GA SA310372206GA SA310372285GA SA310372292GA SA310372298GA SA310382316GA SA310382317GA SA310382323GA SA310382339GA SA310382340GA SA310382342GA SA310382366GA SA310382369GA SA310382374GA SA310382398GA SA310382502GA SA310382528GA SA310382560GA SA310382569GA SA310382579GA SA310382586GA SA310382679GA SA310382680GA SA310382687GA SA310382699GA SA310382701GA SA310382771GA SA310382774GA SA310382775GA SA310382776GA SA310382778GA SA310382780GA SA310382787GA SA310382793GA SA310382795GA SA310382796GA SA310382797GA SA310382801GA SA310382802GA SA310382806GA SA310382808GA SA310382809GA SA310382816GA SA310382818GA SA310382819GA SA310382823GA SA310382824GA SA310382825GA SA310382827GA SA310382828GA SA310382829GA SA310382831GA SA310382834GA SA310382835GA SA310382844GA SA310382848GA SA310392859GA SA310392874GA SA310392876GA SA310403342GA SA310403373GA SA310403375GA SA310403397GA SA310403404GA SA310403408GA SA310403416GA SA310403417GA SA310403424GA SA310403444GA SA310423595GA SA310423676GA SA310423695GA SA310423757GA SA310423776GA SA310423786GA SA310423789GA SA310423797GA SA310423810GA SA310423815GA SA310423817GA SA310423890GA SA310434083GA SA310434135GA SA310434143GA SA310434156GA SA310434162GA SA310434172GA SA310434187GA SA310434209GA SA310454985GA SA310455035GA SA310465321GA SA310465322GA SA310465323GA SA310465325GA SA310465326GA SA310465327GA SA310465328GA SA310465329GA SA310465330GA SA310465331GA SA310465332GA SA310465333GA SA310465334GA SA310465338GA SA310465339GA SA310465341GA SA310465342GA SA310465343GA SA310465346GA SA310465347GA SA310465380GA SA310465381GA SA310465397GA SA310465398GA SA310465403GA SA310465410GA SA310465411GA SA310465412GA SA310465419GA SA310465420GA SA310465427GA SA310465431GA SA310465432GA SA310465433GA SA310465435GA SA310465438GA SA310465439GA SA310465441GA SA310465443GA SA310465445GA SA310465446GA SA310475487GA SA310475515GA SA310475516GA SA310475518GA SA310475544GA SA310475545GA SA310475554GA SA310475572GA SA310475599GA SA310475601GA SA310475603GA SA310475604GA SA310475606GA SA310475607GA SA310475614GA SA310475619GA SA310475621GA SA310475622GA SA310475623GA SA310475624GA SA310475625GA SA310475627GA SA310475633GA SA310475635GA SA310475636GA SA310475640GA SA310475641GA SA310475642GA SA310475643GA SA310475645GA SA310475646GA SA310475647GA SA310475648GA SA310475650GA SA310475651GA SA310475652GA SA310475653GA SA310475654GA SA310475655GA SA310475656GA SA310475657GA SA310475658GA SA310475660GA SA310475661GA SA310475662GA SA310475664GA SA310475665GA SA310475666GA SA310475667GA SA310475668GA SA310475669GA SA310475670GA SA310475671GA SA310475672GA SA310475674GA SA310475675GA SA310475676GA SA310475677GA SA310475678GA SA310475679GA SA310475680GA SA310475683GA SA310475688GA SA310475689GA SA310475690GA SA310475691GA SA310475692GA SA310475693GA SA310475694GA SA310475699GA SA310475708GA SA310475713GA SA310475715GA SA310475716GA SA310475720GA SA310475721GA SA310475722GA SA310475723GA SA310475724GA SA310475725GA SA310475727GA SA310475729GA SA310475735GA SA310475736GA SA310475737GA SA310475738GA SA310475739GA SA310475740GA SA310475741GA SA310475742GA SA310475743GA SA310475744GA SA310475745GA SA310475746GA SA310475747GA SA310475748GA SA310475749GA SA310475750GA SA310475751GA SA310475752GA SA310475753GA SA310475754GA SA310485755GA SA310485756GA SA310485757GA SA310485758GA SA310485759GA SA310485760GA SA310485762GA SA310485764GA SA310485765GA SA310485767GA SA310485771GA SA310485773GA SA310485774GA SA310485775GA SA310485776GA SA310485777GA SA310485778GA SA310485779GA SA310485780GA SA310485781GA SA310485783GA SA310485784GA SA310485787GA SA310485797GA SA310485798GA SA310485799GA SA310485802GA SA310485815GA SA310485821GA SA310485826GA SA310485835GA SA310485841GA SA310485844GA SA310485846GA SA310485850GA SA310485857GA SA310485860GA SA310485868GA SA310485869GA SA310485871GA SA310485874GA SA310485878GA SA310485884GA SA310485885GA SA310485888GA SA310496138GA SA310496149GA SA310496151GA SA310496155GA SA310496206GA SA310496207GA SA310496300GA SA310496346GA SA310496355GA SA310496357GA SA310496508GA SA310496616GA SA310496620GA SA310496629GA SA310496645GA SA310507105GA SA310507134GA SA310507254GA SA310507465GA SA310507501GA SA310507515GA SA310507545GA SA310517714GA SA310517726GA SA310517730GA SA310517735GA SA311039381GA SA311039386GA SA311039395GA SA311039396GA SA311039397GA SA311039402GA SA311039403GA SA311039460GA SA311039482GA SA311039491GA SA311060147GA SA311060173GA SA311060204GA SA311060280GA SA311101178GA SA311101181GA SA311101189GA SA311101213GA SA311101243GA SA311101282GA SA311101384GA SA311101386GA SA311101389GA SA311101393GA SA311101394GA SA311101396GA SA311101408GA SA311101410GA SA311101418GA SA311101420GA SA311142948GA SA311174169GA SA311174184GA SA311174188GA SA311174218GA SA311174220GA SA311174228GA SA311174234GA SA311184517GA SA311184538GA SA311184539GA SA311184547GA SA311184552GA SA311184562GA SA311184589GA SA311184615GA SA311184617GA SA311327822GA SA311327868GA SA311327872GA SA311327886GA SA311327892GA SA312018771GA SA312039150GA SA312039167GA SA312039171GA SA312039182GA SA312039186GA SA312039187GA SA312039189GA SA312039192GA SA312039200GA SA312039203GA SA312039209GA SA312039210GA SA312039212GA SA311307540GA SA311307554GA SA311307562GA SA311317617GA SA311317622GA SA311317633GA SA311317675GA SA312294383GA SA312355455GA SA312355464GA SA312365778GA SA312365783GA SA312365785GA SA312365786GA SA312365788GA SA312365795GA SA312365808GA SA312365826GA SA312457693GA SA312457721GA SA312457726GA SA312457848GA SA312458185GA SA312458244GA SA312458404GA SA312468782GA SA312468797GA SA312468932GA SA312468933GA SA312468940GA SA312468957GA SA312468973GA SA312468983GA SA312468986GA SA312469113GA SA312469114GA SA312469122GA SA312469130GA SA312469136GA SA312469138GA SA312469144GA SA312469145GA SA312469146GA SA312469152GA SA312469158GA SA312469160GA SA312469175GA SA312469183GA SA312469185GA SA312469187GA SA312469195GA SA312469196GA SA312469200GA SA312469201GA SA312469202GA SA312469203GA SA312469204GA SA312469205GA SA312469210GA SA312469211GA SA312469306GA SA312470059GA SA312470062GA SA312480727GA SA312491210GA SA312491250GA SA312491265GA SA312491319GA SA312491320GA SA312491325GA SA312491327GA SA312491328GA SA312491332GA SA312491340GA SA312491344GA SA312491350GA SA312491351GA SA312491360GA SA312491368GA SA312491374GA SA312491407GA SA312491408GA SA312491410GA SA312491422GA SA312491423GA SA312491425GA SA312491431GA SA312491432GA SA312491435GA SA312491436GA SA312491437GA SA312491439GA SA312491443GA SA312491446GA SA312491449GA SA312491456GA SA312491457GA SA312491459GA SA312491460GA SA312491464GA SA312491467GA SA312491468GA SA312491469GA SA312491471GA SA312491472GA SA313013086GA SA313013087GA SA313013089GA SA313013097GA SA313013103GA SA313013109GA SA313013130GA SA313013131GA SA313105881GA SA313116816GA SA313198333GA SA313198334GA SA313198368GA SA313198415GA SA313198420GA SA313230625GA SA313272174GA SA313293029GA SA313293256GA SA313314496GA SA313314501GA SA313419445GA SA313419515GA sa313419519ga SA313420576GA SA313420609GA SA313420610GA SA313451984GA SA313451992GA SA313482765GA SA314025075GA SA314025076GA SA314025097GA SA314025105GA SA314076274GA SA314086476GA SA314139520GA SA314169827GA SA314169842GA SA314169851GA SA314169852GA SA314169901GA SA314169907GA SA314169910GA SA314169912GA SA314201434GA SA314201843GA SA314211924GA SA314211932GA SA314211934GA SA314211938GA SA314211942GA SA314211948GA SA314212011GA SA314212036GA SA314212040GA SA314212094GA SA314232991GA SA314233010GA SA308084641GA SA308084647GA SA308084657GA SA308084764GA SA308084767GA SA308084768GA SA308084772GA SA308084780GA SA308084781GA SA308084782GA SA308094834GA SA309334613GA SA309334640GA SA309334666GA SA309334667GA SA309334670GA SA309334686GA SA309334704GA SA309334754GA SA309344856GA SA309344860GA SA309344867GA SA309344870GA SA312018759GA SA312344985GA SA312355228GA SA312355232GA SA312355237GA SA312355239GA SA312355240GA SA312355242GA SA312355243GA SA312355244GA SA312355245GA SA312355246GA SA312355249GA SA312355251GA SA312355252GA SA312355254GA SA312355439GA SA312355440GA SA312355446GA SA307241688GA SA308388612GA SA308408909GA SA308408914GA SA308408923GA SA308408927GA SA308408931GA SA308408932GA SA308408952GA SA308408956GA SA308409024GA SA308409026GA SA308409035GA SA308409040GA SA308409061GA SA307291937GA SA307291939GA SA307291954GA SA307301990GA SA307301995GA SA308084631GA SA308084666GA SA308084727GA SA308084728GA SA308084731GA SA308084732GA SA308084734GA SA308084746GA SA308094804GA SA308094806GA SA308155924GA SA308155941GA SA308155962GA SA308165998GA SA308166003GA SA308459795GA SA309173142GA SA309173167GA SA309173169GA SA309173172GA SA309173183GA SA309193294GA SA309193296GA SA309193297GA SA309193303GA SA309193305GA SA309193307GA SA309193313GA SA309193317GA SA309193319GA SA309193321GA SA309193323GA SA309193324GA SA309193325GA SA309193351GA SA309193353GA SA309203372GA SA309203416GA SA309203418GA SA309203423GA SA309243659GA SA309243704GA SA309243709GA SA309253729GA SA309253733GA SA309253734GA SA309253735GA SA309253736GA SA309253741GA SA309253742GA SA309253749GA SA309253751GA SA309253752GA SA309253821GA SA309253836GA SA309253838GA SA309253841GA SA309253842GA SA309253843GA SA309253844GA SA309253847GA SA309253864GA SA309253868GA SA309253870GA SA309253875GA SA309253876GA SA309253877GA SA309253884GA SA309253889GA SA309253892GA SA309253893GA SA309253897GA SA309253910GA SA309253912GA SA309385675GA SA309395695GA SA309395702GA SA309395709GA SA309395740GA SA309395804GA SA309446453GA SA309446455GA SA309446495GA SA309446508GA SA309456563GA SA309466736GA SA309466743GA SA309466745GA SA309466747GA SA309466750GA SA309466774GA SA309466794GA SA310260323GA SA310260329GA SA310260338GA SA310260340GA SA310260344GA SA310260345GA SA310260475GA SA310260485GA SA310260492GA SA310260497GA SA310260498GA SA310260500GA SA310321010GA SA310382624GA SA310434202GA SA310496133GA SA310496153GA SA310496298GA SA310506659GA SA310506662GA SA310506664GA SA310506666GA SA310506667GA SA310506669GA SA310506671GA SA310506677GA SA310506685GA SA310506686GA SA310506693GA SA310506699GA SA310506701GA SA310506705GA SA310506712GA SA310506718GA SA310506723GA SA310506742GA SA310506752GA SA310506847GA SA310506850GA SA310506863GA SA310506891GA SA310507033GA SA310507498GA SA310517578GA SA310517700GA SA310517703GA SA310517715GA SA310517759GA SA310517775GA SA310517788GA SA310518622GA SA310518641GA SA310518644GA SA310529143GA SA310529144GA SA310529151GA SA310529154GA SA310529209GA SA311060136GA SA311060138GA SA311060140GA SA311060167GA SA311060208GA SA311060230GA SA311069728GA SA311069731GA SA311069732GA SA311090783GA SA311153208GA SA311163375GA SA311163379GA SA311163380GA SA311163381GA SA311163384GA SA311163391GA SA311163392GA SA311163395GA SA311163399GA SA311163413GA SA311163418GA SA311163419GA SA311163421GA SA312049330GA SA312049335GA SA312049364GA SA312080595GA SA312080611GA SA312121919GA SA312294465GA SA312294570GA SA312375912GA SA312375914GA SA312375927GA SA312375983GA SA312375984GA SA312375985GA SA312375995GA SA312375999GA SA312376007GA SA312376009GA SA312376012GA SA312376014GA SA312376028GA SA312376052GA SA312458107GA SA312470001GA SA312470002GA SA312470003GA SA312470007GA SA312470008GA SA312470016GA SA312470065GA SA312470069GA SA312470076GA SA312470090GA SA312479690GA SA312479693GA SA312479694GA SA312479750GA SA312479777GA SA312479949GA SA312479952GA SA312479958GA SA312479959GA SA312479970GA SA312480094GA SA312480100GA SA312480108GA SA313013094GA SA313023258GA SA313023302GA SA313023303GA SA313023304GA SA313023308GA SA313023312GA SA313023313GA SA313023346GA SA313023347GA SA313023351GA SA313023359GA SA313023362GA SA313023363GA SA313023367GA SA313023371GA SA313023400GA SA313272167GA SA313272367GA SA313272400GA SA313293004GA SA313293006GA SA313324923GA SA313324956GA SA313324957GA SA313324966GA SA313324982GA SA313377649GA SA313377653GA SA313377654GA SA313377655GA SA313377657GA SA313377658GA SA313377659GA SA313377660GA SA313377661GA SA313377664GA SA313377666GA SA313377669GA SA313377670GA SA313377672GA SA313377674GA SA313377675GA SA313377676GA SA313377677GA SA313377678GA SA313377700GA SA313409406GA SA313409425GA SA314273940GA SA309446504GA SA309446509GA SA309466789GA SA310290620GA SA310310959GA SA310321060GA SA310341413GA SA310444314GA SA310444498GA SA310517709GA SA310528892GA SA311039379GA SA311039461GA SA311070580GA SA311090767GA SA311090823GA SA311090828GA SA311090857GA SA311101248GA SA311101252GA SA311101254GA SA311101257GA SA311101260GA SA311132551GA SA311132575GA SA311142751GA SA311163874GA SA311163885GA SA311163912GA SA311163921GA SA311184372GA SA311184398GA SA311184514GA SA311184534GA SA311184575GA SA311184593GA SA311184595GA SA311184600GA SA311246467GA SA311297357GA SA311307536GA SA311317722GA SA311468381GA SA311468392GA SA311508562GA SA312018872GA SA312049328GA SA312049340GA SA312049351GA SA312049355GA SA312059470GA SA312059545GA SA312080493GA SA312080516GA SA312080633GA SA312080682GA SA312121943GA SA312152542GA SA312253362GA SA312294493GA SA312376037GA SA312386645GA SA312386651GA SA312386661GA SA312447468GA SA312458186GA SA312458194GA SA312458199GA SA312458223GA SA312458252GA SA312458454GA SA312469295GA SA312480853GA SA312491462GA SA312501644GA SA312501648GA SA313084772GA SA313084963GA SA313105918GA SA313106394GA SA313137870GA SA313198165GA SA313198472GA SA313208911GA SA313208928GA SA313208931GA SA313208941GA SA313229955GA SA313462208GA SA313462257GA SA313462262GA SA313462313GA SA313462318GA SA313482734GA SA313482737GA SA313482761GA SA313493049GA SA313493089GA SA313493194GA SA314014326GA SA314014373GA SA314076276GA SA314097288GA SA314170193GA SA314212053GA SA314212062GA SA314294862GA SG512300131GA SG512300134GA SA312469059GA SA313240969GA sa313452011ga SA312479829GA SA312479833GA SA310518346GA SA310518387GA SA310518389GA SA310518429GA SA310518435GA SA310518438GA SA310518619GA SA310528945GA SA313452067GA SA314024877GA SA314024905GA SA312396999GA SA311195061GA SA311195068GA SA311236065GA SA311256798GA SA313229426GA SA313378308GA SA313441379GA SA314149662GA SA314149667GA SA314149704GA SA314149747GA SA314201699GA SA314201707GA SA314243214GA SA313431188GA SA313431190GA SA308115268GA SA312091324GA SA307030410GA SA307110555GA SA307221099GA SA307231353GA SA307332220GA SA307332242GA SA307342310GA SA307342326GA SA307342327GA SA307352371GA SA307362388GA SA307382455GA SA308084726GA SA308206679GA SA308388548GA SA308398890GA SA308429360GA SA309122641GA SA309122642GA SA313492962GA SA307020390GA SA307050430GA SA307070439GA SA307080460GA SA307080471GA SA307080475GA SA307080480GA SA307080482GA SA307080487GA SA307080496GA SA307080500GA SA307080505GA SA307080507GA SA307080515GA SA307080529GA SA307080534GA SA307080539GA SA307080540GA SA307080542GA SA307110549GA SA307110551GA SA307110552GA SA307110553GA SA307110561GA SA307110568GA SA307110577GA SA307130597GA SA307140617GA SA307140629GA SA307140630GA SA307140638GA SA307150650GA SA307150657GA SA307150658GA SA307150662GA SA307150664GA SA307150665GA SA307150666GA SA307150667GA SA307150668GA SA307150669GA SA307150670GA SA307150673GA SA307150674GA SA307150676GA SA307221054GA SA307221073GA SA307221143GA SA307231269GA SA307231272GA SA307231283GA SA307231288GA SA307231289GA SA307231308GA SA307231309GA SA307231328GA SA307241524GA SA307291859GA SA307291860GA SA307291892GA SA307302073GA SA307322122GA SA307322124GA SA307322138GA SA307322139GA SA307322140GA SA307322141GA SA307322142GA SA307322143GA SA307322144GA SA307322148GA SA307322149GA SA307322151GA SA307322154GA SA307322161GA SA307322162GA SA307322163GA SA307322164GA SA307322166GA SA307322187GA SA307322190GA SA307322191GA SA307322192GA SA307322193GA SA307322198GA SA307322199GA SA307322202GA SA307322204GA SA307382445GA SA307382519GA SA307392693GA SA307392712GA SA307392756GA SA307392831GA SA307392930GA SA307392946GA SA307392952GA SA307392966GA SA307392968GA SA307392990GA SA307392999GA SA307393002GA SA307393006GA SA307393027GA SA307393029GA SA307393033GA SA307393034GA SA307393036GA SA307393037GA SA307403066GA SA307403069GA SA307403076GA SA307403078GA SA307403080GA SA307403081GA SA307403083GA SA307403084GA SA307403085GA SA307403086GA SA307403088GA SA307403090GA SA307403091GA SA307403092GA SA307403093GA SA307403094GA SA307403096GA SA307403097GA SA307403099GA SA307403100GA SA307403102GA SA307403107GA SA307403109GA SA307403113GA SA307403114GA SA307403117GA SA307403118GA SA307403119GA SA307403121GA SA307403122GA SA307403125GA SA307403130GA SA307403133GA SA307403142GA SA307403144GA SA307403145GA SA307403148GA SA307403153GA SA307403154GA SA307403155GA SA307423139GA SA307423160GA SA307443188GA SA307443195GA SA307443213GA SA307443306GA SA307443311GA SA307443319GA SA307443321GA SA307443322GA SA307443329GA SA307443334GA SA307443336GA SA307443337GA SA307443338GA SA307443339GA SA307443350GA SA307443351GA SA307443352GA SA307443356GA SA307443357GA SA307453422GA SA307453432GA SA307453444GA SA307453453GA SA307453458GA SA307453464GA SA307453474GA SA307453475GA SA307453483GA SA307453590GA SA307514239GA SA307514240GA SA307514247GA SA307514283GA SA307514298GA SA307514299GA SA307514303GA SA307514304GA SA307514313GA SA307514314GA SA307514315GA SA307514318GA SA308084668GA SA308084676GA SA308084678GA SA308084683GA SA308084700GA SA308094856GA SA308094882GA SA308094886GA SA308094892GA SA308094897GA SA308094901GA SA308094902GA SA308094904GA SA308094905GA SA308094906GA SA308094909GA SA308094910GA SA308094912GA SA308094917GA SA308094918GA SA308094920GA SA308095062GA SA308105083GA SA308105084GA SA308105087GA SA308105089GA SA308105096GA SA308105099GA SA308105106GA SA308105113GA SA308105114GA SA308105117GA SA308105118GA SA308105142GA SA308105147GA SA308115261GA SA308115282GA SA308125345GA SA308125351GA SA308125407GA SA308125411GA SA308135461GA SA308135470GA SA308135473GA SA308135515GA SA308135523GA SA308135554GA SA308135558GA SA308135560GA SA308135562GA SA308135564GA SA308135570GA SA308135574GA SA308135575GA SA308135578GA SA308135583GA SA308135589GA SA308135604GA SA308135607GA SA308135610GA SA308135612GA SA308135620GA SA308135623GA SA308135624GA SA308135626GA SA308135628GA SA308135631GA SA308135632GA SA308135634GA SA308135635GA SA308135636GA SA308135637GA SA308135638GA SA308135640GA SA308135641GA SA308135642GA SA308135643GA SA308135644GA SA308135645GA SA308135646GA SA308135647GA SA308135650GA SA308135651GA SA308135653GA SA308135654GA SA308145662GA SA308145674GA SA308145702GA SA308145706GA SA308145709GA SA308145710GA SA308145713GA SA308145716GA SA308145718GA SA308145719GA SA308145720GA SA308145721GA SA308145728GA SA308145733GA SA308145735GA SA308145736GA SA308145737GA SA308145752GA SA308145782GA SA308145783GA SA308145784GA SA308145785GA SA308145788GA SA308145789GA SA308145790GA SA308145791GA SA308145792GA SA308145793GA SA308145794GA SA308145795GA SA308145796GA SA308145798GA SA308145801GA SA308145802GA SA308145804GA SA308145825GA SA308155849GA SA308155935GA SA308166055GA SA308186400GA SA308186421GA SA308186425GA SA308186441GA SA308196458GA SA308196509GA SA308196548GA SA308196556GA SA308196567GA SA308196571GA SA308357967GA SA308357976GA SA308357977GA SA308368111GA SA308368113GA SA308368115GA SA308368116GA SA308368117GA SA308368121GA SA308368122GA SA308368124GA SA308368141GA SA308398662GA SA308398723GA SA308398735GA SA308398736GA SA308409106GA SA308409197GA SA308419300GA SA308429341GA SA308429357GA SA308449508GA SA308459682GA SA308459861GA SA308469886GA SA308479893GA SA308479898GA SA308479904GA SA308481021GA SA308481024GA SA308481037GA SA308481064GA SA308481067GA SA308489935GA SA308489946GA SA308489951GA SA308489965GA SA308489967GA SA308489982GA SA309092473GA SA309092475GA SA309092476GA SA309476911GA SA309476915GA SA309476922GA SA309476926GA SA309476927GA SA309476931GA SA309507336GA SA309507345GA SA309507347GA SA309507349GA SA309507351GA SA309507355GA SA309507359GA SA309507362GA SA309507363GA SA309507364GA SA309507374GA SA309507378GA SA309507384GA SA309507389GA SA309507391GA SA309507394GA SA309507398GA SA309507400GA SA309507401GA SA309507403GA SA309507405GA SA309507408GA SA309507409GA SA309507411GA SA309507415GA SA309527953GA SA310260394GA SA310382363GA SA310382668GA SA310382677GA SA310382833GA SA310382842GA SA310392855GA SA310392872GA SA310392891GA SA310393069GA SA310393215GA SA310403314GA SA310403316GA SA310403347GA SA310423774GA SA310434062GA SA310444495GA SA310465309GA SA310465319GA SA310465377GA SA310475638GA SA310475644GA SA310475673GA SA310475702GA SA310485805GA SA310485809GA SA310485829GA SA310485838GA SA310485861GA SA310485870GA SA310485883GA SA310496132GA SA310496147GA SA310496154GA SA310496208GA SA310496237GA SA310496238GA SA310496289GA SA310496327GA SA310496334GA SA310496335GA SA310496342GA SA310496345GA SA310496348GA SA310496359GA SA310496367GA SA310496377GA SA310496379GA SA310496400GA SA310496413GA SA310496430GA SA310496502GA SA310496507GA SA310496516GA SA310496523GA SA310496619GA SA310496635GA SA310496639GA SA310496640GA SA310496647GA SA310496652GA SA310506660GA SA310506668GA SA310506698GA SA310506720GA SA310506721GA SA310506730GA SA310506740GA SA310506751GA SA310506758GA SA310506764GA SA310506769GA SA310506777GA SA310506779GA SA310506794GA SA310506797GA SA310506804GA SA310506856GA SA310506858GA SA310506866GA SA310506898GA SA310507026GA SA310507035GA SA310507039GA SA310507091GA SA310507159GA SA310507180GA SA310507189GA SA310507194GA SA310507198GA SA310507208GA SA310507434GA SA310507438GA SA310507467GA SA310507468GA SA310507502GA SA310507557GA SA310517642GA SA310517643GA SA310517644GA SA310517648GA SA310517653GA SA310517684GA SA310517734GA SA310517743GA SA310517784GA SA310518145GA SA310518153GA SA310518167GA SA310518171GA SA310518172GA SA310518174GA SA310518175GA SA310518178GA SA310518221GA SA310518226GA SA310518227GA SA310518231GA SA310518260GA SA310518309GA SA310518354GA SA310518425GA SA310518428GA SA310518497GA SA310518500GA SA310518504GA SA310518514GA SA310518517GA SA310518518GA SA310518520GA SA310518524GA SA310518562GA SA310518616GA SA310528784GA SA310528786GA SA310528788GA SA310528789GA SA310528796GA SA310528797GA SA310528811GA SA310528824GA SA310528841GA SA310528851GA SA310528857GA SA310528866GA SA310528873GA SA310528874GA SA310528885GA SA310528893GA SA310528900GA SA310528970GA SA310528971GA SA310528973GA SA310528978GA SA310528983GA SA310528986GA SA310529012GA SA310529078GA SA310529146GA SA310529149GA SA311039368GA SA311039369GA SA311039374GA SA311039376GA SA311039394GA SA311039398GA SA311039400GA SA311039401GA SA311039406GA SA311039407GA SA311039411GA SA311039415GA SA311039428GA SA311039431GA SA311039436GA SA311039443GA SA311039445GA SA311039456GA SA311039462GA SA311039466GA SA311039471GA SA311039476GA SA311039503GA SA311039506GA SA311039512GA SA311039514GA SA311039515GA SA311039517GA SA311039522GA SA311039574GA SA311060018GA SA311060042GA SA311060068GA SA311060078GA SA311060087GA SA311060089GA SA311060094GA SA311060107GA SA311060109GA SA311060155GA SA311060171GA SA311060214GA SA311060228GA SA311060235GA SA311060238GA SA311060240GA SA311060254GA SA311060265GA SA311069646GA SA311069671GA SA311069693GA SA311069719GA SA311069743GA SA311069775GA SA311069776GA SA311069778GA SA311069783GA SA311069791GA SA311069809GA SA311069812GA SA311069852GA SA311080615GA SA311080677GA SA311080678GA SA311090849GA SA311090860GA SA311090870GA SA311090877GA SA311090944GA SA311091003GA SA311091006GA SA311091033GA SA311091047GA SA311101077GA SA311101080GA SA311101085GA SA311101094GA SA311101105GA SA311101106GA SA311101140GA SA311101141GA SA311101241GA SA311101278GA SA311101297GA SA311121753GA SA311121781GA SA311121786GA SA311121797GA SA311121805GA SA311121842GA SA311121913GA SA311121924GA SA311121951GA SA311121970GA SA311121978GA SA311121984GA SA311121985GA SA311121991GA SA311121996GA SA311122000GA SA311122077GA SA311122100GA SA311122142GA SA311122164GA SA311132373GA SA311132379GA SA311132388GA SA311132412GA SA311132424GA SA311132456GA SA311132469GA SA311132474GA SA311132475GA