| Date Initiated by Firm | September 15, 2014 |
| Date Posted | January 13, 2015 |
| Recall Status1 |
Terminated 3 on May 20, 2015 |
| Recall Number | Z-0840-2015 |
| Recall Event ID |
69660 |
| 510(K)Number | K123603 |
| Product Classification |
Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
|
| Product | GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product
Used for general surgical applications and musculoskeletal procedures to visualize. |
| Code Information |
GE OEC Brivo 715 Prime Serial Numbers: B2S1301 6 B2S1301 8 B2S1302 1 B2S13022 B2S13025 B2S13039 B2S13060 B2S13063 B2S13065 B2S14001 B2S14007 B2S14022 B2S14039 B2S1404 0 B2S1404 5 B2S1406 5 B2S1406 6 |
| FEI Number |
1720753
|
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc 384 N Wright Brothers Dr Salt Lake City UT 84116-2862
|
| For Additional Information Contact | 801-536-4516 |
Manufacturer Reason for Recall | mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3). |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
| Action | Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit.
Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8. |
| Quantity in Commerce | 61 units installed in US |
| Distribution | Worldwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = OXO
|