Class 2 Device Recall Discovery IGS 740 Interventional Fluoroscopic Xray Systems

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Discovery IGS 740 Interventional Fluoroscopic Xray Systemssee related information
Date Initiated by FirmMarch 29, 2018
Date PostedOctober 05, 2018
Recall Status1 Terminated 3 on June 15, 2021
Recall NumberZ-2388-2018
Recall Event ID 80177
510(K)NumberK133278 
Product Classification Interventional fluoroscopic x-ray system - Product Code OWB
ProductDiscovery IGS 740 Interventional Fluoroscopic X-ray Systems
Code Information Model Number 5730891- Lot Number (System ID): 00000D4-17-029 (478765NSP), 00000D4-17-031 (712279DIS), 00000D4-17-030 (716845DISC), 00000D4-18-003 (951TVHDISC740), 00000D4-18-001 (352751DISC), 00000D4-17-032 (SA1011VA02);   Model Number 5759342 - 00000D4-17-028 (GON 4610398).
FEI Number 1000284313
Recalling Firm/
Manufacturer
GE Medical Systems, SCS
283 rue de la Miniere
Buc France
Manufacturer Reason
for Recall
The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.
FDA Determined
Cause 2
Process control
ActionGE Healthcare notified customers on approximately 03/29/2018, via an URGENT MEDICAL DEVICE CORRECTION letter. Safety instructions included that before each use, to ensure the IGS system is fully functional as indicated in the GE product labeling and follow each facility's established procedures before and during each use to manage the patient if a repeated X-ray abort error occurs during the procedure or in the case a complete loss of fluoroscopic viewing capability may occur. If the problem persists, customers are instructed to contact their local GE Healthcare Service Representative. Additionally, customers are instructed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or the local Service Representative.
Quantity in Commerce102 devices total
DistributionUS distribution: AR, CA, CA, FL, GA, IA, IL, NC, NH, NJ, NM, NY, PA, SC, TN, A,WI, WV. International Distribution: Argentina, Brazil, China, Colombia, Egypt, France, India, Indonesia, Iran, Italy, Japan, Kuwait, Lithuania, Myanmar, Palestine, Poland, Russia, Saudi Arabia, Singapore, Spain, Tunisia, Turkey, United Kingdom, Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = OWB
-
-