| | Class 2 Device Recall KimVent Closed Suction System for Neonates/Pediatrics |  |
| Date Initiated by Firm | July 29, 2015 |
| Date Posted | August 31, 2015 |
| Recall Status1 |
Terminated 3 on January 12, 2016 |
| Recall Number | Z-2522-2015 |
| Recall Event ID |
71966 |
| Product Classification |
Catheters, suction, tracheobronchial - Product Code BSY
|
| Product | KimVent* Closed Suction System for Neonates/Pediatrics, 6 F, Elbow (Product Code 206); KimVent* Closed Suction System for Neonates/Pediatrics, 8 F, Elbow (Product Code 208); KimVent* Closed Suction System for Pediatrics, 10 F, Elbow (Product Code 210); KimVent* Closed Suction System for Neonates/Pediatrics, 10 F, Elbow (Product Code 2103); KimVent* Closed Suction Systems for Neonates/Pediatrics, 8 F, Elbow (Product Code 8313); KimVent* Closed Suction System for Neonates/Pediatrics, 8 F, Elbow (Product Code 20083)
A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway. |
| Code Information |
Product Code 206 - M5076T503, M5082T501, M5091T504; Product Code 208 - M5089T506, M5096T507, M5103T509; Product Code 210 - M5091T509; Product Code 2103 - M5082T401, M5089T401, M5138T402; Product Code 8313 - M5089T508; Product Code 20083 - M5082T403, M5082T404, M5089T402, M5103T401, M5166T402 |
Recalling Firm/ Manufacturer |
Halyard Health, Inc 5405 Windward Pkwy Alpharetta GA 30004-3894
|
| For Additional Information Contact | Thomas Kozma, PhD 470-448-5681 |
Manufacturer Reason for Recall | The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position. |
FDA Determined Cause 2 | Process control |
| Action | Halyard sent a Product Advisory Notice on July 29, 2015, via FEDEX with a second letter dated August 12, 2015, being issued via FEDEX. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to report any occurrence of a valve leak to Halyard Customer Service at 1-844-425-9273, option 1, option 3. |
| Distribution | Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|