| | Class 2 Device Recall Bimetric, Biogroove, Mallory Head modular stem/stem, PAR5 Shell, Reach Revision stem, Recap, Taperloc |  |
| Date Initiated by Firm | August 21, 2019 |
| Create Date | November 22, 2019 |
| Recall Status1 |
Terminated 3 on July 21, 2021 |
| Recall Number | Z-0523-2020 |
| Recall Event ID |
83783 |
| 510(K)Number | K023409 K030055 |
| Product Classification |
Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
|
| Product | Various Hip systems, Item Nos. 104207
104212
104215
11-162809
11-162810
11-162811
11-162812
11-162813
21-103202
21-103203
21-103204
21-103205
21-103206
21-103207
21-103208
21-103209
21-103210
21-103211
21-103212
21-103215
21-104109
21-104110
21-104111
21-104112
21-104113
21-104114
21-104115
21-104116
21-104117
21-104118
21-104119
21-108110
21-108112
21-108115
21-108117
21-108119
21-108130
21-108132
21-108135
21-108137
21-108139
21-108150
21-108152
21-108155
21-108157
21-108159
21-108251
21-108253
21-108255
21-108257
21-108259
21-108271
21-108273
21-108275
21-108277
21-108279
21-108291
21-108293
21-108295
21-108297
21-108299
21-108391
21-108393
21-108395
21-108397
21-108399
21-109256
21-109260
21-109264
21-109268
21-109272
21-123202
21-123203
21-123204
21-123205
21-123206
21-123207
21-123208
21-123209
21-123210
21-123211
21-123212
21-124309
21-124310
21-124311
21-124312
21-124313
21-124314
21-124315
21-124316
21-124317
US157338
US157340
US157342
US157344
US157346
US157348
US157350
US157352
US157354
US157356
US157358
US157359
US157360
X21-180307
X21-180308
X21-180309
X21-180310
X21-180311
X21-180312
X21-180313
X21-180314
X21-180315
X21-180316
X21-180317
X21-180318
X21-180319
X21-180320
X21-180321
X21-182308
X21-182309
X21-182310
X21-182311
X21-182312
X21-182313
X21-182314
X21-182315
X21-182316
X21-182317
X21-182318
X21-182319
X21-182320
X21-182321
Product Usage: Hip arthroplasty. |
| Code Information |
All lots manufactured between January 2008 - May 2019 272750 155610 197670 341720 341750 174760 838240 504840 360250 735990 174770 272100 742290 414230 133510 723450 642570 642520 504830 020840 115890 679780 642620 679870 344370 642510 790050 360240 708170 707970 707980 788870 788880 642490 360270 742270 669070 642580 642600 642610 569050 888950 548290 823580 788860 888990 360260 708180 790060 642540 133500 708160 766460 642550 790070 669060 707960 888980 012620 824370 969280 943520 974890 974950 969500 589060 788850 504810 504850 969490 973990 969450 995270 504820 708190 969310 969290 969300 974940 956050 838230 642500 974020 974900 974910 974920 974930 969430 974010 613610 974000 568870 591100 341710 458340 956060 670530 056780 956450 723460 969470 969460 974030 360280 679800 874570 685800 797060 200790 679790 622410 642560 960010 056820 073810 790100 074730 969370 212250 974960 099550 969400 974970 166470 974040 956070 133490 974880 230500 230520 067820 953720 969320 230490 418630 074720 308850 969480 343870 167200 969350 247310 679860 230510 969380 969410 264610 343860 195100 343910 368670 368680 380570 265470 056810 368650 343880 368660 056790 328870 969420 969360 959990 322750 969340 328460 193940 053890 236450 042260 310460 344700 344710 344720 056800 328880 328890 344740 617850 343890 343900 591050 236220 230470 451080 617830 617860 344680 230480 328860 344730 969260 825660 328780 195110 450590 473130 342020 723440 720460 959980 416500 344690 497210 515120 571200 515070 429200 585850 736360 440880 473110 451110 451130 473120 344750 265490 450580 529670 554540 265480 515080 637650 515100 637660 515110 515130 590260 529660 554530 637020 590250 344770 723430 209290 458310 730320 714270 590240 590230 530160 656950 515060 590220 451100 571150 368900 455980 545990 509860 656070 656080 451090 709940 617930 753280 656060 384800 515090 656940 408210 408220 195180 328800 782090 709970 188790 762200 711280 892200 656960 349850 871200 617910 591040 895180 871220 782410 895190 871210 932270 827790 817060 523500 328790 761210 842780 554520 709920 709930 909450 709950 753290 787220 842820 761220 817080 842790 842810 709960 656970 762180 762170 402730 846050 049910 733510 591030 977710 387280 617870 591760 959970 960000 956460 909760 169330 134540 035790 971910 017760 971930 981440 018610 981430 782100 787210 971900 017770 753300 574430 591090 767070 455990 767090 617880 497500 591020 946740 320690 959960 999520 872780 262420 981410 971920 909770 113200 169320 261470 767060 767110 246670 176030 909750 949930 035780 073390 334900 571190 571160 731070 550160 130550 305150 305160 087910 261480 305180 305140 246680 305190 288220 113210 087580 113220 288150 711290 288230 305210 395000 591010 340880 305290 304430 087570 304520 305200 314010 484190 229420 340810 767080 304470 483130 304510 339610 304530 288210 288200 314020 441810 418000 340900 340910 307920 339590 483140 241730 842720 418010 842710 417930 767100 205390 229430 288280 516860 304540 304550 617890 617920 484210 304450 441820 328190 287330 340820 501760 501780 675820 288240 305240 304420 603910 134550 288170 288180 484200 340860 340870 305230 340890 304480 565630 313760 339600 313770 340830 168210 981420 340840 292030 292050 304460 501750 340850 288290 292020 340930 305280 305300 484220 288260 288300 304440 813340 714480 627180 784530 675810 516850 340920 484230 501740 288160 392260 541060 340800 692370 692380 288190 501790 784540 288250 675830 288270 307930 552760 726820 950900 918490 764050
714470 567380 340790 541070 675840 399990 763450 763460 803170 730160 803180 803190 709440 804200 772070 304490 763470 803210 873810 730170 763480 803200 772080 906470 417990 417970 969620 969330 859770 969340 004230 784750 969640 973600 990000 966450 841980 567390 933830 873790 989350 873800 989360 989370 620960 042250 009020 027920 933420 009010 009030 041410 691120 305520 841970 099140 061940 061950 009980 969630 989340 933380 907310 921790 099130 139260 115500 099160 061960 648500 990010 061930 068700 541260 990020 174590 009960 115510 212040 198420 198430 198440 198450 967260 099150 225620 174070 139250 958900 174030 139240 098690 041400 098680 174020 954670 198470 098700 094480 009970 041390 174570 174610 156510 198460 791630 174010 814170 174040 174580 967200 004090 174560 225610 212030 363130 198480 198490 198500 288540 211720 211730 288160 288170 034780 174600 225270 377570 342830 292300 292320 268300 469130 174050 174060 288550 378140 221810 505730 329130 505780 272290 174080 329120 198510 377380 472690 447070 288190 309070 329140 329150 377400 329160 288700 727130 907320 377580 340100 248560 062470 604880 472680 621810 469170 505300 578410 527330 523050 527390 578420 527350 292310 578320 578330 517410 517470 578340 383690 578390 578400 578430 288930 444990 527340 366270 473570 489190 528440 528450 528460 578360 340250 814000 608250 527380 607530 629230 650160 703430 730150 703450 470980 272880 527370 703370 650120 703400 607520 517480 578350 703440 703480 703380 703390 703410 703420 426260 474380 730020 730120 406170 971570 489180 730180 795570 819600 750350 795560 750380 955570 795580 795590 907650 991760 908570 726180 748750 760800 487560 627460 716280 991800 726170 608690 799250 991790 814080 990420 955200 473580 955590 990450 021440 118360 953040 263520 990440 263540 991810 150600 971520 918850 070970 991750 346160 882730 152400 442800 229130 990430 205190 472380 762930 705530 991780 030430 230760 152370 230770 070980 374710 398570 346650 757540 070960 030420 462070 230780 398560 456720 494860 346150 346170 346180 346190 346210 555030 669140 346220 605710 605720 533040 692730 721280 692710 729690 992930 398580 721270 721300 626330 692720 692740 770270 721290 106130 605680 605690 654100 886330 533030 605700 770280 030380 030390 216220 668980 829850 794550 054950 794560 078860 857480 857490 913260 941280 941290 941300 216800 030650 030680 030370 729700 857470 769850 263530 030670 332520 030660 682310 290490 030640 078850 941270 136850 008980 668970 030630 113700 174420 288430 274630 761010 008990 136840 976300 274610 334960 939860 553450 751530 242690 274600 288500 334950 274620 274640 274650 446410 446420 423780 485390 461100 423790 446440 423800 659870 525480 334970 334920 525470 334930 533820 659970 758880 804180 730590 758870 659980 730740 730760 730770 899320 689920 730730 730780 672200 730620 694290 298760 400780 758940 730610 995570 899310 829300 125440 730600 758900 101610 080100 209990 125430 079770 355740 393290 829340 884140 884150 884160 730790 209730 829310 400790 286960 125450 151940 209760 286970 286510 730720 080080 829320 606150 652200 704320 911500 891820 935310 561810 860990 606140 665110 861000 774140 606110 606120 606130 607630 907310 607640 286980 652190 849550 849650 848110 962100 981500 120070 849620 073420 848090 962070 962110 849630 849640 438720 829330 962080 188080 182320 182330 232220R 232120 232130 208470 208480 232140 208500 232210 182310 208460 232190 232200 208510 265100 334720 304360 391520 182340 182360 208490 232150 232160 232170 232180 391550 391560 166700 391530 182300 182350 232220 391480 391490 391500 391580 391590 425250 429760 423340 434600 304370 423330 423350 423470 480660 429770 407480 407460 404000 407300 407500 484600R 408260 580480 269210 269230 269250 482320 407490 309310 310050 391570 427170 430300 403360 287440 848120 407470 600230 484650 484820 484530 552930 552940 552990 482790 553040 513620 484580 484590 619110 619120 574820 552910 552960 618560 553020 560310 849580 574290 619020 619130 656800 553000 484640 553010 484630 391510 553050 618900 618890 618880 618920 656830 618950 552920 552950 552970 746210 619050 656840 619000 656790 389480 746220 738260 790900 484570 880410 479880 404510 746440 775000 656770 635420 746400 656860 737990 738270 278960 580410 484600 656510 619080 656820 790150 553030 706820 656810 710500 559820 630340 619030 619140 822850 553910 746190 891480 818650 774980 827330 269240 836830 815560 746200 412080 790370 484620 241710 630330 639440 656780 818630 827310 790270 706830 891490 962760 719700 790280 790290 790390 618860 656850 822900 518190 580200 656380 553870 618750 553920 619040 827830 818640 827300 790330 790340 790350 801680 772930 822910 772940 831940 404030 790380 880480 708430 618910 630350 773410 566140 554080 790360 864650 827480 835770 827320 913580 772920 827290 286940 693400 898600 898610 738280 618970 801700 640680 005680 880520 043940 933430 986700 880530 619060 774990 933440 017510 913570 046800 992660 992680 992690 992710 057950 042510 834500 992730 844550 057650 993750 086760 103780 057380 898580 790040 898570 933360 933460 933410 933480 913560 017910 898770 046810 992670 913590 084660 991480 005720 898620 943680 933370 933380 880500 005700 933420 933450 005230 017920 992700 630360 005040 933400 005270 120180 043710 898590 005710 005350 044440 046840 992740 743610 017900 017500 005360 003480 017520 005370 005380 084610 053410 642670 834510 095090 178860 108460 045710 108470 121020 144400 174710 144430 005690 110710 790050 827880 144420 166240 179600 255070 184170 085460 261480 818660 154320 236400 933470 225030 205050 205060 205080 179540 177960 004050 054710 179220 249470 179330 204480 249490 084620 784620 037350 330310 225050 201260 204510 330520 120870 204490 249530 330490 330510 330260 356710 318590 225040 262200 249510 360940 393380 332740 827820 347380 205040 262190 262180 179530 341370 330480 390240 827940 088180 236560 341390 300470 815570 300660 224870 178070 790030 236550 299410 297860 367950 368000 434420 205070 299180 312920 504990 419300 425800 431110 318580 431090 419010 773460 827950 933390 525430 501250 144410 437090 477820 827870 367760 598200 043540 501380 237070 367990 368130 179610 773470 988180 431100 297870 313240 228760 827890 418580 532230 642620 450840 367770 555060 555070 555100 367700 570850 450860 570520 713150 006040 524020 431080 639460 367690 713200 531820 523940 570660 624850 570430 570630 504490 570680 555090 570400 367720 570440 573690 237060 237040 763400 179630 570510 607560 684320 313470 570550 570700 845730 624890 708990 381140 645010 624910 607660 714070 714270 783110 607830 607610 607590 714420 607600 714450 503900 570500 523930 791300 708970 450870 737220 657150 714280 714400 796310 714130 830670 714140 749120 714410 606870 677640 644960 834870 607580 867880 830630 714300 708960 714430 834880 834890 957860 711090 830650 830640 892550 677320 624870 644970 624860 720180 645000 720200 644980 644950 720210 830680 867890 714150 714290 714440 922700 720170 624880 573700 714220 911600 928750 928910 928920 928760 714080 928930 943290 022830 024260 671790 913490 293720 920290 214240 036690 987080 036710 928950 714090 802560 720160 790600 957950 057170 920280 993980 720190 054490 054510 054480 913980 928770 928940 914560 036540 830660 884800 054500 148200 148900 135560 148890 223370 233920 205100 868410 113850 714190 197790 148850 148760 148830 233890 148880 189840 928960 016890 834900 301360 289520 259770 072340 148770 215300 146160 644990 072390 113860 233870 262950 233840 262920 148780 790610 381350 374070 790620 354830 646340 072520 384930 148870 426270 390750 072490 148860 274330 384910 036700 292750 790630 054470 408330 072400 282150 903260 292740 237050 549300 113870 443510 384870 898100 646380 488880 431610 072380 384940 384970 384880 148750 233950 132930 533410 408150 408320 565030 524450 484250 114000 072510 385010 484260 546910 441430 384950 483800 646350 546930 406800 568620 700690 584840 656670 570540 653910 584850 811480 728190 641310 700660 408140 384900 565020 384890 553270 653890 616330 565040 565050 565090 641330 484240 484290 484630 546900 570550 653940 831870 802830 484310 996710 408340 546920 928690 646310 790640 744340 813810 484100 814340 814350 814360 444210 656940 814380 616340 556720 910590 910610 914370 020680 046920 484270 069670 914400 143950 814410 814420 020670 547620 814330 701880 668600 814050 570530 813840 873170 813800 115530 701920 987470 408130 814060 905030 944240 944250 944270 546890 020700 910600 814400 814010 910620 020840 914360 301970 065660 144300 484110 115940 408350 874770 914350 616480 914390 811110 172850 144290 944260 406810 701990 020750 894690 301980 814370 185560 112660 944230 112670 059850 832050 185540 020690 214520 112470 059860 020820 059870 406820 874820 112620 874790 809380 072500 185520 185590 112690 112500 185530 143920 113560 229310 112480 072360 113480 954720 293160 873130 020830 185550 813830 112630 185580 113530 164230 164570 944280 333880 052430 547610 112700 112710 069540 341380 350560 370760 301960 112680 142830 112610 205110 205240 370560 712470 432260 432150 206430 206380 164560 205250 370780 547640 491230 874800 455720 204300 432270 205100 205170 205370 112510 813820 145160 455940 441290 206440 206400 205340 112520 205190 205130 142970 205390 205180 414500 185570 553190 432250 142800 553130 142690 142700 205770 592340 112530 441360 484280 432190 566340 857450 206410 435820 578890 535800 535850 370490 112490 432160 435800 578740 439270 535710 227910 258070 406790 874780 205230 205350 206420 523090 367580 578640 199860 484090 562420 432280 164530 491220 723700 432170 227900 562380 227930 435810 183920 562400 164540 715680 745270 759990 432180 112570 562290 562410 432240 185820 714920 441350 205330 605300 183900 679730 562280 478340 715610 798180 799100 723680 788220 227920 205380 435360 673650 205090 205150 677580 206390 562390 622590 693650 113170 206370 605320 605310 605290 679760 699230 799200 482030 699520 745160 482050 112580 818740 715590 112560 432140 814390 679720 435480 677650 748490 715410 821070 183910 735510 748480 205760 906200 679770 441330 748510 679780 803230 677540 735560 113720 301940 679750 406410 800690 803220 800680 677660 984520 940540 818770 432300 969600 677550 748500 578870 482760 800670 907680 971840 035800 901370 996820 901760 679710 084930 901350 901410 901770 901790 901220 901780 482750 901750 036220 978640 084910 011210 680330 680370 715120 011800 070420 111730 978500 020770 133110 133120 133170 019530 996550 036200 715130 070470 963400 182430 801180 981900 006130 020760 006340 203870 204020 132700 132740 132800 901800 019630 048310 132880 036010 244320 020820 109980 036130 084900 084920 133130 901740 357420 036160 203900 240920 020790 006350 203820 048340 048350 048380 226370 203990 204000 204040 132690 735580 014230 562300 019820 133100 048300 048360 204050 226380 680280 204060 715100 020780 203800 203840 204010 132820 143120 143090 650340 035990 020750 172910 172840 247010 133160 226290 801190 084940 226320 382700 834110 193680 203860 203880 203890 311620 226390 247030 247060 278640 901400 112550 133140 292870 394890 229170 258130 133180 247050 204030 292850 132710 253330 226300 864750 961140 117440 147510 147490 203830 133150 247020 292550 370730 172950 021040 432290 203780 194360 084950 432230 132720 901380 407800 081880 512340 193450 127240 408270 359600 407840 247090 538760 019420 534860 292680 565260 159720 008240 534500 447520 183930 820110 968210 113710 501760 565080 530760 501770 085970 147500 593540 132670 044410 408060 565070 566190 408260 531150 531280 530770 113730 408250 618410 501740 501750 132810 530750 530730 616140 112540 501730 618220 531290 701380 086450 566160 566200 566210 618210 400790 566180 530710 617650 292830 531110 779130 617690 728510 618190 617490 531180 617590 530720 044580 673870 566230 432820 617480 617470 617510 538790 484080 297900 399350 727050 873030 875500 448160 833100 673970 297860 673890 673860 873010 873040 970500 509090 876910 833090 673880 918330 866990 376060 704410 843970 618230 258120 715110 566580 297870 866970 877440 875490 875510 592280 642470 918320 617800 833170 566170 873020 108610 592300 592350 642400 014300 642460 258080 566590 297890 566570 566600 197380 044630 051070 408120 716080 935220 150370 135690 160620 258100 064680 112370 704400 003060 044570 989480 108600 833770 064690 013970 164350 149820 195020 865300 865360 014000 014270 065500 866980 989460 014010 918350 205910 174290 038080 003000 918340 918360 643100 565090 014250 865350 065490 617660 013980 014030 851490 918370 149810 164610 174620 049070 253260 137660 906850 208620 164340 833780 989450 014260 251800 617500 014290 014020 234880 164650 064840 256480 256000 941200 234870 272890 237880 255880 234860 205890 674460 255870 251790 205900 272900 255990 865380 816830 239560 164360 614820 239540 277560 255860 215990 013960 164620 324680 876990 967570 205880 989620 415680 049080 164640 324650 937140 255970 324730 465290 427180 064670 064700 065320 965180 237920 492130 373110 255340 427240 415690 362500 357250 360530 618460 418440 361320 918520 361330 361340 592310 495540 434780 064710 433160 463680 487120 551600 551630 248620 501720 614590 239550 594380 594390 565560 604440 565550 561730 292690 989470 364070 559490 601370 373100 384290 939300 678610 682080 682110 682120 682140 685650 749750 643220 735310 678640 551580 636060 682090 682150 678590 685680 678650 678660 455630 762840 722080 682130 565580 816850 746350 172280 678620 678630 749810 162970 722090 702260 798040 551640 695750 794310 816770 735320 735330 816800 816830 735300 816880 812000 749880 240930 634040 551610 682070 682100 685670 