|
Class 2 Device Recall T2100 Treadmill |
|
Date Initiated by Firm |
April 02, 2018 |
Create Date |
July 16, 2018 |
Recall Status1 |
Terminated 3 on April 30, 2024 |
Recall Number |
Z-2467-2018 |
Recall Event ID |
80165 |
Product Classification |
Treadmill, powered - Product Code IOL
|
Product |
T2100 Micro flex Drive Treadmill, powered
Product Usage: Powered exercise equipment consist of powered devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. Specifically the T210 Treadmill is intended for use in Exercise Testing, facilitating accurate blood pressure measurements and exercise testing within speed range of 0-13.5 miles per hour.
|
Code Information |
Treadmill Model Numbers 2026182-004, 2026182-002. Known lot/serial numbers: SBC07152197SA, SBC07372859SA, SJV14412914SA, SBC12038975SA, SBC12169265SA, SJV14052101SA, SJV14092190SA, SBC10376789SA, SBC12079098SA, SBC12079100SA, 5460609, 5448740, SBC12049021SA, SBC12059040SA, SBC12139235SA, SBC07112098SA, SBC07132153SA, SBC10226425SA, SBC12129213SA, SBC06331301SA, SBC10336719SA, SBC06261147SA, SBC09485858SA, SBC10376798SA, SBC12069076SA, SBC12119201SA, SBC12149236SA, SBC11298412SA, SBC11308452SA, SBC11308470SA, SBC12038973SA, SBC12099151SA, SBC12119205SA, SBC12129227SA, SJV14212408SA, SBC06180862SA, SJV14382840SA, SJV14372765SA, SBC11148122SA, SBC11288388SA, SBC12089134SA, SBC06040465SA, SBC12049001SA, SBC12049015SA, SBC12049022SA, SBC12049025SA, SBC11418622SA, SBC11418624SA, SBC11418633SA, SJV12380433SA, SBC06170848SA, SBC06170856SA, SBC06170857SA, SBC05490198SA, SBC08514915SA, SJV12430536SA, SJV14412927SA, SBC06030436SA, SBC06040441SA, SBC06210965SA, 8588656, SBC05510384SA, SBC06130722SA, SBC05500340SA, SJV14422942SA, SJV14422947SA, SJV14422949SA, SBC09255443SA, SBC05490221SA, SBC05490222SA, SBC06180891SA, SBC08504866SA, SBC09505911SA, SBC06130749SA, SBC06411468SA, SBC07092029SA, SBC07102056SA, SBC10226427SA, SBC10226440SA, 8607326, SBC10126219SA, SBC12129231SA, 8684256, SBC10146319SA, SBC08113613SA, SBC10447008SA, SJV13491889SA, SBC06050491SA, SBC06060504SA, SBC06060506SA, SBC08514909SA, SBC11087833SA, SBC11308456SA, SBC05450051SA, SBC06451629SA, SJV13090902SA, SJV13100933SA, SJV14242531SA, SBC08364359SA, SBC08113610SA, SBC06070556SA, SBC07483181SA, SBC10046047SA, SBC06020398SA, SBC06421536SA, SBC07262546SA, SBC10306592SA, SBC07362801SA, SBC07372827SA, SBC07372845SA, SBC07071945SA, SBC06241052SA, SBC08394479SA, SJV12500630SA, SJV14112230SA, SJV14112231SA, SJV12230057SA, SBC07443006SA, SJV12230051SA, SBC05500262SA, SJV14122241SA, SBC08173770SA, SBC07483162SA, SJV12280193SA, SBC12209392SA, SBC08233964SA, SBC09235395SA, SBC06210962SA, SBC06501767SA, SBC08073469SA, SBC08444580SA, SBC08504888SA, SBC09235365SA, SBC09235372SA, SBC10116211SA, SBC10487101SA, SBC10487108SA, SBC11037470SA, SBC11107931SA, SJV12310277SA, 49004822, SBC10126261SA, SBC11047579SA, SBC11067690SA, SBC07372858SA, SBC12169271SA, SJV13020725SA, SJV13020748SA, SJV13030762SA, SJV13030764SA, SJV13030765SA, SJV13321453SA, SBC09505909SA, SJV14252581SA, SBC06241066SA, SBC06251096SA, SBC10226442SA, SBC10236455SA, SBC11318483SA, SBC07302628SA, SBC10497150SA, SJV13371543SA, SJV13371584SA, SBC06040447SA, SBC06040453SA, SBC06050474SA, SBC11148121SA, SBC11198194SA, SBC11198195SA, SBC12159257SA, SJV13231183SA, SJV14042060SA, SJV14042071SA, SJV14242559SA, SJV15093259SA, SJV15103276SA, SJV15123323SA, SJV15123324SA, SJV15123325SA, SJV15133326SA, SJV15133328SA, SJV15133335SA, SBC11488845SA, SBC11388575SA, SBC05440028SA, SBC05510374SA, SBC07362820SA, SBC08504878SA, SBC10036029SA, SBC06311238SA, SBC11037463SA, SBC07152191SA, SBC07162242SA, SBC05450073SA, SBC05460079SA, SBC07081995SA, SJV13271281SA, SJV13431709SA, SBC08254038SA, 80147358, 9081830, 9039955, 8831712, SBC12189303SA, SBC11518960SA, SBC06130748SA, SJV13301384SA, SJV13301398SA, SBC07252491SA, SBC08294117SA, SBC08514918SA, 310080281, SBC06321272SA, SBC06231017SA, SBC06501775SA, SBC08093511SA, SBC08193828SA, SBC09285519SA, SJV1347182SA, SBC10106161SA, SBC08384443SA, SBC10527330SA, SJV13020722SA, SJV13020740SA, SJV13511979SA, SBC06080572SA, SJV13421679SA, SBC09064993SA, SBC11168152SA, SK214110302SA, SBC10507209SA, SBC07513326SA, SM914340066SA, SM914340067SA, SM914340068SA, SJV13040779SA, SBC08053411SA, SBC11518959SA, SBC10326632SA, SBC11067677SA, SJV14493180SA, SJV13481867SA, SBC10176317SA, SBC10477061SA, SJV13281298SA, SJV14372805SA, SBC10226428SA, SBC06471687SA, SBC07252501SA, SBC07483146SA, SBC08033380SA, SBC08223923SA, SBC08454638SA, SBC09475822SA, SBC11238324SA, SBC11468781SA, SBC11468782SA, SJV13381587SA, SJV13381589SA, SJV13481865SA, SJV14082173SA, SJV14372766SA, SJV14422943SA, SJV14463085SA, SJV14483150SA, SJV14493181SA, SJV13090914SA, SJV14242543SA, SJV13281302SA, SJV13371576SA, SJV14242544SA, SBC07071973SA, SJV14102226SA, SBC08223924SA, SJV14372788SA, SBC08454646SA, SBC08324217SA, SBC08324220SA, SJV14463082SA, SBC06030417SA, SJV13110964SA, SBC06120680SA, SBC11308442SA, SBC05460107SA, SBC05460110SA, SBC05510360SA, SBC06140770SA, SBC06231013SA, SBC06381421SA, SBC06511793SA, SBC07132171SA, SBC07202323SA, SBC08113632SA, SBC08334237SA, SBC08504882SA, SBC08504894SA, SBC08504895SA, SBC08504897SA, SBC11208216SA, SBC11448717SA, SBC11448722SA, SBC11448727SA, SK212490021SA, SK213400253SA, SK214200313SA, SK214250338SA, SK214250339SA, SM914330013SA, SM914340054SA, SM914340055SA, SM914340056SA, SM914533287SA, SJV13171049SA, SJV14372798SA, SBC08374417SA, SBC08384428SA, SJV12410498SA, SBC10376777SA, SJV12460591SA, SJV14372779SA, SBC09455800SA, SBC10236474SA, SBC12099146SA, SBC12099147SA, SBC12099148SA, SJV12410506SA, SBC11468780SA, SBC10086079SA, SBC10326658SA, SBC05510380SA, SBC06361358SA, SBC07112090SA, SBC07362787SA, SBC07362805SA, SBC07372840SA, SBC07372846SA, SBC08173769SA, SBC08173785SA, SBC08414549SA, SBC08484772SA, SBC09175202SA, SBC09485867SA, SBC10086088SA, SBC10106175SA, SBC10146305SA, SBC10326629SA, SBC10326634SA, SBC10326686SA, SBC10336713SA, SBC10336746SA, SBC10336763SA, SBC11047598SA, SBC11047636SA, SBC11057644SA, SBC11067646SA, SBC11067673SA, SBC11067698SA, SBC11077724SA, SBC11077792SA, SBC11087813SA, SBC11468768SA, SBC11468771SA, SBC11468777SA, SBC11488838SA, SBC11498877SA, SBC12038987SA, SBC12099162SA, SBC12129207SA, SJV12270126SA, SK213070114SA, SBC11067647SA, SJV12410517SA, SBC05500279SA, SBC06301206SA, SBC12049004SA, SBC07312659SA, SBC10507235SA, SJV12380444SA, SJV12400492SA, SJV13140997SA, SJV13421680SA, SJV14232495SA, SJV14412916SA, SJV14433001SA, SJV14493190SA, SBC08484765SA, SBC07312658SA, SBC07092045SA, SJV12430546SA, SJV14102220SA, SJV14102221SA, SBC06301207SA, SBC06331304SA, SBC12089123SA, SJV15093229SA, SBC06321267SA, SBC06321277SA, SBC06361378SA, SBC07302635SA, SBC07302637SA, SBC07453048SA, SBC08203842SA, SBC08284108SA, SBC08454643SA, SBC09225348SA, SBC10086087SA, SBC10236476SA, SBC10256525SA, SBC10326637SA, SBC10336742SA, SBC10416941SA, SBC10457042SA, SBC11067676SA, SBC11107893SA, SBC11228304SA, SBC11228310SA, SBC11268367SA, SBC11468805SA, SBC12079096SA, SJV12240093SA, SJV12250096SA, SJV12350346SA, SJV12440554SA, SJV12460594SA, SJV12480609SA, SJV12500653SA, SJV13070868SA, SJV13080877SA, SJV13291320SA, SJV13351489SA, SJV14032048SA, SJV14082174SA, SJV14092195SA, SJV14232468SA, SJV14362752SA, SJV15053127SA, SJV15053128SA, SBC05470148SA, SBC06140760SA, SJV13141007SA, SBC07493208SA, SBC11037493SA, SBC08203855SA, SBC08474732SA, SBC11138115SA