| Date Initiated by Firm | July 27, 2015 |
| Date Posted | September 03, 2015 |
| Recall Status1 |
Terminated 3 on July 18, 2016 |
| Recall Number | Z-2720-2015 |
| Recall Event ID |
71907 |
| 510(K)Number | K872048 |
| Product Classification |
Flowmeter, blood, cardiovascular - Product Code DPW
|
| Product | Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC
Product Usage:
to measure flow intra-operatively. |
| Code Information |
Serial Numbers: HQC2FMC1801, HQC2FMC 1802, HQC2FMC1805, HQC2FMC1806 |
| FEI Number |
1319030
|
Recalling Firm/ Manufacturer |
Transonic Systems Inc 34 Dutch Mill Rd Ithaca NY 14850-9785
|
| For Additional Information Contact | Mr. David Klementowski 607-257-5300 |
Manufacturer Reason for Recall | Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number. |
FDA Determined Cause 2 | Software Manufacturing/Software Deployment |
| Action | Transonic sent an Urgent Medical Device Recall letters and Response Forms dated July 27, 2015 to customers via Certified Mail, email and/or fax. The letter identified the affected product, problem and actions to be taken. Customers are instructed to contact Transonic Systems' Customer Service at 1-800-353-3569 or 1-607-257-5300 to arrange for replacement (repair/exchange) immediately. |
| Quantity in Commerce | Domestic: 4 units |
| Distribution | Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|