| Class 3 Device Recall Cobalt Chromium Heads |  |
Date Initiated by Firm | June 21, 2016 |
Date Posted | December 01, 2016 |
Recall Status1 |
Terminated 3 on July 01, 2017 |
Recall Number | Z-0689-2017 |
Recall Event ID |
75063 |
510(K)Number | K100866 |
Product Classification |
Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Product Code LZO
|
Product | Cobalt Chromium Heads
Transcend: 26000017, 26000018, 26000019, 26000020, 26000021, 26000022, 26000023, 26000024, 26000025, 26000026, 26000027, 26000028
Conserve:
38013800, 38013804, 38013835, 38014000, 38014004, 38014035, 38014200, 38014204, 38014235, 38014400, 38014404, 38014435, 38014600, 38014604, 38014635, 38014800, 38014804, 38014835, 38015000, 38015004, 38015035, 38015200, 38015204, 38015235, 38015400, 38015404, 38015435,
38015600, 38015604, 38015635, 38AC3600, 38AC3800, 38AC4000, 38AC4200, 38AC4400, 38AC4600, 38AC4800, 38AC5000, 38AC5200, 38AC5400, 38AC5600, 38AM3600, 38AM3604, 38AM3635, 38AM3800, 38AM3804, 38AM3835, 38AM4000, 38AM4004, 38AM4035, 38AM4200,
38AM4204, 38AM4235, 38AM4400, 38AM4404, 38AM4435, 38AM4600, 38AM4604, 38AM4635, 38AM4800, 38AM4804, 38AM4835, 38AM5000, 38AM5004, 38AM5035, 38AM5200, 38AM5204, 38AM5235, 38AM5400, 38AM5404, 38AM5435, 38AM5600, 38AM5604, 38AM5635, PL38HM4400,
PL38HM4404, PL38HM4435, PL38HM4600, PL38HM4604, PL38HM4635, PL38HM4800, PL38HM4804, PL38HM4835, PL38HM5000, PL38HM5004, PL38HM5035, PL38HM5200, PL38HM5204, PL38HM5235 |
Code Information |
Transcend REF 26000017 Serial Numbers: 1392035, 1394262, 1403269, 1404128, 1405348, 1406868, 1416193, 1416410, 416916, 1417966, 1421701, 1424398, 1427395, 1429561, 1432313, 1432727, 1435174, 1436773, 1437714, 1444931, 1463853, 1469307, 1469928, 1472480, 1473802, 1475742, 1483472, 1485543, 1485544, 1491257, 1495872, 1495873, 1497225, 1497226, 1497227, 1497228, 1502990, 1502991, 1504262, 1509260, 1509261, 1513940, 1513941, 1516239, 1519175, 1523291, 1527824, 1527825, 1532934, 1532935, 1532936, 1534591, 1544282, 1544283, 1544284, 1548096, 1548097, 1553540, 1553541, 1555918, 1555919, 1555920, 1564763, 1564764, 1567221, 1567222, 1567223, 1567765, 1574244, 1574245, 1574246, 1586932, 1586937, 1587450, 1587451, 1587452, 1594879, 1602102, 1602103, 1607364, 1607365, 1612602, 1616179, 1618966, 1618967, 1636635, 1638711, 1638712, 1643938, 1643939, 1647724, 1647725, 1648020, 1648021, 1654503, 1654504, 1657402, 1659130, 02465512, 02465513, 02469868, 03211703, 03211704, 03217505, 04479690, 06345789, 06483517, 06485500, 10352345, 11230587, 12358929, 12360234, 015158577, 016305137, 017385213, 018513341, 018523879, 018523880, 026316750, 027411850, 035191682, 035191683, 037414681, 037417015, 037420473, 038542236, 038548053, 046336505, 047426887, 047431539, 049843313, 056332870, 057433631, 058595417, 058595419, 065229799, 066343506, 067442240, 067446422, 068595420, 068595422, 069876963, 069876964, 075255286, 076360909, 077450626, 077455529, 085258369, 086369486, 086370994, 087460631, 087462039, 087469136, 088672406, 088672407, 088672408, 094104768, 096372363, 097480313, 097481152, 098694953, 098703737, 0101010115, 104104769, 105275570, 107483300, 107489009, 108716008, 109938837, 0111278886, 114131980, 114135078, 115290207, 118738338, 118741128, 119978431, 119994169, 125301489, 127496633, 127513339, 127513340, 128739171, 0201069672, 0201069673, 0201069674, 0211297628, 0211299902, 0301089722, 0301091275, 0301096837, 0311323467, 0311325664, 0311330683, 0311332184, 0401100071, 0401100107, 0401101147, 0411332564, 0411344347, 0411344348, 0511345415, 0601148840, 0601148931, 0611371879, 0701179525, 0701179670, 0711370852, 0711371390, 0711381247, 0711390646, 0901193308, 0901199757, 0901200139, 0901200600, 1001235128, 1001236308, 1101235029, 1101254636, 1101254709, 1101258592, 1101258593, 1101259125, 1201266195, 1201267539, 1201267540, 16439381652115 REF 26000018 Serial Numbers: 1389320, 1392131, 1394263, 1397119, 1400719, 1403531, 1406742, 1406967, 1412519, 1413648, 1413871, 1414179, 1414180, 1414896, 1414897, 1416201, 1416673, 1420483, 1430442, 1430443, 1431008, 1431451, 1431452, 1432728, 1435175, 1436383, 1436772, 1439333, 1444099, 1444102, 1444240, 1446872, 1448610, 1457129, 1460999, 1461001, 1466074, 1466075, 1469308, 1470078, 1471689, 1472302, 1474005, 1474006, 1475023, 1476007, 1476100, 1476101, 1477377, 1481146, 1481147, 1483466, 1483467, 1484425, 1484426, 1484767, 1484768, 1487090, 1487091, 1487092, 1489845, 1491259, 1495973, 1495974, 1497250, 1497251, 1498778, 1498779, 1500626, 1501326, 1503278, 1503279, 1503728, 1504263, 1507991, 1507992, 1509298, 1509300, 1511955, 1511956, 1511957, 1513942, 1516240, 1516241, 1516967, 1516968, 1519176, 1523292, 1526399, 1526400, 1526401, 1527827, 1532937, 1532938, 1532939, 1534586, 1534587, 1534588, 1541698, 1541699, 1542310, 1542311, 1542312, 1542313, 1548271, 1548272, 1548797, 1551551, 1553113, 1553114, 1553115, 1554787, 1554788, 1554789, 1561459, 1561460, 1561461, 1561462, 1566790, 1566791, 1567219, 1567220, 1568431, 1568432, 1568433, 1568959, 1569900, 1586629, 1586630, 1587101, 1587102, 1587103, 1587104, 1590827, 1592379, 1592822, 1593585, 1598909, 1598910, 1602100, 1602101, 1607260, 1607262, 1610138, 1612604, 1615574, 1615928, 1618517, 1618518, 1618519, 1622629, 1622630, 1632305, 1632306, 1634568, 1634569, 1638713, 1638714, 1638715, 1643940, 1643941, 1644912, 1646062, 1646063, 1646064, 1647396, 1647397, 1652569, 1652570, 1654233, 1654234, 1656995, 1656996, 1656997, 1656998, 02469869, 03211913, 03217506, 03474751, 04479691, 06224414, 06225575, 06337219, 06341526, 06489178, 07225649, 07489910, 08349027, 08489909, 09226596, 09226597, 10356227, 11229552, 11357108, 12234610, 12357109, 12358930, 015162548, 016305918, 016306451, 017394609, 017400397, 017403582, 018513343, 018513344, 018521820, 018523881, 018523882, 019738626, 027406270, 027410096, 028542237, 028548055, 028548056, 028548062, 035191684, 035191685, 036319426, 037418951, 037420840, 038548054, 046329103, 046331161, 046331162, 047425769, 047426888, 047430967, 055225778, 056347830, 057432058, 057433632, 057436194, 058595423, 058595425, 067438321, 067442366, 067442479, 067446424, 067446425, 067448558, 068595426, 069876965, 069876966, 075255287, 076358119, 076361456, 077455530, 078595428, 078595430, 079891165, 079894758, 085258370, 085262947, 086369386, 086370993, 087459995, 087461380, 087465276, 087466267, 087469137, 088672403, 088672404, 088672405, 096372364, 096373205, 097471746, 098686643, 098686644, 098686645, 099937598, 104112739, 105265818, 106385210, 107481153, 107481154, 107483301, 108719992, 0111291154, 0111291943, 0111292218, 114132907, 114135079, 115290206, 117487787, 118740270, 118744198, 119983784, 119988027, 124153495, 125300048, 125301490, 127496634, 127513342,128738627, 128738628, 129996174, 0201048253, 0201048259, 0201066505, 0201066506, 0201066507, 0211311544, 0301076812, 0301076813, 0311324016, 0311324902, 0311325583, 0311329529, 0311329869, 0401098044, 0401098045, 0401098046, 0401100068, 0401100070, 0411345824, 0411346103, 0501101148, 0501101149, 0501128119, 0601150009, 0601150023, 0611366261, 0611367086, 0611368341, 0611369146, 0611369556, 0701168229, 0701169286, 0701169429, 0711373300, 0711381248, 0711389338, 0711390597, 0801188995, 0801189073, 0801192524, 0801192531, 0901192184, 0901202930, 0901203568, 0901204467, 0901204468, 1101243675, 1101245070, 1101246507, 1101247334, 1101256475, 1101256476, 1101257705, 1101258235, 1101261502, 1101261503, 1201261504, 1201262087, 1201262088, 1201265237, 1201265238, 1291001896 REF 26000019 Serial Numbers: 1401796, 1406150, 1408131, 1408232, 1419783, 1421313, 1421314, 1426641, 1430445, 1431154, 1432726, 1435036, 1435037, 1444788, 1446871, 1453902, 1466324, 1468716, 1472711, 1472712, 1472713, 1480599, 1480600, 1482073, 1483473, 1485542, 1488486, 1488487, 1490426, 1490540, 1494624, 1495078, 1496367, 1496368, 1497252, 1497253, 1503729, 1504265, 1511230, 1511233, 1511235, 1511958, 1511959, 1519173, 1519174, 1523541, 1526396, 1526397, 1526398, 1527826, 1532940, 1532941, 1534589, 1534590, 1542314, 1542315, 1543356, 1546506, 1548098, 1548099, 1553116, 1553117, 1553118, 1555113, 1555114, 1555115, 1561770, 1561771, 1567224, 1567225, 1567226, 1584584, 1584873, 1585000, 1593586, 1596667, 1601734, 1601735, 1601737, 1610139, 1612605, 1615704, 1618965, 1620426, 1622631, 1634730, 1634731, 1640316, 1640317, 1645758, 1645759, 1648022, 1648023, 1652752, 1652753, 1658819, 1658820, 02469870, 04474752, 05483518, 06346612, 06487166, 07348869, 07489911, 10352346, 10352748, 12358931, 015162545, 016305344, 016305905, 016306452, 017395721, 018521821, 018527649, 018528632, 019733515, 025191676, 025191677, 027406271, 036322012, 037415070, 037416897, 038548057, 038548058, 038548059, 047426244, 047426889, 055208759, 057432059, 057434554, 057438322, 058595432, 058595434, 066355137, 067442242, 067446427, 067448563, 075258158, 076358027, 076358747, 076360757, 077455531, 085255288, 086370992, 087461379, 087462400, 088595435, 088685248, 088685249, 096372365, 096374877, 097469510, 097479892, 0101001897, 0101018642, 104109640, 105269552, 105269553, 106373206, 106385212, 107481155, 107483329, 108685247, 108694954, 108703738, 0111283426, 114136099, 114136100, 117489010, 118739174, 118741129, 125300049, 125301516, 127496635, 127513345, 127513346, 127513347, 0201054130, 0201054131, 0201071228, 0201071229, 0211307793, 0211308184, 0301094599, 0301094600, 0301094601, 0301095303, 0301095306, 0311328007, 0311328322, 0401095307, 0411339436, 0411339437, 0411339438, 0511353339, 0511353340, 0601145547, 0601146379, 0601165146, 0601165561, 0611362537, 0611363313, 0611366309, 0611373793, 0711373882, 0711381474, 0711391927, 0711391991 0901196482, 0901196518, 0901196837, 0901197584, 1101236529, 1101237624, 1101238265, 1101239339, 1101255753, 1101255754, 1101256472, 1101256474, 1201269759, 1201269760, 1201275683 REF 26000020 Serial Numbers: 1405523, 1470079, 1502988, 1502989, 1523542, 1533183, 1544281, 1546507, 1553538, 1553539, 1561772, 1561773, 1587637, 1602314, 1615927, 1636055, 1638716, 1640919, 1648024, 1650421, 02469871, 06479693, 07346614, 08337217, 08351329, 11356228, 015162547, 018513348, 026319273, 035198677, 036321904, 037417014, 038548060, 047426243, 047427017, 055208760, 055208761, 056319428, 057432754, 058575583, 066356913, 067442243, 067448143, 068595437, 076360756, 076363042, 087455532, 087462040, 096373203, 097480710, 098694955, 107483330, 0111278279, 117487788, 127504546, 127507024, 128739175, 0211304893, 0211310148, 0301095294, 0511357836, 0711375057 REF 26000021 Serial Numbers: 01210330, 01210331, 01334608, 1389311, 1390744, 1392036, 1392704, 1399207, 1400720, 1404002, 1407102, 1407210, 1407488, 1409744, 1409954, 1415668, 1415669, 1416661, 1416662, 1426063, 1427396, 1431306, 1432314, 1434065, 1434066, 1435285, 1436957, 1438657, 1444329, 1444618, 1446165, 1455905, 1457130, 01460236, 1460645, 1460646, 1462754, 1468905, 1468906, 1470080, 1471505, 1473656, 1475743, 1476102, 1480460, 1482074, 1484469, 1484470, 1484769, 1484770, 1491260, 1491261, 1493552, 1496461, 1500102, 1500103, 1501282, 1503280, 1503730, 1507777, 1507778, 1511960, 1511961, 1511962, 1513943, 1513944, 1516242, 1516243, 1516969, 1516970, 1528275, 1528276, 1534592, 1534593, 1543353, 1543354, 1544279, 1544280, 1546508, 1555116, 1555117, 1555118, 1565734, 1565735, 1565736, 1568429, 1568430, 1581558, 1581559, 