780470 506590 722120 851360 566510 433330 811980 618340 551590 633980 685640 685660 749770 678600 735290 735350 322490 826700 749740 434790 939090 714880 866360 997240 979400 865240 041950 041960 041990 735280 866380 519440 682060 958390 832470 816390 968670 968680 968690 722110 996310 866330 816780 799920 016760 749860 749870 866370 866390 888990 945620 904980 594470 935140 816410 434800 000090 979390 955280 967160 979380 041970 041980 996300 866340 866350 016750 996320 175170 040720 057300 812170 113280 866320 996290 153510 979370 021140 841770 139770 102830 102890 102840 139740 878670 774480 841790 112910 816810 897440 119240 996280 202140 112940 812160 139760 112870 268080 253520 735770 218590 112880 032740 112970 073680 102860 112960 990270 139780 160950 160960 176250 189320 016730 112920 816400 112900 215920 215960 215980 016770 260230 331070 108290 139790 112890 216000 108280 533190 106740 215990 168340 268040 016790 215940 320820 215970 176370 142090 244960 999140 245010 452970 176190 176440 201340 176430 268030 016720 215930 331080 381260 452980 176420 381250 139750 267920 453010 215950 364610 512560 393990 420720 512550 432350 535860 331090 016740 245000 506420 112950 215900 506430 273420 530490 584780 578840 488800 453030 637250 650250 887790 286460 637200 452050 637270 453000 506460 331060 452040 584770 752380 519630 577080 381240 519640 519650 841780 519620 637240 637260 215910 273700 637230 527910 584740 705930 668800 555230 573520 452030 668410 651270 651290 668790 540450 452060 635370 584760 488700 705910 651690 577100 651280 557750 584730 705700 668770 668830 711160 735840 705920 705880 637290 637180 798130 788140 735850 735870 765340 739970 506440 705940 651240 705970 577070 805370 805730 780610 637210 635360 668420 788130 841870 735860 780640 780660 617410 803640 534990 736600 884900 584750 765350 884750 871220 841880 617310 614150 651260 929510 711150 705900 840690 674010 840740 873510 840710 840750 668730 841930 637280 839070 577090 780670 637160 804640 840700 577060 073670 965340 842890 960010 840670 840720 910010 840760 084660 986270 971430 056410 946760 573530 873520 840680 962100 962110 971410 909960 909970 909980 946770 071400 828660 909800 910080 840730 909990 946750 804570 964640 020580 971420 841860 964140 020610 910090 046560 962120 971440 020620 129970 962150 971400 121980 018230 971460 946780 420660 093980 064710 046490 094000 102170 084640 055200 637220 962130 125650 971450 219450 224380 125900 020600 238540 238560 238570 238610 238620 102160 125660 238530 159560 238550 046470 238580 238590 093990 238600 238640 102140 102150 137230 962090 180200 055210 064730 137820 198440 262930 249970 093960 986280 102180 135400 137240 137260 064720 168280 091140 102200 102190 084690 270120 270130 962140 020590 330980 198630 198640 270170 695190 252090 198450 270090 324060 270210 238630 084650 263100 270100 309390 084680 406800 262940 238660 270110 170540 055220 198580 324040 406810 341880 270180 358410 324030 358430 317620 341850 341860 270190 131130 987900 383110 317650 085130 481940 084700 339810 238650 403620 253070 198620 270200 547540 084670 406830 358450 358530 516650 444430 406750 406790 358520 406850 358620 444440 444450 270080 358440 406840 339830 478410 406860 406890 326710 339820 486050 527560 406880 478450 478460 180210 263090 341870 367160 324080 406870 403630 478350 490760 339800 358540 672650 657720 478560 478380 506770 593400 593410 478540 478550 478360 478390 339850 515030 586880 478430 406770 598650 486030 506520 672640 672830 486040 406820 598750 751580 614190 638450 516530 478530 406780 406760 358590 763930 657680 625400 358580 478520 633710 614210 369910 687760 341830 406740 633740 657540 657590 909110 956030 672620 728110 689430 339840 593420 444420 657650 657700 763030 763010 763080 515040 763100 763110 657660 585710 344870 516750 547590 642780 657710 734790 657570 614220 642790 696230 784570 785120 763040 763070 763090 358610 776020 657550 050130 506760 232800 839070 753440 777070 817060 694440 713960 728100 633730 657560 478510 326730 070890 734850 763020 642800 009680 060230 817000 657640 973120 910000 866860 413110 657600 883960 657580 696240 913370 955860 672920 593370 074940 909950 060220 111590 111610 111630 642810 855190 855200 593390 991830 696220 913360 799910 913350 864870 009630 913380 784580 991840 657670 672630 855210 972960 593380 784610 089790 050970 784590 696210 089740 089750 060210 111620 009620 009690 991850 149110 999060 137080 091970 009580 060200 132710 074630 089760 089770 009640 009610 186920 270160 009600 817090 009570 009590 009650 207360 979330 074610 132730 230620 356350 207370 009700 230610 207410 255530 111510 050140 111520 111540 485050 924630 112480 657730 132720 060240 074620 132690 132700 821300 230700 255520 111530 111550 728090 230590 227770 089810 267920 111500 089780 009560 821310 424080 207420 698860 372690 973000 336280 074640 089800 646320R 247430 414390 006950 322740 317580 182800 331140 355950 336220 336230 633720 227780 336270 299930 330830 322750 428020 463320 427980 428040 428070 571160 506400 813650 365070 428060 338630 427960 428000 339000 428030 431360 571250 227790 139060 724260 247440 471060 615180 581450 338640 573780 600790 471090 612860 450100 612850 606670 247470 247420 733590 646320 618780 470450 921890R 305210 618320 617810 859490 246890 612820 818400 646410 746470 617790 646420 247460 571220 571240 612880 863520 571270 761440 761450 931970 600750 746410 612810 427240 571230 775390 000040 471080 247490 999890 201250 201260 133970 450090 826470 954070 999940 646400 573760 813750 450110 912950 746420 752790 728080 110530 884430 999880 999900 612830 921890 931280 600740 073840 931270 931290 761650 761660 427650 925090 954060 024350 106880 133830 571260 612840 931260 161970 746480 065210 090650 884470 646430 851140 884460 884480 099070 190340 612870 843610 746460 967340 616830 775770 190260 617800 167530 889060 133910 830410 050310 884450 106920 168180 883630 106930 999950 999960 999970 073830 931250 035610 073800 090640 951320 883620 134010 133790 746400 021500 156330 106940 073810 863530 134020 167200 201220 206030 106950 140510 122880 122890 166440 134000 044300 999910 133800 133820 133890 134040 141320 201230 133930 413550 501640 394350 446210 133920 205930 156100 330290 789410 330520 019060 990040 174810 134030 133810 133980 485350 141330 657880 330090 388010 610480 201210 133900 141350 307750 249880 330140 537610 161960 307550 307740 954050 205750 307720R 161980 405480 035490 944450 222130 141340 236880 647720 307580 327720 327740 394380 307710 256290 205950 218020 327730 394320 394390 201240 307540 249850 249860 544290 307730 386330 140540 495220 307570 327700 495260 073820 394410 307720 307530 176920 467920 500840 327710 327680 394400 162000 414200 467850 413570 394340 632700 524070 524110 647710 386340 537140 133880 386310 495250 386300 415980 559920 426220 544640 688030 446390 524100 495270 495210 307560 524120 661080 678680 715810 249830 610610 524130 715840 600370 413560 927500 878070 041660 753720 691270 610600 249900 753790 249870 691260 544610 501570 501620 446380 657870 590090 864760 598740 884440 632690 716090 753680 661100 688040 653590 653620 727950 786710 426230 495240 524080 599660 599800 394330 394360 446370 688020 647700 647740 610470 878190 878210 878230 082340 927480 805670 599200 884210 501600 544460 327690 775760 753750 301300 786690 678960 878060 495280 162020 249810 249840 610460 878170 968340 884200 688010 980730 106870 030050 131780 229500 900060 927510 791990 927490 927530 927550 091640 417430 884240 012500 980400 980740 981210 631510 980290 792000 927470 979630 878250 121510 121520 121530 980330 632500 168110 501650 927520 206030 101150 101200 983060 091630 117520 168120 206130 983110 173900 056580 082280 786700 632680 983100 878160 884190 884220 884230 983050 101170 792010 927460 168070 168090 206060 206080 101180 983070 983080 980300 469800 469300 980280 206090 259050 792040 168080 206070 386460 983090 273120 654200R 411300 980720 405490 345000 354200 206120 422920 411310 382320 206050 353470 727120 327220 353480 437680 437690 980270 168100 437660 417390 329100 421950 259040 259060 354120 504450 417410 502720 624960 554130 504470 554100 468700 229400 979920 504430 686370 519740 421940 262500 417350 504460 614510 667330 575000 654360 611050 327210 229430 437950 465170 411290 355390 121500 353460 554140 554170 504480 437700 530210 121540 442120 980390 614500 980310 554150 980750 610980 502710 504490 305320 561680 561810 610950 611020 554120 554180 611030 611040 616080 667280 794940 421930 654200 654180 411320 411360 417380 980370 654170 768400 611000 784840 784860 504500 892310 686150 681210 752460 433760 504400 411390 768410 794960 768450 600390 411340 768340 794730 614490 411410 768420 794670 794720 983120 035520 794760 794770 784830 784850 768430 654160 805080 990020 717350 794710 794740 186970R 876700R 876520 955160 794920 877850 876660 876500 877840 768390 876480 874010 874020 874030 874040 437670 876540 876550 858150 876620 876690 876700 876530 203310 794690 203200 784930 876600 858170 858180 858190 876680 969800 273790 831980 937090 876640 421920 006020 504420 203060 203110 053060 616030 006340 784770 862450 006050 969820 005990 876590 858160 876670 006080 173930 110670 236910 053100 937470 683660 240470 223830 186960 876610 110620 110630 110640 066870 110650 203080 203290 006040 110680 225990 236380 053080 240490 823740 974330 006090 006000 876490 203130 203250 203260 203280 236920 110690 223860 223900 236310 794780 203180 203190 110710 186990 187000 989630 874090 873740 974370 203090 203230 203240 110660 053090 223850 236370 035560 303790 005980 974360 379770 969370 034020 204530 322150 407230 989610 203100 203050 223870 203120 110700 876630 291600 248520 336780 292010 248980 203300 162740 236330 236390 203140 247020 322250 292000 203320 006010 203210 066860 273740 203070 203270 246990 236930 273760 223840 223890 236320 236340 236350 236360 247010 273780 187020 292120 354160 891070 974320 186970 118730 292050 614530 411440 801030 065340 223880 006060 006030 322160 974350 236400 247000 273750 335230 475120R 627780R 388260 435240 053110 823730 423490 273770 186980 187010 408360 336320 273730 408160 511480 423530 460310 408450 439350 447960 379760 236120 273860 408350 368110 408340 203220 379690 379790 475120 475140 388400 491230 408000 511260 459070 447950 459120 053070 484650 069120 614520 335290 459140 989620 006070 654340 504180 626160 459130 616750 484630 206100 511240 387540 336730 335220 627740 407290 627820 273850 081760 596530 697740 627810 627850 475250 627860 596590 627880 335270 627750 627760 276550 387530 387520 123870 697730 627780 627830 627790R 627800R 777880R 928150R 335300 069140 627790 627840 371300 710700 368100 627870 596520 670810 627800 627730 717680 484640 758190 758230 627890 670800 722240 813500 823760 627900 777880 627770 758200 820410 989600 494990 697630 491530 683870 916170 770990 607020 616020 822540 924510 607030 794050 770380 771010 804130 758170 758220 794060 794070 794090 794100 794110 794120 758240 770370 822550 866060 876800 865190 671710 758210 794130 758180 890020 774330 771000 770340 770390 273870 905730 794080 960350 771020 876830 770350 931610 928160 905820 774340 915500 876790 876810 921970 921980 921990 928140 928150 962910 960250 770980 922000 023340 922010 007440 998020 777930 971740 014320 902060 902050 983540 921960 902040 924400 960360 902030 902080 902070 089340 876820 921950 089350 089280 090420 089090 089270 153020 089150 089440 089430 203370 090400 181500 089400 094160 996800 094210 139640 089370 089380 089410 071320 106340 106350 089070 089120 089420 071290 089200 089220 089250 181640 106370 014290 105520 214660 203330 094170 106310 106330 106360 139650 089100 089130 089450 089230 090390 090410 094140 062760 094190 094200 089360 106380 089140 089210 140100 089110 089390 225950 089180 089190 094130 071260 071270 139540 139720 250170 257970 226250 106320 239900 089260 143750 071240 139660 089300 239990 089290 089240 226300 143770 279520 226140 462350 014300 094150 140170 296700 288700 268370 268400 224980 224990 462360 225490 288710 186730 225900 250230 288760 288640 318200 268380 372020 370800 288740 306160 318190 288720 203320 306180 317450 071250 392050 288670 268390 328050 288660 370050 400770 350850 268350 392510 268360 328070 307130 318170 268340 288630 345710 306170 328060 345720 345740 306190 306200 306150 402090 414180 390500 186740 071280 345700 318160 363510 460540 423150 422860 452890 318210 359200 288650 345730 365890 307120 409510 394070 288620 363520 318220 340630 363500 402750 403650 633430 236180 142210 288730 404730 404130 473560 515900 540980 473550 515880 459840 515860 515870 581150 484740 520220 520230 318180 394060 520210 540900 462340 459830 515890 572050 409610 236170 531160 540890 522260 540560 142220 515910 288680 236160 594070 620700 581160 581140 484750 515840 250210 423080 512730 459860 551400 638130 645670 685550 572070 572060 572010 572020 581100 572040 572090 620720 620780 556710 250280 643470 572030 638150 620730 581170 581130 594050 581110 581120 694470 645700 594060 647650 406890 674450 577310 620710 556720 610400 695390 694490 620770 645710 577290 577300 581180 645720 631180 631200 674410 745150 695410 537910 638120 009780 694440 645640 695420 620750 724760 701030 744790 789530 798230 777440 561050 776890 181590 789630 789550 685430 777350 556700 620760 701020 724980 620690 645650 645660 748930 748950 620740 701050 561040 561040R 354354 354684 831654 777340 724740 645630 645680 700980 646455 646847 789650 776630 594040 724770 695400 930020 685630 776440 789640 776960 847790 776900 900890 724750 620790 594030 645620 776840 572080 638140 748940 748960 798200 701000 701040 748990 581190 701060 744720 701060R 776960R 724730 861650 906890 695450 700990 701010 581200 014310 236190 460220 822780 822790 748970 744700 798240 744710 776830 776880 776860 861720 814940 956590 826060 814930 776870 955450 960070 711810 975190 838310 798220 945560 955110 955030 838290 906620 822770 814920 814950 906850 695440 873450 955430 861660 847940 906830 838340 031970 873430 694450 838280 838300 873440 744690 814870 838330 873460 175720 838320 798210 873420 955530 955980 975200 032030 104910 906870 906880 711780 960740 906860 023100 906840 031910 955630 011020 045720 104900 938220 047700 623000 975140 115750 459850 045710 115760 175890 958910 958900 777470 957230 173210 013540 104970 173160 176040 429820 024000 281420 218940 176110 276770 847900 175990 176070 175710 176030 176000 176020 176120 282750 347240 176130 282270 282390 175770 188210 188190 282280 282400 269110 176150 175760 310000 282250 388630 282430 282240 176080 282420 282440 343940 419400 459220 408290 412350 282230 282410 631350 373530 456470 115770 406150 320250 173280 176090 282260 175740 269080 540870 269090 269070 269100 176100 402270 406170 521690 581710 408310 440950 440920 440970 440900 176010 440930 560740 597860 627470 440890 402260 406130 440980 175750 606180 496160 442200 370740 528530 514580 440940 493990 604000 455390 514570 440910 496100 496140 456800 646550 496130 716050 627510 570920 453840 493970 176140 539630 496070 514610 539610 441030 611580 642900 441000 742480 175730 661610 496080 115780 596920 656890 514590 596910 441010 511100 704820 496090 514600 539620 661400 441040 683130 440990 596940 813070 828660 676560 497900 540960 827250 513710 827090 857620 402380 539910 676470 676500 828510 440960 676530 767990 696260 826420 908890 496170 828470 596930 828690 511110 676550 513700 496120 493980 539640 539920 067620 841400 402280 676510 514620 619450 358480 539600 596950 790210 051040 741190 873070 741140 676520 851340 698530 051020 873730 943180 943190 873090 943200 994500 194300 282360 214410 943160 994480 199090 062980 676540 007790 994490 767470 798610 701690 115190 078300 303830 115060 402390 873080 115200 767480 239720 239750 676490 115130 115160 565300 942200 078320 701680 115140 374430 140330 906030 994510 942210 873740 115170 570910 305020 115050 338380 216630 382130 581320 338400 179360 654710 196200 374730 339710 214260 216450 305010 115120 239700 305050 484060 214240 078330 620050 417060 305060 214250 741200 372430 078310 620020 942190 723960 481790 305030 621790 484020 484070 382110 239730 305040 929730 115150 382090 417050 528370 481780 526700 239760 562580 610730 179370 253400 484010 811250 562570 338390 138470 906040 165400 483700 382100 304730 382140 190130 688100 639300 488780 099520 132600 239710 619660 811270 568290 880520 895510 635600 639290 382120 239740 147310 382130 907390 907410 863210 811260 627130 977500 863230 877910 382150 654720 811240 811280 253410 419280 619650 863220 382160 484000 080350 133770 526690 562550 417070 257840 481800 735200 877900 977790 735200R 688050 120670 877920 099530 907480 138460 907370 562560 245850 977760 562590 246440 121210 929690 388290 563680 354780 242570 120660 147340 063590 877990 242580 242600 132650 207230 409920 268620 147320 380530 944830 359720 099540 797970 132610 099550 483710 907400 099510 977770 053380 268890 132630 050480 878000 163730 163740 688110 132640 431590 356390 977780 268750 242590 380050 091470 147350 132660 268910 099560 078340 147330 053360 333570 521820 863360 583760 440310 562930 649470 529820 529810 519440 475470 488600 409870 405540 405470 207220 536810 599050 053390 053370 380040 380070 795610 888170 519390 475410 475440 488590 488620 257380 405460 475380 445050 405530 333590 519450 011100 011120 298430 475450 475420 488700 519430 488630 353430 488570 257370 011130 488690 475460 519420 519400 822340 716040 475390 465740 562850 929320 589960 863350 863380 488560 488720 751960 475370 529790 011110 380030 636290 636300 764780 082410 685280 706430 718560 769430 716070 716030 908500 863340 634780 716060 488710 519460 563910 595550 595590 066070 814410 488610 908490 927140 927160 635530 645020 764790 764800 958600 453760 108560 066060 431740 784880 927150 488580 817440 504810 351520 529800 863370 784890 817430 128050 978840 814420 519470 120940 645200 333580 705420 705420R 082440 049780 345970 645230 131430 634460 814390 178930 099330 645220 178900 178910 331250 178920 645210 228230 380060 178500 370740 705360 228320 346000 423510 389090 389100 327750 228310 142640 327730 877930 607280 389110 379070 424400 510080 504520 506020 696450 661180 454000 584760 779810 422370 167450 694050 519110 252520 645240 595950 492760 252540 595980 596010 654660 475260 049280 595960 492710 596020 142630 903560 142620 142650 252550 624760 814400 423500 595940 595970 691200 596000 691200R 252500 832670 691210 811510 779820 811440 811420 811430 252530 811480 832680 832690 962750 777630 492730 779830 811400 595990 091670 798180 160830 071900 798070 798170 798100 017320 017360 017360R 236490 798090 124850 124890 811490 252510 112440 122320 160780 798080 110760 110780 159130 160710 124840 110760R 962760 110740 160720 160760 798030 159160 160820 983440 171340 160790 171280 171350 170990 122330 110720 110820 297750 171300 160850 171310 110720R 330290 160650 171220 171230 171260 159150 