• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall TherMax Blood Warmer Unit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TherMax Blood Warmer Unitsee related information
Date Initiated by FirmSeptember 19, 2019
Date PostedOctober 30, 2019
Recall Status1 Terminated 3 on December 11, 2020
Recall NumberZ-0247-2020
Recall Event ID 83816
510(K)NumberK190910 
Product Classification Dialyzer, high permeability with or without sealed dialysate system - Product Code KDI
ProductTherMax Blood Warmer Unit
Code Information Product Code: 955630, UDI: 00085412639505, Serial Numbers: 101332, 101335, 101358, 101369, 101402, 101403, 101404, 101406, 101415, 101416, 101429, 101430, 101437, 101439, 101440, 101441, 101443.   Product Code: 955515, Serial Numbers 100261, 100263, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100289, 100290, 100291, 100292, 100294, 100296, 100298, 100299, 100306, 100307, 100308, 100309, 100310, 100312, 100313, 100314, 100315, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100328, 100330, 100341, 100342, 100343, 100344, 100345, 100346, 100347, 100348, 100349, 100356, 100357, 100358, 100359, 100360, 100361, 100362, 100363, 100364, 100365, 100366, 100367, 100368, 100369, 100370, 100391, 100392, 100393, 100395, 100396, 100397,  100398, 100399, 100400, 100401, 100402, 100403, 100404, 100405, 100416, 100417, 100418, 100419, 100420, 100421, 100422, 100423, 100424, 100427, 100428, 100429, 100446, 100450, 100453, 100454, 100455, 100456, 100458, 100459, 100460, 100486, 100487, 100489, 100490, 100496, 100497, 100498, 100499, 100500, 100501, 100502, 100503, 100504, 100505, 100532, 100533, 100535, 100537, 100538, 100539, 100540, 100541, 100545, 100548, 100550, 100551, 100552, 100555, 100581, 100582, 100583, 100584, 100585, 100586, 100587, 100590, 100591, 100595, 100596, 100599, 100601, 100602, 100603, 100605, 100631, 100632, 100633, 100639, 100640, 100641, 100642, 100643, 100644, 100645, 100646, 100649, 100650, 100652, 100653, 100654, 100655, 100756, 100757, 100758, 100759, 100761, 100762, 100763, 100764, 100765, 100766, 100767, 100768, 100769, 100770, 100771, 100772, 100773, 100774, 100775, 100776, 100777, 100778, 100779, 100780, 100781, 100782, 100783, 100784, 100785, 100786, 100787, 100788, 100789, 100790, 100791, 100793, 100794, 100795, 100796, 100798, 100799, 100800, 100801, 100802, 100803, 100804, 100805, 100806, 100807, 100808, 100810, 100811, 100812, 100813, 100814, 100815, 100816, 100817, 100818, 100819, 100820, 100821, 100822, 100823, 100825, 100826, 100827, 100829, 100830, 100856, 100857, 100858, 100859, 100860, 100861, 100863, 100864, 100865, 100868, 100869, 100871, 100872, 100874, 100875, 100878, 100879, 100880, 100881, 100882, 100883, 100884, 100885, 100886, 100887, 100888, 100889, 100890, 100891, 100892, 100893, 100894, 100895, 100896, 100897, 100898, 100899, 100900, 100931, 100932, 100933, 100934, 100936, 100938, 100939, 100940, 100941, 100942, 100943, 100945, 100946, 100947, 100948, 100949, 100950, 100952, 100954, 100955, 100981, 100983, 100984, 100985, 100986, 100987, 100988, 100989, 100990, 100991, 100992, 100997, 100998, 100999, 101000, 101001, 101002, 101003, 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101024, 101026, 101027, 101030, 101081, 101082, 101083, 101084, 101085, 101086, 101088, 101089, 101090, 101091, 101092, 101094, 101095, 101096, 101097, 101098, 101099, 101101, 101103, 101104, 101105, 101106, 101107, 101108, 101109, 101110, 101111, 101112, 101115, 101121, 101122, 101123, 101124, 101126, 101127, 101172, 101173, 101174, 101175, 101176, 101177, 101179, 101180, 101181, 101182, 101183, 101184, 101185, 101186, 101187, 101188, 101189, 101190, 101192, 101229, 101230, 101231, 101233, 101235, 101236, 101237, 101241, 101246, 101247, 101248, 101249, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101271, 101272, 101273, 101274, 101276, 101277, 101278, 101458, 101460, 101469, 101470, 101471, 101472, 101473, 101474, 101475, 101476, 101477, 101478, 101479, 101480, 101481, 101482, 101483, 101484, 101485, 101486, 101487, 101488, 101489, 101490, 101491, 101492, 101508, 101509, 101510, 101511, 101512, 101513, 101514, 101515, 101516, 101517, 101519, 101520, 101521, 101522, 101538, 101539, 101540, 101541, 101542, 101543, 101544, 101545, 101546, 101547, 101548, 101549, 101550, 101551, 101552, 101553, 101554, 101555, 101556, 101557, 101558, 101559, 101560, 101561, 101562, 101593, 101594, 101595, 101596, 101597, 101598, 101599, 101600, 101601, 101602, 101603, 101604, 101605, 101607.
FEI Number 1417572
Recalling Firm/
Manufacturer
Baxter Healthcare Corporation
1 Baxter Pkwy
Deerfield IL 60015-4625
Manufacturer Reason
for Recall
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
FDA Determined
Cause 2
Device Design
ActionBaxter Healthcare notified customers on about 09/19/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that specific TherMax Blood Warmers listed below may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. The affected units may not have an adequate ground path, which could cause the units to become damaged if used with improperly grounded outlets. The damage could cause the TherMax Blood Warmer units to be nonfunctional during clinical use. Instructions included that customers could continue to safely use affected devices with a properly grounded outlet, that a local Baxter representative will contact each facility to arrange for the correction of affected devices, to complete and return the Baxter Customer Reply Form via fax to 224-270-5457 or scanning and e-mailing it to fca@baxter.com, and to notify customers if the product was further distributed. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
Quantity in Commerce504 units total
DistributionNationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = KDI
-
-