SA311132478GA SA311132480GA SA311132484GA SA311132530GA SA311132532GA SA311132535GA SA311132536GA SA311132541GA SA311132547GA SA311132548GA SA311132603GA SA311132615GA SA311132641GA SA311142682GA SA311142685GA SA311142692GA SA311142694GA SA311142702GA SA311142728GA SA311142731GA SA311142734GA SA311142735GA SA311142939GA SA311142968GA SA311142977GA SA311143022GA SA311153041GA SA311153051GA SA311163429GA SA311163457GA SA311163539GA SA311163541GA SA311163556GA SA311163557GA SA311163569GA SA311163570GA SA311163571GA SA311163573GA SA311163575GA SA311163577GA SA311163578GA SA311163590GA SA311163593GA SA311163596GA SA311163599GA SA311163600GA SA311163603GA SA311163604GA SA311163605GA SA311163608GA SA311163611GA SA311163615GA SA311163616GA SA311163618GA SA311163619GA SA311163622GA SA311163631GA SA311163632GA SA311163634GA SA311163635GA SA311163637GA SA311163643GA SA311163644GA SA311163646GA SA311163652GA SA311163653GA SA311163654GA SA311163697GA SA311163701GA SA311163706GA SA311163708GA SA311163709GA SA311163721GA SA311163725GA SA311163727GA SA311163730GA SA311163731GA SA311163777GA SA311163785GA SA311163786GA SA311163787GA SA311163788GA SA311163793GA SA311163795GA SA311163797GA SA311163798GA SA311163802GA SA311163804GA SA311163807GA SA311163810GA SA311163813GA SA311163816GA SA311163819GA SA311163822GA SA311163826GA SA311163827GA SA311163828GA SA311163831GA SA311163834GA SA311163842GA SA311163850GA SA311163851GA SA311163855GA SA311163870GA SA311163908GA SA311163929GA SA311163946GA SA311163961GA SA311163973GA SA311163974GA SA311163979GA SA311163993GA SA311173998GA SA311174001GA SA311174008GA SA311174010GA SA311174011GA SA311174012GA SA311174013GA SA311174014GA SA311174020GA SA311174022GA SA311174023GA SA311174026GA SA311174028GA SA311174029GA SA311174030GA SA311174034GA SA311174036GA SA311174039GA SA311174042GA SA311174043GA SA311174046GA SA311174048GA SA311174050GA SA311174054GA SA311174055GA SA311174057GA SA311174061GA SA311174063GA SA311174068GA SA311174069GA SA311174070GA SA311174074GA SA311174079GA SA311174082GA SA311174086GA SA311174090GA SA311174092GA SA311174094GA SA311174099GA SA311174102GA SA311174104GA SA311174105GA SA311174106GA SA311174109GA SA311174110GA SA311174111GA SA311174112GA SA311174113GA SA311174114GA SA311174115GA SA311174116GA SA311174117GA SA311174118GA SA311174124GA SA311174129GA SA311174135GA SA311174136GA SA311174137GA SA311174138GA SA311174139GA SA311174145GA SA311174146GA SA311174148GA SA311174154GA SA311174155GA SA311174157GA SA311174171GA SA311174173GA SA311174174GA SA311174175GA SA311174183GA SA311174190GA SA311174193GA SA311174204GA SA311184348GA SA311184375GA SA311184384GA SA311184388GA SA311184526GA SA311184546GA SA311184551GA SA311184573GA SA311184583GA SA311184587GA SA311184607GA SA311184684GA SA311184760GA SA311184761GA SA311184810GA SA311194904GA SA311194933GA SA311194934GA SA311194978GA SA311194984GA SA311194993GA SA311195004GA SA311195075GA SA311205306GA SA311205323GA SA311205396GA SA311205564GA SA311215686GA SA311225948GA SA311225954GA SA311225982GA SA311226001GA SA311236018GA SA311236022GA SA311236026GA SA311236031GA SA311236039GA SA311236045GA SA311236048GA SA311236053GA SA311236057GA SA311236070GA SA311236071GA SA311236083GA SA311236123GA SA311236156GA SA311236157GA SA311236158GA SA311236162GA SA311236175GA SA311236176GA SA311236179GA SA311236180GA SA311236192GA SA311236193GA SA311236194GA SA311236195GA SA311236196GA SA311236197GA SA311236198GA SA311236199GA SA311236201GA SA311236203GA SA311236207GA SA311236209GA SA311236215GA SA311236219GA SA311236221GA SA311236223GA SA311236225GA SA311236229GA SA311236257GA SA311236259GA SA311236345GA SA311246389GA SA311246471GA SA311246473GA SA311246533GA SA311246541GA SA311246569GA SA311246574GA SA311246575GA SA311246577GA SA311246578GA SA311246580GA SA311246582GA SA311256850GA SA311256866GA SA311266912GA SA311266914GA SA311266923GA SA311266928GA SA311287032GA SA311287092GA SA311287098GA SA311287176GA SA311297224GA SA311297233GA SA311297235GA SA311307457GA SA311317653GA SA311317654GA SA311317659GA SA311317682GA SA311317714GA SA311317750GA SA311317757GA SA311317762GA SA311327826GA SA311327847GA SA311327890GA SA311337975GA SA311337976GA SA311337980GA SA311337981GA SA311337985GA SA311337986GA SA311337989GA SA311337997GA SA311337999GA SA311338004GA SA311428013GA SA311428026GA SA311428028GA SA311428081GA SA311448172GA SA311448174GA SA311448175GA SA311448176GA SA311448180GA SA311448193GA SA311448196GA SA311448197GA SA311448198GA SA311448199GA SA311448200GA SA311448203GA SA311448204GA SA311448206GA SA311448207GA SA311448208GA SA311448209GA SA311448210GA SA311448211GA SA311458267GA SA311458268GA SA311458269GA SA311458270GA SA311458271GA SA311458273GA SA311458274GA SA311458276GA SA311458277GA SA311458278GA SA311458280GA SA311458281GA SA311458283GA SA311458284GA SA311458285GA SA311458286GA SA311458287GA SA311458288GA SA311458289GA SA311458290GA SA311458291GA SA311458293GA SA311458294GA SA311458295GA SA311458296GA SA311458315GA SA311458318GA SA311458320GA SA311458323GA SA311458324GA SA311458325GA SA311458326GA SA311458327GA SA311458328GA SA311458329GA SA311458330GA SA311458332GA SA311458335GA SA311458336GA SA311458337GA SA311458339GA SA311468378GA SA311468379GA SA311468385GA SA311468405GA SA312029067GA SA312039075GA SA312039079GA SA312039095GA SA312069706GA SA312069724GA SA312069781GA SA312069783GA SA312069786GA SA312069797GA SA312069798GA SA312069807GA SA312069809GA SA312069813GA SA312070194GA SA312070222GA SA312070316GA SA312070337GA SA312070352GA SA312070358GA SA312070362GA SA312070379GA SA312070383GA SA312079965GA SA312080632GA SA312101437GA SA312121776GA SA312121901GA SA312121945GA SA312121954GA SA312121961GA SA312121965GA SA312121968GA SA312121969GA SA312121970GA SA312121991GA SA312122006GA SA312122019GA SA312122023GA SA312122024GA SA312122032GA SA312122034GA SA312122035GA SA312122045GA SA312122047GA SA312122049GA SA312122053GA SA312122061GA SA312122066GA SA312122070GA SA312122072GA SA312122073GA SA312122074GA SA312122075GA SA312122077GA SA312122078GA SA312122088GA SA312132096GA SA312132099GA SA312132111GA SA312132112GA SA312132120GA SA312132169GA SA312457771GA SA312457785GA SA312457801GA SA312457806GA SA312457808GA SA312502043GA SA313116688GA SA313198389GA SA313219081GA SA313240990GA SA313261929GA SA313304156GA sa313409348ga sa313419508ga sa313441488ga sa313441496ga SA313441562GA SA313451821GA sa313452089ga SA313462171GA sa313462181ga SA313462200GA SA313462211GA SA314025081GA SA314025106GA SA314025115GA SA314025121GA SA314025124GA SA314075884GA SA314076020GA SA314076065GA SA314076075GA SA314076154GA sa314076161ga SA314076198GA SA314076230GA SA314076344GA SA314086418GA SA314086446GA SA314086624GA SA314086626GA SA314097073GA SA314097094GA SA314097153GA SA314107965GA SA314107998GA SA314108272GA SA314108280GA SA314108298GA SA314108300GA SA314118360GA SA314118371GA SA314118379GA SA314118389GA SA314118404GA SA314118414GA SA314118511GA SA314118521GA SA314118523GA SA314118562GA SA314119080GA SA314119081GA SA314119083GA SA314119088GA SA314119090GA SA314119102GA SA314119103GA SA314119112GA SA314119121GA SA314119125GA SA314119126GA SA314129140GA SA314129154GA SA314129172GA SA314129256GA SA314129270GA SA314129271GA SA314129282GA SA314129284GA SA314129301GA SA314129303GA SA314129309GA SA314129312GA SA314139366GA SA314139426GA SA314139466GA SA314139475GA SA314139505GA SA314139529GA SA314139542GA SA314139550GA SA314139554GA SA314139570GA SA314139579GA SA314139586GA SA314139590GA SA314149625GA SA314149661GA SA314160002GA SA314160005GA SA314160015GA SA314160034GA SA314160036GA SA314160037GA SA314160041GA SA314160045GA SA314160047GA SA314160050GA SA314160051GA SA314160052GA SA314160059GA SA314169828GA SA314169830GA SA314169831GA SA314169832GA SA314169834GA SA314169835GA SA314169836GA SA314169837GA SA314169838GA SA314169839GA SA314169841GA SA314169847GA SA314169863GA SA314169865GA SA314169872GA SA314169881GA SA314169882GA SA314169895GA SA314169903GA SA314169919GA SA314169932GA SA314170126GA SA314170167GA SA314170174GA SA314170187GA SA314170273GA SA314170285GA SA314170294GA SA314170298GA SA314170306GA SA314170327GA SA314170362GA SA314170405GA SA314170424GA SA314170433GA SA314170454GA SA314180490GA SA314180494GA SA314180523GA SA314180554GA SA314180556GA SA314180560GA SA314180564GA SA314180565GA SA314180575GA SA314180644GA SA314180653GA SA314180658GA SA314180668GA SA314180696GA SA314180698GA SA314180699GA SA314180700GA SA314180701GA SA314180702GA SA314180705GA SA314180706GA SA314180707GA SA314180716GA SA314180721GA SA314180725GA SA314180732GA SA314180741GA SA314180751GA SA314180752GA SA314180763GA SA314180774GA SA314180780GA SA314180781GA SA314190791GA SA314190797GA SA314190798GA SA314190805GA SA314190850GA SA314190857GA SA314190859GA SA314190862GA SA314190870GA SA314190872GA SA314190885GA SA314190908GA SA314190917GA SA314190925GA SA314190926GA SA314190932GA SA314190933GA SA314190936GA SA314190940GA SA314190942GA SA314190952GA SA314190965GA SA314191125GA SA314191130GA SA314191147GA SA314191149GA SA314191158GA SA314191178GA SA314191194GA SA314201252GA SA314201268GA SA314201286GA SA314201365GA SA314201479GA SA314201496GA SA314201501GA SA314201505GA SA314201519GA SA314201612GA SA314201631GA SA314201664GA SA314201668GA SA314201670GA SA314201676GA SA314201679GA SA314201684GA SA314201693GA SA314201694GA SA314201701GA SA314201705GA SA314201709GA SA314201712GA SA314201720GA SA314201721GA SA314201722GA SA314201726GA SA314201728GA SA314201729GA SA314201731GA SA314201733GA SA314201734GA SA314201735GA SA314201736GA SA314201739GA SA314201740GA SA314201780GA SA314201781GA SA314201782GA SA314201787GA SA314201788GA SA314201789GA SA314201792GA SA314201793GA SA314201794GA SA314201797GA SA314201804GA SA314201805GA SA314201811GA SA314201815GA SA314201822GA SA314201831GA SA314201833GA SA314201842GA SA314211846GA SA314211851GA SA314211921GA SA314211953GA SA314212139GA SA314212159GA SA314212201GA SA314212226GA SA314212231GA SA314212250GA SA314212257GA SA314222282GA SA314222287GA SA314222293GA SA314222311GA SA314222312GA SA314222320GA SA314222324GA SA314222326GA SA314222328GA SA314222332GA SA314222337GA SA314222338GA SA314222339GA SA314222341GA SA314222344GA SA314222345GA SA314222446GA SA314222450GA SA314222459GA SA314222469GA SA314222492GA SA314222517GA SA314222541GA SA314222566GA SA314222575GA SA314222578GA SA314222586GA SA314222587GA SA314222609GA SA314222620GA SA314222632GA SA314222636GA SA314222646GA SA314222710GA SA314222718GA SA314222721GA SA314222723GA SA314222728GA SA314222733GA SA314222734GA SA314222740GA SA314222762GA SA314222766GA SA314222772GA SA314222774GA SA314222775GA SA314222778GA SA314222783GA SA314222787GA SA314222788GA SA314222790GA SA314222801GA SA314222806GA SA314222809GA SA314222811GA SA314222812GA SA314222819GA SA314222820GA SA314222824GA SA314222829GA SA314222831GA SA314222838GA SA314222843GA SA314222845GA SA314222846GA SA314222847GA SA314222848GA SA314222849GA SA314222858GA SA314222865GA SA314222871GA SA314222874GA SA314222877GA SA314222904GA SA314222956GA SA314232971GA SA314232973GA SA314232975GA SA314232980GA SA314232985GA SA314232988GA SA314232990GA SA314232995GA SA314233004GA SA314243212GA SA314243222GA SA314243232GA SA314243233GA SA314243237GA SA314243238GA SA314243254GA SA314243261GA SA314243262GA SA314243273GA SA314243279GA SA314243290GA SA314243317GA SA314243418GA SA314243461GA SA314243465GA SA314243467GA SA314253560GA SA314253700GA SA314253702GA SA314253703GA SA314253744GA SA314263783GA SA314263784GA SA314263787GA SA314263788GA SA314263789GA SA314263793GA SA314263794GA SA314263795GA SA314263797GA SA314263799GA SA314263800GA SA314263801GA SA314263804GA SA314263805GA SA314263806GA SA314263807GA SA314263808GA SA314263811GA SA314263812GA SA314273873GA SA314273933GA SA314273973GA SA314273992GA SA314273995GA SA314274004GA SA314274007GA SA314274018GA SA314274019GA SA314274020GA SA314274043GA SA314274047GA SA314274050GA SA314274054GA SA314274056GA SA314274058GA SA314274062GA SA314274064GA SA314274065GA SA314274068GA SA314274075GA SA314274077GA SA314274078GA SA314274080GA SA314274081GA SA314274083GA SA314274091GA SA314274092GA SA314274094GA SA314274097GA SA314274101GA SA314274105GA SA314274111GA SA314274118GA SA314274121GA SA314274127GA SA314274128GA SA314274141GA SA314274186GA SA314274194GA SA314284402GA SA314294794GA SA314294796GA SA314294819GA SA314294832GA SA314294833GA SA314294916GA SA314295155GA SA314295177GA SA314295191GA SA314295199GA SA314295210GA SA314295217GA SA314295220GA SA314295221GA SA314295241GA SA314295245GA SA314295246GA SA314295248GA SA314295253GA SA314295258GA SA314295260GA SA314295269GA SA314295287GA SA312469399GA SA312469406GA SA312469418GA SA312469419GA SA312469422GA SA312469429GA SA312469433GA SA312469435GA SA312469448GA SA312469450GA SA312469455GA SA312469467GA SA312469502GA SA312469505GA SA312469510GA SA312469513GA SA312469515GA SA312469520GA SA312479523GA SA312479525GA SA312479527GA SA312479543GA SA312479545GA SA312479548GA SA312479550GA SA312479551GA SA312479552GA SA312479554GA SA312479557GA SA312479558GA SA312479566GA SA312479568GA SA312479569GA SA312479570GA SA312479575GA SA312479584GA SA312479586GA SA312479588GA SA312479589GA SA312479590GA SA312479592GA SA312479599GA SA312479658GA SA312479671GA SA312501969GA SA312502001GA SA312502050GA SA312502054GA SA313482714GA SA314014422GA SA314014480GA SA314014632GA SA314024875GA sa314024936ga SA314055617GA SA314065737GA SA314065738GA SA314065745GA SA314065758GA SA314065772GA SA314065782GA SA314065790GA SA314075832GA SA314075836GA SA314075843GA SA314075844GA SA314075845GA SA314075847GA SA314075876GA SA314075988GA SA314076027GA SA314076074GA SA314076105GA SA314076126GA SA314076131GA SA314076133GA SA314076142GA SA314076147GA SA314076148GA SA314076158GA SA314076181GA SA314076209GA SA314076211GA SA314076212GA SA314076214GA SA314076221GA SA314076235GA SA314076239GA SA314076243GA SA314076249GA SA314076251GA SA314076256GA SA314076287GA SA314076289GA SA314076290GA SA314076294GA SA314076295GA SA314076296GA SA314076300GA SA314076304GA SA314076305GA SA314076315GA SA314076318GA SA314076319GA SA314076324GA SA314076328GA SA314076330GA SA314076333GA SA314076335GA SA314076342GA SA314076343GA SA314076345GA SA314076346GA SA314086351GA SA314086353GA SA314086355GA SA314086365GA SA314086368GA SA314086371GA SA314086376GA SA314086381GA SA314086382GA SA314086392GA SA314086395GA SA314086396GA SA314086397GA SA314086398GA SA314086399GA SA314086400GA SA314086402GA SA314086403GA SA314086409GA SA314086412GA SA314086413GA SA314086414GA SA314086415GA SA314086417GA SA314086420GA SA314086423GA SA314086424GA SA314086425GA SA314086426GA SA314086428GA SA314086429GA SA314086430GA SA314086431GA SA314086432GA SA314086434GA SA314086436GA SA314086437GA SA314086438GA SA314086439GA SA314086442GA SA314086444GA SA314086447GA SA314086449GA SA314086451GA SA314086453GA SA314086454GA SA314086455GA SA314086456GA SA314086457GA SA314086458GA SA314086460GA SA314086465GA SA314086466GA SA314086467GA SA314086473GA SA314086478GA SA314086479GA SA314086482GA SA314086483GA SA314086484GA SA314086486GA SA314086487GA SA314086489GA SA314086493GA SA314086494GA SA314086495GA SA314086545GA SA314086549GA SA314086550GA SA314086553GA SA314086560GA SA314086561GA SA314086563GA SA314086564GA SA314086567GA SA314086569GA SA314086570GA SA314086571GA SA314086572GA SA314086617GA SA314086630GA SA314086634GA SA314086636GA SA314086648GA SA314086649GA SA314086650GA SA314086654GA SA314086660GA SA314086662GA SA314086663GA SA314086665GA SA314086681GA SA314086696GA SA314086700GA SA314096777GA SA314096807GA SA314096824GA SA314096833GA SA314096835GA SA314096841GA SA314096845GA SA314096848GA SA314096850GA SA314096851GA SA314096859GA SA314096862GA SA314096865GA SA314096873GA SA314096879GA SA314096883GA SA314096886GA SA314096899GA SA314096900GA SA314096903GA SA314096904GA SA314096906GA SA314096907GA SA314096908GA SA314096909GA SA314096910GA SA314096913GA SA314096915GA SA314096916GA SA314096917GA SA314096918GA SA314096919GA SA314096920GA SA314096921GA SA314096922GA SA314096923GA SA314096924GA SA314096925GA SA314096927GA SA314096928GA SA314096931GA SA314096932GA SA314096988GA SA314096996GA SA314097013GA SA314097021GA SA314097027GA SA314097031GA SA314097032GA SA314097045GA SA314097046GA SA314097051GA SA314097244GA SA314097248GA SA314097257GA SA314097259GA SA314097261GA SA314097284GA SA314097294GA SA314097306GA SA314097314GA SA314097338GA SA314107414GA SA314107428GA SA314107430GA SA314107443GA SA314107450GA SA314107463GA SA314107871GA SA314107878GA SA314107921GA SA314108125GA SA314119100GA SA314119108GA SA314129171GA SA314129216GA SA314129307GA SA314139349GA SA314139350GA SA314139351GA SA314139353GA SA314139356GA SA314139364GA SA314139368GA SA314139373GA SA314139376GA SA314139379GA SA314139380GA SA314139381GA SA314139414GA SA314139417GA SA314139418GA SA314139420GA SA314139422GA SA314139443GA SA314139444GA SA314139445GA SA314139446GA SA314139470GA SA314139471GA SA314139492GA SA314139498GA SA314139504GA SA314139509GA SA314139511GA SA314139521GA SA314139522GA SA314139523GA SA314139524GA SA314139532GA SA314139535GA SA314139536GA SA314139549GA SA314139560GA SA314160010GA SA314160039GA SA314160069GA SA314169826GA SA314169833GA SA314169845GA SA314169846GA SA314169849GA SA314169850GA SA314169934GA SA314170121GA SA314170162GA SA314170164GA SA314170165GA SA314170172GA SA314170180GA SA314170189GA SA314170191GA SA314170194GA SA314170202GA SA314170203GA SA314170212GA SA314170232GA SA314170234GA SA314170238GA SA314170297GA SA314170315GA SA314170396GA SA314170417GA SA314170440GA SA314180630GA SA314180638GA SA314180641GA SA314180654GA SA314180657GA SA314191166GA SA314201228GA SA314201480GA SA314201844GA SA314211847GA SA314211848GA SA314211849GA SA314211854GA SA314211857GA SA314222520GA SA314222567GA SA314233078GA SA314243193GA SA314243219GA SA314243220GA SA314243242GA SA314243247GA SA314243396GA SA314243463GA SA314253642GA SA314253752GA SA314263824GA SA314273903GA SA314273944GA SA314273976GA SA314274008GA SA314284412GA SA314284515GA SA314284707GA SA314284723GA SA314284756GA SA314284758GA SA314284761GA SA314294861GA SA314294863GA SA314294866GA SA314294868GA SA314294870GA SA314294872GA SA314294875GA SA314294877GA SA314294883GA SA314295190GA SA307030426GA SA307070445GA SA307080461GA SA307080465GA SA307080490GA SA307080512GA SA307080532GA SA307080535GA SA307150663GA SA307180700GA SA307180703GA SA307180709GA SA307190763GA SA307302065GA SA307392931GA SA307443243GA SA307443364GA SA307443374GA SA307453395GA SA307453397GA SA307453550GA SA308074590GA SA308084648GA SA308105085GA SA308105090GA SA308368243GA SA308378321GA SA308388576GA SA308398715GA SA308398727GA SA308398747GA SA308398750GA SA308398757GA SA308398759GA SA308398760GA SA308398766GA SA308398775GA SA308398782GA SA308398814GA SA308398818GA SA308409089GA SA308409184GA SA308419230GA SA308419255GA SA308419263GA SA308429312GA SA308429359GA SA308439428GA SA308439430GA SA308439436GA SA308439437GA SA308439440GA SA308439441GA SA308439448GA SA308439449GA SA308449512GA SA308449518GA SA308449521GA SA308449575GA SA308449595GA SA308459641GA SA308459669GA SA308459673GA SA308501299GA SA308501338GA SA308501346GA SA308501347GA SA308501385GA SA308511570GA SA309052230GA SA309062276GA SA309173124GA SA309173130GA SA309173131GA SA309173132GA SA309173134GA SA309173135GA SA309173136GA SA309173137GA SA309173138GA SA309173139GA SA309173141GA SA309173143GA SA309173144GA SA309173148GA SA309173152GA SA309173153GA SA309173154GA SA309173156GA SA309173158GA SA309173168GA SA309365085GA SA309365101GA SA309365152GA SA309365259GA SA309365265GA SA309365267GA SA309365281GA SA309365282GA SA309365284GA SA309365317GA SA309365328GA SA309365330GA SA309365334GA SA309375374GA SA309375405GA SA309375438GA SA309375439GA SA309375445GA SA309375446GA SA309375450GA SA309385543GA SA309385551GA SA309385557GA SA309385624GA SA309395730GA SA309395761GA SA309405854GA SA309405870GA SA309405882GA SA309405908GA SA309405916GA SA309405920GA SA309436318GA SA309436350GA SA309436362GA SA309446450GA SA309456535GA SA309456545GA SA309456549GA SA309456676GA SA309466744GA SA309466758GA SA309466760GA SA309466763GA SA309466782GA SA309466784GA SA309466793GA SA309466800GA SA309466848GA SA309476863GA SA309476925GA SA309476937GA SA309476960GA SA309477074GA SA309517670GA SA309517764GA SA309517767GA SA309527909GA SA310038480GA SA310038481GA SA310038495GA SA310038499GA SA310038512GA SA310048565GA SA310048644GA SA310048652GA SA310048672GA SA310048677GA SA310048689GA SA310058703GA SA310058731GA SA310058732GA SA310058734GA SA310058736GA SA310068802GA SA310068807GA SA310068811GA SA310068817GA SA310068821GA SA310068822GA SA310068826GA SA310068829GA SA310068830GA SA310068835GA SA310068840GA SA310068841GA SA310068843GA SA310118965GA SA310118975GA SA310118992GA SA310119004GA SA310119005GA SA310119009GA SA310119018GA SA310139090GA SA310139091GA SA310139097GA SA310139099GA SA310139101GA SA310139102GA SA310139103GA SA310139105GA SA310139108GA SA310139109GA SA310139110GA SA310139111GA SA310139112GA SA310139113GA SA310139116GA SA310139117GA SA310139118GA SA310139123GA SA310139125GA SA310139128GA SA310139129GA SA310139130GA SA310139131GA SA310139136GA SA310139137GA SA310139138GA SA310139139GA SA310139140GA SA310139147GA SA310139170GA SA310139171GA SA310139182GA SA310139196GA SA310139292GA SA310139297GA SA310149347GA SA310159389GA SA310159425GA SA310159445GA SA310159446GA SA310159447GA SA310159448GA SA310159451GA SA310159452GA SA310259895GA SA310259897GA SA310259899GA SA310259901GA SA310259902GA SA310259903GA SA310259905GA SA310259906GA SA310259907GA SA310259908GA SA310259909GA SA310259910GA SA310259931GA SA310259933GA SA310259934GA SA310259935GA SA310259936GA SA310259937GA SA310259940GA SA310259947GA SA310259952GA SA310259962GA SA310259968GA SA310259969GA SA310260227GA SA310260291GA SA310260317GA SA310260356GA SA310260357GA SA310260358GA SA310260361GA SA310260363GA SA310260368GA SA310260369GA SA310260370GA SA310260378GA SA310260380GA SA310260400GA SA310260403GA SA310260419GA SA310260425GA SA310260426GA SA310260427GA SA310260431GA SA310260432GA SA310260433GA SA310260435GA SA310260440GA SA310290651GA SA310290734GA SA310290736GA SA310290738GA SA310290740GA SA310290742GA SA310331088GA SA310331354GA SA310361775GA SA310361782GA SA310361790GA SA310361793GA SA310361806GA SA310361849GA SA310361856GA SA310361883GA SA310361885GA SA310372071GA SA310372073GA SA310372074GA SA310372078GA SA310372081GA SA310372083GA SA310372086GA SA310372087GA SA310372093GA SA310372094GA SA310372103GA SA310372106GA SA310372109GA SA310372111GA SA310372112GA SA310372113GA SA310372119GA SA310372121GA SA310372124GA SA310372126GA SA310372127GA SA310372128GA SA310372129GA SA310372131GA SA310372133GA SA310372134GA SA310372165GA SA310372166GA SA310372169GA SA310372170GA SA310372175GA SA310372181GA SA310372182GA SA310372184GA SA310372185GA SA310372195GA SA310372196GA SA310372208GA SA310372211GA SA310382315GA SA310382319GA SA310382320GA SA310382321GA SA310382322GA SA310382324GA SA310382327GA SA310382328GA SA310382337GA SA310382338GA SA310382346GA SA310382350GA SA310382354GA SA310382356GA SA310382358GA SA310382361GA SA310382362GA SA310382365GA SA310382367GA SA310382368GA SA310382370GA SA310382372GA SA310382373GA SA310382433GA SA310382525GA SA310382526GA SA310382536GA SA310382546GA SA310382549GA SA310382551GA SA310382564GA SA310382567GA SA310382594GA SA310382595GA SA310382597GA SA310382653GA SA310382666GA SA310382669GA SA310382678GA SA310382681GA SA310382683GA SA310382688GA SA310382694GA SA310382702GA SA310382703GA SA310382711GA SA310382772GA SA310382773GA SA310382779GA SA310382781GA SA310382788GA SA310382789GA SA310382799GA SA310382803GA SA310382804GA SA310382807GA SA310382810GA SA310382811GA SA310382812GA SA310382815GA SA310382837GA SA310382841GA SA310382845GA SA310392851GA SA310392852GA SA310392854GA SA310392856GA SA310392857GA SA310392858GA SA310392860GA