• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall VITEK 2 Gram Negative test kit containing colistin (cs01n)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall VITEK 2 Gram Negative test kit containing colistin (cs01n)see related information
Date Initiated by FirmMay 16, 2017
Date PostedJune 27, 2017
Recall Status1 Terminated 3 on June 19, 2019
Recall NumberZ-2770-2017
Recall Event ID 77373
Product Classification Phenotypic test kit, non-susceptible/elevated MIC organisms, cultured isolates - Product Code PTJ
ProductVITEK 2 Gram Negative test kits containing colistin (cs01n)
Code Information REF # (Description): 418114 (AST-GN94), 22258 (AST-N102), 412605 (AST-N194), 412921 (AST-N199), 412862 (AST-N201), 412863 (AST-N202), 412924 (AST-N205), 412918 (AST-N210), 413061 (AST-N212), 413083 (AST-N222), 413112 (AST-N225), 413171 (AST-N236), 413172 (AST-N237), 413204 (AST-N239), 413205 (AST-N240), 413391 (AST-N242), 413394 (AST-N245), 413397 (AST-N248), 413572 (AST-N249), 413722 (AST-N253), 413724 (AST-N255), 413725 (AST-N256), 413753 (AST-N261), 414164 (AST-N272), 414286 (AST-N276), 414492 (AST-N279), 414531 (AST-N280), 414532 (AST-N281), 414972 (AST-N288), 415063 (AST-N292), 415369 (AST-N293), 415433 (AST-N295), 416005 (AST-N299), 416241 (AST-N300), 416590 (AST-N303), 416913 (AST-N308), 416914 (AST-N309), 417426 (AST-N315), 417952 (AST-N318), 418204 (AST-N320), 418489 (AST-N322), 418513 (AST-N235), 418514 (AST-N326), 418630 (AST-N328), 418631 (AST-N329), 418675 (AST-N331), 418676 (AST-N332), 418985 (AST-N335), 419077 (AST-N338), 419341 (AST-N339), 420063 (AST-N343), 420440 (AST-N344), 420867 (AST-N347), 420856 (AST-N348), 421258 (AST-N352), 421297 (AST-N353), 421350 (AST-N354), 421351 (AST-N355, 421451 (AST-N357), 421441 (AST-N358), 421583 (AST-N360), 421584 (AST-N361), 421585 (AST-N362), 421692 (AST-363), 421693 (AST-N364), 421830 (AST-N365), 421853 (AST-N366), 421854 (AST-N367), 421855 (AST-N368), 410025 (AST-XN01), 413230 (AST-XN05)
FEI Number 1950204
Recalling Firm/
Manufacturer
Biomerieux Inc
595 Anglum Rd
Hazelwood MO 63042-2320
For Additional Information Contact
314-731-8694
Manufacturer Reason
for Recall
The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).
FDA Determined
Cause 2
Other
ActionCustomers were notified on or about 05/16/2017 and then notified again with an updated letter. Instructions to customers included requesting that the attached acknowledgement form be completed, distribute the recall notification to customers, and provide the notice to future customers.
Quantity in Commerce1,520,000 cards
DistributionDomestic: NC. Foreign: Argentina, Austria, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, France, Germany, Guatemala, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Netherlands Antilles, Nicaragua, Panama, Paraguay, Peru, Philippines, Poland, Singapore, South Africa, South Korea, Spain, Thailand, Turkey, United Kingdom, Uruguay, Venezuela, & Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-