• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Captivator Polypectomy Snare 27mm, Hexagonal Loop

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Captivator Polypectomy Snare 27mm, Hexagonal Loopsee related information
Date Initiated by FirmJuly 06, 2020
Create DateAugust 05, 2020
Recall Status1 Open3, Classified
Recall NumberZ-2762-2020
Recall Event ID 86004
510(K)NumberK131700 
Product Classification Snare, flexible - Product Code FDI
ProductCaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.
Code Information Outer Package UPN #: M00562341 Inner Package UPN #: M00562340 GTIN: 08714729019350 Lot #s: 25077093, 25164173, 25185062, 25187234, 25226123, 25255681, 25336754, 25403003, 25508379, 25524901 Expiration Dates: 20-Jan-23 through 14-May-23
FEI Number 3005099803
Recalling Firm/
Manufacturer
Boston Scientific Corporation
300 Boston Scientific Way
Marlborough MA 01752-1291
For Additional Information ContactNicole Pshon
508-683-4000
Manufacturer Reason
for Recall
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
FDA Determined
Cause 2
Device Design
ActionBSC will sent the customer communication package on July 6, 2020. The communication package includes a Customer Letter and a Reply Verification Tracking Form.
DistributionWorld wide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = FDI
-
-