SA310392861GA SA310392862GA SA310392863GA SA310392869GA SA310392875GA SA310392877GA SA310392878GA SA310392880GA SA310392881GA SA310392883GA SA310392885GA SA310392888GA SA310392892GA SA310392896GA SA310392906GA SA310392908GA SA310392913GA SA310392916GA SA310392918GA SA310392920GA SA310392921GA SA310392922GA SA310392926GA SA310392927GA SA310392928GA SA310392929GA SA310392932GA SA310393078GA SA310393079GA SA310393081GA SA310393082GA SA310393087GA SA310393091GA SA310393093GA SA310393098GA SA310393099GA SA310393100GA SA310393101GA SA310393102GA SA310393103GA SA310393104GA SA310393105GA SA310393107GA SA310393110GA SA310393111GA SA310393113GA SA310393115GA SA310393132GA SA310393133GA SA310393134GA SA310393135GA SA310393137GA SA310393181GA SA310393189GA SA310393199GA SA310393201GA SA310393204GA SA310393205GA SA310393206GA SA310393207GA SA310393209GA SA310393211GA SA310393214GA SA310393218GA SA310393220GA SA310393223GA SA310393225GA SA310393226GA SA310393228GA SA310393232GA SA310393240GA SA310393241GA SA310393244GA SA310393245GA SA310393247GA SA310393253GA SA310393256GA SA310393264GA SA310393266GA SA310393273GA SA310393286GA SA310393289GA SA310393301GA SA310393307GA SA310403344GA SA310403345GA SA310403376GA SA310403386GA SA310403388GA SA310403392GA SA310403400GA SA310403401GA SA310403406GA SA310403425GA SA310403437GA SA310403445GA SA310403446GA SA310423760GA SA310423762GA SA310423764GA SA310423765GA SA310423775GA SA310423783GA SA310423897GA SA310423901GA SA310423902GA SA310423915GA SA310423918GA SA310423981GA SA310423983GA SA310423996GA SA310424013GA SA310434050GA SA310434072GA SA310434168GA SA310434174GA SA310434190GA SA310434201GA SA310475452GA SA310475471GA SA310475520GA SA310475535GA SA310475538GA SA310475541GA SA310475546GA SA310475547GA SA310475549GA SA310475552GA SA310475556GA SA310475565GA SA310475567GA SA310475568GA SA310475569GA SA310475571GA SA310475574GA SA310475576GA SA310475577GA SA310475580GA SA310475584GA SA310475587GA SA310475588GA SA310475590GA SA310475593GA SA310475594GA SA310475595GA SA310496279GA SA310496458GA SA310496468GA SA310496497GA SA310496648GA SA310506796GA SA310506827GA SA310506828GA SA310506835GA SA310506878GA SA310506994GA SA310506996GA SA310507013GA SA310507030GA SA310507034GA SA310507042GA SA310507044GA SA310507052GA SA310507053GA SA310507060GA SA310507068GA SA310507098GA SA310507099GA SA310507103GA SA310507126GA SA310507142GA SA310507175GA SA310507179GA SA310507182GA SA310507188GA SA310507220GA SA310507224GA SA310507246GA SA310507260GA SA310507271GA SA310507411GA SA310507528GA SA310507532GA SA310518200GA SA310518213GA SA310518381GA SA310518454GA SA310518508GA SA310518510GA SA310518536GA SA310518545GA SA310518550GA SA310518558GA SA310518620GA SA310528734GA SA310528765GA SA310528772GA SA310528790GA SA310528806GA SA310528829GA SA310528853GA SA310528904GA SA310528905GA SA310528907GA SA310528909GA SA310528910GA SA310528946GA SA310529026GA SA310529055GA SA310529056GA SA310529110GA SA310529147GA SA310529155GA SA311039375GA SA311039438GA SA311039440GA SA311039448GA SA311039470GA SA311060001GA SA311060003GA SA311060005GA SA311060006GA SA311060040GA SA311060043GA SA311060044GA SA311060066GA SA311060071GA SA311060099GA SA311060110GA SA311060111GA SA311060149GA SA311060185GA SA311060189GA SA311060205GA SA311060207GA SA311060227GA SA311060267GA SA311060291GA SA311060340GA SA311060347GA SA311060373GA SA311060388GA SA311069680GA SA311069739GA SA311069741GA SA311069766GA SA311069821GA SA311069823GA SA311069825GA SA311069855GA SA311069859GA SA311069862GA SA311069864GA SA311069865GA SA311069866GA SA311069868GA SA311070514GA SA311070543GA SA311070558GA SA311080602GA SA311080604GA SA311080630GA SA311080650GA SA311080654GA SA311080659GA SA311090913GA SA311090948GA SA311090973GA SA311091012GA SA311091029GA SA311101312GA SA311101315GA SA311101329GA SA311101373GA SA311101378GA SA311111599GA SA311111606GA SA311121888GA SA311121968GA SA311122106GA SA311122244GA SA311132284GA SA311132466GA SA311132473GA SA311132516GA SA311132524GA SA311132596GA SA311142695GA SA311142697GA SA311142826GA SA311142970GA SA311153082GA SA311153121GA SA311153128GA SA311153131GA SA311153140GA SA311153151GA SA311153312GA SA311153354GA SA311163368GA SA311163407GA SA311163420GA SA311163713GA SA311163715GA SA311163728GA SA311163733GA SA311163781GA SA311174108GA SA311174160GA SA311174162GA SA311174179GA SA311174201GA SA311174206GA SA311174237GA SA311174239GA SA311184387GA SA311184399GA SA311184400GA SA311184401GA SA311184402GA SA311184410GA SA311184413GA SA311184497GA SA311184520GA SA311184524GA SA311184632GA SA311184754GA SA311184803GA SA311194935GA SA311195009GA SA311195147GA SA311195156GA SA311195255GA SA311195256GA SA311195264GA SA311195268GA SA311195270GA SA311195271GA SA311195287GA SA311195288GA SA311205325GA SA311205329GA SA311205337GA SA311205358GA SA311205416GA SA311205436GA SA311205441GA SA311205572GA SA311205610GA SA311205621GA SA311215683GA SA311215685GA SA311215693GA SA311215694GA SA311215696GA SA311215700GA SA311215702GA SA311215714GA SA311215717GA SA311215723GA SA311215739GA SA311215858GA SA311225895GA SA311225938GA SA311225974GA SA311225977GA SA311225979GA SA311225981GA SA311225986GA SA311225989GA SA311225991GA SA311225995GA SA311225997GA SA311236003GA SA311236034GA SA311236038GA SA311236080GA SA311236084GA SA311236115GA SA311236119GA SA311236121GA SA311236302GA SA311246446GA SA311246459GA SA311246477GA SA311246486GA SA311246488GA SA311246489GA SA311246494GA SA311246514GA SA311246584GA SA311246589GA SA311246610GA SA311246612GA SA311246623GA SA311246630GA SA311246636GA SA311246657GA SA311246661GA SA311246668GA SA311246671GA SA311256683GA SA311256684GA SA311256686GA SA311256690GA SA311256691GA SA311256697GA SA311256700GA SA311256704GA SA311256710GA SA311256746GA SA311256749GA SA311256754GA SA311256764GA SA311256779GA SA311256810GA SA311256817GA SA311256827GA SA311256828GA SA311256845GA SA311256848GA SA311256852GA SA311256859GA SA311256860GA SA311256862GA SA311287173GA SA311297264GA SA311297350GA SA311297352GA SA311307458GA SA311307473GA SA311317661GA SA311317669GA SA311317759GA SA311317766GA SA311317774GA SA311327910GA SA311327920GA SA311327934GA SA311428011GA SA311428057GA SA311428073GA SA311428090GA SA311428094GA SA311448173GA SA311448179GA SA311448181GA SA311448192GA SA311448194GA SA311448195GA SA311448202GA SA311458220GA SA311468399GA SA312018777GA SA312069751GA SA312091056GA SA313084766GA SA313084770GA SA313084949GA SA313095056GA SA313095057GA SA313095083GA SA313095088GA SA313095091GA SA313095101GA SA313095104GA SA313095111GA SA313095117GA SA313095119GA SA313095121GA SA313095125GA SA313095134GA SA313095195GA SA313095233GA SA313095236GA SA313095237GA SA313095261GA SA313095275GA SA313095280GA SA313095281GA SA313095283GA SA313095291GA sa313419493ga SA313419526GA sa313419541ga sa313419546ga sa313420291ga sa313441519ga sa313441696ga sa313441708ga sa313451807ga SA313452129GA SA313482770GA sa314004321ga SA314014362GA SA314014477GA SA314014740GA SA314055714GA SA314055719GA SA314075969GA SA314076000GA SA314086501GA SA314097246GA SA314097267GA SA314107453GA SA314118979GA SA314170184GA SA314170190GA SA314170198GA SA314170200GA SA314170223GA SA314170286GA SA314180489GA SA314180611GA SA314180636GA SA314211853GA SA314253675GA SA314253750GA SA312049273GA SA312049274GA SA312049380GA SA312049398GA SA312049404GA SA312049412GA SA312049435GA SA312049439GA SA312049460GA SA312059484GA SA312059487GA SA312059491GA SA312059492GA SA312059495GA SA312059498GA SA312059504GA SA312059505GA SA312059506GA SA312059508GA SA312059509GA SA312059529GA SA312059532GA SA312059586GA SA312059595GA SA312069805GA SA312069806GA SA312069808GA SA312069817GA SA312069847GA SA312069858GA SA312069866GA SA312069867GA SA312069879GA SA312070302GA SA312079912GA SA312079932GA SA312091291GA SA312091296GA SA312091321GA SA312091327GA SA312091344GA SA312091361GA SA312101369GA SA312101372GA SA312101402GA SA312101405GA SA312101407GA SA312101415GA SA312111663GA SA312121701GA SA312121702GA SA312121741GA SA312121749GA SA312121750GA SA312121806GA SA312121819GA SA312121848GA SA312121895GA SA312121971GA SA312121979GA SA312121987GA SA312132098GA SA312152319GA SA312152335GA SA312152344GA SA312152347GA SA312152361GA SA312152473GA SA312232796GA SA312232799GA SA312232805GA SA312232810GA SA312242995GA SA312242996GA SA312243006GA SA312253069GA SA312253083GA SA312253090GA SA312253116GA SA312253117GA SA312253142GA SA312253161GA SA312253162GA SA312253215GA SA312253221GA SA312253233GA SA312253238GA SA312253239GA SA312253247GA SA312253256GA SA312253272GA SA312253279GA SA312253369GA SA312253425GA SA312253426GA SA312253456GA SA312253459GA SA312253463GA SA312253467GA SA312253475GA SA312263866GA SA312263936GA SA312345056GA SA312345104GA SA312345118GA SA312345120GA SA312345130GA SA312386320GA SA312396972GA SA312396991GA SA312469244GA SA312469463GA SA312470044GA SA312480421GA SA312480441GA SA312480558GA SA312480559GA SA312480571GA SA312480577GA SA312480578GA SA312480595GA SA312480597GA SA312480607GA SA312480700GA SA312480703GA SA312480712GA SA312480713GA SA312480714GA SA312480716GA SA312480718GA SA312480720GA SA312480807GA SA312490893GA SA312490907GA SA312490908GA SA312490914GA SA312502060GA SA312502095GA SA312512257GA SA312512310GA SA312512311GA SA312512321GA SA312512328GA SA312512346GA SA313012804GA SA313013078GA SA313013085GA SA313013099GA SA313013100GA SA313013104GA SA313064248GA SA313084793GA SA313084835GA SA313084840GA SA313084843GA SA313084873GA SA313084962GA SA313084967GA SA313095063GA SA313095074GA SA313095082GA SA313095130GA SA313095133GA SA313095298GA SA313095301GA SA313095316GA SA313095318GA SA313095328GA SA313095331GA SA313095338GA SA313095339GA SA313095346GA SA313095351GA SA313095353GA SA313095357GA SA313095360GA SA313095361GA SA313095362GA SA313095364GA SA313095366GA SA313095370GA SA313095372GA SA313095398GA SA313095457GA SA313095476GA SA313116853GA SA313117144GA SA313127399GA SA313127505GA SA313127509GA SA313127564GA SA313127600GA SA313147971GA SA313147974GA SA313147983GA SA313148020GA SA313198667GA SA313208761GA SA313219020GA sa313230674ga SA313261719GA SA313272078GA SA313292981GA SA313293026GA SA313293034GA SA313293043GA SA313293243GA SA313314449GA SA313314507GA SA313314515GA SA313314590GA SA313314594GA SA313314595GA SA313314596GA SA313314604GA SA313314625GA SA313314766GA SA313314797GA SA313314815GA SA313314834GA SA313314835GA SA313314841GA SA313314846GA SA313314860GA SA313314868GA SA313314871GA SA313314872GA SA313314874GA SA313314877GA SA313314878GA SA313314881GA SA313314884GA SA313314888GA SA313314891GA SA313314892GA SA313324927GA SA313324932GA SA313324994GA SA313335046GA SA313335060GA SA313335089GA SA313335090GA SA313335091GA SA313335094GA SA313335111GA SA313335112GA SA313335124GA SA313335134GA SA313335255GA SA313335263GA SA313335264GA SA313335266GA SA313335267GA SA313335271GA SA313335278GA SA313335283GA SA313335284GA SA313335294GA SA313335297GA SA313335298GA SA313335303GA SA313335311GA SA313335335GA SA313335341GA SA313335343GA SA313335347GA SA313335350GA SA313335351GA SA313335353GA SA313335362GA SA313335366GA SA313335373GA SA313345539GA SA313345567GA SA313345575GA SA313345590GA SA313345593GA SA313345595GA SA313345599GA SA313378331GA SA313378430GA SA313378433GA SA313378434GA SA313378435GA SA313378449GA SA313378453GA SA313378458GA SA313378479GA SA313388492GA SA313388499GA SA313388502GA SA313388504GA SA313388515GA SA313388517GA SA313388518GA SA313388519GA SA313388527GA SA313388528GA SA313388529GA SA313388541GA SA313388544GA SA313388545GA SA313388546GA SA313388548GA SA313388553GA SA313388556GA SA313419518GA SA313419523GA sa313419527ga sa313419528ga sa313419530ga sa313419548ga sa313419549ga SA313419736GA SA313419737GA SA313419738GA SA313419740GA SA313419745GA SA313419746GA SA313419747GA SA313419749GA SA313419750GA SA313419752GA SA313419753GA SA313419754GA SA313419755GA SA313419757GA SA313419761GA SA313419791GA SA313419792GA SA313420540GA SA313430690GA SA313431174GA SA313431178GA SA313431184GA SA313431195GA SA313431197GA SA313431226GA SA313431231GA SA313431235GA SA313441245GA SA313441281GA SA313441284GA SA313441285GA SA313441323GA SA313441327GA SA313441329GA SA313441330GA SA313441331GA SA313441332GA SA313441333GA SA313441336GA SA313441339GA SA313441342GA SA313441343GA SA313441349GA SA313441356GA SA313441364GA SA313441380GA SA313441381GA SA313441387GA SA313441390GA SA313441396GA SA313441419GA SA313441445GA SA313441449GA SA313451779GA SA313451811GA SA313451913GA SA313451948GA SA313451995GA SA313451997GA SA313452008GA SA313452025GA SA313452043GA SA313452071GA SA313452077GA SA313452082GA SA313452094GA SA313452098GA SA313452105GA SA313452115GA SA313452117GA SA313452118GA SA313452124GA SA313452128GA SA313452140GA SA313462217GA SA313462218GA SA313462220GA SA313462222GA SA313462223GA SA313462224GA SA313462227GA SA313462237GA SA313462239GA SA313462241GA SA313462243GA SA313462249GA SA313462255GA SA313462266GA SA313462273GA SA313462278GA SA313462279GA SA313462282GA SA313462302GA SA313462322GA SA313462341GA SA313462346GA SA313462410GA SA313493106GA SA313493203GA SA314004318GA SA314014370GA SA314014372GA SA314014444GA SA314014464GA SA314014744GA SA314014756GA SA314024771GA SA314024772GA SA314024774GA SA314024777GA SA314024780GA SA314024784GA SA314024786GA SA314024788GA SA314024801GA SA314024828GA SA314024830GA SA314024836GA SA314024838GA SA314024839GA SA314024847GA SA314024862GA SA314024869GA SA314024893GA SA314024897GA SA314024907GA SA314024911GA SA314024928GA SA314024941GA SA314025077GA SA314025210GA SA314025232GA SA314035250GA SA314035326GA SA314035327GA SA314035345GA SA314045509GA SA314045510GA SA314045517GA SA314045534GA SA314045537GA SA314055543GA SA314055545GA SA314055563GA SA314055577GA SA314055578GA SA314076067GA SA314096926GA SA314096930GA SA314097062GA SA314097066GA SA314097081GA SA314097082GA SA314107377GA SA314108303GA SA314108305GA SA314108306GA SA314118340GA SA314118341GA SA314118342GA SA314118343GA SA314118344GA SA314118345GA SA314118346GA SA314118348GA SA314118350GA SA314118351GA SA314118353GA SA314118355GA SA314118358GA SA314118359GA SA314118968GA SA314118975GA SA314118995GA SA314119025GA SA314119077GA SA314119122GA SA314129260GA SA314129274GA SA314129285GA SA314180613GA SA314180624GA SA314190951GA SA314191193GA SA314191201GA SA314191206GA SA314201588GA SA314211929GA SA314212017GA SA314212019GA SA314212021GA SA314212028GA SA314212038GA SA314212241GA SA314253681GA SA307080495GA SA307221159GA SA309324383GA SA309324502GA SA309324525GA SA309334769GA SA309334773GA SA309334775GA SA309334787GA SA309334792GA SA309334796GA SA309334798GA SA309334805GA SA309334817GA SA309334818GA SA309334820GA SA309334821GA SA309334822GA SA309334824GA SA309334827GA SA309354890GA SA309354899GA SA309354911GA SA309354929GA SA309354931GA SA309354946GA SA309354951GA SA309354958GA SA309354981GA SA309355018GA SA309365030GA SA309365031GA SA309365035GA SA309365040GA SA309365049GA SA309365052GA SA309365058GA SA309365070GA SA309365087GA SA309365089GA SA309365100GA SA309365106GA SA309365107GA SA309365119GA SA309365121GA SA309365146GA SA309365148GA SA309365149GA SA309365153GA SA309365156GA SA309365159GA SA309365160GA SA309365161GA SA309365169GA SA309365170GA SA309365175GA SA309365176GA SA309365188GA SA309365195GA SA309365200GA SA309365208GA SA309365210GA SA309365211GA SA309365212GA SA309365218GA SA309365219GA SA309365238GA SA309365242GA SA309365244GA SA309365247GA SA309365253GA SA309365255GA SA309365256GA SA309365269GA SA309365313GA SA309365314GA SA309375365GA SA309385617GA SA309385674GA SA309395759GA SA309405817GA SA309405818GA SA309405849GA SA309405863GA SA309405872GA SA309415923GA SA309415932GA SA309415940GA SA309426066GA SA309426067GA SA309426071GA SA309426078GA SA309446401GA SA309456719GA SA309456728GA SA309456729GA SA309466742GA SA309466775GA SA309466777GA SA309466802GA SA309466806GA SA309466811GA SA309466838GA SA309466839GA SA309466855GA SA309476864GA SA309476873GA SA309476876GA SA309476877GA SA309476880GA SA309476882GA SA309476899GA SA309476900GA SA309507404GA SA309507441GA SA309517501GA SA309517524GA SA309517526GA SA309517539GA SA309517627GA SA309517636GA SA309517641GA SA309517642GA SA309517643GA SA309517654GA SA309517839GA SA310028109GA SA310028146GA SA310028156GA SA310028159GA SA310028194GA SA310028198GA SA310028204GA SA310028207GA SA310028210GA SA310028278GA SA310028302GA SA310038374GA SA310038432GA SA310068847GA SA310068849GA SA310068850GA SA310068854GA SA310068873GA SA310068878GA SA310068883GA SA310068905GA SA310068909GA SA310068910GA SA310068928GA SA310068932GA SA310068937GA SA310068939GA SA310139040GA SA310139042GA SA310139043GA SA310139048GA SA310139049GA SA310139050GA SA310139053GA SA310139060GA SA310139064GA SA310139072GA SA310139078GA SA310139079GA SA310259896GA SA310259900GA SA310259945GA SA310260366GA SA310260371GA SA310260391GA SA310290751GA SA310331082GA SA310331083GA SA310331084GA SA310331085GA SA310331086GA SA310331087GA SA310331089GA SA310331090GA SA310331091GA SA310331093GA SA310331094GA SA310331095GA SA310331096GA SA310331097GA SA310331099GA SA310331100GA SA310331101GA SA310331102GA SA310331103GA SA310331105GA SA310331106GA SA310331107GA SA310331108GA SA310331110GA SA310331111GA SA310331112GA SA310331113GA SA310331116GA SA310331119GA SA310331120GA SA310331121GA SA310331122GA SA310331124GA SA310331125GA SA310331129GA SA310331130GA SA310331131GA SA310331132GA SA310331133GA SA310331134GA SA310331135GA SA310331137GA SA310331140GA SA310331141GA SA310331142GA SA310331143GA SA310331144GA SA310331146GA SA310331147GA SA310331148GA SA310331150GA SA310331151GA SA310331152GA SA310331153GA SA310331155GA SA310331156GA SA310331157GA SA310331158GA SA310331161GA SA310331163GA SA310331164GA SA310331165GA SA310331166GA SA310331167GA SA310331168GA SA310331169GA SA310331170GA SA310331171GA SA310331172GA SA310331173GA SA310331174GA SA310331175GA SA310331176GA SA310331177GA SA310331178GA SA310331179GA SA310331180GA SA310331181GA SA310331182GA SA310331183GA SA310331184GA SA310331185GA SA310331186GA SA310331187GA SA310331188GA SA310331189GA SA310331190GA SA310331191GA SA310331192GA SA310331193GA SA310331194GA SA310331195GA SA310331196GA SA310331197GA SA310331198GA SA310331199GA SA310331200GA SA310331201GA SA310331202GA SA310331203GA SA310331204GA SA310331205GA SA310331206GA SA310331207GA SA310331208GA SA310331209GA SA310331210GA SA310331211GA SA310331212GA SA310331213GA SA310475449GA SA310475458GA SA310475468GA SA310475470GA SA310475473GA SA310475479GA SA310475480GA SA310475484GA SA310475499GA SA310475504GA SA310475505GA SA310475508GA SA310475509GA SA310475511GA SA310475522GA SA310475523GA SA310475524GA SA310475533GA SA310496224GA SA310496455GA SA310496505GA SA310496521GA SA310496525GA SA310496621GA SA310496627GA SA310496637GA SA310496641GA SA310496642GA SA310496644GA SA310506688GA SA310506690GA SA310506727GA SA310506735GA SA310506763GA SA310506765GA SA310506770GA SA310507046GA SA310507059GA SA310507063GA SA310507133GA SA310507144GA SA310507151GA SA310507155GA SA310507160GA SA310507193GA SA310507393GA SA310518285GA SA310518300GA SA310518308GA SA310518315GA SA310518317GA SA311142681GA SA311142691GA SA311142701GA SA311142704GA SA311142715GA SA311153057GA SA311297313GA SA311307509GA SA311307580GA SA311307594GA SA311317666GA SA311317673GA SA311317688GA SA311317706GA SA311317711GA SA311317715GA SA311317717GA SA311317719GA SA311317727GA SA311317728GA SA311317739GA SA311317741GA SA311317742GA SA311327900GA SA311327911GA SA311327913GA SA311327932GA SA311337993GA SA311337996GA SA311338003GA SA312028911GA SA312029048GA SA312049322GA SA312049337GA SA312049338GA SA312049343GA SA312049349GA SA312049352GA SA312049455GA SA312059482GA SA312059485GA SA312059488GA SA312059499GA SA312069820GA SA312069825GA SA312069831GA SA312069833GA SA312069848GA SA312069851GA SA312069853GA SA312070118GA SA312070134GA SA312070186GA SA312070251GA SA312070340GA SA312079923GA SA312079939GA SA312079955GA SA312080392GA SA312080562GA SA312080565GA SA312080651GA SA312080685GA SA312080702GA SA312080723GA SA312080730GA SA312080736GA SA312080737GA SA312080739GA SA312080740GA SA312080741GA SA312080745GA SA312080746GA SA312080750GA SA312080752GA SA312080755GA SA312080760GA SA312080762GA SA312080774GA SA312080776GA SA312080780GA SA312080788GA SA312080795GA SA312080799GA SA312080815GA SA312080823GA SA312080838GA SA312080847GA SA312080849GA SA312080851GA SA312080853GA SA312080857GA SA312080859GA SA312080862GA SA312080863GA SA312080865GA SA312080866GA SA312080915GA SA312091100GA SA312091264GA SA312121760GA SA312121761GA SA312121763GA SA312121766GA SA312121769GA SA312121770GA SA312121773GA SA312121775GA SA312121780GA SA312121795GA SA312121808GA SA312121815GA SA312121816GA SA312121820GA SA312121823GA SA312121850GA SA312121852GA SA312122011GA SA312122021GA SA312152349GA SA312152480GA SA312152497GA SA312152547GA SA312232797GA SA312232802GA SA312232815GA SA312232821GA SA312253077GA SA312253119GA SA312263887GA SA312263931GA SA312264009GA SA312284262GA SA312284275GA SA312294622GA SA312294649GA SA312294650GA SA312294655GA SA312294672GA SA312294676GA SA312344933GA SA312345049GA SA312345066GA SA312345107GA SA312345110GA SA312345112GA SA312345113GA SA312345114GA SA312345115GA SA312355253GA SA312376078GA SA312376286GA SA312386323GA SA312386335GA SA312386339GA SA312386342GA SA312386343GA SA312386346GA SA312386386GA SA312386397GA SA312386429GA SA312386497GA SA312386521GA SA312386530GA SA312386561GA SA312386573GA SA312386574GA SA312386576GA SA312386590GA SA312386611GA SA312386612GA SA312386616GA SA312386618GA SA312386619GA SA312386621GA SA312386634GA SA312386635GA SA312386636GA SA312386638GA SA312386639GA SA312386640GA SA312386665GA SA312396787GA SA312396807GA SA312396861GA SA312396910GA SA312396941GA SA312396945GA SA312396980GA SA312397018GA SA312397025GA SA312397035GA SA312397038GA SA312397049GA SA312397050GA SA312397051GA SA312397056GA SA312397060GA SA312397077GA SA312397088GA SA312397092GA SA312397098GA SA312397119GA SA312397137GA SA312397163GA SA312397167GA SA312397172GA SA312397269GA SA312397275GA SA312407276GA SA312407277GA SA312407279GA SA312407284GA SA312407285GA SA312427418GA SA312447421GA SA312447429GA SA312447433GA SA312447434GA SA312447435GA SA312447436GA SA312447437GA SA312447439GA SA312447442GA SA312447443GA SA312447445GA SA312447446GA SA312447449GA SA312457601GA SA312457602GA SA312457603GA SA312457604GA SA312457605GA SA312457606GA SA312457607GA SA312457608GA SA312457609GA SA312457610GA SA312457611GA SA312457612GA SA312457613GA SA312457614GA SA312457615GA SA312457616GA SA312457617GA SA312457618GA SA312457619GA SA312457620GA SA312457621GA SA312457622GA SA312457623GA SA312457624GA SA312457625GA SA312457627GA SA312457628GA SA312457629GA SA312457630GA SA312457631GA SA312457632GA SA312457633GA SA312457634GA SA312457635GA SA312457636GA SA312457637GA SA312457638GA SA312457639GA SA312457640GA SA312457641GA SA312457642GA SA312457643GA SA312457644GA SA312457645GA SA312457646GA SA312457647GA SA312457648GA SA312457649GA SA312457650GA SA312457651GA SA312457652GA SA312457653GA SA312457654GA SA312457655GA SA312457656GA SA312457658GA SA312457659GA SA312457660GA SA312457661GA SA312457662GA SA312457663GA SA312457664GA SA312457665GA SA312457666GA SA312457667GA SA312457668GA SA312457669GA SA312457670GA SA312457671GA SA312457672GA SA312457673GA SA312457674GA SA312457675GA SA312457676GA SA312457677GA SA312457678GA SA312457680GA SA312457681GA SA312457682GA SA312457684GA SA312457687GA SA312457688GA SA312457689GA SA312457690GA SA312457691GA SA312457694GA SA312457695GA SA312457696GA SA312457697GA SA312457698GA SA312457699GA SA312457708GA SA312457709GA SA312457710GA SA312457712GA SA312457713GA SA312457714GA SA312457715GA SA312457716GA SA312457717GA SA312457718GA SA312457720GA SA312457722GA SA312457724GA SA312457725GA SA312457729GA SA312457730GA SA312457731GA SA312457732GA SA312457733GA SA312457734GA SA312457735GA SA312457737GA SA312457738GA SA312457739GA SA312457741GA SA312457742GA