124860 160770 160860 110800 297720 159170 171270 124880 171320 171290 279580 363680 124870 160810 171250 160750 069680 332550 297700 297730 159140 171240 297690 297680 314380 279590 473770 486890 525030 422160 159110 171330 453100 268210 140320 524430 314400 401380 332560 314370 500460 453220 524420 524410 434340 297710 124900 503500 519770 559770 453590 524500 434350 297760 524460 122310 297740 524470 522620 161900 431910 434320 524440 524490 589870 524450 608920 638660 266690 525060 177780 816690 588450 588570 588590 524480 555980 588510 588520 756510 572470 572480 620290 756530 620300 756490 588580 668590 524810 235700 295230 588540 653840 738820 295240 756500 525050 653890 738800 859000 506280 738790 588480 858990 763330 589270 756520 914660 897930 914670 588470 897920 948010 829300 177770 914650 589250 964370 914710 963380 154720 668750 969680 897910 914680 969700 184900 914700 588530 046110 995450 199080 068610 995480 589260 859020 963360 605200 114640 045860 158480 200210 969720 914690 009450 897940 875690 995460 994960 236020 183210 159430 104360 995490 964580 994950 104480 214080 808110 009470 299020 296380 969710 236010 118040 104370 859010 364490 236120 387380 567530 614470 567550 454500 454510 454520 118090 526630 431780 794190 628880 605080 236030 510340 506750 611270 746130 506760 567520 628890 605070 568360 628850 614490 628910 628860 452640 628870 131230 452660 601810 584110 567540 118080 800230 605090 152950 800210 800240 567600 584100 822580 801080 568110 611280 915300 628900 800200 915320 098330 915310 152910 847560 735050 915530 848460 838930 915560 032360 127380 073590 790780 238760 567620 151510 359360 939770 342400 101920 065270 089120 247440 368720 114720 151500 735070 359530 486100 157860 368700 486190 752740 349880 474000 452780 593130 368680 127390 130860 682610 474010 725750 486110 584120 486090 374470 553360 836520 763740 614800 349870 848590 848600 848390 765680 765690 942440 374480 374460 836530 657790 835350 756020 838640 835370 975260 753610 614810 614790 756070 756050 666330 055560 835380 127660 021160 067590 067580 166170 070370 284830 957790 067610 287000 190860 835360 067570 071010 123590 163480 165680 190870 163340 163350 232280 359620 202670 430330 453460 547820 757350 759650 759640 613900 218480 303760 398550 816910 777530 816930 459720 627270 777860 816900 777870 613910 553700 415390 778850 653180 433060 778870 777850 430320 398560 049730 505470 398540 050260 778840 778860 171590 398570 181350 613920 047970 153090 398790 924390 307740 363750 363770 390600 778830 525110 401120 401110 655830 390580 436670 409350 409280 476070 831470 812040 757470 409300 260310 476050 745810 409310 354510 354480 409360 409320 409290 878900 845900 937060 648910 799370 727860 354470 893930 799320 755960 756610 409340 409370 799330 727850 228860 620350 184360 962740 228870 962750 239360 973750 457970 757160 837680 389070 343160 142100 081520 908380 378450 837660 549610 294830 369300 496310 973670 253900 321180 476060 368300 398030 441980 186780 186790 239230 546430 241290 130480 038360 361640 527290 280230 038330 817590 170230 958190 831440 278520 399650 799360 620580 361610 038350 321190 409330 450200 596240 315850 593310 527670 799380 546880 038320 870670 769120 546890 593280 633460 361620 571230 720310 807360 343150 672530 593300 683530 633450 753870 753700 676310 753880 684900 830790 593290 724660 724640 704400 814850 827090 869750 899980 875960 753860 888320 899030 850230 343140 968510 963230 751920 685220 112290 969950 038340 633430 633440 963220 112280 650720 963210 147950 724650 038110 724670 996000 042820 685210 243380 011950 845260 190130 085590 097690 401430 820100 943930 188280 225740 173140 172630 301920 267670 225760 998290 381030 011480 799390 773990 211780 159300 827110 173240 293710 956210 056510 225730 287350 584080 483270 723780 287340 472590 585920 527340 267680 339410 154630 354490 154620 347210 267690 483280 211790 483260 150360 378790 241030 139220 503080 503100 527390 527400 347200 139240 181860 585340 503130 503160 257190 257180 527350 577950 447770 532640 540690 154640 548450 527370 577960 577970 577940 347240 503110 503120 503150 527430 688620 763190 601560 572580 706370 647900 577930 577980 706380 749770 670150 670160 670180 782320 670190 139210 527410 656810 670200 706390 656820 287330 854740 171070 801900 139230 347230 532130 855690 855860 898980 347220 792170 969280 969300 803570 706360 503140 792190 792180 763160 670170 672270 981880 293680 899000 347190 072080 864270 807780 503090 981350 807750 577990 807770 741430 896730 096720 085550 807740 741390 238190 097250 097430 198780 300830 085560 085600 085560R 270210 981340 248390 864260 212210 133120 085580 216050 097590 262790 097580 241710 241730 097570 248410 097610 212220 490140 524000 085590 085590R 248460 622370 607440 248440 440690 557380 622390 212230 622420 557420 470790 470790R 604600 490130 604600R 440680 578620 470800 574630 697320 291390 248400 792200 291080 607570 574640 557410 743130 348810 738840 622690 097600 622710 758790 607590 607450 097560 938680 823770 801680 604590 490110 782560 782560R 782510 700980 490100 801650 801720 935050 085570 868230 868250 949180 968970 988610 823750 943510 782500 291120 801670 801690 782570 868270 988610R 801660 872460 935040 021700 892270 868280 951020 801640 980620 980620R 974130 868350 592940 840540 840570 840580 622670 607580 823740 921150 980660 017730 622700 840620 557390 622660 607560 868260 951940 951030 007140 801730 972590 258510 007120 823760 782540 004370 007110 038580 303350 700990 007090 622410 776570 990510 007100 868240 132750 622360 933330 154840 990500 990470 291110 062150 955620 172310 938540 990490 990520 007130 951850 021630 101940 178940 076380 103370 097440 840630 938510 943340 291100 253100 041270 159660 146640 943940 269940 782520 152940 221670 980650 235740 990480 089310 144490 325240 782530 235780 302000 868290 801740 235780R 235730 235790 301990 235750 251420 251440 235720 059750 336280 267110 381510 251430 323360 323370 302010 129680 418440 474360 474120 235710 381550 089320 323350 251450 938480 395800 221680 235770 301980 143190 840610 938490 622680 222020 728030 840600 395770 705640 395760 553320 413770 585780 221700 501930 089340 323380 302020 413740 546070 395810 501890 501910 429230 613510 600760 221690 613240 728050 413760 432180 655550 705680 501870 301970 413730 728080 343000 624600 671740 415410 579130 415420 655540 694280 501880 721510 721570 694670 722630 671720 501940 694240 671730 694350 703800 008350 490240 395780 600860 694270 655570 381680 432170 246830 980630 703790 735350 626010 782550 323390 579140 988150 703780 413750 501950 566110 501900 960730 187060 008390 187230 899600 187210 901950 008300 008340 267570 901980 899590 081200 081200R 694330 915500 998720 899650 929640 323340 880700 988080 880200 266150 998780 501960 257820 901960 998800 167450 391910 432150 267470 915550 281220 248450 097620 248240 579120 267700 081170 081130 902020 735370 150780 221340 167440 915480 915590 865150 187170 894420 880630 929630 728090 187200 703770 671760 221300 739460 998760 187280 988140 267490 150710 750420 395790 484840 834940 929620 418740 221320 694320 444000 463730 557730 262710 694260 610530 463740 418760 221330 943380 988120 859730 899680 036890 081140 943390 610540 866250 468730 988110 167460 625630 295340 224210 295350 326920 393360 491490 325980 295370 325990 488620 326890 408640 506110 326880 344960 897030 488610 866240 866230 506100 295380 052120 545700 173340 052110 052200 871100 295360 371020 173120 173220 492580 488630 492610 344950 470430 492630 960700 303200 880680 028410 176550 735360 492590 699330 326910 173140 068900 881830 506060 171030 176570 833710 052100 593510 303210 258470 258450 823470 171020 171040 212610 028430 171050 173190 371000 870950 028450 176560 699470 212450 212640 823460 052130 173350 052190 333740 303180 371010 040580 040640 028520 028530 008360 823450 833730 302570 052140 008330 376300 170970 871040 593590 312200 171000 171010 028370 212560 173130 302550 596780 699480 068920 593620 768970 734020 943360 492600 614300 333750 376250 506070 614310 733890 614280 376420 614240 376480 614340 614370 733960 028390 376350 614260 408650 090430 129060 090460 614350 819000 129120 636010 129070 614380 682250 636080 176540 636040 614360 593680 253690 074190 682210 129040 376270 635990 549100 635980 189030 323840 262080 241120 297020 614250 682350 262090 262130 262120 262170 479250 241060 241080 614320 258490 915530 090420 028440 143520 303190 682490 262030 522460 522500 541300 479240 541320 262190 241090 296960 297030 541310 541290 175530 540650 541330 574530 762550 522550 762560 540620 540670 705620 540660 929580 262100 554810 762530 780520 780480 262060 085230 085240 522510 522540 186950 780500 762540 014960 296920 522450 847390 504370 847420 519500 728070 990870 315000 549040 593710 085210 847230 519470 297000 935400 411550 960750 187080 014870 960710 014890 014720 296950 960720 833720 014970 517300 014880 989630 040530 707350 177270 126150 728720 728730 847240 728760 149610 728740 728750 522480 333800 936990 519510 043280 074180 144020 056590 074190 144030 056600 074200 144040 152000 212100 230090 038470 315300 435950 485040 559990 038480
305680 320860 424600 441700 535680 749990 774100 009390 069380 302560 320870 435960 723350 102270 152380 316690 317930 428310 441770 445040 535690 651900 779440 779450 829410 829420 038490 071240 102280 152390 178510 286930 302570 315310 332160 377930 435970 454020 484160 563650 584180 619070 779460 786220 813470 829430 377940 439930 730750 991740 319300 330650 366260 377950 428730 446670 458590 764250 100620 434070 802990 803010 803030 803050 803070 803090 803100 803380 804980 886980 899930 346360 346340 173410 766370 771140 736650 736730 736740 616140 057390 958920 057400 337340 474800 126140 728790 969220 517340 387290 033270 387300 057640 175050 250500 250510 289350 298240 305690 320880 323440 417360 419400 430220 440300 448470 451120 482780 484170 530170 563660 607650 607660 723180 742580 774110 813480 849560 849570 908460 068590 301520 416990 428740 914490 210000 250520 286950 296380 297200 305700 316700 430100 430110 440310 441710 484180 484190 563670 563680 607670 607680 607690 723190 723200 723210 762190 774120 774130 782400 817070 846510 849590 849600 880420 892210 908470 962770 420930 483540 886340 914500 043430 210010 295480 317580 424610 444600 484200 563690 607700 668990 723220 849610 886350 914510 956440 106280 106320 106340 535700 946750 958600 798590 886190 798600 798620 798630 798650 798670 463290 798710 006580 148030 295570 304040 377960 071260 100630 121990 303310 328890 444610 723230 786230 006590 071280 303320 311990 328900 377970 422010 436100 456860 764260 824180 100640 306540 317590 332950 432470 449730 764270 824190 955580 071290 071300 101620 299010 307670 324210 334980 425000 439270 457290 584190 584200 703870 774150 824200 908580 071340 334990 453750 584220 774160 300330 305250 332170 430230 454030 764280 824210 870680 104380 305260 313090 764290 774170 824220 432860 527920 654190 786240 824230 841660 607520 607540 607550 607560 607570 607590 824400 305270 314030 346240 452300 839810 340940 715450 072370 148210 315080 430120 504960 018620 317600 384990 436110 616350 616360 989630 305310 384980 305320 585910 010130 097550 905080 262040 506040 545730 765870 005780 085200 102630 207010 207020 406320 410030 410040 506050 540600 586500 593580 614290 686130 709090 757500 786960 840550 840560 915050 943350 074430 127690 127720 129110 171600 181750 181760 282660 354460 376210 386120 410050 455000 460500 506080 530480 551000 584900 586450 586510 593600 593800 595940 622440 636030 638810 705370 709100 765880 790560 824530 840590 895570 924670 927390 010200 012160 012900 023710 023730 023740 080920 096360 102640 143560 143570 143580 207030 207040 207050 207060 268440 311900 421960 424650 432160 455210 512180 516580 541280 586530 586540 586550 586560 586570 591060 592700 593490 593520 593830 593860 607010 654350 657140 671720 715510 725710 765900 768940 768960 786970 786980 786990 787000 790090 792710 824540 857250 895590 895620 980380 039230 090410 098930 102650 110130 127740 127770 145190 181770 181780 181790 200040 207070 241050 262110 271840 308220 326020 376440 381010 387960 389160 390660 418720 418730 453290 511150 540630 547880 586580 591070 614390 617900 661550 705380 714840 725720 727310 734470 771760 787010 815510 824550 889820 975140 990500 996330 039250 039260 039270 075100 090450 095410 102660 102670 102680 129050 144400 147520 166380 181800 181810 181820 187870 207080 207090 207100 207110 207120 241130 247450 248420 257860 269220 289970 326900 354490 354500 398580 427700 441340 442730 452650 506090 511160
521940 536760 557950 586590 586600 586610 586620 591080 593480 614550 627240 682410 725730 725740 725750 731780 733990 740560 765920 787020 787030 787040 787050 792430 817320 824560 824570 873040 895650 947630 990650 009530 039280 039290 055050 090480 095700 102690 166600 181830 181840 207160 207170 207180 247480 308920 321710 333700 333790 354520 415400 420180 463530 491750 522560 545710 563930 586630 586640 593530 593560 657740 682450 682480 725760 787060 787070 801080 819460 824580 840640 840650 907230 938450 943400 116370 151530 181850 297040 420190 546160 614830 725770 780530 787080 968220 990660 220650 317290 546480 732710 840660 840670 033640 378260 617940 052840 052850 052860 052880 079760 101940 101950 101960 101970 101980 209680 249580 301540 309020 322850 330750 361660 413720 434580 438650 474250 474740 475460 504020 515770 534030 540610 561750 606090 694420 694510 694520 694530 722760 728020 728040 728100 790590 824280 824290 839640 906820 907370 907840 962050 008280 008290 008310 008320 008370 008380 008400 028400 068870 080970 137820 167420 167430 173170 189730 189740 268900 305430 308590 315110 331980 339570 360680 413060 430460 438660 438670 441300 453820 455430 474080 474300 474500 474800 475500 476200 534170 534180 558800 561760 585370 606100 652180 722770 765210 765220 771710 771720 771730 771740 771750 773270 813190 836680 839650 839660 848100 899620 899630 899660 899670 907380 907850 907860 922870 962060 000030 012930 021910 021940 021960 028420 038380 052150 068910 081200 101990 102000 102010 102030 102040 102050 102060 102070 102080 102090 102100 102110 102120 102130 137250 137260 137270 167470 167480 167490 178510 180270 189750 189760 189770 209690 209700 300400 304350 309030 314810 319370 324340 325520 339580 360690 413070 434590 438680 438700 446790 453200 474050 474450 474750 475070 475350 475470 475800 476130 501590 521510 536860 557360 561780 561790 561800 610150 621770 655530 692170 694250 694290 839670 839680 847380 850030 876510 901920 901930 901940 901970 902000 902010 907390 907870 915470 915520 915540 915560 915570 915580 999980 999990 012940 021980 021990 038390 131940 150720 150730 150740 150750 150760 150770 167500 167510 167520 173200 173240 173310 173320 178540 180000 206110 209710 209720 212500 221350 233680 237870 302440 302560 308560 313430 319030 324270 330460 355750 407300 407310 428790 434000 438710 439490 441830 445800 446460 458240 465180 474190 474200 474420 474690 474700 475020 475290 475430 475750 476070 534190 552980 556450 559680 559690 699550 705610 705630 705650 705660 705670 747260 765230 773280 839690 839700 847400 907400 907410 907420 907880 907890 934300 981490 998690 998700 998710 998730 998740 998750 006280 021840 022180 179650 180330 187030 187040 187050 187260 187270 235760 258460 258480 276010 278100 282700 292840 305240 305440 308130 310100 315120 325890 331990 367540 392340 414010 435650 448210 455450 472810 474090 474310 474510 474810 475130 475510 475850 476210 510060 515960 517460 595770 621780 679210 694300 694310 694340 813200 839710 839720 839730 839740 871070 897050 907430 907440 907900 907910 962090 974850 988050 988060 988070 988090 988100 988130 988160 998770 998790 998810 081150 081160 081190 085260 131950 187180 187190 209740 304500 308570 308580 308590 309040 314820 319380 330760 357800 402630 425790 446800 453210 465190 474260 474460 474760 475080 475360 475480 476140 483150 483160 501920 511390 553160 565250 566240 567650 585410 585420 585430 585450 621790