, SBC08213900SA, SBC10326640SA, SBC11047626SA, SBC08093520SA, SBC08213906SA, SBC08354275SA, SBC08354276SA, SBC10086071SA, SBC10086089SA, SBC10166342SA, SBC10196396SA, SBC10326627SA, SBC11037479SA, SBC11037495SA, SBC11037511SA, SBC11037521SA, SBC11037527SA, SBC11037554SA, SBC11047571SA, SBC11047591SA, SBC11047597SA, SBC11047618SA, SBC11047630SA, SBC11067666SA, SBC11067668SA, SBC11077780SA, SBC11087840SA, SBC11107916SA, SBC11117998SA, SBC11128053SA, SBC11128054SA, SBC11128059SA, SBC11128069SA, SBC11138096SA, SBC11138110SA, SBC10336718SA, SBC11037497SA, SBC11037477SA, SBC11067656SA, SBC08474717SA, SBC07493240SA, SBC08093509SA, SBC08093555SA, SBC08053429SA, SBC11077723SA, SBC11117992SA, SBC11128027SA, SBC08093561SA, SBC08504852SA, SBC09245426SA, SBC09485842SA, SBC10106160SA, SBC11047632SA, SBC11128070SA, SJV14473133SA, SBC10447019SA, SBC11128044SA, SBC11097849SA, SBC11107926SA, SBC08093562SA, SBC08213889SA, SBC08213892SA, SBC09455809SA, SBC10386824SA, SBC10386830SA, SBC11037500SA, SBC11118002SA, SBC11037485SA, SBC10446993SA, SBC08113602SA, SBC10246491SA, SBC10376817SA, SBC10507207SA, SBC11047601SA, SBC11107888SA, SBC11107911SA, SBC10426959SA, SBC08193837SA, SBC08213910SA, SBC08213912SA, SBC08213915SA, SBC08354270SA, SBC09054986SA, SBC09195242SA, SBC09305551SA, SBC09475837SA, SBC10086076SA, SBC10186391SA, SBC10226412SA, SBC10326620SA, SBC10446987SA, SBC10507208SA, SBC10517290SA, SBC11037494SA, SBC11037501SA, SBC11037504SA, SBC11037509SA, SBC11037519SA, SBC11037528SA, SBC11037531SA, SBC11037536SA, SBC11047576SA, SBC11047577SA, SBC11047622SA, SBC11047624SA, SBC11067674SA, SBC11067699SA, SBC11067707SA, SBC11077729SA, SBC11087809SA, SBC11097867SA, SBC11117980SA, SBC11128026SA, SBC11138104SA, SBC08213918SA, SBC08354285SA, SBC09395692SA, SBC11067664SA, SBC10086090SA, SBC11067663SA, SJV14382880SA, SJV14392887SA, SBC10136284SA, SBC11087829SA, SBC11128048SA, SBC11097868SA, SBC08354272SA, SBC11118021SA, SBC11107951SA, SBC08484801SA, SJV14392883SA, SBC10396843SA, SBC11037512SA, SBC11047589SA, SBC11087841SA, SBC11107881SA, SBC11107925SA, SBC11128064SA, SBC11117963SA, SBC11117991SA, SJV12270141SA, SBC05450040SA, SJV15123302SA, SJV15123312SA, SJV15123318SA, SJV15123321SA, SBC08093515SA, SBC10446990SA, SBC09265472SA, SBC09465810SA, SBC10086075SA, SBC10396848SA, SBC10507193SA, SBC11037478SA, SBC11037534SA, SBC11047583SA, SBC11047596SA, SBC11047599SA, SBC11047619SA, SBC11067685SA, SBC11117983SA, SBC11118000SA, SBC11138090SA, SBC11138117SA, SBC08223926SA, SBC08354288SA, SBC08414534SA, SBC08434563SA, SBC09415728SA, SBC09475829SA, SBC10336757SA, SBC10446988SA, SBC11037483SA, SBC11037496SA, SBC11037549SA, SBC11037551SA, SBC11047578SA, SBC11047580SA, SBC11047584SA, SBC11047594SA, SBC11047604SA, SBC11067651SA, SBC11067657SA, SBC11067691SA, SBC11077728SA, SBC11077793SA, SBC11097855SA, SBC11097856SA, SBC11097859SA, SBC11097860SA, SBC11097861SA, SBC11107910SA, SBC11117979SA, SBC11118016SA, SBC11128051SA, SBC11128057SA, SBC11128058SA, SBC11128078SA, SBC08053410SA, SBC08213903SA, SBC08213908SA, SBC08354267SA, SBC09195247SA, SBC09225325SA, SBC10046036SA, SBC10086083SA, SBC10096139SA, SBC10376794SA, SBC10507192SA, SBC10517273SA, SBC11037561SA, SBC11047569SA, SBC11047593SA, SBC11067682SA, SBC11077721SA, SBC11077795SA, SBC11087808SA, SBC11117975SA, SBC11117990SA, SBC11117999SA, SBC11128049SA, SBC11128052SA, SBC11128056SA, SBC11128067SA, SBC11128071SA, SBC08354274SA, SBC10086082SA, SBC10436974SA, SBC10446983SA, SBC10446989SA, SBC10446994SA, SBC10446996SA, SBC10446997SA, SBC10446998SA, SBC10446999SA, SBC10447004SA, SBC10447012SA, SBC10447014SA, SBC10447015SA, SBC11017365SA, SBC11037515SA, SBC11037532SA, SBC11047590SA, SBC11107948SA, SBC08073463SA, SBC11057642SA, SBC11118014SA, SBC11037481SA, SBC10396855SA, SBC08484750SA, SBC11107919SA, SBC11047620SA, SBC09185241SA, SBC11047621SA, SBC11077718SA, SBC11097852SA, SBC11117997SA, SBC11128022SA, SBC09054951SA, SBC09475828SA, SBC11128075SA, SBC11107935SA, SBC11128028SA, SBC10236461SA, SJV14392886SA, SBC10326630SA, SBC10376784SA, SBC10507241SA, SBC10507246SA, SBC10507262SA, SBC10517263SA, SBC10517268SA, SBC11107889SA, SBC08053432SA, SBC08083491SA, SBC10386831SA, SBC08354287SA, SBC08444581SA, SBC10166326SA, SBC11047570SA, SBC11047616SA, SBC11077781SA, SBC11097862SA, SBC11128039SA, SBC10447003SA, SBC07503260SA, SBC08053430SA, SBC08063447SA, SBC08484781SA, SBC09065015SA, SBC09195269SA, SBC09455808SA, SBC10086084SA, SBC10396842SA, SBC10507201SA, SBC11037506SA, SBC11037525SA, SBC11037533SA, SBC11047573SA, SBC11047574SA, SBC11047592SA, SBC11077783SA, SBC11118005SA, SBC11128063SA, SBC11128076SA, SBC11138098SA, SBC11158140SA, SBC11097854SA, SBC08213896SA, SBC09215319SA, SBC09295535SA, SBC10336755SA, SBC10497178SA, SBC11047564SA, SBC11067648SA, SBC11067660SA, SBC11087839SA, SBC11097851SA, SBC11107905SA, SBC11118018SA, SBC11128055SA, SBC11138108SA, SBC11138112SA, SBC11138116SA, SBC11118012SA, SBC08093552SA, SBC09195257SA, SBC11067689SA, SBC09265478SA, SBC11077779SA, SBC10336709SA, SBC11087831SA, SBC08414538SA, SBC08494802SA, SBC09485850SA, SBC10166336SA, SBC10226434SA, SBC10326663SA, SBC10326680SA, SBC10507236SA, SBC11067701SA, SBC11107946SA, SBC11117989SA, SBC11138094SA, SBC11138099SA, SBC08213917SA, SBC11138092SA, SBC11138097SA, SBC10166330SA, SBC11037553SA, SBC11138113SA, SBC10416942SA, SBC11128077SA, SBC10166353SA, SBC11138095SA, SBC08093542SA, SBC08474730SA, SBC09305557SA, SBC11037513SA, SBC11067665SA, SBC11087801SA, SJV14392888SA, SBC10126249SA, SBC11118011SA, SBC11128050SA, SBC07493251SA, SBC08434558SA, SBC08464699SA, SBC10236470SA, SBC10507191SA, SBC11037488SA, SBC11077725SA, SBC11107932SA, SBC11118007SA, SBC11118013SA, SBC11128062SA, SBC11138083SA, SBC11138105SA, SJV15053137SA, SJV15183481SA, SBC10507194SA, SBC11037520SA, SBC11037541SA, SBC11107895SA, SBC11107929SA, SBC11107930SA, SBC11118004SA, SBC08474720SA, SBC09215314SA, SBC08354289SA, SBC08474712SA, SBC09475836SA, SBC09485849SA, SBC10086067SA, SBC10106167SA, SBC10126264SA, SBC10236464SA, SBC10396841SA, SBC10446991SA, SBC10497184SA, SBC10507195SA, SBC10507199SA, SBC10507229SA, SBC10507240SA, SBC10507256SA, SBC10517271SA, SBC11037556SA, SBC11047625SA, SBC11067667SA, SBC11067684SA, SBC11067706SA, SBC11077719SA, SBC11077726SA, SBC11077742SA, SBC11087835SA, SBC11087838SA, SBC11097843SA, SBC11107880SA, SBC11117971SA, SBC11118020SA, SBC11128045SA, SBC11128073SA, SBC11138079SA, SBC11138088SA, SBC11138103SA, SBC09054958SA, SBC11037474SA, SJV13020752SA, SJV13231190SA, SBC05440033SA, SBC06110664SA, SBC06301219SA, SBC06411484SA, SBC06461651SA, SBC06501762SA, SBC07071955SA, SBC07102070SA, SBC07162225SA, SBC07372855SA, SBC07372862SA, SBC08073476SA, SBC08073482SA, SBC08083489SA, SBC08233952SA, SBC08233959SA, SBC08243969SA, SBC08243971SA, SBC08444598SA, SBC09085077SA, SBC09185233SA, SBC09505893SA, SBC09525960SA, SBC10096115SA, SBC10096118SA, SBC10166365SA, SBC10336699SA, SBC10376783SA, SBC10376802SA, SBC10376812SA, SBC10376814SA, SBC10426965SA, SBC11017378SA, SBC11037460SA, SBC11117958SA, SBC11117995SA, SBC11138085SA, SBC11178176SA, SBC11318485SA, SBC11388573SA, SBC11458752SA, SBC11498885SA, SBC12049024SA, SBC12089122SA, SBC12089129SA, SJV12250104SA, SJV12300251SA, SJV12370415SA, SJV12370420SA, SJV12400495SA, SJV12440567SA, SJV12460600SA, SJV12520708SA, SJV13020737SA, SJV13020738SA, SJV13090908SA, SJV13140998SA, SJV13141004SA, SJV13141011SA, SJV13151014SA, SJV13151027SA, SJV13181072SA, SJV13391610SA, SJV13391612SA, SJV13441741SA, SJV13501943SA, SJV13511972SA, SJV14032032SA, SJV14032034SA, SJV14052092SA, SJV14052099SA, SJV14222432SA, SJV14222434SA, SJV14282622SA, SJV14282623SA, SJV14372791SA, SJV14443040SA, SJV14453064SA, SJV15023052SA, SJV15033066SA, SJV15033077SA, SJV15083190SA, SJV15123307SA, SJV15123308SA, SJV15123309SA, SBC06110654SA, SBC06311239SA, SBC06311248SA, SBC06311250SA, SBC07152199SA, SBC07162217SA, SBC07162233SA, SBC07342755SA, SBC07342757SA, SBC07362786SA, SBC08374381SA, SBC08374387SA, SBC08464703SA, SBC08474727SA, SBC08484783SA, SBC09085118SA, SBC09095131SA, SBC09215305SA, SBC10126255SA, SBC10226414SA, SBC10226415SA, SBC11067679SA, SJV12310265SA, SJV13421665SA, SJV13511974SA, SJV14032038SA, SBC08113608SA, SBC08314189SA, SBC09044925SA, SBC09044932SA, SBC10296567SA, SBC10296577SA, SBC11128035SA, SBC11218272SA, SBC11218274SA, SBC11248353SA, SJV13441781SA, SJV13441782SA, SK212500046SA, SK212500047SA, SM914340047SA, SK213160157SA, SBC07422941SA, SBC07503304SA, SJV13511978SA, SJV14132276SA, SBC11508951SA, SBC05480173SA, SBC07182282SA, SBC07453046SA, SBC06080580SA, SBC06100643SA, SBC06291195SA, SBC07332721SA, SBC08083490SA, SBC08404506SA, SBC08444593SA, SBC10046042SA, SBC10046043SA, SBC10046050SA, SBC10376803SA, SBC10376804SA, SJV14432975SA, SJV15183492SA, SBC05470128SA, SBC07503270SA, SBC09075033SA, SBC10296571SA, SBC11037505SA, SBC11508929SA, SK213170181SA, SBC07513320SA, SBC07513339SA, SK213160175SA, SK213170180SA, SBC05470132SA, SBC06040449SA, SBC06160803SA, SBC06210953SA, SBC06241084SA, SBC07041853SA, SBC07041858SA, SBC07051876SA, SBC07132147SA, SBC07202329SA, SBC07453009SA, SBC07503281SA, SBC08213914SA, SBC09075048SA, SBC10296561SA, SBC10306581SA, SBC10416936SA, SBC10507198SA, SBC11077776SA, SBC11428648SA, SBC11428652SA, SBC11438686SA, SBC11498904SA, SK213160173SA, SK213170176SA, SK214200311SA, SK214200319SA, SK214200322SA, SK214200323SA, SK214200325SA, SK214200315SA, SBC06070527SA, SBC06040442SA, SBC06130715SA, SBC07503294SA, SBC06070545SA, SBC07192304SA, SBC08213916SA, SBC10296570SA, SBC10296573SA, SBC11248348SA, SK213160174SA, SBC06050490SA, SBC06130716SA, SBC07021826SA, SBC07202335SA, SBC07513314SA, SBC10296563SA, SBC11428650SA, SK213170182SA, SK214200307SA, SM914350089SA, SM915063155SA, SM915073157SA, SM915073158SA, SM915073162SA, SBC06020396SA, SBC06040443SA, SBC06090607SA, SBC06130725SA, SBC10296564SA, SBC11418637SA, SBC07513333SA, SBC09065009SA, SBC07513312SA, SBC10507204SA, SK214200318SA, SBC11248359SA, SBC06170841SA, SBC06461652SA, SBC06501761SA, SJV12240078SA, SJV12240079SA, SBC06331302SA, SBC06421513SA, SBC07021829SA, SBC07031840SA, SBC07302651SA, SBC08033384SA, SBC10276549SA, SK213120143SA, SBC05460100SA, SJV14362726SA, SBC06251134SA, SBC10256533SA, SJV12390464SA, SBC05450052SA, SBC05450053SA, SBC05470116SA, SBC05470121SA, SBC05470124SA, SBC05470125SA, SBC05480163SA, SBC05490209SA, SBC05490240SA, SBC05500248SA, SBC05500269SA, SBC05500270SA, SBC05500278SA, SBC05500297SA, SBC05500301SA, SBC05500302SA, SBC06080595SA, SBC06090605SA, SBC06090611SA, SBC06090620SA, SBC06130730SA, SBC06200937SA, SBC06261137SA, SBC06321281SA, SBC06441589SA, SBC06441596SA, SBC06441598SA, SBC06451612SA, SBC06451621SA, SBC06451623SA, SBC06451628SA, SBC07021830SA, SBC07051879SA, SBC07051881SA, SBC07051885SA, SBC07051887SA, SBC07061905SA, SBC07061909SA, SBC07061913SA, SBC07071944SA, SBC07071948SA, SBC07071953SA, SBC07071954SA, SBC07071971SA, SBC07232420SA, SBC07232422SA, SBC07232426SA, SBC07232431SA, SBC07242432SA, SBC07242434SA, SBC07252444SA, SBC07252445SA, SBC07252450SA, SBC07252458SA, SBC07252528SA, SBC07322707SA, SBC07332718SA, SBC07332720SA, SBC07342749SA, SBC07422929SA, SBC07422932SA, SBC07422933SA, SBC07422936SA, SBC07422937SA, SBC07432971SA, SBC07432982SA, SBC07432983SA, SBC07432985SA, SBC08404499SA, SBC08434566SA, SBC08454620SA, SBC08464690SA, SBC08464704SA, SBC08464705SA, SBC08464709SA, SBC08474734SA, SBC08484746SA, SBC09054983SA, SBC09175185SA, SBC09175195SA, SBC09175197SA, SBC09175198SA, SBC09185221SA, SBC09185231SA, SBC09235357SA, SBC09235358SA, SBC09235359SA, SBC09235360SA, SBC09235363SA, SBC09295531SA, SBC09395684SA, SBC09395685SA, SBC09395695SA, SBC09405716SA, SBC09405717SA, SBC09415726SA, SBC09445755SA, SBC09445756SA, SBC09445767SA, SBC10025983SA, SBC10096117SA, SBC10106152SA, SBC10106157SA, SBC10106162SA, SBC10136279SA, SBC10136280SA, SBC10146312SA, SBC10146316SA, SBC10176368SA, SBC10176380SA, SBC10176384SA, SBC10176387SA, SBC10196399SA, SBC10196407SA, SBC10256538SA, SBC10256539SA, SBC10266541SA, SBC10276544SA, SBC10296579SA, SBC10436977SA, SBC10436979SA, SBC10457041SA, SBC10467049SA, SBC10477057SA, SBC10477059SA, SBC10507214SA, SBC10517288SA, SBC10517316SA, SBC11017371SA, SBC11027411SA, SBC11047585SA, SBC11057640SA, SBC11077737SA, SBC11107956SA, SBC11128029SA, SBC11128042SA, SBC11138102SA, SBC11178167SA, SBC11198196SA, SBC11198197SA, SBC11228292SA, SBC11228293SA, SBC11248350SA, SBC11248355SA, SBC11248356SA, SBC11398599SA, SBC11398600SA, SBC11428654SA, SBC11438678SA, SBC11438681SA, SBC11438689SA, SBC11438690SA, SBC11448733SA, SBC11498894SA, SBC11508912SA, SBC11508914SA, SBC11508915SA, SBC11508916SA, SBC11508923SA, SBC11508934SA, SBC12169286SA, SBC12199334SA, SBC12199339SA, SBC12209382SA, SBC12209383SA, SK212490019SA, SK212500034SA, SK212500035SA, SK212500037SA, SK212500038SA, SK212500039SA, SK213050085SA, SK213060100SA, SK213070109SA, SK213070113SA, SK213070121SA, SK213080129SA, SK213110137SA, SK213120144SA, SK213120145SA, SK213120147SA, SK213120152SA, SK213120153SA, SK213160166SA, SK213200199SA, SK213200200SA, SK213220216SA, SK213220217SA, SK213220218SA, SK213250222SA, SK213480270SA, SK213480273SA, SK214100291SA, SK214110297SA, SK214200326SA, SK214200329SA, SM914330009SA, SM914330018SA, SM914330022SA, SM914330027SA, SM914340035SA, SM914340040SA, SM914340042SA, SM914340043SA.SM914340075SA, SM914340083SA, SM914340086SA, SM914390100SA, SM914390107SA, SM914390109SA, SM914513274SA, SM914513275SA, SM915043109SA, SM915063153SA, SM915073163SA, SJV14453070SA, SJV14412921SA, SJV14453048SA, SJV12510660SA, SJV12480621SA, SJV13070871SA, SJV12510676SA, SBC06080575SA, SJV13140996SA, SJV12460601SA, SJV12510672SA, SBC06090614SA, SBC06100630SA, SBC06060522SA, SBC05490220SA, SJV13060851SA, SJV13100958SA, SJV13090911SA, SJV13110973SA, SJV13231178SA, SJV12460590SA, SJV12510659SA, SBC05460104SA, SJV12510679SA, SBC08334246SA, SBC10246514SA, SBC11087823SA, SJV14052100SA, SJV14082180SA, SJV14202379SA, SJV14412917SA, SJV14422959SA, SJV14432963SA, SBC11288376SA, SBC11077789SA, SBC10246494SA, SBC10126256SA, SBC11268361SA, SJV12270134SA, SJV14282624SA, SBC08033372SA, SJV14062130SA, SJV14062131SA, SBC05490226SA, SBC05490235SA, SBC06421499SA, SBC06431578SA, SBC06441588SA, SBC06491736SA, SBC07182285SA, SBC07182297SA, SBC07212351SA, SBC07222372SA, SBC07222382SA, SBC07222386SA, SBC07362813SA, SBC07372851SA, SBC08063448SA, SBC08193832SA, SBC08203859SA, SBC08304172SA, SBC08334261SA, SBC08364351SA, SBC08434575SA, SBC09054966SA, SBC09225351SA, SBC09415729SA, SBC10085102SA, SBC10096140SA, SBC10106148SA, SBC10136286SA, SBC10226424SA, SBC10416909SA, SBC10416912SA, SBC10497170SA, SBC10507260SA, SBC11218273SA, SJV12220025SA, SJV12300239SA, SJV12300258SA, SJV13291337SA, SJV13301354SA, SJV13311425SA, SJV14172360SA, SJV14212399SA, SJV14222437SA, SJV14232496SA, SBC08063443SA, SBC11318484SA, SBC06481707SA, SBC08