| Date Initiated by Firm | July 12, 2018 |
| Create Date | August 25, 2018 |
| Recall Status1 |
Terminated 3 on July 02, 2020 |
| Recall Number | Z-2935-2018 |
| Recall Event ID |
80581 |
| 510(K)Number | K842416 K983511 |
| Product Classification |
Catheter, hemodialysis, non-implanted - Product Code MPB
|
| Product | Duo-Flow IJ Full Tray; MCDLT114IJ
MCDLT116IJ
MCDLT118IJ |
| Code Information |
MCBA810 MCBS660 MCAX580 MCBH13D MCBKOlO MCBN350 MCBY390 MCCV240 MCDA860 MCDG370 MCDJ730 MCDP710 MCFA850 MCFB360 MCFD630 MCFM700 MCFS710 MCFX790 MKAD790 MKAG470 MKAT090 MKAX630 MCBA820 MCBYS20 MCCP980 MCFD640 MCFQ350 MKAR100 MCAC610 MCAT590 MCCH750 MCDQ920 MCFV690 MCAH750 MCFZ550 |
| FEI Number |
2518902
|
Recalling Firm/ Manufacturer |
Medical Components, Inc dba MedComp 1499 Delp Dr Harleysville PA 19438-2936
|
| For Additional Information Contact | Customer Service 215-256-4201 |
Manufacturer Reason for Recall | The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death. |
FDA Determined Cause 2 | Process control |
| Action | On July 12, 2018, the firm sent a Product Recall letter to its customers. Customers were provided with a full list of product codes and lot numbers and informed that those items had been recalled due to a potential for the suture wings to crack or break.
Customers were informed that implanted catheters with intact suture wings do not need to be removed. However, an alternative method of catheter securement should be initiated for all implanted catheters, due to the potential for the suture wings to crack or break.
Customers were instructed to return all unused recalled product to Medcomp. If product was further distributed, customers were asked to notify their sub-accounts.
Contact your customer service representative for a Returned Goods Authorization (RGA)
number if necessary. The phone number to call is 215-256-4201. Customers are asked to complete and return a customer response form included with the recall letter by fax (215-256-9191) or email (complaints@medcompnet.com). |
| Distribution | Nationwide distribution; worldwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MPB
|