765240 783010 813210 839750 839760 848130 879960 880050 880610 880620 880640 880650 880660 880670 880690 881030 907450 907460 907920 936840 209750 286520 301790 308800 309050 325530 407320 418990 423480 434610 446810 474060 474470 475370 476150 483410 561820 585460 585470 626690 721500 721550 721560 735340 788420 907470 907480 907930 955520 202910 288560 302580 325540 339620 413080 440970 474270 474770 475090 475810 483170 515720 540620 585480 621800 677840 724260 847290 907940 929590 960240 044760 069180 100720 113680 150590 205200 323810 347850 355760 407330 408370 432730 474980 476380 585490 786370 813980 927180 534200 559700 176040 849510 UDI: (01)00880304281752(17)180328(10)969340 (01)00880304282261(17)180428(10)720460 (01)00880304281745(17)180528(10)384800 (01)00880304282216(17)180528(10)515090 (01)00880304281721(17)180628(10)656970 (01)00880304281752(17)180628(10)762180 (01)00880304281783(17)180828(10)484210 (01)00880304182868(17)180928(10)772070 (01)00880304281622(17)181128(10)174580 (01)00880304281592(17)190528(10)668970 (01)00880304281653(17)190828(10)935310 (01)00880304182974(17)191001(10)232220R (01)00880304182882(17)191028(10)304370 (01)00880304839274(17)191101(10)484600R (01)00880304281615(17)200128(10)005690 (01)00880304282223(17)200228(10)205070 (01)00880304282315(17)200228(10)299180 (01)00880304282315(17)200328(10)504490 (01)00880304281615(17)200428(10)624880 (01)00880304281608(17)200928(10)954720 (01)00880304282322(17)200928(10)944280 (01)00880304282186(17)201128(10)679750 (01)00880304182868(17)210128(10)203780 (01)00880304182950(17)210228(10)400790 (01)00880304281653(17)210428(10)272900 (01)00880304282216(17)210428(10)255990 (01)00880304182875(17)210428(10)865380 (01)00880304282162(17)210428(10)816830 (01)00880304282162(17)210428(10)239560 (01)00880304281608(17)210628(10)551590 (01)00880304281615(17)210628(10)633980 (01)00880304281646(17)210828(10)112880 (01)00880304281653(17)211028(10)584730 (01)00880304282209(17)211128(10)841860 (01)00880304182936(17)211228(10)252090 (01)00880304282186(17)211228(10)198450 (01)00880304282193(17)220228(10)585710 (01)00880304182875(17)220428(10)009560 (01)00880304182936(17)220501(10)646320R (01)00880304282186(17)220528(10)606670 (01)00880304182943(17)220601(10)921890R (01)00880304281592(17)220728(10)954050 (01)00880304182929(17)220801(10)307720R (01)00880304182882(17)220828(10)307560 (01)00880304282261(17)221101(10)654200R (01)00880304281608(17)221128(10)305320 (01)00880304282339(17)221128(10)561810 (01)00880304684560(17)221203(10)411360 (01)00880304182905(17)221203(10)417380 (01)00880304182882(17)221204(10)768400 (01)00880304182912(17)221204(10)784840 (01)00880304182912(17)221204(10)784860 (01)00880304182929(17)221204(10)504500 (01)00880304684508(17)221205(10)892310 (01)00880304182882(17)221206(10)686150 (01)00880304684546(17)221211(10)433760 (01)00880304684461(17)221214(10)600390 (01)00880304182905(17)221217(10)794730 (01)00880304684522(17)221219(10)614490 (01)00880304182929(17)230101(10)186970R (01)00880304182929(17)230101(10)876700R (01)00880304182899(17)230102(10)876520 (01)00880304282322(17)230103(10)877850 (01)00880304282315(17)230104(10)877840 (01)00880304684584(17)230108(10)937090 (01)00880304684485(17)230108(10)421920 (01)00880304684560(17)230115(10)823740 (01)00880304684522(17)230115(10)974330 (01)00880304684546(17)230116(10)989630 (01)00880304684607(17)230116(10)874090 (01)00880304684560(17)230116(10)974370 (01)00880304182912(17)230121(10)034020 (01)00880304182899(17)230122(10)204530 (01)00880304684522(17)230122(10)989610 (01)00880304282322(17)230123(10)291600 (01)00880304684546(17)230123(10)248520 (01)00880304684584(17)230123(10)336780 (01)00880304182912(17)230124(10)322250 (01)00880304684560(17)230128(10)118730 (01)00880304684560(17)230128(10)292050 (01)00880304684560(17)230128(10)614530 (01)00880304684584(17)230128(10)411440 (01)00880304281639(17)230128(10)801030 (01)00880304684560(17)230129(10)065340 (01)00880304182882(17)230131(10)322160 (01)00880304182929(17)230201(10)475120R (01)00880304182929(17)230201(10)627780R (01)00880304684508(17)230204(10)823730 (01)00880304182899(17)230212(10)408160 (01)00880304282247(17)230213(10)447960 (01)00880304282353(17)230214(10)236120 (01)00880304684546(17)230215(10)273860 (01)00880304684560(17)230218(10)069120 (01)00880304684560(17)230218(10)614520 (01)00880304684560(17)230218(10)335290 (01)00880304282360(17)230219(10)459140 (01)00880304684584(17)230219(10)989620 (01)00880304684485(17)230219(10)654340 (01)00880304282223(17)230220(10)504180 (01)00880304684522(17)230221(10)626160 (01)00880304282346(17)230221(10)459130 (01)00880304282346(17)230221(10)616750 (01)00880304684546(17)230222(10)511240 (01)00880304684607(17)230222(10)387540 (01)00880304684546(17)230225(10)273850 (01)00880304684546(17)230225(10)081760 (01)00880304684546(17)230226(10)697740 (01)00880304684560(17)230227(10)596590 (01)00880304684461(17)230227(10)335270 (01)00880304282179(17)230227(10)276550 (01)00880304684560(17)230228(10)387530 (01)00880304684522(17)230228(10)387520 (01)00880304684560(17)230228(10)123870 (01)00880304182929(17)230301(10)627790R (01)00880304182929(17)230301(10)627800R (01)00880304182929(17)230301(10)777880R (01)00880304182929(17)230301(10)928150R (01)00880304684560(17)230304(10)335300 (01)00880304684546(17)230305(10)069140 (01)00880304282377(17)230314(10)722240 (01)00880304282223(17)230318(10)820410 (01)00880304684508(17)230318(10)989600 (01)00880304282292(17)230318(10)494990 (01)00880304684461(17)230319(10)697630 (01)00880304684508(17)230319(10)491530 (01)00880304684546(17)230319(10)683870 (01)00880304282223(17)230319(10)916170 (01)00880304684522(17)230319(10)770990 (01)00880304684546(17)230319(10)607020 (01)00880304684560(17)230320(10)616020 (01)00880304684584(17)230320(10)607030 (01)00880304684560(17)230320(10)771010 (01)00880304684584(17)230321(10)866060 (01)00880304684560(17)230321(10)671710 (01)00880304282377(17)230325(10)890020 (01)00880304684546(17)230325(10)774330 (01)00880304684546(17)230326(10)771000 (01)00880304182912(17)230326(10)770340 (01)00880304684584(17)230327(10)273870 (01)00880304684560(17)230327(10)771020 (01)00880304684560(17)230329(10)905820 (01)00880304684584(17)230329(10)774340 (01)00880304684584(17)230405(10)960250 (01)00880304684485(17)230405(10)770980 (01)00880304684522(17)230410(10)007440 (01)00880304282360(17)230411(10)971740 (01)00880304684607(17)230411(10)014320 (01)00880304684560(17)230411(10)902060 (01)00880304684560(17)230411(10)902050 (01)00880304684522(17)230412(10)902040 (01)00880304684522(17)230415(10)902030 (01)00880304684584(17)230415(10)902080 (01)00880304684560(17)230418(10)902070 (01)00880304182875(17)230425(10)921950 (01)00880304182882(17)230426(10)089350 (01)00880304182929(17)230429(10)089280 (01)00880304182936(17)230429(10)090420 (01)00880304182905(17)230429(10)089090 (01)00880304182929(17)230429(10)089270 (01)00880304182912(17)230430(10)153020 (01)00880304182905(17)230501(10)089150 (01)00880304182905(17)230501(10)089440 (01)00880304182905(17)230501(10)089430 (01)00880304182929(17)230501(10)203370 (01)00880304182936(17)230501(10)090400 (01)00880304182936(17)230501(10)181500 (01)00880304182882(17)230501(10)089400 (01)00880304182868(17)230501(10)094160 (01)00880304182868(17)230501(10)996800 (01)00880304182875(17)230501(10)094210 (01)00880304182875(17)230501(10)139640 (01)00880304182882(17)230501(10)089370 (01)00880304182882(17)230501(10)089380 (01)00880304182882(17)230501(10)089410 (01)00880304182899(17)230501(10)071320 (01)00880304182899(17)230501(10)106340 (01)00880304182899(17)230501(10)106350 (01)00880304182905(17)230501(10)089070 (01)00880304182905(17)230501(10)089120 (01)00880304182905(17)230501(10)089420 (01)00880304182912(17)230501(10)071290 (01)00880304182912(17)230501(10)089200 (01)00880304182912(17)230501(10)089220 (01)00880304182929(17)230501(10)089250 (01)00880304182905(17)230502(10)181640 (01)00880304182899(17)230502(10)106370 (01)00880304684522(17)230502(10)014290 (01)00880304684485(17)230502(10)105520 (01)00880304182899(17)230502(10)214660 (01)00880304182905(17)230502(10)203330 (01)00880304182875(17)230502(10)094170 (01)00880304182899(17)230502(10)106310 (01)00880304182899(17)230502(10)106330 (01)00880304182899(17)230502(10)106360 (01)00880304182899(17)230502(10)139650 (01)00880304182905(17)230502(10)089100 (01)00880304182905(17)230502(10)089130 (01)00880304182905(17)230502(10)089450 (01)00880304182912(17)230502(10)089230 (01)00880304182936(17)230502(10)090390 (01)00880304182936(17)230502(10)090410 (01)00880304182868(17)230503(10)094140 (01)00880304182875(17)230503(10)062760 (01)00880304182875(17)230503(10)094190 (01)00880304182875(17)230503(10)094200 (01)00880304182882(17)230503(10)089360 (01)00880304182899(17)230503(10)106380 (01)00880304182905(17)230503(10)089140 (01)00880304182912(17)230503(10)089210 (01)00880304684584(17)230506(10)140100 (01)00880304182905(17)230506(10)089110 (01)00880304182882(17)230506(10)089390 (01)00880304182899(17)230506(10)225950 (01)00880304182912(17)230506(10)089180 (01)00880304182912(17)230506(10)089190 (01)00880304182868(17)230506(10)094130 (01)00880304182899(17)230506(10)071260 (01)00880304182912(17)230506(10)071270 (01)00880304182943(17)230506(10)139540 (01)00880304684607(17)230507(10)139720 (01)00880304182905(17)230508(10)250170 (01)00880304182912(17)230508(10)257970 (01)00880304182882(17)230508(10)226250 (01)00880304182899(17)230508(10)106320 (01)00880304182905(17)230508(10)239900 (01)00880304182929(17)230508(10)089260 (01)00880304282322(17)230508(10)143750 (01)00880304182875(17)230508(10)071240 (01)00880304182905(17)230508(10)139660 (01)00880304182929(17)230508(10)089300 (01)00880304182929(17)230509(10)239990 (01)00880304182929(17)230509(10)089290 (01)00880304182929(17)230510(10)089240 (01)00880304182905(17)230513(10)226300 (01)00880304282360(17)230514(10)143770 (01)00880304182929(17)230515(10)279520 (01)00880304182936(17)230515(10)226140 (01)00880304282179(17)230516(10)462350 (01)00880304684546(17)230516(10)014300 (01)00880304182868(17)230516(10)094150 (01)00880304182905(17)230516(10)140170 (01)00880304182868(17)230517(10)296700 (01)00880304182868(17)230517(10)288700 (01)00880304182899(17)230517(10)268370 (01)00880304182899(17)230517(10)268400 (01)00880304182905(17)230517(10)224980 (01)00880304182912(17)230517(10)224990 (01)00880304282179(17)230520(10)462360 (01)00880304182868(17)230520(10)225490 (01)00880304182868(17)230520(10)288710 (01)00880304182929(17)230520(10)186730 (01)00880304684522(17)230521(10)225900 (01)00880304684584(17)230521(10)250230 (01)00880304182905(17)230522(10)288760 (01)00880304182899(17)230522(10)288640 (01)00880304182905(17)230522(10)318200 (01)00880304182899(17)230523(10)268380 (01)00880304182936(17)230528(10)372020 (01)00880304182905(17)230528(10)370800 (01)00880304182905(17)230528(10)288740 (01)00880304182912(17)230528(10)306160 (01)00880304182882(17)230528(10)318190 (01)00880304182875(17)230529(10)288720 (01)00880304182929(17)230529(10)203320 (01)00880304182912(17)230529(10)306180 (01)00880304182943(17)230529(10)317450 (01)00880304182875(17)230530(10)071250 (01)00880304182912(17)230530(10)392050 (01)00880304182929(17)230530(10)288670 (01)00880304182899(17)230530(10)268390 (01)00880304182912(17)230530(10)328050 (01)00880304182929(17)230530(10)288660 (01)00880304182882(17)230531(10)370050 (01)00880304182882(17)230531(10)400770 (01)00880304182929(17)230531(10)350850 (01)00880304182882(17)230531(10)268350 (01)00880304182875(17)230531(10)392510 (01)00880304182882(17)230531(10)268360 (01)00880304182912(17)230531(10)328070 (01)00880304282377(17)230531(10)307130 (01)00880304182875(17)230531(10)318170 (01)00880304182882(17)230531(10)268340 (01)00880304182882(17)230531(10)288630 (01)00880304182899(17)230531(10)345710 (01)00880304182912(17)230531(10)306170 (01)00880304182912(17)230531(10)328060 (01)00880304182912(17)230531(10)345720 (01)00880304182912(17)230531(10)345740 (01)00880304182936(17)230531(10)306190 (01)00880304182936(17)230531(10)306200 (01)00880304182899(17)230603(10)306150 (01)00880304182929(17)230604(10)402090 (01)00880304282322(17)230604(10)414180 (01)00880304182912(17)230604(10)390500 (01)00880304182936(17)230604(10)186740 (01)00880304182912(17)230604(10)071280 (01)00880304182899(17)230605(10)345700 (01)00880304182868(17)230605(10)318160 (01)00880304182875(17)230605(10)363510 (01)00880304182882(17)230606(10)460540 (01)00880304182929(17)230606(10)423150 (01)00880304182882(17)230606(10)422860 (01)00880304182905(17)230606(10)452890 (01)00880304182905(17)230606(10)318210 (01)00880304282346(17)230606(10)359200 (01)00880304182899(17)230606(10)288650 (01)00880304182912(17)230606(10)345730 (01)00880304182882(17)230607(10)365890 (01)00880304282322(17)230607(10)307120 (01)00880304182912(17)230607(10)409510 (01)00880304182912(17)230611(10)394070 (01)00880304182882(17)230611(10)288620 (01)00880304182905(17)230611(10)318220 (01)00880304182929(17)230611(10)340630 (01)00880304182868(17)230612(10)363500 (01)00880304282148(17)230612(10)402750 (01)00880304282247(17)230612(10)403650 (01)00880304182875(17)230612(10)633430 (01)00880304684560(17)230613(10)236180 (01)00880304684584(17)230613(10)142210 (01)00880304182875(17)230613(10)288730 (01)00880304282377(17)230617(10)404730 (01)00880304182868(17)230617(10)404130 (01)00880304182882(17)230624(10)473560 (01)00880304182899(17)230624(10)515900 (01)00880304182905(17)230625(10)540980 (01)00880304182882(17)230625(10)473550 (01)00880304182899(17)230625(10)515880 (01)00880304182905(17)230625(10)459840 (01)00880304182882(17)230626(10)515860 (01)00880304182882(17)230626(10)515870 (01)00880304182899(17)230626(10)581150 (01)00880304182912(17)230626(10)484740 (01)00880304182936(17)230626(10)520220 (01)00880304182943(17)230626(10)520230 (01)00880304182882(17)230627(10)318180 (01)00880304182882(17)230627(10)394060 (01)00880304182929(17)230627(10)520210 (01)00880304684522(17)230628(10)540900 (01)00880304282155(17)230628(10)462340 (01)00880304182899(17)230628(10)459830 (01)00880304182899(17)230628(10)515890 (01)00880304182899(17)230628(10)572050 (01)00880304182936(17)230628(10)409610 (01)00880304684546(17)230630(10)236170 (01)00880304182912(17)230701(10)531160 (01)00880304182912(17)230701(10)540890 (01)00880304282179(17)230701(10)522260 (01)00880304684546(17)230701(10)540560 (01)00880304684522(17)230701(10)142220 (01)00880304182899(17)230701(10)515910 (01)00880304182929(17)230701(10)288680 (01)00880304684508(17)230701(10)236160 (01)00880304182875(17)230701(10)594070 (01)00880304182905(17)230701(10)620700 (01)00880304182899(17)230702(10)581160 (01)00880304182899(17)230702(10)581140 (01)00880304182912(17)230702(10)484750 (01)00880304182882(17)230702(10)515840 (01)00880304684560(17)230703(10)250210 (01)00880304684607(17)230703(10)423080 (01)00880304684560(17)230703(10)512730 (01)00880304182905(17)230703(10)459860 (01)00880304182905(17)230703(10)551400 (01)00880304182905(17)230703(10)638130 (01)00880304182905(17)230703(10)645670 (01)00880304182899(17)230708(10)685550 (01)00880304182899(17)230708(10)572070 (01)00880304182899(17)230708(10)572060 (01)00880304182882(17)230708(10)572010 (01)00880304182882(17)230708(10)572020 (01)00880304182882(17)230708(10)581100 (01)00880304182899(17)230708(10)572040 (01)00880304182899(17)230708(10)572090 (01)00880304182905(17)230708(10)620720 (01)00880304182912(17)230708(10)620780 (01)00880304182943(17)230708(10)556710 (01)00880304684522(17)230709(10)250280 (01)00880304182875(17)230709(10)643470 (01)00880304182882(17)230709(10)572030 (01)00880304182905(17)230709(10)638150 (01)00880304182912(17)230709(10)620730 (01)00880304182929(17)230709(10)581170 (01)00880304182899(17)230710(10)581130 (01)00880304182875(17)230710(10)594050 (01)00880304182882(17)230710(10)581110 (01)00880304182882(17)230710(10)581120 (01)00880304182882(17)230710(10)694470 (01)00880304182912(17)230710(10)645700 (01)00880304182875(17)230711(10)594060 (01)00880304182875(17)230712(10)647650 (01)00880304182882(17)230712(10)406890 (01)00880304182936(17)230712(10)674450 (01)00880304182929(17)230712(10)577310 (01)00880304182905(17)230712(10)620710 (01)00880304182950(17)230712(10)556720 (01)00880304282308(17)230712(10)610400 (01)00880304182875(17)230712(10)695390 (01)00880304182882(17)230712(10)694490 (01)00880304182912(17)230712(10)620770 (01)00880304182912(17)230712(10)645710 (01)00880304182929(17)230712(10)577290 (01)00880304182929(17)230712(10)577300 (01)00880304182929(17)230712(10)581180 (01)00880304182929(17)230712(10)645720 (01)00880304182899(17)230714(10)631180 (01)00880304182905(17)230715(10)631200 (01)00880304182905(17)230715(10)674410 (01)00880304182875(17)230715(10)745150 (01)00880304182882(17)230715(10)695410 (01)00880304282131(17)230715(10)537910 (01)00880304182905(17)230715(10)638120 (01)00880304282148(17)230715(10)009780 (01)00880304182882(17)230715(10)694440 (01)00880304182899(17)230715(10)645640 (01)00880304182899(17)230715(10)695420 (01)00880304182912(17)230715(10)620750 (01)00880304182899(17)230715(10)724760 (01)00880304182912(17)230715(10)701030 (01)00880304182875(17)230717(10)744790 (01)00880304182899(17)230717(10)789530 (01)00880304182929(17)230717(10)798230 (01)00880304182936(17)230717(10)777440 (01)00880304182936(17)230717(10)561050 (01)00880304182905(17)230717(10)776890 (01)00880304282155(17)230718(10)181590 (01)00880304182882(17)230718(10)789630 (01)00880304182905(17)230718(10)789550 (01)00880304182899(17)230718(10)685430 (01)00880304182912(17)230718(10)777350 (01)00880304182936(17)230718(10)556700 (01)00880304182912(17)230718(10)620760 (01)00880304182912(17)230718(10)701020 (01)00880304282216(17)230718(10)724980 (01)00880304182905(17)230718(10)620690 (01)00880304182905(17)230718(10)645650 (01)00880304182905(17)230718(10)645660 (01)00880304182905(17)230718(10)748930 (01)00880304182905(17)230718(10)748950 (01)00880304182912(17)230718(10)620740 (01)00880304182912(17)230718(10)701050 (01)00880304182936(17)230718(10)561040 (01)00880304182936(17)230718(10)561040R (01)00880304182882(17)230718(10)354354 (01)00880304182882(17)230718(10)354684 (01)00880304182882(17)230718(10)831654 (01)00880304182929(17)230719(10)777340 (01)00880304182882(17)230719(10)724740 (01)00880304182899(17)230719(10)645630 (01)00880304182912(17)230719(10)645680 (01)00880304182912(17)230719(10)700980 (01)00880304182899(17)230719(10)646455 (01)00880304182899(17)230719(10)646847 (01)00880304182905(17)230720(10)789650 (01)00880304182950(17)230720(10)776630 (01)00880304182875(17)230720(10)594040 (01)00880304182899(17)230720(10)724770 (01)00880304182875(17)230721(10)695400 (01)00880304684546(17)230722(10)930020 (01)00880304182899(17)230722(10)685630 (01)00880304182912(17)230722(10)776440 (01)00880304182912(17)230722(10)789640 (01)00880304182936(17)230722(10)776960 (01)00880304182912(17)230722(10)847790 (01)00880304182905(17)230722(10)776900 (01)00880304182905(17)230722(10)900890 (01)00880304182882(17)230722(10)724750 (01)00880304182912(17)230722(10)620790 (01)00880304182875(17)230722(10)594030 (01)00880304182875(17)230722(10)645620 (01)00880304182882(17)230722(10)776840 (01)00880304182899(17)230722(10)572080 (01)00880304182905(17)230722(10)638140 (01)00880304182905(17)230722(10)748940 (01)00880304182905(17)230722(10)748960 (01)00880304182905(17)230722(10)798200 (01)00880304182912(17)230722(10)701000 (01)00880304182912(17)230722(10)701040 (01)00880304182912(17)230722(10)748990 (01)00880304182929(17)230722(10)581190 (01)00880304182929(17)230722(10)701060 (01)00880304182943(17)230722(10)744720 (01)00880304182929(17)230722(10)701060R (01)00880304182936(17)230722(10)776960R (01)00880304182875(17)230723(10)724730 (01)00880304182929(17)230723(10)861650 (01)00880304182905(17)230723(10)906890 (01)00880304182905(17)230723(10)695450 (01)00880304182912(17)230723(10)700990 (01)00880304182912(17)230723(10)701010 (01)00880304182936(17)230723(10)581200 (01)00880304684560(17)230724(10)014310 (01)00880304684584(17)230724(10)236190 (01)00880304282131(17)230725(10)460220 (01)00880304182899(17)230725(10)822780 (01)00880304182899(17)230725(10)822790 (01)00880304182912(17)230725(10)748970 (01)00880304182929(17)230725(10)744700 (01)00880304182929(17)230725(10)798240 (01)00880304182936(17)230725(10)744710 (01)00880304182875(17)230726(10)776830 (01)00880304182899(17)230726(10)776880 (01)00880304182882(17)230727(10)776860 (01)00880304182882(17)230729(10)861720 (01)00880304182882(17)230729(10)814940 (01)00880304282209(17)230802(10)956590 (01)00880304182882(17)230802(10)826060 (01)00880304182882(17)230802(10)814930 (01)00880304182899(17)230802(10)776870 (01)00880304282261(17)230804(10)955450 (01)00880304182899(17)230805(10)960070 (01)00880304182868(17)230805(10)711810 (01)00880304182899(17)230805(10)975190 (01)00880304182905(17)230805(10)838310 (01)00880304182912(17)230805(10)798220 (01)00880304182912(17)230805(10)945560 (01)00880304281684(17)230805(10)955110 (01)00880304281592(17)230806(10)955030 (01)00880304182882(17)230806(10)838290 (01)00880304282209(17)230806(10)906620 (01)00880304182899(17)230806(10)822770 (01)00880304182875(17)230806(10)814920 (01)00880304182899(17)230806(10)814950 (01)00880304182899(17)230806(10)906850 (01)00880304182905(17)230806(10)695440 (01)00880304182905(17)230806(10)873450 (01)00880304281691(17)230807(10)955430 (01)00880304182875(17)230807(10)861660 (01)00880304182882(17)230807(10)847940 (01)00880304182875(17)230807(10)906830 (01)00880304182936(17)230807(10)838340 (01)00880304182912(17)230808(10)031970 (01)00880304182905(17)230808(10)873430 (01)00880304182882(17)230808(10)694450 (01)00880304182882(17)230808(10)838280 (01)00880304182899(17)230808(10)838300 (01)00880304182905(17)230808(10)873440 (01)00880304182929(17)230808(10)744690 (01)00880304182929(17)230808(10)814870 (01)00880304182929(17)230808(10)838330 (01)00880304182929(17)230808(10)873460 (01)00880304182905(17)230809(10)175720 (01)00880304182912(17)230812(10)838320 (01)00880304182912(17)230812(10)798210 (01)00880304182905(17)230812(10)873420 (01)00880304282346(17)230813(10)955530 (01)00880304188167(17)230815(10)955980 (01)00880304182899(17)230815(10)975200 (01)00880304182905(17)230815(10)032030 (01)00880304182882(17)230815(10)104910 (01)00880304182905(17)230815(10)906870 (01)00880304182905(17)230815(10)906880 (01)00880304282322(17)230816(10)711780 (01)00880304182905(17)230816(10)960740 (01)00880304182899(17)230816(10)906860 (01)00880304182882(17)230817(10)023100 (01)00880304182882(17)230817(10)906840 (01)00880304182905(17)230817(10)031910 (01)00880304282360(17)230819(10)955630 (01)00880304182929(17)230819(10)011020 (01)00880304282209(17)230821(10)045720 (01)00880304182882(17)230821(10)104900 (01)00880304182899(17)230821(10)938220 (01)00880304281813(17)230821(10)047700 (01)00880304182899(17)230821(10)623000 (01)00880304182899(17)230821(10)975140 (01)00880304182868(17)230821(10)115750 (01)00880304182905(17)230821(10)459850 (01)00880304282209(17)230822(10)045710 (01)00880304182868(17)230822(10)115760 (01)00880304182868(17)230822(10)175890 (01)00880304282148(17)230824(10)958910 (01)00880304282162(17)230824(10)958900 (01)00880304182882(17)230824(10)777470 (01)00880304282179(17)230826(10)957230 (01)00880304182905(17)230826(10)173210 (01)00880304282339(17)230827(10)013540 (01)00880304182905(17)230828(10)104970 (01)00880304182868(17)230828(10)173160 (01)00880304182929(17)230828(10)176040 (01)00880304282308(17)230829(10)024000 (01)00880304182905(17)230829(10)281420 (01)00880304182868(17)230829(10)218940 (01)00880304182899(17)230829(10)176110 (01)00880304282131(17)230830(10)276770 (01)00880304684560(17)230830(10)847900 (01)00880304182882(17)230830(10)175990 (01)00880304182899(17)230830(10)176070 (01)00880304182905(17)230903(10)175710 (01)00880304182912(17)230904(10)176030 (01)00880304182882(17)230904(10)176000 (01)00880304182882(17)230904(10)176020 (01)00880304182899(17)230905(10)176120 (01)00880304182905(17)230905(10)282750 (01)00880304182899(17)230905(10)347240 (01)00880304182899(17)230905(10)176130 (01)00880304182905(17)230905(10)282270 (01)00880304182912(17)230905(10)282390 (01)00880304182943(17)230905(10)175770 (01)00880304282148(17)230906(10)188210 (01)00880304182882(17)230906(10)188190 (01)00880304182905(17)230909(10)282280 (01)00880304182912(17)230909(10)282400 (01)00880304182936(17)230909(10)269110 (01)00880304182905(17)230910(10)176150 (01)00880304182929(17)230910(10)175760 (01)00880304282315(17)230916(10)310000 (01)00880304182905(17)230916(10)282250 (01)00880304182899(17)230917(10)388630 (01)00880304182929(17)230917(10)282430 (01)00880304182905(17)230917(10)282240 (01)00880304182899(17)230917(10)176080 (01)00880304182929(17)230917(10)282420 (01)00880304182936(17)230917(10)282440 (01)00880304182875(17)230926(10)343940 (01)00880304182905(17)230926(10)419400 (01)00880304182868(17)230926(10)459220 (01)00880304182905(17)230926(10)408290 (01)00880304182929(17)230926(10)412350 (01)00880304182905(17)230927(10)282230 (01)00880304182912(17)230927(10)282410 (01)00880304182899(17)230930(10)631350 (01)00880304182905(17)230930(10)373530 (01)00880304182912(17)230930(10)456470 (01)00880304182868(17)230930(10)115770 (01)00880304182875(17)230930(10)406150 (01)00880304282308(17)230930(10)320250 (01)00880304182929(17)230930(10)173280 (01)00880304182899(17)230930(10)176090 (01)00880304182905(17)230930(10)282260 (01)00880304182912(17)230930(10)175740 (01)00880304182936(17)230930(10)269080 (01)00880304182936(17)231001(10)540870 (01)00880304182936(17)231001(10)269090 (01)00880304182936(17)231001(10)269070 (01)00880304182936(17)231001(10)269100 (01)00880304182899(17)231004(10)176100 (01)00880304282339(17)231004(10)402270 (01)00880304182875(17)231007(10)406170 (01)00880304182905(17)231007(10)521690 (01)00880304182875(17)231008(10)581710 (01)00880304182875(17)231008(10)408310 (01)00880304182905(17)231008(10)440950 (01)00880304182905(17)231009(10)440920 (01)00880304182905(17)231010(10)440970 (01)00880304182875(17)231010(10)440900 (01)00880304182882(17)231010(10)176010 (01)00880304182905(17)231010(10)440930 (01)00880304182905(17)231011(10)560740 (01)00880304182875(17)231012(10)597860 (01)00880304182905(17)231015(10)627470 (01)00880304182875(17)231015(10)440890 (01)00880304282308(17)231015(10)402260 (01)00880304182868(17)231015(10)406130 (01)00880304182912(17)231015(10)440980 (01)00880304182912(17)231016(10)175750 (01)00880304182929(17)231017(10)606180 (01)00880304182882(17)231017(10)496160 (01)00880304684508(17)231017(10)442200 (01)00880304281790(17)231019(10)370740 (01)00880304182912(17)231021(10)528530 (01)00880304182882(17)231021(10)514580 (01)00880304182905(17)231021(10)440940 (01)00880304182929(17)231021(10)493990 (01)00880304684508(17)231022(10)604000 (01)00880304282339(17)231022(10)455390 (01)00880304182875(17)231022(10)514570 (01)00880304182875(17)231022(10)440910 (01)00880304182882(17)231022(10)496100 (01)00880304182882(17)231022(10)496140 (01)00880304182936(17)231023(10)456800 (01)00880304282339(17)231023(10)646550 (01)00880304182882(17)231023(10)496130 (01)00880304182875(17)231024(10)716050 (01)00880304182905(17)231024(10)627510 (01)00880304182899(17)231024(10)570920 (01)00880304282322(17)231024(10)453840 (01)00880304182905(17)231024(10)493970 (01)00880304182905(17)231024(10)176140 (01)00880304182929(17)231024(10)539630 (01)00880304182875(17)231024(10)496070 (01)00880304182899(17)231024(10)514610 (01)00880304182899(17)231024(10)539610 (01)00880304182912(17)231024(10)441030 (01)00880304182875(17)231025(10)611580 (01)00880304182882(17)231025(10)642900 (01)00880304182912(17)231025(10)441000 (01)00880304182875(17)231025(10)742480 (01)00880304182912(17)231025(10)175730 (01)00880304182899(17)231026(10)661610 (01)00880304182882(17)231028(10)496080 (01)00880304182868(17)231028(10)115780 (01)00880304182899(17)231028(10)596920 (01)00880304182882(17)231028(10)656890 (01)00880304182882(17)231028(10)514590 (01)00880304182899(17)231028(10)596910 (01)00880304182912(17)231028(10)441010 (01)00880304282193(17)231029(10)511100 (01)00880304182905(17)231029(10)704820 (01)00880304182882(17)231029(10)496090 (01)00880304182899(17)231029(10)514600 (01)00880304182899(17)231029(10)539620 (01)00880304182905(17)231030(10)661400 (01)00880304182929(17)231030(10)441040 (01)00880304282193(17)231031(10)683130 (01)00880304182912(17)231031(10)440990 (01)00880304182912(17)231031(10)596940 (01)00880304282193(17)231104(10)813070 (01)00880304182943(17)231107(10)828660 (01)00880304282278(17)231108(10)676560 (01)00880304182905(17)231108(10)497900 (01)00880304182905(17)231108(10)540960 (01)00880304182929(17)231108(10)827250 (01)00880304182868(17)231108(10)513710 (01)00880304182905(17)231108(10)827090 (01)00880304182905(17)231108(10)857620 (01)00880304684546(17)231113(10)402380 (01)00880304684485(17)231113(10)539910 (01)00880304281608(17)231114(10)676470 (01)00880304281622(17)231114(10)676500 (01)00880304182899(17)231115(10)828510 (01)00880304182905(17)231116(10)440960 (01)00880304282223(17)231116(10)676530 (01)00880304182882(17)231117(10)767990 (01)00880304282322(17)231118(10)696260 (01)00880304182912(17)231118(10)826420 (01)00880304182868(17)231118(10)908890 (01)00880304182882(17)231118(10)496170 (01)00880304182899(17)231118(10)828470 (01)00880304182905(17)231118(10)596930 (01)00880304182912(17)231118(10)828690 (01)00880304282223(17)231118(10)511110 (01)00880304282247(17)231118(10)676550 (01)00880304182868(17)231118(10)513700 (01)00880304182882(17)231118(10)496120 (01)00880304182912(17)231118(10)493980 (01)00880304182943(17)231121(10)539640 (01)00880304684584(17)231126(10)539920 (01)00880304282216(17)231127(10)067620 (01)00880304182882(17)231127(10)841400 (01)00880304282377(17)231201(10)402280 (01)00880304282209(17)231201(10)676510 (01)00880304182899(17)231201(10)514620 (01)00880304684560(17)231203(10)619450 (01)00880304684508(17)231204(10)358480 (01)00880304182899(17)231204(10)539600 (01)00880304182929(17)231204(10)596950 (01)00880304182905(17)231206(10)790210 (01)00880304182912(17)231206(10)051040 (01)00880304182882(17)231206(10)741190 (01)00880304182882(17)231206(10)873070 (01)00880304182929(17)231206(10)741140 (01)00880304282216(17)231207(10)676520 (01)00880304282193(17)231207(10)851340 (01)00880304182912(17)231209(10)698530 (01)00880304282322(17)231209(10)051020 (01)00880304282322(17)231209(10)873730 (01)00880304182875(17)231209(10)943180 (01)00880304182882(17)231209(10)943190 (01)00880304182936(17)231209(10)873090 (01)00880304182882(17)231210(10)943200 (01)00880304182899(17)231212(10)994500 (01)00880304182936(17)231213(10)194300 (01)00880304182912(17)231213(10)282360 (01)00880304182882(17)231213(10)214410 (01)00880304182899(17)231213(10)943160 (01)00880304182899(17)231213(10)994480 (01)00880304684560(17)231216(10)199090 (01)00880304684508(17)231216(10)062980 (01)00880304282230(17)231217(10)676540 (01)00880304182905(17)231218(10)007790 (01)00880304182936(17)231219(10)994490 (01)00880304282322(17)231220(10)767470 (01)00880304182929(17)231220(10)798610 (01)00880304684560(17)231227(10)701690 (01)00880304182882(17)231227(10)115190 (01)00880304182899(17)231227(10)078300 (01)00880304182899(17)231230(10)303830 (01)00880304684560(17)240102(10)115060 (01)00880304684560(17)240102(10)402390 (01)00880304182899(17)240102(10)873080 (01)00880304182950(17)240102(10)115200 (01)00880304282339(17)240102(10)767480 (01)00880304182899(17)240102(10)239720 (01)00880304182905(17)240102(10)239750 (01)00880304281615(17)240103(10)676490 (01)00880304182868(17)240103(10)115130 (01)00880304182875(17)240103(10)115160 (01)00880304684546(17)240108(10)565300 (01)00880304684546(17)240224(10)942200 (01)00880304182905(17)240108(10)078320 (01)00880304684508(17)240108(10)701680 (01)00880304182875(17)240108(10)115140 (01)00880304182936(17)240109(10)374430 (01)00880304684584(17)240109(10)140330 (01)00880304684560(17)240110(10)906030 (01)00880304182929(17)240110(10)994510 (01)00880304684560(17)240112(10)942210 (01)00880304282353(17)240112(10)873740 (01)00880304182882(17)240113(10)115170 (01)00880304182899(17)240113(10)570910 (01)00880304182899(17)240118(10)305020 (01)00880304684546(17)240120(10)115050 (01)00880304684522(17)240122(10)338380 (01)00880304182936(17)240123(10)216630 (01)00880304182882(17)240123(10)382130 (01)00880304182905(17)240123(10)581320 (01)00880304684584(17)240123(10)338400 (01)00880304684485(17)240123(10)179360 (01)00880304182882(17)240123(10)654710 (01)00880304282209(17)240124(10)196200 (01)00880304684560(17)240127(10)374730 (01)00880304282377(17)240128(10)339710 (01)00880304282155(17)240130(10)214260 (01)00880304684584(17)240130(10)216450 (01)00880304182875(17)240130(10)305010 (01)00880304182868(17)240130(10)115120 (01)00880304182899(17)240130(10)239700 (01)00880304182929(17)240130(10)305050 (01)00880304182936(17)240130(10)484060 (01)00880304282155(17)240131(10)214240 (01)00880304182936(17)240131(10)078330 (01)00880304182899(17)240201(10)620050 (01)00880304182899(17)240201(10)417060 (01)00880304182936(17)240201(10)305060 (01)00880304282155(17)240203(10)214250 (01)00880304182899(17)240203(10)741200 (01)00880304182936(17)240204(10)372430 (01)00880304182899(17)240206(10)078310 (01)00880304182936(17)240207(10)620020 (01)00880304684508(17)240217(10)942190 (01)00880304281615(17)240210(10)723960 (01)00880304182899(17)240212(10)481790 (01)00880304182905(17)240212(10)305030 (01)00880304281615(17)240213(10)621790 (01)00880304182929(17)240213(10)484020 (01)00880304182943(17)240213(10)484070 (01)00880304182882(17)240217(10)382110 (01)00880304182905(17)240217(10)239730 (01)00880304182929(17)240217(10)305040 (01)00880304182929(17)240219(10)929730 (01)00880304182875(17)240219(10)115150 (01)00880304182868(17)240220(10)382090 (01)00880304182868(17)240220(10)417050 (01)00880304182950(17)241002(10)528370 (01)00880304182899(17)240221(10)481780 (01)00880304182912(17)240221(10)526700 (01)00880304182929(17)240221(10)239760 (01)00880304182929(17)240221(10)562580 (01)00880304182875(17)240222(10)610730 (01)00880304684546(17)240222(10)179370 (01)00880304684522(17)240222(10)253400 (01)00880304182929(17)240222(10)484010 (01)00880304182899(17)240222(10)811250 (01)00880304182912(17)240222(10)562570 (01)00880304684560(17)240223(10)338390 (01)00880304684584(17)240225(10)906040 (01)00880304684522(17)240225(10)165400 (01)00880304684522(17)240225(10)483700 (01)00880304182882(17)240225(10)382100 (01)00880304182899(17)240225(10)304730 (01)00880304182905(17)240225(10)382140 (01)00880304182882(17)240226(10)190130 (01)00880304282308(17)240226(10)688100 (01)00880304684560(17)240226(10)639300 (01)00880304182882(17)240226(10)488780 (01)00880304182899(17)240226(10)099520 (01)00880304182899(17)240226(10)132600 (01)00880304182899(17)240226(10)239710 (01)00880304182899(17)240226(10)619660 (01)00880304182912(17)240226(10)811270 (01)00880304684560(17)240227(10)568290 (01)00880304182899(17)240227(10)880520 (01)00880304182899(17)240227(10)895510 (01)00880304182882(17)240227(10)635600 (01)00880304684560(17)240227(10)639290 (01)00880304182882(17)240227(10)382120 (01)00880304182905(17)240228(10)239740 (01)00880304182899(17)240228(10)147310 (01)00880304182899(17)240228(10)382130 (01)00880304182875(17)240229(10)907390 (01)00880304182875(17)240229(10)907410 (01)00880304182882(17)240229(10)863210 (01)00880304182905(17)240302(10)811260 (01)00880304684508(17)240302(10)627130 (01)00880304282216(17)240302(10)977500 (01)00880304182882(17)240302(10)863230 (01)00880304182882(17)240302(10)877910 (01)00880304182905(17)240302(10)382150 (01)00880304182912(17)240302(10)654720 (01)00880304182882(17)240303(10)811240 (01)00880304182912(17)240303(10)811280 (01)00880304684560(17)240304(10)253410 (01)00880304684546(17)240304(10)419280 (01)00880304182868(17)240304(10)619650 (01)00880304182882(17)240304(10)863220 (01)00880304182905(17)240304(10)382160 (01)00880304182929(17)240305(10)484000 (01)00880304182868(17)240306(10)080350 (01)00880304182899(17)240306(10)133770 (01)00880304182905(17)240306(10)526690 (01)00880304182882(17)240306(10)562550 (01)00880304182936(17)240306(10)417070 (01)00880304182868(17)240307(10)257840 (01)00880304182899(17)240307(10)481800 (01)00880304282360(17)240310(10)735200 (01)00880304182868(17)240311(10)877900 (01)00880304182912(17)240311(10)977790 (01)00880304282360(17)240311(10)735200R (01)00880304684508(17)240312(10)688050 (01)00880304182868(17)240312(10)120670 (01)00880304182882(17)240313(10)877920 (01)00880304182929(17)240313(10)099530 (01)00880304182875(17)240313(10)907480 (01)00880304182875(17)240316(10)907370 (01)00880304182905(17)240316(10)562560 (01)00880304182875(17)240317(10)245850 (01)00880304182868(17)240318(10)977760 (01)00880304182936(17)240318(10)562590 (01)00880304182868(17)240319(10)246440 (01)00880304182899(17)240323(10)121210 (01)00880304182899(17)240323(10)929690 (01)00880304182899(17)240323(10)388290 (01)00880304684560(17)240324(10)563680 (01)00880304182905(17)240324(10)354780 (01)00880304281592(17)240324(10)242570 (01)00880304182868(17)240324(10)120660 (01)00880304182905(17)240324(10)147340 (01)00880304684607(17)240327(10)063590 (01)00880304182882(17)240327(10)877990 (01)00880304281608(17)240327(10)242580 (01)00880304282223(17)240327(10)242600 (01)00880304182912(17)240328(10)132650 (01)00880304182905(17)240328(10)207230 (01)00880304182905(17)240329(10)409920 (01)00880304182929(17)240329(10)268620 (01)00880304182905(17)240329(10)147320 (01)00880304182882(17)240330(10)380530 (01)00880304182882(17)240330(10)944830 (01)00880304182905(17)240330(10)359720 (01)00880304182929(17)240330(10)099540 (01)00880304684546(17)240330(10)797970 (01)00880304182899(17)240330(10)132610 (01)00880304182929(17)240330(10)099550 (01)00880304684508(17)240330(10)483710 (01)00880304182875(17)240330(10)907400 (01)00880304182899(17)240330(10)099510 (01)00880304182905(17)240330(10)977770 (01)00880304684560(17)240331(10)053380 (01)00880304182875(17)240331(10)268890 (01)00880304182899(17)240331(10)132630 (01)00880304182905(17)240331(10)050480 (01)00880304182882(17)240401(10)878000 (01)00880304282292(17)240401(10)163730 (01)00880304282315(17)240401(10)163740 (01)00880304282315(17)240401(10)688110 (01)00880304182912(17)240401(10)132640 (01)00880304182929(17)240403(10)431590 (01)00880304182912(17)240403(10)356390 (01)00880304182905(17)240406(10)977780 (01)00880304182875(17)240407(10)268750 (01)00880304281622(17)240408(10)242590 (01)00880304182882(17)240408(10)380050 (01)00880304182905(17)240413(10)091470 (01)00880304182905(17)240413(10)147350 (01)00880304182912(17)240413(10)132660 (01)00880304182875(17)240413(10)268910 (01)00880304182936(17)240413(10)099560 (01)00880304182943(17)240414(10)078340 (01)00880304182905(17)240415(10)147330 (01)00880304684522(17)240417(10)053360 (01)00880304182868(17)240421(10)333570 (01)00880304684522(17)240422(10)521820 (01)00880304182875(17)240520(10)863360 (01)00880304182912(17)240424(10)583760 (01)00880304182899(17)240424(10)440310 (01)00880304182905(17)240424(10)562930 (01)00880304182899(17)240428(10)649470 (01)00880304282247(17)240428(10)529820 (01)00880304282230(17)240428(10)529810 (01)00880304182899(17)240428(10)519440 (01)00880304182905(17)240428(10)475470 (01)00880304182912(17)240428(10)488600 (01)00880304182929(17)240429(10)409870 (01)00880304182912(17)240429(10)405540 (01)00880304182882(17)240429(10)405470 (01)00880304182899(17)240430(10)207220 (01)00880304182875(17)240504(10)536810 (01)00880304182882(17)240504(10)599050 (01)00880304684584(17)240504(10)053390 (01)00880304684546(17)240504(10)053370 (01)00880304182882(17)240504(10)380040 (01)00880304182905(17)240504(10)380070 (01)00880304684546(17)240505(10)795610 (01)00880304684522(17)240505(10)888170 (01)00880304182882(17)240505(10)519390 (01)00880304182899(17)240505(10)475410 (01)00880304182905(17)240505(10)475440 (01)00880304182912(17)240505(10)488590 (01)00880304182929(17)240505(10)488620 (01)00880304182936(17)240505(10)257380 (01)00880304182868(17)240506(10)405460 (01)00880304182899(17)240506(10)475380 (01)00880304684560(17)240507(10)445050 (01)00880304182899(17)240507(10)405530 (01)00880304182905(17)240507(10)333590 (01)00880304182905(17)240507(10)519450 (01)00880304182868(17)240511(10)011100 (01)00880304182868(17)240511(10)011120 (01)00880304182875(17)240512(10)298430 (01)00880304182905(17)240512(10)475450 (01)00880304182899(17)240512(10)475420 (01)00880304182899(17)240512(10)488700 (01)00880304182899(17)240512(10)519430 (01)00880304182936(17)240512(10)488630 (01)00880304684584(17)240514(10)353430 (01)00880304182882(17)240514(10)488570 (01)00880304182936(17)240518(10)257370 (01)00880304182868(17)240518(10)011130 (01)00880304182899(17)240518(10)488690 (01)00880304182905(17)240519(10)475460 (01)00880304182899(17)240520(10)519420 (01)00880304182882(17)240521(10)519400 (01)00880304182875(17)240522(10)822340 (01)00880304182882(17)240522(10)716040 (01)00880304182899(17)240525(10)475390 (01)00880304182868(17)240526(10)465740 (01)00880304182875(17)240526(10)562850 (01)00880304182882(17)240526(10)929320 (01)00880304182905(17)240526(10)589960 (01)00880304182875(17)240526(10)863350 (01)00880304182875(17)240526(10)863380 (01)00880304182882(17)240526(10)488560 (01)00880304182899(17)240526(10)488720 (01)00880304182912(17)240527(10)751960 (01)00880304182899(17)240527(10)475370 (01)00880304282193(17)240527(10)529790 (01)00880304182868(17)240527(10)011110 (01)00880304182868(17)240527(10)380030 (01)00880304182905(17)240527(10)636290 (01)00880304182905(17)240527(10)636300 (01)00880304182905(17)240527(10)764780 (01)00880304281615(17)240528(10)082410 (01)00880304182899(17)240528(10)685280 (01)00880304182905(17)240528(10)706430 (01)00880304182905(17)240528(10)718560 (01)00880304182936(17)240528(10)769430 (01)00880304182882(17)240528(10)716070 (01)00880304182882(17)240528(10)716030 (01)00880304282247(17)240528(10)908500 (01)00880304182875(17)240528(10)863340 (01)00880304182882(17)240528(10)634780 (01)00880304182882(17)240528(10)716060 (01)00880304182899(17)240528(10)488710 (01)00880304182912(17)240528(10)519460 (01)00880304182912(17)240529(10)563910 (01)00880304684584(17)240529(10)595550 (01)00880304684584(17)240529(10)595590 (01)00880304182868(17)240529(10)066070 (01)00880304684560(17)240529(10)814410 (01)00880304182912(17)240529(10)488610 (01)00880304281615(17)240529(10)908490 (01)00880304182875(17)240529(10)927140 (01)00880304182875(17)240529(10)927160 (01)00880304182882(17)240529(10)635530 (01)00880304182912(17)240529(10)645020 (01)00880304182912(17)240529(10)764790 (01)00880304182912(17)240529(10)764800 (01)00880304684522(17)240530(10)958600 (01)00880304282353(17)240530(10)453760 (01)00880304684560(17)240601(10)108560 (01)00880304182875(17)240601(10)066060 (01)00880304182929(17)240601(10)431740 (01)00880304182929(17)240601(10)784880 (01)00880304182875(17)240602(10)927150 (01)00880304182912(17)240603(10)488580 (01)00880304282315(17)240604(10)817440 (01)00880304183001(17)240605(10)351520 (01)00880304282209(17)240605(10)529800 (01)00880304182875(17)240605(10)863370 (01)00880304182929(17)240605(10)784890 (01)00880304282292(17)240608(10)817430 (01)00880304182875(17)240609(10)128050 (01)00880304282223(17)240609(10)978840 (01)00880304684584(17)240609(10)814420 (01)00880304182929(17)240609(10)519470 (01)00880304684485(17)240610(10)120940 (01)00880304684546(17)240611(10)645200 (01)00880304182868(17)240612(10)333580 (01)00880304182967(17)240612(10)705420 (01)00880304182967(17)240613(10)705420R (01)00880304182882(17)240615(10)082440 (01)00880304684584(17)240615(10)049780 (01)00880304282278(17)240615(10)345970 (01)00880304684560(17)240615(10)645230 (01)00880304281639(17)240615(10)131430 (01)00880304182882(17)240615(10)634460 (01)00880304684485(17)240616(10)814390 (01)00880304182875(17)240616(10)178930 (01)00880304684546(17)240617(10)099330 (01)00880304684546(17)240617(10)645220 (01)00880304182875(17)240617(10)178900 (01)00880304182875(17)240618(10)178910 (01)00880304182905(17)240619(10)331250 (01)00880304182875(17)240624(10)178920 (01)00880304684546(17)240625(10)645210 (01)00880304182875(17)240625(10)228230 (01)00880304182905(17)240625(10)380060 (01)00880304182899(17)240627(10)178500 (01)00880304182899(17)240629(10)370740 (01)00880304684508(17)240630(10)705360 (01)00880304182899(17)240630(10)228320 (01)00880304281622(17)240701(10)346000 (01)00880304282223(17)240707(10)423510 (01)00880304182868(17)240707(10)389090 (01)00880304182868(17)240707(10)389100 (01)00880304182882(17)240709(10)327750 (01)00880304182936(17)240709(10)228310 (01)00880304684546(17)240710(10)142640 (01)00880304182882(17)240713(10)327730 (01)00880304182950(17)240713(10)877930 (01)00880304684546(17)240714(10)607280 (01)00880304182868(17)240715(10)389110 (01)00880304684546(17)240716(10)379070 (01)00880304182912(17)240716(10)424400 (01)00880304282223(17)240717(10)510080 (01)00880304281813(17)240720(10)504520 (01)00880304282186(17)240721(10)506020 (01)00880304684546(17)240721(10)696450 (01)00880304182875(17)240721(10)661180 (01)00880304182912(17)240721(10)454000 (01)00880304684546(17)240721(10)584760 (01)00880304182875(17)240721(10)779810 (01)00880304281646(17)240722(10)422370 (01)00880304182929(17)240723(10)167450 (01)00880304282070(17)240723(10)694050 (01)00880304281592(17)240728(10)519110 (01)00880304684546(17)240728(10)252520 (01)00880304684560(17)240728(10)645240 (01)00880304182868(17)240728(10)595950 (01)00880304182899(17)240728(10)492760 (01)00880304684560(17)240729(10)252540 (01)00880304182875(17)240730(10)595980 (01)00880304182899(17)240730(10)596010 (01)00880304282315(17)240731(10)654660 (01)00880304182875(17)240731(10)475260 (01)00880304684584(17)240731(10)049280 (01)00880304182868(17)240731(10)595960 (01)00880304182899(17)240731(10)492710 (01)00880304182899(17)240731(10)596020 (01)00880304684522(17)240803(10)142630 (01)00880304684546(17)240803(10)903560 (01)00880304684508(17)240803(10)142620 (01)00880304684560(17)240805(10)142650 (01)00880304684560(17)240805(10)252550 (01)00880304684560(17)240806(10)624760 (01)00880304684546(17)240806(10)814400 (01)00880304281622(17)240806(10)423500 (01)00880304182868(17)240806(10)595940 (01)00880304182868(17)240806(10)595970 (01)00880304182929(17)240806(10)691200 (01)00880304182899(17)240807(10)596000 (01)00880304182929(17)240807(10)691200R (01)00880304684485(17)240810(10)252500 (01)00880304684508(17)240810(10)832670 (01)00880304182929(17)240810(10)691210 (01)00880304182929(17)240810(10)811510 (01)00880304182899(17)240811(10)779820 (01)00880304182899(17)240811(10)811440 (01)00880304182899(17)240811(10)811420 (01)00880304182899(17)240811(10)811430 (01)00880304684546(17)240812(10)252530 (01)00880304182875(17)240813(10)811480 (01)00880304684546(17)240813(10)832680 (01)00880304684546(17)240813(10)832690 (01)00880304282308(17)240813(10)962750 (01)00880304282346(17)240813(10)777630 (01)00880304182899(17)240813(10)492730 (01)00880304182899(17)240813(10)779830 (01)00880304182899(17)240813(10)811400 (01)00880304182875(17)240814(10)595990 (01)00880304182875(17)240817(10)091670 (01)00880304684560(17)240819(10)798180 (01)00880304182905(17)240822(10)160830 (01)00880304684522(17)240825(10)071900 (01)00880304684508(17)240825(10)798070 (01)00880304684546(17)240825(10)798170 (01)00880304684584(17)240825(10)798100 (01)00880304182868(17)240825(10)017320 (01)00880304182875(17)240825(10)017360 (01)00880304182875(17)240825(10)017360R (01)00880304182912(17)240826(10)236490 (01)00880304684522(17)240826(10)798090 (01)00880304182912(17)240826(10)124850 (01)00880304182912(17)240826(10)124890 (01)00880304182912(17)240826(10)811490 (01)00880304684508(17)240902(10)252510 (01)00880304182899(17)240902(10)112440 (01)00880304281622(17)240902(10)122320 (01)00880304182929(17)240902(10)160780 (01)00880304684522(17)240903(10)798080 (01)00880304182875(17)240904(10)110760 (01)00880304182875(17)240904(10)110780 (01)00880304182899(17)240904(10)159130 (01)00880304182905(17)240904(10)160710 (01)00880304182912(17)240904(10)124840 (01)00880304182875(17)240904(10)110760R (01)00880304182912(17)240905(10)962760 (01)00880304182875(17)240905(10)110740 (01)00880304182905(17)240905(10)160720 (01)00880304182905(17)240905(10)160760 (01)00880304684560(17)240908(10)798030 (01)00880304182899(17)240909(10)159160 (01)00880304182899(17)240910(10)160820 (01)00880304282216(17)240911(10)983440 (01)00880304182943(17)240911(10)171340 (01)00880304182929(17)240915(10)160790 (01)00880304182905(17)240915(10)171280 (01)00880304182943(17)240915(10)171350 (01)00880304182882(17)240916(10)170990 (01)00880304282209(17)240916(10)122330 (01)00880304182868(17)240916(10)110720 (01)00880304182875(17)240916(10)110820 (01)00880304182882(17)240916(10)297750 (01)00880304182912(17)240916(10)171300 (01)00880304182929(17)240916(10)160850 (01)00880304182929(17)240916(10)171310 (01)00880304182868(17)240916(10)110720R (01)00880304182929(17)240917(10)330290 (01)00880304182936(17)240917(10)160650 (01)00880304182899(17)240917(10)171220 (01)00880304182899(17)240917(10)171230 (01)00880304182905(17)240917(10)171260 (01)00880304182899(17)240917(10)159150 (01)00880304182912(17)240917(10)124860 (01)00880304182929(17)240917(10)160770 (01)00880304182936(17)240917(10)160860 (01)00880304182875(17)240918(10)110800 (01)00880304182882(17)240918(10)297720 (01)00880304182899(17)240918(10)159170 (01)00880304182905(17)240918(10)171270 (01)00880304182912(17)240918(10)124880 (01)00880304182929(17)240918(10)171320 (01)00880304182912(17)240919(10)171290 (01)00880304182875(17)240920(10)279580 (01)00880304182905(17)240920(10)363680 (01)00880304182912(17)240920(10)124870 (01)00880304182899(17)240920(10)160810 (01)00880304182905(17)240920(10)171250 (01)00880304182905(17)240920(10)160750 (01)00880304281608(17)240920(10)069680 (01)00880304281639(17)240920(10)332550 (01)00880304182875(17)240920(10)297700 (01)00880304182882(17)240920(10)297730 (01)00880304182899(17)240920(10)159140 (01)00880304182899(17)240920(10)171240 (01)00880304182875(17)240921(10)297690 (01)00880304182868(17)240922(10)297680 (01)00880304182936(17)240922(10)314380 (01)00880304182899(17)240923(10)279590 (01)00880304282216(17)240923(10)473770 (01)00880304282254(17)240925(10)486890 (01)00880304182929(17)240925(10)525030 (01)00880304282230(17)240925(10)422160 (01)00880304182882(17)240925(10)159110 (01)00880304182936(17)240925(10)171330 (01)00880304182899(17)240926(10)453100 (01)00880304182899(17)240926(10)268210 (01)00880304282070(17)240926(10)140320 (01)00880304182899(17)240926(10)524430 (01)00880304182950(17)240926(10)314400 (01)00880304684560(17)240927(10)401380 (01)00880304282254(17)240927(10)332560 (01)00880304182929(17)240928(10)314370 (01)00880304182905(17)240929(10)500460 (01)00880304182912(17)240929(10)453220 (01)00880304182899(17)240929(10)524420 (01)00880304182882(17)240929(10)524410 (01)00880304282193(17)240930(10)434340 (01)00880304182875(17)240930(10)297710 (01)00880304182912(17)240930(10)124900 (01)00880304182875(17)241002(10)503500 (01)00880304182905(17)241002(10)519770 (01)00880304182912(17)241002(10)559770 (01)00880304182936(17)241002(10)453590 (01)00880304182912(17)241003(10)524500 (01)00880304282209(17)241003(10)434350 (01)00880304182882(17)241003(10)297760 (01)00880304182905(17)241006(10)524460 (01)00880304281592(17)241006(10)122310 (01)00880304182882(17)241006(10)297740 (01)00880304182905(17)241006(10)524470 (01)00880304182905(17)241007(10)522620 (01)00880304684522(17)241008(10)161900 (01)00880304282070(17)241008(10)431910 (01)00880304281615(17)241009(10)434320 (01)00880304182899(17)241009(10)524440 (01)00880304182912(17)241009(10)524490 (01)00880304182912(17)241010(10)589870 (01)00880304182905(17)241010(10)524450 (01)00880304182899(17)241013(10)608920 (01)00880304182905(17)241013(10)638660 (01)00880304684485(17)241013(10)266690 (01)00880304684508(17)241013(10)525060 (01)00880304684584(17)241013(10)177780 (01)00880304182912(17)241013(10)816690 (01)00880304182875(17)241013(10)588450 (01)00880304182882(17)241013(10)588570 (01)00880304182912(17)241014(10)588590 (01)00880304182905(17)241014(10)524480 (01)00880304281622(17)241014(10)555980 (01)00880304182882(17)241014(10)588510 (01)00880304182882(17)241014(10)588520 (01)00880304182882(17)241014(10)756510 (01)00880304182905(17)241014(10)572470 (01)00880304182905(17)241014(10)572480 (01)00880304182905(17)241014(10)620290 (01)00880304182912(17)241014(10)756530 (01)00880304182967(17)241014(10)620300 (01)00880304182868(17)241015(10)756490 (01)00880304182882(17)241015(10)588580 (01)00880304281622(17)241016(10)668590 (01)00880304684522(17)241016(10)524810 (01)00880304684560(17)241016(10)235700 (01)00880304684560(17)241016(10)295230 (01)00880304182912(17)241016(10)588540 (01)00880304282216(17)241020(10)653840 (01)00880304281783(17)241020(10)738820 (01)00880304684607(17)241020(10)295240 (01)00880304182875(17)241020(10)756500 (01)00880304182899(17)241021(10)525050 (01)00880304182912(17)241021(10)653890 (01)00880304281776(17)241021(10)738800 (01)00880304182899(17)241021(10)859000 (01)00880304684560(17)241021(10)506280 (01)00880304282223(17)241021(10)738790 (01)00880304182882(17)241021(10)588480 (01)00880304182875(17)241022(10)858990 (01)00880304182875(17)241023(10)763330 (01)00880304684546(17)241023(10)589270 (01)00880304182899(17)241023(10)756520 (01)00880304182882(17)241024(10)914660 (01)00880304182899(17)241024(10)897930 (01)00880304182899(17)241024(10)914670 (01)00880304182875(17)241027(10)588470 (01)00880304182899(17)241027(10)897920 (01)00880304182899(17)241028(10)948010 (01)00880304182899(17)241028(10)829300 (01)00880304684485(17)241029(10)177770 (01)00880304182875(17)241030(10)914650 (01)00880304684546(17)241031(10)589250 (01)00880304282254(17)241103(10)964370 (01)00880304182912(17)241103(10)914710 (01)00880304282254(17)241104(10)963380 (01)00880304182905(17)241105(10)154720 (01)00880304182905(17)241105(10)668750 (01)00880304182875(17)241105(10)969680 (01)00880304182882(17)241107(10)897910 (01)00880304182899(17)241107(10)914680 (01)00880304182899(17)241112(10)969700 (01)00880304182905(17)241113(10)184900 (01)00880304182912(17)241113(10)914700 (01)00880304182912(17)241113(10)588530 (01)00880304182912(17)241113(10)046110 (01)00880304182882(17)241113(10)995450 (01)00880304182899(17)241114(10)199080 (01)00880304182912(17)241114(10)068610 (01)00880304182899(17)241114(10)995480 (01)00880304684584(17)241117(10)589260 (01)00880304182905(17)241118(10)859020 (01)00880304282247(17)241118(10)963360 (01)00880304684522(17)241118(10)605200 (01)00880304281646(17)241118(10)114640 (01)00880304282322(17)241119(10)045860 (01)00880304182875(17)241119(10)158480 (01)00880304182899(17)241119(10)200210 (01)00880304182905(17)241119(10)969720 (01)00880304182905(17)241120(10)914690 (01)00880304282193(17)241120(10)009450 (01)00880304182905(17)241120(10)897940 (01)00880304282322(17)241124(10)875690 (01)00880304182905(17)241124(10)995460 (01)00880304684546(17)241125(10)994960 (01)00880304281769(17)241201(10)236020 (01)00880304684522(17)241201(10)183210 (01)00880304684560(17)241201(10)159430 (01)00880304281639(17)241201(10)104360 (01)00880304182912(17)241201(10)995490 (01)00880304684546(17)241202(10)964580 (01)00880304684522(17)241203(10)994950 (01)00880304684560(17)241204(10)104480 (01)00880304182912(17)241208(10)214080 (01)00880304009035(17)241215(10)808110 (01)00880304282230(17)241215(10)009470 (01)00880304182905(17)241216(10)299020 (01)00880304182905(17)241216(10)296380 (01)00880304182899(17)241220(10)969710 (01)00880304281721(17)241222(10)236010 (01)00880304684546(17)241229(10)118040 (01)00880304282216(17)241229(10)104370 (01)00880304182905(17)241230(10)859010 (01)00880304182905(17)250104(10)364490 (01)00880304282254(17)250105(10)236120 (01)00880304182899(17)250106(10)387380 (01)00880304182929(17)250107(10)567530 (01)00880304182905(17)250108(10)614470 (01)00880304182929(17)250108(10)567550 (01)00880304182899(17)250109(10)454500 (01)00880304182905(17)250109(10)454510 (01)00880304182912(17)250109(10)454520 (01)00880304684560(17)250110(10)118090 (01)00880304684560(17)250112(10)526630 (01)00880304009035(17)250112(10)431780 (01)00880304282216(17)250112(10)794190 (01)00880304182899(17)250112(10)628880 (01)00880304182912(17)250112(10)605080 (01)00880304281776(17)250113(10)236030 (01)00880304282322(17)250113(10)510340 (01)00880304182905(17)250113(10)506750 (01)00880304182912(17)250113(10)611270 (01)00880304281646(17)250114(10)746130 (01)00880304182929(17)250114(10)506760 (01)00880304182882(17)250114(10)567520 (01)00880304182899(17)250114(10)628890 (01)00880304182912(17)250114(10)605070 (01)00880304182905(17)250115(10)568360 (01)00880304182875(17)250120(10)628850 (01)00880304182905(17)250121(10)614490 (01)00880304182912(17)250121(10)628910 (01)00880304182882(17)250123(10)628860 (01)00880304684522(17)250124(10)452640 (01)00880304182899(17)250124(10)628870 (01)00880304009035(17)250126(10)131230 (01)00880304684584(17)250126(10)452660 (01)00880304684546(17)250127(10)601810 (01)00880304684546(17)250128(10)584110 (01)00880304182929(17)250129(10)567540 (01)00880304684607(17)250205(10)118080 (01)00880304182868(17)250209(10)800230 (01)00880304182912(17)250211(10)605090 (01)00880304684560(17)250413(10)152950 (01)00880304182905(17)250212(10)800210 (01)00880304182905(17)250212(10)800240 (01)00880304684485(17)250216(10)567600 (01)00880304684546(17)250216(10)584100 (01)00880304282254(17)250217(10)822580 (01)00880304182929(17)250217(10)801080 (01)00880304684560(17)250218(10)568110 (01)00880304182929(17)250218(10)611280 (01)00880304684522(17)250219(10)915300 (01)00880304182899(17)250226(10)628900 (01)00880304182875(17)250227(10)800200 (01)00880304684584(17)250303(10)915320 (01)00880304684546(17)250303(10)098330 (01)00880304684560(17)250303(10)915310 (01)00880304684584(17)250303(10)152910 (01)00880304684584(17)250303(10)847560 (01)00880304684508(17)250303(10)735050 (01)00880304684485(17)250304(10)915530 (01)00880304282339(17)250304(10)848460 (01)00880304282308(17)250305(10)838930 (01)00880304684546(17)250305(10)915560 (01)00880304684461(17)250306(10)032360 (01)00880304684560(17)250310(10)127380 (01)00880304684584(17)250311(10)073590 (01)00880304282360(17)250313(10)790780 (01)00880304684522(17)250319(10)238760 (01)00880304684508(17)250319(10)567620 (01)00880304684546(17)250323(10)151510 (01)00880304281677(17)250324(10)359360 (01)00880304282216(17)250326(10)939770 (01)00880304684546(17)250326(10)342400 (01)00880304182875(17)250326(10)101920 (01)00880304684607(17)250326(10)065270 (01)00880304684546(17)250327(10)089120 (01)00880304684461(17)250327(10)247440 (01)00880304684546(17)250327(10)368720 (01)00880304684522(17)250327(10)114720 (01)00880304684522(17)250327(10)151500 (01)00880304684546(17)250327(10)735070 (01)00880304282254(17)250401(10)359530 (01)00880304684546(17)250406(10)486100 (01)00880304282322(17)250406(10)157860 (01)00880304684522(17)250408(10)368700 (01)00880304684522(17)250408(10)486190 (01)00880304684546(17)250408(10)752740 (01)00880304684584(17)250408(10)349880 (01)00880304684508(17)250408(10)474000 (01)00880304684485(17)250408(10)452780 (01)00880304684546(17)250409(10)593130 (01)00880304684560(17)250409(10)368680 (01)00880304684560(17)250409(10)127390 (01)00880304684607(17)250416(10)130860 (01)00880304182929(17)250422(10)682610 (01)00880304684584(17)250423(10)474010 (01)00880304684522(17)250424(10)725750 (01)00880304684560(17)250424(10)486110 (01)00880304684560(17)250424(10)584120 (01)00880304684522(17)250424(10)486090 (01)00880304684560(17)250424(10)374470 (01)00880304684560(17)250429(10)553360 (01)00880304684546(17)250430(10)836520 (01)00880304684560(17)250430(10)763740 (01)00880304684584(17)250430(10)614800 (01)00880304684546(17)250430(10)349870 (01)00880304684584(17)250501(10)848590 (01)00880304684560(17)250504(10)848600 (01)00880304684522(17)250505(10)848390 (01)00880304182905(17)250505(10)765680 (01)00880304182929(17)250505(10)765690 (01)00880304684546(17)250511(10)942440 (01)00880304684607(17)250512(10)374480 (01)00880304684546(17)250512(10)374460 (01)00880304684584(17)250512(10)836530 (01)00880304282193(17)250512(10)657790 (01)00880304684546(17)250522(10)835350 (01)00880304684508(17)250522(10)756020 (01)00880304281721(17)250525(10)838640 (01)00880304684607(17)250526(10)835370 (01)00880304684560(17)250526(10)975260 (01)00880304684522(17)250526(10)753610 (01)00880304684607(17)250527(10)614810 (01)00880304684546(17)250527(10)614790 (01)00880304684560(17)250527(10)756070 (01)00880304684560(17)250602(10)756050 (01)00880304684485(17)250603(10)666330 (01)00880304684607(17)250604(10)055560 (01)00880304684621(17)250608(10)835380 (01)00880304684560(17)250608(10)127660 (01)00880304684508(17)250609(10)021160 (01)00880304684461(17)250609(10)067590 (01)00880304684522(17)250609(10)067580 (01)00880304684522(17)250611(10)166170 (01)00880304684546(17)250611(10)070370 (01)00880304684546(17)250611(10)284830 (01)00880304684560(17)250611(10)957790 (01)00880304684485(17)250611(10)067610 (01)00880304684546(17)250611(10)287000 (01)00880304281806(17)250611(10)190860 (01)00880304684546(17)250611(10)835360 (01)00880304684522(17)250611(10)067570 (01)00880304684607(17)250612(10)071010 (01)00880304684508(17)250612(10)123590 (01)00880304182899(17)250612(10)163480 (01)00880304182899(17)250612(10)165680 (01)00880304281813(17)250615(10)190870 (01)00880304282209(17)250617(10)163340 (01)00880304282216(17)250618(10)163350 (01)00880304282322(17)250619(10)232280 (01)00880304281813(17)250630(10)359620 (01)00880304684485(17)250701(10)202670 (01)00880304282247(17)250705(10)430330 (01)00880304684546(17)250710(10)453460 (01)00880304282247(17)250715(10)547820 (01)00880304281790(17)250723(10)757350 (01)00880304282360(17)250723(10)759650 (01)00880304282346(17)250723(10)759640 (01)00880304281769(17)250805(10)613900 (01)00880304684584(17)250806(10)218480 (01)00880304684607(17)250810(10)303760 (01)00880304684546(17)250817(10)398550 (01)00880304182899(17)250817(10)816910 (01)00880304281776(17)250818(10)777530 (01)00880304182905(17)250818(10)816930 (01)00880304684584(17)250819(10)459720 (01)00880304684607(17)250819(10)627270 (01)00880304182899(17)250819(10)777860 (01)00880304182899(17)250819(10)816900 (01)00880304182905(17)250819(10)777870 (01)00880304282193(17)250820(10)613910 (01)00880304684522(17)250822(10)553700 (01)00880304684546(17)250822(10)415390 (01)00880304684546(17)250822(10)778850 (01)00880304684560(17)250823(10)653180 (01)00880304684485(17)250824(10)433060 (01)00880304684584(17)250824(10)778870 (01)00880304182899(17)250824(10)777850 (01)00880304282193(17)250828(10)430320 (01)00880304684546(17)250902(10)398560 (01)00880304684508(17)250910(10)049730 (01)00880304684508(17)250914(10)505470 (01)00880304684485(17)250916(10)398540 (01)00880304684546(17)250916(10)050260 (01)00880304684522(17)250916(10)778840 (01)00880304684560(17)250917(10)778860 (01)00880304684607(17)250917(10)171590 (01)00880304684546(17)250917(10)398570 (01)00880304003293(17)250918(10)181350 (01)00880304282230(17)250918(10)613920 (01)00880304684485(17)250923(10)047970 (01)00880304684607(17)250923(10)153090 (01)00880304282209(17)251005(10)398790 (01)00880304684508(17)251008(10)924390 (01)00880304684546(17)251008(10)307740 (01)00880304684508(17)251012(10)363750 (01)00880304684560(17)251013(10)363770 (01)00880304684560(17)251014(10)390600 (01)00880304684485(17)251016(10)778830 (01)00880304684560(17)251026(10)525110 (01)00880304684508(17)251103(10)401120 (01)00880304684546(17)251104(10)401110 (01)00880304684560(17)251105(10)655830 (01)00880304684447(17)251105(10)390580 (01)00880304282216(17)251105(10)436670 (01)00880304684560(17)251106(10)409350 (01)00880304684508(17)251107(10)409280 (01)00880304684522(17)260118(10)476070 (01)00880304684522(17)260209(10)831470 (01)00880304182899(17)251130(10)812040 (01)00880304684485(17)251209(10)757470 (01)00880304684522(17)251209(10)409300 (01)00880304684607(17)251209(10)260310 (01)00880304684461(17)251209(10)476050 (01)00880304684508(17)251209(10)745810 (01)00880304684522(17)251209(10)409310 (01)00880304684584(17)251209(10)354510 (01)00880304684546(17)251209(10)354480 (01)00880304684584(17)251209(10)409360 (01)00880304684546(17)251215(10)409320 (01)00880304684508(17)251216(10)409290 (01)00880304684522(17)251216(10)878900 (01)00880304684522(17)251216(10)845900 (01)00880304684560(17)251216(10)937060 (01)00880304684522(17)251216(10)648910 (01)00880304684560(17)251217(10)799370 (01)00880304684461(17)251218(10)727860 (01)00880304684522(17)251221(10)354470 (01)00880304282322(17)251221(10)893930 (01)00880304684508(17)251221(10)799320 (01)00880304282223(17)251221(10)755960 (01)00880304684485(17)251222(10)756610 (01)00880304684560(17)251222(10)409340 (01)00880304684584(17)251222(10)409370 (01)00880304684508(17)251222(10)799330 (01)00880304684508(17)251222(10)727850 (01)00880304182905(17)251228(10)228860 (01)00880304282315(17)260105(10)620350 (01)00880304282209(17)260106(10)184360 (01)00880304182899(17)260111(10)962740 (01)00880304182905(17)260111(10)228870 (01)00880304182936(17)260112(10)962750 (01)00880304182899(17)260113(10)239360 (01)00880304182899(17)260113(10)973750 (01)00880304684546(17)260114(10)457970 (01)00880304684560(17)260114(10)757160 (01)00880304684607(17)260114(10)837680 (01)00880304684621(17)260114(10)389070 (01)00880304684546(17)260114(10)343160 (01)00880304684560(17)260118(10)142100 (01)00880304684485(17)260120(10)081520 (01)00880304684584(17)260120(10)908380 (01)00880304282322(17)260121(10)378450 (01)00880304684546(17)260121(10)837660 (01)00880304282322(17)260125(10)549610 (01)00880304684546(17)260125(10)294830 (01)00880304182899(17)260125(10)369300 (01)00880304684461(17)260125(10)496310 (01)00880304684508(17)260125(10)973670 (01)00880304684607(17)260125(10)253900 (01)00880304182929(17)260125(10)321180 (01)00880304684485(17)260126(10)476060 (01)00880304684508(17)260126(10)368300 (01)00880304684522(17)260126(10)398030 (01)00880304684546(17)260126(10)441980 (01)00880304684560(17)260126(10)186780 (01)00880304684607(17)260126(10)186790 (01)00880304182936(17)260126(10)239230 (01)00880304182905(17)260126(10)546430 (01)00880304684584(17)260127(10)241290 (01)00880304684560(17)260127(10)130480 (01)00880304684560(17)260127(10)038360 (01)00880304684607(17)260127(10)361640 (01)00880304182899(17)260127(10)527290 (01)00880304684584(17)260128(10)280230 (01)00880304684508(17)260128(10)038330 (01)00880304684584(17)260128(10)817590 (01)00880304684485(17)260201(10)170230 (01)00880304684485(17)260201(10)958190 (01)00880304684522(17)260201(10)831440 (01)00880304684546(17)260201(10)278520 (01)00880304684584(17)260201(10)399650 (01)00880304684560(17)260201(10)799360 (01)00880304182950(17)260201(10)620580 (01)00880304684546(17)260201(10)361610 (01)00880304684546(17)260201(10)038350 (01)00880304182950(17)260201(10)321190 (01)00880304684546(17)260201(10)409330 (01)00880304684584(17)260202(10)450200 (01)00880304182974(17)260202(10)596240 (01)00880304282322(17)260202(10)315850 (01)00880304182929(17)260202(10)593310 (01)00880304282193(17)260202(10)527670 (01)00880304684584(17)260203(10)799380 (01)00880304282315(17)260203(10)546880 (01)00880304684485(17)260204(10)038320 (01)00880304684522(17)260204(10)870670 (01)00880304282223(17)260204(10)769120 (01)00880304282339(17)260204(10)546890 (01)00880304182899(17)260204(10)593280 (01)00880304182905(17)260204(10)633460 (01)00880304684584(17)260207(10)361620 (01)00880304684546(17)260209(10)571230 (01)00880304684546(17)260209(10)720310 (01)00880304182912(17)260209(10)807360 (01)00880304684522(17)260209(10)343150 (01)00880304684522(17)260209(10)672530 (01)00880304182912(17)260209(10)593300 (01)00880304182899(17)260210(10)683530 (01)00880304182905(17)260210(10)633450 (01)00880304182912(17)260210(10)753870 (01)00880304182905(17)260210(10)753700 (01)00880304182912(17)260210(10)676310 (01)00880304182929(17)260210(10)753880 (01)00880304182943(17)260210(10)684900 (01)00880304182905(17)260211(10)830790 (01)00880304182899(17)260211(10)593290 (01)00880304182905(17)260211(10)724660 (01)00880304182899(17)260211(10)724640 (01)00880304182929(17)260211(10)704400 (01)00880304684560(17)260212(10)814850 (01)00880304182899(17)260215(10)827090 (01)00880304182905(17)260215(10)869750 (01)00880304182912(17)260215(10)899980 (01)00880304282230(17)260215(10)875960 (01)00880304182899(17)260215(10)753860 (01)00880304684508(17)260216(10)888320 (01)00880304684508(17)260217(10)899030 (01)00880304684546(17)260217(10)850230 (01)00880304684485(17)260217(10)343140 (01)00880304182899(17)260217(10)968510 (01)00880304182929(17)260217(10)963230 (01)00880304684560(17)260218(10)751920 (01)00880304684560(17)260218(10)685220 (01)00880304684560(17)260218(10)112290 (01)00880304282360(17)260218(10)969950 (01)00880304684508(17)260219(10)038340 (01)00880304182899(17)260219(10)633430 (01)00880304182905(17)260219(10)633440 (01)00880304182912(17)260221(10)963220 (01)00880304684546(17)260222(10)112280 (01)00880304182912(17)260224(10)650720 (01)00880304182899(17)260224(10)963210 (01)00880304182943(17)260226(10)147950 (01)00880304182905(17)260226(10)724650 (01)00880304684621(17)260228(10)038110 (01)00880304182936(17)260302(10)724670 (01)00880304684485(17)260303(10)996000 (01)00880304182905(17)260307(10)042820 (01)00880304684522(17)260307(10)685210 (01)00880304282292(17)260309(10)011950 (01)00880304282292(17)260309(10)845260 (01)00880304182943(17)260310(10)190130 (01)00880304182899(17)260311(10)085590 (01)00880304182912(17)260314(10)097690 (01)00880304684560(17)260316(10)401430 (01)00880304282377(17)260316(10)820100 (01)00880304684508(17)260323(10)943930 (01)00880304182912(17)260324(10)188280 (01)00880304182929(17)260324(10)225740 (01)00880304182943(17)260325(10)173140 (01)00880304182899(17)260329(10)172630 (01)00880304182936(17)260329(10)301920 (01)00880304182868(17)260329(10)267670 (01)00880304182936(17)260330(10)225760 (01)00880304684485(17)260331(10)998290 (01)00880304684546(17)260331(10)381030 (01)00880304684522(17)260331(10)011480 (01)00880304684584(17)260401(10)799390 (01)00880304684621(17)260401(10)773990 (01)00880304282308(17)260402(10)211780 (01)00880304684560(17)260404(10)159300 (01)00880304182899(17)260404(10)827110 (01)00880304182905(17)260405(10)173240 (01)00880304182905(17)260405(10)293710 (01)00880304182912(17)260405(10)956210 (01)00880304684584(17)260407(10)056510 (01)00880304182905(17)260408(10)225730 (01)00880304182950(17)260413(10)287350 (01)00880304182912(17)260414(10)584080 (01)00880304182912(17)260414(10)483270 (01)00880304684546(17)260414(10)723780 (01)00880304182929(17)260414(10)287340 (01)00880304684560(17)260415(10)472590 (01)00880304182899(17)260415(10)585920 (01)00880304182875(17)260415(10)527340 (01)00880304182899(17)260415(10)267680 (01)00880304684522(17)260418(10)339410 (01)00880304684546(17)260418(10)154630 (01)00880304684546(17)260418(10)354490 (01)00880304684546(17)260418(10)154620 (01)00880304684522(17)260418(10)347210 (01)00880304182912(17)260418(10)267690 (01)00880304182936(17)260418(10)483280 (01)00880304282346(17)260418(10)211790 (01)00880304182905(17)260418(10)483260 (01)00880304684560(17)260419(10)150360 (01)00880304684560(17)260419(10)378790 (01)00880304684584(17)260419(10)241030 (01)00880304684607(17)260419(10)139220 (01)00880304182882(17)260419(10)503080 (01)00880304182882(17)260419(10)503100 (01)00880304182882(17)260419(10)527390 (01)00880304182882(17)260419(10)527400 (01)00880304684560(17)260420(10)347200 (01)00880304684584(17)260420(10)139240 (01)00880304684522(17)260421(10)181860 (01)00880304182912(17)260421(10)585340 (01)00880304182899(17)260421(10)503130 (01)00880304182912(17)260421(10)503160 (01)00880304684546(17)260421(10)257190 (01)00880304684546(17)260421(10)257180 (01)00880304182875(17)260421(10)527350 (01)00880304182905(17)260421(10)577950 (01)00880304684546(17)260422(10)447770 (01)00880304684546(17)260422(10)532640 (01)00880304684546(17)260425(10)540690 (01)00880304684560(17)260425(10)154640 (01)00880304684560(17)260425(10)548450 (01)00880304182875(17)260425(10)527370 (01)00880304182905(17)260425(10)577960 (01)00880304182905(17)260425(10)577970 (01)00880304182905(17)260426(10)577940 (01)00880304684546(17)260426(10)347240 (01)00880304182899(17)260426(10)503110 (01)00880304182899(17)260426(10)503120 (01)00880304182912(17)260426(10)503150 (01)00880304182950(17)260426(10)527430 (01)00880304684546(17)260428(10)688620 (01)00880304182882(17)260428(10)763190 (01)00880304182899(17)260428(10)601560 (01)00880304182905(17)260428(10)572580 (01)00880304182899(17)260428(10)706370 (01)00880304182905(17)260428(10)647900 (01)00880304182905(17)260428(10)577930 (01)00880304182912(17)260428(10)577980 (01)00880304182912(17)260428(10)706380 (01)00880304182967(17)260429(10)749770 (01)00880304182868(17)260429(10)670150 (01)00880304182868(17)260502(10)670160 (01)00880304182875(17)260502(10)670180 (01)00880304182899(17)260503(10)782320 (01)00880304182882(17)260503(10)670190 (01)00880304684522(17)260503(10)139210 (01)00880304182882(17)260503(10)527410 (01)00880304182967(17)260503(10)656810 (01)00880304182882(17)260503(10)670200 (01)00880304182912(17)260503(10)706390 (01)00880304182967(17)260503(10)656820 (01)00880304182905(17)260506(10)287330 (01)00880304182899(17)260509(10)854740 (01)00880304684546(17)260509(10)171070 (01)00880304182868(17)260509(10)801900 (01)00880304684508(17)260509(10)139230 (01)00880304684508(17)260509(10)347230 (01)00880304684508(17)260510(10)532130 (01)00880304684522(17)260510(10)855690 (01)00880304684560(17)260510(10)855860 (01)00880304684584(17)260510(10)898980 (01)00880304684560(17)260511(10)347220 (01)00880304182912(17)260511(10)792170 (01)00880304182905(17)260512(10)969280 (01)00880304182875(17)260512(10)969300 (01)00880304182882(17)260512(10)803570 (01)00880304182899(17)260512(10)706360 (01)00880304182912(17)260513(10)503140 (01)00880304182905(17)260513(10)792190 (01)00880304182899(17)260513(10)792180 (01)00880304182899(17)260516(10)763160 (01)00880304182875(17)260516(10)670170 (01)00880304182912(17)260516(10)672270 (01)00880304684546(17)260517(10)981880 (01)00880304684546(17)260518(10)293680 (01)00880304684560(17)260518(10)899000 (01)00880304684560(17)260518(10)347190 (01)00880304182943(17)260519(10)072080 (01)00880304684560(17)260521(10)864270 (01)00880304182943(17)260521(10)807780 (01)00880304182882(17)260521(10)503090 (01)00880304182882(17)260521(10)981350 (01)00880304182905(17)260521(10)807750 (01)00880304182912(17)260521(10)577990 (01)00880304182912(17)260521(10)807770 (01)00880304182912(17)260524(10)741430 (01)00880304684560(17)260524(10)896730 (01)00880304182868(17)260601(10)096720 (01)00880304182875(17)260601(10)085550 (01)00880304182899(17)260601(10)807740 (01)00880304182912(17)260601(10)741390 (01)00880304182868(17)260608(10)238190 (01)00880304182868(17)260608(10)097250 (01)00880304182912(17)260608(10)097430 (01)00880304182912(17)260608(10)198780 (01)00880304182912(17)260608(10)300830 (01)00880304182875(17)260608(10)085560 (01)00880304182905(17)260608(10)085600 (01)00880304182875(17)260608(10)085560R (01)00880304182882(17)260609(10)270210 (01)00880304182882(17)260609(10)981340 (01)00880304684522(17)260609(10)248390 (01)00880304684522(17)260609(10)864260 (01)00880304684447(17)260609(10)212210 (01)00880304684560(17)260610(10)133120 (01)00880304182905(17)260610(10)085580 (01)00880304684461(17)260612(10)216050 (01)00880304684522(17)260613(10)097590 (01)00880304182936(17)260616(10)262790 (01)00880304684508(17)260617(10)097580 (01)00880304182868(17)260621(10)241710 (01)00880304182912(17)260621(10)241730 (01)00880304684508(17)260621(10)097570 (01)00880304684546(17)260621(10)248410 (01)00880304684560(17)260623(10)097610 (01)00880304684461(17)260628(10)212220 (01)00880304282339(17)260628(10)490140 (01)00880304282292(17)260630(10)524000 (01)00880304182905(17)260701(10)085590 (01)00880304182905(17)260701(10)085590R (01)00880304684584(17)260708(10)248460 (01)00880304684461(17)260713(10)622370 (01)00880304182882(17)260713(10)607440 (01)00880304684560(17)260714(10)248440 (01)00880304684485(17)260715(10)440690 (01)00880304684461(17)260715(10)557380 (01)00880304684485(17)260715(10)622390 (01)00880304684485(17)260715(10)212230 (01)00880304684508(17)260716(10)622420 (01)00880304684522(17)260716(10)557420 (01)00880304182882(17)260716(10)470790 (01)00880304182882(17)260716(10)470790R (01)00880304281790(17)260718(10)604600 (01)00880304282315(17)260718(10)490130 (01)00880304281790(17)260718(10)604600R (01)00880304684485(17)260719(10)440680 (01)00880304684621(17)260720(10)578620 (01)00880304182882(17)260724(10)470800 (01)00880304684508(17)260726(10)574630 (01)00880304282315(17)260727(10)697320 (01)00880304684621(17)260727(10)291390 (01)00880304684546(17)260728(10)248400 (01)00880304182905(17)260728(10)792200 (01)00880304182899(17)260728(10)291080 (01)00880304684546(17)260730(10)607570 (01)00880304684508(17)260730(10)574640 (01)00880304684522(17)260730(10)557410 (01)00880304182905(17)260801(10)743130 (01)00880304282322(17)260801(10)348810 (01)00880304282322(17)260801(10)738840 (01)00880304684560(17)260802(10)622690 (01)00880304684546(17)260802(10)097600 (01)00880304684584(17)260803(10)622710 (01)00880304684621(17)260805(10)758790 (01)00880304684560(17)260805(10)607590 (01)00880304182899(17)260808(10)607450 (01)00880304684485(17)260808(10)097560 (01)00880304684584(17)260808(10)938680 (01)00880304182912(17)260808(10)823770 (01)00880304182899(17)260809(10)801680 (01)00880304281783(17)260810(10)604590 (01)00880304182882(17)260810(10)490110 (01)00880304182912(17)260810(10)782560 (01)00880304182912(17)260810(10)782560R (01)00880304182875(17)260811(10)782510 (01)00880304684607(17)260811(10)700980 (01)00880304182882(17)260811(10)490100 (01)00880304182882(17)260811(10)801650 (01)00880304182899(17)260811(10)801720 (01)00880304182912(17)260811(10)935050 (01)00880304182875(17)260811(10)085570 (01)00880304182875(17)260811(10)868230 (01)00880304182899(17)260811(10)868250 (01)00880304182875(17)260811(10)949180 (01)00880304182882(17)260812(10)968970 (01)00880304182882(17)260812(10)988610 (01)00880304182912(17)260812(10)823750 (01)00880304282308(17)260812(10)943510 (01)00880304182875(17)260812(10)782500 (01)00880304182899(17)260812(10)291120 (01)00880304182899(17)260812(10)801670 (01)00880304182899(17)260812(10)801690 (01)00880304182912(17)260812(10)782570 (01)00880304182912(17)260812(10)868270 (01)00880304182882(17)260812(10)988610R (01)00880304182882(17)260813(10)801660 (01)00880304182882(17)260813(10)872460 (01)00880304182882(17)260813(10)935040 (01)00880304182899(17)260813(10)021700 (01)00880304182899(17)260813(10)892270 (01)00880304182912(17)260813(10)868280 (01)00880304182912(17)260813(10)951020 (01)00880304182882(17)260814(10)801640 (01)00880304182882(17)260814(10)980620 (01)00880304182882(17)260814(10)980620R (01)00880304684485(17)260815(10)974130 (01)00880304684546(17)260815(10)868350 (01)00880304182899(17)260815(10)592940 (01)00880304684447(17)260815(10)840540 (01)00880304684485(17)260815(10)840570 (01)00880304684485(17)260815(10)840580 (01)00880304684522(17)260815(10)622670 (01)00880304182882(17)260816(10)921150 (01)00880304182882(17)260816(10)980660 (01)00880304684584(17)260817(10)017730 (01)00880304684560(17)260817(10)622700 (01)00880304684560(17)260817(10)840620 (01)00880304684522(17)260818(10)557390 (01)00880304684522(17)260818(10)622660 (01)00880304684546(17)260818(10)607560 (01)00880304182899(17)260818(10)868260 (01)00880304684621(17)260819(10)951940 (01)00880304182912(17)260819(10)951030 (01)00880304182943(17)260819(10)007140 (01)00880304182950(17)260819(10)801730 (01)00880304182905(17)260820(10)972590 (01)00880304182950(17)260820(10)258510 (01)00880304182936(17)260820(10)007120 (01)00880304182912(17)260820(10)823760 (01)00880304182905(17)260821(10)782540 (01)00880304182882(17)260822(10)004370 (01)00880304182882(17)260822(10)007110 (01)00880304182912(17)260823(10)038580 (01)00880304684607(17)260824(10)700990 (01)00880304182875(17)260825(10)007090 (01)00880304684508(17)260825(10)622410 (01)00880304182950(17)260826(10)776570 (01)00880304182882(17)260826(10)990510 (01)00880304182882(17)260826(10)007100 (01)00880304182875(17)260826(10)868240 (01)00880304182936(17)260828(10)132750 (01)00880304684447(17)260829(10)622360 (01)00880304684485(17)260829(10)933330 (01)00880304182899(17)260829(10)154840 (01)00880304182882(17)260829(10)990500 (01)00880304182875(17)260829(10)990470 (01)00880304182899(17)260829(10)291110 (01)00880304182899(17)260830(10)062150 (01)00880304182899(17)260830(10)955620 (01)00880304182882(17)260830(10)172310 (01)00880304684508(17)260830(10)938540 (01)00880304182875(17)260830(10)990490 (01)00880304182882(17)260830(10)990520 (01)00880304182936(17)260830(10)007130 (01)00880304684621(17)260831(10)951850 (01)00880304182882(17)260831(10)021630 (01)00880304182882(17)260831(10)101940 (01)00880304182882(17)260831(10)178940 (01)00880304182882(17)260831(10)076380 (01)00880304182905(17)260831(10)103370 (01)00880304182936(17)260831(10)097440 (01)00880304684560(17)260831(10)840630 (01)00880304684560(17)260831(10)938510 (01)00880304182875(17)260831(10)943340 (01)00880304182899(17)260831(10)291100 (01)00880304182875(17)260901(10)253100 (01)00880304182899(17)260901(10)041270 (01)00880304182882(17)260901(10)159660 (01)00880304182936(17)260901(10)146640 (01)00880304684560(17)260901(10)943940 (01)00880304182875(17)260901(10)269940 (01)00880304182875(17)260901(10)782520 (01)00880304182882(17)260901(10)152940 (01)00880304182905(17)260901(10)221670 (01)00880304182882(17)260901(10)980650 (01)00880304182905(17)260901(10)235740 (01)00880304182875(17)260901(10)990480 (01)00880304182929(17)260901(10)089310 (01)00880304182875(17)260902(10)144490 (01)00880304182905(17)260902(10)325240 (01)00880304182875(17)260902(10)782530 (01)00880304182905(17)260902(10)235780 (01)00880304182929(17)260902(10)302000 (01)00880304182943(17)260902(10)868290 (01)00880304182950(17)260902(10)801740 (01)00880304182905(17)260902(10)235780R (01)00880304182905(17)260905(10)235730 (01)00880304182905(17)260905(10)235790 (01)00880304182929(17)260905(10)301990 (01)00880304182905(17)260906(10)235750 (01)00880304182929(17)260906(10)251420 (01)00880304182936(17)260906(10)251440 (01)00880304182905(17)260912(10)235720 (01)00880304684522(17)260915(10)059750 (01)00880304684546(17)260915(10)336280 (01)00880304182899(17)260916(10)267110 (01)00880304182905(17)260916(10)381510 (01)00880304182929(17)260919(10)251430 (01)00880304182868(17)260919(10)323360 (01)00880304182868(17)260919(10)323370 (01)00880304182943(17)260919(10)302010 (01)00880304684522(17)260920(10)129680 (01)00880304182868(17)260920(10)418440 (01)00880304182929(17)260920(10)474360 (01)00880304182936(17)260920(10)474120 (01)00880304182905(17)260920(10)235710 (01)00880304182905(17)260920(10)381550 (01)00880304182929(17)260920(10)089320 (01)00880304182868(17)260920(10)323350 (01)00880304182936(17)260920(10)251450 (01)00880304684584(17)260920(10)938480 (01)00880304182899(17)260920(10)395800 (01)00880304182905(17)260920(10)221680 (01)00880304182905(17)260920(10)235770 (01)00880304182905(17)260920(10)301980 (01)00880304684560(17)260921(10)143190 (01)00880304684560(17)260921(10)840610 (01)00880304684508(17)260926(10)938490 (01)00880304684522(17)260926(10)622680 (01)00880304684607(17)260928(10)222020 (01)00880304684560(17)260929(10)840600 (01)00880304182899(17)260929(10)395770 (01)00880304182899(17)261005(10)395760 (01)00880304182875(17)261029(10)553320 (01)00880304182936(17)261029(10)413770 (01)00880304182899(17)261030(10)585780 (01)00880304182905(17)261031(10)221700 (01)00880304182912(17)261031(10)501930 (01)00880304282292(17)261031(10)089340 (01)00880304182868(17)261101(10)323380 (01)00880304182943(17)261101(10)302020 (01)00880304182899(17)261101(10)413740 (01)00880304182868(17)261102(10)546070 (01)00880304182899(17)261102(10)395810 (01)00880304182912(17)261102(10)501890 (01)00880304182912(17)261102(10)501910 (01)00880304182905(17)261103(10)429230 (01)00880304182912(17)261103(10)613510 (01)00880304182899(17)261103(10)600760 (01)00880304182905(17)261103(10)221690 (01)00880304182936(17)261103(10)613240 (01)00880304182868(17)261103(10)728050 (01)00880304182899(17)261104(10)413760 (01)00880304684546(17)261104(10)432180 (01)00880304182882(17)261104(10)655550 (01)00880304182899(17)261104(10)705680 (01)00880304182912(17)261104(10)501870 (01)00880304182905(17)261105(10)301970 (01)00880304182899(17)261105(10)413730 (01)00880304182929(17)261106(10)728080 (01)00880304684522(17)261107(10)343000 (01)00880304684546(17)261107(10)624600 (01)00880304182905(17)261107(10)671740 (01)00880304282322(17)261107(10)415410 (01)00880304282322(17)261107(10)579130 (01)00880304282346(17)261107(10)415420 (01)00880304182882(17)261107(10)655540 (01)00880304182882(17)261107(10)694280 (01)00880304182912(17)261107(10)501880 (01)00880304182929(17)261107(10)721510 (01)00880304182929(17)261107(10)721570 (01)00880304182875(17)261108(10)694670 (01)00880304182912(17)261108(10)722630 (01)00880304182905(17)261108(10)671720 (01)00880304182912(17)261108(10)501940 (01)00880304182882(17)261108(10)694240 (01)00880304182905(17)261108(10)671730 (01)00880304182905(17)261108(10)694350 (01)00880304182950(17)261108(10)703800 (01)00880304182929(17)261109(10)490240 (01)00880304182899(17)261109(10)395780 (01)00880304182899(17)261111(10)600860 (01)00880304182882(17)261111(10)694270 (01)00880304182882(17)261112(10)655570 (01)00880304684508(17)261115(10)381680 (01)00880304684546(17)261115(10)432170 (01)00880304684522(17)261115(10)246830 (01)00880304182882(17)261116(10)980630 (01)00880304182950(17)261117(10)703790 (01)00880304182929(17)261117(10)735350 (01)00880304182950(17)261118(10)626010 (01)00880304182905(17)261119(10)782550 (01)00880304182868(17)261119(10)323390 (01)00880304282353(17)261123(10)579140 (01)00880304182905(17)261125(10)988150 (01)00880304182950(17)261130(10)703780 (01)00880304182899(17)261201(10)413750 (01)00880304182912(17)261201(10)501950 (01)00880304182967(17)261205(10)566110 (01)00880304182912(17)261207(10)501900 (01)00880304282346(17)261209(10)960730 (01)00880304182905(17)261209(10)187060 (01)00880304182875(17)261210(10)008390 (01)00880304182912(17)261210(10)187230 (01)00880304182875(17)261212(10)899600 (01)00880304182912(17)261212(10)187210 (01)00880304182882(17)261213(10)901950 (01)00880304182875(17)261213(10)008300 (01)00880304182875(17)261213(10)008340 (01)00880304182875(17)261213(10)267570 (01)00880304182882(17)261213(10)901980 (01)00880304182875(17)261213(10)899590 (01)00880304182912(17)261213(10)081200 (01)00880304182912(17)261213(10)081200R (01)00880304182905(17)261214(10)694330 (01)00880304182899(17)261216(10)998720 (01)00880304182943(17)261218(10)929640 (01)00880304182868(17)261218(10)323340 (01)00880304182905(17)261220(10)988080 (01)00880304182912(17)261220(10)880200 (01)00880304182882(17)261221(10)266150 (01)00880304182905(17)261221(10)998780 (01)00880304182912(17)261221(10)501960 (01)00880304182875(17)261227(10)257820 (01)00880304182882(17)261227(10)901960 (01)00880304182905(17)261228(10)998800 (01)00880304182882(17)261228(10)167450 (01)00880304182899(17)261229(10)391910 (01)00880304684508(17)261229(10)432150 (01)00880304182905(17)261229(10)267470 (01)00880304182882(17)261229(10)915550 (01)00880304182912(17)261230(10)281220 (01)00880304684584(17)261230(10)248450 (01)00880304684584(17)261230(10)097620 (01)00880304182875(17)261230(10)248240 (01)00880304282315(17)261230(10)579120 (01)00880304182882(17)270102(10)267700 (01)00880304182912(17)270103(10)081170 (01)00880304182912(17)270109(10)081130 (01)00880304182882(17)270109(10)902020 (01)00880304182929(17)270110(10)735370 (01)00880304182899(17)270111(10)150780 (01)00880304182899(17)270112(10)221340 (01)00880304182882(17)270114(10)915480 (01)00880304182882(17)270115(10)915590 (01)00880304182967(17)270115(10)865150 (01)00880304182912(17)270115(10)187170 (01)00880304182950(17)270115(10)894420 (01)00880304182912(17)270115(10)880630 (01)00880304182943(17)270115(10)929630 (01)00880304182868(17)270116(10)728090 (01)00880304182912(17)270116(10)187200 (01)00880304182936(17)270116(10)703770 (01)00880304182905(17)270116(10)671760 (01)00880304182899(17)270116(10)221300 (01)00880304182882(17)270126(10)739460 (01)00880304182899(17)270126(10)998760 (01)00880304182905(17)270126(10)187280 (01)00880304182905(17)270126(10)988140 (01)00880304182875(17)270126(10)267490 (01)00880304182899(17)270127(10)150710 (01)00880304282339(17)270128(10)750420 (01)00880304182899(17)270201(10)395790 (01)00880304182882(17)270203(10)484840 (01)00880304182899(17)270203(10)834940 (01)00880304182936(17)270203(10)929620 (01)00880304684546(17)270203(10)418740 (01)00880304182899(17)270203(10)221320 (01)00880304182905(17)270203(10)694320 (01)00880304684508(17)270204(10)444000 (01)00880304684508(17)270205(10)463730 (01)00880304182905(17)270206(10)557730 (01)00880304182912(17)270206(10)262710 (01)00880304182882(17)270206(10)694260 (01)00880304684560(17)270210(10)463740 (01)00880304684584(17)270214(10)418760 (01)00880304182899(17)270217(10)221330 (01)00880304684546(17)270301(10)943380 (01)00880304182905(17)270306(10)988120 (01)00880304684546(17)270310(10)859730 (01)00880304182875(17)270311(10)899680 (01)00880304684461(17)270312(10)036890 (01)00880304182912(17)270316(10)081140 (01)00880304684584(17)270320(10)943390 (01)00880304684560(17)270323(10)610540 (01)00880304684607(17)270324(10)866250 (01)00880304182936(17)270325(10)468730 (01)00880304182905(17)270327(10)988110 (01)00880304182882(17)270327(10)167460 (01)00880304282315(17)270407(10)625630 (01)00880304282292(17)270407(10)295340 (01)00880304282230(17)270407(10)224210 (01)00880304282308(17)270409(10)295350 (01)00880304684560(17)270410(10)326920 (01)00880304282322(17)270412(10)393360 (01)00880304282193(17)270416(10)491490 (01)00880304684546(17)270417(10)325980 (01)00880304282346(17)270423(10)295370 (01)00880304684546(17)270425(10)325990 (01)00880304684546(17)270429(10)488620 (01)00880304684560(17)270430(10)326890 (01)00880304684584(17)270505(10)408640 (01)00880304684560(17)270511(10)506110 (01)00880304684508(17)270513(10)326880 (01)00880304282315(17)270519(10)344960 (01)00880304182899(17)270520(10)897030 (01)00880304684522(17)270524(10)488610 (01)00880304684522(17)270525(10)866240 (01)00880304684522(17)270525(10)866230 (01)00880304282353(17)270526(10)295380 (01)00880304182875(17)270607(10)052120 (01)00880304684584(17)270616(10)545700 (01)00880304182899(17)270616(10)173340 (01)00880304182875(17)270617(10)052110 (01)00880304182936(17)270617(10)052200 (01)00880304182905(17)270617(10)871100 (01)00880304282322(17)270619(10)295360 (01)00880304281615(17)270620(10)371020 (01)00880304182868(17)270622(10)173120 (01)00880304182899(17)270623(10)173220 (01)00880304684546(17)270701(10)492580 (01)00880304684607(17)270706(10)488630 (01)00880304684560(17)270706(10)492610 (01)00880304282377(17)270706(10)344950 (01)00880304684560(17)270717(10)470430 (01)00880304684560(17)270717(10)492630 (01)00880304282292(17)270720(10)960700 (01)00880304182905(17)270720(10)303200 (01)00880304182912(17)270721(10)880680 (01)00880304182882(17)270723(10)028410 (01)00880304282360(17)270724(10)176550 (01)00880304182929(17)270724(10)735360 (01)00880304684546(17)270725(10)492590 (01)00880304182899(17)270727(10)699330 (01)00880304684560(17)270728(10)326910 (01)00880304182868(17)270728(10)173140 (01)00880304182875(17)270729(10)068900 (01)00880304282193(17)270731(10)881830 (01)00880304684485(17)270802(10)506060 (01)00880304281776(17)270802(10)171030 (01)00880304282308(17)270802(10)176570 (01)00880304282339(17)270802(10)833710 (01)00880304182875(17)270802(10)052100 (01)00880304684522(17)270803(10)593510 (01)00880304182912(17)270805(10)303210 (01)00880304182905(17)270805(10)258470 (01)00880304182899(17)270806(10)258450 (01)00880304281738(17)270807(10)823470 (01)00880304281769(17)270807(10)171020 (01)00880304281783(17)270807(10)171040 (01)00880304182905(17)270807(10)212610 (01)00880304182882(17)270807(10)028430 (01)00880304281790(17)270808(10)171050 (01)00880304182882(17)270808(10)173190 (01)00880304281608(17)270809(10)371000 (01)00880304182899(17)270809(10)870950 (01)00880304182912(17)270809(10)028450 (01)00880304282308(17)270810(10)176560 (01)00880304182882(17)270810(10)699470 (01)00880304182899(17)270810(10)212450 (01)00880304182905(17)270810(10)212640 (01)00880304282223(17)270810(10)823460 (01)00880304182875(17)270811(10)052130 (01)00880304182912(17)270811(10)173350 (01)00880304182912(17)270811(10)052190 (01)00880304684584(17)270812(10)333740 (01)00880304182905(17)270812(10)303180 (01)00880304281615(17)270814(10)371010 (01)00880304282315(17)270814(10)040580 (01)00880304282315(17)270814(10)040640 (01)00880304182868(17)270814(10)028520 (01)00880304182868(17)270814(10)028530 (01)00880304182875(17)270814(10)008360 (01)00880304281639(17)270814(10)823450 (01)00880304282292(17)270816(10)833730 (01)00880304182899(17)270816(10)302570 (01)00880304182882(17)270817(10)052140 (01)00880304182875(17)270817(10)008330 (01)00880304684522(17)270818(10)376300 (01)00880304281721(17)270818(10)170970 (01)00880304182899(17)270819(10)871040 (01)00880304684508(17)270820(10)593590 (01)00880304182967(17)270820(10)312200 (01)00880304281745(17)270821(10)171000 (01)00880304281752(17)270821(10)171010 (01)00880304182875(17)270821(10)028370 (01)00880304182905(17)270828(10)212560 (01)00880304182868(17)270830(10)173130 (01)00880304182899(17)270830(10)302550 (01)00880304684560(17)270901(10)596780 (01)00880304182882(17)270904(10)699480 (01)00880304182882(17)270908(10)068920 (01)00880304684584(17)270912(10)593620 (01)00880304684560(17)270914(10)768970 (01)00880304684560(17)270920(10)734020 (01)00880304684546(17)270921(10)943360 (01)00880304684546(17)270921(10)492600 (01)00880304684508(17)270921(10)614300 (01)00880304684584(17)270921(10)333750 (01)00880304684508(17)270922(10)376250 (01)00880304684508(17)270922(10)506070 (01)00880304684508(17)270922(10)614310 (01)00880304684546(17)270922(10)733890 (01)00880304684485(17)270922(10)614280 (01)00880304684546(17)270922(10)376420 (01)00880304684461(17)270922(10)614240 (01)00880304684546(17)270922(10)376480 (01)00880304684546(17)270922(10)614340 (01)00880304684546(17)270922(10)614370 (01)00880304684546(17)270922(10)733960 (01)00880304182875(17)270922(10)028390 (01)00880304684546(17)270924(10)376350 (01)00880304684485(17)270925(10)614260 (01)00880304684607(17)271012(10)408650 (01)00880304684560(17)271227(10)090430 (01)00880304684560(17)271024(10)129060 (01)00880304684560(17)271025(10)090460 (01)00880304684546(17)271027(10)614350 (01)00880304684522(17)271110(10)819000 (01)00880304684522(17)271110(10)129120 (01)00880304684508(17)271118(10)636010 (01)00880304684560(17)271119(10)129070 (01)00880304684546(17)271120(10)614380 (01)00880304684485(17)271120(10)682250 (01)00880304684522(17)271120(10)636080 (01)00880304684522(17)271129(10)636040 (01)00880304684546(17)271129(10)614360 (01)00880304684485(17)271129(10)593680 (01)00880304684546(17)271207(10)253690 (01)00880304839311(17)271211(10)074190 (01)00880304684461(17)271214(10)682210 (01)00880304684560(17)271214(10)129040 (01)00880304684522(17)271216(10)376270 (01)00880304684485(17)271218(10)635990 (01)00880304282254(17)271229(10)549100 (01)00880304684485(17)280111(10)635980 (01)00880304684546(17)280102(10)189030 (01)00880304684461(17)280105(10)323840 (01)00880304684485(17)280111(10)262080 (01)00880304684560(17)280112(10)241120 (01)00880304684560(17)280112(10)297020 (01)00880304684461(17)280115(10)614250 (01)00880304684560(17)280116(10)682350 (01)00880304684508(17)280116(10)262090 (01)00880304684546(17)280119(10)262120 (01)00880304684584(17)280119(10)262170 (01)00880304282230(17)280122(10)479250 (01)00880304684546(17)280123(10)241060 (01)00880304684546(17)280123(10)241080 (01)00880304684522(17)280123(10)614320 (01)00880304182905(17)280123(10)258490 (01)00880304182882(17)280126(10)915530 (01)00880304684546(17)280126(10)090420 (01)00880304182912(17)280126(10)028440 (01)00880304684522(17)280129(10)143520 (01)00880304182905(17)280129(10)303190 (01)00880304684607(17)280130(10)682490 (01)00880304684461(17)280130(10)262030 (01)00880304684508(17)280201(10)522460 (01)00880304684546(17)280202(10)522500 (01)00880304684546(17)280202(10)541300 (01)00880304282209(17)280202(10)479240 (01)00880304684546(17)280205(10)541320 (01)00880304684584(17)280205(10)262190 (01)00880304684560(17)280205(10)241090 (01)00880304684560(17)280207(10)296960 (01)00880304684607(17)280207(10)297030 (01)00880304684546(17)280208(10)541310 (01)00880304684522(17)280208(10)541290 (01)00880304684485(17)280212(10)175530 (01)00880304684560(17)280212(10)540650 (01)00880304684546(17)280220(10)541330 (01)00880304684560(17)280221(10)574530 (01)00880304684584(17)280221(10)762550 (01)00880304684584(17)280221(10)522550 (01)00880304684584(17)280221(10)762560 (01)00880304684546(17)280222(10)540620 (01)00880304684584(17)280222(10)540670 (01)00880304182899(17)280223(10)705620 (01)00880304684560(17)280223(10)540660 (01)00880304182912(17)280223(10)929580 (01)00880304684508(17)280224(10)262100 (01)00880304182912(17)280225(10)554810 (01)00880304684560(17)280226(10)762530 (01)00880304684546(17)280226(10)780520 (01)00880304684522(17)280228(10)780480 (01)00880304684485(17)280302(10)262060 (01)00880304182899(17)280302(10)085230 (01)00880304182905(17)280308(10)085240 (01)00880304684560(17)280312(10)522510 (01)00880304684560(17)280314(10)522540 (01)00880304684546(17)280316(10)780500 (01)00880304684560(17)280321(10)762540 (01)00880304281684(17)280326(10)014960 (01)00880304684508(17)280330(10)296920 (01)00880304684508(17)280418(10)522450 (01)00880304182882(17)280418(10)847390 (01)00880304182899(17)280419(10)504370 (01)00880304182899(17)280420(10)847420 (01)00880304684546(17)280421(10)519500 (01)00880304182868(17)280423(10)728070 (01)00880304182936(17)280427(10)315000 (01)00880304684485(17)280430(10)593710 (01)00880304182882(17)280430(10)085210 (01)00880304684546(17)280501(10)519470 (01)00880304684560(17)280510(10)297000 (01)00880304182899(17)290218(10)935400 (01)00880304182905(17)280519(10)411550 (01)00880304684522(17)280821(10)296950 |
Recalling Firm/ Manufacturer |
Zimmer Biomet, Inc. 56 E Bell Dr Warsaw IN 46582-6989
|
| For Additional Information Contact | 411 Technical Services 574-371-3071 |
Manufacturer Reason for Recall | Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue. |
FDA Determined Cause 2 | Environmental control |
| Action | On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue.
Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties.
If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. |
| Quantity in Commerce | 219988 units (total) |
| Distribution | US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = LPH
|
|
|
|