SA312457743GA SA312457744GA SA312457745GA SA312457746GA SA312457747GA SA312457748GA SA312457749GA SA312457750GA SA312457751GA SA312457752GA SA312457753GA SA312457754GA SA312457755GA SA312457756GA SA312457757GA SA312457758GA SA312457760GA SA312457761GA SA312457762GA SA312457763GA SA312457764GA SA312457765GA SA312457766GA SA312457767GA SA312457769GA SA312457770GA SA312457772GA SA312457773GA SA312457774GA SA312457775GA SA312457776GA SA312457777GA SA312457778GA SA312457779GA SA312457780GA SA312457781GA SA312457782GA SA312457783GA SA312457784GA SA312457787GA SA312457788GA SA312457789GA SA312457790GA SA312457792GA SA312457793GA SA312457794GA SA312457796GA SA312457797GA SA312457798GA SA312457799GA SA312457802GA SA312457803GA SA312457804GA SA312457809GA SA312457813GA SA312457815GA SA312457816GA SA312457817GA SA312457818GA SA312457819GA SA312457821GA SA312457822GA SA312457823GA SA312457824GA SA312457825GA SA312457826GA SA312457827GA SA312457828GA SA312457829GA SA312457830GA SA312457831GA SA312457832GA SA312457833GA SA312457834GA SA312457835GA SA312457836GA SA312457837GA SA312457838GA SA312457839GA SA312457840GA SA312457841GA SA312457842GA SA312457843GA SA312457844GA SA312457845GA SA312457846GA SA312457847GA SA312457849GA SA312457850GA SA312457851GA SA312457852GA SA312457853GA SA312457854GA SA312457855GA SA312457856GA SA312457857GA SA312457859GA SA312457860GA SA312457861GA SA312457862GA SA312457863GA SA312457864GA SA312457865GA SA312457866GA SA312457867GA SA312457868GA SA312457869GA SA312457870GA SA312457871GA SA312457872GA SA312457873GA SA312457874GA SA312457875GA SA312457876GA SA312457877GA SA312457879GA SA312457880GA SA312457881GA SA312457882GA SA312457883GA SA312457884GA SA312457885GA SA312457886GA SA312457887GA SA312457888GA SA312457890GA SA312457892GA SA312457893GA SA312457894GA SA312457895GA SA312457896GA SA312457897GA SA312457898GA SA312457899GA SA312457900GA SA312457901GA SA312457902GA SA312457903GA SA312457904GA SA312457905GA SA312457906GA SA312457907GA SA312457908GA SA312457909GA SA312457911GA SA312457912GA SA312457913GA SA312457916GA SA312457917GA SA312457918GA SA312457919GA SA312457920GA SA312457921GA SA312457922GA SA312457924GA SA312457925GA SA312457926GA SA312457928GA SA312457929GA SA312457930GA SA312457931GA SA312457932GA SA312457933GA SA312457934GA SA312457935GA SA312457936GA SA312457937GA SA312457938GA SA312457939GA SA312457940GA SA312457941GA SA312457942GA SA312457943GA SA312457944GA SA312457945GA SA312457946GA SA312457947GA SA312457948GA SA312457949GA SA312457950GA SA312457951GA SA312457952GA SA312457953GA SA312457954GA SA312457955GA SA312457956GA SA312457957GA SA312457958GA SA312457959GA SA312457960GA SA312457961GA SA312457962GA SA312457963GA SA312457965GA SA312457966GA SA312457967GA SA312457970GA SA312457971GA SA312457972GA SA312457973GA SA312457974GA SA312457975GA SA312457976GA SA312457977GA SA312457978GA SA312457979GA SA312457981GA SA312457982GA SA312457983GA SA312457986GA SA312457990GA SA312457994GA SA312457996GA SA312457997GA SA312457999GA SA312458000GA SA312458002GA SA312458003GA SA312458004GA SA312458005GA SA312458006GA SA312458008GA SA312458009GA SA312458010GA SA312458011GA SA312458012GA SA312458013GA SA312458016GA SA312458017GA SA312458019GA SA312458022GA SA312458023GA SA312458024GA SA312458025GA SA312458026GA SA312458028GA SA312458030GA SA312458032GA SA312458034GA SA312458037GA SA312458038GA SA312458040GA SA312458041GA SA312458042GA SA312458043GA SA312458044GA SA312458045GA SA312458048GA SA312458049GA SA312458050GA SA312458051GA SA312458052GA SA312458053GA SA312458054GA SA312458056GA SA312458057GA SA312458058GA SA312458059GA SA312458061GA SA312458062GA SA312458063GA SA312458066GA SA312458067GA SA312458068GA SA312458069GA SA312458070GA SA312458071GA SA312458074GA SA312458076GA SA312458077GA SA312458079GA SA312458080GA SA312458082GA SA312458084GA SA312458085GA SA312458086GA SA312458087GA SA312458088GA SA312458089GA SA312458090GA SA312458091GA SA312458093GA SA312458095GA SA312458097GA SA312458098GA SA312458100GA SA312458101GA SA312458102GA SA312458105GA SA312458106GA SA312458108GA SA312458109GA SA312458110GA SA312458112GA SA312458113GA SA312458116GA SA312458118GA SA312458120GA SA312458121GA SA312458122GA SA312458124GA SA312458125GA SA312458126GA SA312458127GA SA312458179GA SA312458182GA SA312458183GA SA312458184GA SA312458188GA SA312458190GA SA312458191GA SA312458193GA SA312458196GA SA312458197GA SA312458198GA SA312458201GA SA312458202GA SA312458204GA SA312458205GA SA312458206GA SA312458207GA SA312458208GA SA312458209GA SA312458210GA SA312458212GA SA312458213GA SA312458215GA SA312458216GA SA312458217GA SA312458218GA SA312458219GA SA312458220GA SA312458221GA SA312458222GA SA312458224GA SA312458227GA SA312458229GA SA312458230GA SA312458231GA SA312458234GA SA312458235GA SA312458236GA SA312458237GA SA312458238GA SA312458239GA SA312458240GA SA312458241GA SA312458242GA SA312458243GA SA312458247GA SA312458248GA SA312458249GA SA312458250GA SA312458251GA SA312458254GA SA312458255GA SA312458257GA SA312458261GA SA312458263GA SA312458265GA SA312458266GA SA312458267GA SA312458394GA SA312458395GA SA312458396GA SA312458399GA SA312458400GA SA312458401GA SA312458402GA SA312458403GA SA312458405GA SA312458406GA SA312458407GA SA312458408GA SA312458409GA SA312458410GA SA312458411GA SA312458412GA SA312458413GA SA312458414GA SA312458415GA SA312458416GA SA312458417GA SA312458421GA SA312458422GA SA312458424GA SA312458426GA SA312458428GA SA312458429GA SA312458430GA SA312458431GA SA312458432GA SA312458434GA SA312458435GA SA312458456GA SA312458457GA SA312458458GA SA312458459GA SA312458461GA SA312458462GA SA312458463GA SA312458465GA SA312468532GA SA312468533GA SA312468534GA SA312468537GA SA312468538GA SA312468540GA SA312468541GA SA312468542GA SA312468545GA SA312468547GA SA312468550GA SA312468551GA SA312468552GA SA312468553GA SA312468559GA SA312468562GA SA312468563GA SA312468564GA SA312468565GA SA312468566GA SA312468567GA SA312468568GA SA312468570GA SA312468571GA SA312468572GA SA312468575GA SA312468577GA SA312468579GA SA312468581GA SA312468582GA SA312468583GA SA312468584GA SA312468585GA SA312468587GA SA312468588GA SA312468589GA SA312468591GA SA312468772GA SA312468774GA SA312468775GA SA312468776GA SA312468777GA SA312468778GA SA312468779GA SA312468780GA SA312468781GA SA312468784GA SA312468786GA SA312468787GA SA312468791GA SA312468792GA SA312468793GA SA312468794GA SA312468796GA SA312468798GA SA312468800GA SA312468801GA SA312468803GA SA312468805GA SA312468806GA SA312468807GA SA312468813GA SA312468814GA SA312468816GA SA312468817GA SA312468818GA SA312468819GA SA312468820GA SA312468824GA SA312468826GA SA312468831GA SA312468838GA SA312468841GA SA312468844GA SA312468849GA SA312468850GA SA312468851GA SA312468852GA SA312468853GA SA312468857GA SA312468858GA SA312468859GA SA312468861GA SA312468862GA SA312468864GA SA312468867GA SA312468902GA SA312468907GA SA312468924GA SA312468925GA SA312468944GA SA312468946GA SA312468953GA SA312468954GA SA312468959GA SA312468962GA SA312468963GA SA312468964GA SA312468967GA SA312468968GA SA312468969GA SA312468970GA SA312468971GA SA312468974GA SA312468975GA SA312468978GA SA312468979GA SA312468980GA SA312468981GA SA312468982GA SA312468984GA SA312468987GA SA312468988GA SA312468989GA SA312468990GA SA312468993GA SA312468995GA SA312468996GA SA312468999GA SA312469000GA SA312469002GA SA312469003GA SA312469004GA SA312469005GA SA312469006GA SA312469007GA SA312469009GA SA312469010GA SA312469011GA SA312469042GA SA312469046GA SA312469047GA SA312469049GA SA312469050GA SA312469051GA SA312469052GA SA312469060GA SA312469061GA SA312469062GA SA312469064GA SA312469066GA SA312469067GA SA312469068GA SA312469070GA SA312469071GA SA312469103GA SA312469104GA SA312469105GA SA312469106GA SA312469107GA SA312469108GA SA312469110GA SA312469111GA SA312469115GA SA312469116GA SA312469117GA SA312469118GA SA312469119GA SA312469120GA SA312469121GA SA312469123GA SA312469124GA SA312469125GA SA312469127GA SA312469128GA SA312469129GA SA312469131GA SA312469132GA SA312469135GA SA312469137GA SA312469139GA SA312469140GA SA312469141GA SA312469142GA SA312469143GA SA312469147GA SA312469149GA SA312469150GA SA312469151GA SA312469153GA SA312469154GA SA312469156GA SA312469157GA SA312469162GA SA312469163GA SA312469164GA SA312469166GA SA312469167GA SA312469169GA SA312469170GA SA312469171GA SA312469172GA SA312469174GA SA312469176GA SA312469177GA SA312469181GA SA312469182GA SA312469184GA SA312469186GA SA312469188GA SA312469189GA SA312469190GA SA312469191GA SA312469192GA SA312469193GA SA312469194GA SA312469197GA SA312469198GA SA312469199GA SA312469206GA SA312469207GA SA312469213GA SA312469215GA SA312469217GA SA312469218GA SA312469221GA SA312469282GA SA312469283GA SA312469284GA SA312469285GA SA312469286GA SA312469287GA SA312469289GA SA312469290GA SA312469291GA SA312469292GA SA312469293GA SA312469294GA SA312469296GA SA312469297GA SA312469298GA SA312469299GA SA312469300GA SA312469301GA SA312469302GA SA312469303GA SA312469307GA SA312469308GA SA312469309GA SA312469311GA SA312469342GA SA312469343GA SA312469344GA SA312469345GA SA312469346GA SA312469348GA SA312469349GA SA312469350GA SA312469351GA SA312469352GA SA312469353GA SA312469355GA SA312469356GA SA312469357GA SA312469358GA SA312469359GA SA312469360GA SA312469362GA SA312469363GA SA312469365GA SA312469366GA SA312469368GA SA312469370GA SA312469371GA SA312479553GA SA312479594GA SA312479685GA SA312479696GA SA312479742GA SA312479754GA SA312479788GA SA312479791GA SA312479793GA SA312479794GA SA312479795GA SA312479797GA SA312479801GA SA312479811GA SA312479817GA SA312479818GA SA312480118GA SA312480119GA SA312480120GA SA312480333GA SA312480334GA SA312480336GA SA312480341GA SA312480344GA SA312480345GA SA312480350GA SA312480354GA SA312480357GA SA312480358GA SA312480359GA SA312480361GA SA312480362GA SA312480379GA SA312480389GA SA312480410GA SA312480867GA SA312501561GA SA312501795GA SA312501990GA SA312502128GA SA312512177GA SA312512182GA SA312512191GA SA312512193GA SA312512258GA SA312512273GA SA312512275GA SA312512295GA SA312512296GA SA312512300GA SA312512307GA SA312512332GA SA312512352GA SA312522575GA SA313012727GA SA313012728GA SA313012730GA SA313012731GA SA313012734GA SA313012735GA SA313012737GA SA313012738GA SA313012740GA SA313012744GA SA313012747GA SA313012748GA SA313012749GA SA313012755GA SA313012756GA SA313012757GA SA313012758GA SA313012759GA SA313012761GA SA313012765GA SA313012768GA SA313012772GA SA313012773GA SA313012776GA SA313012781GA SA313012792GA SA313012793GA SA313012796GA SA313012798GA SA313012799GA SA313012800GA SA313012801GA SA313012802GA SA313012803GA SA313012805GA SA313012808GA SA313012809GA SA313012812GA SA313012813GA SA313012821GA SA313012823GA SA313012829GA SA313012830GA SA313012832GA SA313012839GA SA313012840GA SA313012848GA SA313012913GA SA313012932GA SA313012933GA SA313012935GA SA313012936GA SA313012937GA SA313012938GA SA313012939GA SA313012940GA SA313012941GA SA313012942GA SA313012944GA SA313012945GA SA313012946GA SA313012947GA SA313012948GA SA313012952GA SA313012953GA SA313012954GA SA313012957GA SA313012959GA SA313012961GA SA313012965GA SA313012971GA SA313012973GA SA313012976GA SA313012977GA SA313012981GA SA313012982GA SA313012984GA SA313012986GA SA313012987GA SA313012988GA SA313012989GA SA313012990GA SA313012991GA SA313023252GA SA313023257GA SA313023354GA SA313023355GA SA313023357GA SA313023378GA SA313023391GA SA313023403GA SA313023419GA SA313023425GA SA313023428GA SA313033528GA SA313033531GA SA313033587GA SA313033588GA SA313033589GA SA313033590GA SA313033593GA SA313033605GA SA313033610GA SA313033612GA SA313033621GA SA313033625GA SA313033626GA SA313033628GA SA313033630GA SA313033631GA SA313033648GA SA313033650GA SA313033652GA SA313033654GA SA313033715GA SA313033749GA SA313033750GA SA313033751GA SA313033753GA SA313033754GA SA313033756GA SA313033757GA SA313033765GA SA313033770GA SA313033771GA SA313033772GA SA313033773GA SA313033775GA SA313033776GA SA313033807GA SA313033809GA SA313033813GA SA313033814GA SA313033815GA SA313033816GA SA313033818GA SA313033822GA SA313033825GA SA313033826GA SA313033830GA SA313033831GA SA313033833GA SA313033834GA SA313033835GA SA313043868GA SA313043873GA SA313043874GA SA313043877GA SA313043880GA SA313043883GA SA313043884GA SA313043907GA SA313043909GA SA313043918GA SA313043919GA SA313043920GA SA313043922GA SA313043928GA SA313043933GA SA313043934GA SA313043940GA SA313043943GA SA313043945GA SA313043961GA SA313043965GA SA313043966GA SA313043968GA SA313043969GA SA313044020GA SA313044069GA SA313044074GA SA313044077GA SA313044078GA SA313044079GA SA313044080GA SA313044094GA SA313044096GA SA313044120GA SA313044121GA SA313044122GA SA313064177GA SA313064178GA SA313064179GA SA313064181GA SA313064182GA SA313064185GA SA313064199GA SA313064202GA SA313064231GA SA313064251GA SA313064425GA SA313084446GA SA313084448GA SA313084449GA SA313084451GA SA313084453GA SA313084454GA SA313084455GA SA313084456GA SA313084457GA SA313084458GA SA313084459GA SA313084460GA SA313084463GA SA313084465GA SA313084466GA SA313084467GA SA313084468GA SA313084469GA SA313084471GA SA313084472GA SA313084474GA SA313084475GA SA313084476GA SA313084479GA SA313084481GA SA313084483GA SA313084484GA SA313084485GA SA313084487GA SA313084491GA SA313084493GA SA313084500GA SA313084501GA SA313084503GA SA313084508GA SA313084509GA SA313084510GA SA313084511GA SA313084512GA SA313084513GA SA313084514GA SA313084516GA SA313084518GA SA313084519GA SA313084525GA SA313084527GA SA313084528GA SA313084530GA SA313084532GA SA313084534GA SA313084535GA SA313084654GA SA313084687GA SA313084688GA SA313084689GA SA313084691GA SA313084692GA SA313084694GA SA313084697GA SA313084699GA SA313084703GA SA313084704GA SA313084707GA SA313084708GA SA313084710GA SA313084712GA SA313084714GA SA313084745GA SA313084749GA SA313084753GA SA313084760GA SA313084765GA SA313084767GA SA313084771GA SA313084776GA SA313084777GA SA313084778GA SA313084787GA SA313084794GA SA313084807GA SA313084816GA SA313084817GA SA313084818GA SA313084819GA SA313084958GA SA313084959GA SA313084961GA SA313084973GA SA313084975GA SA313084976GA SA313084977GA SA313084978GA SA313084982GA SA313094992GA SA313094996GA SA313094997GA SA313095000GA SA313095037GA SA313095118GA SA313095171GA SA313095256GA SA313095260GA SA313095441GA SA313095493GA SA313095497GA SA313095526GA SA313095601GA SA313095639GA SA313105651GA SA313105656GA SA313105657GA SA313105661GA SA313105675GA SA313105684GA SA313105694GA SA313105939GA SA313106005GA SA313106026GA SA313116516GA SA313116526GA SA313116692GA SA313116700GA SA313116705GA SA313116723GA SA313116724GA SA313116725GA SA313127541GA SA313282471GA SA313282476GA SA313282479GA SA313282538GA SA313282558GA SA313282570GA SA313282583GA SA313282857GA SA313292901GA SA313292902GA SA313292951GA SA313292971GA SA313292974GA SA313292980GA SA313292988GA SA313292994GA SA313293000GA SA313293002GA SA313293010GA SA313293014GA SA313293022GA SA313293025GA SA313293028GA SA313293036GA SA313293038GA SA313293047GA SA313293054GA SA313293057GA SA313293060GA SA313293065GA SA313293067GA SA313293070GA SA313293077GA SA313293078GA SA313293079GA SA313293090GA SA313293097GA SA313293103GA SA313293110GA SA313293115GA SA313293118GA SA313293133GA SA313293147GA SA313293148GA SA313293149GA SA313293153GA SA313293158GA SA313293159GA SA313293163GA SA313293171GA SA313293173GA SA313293177GA SA313293184GA SA313293191GA SA313293192GA SA313293195GA SA313293200GA SA313293204GA SA313293212GA SA313293214GA SA313293217GA SA313293219GA SA313293221GA SA313293232GA SA313293233GA SA313293236GA SA313293237GA SA313293238GA SA313293242GA SA313293248GA SA313293251GA SA313293253GA SA313293254GA SA313293260GA SA313293266GA SA313293267GA SA313293271GA SA313293276GA SA313293288GA SA313293293GA SA313314436GA SA313314451GA SA313314455GA SA313314471GA SA313314475GA SA313314476GA SA313314477GA SA313314479GA SA313314483GA SA313314490GA SA313314498GA SA313314499GA SA313314502GA SA313314504GA SA313314508GA SA313314511GA SA313314516GA SA313314519GA SA313314524GA SA313314555GA SA313314556GA SA313324909GA SA313324934GA SA313335001GA SA313335016GA SA313335018GA SA313335026GA SA313335027GA SA313335028GA SA313335029GA SA313335034GA SA313335036GA SA313335037GA SA313335038GA SA313335041GA SA313335140GA SA313335148GA SA313335153GA SA313335156GA SA313335159GA SA313335160GA SA313335163GA SA313335195GA SA313335198GA SA313335205GA SA313335211GA SA313335229GA SA313335246GA SA313335252GA SA313335299GA SA313335300GA SA313335307GA SA313335327GA SA313335357GA SA313335360GA SA313335413GA SA313335419GA SA313335432GA SA313345531GA SA313345547GA SA313345553GA SA313345720GA SA313345755GA SA313345758GA SA313356314GA SA313356353GA SA313356361GA SA313356388GA SA313366786GA SA313367501GA SA313367558GA SA313367562GA SA313367570GA SA313367571GA SA313367574GA SA313367578GA SA313367581GA SA313367584GA SA313367585GA SA313378035GA SA313378149GA SA313378307GA SA313378369GA SA313378379GA SA313378383GA SA313378385GA SA313378405GA SA313378413GA SA313378438GA SA313378460GA SA313388561GA SA313388685GA SA313388698GA SA313409420GA SA313419435GA SA313419436GA SA313419437GA SA313419452GA SA313419459GA SA313419585GA SA313420032GA SA313420220GA SA313420225GA SA313420226GA SA313420227GA SA313420228GA SA313420235GA SA313420247GA SA313420250GA SA313420270GA SA313420271GA SA313420272GA SA313420273GA SA313420275GA SA313420276GA SA313420280GA SA313420281GA SA313420283GA SA313420285GA SA313420288GA SA313420292GA SA313420294GA SA313420296GA SA313420298GA SA313420299GA SA313420300GA SA313430617GA SA313430695GA SA313430698GA SA313430703GA SA313430710GA SA313430712GA SA313430714GA SA313430728GA SA313430729GA SA313430748GA SA313430769GA SA313430894GA SA313430895GA SA313430899GA SA313430904GA SA313430917GA SA313430939GA SA313430943GA SA313430944GA SA313430976GA SA313431077GA SA313431107GA SA313431182GA SA313431183GA SA313431187GA SA313431194GA SA313431204GA SA313431208GA SA313441340GA SA313441376GA SA313441497GA SA313441564GA SA313441567GA SA313441630GA SA313441631GA SA313441635GA SA313441637GA SA313441638GA SA313441642GA SA313441645GA SA313441646GA SA313441652GA SA313441657GA SA313441658GA SA313441689GA SA313441698GA SA313441705GA SA313451816GA SA313451831GA SA313451847GA SA313451848GA SA313451855GA SA313451877GA SA313451883GA SA313451887GA SA313451897GA SA313451898GA SA313451906GA SA313451907GA SA313451909GA SA313451916GA SA313451922GA SA313451925GA SA313451939GA SA313451943GA SA313451953GA SA313451960GA SA313451962GA SA313451964GA SA313451973GA SA313451982GA SA313452013GA SA313452019GA SA313452029GA SA313452035GA SA313452041GA SA313452042GA SA313452056GA SA313452134GA SA313452136GA SA313452149GA SA313462159GA SA313462173GA SA313462182GA SA313462197GA SA313462204GA SA313462213GA SA313462214GA SA313462238GA SA313462287GA SA313462291GA SA313462307GA SA313462320GA SA313462327GA SA313462337GA SA313462338GA SA313462444GA SA313462560GA SA313482608GA SA313482614GA SA313482615GA SA313482620GA SA313482621GA SA313482629GA SA313482630GA SA313482633GA SA313482635GA SA313482637GA SA313482651GA SA313482653GA SA313482654GA SA313482656GA SA313482658GA SA313482659GA SA313482661GA SA313482664GA SA313482669GA SA313482675GA SA313482679GA SA313482685GA SA313482698GA SA313482700GA SA313503607GA SA313513829GA SA313513837GA SA313513846GA SA313513849GA SA313513851GA SA313513866GA SA313513870GA SA313513871GA SA313513873GA SA313513878GA SA313513888GA SA313513892GA SA313513893GA SA313513916GA SA313513923GA SA313513926GA SA313513940GA SA313513945GA SA313523954GA SA314014360GA SA314014468GA SA314014482GA SA314014494GA SA314014502GA SA314014513GA SA314014517GA SA314014520GA SA314014522GA SA314014523GA SA314014524GA SA314014527GA SA314014540GA SA314014546GA SA314014550GA SA314014551GA SA314014554GA SA314014561GA SA314014567GA SA314014572GA SA314014575GA SA314014578GA SA314014579GA SA314014583GA SA314014584GA SA314014587GA SA314014595GA SA314014598GA SA314014599GA SA314014600GA SA314014601GA SA314014622GA SA314014649GA SA314014650GA SA314014651GA SA314014652GA SA314014654GA SA314014658GA SA314014736GA SA314024956GA SA314025011GA SA314025023GA SA314025029GA SA314025030GA SA314025054GA SA314025064GA SA314025065GA SA314025102GA SA314025129GA SA314025174GA SA314025176GA SA314025180GA SA314025182GA SA314025184GA SA314025185GA SA314025188GA SA314025189GA SA314025191GA SA314025192GA SA314025193GA SA314025194GA SA314025195GA SA314025198GA SA314025199GA SA314025200GA SA314025203GA SA314025207GA SA314025228GA SA314035234GA SA314035236GA SA314035237GA SA314035240GA SA314035241GA SA314035243GA SA314035244GA SA314035245GA SA314035247GA SA314035248GA SA314035251GA SA314035252GA SA314035253GA SA314035255GA SA314035256GA SA314035258GA SA314035259GA SA314035260GA SA314035262GA SA314035264GA SA314035270GA SA314035271GA SA314035272GA SA314035273GA SA314035275GA SA314035278GA SA314035280GA SA314035287GA SA314035290GA SA314035298GA SA314035299GA SA314035300GA SA314035302GA SA314035310GA SA314035314GA SA314035316GA SA314035320GA SA314035330GA SA314035332GA SA314035335GA SA314035338GA SA314035354GA SA314035406GA SA314035407GA SA314045483GA SA314045502GA SA314045521GA SA314055539GA SA314055548GA SA314055554GA SA314055562GA SA314055565GA SA314055598GA SA314055603GA SA314055607GA SA314055619GA SA314055626GA SA314055647GA SA314055649GA SA314055665GA SA314055666GA SA314055667GA SA314055672GA SA314055673GA SA314055674GA SA314055679GA SA314055680GA SA314055684GA SA314055701GA SA314055702GA SA314055706GA SA314055710GA SA314065736GA SA314065739GA SA314065741GA SA314065743GA SA314065773GA SA314065776GA SA314065779GA SA314065783GA SA314065787GA SA314065792GA SA314075800GA SA314075856GA SA314075949GA SA314075972GA SA314075984GA SA314076117GA SA314086645GA SA314086682GA SA314086684GA SA314086686GA SA314086695GA SA314086713GA SA314096878GA SA314096891GA SA314096894GA SA314096897GA SA314096914GA SA314096953GA SA314096956GA SA314096965GA SA314096968GA SA314097017GA SA314097072GA SA314097084GA SA314097088GA SA314097101GA SA314097102GA SA314097241GA SA314097245GA SA314097291GA SA314097353GA SA314107407GA SA314107416GA SA314107426GA SA314107427GA SA314107438GA SA314107446GA SA314107449GA SA314107452GA SA314107454GA SA314107456GA SA314107457GA SA314107458GA SA314107462GA SA314107465GA SA314107466GA SA314107469GA SA314107534GA SA314107570GA SA314107611GA SA314107660GA SA314107693GA SA314107779GA SA314107780GA SA314107899GA SA314107974GA SA314118460GA SA314118516GA SA314118524GA SA314118628GA SA314118641GA SA314118660GA SA314118733GA SA314118734GA SA314118847GA SA314118872GA SA314118877GA SA314118886GA SA314119101GA SA314129184GA SA314129203GA SA314129226GA SA314129234GA SA314129235GA SA314129242GA SA314129254GA SA314129255GA SA314129302GA SA314139447GA SA314139510GA SA314139540GA SA314149610GA SA314149611GA SA314149617GA SA314149626GA SA314149630GA SA314149634GA SA314149649GA SA314149655GA SA314149660GA SA314169823GA SA314169870GA SA314169894GA SA314170280GA SA314170281GA SA314170467GA SA314180541GA SA314190902GA SA314190923GA SA314191123GA SA314191127GA SA314191142GA SA314191160GA SA314191173GA SA314201383GA SA314201384GA SA314201409GA SA314201411GA SA314201416GA SA314201430GA SA314201438GA SA314201440GA SA314201471GA SA314201473GA SA314201486GA SA314201488GA SA314201492GA SA314201494GA SA314201495GA SA314201607GA SA314201716GA SA314201764GA SA314211926GA SA314211943GA SA314211947GA SA314212008GA SA314212024GA SA314212034GA SA314212039GA SA314212049GA SA314212057GA SA314212096GA SA314212105GA SA314212163GA SA314212175GA SA314212187GA SA314212199GA SA314212213GA SA314212217GA SA314212220GA SA314212225GA SA314212229GA SA314212233GA SA314212234GA SA314212235GA SA314212248GA SA314212255GA SA314222277GA SA314222279GA SA314222280GA SA314222284GA SA314222285GA SA314222288GA SA314222290GA SA314222291GA SA314222294GA SA314222299GA SA314222307GA SA314222308GA SA314222310GA SA314222314GA SA314222318GA SA314222319GA SA314222321GA SA314222327GA SA314222329GA SA314222330GA