• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BIOTRONIK ITREVIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall BIOTRONIK ITREVIA see related information
Date Initiated by Firm March 08, 2021
Create Date April 21, 2021
Recall Status1 Terminated 3 on April 25, 2023
Recall Number Z-1452-2021
Recall Event ID 87549
PMA Number P050023S079 
Product Classification Defibrillator, implantable, dual-chamber - Product Code MRM
Product BIOTRONIK ITREVIA 7, VR-T DF-4, REF 393041, UDI: 04035479129569 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
Code Information Serial Number 60758873 60790507 60790508 60790509 60793100 60793104 60793112 60793113 60793114 60793115 60793120 60793132 60793151 60793446 60793447 60797807 60797810 60797811 60797813 60797818 60797820 60797824 60797825 60797829 60797832 60797833 60797835 60797838 60797840 60797841 60797842 60797853 60797854 60797856 60798617 60798618 60798620 60798633 60798637 60798641 60799910 60799917 60799921 60799924 60799925 60799926 60799929 60799935 60799937 60799938 60799947 60800996 60800998 60802135 60802141 60802144 60802568 60802575 60802579 60805453 60805593 60806318 60806320 60807087 60807090 60807091 60807092 60807093 60807095 60807096 60807186 60807187 60807189 60807198 60807199 60807203 60808405 60808407 60808412 60808413 60808416 60808426 60808428 60808430 60808432 60808433 60808436 60810738 60810741 60810742 60810744 60810747 60810748 60810750 60810753 60810755 60810758 60810759 60810762 60810765 60810766 60810767 60810770 60810774 60810776 60811841 60811842 60811850 60811982 60811983 60811985 60811990 60811999 60812000 60812002 60812003 60812009 60812016 60812018 60812683 60812686 60812688 60812697 60812701 60812775 60812777 60812779 60812785 60812791 60813683 60813684 60813685 60813690 60813691 60813692 60813693 60813694 60813695 60813696 60813698 60813700 60813701 60814174 60814175 60814178 60814180 60814182 60814184 60814186 60814189 60814190 60814195 60814819 60814822 60814835 60814840 60814841 60814843 60814844 60814845 60814848 60814849 60814852 60814854 60815356 60815357 60815359 60815361 60815363 60815365 60815366 60819234 60819240 60819259 60819260 60820995 60820996 60820998 60820999 60821003 60821005 60821006 60821011 60821014 60821017 60821020 60821027 60821030 60821031 60821032 60821033 60821034 60821035 60821038 60821039 60821040 60821042 60821043 60821046 60821247 60821248 60821250 60821256 60822034 60822039 60822043 60822307 60822311 60822314 60822329 60823021 60823030 60823031 60823033 60823037 60823055 60823057 60823071 60823072 60823073 60823076 60823077 60823078 60823080 60823182 60823184 60823203 60823207 60823998 60824000 60824002 60824012 60824013 60824014 60824015 60824018 60824019 60824021 60824022 60824025 60824027 60824028 60824030 60824034 60824035 60824038 60824040 60824044 60824046 60824217 60824241 60824257 60825786 60825795 60825796 60825804 60825851 60826562 60826565 60826585 60826619 60826623 60826634 60826643 60828285 60828286 60828287 60828296 60828326 60828332 60828338 60828350 60828359 60828391 60828393 60828395 60828396 60828397 60829692 60829697 60829704 60831259 60831260 60831261 60831263 60831269 60831271 60831273 60832589 60832592 60832597 60832626 60832651 60833222 60833502 60833520 60833524 60833528 60833530 60833536 60834928 60834930 60834932 60834933 60834935 60834952 60834954 60834956 60835265 60835790 60835803 60835808 60835809 60835817 60835818 60835822 60835828 60835829 60835841 60835860 60835862 60837079 60837089 60837094 60837296 60837301 60837304 60837310 60837313 60837318 60837330 60837342 60838350 60838362 60838363 60838366 60838523 60838524 60838525 60838532 60838534 60838547 60839796 60839798 60839800 60839801 60840333 60840341 60840342 60841043 60841044 60841047 60841048 60841049 60841051 60841052 60841058 60841060 60843405 60843406 60843407 60843413 60843443 60843458 60843460 60844898 60844912 60844927 60844929 60844936 60844937 60844951 60845160 60845162 60845165 60845168 60845169 60845186 60845270 60845272 60845281 60845286 60845297 60845298 60845300 60845304 60845306 60847596 60847615 60847628 60847632 60848923 60848924 60848928 60848930 60848932 60848933 60848934 60848940 60848944 60848947 60848950 60848952 60849391 60849393 60849407 60849408 60850920 60850929 60852100 60852101 60852102 60852107 60852108 60852109 60852110 60852111 60852113 60852114 60852116 60852118 60852120 60852122 60852123 60852124 60852127 60852129 60852130 60852131 60852135 60852137 60852449 60852455 60852458 60852459 60852461 60852462 