| Date Initiated by Firm | June 06, 2019 |
| Create Date | June 28, 2019 |
| Recall Status1 |
Terminated 3 on February 09, 2021 |
| Recall Number | Z-1895-2019 |
| Recall Event ID |
83082 |
| 510(K)Number | K181407 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Siemens Artis zee ceiling -Interventional Fluoroscopic X-Ray System
Model Number: 10094137 |
| Code Information |
Serial Numbers: 147208 146637 146395 146728 146942 146697 146663 146731 146945 146774 146784 146866 147186 146867 146692 146869 146759 146776 146775 146797 146871 146801 146857 146803 146886 146758 146806 146851 146929 147157 146971 146919 146969 146973 146925 147695 146948 146964 147046 146861 147123 146883 147216 147866 146888 146955 147214 146921 147931 147100 146928 146946 147030 146983 146899 147089 146975 146963 146986 146960 146898 146937 146918 147047 146977 146993 147035 146982 147075 147084 146991 147114 147023 147233 147049 147036 147093 147009 147500 146996 147029 147174 147766 147113 147241 147083 147087 147124 147120 147119 147172 147692 147693 147775 147165 147167 147303 147079 147203 147107 147168 147144 147237 147240 147236 147239 147243 147234 147190 147175 147515 147559 147250 147254 147253 147252 147251 147649 147612 147557 147507 147516 147585 147528 147573 147264 147620 147558 147526 147604 147646 147581 147608 147849 147583 147553 147601 147659 147644 147701 147666 147635 147625 147905 147660 147842 147577 147587 147621 147670 147700 147662 147718 147737 147887 147676 147717 147783 147709 147364 125111 147734 147753 147893 147763 147956 147958 147957 147373 147955 147749 147760 147761 147821 147889 147847 147833 125003 147773 147800 147838 147816 147862 147921 147795 147926 147928 147925 147940 147880 147841 148274 147938 147891 148282 148242 148240 148239 125230 147894 147909 147882 148327 147896 147937 147892 147919 147910 147982 147923 147936 147954 147934 147952 147983 147992 125008 148375 147974 125010 125105 125106 125109 125100 148001 147977 148010 148230 125112 148022 148020 148235 148220 148029 148278 148025 148030 148285 125239 125208 125219 125212 148254 148318 148335 125115 125214 148352 148341 148326 125228 125227 125222 125210 125250 125226 125245 125254 148348 148316 125243 125240 148333 125244 125235 125232 148409 148407 148349 148384 148404 148382 148400 125253 125249 148423 |
| FEI Number |
2240869
|
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern
PA
19355-1418
|
| For Additional Information Contact | SAME 610-219-4834 |
Manufacturer Reason for Recall | Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel |
FDA Determined Cause 2 | Device Design |
| Action | Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced. |
| Quantity in Commerce | 276 |
| Distribution | Nationwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB
|