• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CheckFlo Performer Introducer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CheckFlo Performer Introducersee related information
Date Initiated by FirmFebruary 01, 2017
Create DateMarch 02, 2017
Recall Status1 Terminated 3 on September 28, 2018
Recall NumberZ-1387-2017
Recall Event ID 76360
510(K)NumberK132592 
Product Classification Introducer, catheter - Product Code DYB
ProductCheck-Flo Performer Introducer, Individual product is packaged in a Tyvek-film sterilizable outer pouch. Introducers are intended for introduction of balloons, closed and non-tapered end catheters or other diagnostic and interventional devices.
Code Information Catalog Identifier: RCFW-4.0(5.0,6.0,7.0,8.0)- / -BV  Codes consist of Lot #, RPN #, Manufacture Date and Expiration Date: 4826254, RCFW-4.0-35-BV, 2014/03, 3/1/2017; 504036, RCFW-4.0-35-BV, 6/1/2014, 6/1/2017; 5364287, RCFW-4.0-35-BV, 10/15/2014, 10/15/2017; 5379911, RCFW-4.0-35-BV, 10/21/2014, 10/21/2017; 5400933, RCFW-4.0-35-BV, 10/29/2014, 10/29/2017; 5435105, RCFW-4.0-35-BV, 11/11/2014, 11/11/2017; 5462040, RCFW-4.0-35-BV, 11/20/2014, 11/20/2017; 5474624, RCFW-4.0-35-BV, 11/25/2014, 11/25/2017; NS5424653, RCFW-4.0-35-BV, 11/6/2014, 11/6/2017; NS5424654, RCFW-4.0-35-BV, 11/6/2014, 11/6/2017; 5213872, RCFW-5.0-35-BV, 8/14/2014, 8/14/2017; 5273219, RCFW-5.0-35-BV, 9/10/2014, 9/10/2017; 5302128, RCFW-5.0-35-BV, 9/23/2014, 9/23/2017;  5354215, RCFW-5.0-35-BV, 10/10/2014, 10/10/2017;  5400931, RCFW-5.0-35-BV, 10/29/2014, 10/29/2017;  5523548, RCFW-5.0-35-BV, 12/17/2014, 12/17/2017;  5681717, RCFW-5.0-35-BV, 2/26/2015, 2/26/2018; 5690760, RCFW-5.0-35-BV, 3/3/2015, 3/3/2018; 5701193, RCFW-5.0-35-BV, 3/6/2015, 3/6/2018; 6004897, RCFW-5.0-35-BV, 7/2/2015, 7/2/2018; 5260343, RCFW-6.0-35-BV, 9/4/2014, 9/4/2017; 5275648, RCFW-6.0-35-BV, 9/11/2014, 9/11/2017; 5308224, RCFW-6.0-35-BV, 9/24/2014, 9/24/2017; 5312947, RCFW-6.0-35-BV, 9/25/2014, 9/25/2017; 5368247, RCFW-6.0-35-BV, 10/16/2014, 10/16/2017; 5383333, RCFW-6.0-35-BV, 10/22/2014, 10/22/2017; 5444866, RCFW-6.0-35-BV, 11/13/2014, 11/13/2017; 5585193, RCFW-6.0-35-BV, 1/16/2015, 1/16/2018; 5681719, RCFW-6.0-35-BV, 2/26/2015, 2/26/2018; 5690162, RCFW-6.0-35-BV, 3/3/2015, 3/3/2018; 5692849, RCFW-6.0-35-BV, 3/4/2015, 3/4/2018; 5974636, RCFW-6.0-35-BV, 6/20/2015, 6/20/2018; 4906939, RCFW-7.0-35-BV, 2014/04, 4/1/2017; 4987018, RCFW-7.0-35-BV, 5/1/2014, 5/1/2017; 5228810, RCFW-7.0-35-BV, 8/21/2014, 8/21/2017; 5275669, RCFW-7.0-35-BV, 9/11/2014, 9/11/2017; 5301927, RCFW-7.0-35-BV, 9/22/2014, 9/22/2017; 5301930, RCFW-7.0-35-BV, 9/22/2014, 9/22/2017; 5312948, RCFW-7.0-35-BV, 9/25/2014, 9/25/2017; NS5831881,RCFW-7.0-38-75-RB-HLA-091100-BV, 4/23/2015, 4/23/2018;  NS5878399, RCFW-7.0-38-75-RB-HLA-091100-BV, 5/12/2015, 5/12/2018; NS5890383, RCFW-7.0-38-75-RB-HLA-091100-BV,5/18/2015, 5/18/2018; NS5922577,RCFW-7.0-38-75-RB-HLA-091100-BV,6/1/2015, 6/1/2018;  NS6000260, RCFW-7.0-38-75-RB-HLA-091100-BV, 6/30/2015, 6/30/2018; NS6091893, RCFW-7.0-38-75-RB-HLA-091100-BV, 8/4/2015, 8/4/2018; NS6264206, RCFW-7.0-38-75-RB-HLA-091100-BV, 10/6/2015, 10/6/2018;  F4845598, RCFW-8.0-35-BV, 2014/03, 3/1/2017;  NS5312949, RCFW-8.0-35-BV, 9/25/2014, 9/25/2017; NS5596871, RCFW-8.0-35-BV, 1/21/2015, 1/21/2018; NS5755064, RCFW-8.0-35-BV, 3/26/2015, 3/26/2018;  NS5793423, RCFW-8.0-35-BV, 4/9/2015, 4/9/2018;  NS5831884,RCFW-8.0-38-75-RB-HLA-091100-BV, 