• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall TOPAZ MicroDebrider

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TOPAZ MicroDebridersee related information
Date Initiated by FirmMarch 03, 2016
Date PostedApril 05, 2016
Recall Status1 Terminated 3 on May 17, 2017
Recall NumberZ-1321-2016
Recall Event ID 73525
510(K)NumberK080282 
Product Classification Electrosurgical, cutting & coagulation & accessories - Product Code GEI
ProductTOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01
Code Information 1031506, 1031507, 1032123, 1032325, 1032741, 1032742, 1033443, 1033965, 1033966, 1034397, 1039120, 1041765, 1044089, 1044090, 1045862, 1048237, 1048933, 1049185, 1052071, 1056212, 1060586, 1060979, 1063394, 1064599, 1067696, 1067748, 1069174, 1071358, 1074125, 1075714, 1070521, 1070812, 1071974, 1078025, 1077499, 1082811, 1084095, 1084094, 1085458, 1086436, 1086437, 1087917, 1094313, 1097573, 1103824, 1107183, 1107297, 1113562, 1113563, 1114534, 1114535, 1114536, 1115896, 1031518, 1032308, 1032309, 1032310, 1032755, 1032756, 1033972, 1033973, 1034225, 1034226, 1035283, 1035284, 1038547, 1043507, 1046556, 1047064, 1048722, 1049266, 1051812, 1052070, 1054443, 1055730, 1059865, 1063661, 1067691, 1067273, 1067747, 1068019, 1069172, 1071357, 1072853, 1072854, 1070497, 1070808, 1071971, 1076849, 1077450, 1079140, 1079220, 1078115, 1081753, 1082361, 1082810, 1086380, 1094312, 1094312, 1094721, 1097571, 1101041, 1107295, 1109045, 1111935, 1111936, 1113534, 1113535, 1113536, 1114532, 1115888, 1115891, 1116959, 1116130, 1116720, 1117209, 1117214, 1031341, 1032311, 1033761, 1034446, 1035691, 1035692, 1041775, 1043962, 1046557, 1048236, 1050067, 1052651, 1057008, 1058808, 1059867, 1063665, 1064594, 1067695, 1068021, 1071319, 1071320, 1073441, 1074876, 1077498, 1080520, 1114533, 1115893, 1031340, 1032962, 1033266, 1045359, 1047065, 1051331, 1056849, 1064592, 1067694, 1070709
Recalling Firm/
Manufacturer
ArthroCare Corporation
7000 W William Cannon Dr Bldg 1
Austin TX 78735-8509
For Additional Information ContactField Actions Department
901-566-7975
Manufacturer Reason
for Recall
During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.
FDA Determined
Cause 2
Package design/selection
ActionConsignees were notified via letter on 3/3/2016. Consignees were instructed to quarantine the recalled products and complete an Inventory Return Certification Form.
Quantity in Commerce26,903 devices
DistributionWorldwide Distribution-US (nationwide) and the following countries-Austria, Belgium, Bulgaria, Canada, Denmark, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, China, Hong Kong, India, Lebanon, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arabic Emirates, Vietnam, Australia, Japan, New Zealand, Brazil, and Chile.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = GEI
-
-