1581864, 1581865, 1590482, 1592574, 1594031, 1596668, 1597202, 1600813, 1600814, 1603004, 1610985, 1610986, 1611451, 1615929, 1619776, 1619777, 1634973, 1634974, 1640318, 1655444, 1655445, 1656992, 02465514, 02469872, 03210332, 03217507, 03217508, 04335947, 04339966, 04479694, 05483519, 06346617, 07489913, 08489912, 09228845, 09229554, 09351564, 09351689, 10356229, 11232604, 11355135, 12357450, 12359154, 12359799, 015162546, 016305126, 016314361, 018521822, 026316751, 026317142, 027394617, 027410097, 028521823, 028527650, 028528633, 028542658, 035191678, 035191679, 037415576, 037419471, 037419472, 038564682, 039824493, 039834733, 039837391, 039839758, 047424279, 047427018, 047431541, 049840545, 049843257, 049843791, 049843792, 049844822, 056343508, 057435218, 058595438, 058595439, 066347956, 067440502, 067442480, 067446428, 067448144, 068595440, 068595441, 069876612, 075250526, 076364822, 077459486, 078625404, 079890262, 079894024, 085267368, 087455533, 087466785, 087470688, 087471745, 088625406, 088672409, 088672411, 088685245, 096373201, 097480717, 098672410, 098685244, 098685246, 105258371, 106380199, 107483331, 109945428, 109960618, 0111277103 0111288491, 0111289477, 116385214, 117495022, 117495713, 118736946, 118740273, 119978890, 124140851, 125300062, 125304289, 126392066, 126394627, 127504547, 127507025, 0201035715, 0201044676, 0201050774, 0201059744, 0301072896, 0301072897, 0301081076, 0301081077, 0301085935, 0311310059, 0311310060, 0311315004, 0411339095, 0411339096, 0411339097, 0411339098, 0501115949, 0501116919, 0501118998, 0501119678, 0511354516, 0511354517, 0511357202, 0601142471, 0601148205, 0601151515, 0611362538, 0611365401, 0611366310, 0611374826, 0701162968, 0701163563, 0701171826, 0711388906, 0801188058, 0901197221, 0901206707, 0901206708, 1001218546, 1001228991, 1101243617, 1101251429, 1101256240, 1101256241, 1101256242, 1101256243, 1101260335, 1201267601, 1201267602, 1201273043, 1291000330, 1291017281, 1291017718, 1291017719, 16008131604573 REF 26000022 Serial Numbers: 01210333, 01210744, 01210745, 01334612, 1392476, 1392705, 1396837, 1396841, 1399341, 1400518, 1402203, 1404278, 1404489, 1407215, 1407487, 1409305, 1409515, 1414006, 1414824, 1414825, 1416411, 1419552, 1419849, 1421315, 1421316, 1425082, 1426584, 1426585, 1428862, 1428863, 1431009, 1431152, 1431153, 1433645, 1433646, 1435286, 1435680, 1436958, 1437263, 1437264, 1440450, 1441515, 1444036, 1444195, 1444785, 1444786, 1444975, 1447129, 1448611, 1456416, 1457554, 1460129, 1460130, 1462756, 1464086, 1468466, 1468718, 1470779, 1471688, 1473070, 1476103, 1477378, 1483275, 1483276, 1485330, 1485331, 1485364, 1485365, 1485541, 1487539, 1493551, 1497572, 1498036, 1498037, 1501036, 1501037, 1501327, 1503281, 1503731, 1506325, 1506326, 1507935, 1507936, 1510461, 1510462, 1510463, 1513945, 1513946, 1513947, 1515635, 1515636, 1515694, 1515695, 1519177, 1523543, 1528502, 1528503, 1532401, 1532402, 1536699, 1536700, 1536701, 1541033, 1541034, 1541035, 1544285, 1544286, 1544287, 1546973, 1546974, 1554790, 1554791, 1554792, 1556779, 1556780, 1556781, 1556782, 1561463, 1561464, 1561465, 1561466, 1567227, 1567228, 1567229, 1569901, 1569902, 1569903, 1569904, 1582251, 1582252, 1590077, 1590461, 1590462, 1591051, 1592576, 1593587, 1594032, 1594881 1597203, 1599335, 1600811, 1600812, 1601731, 1601732, 1606609, 1606610, 1607592, 1608836, 1610561, 1610562, 1611452, 1615572, 1615705, 1616178, 1620678, 1620679, 1620680, 1634975, 1634976, 1637238, 1648025, 1648026, 1648619, 1648620, 1655446, 1655447, 1658514, 1658515, 02460237, 02461648, 02468619, 02469873, 03211705, 03211906, 03211910, 03217509, 03217510, 03335948, 03339965, 03474753, 04217269, 04479695, 05483520, 06219768, 06219769, 06341527, 06345236, 06346616, 06489179, 07489914, 08348870, 08349026, 09229551, 09351690, 10351707, 10356230, 11232605, 12355136, 12357110, 015162435, 015162436, 016294652, 016303114, 016306455, 018521824, 018521825, 018523883, 018523884, 027412434, 035191680, 035191681, 035198491, 036322010, 037414680, 037416655, 037419752, 037420843, 038544856, 038544857, 038548061, 039819344, 039821131, 039825300, 039836605, 039837392, 046319425, 047426891, 047429086, 049832652, 049838305, 049840544, 049843310, 056331164, 056343507, 057435219, 058595442, 058595443, 065244421, 066347949, 067440182, 067442481, 067442482, 067446429, 068595444, 068595445, 068595446, 069872992, 069872993, 069876967, 069876968, 075245202, 076360910, 076363530, 076365178, 076365891, 077450128, 077455534, 078625407, 078625408, 079894023, 079894759, 085258372, 086369487, 087461047, 087462399, 087466268, 087469126, 087470003, 088672400, 088672401, 088672402, 089904047, 094104770, 096371131, 096373883, 096374876, 097480718, 098686646, 098686647, 098686648, 098694956, 099929426, 099937599, 0101000331, 0101016260, 0101029687, 105265819, 105275571, 105286916, 107483332, 108706225, 108706226, 108713332, 109944907, 109955347, 109958793, 0111277102, 114112742, 116385211, 117487789, 117496636, 118738629, 118738644, 118739655, 118744193, 119976423, 119976424, 124145838, 125307855, 126392067, 126395276, 126395722, 127504548, 127505594, 127507026, 129983786, 129998778, 0201029688, 0201053792, 0201053814, 0201062914, 0201062915, 0201062916, 0201062917, 0211300500, 0211301131, 0211302254, 0301078356, 0301080502, 0301089723, 0301089724, 0301089725, 0301089726, 0311318061, 0311318866, 0311321116, 0311322523, 0401106430, 0401108732, 0401109393, 0401109394, 0401112109, 0411333202, 0411339439, 0411339440, 0411339441, 0411339442, 0411339443, 0411346844, 0411346845, 0501115011, 0501115559, 0501124773, 0501124864, 0501131797, 0501132629, 0501142048, 0501142198, 0511346962, 0511353341, 0601147116, 0601147180, 0601160630, 0611353342, 0611359114, 0611371891, 0611371941, 0701161822, 0701172647, 0701173465, 0701173482, 0701180405, 0701181072, 0711372612, 0711380277, 0711389319, 0711389337, 0711390551, 0711390596, 0801188152, 0801188638, 0901195838, 0901205530, 0901205532, 0901206125, 1101247390, 1101248665, 1101249266, 1101249421, 1101250724, 1101252573, 1101253677, 1101253729, 1201270046, 1201270047, 1201270048, 1201270049, 1201271744, 1201271756 REF 26000023 Serial Numbers: 01210328, 01210329, 01210334, 01334607, 1391134, 1391967, 1392022, 1399342, 1399523, 1410249, 1410250, 1414894, 1414895, 1419850, 1420737, 1425873, 1427398, 1431010, 1431453, 1435893, 1435894, 1437326, 1444448, 1444787, 1445941, 1448012, 1469526, 1470780, 1472881, 1473067, 1476607, 1479549, 1479550, 1480095, 1482752, 1484771, 1485538, 1489820, 1490428, 1491266, 1491268, 1496462, 1498038, 1498039, 1501283, 1501284, 1503282, 1503732, 1509295, 1509296, 1513948, 1513949, 1513950, 1516244, 1516245, 1523544, 1528086, 1528087, 1540934, 1540935, 1540936, 1543357, 1543358, 1548273, 1555119, 1555120, 1555121, 1557969, 1557970, 1566398, 1566399, 1566400, 1568427, 1568428, 1570710, 1570711, 1570712, 1590484, 1592575, 1594033, 1600184, 1600810, 1607802, 1607803, 1611262, 1611450, 1615573, 1622632, 1622633, 1640627, 1648367, 1648626, 1656993, 1656994, 02210336, 02461298, 02461649, 03210335, 04335950, 04339969, 04479696, 05483521, 06219765, 06219766, 07490491, 07492049, 08351706, 09226599, 09353291, 09353668, 10356231, 11226598, 11358852, 12235138, 12355137, 12359801, 015158578, 015165779, 016306454, 016314362, 017398847, 017403583, 018521826, 018523885, 025170081, 025191038, 025191196, 027406272, 027410274, 036319427, 036322013, 037418952, 037420844, 038544858, 039821130, 039824494, 039832653, 045198492, 047426892, 048575584, 048585743, 049838306, 049841551, 057432755, 057436426, 058594216, 058595447, 058595448, 065245201, 066355138, 067442483, 067446430, 067450512, 068595466, 069879081, 075249042, 076358028, 076360758, 076363043, 077455535, 078625409, 078625410, 079890265, 079890267, 079890269, 079891166, 079894760, 084100343, 085258373, 086364823, 087459184, 087462044, 088679474, 089927667, 096373885, 097479891, 097480719, 098693693, 099931101, 099931102, 0101022067, 0101022068, 104109600, 105265820, 105277296, 107483333, 108693692, 108694957, 109955348, 0111275709, 0111286672, 114132908, 116389340, 117495023, 118706227, 118743308, 119976425, 119976426, 124153494, 125300061, 125303468, 125306189, 126391601, 127504549, 127505595, 127507027, 128739654, 129998780, 129998781, 0201053794, 0201053815, 0201071230, 0201071231, 0301080504, 0301089727, 0311314247, 0311316420, 0311329145, 0311329525, 0311329868, 0311330086, 0311330556, 0311332183, 0401105191, 0401105658, 0401110401, 0401111498, 0401111632, 0411339444, 0411339445, 0501120033, 0501125521, 0501137716, 0501138318, 0501138642, 0601138735, 0601164238, 0601165033, 0611370671, 0711370344, 0711370672, 0711379604, 0711380154, 0801182759, 0801182761, 0801193309, 0801193417, 0801194319, 1101240494, 1101240495, 1101240526, 1101241565, 1101242470, 1101243259, 1101243260, 1201267537, 1201267538, 1291016261 REF 26000024 Serial Numbers: 01210341, 01334609, 1406711, 1435113, 1449965, 1473068, 1501285, 1504260, 1506157, 1506158, 1516246, 1523545, 1532942, 1540932, 1540933, 1546509, 1546972, 1564765, 1564766, 1567230, 1567231, 01583522, 1600185, 1608033, 1611453, 1634732, 1635410, 1639319, 1639320, 1645174, 1648368, 1648627, 03474754, 04335949, 04335951, 04479697, 05485501, 06219767, 08346615, 09226600, 09229553, 09497828, 10356232, 12359155, 12359802, 016305926, 01630645, 018521827, 018521828, 018523886, 025179480, 026318355, 037416746, 047426893, 047431540, 048585738, 055225777, 056343509, 057432756, 057436489, 066347957, 067445022, 075245203, 077450627, 078595467, 085258374, 086364821, 087455536, 087469055, 095264009, 096374875, 105277299, 107486693, 108660967, 109947893, 109964117, 118733325, 124136101, 124158576, 126389857, 126395277, 127512693, 128733324, 128738339, 129998782, 0201044379, 0211308909, 0301078357, 0301100065, 0301100067, 0311316582, 0401106035, 0501127371, 0611373287, 0711373247, 0901194832, 0901194834, 1201266194, 1201274863, 1291016262 REF 26000025 Serial Numbers: 1395467, 1396575, 1397861, 1398723, 1398896, 1399773, 1399814, 1400721, 1402023, 1404222, 1404490, 1405724, 1405892, 1410826, 1410992, 1416412, 1417307, 1417308, 1417777, 1417778, 1426064, 1426586, 1427399, 1429434, 1430350, 1430351, 1432723, 1432724, 1432725, 1434063, 1434260, 1434261, 1441514, 1447130, 1452147, 1452148, 1456731, 1459787, 1459807, 01460238, 1461406, 1461407, 1463852, 1464781, 1464782, 1468041, 1470963, 1471176, 1472042, 1473069, 1475024, 1476982, 1479895, 1479896, 1480428, 1480461, 1483471, 1484772, 1484773, 1490430, 1490539, 1497527, 1497528, 1497573, 1497574, 1497575, 1502049, 1505250, 1505251, 1505592, 1505876, 1505877, 1506699, 1506700, 1509303, 1509305, 1513951, 1513952, 1513953, 1520540, 1520541, 1520542, 1521200, 1521201, 1521731, 1522475, 1523886, 1525248, 1528616, 1528617, 1528618, 1529767, 1530737, 1530738, 1531372, 1531559, 1533184, 1533185, 1539540, 1539541, 1539542, 1539543, 1543355, 1544022, 1544288, 1544289, 1546216, 1548100, 1548101, 1548793, 1548794, 1550353, 1550354, 1553544, 1553545, 1553546, 1554785, 1554786, 1556783, 1556784, 1556785, 1556786, 1560840, 1560841, 1565737, 1565738, 1565739, 1565740, 1570920, 1570921, 1570922, 1576847, 1576848, 1576849, 1576850, 1582750, 1582751, 1582752, 1583512, 1589538, 1589725, 1590828, 1591512, 1592284, 1592377, 1592577, 1598152, 1598153, 1603957, 1603958, 1605902, 1605903, 1608031, 