Class 2 Device Recall SYNCHRON Systems Microalbumin (MA) Reagent

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall SYNCHRON Systems Microalbumin (MA) Reagentsee related information
Date Initiated by FirmApril 24, 2017
Create DateMay 22, 2017
Recall Status1 Terminated 3 on December 30, 2017
Recall NumberZ-2124-2017
Recall Event ID 77138
510(K)NumberK082251 
Product Classification Albumin, antigen, antiserum, control - Product Code DCF
ProductMicroalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine
Code Information M609458
FEI Number 2050012
Recalling Firm/
Manufacturer
Beckman Coulter Inc.
250 S Kraemer Blvd
Brea CA 92821-6232
Manufacturer Reason
for Recall
Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.
FDA Determined
Cause 2
Under Investigation by firm
ActionBeckman Coulter sent an Urgent Medical Device Recall letter dated April 24, 2017, to affected customers to inform them that Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2. Customers are informed of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to Customer Support Center via http://www.beckmancoulter.com/customersupport/support, call 1-800-854-3633 in the United States and Canada, and outside US and Canada they are instructed to contact their local Beckman Coulter Representative.
Quantity in Commerce3,158 units total (1,957 units in US)
DistributionWorldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Guatemala, India, Italy, Japan, Mayotte, Mexico, Netherlands, Oman, Philippines, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DCF
-
-