|
Class 2 Device Recall Leksell Vantage Stereotactic System |
|
Date Initiated by Firm |
June 10, 2020 |
Create Date |
July 10, 2020 |
Recall Status1 |
Completed |
Recall Number |
Z-2442-2020 |
Recall Event ID |
85836 |
510(K)Number |
K171123
|
Product Classification |
Neurological stereotaxic Instrument - Product Code HAW
|
Product |
Leksell Vantage Stereotactic System, UDI/GTIN 7340048304887 |
Code Information |
Serial numbers: SH00026, SH00049, SH00061, SH00082, SH00092, SH00111, SH00028, SH00050, SH00072, SH00084, SH00094, SH00118, SH00029, SH00052, SH00073, SH00085, SH00095, SH00121, SH00030, SH00054, SH00074, SH00086, SH00097, SH00123, SH00032, SH00055, SH00077, SH00087, SH00099, SH00127, SH00033, SH00057, SH00078, SH00088, SH00100, SH00162, SH00041, SH00058, SH00079, SH00089, SH00102, SH00044, SH00059, SH00080, SH00090, SH00105, SH00045, SH00060, SH00081, SH00091, SH00108 |
Recalling Firm/ Manufacturer |
Elekta Instrument AB Kungstensgatan 18 Stockholm Sweden
|
Manufacturer Reason for Recall |
The locking mechanism at the interface of the Instrument Carrier and the Leksell Vantage Arc may not function properly.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm notified their consignees on 06/10/2020 by email. The notice explained the problem and provided directions on the removal of the spring. The firm offered assistance with the removal and provided the following contact: Neurosupport@elekta.com. |
Quantity in Commerce |
51 devices |
Distribution |
MA, IL, NH, NC, UT, OH, FL, MN, TX, NY, LA, AZ, MT, WV, Austria, Croatia, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Kuwait, Lithuania, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Switzerland, United Kingdom |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database |
510(K)s with Product Code = HAW and Original Applicant = Elekta Instrument AB
|
|
|
|