SA314222331GA SA314222334GA SA314222336GA SA314222340GA SA314222342GA SA314222343GA SA314222356GA SA314222360GA SA314222366GA SA314222370GA SA314222379GA SA314222380GA SA314222384GA SA314222387GA SA314222388GA SA314222390GA SA314222402GA SA314222403GA SA314222408GA SA314222409GA SA314222412GA SA314222413GA SA314222414GA SA314222415GA SA314222420GA SA314222421GA SA314222426GA SA314222427GA SA314222431GA SA314222453GA SA314222457GA SA314222470GA SA314222474GA SA314222491GA SA314222494GA SA314222589GA SA314222595GA SA314222605GA SA314232979GA SA314233060GA SA314243215GA SA314243225GA SA314243230GA SA314243234GA SA314243248GA SA314243249GA SA314243251GA SA314243252GA SA314253542GA SA314253573GA SA314253701GA SA314253735GA SA314263785GA SA314263786GA SA314263791GA SA314263796GA SA314263798GA SA314263802GA SA314263831GA SA314274147GA SA314274148GA SA314274180GA SA314274188GA SA314274192GA SA314274224GA SA314274227GA SA314274302GA SA314274308GA SA314274322GA SA314284588GA SA314284683GA SA314284694GA SA314284729GA SA314284731GA SA314284736GA SA314284738GA SA314284745GA SA314284746GA SA314294852GA SA314294854GA SA314294857GA SA314294864GA SA314294865GA SA314294869GA SA314294871GA SA314294874GA SA314294876GA SA314294879GA SA314294882GA SA314294884GA SA314294885GA SA314294887GA SA314294889GA SA314295134GA SA311428032GA SA312039088GA SA312039132GA SA312294530GA SA312294615GA SA312294636GA SA312294651GA SA312294659GA SA312294664GA SA312386505GA SA312386509GA SA312386526GA SA312407287GA SA312407299GA SA312458200GA SA312458211GA SA312458214GA SA312458253GA SA312458259GA SA312458260GA SA312458262GA SA312458418GA SA312458419GA SA312458423GA SA312479667GA SA312501673GA SA313064171GA SA313064235GA SA313064240GA SA313064241GA SA313064282GA SA313064290GA SA313064296GA SA313064338GA SA313064340GA SA313064342GA SA313064349GA SA313064350GA SA313064373GA SA313064382GA SA313064387GA SA313084702GA SA313084768GA SA313084834GA SA313084860GA SA313084981GA SA313095058GA SA313095095GA SA313095556GA SA313095577GA SA313095578GA SA313095627GA SA313095631GA SA313105655GA SA313105663GA SA313105664GA SA313105665GA SA313105671GA SA313105676GA SA313105677GA SA313105679GA SA313105680GA SA313105685GA SA313105686GA SA313127707GA SA313137746GA SG512300132GA SG512300133GA SA307392995GA SA312049421GA SA312111547GA SA312111550GA SA312111553GA SA312111555GA SA312111574GA SA312111576GA SA312111579GA SA312111583GA SA312253284GA SA312264032GA SA313033566GA SA313033600GA SA313033603GA SA313033604GA SA313033606GA SA313441568GA SA313452045GA SA313452102GA SA313452114GA SA313462199GA SA313462231GA SA313462240GA SA313462242GA SA313462246GA SA313462248GA SA313462250GA SA313462253GA SA313462267GA SA313462269GA SA313462323GA SA313462326GA SA313462332GA SA307231242GA SA308357888GA SA308358030GA SA308368060GA SA308368211GA SA308368231GA SA308368232GA SA308378280GA SA308378281GA SA308378306GA SA308388481GA SA308388550GA SA308388554GA SA308388558GA SA308388563GA SA308388564GA SA308388566GA SA308388571GA SA308388572GA SA308388573GA SA308388574GA SA308388575GA SA308388581GA SA308388589GA SA308388591GA SA308388592GA SA308388594GA SA308388597GA SA308388598GA SA308388599GA SA308388601GA SA308388617GA SA308388621GA SA308388623GA SA308388624GA SA308388630GA SA308388631GA SA308388633GA SA308388634GA SA308388636GA SA308388637GA SA308388638GA SA308398664GA SA308398666GA SA308398667GA SA308398668GA SA308398669GA SA308398671GA SA308398672GA SA308398673GA SA308398674GA SA308398675GA SA308398678GA SA308398679GA SA308398680GA SA308398682GA SA308398684GA SA308398685GA SA308398686GA SA308398688GA SA308398690GA SA308398691GA SA308398692GA SA308398706GA SA308398709GA SA308398710GA SA308398786GA SA308398787GA SA308398788GA SA308398789GA SA308398790GA SA308398791GA SA308398792GA SA308398793GA SA308398794GA SA308398796GA SA308398797GA SA308398798GA SA308398799GA SA308398800GA SA308398801GA SA308398804GA SA308398805GA SA308398806GA SA308398807GA SA308398808GA SA308398809GA SA308398811GA SA308398816GA SA308398819GA SA308398820GA SA308398823GA SA308398829GA SA308398831GA SA308398832GA SA308398833GA SA308398834GA SA308398835GA SA308398836GA SA308398837GA SA308398838GA SA308398839GA SA308398840GA SA308398842GA SA308398843GA SA308398844GA SA308398845GA SA308398846GA SA308398849GA SA308398850GA SA308398851GA SA308398852GA SA308398859GA SA308398904GA SA308408918GA SA308408919GA SA308408920GA SA308408922GA SA308408926GA SA308408928GA SA308408933GA SA308408936GA SA308408939GA SA308408940GA SA308408942GA SA308408943GA SA308408944GA SA308408947GA SA308408950GA SA308408951GA SA308408953GA SA308408954GA SA308408955GA SA308408958GA SA308408959GA SA308408961GA SA308408962GA SA308408963GA SA308409009GA SA308409011GA SA308409014GA SA308409018GA SA308409021GA SA308409029GA SA308409051GA SA308409052GA SA308409062GA SA308409066GA SA308409067GA SA308409170GA SA308409174GA SA308409177GA SA308409188GA SA308409200GA SA308409212GA SA308419218GA SA308419229GA SA308419231GA SA308419232GA SA308419233GA SA308419236GA SA308419239GA SA308419240GA SA308419242GA SA308419243GA SA308419257GA SA308419258GA SA308419259GA SA308419260GA SA308419262GA SA308419264GA SA308419266GA SA308419268GA SA308429327GA SA308429331GA SA308429342GA SA308429349GA SA308429350GA SA308429351GA SA308429353GA SA308429354GA SA308429363GA SA308501400GA SA308501403GA SA308501418GA SA308511434GA SA308511443GA SA308511445GA SA308511463GA SA308511482GA SA308511494GA SA308511497GA SA308511498GA SA308511507GA SA308511516GA SA308511528GA SA308511543GA SA308511544GA SA308511546GA SA308511549GA SA308511553GA SA308511632GA SA308511635GA SA308511642GA SA308511688GA SA308511691GA SA308511692GA SA308511694GA SA308511700GA SA308511701GA SA308511706GA SA308511707GA SA308511712GA SA308511714GA SA308511716GA SA308511718GA SA308511720GA SA308511721GA SA308511722GA SA308511726GA SA308511729GA SA308511730GA SA308511731GA SA308511732GA SA308511733GA SA308511734GA SA308511735GA SA308511736GA SA308511737GA SA308511738GA SA308511739GA SA308511740GA SA308511741GA SA308511742GA SA308511743GA SA308511744GA SA308511745GA SA308511747GA SA308511748GA SA308511749GA SA308511768GA SA308511772GA SA309011867GA SA309011882GA SA309031996GA SA309042157GA SA309042162GA SA309042163GA SA309042164GA SA309042165GA SA309042168GA SA309042171GA SA309042172GA SA309042174GA SA309132801GA SA309132805GA SA309132812GA SA309132814GA SA309132817GA SA309132822GA SA309132823GA SA309173113GA SA309173115GA SA309173120GA SA309173164GA SA309173173GA SA309173176GA SA309173177GA SA309173179GA SA309173181GA SA309173182GA SA309173188GA SA309173189GA SA309243642GA SA309243643GA SA309243645GA SA309243646GA SA309243650GA SA309243651GA SA309243652GA SA309243653GA SA309243654GA SA309243655GA SA309253865GA SA309314291GA SA309324300GA SA309324304GA SA309324332GA SA309354923GA SA309354932GA SA309354934GA SA309354954GA SA309354960GA SA309354973GA SA309354984GA SA309354992GA SA309354996GA SA309354997GA SA309354999GA SA309355001GA SA309355005GA SA309355011GA SA309365066GA SA309365081GA SA309395696GA SA309395697GA SA309395699GA SA309395701GA SA309395711GA SA309395733GA SA309395744GA SA309395751GA SA309395791GA SA309395801GA SA309395808GA SA309405826GA SA309405840GA SA309446525GA SA309446529GA SA309446530GA SA309456548GA SA309466776GA SA310199699GA SA310361689GA SA310361903GA SA310361913GA SA310372009GA SA310372028GA SA310372032GA SA310372034GA SA310506709GA SA310507087GA SA310517569GA SA310517733GA SA310517781GA SA310517819GA SA310517820GA SA310517822GA SA310517829GA SA310517834GA SA310517835GA SA310517843GA SA310517879GA SA310517984GA SA310517989GA SA310517992GA SA310517999GA SA310518006GA SA310518008GA SA310518011GA SA310528877GA SA313462203GA SA313462209GA SA313013129GA SA312242984GA SA312491195GA SA313033768GA SA313043979GA SA313420572GA SA313064247GA SA314097240GA SA314097249GA SA314097254GA SA314097258GA SA314097266GA SA314097295GA SA314097296GA SA314097297GA SA314097301GA SA314097304GA SA314097305GA SA314097308GA SA314097334GA SA314097335GA SA314097336GA SA314097339GA SA307312104GA SA307453539GA SA310038454GA SA312049445GA SA312059624GA SA312059628GA SA312069729GA SA312069773GA SA312253222GA SA312253230GA SA312386312GA SA312491280GA SA313013101GA SA307322203GA SA307514339GA SA307514345GA SA308094931GA SA308094939GA SA308094950GA SA308094951GA SA308449567GA SA308459856GA SA308479901GA SA308489919GA SA308489924GA SA308489932GA SA308489933GA SA308489936GA SA308489941GA SA308489942GA SA308489958GA SA308489975GA SA308489980GA SA308489981GA SA308489985GA SA309072395GA SA309072396GA SA309173108GA SA309173109GA SA309173112GA SA309173114GA SA309385522GA SA309385525GA SA309385664GA SA310423792GA SA310475477GA SA310475483GA SA310475490GA SA310475526GA SA310475528GA SA310475596GA SA310496395GA SA310496407GA SA310496409GA SA310496420GA SA310496424GA SA310496429GA SA310496444GA SA310496445GA SA310496449GA SA310496451GA SA310496484GA SA310506704GA SA310506719GA SA310528921GA SA311132617GA SA311132631GA SA311163711GA SA311163716GA SA311174199GA SA312355380GA SA312355385GA SA312355386GA SA312355395GA SA312355436GA SA312355441GA SA312355443GA SA312355444GA SA312355445GA SA312355450GA SA312469077GA SA312479561GA SA312479616GA SA312480134GA SA312480248GA SA312480297GA SA312480801GA SA312491029GA SA312491379GA SA312501702GA SA312501704GA SA312501705GA SA312501736GA SA312501749GA SA312501773GA SA312501852GA SA312501887GA SA312501888GA SA312502152GA SA312522386GA SA312522408GA SA312522430GA SA312522471GA SA312522499GA SA312522502GA SA312522506GA SA312522508GA SA312522521GA SA312522525GA SA312522527GA SA312522528GA SA312522531GA SA312522532GA SA312522533GA SA312522551GA SA312522557GA SA312522558GA SA312522559GA SA312522594GA SA312522597GA SA312522646GA SA312522647GA SA311132299GA SA311428072GA SA312294373GA SA312294415GA SA312294468GA SA312294469GA SA312294516GA SA312294526GA SA312294531GA SA312294558GA SA312365776GA SA312365780GA SA312365791GA SA312365801GA SA312365807GA SA308266823GA SA308266828GA SA308266889GA SA308266891GA SA308286904GA SA308286908GA SA308286917GA SA308286926GA SA308286927GA SA308286934GA SA308286946GA SA308286947GA SA308286952GA SA308286956GA SA308286957GA SA308286991GA SA308286992GA SA308286997GA SA308286998GA SA308286999GA SA308287002GA SA308287014GA SA308287015GA SA308327376GA SA309263970GA SA310118994GA SA310139214GA SA310139217GA SA310139220GA SA310139223GA SA310139224GA SA310139228GA SA310139233GA SA310139235GA SA310139289GA SA310139307GA SA310149343GA SA310423750GA SA311069707GA SA311132540GA SA311142711GA SA311195199GA SA314180639GA SA314180678GA SA314180730GA SA314180734GA SA314180739GA SA314180765GA SA314180767GA SA314180776GA SA314180784GA SA314191195GA SA312049367GA SA312049375GA SA312049376GA SA312049379GA SA312049381GA SA312049384GA SA312049413GA SA312049420GA SA312490969GA SA312490971GA SA312491012GA SA312491048GA SA309436321GA SA309436329GA SA309436363GA SA309446428GA SA309446438GA SA309446443GA SA309446485GA SA309446522GA SA309466767GA SA312018864GA SA313282543GA SA313282848GA SA313282851GA SA313282865GA SA313282867GA SA313292897GA SA313292899GA SA313292953GA SA313292965GA SA313292966GA SA313292969GA SA313314747GA SA313314755GA SA313314764GA SA313324901GA SA313324978GA SA313335008GA SA313420211GA SA313420231GA SA313420238GA SA313420253GA SA313420255GA SA313420259GA SA313420265GA SA313420266GA SA313420334GA SA313420339GA SA313420342GA SA313420364GA SA313420365GA SA313420370GA SA313420394GA SA313420405GA SA313420545GA SA313420551GA SA313420602GA SA313420603GA SA313430618GA SA313430629GA SA313430630GA SA313430631GA SA313430632GA SA313430633GA SA313430634GA SA313430635GA SA313430638GA SA313430639GA SA313430640GA SA313430641GA SA313430642GA SA313430643GA SA313430652GA SA313430653GA SA313430655GA SA313430656GA SA313430658GA SA313430660GA SA313430668GA SA313430670GA SA313430674GA SA313430675GA SA313430681GA SA313430682GA SA313430689GA SA313431086GA SA313431134GA SA313431137GA SA313431138GA SA313431140GA SA313431141GA SA313431142GA SA313431145GA SA313431146GA SA313431149GA SA313431150GA SA313431156GA SA313431159GA SA313431166GA SA313431198GA SA313431207GA SA313431211GA SA313431212GA SA313431213GA SA313431219GA SA313431225GA SA313431227GA SA313431242GA SA313441275GA SA313441283GA SA313441316GA SA313441513GA SA313451797GA SA313493035GA SA314024840GA SA314024843GA SA314025157GA SA314025178GA SA314025190GA SA314025197GA SA314025208GA SA314025220GA SA314035348GA SA314035353GA SA314035358GA SA314045491GA SA314076111GA SA314107493GA SA314108267GA SA314118945GA SA314118963GA SA314118980GA SA314119003GA SA314119022GA SA314119026GA SA314119030GA SA314119038GA SA314129306GA SA314129308GA SA314201529GA SA314211958GA SA314211985GA SA314212035GA SA308449588GA SA308449594GA SA308459640GA SA308459647GA SA308459671GA SA308459677GA SA308459684GA SA308459859GA SA308469887GA SA308469889GA SA308479908GA SA308481000GA SA308481004GA SA308481019GA SA308481025GA SA308481045GA SA308481047GA SA308481049GA SA308481051GA SA308489920GA SA308489925GA SA308489926GA SA308489929GA SA308489930GA SA308489931GA SA308489939GA SA308489940GA SA308489944GA SA308489945GA SA308489952GA SA308489953GA SA308489955GA SA308489956GA SA308489966GA SA308489968GA SA308489970GA SA308489972GA SA308489977GA SA308489978GA SA308489984GA SA308489986GA SA308489987GA SA308489988GA SA308489989GA SA308489992GA SA308489996GA SA308489999GA SA310270583GA SA310290741GA SA310290743GA SA310290744GA SA310444286GA SA310506715GA SA310507112GA SA310507403GA SA310507453GA SA310507457GA SA310507520GA SA310507544GA SA311090964GA SA311101088GA SA311101090GA SA311101135GA SA311101139GA SA311101157GA SA311101457GA SA311101466GA SA311111620GA SA311111634GA SA311111635GA SA311111639GA SA311111643GA SA311111644GA SA311111645GA SA311111646GA SA311111647GA SA311111648GA SA311111651GA SA311111652GA SA311111653GA SA311111656GA SA311111658GA SA311111660GA SA311111664GA SA311121980GA SA311122068GA SA311122070GA SA311122075GA SA311122078GA SA311122081GA SA311122083GA SA311122090GA SA311122091GA SA311122092GA SA311122097GA SA311122110GA SA311122121GA SA311122122GA SA311122137GA SA311122139GA SA311122144GA SA311122146GA SA311122151GA SA311122152GA SA311122157GA SA311122158GA SA311122162GA SA311122165GA SA311132428GA SA311132493GA SA311132538GA SA311132542GA SA311132544GA SA311132601GA SA311132602GA SA311132608GA SA311132609GA SA311132611GA SA311132621GA SA311132622GA SA311132624GA SA311132625GA SA311132626GA SA311132627GA SA311132628GA SA311132629GA SA311132632GA SA311132633GA SA311132637GA SA311132638GA SA311132640GA SA311132642GA SA311132643GA SA311132644GA SA311132645GA SA311132647GA SA311132654GA SA311132656GA SA311132657GA SA311132661GA SA311132664GA SA311132665GA SA311132670GA SA311132671GA SA311132672GA SA311132673GA SA311142687GA SA311142699GA SA311142843GA SA311163377GA SA311163396GA SA311163402GA SA311163861GA SA311163864GA SA311163866GA SA311163875GA SA311163879GA SA311163895GA SA311163910GA SA311184866GA SA311195063GA SA311195065GA SA311195114GA SA311205293GA SA311225953GA SA311236006GA SA311236007GA SA311236008GA SA311236014GA SA311236020GA SA311236021GA SA311236024GA SA311236042GA SA311236046GA SA311236052GA SA311236056GA SA311236165GA SA311236188GA SA311236189GA SA311236200GA SA311236202GA SA311236204GA SA311236205GA SA311236208GA SA311236213GA SA311236214GA SA311236220GA SA311236224GA SA311236226GA SA311236230GA SA311236232GA SA311236253GA SA311236255GA SA311236258GA SA311236280GA SA311236364GA SA311246449GA SA311246476GA SA311246478GA SA311246495GA SA311246498GA SA311246501GA SA311256802GA SA311256806GA SA311256813GA SA311256849GA SA311256854GA SA311256861GA SA311256864GA SA311256867GA SA311266913GA SA311266916GA SA311266917GA SA311266918GA SA311266920GA SA311266921GA SA311266924GA SA311266926GA SA311266927GA SA311266929GA SA311266930GA SA311266931GA SA311266932GA SA311266933GA SA311266934GA SA311266936GA SA311266937GA SA311266939GA SA311266940GA SA311266943GA SA311266944GA SA311266945GA SA311266948GA SA311266949GA SA311266950GA SA311307576GA SA311327915GA SA311468430GA SA311468435GA SA311468447GA SA311488453GA SA312079885GA SA312079886GA SA312079913GA SA312079931GA SA312079947GA SA312079976GA SA312079981GA SA312079984GA SA312101429GA SA312111500GA SA312111505GA SA312192588GA SA312192597GA SA312192600GA SA312192723GA SA312253197GA SA312264039GA SA312470071GA SA312470082GA SA312470083GA SA312470089GA SA312480103GA SA312480104GA SA313335182GA SA313335219GA SA313335407GA SA307080485GA SA307080537GA SA307210823GA SA307221112GA SA307221150GA SA307231195GA SA307231213GA SA307231361GA SA307231407GA SA307241555GA SA307241596GA SA307251793GA SA307291866GA SA307291868GA SA307301997GA SA307393008GA SA307393015GA SA307393038GA SA307423125GA SA307443189GA SA307443219GA SA307443308GA SA307443354GA SA307453443GA SA307453469GA SA307453472GA SA307453520GA SA307453530GA SA307453560GA SA307483802GA SA307493886GA SA307493889GA SA307493968GA SA307493978GA SA307493979GA SA307493986GA SA307494036GA SA307494046GA SA307504052GA SA307504053GA SA307504054GA SA307504055GA SA307504056GA SA307504057GA SA307504058GA SA307504059GA SA307504060GA SA307504061GA SA307504062GA SA307504063GA SA307504064GA SA307504065GA SA307504101GA SA307504103GA SA307504113GA SA307504116GA SA307504121GA SA307504187GA SA307504222GA SA307504225GA SA307504226GA SA307504227GA SA307504228GA SA307504229GA SA307504230GA SA307504231GA SA307504234GA SA307514236GA SA307514237GA SA307514241GA SA307514242GA SA307514243GA SA307514244GA SA307514246GA SA307514249GA SA307514250GA SA307514251GA SA307514252GA SA307514253GA SA307514254GA SA307514255GA SA307514257GA SA307514258GA SA307514259GA SA307514260GA SA307514261GA SA307514262GA SA307514263GA SA307514264GA SA307514265GA SA307514266GA SA307514267GA SA307514268GA SA307514269GA SA307514270GA SA307514271GA SA307514272GA SA307514273GA SA307514274GA SA307514275GA SA307514278GA SA307514279GA SA307514280GA SA307514281GA SA307514282GA SA307514284GA SA307514285GA SA307514286GA SA307514287GA SA307514288GA SA307514292GA SA307514293GA SA307514294GA SA307514295GA SA307514296GA SA307514297GA SA307514300GA SA307514301GA SA307514305GA SA307514306GA SA307514307GA SA307514309GA SA307514310GA SA307514317GA SA307514362GA SA307514409GA SA307514413GA SA307524460GA SA307524461GA SA307524462GA SA307524464GA SA307524465GA SA307524466GA SA307524467GA SA307524469GA SA307524471GA SA307524472GA SA307524476GA SA307524477GA SA307524479GA SA308044484GA SA308044486GA SA308044487GA SA308044488GA SA308044489GA SA308044490GA SA308044491GA SA308044492GA SA308044494GA SA308044495GA SA308044497GA SA308044498GA SA308044499GA SA308044500GA SA308044502GA SA308044503GA SA308044504GA SA308044505GA SA308044506GA SA308044507GA SA308044509GA SA308044510GA SA308044512GA SA308044515GA SA308044516GA SA308044518GA SA308044519GA SA308044524GA SA308044526GA SA308044529GA SA308044530GA SA308044531GA SA308044532GA SA308044533GA SA308044534GA SA308074581GA SA308074600GA SA308074601GA SA308074607GA SA308074608GA SA308084614GA SA308084656GA SA308084669GA SA308084672GA SA308084698GA SA308084709GA SA308084730GA SA308084745GA SA308084748GA SA308084757GA SA308084765GA SA308084771GA SA308084785GA SA308094798GA SA308094799GA SA308094802GA SA308094812GA SA308094817GA SA308094821GA SA308094822GA SA308094823GA SA308094824GA SA308094825GA SA308094826GA SA308094827GA SA308094828GA SA308094829GA SA308094830GA SA308094831GA SA308094833GA SA308094835GA SA308094866GA SA308094881GA SA308094923GA SA308094924GA SA308094926GA SA308094928GA SA308094932GA SA308094936GA SA308094945GA SA308094956GA SA308094961GA SA308094962GA SA308094963GA SA308094969GA SA308094971GA SA308094972GA SA308094974GA SA308094975GA SA308094976GA SA308094977GA SA308094982GA SA308094983GA SA308094985GA SA308094986GA SA308094987GA SA308094988GA SA308094990GA SA308094992GA SA308094993GA SA308094994GA SA308094996GA SA308094997GA SA308095001GA SA308095003GA SA308095005GA SA308095006GA SA308095007GA SA308095008GA SA308095009GA SA308095010GA SA308095011GA SA308095012GA SA308095045GA SA308095046GA SA308095049GA SA308095050GA SA308095052GA SA308095053GA SA308105151GA SA308105152GA SA308105153GA SA308105154GA SA308105155GA SA308105157GA SA308105158GA SA308105159GA SA308105160GA SA308105161GA SA308105162GA SA308105163GA SA308105164GA SA308105166GA SA308105168GA SA308105169GA SA308105170GA SA308105171GA SA308105172GA SA308105174GA SA308105175GA SA308105177GA SA308105179GA SA308105180GA SA308105181GA SA308105186GA SA308105187GA SA308105188GA SA308105189GA SA308105190GA SA308105194GA SA308105195GA SA308105197GA SA308105199GA SA308105201GA SA308105202GA SA308105204GA SA308105205GA SA308105207GA SA308105209GA SA308105210GA SA308105211GA SA308105212GA SA308105213GA SA308105216GA SA308105218GA SA308105219GA SA308105221GA SA308105222GA SA308105225GA SA308105226GA SA308105227GA SA308105228GA SA308105229GA SA308105230GA SA308115231GA SA308115232GA SA308115233GA SA308115236GA SA308115238GA SA308115239GA SA308115240GA SA308115241GA SA308115243GA SA308115244GA SA308115245GA SA308115246GA SA308115249GA SA308115251GA SA308115253GA SA308115254GA SA308115259GA SA308115260GA SA308125316GA SA308125317GA SA308125318GA SA308125319GA SA308125383GA SA308145751GA SA308155844GA SA308155855GA SA308155856GA SA308155861GA SA308155862GA SA308155870GA SA308155889GA SA308155890GA SA308155891GA SA308155894GA SA308155895GA SA308155896GA SA308155902GA SA308155907GA SA308155910GA SA308155911GA SA308155912GA SA308155913GA SA308155922GA SA308155926GA SA308155930GA SA308155934GA SA308155952GA SA308155976GA SA308165992GA SA308166038GA SA308166065GA SA308166087GA SA308166089GA SA308166093GA SA308166099GA SA308166104GA SA308166105GA SA308166106GA SA308166117GA SA308186339GA SA308196476GA SA308196524GA SA308206648GA SA308206665GA SA308266782GA SA308266783GA SA308266785GA SA308266786GA SA308266787GA SA308266790GA SA308266791GA SA308266792GA SA308266793GA SA308266796GA SA308266797GA SA308266799GA SA308266800GA SA308266801GA SA308266802GA SA308266803GA SA308266804GA SA308266805GA SA308266806GA SA308266807GA SA308266808GA SA308266809GA SA308266810GA SA308266812GA SA308266883GA SA308337686GA SA308337704GA SA308347754GA SA308347756GA SA308347767GA SA308347789GA SA308347791GA SA308347800GA SA308347829GA SA308347844GA SA308347845GA SA308347846GA SA308347847GA SA308347848GA SA308347850GA SA308347851GA SA308357943GA SA308357944GA SA308357945GA SA308357946GA SA308368070GA SA308378286GA SA308378297GA SA308388627GA SA308398712GA SA308409085GA SA308409161GA SA308409173GA SA308409189GA SA308419250GA SA308419267GA SA308419272GA SA308419274GA SA308429347GA SA308459849GA SA308459857GA SA308459874GA SA308491152GA SA308491153GA SA308491154GA SA308491155GA SA308491157GA SA308491158GA SA308491160GA SA308491162GA SA308491163GA SA308511770GA SA308521816GA SA309021975GA SA309021987GA SA309032009GA SA309032012GA SA309032013GA SA309032017GA SA309032021GA SA309032053GA SA309032054GA SA309032055GA SA309032056GA SA309032057GA SA309032058GA SA309032059GA SA309032060GA SA309032061GA SA309032062GA SA309032063GA SA309032065GA SA309032067GA SA309032068GA SA309032069GA SA309042132GA SA309052223GA SA309112570GA SA309122693GA SA309132803GA SA309132854GA SA309132865GA SA309142877GA SA309142895GA SA309142901GA SA309142906GA SA309142907GA SA309142989GA SA309153034GA SA309153038GA SA309153046GA SA309153048GA SA309153053GA SA309193332GA SA309193333GA SA309193338GA SA309193342GA SA309193345GA SA309193347GA SA309193350GA SA309193355GA SA309193359GA SA309193361GA SA309203367GA SA309203374GA SA309203377GA SA309203381GA SA309203389GA SA309203397GA SA309203398GA SA309203399GA SA309203411GA SA309203413GA SA309203414GA SA309203415GA SA309203419GA SA309203422GA SA309203426GA SA309223436GA SA309223437GA SA309223438GA SA309223440GA SA309223441GA SA309223446GA SA309223447GA SA309223448GA SA309223451GA SA309223454GA SA309223460GA SA309223461GA SA309223463GA SA309223464GA SA309223467GA