• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Ultra 8 IAB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Ultra 8 IABsee related information
Date Initiated by FirmFebruary 10, 2016
Date PostedMarch 11, 2016
Recall Status1 Terminated 3 on May 12, 2021
Recall NumberZ-1056-2016
Recall Event ID 73266
510(K)NumberK000729 
Product Classification System, balloon, intra-aortic and control - Product Code DSP
ProductUltra 8 IAB: 8Fr 30cc; Product Code: IAB-05830-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.
Code Information Batch: 18F14A0031 18F14A0032 18F14A0033 18F14D0066 18F14E0018 18F14E0072 18F14F0067 18F14F0079 18F14G0045 18F14G0046 18F14G0081 18F14H0037 18F14J0023 18F14L0014 18F14M0033 18F15A0010 18F15B0015 18F15C0004 18F15C0023 18F15D0025 18F15D0052 18F15E0020 18F15F0002 18F15F0039 18F15G0005 18F15G0015 18F15H0006 18F15H0037 18F15H0051 18F15J0026 18F15L0007 18S14G0046 
FEI Number 3010532612
Recalling Firm/
Manufacturer
Arrow International, Inc., Division of Teleflex Medical Inc.
9 Plymouth St
Everett MA 02149-1814
For Additional Information Contact
617-389-6400
Manufacturer Reason
for Recall
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
FDA Determined
Cause 2
Process control
ActionThe firm, Teleflex Arrow International, sent an "Urgent Medical Device Recall Notification" letter dated 2/11/2016 via FedEx to its Customers. The letter described the product, problem and actions to be taken. The customers were informed of the reason for recall and were instructed to take the following actions: 1. If you have any products referenced in the attached enclosure, immediately discontinue use and quarantine those products appropriately. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507 Attn: Customer Service or email it to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email it to recalls@teleflex.com. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
Quantity in Commerce13,405 US and 33,735 OUS in total
DistributionWorldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DSP
-
-