1608837, 1610134, 1613897, 1613898, 1613899, 1613900, 1615401, 1617609, 1617610, 1621175, 1621176, 1633866, 1633867, 1637239, 1637240, 1637400, 1637401, 1640920, 1641177, 1641266, 1643942, 1643943, 1645360, 1645361, 1649376, 1649377, 1649378, 1653527, 1653941, 1655093, 1655094, 1655095, 1660417, 1660418, 1660419, 02461297, 03210340, 04210347, 04474755, 04479698, 06485502, 08348872, 09494170, 12356233, 016311730, 016311731, 017398862, 017404755, 018523887, 025179482, 037420327, 038564678, 038566589, 038567197, 039809783, 039828327, 039831048, 047426894, 048585740, 049839757, 049842582, 055211009, 057432757, 057438323, 058595338, 058605218, 065227870, 067442484, 068625412, 069870369, 075250528, 077455256, 078651460, 078660968, 079890271, 079891167, 087465273, 087469506, 087471663, 088679475, 089905087, 089912891, 089915882, 089924905, 089926242, 089929627, 097473752, 099937600, 0101013692, 0101020921, 0101020922, 101046898, 0101048291, 0101048292, 104109643, 105258375, 105275573, 107482923, 107486694, 108688676, 108694958, 108713333, 109944908, 0111293922, 0111295010, 114131981, 117506038, 117506039, 119970864, 119986953, 124135080, 124135081, 125294651, 125307997, 126396172, 127507028, 128743303, 129989231, 0201071232, 0201074285, 0211296735, 0211297626, 0301071233, 0301074284, 0311324901, 0311325545, 0311325581, 0401099033, 0401101150, 0401104097, 0401112925, 0401121017, 0401121190, 0401121935, 0411334830, 0411334831, 0411335373, 0411335374, 0411335375, 0411335376, 0501128425, 0501128927, 0501136396, 0501137369, 0601147954, 0601154260, 0601154738, 0601158914, 0601159616, 0611369557, 0611370352, 0611371971, 0611371972, 0611372616, 0701171293, 0701171921, 0701171982, 0701181919, 0701182353, 0711388280, 0711388281, 0711389894, 0711389895, 0901192133, 0901200718, 0901200724, 0901201439, 0901212159, 1001212319, 1201260895, 1201260897, 1201260899, 1201262085, 1201262086, 1201265234, 1201265235, 1201275684, 1291013693 REF 26000026 Serial Numbers: 1395468, 1396342, 1396576, 1396577, 1398013, 1399012, 1399092, 1399343, 1399780, 1399781, 1400013, 1400722, 1402502, 1406151, 1407809, 1407837, 1407993, 1409086, 1411158, 1411233, 1413507, 1413958, 1416659, 1416660, 1418715, 1420561, 1420562, 1421012, 1421014, 1424399, 1425874, 1425875, 1427400, 1427401, 1427724, 1429435, 1429436, 1431155, 1431156, 1434064, 1434106, 1434262, 1434263, 1434947, 1435284, 1436537, 1437780, 1438128, 1445942, 1448771, 1452342, 1453180, 1453903, 1456733, 1457555, 1458651, 1459358, 1459550, 1462757, 1463679, 1466554, 1468039, 1468040, 1469680, 1470781, 1471177, 1471506, 1472043, 1473328, 1474204, 1475084, 1476462, 1476777, 1476778, 1479030, 1479031, 1479551, 1479552, 1483274, 1483468, 1483469, 1483470, 1490527, 1490530, 1497491, 1497576, 1497577, 1497579, 1497580, 1497582, 1501328, 1502050, 1502051, 1502052, 1505252, 1505253, 1505254, 1505255, 1505593, 1505594, 1505878, 1506697, 1506698, 1507570, 1507571, 1508081, 1508082, 1509239, 1509240, 1509307, 1509308, 1510887, 1511963, 1511964, 1511965, 1520543, 1520544, 1520545, 1521418, 1521419, 1521420, 1521421, 1521732, 1522474, 1523631, 1523632, 1523633, 1523885, 1525249, 1525250, 1529221, 1529222, 1529223, 1529224, 1530739, 1530740, 1530741, 1531560, 1531561, 1533186, 1533187, 1533188, 1534594, 1534595, 1539544, 1539545, 1539546, 1539547, 1540823, 1540824, 1540825, 1543359, 1543360, 1543361, 1544290, 1544291, 1546217, 1547297, 1548102, 1548103, 1548795, 1548796, 1550351, 1550352, 1551552, 1552126, 1552127, 1554782, 1554783, 1554784, 1556787, 1556788, 1556789, 1556790, 1560838, 1560839, 1560940, 1560941, 1560942, 1566401, 1566402, 1566786, 1566787, 1566788, 1566789, 1570923, 1570924, 1572013, 1572014, 1572015, 1577848, 1577849, 1577850, 1577851, 1578039, 1578040, 1578328, 1578329, 1588738, 1588739, 1588740, 1588741, 1589114, 1589126, 1589127, 1593366, 1597204, 1598154, 1598155, 1599336, 1603005, 1603006, 1603007, 1603697, 1603959, 1605900, 1605901, 1607569, 1607570, 1608032, 1608838, 1609813, 1609814, 1614087, 1614088, 1614089, 1614985, 1614986, 1614987, 1614988, 1615400, 1615926, 1618513, 1618514, 1618515, 1618516, 1619775, 1621177, 1621178, 1622634, 1623480, 1623481, 1633691, 1633692, 1637402, 1637403, 1641265, 1643612, 1643613, 1643614, 1644908, 1644909, 1646058, 1646400, 1649158, 1649159, 1649160, 1650957, 1650958, 1652571, 1653580, 1653940, 1654505, 1654506, 1654507, 1655091, 1655092, 1657876, 1661716, 1661717, 02461650, 03211706, 03474756, 04338849, 04342275, 04479699, 08346613, 08489180, 08494179, 09227484, 09497825, 09499461, 10356234, 12357091, 015158579, 015170083, 016303113, 016307998, 018521829, 018521830, 027408822, 027410505, 028527651, 029809784, 038564680, 039821129, 039828329, 039828331, 039833520, 047424093, 047426440, 048575585, 048585745, 048594217, 049836606, 049839163, 049840542, 055211010, 057438324, 058595468, 058595469, 058605219, 058605220, 065244422, 067442485, 068625413, 069862327, 069871595, 069879080, 075227871, 076362895, 077455258, 077455265, 078651462, 078660969, 079885335, 079901147, 085258376, 085262948, 087465274, 087467823, 087469507, 088679476, 088688677, 088688678, 089885334, 089904048, 089922648, 089926243, 097470852, 097473753, 097478810, 099929469, 099929470, 099937601, 0101022069, 0101022070, 0101029689, 0101029692, 0101044381, 0101044382, 104115078, 104129606, 104130623, 106373884, 107482924, 107486695, 108694959, 108706229, 108713693, 109943609, 109970865, 109970866, 0111276616, 0111276617, 0111287723, 114131982, 114132905, 114132906, 114140850, 115275572, 116388902, 117500340, 119972079, 119973176, 119986954, 119986955, 119989232, 124145839, 126391570, 126396174, 126397419, 127504550, 127507029, 128742803, 129990537, 0201059390, 0201059391, 0201059392, 0201059394, 0201074286, 0211306009, 0211306393, 0211307060, 0301074287, 0301084473, 0301084475, 0311319152, 0311320056, 0311320057, 0311322522, 0311324014, 0311331668, 0401099034, 0401101151, 0401101152, 0401104098, 0401112937, 0401122034, 0411331739, 0411332563, 0411333201, 0411334772, 0411344343, 0411344344, 0411345497, 0501116687, 0501123240, 0501123311, 0501127070, 0501127370, 0501133074, 0501134099, 0501134957, 0501143656, 0501143754, 0601145033, 0601157869, 0611358067, 0611358252, 0611359163, 0611376329, 0611378284, 0611378752, 0611378940, 0701157306, 0701170750, 0701170801, 0701177942, 0701178446, 0701178474, 0711388282, 0711388907, 0711389148, 0711390647, 0711392706, 0801189095, 0801189691, 0801189731, 0801197634, 0801198120, 0901198291, 0901199318, 0901201662, 0901202102, 0901202875, 0901213558, 0901213853, 0901214167, 1001214166, 1101243676, 1101245036, 1101246508, 1101247356, 1101248453, 1101248667, 1101249422, 1101250726, 1101252574, 1101253678, 1101253730, 1291013696, 1291013697, 16436121646767 REF 26000027 Serial Numbers: 1395469, 1396578, 1397181, 1397182, 1399209, 1399524, 1400519, 1401808, 1404831, 1405271, 1408373, 1408406, 1411594, 1412034, 1415989, 1415990, 1417309, 1417310, 1417959, 1417961, 1422753, 1422754, 1427402, 1430352, 1430353, 1431454, 1435173, 1435283, 1435895, 1435896, 1438129, 1446145, 1449059, 1455939, 1455940, 1458005, 1458006, 1462443, 1462690, 1464087, 1466555, 1469681, 1469922, 1470782, 1471690, 1472268, 1476179, 1476180, 1476981, 1481230, 1481231, 1483277, 1484468, 1484776, 1490517, 1497493, 1497529, 1497581, 1498055, 1498784, 1498785, 1502053, 1505151, 1505152, 1505153, 1505177, 1505879, 1505880, 1507568, 1507569, 1508079, 1508080, 1511966, 1511967, 1511968, 1513954, 1513955, 1521202, 1521729, 1521730, 1522471, 1522472, 1523634, 1523635, 1523887, 1528949, 1528951, 1528953, 1528954, 1530742, 1530743, 1531373, 1531562, 1533189, 1533190, 1534596, 1536702, 1536704, 1540820, 1540821, 1540822, 1543362, 1543985, 1543986, 1546218, 1546219, 1548274, 1548275, 1552128, 1552129, 1554779, 1554780, 1554781, 1560032, 1560033, 1560034, 1560035, 1561774, 1561775, 1561776, 1565992, 1567763, 1567764, 1572471, 1572472, 1572473, 1576047, 1576048, 1576049, 1576050, 1587914, 1587915, 1587917, 1590829, 1592378, 1593368, 1597205, 1598156, 1598157, 1600182, 1600183, 1603961, 1603962, 1604754, 1605893, 1607588, 1607590, 1609169, 1610136, 1614758, 1614759, 1618962, 1618963, 1618964, 1621629, 1623482, 1623483, 1623484, 1634154, 1634155, 1639541, 1640921, 1640922, 1643615, 1643616, 1644911, 1646059, 1649611, 1649612, 1649613, 1654237, 1654238, 1654239, 1657877, 02461294, 02461651, 03210338, 03210348, 04479700, 06342276, 07489181, 08348868, 09494180, 09497826, 10352347, 12360241, 017399379, 017404756, 027407673, 028523888, 028527652, 035179481, 035198676, 038552347, 039809786, 039830315, 039832650, 039832651, 046330150, 047426441, 048585764, 049841552, 056340859, 057434553, 057438325, 058594218, 058595471, 066355139, 067442486, 068595470, 068605221, 069870370, 075250530, 076358512, 077455261, 078605222, 079890273, 079890275, 085258377, 087465275, 087467824, 087469508, 087470853, 088683522, 088683523, 089901148, 089912892, 089915884, 089922649, 097473754, 097478811, 099937602, 0101013698, 0101013699, 0101029693, 0101029696, 105269554, 105269555, 106366811, 107482925, 107486696, 107486697, 108713334, 108713694, 109949207, 114135083, 118733326, 119973178, 119986956, 124140852, 126390045, 127500341, 127504551, 127507030, 128737887, 128742804, 129990538, 0201053796, 0201053797, 0201066508, 0201071234, 0201071235, 0301081713, 0311314214, 0311317715, 0311325658, 0311329527, 0401096839, 0401104099, 0401114367, 0401114509, 0411329867, 0411330087, 0411330599, 0411344036, 0411344037, 0501126458, 0501126876, 0501136266, 0501139842, 0601148122, 0601155524, 0601156525, 0611368954, 0611368955, 0701169799, 0701170011, 0701170120, 0711374908, 0711387924, 0711387925, 0711388283, 0711390745, 0711392707, 0801190503, 0801191385, 0901190508, 0901191344, 0901209976, 0901210580, 0901211643, 1101258587, 1101259118, 1101259122, 1101261500, 1201261499, 1201261501, 1201262080, 1201262082, 1201262083, 1201265229, 1201265232 REF 26000028 Serial Numbers: 1391135, 1400287, 1406168, 1412299, 1415479, 1417633, 1420013, 1435172, 1438656, 01457092, 1457966, 1462758, 1470964, 1472301, 1473329, 1485537, 1489846, 1498786, 1502054, 1505154, 1505595, 1507566, 1507567, 1510370, 1510464, 1511227, 1511228, 1522473, 1524014, 1531563, 1534597, 1539548, 1539549, 1543363, 1543364, 1546220, 1548276, 1548277, 1553774, 1553775, 1560036, 1560037, 1561777, 1561778, 1561779, 1572474, 1572475, 1572476, 1588490, 1588491, 1593367, 1600180, 1600181, 1605894, 1606608, 1609170, 1614090, 1615399, 1616177, 1617417, 1618062, 1621875, 1623485, 1633439, 1633440, 1636636, 1636637, 1641264, 1643611, 1644910, 1645359, 1646060, 1646061, 1648946, 1648947, 1650955, 1650956, 1653939, 1654235, 1654236, 1657878, 1661713, 1661714, 04339973, 04479701, 06210349, 06345233, 07487219, 08348871, 09489182, 09494181, 12356235, 015170084, 016308000, 017404757, 018507031, 028523889, 039825936, 047426442, 057438326, 058602426, 069870373, 077455264, 078657274, 078660970, 085262946, 087468371, 087469509, 088679477, 089905088, 089915885, 107489490, 108700628, 109938838, 109947891, 109960510, 117495714, 119986824, 119986957, 124135082, 126396173, 128755754, 0201048290, 0301074288, 0301081719, 0301081720, 0301089732, 0311316620, 0411343205, 0411343206, 0548352301, 0611365946, 0611374907, 0701178782, 0901196481, 0901208792, 0901208996, 16453591651601 Conserve REF 38013800 Serial