• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Alteon Neck Preserving Stem

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Alteon Neck Preserving Stemsee related information
Date Initiated by FirmNovember 07, 2016
Create DateJanuary 17, 2017
Recall Status1 Terminated 3 on August 05, 2019
Recall NumberZ-1032-2017
Recall Event ID 75989
510(K)NumberK141821 
Product Classification Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Product Code LZO
ProductAlteon Neck Preserving Stem, Standard Offset, Plasma Coated, Size 6
Code Information Serial No. 3787445-3787447, 3868023-3868072, 4209114-4209135, 4318092-4318102, 4350860-4350869, 4386939, 4420620-4420644, 4558261-4458262; Catalog No. 162-00-06; UDI: 10885862262776
FEI Number 1038671
Recalling Firm/
Manufacturer
Exactech, Inc.
2320 NW 66th Ct
Gainesville FL 32653-1630
For Additional Information ContactKaya Davis
352-377-1140
Manufacturer Reason
for Recall
Products were labeled with incorrect Global Trade Item Numbers. The GTIN on the label identifies the devices as another.
FDA Determined
Cause 2
Under Investigation by firm
ActionIn the recall notice sent to agents, dated 11/01/2016, recipients are instructed to cease product distribution, extend knowledge of the recall to their accounts, identify and quarantine any product in inventory, and complete and return the Recall Inventory Response Form. Consignees are instructed to contact their Exactech Representative within 5 days of receipt of the notice to confirm product quantities at each location. Consignees with any questions are instructed to contact Kaya Davis at kaya@exac.com or 1-800-392-2832.
Quantity in Commerce98
DistributionAZ, CA, CO, FL, IL, KS, MD, NJ, NY, OH, OK, TN, & VA; Germany, Italy, Japan, Spain, & Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = LZO
-
-