 Class 1 Device Recall The Plum A Infusion Systemsee related information
Date Initiated by FirmJuly 02, 2014
Date PostedFebruary 23, 2015
Recall Status1 Terminated 3 on May 24, 2017
Recall NumberZ-1072-2015
Recall Event ID 69336
510(K)NumberK070398 
Product Classification Pump, infusion - Product Code FRN
ProductThe Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one line out. The pump can be used for standard, piggyback, or concurrent delivery. Therapy modes include: Standard Infusions, Multistep Programming, Loading Dose, and Dose Calculation. The Plum A+ is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual.
Code Information 1) List Number: 12391; Serial Numbers: 13840011 , 13840048 , 13840153 , 13840167 , 13840498 , 13840591 , 13840629 , 13840640 , 13840943 , 13841068 , 13841130 , 13841271 , 13842284 , 13842383 , 13845291 , 13845368 , 13845404 , 13845429 , 13845450 , 13845495 , 13845510 , 13845560 , 13845599 , 13845686 , 13845697 , 13845701 , 13846352 , 13846723 , 13846727 , 13846788 , 13846801 , 13846803 , 13846804 , 13846807 , 13846826 , 13846827 , 13846847 , 13846852 , 13846862 , 13846877 , 13846878 , 13846902 , 13846971 , 13847066 , 13847663 , 13847677 , 13847698 , 13847702 , 13847707 , 13847719 , 13847777 , 13847802 , 13847804 , 13847805 , 13847812 , 13847815 , 13847836 , 13847838 , 13847840 , 13847841 , 13847847 , 13847873 , 13847887 , 13847893 , 13847898 , 13847901 , 13847905 , 13847907 , 13847916 , 13847942 , 13847947 , 13847956 , 13847960 , 13847965 , 13847973 , 13847984 , 13847988 , 13848031 , 13848043 , 13848044 , 13848045 , 13848046 , 13848047 , 13848079 , 13848082 , 13848114 , 13848128 , 13848132 , 13848140 , 13848141 , 13848159 , 13848166 , 13848179 , 13848181 , 13848184 , 13848190 , 13848195 , 13848201 , 13848203 , 13848204 , 13848207 , 13848210 , 13848220 , 13848222 , 13848244 , 13848249 , 13848250 , 13848264 , 13848269 , 13848271 , 13848278 , 13848306 , 13848314 , 13848322 , 13848323 , 13848328 , 13848337 , 13848340 , 13848355 , 13848362 , 13848368 , 13848379 , 13848394 , 13848410 , 13848431 , 13848433 , 13848438 , 13848452 , 13848456 , 13848482 , 13848514 , 13848528 , 13848556 , 13848639 , 13848661 , 13848708 , 13848812 , 13849002 , 13850656 , 13850659 , 13850674 , 13850698 , 13850721 , 13850952 , 13850967 , 13850977 , 13850979 , 13850980 , 13850983 , 13850998 , 13851001 , 13851002 , 13851011 , 13851014 , 13851048 , 13851055 , 13851058 , 13851088 , 13851095 , 13851111 , 13851144 , 13851252 , 13851900 , 13851914 , 13852452 , 13852877 , 13852889 , 13852907 , 13852952 , 13852959 , 13852963 , 13853061 , 13853147 , 13853325 , 13853349 , 13853435 , 13853470 , 13853584 , 13854149 , 13854212 , 13854219 , 13854276 , 13854277 , 13854295 , 13840558 , 13840638 , 13840834 , 13841113 , 13841623 , 13842119 , 13842220 , 13842298 , 13847972 , 13849651 , 13850661 , 13851038 , 13854167 , 13854487 , 14100242 , 14102019 , 14102288 , 14102308 , 14102334 , 14102340 , 14102342 , 14102347 , 14102359 , 14102497 , 14102677 , 14103117 , 14103175 , 14103738 , 14104011 , 14104012 , 14104027 , 14104553 , 14120009 , 14120010 , 14120020 , 14120030 , 14120034 , 14120061 , 14120063 , 14120082 , 14120441 , 14120526 , 14120530 , 14120580 , 14120599 , 14120600 , 14120601 , 14120946 , 14120963 , 14120985 , 14120992 , 14121009 , 14121011 , 14121036 , 14121045 , 14121080 , 14123158 , 14123239 , 14123254 , 14123266 , 14123270 , 14123272 , 14123273 , 14123376 , 14123380 , 14123386 , 14124013 , 14124096 , 14124100 , 14124102 , 14124103 , 14124113 , 14124115 , 14124120 , 14124136 , 14124157 , 14124180 , 14124223 , 14124245 , 14124251 , 14124254 , 14124257 , 14124261 , 14124280 , 14124306 , 14124309 , 14124315 , 14124317 , 14124318 , 14124319 , 14124327 , 14124336 , 14124338 , 14124344 , 14124345 , 14124350 , 14124370 , 14124372 , 14124377 , 14124378 , 14124379 , 14124382 , 14124383 , 14124386 , 14124388 , 14124389 , 14124390 , 14124392 , 14124406 , 14124407 , 14124408 , 14124409 , 14124424 , 14124428 , 14124433 , 14124435 , 14124444 , 14124446 , 14124451 , 14124452 , 14124453 , 14124455 , 14124456 , 14124462 , 14124475 , 14124486 , 14124487 , 14124488 , 14124489 , 14124493 , 14124494 , 14124505 , 14124507 , 14124509 , 14124512 , 14124536 , 14124542 , 14124543 , 14124806 , 14124864 , 14124871 , 14125151 , 14125197 , 14125268 , 14125360 , 14125969 , 14126165 , 14127608 , 14128306 , 14128562 , 14128567 , 14128628 , 14128656 , 14129164 , 13849118 , 13849119 , 13850020 , 13850095 , 13850096 , 13850127 , 13850412 , 13850751 , 13851857 , 13852152 , 13852274 , 13852354 , 13852409 , 13852444 , 13852462 , 13852463 , 13852475 , 13852566 , 13852575 , 13852594 , 13852603 , 13852733 , 13852786 , 13852891 , 13853075 , 13853081 , 13853102 , 13853140 , 13853156 , 13853178 , 13853190 , 13853218 , 13853244 , 13853251 , 13853285 , 13853287 , 13853328 , 13853452 , 13853483 , 13853489 , 13853516 , 13853537 , 13853544 , 13853547 , 13853549 , 13853550 , 13853563 , 13853568 , 13853575 , 13853594 , 13853600 , 13853604 , 13853610 , 13853656 , 13853657 , 13853659 , 13853660 , 13853674 , 13853677 , 13853681 , 13853685 , 13853686 , 13853688 , 13853693 , 13853698 , 13853701 , 13853703 , 13853705 , 13853711 , 13853718 , 13853720 , 13853721 , 13853724 , 13853727 , 13853731 , 13853737 , 13853742 , 13853743 , 13853745 , 13853748 , 13853752 , 13853756 , 13853763 , 13853772 , 13853773 , 13853775 , 13853781 , 13853784 , 13853785 , 13853786 , 13853790 , 13853791 , 13853793 , 13853801 , 13853802 , 13853803 , 13853809 , 13853810 , 13853815 , 13853816 , 13853820 , 13853821 , 13853824 , 13853831 , 13853834 , 13853836 , 13853838 , 13853839 , 13853841 , 13853845 , 13853862 , 13853872 , 13853874 , 13853878 , 13853901 , 13853933 , 13853941 , 13853946 , 13853951 , 