 Class 2 Device Recall NAMIC Stopcockssee related information
Date Initiated by FirmAugust 06, 2021
Create DateSeptember 24, 2021
Recall Status1 Terminated 3 on August 25, 2025
Recall NumberZ-2575-2021
Recall Event ID 88516
510(K)NumberK782095 K842829 
Product Classification Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass - Product Code DTL
ProductNAMIC Stopcocks: (1) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015002, UPN H965700150021, Three-way Stopcock with Rotating Collar, Rx only, Sterile; (2) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015003, UPN H965700150031, Three-Way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (3) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015012, UPN H965700150121, Four-way Stopcock with Rotating Collar, Rx only, Sterile; (4) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015013, UPN H965700150131, Four-way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (5) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015015, UPN H965700150151, One-way Stopcock with Collar, Rx only, Sterile; (6) NAMIC STOPCOCK, 400 psi (28 bar), REF 70035002, UPN H965700350021, Three-way Stopcock with Rotating Collar, Rx only, Sterile; (7) NAMIC STOPCOCK, 400 psi (28 bar), REF 70035003, UPN H965700350031, Three-way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (8) NAMIC STOPCOCK, 400 psi (28 bar), REF 70035008, UPN H965700350081, Three-way Stopcock with Rotating Adaptor, Rx only, Sterile; (9) NAMIC STOPCOCK, 400 psi (28 bar), REF 70035009, UPN H965700350091, Three-way Stopcock with Rotating Adaptor, Rx only, Sterile; (10) NAMIC STOPCOCK, 1050 psi (72 bar), REF 70055003, UPN H965700550031, Three-way Stopcock with Fixed Male Luer Lock, Rx only, Sterile; (11) NAMIC STOPCOCK, 1050 psi (72 bar), REF 70055008, UPN H965700550081, Three-way Stopcock with Rotating Adapter, Rx only, Sterile; (12) NAMIC STOPCOCK, 1050 psi (72 bar), REF 70055009, UPN H965700550091, Three-way Stopcock with Rotating Adaptor, Rx only, Sterile; and (13) NAMIC STOPCOCK, 1050 psi (72 bar), REF 70055017, UPN H965700550171, One-way Stopcock with Fixed Male Luer Lock, Rx only, Sterile.
Code Information Expiration dates 12/31/2023-7/12/2024. (1) REF 70015002 - Lot numbers 0000002034, 0000004540, 0000004748, 0000004762, 0000005986, 0000006006, 0000008507, 0000008561, 0000010311, 0000011009, 0000011021, 0000012601, 0000012667, 0000013628, 0000013629, 0000014168, 0000015265, 0000015855, 0000016237, 0000017369; (2) REF 70015003 - Lot numbers 0000002035,0000004399, 0000004781, 0000004859, 0000005958, 0000014171, 0000015997; UPN H965700150031 - Lot number 5658899; (3) REF 70015012 - Lot numbers 0000012075, 0000013640, 0000015290; (4) REF 70015013 - Lot number 0000008503, UPN H965700150131 - Lot numbers 5656213, 5662009; (5) REF 70015015 - Lot numbers 0000005117, 0000007641, 0000007718, 0000011557, 0000014654, 0000017017; UPN H965700150151 - Lot numbers 5655470, 5656214, 5659327; (6) REF 70035002 - Lot numbers 0000002036, 0000002455, 0000008408, 0000008440, 0000008476, 0000010488, 0000015089, 0000017049; UPN H965700350021 - Lot numbers 5654454; 5656220; (7) REF 70035003 - Lot numbers 0000006313, 0000008178, 0000015121; (8) REF 70035008 - Lot numbers 0000003306, 0000010735, 0000014795, 0000016525; UPN H965700350081 - Lot numbers 5653290, 5656221; (9) REF 70035009 - Lot numbers 0000007115, 0000007611, 0000007616, 0000009644, 0000010737, 0000010748, 0000012633, 0000012634, 0000014808, 0000015019, 0000015024, 0000017448; UPN H965700350091 - Lot numbers 5653289, 5658896, 5659333, 5659887, 5659889; (10) REF 70055003 - Lot numbers 0000007025, 0000012635, 0000016518; UPN H965700550031 - Lot numbers 5653288, 5658901, 5662967; (11) REF 70055008 - Lot numbers 0000007081, 0000014807; UPN H965700550081 - Lot numbers 5653287, 5656299; (12) REF 70055009 - Lot numbers 0000003281, 0000003284, 0000007024, 0000007600, 0000007601, 0000009626, 0000009627, 0000010746, 0000012300, 0000012301, 0000012302, 0000014982, 0000015022, 0000016270, 0000016520; UPN H965700550091 - Lot numbers 5653286, 5657315, 5658895, 5658897, 5659900, 5659903, 5659905; and (13) REF 70055017 - Lot numbers 0000007147, 0000010747, 0000015023; UPN H965700550171 - Lot number 5661997.
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries Inc
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactMs. Karin Johnson
866-359-1704
Manufacturer Reason
for Recall
Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions.
FDA Determined
Cause 2
Process design
ActionThe recalling firm issued letters dated and mailed on 8/6/2021 notifying consignees of the issue and requesting they immediately check their stock for the affected product and destroy it.
Quantity in Commerce816,420 manifolds and stopcocks
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Australia, Japan, Latvia, Lebanon, Malaysia, Netherlands, Pakistan, People's Republic of China, Panama, Singapore, South Africa, South Korea, Thailand, and United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DTL
510(K)s with Product Code = DTL
-
-