364314SA, SBC12099166SA, SBC12109168SA, SBC07071941SA, SBC07222395SA, SBC09044940SA, SBC10246502SA, SJV13421663SA, SJV13421666SA, SJV14092198SA, SJV12340316SA, SJV14463089SA, SBC07222385SA, SBC08284111SA, SBC10326633SA, SJV14212413SA, SJV15033063SA, SBC07182267SA, SBC10246505SA, SBC10326651SA, SJV14503226SA, SBC07222367SA, SBC08093536SA, SBC09065008SA, SBC09285515SA, SBC10477055SA, SJV12220037SA, SJV12340320SA, SJV12350345SA, SJV13100919SA, SJV13321446SA, SJV13461810SA, SJV13471834SA, SJV13481864SA, SJV14032037SA, SJV14092194SA, SJV14122255SA, SJV14232498SA, SBC07081996SA, SBC10406907SA, SBC10416919SA, SJV14433004SA, SBC11218278SA, SBC11228285SA, SJV15083167SA, SJV15083168SA, SBC11087818SA, SJV12290200SA, SJV12290201SA, SJV13301373SA, SBC08163758SA, SJV14372786SA, SJV12270135SA, SJV14102208SA, SJV12220028SA, SBC08404493SA, SJV13441756SA, SBC11308463SA, SBC10096142SA, SJV13271287SA, SJV13461791SA, SJV13461793SA, SJV13461802SA, SJV13461807SA, SJV14372773SA, SJV14433000SA, SBC09475813SA, sjv13361533sa, sjv13371564sa, SJV14122240SA, SBC11087836SA, SBC07182270SA, SBC09085083SA, SBC10116194SA, SJV12220032SA, SBC070819965A, SBC07442994SA, SBC07082004SA, SBC07082006SA, SBC07082007SA, SBC07082009SA, SBC07092013SA, SBC07352781SA, SBC07372826SA, SBC09265462SA, SBC09265476SA, SBC09265481SA, SBC09265484SA, SBC09265489SA, SBC09305561SA, SBC09345606SA, SBC09365650SA, SBC11107870SA, SBC11107884SA, SBC11288374SA, SBC11378552SA, SBC11378557SA, SJV12310264SA, SJV13441732SA, SJV14082182SA, SJV14483135SA, SJV14483136SA, SBC10246490SA, SJV12280179SA, SBC09064999SA, SBC06080587SA, SBC06361370SA, SBC07352778SA, SBC08033386SA, SBC09195250SA, SBC09195270SA, SBC10046038SA, SBC10406892SA, SBC10517315SA, SBC11077746SA, SBC12059047SA, SJV12220002SA, SJV12240067SA, SJV13070869SA, SJV13261267SA, SJV14062138SA, SJV14282600SA, SJV14372800SA, SJV14432982SA, SJV144329985PA, SJV14432998SA, SJV14463094SA, SJV14473115SA, SJV14473131SA, SJV15043114SA, SJV15173436SA, SJV13321454SA, SJV12280184SA, SJV13321439SA, SJV14232499SA, SJV14422939SA, SBC07332724SA, SBC08504881SA, SBC09305559SA, SBC10116198SA, SJV12310282SA, SJV13281304SA, SBC08203877SA, SBC09475827SA, SBC11097847SA, SJV13431701SA, SJV13461808SA, SJV14433002SA, SBC10246513SA, SBC10246515SA, SJV14112236SA, SBC08354273SA, SBC09345611SA, SJV14142318SA, SJV15173432SA, SBC06231001SA, SBC07222403SA, SBC08233966SA, SBC11378556SA, SBC08203856SA, SBC09255439SA, SJV12350360SA, SBC08043395SA, SBC11087824SA, SBC11498899SA, SBC07182281SA, SJV14372780SA, SBC06481701SA, SBC09075067SA, SBC09475823SA, SBC10096137SA, SBC11087815SA, SBC11218262SA, SBC11288377SA, SJV13311412SA, SJV13341464SA, SJV13351476SA, SJV13351483SA, SJV13411623SA, SJV14372793SA, SJV15043115SA, SJV15053124SA, SJV15133341SA, SJV15163429SA, SBC07102055SA, SJV13371566SA, SJV13381592SA, SBC10136294SA, SBC12038969SA, SJV14222446SA, SBC08193827SA, SBC10326654SA, SBC11378554SA, SJV13431719SA, SJV13461788SA, SBC11208227SA, SBC07092016SA, SBC08374384SA, SBC11308455SA, SBC08203869SA, SJV12270149SA, SJV14102224SA, SBC11378553SA, SBC11468794SA, SJV13251241SA, SBC06481712SA, SBC07342753SA, SBC10096143SA, SJV14102205SA, SJV14282591SA, SJV14282599SA, SBC10527350SA, SBC08414514SA, SBC12109170SA, SBC12109171SA, SBC06361355SA, SBC064917145SA, SBC07202311SA, SBC07212355SA, SBC07222375SA, SBC07222384SA, SBC08073456SA, SBC08083506SA, SBC08133675SA, SBC08173786SA, SBC08203860SA, SBC08203872SA, SBC08203879SA, SBC08213886SA, SBC08384433SA, SBC08414508SA, SBC08464686SA, SBC08464706SA, SBC08494842SA, SBC09064994SA, SBC09255451SA, SBC10126237SA, SBC10186394SA, SBC10246482SA, SBC10316605SA, SBC10336765SA, SBC10406906SA, SBC10517265SA, SBC11107878SA, SJV12220009SA, SJV12290203SA, SJV13151036SA, SJV13351490SA, SJV13371568SA, SJV13381596SA, SJV13431715SA, SJV13441733SA, SJV13441735SA, SJV14042068SA, SJV14082175SA, SJV14102209SA, SJV14132291SA, SJV14172368SA, SJV14463102SA, SJV14483173SA, SJV14493209SA, SBC09225354SA, SJV15133342SA, SBC09335585SA, SBC10186393SA, SJV14082181SA, SBC07222374SA, SBC11288381SA, SBC12099161SA, SBC09225346SA, SJV13481869SA, SBC06421506SA, SJV14483152SA, SBC11288397SA, SBC08043396SA, SJV13221169SA, SJV13522005SA, SJV14142298SA, SBC06311258SA, SBC06421525SA, SBC06441591SA, SBC06491732SA, SBC07071957SA, SBC07172262SA, SBC07222368SA, SBC07222370SA, SBC07222377SA, SBC07312672SA, SBC07372842SA, SBC07372849SA, SBC07412910SA, SBC07463095SA, SBC08304148SA, SBC08414529SA, SBC09054946SA, SBC09215315SA, SBC09265480SA, SBC09265483SA, SBC09485855SA, SBC10096146SA, SBC10106149SA, SBC10106150SA, SBC10386834SA, SBC11027415SA, SBC11178156SA, SBC11218247SA, SBC11298411SA, SBC11378537SA, SBC11468795SA, SBC12209396SA, SJV12370391SA, SJV12370397SA, SJV13261265SA, SJV13341462SA, SJV13431720SA, SJV13461813SA, JV14092187SA, SJV14302672SA, SJV14412913SA, SJVI2510676SA, SJVI2510680SA, SBC08073462SA, SBC09425740SA, SJV13431695SA, SJV13431698SA, SJV13431699SA, SJV13431700SA, SJV14102207SA, SJV14102210SA, SBC11037523SA, SBC09235384SA, SBC06231028SA, SBC08414544SA, SBC07382893SA, SBC11178178SA, SJV13411653SA, SJV14483151SA, SBC06070539SA, SBC06481706SA, SBC07041866SA, SBC07372828SA, SBC08454648SA, SBC09455803SA, SBC10086102SA, SBC10096141SA, SBC11097846SA, SBC11107903SA, SBC12069065SA, SBC12069071SA, SBC12099145SA, SJV14212401SA, SJV14212407SA, SJV12220005SA, SBC10326673SA, SBC09054969SA, SBC10126230SA, SBC11498878SA, SBC09065016SA, SBC10256526SA, SJV14102222SA, SJV14142315SA, SBC09054961SA, SBC08364303SA, SBC09044923SA, SBC10036010SA, SBC10086072SA, SBC11067669SA, SBC11208215SA, SBC11218258SA, SJV13231197SA, SJV14372785SA, SBC10376815SA, SBC10246506SA, SBC06491743SA, SBC10126235SA, SJV13110976SA, SBC07352772SA, SBC07352773SA, SBC11488851SA, SBC11378558SA, SJV13441750SA, SJV14453078SA, SJV14292646SA, SBC06471674SA, SJV13231177SA, SBC11228300SA, SJV13231199SA, SJV14232471SA, SJV13391602SA, SJV12480604SA, SJV14142297SA, SJV14483174SA, SJV13291318SA, SJV13291347SA, SJV13291349SA, SJV14432996SA, SJV14232483SA, SBC06331303SA, SBC08464698SA, SJV12440552SA, SJV14082159SA, SBC10477052SA, SBC05490201SA, SJV12510680SA, SBC06080569SA, SJV12510683SA, SJV13109435SA, SJV13211155SA, SJV13050805SA, SJV13110966SA, SJV13050803SA, SJV13181067SA, SJV13100957SA, SBC06070553SA, SJV12460603SA, SBC05450042SA, SBC05510371SA, SBC05510372SA, SBC05510381SA, SBC06050492SA, SBC06070531SA, SBC06080585SA, SBC06251101SA, SBC06251103SA, SBC06311224SA, SBC06311240SA, SBC06361381SA, SBC06371396SA, SBC06371405SA, SBC06471686SA, SBC06471689SA, SBC06481694SA, SBC06481697SA, SBC06481699SA, SBC06481703SA, SBC06481709SA, SBC07092018SA, SBC07092020SA, SBC07092025SA, SBC07092035SA, SBC07102060SA, SBC07172264SA, SBC07232418SA, SBC07252461SA, SBC07252468SA, SBC07342730SA, SBC07342742SA, SBC07342744SA, SBC07352771SA, SBC07372841SA, SBC07372850SA, SBC07453029SA, SBC07453036SA, SBC07453037SA, SBC07453038SA, SBC07453043SA, SBC07453055SA, SBC07453061SA, SBC07463074SA, SBC07463075SA, SBC07463082SA, SBC07463089SA, SBC07463091SA, SBC07463092SA, SBC07463099SA, SBC07463102SA, SBC07463103SA, SBC08073470SA, SBC08073471SA, SBC08073473SA, SBC08073475SA, SBC08073477SA, SBC08073478SA, SBC08073480SA, SBC08073484SA, SBC08073485SA, SBC08093508SA, SBC08203868SA, SBC08203875SA, SBC08203876SA, SBC08203881SA, SBC08233929SA, SBC08233930SA, SBC08233941SA, SBC08233942SA, SBC08233943SA, SBC08233944SA, SBC08233947SA, SBC08233948SA, SBC08233949SA, SBC08233