13853970 , 13853971 , 13853977 , 13853984 , 13853997 , 13854004 , 13854005 , 13854011 , 13854023 , 13854032 , 13854033 , 13854055 , 13854068 , 13854071 , 13854089 , 13854094 , 13854108 , 13854109 , 13854115 , 13854121 , 13854123 , 13854125 , 13854214 , 13854220 , 13854237 , 13854658 , 13854721 , 13854739 , 13854744 , 13854773 , 13854784 , 13854786 , 13854809 , 13854847 , 13854865 , 13854883 , 13855287 , 13855566 , 13855568 , 13855569 , 13855586 , 13855591 , 13855595 , 13855600 , 13855607 , 13855628 , 13855635 , 13855640 , 13855703 , 13855713 , 13855767 , 13855793 , 13855794 , 13855796 , 13855808 , 13855813 , 13855820 , 13855832 , 13855837 , 13855855 , 13855859 , 13855870 , 13855875 , 13855880 , 13855881 , 13855883 , 13855885 , 13855901 , 13855911 , 13855932 , 13855935 , 13855937 , 13855946 , 13855952 , 13855953 , 13856152 , 13856477 , 13857148 , 13857190 , 13857330 , 13857542 , 13857575 , 13857584 , 13857587 , 13857589 , 13857590 , 13857732 , 13857743 , 13857744 , 13857753 , 13857809 , 13857814 , 13857815 , 13857818 , 13857819 , 13857820 , 13857821 , 13857822 , 13857824 , 13857825 , 13857826 , 13857827 , 13857828 , 13857829 , 13857831 , 13858282 , 13858288 , 13858418 , 13858436 , 13858473 , 13858484 , 13858487 , 13858492 , 13858494 , 13858495 , 13858496 , 13858498 , 13858499 , 13858500 , 13858509 , 13858510 , 13858514 , 13858535 , 13858538 , 13858540 , 13858542 , 13858543 , 13858545 , 13858550 , 13858553 , 13858555 , 13858558 , 13858565 , 13858576 , 13858583 , 13858585 , 13858597 , 13858601 , 13858602 , 13858604 , 13858605 , 13858608 , 13858614 , 13858616 , 13858617 , 13858618 , 13858619 , 13858622 , 13858623 , 13858625 , 13858629 , 13858630 , 13858632 , 13858633 , 13858634 , 13858635 , 13858636 , 13858638 , 13858640 , 13858642 , 13858643 , 13858644 , 13858645 , 13858646 , 13858648 , 13858650 , 13858653 , 13858657 , 13858660 , 13858665 , 13858666 , 13858667 , 13858669 , 13858670 , 13858671 , 13858673 , 13858674 , 13858676 , 13858700 , 13858706 , 13858708 , 13858710 , 13858713 , 13858714 , 13858718 , 13858722 , 13858724 , 13858725 , 13858726 , 13858727 , 13858730 , 13858734 , 13858736 , 13858747 , 13858749 , 13858751 , 13858754 , 13858759 , 13858764 , 13858773 , 13858774 , 13858777 , 13858782 , 13858783 , 13858785 , 13858786 , 13858787 , 13858788 , 13858792 , 13858793 , 13858798 , 13858801 , 13858805 , 13858810 , 13858816 , 13858817 , 13858821 , 13858822 , 13858823 , 13858825 , 13858827 , 13858828 , 13858829 , 13858830 , 13858831 , 13858832 , 13858834 , 13858835 , 13858836 , 13858843 , 13858852 , 13858855 , 13858856 , 13858860 , 13858862 , 13858863 , 13858864 , 13858869 , 13858870 , 13858876 , 13858878 , 13858884 , 13858891 , 13858901 , 13858908 , 13858914 , 13858915 , 13858917 , 13858919 , 13858923 , 13858924 , 13858925 , 13858926 , 13858929 , 13858937 , 13858942 , 13858944 , 13858950 , 13858951 , 13858960 , 13858963 , 13858968 , 13858970 , 13858971 , 13858973 , 13858974 , 13858978 , 13858982 , 13858984 , 13858987 , 13858991 , 13858992 , 13859003 , 13859004 , 13859013 , 13859017 , 13859023 , 13859027 , 13859029 , 13859031 , 13859034 , 13859035 , 13859037 , 13859038 , 13859041 , 13859042 , 13859043 , 13859044 , 13859045 , 13859046 , 13859057 , 13859080 , 13859087 , 13859092 , 13859093 , 13859100 , 13859111 , 13859112 , 13859113 , 13859116 , 13859118 , 13859120 , 13859123 , 13859124 , 13859126 , 13859132 , 13859138 , 13859141 , 13859147 , 13859153 , 13859154 , 13859157 , 13859158 , 13859159 , 13859162 , 13859164 , 13859166 , 13859168 , 13859171 , 13859175 , 13859178 , 13859181 , 13859183 , 13859194 , 13859195 , 13859209 , 13859213 , 13859214 , 13859250 , 13859251 , 13859263 , 13859264 , 13859268 , 13859269 , 13859270 , 13859283 , 13859288 , 13859292 , 13859294 , 13859299 , 13859300 , 13859301 , 13859307 , 13859308 , 13859312 , 13859324 , 13859325 , 13859358 , 13859371 , 13859372 , 13859374 , 13859375 , 13859377 , 13859380 , 13859382 , 13859383 , 13859396 , 13859401 , 13859402 , 13859404 , 13859408 , 13859419 , 13859425 , 13859430 , 13859431 , 13859441 , 13859443 , 13859444 , 13859447 , 13859460 , 13859471 , 13859478 , 13859493 , 13859718 , 13859720 , 13859739 , 13859744 , 13859752 , 13859768 , 13859772 , 13859773 , 13859777 , 13859781 , 13859802 , 13859803 , 13859804 , 13859807 , 13859815 , 13859818 , 13859822 , 13859824 , 13859828 , 13859829 , 13859830 , 13859833 , 13859834 , 13859836 , 13859838 , 13859839 , 13859841 , 13859842 , 13859843 , 13859848 , 13859850 , 13859851 , 13859852 , 13859854 , 13859858 , 13859859 , 13859861 , 13859863 , 13859864 , 13859865 , 13859870 , 13859871 , 13859872 , 13859873 , 13859875 , 13859876 , 13859887 , 13859888 , 13859889 , 13859890 , 13859893 , 13859895 , 13859897 , 13859899 , 13859901 , 13859902 , 13859903 , 13859905 , 13859910 , 13859911 , 13859914 , 13859915 , 13859917 , 13859920 , 13859921 , 13859922 , 13859923 , 13859924 , 13859925 , 13859928 , 13859929 , 13859930 , 13859940 , 13859941 , 13859943 , 13859949 , 13859952 , 13859954 , 13859955 , 13859957 , 13859959 , 13859961 , 13859963 , 13859965 , 13859966 , 13859979 , 13859980 , 13859982 , 13859990 , 13859995 , 13859998 , 13860000 , 13860009 , 13860010 , 13860020 , 13860049 , 13860065 , 13860092 , 13860094 , 13860104 , 13860108 , 13860116 , 13860117 , 13860118 , 13860119 , 13860123 , 13860125 , 13860135 , 13860136 , 13860139 , 13860141 , 13860144 , 13860161 , 13860164 , 13860165 , 13860166 , 13860168 , 13860171 , 13860173 , 13860176 , 13860177 , 13860178 , 13860179 , 13860184 , 13860185 , 13860188 , 13860192 , 13860193 , 13860195 , 13860197 , 13860203 , 13860206 , 13860213 , 13860218 , 13860273 , 13860336 , 13860408 , 13860419 , 13860429 , 13860430 , 13860433 , 13860510 , 13860512 , 13860513 , 13860551 , 13860555 , 