Numbers: 01336252, 01336648, 01465788, 01467492, 1561766, 02338825, 02339631, 02470528, 02470533, 02473240, 03477697, 04481833, 05342022, 07484291, 08348056, 09353669, 10232864, 10355850, 11234642, 12360787, 017397516, 018523897, 037420861, 037423780, 038568859, 039825303, 039838301, 048575671, 049850492, 057432005, 058594224, 065233869, 067449131, 076362909, 085257534, 086370585, 087460705, 088685275, 105270444, 116388901, 127515855, 128746770, 0201030497, 0201044688, 0211296900, 0211299343, 0301083872, 0401117838, 0601145349, 0801192556, 0901197216 REF 38013804 Serial Numbers: 01336643, 01461784, 01467493, 02338605, 02339237, 02339632, 02470529, 03476222, 04481835, 05342021, 06483920, 06485107, 06485503, 08348055, 09353670, 10231641, 11234641, 12360249, 017398845, 028544861, 037420822, 037423183, 047431575, 048577483, 049840536, 058594223, 067444759, 067449132, 078608779, 085257535, 086370586, 088685276, 105285957, 107481164, 116389030, 117496647, 128752498, 0411322524, 0601145350, 0701165734 REF 38013835 Serial Numbers: 01336255, 1394264, 01465477, 01467494, 02338611, 02339630, 02470530, 03341483, 03477702, 04481837, 05342023, 07485504, 08347658, 09352749, 10231504, 10233020, 10355852, 11355957, 12361305, 037420869, 038568860, 047427004, 049840535, 058594222, 067444760, 085260582, 087459469, 088669038, 095270445, 098690182, 116389877, 117495024, 127515856, 128744507, 0548392101, 0601145351, 0601161457, 0901192558, 0901197218 REF 38014000 Serial Numbers: 01336253, 01336649, 1374235, 01465478, 01465789, 01466539, 01467495, 01469311, 1490572, 1502992, 1581556, 02338827, 02339635, 02470531, 03341480, 03477698, 04341945, 04481838, 06485505, 07341617, 07347292, 07483918, 10232865, 10354558, 11234910, 12360252, 016306458, 018515857, 018523898, 019785728, 019789145, 027405945, 028536056, 029806011, 037418660, 037420871, 038568861, 039822086, 039839161, 039839162, 049853371, 055226999, 057432062, 057435216, 058602436, 059864554, 059869521, 065227204, 067441994, 067447212, 068605242, 076360810, 077453355, 078635570, 085255783, 086369817, 087466227, 087471545, 088682824, 098693402, 105271450, 106375750, 115297572, 117497153, 126394101, 127501987, 128747419, 0201036774, 0201036775, 0201072753, 0201072754, 0401108634, 0401109396, 0411323468, 0711373794, 0711374911, 0801185466, 0901204267, 0901205077 REF 38014004 Serial Numbers: 01336250, 01336645, 1414567, 01461785, 01465481, 01467496, 1649368, 02469312, 02470532, 2639633, 03341477, 04341612, 04341941, 04481840, 06483919, 07347288, 10232862, 10354559, 11234911, 11359156, 12360788, 015176785, 015176787, 016306460, 017403689, 018519213, 019780240, 036305869, 037407225, 039815508, 048577485, 049841538, 049848157, 057432820, 057436026, 059869522, 065227004, 068608782, 085255233, 085255780, 087460706, 088669037, 088690183, 096372713, 116392086, 117497154, 124139818, 124140865, 127497155, 128751148, 0611360339, 0801192560 REF 38014035 Serial Numbers: 01336256, 1417128, 1454806, 01461786, 1468903, 01469313, 1475846, 1476253, 02338612, 02339244, 02473431, 02473525, 03341484, 04341621, 04481842, 06483917, 07346551, 09352348, 10233021, 10354693, 11233655, 11234643, 12360789, 12495573, 017400575, 018523899, 038565206, 038568862, 039815509, 045213408, 046315320, 046334759, 047425750, 049855231, 057433176, 066342257, 068625439, 068635572, 076360815, 077453356, 078635571, 079892758, 085255785, 088672396, 097459154, 097477578, 098693407, 0101017300, 0101027432, 104130609, 105273942, 106378167, 127513351, 0401102851, 0501124767, 0611360340, 0701165735, 0901197219 REF 38014200 Serial Numbers: 01336254, 01336651, 1383899, 1450801, 1452150, 01467802, 1468904, 01469314, 1471691, 1474067, 1475847, 1486196, 1647723, 02338829, 02473318, 02473434, 02473526, 03341481, 03341946, 03477700, 05341618, 05483752, 05483914, 07347287, 08351687, 09353671, 10232866, 11233657, 11234912, 11357112, 12359713, 12361304, 015173214, 016303101, 018521832, 019789146, 025182500, 026302357, 026313046, 027405896, 027407226, 028530855, 029798207, 029808009, 035200369, 037420824, 039821653, 039821654, 039827548, 047426896, 048577500, 048594221, 049849871, 049850490, 049852500, 057433177, 057435217, 065227003, 066352027, 066354242, 066354706, 067445906, 068625441, 069876959, 069881894, 075250772, 075253434, 076362910, 077457677, 078625442, 078657132, 085252993, 086365774, 086368897, 087460707, 087466228, 087469135, 088672394, 088672395, 096375115, 097472432, 104104762, 104115109, 105271451, 107489494, 108716197, 115296910, 116389031, 116392085, 117496648, 117496649, 118740239, 119983797, 124140855, 126395716, 127501988, 127519214, 128740240, 128763683, 128763685, 129998822, 0201036776, 0201064278, 0201064279, 0201072756, 0301072755, 0301072757, 0401106377, 0411327097, 0511351298, 0511352156, 0601158630, 0611351300, 0701158960, 0801188157, 0801192107, 0901192472, 0901193286, 1001231167, 1001231749 REF 38014204 Serial Numbers: 01336647, 1384696, 01465482, 01469315, 1474381, 1474535, 1649369, 02338824, 02339636, 02473527, 03341478, 03341613, 04482820, 05341942, 05483915, 06483697, 07347290, 08351688, 09353672, 10232863, 11233654, 11234913, 11357353, 11359157, 12360253, 015140867, 015176792, 016303102, 017405902, 028544862, 029798204, 035200370, 037407227, 038547962, 039815510, 039834731, 048568863, 048575681, 049847487, 049849872, 056340860, 056345833, 057431797, 057439226, 059863220, 065227005, 066354553, 068602438, 069879076, 075255779, 076361464, 077449133, 085255232, 087460708, 087467552, 088648549, 088672397, 096372714, 097472433, 104104763, 104106791, 105275578, 106379491, 108690184, 115287584, 117495025, 117496651, 118744194, 126394102, 126398008, 128743305, 128763682, 0311310058, 0411315011, 0601148208, 0701173504 REF 38014235 Serial Numbers: 01336257, 01336653, 1383276, 1383883, 01465480, 01469316, 02339245, 02473528, 03341485, 04482821, 05341948, 05483916, 07347289, 09352349, 09495574, 10231529, 10233142, 10233143, 11233656, 11234914, 11357567, 12359714, 12361303, 015145840, 016302010, 019767192, 027410607, 028552902, 029789337, 035200368, 037415866, 038571500, 039815394, 039828959, 049839754, 055227001, 056338866, 056345834, 057432818, 067445905, 068602440, 069879077, 075227206, 076362911, 077457316, 086372051, 088669036, 095273947, 097474459, 0101016267 104130619, 105285958, 106385457, 108685272, 109949750, 0111287777, 117495027, 117496667, 126394103, 127513352, 0301083873, 0301083874, 0311325584, 0401105660, 0501117711, 0501128449, 0501131024, 0711352902 REF 38014400 Serial Numbers: 01336832, 01336907, 01337112, 1381688, 1382261, 1384741, 01461663, 01465534, 01468468, 02338767, 02339444, 02340841, 02470399, 02473319, 03340996, 04480123, 04480855, 07344293, 07349405, 07485506, 09353673, 09495575, 10230893, 10354694, 11234389, 12235100, 12360255, 12361306, 015158580, 015167732, 016310202, 017398864, 018507049, 019785730, 026302358, 028527662, 029785729, 029789339, 029789342, 035193943, 035194618, 037411631, 037415536, 038547963, 039823322, 046329111, 047424301, 047426472, 047430965, 049841539, 049847492, 049850493, 049850495, 049851824, 055227006, 056337057, 057436195, 058602442, 058602444, 059857913, 059859140, 059864556, 066349389, 066351972, 066355933, 067445904, 068608783, 069876960, 069881895, 075249050, 075250536, 076358752, 076366556, 077449077, 077450715,078635573, 078657134, 078672393, 085252991, 086368890, 086369816, 086371419, 087459996, 087464162, 088672392, 095273944, 095275579, 097470658, 097475394, 098693403, 098700614, 098701782, 104106789, 106376552, 107480653, 107489019, 108716198, 114135084, 115287546, 115287547, 116392084, 117495026, 117496650, 117501989, 118742673, 118748676, 124140858, 125300600, 126392481, 126395717, 128745407, 128748675, 0201001912, 0201066515, 0211288441, 0301072758, 0301072759, 0301081725, 0301081726, 0401108181, 0411326366, 0411326680, 0601141672, 0601142039, 0601159840, 0601165147, 0801198290, 0901197586, 1001214887, 1101241577 REF 38014404 Serial Numbers: 01336830, 01336905, 01465483, 02338762, 02339439, 02465535, 02472353, 02473553, 03341021, 04480124, 04480857, 05340994, 06485507, 08349406, 10230888, 10354695, 11234398, 12235394, 12360257, 016310200, 016311873, 017399471, 025193944, 028547964, 029802357, 035199114, 035199115, 035200364, 035200365, 037412686, 039830313, 047425771, 047426473, 048552379, 048577504, 049843294, 049851163, 056340862, 056342430, 057430966, 059859141, 066346316, 067445903, 067448528, 069879078, 069881896, 076362305, 078641614, 078657131, 078672388, 078672389, 085255782, 085260579, 086368436, 086371010, 086371420, 087460709, 087469499, 094104764, 098693408, 104106788, 105273951, 105277333, 106376551, 114135086, 116389359, 116392083, 117489495, 117496668, 118742676, 124140861, 125287592, 126395718, 127501990, 128746771, 128753687, 0301089472, 0401089473, 0601148935, 0601151658, 0801190510, 1001191346 REF 38014435 Serial Numbers: 01336833, 01336910, 1383881, 01465484, 01468469, 02338770, 02340843, 02473529, 02473551, 03476732, 04341023, 05483753, 06485508, 07347294, 10231600, 10354560, 11234400, 11357568, 12235118, 12360259, 015173215, 016309000, 017395266, 027409853, 029795564, 035194615, 035203185, 037420875, 038552903, 038573156, 039816373, 039830314, 046335924, 047426474, 048571502, 049848066, 059853372, 059859722, 059867733, 059869520, 065248214, 066348363, 066355934, 067444761, 067448994, 068625447, 069871614, 069879079, 076361925, 078625449, 078662885, 078669035, 079887541, 085255784, 085269116, 086366557, 086368437, 086369820, 095271452, 097469498, 097476275, 097476520, 098693405, 106375751, 115287548, 116389876, 117497156, 118739656, 124139823, 127497157, 0401100080, 0401111074, 0411324903, 0701165563, 0901213595, 1201272364 REF 38014600 Serial Numbers: 01336908, 01337113, 1392034, 1399208, 1417964, 1455082, 01465536, 01468470, 1554793, 02338769, 02339445, 02472354, 02473437, 02473552, 03476735, 04340997, 04341022, 04480362, 04480859, 05341947, 06485509, 10230894, 10355958, 11234390, 11358409, 12235395, 12360317, 015167729, 016306329, 016308714, 016310201, 017399547, 019733307, 019767191, 025182501, 027409856, 027415535, 028519215, 029798205, 035199116, 035200366, 036318356, 037423408, 038547965, 039813029, 039813030, 039813031, 039813032, 039823321, 039828323, 039832643, 039832644, 039832694, 045213409, 047426475, 047426476, 048571503, 048594220, 049849873, 049851825, 049853374, 049853375, 056331311, 056340965, 056343510, 058602446, 059859723, 059864560, 065233870, 066345835, 066349382, 066356210, 067441232, 067447341, 068605243, 069876961, 069880129, 075245330, 075252992, 075253435, 076360809, 076361926, 076364297, 077449022, 077451674, 077455287, 077458075, 078608784, 079867734, 085257537, 085267660, 086368438, 086371421, 087462404, 087464163, 088669034, 088676886, 088676887, 094115107, 095271453, 095275580, 096374344, 096376553, 097474406, 0101016268, 104106786, 105277323, 105287561, 108693406, 108716199, 114135087, 115287562, 116389875, 116392482, 117483297, 117497158, 118734978, 118740926, 118740927, 119978432, 119987981, 124140853, 124145841, 125300630, 126395719, 126397517, 127501997, 127507050, 128733308, 128767190, 0201064281, 0201064282, 0301098053, 0401098052, 0401099040, 0401101153, 0411317716, 0411321262, 0501137431, 0601148282, 0601153394, 0601164206, 0601164680, 0611353348, 0701178785, 0701181826, 0711378881, 