Class 2 Device Recall Prestige II, Prestige SI, Prestige VH |
|
Date Initiated by Firm |
March 29, 2019 |
Date Posted |
June 10, 2019 |
Recall Status1 |
Terminated 3 on April 15, 2021 |
Recall Number |
Z-1751-2019 |
Recall Event ID |
82668 |
510(K)Number |
K943805
|
Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code JAA
|
Product |
Prestige II, Prestige SI, Prestige VH
Prestige system is a new multi-purpose conventional and digital R/F remote controlled system. |
Code Information |
Prestige II 030133RX32; 080053RX11; 080083RX17; 082408010027; 082408030060A; 082408040021; 082408040036; 082408040037; 082408040044; 082408040045; 082408040047; 082408040048; 082408060005; 082408080006; 082408100070; 082408120030; 082408120032; 082408120033; 082408120037; 082408160055; 082408160056; 082408190006; 082408210030; 082408270027; 0850080431; 0850080628; 0850080701; 0853080153; 0859080008; 209524GR2; 209577PRES; 24646PRESTIGE; 306766BRF2; 3347934R4; 3641PRESTIGE; 388008783XRMET; 38808761XRMET; 38808793XRAND; 38808793XRAND; 38842588XRMET; 38875194XRMET; 38875194XROCC; 3888773XRMET; 38894987XRCTR; 4813MPH; 48660; 519767REM; 519843REM; 540332RM7XXXXX; 60001PRESHM; 600126RX02; 705567VH; 705848REM; 757827RM6; 773854PREST; 84917; 84929; 89763; 902863AREM2; 905871REM; A5371104; A5753801; B5363103; BG4029RX03; BR48660; CS1019RX01; CS1156RX04; DZ1061RX01; EG1251RX01; HC5651XR04; M1828910; M4144612; M4163020; M4198029; M4484605; M5505607; M6042402; PL1507RX01; PPR31321; RU1152RX01; RU1884RX01; SA1231RX03; and X6004701. Prestige SI 2473; 010028RX09; 030024RX13; 030057RX09; 030115RX05; 030309RX08; 030309RX09; 030392RX04; 030488RX01; 040158RX02; 0530863PRE; 06004FLU03; 080100RX04; 080100RX05; 082408050009; 0850080345; 0853080139; 100800RX01; 110018RX05; 160015RX01; 190800RX01; 203573RAD7; 306359ARF1; 38821595XRMET; 38824949XRSOR; 38839875XRSOR; 412343F1; 412343F2; 504380LP2; 506357RF1; 506648RM12A; 514263PRESTIGE; 558039PRESTXT; 600036RX01; 615791W2; 707546XRAY2; 722015002; 808599PRESSI; 902428CRM3349A; 905895RM2; 909580PRESTIGE; 920051; 960025RX01; A1082408; A1111504; A1248010; A1265010; A1300409; A1350106; A1420001; A1532607; A1558706; A1586508; A1586513; A1599005; A1662507; A1767009; A1795401; A1898509; A1962801; A2032101; A2035102; A2072101; A2759005; A2788801; A5109204; A5131828; A5143004; A5145912; A5155124; A5156906; A5186206; A5196319; A5197202; A5216505; A5221505; A5238508; A5269712; A5270003; A5290803; A5323803; A5324803; A5330901; A5354602; A5356913; A5356917; A5358101; A5389206; A5395703; A5396001; A5397405; A5400801; A5400902; A5401303; A5419502; A5637603; A5674401; A5727602; A5807111; A5811711; A5811808; A5814112; A5814401; A6007701; A6014906; A6022104; A6035911; A6038301; A6039707; A6047607; A9884911; A9884912; AP3408XR07; AP4502XR06; B1098609; B1111504; B1222409; B1265010; B1303706; B1485001; B1500413; B1852311; B1989906; B2405502; B2485208; B5115215; B5119102; B5125307; B5127223; B5144909; B5174302; B5174303; B5174304; B5346201; B5374209; B5810209; B6047606; B98689601; BA4001RX01; BA4014RX01; C1074204; C1265006; C1282404; C1607006; C1737601; C5119102; C5129901; C9862608; CE268642; CI1040RX01; CY1010RX01; D1502404; D1598808; D1648601; D4185515; D4185521; DEF99605; DP418833; DPN25102; DPN28001; DPN29001; DZ1014RX01; DZ1024RX01; E1680813; E5821402; EG1070RX01; EG1101RX01; F1669601; FEG60900; FPG22203; GP178304; H13061RF06; H2053RF07; H2197RF05; H2197RF06; J9546503; K1680801; LB1064RX03; LB1065RX01; M1011301; M1013607; M1022102; M1025510; M1033510; M1039405; M1041605; M1074204; M1076846; M1096508; M1098609; M1111504; M1118409; M1137308; M1143415; M1143910; M1192102; M1198105; M1214310; M1225208; M1276501; M1329202; M1388607; M1413612; M1455828; M1469807; M1478607; M1480502; M1522002; M1528409; M1532306; M1558705; M1596410; M1599216; M1625914; M1631806; M1634011; M1653809; M1662008; M1700914; M1711501; M1750209; M1760309; M177858605; M1790301; M1803405; M1806510; M1843404; M1890305; M1962812; M2004601; M2005916; M2050503; M2148216; M2184403; M220648601; M2228001; M224352303; M2324822; M2326003; M2521008; M2565202; M2569117; M2620201; M2624211; M2634202; M2635601; M2718125; M2844616; M2845211; M2845606; M2845611; M2845701; M2864108; M2978303; M4012002; M4020525; M4031057; M4035208; M4041509; M4047511; M4160103; M4172407; M4188810; M4190721; M4209911; M4265604; M4400706; M4460813; M4480505; M4481305; M4485913; M4491604; M4494703; M4548517; M4548519; M470416402; M4859008; M5480805; M5485805; M6001908; M6005406; M6006101; M6012910; M6021006; M6023801; M6024901; M6035913; M6039701; M6042507; M6046306; M6058808; M62631501; M66000404; M9171010; M9792805; MA1219RX01; NC1003RX01; PER09204; PF1017RX02; PL1113RX01; PPR11604; PPR15304; PPR40803; PPR64613; RU1021RX04; RU1032RX02; RU1076RX01; RU1332RX01; RU1591RX02; RU2296RX01; RU2380RX01; RU2417RX01; RU2433RX01; RU2450RX01; RU2460RX01; RU2461RX01; RU2498RX01; RU2592RX01; RU2658RX01; TN1007RX02; TN1009RX02; TN1027RX01; TN1045RX01; TN1055RX02; TN1063RX01; TN1064RX01; TN1072RX01; TN1113RX01; TN1192XR01; TN1211RX01; TN1241RX01; TN1242RX01; TN2001RX02; UK5025RX02; X103911801; X116483805; X128275410; X382885701; X626315001; YP158307; and