13860556 , 13862260 , 13862995 , 13863193 , 13866591 , 13866658 , 13866840 , 13867212 , 13867401 , 13867406 , 13867414 , 13867440 , 13867791 , 13867794 , 13867812 , 13867821 , 13867825 , 13868061 , 13868456 , 13872381 , 13872417 , 13872472 , 13872582 , 13872701 , 13872702 , 13872726 , 13872960 , 13872979 , 13872983 , 13872997 , 13874910 , 13875413 , 13875414 , 13875435 , 13878156 , 13878188 , 13878189 , 13878194 , 13878195 , 13878196 , 13878197 , 13878203 , 13878209 , 13878372 , 13878420 , 13878421 , 13878430 , 13878433 , 13878442 , 13878448 , 13878449 , 13878472 , 13878486 , 13878496 , 13878506 , 13879080 , 13879087 , 13880737 , 13840139 , 13840390 , 13840606 , 13840688 , 13841334 , 13841349 , 13841896 , 13841965 , 13842250 , 13842255 , 13842873 , 13843106 , 13843991 , 13845035 , 13845087 , 13845144 , 13845207 , 13846328 , 13846531 , 13846546 , 13846596 , 13846816 , 13846819 , 13846824 , 13846838 , 13846859 , 13846897 , 13847076 , 13847612 , 13847745 , 13847782 , 13847844 , 13847927 , 13848013 , 13848022 , 13848158 , 13848276 , 13848304 , 13848324 , 13848466 , 13848545 , 13848610 , 13848893 , 13849617 , 13850073 , 13850572 , 13850654 , 13850995 , 13851176 , 13851228 , 13851722 , 13851729 , 13851769 , 13851920 , 13852121 , 13852324 , 13852730 , 13852910 , 13852948 , 13852974 , 13853007 , 13853037 , 13853046 , 13853129 , 13853169 , 13853252 , 13853436 , 13853517 , 13853876 , 13854060 , 13854230 , 13854257 , 13854274 , 13854475 , 13856210 , 13856540 , 13856627 , 13856687 , 13857007 , 13857018 , 13857110 , 13857147 , 13857178 , 13857239 , 13857283 , 13857817 , 13858077 , 13858551 , 13858557 , 13858584 , 13858606 , 13858621 , 13858626 , 13858631 , 13858668 , 13858712 , 13858717 , 13858728 , 13858738 , 13858945 , 13858979 , 13859011 , 13859032 , 13859040 , 13859110 , 13859170 , 13859197 , 13859293 , 13859328 , 13859376 , 13859429 , 13859440 , 13859729 , 13859769 , 13859805 , 13859819 , 13859827 , 13859832 , 13859857 , 13859878 , 13859926 , 13859927 , 13859944 , 13859951 , 13859971 , 13859999 , 13860024 , 13860262 , 13860264 , 13860381 , 13860567 , 13860933 , 13863401 , 13863589 , 13864071 , 13864736 , 13864989 , 13865199 , 13865344 , 13865441 , 13865527 , 13866250 , 13866368 , 13866396 , 13866399 , 13866455 , 13866463 , 13866912 , 13867797 , 13868025 , 13868359 , 13868946 , 13869182 , 13869282 , 13869453 , 13869729 , 13870421 , 13872736 , 13872965 , 13875195 , 13877576 , 13878206 , 13878207 , 13878246 , 13878251 , 13878264 , 13878303 , 13878396 , 13878414 , 13878425 , 13878437 , 13878440 , 13878446 , 13878453 , 13878468 , 13878494 , 13878809 , 13878998 , 14101362 , 14101617 , 14101658 , 14102071 , 14102132 , 14102314 , 14102690 , 14102764 , 14102774 , 14102827 , 14102911 , 14102938 , 14102947 , 14102991 , 14103047 , 14103058 , 14103097 , 14103295 , 14103339 , 14103360 , 14104182 , 14104296 , 14104354 , 14104368 , 14120234 , 14120346 , 14121021 , 14121063 , 14122916 , 14123869 , 14124060 , 14124137 , 14124196 , 14124434 , 14124445 , 14124669 , 14124744 , 14124769 , 14125056 , 14125683 , 14125790 , 14125878 , 14125890 , 14125900 , 14125927 , 14126132 , 14126285 , 14126369 , 14126505 , 14126603 , 14126799 , 14127000 , 14127020 , 14127138 , 14127530 , 14127541 , 14127695 , 14127789 , 14128252 , 14128316 , 14129442 , 14129445 , 14129641 , 14129979 , 14130266 , 14130386 , 14130529 , 14130542 , 14130546 , 14130715 , 14130800 , 14130816 , 14130830 , 14130832 , 14130839 , 14130861 , 14130862 , 14130874 , 14130918 , 14130923 , 14130976 , 14130994 , 14131086 , 14131089 , 14131133 , 14131139 , 14131213 , 14131240 , 14131322 , 14131416 , 14131488 , 14131650 , 14131727 , 14131863 , 14132092 , 14132812 , 14132895 , 14132927 , 14132957 , 14132958 , 14132977 , 14132989 , 14133279 , 14133338 , 14133423 , 14133636 , 14133709 , 14133724 , 14133752 , 14133814 , 14134439 , 14134557 , 14134590 , 14134598 , 14134619 , 14134636 , 14134642 , 14134649 , 14134661 , 14134666 , 14134670 , 14134671 , 14134676 , 14134683 , 14134684 , 14134722 , 14134723 , 14134724 , 14134727 , 14134734 , 14134750 , 14134765 , 14135295 , 13873549 , 13880768 , 13880838 , 13880875 , 13880894 , 13880936 , 13881087 , 13881097 , 13881875 , 13881933 , 13882073 , 13882277 , 13882325 , 13882527 , 13882572 , 13883083 , 13884072 , 13884151 , 13884156 , 13884177 , 13884187 , 13884207 , 13884217 , 13884226 , 13884233 , 13884247 , 13884263 , 13884273 , 13884385 , 13884584 , 13884596 , 13884676 , 13884680 , 13884755 , 13884762 , 13884832 , 13884844 , 13884859 , 13884872 , 13884874 , 13884883 , 13884887 , 13884894 , 13884896 , 13884898 , 13884909 , 13884911 , 13884915 , 13884916 , 13884917 , 13884919 , 13884920 , 13884921 , 13884922 , 13884925 , 13884926 , 13884929 , 13884930 , 13884931 , 13884935 , 13884936 , 13884937 , 13884938 , 13884939 , 13884940 , 13884941 , 13884942 , 13884947 , 13884953 , 13884954 , 13884955 , 13884956 , 13884958 , 13884960 , 13884972 , 13884978 , 13884996 , 13884997 , 13885017 , 13885019 , 13885021 , 13885022 , 13885023 , 13885025 , 13885026 , 13885028 , 13885029 , 13885031 , 13885033 , 13885034 , 13885036 , 13885037 , 13885038 , 13885039 , 13885041 , 13885043 , 13885045 , 13885046 , 13885048 , 13885050 , 13885051 , 13885052 , 13885053 , 13885054 , 13885055 , 13885056 , 13885057 , 13885059 , 13885060 , 13885062 , 13885063 , 13885064 , 13885065 , 13885067 , 13885068 , 13885069 , 13885070 , 13885071 , 13885074 , 13885075 , 13885077 , 13885079 , 13885080 , 13885082 , 13885083 , 13885085 , 13885086 , 13885087 , 13885088 , 13885089 , 13885090 , 13885093 , 13885094 , 13885096 , 13885097 , 13885098 , 13885099 , 13885100 , 13885101 , 13885102 , 13885103 , 13885104 , 13885105 , 13885108 , 13885109 , 13885110 , 13885111 , 13885113 , 13885114 , 13885115 , 