SA309223473GA SA309223474GA SA309223475GA SA309223483GA SA309223487GA SA309223488GA SA309223491GA SA309223493GA SA309223494GA SA309223495GA SA309223496GA SA309223497GA SA309223498GA SA309223499GA SA309223500GA SA309223501GA SA309223504GA SA309223506GA SA309223510GA SA309223511GA SA309223512GA SA309223513GA SA309233521GA SA309233535GA SA309233537GA SA309233538GA SA309233540GA SA309233541GA SA309233542GA SA309233543GA SA309233545GA SA309233548GA SA309233549GA SA309233551GA SA309243699GA SA309243702GA SA309243705GA SA309243707GA SA309243711GA SA309243712GA SA309253721GA SA309253727GA SA309253745GA SA309253746GA SA309253754GA SA309365080GA SA309365109GA SA309365115GA SA309365117GA SA309365164GA SA309365172GA SA309365177GA SA309365191GA SA309365192GA SA309365194GA SA309365196GA SA309365201GA SA309365202GA SA309365222GA SA309375503GA SA309405819GA SA309426155GA SA309436327GA SA309436331GA SA309446392GA SA309446437GA SA309446482GA SA309446515GA SA309446516GA SA309446528GA SA309456718GA SA309456720GA SA309456731GA SA309466735GA SA309466739GA SA309466746GA SA309466754GA SA309466759GA SA309466769GA SA309466780GA SA309466795GA SA309466797GA SA309466799GA SA309476861GA SA309476866GA SA309476867GA SA309476869GA SA309476870GA SA309476881GA SA309476883GA SA309476884GA SA309476887GA SA309476888GA SA309476889GA SA309476890GA SA309476891GA SA309476898GA SA309476950GA SA309476993GA SA309476995GA SA309476996GA SA309477003GA SA309477008GA SA309477032GA SA309477136GA SA309517710GA SA309527915GA SA309527931GA SA309527938GA SA309538038GA SA310028193GA SA310028200GA SA310028219GA SA310028224GA SA310028271GA SA310028276GA SA310038354GA SA310038372GA SA310038377GA SA310038389GA SA310038406GA SA310038410GA SA310038413GA SA310038414GA SA310038415GA SA310038419GA SA310038425GA SA310038426GA SA310038435GA SA310038438GA SA310038439GA SA310038443GA SA310038446GA SA310038453GA SA310038457GA SA310038459GA SA310038472GA SA310038473GA SA310038474GA SA310048584GA SA310048585GA SA310048588GA SA310048589GA SA310048590GA SA310048593GA SA310048596GA SA310048597GA SA310048622GA SA310058741GA SA310058747GA SA310068820GA SA310068828GA SA310260005GA SA310260364GA SA310260413GA SA310260416GA SA310310877GA SA310310886GA SA310310896GA SA310310914GA SA310310915GA SA310310917GA SA310310930GA SA310310933GA SA310310944GA SA310310951GA SA310310953GA SA310310954GA SA310310957GA SA310310961GA SA310310964GA SA310310987GA SA310321023GA SA310331226GA SA310331250GA SA310331257GA SA310331260GA SA310331267GA SA310331268GA SA310331312GA SA310331314GA SA310331315GA SA310331317GA SA310331319GA SA310331320GA SA310331321GA SA310331325GA SA310331355GA SA310341383GA SA310361657GA SA310361836GA SA310361879GA SA310372075GA SA310372080GA SA310372095GA SA310372100GA SA310372110GA SA310372115GA SA310372117GA SA310372120GA SA310372130GA SA310372136GA SA310372173GA SA310372174GA SA310372176GA SA310372197GA SA310372198GA SA310372201GA SA310372207GA SA310382344GA SA310382360GA SA310382456GA SA310382539GA SA310382692GA SA310382846GA SA310392893GA SA310392901GA SA310392907GA SA310392909GA SA310392917GA SA310392930GA SA310393089GA SA310393096GA SA310393112GA SA310393136GA SA310393185GA SA310393278GA SA310393285GA SA310393287GA SA310393293GA SA310393297GA SA310393298GA SA310393300GA SA310393302GA SA310393305GA SA310393308GA SA310413479GA SA310413481GA SA310413505GA SA310413518GA SA310413522GA SA310423576GA SA310423658GA SA310423702GA SA310423829GA SA310423831GA SA310423834GA SA310423851GA SA310423970GA SA310424042GA SA310434145GA SA310444311GA SA310444506GA SA310444510GA SA310454997GA SA310455078GA SA310455166GA SA310465371GA SA310465376GA SA310465379GA SA310475617GA SA310475726GA SA310496127GA SA310496218GA SA310496278GA SA310496282GA SA310496285GA SA310496309GA SA310496321GA SA310496617GA SA310506767GA SA310506776GA SA310506793GA SA310506798GA SA310506802GA SA310506815GA SA310506823GA SA310506825GA SA310506829GA SA310506833GA SA310507054GA SA310507061GA SA310507147GA SA310507150GA SA310507165GA SA310507167GA SA310507172GA SA310507173GA SA310507191GA SA310507192GA SA310507207GA SA310507209GA SA310507210GA SA310507211GA SA310507214GA SA310507229GA SA310507234GA SA310507245GA SA310507251GA SA310507261GA SA310507264GA SA310507265GA SA310507269GA SA310507273GA SA310507274GA SA310507277GA SA310507282GA SA310507283GA SA310507285GA SA310507341GA SA310507350GA SA310507358GA SA310507359GA SA310507364GA SA310507365GA SA310507369GA SA310507394GA SA310507513GA SA310507514GA SA310507542GA SA310517561GA SA310517571GA SA310517579GA SA310517580GA SA310517582GA SA310517584GA SA310517586GA SA310517598GA SA310517603GA SA310517609GA SA310517615GA SA310517620GA SA310517628GA SA310517630GA SA310517647GA SA310517655GA SA310517658GA SA310517716GA SA310517748GA SA310517749GA SA310517750GA SA310517779GA SA310517816GA SA310518151GA SA310518155GA SA310518173GA SA310518183GA SA310518188GA SA310518199GA SA310518239GA SA310528860GA SA310528876GA SA310528882GA SA310528964GA SA310528966GA SA310528974GA SA310528988GA SA310528991GA SA310528993GA SA310528996GA SA310539350GA SA311039371GA SA311039373GA SA311039404GA SA311039435GA SA311039455GA SA311039465GA SA311039474GA SA311039481GA SA311039489GA SA311059586GA SA311059594GA SA311060010GA SA311060014GA SA311060033GA SA311060045GA SA311060126GA SA311060148GA SA311060154GA SA311060157GA SA311060163GA SA311060232GA SA311060274GA SA311069626GA SA311069628GA SA311069629GA SA311069631GA SA311069632GA SA311069633GA SA311069634GA SA311069635GA SA311069642GA SA311069643GA SA311069647GA SA311069664GA SA311069673GA SA311069686GA SA311069698GA SA311069723GA SA311069834GA SA311070536GA SA311080672GA SA311080673GA SA311080675GA SA311080681GA SA311080686GA SA311080691GA SA311080693GA SA311080696GA SA311080697GA SA311090792GA SA311090934GA SA311090938GA SA311091007GA SA311101098GA SA311101272GA SA311101274GA SA311101279GA SA311101403GA SA311101424GA SA311111614GA SA311121779GA SA311121849GA SA311121852GA SA311121855GA SA311121857GA SA311121859GA SA311121860GA SA311121862GA SA311121864GA SA311121865GA SA311121866GA SA311121871GA SA311121873GA SA311121874GA SA311121875GA SA311121879GA SA311121882GA SA311121883GA SA311121884GA SA311121886GA SA311121887GA SA311121893GA SA311121896GA SA311121898GA SA311121915GA SA311121916GA SA311122013GA SA311122015GA SA311122018GA SA311122021GA SA311122026GA SA311122029GA SA311122044GA SA311122046GA SA311122053GA SA311122055GA SA311122141GA SA311122230GA SA311122248GA SA311122251GA SA311132287GA SA311132313GA SA311132314GA SA311132322GA SA311132352GA SA311132377GA SA311132387GA SA311132398GA SA311132546GA SA311132554GA SA311142755GA SA311142773GA SA311142776GA SA311142778GA SA311142794GA SA311142796GA SA311142918GA SA311142947GA SA311153033GA SA311153077GA SA311153258GA SA311153307GA SA311163383GA SA311163404GA SA311163438GA SA311163450GA SA311163452GA SA311163531GA SA311163535GA SA311163550GA SA311163566GA SA311163568GA SA311163588GA SA311163597GA SA311163648GA SA311163699GA SA311163792GA SA311163805GA SA311163809GA SA311163814GA SA311163821GA SA311163833GA SA311163835GA SA311163854GA SA311163856GA SA311163865GA SA311163869GA SA311163888GA SA311163890GA SA311163893GA SA311163914GA SA311163947GA SA311163951GA SA311163955GA SA311163962GA SA311163980GA SA311174045GA SA311174047GA SA311174058GA SA311174065GA SA311174127GA SA311184379GA SA311184383GA SA311184457GA SA311184464GA SA311184470GA SA311184471GA SA311184473GA SA311184476GA SA311184482GA SA311184486GA SA311184488GA SA311184490GA SA311184501GA SA311184508GA SA311184510GA SA311184511GA SA311184512GA SA311184521GA SA311184690GA SA311184722GA SA311184725GA SA311184729GA SA311184730GA SA311184732GA SA311184735GA SA311184736GA SA311184737GA SA311184739GA SA311184751GA SA311184752GA SA311184753GA SA311184763GA SA311184768GA SA311184769GA SA311184776GA SA311184788GA SA311184790GA SA311184791GA SA311184797GA SA311184798GA SA311184799GA SA311184805GA SA311184808GA SA311184816GA SA311184825GA SA311184869GA SA311184871GA SA311194931GA SA311194962GA SA311194973GA SA311194987GA SA311195007GA SA311195042GA SA311195043GA SA311195047GA SA311195048GA SA311195049GA SA311195051GA SA311195052GA SA311195062GA SA311195074GA SA311195076GA SA311195085GA SA311195087GA SA311195170GA SA311195171GA SA311195172GA SA311205311GA SA311205313GA SA311205317GA SA311205322GA SA311205328GA SA311205339GA SA311205393GA SA311205403GA SA311205404GA SA311205405GA SA311205406GA SA311205407GA SA311205413GA SA311205415GA SA311205417GA SA311205419GA SA311205421GA SA311205422GA SA311205428GA SA311205431GA SA311205433GA SA311205435GA SA311205438GA SA311205570GA SA311205576GA SA311205583GA SA311205586GA SA311205594GA SA311205595GA SA311205596GA SA311205598GA SA311205599GA SA311205600GA SA311205603GA SA311205606GA SA311205612GA SA311205620GA SA311225897GA SA311225928GA SA311225929GA SA311225930GA SA311225951GA SA311225957GA SA311236041GA SA311236044GA SA311236373GA SA311246451GA SA311246455GA SA311246458GA SA311246464GA SA311246475GA SA311246491GA SA311246505GA SA311246508GA SA311246519GA SA311246532GA SA311246535GA SA311246536GA SA311246538GA SA311246539GA SA311246540GA SA311246548GA SA311246550GA SA311246554GA SA311246555GA SA311246556GA SA311246557GA SA311246563GA SA311246579GA SA311246590GA SA311246667GA SA311256722GA SA311256724GA SA311256736GA SA311256811GA SA311256820GA SA311256824GA SA311256846GA SA311256853GA SA311256863GA SA311266877GA SA311266885GA SA311266887GA SA311266888GA SA311266904GA SA311266905GA SA311266908GA SA311297310GA SA311307452GA SA311307490GA SA311458307GA SA311458308GA SA311458309GA SA311458311GA SA311458312GA SA311458313GA SA311458314GA SA311458316GA SA311458319GA SA311458331GA SA311458334GA SA311468412GA SA311468413GA SA311468416GA SA311468418GA SA311468421GA SA311468422GA SA311468425GA SA311468427GA SA311468429GA SA311468431GA SA311468436GA SA311468437GA SA311468440GA SA311468441GA SA311468442GA SA311468444GA SA311468449GA SA311488463GA SA311488466GA SA311488467GA SA312018906GA SA312028949GA SA312028967GA SA312028990GA SA312028996GA SA312029068GA SA312029072GA SA312039115GA SA312039188GA SA312049323GA SA312049325GA SA312049333GA SA312049334GA SA312049357GA SA312049366GA SA312049422GA SA312049424GA SA312049444GA SA312059478GA SA312059629GA SA312069705GA SA312069788GA SA312069791GA SA312069800GA SA312069819GA SA312069821GA SA312069824GA SA312069827GA SA312069830GA SA312069835GA SA312069840GA SA312069845GA SA312069854GA SA312069859GA SA312070078GA SA312070268GA SA312079889GA SA312079893GA SA312079896GA SA312079898GA SA312079901GA SA312079904GA SA312079905GA SA312079906GA SA312079909GA SA312079910GA SA312079911GA SA312079925GA SA312079949GA SA312080468GA SA312080475GA SA312080480GA SA312080518GA SA312080519GA SA312080527GA SA312080643GA SA312080699GA SA312080786GA SA312080789GA SA312080825GA SA312080841GA SA312080850GA SA312080854GA SA312080861GA SA312080889GA SA312080895GA SA312080902GA SA312080904GA SA312080906GA SA312080907GA SA312080908GA SA312091029GA SA312091047GA SA312091049GA SA312091065GA SA312091138GA SA312091140GA SA312091150GA SA312091318GA SA312091351GA SA312101367GA SA312101368GA SA312101371GA SA312101373GA SA312101381GA SA312101382GA SA312101383GA SA312101384GA SA312101385GA SA312101387GA SA312101388GA SA312101392GA SA312101393GA SA312111677GA SA312121744GA SA312121942GA SA312121947GA SA312121995GA SA312121999GA SA312122046GA SA312152494GA SA312232865GA SA312232893GA SA312232903GA SA312232909GA SA312253140GA SA312253145GA SA312253150GA SA312253157GA SA312253163GA SA312253203GA SA312253210GA SA312253264GA SA312253294GA SA312253298GA SA312253318GA SA312253363GA SA312253364GA SA312253440GA SA312253454GA SA312253476GA SA312253591GA SA312263632GA SA312263639GA SA312263668GA SA312263677GA SA312263693GA SA312263763GA SA312263796GA SA312263815GA SA312263819GA SA312263911GA SA312264036GA SA312284117GA SA312284230GA SA312284300GA SA312294435GA SA312294474GA SA312294479GA SA312294528GA SA312294774GA SA312344873GA SA312344896GA SA312344910GA SA312344916GA SA312344923GA SA312344930GA SA312344937GA SA312344943GA SA312344949GA SA312344950GA SA312344961GA SA312344975GA SA312344978GA SA312345019GA SA312345021GA SA312345023GA SA312345024GA SA312345025GA SA312345027GA SA312345029GA SA312345032GA SA312345035GA SA312345096GA SA312345101GA SA312345161GA SA312345179GA SA312365554GA SA312365613GA SA312365622GA SA312365623GA SA312365631GA SA312365635GA SA312365639GA SA312365640GA SA312365644GA SA312365668GA SA312365689GA SA312365712GA SA312365732GA SA312365834GA SA312365843GA SA312365854GA SA312375952GA SA312375955GA SA312375956GA SA312375957GA SA312375958GA SA312375959GA SA312375960GA SA312375961GA SA312375963GA SA312375964GA SA312375965GA SA312375966GA SA312375967GA SA312375968GA SA312375969GA SA312375970GA SA312375971GA SA312375973GA SA312375974GA SA312375976GA SA312375977GA SA312375978GA SA312375979GA SA312375980GA SA312375986GA SA312376006GA SA312376023GA SA312376104GA SA312376105GA SA312376107GA SA312376108GA SA312376111GA SA312376112GA SA312376113GA SA312376114GA SA312376115GA SA312376117GA SA312376118GA SA312376121GA SA312376124GA SA312376127GA SA312376166GA SA312376168GA SA312376191GA SA312376227GA SA312376238GA SA312376293GA SA312376299GA SA312376303GA SA312376307GA SA312376310GA SA312378023GA SA312378039GA SA312378041GA SA312378044GA SA312378045GA SA312378050GA SA312378054GA SA312378065GA SA312386313GA SA312386318GA SA312386319GA SA312386321GA SA312386322GA SA312386324GA SA312386325GA SA312386327GA SA312386328GA SA312386340GA SA312386341GA SA312386344GA SA312386345GA SA312386348GA SA312386349GA SA312386351GA SA312386352GA SA312386354GA SA312386355GA SA312386357GA SA312386358GA SA312386371GA SA312386375GA SA312386392GA SA312386395GA SA312386398GA SA312386483GA SA312386501GA SA312386529GA SA312386552GA SA312386562GA SA312386567GA SA312396775GA SA312396781GA SA312396853GA SA312396856GA SA312396873GA SA312396884GA SA312396888GA SA312396897GA SA312396908GA SA312396943GA SA312396944GA SA312396946GA SA312396949GA SA312396953GA SA312396954GA SA312396955GA SA312396957GA SA312396960GA SA312396962GA SA312396964GA SA312396967GA SA312396968GA SA312397037GA SA312397040GA SA312397043GA SA312397048GA SA312397053GA SA312397057GA SA312397058GA SA312397059GA SA312397091GA SA312397094GA SA312397095GA SA312397096GA SA312397097GA SA312397099GA SA312397102GA SA312397105GA SA312397107GA SA312397109GA SA312397110GA SA312397257GA SA312417306GA SA312417311GA SA312417316GA SA312417317GA SA312417318GA SA312417323GA SA312417328GA SA312417329GA SA312417332GA SA312417344GA SA312417346GA SA312417349GA SA312417362GA SA312417364GA SA312417365GA SA312417366GA SA312417374GA SA312417382GA SA312417386GA SA312417392GA SA312417397GA SA312417410GA SA312447570GA SA312468598GA SA312468620GA SA312468658GA SA312468659GA SA312468660GA SA312468661GA SA312468669GA SA312468703GA SA312468720GA SA312469262GA SA312469400GA SA312469401GA SA312469403GA SA312469405GA SA312469407GA SA312469408GA SA312469410GA SA312469412GA SA312469414GA SA312469423GA SA312469424GA SA312469427GA SA312469430GA SA312469432GA SA312469441GA SA312469493GA SA312469494GA SA312469496GA SA312469497GA SA312469498GA SA312469503GA SA312469516GA SA312469517GA SA312479528GA SA312479529GA SA312479530GA SA312479531GA SA312479533GA SA312479534GA SA312479535GA SA312479536GA SA312479539GA SA312479541GA SA312479544GA SA312479549GA SA312479572GA SA312479573GA SA312479581GA SA312479600GA SA312479610GA SA312479617GA SA312479621GA SA312479626GA SA312479627GA SA312479649GA SA312479695GA SA312479803GA SA312479804GA SA312479814GA SA312479819GA SA312479825GA SA312479837GA SA312479838GA SA312479839GA SA312479842GA SA312479844GA SA312479845GA SA312479850GA SA312479930GA SA312480097GA SA312480124GA SA312480125GA SA312480126GA SA312480129GA SA312480130GA SA312480132GA SA312480135GA SA312480136GA SA312480138GA SA312480143GA SA312480148GA SA312480150GA SA312480151GA SA312480153GA SA312480155GA SA312480156GA SA312480157GA SA312480158GA SA312480159GA SA312480163GA SA312480164GA SA312480166GA SA312480168GA SA312480169GA SA312480170GA SA312480171GA SA312480172GA SA312480173GA SA312480174GA SA312480179GA SA312480180GA SA312480181GA SA312480182GA SA312480186GA SA312480188GA SA312480195GA SA312480196GA SA312480198GA SA312480210GA SA312480212GA SA312480213GA SA312480214GA SA312480225GA SA312480230GA SA312480232GA SA312480238GA SA312480242GA SA312480244GA SA312480265GA SA312480266GA SA312480428GA SA312480429GA SA312480433GA SA312480435GA SA312480443GA SA312480444GA SA312480461GA SA312480475GA SA312480506GA SA312480508GA SA312480521GA SA312480527GA SA312480529GA SA312480534GA SA312480543GA SA312480545GA SA312480549GA SA312480550GA SA312480556GA SA312480557GA SA312480564GA SA312480567GA SA312480569GA SA312480574GA SA312480580GA SA312480584GA SA312480586GA SA312480588GA SA312480589GA SA312480590GA SA312480592GA SA312480593GA SA312480594GA SA312480599GA SA312480605GA SA312480606GA SA312480608GA SA312480611GA SA312480613GA SA312480617GA SA312480621GA SA312480623GA SA312480625GA SA312480626GA SA312480627GA SA312480634GA SA312480635GA SA312480636GA SA312480637GA SA312480639GA SA312480640GA SA312480641GA SA312480643GA SA312480647GA SA312480652GA SA312480659GA SA312480692GA SA312480696GA SA312480697GA SA312480698GA SA312480699GA SA312480701GA SA312480702GA SA312480706GA SA312480707GA SA312480710GA SA312490883GA SA312490920GA SA312490968GA SA312490970GA SA312490973GA SA312490975GA SA312490981GA SA312490985GA SA312490986GA SA312490991GA SA312490992GA SA312491002GA SA312491014GA SA312491020GA SA312491021GA SA312491028GA SA312491032GA SA312491037GA SA312491044GA SA312491079GA SA312491093GA SA312491098GA SA312491124GA SA312491127GA SA312491131GA SA312491139GA SA312491140GA SA312491145GA SA312491156GA SA312491159GA SA312491160GA SA312491164GA SA312491165GA SA312491173GA SA312491177GA SA312491178GA SA312491183GA SA312491186GA SA312491189GA SA312491190GA SA312491191GA SA312491192GA SA312491193GA SA312491194GA SA312491197GA SA312491201GA SA312491203GA SA312491205GA SA312491209GA SA312491214GA SA312491215GA SA312491221GA SA312491223GA SA312491229GA SA312491230GA SA312491235GA SA312491236GA SA312491238GA SA312491241GA SA312491242GA SA312491243GA SA312491246GA SA312491251GA SA312491253GA SA312491254GA SA312491261GA SA312491263GA SA312491267GA SA312491268GA SA312491270GA SA312491271GA SA312491273GA SA312491275GA SA312491276GA SA312491277GA SA312491278GA SA312491281GA SA312491283GA SA312491287GA SA312491292GA SA312491294GA SA312491302GA SA312491307GA SA312491311GA SA312491342GA SA312491355GA SA312491361GA SA312491380GA SA312491381GA SA312491387GA SA312491391GA SA312491393GA SA312491394GA SA312491396GA SA312491397GA SA312491400GA SA312491401GA SA312491405GA SA312491411GA SA312491451GA SA312501742GA SA312501760GA SA312501767GA SA312501770GA SA312501787GA SA312501788GA SA312501790GA SA312501792GA SA312501807GA SA312501839GA SA312501854GA SA312501857GA SA312501860GA SA312501861GA SA312501862GA SA312501902GA SA312502039GA SA312502068GA SA312502074GA SA312502076GA SA312502077GA SA312502080GA SA312502083GA SA312502098GA SA312502137GA SA312502142GA SA312502144GA SA312502156GA SA312502157GA SA312502159GA SA312502168GA SA312502170GA SA312512176GA SA312512178GA SA312522388GA SA312522389GA SA312522393GA SA312522399GA SA312522400GA SA312522404GA SA312522406GA SA312522407GA SA312522409GA SA312522411GA SA312522414GA SA312522415GA SA312522418GA SA312522419GA SA312522422GA SA312522423GA SA312522424GA SA312522425GA SA312522426GA SA312522427GA SA312522428GA SA312522429GA SA312522433GA SA312522434GA SA312522437GA SA312522439GA SA312522486GA SA312522487GA SA312522489GA SA312522490GA SA312522491GA SA312522493GA SA312522494GA SA312522500GA SA312522545GA SA312522549GA SA313012911GA SA313023195GA SA313023332GA SA313023340GA SA313023472GA SA313033664GA SA313043872GA SA313043887GA SA313043888GA SA313043904GA SA313043915GA SA313043927GA SA313043929GA SA313043931GA SA313043938GA SA313043941GA SA313043974GA SA313043975GA SA313043980GA SA313043981GA SA313043983GA SA313043984GA SA313043987GA SA313043989GA SA313043991GA SA313043997GA SA313043999GA SA313044000GA SA313044001GA SA313044002GA SA313044125GA SA313064184GA SA313064186GA SA313064191GA SA313064193GA SA313064197GA SA313064267GA SA313064270GA SA313064271GA SA313064275GA SA313064278GA SA313064281GA SA313064284GA SA313064285GA SA313064287GA SA313064288GA SA313064292GA SA313064306GA SA313064310GA SA313064319GA SA313064343GA SA313064345GA SA313064347GA SA313064352GA SA313064353GA SA313064355GA SA313064357GA SA313064361GA SA313064365GA SA313064366GA SA313064367GA SA313064371GA SA313064372GA SA313064375GA SA313064377GA SA313064379GA SA313064380GA SA313064381GA SA313064385GA SA313064392GA SA313064403GA SA313064427GA SA313064433GA SA313064440GA SA313064442GA SA313064443GA SA313084464GA SA313084473GA SA313084630GA SA313084802GA SA313084934GA SA313084935GA SA313084951GA SA313094984GA SA313095048GA SA313095062GA SA313095079GA SA313095084GA SA313095102GA SA313095112GA SA313095161GA SA313095257GA SA313095273GA SA313095276GA SA313095290GA SA313095324GA SA313095326GA SA313095330GA SA313095352GA SA313095379GA SA313095385GA SA313095424GA SA313095433GA SA313095442GA SA313095454GA SA313095562GA SA313095566GA SA313095573GA SA313095576GA SA313095591GA SA313095597GA SA313095600GA SA313095602GA SA313095603GA SA313095606GA SA313095609GA SA313095610GA SA313095612GA SA313095613GA SA313095626GA SA313095635GA SA313095638GA SA313095643GA SA313105653GA SA313105754GA SA313116616GA SA313116673GA SA313116678GA SA313116680GA SA313116689GA SA313116696GA SA313116698GA SA313116702GA SA313116704GA SA313116706GA SA313116709GA SA313116710GA SA313116716GA SA313116721GA SA313116764GA SA313116767GA SA313116775GA SA313116791GA SA313116883GA SA313117039GA SA313117040GA SA313117046GA SA313117048GA SA313117052GA SA313117054GA SA313117057GA SA313127302GA SA313127304GA SA313127306GA SA313127307GA SA313127308GA SA313127309GA SA313127310GA SA313127311GA SA313127312GA SA313127313GA SA313127315GA SA313127319GA SA313127320GA SA313127321GA SA313127322GA SA313127327GA SA313127328GA SA313127329GA SA313127362GA SA313127366GA SA313127367GA SA313127369GA SA313127370GA SA313127371GA SA313127372GA SA313127373GA SA313127374GA SA313127375GA SA313127378GA SA313127381GA SA313127382GA SA313127383GA SA313127384GA SA313127387GA SA313127388GA SA313127389GA SA313127398GA SA313127400GA SA313127403GA SA313127404GA SA313127407GA SA313127410GA SA313127411GA SA313127414GA SA313127483GA SA313127485GA SA313127487GA SA313127488GA SA313127492GA SA313127495GA SA313127496GA SA313127497GA SA313127498GA SA313127501GA SA313127502GA SA313127503GA SA313304120GA SA313356466GA SA313367503GA SA313367510GA SA313367511GA SA313367517GA SA313367520GA SA313367525GA SA313367526GA SA313367529GA SA313367531GA SA313367534GA SA313367535GA SA313367538GA SA313367539GA SA313367541GA SA313367542GA SA313367545GA SA313367551GA SA313367557GA SA313367559GA SA313367560GA SA313367561GA SA313367564GA SA313367567GA SA313367568GA SA313367577GA SA313367583GA SA313377651GA SA313377652GA SA313377656GA SA313377662GA SA313377663GA SA313377668GA SA313377671GA SA313377673GA SA313377682GA SA313377683GA SA313377684GA SA313377686GA SA313377690GA SA313377692GA SA313377696GA SA313377697GA SA313377698GA SA313377713GA SA313377714GA SA313377715GA SA313377718GA SA313378028GA SA313378031GA SA313378037GA SA313378042GA SA313378043GA SA313378045GA SA313378064GA SA313378067GA SA314065786GA SA314086568GA SA314086633GA SA314086685GA SA314086691GA SA314086733GA SA314097044GA SA314097330GA SA314107387GA SA314107468GA SA314107656GA SA314107853GA SA314107860GA SA314107894GA SA314107900GA SA314107920GA SA314107925GA SA314107962GA SA314107978GA SA314107980GA SA314107982GA SA314108031GA SA314108101GA SA314108138GA SA314108139GA SA314108140GA SA314108141GA SA314108142GA SA314108143GA SA314108144GA SA314108145GA SA314108146GA SA314108148GA SA314108149GA SA314108150GA