4/23/2015, 4/23/2018; NS5996958 RCFW-8.0-38-75-RB-HLA-091100-BV 6/29/2015 6/29/2018;  NS5996959 RCFW-8.0-38-75-RB-HLA-091100-BV 6/29/2015 6/29/2018;  NS6002168 RCFW-8.0-38-75-RB-HLA-091100-BV 7/1/2015 7/1/2018;  NS6014785 RCFW-8.0-38-75-RB-HLA-091100-BV 7/7/2015, 7/7/2018; NS6014786, RCFW-8.0-38-75-RB-HLA-091100-BV, 7/7/2015, 7/7/2018; NS6022050,RCFW-8.0-38-75-RB-HLA-091100-BV, 7/10/2015, 7/10/2018; NS6031799,RCFW-8.0-38-75-RB-HLA-091100-BV, 7/14/2015, 7/14/2018; NS6083872, RCFW-8.0-38-75-RB-HLA-091100-BV, 7/31/2015, 7/31/2018; NS6091894, RCFW-8.0-38-75-RB-HLA-091100-BV, 8/4/2015, 8/4/2018; NS6140704, RCFW-8.0-38-75-RB-HLA-091100-BV, 8/20/2015, 8/20/2018; NS6225353, RCFW-8.0-38-75-RB-HLA-091100-BV, 9/22/2015, 9/22/2018; NS6295130, RCFW-8.0-38-75-RB-HLA-091100-BV, 10/19/2015, 10/19/2018 ; NS6329426, RCFW-8.0-38-75-RB-HLA-091100-BV, 10/28/2015, 10/28/2018 
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook Inc.
750 N Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactCook Medical Customer Relations Departme
812-339-2235
Manufacturer Reason
for Recall
Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.
FDA Determined
Cause 2
Under Investigation by firm
ActionCook Inc. is initiating a voluntary recall of multiple products following the identification of an increase in reports of blood loss associated with devices using a specific hemostatic valve design referred to as the blue valve or polyisoprene valve. Potential adverse events that may occur if these devices were used in the arterial system (Flexor Radial Access Sets) include delay in procedure and blood loss. If devices were used in the central venous system, adverse events that may occur include delay in procedure, blood loss, or air embolism. The firm is distributing notices via mail courier on 02/01/2017. Customers are being instructed to share the recall notice with appropriate personnel, including down to the user level, within the customer's organization or to any organization where the potentially affected devices have been transferred. 1. Examine inventory immediately to determine if customers have affected product and quarantine affected product(s). 2. Return affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. NOTE: Returned un-affected product will not be credited. 3. Even if customers do not have affected product on hand, customers must still complete the Acknowledgement and Receipt Form and return via fax 812.339.7316 or email to Fieldactionsna@cookmedical.com. 4. Report adverse events to COOK Medical Customer Relations at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time), or by email to: customerrelationsna@cookmedical.com. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA: Online at: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail) Call the FDA at: 1-800-FDA-1088 Effectiveness checks of this recall will be done by response form mailed. All recalled products will be destroyed at the regional distribution warehouses. A cer
Quantity in Commerce884
DistributionWorldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to Canada, Armenia, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Spain, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Kenya, Kuwait, Lithuania, Mauritius, Netherlands, Norway, Oman, Poland, Reunion, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom VA/DOD: Yes
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DYB
-
-