|
Class 2 Device Recall Mako Integrated Cutting System (MICS) Handpiece |
 |
Date Initiated by Firm |
October 23, 2020 |
Date Posted |
November 19, 2020 |
Recall Status1 |
Completed |
Recall Number |
Z-0472-2021 |
Recall Event ID |
86700 |
510(K)Number |
K142350
|
Product Classification |
Orthopedic stereotaxic instrument - Product Code OLO
|
Product |
Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). |
Code Information |
(01) 00848486030193 (21), Serial numbers: 4200052, 4200095, 4200112, 4200117, 4200119, 4200167, 4200190, 4200193, 4200202, 4200203, 4200204, 4200226, 4200233, 4200235, 4200248, 4200250, 4200257, 4200269, 4200270, 4200288, 4200293, 4200305, 4200350, 4200373, 4200384, 4200387, 4200391, 4200397, 4200398, 4200400, 4200409, 4200436, 4200457, 4200459, 4200477, 4200488, 4200512, 4200540, 4200548, 4200550, 4200558, 4200624, 4200645, 4200653, 4200666, 4200717, 4200766, 4200774, 4200785, 4200789, 4200801, 4200806, 4200807, 4200816, 4200843, 4200856, 4200868, 4200904, 4200917, 4201000, 4201040, 4201068, 4201089, 4201102, 4201103, 4201121, 4201167, 4201218, 4201236, 4201238, 4201253, 4201296, 4201302, 4201366, 4201367, 4201376, 4201385, 4201408, 4201412, 4201442, 4201452, 4201454, 4201477, 4201490, 4201630, 4201661, 4201714, 4201752, 4201829, 4201831, 4201851, 4202009, 4202020, 4202038, 4202044, 4202062, 4202063, 4202122, 4202146, 4202162, 4202175, 4202315, 4202333, 4202626, 4202631, 4202659, 4202729, 4202742, 4202882, 4202987, 4203180, 4203198, 4203289, 4203424, 4203441, 4203649, 4203709, 4203728, 4203730, 4203731, 4203736, 4203807, 4203826, 4204015, 4204164, 4204389, 4204448, 4204539, 4204553, 4204577, 4204673, 4204812, 4204934, 4205093, 4205190, 4205380, 4205426, 4206219, 4206271, 4209023, 4209025, 4209044, 4209045, 4209046, 4209065, 4209066, 4209067, 4209068, 4209071, 4209149, 4209151, 4209155, 4209156, 4209157, 4209162, 4209164, 4209166, 4209167, 4209171, 4209185, 4209189, 4209210, 4209211, 4209213, 4209220, 4209244, 4209250, 4209252, 4209255, 4209256, 4209259, 4209261, 4209262, 4209267, 4209268, 4209271, 4209298, 4209299, 4209303, 4209304, 4209306, 4209308, 4209312, 4209317, 4209318, 4209345, 4209346, 4209357, 4209360, 4209363, 4209364, 4209365, 4209387, 4209388, 4209389, 4209398, 4209441, 4209443, 4209460, 4209463, 4209464, 4209466, 4209471, 4209472, 4209473, 4209474, 4209475, 4209476, 4209477, 4209486, 4209487, 4209491, 4209493, 4209494, 4209495, 4209496, 4209500, 4209519, 4209522, 4209526, 4209529, 4209531, 4209532, 4209533, 4209535, 4209541, 4209542, 4209543, 4209546, 4209597, 4209598, 4209599, 4209600, 4209601, 4209602, 4209604, 4209605, 4209606, 4209607, 4209608, 4209609, 4209610, 4209611, 4209612, 4209613, 4209614, 4209615, 4209616, 4209621, 4209628, 4209632, 4209634, 4209635, 4209637, 4209639, 4209640, 4209647, 4209667, 4209671, 4209685, 4209694, 4209698, 4209734, 4209764, 4209771, 4209774, 4209775, 4209776, 4209777, 4209778, 4209779, 4209780, 4209781, 4209782, 4209783, 4209784, 4209785, 4209786, 4209787, 4209788, 4209789, 4209790, 4209791, 4209792, 4209793, 4209794, 4209795, 4209796, 4209797, 4209798, 4209799, 4209800, 4209801, 4209802, 4209803, 4209804, 4209805, 4209806, 4209807, 4209808, 4209809, 4209810, 4209811, 4209812, 4209813, 4209814, 4209815, 4209816, 4209817, 4209818, 4209819, 4209820, 4209821, 4209822, 4209823, 4209834, 4209849, 4209859, 4209860, 4209863, 4209927, 4209928, 4209929, 4209930, 4209931, 4209932, 4209933, 4209934, 4209935, 4209936, 4209937, 4209938, 4209939, 4209940, 4209941, 4209942, 4209943, 4209945, 4209946, 4209947, 4209948, 4209949, 4209950, 4209951, 4209952, 4209954, 4209955, 4209956, 4209957, 4209959, 4209960, 4209961, 4209962, 4209963, 4209964, 4209965, 4209966, 4209967, 4209968, 4209969, 4209985, 4209986, 4209987, 4209988 Updated information submitted: Serial numbers removed from 12/23/20 letter: 4200112, 4200117, 4200202, 4201412, 4201490, 4201661; reason: These devices are still in scope and exhibit the issue. However, it was found that none of these devices were distributed to a customer. Therefore, they were removed from the attachment in the letter and are within Stryker control. Serial Numbers added to the recall letter: 4209292, 4209646, 4209841, 4202629, 4203963; reason: These devices were originally thought to be in Stryker control, but were shipped to customers. Serial Numbers added to the recall letter: 4209917, 4209922, 4209924; reason: These devices were shipped to Stryker OUS distributor and were thought to be distributed to customers. (Firm later confirmed that the units are under Stryker control.) |
Recalling Firm/ Manufacturer |
Mako Surgical Corporation 2555 Davie Rd Ste 110 Davie FL 33317-7424
|
For Additional Information Contact |
Loriann Russo 201-831-5272
|
Manufacturer Reason for Recall |
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
The firm issued a recall notice by UPS 1-day air through Sterilcycle, Inc. on 10/23/2020. The notice provided the following:
"Actions needed:
1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware.
2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility.
3. Quarantine and discontinue use of the recalled devices identified in the affected product list. Return the devices back to Stryker.
4. Hospitals/Branches/Agencies: Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form and fax a copy to 1-855-639-0611 or email to strykerorth6286@stericycle.com.
5. Hospitals Only: Please contact your Local Sales Office or your Stryker Sales Representative directly for product returns and inventory questions.
6. Branches/Agencies Only: Please forward this Urgent Medical Device Recall to the individuals or organizations who have consigned product, if applicable. Return all affected devices available at your location to the following address.
Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team
325 Corporate Drive
Dock M-East
Mahwah, NJ 07430
Ref. PFA 2514204
If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com."
On 12/23/2020, the firm sent an updated letter with amended serial numbers. |
Quantity in Commerce |
367 units |
Distribution |
Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|
510(K) Database |
510(K)s with Product Code = OLO and Original Applicant = GYRUS ACMI INC.
|
|
|
|