SA314108151GA SA314108153GA SA314108154GA SA314108155GA SA314108156GA SA314108157GA SA314108158GA SA314108159GA SA314108160GA SA314108161GA SA314108163GA SA314108164GA SA314108166GA SA314108167GA SA314108168GA SA314108169GA SA314108170GA SA314108171GA SA314108172GA SA314108173GA SA314108174GA SA314108175GA SA314108176GA SA314108177GA SA314108178GA SA314108179GA SA314108180GA SA314108181GA SA314108182GA SA314108184GA SA314108185GA SA314108186GA SA314108187GA SA314108188GA SA314108189GA SA314108190GA SA314108191GA SA314108194GA SA314108195GA SA314108196GA SA314108197GA SA314108198GA SA314108199GA SA314108200GA SA314108202GA SA314108203GA SA314108205GA SA314108206GA SA314108207GA SA314108208GA SA314108209GA SA314108210GA SA314108211GA SA314108212GA SA314108213GA SA314108214GA SA314108216GA SA314108217GA SA314108218GA SA314108219GA SA314108220GA SA314108221GA SA314108223GA SA314108241GA SA314108242GA SA314108246GA SA314108247GA SA314108271GA SA314108276GA SA314108277GA SA314108278GA SA314108282GA SA314108288GA SA314108289GA SA314108292GA SA314108296GA SA314108297GA SA314108307GA SA314118367GA SA314118369GA SA314118372GA SA314118374GA SA314118376GA SA314118378GA SA314118381GA SA314118383GA SA314118386GA SA314118387GA SA314118390GA SA314118392GA SA314118394GA SA314118397GA SA314118398GA SA314118400GA SA314118401GA SA314118402GA SA314118403GA SA314118405GA SA314118409GA SA314118410GA SA314118411GA SA314118415GA SA314118416GA SA314118417GA SA314118418GA SA314118419GA SA314118476GA SA314118510GA SA314118512GA SA314118513GA SA314118514GA SA314118515GA SA314118518GA SA314118519GA SA314118520GA SA314118522GA SA314118526GA SA314118528GA SA314118530GA SA314118531GA SA314118536GA SA314118537GA SA314118538GA SA314118544GA SA314118545GA SA314118546GA SA314118547GA SA314118548GA SA314118549GA SA314118550GA SA314118551GA SA314118552GA SA314118553GA SA314118554GA SA314118555GA SA314118556GA SA314118557GA SA314118558GA SA314118561GA SA314118563GA SA314118564GA SA314118565GA SA314118566GA SA314118567GA SA314118568GA SA314118569GA SA314118570GA SA314118571GA SA314118572GA SA314118573GA SA314118586GA SA314118618GA SA314118656GA SA314118735GA SA314118820GA SA314118835GA SA314118850GA SA314118856GA SA314118874GA SA314118876GA SA314118878GA SA314118879GA SA314118880GA SA314118881GA SA314118883GA SA314118884GA SA314118885GA SA314118887GA SA314118888GA SA314118889GA SA314118890GA SA314118891GA SA314118892GA SA314118893GA SA314118896GA SA314118897GA SA314118898GA SA314118899GA SA314118900GA SA314118901GA SA314118902GA SA314118904GA SA314118905GA SA314118906GA SA314118907GA SA314118908GA SA314118909GA SA314118910GA SA314118911GA SA314118912GA SA314118913GA SA314118914GA SA314118915GA SA314118916GA SA314118917GA SA314118918GA SA314118919GA SA314118920GA SA314118921GA SA314118922GA SA314118923GA SA314118924GA SA314118925GA SA314118926GA SA314118927GA SA314118928GA SA314118929GA SA314118930GA SA314118931GA SA314118932GA SA314118933GA SA314118934GA SA314118935GA SA314118936GA SA314118937GA SA314118938GA SA314118939GA SA314118940GA SA314118941GA SA314118942GA SA314118943GA SA314118944GA SA314118946GA SA314118947GA SA314118948GA SA314118950GA SA314118952GA SA314118953GA SA314118954GA SA314118955GA SA314118956GA SA314118959GA SA314118960GA SA314118965GA SA314118966GA SA314118969GA SA314118970GA SA314118976GA SA314118977GA SA314118981GA SA314118985GA SA314119007GA SA314119012GA SA314119015GA SA314119017GA SA314119018GA SA314119024GA SA314119116GA SA314119133GA SA314129229GA SA314129237GA SA314129238GA SA314129240GA SA314129244GA SA314129246GA SA314129248GA SA314129250GA SA314129252GA SA314222530GA SA314222596GA SA314222597GA SA314222598GA SA314222712GA SA314222716GA SA314222717GA SA314222720GA SA314222730GA SA314222731GA SA314222741GA SA314222763GA SA314222767GA SA314222768GA SA314222769GA SA314222770GA SA314222773GA SA314222777GA SA314222779GA SA314222782GA SA314222784GA SA314222785GA SA314222793GA SA314222794GA SA314222795GA SA314222796GA SA314222798GA SA314222799GA SA314222800GA SA314222802GA SA314222804GA SA314222805GA SA314222807GA SA314222810GA SA314222814GA SA314222817GA SA314222821GA SA314222823GA SA314222825GA SA314222826GA SA314222828GA SA314222830GA SA314222833GA SA314222834GA SA314222835GA SA314222836GA SA314222839GA SA314222841GA SA314222842GA SA314222844GA SA313095446GA SA313117145GA SA313127517GA SA313127735GA SA313127740GA SA313198747GA SA313272058GA SA313282563GA SA313282637GA SA313282875GA SA313293285GA SA313304227GA SA313314434GA SA313314757GA SA313314777GA SA313314811GA SA313314827GA SA313314836GA SA313314854GA SA313324939GA SA313324993GA SA313335004GA SA313335048GA SA313335110GA SA313335411GA SA313335430GA SA313345535GA SA313345549GA SA313345577GA SA313345716GA SA313345718GA SA313345730GA SA313345943GA SA313345990GA SA313356180GA SA313356264GA SA313356673GA SA313356696GA SA313366717GA SA313366719GA SA313366722GA SA313366756GA SA313366821GA SA313377907GA SA313378222GA SA313378227GA SA313378230GA SA313378232GA SA313378234GA SA313378236GA SA313378237GA SA313378241GA SA313378242GA SA313378243GA SA313378244GA SA313378245GA SA313378246GA SA313378247GA SA313378248GA SA313378249GA SA313378250GA SA313378251GA SA313378256GA SA313378257GA SA313378258GA SA313378263GA SA313378264GA SA313378265GA SA313378266GA SA313378268GA SA313378269GA SA313378275GA SA313378276GA SA313378279GA SA313378281GA SA313378282GA SA313378284GA SA313378288GA SA313378289GA SA313378290GA SA313378291GA SA313378294GA SA313378296GA SA313378298GA SA313378301GA SA313378310GA SA313378311GA SA313378313GA SA313378314GA SA313378316GA SA313378318GA SA313378321GA SA313378322GA SA313378326GA SA313378327GA SA313378336GA SA313378338GA SA313378339GA SA313378340GA SA313378341GA SA313378345GA SA313378346GA SA313378351GA SA313378352GA SA313378353GA SA313378356GA SA313378358GA SA313378359GA SA313378361GA SA313378368GA SA313378371GA SA313378374GA SA313378376GA SA313378377GA SA313378380GA SA313378381GA SA313378382GA SA313378384GA SA313378387GA SA313378388GA SA313378392GA SA313378401GA SA313378402GA SA313378403GA SA313378404GA SA313378406GA SA313378407GA SA313378408GA SA313378410GA SA313378411GA SA313378412GA SA313378415GA SA313378417GA SA313378418GA SA313378419GA SA313378420GA SA313378421GA SA313378423GA SA313378424GA SA313378425GA SA313378426GA SA313378427GA SA313378428GA SA313378432GA SA313378437GA SA313378440GA SA313378441GA SA313378442GA SA313378443GA SA313378444GA SA313378445GA SA313378447GA SA313378448GA SA313378450GA SA313378451GA SA313378452GA SA313378454GA SA313378457GA SA313378465GA SA313378466GA SA313378472GA SA313378475GA SA313378478GA SA313378481GA SA313388493GA SA313388494GA SA313388496GA SA313388497GA SA313388498GA SA313388500GA SA313388501GA SA313388506GA SA313388514GA SA313388516GA SA313388520GA SA313388521GA SA313388525GA SA313388532GA SA313388533GA SA313388534GA SA313388536GA SA313388538GA SA313388539GA SA313388542GA SA313388543GA SA313388547GA SA313388550GA SA313388551GA SA313388554GA SA313388555GA SA313388557GA SA313388559GA SA313388560GA SA313388683GA SA313388684GA SA313388686GA SA313388687GA SA313388688GA SA313388692GA SA313388697GA SA313388700GA SA313388701GA SA313388705GA SA313388706GA SA313388708GA SA313420031GA SA313420035GA SA313420048GA SA313420049GA SA313420278GA SA313420321GA SA313420322GA SA313420323GA SA313420357GA SA313420361GA SA313420377GA SA313420383GA SA313420390GA SA313420392GA SA313420395GA SA313420397GA SA313420398GA SA313420402GA SA313420408GA SA313420410GA SA313420411GA SA313420415GA SA313420417GA SA313420433GA SA313420434GA SA313420451GA SA313420460GA SA313420461GA SA313420470GA SA313420472GA SA313420477GA SA313420495GA SA313420501GA SA313420546GA SA313430619GA SA313430626GA SA313430671GA SA313430680GA SA313430694GA SA313430696GA SA313430701GA SA313430702GA SA313430704GA SA313430716GA SA313430717GA SA313430720GA SA313430722GA SA313430732GA SA313430734GA SA313430737GA SA313430739GA SA313430740GA SA313430741GA SA313430742GA SA313430744GA SA313430747GA SA313430749GA SA313430758GA SA313430760GA SA313430765GA SA313430770GA SA313430773GA SA313430778GA SA313430786GA SA313430787GA SA313430788GA SA313430896GA SA313430898GA SA313430902GA SA313430903GA SA313430942GA SA313430945GA SA313430946GA SA313430949GA SA313430950GA SA313430951GA SA313430972GA SA313430980GA SA313430981GA SA313430983GA SA313430985GA SA313430986GA SA313431021GA SA313431094GA SA313431108GA SA313431118GA SA313431120GA SA313431121GA SA313431169GA SA313431185GA SA313431186GA SA313431189GA SA313431192GA SA313431205GA SA313431214GA SA313441518GA SA313441523GA SA313451822GA SA313451834GA SA313451968GA SA313452069GA SA313452139GA SA313462321GA SA313462400GA SA313482709GA SA313482711GA SA313482725GA SA313493018GA SA313493080GA SA313493412GA SA313493421GA SA313493440GA SA313493453GA SA313493461GA SA313493462GA SA313493469GA SA313513838GA SA313513868GA SA313513874GA SA313513877GA SA313513879GA SA313513881GA SA313513882GA SA313513884GA SA313513887GA SA313513891GA SA313523948GA SA314014411GA SA314014460GA SA314014504GA SA314014609GA SA314014647GA SA314014703GA SA314014763GA SA314024791GA SA314024853GA SA314024856GA SA314024861GA SA314024866GA SA314024895GA SA314024898GA SA314024906GA SA314024917GA SA314024924GA SA314025007GA SA314025110GA SA314025112GA SA314025113GA SA314025117GA SA314025119GA SA314025120GA SA314025122GA SA314025125GA SA314025126GA SA314025132GA SA314025134GA SA314025135GA SA314025141GA SA314025167GA SA314025212GA SA314035391GA SA314035403GA SA314035405GA SA314045512GA SA314045532GA SA314055552GA SA314055574GA SA314055594GA SA314055693GA SA314055711GA SA314055723GA SA314055732GA SA314075965GA SA314086401GA SA314086416GA SA314086427GA SA314086459GA SA314086462GA SA314086463GA SA314086470GA SA314086485GA SA314086488GA SA314086505GA SA314086546GA SA314086552GA SA314086562GA SA314086574GA SA314086575GA SA314086576GA SA314086583GA SA314086586GA SA314086592GA SA314086598GA SA314086601GA SA314086603GA SA314086607GA SA314086611GA SA314086619GA SA314086621GA SA314086622GA SA314086625GA SA314086631GA SA314086632GA SA314086637GA SA314086653GA SA314086657GA SA314086690GA SA314096763GA SA314096827GA SA314096832GA SA314096836GA SA314096844GA SA314096847GA SA314096856GA SA314096858GA SA314096860GA SA314096872GA SA314096874GA SA314096875GA SA314096880GA SA314096881GA SA314096882GA SA314096884GA SA314096885GA SA314096887GA SA314096888GA SA314096889GA SA314096890GA SA314096893GA SA314096895GA SA314096896GA SA314096898GA SA314096905GA SA314096912GA SA314096940GA SA314096941GA SA314096945GA SA314096954GA SA314096961GA SA314096963GA SA314096971GA SA314096974GA SA314096977GA SA314096979GA SA314096982GA SA314096986GA SA314096990GA SA314096991GA SA314096994GA SA314097000GA SA314097002GA SA314097004GA SA314097008GA SA314097012GA SA314097022GA SA314097026GA SA314097029GA SA314097030GA SA314097039GA SA314097040GA SA314097041GA SA314097043GA SA314097048GA SA314097049GA SA314097060GA SA314097068GA SA314097074GA SA314097075GA SA314097077GA SA314097079GA SA314097083GA SA314097089GA SA314097093GA SA314097098GA SA314097106GA SA314097108GA SA314097109GA SA314097112GA SA314097114GA SA314097116GA SA314097150GA SA314097166GA SA314097247GA SA314097252GA SA314097265GA SA314097289GA SA314107375GA SA314107379GA SA314107386GA SA314107425GA SA314107434GA SA314107448GA SA314118620GA SA314139485GA SA314139487GA SA314139489GA SA314139500GA SA314139519GA SA314139539GA SA314139559GA SA314139563GA SA314139566GA SA314139567GA SA314139576GA SA314139589GA SA314139593GA SA314139597GA SA314139599GA SA314149613GA SA314149615GA SA314149621GA SA314160042GA SA314170464GA SA314180475GA SA314180481GA SA314180492GA SA314180501GA SA314180520GA SA314180521GA SA314180524GA SA314180672GA SA314191019GA SA314191045GA SA314201595GA SA314201608GA SA314201609GA SA314201665GA SA314201753GA SA314211957GA SA314211982GA SA314211990GA SA314212007GA SA314212125GA SA314212132GA SA314212145GA SA314212146GA SA314222912GA SA314233007GA SA314233020GA SA314233021GA SA314233027GA SA314233041GA SA314233044GA SA314233046GA SA314233057GA SA314243211GA SA314243223GA SA314243235GA SA314243236GA SA314243241GA SA314243253GA SA314243255GA SA314243263GA SA314243264GA SA314243265GA SA314243269GA SA314243286GA SA314243295GA SA314253541GA SA314253554GA SA314253562GA SA314253564GA SA314253567GA SA314253586GA SA314253600GA SA314253605GA SA314253608GA SA314253612GA SA314253614GA SA314253617GA SA314253621GA SA314253624GA SA314253626GA SA314253629GA SA314253631GA SA314253643GA SA314253653GA SA314253658GA SA314253659GA SA314253670GA SA314253678GA SA314273955GA SA314274110GA SA314274221GA SA314294903GA SA314294910GA SA314294937GA SA314294944GA SA314294950GA SA314294951GA SA314294953GA SA314294955GA SA314294956GA SA314294960GA SA314294964GA SA314294966GA SA314294968GA SA314294969GA SA314294971GA SA314294973GA SA314294976GA SA314294985GA SA314294986GA SA314294987GA SA314294988GA SA314294991GA SA314294994GA SA314294996GA SA314294997GA SA314294998GA SA314294999GA SA314295000GA SA314295005GA SA314295009GA SA314295011GA SA314295012GA SA314295013GA SA314295015GA SA314295016GA SA314295019GA SA314295055GA SA314295133GA SA314295141GA SA314295142GA SA314295152GA SA314295153GA SA314295161GA SA314295162GA SA314295164GA SA314295167GA SA314295170GA SA314295180GA SA314295186GA SA314295218GA SA314295229GA SA314295236GA SA314295244GA SA314295249GA SA314295251GA SA314295252GA SA314295262GA SA314295263GA SA314295266GA SA314295270GA SA314295274GA SA314295275GA SA314295421GA SA310455168GA SA311195116GA SA311195150GA SA313431239GA SA313044030GA SA310028250GA SA310028261GA SA310028297GA SA310028303GA SA310038383GA SA310038468GA SA310068864GA SA310068879GA SA311101295GA SA311111650GA SA311121762GA SA311246420GA SA311246427GA SA311246432GA SA311246435GA SA311246439GA SA311246441GA SA311297274GA SA311297288GA SA311307413GA SA311307414GA SA311307422GA SA311307426GA SA311307429GA SA311307430GA SA311307434GA SA311307435GA SA311307583GA SA311307584GA SA311307591GA SA311458214GA SA311458218GA SA311458219GA SA311458225GA SA311458231GA SA311458243GA SA311458256GA SA311458257GA SA311458262GA SA311458263GA SA311458343GA SA311458355GA SA311458356GA SA311458360GA SA311458364GA SA311458365GA SA311458368GA SA311518728GA SA311518729GA SA312080490GA SA312080546GA SA312080557GA SA312080578GA SA312080634GA SA312080867GA SA312080868GA SA312080869GA SA312080883GA SA312080910GA SA312101447GA SA312111515GA SA312111551GA SA312111562GA SA312111570GA SA312111657GA SA312111658GA SA312111659GA SA312111661GA SA312111662GA SA312111665GA SA312111668GA SA312111670GA SA312111671GA SA312111672GA SA312111674GA SA312111675GA SA312111676GA SA312111678GA SA312111679GA SA312111680GA SA312111681GA SA312111683GA SA312111684GA SA312111686GA SA312111687GA SA312111691GA SA312121692GA SA312121693GA SA312121696GA SA312121697GA SA312121699GA SA312121710GA SA312121714GA SA312121720GA SA312121722GA SA312121723GA SA312121724GA SA312121725GA SA312121728GA SA312121730GA SA312121731GA SA312121733GA SA312121734GA SA312121735GA SA312121737GA SA312121738GA SA312121743GA SA312121745GA SA312121747GA SA312263904GA SA312263905GA SA312263910GA SA312263921GA SA312263922GA SA312263925GA SA312263933GA SA312263934GA SA312263937GA SA312263938GA SA312263940GA SA312263943GA SA312263944GA SA312263946GA SA312263948GA SA312263980GA SA312263981GA SA312263984GA SA312263985GA SA312263986GA SA312263987GA SA312263988GA SA312263989GA SA312263991GA SA312263992GA SA312263993GA SA312263994GA SA312263995GA SA312263996GA SA312263997GA SA312263998GA SA312263999GA SA312264001GA SA312264003GA SA312264004GA SA312264006GA SA312264008GA SA312458258GA SA312458420GA SA312458425GA SA312458471GA SA312468548GA SA312468549GA SA312468557GA SA313084847GA SA312091089GA SA312152489GA SA312152514GA SA312152541GA SA312152558GA SA312253325GA SA312294380GA SA312294419GA SA312294453GA SA312294458GA SA312294532GA SA312294559GA SA312294561GA SA312294564GA SA312294611GA SA312386379GA SA312386412GA SA313033738GA SA313033752GA SA313033755GA SA313033758GA SA313033759GA SA313033761GA SA313033763GA SA313033764GA SA313033767GA SA313033769GA SA313033842GA SA314075945GA SA314139514GA SA314139555GA SA314160009GA SA314160016GA SA314169915GA SA314169918GA SA314169922GA SA314169928GA SA314169942GA SA314170274GA SA314170282GA SA314170346GA SA314170347GA SA314170348GA SA314170350GA SA314170354GA SA314170356GA SA314170360GA SA314170398GA SA314170399GA SA314170400GA SA314170403GA SA314170407GA SA314170408GA SA314170411GA SA314170412GA SA314170416GA SA314170429GA SA314170439GA SA314170441GA SA314170457GA SA314180500GA SA314180569GA SA314180761GA SA314212050GA SA314212058GA SA314212104GA SA314212108GA SA314212110GA SA314212112GA SA314212113GA SA314212114GA SA314212116GA SA314212228GA SA314212236GA SA314212258GA SA314222292GA SA314222309GA SA314253705GA SA314253706GA SA314253707GA SA314253708GA SA314253710GA SA314253711GA SA314253718GA SA314253720GA SA314253721GA SA314253723GA SA314253726GA SA314253736GA SA314253742GA SA314253747GA SA313272191GA SA314160055GA SA314086666GA SA314086734GA SA314096902GA SA314108165GA SA314108204GA SA314129227GA SA314129230GA SA314129231GA SA314129232GA SA314129233GA SA314129236GA SA314129239GA SA314129241GA SA314129243GA SA314129245GA SA314129247GA SA314129249GA SA314129251GA SA314159800GA SA314159819GA SA314160064GA SA314160066GA SA314160068GA SA314160070GA SA314160071GA SA314160072GA SA314160075GA SA314160076GA SA314160078GA SA314160080GA SA314160081GA SA314160083GA SA314160084GA SA314160086GA SA314160090GA SA314160100GA SA314160116GA SA312091255GA SA307423166GA SA308044528GA SA308481003GA SA308481039GA SA308481040GA SA308481042GA SA308481062GA SA308481066GA SA308481069GA SA308489934GA SA308489938GA SA308489947GA SA308489949GA SA308489969GA SA309365320GA SA309405853GA SA309507424GA SA310159364GA SA310159366GA SA310159382GA SA310159459GA SA310199690GA SA310199712GA SA310199720GA SA310496338GA SA310496339GA SA310507137GA SA310529221GA SA311039388GA SA311039399GA SA311039420GA SA311039437GA SA311039447GA SA311039454GA SA311070513GA SA311070541GA SA311070562GA SA311070564GA SA311080728GA SA311080732GA SA311080734GA SA311090773GA SA311090778GA SA311090779GA SA311090780GA SA311090800GA SA311090804GA SA311090835GA SA311090842GA SA311090850GA SA311090853GA SA311090856GA SA311090861GA SA311090887GA SA311174167GA SA311184362GA SA311184532GA SA311184545GA SA311184556GA SA311184565GA SA311184567GA SA311184568GA SA311184591GA SA311184597GA SA311184599GA SA311184603GA SA311184606GA SA311205394GA SA311205418GA SA311236019GA SA311236027GA SA311236033GA SA311236051GA SA311236059GA SA311236173GA SA311236181GA SA311236266GA SA311236272GA SA311236352GA SA311236357GA SA311236359GA SA311236366GA SA311236368GA SA311236377GA SA311246469GA SA311246504GA SA311246511GA SA311266935GA SA311266941GA SA311297241GA SA312152482GA SA312152556GA SA312365777GA SA312365784GA SA312365799GA SA312365824GA SA313106392GA SA313106395GA SA313106411GA SA313106412GA SA313106414GA SA313106415GA SA313106416GA SA313106418GA SA313106423GA SA313116450GA SA313116456GA SA310038429GA SA311184614GA SA310392884GA SA310393182GA SA310393183GA SA310393208GA SA310393221GA SA311246421GA SA311307431GA SA307241671GA SA307241699GA SA307251787GA SA307251799GA SA307524463GA SA308044485GA SA308044508GA SA308044514GA SA308084626GA SA308084628GA SA308084632GA SA308084633GA SA308084651GA SA308439447GA SA309385535GA SA309385564GA SA309405814GA SA309405815GA SA309405816GA SA309405822GA SA309405823GA SA309405828GA SA309405846GA SA309405847GA SA309405855GA SA309446493GA SA309446511GA SA309456547GA SA309466748GA SA309466773GA SA309466781GA SA309466783GA SA309517683GA SA310038522GA SA310331216GA SA310331235GA SA310331237GA SA310496328GA SA310496365GA SA310496459GA SA310496482GA SA310506857GA SA310507109GA SA310507435GA SA310507531GA SA310529152GA SA311142961GA SA311174238GA SA311174242GA SA311184832GA SA311195198GA SA311205459GA SA311205468GA SA311205470GA SA311205485GA SA311205493GA SA311205496GA SA311205497GA SA311225985GA SA311236093GA SA311236096GA SA311236289GA SA311236311GA SA311236333GA SA311266884GA SA311317644GA SA311327870GA SA311327871GA SA311327876GA SA311327880GA SA311458213GA SA311458259GA SA311458342GA SA311458349GA SA311458359GA SA311468386GA SA311468390GA SA311468400GA SA312069645GA SA312069658GA SA312069681GA SA312069727GA SA312069735GA SA312069744GA SA312069747GA SA312069748GA SA312069752GA SA312069756GA SA312069763GA SA312069771GA SA312069774GA SA312079951GA SA312091093GA SA312091101GA SA312469155GA SA307140603GA SA307140616GA SA307140620GA SA307140625GA SA307150648GA SA307150652GA SA307200780GA SA307220871GA SA307220891GA SA307220966GA SA307220970GA SA307220997GA SA307221019GA SA307221022GA SA307221032GA SA307221090GA SA307221093GA SA307221111GA SA307221147GA SA307221148GA SA307221166GA SA307221167GA SA307231190GA SA307231193GA SA307231198GA SA307231357GA SA307231367GA SA307231371GA SA307231401GA SA307231409GA SA307231446GA SA307231447GA SA307241517GA SA307241584GA SA307241675GA SA307241693GA SA307241700GA SA307241704GA SA307251709GA SA307251717GA SA307251723GA SA307251728GA SA307251749GA SA307312105GA SA307312107GA SA307312110GA SA307322112GA SA307322114GA SA307322116GA SA307322117GA SA307322118GA SA307322119GA SA307322120GA SA307322121GA SA307362416GA SA307362418GA SA307362427GA SA307362436GA SA307392723GA SA307393017GA SA307413109GA SA307413110GA SA307423155GA SA307423158GA SA307423159GA SA307423163GA SA307423164GA SA307423169GA SA307423172GA SA307423173GA SA307423174GA SA307423177GA SA307423180GA SA307423182GA SA307423183GA SA307423184GA SA307443220GA SA307504205GA SA307514341GA SA308084617GA SA308084644GA SA308084760GA SA308084773GA SA308094964GA SA308105104GA SA308105105GA SA308105112GA SA308105146GA SA308105148GA SA308105149GA SA308105150GA SA308105182GA SA308105183GA SA308105184GA SA308105192GA SA308105200GA SA308105206GA SA308105208GA SA308105215GA SA308125362GA SA308135543GA SA308145659GA SA308145677GA SA308196504GA SA308206734GA SA308206737GA SA308206742GA SA308206743GA SA308398744GA SA308398768GA SA308449482GA SA308449492GA SA308449493GA SA308449497GA SA309132800GA SA309142965GA SA309142968GA SA309142982GA SA309153071GA SA309153073GA SA309153076GA SA309153080GA SA309153089GA SA309304222GA SA309324434GA SA309334610GA SA309334684GA SA309334745GA SA309344861GA SA309344866GA SA309354921GA SA309354927GA SA309354940GA SA309395729GA SA309395780GA SA309426112GA SA309436186GA SA309436278GA SA309436279GA SA309507357GA SA309507368GA SA309507369GA SA309507372GA SA309507375GA SA309507399GA SA309507412GA SA309507414GA SA309507416GA SA309507419GA SA309507425GA SA309507433GA SA309507436GA SA309507437GA SA309507439GA SA309507443GA SA309507444GA SA309507447GA SA309507449GA SA309517450GA SA309517451GA SA309517455GA SA309517457GA SA309517645GA SA309528006GA SA309538019GA SA310028113GA SA310028119GA SA310028121GA SA310028143GA SA310028168GA SA310028172GA SA310028237GA SA310028242GA SA310028281GA SA310028299GA SA310028301GA SA310028313GA SA310038356GA SA310038362GA SA310038366GA SA310038378GA SA310038417GA SA310038420GA SA310038421GA SA310038424GA SA310038464GA SA310038465GA SA310038467GA SA310038470GA SA310038475GA SA310038508GA SA310038523GA SA310048623GA SA310048637GA SA310048653GA SA310048661GA SA310048662GA SA310048671GA SA310048674GA SA310048685GA SA310048688GA SA310058702GA SA310058706GA SA310068869GA SA310068949GA SA310119008GA SA310119010GA SA310119011GA SA310119012GA SA310119019GA SA310260540GA SA310290637GA SA310321043GA SA310321067GA SA310341414GA SA310341420GA SA310341457GA SA310341461GA SA310341463GA SA310341464GA SA310351527GA SA310351540GA SA310351544GA SA310351553GA SA310361609GA SA310361628GA SA310361629GA SA310361631GA SA310361635GA SA310361642GA SA310361645GA SA310361656GA SA310361658GA SA310361661GA SA310361662GA SA310361663GA SA310361664GA SA310361665GA SA310361670GA SA310361675GA SA310361676GA SA310361682GA