954SA, SBC08233957SA, SBC08334248SA, SBC08354265SA, SBC08354271SA, SBC08374412SA, SBC08384435SA, SBC08384447SA, SBC08384450SA, SBC08384453SA, SBC08394468SA, SBC08394473SA, SBC08394474SA, SBC08394480SA, SBC08394481SA, SBC08404502SA, SBC08414517SA, SBC08414535SA, SBC08414536SA, SBC08414542SA, SBC08414545SA, SBC09064989SA, SBC09064992SA, SBC09065018SA, SBC09075075SA, SBC09205283SA, SBC09215310SA, SBC09225331SA, SBC09225347SA, SBC09235368SA, SBC09235376SA, SBC09235382SA, SBC09255458SA, SBC09345613SA, SBC09415738SA, SBC09475838SA, SBC09485869SA, SBC09485870SA, SBC10126227SA, SBC10126251SA, SBC10126258SA, SBC10146321SA, SBC10166332SA, SBC10166333SA, SBC10166345SA, SBC10166357SA, SBC10176372SA, SBC10176378SA, SBC10176385SA, SBC10186392SA, SBC10196409SA, SBC10386833SA, SBC10386836SA, SBC10386838SA, SBC10406905SA, SBC10426949SA, SBC10426963SA, SBC10446984SA, SBC10446985SA, SBC10446986SA, SBC10477066SA, SBC10487119SA, SBC11037516SA, SBC11037518SA, SBC11037543SA, SBC11037548SA, SBC11037562SA, SBC11047565SA, SBC11047617SA, SBC11057643SA, SBC11067649SA, SBC11067653SA, SBC11067662SA, SBC11077790SA, SBC11087827SA, SBC11097842SA, SBC11097865SA, SBC11107891SA, SBC11107953SA, SBC11117993SA, SBC11118003SA, SBC11118009SA, SBC11128023SA, SBC11128033SA, SBC11138084SA, SBC11288403SA, SBC11498907SA, SBC12059033SA, SBC12059035SA, SBC12059042SA, SBC12059045SA, SBC12069063SA, SBC12069069SA, SBC12079080SA, SBC12079101SA, SBC12099165SA, SBC12109175SA, SBC12149238SA, SBC12149240SA, SBC12209395SA, SJV12220003SA, SJV12220020SA, SJV12220024SA, SJV12220030SA, SJV12220033SA, SJV12220034SA, SJV12220035SA, SJV12250117SA, SJV12250120SA, SJV12260121SA, SJV12260122SA, SJV12260123SA, SJV12280186SA, SJV12290209SA, SJV12290227SA, SJV12300233SA, SJV12300235SA, SJV12330287SA, SJV12340309SA, SJV12340310SA, SJV12340311SA, SJV12340312SA, SJV12340315SA, SJV12340317SA, SJV12340318SA, SJV12340319SA, SJV12340329SA, SJV12340331SA, SJV12340332SA, SJV12350334SA, SJV12350341SA, SJV12350354SA, SJV12360367SA, SJV12360371SA, SJV12360376SA, SJV12360381SA, SJV12380436SA, SJV12380459SA, SJV13211149SA, SJV13211159SA, SJV13211160SA, SJV13211163SA, SJV13211164SA, SJV13231179SA, SJV13241211SA, SJV13241215SA, SJV13241220SA, SJV13241221SA, SJV13291346SA, SJV13301391SA, SJV13321448SA, SJV13341466SA, SJV13341467SA, SJV13351481SA, SJV13351493SA, SJV13361495SA, SJV13361496SA, SJV13361505SA, SJV13361526SA, SJV13361529SA, SJV13371542SA, SJV13371567SA, SJV13371573SA, SJV13411618SA, SJV13411619SA, SJV13431693SA, SJV13441744SA, SJV13441749SA, SJV13441765SA, SJV13471849SA, SJV13481851SA, SJV13481857SA, SJV13481859SA, SJV14062140SA, SJV14062145SA, SJV14062147SA, SJV14062153SA, SJV14062154SA, SJV14092196SA, SJV14102202SA, SJV14152333SA, SJV14152334SA, SJV14152337SA, SJV14152339SA, SJV14152340SA, SJV14212386SA, SJV14212415SA, SJV14212421SA, SJV14222439SA, SJV14222442SA, SJV14222450SA, SJV14222452SA, SJV14222453SA, SJV14222455SA, SJV14222456SA, SJV14232461SA, SJV14362753SA, SJV14372792SA, SJV14422948SA, SJV14432961SA, SJV14433003SA, SJV14443017SA, SJV14483149SA, SJV14513264SA.SJV15023043SA, SJV15083185SA, SJV15103270SA, SJV15123295SA, SJV15123296SA, SJV15123306SA, SJV15123311SA, SJV15123315SA, SJV15123316SA, SJV15123317SA, SJV15123319SA, SJV15123320SA, SJV15133343SA, SJV15133345SA, SJV15133346SA, SBC06130714SA, SJV13080868SA, SBC11508921SA, SBC05500324SA, SBC10507226SA, SBC11027434SA, SBC12199341SA, SJV14453061SA, SJV14453066SA, SJV14513271SA, SBC05460098SA, SBC05470137SA, SBC05480170SA, SBC05480178SA, SBC05480180SA, SBC05490208SA, SBC06030414SA, SBC06070543SA, SBC06120705SA, SBC06170834SA, SBC06241078SA, SBC06411490SA, SBC06421501SA, SBC06421508SA, SBC06421514SA, SBC06461666SA, SBC06471672SA, SBC07021825SA, SBC07051883SA, SBC07061902SA, SBC07061906SA, SBC07432960SA, SBC07442986SA, SBC07442987SA, SBC07442991SA, SBC07442993SA, SBC07443000SA, SBC07443008SA, SBC07453012SA, SBC08324200SA, SBC08334232SA, SBC08334241SA, SBC08334251SA, SBC09054963SA, SBC09175184SA, SBC09175191SA, SBC09175192SA, SBC09185209SA, SBC09185215SA, SBC09405712SA, SBC09445777SA, SBC10166344SA, SBC10166355SA, SBC10376795SA, SBC10416908SA, SBC10416925SA, SBC10416927SA, SBC10477077SA, SBC11027391SA, SBC11027392SA, SBC11027418SA, SBC11037486SA, SBC11037491SA, SBC11077734SA, SBC11107936SA, SBC11158141SA, SBC11188189SA, SBC11208205SA, SBC11428644SA, SBC12109189SA, SJV13301372SA, SJV13321434SA, SJV13321435SA, SJV13351472SA, SJV13511958SA, SJV13511961SA, SJV13511968SA, SJV14362723SA, SJV14362730SA, SK213040074SA, SK213070116SA, SM914330024SA, SM915023280SA, M915023281SA, SM915023284SA, SK213170183SA, SBC10336691SA, SBC11438712SA, SBC09285518SA, SBC11067705SA, SBC11077730SA, SJV13040782SA, SBC07432969SA, SBC07432976SA, SBC07442992SA, SBC07443003SA, SBC08324228SA, SBC08334238SA, SBC08334245SA, SBC08354286SA, SBC09054982SA, SBC10136269SA, SBC11027433SA, SBC11027436SA, SBC11027438SA, SBC11027440SA, SBC11027441SA, SBC11027443SA, SBC11027454SA, SBC11037507SA, SBC11077735SA, SBC11087828SA, SBC11117996SA, SBC11128030SA, SBC11128032SA, SBC11128038SA, SBC11128040SA, SBC11128041SA, SBC11208200SA, SBC11288383SA, SBC11288390SA, SBC11288399SA, SBC11298404SA, SBC11308464SA, SBC11308465SA, SBC11308469SA, SBC11318480SA, SBC11318481SA, SBC11318486SA, SBC11318487SA, SBC11318488SA, SBC11318489SA, SBC11318490SA, SBC11358496SA, BC11358497SA, SBC11358498SA, SBC11358500SA, SBC11358501SA, SBC11358502SA, SBC11358505SA, SBC11358506SA, SBC11358508SA, SBC11358509SA, SBC11358510SA, SBC11358511SA, SBC11358513SA, SBC11358514SA, SBC11358515SA, SBC11358516SA, SBC11358519SA, SBC11358520SA, SBC11358524SA, SBC11368525SA, SBC11368526SA, SBC11368527SA, SBC11368528SA, SBC11368529SA, SBC11368531SA, SBC12169289SA, SBC12229441SA, SJV12280182SA, SJV12280183SA, SJV12290220SA, SJV12330292SA, SJV12370414SA, SJV12400473SA, SJV12400475SA, SJV12400477SA, SJV12400479SA, SJV13080893SA, sjv13431694sa, sjv13441767sa, SJV14132277SA, SJV13070866SA, SJV13070877SA, SJV13070888SA, SJV13171048SA, SJV12280185SA, SJV13080898SA, SJV13171045SA, SJV12310281SA, SBC11508943SA, SJV13070872SA, SJV13080881SA, SBC10136270SA, SBC11438684SA, SBC09535968SA, SBC10025978SA, SBC11057637SA, SBC11077782SA, SBC12189314SA, SJV12340327SA, SJV12500635SA, SBC08354284SA, SBC10126259SA, SBC10326688SA, SJV12510674SA, SJV14062128SA, SJV14132285SA, SJV14222454SA, SBC06491740SA, SBC08254051SA, SBC09375671SA, SBC08264082SA, SBC10126257SA, SBC10497142SA, SBC11047623SA, SBC11087820SA, SBC11418628SA, SJV12380425SA, SJV13291314SA, SJV14262582SA, SJV14282608SA, SBC07302626SA, SBC10126262SA, SBC10246493SA, SBC12099150SA, SBC12189299SA, SJV12280171SA, SJV12510665SA, SJV13211153SA, SJV14282621SA, SBC10487099SA, SBC06391446SA, SBC07483149SA, SBC08414513SA, SBC09185224SA, SJV14222430SA, SJV14222451SA, SBC08093517SA, SBC08334247SA, SBC08414548SA, SBC10326621SA, SJV12520709SA, SJV13241214SA, SBC06231030SA, SBC06411478SA, SBC07152200SA, SBC07302634SA, SBC08203884SA, SBC08233960SA, SBC08404507SA, SBC08414532SA, SBC08414543SA, SBC09485854SA, SBC10236468SA, SBC10316612SA, SBC10326669SA, SBC10487124SA, SBC12059032SA, SBC12189315SA, SJV12270155SA, SJV12380455SA, SJV12420522SA, SJV13221171SA, SJV13291341SA, SJV14142306SA, SJV14232464SA, SJV14242534SA, SJV14372782SA, SBC10025971SA, SBC10497143SA, SJV12220015SA, SJV13241212SA, SJV14282620SA, SBC07252539SA, SBC08093556SA, SBC10116205SA, SJV12430538SA, SJV14032025SA, SJV15123293SA, SBC07182274SA, SBC10086080SA, SJV14282606SA, SBC07503272SA, SBC09165183SA, SBC09265468SA, SBC10025974SA, SBC10326624SA, SBC11077794SA, SJV12270163SA, SJV12370389SA, SJV12510670SA, SBC07071937SA, SBC07302653SA, SBC10126260SA, SBC05450062SA, SBC05450064SA, SBC06030415SA, SBC06341318SA, SBC06351329SA, SBC06411485SA, SBC06421529SA, SBC07061929SA, SBC07061931SA, SBC07222396SA, SBC07302640SA, SBC07422925SA, SBC08033378SA, SBC08083486SA, SBC08223928SA, SBC08294134SA, SBC08304169SA, SBC08414510SA, SBC08434552SA, SBC09054985SA, SBC09165165SA, SBC10025987SA, SBC10096112SA, SBC10126254SA, SBC10166328SA, SBC10196408SA, SBC10487111SA, SBC10517320SA, SBC10537353SA, SBC11027452SA, SBC11037459SA, SBC11107947SA, SBC12089115SA, SBC12089132SA, SBC12199333SA, SJV12220004SA, SJV12310271SA, SJV12340323SA, SJV13211154SA, SJV13261260SA, SJV13261261SA, SJV13261263SA, SJV13291343SA, SJV14032027SA, SJV14142307SA, SJV14242526SA, SJV14372770SA, SJV14372771SA, SBC12089128SA, SJV13040792SA, SJV14242533SA, SBC06461661SA, SBC07061910SA, SBC07182271SA, SBC07503280SA, SBC07513335SA, SBC08133691SA, SBC08153734SA, SBC08193822SA, SBC08474721SA, SBC08494844SA, SBC09044939SA, SBC09044942SA, SBC09065024SA, SBC09075045SA, SBC10096120SA, SBC10136276SA, SBC11218233SA, SBC11218280SA, SBC11428660SA, SBC11438691SA, SJV13070886SA, SJV13201131SA, SJV13201133SA, SJV14132269SA, SBC05510355SA, SBC07503283SA, SBC08334236SA, SBC09075034SA, SBC09185210SA, SBC10416913SA, SBC11077732SA, SBC12219416SA, SK213480265SA, SK214060270SA, SK214250342SA, SBC05500284SA, SBC06050473SA, SBC08133694SA, SK214060272SA, M914340052SA, SBC09054968SA, SBC06060510SA, SBC07071961SA, SBC07132170SA, SBC08083504SA, SBC08103568SA, SBC08384434SA, SBC08454622SA, SBC09415722SA, SBC09415735SA, SBC10166346SA, SBC10166348SA, SBC10166352SA, SK213070117SA, SK213480277SA, SBC05470119SA, SBC05470141SA, SBC05470143SA, SBC05470146SA, SBC05470147SA, SBC05490245SA, SBC06120667SA, SBC06220986SA, SBC06231035SA, SBC06231036SA, SBC06231037SA, SBC06231038SA, SBC06361376SA, SBC06361380SA, SBC06361382SA, SBC06371384SA, SBC06371387SA, SBC06371390SA, SBC06431563SA, SBC06501751SA, SBC06501755SA, SBC07081993SA, SBC07152203SA, SBC07182294SA, SBC07212343SA, SBC07212356SA, SBC07222394SA, SBC07262554SA, SBC07342737SA, SBC07342746SA, SBC07342756SA, SBC07352763SA, SBC07352764SA, SBC07372834SA, SBC07463094SA, SBC07463101SA, SBC07483137SA, SBC08053422SA, SBC08193834SA, SBC08193835SA, SBC08203844SA, SBC08203848SA, SBC08203849SA, SBC08203852SA, SBC08404504SA, SBC08444619SA, SBC08454624SA, SBC08454633SA, SBC08454649SA, SBC08494837SA, SBC09054976SA, SBC09085084SA, SBC09205282SA, SBC09215320SA, SBC09225330SA, SBC09235373SA, SBC09235375SA, SBC09235377SA, SBC09235383SA, SBC09275493SA, SBC09275498SA, SBC09315564SA, SBC09325576SA, SBC09335582SA, SBC09335587SA, SBC09335588SA, SBC09335591SA, SBC09335594SA, SBC09335596SA, SBC09335601SA, SBC09375679SA, SBC09375681SA, SBC10336693SA, SBC10336730SA, SBC10336733SA, SBC10336750SA, SBC10336762SA, SBC10346768SA, SBC10386832SA, SBC10396854SA, SBC10396858SA, SBC10396862SA, SBC10396873SA, SBC10396875SA, SBC10396881SA, SBC10416914SA, SBC10416921SA, SBC10416939SA, SBC11208226SA, SBC11218244SA, SBC11218252SA, SBC11218253SA, SBC11218254SA, SBC11218261SA, SBC11218269SA, SBC11218270SA, SBC11388559SA, SBC11388560SA, SBC11388562SA, SBC11388564SA, SBC11388566SA, SBC11388569SA, SBC11398613SA, SBC11458753SA, SBC11468775SA, SBC11468779SA, SBC11468786SA, SBC11468793SA, SBC11468803SA, SBC11468806SA, SBC11498906SA, SBC11508946SA, SBC11518954SA, SBC11518955SA, SBC12109176SA, SJV12270162SA, SJV12280164SA, SJV12280168SA, SJV12280169SA, SJV12280174SA, SJV12370417SA, SJV12380432SA, SJV12380451SA, SJV12400489SA, SJV12410505SA, SJV12410516SA, SJV12510675SA, SJV12510678SA, SJV12510682SA, SJV13050794SA, SJV13050810SA, SJV13060849SA, SJV13060856SA, SJV13060862SA, SJV13151030SA, SJV13171046SA, SJV13171050SA, SJV13171055SA, SJV13171059SA, SJV13191077SA, SJV13191078SA, SJV13211146SA, SJV13231198SA, SJV13301358SA, SJV13301364SA, SJV13301376SA, SJV13301380SA, SJV13301381SA, SJV13311408SA, SJV13391604SA, SJV13391605SA, SJV13421657SA, SJV13421658SA, SJV13421660SA, SJV13441745SA, SJV13441764SA, SJV13441769SA, SJV13441771SA, SJV13441772SA, SJV14222436SA, SJV14222440SA, SJV14222441SA, SJV14222443SA, SJV14222444SA, SJV14232466SA, SJV14232479SA, SJV14232480SA, SJV14232481SA, SJV14232482SA, SJV14242551SA, SJV14282605SA, SJV14282607SA, SJV14282609SA, SJV14282611SA, SJV14282613SA, SJV14372795SA, SJV14422933SA, SJV14422945SA, SJV14443019SA, SJV14443034SA, SJV14443038SA, SJV14463090SA, SJV14463093SA, SJV14463095SA, SJV15093225SA, SJV15093232SA, SJV15093238SA, SJV15093264SA, SJV15103274SA, SJV13361524SA, SBC07453021SA, SJV13100928SA, SBC06020387SA, SBC06030416SA, SBC05500320SA, SBC06020391SA, SBC06241060SA, SBC06030439SA, SBC06441600SA, SBC11037539SA, SBC07463085SA, SBC07463079SA, SBC08414521SA, SBC08384441SA, SBC07212350SA, SBC07453060SA, SBC09044938SA, SBC09365643SA, SBC11107928SA, SBC08083501SA, SBC05450066SA, SBC05500250SA, SBC06020386SA, SJV13491884SA, SBC10497134SA, SBC07463105SA, SBC06140766SA, SBC06150779SA, SBC06150780SA, SBC11037545SA, SBC11128061SA, SBC11468802SA, SBC05490197SA, SBC07463069SA, SBC07453052SA, SBC11047610SA, SBC12219415SA, SBC08093553SA, SBC08103582SA, SBC11027458SA, SBC11358499SA, SBC11358512SA, SJV12480619SA, SK214200331SA, SBC12169282SA, SJV12450573SA, SBC08053413SA, SJV13351477SA, SBC08324205SA, SJV12450572SA, SJV12300260SA, SBC10166335SA, SBC11508948SA, SJV12280188SA, 475552, SJV14433013SA, SJV14443026SA, SJV14443035SA, SBC06321276SA, SBC10086081SA, SBC10096119SA, SBC10096133SA, SBC11027405SA, SBC11168147SA, SBC11308468SA, SJV13381588SA, SBC09085091SA, SBC05470145SA, SBC06030438SA, SBC07061926SA, SBC07081974SA, SBC07122130SA, SBC07372836SA, SBC07382886SA, SBC08193810SA, SBC08304170SA, SBC08324195SA, SBC09215303SA, SBC11037473SA, SBC12010009SA, SJV12430528SA, SJV12430529SA, SJV13100952SA, SJV13100961SA, SJV13291329SA, SJV14282594SA, SBC05490243SA, SBC07122139SA, SBC07122141SA, SBC07122144SA, SBC07122146SA, SBC07453058SA, SBC08294113SA, SBC08374401SA, SBC08444585SA, SBC08444586SA, SBC08444587SA, SBC08444589SA, SBC08444591SA, SBC08444592SA, SBC08444594SA, SBC08444597SA, SBC10126224SA, SBC10487118SA, SBC11037490SA, SJV13301393SA, SJV13301397SA, SJV13321441SA, SJV13321449SA, SJV13391613SA, SJV14032029SA, SBC06100641SA, SBC06110651SA, SBC06190907SA, SJV12480611SA, SJV14372796SA, SBC06231025SA, SBC06231040SA, SBC06231041SA, SBC06231042SA, SBC06231043SA, SBC06241045SA, SBC06311261SA, SBC06361367SA, SBC06441584SA, SBC06471681SA, SBC06501747SA, SBC07071966SA, SBC07081975SA, SBC07081982SA, SBC07092010SA, SBC07092017SA, SBC07092019SA, SBC07092026SA, SBC07092039SA, SBC07182298SA, SBC07192301SA, SBC07202316SA, SBC07212353SA, SBC07232415SA, SBC07252520SA, SBC07252521SA, SBC07252529SA, SBC07282575SA, SBC07282580SA, SBC07322680SA, SBC07322699SA, SBC07322705SA, SBC07332726SA, SBC08073450SA, SBC08113607SA, SBC08133648SA, SBC08133660SA, SBC08133670SA, SBC08133679SA, SBC08203847SA, SBC08203857SA, SBC08203858SA, SBC08294120SA, SBC08324191SA, SBC08334260SA, SBC08444606SA, SBC08444607SA, SBC08454630SA, SBC08454632SA, SBC08454650SA, SBC08464651SA, SBC08464652SA, SBC09054967SA, SBC09065011SA, SBC09065021SA, SBC09075047SA, SBC09205301SA, SBC09215309SA, SBC09225328SA, SBC09225332SA, SBC09225336SA, SBC09225337SA, SBC09225339SA, SBC09225352SA, SBC09285513