 Class 2 Device Recall Medtronic Wireless Rechargersee related information
Date Initiated by FirmNovember 04, 2021
Create DateNovember 24, 2021
Recall Status1 Open3, Classified
Recall NumberZ-0293-2022
Recall Event ID 88953
PMA NumberP970004S302 
Product Classification Stimulator, electrical, implantable, for incontinence - Product Code EZW
ProductMedtronic Wireless Recharger Model WR9220 included in the RS5200 Recharger Kit used by the InterStim Micro (Model 97810) implantable neurostimulator.
Code Information UDI 00763000186265
FEI Number 2182207
Recalling Firm/
Manufacturer
Medtronic Neuromodulation
7000 Central Ave Ne
Minneapolis MN 55432-3568
Manufacturer Reason
for Recall
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
FDA Determined
Cause 2
Other
ActionThe recalling firm issued undated letters on 11/4/2021 via USPS Priority Shipping. The recalling firm has three different letters addressed to the Healthcare Professional depending upon the type of device the wireless recharger is used with (InterStim Micro, product number RS5200; Activa RC, product number RS6200; or RestoreUltra & RestoreSensor, Product number RS7200, which is currently not distributed in the U.S. at this time). The letter gives an issue overview and provides requested actions. Those actions include reinforcing the instructions in the enclosed Patient Quick User Guide with their patients and to attempt to reset the WR if the device appears unresponsive. If the problem is not resolved, then contact information is provided for Medtronic Technical Services. A customer confirmation form dated November 2021 was enclosed to acknowledge the consignee has received the letter.
Quantity in Commerce10,081 devices
DistributionUS Nationwide, including PR. There was government distribution but no military distribution was identified.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.