 Class 2 Device Recallsee related information
Date Initiated by FirmOctober 24, 2017
Create DateAugust 08, 2018
Recall Status1 Terminated 3 on May 18, 2023
Recall NumberZ-2699-2018
Recall Event ID 79315
510(K)NumberK060038 
Product Classification Prosthesis, Elbow, Semi-Constrained, Cemented - Product Code JDB
ProductRadial Head Assy Size #4, Rad. Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.
Code Information 410-0006
FEI Number 3003070421
Recalling Firm/
Manufacturer
Howmedica Osteonics Corp.
325 Corporate Dr
Mahwah NJ 07430-2006
For Additional Information Contact
201-831-5000
Manufacturer Reason
for Recall
Revision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn October 27, 2017 an URGENT MEDICAL RECALL NOTIFICATION was issued to customers titled Product Field Action PR 1642784. The letter requests that recipients notify all necessary individuals within their organization of the recall and that a Stryker trauma representative will collect any affected product from the customer location. Questions or concerns can be directed to 201-831-6693
Quantity in Commerce107
DistributionDistribution US Nationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.