| | Class 2 Device Recall RUSCH |  |
| Date Initiated by Firm | November 22, 2016 |
| Create Date | December 16, 2016 |
| Recall Status1 |
Terminated 3 on November 24, 2017 |
| Recall Number | Z-0771-2017 |
| Recall Event ID |
75781 |
| 510(K)Number | K010420 |
| Product Classification |
Catheter, urological - Product Code KOD
|
| Product | RUSCH, Pocket PAC IC Intermittent Cath Kit, 8 FR, 12 FR, 14 FR, 10 FR and 16 FR, Rx Only, Sterile,
Intermittent self-catheterization |
| Code Information |
Product Code: 10096080, Lot number: 74F1501853 (8 FR); Product Code: 10096120, Lot numbers: 74E1602795, 74F1503201, 74G1600723 & 74J1500576 (12 FR); Product Code: 10096140, Lot numbers: 74A1600023, 74A1601105, 74A1602234, 74D1602007, 74E1602235, 74E1602796, 74F1501855, 74G1600486, 74H1500530 & 74L1500363 (14 FR); Product Code: 10096100, Lot numbers: 74A1600799, 74A1600800, 74A1602236, 74A1603122, 74D1600231, 74D1600923, 74D1601540, 74D1602136, 74E1600845, 74E1601768, 74E1602237, 74F1501857, 74F1501962, 74F1601966, 74F1601988, 74H1500534, 74H1500978, 74J1501157, 74K1502306, 74L1500364, 74L1501303, 74L1501938, 74M1500227 and 74M1501087 (10 FR) and Product Code: 10096160, Lot numbers: 74A1600024, 4A1600801, 74A1601106, 74A1603123, 74D1601541, 74E1600846, 74E1602797, 74G1600724, 74H1500979, 74J1501158, 74K1501837, 74L1501202 &74L1501939 (16 FR). |
| FEI Number |
3005747797
|
Recalling Firm/ Manufacturer |
Teleflex Medical 2917 Weck Dr Research Triangle Park NC 27709-0186
|
| For Additional Information Contact | Alice Harper 610-378-0131 |
Manufacturer Reason for Recall | Labeling Inconsistency: The products have been
labeled with the incorrect expiration date which exceeds the actual, validated 35-month shelf life.
|
FDA Determined Cause 2 | Error in labeling |
| Action | Teleflex sent an Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers with affected stock, immediately discontinue use and quarantine any affected products. To return product, customers were advised to complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recall@teleflex.com. Customers with no affected stock should also complete the enclosed Recall Acknowledgment Form to confirm receipt of the letter. Customers with questions should contact Customer Service at 1-866-246-6990. |
| Quantity in Commerce | 102,843 units |
| Distribution | Nationwide Distribution to CA, CO, FL, GA, IL, KS, KY, MA, MO, MS, NC, NJ, NY, OH, OK, OR, PA and TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KOD
|
|
|
|