Date Initiated by Firm | April 04, 2016 |
Create Date | April 15, 2017 |
Recall Status1 |
Terminated 3 on June 22, 2020 |
Recall Number | Z-1828-2017 |
Recall Event ID |
76844 |
510(K)Number | K110006 K913929 |
Product Classification |
System, image management, ophthalmic - Product Code NFJ
|
Product | Merge Eye Station and Merge Eye Care PACS. |
Code Information |
All versions |
Recalling Firm/ Manufacturer |
Merge Healthcare, Inc. 900 Walnut Ridge Dr Hartland WI 53029-8347
|
For Additional Information Contact | Merge Service and Support 877-741-5369 |
Manufacturer Reason for Recall | During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently. |
FDA Determined Cause 2 | Software in the Use Environment |
Action | The recalling firm issued a letter dated 3/28/2016 via email the week of 4/4/2016 or via certified mail if they did not have an email address for the customer. A second letter dated 7/11/2016 was issued to nonresponders. |
Quantity in Commerce | 1,627 sites potentially have the affected versions |
Distribution | Distribution was nationwide to medical facilities. Foreign distribution was made to Canada, as well as to other countries. There was also government distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = NFJ
|