SA, SBC09365635SA, SBC09365649SA, SBC09495876SA, SBC09495877SA, SBC09495878SA, SBC09495879SA, SBC09495883SA, SBC10126226SA, SBC10136301SA, SBC10326678SA, SBC10336711SA, SBC10336722SA, SBC10336727SA, SBC10336736SA, SBC10336754SA, SBC10336758SA, SBC10376779SA, SBC10406903SA, SBC10416920SA, SBC10507237SA, SBC10507238SA, SBC10507248SA, SBC10507259SA, SBC10507261SA, SBC10517267SA, SBC10517269SA, SBC10517276SA, SBC10517281SA, SBC10517284SA, SBC11017373SA, SBC11017374SA, SBC11027380SA, SBC11027428SA, SBC11037492SA, SBC11037502SA, SBC11047586SA, SBC11047611SA, SBC11067703SA, SBC11087812SA, SBC11107941SA, SBC11128047SA, SBC11178170SA, SBC11238314SA, SBC11238327SA, SBC11238330SA, SBC11238331SA, SBC11238333SA, SBC11238339SA, SBC11238341SA, SBC11238343SA, SBC11238344SA, SBC11238345SA, SBC11238346SA, SBC12038981SA, SBC12049019SA, SBC12079092SA, SBC12109177SA, SBC12109178SA, SBC12109179SA, SBC12109181SA, SBC12109183SA, SBC12109184SA, SBC12119198SA, SBC12119199SA, SJV12340308SA, SJV13321452SA, SJV13321455SA, SJV13341460SA, SJV13341461SA, SJV13361502SA, SJV13361503SA, SJV13361520SA, SJV13371544SA, SJV13371545SA, SJV13371556SA, SJV13371557SA, SJV13371572SA, SJV13371575SA, SJV13371581SA, SJV13381595SA, SJV13461809SA, SJV13481856SA, SJV13481871SA, SJV13481873SA, SJV13491877SA, SJV13491881SA, SJV13501940SA, SJV13511949SA, SJV13511956SA, SJV14232459SA, SJV14232473SA, SJV14232474SA, SJV14232476SA, SJV14232477SA, SJV14232488SA, SJV14232494SA, SJV14242560SA, SJV14252564SA, SJV14262585SA, SJV14262586SA, SJV14282587SA, SJV14282588SA, SJV14282589SA, SJV14282590SA, SJV14282602SA, SJV14282603SA, SJV14372763SA, SJV14372764SA, SJV14372769SA, SJV14372777SA, SJV14372790SA, SJV14463097SA, SBC11017367SA, SBC11208206SA, SBC11398605SA, SBC11428640SA, SBC07102072SA, SBC07503299SA, SBC07503302SA, SBC08103581SA, SBC08213909SA, SBC08444600SA, SBC08464697SA, SBC08504890SA, SBC09075056SA, SBC09075057SA, SBC09165179SA, SBC10186388SA, SBC10186389SA, SBC10336712SA, SBC10386827SA, SBC10396845SA, SBC10447022SA, SBC10447023SA, SBC10487092SA, SBC11017370SA, SBC11037552SA, SBC11067712SA, SBC11067713SA, SBC11077740SA, SBC11158144SA, SBC11188183SA, SBC11208202SA, SBC11208213SA, SBC11228290SA, SBC11358495SA, SBC11398590SA, SBC11398591SA, SBC11448735SA, SBC11448736SA, SBC11498888SA, SBC11508918SA, SBC11508927SA, SBC12169281SA, SBC12199346SA, SBC12209356SA, SK212470003SA, SK212480009SA, SK212480010SA, SK212500027SA, SK212500036SA, SK213040069SA, SK213040070SA, SK213040071SA, SK213040072SA, SK213040073SA, SK213060098SA, SK213060099SA, SK213070102SA, SK213070105SA, SK213070106SA, SK213070123SA, SK213070124SA, SK213120150SA, SK213120151SA, SK213170186SA, SK213200193SA, SK213200194SA, SK213200195SA, SK213220201SA, SK213400239SA, SK213400240SA, SK213480268SA, SK213480281SA, SK213480282SA, SK214060269SA, SK214060271SA, SK214060274SA, SK214200321SA, SK214200324SA, SK214200328SA, SM914340036SA, SM914340037SA, SM914340038SA, SM914340039SA, SM915073160SA, SM915083216SA, SJV15033073SA, SBC06281160SA, SBC05450058SA, SBC06190899SA, SBC06261148SA, SJV13301353SA, SJV14463092SA, SBC10126253SA, SBC11047605SA, SBC10136272SA, SBC10136285SA, SBC10477071SA, SBC09415723SA, SBC11128036SA, SBC11248351SA, SK213180190SA, SM915053122SA, SBC06090612SA, SBC06251095SA, SBC06441595SA, SBC07222376SA, SBC07432975SA, SBC08113589SA, SBC08254016SA, SBC08354293SA, SBC08394470SA, SBC09445749SA, SBC09445778SA, SBC10136278SA, SBC10426964SA, SBC10447018SA, SBC11218234SA, SBC11438687SA, SBC11438688SA, SBC12109187SA, SBC12209361SA, SJV13080864SA, SJV13201120SA, SJV14172366SA, SJV14202372SA, SJV14202375SA, SJV14473119SA, SBC05470123SA, SBC05490205SA, SBC05450054SA, SBC05490231SA, SBC06170845SA, SBC06361375SA, SBC06491717SA, SBC06491727SA, SBC06501752SA, SBC07081999SA, SBC07092014SA, SBC07352777SA, SBC07362804SA, SBC07453028SA, SBC07463068SA, SBC07463086SA, SBC08083498SA, SBC08173788SA, SBC08173789SA, SBC08354297SA, SBC08474728SA, SBC08474731SA, SBC08494817SA, SBC08504853SA, SBC09225350SA, SBC09485871SA, SBC10025976SA, SBC10126231SA, SBC10136291SA, SBC10176375SA, SBC10186390SA, SBC10246483SA, SBC10306586SA, SBC10326648SA, SBC10336706SA, SBC10336715SA, SBC10336735SA, SBC10396852SA, SBC10447017SA, SBC10447026SA, SBC10507220SA, SBC10507239SA, SBC10507247SA, SBC11077722SA, SBC11087837SA, SBC11097844SA, SBC11097869SA, SBC11107917SA, SBC11107924SA, SBC11128025SA, SBC11498902SA, SBC12069067SA, SBC12099140SA, SBC12119202SA, SBC12129209SA, SBC12129219SA, SBC12129224SA, SBC12209371SA, SBC12219410SA, SBC12219418SA, SJV12270140SA, SJV12270160SA, SJV12310267SA, SJV12330297SA, SJV12370411SA, SJV12400484SA, SJV12420525SA, SJV12420526SA, SJV12490625SA, SJV13151017SA, SJV13151033SA, SJV13191110SA, SJV13221168SA, SJV13311410SA, SJV13311415SA, SJV13461796SA, SJV13471816SA, SJV13471823SA, SJV13471828SA, SJV14122244SA, SJV14132283SA, SJV14172352SA, SJV14212392SA, SJV14212394SA, SJV14212395SA, SJV14222447SA, SJV14222449SA, SJV14242507SA, SJV14242512SA, SJV14372768SA, SJV14372774SA, SJV14372775SA, SJV14372781SA, SJV14372794SA, SJV14382859SA, SJV14382871SA, SJV14392882SA, SJV14422940SA, SJV14463091SA, SJV15023035SA, SJV15063143SA, SJV15093260SA, SJV15133339SA, SJV15163431SA, SBC06180894SA, SBC07352775SA, SBC11087803SA, SK214100295SA, SBC07112089SA, SBC07162214SA, SBC07322674SA, SBC07503286SA, SBC08043399SA, SBC09395690SA, SBC09415731SA, SJV13511969SA, SJV13511980SA, SBC08324229SA, SBC08334242SA, SBC09455794SA, SBC12199336SA, SJV12510693SA, SJV13140981SA, SBC05470139SA, SBC06020401SA, SBC06040454SA, SBC06040460SA, SBC06090613SA, SBC06130736SA, SBC06130738SA, SBC06451613SA, SBC07061920SA, SBC07392904SA, SBC09044921SA, SBC09065022SA, SBC09175186SA, SBC09295528SA, SBC10136268SA, SBC10296558SA, SBC10296575SA, SBC10326667SA, SBC10497179SA, SBC11047615SA, SBC11067715SA, SBC11208223SA, SBC11248358SA, SBC11248360SA, SBC11498892SA, SBC11498910SA, SBC12169284SA, SBC12219422SA, SJV12290231SA, SJV12510694SA, SJV13040781SA, SJV13060841SA, SJV13060845SA, SJV13060846SA, SJV13070885SA, SJV13080867SA, SJV13100926SA, SJV13140979SA, SJV13321433SA, SJV13351474SA, SJV13441737SA, SJV13441755SA, SJV13441759SA, SJV13511971SA, SJV13511981SA, SM914330012SA, SM914330017SA, SM914330028SA, SM914340057SA, SM914350091SA, SM914390104SA, SM914493184SA, SM914493187SA, SBC10025991SA, SBC10086092SA, SK213400243SA, SBC07102069SA, BC08163762SA, SJV13110968SA, SJV14092192SA, SBC06050495SA, SBC06421516SA, SBC06431566SA, SBC07011813SA, SBC07061915SA, SK213110140SA, SK213220208SA, SK213030049SA, SK213080132SA, SBC09455787SA, SBC06421510SA, SK212500042SA, SBC11107937SA, SBC06030435SA, SBC09455786SA, SK213110139SA, SBC07031832SA, SBC10056055SA, SBC06281151SA, SBC10326661SA, SBC11067686SA, SBC09365633SA, SBC09195264SA, SBC07162212SA, SBC09075049SA, SBC09285510SA, SBC10457029SA, SJV12500641SA, SBC06281156SA, SBC11268368SA, SJV12240088SA, SJV12430535SA, SJV12240077SA, SJV12300248SA, SJV12220014SA, SJV12240065SA, SJV12240086SA, SJV12280190SA, SJV13301387SA, SJV13431687SA, SJV13431724SA, SJV14473132SA, SJV13281306SA, SJV14493182SA, SJV15133338SA, SJV14132282SA, SJV12440553SA, SJV13050825SA, SJV14292631SA, SJV14292632SA, SJV15083182SA, SJV12240080SA, SJV13211156SA, SJV14132289SA, SJV15133347SA, SBC11238337SA, SBC11308467SA, SJV12300245SA, SJV14022013SA, SJV14032039SA, SJV14132288SA, SJV14463100SA, SJV15093230SA, SJV15093239SA, SJV14473112SA, SJV12350365SA, SJV15053140SA, SJV13471840SA, SBC07061908SA, SBC06281161SA, SJV12500643SA, SBC06080583SA, SBC11027457SA, SBC06281155SA, SBC10326664SA, SBC07252509SA, SBC08254052SA, SBC08334240SA, SJV13060843SA, SBC07362791SA, SBC10507217SA, SBC11067687SA, SBC11067717SA, SJV12460602SA, SBC06170838SA, SBC05480155SA, SBC08043390SA, SBC09305555SA, SBC07252507SA, SBC07382866SA, SBC08053434SA, SBC08233934SA, SBC08233946SA, SBC08233950SA, SBC08294133SA, SBC08494810SA, SBC09054956SA, SBC09375656SA, SBC11468772SA, SBC10416911SA, SBC08304168SA, SBC07473108SA, SBC08193805SA, SBC08304161SA, SBC09085125SA, SBC09195268SA, SBC10457030SA, SBC11468789SA, SBC07302633SA, SBC08264083SA, SBC08304150SA, SBC05480154SA, SBC10036019SA, SJV12500637SA, SJV13060834SA, SBC07011819SA, SBC07182269SA, SBC08314182SA, SBC09195265SA, SBC10497191SA, SBC12099159SA, SBC06461649SA, SBC11308473SA, SBC08354264SA, SBC08474719SA, SBC07362792SA, SBC07322692SA, SBC08233938SA, SJV14473103SA, SBC08364337SA, SBC09185240SA, SBC06291185SA, SBC08233932SA, SBC06060519SA, SBC08314187SA, SBC06421523SA, SBC06451639SA, SBC06471685SA, SBC06481716SA, SBC07081994SA, SBC07082005SA, SBC07142186SA, SBC07172250SA, SBC07212364SA, SBC07252496SA, SBC07252518SA, SBC07252527SA, SBC07262553SA, SBC07292610SA, SBC07342754SA, SBC07342758SA, SBC07352761SA, SBC07432984SA, SBC07493216SA, SBC08023363SA, SBC08023364SA, SBC08033374SA, SBC08053417SA, SBC08113627SA, SBC08113630SA, SBC08113640SA, SBC08133647SA, SBC08364305SA, SBC08374383SA, SBC08374415SA, SBC08404492SA, SBC08484785SA, SBC08484791SA, SBC08494809SA, SBC09185232SA, SBC09255431SA, SBC09255456SA, SBC09265477SA, SBC09265488SA, SBC09415733SA, SBC09415734SA, SBC10025977SA, SBC10025984SA, SBC10036011SA, SBC10036028SA, SBC10096109SA, SBC10106186SA, SBC10126223SA, SBC10336698SA, SBC10336756SA, SBC10406893SA, SBC10447024SA, SBC10507245SA, SBC10507250SA, SBC11017375SA, SBC11027419SA, SBC11037480SA, SBC11047587SA, SBC11047588SA, SBC11047588SABOP, SBC11057641SA, SBC11067697SA, SBC11067714SA, SBC11087816SA, SBC11087826SA, SBC11117978SA, SBC11128046SA, SBC11138109SA, SBC11418627SA, SBC12038982SA, SBC12038986SA, SBC12038988SA, SBC12069061SA, SBC12139232SA, SBC12139233SA, SJV12270147SA, SJV12270148SA, SJV12280173SA, SJV12290194SA, SJV12350359SA, SJV12450577SA, SJV12450579SA, SJV12450580SA, SJV12460583SA, SJV12490622SA, SJV13050813SA, SJV13050828SA, SJV13070870SA, SJV13100934SA, SJV13140989SA, SJV13151035SA, SJV13191099SA, SJV13231176SA, SJV13231180SA, SJV13381590SA, SJV13431697SA, SJV13431723SA, SJV13431727SA, SJV13461790SA, SJV13461795SA, SJV13461800SA, SJV13471846SA, SJV13481854SA, SJV13481855SA, SJV13481868SA, SJV13522006SA, SJV14032047SA, SJV14082179SA, SJV14092197SA, SJV14232478SA, SJV14282619SA, SJV14292629SA, SJV14372806SA, SJV14372808SA, SJV14372809SA, SJV14382846SA, SJV14453073SA, SJV14473120SA, SJV14513257SA, SJV15123292SA, SJV15123299SA, SJV15123310SA, SBC08093518SA, SJV12460588SA, SBC05500335SA, SBC06120686SA, SBC05510383SA, SBC06080602SA, SBC11077731SA, 479204, 571874, 569443, 449303, SBC11518962SA, SBC08053408SA, SBC06331306SA, SBC08334255SA, SBC10416937SA, SBC11168151SA, SBC05500338SA, SBC08133680SA, SBC11037546SA, SBC08093558SA, SBC08444595SA, SBC09185237SA, SBC11428656SA, SJV13501903SA, SJV14382833SA, SJV14382872SA, SJV14382875SA, SBC05510354SA, SBC05470118SA, SBC05480190SA, SBC06040467SA, SBC06241086SA, SBC06441587SA, SBC06451641SA, SBC06471667SA, SBC07071942SA, SBC07092046SA, SBC07112085SA, SBC07132150SA, SBC07503287SA, SBC08103571SA, SBC08103573SA, SBC08103574SA, SBC08364340SA, SBC08374366SA, SBC08394483SA, SBC08444612SA, SBC08464655SA, SBC08464677SA, SBC09175204SA, SBC09445741SA, SBC10166327SA, SBC10416915SA, SBC10416930SA, SBC10416938SA, SBC10426948SA, SBC10426952SA, SBC10426958SA, SBC10436968SA, SBC10507211SA, SBC10507212SA, SBC10507232SA, SBC11208220SA, SBC11208230SA, SBC11388582SA, SBC11398592SA, SBC11428649SA, SBC11428661SA, SBC11428662SA, SJV13070887SA, SJV13080872SA, SJV13080892SA, SJV13100945SA, SJV13501920SA, SJV13511957SA, SJV13511965SA, SJV13511989SA, SJV14362733SA, SJV14382838SA, SJV14382869SA, SJV14382876SA, SJV14382877SA, SJV14382878SA, SJV14382879SA, SJV14453063SA, SJV14483142SA, SK213030050SA, SK213400262SA, SM915023060SA, SBC10507202SA, SJV13080870SA, SJV13511987SA, SBC09275491SA, SBC10396851SA, SBC11468784SA, SJV14402904SA, SJV15173467SA, SJV15183485SA, SJV13361499SA, SBC05500307SA, SBC06231024SA, SBC06511809SA, SBC07182287SA, SBC07222409SA, SBC07222411SA, SBC07322702SA, SBC07362809SA, SBC07483157SA, SBC07483160SA, SBC07493210SA, SBC07503256SA, SBC07503293SA, SBC08023358SA, SBC08143701SA, SBC08143703SA, SBC08143709SA, SBC08264064SA, SBC08264072SA, SBC08264097SA, SBC08284112SA, SBC08294114SA, SBC08294119SA, SBC08294121SA, SBC08304177SA, SBC08324201SA, SBC08504862SA, SBC10266540SA, SBC10276550SA, SJV13020719SA, SJV13020720SA, SJV14042076SA, SJV14042077SA, SJV15063149SA, SBC06431568SA, SBC07262557SA, SBC08354281SA, SBC10336723SA, SBC11107883SA, SJV12520713SA, SJV14372776SA, SJV15023032SA, SBC08374407SA, SBC05500267SA, SBC06060521SA. |
Recalling Firm/ Manufacturer |
GE Medical Systems Ultrasound & Primary Care Diagnostics, LL 9900 W Innovation Dr Milwaukee WI 53226-4856
|
For Additional Information Contact |
414-721-4548
|
Manufacturer Reason for Recall |
A performance issue with customer owned spare parts, T2100 Microflex drive (2026182-002 or 2026182-004), was not addressed with a previous safety correction. If these parts were installed from customer owned stock on the T2100 Treadmill, uncontrolled walking belt motion during a stress exercise test could occur.
|
FDA Determined Cause 2 |
Device Design |
Action |
GE Healthcare sent an Urgent Medical Device Correction letter dated March 30, 2018 to customers. The letter identified the affected product, problem and actions to be taken. The letter stated that GE Healthcare will correct all affected products at no cost. Customers were asked to complete and return the attached "Customer Response" form via email to DCAR.30085@ge.com and a GE Healthcare representative will contact you to arrange for a correction if necessary. For questions contact GE Healthcare Service at 1-800-437-1171 or your local service Representative. |
Quantity in Commerce |
3,721 devices total |
Distribution |
Worldwide Distribution - US Nationwide in the states of AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI.
International distribution to Kenya DEMOCRATIC REPUBLIC OF CONGO, ALGERIA, ARGENTINA, Australia, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGIUM, BHUTAN, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, BURMA(MYANMAR), CAMBODIA, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEMOCRATIC REPUBLIC OF CONGO, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, GEORGIA, GERMANY, GREECE, GUAM, Hong Kong, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JORDAN, AZAKHSTAN, KENYA, KOREA, KUWAIT, LEBANON, LIBYA, MACEDONIA, ALAYSIA, MAURITIUS, MEXICO, MOLDOVA, MOROCCO, NAMIBIA, NEPAL, NEW CALEDONIA, NEW ZEALAND, Nigeria, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PUERTO RICO, QATAR, REUNION, ROMANIA, RUSSIA, SERBIA--MONTENEGRO, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SUDAN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UGANDA, UKRAINE, UNITED ARAB EMIRATES, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, Virgin Islands, YEMEN. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|