| Date Initiated by Firm | June 07, 2017 |
| Create Date | December 08, 2017 |
| Recall Status1 |
Terminated 3 on April 25, 2019 |
| Recall Number | Z-0224-2018 |
| Recall Event ID |
77661 |
| 510(K)Number | K142183 |
| Product Classification |
Introducer, catheter - Product Code DYB
|
| Product | Glidesheath Slender Hydrophilic Coated Introducer Sheath, 5FR - 16CM (0.021), REF/Product Code RM*RS5F16PA, STERILE, Rx ONLY |
| Code Information |
Lot s 161107, 170209, 170307, 170406 |
| FEI Number |
2243441
|
Recalling Firm/ Manufacturer |
Terumo Medical Corp 2101 Cottontail Ln Somerset NJ 08873-1277
|
| For Additional Information Contact | 908-302-4916 |
Manufacturer Reason for Recall | An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | URGENT VOLUNTARY MEDICAL DEVICE RECALL Letters (dated June 7, 2017) and Response Forms were sent to the consignees via UPS on June 7, 2017. |
| Distribution | Nationwide and Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = DYB
|