0711378941, 1001214141 REF 38014604 Serial Numbers: 01336906, 01461664, 01467803, 02338765, 02339440, 02339917, 02468471, 02473549, 03476734, 04340995, 04480860, 05341943, 07485511, 08346552, 08348588, 10230889, 10354561, 11234399, 12360260, 015167728, 016297586, 016308999, 016311874, 017399756, 019785738, 027405946, 029795563, 035194616, 035203186, 036318357, 036327779, 039833518, 045213406, 046329112, 047424302, 047426477, 048577489, 048577490, 049844819, 049844820, 055227007, 056339150, 056342431, 056343511, 059861570, 065216561, 066345836, 067436427, 067447392, 068602448, 069876962, 075248219, 075255234, 076357553, 076364298, 076365412, 077453357, 077457317, 078635574, 079894025, 085255781, 087460710, 087464123, 088672398, 088672399, 095267657, 096374343, 104106785, 104115108, 105277227, 106380746, 115286917, 117497159, 119976437, 124140857, 126394336, 126397072, 127497160, 127507051, 128751977, 0401099038, 0601137718, 0601160504, 1001207004 REF 38014635 Serial Numbers: 01336835, 01336911, 1416658, 01461822, 01467804, 02338771, 02339450, 02339918, 02470400, 02472355, 03476737, 04341949, 04480363, 05341024, 06483698, 08495576, 10231602, 11234391, 11355959, 12235396, 015160407, 015166269, 015168556, 016302013, 016309001, 016313048, 017400225, 019780241, 028501992, 028536057, 029792978, 029795762, 037420876, 037422921, 037423589, 039838302, 045208767, 048594219, 049841540, 055227205, 056335923, 056342432, 057426478, 059857912, 059861572, 065248211, 066345837, 066354243, 066356211, 067447342, 067447777, 068605244, 069869523, 075250537, 078657133, 085257536, 086368439, 086370587, 086373093, 087458342, 087460711, 094106784, 095267370, 096374648, 098682825, 098690186, 105275577, 114130620, 117497161, 117497162, 118742675, 118751149, 126394337, 126397518, 0201016269, 0401089748, 0401089749, 0501114408, 0511353507, 0601152516, 0801181963, 0901209979 REF 38014800 Serial Numbers: 01336909, 01337114, 1383901, 1410827, 01461823, 1486662, 1495079, 02339446, 02341000, 02468472, 02469865, 02473439, 03341245, 03476713, 04341619, 04482816, 05344285, 05484321, 07346546, 07347291, 07487160, 07495578, 08495577, 8686045, 09352750, 10230896, 10230988, 10354562, 12235103, 12360261, 12361302, 015173212, 016305871, 016308716, 018506077, 019733514, 019769902, 025182502, 26313045, 027410231, 029791226, 029795566, 035200367, 037412685, 037418047, 037418978, 038527663, 039819345, 039819346, 039824496, 039836604, 047424303, 047426899, 047430769, 048571505, 049843297, 049847489, 049848945, 049849874, 049851826, 055226995, 055227194, 056331312, 056342433, 058602504, 059859724, 059861571, 059863222, 059864563, 066347882, 066354554, 067438330, 067441233, 067447778, 069881897, 075210117, 075248252, 075251169, 077458448, 078608785, 085260578, 086364914, 086366559, 087466674, 087468803, 088676884, 088676885, 094108458, 095273945, 095277334, 096374339, 096378765, 097475897, 097476288, 098693404, 098700617, 104106783, 105275581, 106379024, 114135088, 114135090, 117497163, 117497164, 118734979, 118745572, 118745573, 124140868, 125300650, 125302633, 126394338, 126398009, 127501993, 128753069, 0201016287, 0201067285, 0201067286, 0301084479, 0301095302, 0301096842, 0401096843, 0401101154, 0401101156, 0601162976, 0711383129, 0801182801, 0901194840, 0901206933, 1001206430, 1001217379, 1001223350, 1001224746, 1001225229, 080118280110 REF 38014804 Serial Numbers: 01337109, 01461824, 01468473, 02338606, 02339441, 02340998, 02471132, 03341237, 03473550, 04341614, 04483159, 05344287, 07346549, 07348589, 07489310, 08495579, 08495580, 10230890, 10230989, 10354563, 11357569, 12360263, 015166267, 015167727, 017398010, 018507053, 028531106, 029802361, 035194620, 037413265, 037420877, 039820465, 039836602, 039836603, 045213410, 047430272, 048530854, 056331313, 056342434, 056345838, 058602507, 059860581, 065233871, 065248215, 067441995, 067448020, 075252705, 076357554, 076363155, 077451731, 077458835, 078625450, 078625451, 079881902, 085260576, 086366558, 086371422, 087464164, 088676888, 088676889, 088682823, 095273946, 104130611, 105277300, 106379492, 114131983, 116389360, 117489496, 117496652, 125297570, 125302355, 125302644, 126392483, 0201014732, 0401084481, 0401084482, 0701169433, 1001212158 REF 38014835 Serial Numbers: 1380278, 01461825, 02338613, 02339451, 02341002, 02469862, 02471133, 03341249, 04341622, 04480364, 04483160, 05344288, 07347293, 07487851, 08495581, 10230912, 10232491, 10354564, 11234401, 11355960, 12235107, 12358860, 015168557, 016302645, 016313043, 017397073, 017400305, 019780242, 025185570, 029792979, 037420823, 037423396, 038552904, 038568864, 039816371, 039828321, 039828322, 047426900, 049843777, 049843779, 055225779, 056340863, 056345323, 057434552, 058602509, 059860407, 065248213, 067438331, 069871616, 075249040, 076361927, 077453358, 078625452, 078625454, 085257538, 086368440, 086369818, 087467534, 088672390, 088672391, 094106780, 095273948, 096375114, 097474166, 105277345, 106380745, 114130621, 114135091, 117495028, 125290519, 127496653, 0401091965, 0401091966, 0601161845, 0801183925, 0901209919 REF 38015000 Serial Numbers: 01337115, 1393712, 1430418, 1439985, 1452149, 1455083, 01461665, 01467805, 1491272, 02338609, 02339447, 02339920, 02473320, 03341247, 03342029, 04477704, 04483162, 06344286, 06347295, 07489311, 08495582, 10230897, 10230904, 10354565, 11357570, 015140862, 015166268, 015168555, 016303466, 017399542, 017400469, 018521833, 026310203, 027406554, 029792980, 029802358, 029802359, 035199117, 035200363, 037414646, 037420878, 038571506, 039824497, 039831047, 04742690, 048575688, 049851164, 049851165, 049852498, 055226996, 057432246, 058596749, 059863221, 065227202, 066352028, 067438336, 075252703, 075252704, 076356212, 077448993, 077458450, 078602511, 079871617, 079894026, 086366560, 086368896, 087464019, 087470002, 088665029, 094106779, 095267665, 096373702, 104130612, 105275576, 105277322, 106378766, 106379025, 108682827, 108700611, 108700612, 114135092, 117495029, 117496654, 118737879, 118743304, 119993336, 125290518, 126393905, 126395720, 127501994, 128755751, 0201014733, 0201016288, 0301084483, 0301084485, 0301099039, 0401105192, 0501133646, 0501134958, 0701168074, 0711378282, 0711378283, 0801184993, 0801189728, 1001214888, 1101242472, 1201269269, 1201273376, 040110519210 REF 38015004 Serial Numbers: 01337110, 01461826, 01467806, 02338607, 02339442, 02340999, 02469863, 03341239, 03341615, 03476224, 05483161, 06342027, 06344290, 07489313, 08495583, 10230891, 12360264, 015173216, 016303467, 017399475, 019783544, 025182503, 026310204, 027404971, 029802360, 037420879, 038552905, 039829629, 045208777, 045213407, 047430277, 048571507, 049848067, 049852501, 056340864, 065248212, 067448255, 068596794, 071172648, 075249041, 076356975, 077460111, 079888791, 088682826, 094106778, 095267369, 095275498, 096372715, 104130614, 105275574, 105275575, 106378767, 114135093, 117489497, 118744196, 124145842, 126394339, 127496655, 129998813, 0301041071, 0601138641, 0701168231, 0701171296 REF 38015035 Serial Numbers: 01337117, 01461827, 02338614, 02339452, 02339921, 02469724, 02473432, 03341250, 03341623, 03476714, 05483754, 06342031, 06344292, 07489314, 09495584, 10232417, 12360266, 015140870, 015167731, 016305919, 016306459, 027410460, 029783545, 035194617, 037420880, 037423397, 038552906, 039822085, 049843775, 049843776, 055225781, 057431923, 057432064, 058594667, 058596795, 069880130, 075248217, 075248250, 076356911, 087466204, 088682828, 095273950, 096372050, 097471978, 097478061, 104130616, 106380744, 114131984, 118745574, 125290551, 126395038, 126398011, 127510026, 0301044689, 0601156629, 0801188066 REF 38015200 Serial Numbers: 01337116, 1380155, 01461666, 01465486, 01467807, 02338610, 02339448, 02341001, 02469864, 02471134, 02471135, 03341248, 03341625, 04480862, 07489318, 08342030, 08495585, 09495586, 10230898, 10230911, 12358861, 015140864, 016287594, 017400091, 018521834, 019773567, 025176793, 026308715, 029783546, 029798206, 037420881, 037423398, 039820463, 039820464, 045208778, 045213412, 048577492, 049843772, 049848944, 055227008, 056331314, 057432247, 058577494, 058596797, 059843771, 069863223, 075248216, 076362314, 077453359, 078657130, 085260580, 086368442, 087459153, 087464165, 088685271, 094106776, 095267662, 096373828, 097471749, 097472434, 098700616, 105275582, 108701783, 114139820, 116389878, 118749030, 126396165, 127512695, 0301080511, 0301080512, 0401103157, 0411324084, 0601139755, 0601155505, 0601156389, 0711379606, 0801189097, 0901189896, 1291001913 REF 38015204 Serial Numbers: 01338608, 01461828, 02339443, 02469725, 02473435, 03341242, 03341616, 04480365, 05484322, 08342028, 08344297, 09495587, 09495588, 10230892, 015158581, 015176794, 016306330, 016306456, 019783547, 026315807, 027406478, 028497165, 028536058, 037420882, 038552380, 039825302, 047426902, 055226998, 055227203, 058594666, 067444762, 068596796, 069868797, 077458449, 085260577, 094106775, 095273121, 098690185, 0101010919, 0101044690, 105285959, 106373703, 107483294, 117496656, 124139825, 126396166, 128737884, 128747418, 0301050810 REF 38015235 Serial Numbers: 01337279, 01338615, 01461829, 01465537, 02339453, 02469866, 03341251, 03341624, 03476225, 04482822, 05484323, 06342032, 06344296, 07489320, 08495590, 09495589, 10230990, 10232418, 12358862, 12361307, 015145844, 015176790, 017403690, 018521835, 026306457, 027413226, 037420883, 038568865, 039829628, 045208768, 057426903, 058594665, 065227009, 066352033, 068596798, 076356931, 077453360, 085251170, 094117673, 095273124, 096373701, 098703645, 104130617, 105285960, 107470929, 124139827, 127497166, 128739172, 128755752, 129996003, 0301017301, 0301044691, 0301096844, 0901197220 REF 38015400 Serial Numbers: 1394265, 01461667, 01465538, 02472356, 02473433, 03341026, 03342024, 03476841, 04477705, 04482817, 04483163, 06342687, 08495591, 10231426, 10354696, 12360790, 015140871, 025185571, 027409857, 028521836, 029799070, 036308717, 038536059, 039832645, 045213405, 047424722, 047425772, 048575698, 057435632, 059842579, 065233873, 068605246, 068625455, 077453361, 078657670, 085257539, 086371132, 094104766, 095271454, 097466276, 097472435, 0101017302, 0101044692, 104106773, 105285961, 126396167, 127512696, 128751979, 0301050859, 0401106376, 1001218549 REF 38015404 Serial Numbers: 01461668, 01465539, 02338565, 02340839, 02469867, 03341944, 03476777, 04342025, 04480125, 04482818, 04483164, 06342688, 06483755, 09495592, 10231425, 11359158, 017399168, 028496657, 037414645, 038552381, 039825304, 047430280, 048571509, 055213411, 058575699, 066347883, 075249051, 076362912, 077451767, 078657135, 079843291, 085260581, 095273125, 096376129, 105285962, 114139822, 117495030, 124145843, 124155169, 128751978 REF 38015435 Serial Numbers: 01336260, 01461669, 02338566, 02469726, 02471450, 02472357, 03341025, 03341950, 03476715, 04480126, 04480864, 04483165, 06342026, 06484324, 07349404, 10231427, 10354697, 12359715, 015173211, 017399850, 035194619, 037420885, 037422920, 038523900, 039836600, 047424304, 055227002, 058594664, 068605247, 075227207, 076361468, 077453362, 097468338, 098690181, 115285963, 124139855, 125294660, 125303923, 126394104, 127512697, 128744195 REF 38015600 Serial Numbers: 01461830, 01469727, 02470534, 02473436, 03476842, 04480127, 04480366, 04482819, 06483166, 09495593, 10355620, 015173213, 017397074, 