• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Guardian Sensor (3)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Guardian Sensor (3)see related information
Date Initiated by FirmFebruary 26, 2021
Create DateMay 17, 2021
Recall Status1 Open3, Classified
Recall NumberZ-1618-2021
Recall Event ID 87476
Product Classification Sensor, glucose, invasive - Product Code MDS
ProductMedtronic, REF: MMT-7008D, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 1PK; GTIN: 00763000252045 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.
Code Information All Lot numbers within active shelf life
FEI Number 3003166194
Recalling Firm/
Manufacturer
Medtronic Inc.
18000 Devonshire St
Northridge CA 91325-1219
For Additional Information ContactPamela Reese
818-576-4700
Manufacturer Reason
for Recall
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn February 26, 2021, Medtronic posted the "Urgent Recall Correction" notification to their website. On March 2, 2201, Medtronic provided U.S HCPs, Patients, Distributors and globally affected Clinical trial sites with a copy of the "Urgent Recall Correction" Notification. On March 8, 2021, Medtronic provided OUS patients, HCP and distributions with a copy of the "Urgent Recall Correction" notification Communication was provided to users via: email, fax, and/or mail. In addition to informing affected users about the correction, Medtronic ask consignees to take the following actions: Do not use continuous glucose monitoring (CGM) while taking hydroxyurea. Only use their BG meter to monitor glucose levels while taking hydroxyurea. Consult with the HCP if they are taking hydroxyurea and using a CGM. When using an insulin pump system, they should disable the CGM feature by going to the sensor settings and refer to the user guide that came with the insulin pump for instructions. Always check the label of any medication and confirm with you whether or not hydroxyurea or hydroxycarbamide is an active ingredient. If further distributed, please provide those consignees with a copy of the notification. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Complete and submit the report Online: www.fda.gov/medwatch/report.htm " Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Quantity in Commerce207 units
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-