| Date Initiated by Firm | July 02, 2021 |
| Create Date | September 10, 2021 |
| Recall Status1 |
Terminated 3 on November 02, 2023 |
| Recall Number | Z-2443-2021 |
| Recall Event ID |
88389 |
| Product Classification |
Dispenser, solid medication - Product Code NXB
|
| Product | The Hero Model 100 (H-100) is a system consisting of a smart pill dispenser and a medication management app. |
| Code Information |
Model 100 (H-100)
Catalog Number: F00002 Hero 100 V1.4 |
Recalling Firm/ Manufacturer |
Hero Health 20902 Corsair Boulevard Hayward CA 94545
|
| For Additional Information Contact | 855-855-9962 |
Manufacturer Reason for Recall | Due to dispenser prompting for a missed dose that had already been dispensed |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | On 07/02/2021, the firm initially sent out an "URGENT NOTICE" via email notifying customers of an product update correction. Customers were informed that the firm had become aware that a recent firmware update (v2.4.12) that users have experienced an error, which caused improper recording of a dispensed dose as having not been dispensed. This error may prompt a user to dispense the same dose again, and the mobile app will indicate that the previously dispensed dose has not been taken yet.
Customers are instructed to:
-If the pill dispenser prompts you to re-dispense a scheduled dose which has already been dispensed today, select Skip.
-Use extra caution when dispensing doses prior to the scheduled firmware update and verify that the pill dispenser has a working internet connection. This will ensure that the firmware update (v2.4.14) will be automatically applied on all potentially affected online pill dispensers (at approximately 2:00 AM on July 3, 2021) to correct the issue.
-To manually update firmware immediately, follow these steps:
1) Ensure the Pill Dispenser is connected to WiFi (Check for the WiFi icon on the main screen).
2) Go to the Device menu, by pressing the Center button on the main screen.
3) Scroll down and select Device Settings .
4) Scroll down and select Firmware Update.
5) Select I m Ready.
The update could take several minutes. It is recommended to run the firmware update at least 30 minutes before a scheduled dose.
-If the pill dispenser is not connected to WiFi, the firmware update cannot be completed. In this event, contact the Recalling Firm's Member support for assistance before continuing to use the pill dispenser to ensure the update is applied as soon as possible
For question or further assistance, contact the Recalling firm at 855-855-9962 or via email at hello@herohealth.com (24/7).
|
| Quantity in Commerce | 12,039 units which had software version V2.04.012 |
| Distribution | U.S. Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|