• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Zimmer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Zimmer see related information
Date Initiated by Firm January 11, 2016
Date Posted March 12, 2016
Recall Status1 Terminated 3 on April 13, 2017
Recall Number Z-1131-2016
Recall Event ID 72900
510(K)Number K840643  
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
Product Product 33 consists of all product under product code: JDI and same usage:
Item no:
662406515 HGP II ACETABULAR CUP BON
662406520 HGP II ACETABULAR CUP BON
662406525 HGP II ACETABULAR CUP BON
662406530 HGP II ACETABULAR CUP BON
662406535 HGP II ACETABULAR CUP BON
662406540 HGP II ACETABULAR CUP BON
662406550 HGP II ACETABULAR CUP BON
662406560 HGP II ACETABULAR CUP BON

For use in total hip arthroplasty
Code Information lot no.: 61674472 61906696 62016799 62332483 61674473 61902194 61906697 62326386 61674476 61953546 62161131 62172656 62326385 61953547 62011539 62017894 62130757 61886072 61953548 62017895 62172659 61886073 61948763 61953551 62172661 61874793 61948764 61996232 62017897 62172663 61874794 61953554 62286122 
Recalling Firm/
Manufacturer
Zimmer Manufacturing B.V.
Turpeaux Industrial Park Rd 1 km 123.4
Mercedita PR 00715
For Additional Information Contact Consumer Relations Call Center
800-447-5633
Manufacturer Reason
for Recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
FDA Determined
Cause 2
Packaging
Action Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity in Commerce 2236
Distribution US Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JDI and Original Applicant = ZIMMER, INC.
-
-