| | Class 2 Device Recall QuickFlap |  |
| Date Initiated by Firm | June 22, 2018 |
| Create Date | August 29, 2018 |
| Recall Status1 |
Terminated 3 on July 08, 2019 |
| Recall Number | Z-2910-2018 |
| Recall Event ID |
80371 |
| 510(K)Number | K120352 |
| Product Classification |
Plate, cranioplasty, preformed, alterable - Product Code GWO
|
| Product | AXS PP, 3 X 2-HOLE DOG BONE PLATES WITH 1.5X4MM SELF DRILLING SCREWS, Part Number 12-01540S, for use on the cranium |
| Code Information |
1000201514, 1000203740, 1000216825, 1000220166, 1000233195, 1000237514, 1000238415, 1000245178, 1000252563, 1000254189, 1000254403, 1000255709, 1000256557, 1000257448, 1000258647, 1000258648, 1000258649, 1000259406, 1000259407, 1000260281, 1000260282, 1000261023, 1000261024, 1000261687, 1000261688, 1000262515, 1000262516, 1000262517, 1000263497, 1000264403, 1000264404, 1000264405, 1000265422, 1000268571, 1000268572, 1000201514, 1000287394, 1000292115, 1000293285, |
| FEI Number |
3002821504
|
Recalling Firm/ Manufacturer |
Stryker Leibinger GmbH & Co. KG
Botzinger Str. 41
Freiburg Im Breisgau
Germany
|
| For Additional Information Contact | Julie Brown 269-389-3946 |
Manufacturer Reason for Recall | After a three year term, there is a potential for the external packaging (sterile barrier) to become compromised. |
FDA Determined Cause 2 | Package design/selection |
| Action | URGENT MEDICAL DEVICE RECALL letters dated 6/22/18 were sent to customers with the following requested actions to be taken:
1. Please inform users of this Medical Device Removal and pass this notice to all appropriate
individuals within your organization.
2. Return all affected products available at your location to:
Stryker CMF
Attn: Recall Coordinator
6300 Sprinkle Rd
Portage, MI 49002
REF: PFA #1739180
Or, Contact Stryker Customer Service and refer to PFA #1739180 with any questions about returning the product to us.
3. Complete and sign the enclosed Business Reply Form and fax a copy to: (877) 648-7114 or email a copy to CMF-custserv@stryker.com.
4. Keep a copy of the completed and executed Customer response Form for your records. |
| Quantity in Commerce | 22,831 total |
| Distribution | US distribution in the states: AZ, FL, NH, and OH, Australia, Canada, France, New Zealand, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = GWO
|
|
|
|