YP607003. Prestige VH 1773; 00107FLU02; 00336FLU02; 00532FLU03; 010028RX11; 030133RX28; 030137RX08; 030220RX10;080076RX01; 080100RX13; 0853080148; 0853080152; 160012RX03; 190282RX11; 190905RX01; 306359BRF1; 30869; 318473RD3; 38897739XRMET; 418659PRESTIGE; 421PRESTIGE; 536692PRESTIGE; 55085PRESTIGE; 600001RX01; 724284R3; 75648PREST; 808599PRESVH; 819523PRESTIGE; 819879PRESTIGE; 852626PRESTIGE; 860646RF3; 902496REM; 902567PRESTIGE; A1041408; A1279206; A1737608; A1969506; A4463304; A5145724; A5203304; A5216811; A5216812; A5216813; A5216814; A5389610; A5401508; A5414301; A5416101; A5684502; A5811413; A6014902; A6033503; A6064710; A9190111; A9222217; B1039601; B1081908; B1680807; B1720210; B4195715; B5111113; B5266904; B5806715; B5811414; B9185602; B9185603; BG4557RX02; C1248003; C1790301; C1929402; CE673215; CE755A08; D4185519; D5195608; D9956808; DE641212; DE669613; DEG27403; DPM88102; E1484007; EG1039RX04; EG1060RX02; EG1205RX01; EG1214RX01; F5119108; FE416035; FE443837; FEC28A09; FP182A08; FPF64404; FPG26601; FPG30509; FPG44301; FPG58804; G5119101; GP1046RX06; H2036RF14; H2077VAS06; H4858958; K4858901; LB1045RX01; M1013504; M1013608; M1152607; M1185409; M1185412; M1282411; M1622804; M1658910; M1680808; M1796317; M1974117; M1997006; M2742217; M2755002; M2802745; M2802755; M4030501; M4031082; M4159445; M4162211; M4162718; M4175810; M4189606; M4191021; M4192006; M4195701; M4196906; M4205406; M4207303; M4209920; M4210233; M4210424; M4210507; M4398206; M4401013; M4464618; M4474709; M4476119; M4477112; M4482788; M4495709; M4500708; M4501203; M4515907; M5460623; M5484813; M6016708; M6018604; M6049211; M6053304; M9865703; ME244360; ME246208; ME667519; ME799809; MP239201; N4192923; NE655313; NO1108RX01; NPV20503; P4172815; P4172816; P4200408; PER51102; PL1096RX01; PL1636RX01; R4211467; R4211473; RU1032RX05; RU1318RX02; RU1457RX01; RU1467RX01; RU1704RX02; RU1944RX01; RU2154RX01; SA1049RX05; SA1050RX05; SA1159RX02; TN1015RX01; TN1118RX01; TN1122RX01; UK1034RX03; UP644104; X62005701; XE0007; XE0018; XE0019; XE0021; XE0022; and XE0024. |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
GE Healthcare Service 800-437-1171
|
Manufacturer Reason for Recall |
There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
|
FDA Determined Cause 2 |
Device Design |
Action |
On March 29, 2019, GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated March 29, 2019 to its consignees/customers. The letter described the product, problem and actions to be taken. Customers
were instructed to do the following:
Please follow the precautions below until the monitor feet on your unit are replaced:
-The CRT monitor and suspension should be inspected and used with caution. Make sure the monitors are secured properly to the tray before each use.
-Immediately stop using the system if any looseness of the monitor is noticed and contact your GE service representative.
-Monitor should not be positioned directly above the patient.
-Before service or preventative maintenance of the monitor is attempted, review the service and preventive maintenance procedures.
Also the customer were instructed to complete and return the MEDICAL DEVICE NOTIFICATION ACKNOWLEDGEMENT RESPONSE REQUIRED form by scanning or taking a photo of the completed form e-mailing to: Recall_10927_reply_form@ge.com. F/U visits by GE Healthcare will be made to replace the plastic feet.
If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
Quantity in Commerce |
825 (38 US and 787 OUS) in total |
Distribution |
Worldwide Distribution: US (nationwide) to states of: AL, CA, CT, FL, GA, HI, IL, LA, MD, MI, MN, MO, NH, NY, OK, PA, TN, and VA; and countries of: Algeria, Argentina, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Egypt , France, French Polynesia, Germany, Greece, Guadeloupe, Guyana, Hungary, Italy, Japan, Korea, Lebanon, Mexico, Morocco, New Caledonia, Norway , Poland, Portugal, Russia, Saudi Arabia,Slovakia, Spain, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom, Venezuela, and Vietnam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAA and Original Applicant = GENERAL ELECTRIC CO.
|
|
|
|