 Class 2 Device Recall EVOS MEDIAL DISTAL FEMUR PLATEsee related information
Date Initiated by FirmNovember 01, 2021
Create DateDecember 10, 2021
Recall Status1 Terminated 3 on June 15, 2026
Recall NumberZ-0364-2022
Recall Event ID 89031
510(K)NumberK201918 
Product Classification Plate, fixation, bone - Product Code HRS
Productsmith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201
Code Information Batch Number: 20HM10255, 20HM10256, 20HM10257, UDI
FEI Number 1020279
Recalling Firm/
Manufacturer
Smith & Nephew, Inc.
1450 E Brooks Rd
Memphis TN 38116-1804
For Additional Information Contact
901-396-2121
Manufacturer Reason
for Recall
Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
FDA Determined
Cause 2
Under Investigation by firm
ActionThe firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.
Quantity in Commerce35 units
DistributionUS Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.