| | Class 2 Device Recall Improve Blood Collection Sets |  |
| Date Initiated by Firm | October 19, 2016 |
| Create Date | August 12, 2018 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2790-2018 |
| Recall Event ID |
80519 |
| 510(K)Number | K123987 |
| Product Classification |
Needle, hypodermic, single lumen - Product Code FMI
|
| Product | Improve Blood Collection Sets with Multiple Sample Luer Adapter - Sterile/EO
Product Usage:
The Blood Collection Set are single-use, sterile, winged needle bonded to a flexible tubing with luer adapter. The needle consists of (1) a winged needle, (2) a flexible tubing, (3) a luer adaptor. The products and evacuated blood collection tube/ syringe are used together for the collection of venous blood. |
| Code Information |
Serial Number 102030132 Lots #'s: 20150210 and 20150410 |
| FEI Number |
3008449424
|
Recalling Firm/ Manufacturer |
Guangzhou Improve Medical Instruments Co., Ltd. NO. 102 Kaiyuan Avenue Science City Guangzhou China
|
Manufacturer Reason for Recall | Complaints about the needle is not smooth, with discoloration, a kind of gel agglomeration on the needle tip. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Several e-mails were sent between the manufacturer to the US distributor in 2016 |
| Quantity in Commerce | 700,000 |
| Distribution | Puerto Rico |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = FMI
|
|
|
|