| | Class 1 Device Recall INRatio Prothrombin Time (PT) Monitoring System |  |
| Date Initiated by Firm | July 11, 2016 |
| Date Posted | August 19, 2016 |
| Recall Status1 |
Terminated 3 on July 20, 2017 |
| Recall Number | Z-2355-2016 |
| Recall Event ID |
74665 |
| 510(K)Number | K020679 |
| Product Classification |
Test, time, prothrombin - Product Code GJS
|
| Product | INRatio Prothrombin Time (PT) Monitoring System
Model Number: 0100007
Product Usage:
The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test strip area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the Prothrombin time test on the blood sample. The other two channels contain the reagents. to perfom the Prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control test are required for the INRatio System. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set range, it indicates a problem with the test and an error message is reported on the display instead of a result. |
| Code Information |
All Lots All Serial Numbers |
| FEI Number |
1000125596
|
Recalling Firm/ Manufacturer |
Alere San Diego, Inc. 9975 Summers Ridge Rd San Diego CA 92121-2997
|
| For Additional Information Contact | Jackie Lustig 781-341-4009 |
Manufacturer Reason for Recall | Alere San Diego is initiating a voluntary recall for the Alere INRatio/INRatio2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio/INRatio2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory. |
FDA Determined Cause 2 | Other |
| Action | Alere issued a press release on July 11, 2016. An Urgent Medical Device Letter dated July 26, 2016 was sent to affected customers on July 29, 2016. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to consult with their healthcare provider as soon as possible to transition to an alternate method of PT/INR testing and to contact the firm regarding return or disposal of unused product and obtaining device replacement from a different manufacturer. For questions contact the Alere INRatid1 Recall Hotline at 1-866-723-2535. |
| Distribution | Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = GJS
|
|
|
|