| Class 2 Device Recall AB Sciex API 3200MD, 3200MD QTRAP, Triple Quad, QTRAP 4500MD LC/MS/MS Systems | |
Date Initiated by Firm | February 24, 2016 |
Date Posted | April 29, 2016 |
Recall Status1 |
Terminated 3 on February 06, 2017 |
Recall Number | Z-1586-2016 |
Recall Event ID |
73433 |
Product Classification |
Mass spectrometer, clinical use - Product Code DOP
|
Product | Analyst MD Version 1.6.1 and 1.6.2 Software used with the following instruments: API 3200MD" LC/MS/MS System, Instrument Part Number: 5024501; 3200MD QTRAP LC/MS/MS System, Instrument Part Number: 5024500; Triple Quad" 4500MD LC/MS/MS System, Instrument Part Number: 5031257; QTRAP 4500MD LC/MS/MS System, Instrument Part Number: 5031231
3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use. |
Code Information |
Software Versions: Version 1.6.1 and 1.6.2 |
Recalling Firm/ Manufacturer |
Ab Sciex 500 Old Connecticut Path Framingham MA 01701-4574
|
For Additional Information Contact | 289-982-2531 |
Manufacturer Reason for Recall | Wrong quantitative results may be displayed in a report from the device, which may potentially lead to an incorrect patient diagnosis. |
FDA Determined Cause 2 | Software design |
Action | Sciex sent an Urgent Medical Device Correction letter dated February 29, 2016, to all affected US customers via Fedex on March 2, 2016. International customers were notified beginning on March 3, 2016. The letter informed customers of the issue and requests that they take temporary steps to eliminate the potential for erroneous events. In addition, a software update is in development and is expected to be available in approximately 4 weeks. Customers with questions were instructed to contact SCIEX at 1-289-982-2531. |
Quantity in Commerce | 279 |
Distribution | Worldwide Distribution - US (Nationwide) and Internationally to the following countries:
AUSTRIA
CANADA
CHINA
FRANCE
GERMANY
INDIA
IRELAND
ITALY
KOREA
MALAYSIA
NETHERLANDS
NORWAY
RUSSIA
SINGAPORE
SPAIN
SWITZERLAND
UNITED KINGDOM |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|