• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall American Access; Nikkiso; Lifeline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall American Access; Nikkiso; Lifelinesee related information
Date Initiated by FirmJuly 12, 2018
Create DateAugust 25, 2018
Recall Status1 Terminated 3 on July 02, 2020
Recall NumberZ-2938-2018
Recall Event ID 80581
510(K)NumberK842416 K983511 
Product Classification Catheter, hemodialysis, non-implanted - Product Code MPB
ProductDuo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C T118M T118ME T119M
Code Information MHJV260 MHMJ380 MHRC840 MHXF770 MJAZ910 MJDV230 MJLF440 MLCT530 MCAH650 MCAQ240 MCAW220 MCBC440 MCCJ920 MCCP260 MCDC910 MCDG320 MCDL590 MCDW940 MCFA540 MCFG000 MCFM820 MCFR060 MCFV060 MKAA520 MKAL360 MKAN510 MKAQ280 MKAZ530 MCAA590 MCAD760 MCAK990 MCBM860 MCBQ500 MCCA150 MCCG660 MCDD810 MCDG250 MCDJ980 MCDP910 MCDV760 MKAD560 MKAL370 MKBC340  MBZS180 MCAH050 MCAL000 MCBM870 MCBT030 MCCG580 MCCP590 MCDD570 MCDR280 MCDX060 MCFQ250 MCFS300 MCFT650 MCFX880 MKAB600 MKAM870  MBZY670 MCCT820 MCFQ150 MKBB070  MJLP070 MLDH060  MHLM190 MHRX850 MHSY930 MHYG710 MJCV440  MBYR870 MBZM650 MCAH670 MCAJ630 MCAQ250 MCAQ870 MCBA060 MCBD890 MCBX380 MCDF830 MCDL020 MCDR290 MCDX070 MCFA570 MCFD860 MCFH790 MKAF320 MKAG530 MKAK390 MKAL400 MKAM880 MKBA670  MCCT830  MHLJ070 MHRD620 MHRX860 MJKJ070  MCBC830 MCFK400  
FEI Number 2518902
Recalling Firm/
Manufacturer
Medical Components, Inc dba MedComp
1499 Delp Dr
Harleysville PA 19438-2936
For Additional Information ContactCustomer Service
215-256-4201
Manufacturer Reason
for Recall
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
FDA Determined
Cause 2
Process control
ActionOn July 12, 2018, the firm sent a Product Recall letter to its customers. Customers were provided with a full list of product codes and lot numbers and informed that those items had been recalled due to a potential for the suture wings to crack or break. Customers were informed that implanted catheters with intact suture wings do not need to be removed. However, an alternative method of catheter securement should be initiated for all implanted catheters, due to the potential for the suture wings to crack or break. Customers were instructed to return all unused recalled product to Medcomp. If product was further distributed, customers were asked to notify their sub-accounts. Contact your customer service representative for a Returned Goods Authorization (RGA) number if necessary. The phone number to call is 215-256-4201. Customers are asked to complete and return a customer response form included with the recall letter by fax (215-256-9191) or email (complaints@medcompnet.com).
DistributionNationwide distribution; worldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MPB
-
-