| | Class 2 Device Recall Alinity ci series |  |
| Date Initiated by Firm | June 19, 2020 |
| Create Date | July 24, 2020 |
| Recall Status1 |
Terminated 3 on April 10, 2024 |
| Recall Number | Z-2701-2020 |
| Recall Event ID |
85914 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Alinity ci series System Control Module (SCM); LN 3R70-01
Alinity ci series software versions, 3.1.1 and below |
| Code Information |
All serial numbers of System Control Module |
| FEI Number |
3002809144
|
Recalling Firm/ Manufacturer |
Abbott Gmbh & Co. KG
Max-Planck-Ring 2
65205
Wiesbaden
Germany
|
| For Additional Information Contact | Dana Schmidt 224-667-8598 |
Manufacturer Reason for Recall | Quality Control results were not properly evaluated for QC failures. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On June 19, 2020, Abbott issued Product Correction letters to all installed Alinity ciseries instrument customers:
customers were asked to do the following action:
1. Your Abbott representative will schedule a mandatory upgrade of your Alinity ciseries to software version 3.2.0. Until software version 3.2.0 is installed on your Alinity ciseries, Abbott recommends the following workarounds for issues 1 and 2:
2. In order to restore the QC analysis function for all assay files that have been previously updated and for all future assay file revisions until software 3.2.0 is installed on your system, it will be necessary to remove and then readd the assay to all control configurations containing the assay.
3. See Appendix A for instructions to remove and readd an assay from a control that contains more than one assay.
4. See Appendix B for instructions to remove and readd an assay from a control that contains only one assay.
5. These procedures will delete historical QC data from the Quality Control Summary screen and LeveyJennings (Graph) screen. You may choose to print the QC Levey Jennings Report, the QC Summary Report and to archive control results before performing this option. QC results will remain on the Control tab of the Results screen.Assay calibration data, calibrator configuration data and assay parameter configuration data are retained.
6. Please complete the included Abbott Customer Reply Form.
7. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter.
8.Please retain this letter for your laboratory records.
For any questions, U.S. Customers can contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., were asked to contact local area Customer Service |
| Quantity in Commerce | 3166 |
| Distribution | Worldwide distribution - US Nationwide distribution and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BERMUDA, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BULGARIA, BURKINA FASO, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN,
JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MADAGASCAR, MALAYSIA,
MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, OMAN
PAKISTAN, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, RUSSIA, SAUDI ARABIA,
SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, UKRAINE, URUGUAY, VIETNAM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|