Class 2 Device Recall Ion Torrent Dx No Template Control Kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Ion Torrent Dx No Template Control Kitsee related information
Date Initiated by FirmMarch 16, 2020
Create DateApril 14, 2020
Recall Status1 Terminated 3 on March 26, 2021
Recall NumberZ-1707-2020
Recall Event ID 85203
PMA NumberP160045 
Product Classification Next generation sequencing oncology panel, somatic or germline variant detection system - Product Code PQP
ProductIon Torrent Dx No Template Control Kit- IVD for detection of DNA and RNA isolated from non-small cell lung cancer (NSCLC) tumor specimens Catalog Number: A32444
Code Information Lot Codes: S119_2174740* S119_2108429* P119_2149308 N119_2078115 S218_2069396 V118_1991155 T118_1990947 S217_2008385 S118_1988604 P118_1996942 N118_1974439 S117_1988308  * Mislabeled with a 24-month expiration date. However, the approved shelf life of the vials is also 9 months. Please note, the kit level labelling for the two impacted lots of Ion Torrent Dx No Template Control kit shows a 12-month expiration date, which is also incorrect  UDI: (01)10190302006156
FEI Number 3003335080
Recalling Firm/
Manufacturer
Life Technologies Corporation
7335 Executive Way
Frederick MD 21704-8354
Manufacturer Reason
for Recall
Extended Expiration Date prior to clearance
FDA Determined
Cause 2
Incorrect or no expiration date
ActionThermo Fisher issued notification on 3/16/20 via email followed by hard copy on 3/18/20, stating reason for recall, health risk and action to take: Review the expiration date of the products referenced in the table included in the Customer Notification and take the following actions: "Notify all Oncomine Dx Target Test users in your facility of this issue. "Review the expiration date of the products referenced in the table above and using Attachment 1 of this letter, take the following actions. Attachment 1 provides the corrected expiration dates per lot. For products that are past the 9-month corrected expiration date per Attachment 1: Discontinue use of the product, record the kit lot numbers and quantities impacted on the attached Customer Response Sheet, and email the completed response sheet to gabriella.armin@thermofisher.com. For products that are still within the 9-month corrected expiration date per Attachment 1: Correct the expiration dating on the product labelling within your inventory by reviewing Attachment 1 to determine the correct expiration date per kit lot. Strike through the incorrect expiration date and add the new one using permanent ink or amend the label by adding an additional label correcting the expiration date on the labelling. The updated labelled product can continue to be used until the 9-month expiration date, after which any remaining product should be discarded. Note: If affected product was used past the 9-month corrected expiration date per Attachment 1, but before the 12-month expiration date on the kit labels, there is no anticipated risk of generating inaccurate/incorrect results because the system will inform the user if the assay did not meet run passing criteria and test results wont be generated. Therefore, no action on your part is required. If you have any technical questions or concerns, please contact Technical Support at [NA --800-955-6288 option 4 or techsupport@thermofisher.com]
Quantity in Commerce79 units
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
PMA DatabasePMAs with Product Code = PQP
-
-