60852464 60852465 60852466 60852475 60852476 60853376 60853398 60853400 60853403 60853419 60853424 60853430 60853432 60853461 60854791 60854792 60854798 60854799 60854800 60855555 60855556 60855557 60855560 60855561 60855563 60855564 60855568 60855571 60855572 60855573 60855574 60855575 60855576 60855579 60855580 60855581 60855582 60855583 60855771 60855780 60855782 60855785 60855786 60855787 60857000 60857003 60857007 60857009 60857012 60857013 60857014 60857015 60857016 60857594 60857608 60857610 60857611 60857614 60863671 60863672 60863674 60863675 60863685 60863687 60864383 60864522 60864527 60864538 60864564 60865015 60865017 60865020 60865029 60865030 60865033 60865046 60865173 60865175 60865176 60865179 60865187 60865196 60865197 60865208 60866545 60866568 60866572 60867643 60867649 60867665 60867667 60867668 60867788 60867790 60867791 60867795 60868576 60869062 60869085 60869093 60869485 60869489 60869505 60869509 60875564 60875612 60875619 60876213 60876520 60877067 60877104 60877107 60877111 60877113 60878975 60878976 60878981 60878984 60878985 60878988 60878990 60878993 60878995 60879008 60879009 60879010 60879018 60879823 60879826 60880932 60880947 60880978 60880979 60880987 60881016 60883599 60883612 60883616 60883647 60883654 60883656 60883662 60883664 60883676 60883677 60883996 60884626 60887000 60888332 60888335 60888342 60888349 60888351 60888366 60888610 60888622 60888624 60888626 60888628 60888631 60888633 60889957 60889971 60889975 60889978 60891122 60891123 60891124 60891125 60891129 60891139 60891141 60891142 60891144 60891145 60891146 60891148 60892405 60892414 60893247 60895968 60895969 60895970 60896214 60896221 60896274 60898481 60901108 60901799 60901808 60901813 60901821 60901825 60901828 60902059 60902062 60902065 60902072 60904661 60905436 60905438 60905439 60905440 60905443 60905448 60905451 60907613 60907631 60907638 60907647 60907648 60907651 60909387 60909388 60909390 60909394 60910780 60914336 60914349 60914404 60914869 60915274 60915296 60917836 60920540 60920544 60920572 60920573 60920583 60920585 60920588 60921111 60921113  
Recalling Firm/
Manufacturer
BIOTRONIK Inc
6024 Jean Rd
Lake Oswego OR 97035-5571
For Additional Information Contact BIOTRONIK Technical Services
1800-547-0394
Manufacturer Reason
for Recall
There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
FDA Determined
Cause 2
Device Design
Action On 3/8/21, BIOTRONIK issued a "Urgent Field Safety" recall notification to all affected consignees via UPS. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. To determine if a specific device is affected by this corrective action you can lookup the device at the following web address: www.biotronik.com/devicelookup. 2. Devices in stock: Do not implant any potentially affected devices. Local BIOTRONIK representatives will replace affected devices in hospital inventory. 3. Continue with the standard patient follow-up schedule. -During follow-ups: Verify the status of the device and battery during in-office or Home Monitoring follow-ups. Please note that unresponsive devicesor those that are not transmitting data may be experiencing this issue and your BIOTRONIK representative should be informed if you observe any unusual device behavior. -Home Monitoring should be utilized whenever possible as it provides timely ERI warnings to reduce the risk of sudden loss of therapy. If you do not yet use Home Monitoring, please consider if this option is appropriate for you and your patients. BIOTRONIK will provide Cardio Messenger devices to monitor implants affected by this advisory. If you would like to register for Home Monitoring, please contact your local BIOTRONIK representative. 4. If there is an unexpected ERI notification for a device that is subject to this advisory, a timely replacement should be considered based on the patients underlying conditions: - For patients that are not pacemaker dependent, or patients with a primary prevention ICD, device replacement within one week after ERI notification is recommended. - For pacemaker dependent patients, replacement of the device is recommended immediately after ERI notification. 5. In consultation with our medical advisory board, BIOTRONIK does not recommend prophylactic replacement. The risk of complications for ICD exchange1-3 outweighs
Quantity in Commerce 670 units
Distribution US: Nationwide OUS: Worldwide
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA Database PMAs with Product Code = MRM and Original Applicant = BIOTRONIK, INC.
-
-