13885116 , 13885118 , 13885119 , 13885120 , 13885121 , 13885122 , 13885123 , 13885124 , 13885126 , 13885127 , 13885128 , 13885131 , 13885133 , 13885134 , 13885135 , 13885139 , 13885140 , 13885141 , 13885142 , 13885143 , 13885144 , 13885145 , 13885146 , 13885147 , 13885148 , 13885149 , 13885150 , 13885151 , 13885152 , 13885153 , 13885154 , 13885172 , 13885173 , 13885192 , 13885220 , 13885224 , 13885239 , 13885262 , 13885272 , 13885312 , 13885509 , 13885923 , 13886164 , 13886797 , 13886827 , 13886839 , 13891249 , 13891774 , 13891984 , 13892347 , 13892826 , 13892831 , 13892836 , 13892837 , 13892844 , 13895238 , 13895257 , 13895265 , 13895274 , 13895358 , 13895368 , 13895377 , 13895388 , 13895575 , 13895581 , 13895582 , 13895587 , 13895591 , 13895606 , 13895804 , 13895816 , 13895887 , 13896065 , 13896873 , 13896888 , 13896906 , 13897076 , 13898436 , 13898667 , 13898680 , 13898681 , 13898786 , 13898849 , 13898864 , 13898866 , 13898891 , 13898896 , 13898912 , 13899457 , 13899525 , 13899535 , 13899665 , 13899671 , 13899701 , 13900268 , 13900397 , 13900474 , 13900512 , 13900515 , 13900519 , 13900523 , 13900524 , 13900526 , 13900528 , 13900531 , 13900532 , 13900536 , 13900540 , 13900552 , 13900553 , 13900557 , 13900559 , 13900564 , 13900682 , 13900849 , 13900871 , 13901071 , 13901073 , 13901998 , 13902153 , 13902170 , 13902202 , 13902206 , 13902213 , 13902263 , 13902274 , 13902275 , 13902276 , 13902279 , 13902280 , 13902283 , 13902285 , 13902293 , 13902677 , 13902890 , 13903198 , 13903247 , 13903353 , 13903689 , 13903692 , 13903693 , 13903697 , 13903723 , 13903724 , 13903725 , 13903732 , 13903748 , 13903752 , 13903760 , 13903762 , 13903789 , 13904177 , 13904321 , 13904424 , 13904756 , 13905003 , 13905005 , 13905006 , 13905017 , 13905024 , 13905042 , 13905056 , 13905068 , 13905075 , 13905131 , 13905136 , 13905140 , 13905143 , 13905195 , 13905202 , 13905206 , 13905623 , 13905926 , 13906173 , 13906380 , 13906566 , 13906589 , 13906590 , 13906702 , 13906707 , 13906907 , 13906975 , 13907169 , 13907198 , 13907207 , 13907293 , 13908785 , 13911042 , 13911193 , 13911227 , 13911228 , 13911238 , 13911239 , 13911240 , 13911243 , 13911253 , 13911290 , 13911874 , 13911883 , 13911897 , 13911937 , 13911942 , 13911949 , 13911955 , 13911956 , 13911968 , 13911969 , 13911978 , 13912122 , 13912145 , 13912178 , 13913889 , 13913947 , 13913991 , 13914133 , 13914153 , 13914154 , 13914158 , 13914251 , 13914273 , 13914278 , 13914401 , 13914402 , 13914404 , 13914437 , 13914479 , 13914485 , 13914504 , 13914575 , 13914687 , 13914691 , 13914708 , 13914720 , 13914736 , 13914740 , 13914750 , 13914752 , 13914765 , 13914766 , 13914784 , 13914798 , 13914809 , 13914815 , 13914824 , 13914836 , 13914852 , 13914872 , 13914881 , 13914885 , 13914896 , 13914906 , 13914920 , 13914931 , 13914932 , 13914939 , 13914967 , 13914986 , 13915012 , 13915016 , 13915020 , 13915031 , 13915033 , 13915038 , 13915046 , 13915051 , 13915053 , 13915058 , 13915065 , 13915066 , 13915070 , 13915076 , 13915087 , 13915091 , 13915106 , 13915109 , 13915112 , 13915117 , 13915119 , 13915122 , 13915134 , 13915135 , 13915145 , 13915146 , 13915147 , 13915148 , 13915150 , 13915161 , 13915174 , 13915175 , 13915181 , 13915182 , 13915185 , 13915186 , 13915187 , 13915190 , 13915191 , 13915195 , 13915204 , 13915213 , 13915222 , 13915229 , 13915247 , 13915248 , 13915255 , 13915262 , 13915270 , 13915273 , 13915278 , 13915288 , 13915290 , 13915292 , 13915296 , 13915298 , 13915305 , 13915307 , 13915308 , 13915309 , 13915316 , 13915318 , 13915326 , 13915330 , 13915336 , 13915337 , 13915340 , 13915350 , 13915353 , 13915358 , 13915361 , 13915364 , 13915366 , 13915371 , 13915376 , 13915378 , 13915385 , 13915386 , 13915387 , 13915389 , 13915390 , 13915393 , 13915402 , 13915408 , 13915411 , 13915418 , 13915419 , 13915423 , 13915431 , 13915432 , 13915434 , 13915438 , 13915443 , 13915444 , 13915449 , 13915460 , 13915461 , 13915467 , 13915469 , 13915482 , 13915483 , 13915485 , 13915495 , 13915496 , 13915497 , 13915498 , 13915502 , 13915506 , 13915507 , 13915509 , 13915512 , 13915520 , 13915523 , 13915524 , 13915533 , 13915538 , 13915539 , 13915540 , 13915542 , 13915543 , 13915546 , 13915548 , 13915549 , 13915550 , 13915555 , 13915558 , 13915561 , 13915562 , 13915563 , 13915564 , 13915571 , 13915573 , 13915579 , 13915590 , 13915592 , 13915598 , 13915599 , 13915604 , 13915612 , 13915614 , 13915618 , 13915624 , 13915625 , 13915627 , 13915630 , 13915633 , 13915634 , 13915643 , 13915650 , 13915651 , 13915657 , 13915658 , 13915659 , 13915661 , 13915662 , 13915667 , 13915671 , 13915673 , 13915679 , 13915680 , 13915682 , 13915688 , 13915692 , 13915695 , 13915700 , 13915701 , 13915717 , 13915721 , 13915722 , 13915727 , 13915732 , 13915742 , 13915744 , 13915752 , 13915759 , 13915766 , 13915767 , 13915788 , 13915802 , 13915807 , 13915812 , 13915815 , 13915816 , 13915817 , 13915821 , 13915827 , 13915832 , 13915833 , 13915960 , 13916093 , 13916119 , 13916122 , 13916125 , 13916126 , 13916129 , 13916141 , 13916156 , 13916175 , 13916180 , 13916182 , 13916187 , 13916203 , 13916242 , 13916256 , 13916282 , 13916283 , 13916285 , 13916295 , 13916302 , 13916305 , 13916333 , 13916336 , 13916346 , 13917089 , 13917541 , 13917578 , 13917596 , 13918458 , 13918460 , 13918545 , 13918557 , 13918566 , 13918567 , 13918569 , 13918570 , 13918571 , 13918574 , 13918601 , 13918608 , 13918617 , 13918619 , 13918626 , 13918628 , 13918637 , 13918646 , 13918663 , 13918744 , 13918746 , 13918798 , 13918822 , 13918886 , 13918946 , 13919079 , 13919086 , 13919157 , 13919159 , 13919203 , 13919232 , 13919261 , 13919267 , 13919270 , 13919300 , 13919302 , 13919303 , 13919318 , 13919331 , 13919356 , 13919362 , 13919377 , 13919469 , 13919478 , 13919488 , 13919496 , 13919497 , 13919500 , 13919508 , 13919514 , 13919516 , 13919519 , 13919520 , 13919523 , 13919524 , 13919525 , 13919526 , 13919533 , 13919534 , 13919536 , 13919544 , 13919547 , 13919552 , 13919556 , 13919557 , 13919563 , 13919564 , 13919565 , 13919571 , 13919574 , 13919585 , 13919600 , 13919610 , 13919618 , 13919639 , 13919641 , 13919649 , 13919650 , 13919654 , 13919657 , 13919659 , 13919660 , 13919662 , 13919665 , 13919666 , 13919667 , 13919669 , 13919670 , 13919683 , 13919684 , 13919686 , 13919704 , 13919719 , 13919722 , 13919738 , 13919748 , 13919759 , 13919773 , 13919797 , 13919798 , 13919893 , 13919896 , 13919898 , 13919899 , 13919904 , 13919907 , 13919916 , 13919928 , 13919932 , 13919935 , 13919997 , 13919998 , 13920001 , 13920014 , 13920015 , 13920022 , 13920032 , 13920034 , 13920037 , 13920038 , 13920040 , 13920041 , 13920042 , 13920043 , 13920044 , 13920058 , 13920059 , 13920064 , 13920070 , 13920071 , 13920079 , 13920084 , 13920086 , 13920088 , 13920098 , 13920100 , 13920103 , 13920106 , 13920110 , 13920112 , 13920114 , 13920115 , 13920128 , 13920129 , 13920131 , 13920132 , 13920133 , 13920136 , 13920137 , 13920140 , 13920142 , 13920143 , 13920144 , 13920145 , 13920146 , 13920147 , 13920148 , 13920149 , 13920150 , 13920152 , 13920153 , 13920155 , 13920157 , 13920161 , 13920162 , 13920165 , 13920167 , 13920168 , 13920170 , 13920172 , 13920173 , 13920174 , 13920176 , 13920178 , 13920180 , 13920181 , 13920185 , 13920186 , 13920187 , 13920189 , 13920190 , 13920191 , 13920192 , 13920195 , 13920197 , 13920198 , 13920199 , 13920201 , 13920202 , 13920204 , 13920205 , 13920208 , 13920209 , 13920211 , 13920212 , 13920214 , 13920217 , 13920218 , 13920223 , 13920226 , 13920227 , 13920229 , 13920231 , 13920232 , 13920239 , 13920240 , 13920241 , 13920247 , 13920250 , 13920253 , 13920255 , 13920258 , 13920261 , 13920262 , 13920264 , 13920266 , 13920268 , 13920271 , 13920272 , 13920275 , 13920276 , 13920278 , 13920289 , 13920299 , 13920312 , 13920321 , 13920325 , 13920326 , 13920328 , 13920330 , 13920331 , 13920332 , 13920333 , 13920336 , 13920337 , 13920338 , 13920339 , 13920340 , 13920347 , 13920349 , 13920351 , 13920355 , 13920369 , 13920385 , 13921906 , 13922000 , 13922050 , 13922064 , 13922826 , 13922832 , 13922870 , 13922952 , 13922958 , 13922974 , 13923066 , 13923068 , 13923111 , 13923129 , 13923143 , 13923166 , 13923462 , 13923501 , 13923515 , 13923658 , 13923694 , 13840337 , 13841014 , 13841392 , 13841946 , 13842115 , 13842594 , 13842837 , 13845006 , 13845624 , 13846042 , 13846537 , 13846720 , 13846829 , 13846901 , 13846915 , 13847040 , 13847243 , 13847542 , 13847736 , 13847801 , 13847820 , 13847856 , 13847899 , 13847944 , 13847980 , 13847983 , 13848067 , 13848110 , 13848123 , 13848134 , 13848217 , 13848252 , 13848258 , 13848268 , 13848293 , 13848307 , 13848437 , 13848441 , 13850390 , 13850697 , 13850864 , 13850994 , 13851023 , 13851049 , 13851056 , 13851115 , 13851170 , 13851256 , 13851257 , 13851299 , 13851521 , 13851995 , 13852797 , 13853382 , 13853434 , 13853661 , 13854014 , 13854050 , 13855369 , 13855958 , 13856752 , 13857029 , 13857804 , 13858513 , 13858530 , 13858549 , 13858556 , 13858612 , 13858641 , 13858656 , 13858661 , 13858740 , 13858778 , 13858806 , 13858826 , 13858853 , 13858861 , 13858932 , 13858967 , 13858972 , 13858980 , 13858988 , 13858990 , 13858998 , 13859010 , 13859099 , 13859163 , 13859167 , 13859210 , 13859439 , 13859448 , 13859826 , 13859846 , 13859868 , 13859869 , 13859874 , 13859879 , 13859882 , 13859891 , 13859894 , 13859936 , 13859946 , 13859948 , 13859964 , 13859967 , 13859972 , 13860073 , 13860083 , 13860138 , 13860169 , 13860181 , 13860187 , 13860205 , 13860208 , 13860339 , 13860435 , 13860816 , 13860920 , 13862666 , 13863674 , 13864589 , 13868277 , 13868369 , 13868412 , 13871261 , 13871289 , 13872546 , 13872956 , 13873629 , 13876134 , 13876188 , 13878152 , 13878190 , 13878201 , 13878403 , 13878427 , 13878435 , 13878465 , 13878481 , 13878926 , 13880635 , 13880667 , 13880751 , 13880819 , 13880873 , 13880906 , 13880996 , 13881005 , 13881012 , 13881036 , 13881047 , 13881064 , 13881068 , 13881083 , 13881094 , 13881102 , 13881131 , 13881137 , 13881160 , 13881765 , 13882184 , 13882218 , 13882385 , 13882552 , 13882672 , 13882715 , 13882974 , 13883331 , 13883457 , 13883999 , 13884272 , 13884274 , 13884275 , 13884276 , 13884280 , 13884298 , 13884306 , 13884307 , 13884312 , 13884313 , 13884314 , 13884315 , 13884317 , 13884328 , 13884340 , 13884341 , 13884370 , 13884371 , 13884372 , 13884375 , 13884377 , 13884378 , 13884379 , 13884380 , 13884381 , 13884384 , 13884387 , 13884388 , 13884390 , 13884392 , 13884394 , 13884397 , 13884398 , 13884401 , 13884403 , 13884404 , 13884406 , 13884407 , 13884408 , 13884409 , 13884410 , 13884414 , 13884498 , 13884577 , 13884658 , 13884826 , 13884886 , 13884907 , 13884918 , 13884923 , 13884962 , 13885018 , 13885020 , 13885024 , 13885032 , 13885035 , 13885072 , 13885076 , 13885078 , 13885084 , 13885106 , 13885117 , 13885125 , 13885129 , 13885130 , 13885159 , 13885168 , 13885175 , 13885188 , 13885209 , 13885216 , 13885221 , 13885226 , 13885227 , 13885253 , 13885258 , 13885405 , 13885409 , 13885432 , 13885455 , 13885643 , 13885711 , 13885735 , 13885748 , 13885751 , 13885754 , 13885758 , 13885761 , 13886132 , 13886137 , 13886163 , 13886488 , 13886666 , 13886807 , 13886964 , 13886997 , 13887080 , 13887133 , 13887196 , 13887350 , 13887502 , 13887576 , 13887598 , 13887604 , 13887660 , 13887744 , 13887774 , 13887893 , 13887895 , 13887948 , 13887972 , 13888081 , 13888214 , 13888366 , 13888368 , 13888371 , 13888372 , 13888375 , 13888378 , 13888473 , 13888664 , 13888668 , 13888669 , 13888670 , 13888672 , 13888674 , 13888675 , 13888699 , 13888725 , 13888777 , 13889211 , 13889255 , 13889373 , 13889374 , 13889420 , 13889621 , 13889663 , 13889692 , 13889703 , 13889712 , 13889713 , 13889714 , 13889717 , 13889718 , 13889719 , 13889720 , 13889723 , 13889725 , 13889728 , 13889730 , 13889731 , 13889732 , 13889735 , 13889738 , 13889739 , 13889742 , 13889745 , 13889749 , 13889751 , 13889764 , 13889765 , 13889767 , 13889769 , 13889771 , 13889780 , 13891106 , 13891119 , 13891276 , 13891964 , 13892004 , 13892034 , 13892277 , 13892286 , 13892294 , 13892336 , 13892354 , 13892387 , 13892390 , 13892393 , 13892394 , 13892401 , 13892402 , 13892403 , 13892406 , 13892642 , 13893088 , 13894187 , 13894461 , 13894525 , 13894561 , 13894635 , 13894650 , 13894685 , 13894767 , 13895088 , 13895183 , 13895191 , 13895296 , 13895346 , 13895418 , 13895584 , 13895645 , 13895676 , 13895712 , 13895978 , 13896125 , 13896166 , 13896202 , 13896409 , 13896429 , 13896439 , 13896505 , 13896514 , 13896526 , 13896528 , 13896530 , 13896666 , 13896845 , 13896856 , 13896858 , 13896861 , 13896902 , 13897018 , 13897058 , 13897067 , 13897098 , 13897129 , 13897237 , 13897244 , 13898263 , 13898678 , 13898684 , 13898685 , 13898688 , 13898689 , 13898690 , 13898692 , 13898693 , 13898695 , 13898698 , 13898700 , 13898701 , 13898705 , 13898721 , 13898727 , 13898749 , 13898773 , 13898782 , 13898927 , 13899136 , 13899322 , 13899464 , 13899474 , 13899480 , 13899521 , 13900276 , 13900317 , 13900323 , 13900335 , 13900369 , 13900410 , 13900416 , 13900434 , 13900473 , 13900530 , 13900541 , 13900582 , 13900585 , 13901911 , 13901987 , 13902056 , 13902154 , 13902204 , 13902264 , 13902273 , 13902288 , 13902331 , 13902462 , 13902474 , 13902774 , 13902781 , 13902906 , 13903122 , 13903142 , 13903145 , 13903147 , 13903150 , 13903158 , 13903169 , 13903175 , 13903177 , 13903183 , 13903187 , 13903208 , 13903214 , 13903215 , 13903216 , 13903218 , 13903273 , 13903299 , 13903330 , 13903572 , 13903655 , 13903668 , 13903700 , 13903706 , 13903721 , 13903726 , 13903728 , 13903733 , 13903736 , 13903739 , 13903785 , 13903792 , 13903867 , 13903893 , 13903930 , 13903944 , 13904100 , 13904102 , 13904104 , 13904106 , 13904108 , 13904111 , 13904312 , 13904324 , 13904354 , 13904380 , 13904384 , 13904397 , 13904597 , 13904695 , 13904809 , 13904837 , 13904888 , 13904922 , 13905036 , 13905053 , 13905054 , 13905218 , 13905336 , 13905425 , 13905511 , 13905548 , 13905734 , 13905804 , 13905860 , 13905899 , 13906032 , 13906063 , 13906125 , 13906319 , 13906327 , 13906486 , 13906517 , 13906854 , 13907109 , 13907156 , 13907367 , 13907417 , 13907539 , 13907672 , 13907878 , 13908057 , 13908083 , 13908199 , 13908217 , 13908230 , 13908765 , 13908769 , 13908804 , 13908991 , 13909181 , 13909197 , 13909213 , 13909493 , 13909533 , 13909588 , 13909600 , 13909631 , 13909649 , 13909771 , 13909773 , 13909898 , 13910167 , 13910520 , 13910637 , 13910645 , 13910749 , 13911017 , 13911031 , 13911091 , 13911097 , 13911190 , 13911233 , 13911234 , 13911250 , 13911258 , 13911276 , 13911287 , 13911288 , 13911304 , 13911316 , 13911317 , 13911323 , 13911342 , 13911348 , 13911872 , 13911877 , 13911879 , 13912065 , 13912223 , 13912255 , 13912347 , 13912398 , 13912443 , 13912540 , 13912573 , 13912649 , 13912671 , 13912681 , 13912731 , 13912779 , 13912798 , 13913687 , 13913771 , 13913823 , 13913825 , 13913839 , 13913883 , 13913993 , 13914095 , 13914326 , 13914328 , 13914354 , 13914370 , 13914382 , 13914389 , 13914413 , 13914418 , 13914476 , 13914481 , 13914525 , 13914557 , 13914566 , 13914660 , 13914676 , 13914686 , 13914689 , 13914702 , 13914707 , 13914724 , 13914733 , 13914749 , 13914751 , 13914753 , 13914754 , 13914755 , 13914758 , 13914785 , 13914789 , 13914792 , 13914793 , 13914811 , 13914817 , 13914839 , 13914840 , 13914842 , 13914843 , 13914845 , 13914857 , 13914864 , 13914868 , 13914876 , 13914889 , 13914890 , 13914897 , 13914898 , 13914901 , 13914910 , 13914911 , 13914913 , 13914930 , 13914938 , 13914958 , 13914978 , 13914992 , 13914997 , 13915001 , 13915006 , 13915007 , 13915008 , 13915010 , 13915011 , 13915018 , 13915021 , 13915022 , 13915025 , 13915039 , 13915044 , 13915047 , 13915050 , 13915055 , 13915057 , 13915062 , 13915064 , 13915067 , 13915072 , 13915075 , 13915078 , 13915079 , 13915081 , 13915082 , 13915085 , 13915086 , 13915088 , 13915092 , 13915098 , 13915100 , 13915103 , 13915104 , 13915105 , 13915111 , 13915113 , 13915118 , 13915121 , 13915124 , 13915126 , 13915127 , 13915129 , 13915130 , 13915131 , 13915140 , 13915141 , 13915142 , 13915152 , 13915153 , 13915154 , 13915155 , 13915156 , 13915158 , 13915160 , 13915162 , 13915165 , 13915167 , 13915171 , 13915172 , 13915177 , 13915193 , 13915194 , 13915196 , 13915197 , 13915198 , 13915199 , 13915200 , 13915202 , 13915205 , 13915208 , 13915211 , 13915212 , 13915214 , 13915215 , 13915217 , 13915223 , 13915225 , 13915231 , 13915232 , 13915233 , 13915236 , 13915237 , 13915239 , 13915240 , 13915241 , 13915242 , 13915243 , 13915244 , 13915245 , 13915250 , 13915251 , 13915252 , 13915254 , 13915257 , 13915258 , 13915259 , 13915261 , 13915263 , 13915269 , 13915272 , 13915275 , 13915276 , 13915279 , 13915282 , 13915283 , 13915284 , 13915289 , 13915302 , 13915303 , 13915304 , 13915311 , 13915315 , 13915317 , 13915319 , 13915322 , 13915323 , 13915324 , 13915329 , 13915333 , 13915339 , 13915341 , 13915342 , 13915343 , 13915345 , 13915346 , 13915349 , 13915351 , 13915352 , 13915355 , 13915360 , 13915362 , 13915365 , 13915367 , 13915368 , 13915372 , 13915379 , 13915384 , 13915388 , 13915391 , 13915395 , 13915396 , 13915397 , 13915399 , 13915401 , 13915404 , 13915409 , 13915413 , 13915415 , 13915417 , 13915424 , 13915425 , 13915429 , 13915435 , 13915439 , 13915442 , 13915445 , 13915448 , 13915451 , 13915454 , 13915456 , 13915457 , 