| | Class 2 Device Recall Abdominoplasty Tray |  |
| Date Initiated by Firm | April 22, 2015 |
| Create Date | October 20, 2016 |
| Recall Status1 |
Terminated 3 on October 25, 2016 |
| Recall Number | Z-0140-2017 |
| Recall Event ID |
71968 |
| Product Classification |
Light, surgical, ceiling mounted - Product Code FSY
|
| Product | Abdominoplasty Tray, there are two Kit part numbers: PSS1782 and PSS1782(A.
convenience custom kits used for general surgery in hospital operating room |
| Code Information |
Lot Numbers/Expiration Dates: 62789 11/7/2015 61201 11/8/2015 63413 11/9/2015 67096 11/9/2015 64598 1/29/2016 74803 6/26/2016 57510 6/27/2016 67607 8/21/2016 69801 9/15/2016 71002 10/6/2016 68446 10/9/2016 71403 12/24/2016 72095 1/14/2017 78759 11/5/2017 81077 2/13/2018 |
Recalling Firm/ Manufacturer |
Windstone Medical Packaging, Inc. 1602 4th Ave N Billings MT 59101-1521
|
| For Additional Information Contact | Vicki Davis 406-259-6387 |
Manufacturer Reason for Recall | The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES, Part # 13140, of the custom kit were recalled because gloves may contain splits or holes compromising sterility. |
FDA Determined Cause 2 | Unknown/Undetermined by firm |
| Action | Aligned Medical Solutions (AMS) sent the Medtronic Urgent Medical Device for Devon Light Glove letter, dated April 23, 2015 to all affected consignees.
Consignees were informed that the sterile kits assembled and distributed by AMS contain the recalled Devon COVER LIGHT HANDLE DISPOSABLE Glove (AMS part #13140 or Medtronic part # 571711) due to splits or holes.
Consignees were advised to identify any remaining affected kits and follow the over-labeling instruction. At the time the kit is opened for use, any light gloves should be identified and disposed of pursuant to the medical waste policies in effect at your institution.
AMS can provide replacement product if needed or you can contact your local Medtronic sales rep for sterile replacements as well.
Due to the manufacturing process and finished sterilized packs in inventory, or at your facility, over labeling of inventory and/or future shipment affected by this recall will be labeled with a sticker attached to each pack. This sticker will read:
RECALL NOTICE
Medtronic announced the recall l of Devon Light Glove contained in this kit
after it was sealed and sterilized. All other components in the kit are not
affected by this recall.
1. At the time the kit is opened for use any Light Glove manufactured by Medtronic should be identified and set aside
2. The recalled light glove should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution.
Consignees should complete the attached Reply Form and fax to AMS at 406-256-9875.
For any questions about this recall, please call 321-527-7714.
For questions regarding this recall call 404-259-6387. |
| Quantity in Commerce | 192 kits |
| Distribution | Nationwide Distribution to California, Florida, Georgia, Illinois, Kansas, Louisiana, Michigan, New Jersey, Nevada, Texas, Virginia, and Wyoming. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|