| | Class 2 Device Recall SYNCHRON Systems Microalbumin (MA) Reagent |  |
| Date Initiated by Firm | April 24, 2017 |
| Create Date | May 22, 2017 |
| Recall Status1 |
Terminated 3 on December 30, 2017 |
| Recall Number | Z-2124-2017 |
| Recall Event ID |
77138 |
| 510(K)Number | K082251 |
| Product Classification |
Albumin, antigen, antiserum, control - Product Code DCF
|
| Product | Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL
MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine |
| Code Information |
M609458 |
| FEI Number |
2050012
|
Recalling Firm/ Manufacturer |
Beckman Coulter Inc.
250 S Kraemer Blvd
Brea
CA
92821-6232
|
Manufacturer Reason for Recall | Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Beckman Coulter sent an Urgent Medical Device Recall letter dated April 24, 2017, to affected customers to inform them that Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2. Customers are informed of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to Customer Support Center via http://www.beckmancoulter.com/customersupport/support, call 1-800-854-3633 in the United States and Canada, and outside US and Canada they are instructed to contact their local Beckman Coulter Representative. |
| Quantity in Commerce | 3,158 units total (1,957 units in US) |
| Distribution | Worldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Guatemala, India, Italy, Japan, Mayotte, Mexico, Netherlands, Oman, Philippines, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, Taiwan, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = DCF
|
|
|
|