13915458 , 13915459 , 13915466 , 13915470 , 13915475 , 13915476 , 13915480 , 13915481 , 13915486 , 13915489 , 13915490 , 13915491 , 13915494 , 13915499 , 13915501 , 13915503 , 13915504 , 13915508 , 13915510 , 13915511 , 13915513 , 13915514 , 13915515 , 13915521 , 13915522 , 13915526 , 13915527 , 13915528 , 13915530 , 13915531 , 13915532 , 13915534 , 13915536 , 13915537 , 13915545 , 13915551 , 13915552 , 13915553 , 13915554 , 13915556 , 13915557 , 13915559 , 13915568 , 13915572 , 13915574 , 13915578 , 13915583 , 13915589 , 13915591 , 13915600 , 13915602 , 13915603 , 13915606 , 13915607 , 13915608 , 13915610 , 13915616 , 13915617 , 13915619 , 13915621 , 13915628 , 13915629 , 13915632 , 13915638 , 13915641 , 13915642 , 13915645 , 13915646 , 13915647 , 13915648 , 13915655 , 13915656 , 13915660 , 13915665 , 13915668 , 13915674 , 13915686 , 13915689 , 13915690 , 13915696 , 13915697 , 13915698 , 13915704 , 13915707 , 13915708 , 13915710 , 13915713 , 13915714 , 13915715 , 13915719 , 13915725 , 13915728 , 13915730 , 13915734 , 13915735 , 13915737 , 13915738 , 13915740 , 13915747 , 13915750 , 13915756 , 13915763 , 13915764 , 13915772 , 13915773 , 13915780 , 13915786 , 13915804 , 13915806 , 13915811 , 13915822 , 13915830 , 13915871 , 13916034 , 13916062 , 13916104 , 13916113 , 13916121 , 13916142 , 13916169 , 13916172 , 13916184 , 13916275 , 13916345 , 13916364 , 13916413 , 13916577 , 13916623 , 13916633 , 13916668 , 13916769 , 13916798 , 13916863 , 13916898 , 13917073 , 13917365 , 13917407 , 13917456 , 13917611 , 13917673 , 13917719 , 13918427 , 13918529 , 13918568 , 13918648 , 13918857 , 13919077 , 13919298 , 13919322 , 13919440 , 13919470 , 13919538 , 13919539 , 13919562 , 13919614 , 13919653 , 13919724 , 13919867 , 13920111 , 13920113 , 13920139 , 13920175 , 13920193 , 13920215 , 13920216 , 13920236 , 13920287 , 13920493 , 13920564 , 13920656 , 13920939 , 13920960 , 13921280 , 13921292 , 13921385 , 13921386 , 13921390 , 13921391 , 13921394 , 13921396 , 13921398 , 13921399 , 13921400 , 13921615 , 13921833 , 13922058 , 13922134 , 13922728 , 13923095 , 13923099 , 13923112 , 13923114 , 14100034 , 14100048 , 14100417 , 14100499 , 14100648 , 14101720 , 14102168 , 14102269 , 14102321 , 14102363 , 14102566 , 14102662 , 14103428 , 14103767 , 14104178 , 14104569 , 14104702 , 14120029 , 14120354 , 14121004 , 14121035 , 14121149 , 14121158 , 14122015 , 14123053 , 14123114 , 14123172 , 14123275 , 14123438 , 14124026 , 14124138 , 14124202 , 14124256 , 14124313 , 14124376 , 14124414 , 14124422 , 14124550 , 14126061 , 14126269 , 14126764 , 14128187 , 14128617 , 14128690 , 14128792 , 14128941 , 14129135 , 14129215 , 14129361 , 14129514 , 14129646 , 14129969 , 14130169 , 14130370 , 14130374 , 14131026 , 14131544 , 14133225 , 14133283 , 14133512 , 14133521 , 14134348 , 14134384 , 14134518 , 14134597 , 14134605 , 14134665 , 14134694 , 14134749 , 14134860 , 14134990 , 14135032 , 14135040 , 14135067 , 14135105 , 14135135 , 14135148 , 14135258 , 14135373 , 14135548 , 14135640 , 14135845 , 14135858 , 14135991 , 14136172 , 14136189 , 14136228 , 14136287 , 14136378 , 14136439 , 14136848 , 14136890 , 14136911 , 14136959 , 14136967 , 14136976 , 14136998 , 14136999 , 14137017 , 14137019 , 14137032 , 14137045 , 14137046 , 14137049 , 14137059 , 14137140 , 14137307 , 14137314 , 14137425 , 14137447 , 14137460 , 14137549 , 14137570 , 14137602 , 17427525 , 17427571 , 17428143 , 17428478 , 17428516 , 17428567 , 17428705 , 17428707 , 17428740 , 17429030 , 17429068 , 17429070 , 17429343 , 17429424 , 17429442 , 17429506 , 17429535 , 17429547 , 17429559 , 17429584 , 17429615 , 17429618 , 17429633 , 17429636 , 17429640 , 17429641 , 17429642 , 17429662 , 17429682 , 17429692 , 17429699 , 17429721 , 17429725 , 17429726 , 17429733 , 17429736 , 17429741 , 17429753 , 17429754 , 17429756 , 17429759 , 17429760 , 17429761 , 17429771 , 17429774 , 17429775 , 17429777 , 17429778 , 17429781 , 17429872 , 17429995 , 17430413 , 17430419 , 17430469 , 17430632 , 17430634 , 17431904 , 17432084 , 17432207 , 17432242 , 17432246 , 17432263 , 17432331 , 17432374 , 17432380 , 17432387 , 17432416 , 17432427 , 17432508 , 17432641 , 17432688 , 17432693 , 17432910 , 17432920 , 17432968 , 17432996 , 17433140 , 17433150 , 17433713 , 17433752 , 17433942 , 17433972 , 17434015 , 17434220 , 17434229 , 17434234 , 17434246 , 17434253 , 17434258 , 17434351 , 17434412 , 17434493 , 17434572 , 17434622 , 17434632 , 17434717 , 17434729 , 17434739 , 17434754 , 17434758 , 17434986 , 17435301 , 17435402 , 17435403 , 17435418 , 17435433 , 17435435 , 17435438 , 17435442 , 17435443 , 17435468 , 17435478 , 17435505 , 17435516 , 17435530 , 17435531 , 17435534 , 17435538 , 17435556 , 17435656 , 17435688 , 17435701 , 17435724 , 17435830 , 17435834 , 17435836 , 17435838 , 17435855 , 17435875 , 17435895 , 17435929 , 17435931 , 17435941 , 17436023 , 17436097 , 17436104 , 17436117 , 17436138 , 17436141 , 17436147 , 17436149 , 17436152 , 17436153 , 17436155 , 17436156 , 17436157 , 17436158 , 17436159 , 17436161 , 17436162 , 17436166 , 17436167 , 17436171 , 17436172 , 17436174 , 17436175 , 17436177 , 17436179 , 17436181 , 17436183 , 17436186 , 17436189 , 17436190 , 17436193 , 17436194 , 17436198 , 17436199 , 17436203 , 17436209 , 17436211 , 17436222 , 17436238 , 17436240 , 17436242 , 17436244 , 17436245 , 17436247 , 17436249 , 17436251 , 17436254 , 17436256 , 17436263 , 17436264 , 17436265 , 17436266 , 17436267 , 17436268 , 17436273 , 17436274 , 17436275 , 17436276 , 17436277 , 17436278 , 17436280 , 17436285 , 17436287 , 17436288 , 17436289 , 17436292 , 17436294 , 17436295 , 17436296 , 17436299 , 17436300 , 17436301 , 17436303 , 17436304 , 17436305 , 17436310 , 17436311 , 17436312 , 17436313 , 17436314 , 