025185572, 028547966, 029773599, 038568866, 047423413, 048575701, 055225780, 057433178, 069836601, 078641618, 085257540, 086370950, 087464166, 106374338, 108693401, 114130618, 115285964, 115294661, 117495031 REF 38015604 Serial Numbers: 01461831, 01469728, 02473438, 02473530, 04480128, 04480865, 06484325, 09495594, 10355621, 12359716, 015140860, 015176791, 017395254, 017395255, 038552907, 045212675, 045216577, 048565042, 048577518, 055227011, 068625457, 077452352, 079852082, 096374225, 105273943, 108690175, 115287563, 128739173, 128753070 REF 38015635 Serial Numbers: 02469729, 02473440, 03476716, 04480867, 06484326, 09495595, 09495596, 10355622, 11359159, 015176789, 015176797, 017395256, 017395258, 018518301, 048577495, 056340329, 065227010, 067433179, 068625459, 077452351, 086371133, 105273949, 108690176, 127510027, 128752499 REF 38AC3600 Serial Numbers: 1393001, 1397482, 1398072, 1399095, 1404117, 1404493, 1404495, 1417636, 417637, 1419883, 1419884, 1422926, 1422927, 1424026, 1424027, 1427726, 1427727, 1434686, 1437246, 447097, 1448017, 1449308, 1450189, 1455908, 1457556, 1458011, 1461381, 1462438, 1476115, 1476495, 1476496, 1480131, 1482204, 1482507, 1483839, 1488742, 1488743, 1493474, 1495887, 1496143, 1497348, 1502529, 1502530, 1506859, 1506860, 1508795, 1511240, 1513856, 1513857, 1522498, 1522499, 1529773, 1532403, 1533876, 1536694, 1536695, 1536696, 1542347, 1542348, 1542349, 1543428, 1544116, 1560851, 1580772, 1580773, 1587443, 1593840, 1604512, 1606605, 1610983, 1613901, 1617423, 1620665, 1634156, 1634157, 1638020, 1647087, 1656682, 1659576, 019786468, 029807349, 038547863, 038578399, 039837410, 049846992, 059855434, 059857904, 069869511, 069880126, 079901175, 079901176, 089924906, 099932248, 099939753, 0101018644, 108730112, 109966096, 109969843, 118739168, 118742677, 119976429, 119976430, 119986961, 119992248, 129996175, 0201043275, 0201049059, 0201064312, 0201064313, 0211290529, 211295980, 0211296909, 0301087820, 0311316421, 0311316583, 0401105659, 0401111497, 0401121938, 0401121955, 0401122160, 0411324904, 0411341127, 0411341128, 0501132418, 0501132608, 0611366498, 0611366499, 0611366500, 0611378274, 0701178248, 0701179352, 0701180199, 0801198736, 0801199186, 0901203537, 0901204223, 0901204286, 1001210584, 1001219290, 1001228041, 1101235132, 1101238264, 1201262089, 1201262091, 1201265239, 1201265240, 1291010230 REF 38AC3800 Serial Numbers: 1396759, 1398073, 1399549, 1404433, 1416194, 1416478, 1417313, 1417314, 1417315, 1426350, 1434685, 1437247, 1451061, 1459879, 1461380, 1482205, 1483840, 1483841, 1493316, 1495576, 1506925, 1506926, 1513858, 1517450, 1529776, 1532404, 1542922, 1572004, 1587442, 1610563, 1634158, 029807350, 038578400, 059837411, 059857905, 069869509, 079901177, 079901178, 0101016207, 0101029699, 0101046899, 108730113, 0111282007, 0111294098, 118739169, 118742678, 119975130, 119976431, 119983788, 129997987, 0201053801, 0211307016, 0301067283, 0301080506, 0301096845, 0301096846, 0311324008, 0411342684, 0411343980, 0501123293, 0501123312, 0501124120, 0501136359, 0501136360, 0701187213, 0711367087, 0711367088, 0711393002, 0801187575, 0801188005, 0901202855, 0901202931, 0901208794, 0901208998, 1001220000, 1001220309, 1001220708, 1001237626, 090120899810 REF 38AC4000 Serial Numbers: 1393003, 1397580, 1398074, 1400509, 1404280, 1404502, 1406087, 1412304, 1412545, 1412588, 1412846, 1414064, 1414935, 1415270, 1419886, 1419887, 1422368, 1422369, 1424037, 1424038, 1426349, 1434683, 1434684, 1435350, 1436961, 1438774, 1446746, 1447470, 1449064, 1450136, 1451060, 1455909, 1457250, 1458010, 1461404, 1462169, 1462420.1478782, 1479036, 1480374, 1480375, 1484738, 1487558, 1487559, 1499248, 1499249, 1507352, 1507555, 1513859, 1517646, 1522500, 1529780, 1533877, 1536697, 1542924, 1544115, 1560849, 1560850, 1604977, 1610560, 1612607, 1652572, 1658821, 029807351, 038578401, 039833646, 039837412, 059857906, 069869507, 069879075, 079901179, 079901180, 0101016264, 0101030558, 0101048254, 0101048281, 0111291786, 118733840, 118739170, 119974118, 119976432, 119976433, 119987980, 119993327, 0201066509, 0211308910, 0211310149, 0301078359, 0301079664, 0301079665, 0311317396, 0311324978, 0311326684, 0501127363, 0501131312, 0501131832, 0511346555, 0511346556, 0611376331, 0611378272, 0611378273, 0701171611, 0701171838, 0801191517, 0801200097, 0801200726, 0801201509, 0901200884, 1001217387, 1001224434, 1001232783, 1101234146, 1201259135, 1201259139, 1291007838 REF 38AC4200 Serial Numbers: 1395680, 1397578, 1398075, 1404211, 1404500, 1406088, 1416195, 1416479, 1416480, 1417316, 1426351, 1433638, 1435352, 1440048, 1447135, 1449066, 1450803, 1457557, 1461405, 1480376, 1483842, 1483843, 1493058, 1493475, 1503304, 1507556, 1513860, 1513862, 1536698, 1542925, 1580774, 1587657, 1612608, 1652573, 1656683, 029807352, 038578402, 049838201, 059854132, 059859720, 069869506, 069879074, 079901181, 079901182, 0101020923, 0101030559, 0101043276, 0111290062, 118730115, 118743300, 119983791, 119993328, 128744191, 0201053802, 0211297586, 0211314248, 0301085770, 0301089742, 0301089743, 0301089744, 0311323469, 0411335377, 0411335378, 0411346846, 0411347584, 0501124768, 0501127075, 0501127361, 0511348849, 0611368272, 0611368273, 0611368274, 0701177939, 0701182354, 0701182550, 0801197636, 0801198416, 0901204469, 0901205069, 0901205704, 1001231515, 1001231516, 1101239256, 1201260902, 1291009549 REF 38AC4400 Serial Numbers: 1398076, 1419880, 1421704, 1433637, 1440182, 1449309, 1459880, 1493315, 1517453, 1533878, 1542923, 029806010, 039826180, 039838202, 048578403, 049852508, 059857907, 059867730, 069876958, 079901183, 079901186, 0101018645, 0101030560, 118733841, 129998796, 0201049075, 0211302283, 0211304018, 0301076814, 0301076815, 0401106599, 0501119003, 0501119677, 0501121086, 0611367089, 0701173450, 0701181828, 0701181831, 0801201594, 0801202105, 0901211308, 1001226358, 1001228120, 1001229792, 1291007900 REF 38AC4600 Serial Numbers: 1397120, 1398077, 1417318, 1423213, 1435353, 1515691, 029807354, 048578404, 049838203, 049854131, 059855435, 059857908, 069867731, 069876957, 079901187, 079901188, 0111283143, 118731297, 118743301, 118744192, 0201041070, 0201059407, 0211304809, 0301078360, 0301079666, 0301079667, 0411343982, 0501129972, 0501131153, 0611364788, 0611364789, 0701170194, 0701170751, 0801194819, 0801194931, 0901195618, 0901211949, 0901212266, 1001226443, 1201260901 REF 38AC4800 Serial Numbers: 1396838, 1414898, 1419307, 1426348, 1427533, 1427967, 029733843, 029806009, 038578405, 039826183, 049838204, 059852504, 059857909, 069867732, 069876956, 069881865, 079901189, 079901190, 118728136, 0301068678, 0301071238, 0311313422, 0311319143, 0401108735, 0401109406, 0501134098, 0501136264, 0701186227, 0711389260, 0801186455, 0801188081, 0801188460, 0901208344, 1001206434, 1001206811, 1001213551 REF 38AC5000 Serial Numbers: 1397205, 1399550, 029807356, 038578406, 059838218, 059852506, 059860582, 069864573, 069876955, 079881864, 079901191, 079901192, 118715111, 118737877, 0211295494, 0211304894, 0301096480, 0311318361, 0401096481, 0401096998, 0401096999, 0801188772, 0801199295, 0801199945, 0901199320, 1001227077, 1201272366 REF 38AC5200 Serial Numbers: 1397763, 029807357, 038578407, 039833647, 059854133, 059860583, 069864571, 069876954, 069880127, 079901193, 079901194, 108731298, 0211311479, 0311311815, 0311317717, 0401118056, 0401118878, 0701172476, 0801193419, 0801193740, 0901206138, 0901206248, 1101237605 REF 38AC5400 Serial Numbers: 1396839, 029807358, 038578408, 069861564, 069864569, 069872471, 069872472, 069880128, 079901195, 079901196, 118731299, 118733844, 118745542, 0801192523, 0801192534, 0901209980, 1001209497 REF 38AC5600 Serial Numbers: 019745543, 029807359, 048547867, 048578409, 069861567, 069864568, 069872431, 069872432, 069883953, 079901197, 079901198, 0801188461, 0801196131, 0801196455, 0801196513, 0901197095 REF 38AM3600 Serial Numbers: 1397865, 1400712, 1402245, 1407216, 1412007, 1412590, 1412592, 1417319, 1426346, 1432680, 1435899, 1447471, 1448614, 1450199, 1456415, 1457005, 1459881, 1462437, 1462717, 1476771, 1476772, 1483572, 1488744, 1488745, 1497257, 1497258, 1501977, 1501978, 1508511, 1517647, 1523002, 1525242, 1530125, 1534147, 1542177, 1558316, 1560847, 1572005, 1572006, 1580083, 1604978, 1610984, 1634159, 1638021, 1647088, 1657881, 016268864, 019755725, 026270446, 026286922, 028533957, 028544866, 028552356, 029786477, 037414661, 037415051, 038562882, 039816366, 039828317, 046337958, 049849868, 057438209, 058597857, 058602428, 067445889, 068605226, 068628914, 069871612, 069876949, 075204838, 075229804, 077451765, 078648643, 078660991, 079893234, 085261216, 085265547, 088670399, 088681290, 089911141, 089922082, 098690253, 098700790, 108715112, 108728960, 117506075, 118722757, 119976434, 0301058384, 0411335255, 0411339017, 0511359164, 0511359746, 0601148210, 0601155557, 0611362495, 0611363290, 0801181827, 0801182732, 0901213858, 1001214137, 1101241342, 1201242473 REF 38AM3604 Serial Numbers: 1398726, 1401119, 1405718, 1406645, 1409292, 1409403, 1409960, 1413650, 1413651, 1414570, 1416481, 1432652, 1432679, 1435900, 1438199, 1448380, 1450191, 1452346, 1456424, 1459882, 1460664, 1462718, 1477349, 1477536, 1484370, 1484739, 1492232, 1493476, 1496144, 1501675, 1501808, 1507557, 1507762, 1513861, 1513863, 1542178, 1543429, 1555138, 1555139, 1555140, 1572007, 1612609, 1634160, 1638022, 1647086, 1656680, 017395494, 028548127, 029792975, 029798201, 037422290, 038552357, 039821652, 039832450, 046337625, 048585771, 056339972, 057437831, 057438631, 058617072, 066342397, 067445890, 068628867, 068635564, 069871613, 069880125, 075204839, 078648087, 079890283, 085261218, 085269473, 088662880, 098695710, 107480810, 108722758, 117508399, 118744190,125270447, 128755726, 0211299336, 0301041083, 0301085942, 0511355230, 0601156630, 0611355263, 0701171827, 0711370670, 0801188062, 0883681291, 1001214996, 1101251437, 1291010896 REF 38AM3635 Serial Numbers: 1402518, 1406871, 1409961, 1450676, 1451089, 1459883, 1462719, 1468924, 1477350, 1477671, 1497349, 1507780, 1525243, 1534148, 1541688, 1541689, 1560848, 1580081, 1580082, 1606606, 1617420, 1634733, 1638461, 1656681, 017395495, 028552897, 029798202, 036276112, 036287554, 039825296, 039830309, 047422771, 048580387, 056339980, 057438208, 058622133, 067443016, 068625420, 068634572, 069869336, 075204840, 077450129, 078647840, 078660992, 085261219, 085261220, 098695712, 099934272, 118739167, 125271455, 127510021, 128755727, 0201017298, 0301081062, 0511363319, 0601145040, 1001214997, 1101251436 REF 38AM3800 Serial Numbers: 1402556, 1405255, 1406089, 1414899, 1415271, 1415272, 1425078, 1425711, 1433461, 1438198, 1483573, 1523000, 1524616, 1528094, 1536233, 1541690, 1541691, 1620673, 1640625, 017394108, 018507033, 019782605, 026286923, 028544867, 029797063, 037413917, 037417332, 038552358, 039816367, 039828318, 046335286, 047422868, 048568848, 049842576, 056339981, 057436637, 058602429, 058605227, 065232327, 067445545, 068625422, 069861050, 069875060, 075204841, 078648085, 078660994, 078669047, 079876950, 079893235, 085261221, 085269849, 089922083, 095271456, 098695713, 099934903, 107481156, 108722759, 117482910, 118741126, 128755728, 129994170, 0301072613, 0301072614, 0301072615, 0601159622, 0611376332, 0611378301, 0711378874, 0801192533, 0801193418, 0801193739, 0901206247, 0901206432, 