SA310361683GA SA310361685GA SA310361686GA SA310361687GA SA310382840GA SA310393176GA SA310423596GA SA310444256GA SA310455011GA SA310455022GA SA310475465GA SA310475531GA SA310475553GA SA310475555GA SA310475558GA SA310475562GA SA310475564GA SA310475578GA SA310475589GA SA310475602GA SA310475616GA SA310475618GA SA310475620GA SA310475687GA SA310475700GA SA310475704GA SA310475705GA SA310475706GA SA310475714GA SA310475717GA SA310475718GA SA310485761GA SA310485763GA SA310485766GA SA310485770GA SA310485772GA SA310485791GA SA310485813GA SA310485820GA SA310485832GA SA310485839GA SA310485842GA SA310485845GA SA310485858GA SA310485862GA SA310485873GA SA310485880GA SA310496315GA SA310496319GA SA310496325GA SA310496350GA SA310496369GA SA310496389GA SA310496440GA SA310496466GA SA310496475GA SA310496486GA SA310496488GA SA310496493GA SA310506755GA SA310506768GA SA310506774GA SA310506783GA SA310506799GA SA310506803GA SA310506807GA SA310506816GA SA310506842GA SA310507124GA SA310507178GA SA310507195GA SA310507204GA SA310507213GA SA310507222GA SA310507226GA SA310507233GA SA310507249GA SA310507252GA SA310507253GA SA310507255GA SA310507257GA SA310507258GA SA310507280GA SA310507284GA SA310507289GA SA310507293GA SA310507294GA SA310507297GA SA310507299GA SA310507300GA SA310507306GA SA310507312GA SA310507315GA SA310507318GA SA310507321GA SA310507322GA SA310507323GA SA310507327GA SA310507331GA SA310507333GA SA310507334GA SA310507335GA SA310507339GA SA310507340GA SA310507342GA SA310507343GA SA310507345GA SA310507347GA SA310507349GA SA310507352GA SA310507354GA SA310507355GA SA310507360GA SA310507361GA SA310507362GA SA310507363GA SA310507378GA SA310507418GA SA310507424GA SA310507437GA SA310507459GA SA310507462GA SA310507470GA SA310507474GA SA310517599GA SA310517601GA SA310517611GA SA310517617GA SA310517623GA SA310517631GA SA310517639GA SA310517649GA SA310517675GA SA310517677GA SA310517694GA SA310518158GA SA310518165GA SA310518189GA SA310518206GA SA310518211GA SA311039509GA SA311059583GA SA311069740GA SA311069757GA SA311080676GA SA311090746GA SA311090749GA SA311090914GA SA311090915GA SA311090916GA SA311090921GA SA311090922GA SA311090924GA SA311091036GA SA311091046GA SA311091052GA SA311121731GA SA311121732GA SA311121878GA SA311121881GA SA311122143GA SA311122169GA SA311122173GA SA311122174GA SA311122177GA SA311122179GA SA311122187GA SA311122188GA SA311122190GA SA311122192GA SA311122193GA SA311122196GA SA311122206GA SA311122207GA SA311122212GA SA311122221GA SA311122224GA SA311122226GA SA311122227GA SA311122235GA SA311122236GA SA311122238GA SA311122257GA SA311122258GA SA311122259GA SA311122262GA SA311122267GA SA311132277GA SA311132280GA SA311132290GA SA311132291GA SA311132295GA SA311132297GA SA311132318GA SA311132321GA SA311132330GA SA311132340GA SA311132348GA SA311132367GA SA311132399GA SA311132404GA SA311132406GA SA311132423GA SA311132432GA SA311132436GA SA311132451GA SA311142881GA SA311142888GA SA311142914GA SA311142923GA SA311142928GA SA311142956GA SA311153081GA SA311153085GA SA311153087GA SA311153103GA SA311153107GA SA311153109GA SA311153122GA SA311153123GA SA311153130GA SA311153133GA SA311153137GA SA311153142GA SA311153160GA SA311153173GA SA311153182GA SA311153207GA SA311153214GA SA311153218GA SA311153220GA SA311153222GA SA311184542GA SA311205454GA SA311215690GA SA311215695GA SA311215698GA SA311215715GA SA311215731GA SA311215740GA SA311246487GA SA311256808GA SA311256837GA SA311256858GA SA311307524GA SA311307538GA SA311307539GA SA311468376GA SA311468389GA SA311468406GA SA311468410GA SA311518733GA SA312070016GA SA312070029GA SA312080405GA SA312080407GA SA312080418GA SA312080428GA SA312080430GA SA312080568GA SA312080582GA SA312080585GA SA312080599GA SA312080600GA SA312080601GA SA312080605GA SA312080607GA SA312080613GA SA312080616GA SA312080620GA SA312080638GA SA312080639GA SA312080640GA SA312080652GA SA312080657GA SA312080660GA SA312080664GA SA312080696GA SA312080710GA SA312080733GA SA312080734GA SA312080735GA SA312091366GA SA312101436GA SA312111461GA SA312111463GA SA312111465GA SA312111471GA SA312111506GA SA312111507GA SA312111673GA SA312121825GA SA312121908GA SA312121977GA SA312132137GA SA312152312GA SA312152316GA SA312152318GA SA312152327GA SA312152329GA SA312152356GA SA312152360GA SA312152364GA SA312152368GA SA312152469GA SA312152470GA SA312152472GA SA312152481GA SA312152487GA SA312152508GA SA312152543GA SA312152544GA SA312152550GA SA312232814GA SA312242983GA SA312242985GA SA312242993GA SA312243016GA SA312243043GA SA312253082GA SA312284177GA SA312284271GA SA312294441GA SA312294569GA SA312294619GA SA312355226GA SA312355229GA SA312355233GA SA312355247GA SA312355248GA SA312376199GA SA312376213GA SA312396805GA SA312397216GA SA312397251GA SA312407293GA SA312407298GA SA312458398GA SA312469305GA SA312479816GA SA312479951GA SA312480346GA SA312480404GA SA312490889GA SA312490967GA SA312491260GA SA312491395GA SA312491448GA SA312491532GA SA312512306GA SA313013105GA SA313023341GA SA313023358GA SA313023514GA SA313033578GA SA313033581GA SA313033584GA SA313033611GA SA313033613GA SA313033616GA SA313033617GA SA313033618GA SA313033622GA SA313033662GA SA313033673GA SA313033675GA SA313033681GA SA313033683GA SA313033684GA SA313064410GA SA313084477GA SA313084490GA SA313084507GA SA313084565GA SA313084653GA SA313084709GA SA313084915GA SA313084945GA SA313095017GA SA313095081GA SA313095185GA SA313095269GA SA313095466GA SA313095472GA SA313095477GA SA313095570GA SA313095574GA SA313095575GA SA313095583GA SA313095599GA SA313095614GA SA313095632GA SA313105654GA SA313105662GA SA313105668GA SA313105673GA SA313105674GA SA313105678GA SA313105682GA SA313105701GA SA313105704GA SA313105714GA SA313105758GA SA313105899GA SA313105903GA SA313105907GA SA313105913GA SA313105920GA SA313105922GA SA313105949GA SA313105972GA SA313105973GA SA313105981GA SA313105985GA SA313105991GA SA313106023GA SA313106029GA SA313106113GA SA313106124GA SA313106206GA SA313106266GA SA313106296GA SA313106336GA SA313106428GA SA313106439GA SA313106445GA SA313116447GA SA313116460GA SA313116464GA SA313116473GA SA313116491GA SA313116494GA SA313116502GA SA313116515GA SA313116529GA SA313116583GA SA313116598GA SA313116601GA SA313116603GA SA313116607GA SA313116620GA SA313116625GA SA313116629GA SA313116632GA SA313116633GA SA313116634GA SA313116640GA SA313116641GA SA313116647GA SA313116648GA SA313116650GA SA313116654GA SA313116658GA SA313116660GA SA313116666GA SA313116668GA SA313116733GA SA313116741GA SA313116757GA SA313116758GA SA313116952GA SA313116969GA SA313116976GA SA313116994GA SA313116998GA SA313117003GA SA313117072GA SA313117089GA SA313117090GA SA313117099GA SA313117101GA SA313117163GA SA313117191GA SA313117192GA SA313117194GA SA313117205GA SA313127216GA SA313127219GA SA313127221GA SA313127224GA SA313127227GA SA313127230GA SA313127234GA SA313127236GA SA313137773GA SA313137854GA SA313137864GA SA313137871GA SA313188052GA SA313188127GA SA313198223GA SA313198296GA SA313198301GA SA313198374GA SA313198431GA SA313198460GA SA313198585GA SA313198591GA SA313198595GA SA313198596GA SA313198599GA SA313198609GA SA313198625GA SA313198637GA SA313198645GA SA313198650GA SA313198686GA SA313198707GA SA313208779GA SA313208788GA SA313208826GA SA313208847GA SA313208900GA SA313208904GA SA313208905GA SA313208910GA SA313208913GA SA313208921GA SA313208924GA SA313208925GA SA313208933GA SA313208939GA SA313208940GA SA313208943GA SA313208944GA SA313208977GA SA313208981GA SA313208983GA SA313208984GA SA313208985GA SA313208986GA SA313208987GA SA313208988GA SA313208989GA SA313208993GA SA313208995GA SA313209001GA SA313209002GA SA313209003GA SA313209004GA SA313209005GA SA313209006GA SA313209009GA SA313209010GA SA313219021GA SA313219024GA SA313219062GA SA313219068GA SA313219076GA SA313219149GA SA313219156GA SA313229319GA SA313229324GA SA313229326GA SA313229330GA SA313229331GA SA313229332GA SA313229333GA SA313229336GA SA313229373GA SA313229403GA SA313229408GA SA313229412GA SA313229423GA SA313229424GA SA313229425GA SA313229428GA SA313229430GA SA313229436GA SA313229437GA SA313229443GA SA313229444GA SA313229445GA SA313229449GA SA313229451GA SA313229489GA SA313229490GA SA313229491GA SA313229494GA SA313229495GA SA313229496GA SA313229497GA SA313229500GA SA313229501GA SA313229502GA SA313229503GA SA313229504GA SA313229505GA SA313229506GA SA313229512GA SA313229514GA SA313229516GA SA313229519GA SA313229550GA SA313229552GA SA313229553GA SA313229558GA SA313229559GA SA313229561GA SA313229562GA SA313229564GA SA313229565GA SA313229566GA SA313229567GA SA313229568GA SA313229570GA SA313229571GA SA313229576GA SA313229577GA SA313229791GA SA313229793GA SA313229795GA SA313229798GA SA313229799GA SA313229800GA SA313229801GA SA313229802GA SA313229803GA SA313229806GA SA313229809GA SA313229811GA SA313229813GA SA313229842GA SA313230477GA SA313230576GA SA313230608GA SA313230622GA SA313230635GA SA313230638GA SA313230639GA SA313230640GA SA313230641GA SA313230643GA SA313230647GA SA313230648GA SA313230650GA SA313230654GA SA313230660GA SA313230667GA SA313230669GA SA313230670GA SA313230683GA SA313230684GA SA313230691GA SA313230692GA SA313230694GA SA313230695GA SA313230696GA SA313230700GA SA313230701GA SA313230702GA SA313230705GA SA313230708GA SA313230709GA SA313230710GA SA313240904GA SA313240912GA SA313240913GA SA313240914GA SA313240921GA SA313240922GA SA313240923GA SA313240925GA SA313240932GA SA313240935GA SA313240938GA SA313240945GA SA313240949GA SA313240950GA SA313240952GA SA313240961GA SA313240962GA SA313240964GA SA313240965GA SA313240967GA SA313240971GA SA313240972GA SA313240977GA SA313240980GA SA313240982GA SA313240984GA SA313240987GA SA313261730GA SA313261750GA SA313261874GA SA313261889GA SA313271970GA SA313272047GA SA313272050GA SA313272128GA SA313272148GA SA313272150GA SA313272161GA SA313272186GA SA313272192GA SA313272274GA SA313272279GA SA313272286GA SA313272360GA SA313272398GA SA313282859GA SA313292944GA SA313292948GA SA313293023GA SA313293024GA SA313293027GA SA313293032GA SA313293040GA SA313293041GA SA313293042GA SA313293050GA SA313293062GA SA313293084GA SA313293085GA SA313293121GA SA313293131GA SA313293144GA SA313293155GA SA313293161GA SA313293164GA SA313293189GA SA313293193GA SA313293201GA SA313293208GA SA313293220GA SA313293223GA SA313293225GA SA313293230GA SA313293235GA SA313293241GA SA313303941GA SA313314463GA SA313314521GA SA313314535GA SA313314538GA SA313314540GA SA313314545GA SA313314557GA SA313314558GA SA313314559GA SA313314563GA SA313314564GA SA313314565GA SA313314567GA SA313314569GA SA313314571GA SA313314573GA SA313314576GA SA313314577GA SA313314578GA SA313314581GA SA313314582GA SA313314796GA SA313314805GA SA313314814GA SA313314825GA SA313314831GA SA313314832GA SA313314839GA SA313314840GA SA313314844GA SA313314849GA SA313314850GA SA313314853GA SA313314855GA SA313314857GA SA313314859GA SA313314861GA SA313314864GA SA313314866GA SA313314869GA SA313314870GA SA313314875GA SA313314879GA SA313314880GA SA313314882GA SA313314885GA SA313314886GA SA313314889GA SA313314890GA SA313314893GA SA313314894GA SA313335015GA SA313335019GA SA313335023GA SA313335033GA SA313335155GA SA313335172GA SA313335179GA SA313335189GA SA313335193GA SA313335206GA SA313335233GA SA313335241GA SA313345576GA SA313345622GA SA313419743GA SA313419751GA SA313419756GA SA313419762GA SA313419763GA SA313420101GA SA313420106GA SA313420116GA SA313420141GA SA313420148GA SA313420246GA SA313420310GA SA313420317GA SA313420318GA SA313420330GA SA313420332GA SA313420338GA SA313420341GA SA313420344GA SA313420345GA SA313420349GA SA313420356GA SA313420362GA SA313420367GA SA313420368GA SA313420379GA SA313420380GA SA313420387GA SA313420388GA SA313420399GA SA313420401GA SA313420407GA SA313420409GA SA313420412GA SA313420413GA SA313420416GA SA313420419GA SA313420420GA SA313420422GA SA313420424GA SA313420425GA SA313420426GA SA313420429GA SA313420435GA SA313420437GA SA313420439GA SA313420441GA SA313420445GA SA313420446GA SA313420447GA SA313420449GA SA313420453GA SA313420454GA SA313420455GA SA313420462GA SA313420463GA SA313420464GA SA313420466GA SA313420467GA SA313420469GA SA313420474GA SA313420478GA SA313420481GA SA313420484GA SA313420485GA SA313420486GA SA313420489GA SA313420490GA SA313420491GA SA313420492GA SA313420493GA SA313420494GA SA313420496GA SA313420497GA SA313420499GA SA313420500GA SA313420503GA SA313420510GA SA313420511GA SA313420512GA SA313420515GA SA313420516GA SA313420517GA SA313420519GA SA313420521GA SA313420524GA SA313420525GA SA313420528GA SA313420530GA SA313420532GA SA313420535GA SA313420538GA SA313420539GA SA313420541GA SA313420543GA SA313420547GA SA313420549GA SA313420552GA SA313420553GA SA313420554GA SA313420560GA SA313420561GA SA313420563GA SA313420567GA SA313420573GA SA313420580GA SA313420598GA SA313420601GA SA313420606GA SA313420608GA SA313420613GA SA313430636GA SA313430637GA SA313430644GA SA313430651GA SA313431193GA SA313431210GA SA313431241GA SA313441303GA SA313441322GA SA313441328GA SA313441346GA SA313441348GA SA313441351GA SA313441353GA SA313441355GA SA313441360GA SA313441366GA SA313441536GA SA313441540GA SA313441541GA SA313441542GA SA313441544GA SA313441548GA SA313441549GA SA313441691GA SA313441695GA SA313441703GA SA313451709GA SA313451717GA SA313451718GA SA313451719GA SA313451721GA SA313451781GA SA313451782GA SA313451783GA SA313451905GA SA313451908GA SA313451928GA SA313451935GA SA313451955GA SA313451957GA SA313451969GA SA313451998GA SA313452018GA SA313452020GA SA313452022GA SA313452023GA SA313452030GA SA313452033GA SA313452059GA SA313452061GA SA313452063GA SA313452065GA SA313452073GA SA313452087GA SA313452113GA SA313452138GA SA313462415GA SA313462423GA SA313482865GA SA313492956GA SA313523970GA SA314004103GA SA314004109GA SA314004110GA SA314004116GA SA314004118GA SA314004119GA SA314004122GA SA314004124GA SA314004129GA SA314004183GA SA314004184GA SA314004186GA SA314004191GA SA314014325GA SA314014336GA SA314014340GA SA314014343GA SA314014357GA SA314014363GA SA314014365GA SA314014387GA SA314014388GA SA314014399GA SA314014443GA SA314014448GA SA314014451GA SA314014453GA SA314014467GA SA314014473GA SA314014548GA SA314014608GA SA314014754GA SA314024965GA SA314025015GA SA314025047GA SA314025068GA SA314025070GA SA314025072GA SA314025079GA SA314025087GA SA314025090GA SA314025093GA SA314055555GA SA314055576GA SA314075986GA SA314075996GA SA314076004GA SA314076010GA SA314076011GA SA314076013GA SA314076015GA SA314076016GA SA314076023GA SA314076026GA SA314076031GA SA314076068GA SA314076089GA SA314076092GA SA314076108GA SA314076109GA SA314076110GA SA314076116GA SA314086503GA SA314086506GA SA314086511GA SA314086521GA SA314097354GA SA314118958GA SA314118999GA SA314119113GA SA314119130GA SA314129182GA SA314129190GA SA314129212GA SA314129217GA SA314139390GA SA314139397GA SA314139402GA SA314139458GA SA314139486GA SA314139490GA SA314139537GA SA314139557GA SA314139561GA SA314139569GA SA314139578GA SA314139581GA SA314139591GA SA314149757GA SA314160006GA SA314160014GA SA314160044GA SA314160046GA SA314160048GA SA314169917GA SA314170183GA SA314170185GA SA314170195GA SA314170197GA SA314170199GA SA314170206GA SA314170209GA SA314170213GA SA314170219GA SA314170221GA SA314170222GA SA314170224GA SA314170226GA SA314170227GA SA314170228GA SA314170229GA SA314170233GA SA314170235GA SA314170236GA SA314170237GA SA314170239GA SA314170272GA SA314170275GA SA314170276GA SA314170283GA SA314170284GA SA314170289GA SA314170290GA SA314170291GA SA314170299GA SA314170300GA SA314170302GA SA314170303GA SA314170304GA SA314170305GA SA314170307GA SA314170308GA SA314170310GA SA314170342GA SA314170343GA SA314170344GA SA314170345GA SA314170349GA SA314170351GA SA314170352GA SA314170353GA SA314170357GA SA314170358GA SA314170359GA SA314170366GA SA314170409GA SA314170423GA SA314170432GA SA314170435GA SA314170438GA SA314170442GA SA314170444GA SA314170445GA SA314170446GA SA314170462GA SA314170463GA SA314170468GA SA314170471GA SA314180474GA SA314180482GA SA314180483GA SA314180485GA SA314180486GA SA314180487GA SA314180488GA SA314180491GA SA314180495GA SA314180496GA SA314180499GA SA314180502GA SA314180505GA SA314180507GA SA314180517GA SA314180526GA SA314180532GA SA314180536GA SA314180540GA SA314180542GA SA314180550GA SA314180553GA SA314180570GA SA314180574GA SA314180578GA SA314180579GA SA314180580GA SA314180581GA SA314180582GA SA314180586GA SA314180587GA SA314180588GA SA314180593GA SA314180594GA SA314180595GA SA314180597GA SA314180598GA SA314180600GA SA314180602GA SA314180605GA SA314180608GA SA314180617GA SA314180619GA SA314180640GA SA314180655GA SA314180656GA SA314180659GA SA314180660GA SA314180661GA SA314180663GA SA314190881GA SA314191143GA SA314191162GA SA314201381GA SA314201485GA SA314201490GA SA314201498GA SA314201513GA SA314201515GA SA314201591GA SA314201616GA SA314201617GA SA314212014GA SA314212022GA SA314212023GA SA314212030GA SA314212032GA SA314212041GA SA314212045GA SA314212052GA SA314212095GA SA314212097GA SA314212098GA SA314212099GA SA314212100GA SA314212101GA SA314212103GA SA314212106GA SA314212107GA SA314212111GA SA314212115GA SA314212118GA SA314212119GA SA314212120GA SA314212121GA SA314212122GA SA314212123GA SA314212184GA SA314212192GA SA314212202GA SA314212214GA SA314212222GA SA314212232GA SA314212238GA SA314212246GA SA314222475GA SA314222513GA SA314253498GA SA314253582GA SA314253583GA SA314253594GA SA314253615GA SA314253616GA SA314253648GA SA314253651GA SA314253663GA SA314253668GA SA314253677GA SA314253685GA SA314253686GA SA314253713GA SA314253719GA SA314253729GA SA314253733GA SA314253738GA SA314273905GA SA314273914GA SA314273919GA SA314273920GA SA314273937GA SA314273947GA SA314273957GA SA314273964GA SA314273966GA SA314273967GA SA314273971GA SA314273975GA SA314273978GA SA314273979GA SA314273982GA SA314273989GA SA314284521GA SA314284647GA SA314284743GA SA314284748GA SA314284759GA SA314294827GA SA314294837GA SA314294859GA SA314294878GA SA314295108GA SA314295111GA SA314295114GA SA314295119GA SA314295127GA SA314295128GA SA314076140GA SA309032112GA SA309042156GA SA309042158GA SA309042166GA SA309042204GA SA309042209GA SA309042212GA SA309527940GA SA311090830GA SA311153186GA SA311195136GA SA311236331GA SA311246391GA SA311246394GA SA311246396GA SA311246401GA SA311246402GA SA311246403GA SA311246405GA SA311246406GA SA311246407GA SA311246413GA SA311246414GA SA311246415GA SA311246428GA SA311246431GA SA311246434GA SA311246448GA SA311297331GA SA311297332GA SA311307420GA SA311307433GA SA311307442GA SA311428096GA SA311428105GA SA311428112GA SA311428115GA SA311428119GA SA311428121GA SA311428124GA SA311428125GA SA311428129GA SA311428131GA SA311428134GA SA311428142GA SA311428144GA SA311428145GA SA311428148GA SA311428154GA SA311428158GA SA311428159GA SA312049234GA SA312386503GA SA313116651GA SA313282876GA SA313282882GA SA313292903GA SA313292940GA SA313292943GA SA313292949GA SA313314807GA SA313314808GA SA313314816GA SA313314819GA SA313314820GA SA313314823GA SA313335142GA SA313335149GA SA313335377GA SA313335390GA SA313335391GA SA313335400GA SA313335412GA SA313335426GA SA313335428GA SA313335435GA SA313335439GA SA313335442GA SA313335445GA SA313335446GA SA313335449GA SA313335451GA SA313335454GA SA313335455GA SA313335458GA SA313335495GA SA313335499GA SA313335506GA SA313335513GA SA313452079GA SA313452151GA SA314086380GA SA314119109GA SA314211945GA SA312069842GA SA312070063GA SA312070088GA SA312070137GA SA312111527GA SA312111660GA SA312121727GA SA312121736GA SA312355504GA SA312355509GA SA312355518GA SA312355521GA SA312457704GA SA312457805GA SA312458225GA SA312468543GA SA312468573GA SA312468600GA SA312468676GA SA312468802GA SA312468815GA SA312468821GA SA312468842GA SA312468854GA SA312468856GA SA312468866GA SA312468871GA SA312468872GA SA312469232GA SA312469421GA SA312469444GA SA312469500GA SA312470086GA SA312479577GA SA312479868GA SA312480235GA SA312480349GA SA312490874GA SA312491322GA SA312491335GA SA312491347GA SA312491349GA SA312491358GA SA312491371GA SA312491455GA SA312491466GA SA312491478GA SA312491489GA SA312491497GA SA312491501GA SA312491515GA SA312491521GA SA312491528GA SA312491530GA SA312491534GA SA312491544GA SA312491545GA SA312501892GA SA312522396GA SA312522596GA SA312522603GA SA313012774GA SA313012912GA SA313012956GA SA313012968GA SA313012979GA SA313013012GA SA313013080GA SA313013096GA SA313023188GA SA313023191GA SA313023196GA SA313023203GA SA313023213GA SA313023219GA SA313023225GA SA313023238GA SA313023318GA SA313023319GA SA313023320GA SA313023321GA SA313023324GA SA313023325GA SA313023326GA SA313023327GA SA313023328GA SA313023329GA SA313023330GA SA313023331GA SA313023333GA SA313023334GA SA313043939GA SA313043982GA SA313043986GA SA313043992GA SA313043993GA SA313043995GA SA313044003GA SA313044124GA SA313044126GA SA313044127GA SA313044128GA SA313044130GA SA313064165GA SA313064204GA SA313064213GA SA313064215GA SA313064219GA SA313064221GA SA313064224GA SA313064227GA SA313064228GA SA313064264GA SA313064286GA SA313064289GA SA313064291GA SA313064304GA SA313064308GA SA313064312GA SA313064313GA SA313064317GA SA313064328GA SA313064329GA SA313064332GA SA313064335GA SA313064384GA SA313064388GA SA313064389GA SA313064390GA SA313064395GA SA313064396GA SA313064397GA SA313064398GA SA313064399GA SA313064400GA SA313064404GA SA313064405GA SA313064406GA SA313064407GA SA313064408GA SA313064409GA SA313064411GA SA313064412GA SA313064413GA SA313064414GA SA313064416GA SA313064417GA SA313064418GA SA313064419GA SA313064420GA SA313064421GA SA313064422GA SA313064423GA SA313064424GA SA313095299GA SA313095302GA SA313095307GA SA313095550GA SA313095604GA SA313106065GA SA313106291GA SA313116790GA SA313116954GA SA313117020GA SA313117210GA SA313127211GA SA313127218GA SA313127232GA SA313127235GA SA313127238GA SA313127240GA SA313127433GA SA313127438GA SA313127441GA SA313127442GA SA313127447GA SA313127453GA SA313127455GA SA313127457GA SA313127458GA SA313127467GA SA313127468GA SA313127682GA SA313127712GA SA313127727GA SA313127736GA SA313137779GA |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact | Technical Support 800-558-7044 |
Manufacturer Reason for Recall | GE Healthcare has recently become aware of a potential safety issue to the ECG calculations following a disconnect/reconnect cycle with the Patient Data Module, when used with the Carescape Bx50 monitors.
If the PDM is quickly disconnected and reconnected (a less than 10 second cycle) from a Carescape Bx50 monitor following a Patient Discharge or during active monitoring, the ECG waveform an |
FDA Determined Cause 2 | Software design |
Action | Consignees were sent on 9/19/2014 a GE Healthcare "Urgent Medical Device Correction" letter Ref# 336107 dated September 18, 2014.
The letter was addressed to Director of Biomedical Engineering, Chief of Nursing and Healthcare Administrator / Risk Manager. The letter described the Safety Issue, Safety Instructions , Affected Product Details, Product Correction and Contact Information. For questions they can contact Technical Support at 1-800-558-70444 or your local Service Representative. |
Quantity in Commerce | 68,452 (39,505 units US, 24,711 units OUS, 4,236 units unidentified/unknown) |
Distribution | Worldwide Distribution: US (nationwide) including DC, PR, Guantanamo except DE, HI, ND, SD & WY; and countries of: SOUTH AFRICA, KOREA, UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, TRINIDAD TOBAGO, THAILAND, TANZANIA, TAIWAN, SWITZERLAND, SWEDEN, SPAIN, SLOVAKIA, SINGAPORE, SAUDI ARABIA, RUSSIA, QATAR, BRASIL, PORTUGAL, POLAND, PHILIPPINES, PERU, PANAMA, PARAGUAY, OMAN, NEW ZEALAND, NORWAY, NETHERLAND, NEPAL, MYANMAR, MEXICO, MOROCCO, MALDOVA, MAURITIUS, MALAYSIA, LITHUANIA, LEBANON, BAHRAIN, KUWAIT, REPUBLIC OF KOREA, JAPAN, JORDAN, JAMAICA, ITALY, ISRAEL, IRELAND, INDONESIA, INDIA, ICELAND, HUNGARY, HONG KONG, GUATEMALA, GREECE, GERMANY, FRANCE, FINLAND, ESTONIA, EGYPT,ECUADOR, DENMARK, CECH REPUBLIC, COSTA RICA, COLOMBIA,, CHINA, CANADA, BULGARIA, BELGIUM BANGLADESH, AUSTRIA, AUSTRALIA, and ARGENTINA. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MHX
|
|
|
|