• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Canon Aquilion Precision, Model TSX304A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Canon Aquilion Precision, Model TSX304Asee related information
Date Initiated by FirmMarch 25, 2019
Date PostedApril 10, 2020
Recall Status1 Open3, Classified
Recall NumberZ-1725-2020
Recall Event ID 84954
510(K)NumberK173468 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductCanon Aquilion Precision, Model TSX-304A
Code Information Canon Aquilion Precision, Whole-Body X-Ray CT Scanner, Model TSX-304A, Serial number 2DB1892018.
Recalling Firm/
Manufacturer
Canon Medical System, USA, INC.
2441 Michelle Dr
Tustin CA 92780-7047
For Additional Information ContactMs. Terry Schultz
714-669-1421
Manufacturer Reason
for Recall
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.
FDA Determined
Cause 2
Software design
ActionRecall letters dated 3/25/2019 were issued on 3/25/2019 via USPS return receipt mail.
Quantity in Commerce1 device
DistributionDistribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JAK
-
-