| | Class 2 Device Recall Philips Physiomonitoring system |  |
| Date Initiated by Firm | March 29, 2021 |
| Create Date | April 29, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1517-2021 |
| Recall Event ID |
87732 |
| 510(K)Number | K101571 |
| Product Classification |
Monitor, physiological, patient (without arrhythmia detection or alarms) - Product Code MWI
|
| Product | Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by professional healthcare providers for complete physiologic/hemodynamic monitoring
Service numbers associated with FC2010:
Xper Flex Cardio 2010 Rev A 453564241901
Xper Flex Cardio 2010 Rev B 453564483321
Flex Cardio FC2010 Rev C 453564621791
Xper Flex Cardio 2010 REV C RUEXCHANGE 453564675021
Xper Flex Cardio 2010 Rev D 453564669081
Xper Flex Cardio 2010 Rev D Russia 453564674581
Xper Flex Cardio FC2010 Rev E 453564845841
Xper Flex Cardio 2010 Rev E Russia 453564845881
Service numbers associated with FC2020:
Xper Flex Cardio 2020 Rev A 453564241911
Xper Flex Cardio 2020 Rev B 453564483331
Flex Cardio FC2020 Rev C 453564621801
Xper Flex Cardio 2020 Rev D 453564675001
Xper Flex Cardio FC2020 Rev E 453564845861
Software version number: 1.5.19 and 1.6.8 |
| Code Information |
Service Number/ UDI # 453564241901 (01)00884838082113; 453564241911 (01)00884838082113; 453564483321 (01)00884838082113; 453564483331 (01)00884838083516; 453564621791 (01)00884838084902; 453564621801 (01)00884838084919; 453564845881 (01)00884838095052; 453564675021 (01)00884838103245; 453564669081 (01)00884838083820; 453564675001 (01)00884838086005; 453564674581 (01)00884838083820; 453564845841 (01)00884838093645; 453564845861 (01)00884838093652 |
Recalling Firm/ Manufacturer |
Philips North America Llc 222 Jacobs St Cambridge MA 02141-2289
|
| For Additional Information Contact | Philips Customer Services 800-722-9377 |
Manufacturer Reason for Recall | Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric
values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform,
erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor |
FDA Determined Cause 2 | Software design |
| Action | Philps issued Urgent Field Safety Field Notice-Medical Device Correction to or U.S. customers, the letters were sent via certified mail on 3/29/21.
Philips Markets Organizations are responsible for distributing the letters outside of the U.S.
Letter states reason for recall, health risk and action to take:
Flex Cardio System.
Because the issues listed above can be promptly identified by a qualified health care professional who has reviewed this notice and is closely observing the monitored patient, the Xper Flex Cardio may continue to be used if this step is taken.
Resetting the device as described in the IFU (Section 4, page 87), is likely to
return the device to normal operation, which takes approximately 3-5 minutes.
To reset your Flex Cardio device, close any patient cases and then turn the
power switch off (see back of device for the power switch). Wait 510 seconds
then turn the power switch back on. Allow the device to fully restart and
restore the monitoring display.
Please complete and return the Customer Reply Form included on the last page of this communication indicating your receipt and understanding of this
information.
Philips will provide a software update for the Xper Flex Cardio to correct all but,
issue #7 above at no charge. A Philips representative will contact you when the
software is available for installation.
Philips has also added directions to the IFU for the Xper Flex Cardio on how to
reset the device in the event that the user observes issue #7 above. An IFU
addendum with the directions will be provided to all affected users when it
becomes available.
If you need any further information or support concerning this issue, please
contact your local Philips representative: 800-669-1328 option 2, then option 3. |
| Quantity in Commerce | 9208 units WW |
| Distribution | Nationwide
Foreign:
Argentina
Australia
Austria
Bangladesh
Belgium
Bulgaria
Canada
Chile
China
Croatia
Cyprus
Czech Republic
Denmark
Dominican Republic
Ecuador
Egypt
Estonia
Finland
France
Germany
Greece
Hungary
Iceland
India
Indonesia
Iran
Iraq
Ireland
Israel
Italy
Jordan
Kazakhstan
Kenya
Korea, Republic of
Kuwait
Latvia
Lebanon
Libya
Lithuania
Malaysia
Mauritius
Mexico
Mongolia
Morocco
Myanmar
Nepal
Netherlands
New Caledonia
New Zealand
Nigeria
Norway
Oman
Pakistan
Palestine, State of
Panama
Peru
Philippines
Poland
Portugal
Puerto Rico
R¿union
Romania
Russian Federation
Saudi Arabia
Serbia
Singapore
Slovenia
South Africa
Spain
Sri Lanka
Sweden
Switzerland
Syrian Arab Republic
Taiwan
Thailand
Tunisia
Turkey
United Arab Emirates
United Kingdom
Uzbekistan
Viet Nam
Yemen |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = MWI
|
|
|
|