1001215453, 1001215454, 1101232520 REF 38AM3804 Serial Numbers: 1399014, 1402033, 1407997, 1415233, 1415273, 1428229, 1440003, 1448975, 1463014, 1528092, 1530126, 1542179, 017387782, 027412024, 028544868, 029800420, 037417330, 039822077, 039832452, 046335287, 046335929, 047426822, 048562883, 057429771, 057436638, 058592680, 065232328, 067438199, 067445175, 068636362, 075204842, 078648088, 079899197, 085261222, 085269850, 087450552, 088648086, 089929112, 095271457, 098700753, 099941217, 099942632, 0101017299, 107492864, 108715113, 116387842, 118745513, 127513349, 128752305, 0211299342, 0301073849, 0301081063, 0401107031, 0601145352, 0601148833, 0701171828, 0711378942, 0901206717, 1001218547, 1201251438 REF 38AM3835 Serial Numbers: 1402175, 1407604, 1415234, 1415274, 1484963, 1485338, 1508794, 1523001, 1528096, 1536234, 1587658, 017394242, 019789324, 028542258, 029800422, 037417331, 038507034, 039826198, 046335288, 048568849, 049841534, 056339979, 057431547, 058608765, 065232329, 067436639, 068636363, 069869337, 075204843, 078648089, 085261223, 085265548, 085269851, 089929113, 095271458, 097450984, 098695709, 0101010910, 108715114, 109960595, 117506076, 118743299, 119978898, 128755729, 129996001, 0301073850, 0301081064, 0401107032, 0601158803, 0711388288, 0801188063, 0901206718, 1001218548, 1201256298 REF 38AM4000 Serial Numbers: 1395681, 1402674, 1403983, 1406091, 1412458, 1412596, 1412847, 1413859, 1414571, 1417320, 1425079, 1425352, 1425353, 1428254, 1428255, 1437715, 1493290, 1496145, 1507781, 1513864, 1541692, 1560038, 1580775, 1640324, 1656679, 017395737, 017400510, 019776464, 019792977, 026287553, 028507035, 028542259, 037414650, 037420808, 038552359, 039808002, 039822080, 039829075, 046330367, 046330368, 047426450, 048575648, 049849870, 056339969, 057431553, 058592681, 059871596, 065245204, 065245205, 067439224, 068608766, 068625424, 068636364, 069880124, 069890285, 075204844, 078636365, 078648131, 078670401, 079911139, 085261224, 087447835, 087470005, 088691654, 089924907, 095270448, 096370676, 098700792, 099932180, 0101012378, 107481157, 108705526, 0111282706, 116385195, 117482911, 118728901, 118741127, 119988028, 126388859, 128755730, 0201052604, 0201059408, 0311327204, 0311327683, 0401118103, 0401118629, 0511354378, 0511354379, 0511362496, 0511363320, 0601151661, 0601152844, 0801188080, 0901192046, 0901200721, 0901200723, 0901214189, 0901214987, 1001240496 REF 38AM4004 Serial Numbers: 1378270, 1396840, 1402721, 1404021, 1408926, 1412599, 1415232, 1415275, 1416482, 1417639, 1417640, 1428527, 1437716, 1459884, 1525244, 1543430, 1606607, 1640325, 017395783, 019789330, 019789332, 028534018, 028542260, 029802351, 036311888, 036311889, 037415557, 037418340, 038552360, 039816368, 046337959, 047426451, 048573076, 049843763, 049843765, 057431554, 058608767, 059871597, 067439225, 068625240, 068625426, 075204845, 075245206, 075246820, 078648090, 078660995, 079891169, 085247558, 085247559, 087448657, 087469500, 088691655, 098700791, 098705525, 099937603, 106379217, 107483296, 107489011, 0111286050, 0111289478, 118733511, 118733512, 125270449, 126390195, 128755731, 128769919, 0201053803, 0501126063, 0501126744, 0501127073, 0511351997, 0511352000, 0801185467, 0801185857, 0801186454, 0901211477, 0901212227, 1101239255, 1291007901, 16403251643272 REF 38AM4035 Serial Numbers: 1402722, 1409962, 1412006, 1412848, 1415276, 1415277, 1428528, 1439728, 1580776, 1640794, 017403688, 018502564, 018536019, 019771055, 029797036, 036313458, 37416895, 037421449, 038565198, 039825297, 039830310, 046313457, 046336279, 047426452, 049851162, 056339978, 058592682, 059869338, 067438327, 067446442, 068647841, 075204846, 078648091, 078662881, 079909594, 085246821, 085246822, 085247570, 085247571, 097476149, 098691656, 098706234, 0101010911, 109950529, 116388070, 117497149, 119978899, 125270450, 126394326, 128752844, 0201059750, 0201065516, 0401092467, 0511348850, 0511350543, 0511360349, 0601145353, 0601161356, 0801197209, 1201251439, 1201256299 REF 38AM4200 Serial Numbers: 1401998, 1403535, 1409963, 1411527.1412849, 1418329, 1420484, 1426116, 1436362, 1439651, 1464810, 1508512, 1520217, 1541696, 1560039, 1617421, 016306195, 017395937, 017400511, 018496639, 018536020, 019771056, 019779432, 019791229, 026309774, 026309775, 027412846, 028527660, 029805760, 037417337, 037423320, 038562884, 039825298, 039832454, 046331322, 047426453, 048575649, 049840531, 049846673, 049854862, 056339871, 057436436, 057438012, 058602430, 059861054, 059871598, 065236108, 066347965, 067442932, 067447593, 068608768, 069880123, 069888794, 075204847, 078625427, 078648092, 078670402, 079895129, 085261225, 085269852, 087452394, 087470006, 089915887, 089928629, 095271459, 096374629, 096378756, 098695711, 098700752, 098706235, 099934273, 099944910, 0101027424, 0101030561, 107481281, 116385196, 116388151, 117482912, 117507036, 118744503, 119973076, 119982338, 119993329, 126393863, 128755732, 0201013704, 0501124157, 0501124158, 0501125126, 0511360448, 0511361491, 0701170054, 0701170122, 0701170800, 0801197607, 0801197635, 0801198415, 0901200633, 0901216446, 0901216602, 1001217228 REF 38AM4204 Serial Numbers: 1399211, 1403274, 1406092, 1412448, 1412850, 1419662, 1436478, 1452151, 1541697, 017402172, 017405959, 018533958, 019778611, 028536021, 029793768, 037417338, 037421526, 038548128, 039809750, 039825299, 039832453, 039832695, 046336289, 049851822, 057435193, 058592683, 058605228, 059861569, 059871599, 066347962, 067439766, 068625241, 068625429, 068635563, 069881888, 069881889, 075204848, 075236109, 077447834, 078648093, 078662882, 085261226, 085268855, 086365013, 087466392, 088688768, 095269853, 095271460, 097476150, 098688769, 099931105, 099939754, 0101027425, 106379218, 107483290, 117489012, 118737874, 118737875, 119985907, 126388295, 126388399, 128755733, 129998799, 0311332185, 0401109591, 0401119002, 0401121015, 0401121088, 0411331605, 0511357837, 0511358253, 0601164913, 0601165148, 0601167202, 0901203163, 0901204222, 0901204285 REF 38AM4235 Serial Numbers: 1402984, 1412037, 1419661, 1426117, 1436477, 1457006, 1542180, 1560040, 1560041, 017395582, 018507037, 019783430, 027413465, 028536025, 029805761, 038548129, 038552361, 039835272, 046329106, 046339867, 047425220, 047425221, 048568850, 048575652, 049840532, 049856537, 056340911, 056341043, 057437516, 058608776, 059875062, 067443989, 068625242, 069885338, 075204849, 075236110, 077449501, 078648094, 079890288, 085261227, 085261228, 085265549, 088690237, 095271461, 098695748, 098700793, 098707794, 099934904, 0101010912, 0101041082, 107481159, 117482913, 119978900, 126388842, 128755734, 0301072899, 0401092468, 0401121937, 0601149888, 0601162474, 0711365948, 0711365949, 0901197212, 0901206719, 1201256300, 07113659481438578 REF 38AM4400 Serial Numbers: 1402458, 1409012, 1418330, 1423214, 017399752, 017403037, 017404793, 018507038, 019771058, 019784316, 019784317, 26306196, 026309014, 026309015, 028536026, 028548130, 029816369, 037415606, 037418517, 038568851, 039809760, 039835279, 039835280, 046330369, 046336273, 047426454, 047430359, 049848942, 055208892, 055219123, 055236111, 056339976, 057436078, 058592684, 058592685,059856535, 059862976, 065236625, 066355224, 067442283, 068619017, 068635565, 069876951, 069884533, 069886183, 077444899, 077448655, 078648099, 078670398, 079890291, 086365015, 088687709, 088688766, 088688767, 089915886, 089927670, 096371018, 097453677, 097471657, 098705523, 099932181, 099935698, 0101018646, 106378757, 106382898, 107483291, 108715115, 109949746, 116386565, 117489013, 117496666, 118736944, 125284390, 126387629, 126392069, 127502565, 128752845, 128755735, 129993331, 0201030562, 0401109590, 0401116682, 0401117710, 0401122159, 0401123278, 0411329530, 0411330542, 0511353349, 0601154025, 0601154488, 0601155351, 0711378271, 0801183815, 0801184363, 0801196454, 0801196512, 0801196551, 0901214986, 0901215449, 1101240358, 1291001846, 08011965511306747 REF 38AM4404 Serial Numbers: 1403389, 1419660, 1425712, 017391072, 017399540, 017402315, 018502566, 019771054, 019780124, 019791228, 027404727, 028548131, 028552362, 029809782, 036313445, 036313446, 037415607, 039819959, 039826188, 046337960, 047430238, 048585773, 048592686, 049839750, 049846674, 055208898, 055219124, 055236112, 056339977, 057430357, 057436640, 058592687, 059851823, 059861056, 059862979, 066349007, 067444898, 068619018, 069875064, 069879072, 069885339, 069888793, 078648132, 078648133, 079893236, 086361355, 086365301, 086372054, 087447146, 087450993, 088662883, 089924908, 097470693, 098695704, 098695705, 099931106, 099934274, 099939755, 0101019955, 107485758, 108707804, 115276118, 116381039, 116385197, 116387405, 117489014, 125277872, 125284391, 127496640, 128739653, 128752860, 128755736, 129995216, 0311329123, 0401114366, 0401115039, 0401116683, 0411334833, 0411345043, 0501138317, 0511355907, 0511356433, 0511357203, 0701179430, 0701179480, 0701180198, 0901213550, 0901213855, 1001206137 REF 38AM4435 Serial Numbers: 017399807, 017402395, 019771057, 019783431, 026308012, 029800423, 037415279, 037415608, 038568852, 039820462, 046328089, 046336280, 047430239, 047430358, 048575654, 049841533, 049846061, 055208899, 055219125, 056339973, 058592688, 059861568, 065236113, 066347963, 067441228, 067445553, 068636366, 069876952, 069886185, 078648134, 079890293, 086364259, 086365940, 087448312, 087453679, 087471734, 088648135, 088691653, 089928624, 095268853, 095276116, 098705524, 099937606, 099944909, 0101018647, 106378192, 106382899, 107492865, 108715116, 118746769, 126386575, 126391797, 127496641, 127507039, 128755737, 129993332, 0201049076, 0401098047, 0401098048, 0501123294, 0501124121, 0701180924, 0701181123, 0701181698, 0901208343, 1001206809, 1001220902, 1001224433, 1001226444, 1001228119, 1291007902 REF 38AM4600 Serial Numbers: 1393004, 1399783, 1407812, 1436959, 017399753, 017400207, 017403036, 019771059, 019785719, 026309784, 026309785, 027412441, 027413271, 027413272, 028533961, 028536027, 028552363, 029797064, 037415595, 039819960, 039835283, 046328091, 046338149, 047430360, 048568853, 049843277, 049843278, 055214149, 056341797, 056343986, 057435194, 057436642, 057437153, 058592689, 058592690, 059856536, 059862346, 059868804, 065226085, 065241653, 065245200, 067444092, 068608777, 068625431, 068635567, 068648136, 069876953, 069884534, 069886188, 075243194, 076355222, 076359965, 077451672,078648137, 079891170, 087447373, 087464092, 087466395, 088662884, 088688770, 088688771, 096376539, 096379219, 097474648, 099937604, 106381040, 106382900, 107483292, 116389189, 116389225, 117489015, 118732501, 119967198, 119988030, 125277869, 125281532, 126385200, 127496642, 128747412, 128747413, 0201030563, 0401109589, 0401112134, 0401112419, 0401112951, 0501132417, 0501132607, 0801194818, 0801194930, 0801195616, 0901196130, 0901217384, 1291010231 REF 38AM4604 Serial Numbers: 1393005, 1407839, 017399552, 017400318, 017402398, 018523878, 019776465, 019791230, 028523892, 028536028, 037415281, 037416559, 039809763, 039822084, 039835275, 046328090, 046333741, 046338150, 048552364, 049846059, 049851821, 055214150, 056341389, 056346219, 057430361, 057436435, 057436641, 058592691, 059862348, 065226086, 065240416, 065241654, 065243195, 067445174, 068592692, 068648138, 068648139, 069868806, 069879073, 069885340, 069889889, 076358121, 076361356, 