|
Class 2 Device Recall ADVIA Chemistry Triglyceride_2 concentrated |
|
Date Initiated by Firm |
March 07, 2016 |
Create Date |
May 14, 2016 |
Recall Status1 |
Terminated 3 on December 01, 2017 |
Recall Number |
Z-1670-2016 |
Recall Event ID |
73635 |
510(K)Number |
K133067
|
Product Classification |
Lipase hydrolysis/glycerol kinase enzyme, triglycerides - Product Code CDT
|
Product |
ADVIA Chemistry Triglyceride_2, concentrated; TRIG_c; Catalog # SMN # 10697575.
For in vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma on ADVIA Chemistry systems. |
Code Information |
Kit lots 348297 & 359932. |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact |
Loriann Russo 914-524-2287
|
Manufacturer Reason for Recall |
Siemens Healthcare Diagnostics has confirmed the ADVIA Chemistry Triglyceride concentrated reagent (TRIG_c) does not meet Instructions For Use (IFU) linearity claims at the upper limit of the assay range as it approaches end of shelf life for lots # 348297 and 359932.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Siemens mailed an Urgent Field Safety Notice, dated March 2016, to customers confirming the ADVIA Triglyceride concentrated reagent (TRIG_c) does not meet Instructions for Use (IFU) linearity claim at the upper limit of the assay range as it approaches end of shelf life for lots # 348297 and 359932. Siemens asked customers to discontinue use of and discard kits from the two referenced lots, complete and return the Field Correction Effectiveness Check attached to the letter within 30 days. Customers were to review their inventory to determine their laboratory 's replacement needs. Customers are asked to retain the letter for their records and forward a copy to anyone that received the product. Customers may contact Siemens Customers Service or their local technical support. |
Quantity in Commerce |
7,433 units (4,228 for lot 348297 and 3,199 for lot # 359932) |
Distribution |
Worldwide Distribution -- USA, including the states of CA, TN, UT, OK, FL, MI, MN, NY, CT, NC, KS, TX, CO, WA, NM, and OR; and, the countries of Austria, Australia, Belgium, Brazil, Bulgaria, Denmark, Estonia, France, Germany. Great Britain, Italy, Latvia, Netherland, Norway, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Czech Republic, India, New Zealand, Thailand, Turkey, and Russian Fed. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = CDT and Original Applicant = SIEMENS HEALTHCARE DIAGNOSTICS, INC.
|
|
|
|