17436316 , 17436317 , 17436318 , 17436319 , 17436320 , 17436321 , 17436322 , 17436323 , 17436324 , 17436325 , 17436326 , 17436327 , 17436328 , 17436329 , 17436331 , 17436332 , 17436333 , 17436335 , 17436336 , 17436337 , 17436342 , 17436343 , 17436344 , 17436345 , 17436347 , 17436348 , 17436350 , 17436353 , 17436355 , 17436356 , 17436358 , 17436360 , 17436361 , 17436362 , 17436366 , 17436367 , 17436368 , 17436369 , 17436372 , 17436373 , 17436374 , 17436376 , 17436379 , 17436380 , 17436381 , 17436382 , 17436383 , 17436384 , 17436385 , 17436386 , 17436387 , 17436388 , 17436389 , 17436391 , 17436392 , 17436393 , 17436394 , 17436395 , 17436396 , 17436399 , 17436403 , 17436406 , 17436411 , 17436412 , 17436413 , 17436414 , 17436415 , 17436416 , 17436417 , 17436418 , 17436420 , 17436421 , 17436422 , 17436423 , 17436425 , 17436426 , 17436428 , 17436429 , 17436430 , 17436433 , 17436434 , 17436436 , 17436438 , 17436439 , 17436440 , 17436441 , 17436442 , 17436443 , 17436445 , 17436447 , 17436449 , 17436450 , 17436452 , 17436453 , 17436454 , 17436455 , 17436457 , 17436459 , 17436461 , 17436464 , 17436466 , 17436470 , 17436475 , 17436477 , 17436479 , 17436480 , 17436489 , 17436493 , 17436496 , 17436498 , 17436499 , 17436501 , 17436502 , 17436506 , 17436507 , 17436508 , 17436509 , 17436511 , 17436512 , 17436513 , 17436515 , 17436516 , 17436518 , 17436520 , 17436522 , 17436524 , 17436527 , 17436529 , 17436530 , 17436532 , 17436533 , 17436534 , 17436536 , 17436537 , 17436539 , 17436541 , 17436549 , 17436552 , 17436555 , 17436559 , 17436560 , 17436578 , 17436579 , 17436584 , 17436585 , 17436587 , 17436589 , 17436591 , 17436592 , 17436595 , 17436596 , 17436597 , 17436600 , 17436603 , 17436605 , 17436606 , 17436610 , 17436614 , 17436615 , 17436621 , 17436623 , 17436625 , 17436627 , 17436628 , 17436629 , 17436632 , 17436634 , 17436635 , 17436639 , 17436640 , 17436663 , 17436672 , 17436678 , 17436679 , 17436680 , 17436688 , 17436689 , 17436702 , 17436727 , 17436868 , 17436874 , 17436947 , 17436966 , 17436971 , 17436983 , 17436988 , 17436996 , 17436998 , 17437008 , 17437009 , 17437014 , 17437029 , 17437033 , 17437036 , 17437044 , 17437103 , 17437213 , 17437327 , 17437526 , 17441845 , 19766078 , 19766079 , 19766083 , 19766095 , 19766111 , 19766113 , 19766115 , 19766119 , 19766120 , 19766121 , 19766125 , 19766129 , 19766130 , 19766133 , 19766136 , 19766137 , 19766138 , 19766142 , 19766149 , 19766151 , 19766154 , 19766159 , 19766160 , 19766163 , 19766172 , 19766176 , 19766179 , 19766180 , 19766181 , 19766184 , 19766191 , 19766201 , 19766204 , 17427436 , 17427549 , 17427619 , 17427717 , 17427728 , 17427729 , 17427734 , 17427735 , 17427754 , 17427769 , 17427776 , 17427779 , 17428553 , 17428615 , 17428625 , 17428892 , 17429096 , 17429120 , 17429204 , 17429597 , 17430008 , 17430212 , 17430224 , 17430998 , 17431114 , 17431152 , 17431224 , 17431293 , 17431780 , 17431879 , 17431903 , 17432194 , 17432261 , 17432341 , 17432413 , 17432430 , 17432511 , 17432518 , 17432519 , 17432540 , 17432672 , 17432686 , 17432712 , 17432732 , 17432790 , 17432814 , 17432941 , 17432943 , 17432950 , 17433007 , 17433014 , 17433052 , 17433156 , 17433158 , 17433194 , 17433292 , 17433382 , 17433425 , 17433555 , 17433805 , 17434099 , 17434248 , 17434402 , 17434545 , 17434554 , 17434591 , 17435028 , 17435145 , 17435150 , 17435308 , 17435413 , 17435444 , 17435465 , 17435480 , 17435581 , 17435587 , 17435649 , 17435657 , 17435668 , 17435684 , 17435773 , 17435798 , 17435899 , 17435956 , 17435959 , 17435979 , 17436163 , 17436260 , 17436302 , 17436315 , 17436334 , 17436377 , 17436427 , 17436437 , 17436463 , 17436467 , 17436487 , 17436492 , 17436495 , 17436526 , 17436588 , 17436590 , 17436601 , 17436653 , 17436686 , 17436706 , 17436724 , 17436725 , 17436876 , 17436919 , 17437100 , 17437151 , 17437462 , 17437555 , 17437807 , 17437860 , 17437908 , 17437925 , 17438237 , 17441419 , 17441434 , 17441626 , 17441674 , 17443702 , 19766085 , 19766112 , 19766117 , 19766123 , 19766127 , 19766135 , 19766164 , 19766166 , 19766173 , 19766220
FEI Number 3013319212
Recalling Firm/
Manufacturer
Hospira Inc.
275 N Field Dr
Lake Forest IL 60045-2579
For Additional Information ContactHospira Global Complaint Management
800-441-4100
Manufacturer Reason
for Recall
One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
FDA Determined
Cause 2
Component change control
ActionHospira sent an Urgent Medical Device Correction letter dated July 2, 2014, to all affected customers. The letter notified customers that Hospira will be contacting them regarding the completion of an audible alarm test and will replace any alarm assemblies that fail to audible alarm test. Additionally, the letter included instructions for customers to: 1) inform potential users of this notification; 2) complete and return the attached reply form; and, 3) if the products were further distributed, notify those accounts of this recall and ask them to contact Stericycle at 888-912-7350 (Monday - Friday, 8:00 AM - 5:00 PM ET) to receive a reply form. A second URGENT MEDICAL DEVICE CORRECTION letter dated July 16, 2014 was sent to all customers of record because the initial letters may have contained errors in the attached data files listing the potentially affected devices at the customer's' facility. The instructions to the customer were unchanged from the original letter. For questions regarding this recall call 800-441-4100.
Quantity in Commerce3,939 infusion pumps
DistributionWorldwide Distribution - USA including AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and Brazil, Canada, Colombia, Ecuador, France, Israel, Italy, Mexico, Oman, Peru, Saudi Arabia, Spain, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FRN
-
-