  Class 2 Device Recall Customed see related information
Date Initiated by Firm January 07, 2016
Date Posted April 14, 2016
Recall Status1 Terminated 3 on May 03, 2017
Recall Number Z-1461-2016
Recall Event ID 72992
Product Classification General surgery tray (kit) - Product Code LRO
Product Catalog Number: 9002829
BASIC SURG PACK
Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Code Information LOTS: SL15187 (150719782) 31-JUL-16 SL15191-01 (150719783) 31-JUL-16 SL15244 (150820308) 31-AUG-16 SL15264 (150820309) 30-SEP-16 SL15273 (150820307) 30-SEP-16 SL15274 (150820307) 30-SEP-16 
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
Action Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016.
Quantity in Commerce 152
Distribution Distributed only in Puerto Rico.