086366642, 086366777, 086366778, 087447374, 087451764, 088648140, 088670400, 096369173, 097471440, 097471731, 099927671, 099935700, 0101018648, 108719490, 115277868, 115281522, 115284399, 116384883, 116384884, 118738621, 118738622, 119975131, 126392074, 127510022, 127513350, 128755738, 128769920, 0401118877, 0411332565, 0501127366, 0501127368, 0501129968, 0511361492, 0611371856, 0611371901, 0701183534, 0801182746, 0901204998, 0901205662, 1001220308, 1291007903 REF 38AM4635 Serial Numbers: 017399535, 017400403, 019785717, 027405323, 027406264, 028523893, 028536029, 029802352, 037416784, 039822081, 046332875, 048568854, 049841532, 049846060, 055214151, 056339974, 057431543, 057436079, 059861057, 065226087, 065240419, 065242427, 065243196, 066347964 067444091, 068592693, 069871607, 069885341, 069889890, 078648141, 086359499, 086370239, 088636367, 088691652, 096374630, 097448313, 098694783, 098706236, 106379846, 107476151, 115277867, 116385201, 116388152, 117482914, 125281535, 125284398, 126391041, 127496643, 128752846, 128755739, 0501131025, 0501131310, 0701165733, 0701171829, 0801197215 REF 38AM4800 Serial Numbers: 1393006, 1412154, 1426115, 016308025, 016313447, 016313448, 016313450, 016313451, 017399902, 017402626, 017402986, 018521831, 018527579, 019767195, 019779435, 027405802, 028523894, 028536030, 028542261, 029792976, 037415559, 038552365, 039808006, 039809762,039826189, 039828956, 046337961, 047425222, 048568855, 048592694, 049848941, 055216746, 057431555, 058592695, 058592696, 059856126, 059862978, 065236623, 065237424, 065240421, 066347217, 067438210, 067443017, 067445945, 067448314, 068648142, 068648149, 069875066, 069881890, 069881891, 076361358, 076364260, 078648150, 085256579, 086366641, 086366779, 087454154, 088670406, 088673569, 097471644, 099935701, 0101029702, 106372053, 106379845, 106380737, 109949748, 116384881, 116384882, 125281533, 126390142, 126394532, 126395496, 127510023, 128738625, 128746768, 128755741, 0401115945, 0401115990, 0601150025, 0601151289, 0801184567, 0801185113, 0901210307, 1001210583, 1001220707 REF 38AM4804 Serial Numbers: 1393007, 1398379, 1419560, 1420485, 1420698, 1420738, 017400078, 017400138, 019767196, 019780123, 026309791, 026309792, 027405322, 028523895, 028536031, 029797062, 037415602, 037418973, 037421525, 038552366, 038568856, 039809761, 039826194, 046331323, 046339234, 049831043, 049851161, 055216748, 056342400, 057431556, 057438211, 058580415, 059862343, 059864564, 065236624, 065240423, 065242424, 065243193, 067444800, 068605229, 068625433, 068635568, 069881893, 076357863, 078648163, 079881892, 085256578, 087468953, 088670404, 088673567, 088673568, 096369175, 096377460, 097448315, 098700926, 098701775, 0101027426, 106380256, 107481160, 107482915, 116386563, 116387403, 116390597, 125281530, 126395393, 127512694, 128755743, 0301068679, 0301071239, 0411333203, 0601156526, 0601157554, 0801185230, 0801185231, 1001214136, 1001214188, 14206981435332 REF 38AM4835 Serial Numbers: 019767197, 019779431, 026313452, 026313453, 027405546, 028507040, 029799053, 037415603, 037421400, 038552367, 039822079, 039828957, 046336291, 046337962, 048575656, 055216750, 055229818, 057431557, 057438212, 058592703, 059864565, 065236074, 065242429, 066347656, 067444090, 075242430, 078648164, 078648312, 085256577, 086361357, 086367650, 086367986, 087447214, 087451673, 088673571, 088673572, 089909595, 105277870, 106369174, 107483293, 109960600, 116385202, 116388290, 116390097, 117489016, 118747415, 119978901, 125281536, 126394581, 126395233, 127496644, 0301081066, 0401102850, 0601145354, 0601163565, 0801188064, 1201260341, 1291010913 REF 38AM5000 Serial Numbers: 016311876, 016311877, 017399921, 019782601, 027404846, 027406265, 028523896, 028552368, 029802353, 037415068, 037418979, 037423123, 038562885, 039822083, 046329107, 046339235, 047426456, 048580416, 049843741, 049843745, 056341798, 057431558, 058592704, 059864566, 065226091, 065229807, 065243285, 065243286, 065243287, 066349006, 067441229, 068619019, 069881866, 069881867, 077445943, 078648165, 078670405, 085256576, 086367987, 086371019,087450130, 088688772, 096374628, 097474919, 098698181, 0101013705, 106380254, 115277871, 115281534, 116382901, 116386567, 117497150, 126388401, 126392111, 126395029, 127502567, 127507041, 128755744, 0401112420, 0401114511, 0501131831, 0601168067, 0601168232, 0901200883, 0901202103, 1001215451, 1001216445 REF 38AM5004 Serial Numbers: 016281531, 017395028, 017400140, 019767198, 019785718, 028533963, 028536032, 028552369, 037414687, 037419584, 037419991, 038577479, 039808004, 039822082, 046328092, 046338007, 048580417, 049839751, 056342908, 057435195, 057436343, 059854861, 059862977, 059864567, 065226092, 065229812, 065243289, 065243290, 067439767, 068625435, 068625437, 068648166, 069886219, 075243288, 076357872, 077447836, 078670403, 079893238, 085256575, 086367646, 086368162, 087466397, 088670476, 096367647, 096368163, 097475878, 098688773, 099937605, 0101016265, 105277873, 107482916, 116384885, 117489017, 118738624, 126389825, 126392110, 126395257, 128755746, 0301067284, 0301086926, 0401098180, 0701169302, 0701169437, 0901202876, 1001202826, 1001216634, 1001216896, 1001217154 REF 38AM5035 Serial Numbers: 016281537, 017400079, 017405530, 018496645, 019767199, 019778610, 027413466, 028536033, 028552370, 029806008, 037416230, 039822078, 046330370, 046339233, 049843279, 056341799, 057431559, 058592705, 059861052, 065226094, 065229819, 065243291, 065243293, 067438834, 067444903, 067449063, 069881868, 069881869, 075245944, 075245945, 076358120, 077450104, 078648167, 079895130, 085256574, 086370240, 087448656, 088670407, 095277874, 096368882, 106375078, 106381041, 107481161, 108707542, 116386583, 117482917, 118746766, 126390145, 126394746, 0401092469, 0801188065 REF 38AM5200 Serial Numbers: 017388858, 017395372, 017400404, 019764339, 019767200, 019785720, 026285592, 027405137, 028544960, 029802354, 029806007, 036313051, 036313052, 037416021, 037419506, 037420432, 038552371, 046333742, 046339238, 048571495, 055218977, 055229806, 057430366, 058592707, 065226103, 066343987, 066349008, 067439768, 075246711, 078648168, 078648169, 085256573, 086364258, 086367648, 086367649, 087446440, 089924696, 089929114 095276117, 096368883, 097450553, 106380252, 106382902, 106384880, 107482918, 0111276684, 119972207, 127496646, 0301072900, 0301081067, 0801185312 REF 38AM5204 Serial Numbers: 019766269, 019779433, 026306197, 026313049, 027400326, 027405734, 028548140, 028552372, 029802350, 036313050, 037416197, 046330372, 046339240, 048592708, 055218978, 057430365, 058592867, 059843280, 065229811, 065245197, 066347840, 067441227, 075226106, 075245198, 076355223, 076361359, 077450992, 078648170, 078648171, 085256572, 086365014, 089924697, 097445942, 106372052, 106382903, 109960621, 119978902, 126385203, 126392094, 126395497, 127482920, 0301072901, 0401092470, 0801185313 REF 38AM5235 Serial Numbers: 019791227, 026286040, 027404926, 027405801, 028548141, 028552373, 037416229, 037420809, 039808003, 039828954, 046339237, 055218979, 056339734, 057430364, 058592709, 065226107, 065245199, 065246853, 065247555, 067442933, 067447837, 075246712, 076360860, 078648172, 085256571, 096367651, 096368884, 106382904, 107481162, 107482921, 109960622, 126385198, 126392095, 126395734, 128752859, 0301044685, 0301081068, 0801185314 REF 38AM5400 Serial Numbers: 017400731, 017405547, 018519212, 019766270, 019782604, 026285593, 028527664, 028536034, 028542262, 036313455, 036313456, 037416022, 038552374, 038568857, 046337963, 056340910, 057430363, 059849864, 065230415, 067438329, 068592710, 078648173, 085257622, 085265550, 086365012, 088670408, 095271462, 097446441, 106377461, 106382905, 125204850, 126386582, 126392109, 126395779, 127510024, 0801185315 REF 38AM5404 Serial Numbers: 017400995, 017402087, 019779434, 028507042, 028536035, 028548142, 029795763, 036326034, 036328087, 037416023, 038552375, 039831044, 048562886, 056339971, 057432116, 065230416, 067442284, 068641620, 078648174, 085257623, 085265551, 095271463, 096370675, 097449502, 098688774, 099941245, 106380738, 106384893, 116387767, 116387898, 117482922, 117509481, 118738623, 125305236, 125305237, 128747414 REF 38AM5435 Serial Numbers: 017394109, 017401815, 026286041, 028507043, 029799052, 036329108, 037416196, 038552376, 038571496, 046336290, 048586170, 057432117, 065230417, 067441990, 068605230, 068635569, 078648175, 085205659, 085257625, 085265866, 095271465, 095276119, 096368885, 097447168, 106379214, 106385199, 116388963, 117481163, 128752861, 0301041084, 05741337101 REF 38AM5600 Serial Numbers: 016204851, 016271466, 016285594, 017400513, 018527661, 019766271, 027405957, 028536036, 028542263, 028548143, 029787449, 036328088, 037416231, 038552377, 039828955, 046336281, 056339970, 057429770, 065223267, 067439765, 068592750, 078648176, 085257627, 085265552, 086368886, 096375079, 097447375, 106380251, 126388067, 126395778, 128765511 REF 38AM5604 Serial Numbers: 017395938, 018528569, 028536038, 028542264, 028548145, 029783432, 036309800, 036309801, 037413467, 037421527, 038552378, 046334710, 056339975, 057437830, 065223268, 075205661, 078592751, 085257628, 088648177, 097446439, 105271467, 106380250, 126387599, 126394494 REF 38AM5635 Serial Numbers: 017405908, 018533966, 019771053, 028536039, 028548146, 029793767, 036286042, 037413919, 037417333, 046334711, 047426457, 048587122, 056339968, 057436166, 065223269, 067438247, 067445944, 075247556, 075247557, 075247569, 078648178, 085247568, 088679485, 096378165, 097448660, 108700622, 115271468, 116387319, 125276120, 126394287, 1101232521 REF PL38HM4400 Serial Numbers: 035176036, 035177466, 035177467, 035185563 REF PL38HM4404 Serial Numbers: 035178044, 035185564 REF PL38HM4435 Serial Numbers: 035178045 REF PL38HM4600 Serial Numbers: 035179372, 035185815 REF PL38HM4604 Serial Numbers: 035179373 REF PL38HM4635 Serial Numbers: 035179374 REF PL38HM4800 Serial Numbers: 035179927 REF PL38HM4804 Serial Numbers: 035179928, 035185566 REF PL38HM4835 Serial Numbers: 035179929, 035185567 REF PL38HM5000 Serial Numbers: 035181982, 035194833 REF PL38HM5004 Serial Numbers: 035181984 REF PL38HM5035 Serial Numbers: 035181986 REF PL38HM5200 Serial Numbers: 035181988 REF PL38HM5204 Serial Numbers: 035181990 REF PL38HM5235 Serial Numbers: 035181993 |
Recalling Firm/ Manufacturer |
MicroPort Orthopedics Inc. 5677 Airline Rd Arlington TN 38002-9501
|
For Additional Information Contact | Rachael Wise 901-867-4771 |
Manufacturer Reason for Recall | Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R)
metal acetabular liner systems. |
FDA Determined Cause 2 | Device Design |
Action | MicroPort sent an Urgent Field Safety Notice dated June 22, 2016, to all affected customers notifying them of the change in the instructions for use on July 7th, 2016 by email. The firm also notified US hospitals where the implantation surgeries occurred within the last twelve (12) months by letter on July 13th, 2016 delivered by FedEx Priority. Customers were asked to complete and return the enclosed form even if they do not have any affected product to return by fax to 1-901-451-6032 or PostMarket@ortho.microport.com.
For questions regarding this recall call 901-867-4771. |
Quantity in Commerce | 108,362 units |
Distribution | Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LZO
|
|
|
|